2 | HoverTech International HMManual, Rev. D
Intended Use and Precautions
INTENDED USE
The HoverMatt® Air Transfer System is used to assist caregivers with patient
transfers, positioning, turning and proning. The HoverTech Air Supply
inflates the HoverMatt to cushion and cradle the patient, while air
simultaneously escapes from the holes on the underside, reducing the
force needed to move the patient by 80-90%.
INDICATIONS
• Patients unable to assist in their own lateral transfer
• Patients whose weight or girth poses a potential health risk for the
caregivers responsible for repositioning or laterally transferring said
patients
CONTRAINDICATIONS
• Patients who are experiencing thoracic, cervical or lumbar fractures
that are deemed unstable, unless using in conjunction with a spinal
board on top of the HoverMatt (follow your state’s protocol regarding
use of spinal boards)
INTENDED CARE SETTINGS
• Hospitals, long term or extended care facilities
PRECAUTIONS HOVERMATT
• Caregivers must verify that all caster brakes have been engaged prior
to transfer.
• For safety, always use two people during patient transfer.
• Additional caregivers are recommended when moving a patient over
750 lbs/340kg.
• Never leave patient unattended on an inflated device.
• Use this product only for its intended purpose as described in this
manual.
• Only use attachments and/or accessories that are authorized by HoverTech International.
• When transferring to a low air loss bed, set the bed mattress air flow to
the highest level for a firm transfer surface.
• Never attempt to move a patient on an uninflated HoverMatt.
• WARNING: In the OR – To prevent the patient from slipping, always
deflate the HoverMatt and secure the patient and HoverMatt to the
OR table prior to moving the table into an angled position.
PRECAUTIONS AIR SUPPLY
• Not for use in the presence of flammable anesthetics or in a hyperbaric
chamber or oxygen tent.
• Route the power cord in a manner to ensure freedom from hazard.
• Avoid blocking the air intakes of the air supply.
• When using the HoverMatt in the MRI environment, a 25 ft. specialty
MRI hose is required (available for purchase).
• CAUTION: Avoid electric shock. Do not open air supply.
• WARNING: Reference product specific user manuals for operating
instructions.
Symbol Reference
DECLARATION OF CONFORMITY
TO MEDICAL DEVICE DIRECTIVE