Hospira Plum A+ Service Manual Plum A+ Service Manual

Page 1
INFUSION SYSTEM
For use with list numbers 12348-04 and 12618-04
RR
3
Technical Service Manual
430-95424-002 (Rev. 01/06)
Page 2
©
Hospira, Inc.
This document and the subject matter disclosed herein are proprietary information. Hospira retains all the exclusive rights of dissemination, reproduction, manufacture, and sale. Any party using this document accepts it in confidence, and agrees not to duplicate it in whole or in part nor disclose it to others without the written consent of Hospira.
430-95424-002 (Rev. 01/06) Plum A+®3 Infusion System
Page 3
Change History
Part Number Description of Change
430-95424-001 (Rev. 09/03)
430-95424-002 (Rev. 01/06)
Original Issue
Second Issue
Updated Section 1, Section 5, Section 6, and Section 8
Updated battery information
Incorporated Hospira name change and updated to current style
Updated graphics throughout
Updated back page
Technical Service Manual 430-95424-002 (Rev. 01/06)
Page 4
CHANGE HISTORY
This page intentionally left blank.
430-95424-002 (Rev. 01/06) Plum A+®3 Infusion System
Page 5

Contents

Section 1
INTRODUCTION . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
1.1 SCOPE . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
1.2 CONVENTIONS . . . . . . . . . . . . . . . . . . . . . . . . 1-2
1.3 COMPONENT DESIGNATORS . . . . . . . . . . . . . . . . . . 1-2
1.4 ACRONYMS AND ABBREVIATIONS . . . . . . . . . . . . . . . . 1-3
1.5 USER QUALIFICATION . . . . . . . . . . . . . . . . . . . . . 1-5
1.6 ARTIFACTS . . . . . . . . . . . . . . . . . . . . . . . . . 1-5
1.7 INSTRUMENT INSTALLATION PROCEDURE . . . . . . . . . . . . 1-5
1.7.1 UNPACKING . . . . . . . . . . . . . . . . . . . . . . 1-6
1.7.2 INSPECTION . . . . . . . . . . . . . . . . . . . . . . 1-6
1.7.3 SELF TEST. . . . . . . . . . . . . . . . . . . . . . . . 1-6
1.8 BIOMED SETTINGS . . . . . . . . . . . . . . . . . . . . . . 1-8
1.8.1 IV PARAMETERS . . . . . . . . . . . . . . . . . . . . . 1-10
1.8.2 ALARMS LOG . . . . . . . . . . . . . . . . . . . . . . 1-11
1.8.3 SETTING THE TIME AND DATE . . . . . . . . . . . . . . . 1-12
Section 2
WARRANTY . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
Section 3
SYSTEM OPERATING MANUAL . . . . . . . . . . . . . . . . . . . . 3-1
Section 4
THEORY OF OPERATION . . . . . . . . . . . . . . . . . . . . . . . 4-1
4.1 GENERAL DESCRIPTION . . . . . . . . . . . . . . . . . . . . 4-1
4.2 ELECTRONIC SUBSYSTEM OVERVIEW . . . . . . . . . . . . . . . 4-2
4.2.1 CPU SUBSYSTEM . . . . . . . . . . . . . . . . . . . . . 4-3
4.2.1.1 CPU . . . . . . . . . . . . . . . . . . . . . . . 4-3
4.2.1.2 SYSTEM MEMORY ADDRESS MAP . . . . . . . . . . . 4-3
4.2.1.3 PROGRAMMABLE READ-ONLY MEMORY . . . . . . . . 4-4
4.2.1.4 STATIC RANDOM ACCESS MEMORY . . . . . . . . . . 4-4
4.2.1.5 CONTROL LOGIC . . . . . . . . . . . . . . . . . 4-4
4.2.1.6 LCD CONTROLLER . . . . . . . . . . . . . . . . . 4-4
4.2.1.7 LCD BACKLIGHT CONTROL . . . . . . . . . . . . . 4-5
4.2.1.8 LCD CONTRAST CONTROL. . . . . . . . . . . . . . 4-5
4.2.1.9 REAL-TIME CLOCK . . . . . . . . . . . . . . . . . 4-5
4.2.1.10 VOLTAGE MONITOR WATCHDOG TIMER . . . . . . . . 4-6
4.2.1.11 ANALOG-TO-DIGITAL CONVERTER . . . . . . . . . . 4-6
4.2.1.12 DIGITAL-TO-ANALOG CONVERTER . . . . . . . . . . 4-8
4.2.1.13 FRONT PANEL KEYPAD MATRIX . . . . . . . . . . . 4-8
4.2.1.14 FRONT PANEL [ON/OFF] KEY . . . . . . . . . . . . . 4-8
4.2.1.15 FRONT PANEL LED INDICATORS . . . . . . . . . . . 4-8
4.2.1.16 KEYPAD LOCKOUT INTERFACE . . . . . . . . . . . . 4-9
4.2.1.17 NURSE CALL INTERFACE . . . . . . . . . . . . . . 4-9
4.2.1.18 AUDIBLE INDICATORS . . . . . . . . . . . . . . . 4-9
4.2.1.19 BARCODE READER INTERFACE . . . . . . . . . . . . 4-9
4.2.1.20 DATAPORT INTERFACE . . . . . . . . . . . . . . . 4-10
4.2.1.21 POWER SUPPLY INTERFACE . . . . . . . . . . . . . 4-10
4.2.1.22 MECHANISM INTERFACE . . . . . . . . . . . . . . 4-11
Technical Service Manual iii 430-95424-002 (Rev. 01/06)
Page 6
CONTENTS
4.2.2 POWER SUPPLY SUBSYSTEM . . . . . . . . . . . . . . . . 4-13
4.2.2.1 MAIN SWITCHING REGULATOR . . . . . . . . . . . 4-13
4.2.2.2 MAIN REGULATOR FAULT DETECTION . . . . . . . . 4-14
4.2.2.3 SYSTEM POWER . . . . . . . . . . . . . . . . . . 4-14
4.2.2.4 AUXILIARY SUPPLIES . . . . . . . . . . . . . . . . 4-14
4.2.2.5 POWER CONTROL . . . . . . . . . . . . . . . . . 4-15
4.2.2.6 BATTERY VOLTAGE MEASUREMENT . . . . . . . . . 4-16
4.2.2.7 BATTERY CHARGE/DISCHARGE CURRENT
MEASUREMENT . . . . . . . . . . . . . . . . . . 4-16
4.2.2.8 BATTERY CHARGER . . . . . . . . . . . . . . . . 4-16
4.2.3 MECHANISM SUBSYSTEM . . . . . . . . . . . . . . . . . 4-17
4.2.3.1 MOTORS/MOTOR DRIVE . . . . . . . . . . . . . . 4-17
4.2.3.2 MOTOR POSITION SENSORS . . . . . . . . . . . . . 4-18
4.2.3.3 V2_5 REFERENCE VOLTAGE . . . . . . . . . . . . . 4-19
4.2.3.4 AIR SENSORS . . . . . . . . . . . . . . . . . . . 4-19
4.2.3.5 PRESSURE SENSORS . . . . . . . . . . . . . . . . 4-21
4.2.3.6 PRESSURE SENSOR CALIBRATION . . . . . . . . . . 4-23
4.2.3.7 CASSETTE TYPE/PRESENCE SELECTION. . . . . . . . . 4-23
4.2.3.8 SERIAL EEPROM . . . . . . . . . . . . . . . . . . 4-24
4.3 PRINTED WIRING ASSEMBLIES . . . . . . . . . . . . . . . . . 4-24
4.3.1 POWER SUPPLY PWA . . . . . . . . . . . . . . . . . . . 4-24
4.3.2 PERIPHERAL PWA . . . . . . . . . . . . . . . . . . . . 4-25
4.3.3 PERIPHERAL INTERFACE PWA . . . . . . . . . . . . . . . 4-25
4.3.4 CPU PWA . . . . . . . . . . . . . . . . . . . . . . . . 4-26
4.3.5 DRIVER PWA . . . . . . . . . . . . . . . . . . . . . . 4-26
4.3.6 SWITCH PWA . . . . . . . . . . . . . . . . . . . . . . 4-27
4.3.7 APP PWA . . . . . . . . . . . . . . . . . . . . . . . . 4-27
4.4 REMOTE MOUNTED PERIPHERALS . . . . . . . . . . . . . . . . 4-28
4.4.1 LCD . . . . . . . . . . . . . . . . . . . . . . . . . . 4-28
4.4.2 SEALED LEAD ACID BATTERY . . . . . . . . . . . . . . . 4-28
4.4.3 BARCODE READER WAND . . . . . . . . . . . . . . . . . 4-28
4.5 MECHANICAL OVERVIEW . . . . . . . . . . . . . . . . . . . 4-28
4.5.1 CASSETTE . . . . . . . . . . . . . . . . . . . . . . . 4-29
4.5.2 MECHANISM ASSEMBLY . . . . . . . . . . . . . . . . . 4-32
4.5.2.1 MOTOR AND VALVE ASSEMBLIES. . . . . . . . . . . 4-32
4.5.2.2 A/B VALVE SUBSYSTEM . . . . . . . . . . . . . . . 4-32
4.5.2.3 INLET/OUTLET VALVE SUBSYSTEM . . . . . . . . . . 4-33
4.5.2.4 PLUNGER DRIVE SUBSYSTEM . . . . . . . . . . . . 4-33
Section 5
MAINTENANCE AND SERVICE TESTS . . . . . . . . . . . . . . . . . . 5-1
5.1 ROUTINE MAINTENANCE . . . . . . . . . . . . . . . . . . . 5-1
5.1.1 CLEANING . . . . . . . . . . . . . . . . . . . . . . . 5-1
5.1.2 SANITIZING . . . . . . . . . . . . . . . . . . . . . . 5-2
5.2 PERFORMANCE VERIFICATION TEST . . . . . . . . . . . . . . . 5-2
5.2.1 EQUIPMENT REQUIRED . . . . . . . . . . . . . . . . . . 5-3
5.2.2 INSPECTION . . . . . . . . . . . . . . . . . . . . . . 5-3
5.2.3 TEST SETUP . . . . . . . . . . . . . . . . . . . . . . . 5-4
5.2.4 SELF TEST . . . . . . . . . . . . . . . . . . . . . . . 5-4
5.2.5 CASSETTE ALARM TEST . . . . . . . . . . . . . . . . . . 5-6
5.2.6 FREE FLOW TEST . . . . . . . . . . . . . . . . . . . . . 5-6
5.2.7 DISPLAY TEST . . . . . . . . . . . . . . . . . . . . . . 5-6
5.2.8 KEYPAD VERIFICATION/FUNCTIONAL TEST . . . . . . . . . . 5-7
5.2.9 ALARM LOUDNESS TEST . . . . . . . . . . . . . . . . . 5-7
5.2.10 LOCKOUT SWITCH TEST. . . . . . . . . . . . . . . . . . 5-8
5.2.11 PROXIMAL OCCLUSION TEST. . . . . . . . . . . . . . . . 5-9
®
430-95424-002 (Rev. 01/06) iv Plum A+
3 Infusion System
Page 7
CONTENTS
5.2.12 PROXIMAL AIR-IN-LINE TEST . . . . . . . . . . . . . . . . 5-9
5.2.13 DISTAL AIR-IN-LINE TEST . . . . . . . . . . . . . . . . . 5-10
5.2.14 DISTAL OCCLUSION TEST . . . . . . . . . . . . . . . . . 5-11
5.2.15 DELIVERY ACCURACY TEST . . . . . . . . . . . . . . . . 5-12
5.2.16 NURSE CALL TEST . . . . . . . . . . . . . . . . . . . . 5-13
5.2.17 ELECTRICAL SAFETY TEST . . . . . . . . . . . . . . . . . 5-13
5.2.18 END OF THE PVT . . . . . . . . . . . . . . . . . . . . . 5-13
5.5 PERIODIC MAINTENANCE INSPECTION . . . . . . . . . . . . . . 5-14
5.6 BATTERY OPERATION OVERVIEW . . . . . . . . . . . . . . . . 5-14
Section 6
TROUBLESHOOTING . . . . . . . . . . . . . . . . . . . . . . . . 6-1
6.1 TECHNICAL ASSISTANCE. . . . . . . . . . . . . . . . . . . . 6-1
6.2 WARNING MESSAGES . . . . . . . . . . . . . . . . . . . . . 6-1
6.3 ALARM MESSAGES AND ERROR CODES . . . . . . . . . . . . . . 6-2
6.3.1 OPERATIONAL ALARM MESSAGES . . . . . . . . . . . . . 6-2
6.3.2 ERROR CODES REQUIRING TECHNICAL SERVICE . . . . . . . . 6-7
6.4 TROUBLESHOOTING PROCEDURES. . . . . . . . . . . . . . . . 6-13
Section 7
REPLACEABLE PARTS AND REPAIRS . . . . . . . . . . . . . . . . . . 7-1
7.1 REPLACEABLE PARTS . . . . . . . . . . . . . . . . . . . . . 7-1
7.2 REPLACEMENT PROCEDURES . . . . . . . . . . . . . . . . . . 7-3
7.2.1 SAFETY AND EQUIPMENT PRECAUTIONS . . . . . . . . . . . 7-3
7.2.2 REQUIRED TOOLS AND MATERIALS . . . . . . . . . . . . . 7-3
7.2.3 RUBBER FOOT PAD REPLACEMENT . . . . . . . . . . . . . 7-4
7.2.4 BATTERY, BATTERY DOOR, AND DOOR PAD REPLACEMENT . . . 7-5
7.2.5 AC POWER CORD, RETAINER, AND VELCRO STRAP
REPLACEMENT . . . . . . . . . . . . . . . . . . . . . 7-7
7.2.6 SEPARATING THE FRONT ENCLOSURE, REAR ENCLOSURE,
AND MAIN CHASSIS ASSEMBLY . . . . . . . . . . . . . . . 7-8
7.2.7 PERIPHERAL INTERFACE ASSEMBLY REPLACEMENT . . . . . . 7-10
7.2.8 PERIPHERAL PWA REPLACEMENT . . . . . . . . . . . . . . 7-11
7.2.9 PERIPHERAL COMPONENT REPLACEMENT . . . . . . . . . . 7-13
7.2.9.1 VOLUME CONTROL KNOB REPLACEMENT . . . . . . . 7-13
7.2.9.2 PERIPHERAL COVER REPLACEMENT . . . . . . . . . . 7-14
7.2.10 FRONT/REAR ENCLOSURE GASKET REPLACEMENT . . . . . . . 7-16
7.2.11 LOWER FRONT ENCLOSURE GASKET REPLACEMENT . . . . . . 7-16
7.2.11.1 EMI GASKET REPLACEMENT . . . . . . . . . . . . . 7-17
7.2.11.2 KEYPAD GASKET REPLACEMENT . . . . . . . . . . . 7-18
7.2.11.3 TOP SEAL GASKET REPLACEMENT . . . . . . . . . . 7-18
7.2.12 REAR ENCLOSURE ASSEMBLY COMPONENT REPLACEMENT . . . 7-19
7.2.12.1 POLE CLAMP ASSEMBLY AND BACKING PLATE
REPLACEMENT . . . . . . . . . . . . . . . . . . 7-21
7.2.12.2 INTERNAL AC POWER CORD REPLACEMENT . . . . . . 7-22
7.2.12.3 AC CONNECTOR REPLACEMENT . . . . . . . . . . . 7-22
7.2.12.4 FUSE REPLACEMENT . . . . . . . . . . . . . . . . 7-23
7.2.12.5 REAR ENCLOSURE GASKET REPLACEMENT . . . . . . . 7-24
7.2.13 MINIPOLE ASSEMBLY REPLACEMENT . . . . . . . . . . . . 7-24
7.2.13.1 COTTER RING REPLACEMENT . . . . . . . . . . . . 7-25
7.2.13.2 BAG HANGER REPLACEMENT . . . . . . . . . . . . 7-26
7.2.13.3 CLUTCH HOUSING REPLACEMENT . . . . . . . . . . 7-26
7.2.13.4 CLUTCH SPRING REPLACEMENT . . . . . . . . . . . 7-27
7.2.14 MAIN CHASSIS ASSEMBLY COMPONENT REPLACEMENT. . . . . 7-27
7.2.14.1 POWER SUPPLY PWA REPLACEMENT . . . . . . . . . . 7-30
Technical Service Manual v 430-95424-002 (Rev. 01/06)
Page 8
CONTENTS
7.2.14.2 KEYPAD REPLACEMENT. . . . . . . . . . . . . . . 7-30
7.2.14.3 DISPLAY ASSEMBLY REPLACEMENT . . . . . . . . . . 7-31
7.2.14.4 CPU/DRIVER CABLE REPLACEMENT . . . . . . . . . . 7-32
7.2.14.5 MOTOR POWER CABLE REPLACEMENT. . . . . . . . . 7-35
7.2.14.6 CPU PWA REPLACEMENT . . . . . . . . . . . . . . 7-35
7.2.14.7 PIEZO ALARM ASSEMBLY REPLACEMENT. . . . . . . . 7-36
7.2.14.8 MECHANISM ASSEMBLY REPLACEMENT . . . . . . . . 7-37
7.2.14.9 CASSETTE DOOR AND FLUID SHIELD REPLACEMENT . . . 7-38
7.2.14.10 OPENER HANDLE ASSEMBLY REPLACEMENT . . . . . . 7-41
Section 8
SPECIFICATIONS . . . . . . . . . . . . . . . . . . . . . . . . . . 8-1
Section 9
DRAWINGS . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-1
Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . I-1

Figures

Figure 1-1. Display and Keypad . . . . . . . . . . . . . . . . . . . . . 1-7
Figure 1-2. Biomed Settings . . . . . . . . . . . . . . . . . . . . . . 1-9
Figure 1-3. IV Parameters . . . . . . . . . . . . . . . . . . . . . . . 1-10
Figure 1-4. Alarms Log . . . . . . . . . . . . . . . . . . . . . . . . 1-11
Figure 1-5. Setting the Time and Date. . . . . . . . . . . . . . . . . . . 1-12
Figure 1-6. Common IV Parameters . . . . . . . . . . . . . . . . . . . 1-13
Figure 1-7. Macro IV Parameters . . . . . . . . . . . . . . . . . . . . 1-13
Figure 4-1. Electronic Functional Diagram . . . . . . . . . . . . . . . . . 4-2
Figure 4-2. Serial Interface to ADC. . . . . . . . . . . . . . . . . . . . 4-7
Figure 4-3. System Startup and Shutdown Timing, Battery Powered . . . . . . . . 4-15
Figure 4-4. Stepper Motor Coils . . . . . . . . . . . . . . . . . . . . . 4-18
Figure 4-5. Air Sensor Block Diagram. . . . . . . . . . . . . . . . . . . 4-20
Figure 4-6. Pressure Sensor Excitation and Amplifier Block Diagram . . . . . . . 4-22
Figure 4-7. Major Elements of the Dual-Channel Cassette . . . . . . . . . . . 4-30
Figure 4-8. Fluid Path in the Cassette . . . . . . . . . . . . . . . . . . . 4-31
Figure 4-9. Mechanism Valve Pins and Sensor Locations . . . . . . . . . . . . 4-33
Figure 5-1. Display and Keypad . . . . . . . . . . . . . . . . . . . . . 5-5
Figure 5-2. Rear View . . . . . . . . . . . . . . . . . . . . . . . . 5-8
Figure 5-3. Special Cassettes with Bubble Sensor Tips Removed . . . . . . . . . 5-10
Figure 5-4. Distal Occlusion Test Setup . . . . . . . . . . . . . . . . . . 5-12
Figure 7-1. Bottom View . . . . . . . . . . . . . . . . . . . . . . . 7-4
Figure 7-2. AC Power Cord, Retainer, Velcro Strap, and Battery Assembly . . . . . 7-6
Figure 7-3. Separating the Front Enclosure, Rear Enclosure, and Main Chassis
Assembly . . . . . . . . . . . . . . . . . . . . . . . . 7-9
Figure 7-4. Screw Placement Sequence . . . . . . . . . . . . . . . . . . 7-10
Figure 7-5. Peripheral Interface Assembly and Peripheral PWAs . . . . . . . . . 7-12
Figure 7-6. Peripheral Interface Assembly Components . . . . . . . . . . . . 7-15
Figure 7-7. Lower Front Enclosure Gaskets . . . . . . . . . . . . . . . . . 7-17
Figure 7-8. External Rear Enclosure Assembly Components . . . . . . . . . . . 7-19
Figure 7-9. Internal Rear Enclosure Assembly Components . . . . . . . . . . . 7-20
Figure 7-10. Minipole Assembly . . . . . . . . . . . . . . . . . . . . . 7-25
430-95424-002 (Rev. 01/06) vi Plum A+
®
3 Infusion System
Page 9
CONTENTS
Figure 7-11. Main Chassis Components (1 of 2) . . . . . . . . . . . . . . . . 7-28
Figure 7-11. Main Chassis Components (2 of 2) . . . . . . . . . . . . . . . . 7-29
Figure 7-12. CPU/Driver Cable Routing . . . . . . . . . . . . . . . . . . 7-33
Figure 7-13. Ferrite Tape Positioning (1 of 2) . . . . . . . . . . . . . . . . . 7-34
Figure 7-13. Ferrite Tape Positioning (2 of 2) . . . . . . . . . . . . . . . . . 7-34
Figure 7-14. Fluid Shield Replacement . . . . . . . . . . . . . . . . . . . 7-39
Figure 7-15. Cassette Door and Opener Handle Assembly Replacement . . . . . . . 7-40
Figure 9-1. Illustrated Parts Breakdown . . . . . . . . . . . . . . . . . . 9-5
Figure 9-2. Front Enclosures, Rear Enclosure, and Main Chassis Assembly . . . . . 9-9
Figure 9-3. Front Enclosure Assemblies . . . . . . . . . . . . . . . . . . 9-11
Figure 9-4. Rear Enclosure Assembly . . . . . . . . . . . . . . . . . . . 9-13
Figure 9-5. Peripheral Interface Assembly . . . . . . . . . . . . . . . . . 9-17
Figure 9-6. Main Chassis Assembly . . . . . . . . . . . . . . . . . . . 9-19
Figure 9-7. CPU PWA, Display, and Keypad . . . . . . . . . . . . . . . . 9-23
Figure 9-8. CPU PWA and Main Chassis . . . . . . . . . . . . . . . . . . 9-25
Figure 9-9. AC Power Cord, Retainer, Batteries, and Minipole . . . . . . . . . . 9-27
Figure 9-10. Mechanism Assembly . . . . . . . . . . . . . . . . . . . . 9-29
Figure 9-11. Power Supply PWA Schematic . . . . . . . . . . . . . . . . . 9-31
Figure 9-12. Peripheral Interface PWA Schematic . . . . . . . . . . . . . . . 9-51
Figure 9-13. Peripheral PWA Schematic . . . . . . . . . . . . . . . . . . 9-63
Figure 9-14. CPU PWA Schematic. . . . . . . . . . . . . . . . . . . . . 9-71
Figure 9-15. Driver PWA Schematic . . . . . . . . . . . . . . . . . . . . 9-91
Figure 9-16. Switch PWA Schematic . . . . . . . . . . . . . . . . . . . . 9-97
Figure 9-17. APP PWA Schematic . . . . . . . . . . . . . . . . . . . . . 9-99

Tables

Table 1-1. Conventions. . . . . . . . . . . . . . . . . . . . . . . . 1-2
Table 1-2. System Configuration Data . . . . . . . . . . . . . . . . . . 1-9
Table 4-1. Analog Inputs . . . . . . . . . . . . . . . . . . . . . . . 4-7
Table 4-2. Keypad Map. . . . . . . . . . . . . . . . . . . . . . . . 4-8
Table 4-3. CPU-Power Supply Interface . . . . . . . . . . . . . . . . . . 4-10
Table 4-4. CPU-Mechanism Interface Signals . . . . . . . . . . . . . . . . 4-11
Table 4-5. Power Supply PWA Interface Connections . . . . . . . . . . . . . 4-24
Table 4-6. Peripheral PWA Interface Connections . . . . . . . . . . . . . . 4-25
Table 4-7. Peripheral Interface PWA Interface Connections . . . . . . . . . . . 4-25
Table 4-8. CPU PWA Interface Connections . . . . . . . . . . . . . . . . 4-26
Table 4-9. Driver PWA Interface Connections. . . . . . . . . . . . . . . . 4-27
Table 4-10. APP PWA Interface Connections . . . . . . . . . . . . . . . . 4-27
Table 5-1. Cleaning Solutions . . . . . . . . . . . . . . . . . . . . . 5-2
Table 6-1. Warning Messages . . . . . . . . . . . . . . . . . . . . . 6-2
Table 6-2. Operational Alarm Messages and Corrective Actions . . . . . . . . . 6-3
Table 6-3. Error Codes Requiring Technical Service . . . . . . . . . . . . . 6-7
Table 6-4. Troubleshooting with the PVT . . . . . . . . . . . . . . . . . 6-13
Table 9-1. Drawings . . . . . . . . . . . . . . . . . . . . . . . . . 9-1
Table 9-2. IPB for the Infuser . . . . . . . . . . . . . . . . . . . . . 9-2
Technical Service Manual vii 430-95424-002 (Rev. 01/06)
Page 10
CONTENTS
This page intentionally left blank.
430-95424-002 (Rev. 01/06) viii Plum A+
®
3 Infusion System
Page 11
Section 1

INTRODUCTION

The Hospira Plum A+®3 infusion system is designed to meet the growing demand for hospital wide device standardization. The system consists of three component infusers, which are designated line 1, line 2, and line 3. By incorporating three lines into one unit, the Plum A+ delivery capabilities. The Plum A+ needs. Compatibility with the LifeCare accessories make the Plum A+
1.1

SCOPE

This manual is organized into the following sections:
Section 1 Introduction
Section 2 Warranty
Section 3 System Operating Manual
Section 4 Theory of Operation
Section 5 Maintenance and Service Tests
Section 6 Troubleshooting
Section 7 Replaceable Parts and Repairs
Section 8 Specifications
Section 9 Drawings
Appendices
Index
Technical Service Bulletins
®
3 provides three primary lines, three secondary lines, and piggyback fluid
®
3 serves a wide range of general floor and critical care
®
3 infusion system convenient and cost-effective.
®
5000 PlumSet® administration sets and
If a problem in device operation cannot be resolved using the information in this manual,
contact Hospira (see Section 6.1).
Specific instructions for operating the device are contained in the Plum A+ System Operating Manual. Provision is made for the inclusion of the system operating manual in Section 3 of this manual.
®
and Plum A+®3
Note: The terms “infusion system”, “infuser”, and “device” are used interchangeably
throughout the manual.
Note: Figures are rendered as graphic representations to approximate actual product.
Therefore, figures may not exactly reflect the product.
Note: Screen representations are examples only, and do not necessarily reflect the
most current software version.
Technical Service Manual 1 - 1 430-95424-002 (Rev. 01/06)
Page 12
SECTION 1 INTRODUCTION
1.2

CONVENTIONS

The conventions listed in Table 1-1 are used throughout this manual.
Table 1-1. Conventions
Convention Application Example
Italic Reference to a section, figure,
table, or publication
[ALL CAPS] In-text references to keys
and touchswitches ALL CAPS Screen displays CASSETTE TEST IN PROGRESS Bold Emphasis CAUTION: Use proper ESD grounding
Throughout this manual, warnings, cautions, and notes are used to emphasize important information as follows:
WARNING:
CAUTION: A CAUTION usually appears in front of a procedure or statement. It contains information that could prevent hardware failure, irreversible damage to equipment, or loss of data.
A WARNING CONTAINS SPECIAL SAFETY EMPHASIS AND MUST BE OBSERVED AT ALL TIMES. FAILURE TO OBSERVE A WARNING MAY RESULT IN PATIENT INJURY AND BE LIFE-THREATENING.
(see Section 6.1)
[START]
techniques when handling components.
Note: A note highlights information that helps explain a concept or procedure.
1.3

COMPONENT DESIGNATORS

Components are indicated by alpha-numeric designators, as follows:
Battery BT Diode D Resistor R
Capacitor C Fuse F Switch SW
Crystal Y Integrated Circuit U Transistor Q
The number following the letter is a unique value for each type of component (e.g., R1, R2).
Note: Alpha-numeric designators may be followed with a dash (-) number that
indicates a pin number for that component. For example, U15-13 is pin 13 of the encoder chip [U15] on the interface PWA.
®
430-95424-002 (Rev. 01/06) 1 - 2 Plum A+
3 Infusion System
Page 13
1.4 ACRONYMS AND ABBREVIATIONS
1.4

ACRONYMS AND ABBREVIATIONS

Acronyms and abbreviations used in this manual are as follows:
A Ampere
AC Alternating current
A/D Analog-to-digital
ADC Analog-to-digital converter
APP Air, pressure, and pin
BCR Barcode reader
CCA Clinical care area
CCFT Cold cathode fluorescent tube
CMOS Complementary metal-oxide semiconductor
CPU Central processing unit
DAC Digital-to-analog converter
DC Direct current
DIP Dual in-line package
DMA Direct memory access
DMM Digital multimeter
DPM Digital pressure meter
ECG Electrocardiograph
EEG Electroencephalogram
EEPROM Electrically erasable/programmable read-only memory
EMG Electromyogram
EMI Electromagnetic interference
ESD Electrostatic discharge
ETO Ethylene oxide
FPGA Field programmable gate array
FSR Force sensing resistor
hr Hour
Hz Hertz
ID Identification
I/O Input/output
IPB Illustrated parts breakdown
IV Intravenous
KB Kilobyte
kHz Kilohertz
KVO Keep vein open
lbs Pounds
Technical Service Manual 1 - 3 430-95424-002 (Rev. 01/06)
Page 14
SECTION 1 INTRODUCTION
MOSFET Metal-oxide semiconductor field-effect transistor
Op-amp Operational amplifier
LCD Liquid crystal display
LED Light emitting diode
L/S Line select
MB Megabyte
MHz Megahertz
min Minute
mL Milliliter
mL/hr Milliliter per hour
MMIO Memory-mapped input/output
ms Millisecond
nF nanofarad
pF picofarad
PROM Programmable read-only memory
PVT Performance verification test
PWA Printed wiring assembly
PWM Pulse width modulator
RAM Random-access memory
rms Root-mean-square
RTC Real-time clock
SCC Serial communication controller
SCP Serial communication port
SMT Surface mount technology
SPI Serial peripheral interface
SRAM Static random access memory
TQFP Thin quad flat pack
V Volt
V
Volts AC
AC
V
Collector supply voltage
CC
VCO Voltage-controlled oscillator
Volts DC
V
DC
VSC 5 V
supply circuitry
DC
VSO Voltage sweep oscillator
VTBI Volume to be infused
WDI Watchdog input
µA Microampere
µL Microliter
µV Microvolt
µsec Microsecond
430-95424-002 (Rev. 01/06) 1 - 4 Plum A+
®
3 Infusion System
Page 15
1.5 USER QUALIFICATION
1.5

USER QUALIFICATION

The infusion system is intended for use at the direction or under the supervision of licensed physicians or certified healthcare professionals who are trained in the use of the infusion system and the administration of parenteral and enteral fluids and drugs, and whole blood or red blood cell components. Training should emphasize preventing related IV complications, including appropriate precautions to prevent accidental infusion of air. The epidural route can be used to provide anesthesia or analgesia.
1.6

ARTIFACTS

Nonhazardous, low-level electrical potentials are commonly observed when fluids are administered using infusion devices. These potentials are well within accepted safety standards, but may create artifacts on voltage-sensing equipment such as ECG, EMG, and EEG machines. These artifacts vary at a rate that is associated with the infusion rate. If the monitoring machine is not operating correctly or has loose or defective connections to its sensing electrodes, these artifacts may be accentuated so as to simulate actual physiological signals. To determine if the abnormality in the monitoring equipment is caused by the infuser instead of some other source in the environment, set the device so that it is temporarily not delivering fluid. Disappearance of the abnormality indicates that it was probably caused by electronic noise generated by the infuser. Proper setup and maintenance of the monitoring equipment should eliminate the artifact. Refer to the appropriate monitoring system documentation for setup and maintenance instructions.
1.7

INSTRUMENT INSTALLATION PROCEDURE

CAUTION: Infusion system damage may occur unless proper care is exercised during product unpacking and installation. The battery may not be fully charged upon receipt of the infuser. Do not place the infuser in service if it fails the self test.
CAUTION: Infusion system performance may be degraded by electromagnetic interference (EMI) from devices such as electrosurgical units, cellular phones, and two-way radios. Operation of the infusion system under such conditions should be avoided.
Accessory equipment connected to the analog and digital interfaces must be certified according to the respective IEC standards (e.g., IEC 60950 for data processing equipment, and IEC 60601-1 for medical equipment). Furthermore, all configurations shall comply with the system standard IEC 60601-1-1. Any person who connects additional equipment to the signal input or output part configures a medical system, and is therefore responsible for ensuring that the system complies with the requirements of IEC 60601-1-1. If in doubt,
contact Hospira Technical Support Operations (see Section 6.1).
The instrument installation procedure consists of unpacking, inspection, and self test.
Technical Service Manual 1 - 5 430-95424-002 (Rev. 01/06)
Page 16
SECTION 1 INTRODUCTION
1.7.1

UNPACKING

Inspect the shipping container as detailed in Section 1.7.2. Use care when unpacking the infusion system. Retain the packing slip and save all packing material in the event it is necessary to return the infuser to the factory. Verify the shipping container contains a copy of the system operating manual.
1.7.2

INSPECTION

Inspect the shipping container for damage. Should any damage be found, contact the delivering carrier immediately.
CAUTION: Inspect the infuser for evidence of damage. Do not use the device if it appears to be damaged. Should damage be found, contact Hospira (see Section 6.1).
Inspect the infusion system periodically for signs of defects such as worn accessories, broken connections, or damaged cable assemblies. Also inspect the infuser after repair or during cleaning. Replace any damaged or defective external parts.
1.7.3

SELF TEST

CAUTION: Do not place the infuser in service if the self test fails.
Note: Do not place the infuser in service if the battery is not fully charged. To make
certain the battery is fully charged, connect the infuser to AC power for six hours
(see Section 8).
Note: If an alarm condition occurs during the self test, cycle the power and repeat the
self test. If the alarm condition recurs, note the message and take corrective action
(see Section 6). Repeat the self test. If the alarm condition continues to recur, remove
the infuser from service and contact Hospira.
Note: Records prior to the date the infuser is received may be from the manufacturing
process. Disregard any events from dates prior to receipt of the infuser.
Note: When performing the self test, line 1, line 2, and line 3 must be tested.
However, if appropriate, the test may be performed on all lines concurrently.
To perform the self test, see Figure 1-1, and proceed as follows:
1. Connect the AC power cord to a grounded AC outlet. Verify the charging/line
indicator CHARGE illuminates and an alarm beep sounds.
2. Without a cassette installed, press [ON/OFF] to turn on the infuser.
3. The LCD screen briefly displays the SELF TEST screen (see Figure 1-1).
Note: If the SELF TEST screen does not appear, contact Hospira.
4. After the self test is complete, the message “INSERT PLUM SET CLOSE LEVER” appears.
5. Verify the time and date. To set the time and date, see Section 1.8.3.
®
430-95424-002 (Rev. 01/06) 1 - 6 Plum A+
3 Infusion System
Page 17
1.7 INSTRUMENT INSTALLATION PROCEDURE
ON/OFF
6. Open the cassette door and insert a primed cassette. Close the cassette door. The cassette test is complete when the “CASSETTE TEST IN PROGRESS” message disappears.
Note: The message “MECHANISM INITIALIZATION IN PROGRESS” may briefly
appear prior to the “CASSETTE TEST IN PROGRESS” message.
7. The “CLEAR SETTINGS?” message may appear. Press the [YES] softkey.
8. Press [ON/OFF] to turn off the infuser.
LINE FLOW
INDICATORS
A B
HOSPIRA Plum A+
STATUS REGION
MESSAGE REGION
SOFTKEY LABEL REGION
LINE INDICATOR
Version X.XX MM/DD/YY
START
STOP
CHARGE
ON/OFF
Copyright Hospira Inc.
2005
System Self Test
In Progress
1
4
2 5
78
CLEAR
0
3 6
9
.
WORKING
REGION
SOFTKEYS
SELECT
KEYPAD
SILENCE
05K01002
Figure 1-1. Display and Keypad
Technical Service Manual 1 - 7 430-95424-002 (Rev. 01/06)
Page 18
SECTION 1 INTRODUCTION
1.8

BIOMED SETTINGS

The biomed settings screens contain the following options that can be changed or reviewed by qualified personnel:
- IV screen parameters
- Alarms log
- Set time and date
All infusers (new or refurbished) are shipped with factory settings (see Table 1-2).
Note: Biomed screens do not time out for the Infuser Idle alarm or No Action alarm.Note: The battery will not be detected in the biomed service mode.
To access the biomed settings, proceed as follows:
1. Open the door and turn on the device. The infusion system will perform a self test.
2. After the self test is complete, the message “INSERT PLUM SET CLOSE LEVER”
appears. Press the decimal [.] key, then [START], and verify the BIOMED SETTINGS screen is displayed (see Figure 1-2)
.
Note: The [CHANGE BATTERY] softkey does not appear on earlier versions
of the Plum A+
®
3.
430-95424-002 (Rev. 01/06) 1 - 8 Plum A+
®
3 Infusion System
Page 19
1.8 BIOMED SETTINGS
Table 1-2. System Configuration Data
Data Options Range Factory Setting
Maximum macro IV mode delivery rate 0.1 - 99.9 mL/hr and
100 - 999 mL/hr
Macro distal occlusion alarm (pressure level)
Deliver together enable Concurrent or Piggyback Piggyback Delayed start/standby enable Yes or No Yes Continue rate Rate or KVO KVO Nurse callback default Yes or No No Time (24 hr) 00:00 - 23:59 in
Date 1/1/2002 - 12/31/2098 Factory date
1 to 15 psi 6 psi
one minute increments
999 mL/hr
Factory time
BIOMED SETTINGS
IV Screen Parameters
Alarm Log
Set Time and Date
Select, then Choose
Change Battery
Figure 1-2. Biomed Settings
Choose
05K03010
Technical Service Manual 1 - 9 430-95424-002 (Rev. 01/06)
Page 20
SECTION 1 INTRODUCTION
1.8.1

IV PARAMETERS

The IV parameters screen contains common IV parameters and macro IV parameters
(see Figure 1-3).
To change the IV parameters, see Figure 1-6 and Figure 1-7, then proceed as follows:
1. Access the biomed settings screen as described in Section 1.8.
2. Select IV Screen Parameters, and press [CHOOSE].
3. Select the parameters to be changed, and press [CHOOSE].
4. Using the [CHANGE VALUE] softkey, select the desired value, and press [ENTER].
5. Repeat step 3 and step 4 for each parameter to be changed.
6. If there are no other changes, turn off the infuser.
BIOMED SETTINGS
IV Parameters
Common IV Parameters
Macro IV Parameters
Select, then Choose
Choose Back
Figure 1-3. IV Parameters
00H03003
430-95424-002 (Rev. 01/06) 1 - 10 Plum A+
®
3 Infusion System
Page 21
1.8 BIOMED SETTINGS
1.8.2

ALARMS LOG

Note: The alarms log will retain the latest 40 alarm and malfunction codes, listed
in order from the most current to the oldest.
To view the alarms log, see Figure 1-2 and Figure 1-4, then proceed as follows:
1. Access the biomed settings screen as described in Section 1.8.
2. Select Alarms Log, and press [CHOOSE}.
3. Use the [PAGE UP] and [PAGE DOWN] softkeys to view the alarms log.
4. Press the [BACK] softkey to exit the alarms log and return to the main biomed settings screen.
ALARMS LOG
6/23/04 01:43:01 E437 S/W Failure # 202 6/23/04 09:18:10 N190 Neg. Prox. Occl. A 6/22/04 23:44:11 N102 Infuser Idle 2 minutes 6/22/04 21:43:14 N161 Line A VTBI complete 6/22/04 11:44:20 N106 Distal occlusion 6/22/04 09:43:07 N161 Line A VTBI complete 6/22/04 06:23:20 N160 Line B VTBI complete 6/22/04 03:40:13 N101 No action alarm
Page
Up
Figure 1-4. Alarms Log
Page
Down
Back
05K03008
Technical Service Manual 1 - 11 430-95424-002 (Rev. 01/06)
Page 22
SECTION 1 INTRODUCTION
1.8.3

SETTING THE TIME AND DATE

Note: The infusion system will automatically display February 29 on leap years.Note: Daylight savings and time zone changes must be made manually.
To set the time and date, see Figure 1-5, then proceed as follows:
1. Access the biomed settings screen as described in Section 1.8.
2. Select Set Time and Date, and press [CHOOSE].
3. Select the parameter to be changed.
4. Enter the desired value.
5. Repeat step 3 and step 4 for each parameter to be changed.
6. Verify the time and date are correct, then press [ENTER] to return to the biomed settings screen.
7. If there are no other changes, turn off the infuser.
BIOMED SETTINGS
Set Time and Date
Time : 22 hr:min Year 2005 Month 02 Day 14
Enter value using keypad
Figure 1-5. Setting the Time and Date
14
Enter
Cancel/
Back
05K13004
430-95424-002 (Rev. 01/06) 1 - 12 Plum A+
®
3 Infusion System
Page 23
BIOMED SETTINGS
Common IV Parameters
1.8 BIOMED SETTINGS
Continue Rate Deliver Together
KVO
Concurrent
Enable Delay/Standby Callback Default
Select using Change Value
Change
Value
Figure 1-6. Common IV Parameters
Enter
Cancel/
BIOMED SETTINGS
Yes
No
Back
02K03004
Macro IV Parameters
Default Distal Press Max Rate
Enter Value using keypad
Enter
Figure 1-7. Macro IV Parameters
6.0 psi
999 mL/hr
Cancel/
Back
00K13005
Technical Service Manual 1 - 13 430-95424-002 (Rev. 01/06)
Page 24
SECTION 1 INTRODUCTION
This page intentionally left blank.
430-95424-002 (Rev. 01/06) 1 - 14 Plum A+
®
3 Infusion System
Page 25
Section 2

WARRANTY

Subject to the terms and conditions herein, Hospira, Inc., herein referred to as Hospira, warrants that (a) the product shall conform to Hospira's standard specifications and be free from defects in material and workmanship under normal use and service for a period of one year after purchase, and (b) the replaceable battery shall be free from defects in material and workmanship under normal use and service for a period of 90 days after purchase. Hospira makes no other warranties, express or implied, as to merchantability, fitness for a particular purpose, or any other matter.
Purchaser's exclusive remedy shall be, at Hospira's option, the repair or replacement of the product. In no event shall Hospira's liability arising out of any cause whatsoever (whether such cause be based in contract, negligence, strict liability, other tort, or otherwise) exceed the price of such product, and in no event shall Hospira be liable for incidental, consequential, or special damages or losses or for lost business, revenues, or profits. Warranty product returned to Hospira must be properly packaged and sent freight prepaid.
The foregoing warranty shall be void in the event the product has been misused, damaged, altered, or used other than in accordance with product manuals so as, in Hospira's judgment, to affect its stability or reliability, or in the event the serial or lot number has been altered, effaced, or removed.
The foregoing warranty shall also be void in the event any person, including the Purchaser, performs or attempts to perform any major repair or other service on the product without having been trained by an authorized representative of Hospira and using Hospira documentation and approved spare parts. For purposes of the preceding sentence, “major repair or other service” means any repair or service other than the replacement of accessory items such as batteries, flow detectors, detachable AC power cords, and patient pendants.
In providing any parts for repair or service of the product, Hospira shall have no responsibility or liability for the actions or inactions of the person performing such repair or service, regardless of whether such person has been trained to perform such repair or service. It is understood and acknowledged that any person other than an Hospira representative performing repair or service is not an authorized agent of Hospira.
Technical Service Manual 2 - 1 430-95424-002 (Rev. 01/06)
Page 26
SECTION 2 WARRANTY
This page intentionally left blank.
430-95424-002 (Rev. 01/06) 2 - 2 Plum A+
®
3 Infusion System
Page 27
Section 3

SYSTEM OPERATING MANUAL

A copy of the system operating manual is included with every Plum A+®3 infusion system. Insert a copy here for convenient reference. If a copy of the system operating manual is not available, contact Hospira Technical Support Operations (see Section 6.1).
Technical Service Manual 3 - 1 430-95424-002 (Rev. 01/06)
Page 28
SECTION 3 SYSTEM OPERATING MANUAL
This page intentionally left blank.
430-95424-002 (Rev. 01/06) 3 - 2 Plum A+
®
3 Infusion System
Page 29
Section 4

THEORY OF OPERATION

This section describes the Plum A+®3 theory of operation. The theory of operation details the general description, electronic subsystem overview, printed wiring assemblies, remote mounted peripherals, and mechanical overview of the infusion system. Related drawings are provided in Section 9.
4.1

GENERAL DESCRIPTION

The infusion system includes the following features:
- Dose calculation
- Loading dose
-Multi-step programming
- Therapy selection
-Nurse call
- Delayed start setting
- Standby mode
- Drug label library
- Piggyback and concurrent delivery modes
-Titration
- 0.1-99.9 mL/hr flow rate range for both lines (in 0.1 mL/hr increments)
- 100-999 mL/hr flow rate range for both lines (in 1.0 mL/hr increments)
- Anti free-flow protection
- Air removal/backpriming
- Air detection (proximal and distal)
- Serial communication
-Alarm history
- Volumes infused (A, B, total volumes)
- KVO at dose end (1.0 mL/hr or less depending on delivery rate) or continue rate to continue
- Variable distal pressure setting
- Nonpulsatile volumetric accuracy
- Microprocessor control
-Large display
- Panel back illumination on mains power
- Lockout switch
- Standard fullfill, partfill, syringe, and vial use
- Enteral and parenteral fluid delivery
- Blood and blood product delivery
- Battery gauge
Technical Service Manual 4 - 1 430-95424-002 (Rev. 01/06)
- Wide range of standard and specialty administration sets
Page 30
SECTION 4 THEORY OF OPERATION
Alarms include the following:
- Distal occlusion
- Proximal occlusion
- Proximal air-in-line
- Distal air-in-line
-Low battery
- Door open while pumping
- Lockout violation
-VTBI complete
- Valve/cassette test failure
-Nurse call
- No action alarm
- Infuser idle for two minutes
4.2

ELECTRONIC SUBSYSTEM OVERVIEW

This section describes the function and electronic circuitry of three main subsystems in the infusion system: CPU subsystem, power supply subsystem, and mechanism subsystem (see Figure 4-1). Schematic diagrams of PWAs are located in Section 9.
Note: An asterisk (*) denotes an active low or negative true logic signal.
Battery
Power Supply PWA
Peripheral PWA
Pressure Sensors
SYSTEM 1
Pressure Sensors
SYSTEM 2
Pressure Sensors
SYSTEM 3
Plunger, LS, & IO Motors
MECHANISM
Battery
Plunger, LS, & IO Motors
MECHANISM
Battery
Plunger, LS, & IO Motors
MECHANISM
APP PWA
Switches PWA
Flex FSR
Power Supply PWA
APP PWA
Switches PWA
Flex FSR
Power Supply PWA
APP PWA
Switches PWA
Flex FSR
Driver PWA
Driver
PWA
Driver
PWA
CPU PWA
CPU PWA
CPU PWA
LCD Display
Keypad
LEDs
On/Off Switch
FRONT PANEL
Main Piezo Buzzer
Peripheral PWA
LCD Display
Keypad
LEDs
On/Off Switch
FRONT PANEL
Main Piezo Buzzer
Peripheral Interface PWA
LCD Display
Keypad
LEDs
On/Off Switch
FRONT PANEL
Main Piezo Buzzer
Figure 4-1. Electronic Functional Diagram
430-95424-002 (Rev. 01/06) 4 - 2 Plum A+
05K07015
®
3 Infusion System
Page 31
4.2 ELECTRONIC SUBSYSTEM OVERVIEW
4.2.1

CPU SUBSYSTEM

The CPU subsystem contains the main microcontroller, which is responsible for controlling the display/keyboard interface, external communications interfaces, barcode reader (BCR) interface, and system management.
The CPU subsystem provides the following functions:
- External memory devices access
-LCD interfaces
-Real-time clock generator interface
-System watchdog
- Analog-to-digital and digital-to-analog converter interface
- Keypad interfaces
- Control and monitor status signals, such as LEDs, audible alarms, volume control, nurse call switch, and lockout switch
- Serial communication with host computer (DataPort) and barcode reader
- Power supply subsystem interface
- Mechanism subsystem interface
4.2.1.1
CPU
The central processing unit is a Motorola MC68302 CPU. The CPU has a closely coupled 16 bit data bus and 24 bit address bus, MC68000 microprocessor core, a system integration block for peripherals, and an RISC communications processor. The MC68302 is packaged in a 144 pin thin quad flat pack (TQFP) package and operates from a 3.3 V power supply. The on-chip peripheral devices are isolated from the system through the dual port RAM. The 1152 byte dual port RAM has 576 bytes of system RAM and 576 bytes of parameter RAM, which contains various peripheral registers, parameters, and the buffer descriptors for each of the three serial communication controller (SCC) channels and the serial communication port (SCP) channels. The 24 bit address bus is capable of accessing up to 16 MB of data.
4.2.1.2
DC
SYSTEM MEMORY ADDRESS MAP
The CPU has a 24 bit address bus when combined with UDS*/A0. The address bus is a bi-directional, three state bus capable of addressing 16 MB of data that is configured as 16 bits per word (including the IMP internal address space). Each of the four programmable chip-select lines has two registers that define the starting address of a particular address space and the block size.
Technical Service Manual 4 - 3 430-95424-002 (Rev. 01/06)
Page 32
SECTION 4 THEORY OF OPERATION
4.2.1.3
PROGRAMMABLE READ-ONLY MEMORY
The CPU subsystem has two 512 K x 8 bit programmable read-only memory (PROM) memory devices, which provide a total of 1024 KB. The PROM space is expandable up to 2 MB. The PROM memory devices operate off the 3.3 V
supply. The CPU chip-select
DC
0 pin (CS0*), is connected to the PROM chip-enable (CE*) pin (signal CSROM*). This special chip-select signal can support bootstrap operation after reset. The interface to the CPU is the 16 bit data bus, and a 19 bit address bus. The address bus is connected to the ADDR<19:1> lines, and the data bus is connected to the DATA<15:0> lines.
4.2.1.4
STATIC RANDOM ACCESS MEMORY
There are two 512 K x 8 bit CMOS static random access memory (SRAM) devices, which provide a total of 1024 KB of data memory. During an SRAM read or write cycle, the chip-enable (CE*) is controlled by the CPU chip-select pin 1 (CS1*, signal name (CSRAM*)). The SRAM space is expandable up to 2 MB. The SRAM operates off the 3.3 V The CPU subsystem includes the additional SRAM for video buffer and real-time clock
(see Section 4.2.1.6 and Section 4.2.1.9).
supply.
DC
4.2.1.5
CONTROL LOGIC
The CPU PWA uses field programmable gate arrays (FPGA), which are high density, high speed, I/O intensive general purpose devices. They are used to implement all the digital control functions, including: memory-map address decoding; memory read-write enable; direct memory access (DMA) request; I/O status signals; chip-select control; motor control; sensor select; and power up/system reset control.
4.2.1.6
LCD CONTROLLER
The liquid crystal display (LCD) controller is used to interface the LCD to the CPU. The device displays layered text and graphics, scrolls the display in any direction, and partitions the display into multiple screens. It stores bit-mapped graphic data in external frame buffer memory. The display controller functions include: transferring data from the controlling microprocessor to the buffer memory; reading memory data; converting data to display pixels; and generating timing signals for the buffer memory and LCD panel. The LCD controller accesses 32 KB of frame buffer SRAM (video) via the controller’s video address and data busses (VA<14:0> and VD<7:0>). The LCD controller external clock frequency is 8 MHz. The LCD controller and the display memory are operated off the 3.3 V interface with the 5 V
supply. The output signal levels are shifted up to 5 VDC by buffers for
DC
LCD panel.
DC
The interface to the CPU is through the lower 8 bits of the data bus, which is connected to DATA<7:0> lines, address line A1, and LCD chip-select signal CSLCD* (CS2*). This controller is also configured as 8080 family compatible interface device with all the control signals, such as WRLCD* (WR*) and RDLCD* (RD*), generated by the FPGA logic.
®
430-95424-002 (Rev. 01/06) 4 - 4 Plum A+
3 Infusion System
Page 33
4.2 ELECTRONIC SUBSYSTEM OVERVIEW
4.2.1.7
LCD BACKLIGHT CONTROL
The LCD panel is backlit by a cold cathode fluorescent tube (CCFT) lamp. The CCFT lamp requires 300 V to deliver a current regulated sine wave to the lamp. A switching regulator regulates the CCFT current by monitoring feedback pin 3, and varies its output duty cycle to drive a DC/AC inverter. Intensity control is achieved by superimposing a DC control signal with the feedback signal. The DC control signal is sourced by a voltage divider consisting of a digitally controlled non-volatile potentiometer and three series diodes.
The CPU can adjust LCD backlight intensity by selecting the digitally controlled non-volatile potentiometer and controlling TUBU/D and TUBINC* signals. The potentiometer has a five bit up/down counter with non-volatile memory. It is used to store one of 31 settings of the potentiometer. Each count represents 323with a range of 323 to 10 KΩ. The current counter value is stored in non-volatile memory after CSTUB* is returned high while the TUBINC* input is also high. The current counter value is not stored if CSTUB* is returned high and TUBINC* is low. The CCFT intensity is directly proportional to the CCFT current, where 0 mA is maximum intensity. The CCFT current is inversely proportional to the counter value.
to operate; a current controlled DC-to-AC voltage inverter circuit is used
rms
is minimum intensity and 5 mA
rms
rms
4.2.1.8
LCD CONTRAST CONTROL
A digitally adjustable LCD bias supply is used to control the LCD contrast over a range of -24 converter (DAC). The CPU provides two signals, LCDADJ (ADJ) and LCDCTL (CTL), to interface with this device. On power up or after a reset, the counter sets the DAC output to the mid-range value. Each rising edge of LCDADJ increments the DAC output. When incremented beyond full scale, the counter rolls over and sets the DAC to the minimum value. Therefore, a single pulse applied to LCDADJ increases the DAC set point by one step, and 63 pulses decrease the set point by one step.
4.2.1.9
to -8 VDC. It is digitally adjustable in 64 equal steps by an internal digital-to-analog
REAL-TIME CLOCK
The watchdog timekeeper chip includes a complete real-time clock/calendar (RTC), watchdog timer, alarm, and interval timer. The time/date information includes hundredths of seconds, seconds, minutes, hours, date, month, and year. The date at the end of the month is automatically adjusted for months with less than 31 days, including correction for leap year. The watchdog timekeeper operates in either 24-hour or 12-hour format with an AM/PM indicator. The device can be programmed to set up an interval timer, and it can generate an alarm every day, hour, or minute. These alarm functions ma y be use d to schedule real-time related activities. A parallel resonant 32,768 Hz crystal oscillator drives the internal time base.
The external interface is a separate (non-multiplexed) 8 bit data bus and 6 bit address bus, with a contiguous address space of 64 bytes. When system power is turned off, a battery voltage input is available, which makes the RTC data non-volatile. The address bus is connected to the ADDR<6:1> lines, and the data bus is connected to DATA<7:0> lines. Since the CPU accesses are 16 bits wide, the RTC data is on the lower byte of the word. The RTC chip-enable pin (CE*) is active low enabled for read and write operations. It is driven by the FPGA control logic, chip-select RTC signal (CSRTC*), which involves address decoding circuitry (see Section 4.2.1.2).
Technical Service Manual 4 - 5 430-95424-002 (Rev. 01/06)
Page 34
SECTION 4 THEORY OF OPERATION
4.2.1.10
VOLTAGE MONITOR WATCHDOG TIMER
It is important to protect the system during power transitions, and the CPU is reset after the V automatic reset output during power up, power down, or brownout conditions. When the V
CC
and holds the microprocessor in reset for approximately 200 ms after V the threshold. The supervisory circuit includes a chip-select inhibit circuit, which is used to disable access to the real-time clock’s non-volatile SRAM during power transitions and power down mode.
This device also provides a watchdog timer function to monitor the activity of the microprocessor. To service the watchdog timer immediately after reset, the device has a longer time-out period (1.6 second minimum) right after a reset. The normal time-out period (70 ms minimum) is effective after the first transition of watchdog input (WDI) after RESET* is inactive. If the microprocessor does not toggle WDI within the time-out period, both RESET* and watchdog out (WDO*) outputs are asserted low. The RESET* remains active low for a minimum of 140 ms and it resets the CPU. The WDO* remains low as long as the WDI remains either high or low for longer than the watchdog time-out period. After a reset, the software reads this memory-mapped bit to determine if the latest reset was a watchdog time-out.
power supply is applied. The microprocessor supervisory circuit generates an
CC
falls below the reset threshold voltage of 2.9 VDC, the reset signal (RESET*) goes low
rises above
CC
4.2.1.11
ANALOG-TO-DIGITAL CONVERTER
The analog-to-digital converter (ADC) monitors the proximal pressure sensor, distal pressure sensor, proximal air sensor, distal air sensor, battery charge/discharge current, battery voltage, buzzer test signal, LCD contrast voltage, CCFT test signal, and two chopper motor drive reference voltages. The ADC is an advanced 10 bit accurate, 11 channel, switched-capacitor, successive-approximation device. It has three inputs and a three-state output (chip-select, I/O clock, address input, and data out) that provide a direct four-wire interface to the serial communication port of the CPU. The ADC is designed to be used in conjunction with multiple serial devices on a common bus; consequently, the data-out pin is driven only when the chip-select (CS*) pin is asserted. Figure 4-2 illustrates the serial interface between the ADC and the CPU.
In addition to a high-speed ADC and versatile control capability, this device has an on-chip 14 channel multiplexer that can select any one of 11 analog inputs or any one of three internal self test voltages. The sample-and-hold function is automatic. The end-of-conversion (EOC) output goes high to indicate that conversion is complete. The CPU polls the EOC signal.
Channel selection and conversion results are transferred through the SCP pins. A serial transfer synchronizing clock (SPCLK) must be fed into the I/O clock input pin when the CS* pin is driven low. The address to be converted is serially transmitted into the address pin, and the conversion results are serially shifted out the data-out pin. Typical access time is 21 µsec. The APP PWA is the source of the 2.5 V inputs are selected by the channel multiplexer according to the input address
(see Table 4-1). The input multiplexer is a break-before-make type to reduce input-to-input
noise injection resulting from channel switching.
reference voltage. The analog
DC
430-95424-002 (Rev. 01/06) 4 - 6 Plum A+
®
3 Infusion System
Page 35
4.2 ELECTRONIC SUBSYSTEM OVERVIEW
ANALOG INPUTS
U4
A0
A1
A2
A3
A4
A5
A6
A7
A8
A9
A10
GND
CS*
I/O CLOCK
ADDRESS
DATA OUT
EOC
REF+
REF-
GDIG
Figure 4-2. Serial Interface to ADC
CPU
I/O
SPCLK
SPTXD SCP
SPRXD
I/O (OR INTERRUPT)
2.5V
GANA
98K01022
Table 4-1. Analog Inputs
Signal Name Analog Input Address (hex) Description
PRPRS A0 $00 Proximal pressure sensor
DIPRS A1 $01 Distal pressure sensor
PXAIR A2 $02 Proximal air sensor
DIAIR A3 $03 Distal air sensor
IBATT A4 $04 Battery current
VBATT A5 $05 Battery voltage
BUZTST A6 $06 Buzzer test voltage
LCDTST A7 $07 LCD contrast test voltage
TUBTST A8 $08 CCFT intensity test voltage
MI_STA A9 $09 Motor current A control
MI_STB A10 $0A Motor current B control
$0B (V
$0C V
ref(+)
ref(-)
- V
ref(-)
) / 2
$0D V
Technical Service Manual 4 - 7 430-95424-002 (Rev. 01/06)
ref(+)
Page 36
SECTION 4 THEORY OF OPERATION
4.2.1.12
DIGITAL-TO-ANALOG CONVERTER
The dual 8 bit digital-to-analog converter (DAC) generates two analog signals to control the phase A and phase B motor coil currents. The interface between the DAC device and the CPU is the 8 bit data bus, which is connected to DATA15:8. All the control signals for this DAC are generated by FPGA logic devices. Buffer amplifier/ground compensation circuits condition the DAC outputs.
4.2.1.13
FRONT PANEL KEYPAD MATRIX
A 5 x 5 membrane switch keypad matrix is located on the front panel. The keypad column lines (COL4:0) are driven by open collector type memory mapped input ports, while the keypad row lines (ROW4:0), are read by memory mapped input ports (see Table 4-2). The keypad strobing, scanning, and switch de-bouncing is accomplished by software. The keypad interface is designed with ESD protection.
Table 4-2. Keypad Map
COL 0 COL 1 COL 2 COL 3 COL 4
Row 4 Softkey 1 Softkey 2 Softkey 3 Softkey 4
Row 3 Start123[
Row 2 Stop456
Row 1 789[
Row 0 On/Off Clear 0
4.2.1.14
.
]
]
Silence
FRONT PANEL [ON/OFF] KEY
The [ON/OFF] key on the front panel provides a start up (STRTUP) signal to wake up the power supply when the system is shutdown. When activated during normal operation, the [ON/OFF] key interrupts (STRUPD*) the CPU, signaling a request for shutdown.
4.2.1.15
FRONT PANEL LED INDICATORS
The CPU drives the three light emitting diode (LED) indicators embedded in the front panel. Two memory mapped I/O signals activate the two LED lights used to indicate which channel is in delivery mode (LEDAE*, LEDBE*). The AC power on LED indicates the status of AC power (LEDAC) and that the system is in the battery charge mode. A buffered AC on signal (BACON) drives the LED and is active only when AC power is present.
430-95424-002 (Rev. 01/06) 4 - 8 Plum A+
®
3 Infusion System
Page 37
4.2 ELECTRONIC SUBSYSTEM OVERVIEW
4.2.1.16
KEYPAD LOCKOUT INTERFACE
A lockout switch (SW1) on the peripheral interface PWA locks the front panel keypad for all three infusers. A memory mapped input port (LOTSW*) reads the switch. The switch serves as a lockout request and software performs the lockout.
4.2.1.17
NURSE CALL INTERFACE
A nurse call relay switch on the peripheral interface PWA indicates alarm conditions to a remote operator. A memory-mapped output signal (NURSE) activates the relay during alarm conditions. The relay has both normally open and normally closed contacts. A jumper on the peripheral interface board selects the contact type. The factory setting is normally open.
4.2.1.18
AUDIBLE INDICATORS
There are two audible indicators on the CPU subsystem. Three loud, main audible indicators are mounted on the main chassis. This main alarm is used to alert the operator to alarm conditions. A keypad beeper, with lower power and a distinctly different tone, is used to provide audible feedback to the operator. The keypad beeper is driven by a memory-mapped output (KEYALM). It is used to indicate keypad activation, and confirmation to the operator.
The main alarm has an adjustable volume control on the peripheral interface PWA, mounted on the rear of the device. The main alarm can be activated by either a memory-mapped control (MAINALM), the reset pulse(s), or by a power failure alarm latch. The main alarm will sound a chirp for every reset pulse sent by the watchdog timer IC. Continuous chirping indicates a stuck processor.
The alarm is activated continuously during power failure. If the control software does not shut down power in a proper sequence, a latch on the CPU PWA, powered by a backup supply (0.1 F supercap), will activate a continuous alarm. This continuous alarm sounds until either the backup supply is discharged or the user resets the latch by pressing the [ON/OFF] key. Reliable operation of the main alarm is assured by software monitoring of a buzzer test signal (FBUZTST) via the ADC.
4.2.1.19
BARCODE READER INTERFACE
Note: The barcode reader feature will not be present on later versions of the Plum A+
®
3.
The CPU communicates with a barcode wand that is connected to the peripheral PWA from the rear of the infuser. The barcode wand reads and decodes a Code 128 barcode symbology and outputs the barcode data via an RS-232 port using an asynchronous, serial ASCII format.The software controls power to the barcode reader and to the interface circuits via memory-mapped outputs BARPWR and COMPWR*. The barcode reader is isolated from the main system by an optical data path on the peripheral PWA and an isolated power supply.
Technical Service Manual 4 - 9 430-95424-002 (Rev. 01/06)
Page 38
SECTION 4 THEORY OF OPERATION
4.2.1.20
DATAPORT INTERFACE
The CPU communicates with an external computer by way of a DataPort interface. The DataPort interface provides for remote monitoring of up to four infusers using a host computer with a modified RS-232-D serial interface. Infusers are either connected directly to the host or in a daisy chain configuration using junction boxes that provide a 5 bit hard ID via DIP switches on the junction box. The DIP switches are buffered and read by the CPU via the memory-mapped input/output (MMIO) port.
The DataPort system conforms to the EIA-232-D standard, with the following exceptions:
- DataPort uses non-standard DB-15 and 6 pin modular connectors in addition to the standard DB-25 and DB-9 connectors
- With DataPort, more than one infuser is allowed on the line
- The minimum line impedance is 2 KΩ (EIA-232-D standard: 3 KΩ min.)
- The maximum line impedance is 30 KΩ (EIA-232-D standard: 7 KΩ max.)
- The maximum line capacitance is 13 nF (EIA-232-D standard: 2,500 pF)
The communications default is 1200 BAUD, no parity, 8 data bits and 1 stop bit. The BAUD rate is selectable (1200, 2400, 4800, and 9600). The data format on the serial port is a 10 bit frame with asynchronous start and stop. The CTS line is held high and the RTS line is disconnected.
The DataPort is isolated from the main system by an optical data path on the peripheral PWA and an isolated power supply.
4.2.1.21
POWER SUPPLY INTERFACE
The CPU subsystem interfaces the power supply subsystem by providing the MMIO signals needed for power control and battery management. Additionally, the CPU subsystem measures the battery terminal voltage and charge/discharge current via the ADC.
See Table 4-3 for CPU-power supply interface signals.
Table 4-3. CPU-Power Supply Interface
Signal Name Typ e Description
PWRHLD D, O Holds system power on
STRTUP A, I Startup pulse from the [ON/OFF] key
STRUPD* D, I Digital startup pulse, used as interrupt to the CPU
V3_3 P 3.3 V system power
V5_0/VANA P 5 V analog and interface power
VMOT P Raw, unregulated charger voltage or battery voltage
V2_7 P 2.7 V backup power for RTC and non-volatile SRAM
VSC P Full time 5 V supply, backed up by supercap
®
430-95424-002 (Rev. 01/06) 4 - 10 Plum A+
3 Infusion System
Page 39
4.2 ELECTRONIC SUBSYSTEM OVERVIEW
Table 4-3. CPU-Power Supply Interface
Signal Name Typ e Description
V12_0 P 12 V, low current supply for audio alarm
OVRVLT* D, I Signal that indicates overvoltage, regulation problem
on the power supply main regulator
BACON D, I Buffered AC on signal
IBATT A, I Voltage proportional to integration of battery
charge/discharge current
VBATT A, I Divided battery terminal voltage
CHG* D, O Battery charger enable
VFLOAT* D, O Set the main regulator voltage to battery float charge level
ITGRST D, O Reset the charge current integrator
Legend: P = Power; A = Analog; D = Digital; I = Input; O = Output
4.2.1.22
MECHANISM INTERFACE
The CPU subsystem provides the MMIO ports for interface to the mechanism subsystem, in addition to the analog interface mentioned in the ADC and DAC sections
(see Section 4.2.1.11 and Section 4.2.1.12).
See Table 4-4 for CPU-mechanism interface signals.
Table 4-4. CPU-Mechanism Interface Signals
Signal Name Type Description
MI_STA A, O Motor current set for phase A
MI_STB A, O Motor current set for phase B
GDAC A, O Ground signal from chopper (for compensation)
M_PHA D, O Motor phase A
M_PHB D, O Motor phase B
M_SEL1, M_SEL0 D, O Motor select bits
FLCAME D, O I/O and L/S cam flag sensors enable
FLPINE D, O L/S pin motion detectors enable
FLPLE D, O Plunger motor sensor pair enable
FLLS_C D, I Flag, L/S valve cam sensor
FLIO_C D, I Flag, I/O valve cam sensor
FLLS_A D, I Flag, L/S valve A pin detector
Technical Service Manual 4 - 11 430-95424-002 (Rev. 01/06)
Page 40
SECTION 4 THEORY OF OPERATION
Table 4-4. CPU-Mechanism Interface Signals
Signal Name Type Description
FLLS_B D, I Flag, L/S valve B pin detector
FLPLRO D, I Flag, plunger rotation sensor
FLPLTR D, I Flag, plunger translation sensor
PXPRE D,O Proximal pressure sensor enable
PXPRS A, I Proximal pressure sensor
DIPRE D, O Distal pressure sensor enable
DIPRS D, O Distal pressure sensor
PXARE D, O Proximal air sensor enable
PXAIR A, I Proximal air sensor
DIARE D, O Distal air sensor enable
DIAIR A, I Distal air sensor
CASPR* D, I Cassette present
CASS2*, CASS1*, CASSO* D, I Cassette type coding: Macro (111), Micro (010),
all others are invalid
SPCLK D, O SCP clock output
SPRXD D, I SCP receive data
SPTXD D, O SCP transmit data
CSSEP* D, O Chip select, EEPROM
V5_0 P 5 V supply for interface power
V3_3 P 3.3 V supply for logic power
GDIG P Digital ground
VANA P 5 V supply for analog power
GANA P Analog ground
VMOT, GMOT P Motor power is directly from power supply PWA
V2_5 A, I Reference voltage for ADC and DAC
Legend: P = Power; A = Analog; D = Digital; I = Input; O = Output
430-95424-002 (Rev. 01/06) 4 - 12 Plum A+
®
3 Infusion System
Page 41
4.2 ELECTRONIC SUBSYSTEM OVERVIEW
4.2.2

POWER SUPPLY SUBSYSTEM

The power supply subsystem provides DC power to system circuits and interface software controlled power and battery management.
The power supply subsystem provides for the following functions:
- Main switching regulator
-AC power detection
- Main regulator fault detection
- System power (secondary regulators)
- Auxiliary supplies
-Power control
- Battery charging circuitry
- Battery terminal voltage measurement
- Battery charge/discharge current measurement
4.2.2.1
MAIN SWITCHING REGULATOR
The main source of power for the infuser is the AC line. The main switching regulator is a pulse width modulated, AC-to-DC converter which provides the system an isolated DC voltage of 6.9 V preceded by line fuses F1 and F2, surge suppressor VR1, and a line filter. The bridge rectifier U14 and capacitors C52 and C53 provide the DC voltage required for the switching circuit. Voltage regulator U13 provides the pulse width modulator (PWM) device U12 startup supply voltage. After startup, supply voltage for U12 is supplied by half wave rectifier circuitry CR14, R76, and C51.
The PWM oscillation frequency is approximately 40 kHz, determined by external resistor R72 and capacitor C45. U12 controls the power delivered by varying the duty cycle of the power metal-oxide-semiconductor field-effect transistor (MOSFET) Q9, which drives T2. A half-wave rectifier rectifies the transformer’s secondary voltage, which provides the raw DC voltage for the battery charger and system power. There are three feedback mechanisms that maintain control: a main loop for normal control, a secondary loop for overvoltage protection, and a current limit loop.
4.2.2.1.1
(or 7.5 VDC in battery charger boost mode). The main regulator is
DC
Main Loop
The main loop uses an optical feedback path to regulate the charger voltage (BATPOS) at 6.9 V control of the VFLOAT* line). A shunt regulator and opto-isolator provide feedback to the PWM error amplifier.
Technical Service Manual 4 - 13 430-95424-002 (Rev. 01/06)
(except during boost charge, when the limit is raised to 7.5 VDC by software
DC
Page 42
SECTION 4 THEORY OF OPERATION
4.2.2.1.2
Secondary Loop
Diode CR10 and opto-isolator U10 provide overvoltage protection. CR10 conducts and activates U10 when secondary voltage exceeds approximately 10 V
. The duty cycle of
DC
U12 is reduced until the excessive voltage is removed.
4.2.2.1.3
Current Limit Loop
The current limit loop is activated when the primary current, sensed by R71, exceeds 3 A. Resistor R70 and capacitor C46 filter the voltage across R71 and feed it back to the current sense input (1.5 V
threshold) of U12. The duty cycle of U12 is reduced until the excessive
DC
load is removed.
4.2.2.2
MAIN REGULATOR FAULT DETECTION
If the switching regulator’s main loop fails, the secondary voltage limit loop takes over. However, the battery charger and motors must be disabled, and an alarm must be generated. A comparator is used to monitor the raw DC (+BUSS) for overvoltage. A 3.3 V logic signal (OVRVLT*) is provided to the CPU subsystem.
DC
4.2.2.3
SYSTEM POWER
Along with the unregulated VMOT supply, a secondary switching regulator provides system power. The secondary switching regulator includes IC U4, transformer T1, and transistors Q4 and Q5. The regulator is a triple output, wide supply range, fly-back converter that provides regulated 3.3 V transformer T1. The regulator operates over an input range of 4 V
, 5 VDC, and 12 VDC outputs from the five winding
DC
to 10 VDC and provides
DC
output current limit as well as voltage overshoot limit. Primary feedback is metered through a bias arrangement on transistor Q3. A Schottky rectifier diode CR4 provides feedback in the event of V3_3 or V12_0 failure, and transistor Q10 provides feedback in the event of V5_0 failure. The positive terminal of the battery provides the raw DC voltage, VMOT, for the motors and backlight of the display.
4.2.2.4
AUXILIARY SUPPLIES
The power supply subsystem provides full time 5 V active when battery or AC voltage is present. The full time 5 V low dropout voltage regulator U6, whose power source is directly from the battery and is backed up by a 0.1 F capacitor. VSC is used for the ON/OFF switch and a power failure alarm latch. The full time 2.7 V
supply (V2_7) is derived from VSC and is used to supply
DC
the ultra-low current needed to power the real-time clock and non-volatile SRAM during shutdown.
and 2.7 VDC supplies, which are
DC
supply (VSC) uses a linear
DC
430-95424-002 (Rev. 01/06) 4 - 14 Plum A+
®
3 Infusion System
Page 43
4.2 ELECTRONIC SUBSYSTEM OVERVIEW
4.2.2.5
POWER CONTROL
The infuser will operate in one of three modes: normal, standby, or shutdown. During normal operation, the user interface is active and either on battery or AC line power. During standby mode the user interface is inactive while the CPU is still operating, servicing the battery management and waiting for a startup interrupt. Shutdown mode iswhen system power is off. Shutdown mode only occurs during battery operation; otherwise, +BUSS holds the system power on.
The infuser is activated when the [ON/OFF] key is pressed or the AC line is plugged in. The [ON/OFF] key activates the STRTUP signal, triggering a three second one-shot circuit that will temporarily turn the system power on. This three second one-shot period allows the CPU enough time to power up, initialize, and turn on the PWRHLD signal. The CPU monitors the STRTUP signal, via interrupt, to signal a user request for turning off the infuser.
Figure 4-3 illustrates the system startup/shutdown sequence while battery powered.
System power is always on while AC powered.
STRTUP
ONE-SHOT
V3_3, V5_0, V12V
PWRHLD
Figure 4-3. System Startup and Shutdown Timing, Battery Powered
3 SEC
98K01021
Technical Service Manual 4 - 15 430-95424-002 (Rev. 01/06)
Page 44
SECTION 4 THEORY OF OPERATION
4.2.2.6
BATTERY VOLTAGE MEASUREMENT
The battery terminal voltage (BATPOS - BATNEG) is measured with a differential amplifier consisting of U1, R1, R2, R4, R7, and R8. It has a gain of 0.317 to generate a single ended VBATT signal. The VBATT signal is then provided to the CPU A/D converter as input for the battery management algorithms.
4.2.2.7
BATTERY CHARGE/DISCHARGE CURRENT MEASUREMENT
The battery management algorithms measure battery charge/discharge current for battery capacity estimation and charger control. The charge/discharge current is measured by integrating the voltage across current sense resistor R57. An operational amplifier (op-amp) integrator circuit, consisting of U2, C5, R12, R13, R19, and R20, provides a voltage proportional to the integration of battery current (IBATT) over a CPU controlled measurement period. The IBATT signal is fed to the CPU A/D converter, where it is sampled at the end of the measurement period. The battery management algorithm further accumulates the charge/discharge current for battery capacity estimation. The op-amp integrator is reset by the CPU system at the beginning of each measurement period by parallel analog switches U3, controlled by the CPU’s ITGRST signal. The battery management algorithm periodically calibrates the op-amp integrator.
4.2.2.8
BATTERY CHARGER
The software battery management algorithm controls the battery charger. The charging scheme is a current limit/two stage voltage limit charger. The charge current is limited to
1.3 A and the voltage is limited to either 6.9 V
The source of the charge current is power MOSFET transistor Q7 operating in the linear mode. Charge current passes through a current sense resistor R57, where it develops a feedback signal for the charger control amplifier consisting of U7, Q6, and associated parts. The feedback signal is compared against a 2.5 V A .5 A fuse (F4) protects against damage due to a short circuit. The battery management algorithm maintains on/off control of the charger by the charger enable signal CHG*. When set high, CHG* activates a comparator U7, which overrides the feedback signal and disables the charger. Excessive voltage on the BATNEG terminal indicates that there is a shorted battery cell, and will disable the charger through the same comparator.
or 7.5 VDC.
DC
voltage reference U8.
DC
430-95424-002 (Rev. 01/06) 4 - 16 Plum A+
®
3 Infusion System
Page 45
4.2 ELECTRONIC SUBSYSTEM OVERVIEW
4.2.3

MECHANISM SUBSYSTEM

The mechanism subsystem includes the electronics and electromechanical components that interface with the pumping mechanism.
The mechanism subsystem provides the following functions:
- Chopper motor drive for three stepper motors (plunger, L/S valve, I/O valve)
- Four motor position sensors (flag detectors)
- Precision voltage reference
- Two air sensors (distal, proximal)
- Two pressure sensors (distal, proximal)
- Cassette presence and type detection
- Serial electrically erasable PROM (EEPROM)
See Table 4-4 for mechanism interface signals.
4.2.3.1
MOTORS/MOTOR DRIVE
The infuser uses three stepper motors for pumping: one for fluid displacement and two for cassette valve actuation. The stepper motors are driven, under step-by-step control from software, by a unipolar chopper drive.
4.2.3.1.1
Stepper Motors
Each motor is named by its function:
- Plunger motor for driving the plunger screw
- I/O valve motor for moving the input-output valve pins
- L/S valve motor for moving the line select valve pins A and B
All three motors are four phase stepper types. One electrical revolution is accomplished after four motor steps (phases) are completed. The step-angle (the number of steps per shaft revolution) resolutions are 3.6 and 7.5
The unipolar motor windings have a center tap connected on each of the two coils as shown in Figure 4-4. Unidirectional current enters the center tap and is steered to one end of the coil or the other end by the driver electronics, creating positive or negative flux lines in the motor coil. With two coils each with a choice of flux polarity, four electrical combinations or phases are possible.
° /step (48 steps/rev) for the I/O and L/S valve motors.
° /step (100 steps/rev) for the plunger motor,
Technical Service Manual 4 - 17 430-95424-002 (Rev. 01/06)
Page 46
SECTION 4 THEORY OF OPERATION
A
ACOM
A
BCOM
B B
98K01020
Figure 4-4. Stepper Motor Coils
4.2.3.1.2
Chopper Motor Drive
The infuser stepper motor drive is a chopper drive, which is a pulse width modulation of the coil current in each motor winding. Current is switched on and off to maintain a predetermined coil current independent of supply voltage and motor speed. The motor winding inductance acts as a filter to smooth out the switching currents, slowing the current rise when turned on and storing a decaying current when turned off. Each motor coil is modulated independently, allowing different coil currents in the two motor windings. The coil current is sensed and compared to a reference input for each winding. Modulation circuits correct for any error between the sensed current and the reference. This reference input can be changed to set a different coil current.
4.2.3.2
MOTOR POSITION SENSORS
Motor position is estimated by counting the motor steps, relative to a position reference. Optical switches and flags serve as position references, which are used to find the motor home positions and to verify proper motion. Flag positions are anticipated by software.
Optical switch flag sensors are used for tracking the following:
- Plunger motor rotational position (coupler flag)
- Plunger translational (linear) position
- I/O valve motor rotational position (cam flag)
- L/S valve motor rotational position (cam flag)
430-95424-002 (Rev. 01/06) 4 - 18 Plum A+
®
3 Infusion System
Page 47
4.2 ELECTRONIC SUBSYSTEM OVERVIEW
Each optical switch consists of an infrared LED, which shines through a rectangular aperture, across a slot, to illuminate a photo-transistor. The photo-transistor is activated as long as the beam is on and not blocked (by a flag in the slot). The optical switches are distributed throughout the mechanism, near their associated flags. The motor rotational optical switches are mounted on the driver PWA along with the control circuitry. The plunger translational optical switch is mounted remotely on the switch PWA. The switches are used intermittently to save power.
There are two control signals that enable associated switch pairs:
- FLCAME flag valve motor cam sensor enable
- FLPLE flag plunger motor rotation and translation sensors enable
Each of these control signals enables a constant current source which turns on the associated switch’s infrared LEDs. The photo transistor states are sensed by Schmidt trigger inverters (U11 on driver PWA) which provide a 3.3 V logic high when the optical path is blocked or a logic low when the optical path is clear. The Schmidt trigger output is high when the sensor is disabled.
The following output signals are provided to the CPU subsystem:
- FLIO_C flag I/O valve motor cam sensor
- FLLS_C flag L/S valve motor cam sensor
- FLPLRO flag plunger motor rotation sensor
- FLPLTR flag plunger motor transition sensor
4.2.3.3
V2_5 REFERENCE VOLTAGE
A precision 2.5 VDC reference voltage is generated on the APP PWA for use by the pressure sensor excitation circuits, the air sensor amplifier circuits, and the ADC and DAC reference voltage. The precision 2.5 V is V2_5.
4.2.3.4
reference is buffered by a voltage follower. The signal name
DC
AIR SENSORS
The mechanism subsystem includes two air sensors, used to detect air passage into (proximal) or out of (distal) the cassette. Both sensors are piezoelectric crystal transmitter receiver pairs. Liquid between the transmitter and receiver will conduct the ultrasonic signal, while air will not (see Figure 4-5).
Technical Service Manual 4 - 19 430-95424-002 (Rev. 01/06)
Page 48
SECTION 4 THEORY OF OPERATION
VSO
CPU
VCO/PLL
Figure 4-5. Air Sensor Block Diagram
4.2.3.4.1
Transmitter Circuitry
A/D
G_TX
OUTPUT
AMP
XTL TX
XTL RX
G_RX
PEAK
DETECTOR
98K01019
The transmitter circuitry consists of a voltage sweep oscillator, a voltage-controlled oscillator (VCO), and a transmitter amplifier, and are located on the APP PWA.
The voltage sweep oscillator circuit oscillates at approximately 12 kHz at 50 percent duty cycle. The output of the sweep oscillator is between +2 V
and +3 VDC, and is used to
DC
sweep the VCO. The VCO sweeps through the sensor’s peak coupling frequency, which is between 3 MHz and 6 MHz. A resistor and capacitor are used to configure the VCO center frequency. The VCO is enabled when the CPU asserts either DIARE or PXARE control signals.
The transmitter amplifier consists of a push-pull, emitter-follower, complementary pair of transistors. The transmitter amplifier drives both proximal and distal sensors simultaneously.
4.2.3.4.2
Receiver Circuitry
When the cassette’s test port is filled with fluid, the transmitted signal will be coupled to an identical piezoelectric crystal, where it is amplified and detected by the receiver circuitry. The receiver circuitry consists of an amplifier, a peak detector, and an adjustable gain buffer stage. There is a separate, symmetrical receiver circuit for each channel (proximal and distal). Component references (called out in this design description) will be made to the distal channel only.
430-95424-002 (Rev. 01/06) 4 - 20 Plum A+
®
3 Infusion System
Page 49
4.2 ELECTRONIC SUBSYSTEM OVERVIEW
The first amplifier includes two, directly coupled common emitter stages, biased from the V2_5 supply. DIARE and PXARE are used to enable the distal and proximal sensors, respectively. The detector stage consists of an emitter follower, charging a 400 µsec time constant, refreshed every 40 microseconds (twice per VCO sweep).
The peak detector output is buffered by an op-amp configured as a basic non-inverting amplifier with a trimming potentiometer (R31) for gain adjustment. Each sensor has an independent gain adjustment. The two air sensor, gain-trimming potentiometers are accessible for calibration in an assembled mechanism.
These final signals are read by the CPU subsystem via the ADC.
- PXAIR proximal air sensor output
- DIAIR distal air sensor output
4.2.3.5
PRESSURE SENSORS
The mechanism subsection contains two strain gauge-type pressure sensors, one at the proximal and the other at the distal cassette ports. Electrically, the strain gauge is a Wheatstone bridge made of four strain gauge resistors. When the bridge is electrically excited, the bridge will output a millivolt level signal proportional to the applied pressure. The output signal is amplified and offset adjusted before being read by the ADC. Each pressure sensor circuit includes an excitation voltage supply, sensor amplifiers, and a low pass filter.
The pressure sensor circuitry is on the APP PWA. Each of the two channels has an identical topology, but different gain and filter response. A block diagram of this circuit is shown in Figure 4-6.
Note: Component references are made to the distal channel only.
Technical Service Manual 4 - 21 430-95424-002 (Rev. 01/06)
Page 50
SECTION 4 THEORY OF OPERATION
PRESSURE
SENSOR
ENABLE
(FROM CPU)
BRIDGE
EXCITATION
3.75 V
REFERENCE
2.5 V
WHEATSTONE
BRIDGE
DIFFERENTIAL
AMPLIFIER
AND OFFSET
ADJUST
OUTPUT
AMPLIFIER
AND FILTER
PRESSURE SIGNAL OUTPUT (TO CPU)
Figure 4-6. Pressure Sensor Excitation and Amplifier Block Diagram
430-95424-002 (Rev. 01/06) 4 - 22 Plum A+
05K01018
®
3 Infusion System
Page 51
4.2 ELECTRONIC SUBSYSTEM OVERVIEW
4.2.3.5.1
Bridge Excitation Supply
The bridge excitation voltage is 3.75 VDC, and is derived from the 2.5 VDC reference signal (V2_5), gained 1.5 times by an amplifier. The CPU subsystem may independently enable power to each pressure sensor bridge.
These enable signals are active high 3.3 V logic level inputs:
- PXPRE proximal pressure sensor enable
- DIPRE distal pressure sensor enable
4.2.3.5.2
Amplifier and Low Pass Filter
The pressure sensor amplifiers include a high gain differential pre-amplifier, followed by a second stage non-inverting amplifier with low gain. A trimming potentiometer is adjusted to minimize any offset in the impedance of the bridge.
A two-pole filter is used to filter the pressure signals. The first pole is formed by a capacitor (C39, multiplied by 230 due to Miller effect) and a Thevenin resistance (seen at U4-2). The second pole is the RC filter at the ADC input, which is located on the CPU PWA.
These output signals to the A/D converter in the CPU PWA are:
- PXPRS proximal pressure signal
- DIPRS distal pressure signal
4.2.3.6
PRESSURE SENSOR CALIBRATION
Pressure sensors are calibrated for offset and gain during mechanism calibration. A trimming potentiometer is used to adjust the initial, zero pressure offset. The proximal and distal pressure sensors have independent offset adjustments. The final system gain (cassette pressure to corrected amplifier output) is adjusted in software. During mechanism calibration, each channel’s gain (amplifier output/cassette pressure) will be measured, and stored in the serial EEPROM on the driver PWA.
4.2.3.7
CASSETTE TYPE/PRESENCE SELECTION
The mechanism subsystem includes four force sensing resistor (FSR) switches, which are coupled to the cassette. Three FSRs are used for cassette type decoding and one is used for cassette present detection. The FSR is a polymer thick film device, which exhibits a decrease in resistance with any increase in force applied to the active surface. The FSRs have a resistance that is either very large (> 1 MΩ) or relatively small (< 100 KΩ). The large resistance is defined as a logical ‘0’, and the small resistance is defined as logical ‘1’. Each FSR is arranged in a voltage divider configuration with a fixed resistor, followed by a comparator with hysteresis. The comparator circuits are located on the CPU PWA. The comparators are designed to trip as the FSR’s resistance falls below 120 KΩ.
Technical Service Manual 4 - 23 430-95424-002 (Rev. 01/06)
Page 52
SECTION 4 THEORY OF OPERATION
4.2.3.8
SERIAL EEPROM
The driver PWA holds the 8 K x 8 bit, serial EEPROM, which is used to store event, alarm, malfunction, and calibration data specific to the pumping mechanism. It is accessed through a serial peripheral interface (SPI) compatible interface, which is a high-speed serial interface to the CPU. The CPU PWA accesses this device through its SCP serial interface. This interface is a subset of the SPI, and consists of clock (SPCLK), data in (SPRXD), and data out (SPTXD) pins. This device is in the driver PWA to allow the calibration data to stay with the mechanism.
4.3

PRINTED WIRING ASSEMBLIES

Infusion system electronics are packaged into six printed wiring assemblies (PWA) and several remote mounted peripherals (see Section 4.4). The following sections provide a brief description of the functional interfaces of each PWA.
4.3.1

POWER SUPPLY PWA

The power supply PWA (see Figure 9-11) contains the following functions of the power supply subsystem:
- Main switching regulator
-AC power detection
- Main regulator fault detection
-System power
The power supply PWA is a four layer board, with primarily surface mount technology (SMT) components. The board is fully testable from the bottom side. An insulating tape covers the back of the power supply PWA. Open system troubleshooting should be done under battery power. If connection to the AC line is required, an isolation transformer should be used since AC line potentials are present on the power supply PWA.
See Section 4.2.2 for a functional description, and see Table 4-5 for power supply PWA interface connections.
Table 4-5. Power Supply PWA Interface Connections
Connector Type Interface
P2 30 pin receptacle Board-to-board connection to CPU PWA
- Auxiliary supplies
-Power control
- Battery management
J16 4 pin header Motor power connection to driver PWA
J21 3 pin receptacle AC power cord connection
J22 2 pin header Battery cable connection
®
430-95424-002 (Rev. 01/06) 4 - 24 Plum A+
3 Infusion System
Page 53
4.3 PRINTED WIRING ASSEMBLIES
4.3.2

PERIPHERAL PWA

The peripheral PWA (see Figure 9-13) contains part of the CPU subsystem circuitry, including system program and data memories (PROM and SRAM), and external communication interface circuits. The peripheral PWA is designed to be field replaceable, to facilitate software upgrades or additional external interfaces.
The peripheral PWA is a four layer board, including one ground plane, one power plane, and two signal layers. In its initial configuration, all of the components are mounted on the top side.
See Table 4-6 for peripheral PWA interface connections.
Table 4-6. Peripheral PWA Interface Connections
Connector Type Interface
P1 96 pin receptacle Board-to-board connection to CPU PWA
J26 15 pin D-sub DataPort
J27 9 pin D-sub Barcode reader connection
J28 3 pin phone jack Nurse call jack
4.3.3

PERIPHERAL INTERFACE PWA

The peripheral interface PWA (see Figure 9-12) contains part of the CPU subsystem circuitry, including system program and data memories (PROM and SRAM), external communication interface circuits, and rear instrument user controls.
See Table 4-7 for peripheral interface PWA interface connections.
Table 4-7. Peripheral Interface PWA Interface Connections
Connector Type Interface
P1 96 pin receptacle Board-to-board connection to CPU PWA
J29 and J30 50 pin plug Board-to-board connection to peripheral PWA
J26 15 pin D-sub DataPort
J27 9 pin D-sub Barcode reader connection
J28 3 pin phone jack Nurse call jack
Technical Service Manual 4 - 25 430-95424-002 (Rev. 01/06)
Page 54
SECTION 4 THEORY OF OPERATION
4.3.4

CPU PWA

The CPU PWA (see Figure 9-14) contains most of the CPU subsystem functions, with the exception of main memory and communications ports, which are located on the peripheral PWA. The CPU PWA also accommodates system interconnect.
The CPU PWA is an eight layer board, with one ground plane, one power plane, and six signal layers. The CPU PWA primarily contains SMT components. Most of the components are on the top side, while the bottom side holds wave-solder compatible SMT resistors and capacitors.
See Section 4.2.1 for a functional description, and see Table 4-8 for CPU PWA interface connections.
Table 4-8. CPU PWA Interface Connections
Connector Type Interface
J7 96 pin header Connection to peripheral PWA (CPU bus,
rear panel I/O, and communication ports)
J2 30 pin header Connection to power supply PWA
J3 50 pin SMT Ribbon cable connection to driver PWA
(mechanism)
J4 21 pin header Front panel connector (keypad, LEDs,
on/off switch)
J5 14 pin SMT Flat flex cable to LCD panel
J20 4 pin header CCFT backlight connector
J24 2 pin header Main audible alarm connector
4.3.5

DRIVER PWA

The driver PWA (see Figure 9-15) contains the mechanism subsystem’s motor drive circuitry, motor position sensors, and serial EEPROM. The driver PWA is mounted in the mechanism sub-chassis.
The driver PWA is a four-layer PWB, with one ground plane, one power plane, and two signal layers. The driver PWA primarily uses SMT components. Most of the components are located on the top side of the board, while the bottom side holds wave-solder compatible resistors and capacitors.
See Table 4-9 for driver PWA interface connections.
430-95424-002 (Rev. 01/06) 4 - 26 Plum A+
®
3 Infusion System
Page 55
4.3 PRINTED WIRING ASSEMBLIES
Table 4-9. Driver PWA Interface Connections
Connector Type Interface
J7 6 pin header Plunger motor
J8 6 pin header Input/output motor
J9 6 pin header Line select motor
J10 20 pin SMT Flat flex cable to APP PWA
J11 50 pin header Ribbon cable to CPU PWA
J12 6 pin SMT FSR flex circuit
J13 4 pin header Motor power, from power supply PWA
J14 8 pin SMT Flat flex cable to switch PWA
4.3.6

SWITCH PWA

The switch PWA (see Figure 9-16) contains the plunger translation position sensor, which is one of four position sensors in the system. The switch PWA is located at the side of the mechanism sub-chassis, and connects to the driver PWA.
4.3.7

APP PWA

The APP PWA (see Figure 9-17) is mounted in the mechanism sub-chassis, and contains the following mechanism subsystem circuitry:
- Proximal and distal air sensors and circuitry
- Proximal and distal pressure sensor amplifiers and excitation
- V2_5 precision voltage reference
- Pin detector optical switch module
The APP PWA is a four layer board, with one ground plane, one power plane, and two signal layers. The APP PWA uses SMT components, mounted on both sides of the board. The air sensors and the pin detector module are board mounted.
See Table 4-10 for APP PWA interface connections.
Table 4-10. APP PWA Interface Connections
Connector Type Interface
J15 20 pin SMT Flat flex cable to driver PWA
J11 10 pin SMT Pressure sensor connector
Technical Service Manual 4 - 27 430-95424-002 (Rev. 01/06)
Page 56
SECTION 4 THEORY OF OPERATION
4.4

REMOTE MOUNTED PERIPHERALS

The following sections describe the major remote mounted peripherals.
4.4.1
LCD
The infuser uses a graphic LCD module with a CCFT. The CCFT provides a backlight source for the LCD. The LCD requires a nominal -16 V is controlled by the CPU. The infuser’s graphic display data is shifted out to the LCD by the CPU LCD controller, which interfaces directly with the CPU (see Section 4.2.1.6). The display is configured as a 240 x 240 dot matrix with a viewing angle of approximately 60
° .
4.4.2

SEALED LEAD ACID BATTERY

supply for contrast control, which
DC
The infuser uses a nominal 6 VDC rechargeable sealed lead acid battery with 4 amp-hour capacity.
4.4.3

BARCODE READER WAND

Note: The barcode reader feature will not be present on later versions of the Plum A+
The barcode reader wand connects to the BCR port J27 on the peripheral PWA. The BCR wand interfaces through the infuser’s optically isolated, TTL logic level, asynchronous interface. The BCR wand is also capable of interfacing at RS-232 levels. The infuser provides an isolated +5 V of the BCR wand is swiped across a barcode label, the reflected light is scanned and processed. After a successful scan, the data is sent over the communication interface to the CPU.
regulator to power the BCR wand. When the LED at the tip
DC
®
3.
4.5

MECHANICAL OVERVIEW

The principal mechanical elements of the infuser include the cassette and the mechanism assembly. When a cassette is locked into the operating position and the [ON/OFF] switch is pressed, the infuser performs a self test to verify the integrity of the internal systems. The operation of the mechanism assembly moves a plunger, causing a pumping action. A valve motor selects the A or B valve, depending on the command. An additional valve motor alternately opens and closes an inlet valve and outlet valve to control fluid flow through the cassette pumping chamber.
The following sections detail the cassette and the mechanism assembly.
430-95424-002 (Rev. 01/06) 4 - 28 Plum A+
®
3 Infusion System
Page 57
4.5 MECHANICAL OVERVIEW
4.5.1

CASSETTE

The cassette (see Figure 4-7 and Figure 4-8) operates on a fluid displacement principle to volumetrically deliver fluid. See the system operating manual for a description of the major cassette functions.
The pumping cycle begins when the outlet valve is opened and the inlet valve is closed. The plunger extends to deflect the cassette diaphragm and expel fluid. At the end of the pumping stroke, the outlet valve is closed, the inlet opens, the appropriate A or B valve opens, and the plunger retracts to allow fluid to refill the pumping chamber. After the pumping chamber is filled, the inlet and outlet valves are reversed, the A and B valves are closed, and the cycle repeats.
The cassette contains two chambers: an upper air trap chamber and a pumping chamber. The two chambers are separated by an inlet valve and operate together to detect air. The air trap chamber receives fluid from the intravenous (IV) container through either the A or B valve. The air trap chamber collects air bubbles from the IV line and container to prevent them from entering the pumping chamber and can collect a substantial amount of air.
A proximal air-in-line sensor (bubble detector) is located between the A/B valves and the upper air-trap chamber. The proximal air-in-line sensor detects air entering the upper air-trap chamber and initiates an audible alarm if the predetermined air collection threshold is exceeded. Similarly, a second air-in-line sensor located distal to the pumping chamber initiates an audible alarm if a predetermined amount of air is detected.
The pumping chamber receives fluid from the upper air-trap chamber through an inlet valve. A pressure sensor located in the upper air-trap chamber monitors pressure on the proximal side of the cassette. When the diaphragm covering the pumping chamber is deflected by the plunger, the pumping chamber expels fluid through an outlet valve. A pressure sensor located distal to the pumping chamber monitors pressure on the distal side of the cassette.
A flow regulator is incorporated into the cassette distal end. This flow regulator is used to manually control flow when the cassette is not inserted in the infuser. When the cassette is properly inserted into the infuser and the door is closed, a mechanism opens the flow regulator to allow the infuser to control fluid flow. When the door is opened, the same mechanism closes the flow regulator to disable fluid flow.
Technical Service Manual 4 - 29 430-95424-002 (Rev. 01/06)
Page 58
SECTION 4 THEORY OF OPERATION
RIGHT VIEW
SECONDARY PORT (Y-RESEAL OR LOCKING CAP)
B VALVE
AIR-IN-LINE
SENSOR
(PROXIMAL)
PRESSURE SENSOR
(PROXIMAL)
AIR TRAP
CHAMBER
INLET VALVE
FROM PRIMARY
CONTAINER
REAR VIEW
(CONTROL NOT SHOWN)
A VALV E
AIR-IN-LINE
SENSOR (DISTAL)
PRESSURE SENSOR
(DISTAL)
OUTLET VALVE
PUMPING CHAMBER
PRECISION GRAVITY
FLOW REGULATOR
FINGER
GRIP
OUTLET
TO
PATIENT
LEFT VIEW
Figure 4-7. Major Elements of the Dual-Channel Cassette
430-95424-002 (Rev. 01/06) 4 - 30 Plum A+
05K13017
®
3 Infusion System
Page 59
LINE A
4.5 MECHANICAL OVERVIEW
A VALV E
AIR TRAP CHAMBER
INLET VALVE
OUTLET VALVE
LINE B
B VALVE
AIR-IN-LINE SENSOR (PROXIMAL)
PRESSURE SENSOR (PROXIMAL)
PUMPING CHAMBER
PRESSURE SENSOR (DISTAL)
AIR-IN-LINE SENSOR (DISTAL)
PRECISION GRAVITY
FLOW REGULATOR (AND SHUT OFF)
98G01001
Figure 4-8. Fluid Path in the Cassette
Technical Service Manual 4 - 31 430-95424-002 (Rev. 01/06)
Page 60
SECTION 4 THEORY OF OPERATION
4.5.2

MECHANISM ASSEMBLY

The mechanism assembly is a fully self-contained unit consisting of the motor and valve assemblies, A/B valve subsystem, inlet/outlet valve subsystem, plunger drive subsystem, air bubble (ultrasonic) sensor assemblies, cassette door, and pressure sensor assemblies. The motor and valve assemblies, A/B valve subsystem, inlet/outlet valve subsystem, and plunger drive subsystem are detailed in the following sections.
During infuser operation, the mechanism assembly plunger motor drives a lead screw that is coupled to the plunger. The motor action and lead screw move the plunger forward to cause the delivery of approximately 0.33 mL of fluid per cycle. The plunger motion is synchronized to the valve motors to provide controlled fluid delivery.
4.5.2.1
MOTOR AND VALVE ASSEMBLIES
The mechanism assembly pumping action is controlled by three stepper motors. The first stepper motor, in conjunction with an associated valve assembly, activates the A or B valve of the cassette, depending on the command. The second stepper motor alternately opens and closes the inlet and outlet valve to control fluid delivery through the cassette pumping chamber. A third stepper motor controls plunger movement.
4.5.2.2
A/B VALVE SUBSYSTEM
The A/B valve subsystem includes a motor designed to rotate a cam. When the cam is positioned at top-dead-center (home position), both valves are closed. Clockwise rotation (when viewed from the motor side) from the home position opens the A valve, while the B valve remains closed. Counterclockwise rotation opens the B valve, while the A valve remains closed (see Figure 4-9).
The A/B valve subsystem consists of a stepper motor with attached cam and integral cam flag, A and B rockers and valve pins, and a pin detector assembly. The cam flag passes through an interrupter module as it rotates with the cam. Valve home position is determined by this cam flag/interrupter module combination through predetermined factory calibration data. During operation, if the cam flag passes through the interrupter module at the incorrect time sequence, a motor phase loss is detected. The rocker is the connecting link between the cam and the valve pin.
430-95424-002 (Rev. 01/06) 4 - 32 Plum A+
®
3 Infusion System
Page 61
4.5 MECHANICAL OVERVIEW
A VALV E
REGULATOR ACTUATOR
AIR-IN-LINE SENSOR (DISTAL)
PRESSURE SENSOR (DISTAL)
CASSETTE LOCATOR
OUTLET VALVE
PLUNGER
INLET VALVE
Figure 4-9. Mechanism Valve Pins and Sensor Locations
4.5.2.3
INLET/OUTLET VALVE SUBSYSTEM
B VALVE
AIR-IN-LINE SENSOR
(PROXIMAL)
PRESSURE SENSOR
(PROXIMAL)
FORCE SENSING
RESISTOR
02K01023
The inlet/outlet valve subsystem is similar in function and build to the A/B valve subsystem (see Section 4.5.2.2).
4.5.2.4
PLUNGER DRIVE SUBSYSTEM
The main components of the plunger drive subsystem are: plunger, lead screw and coupler, and stepper motor. When the infuser is turned on, the plunger moves from the retracted, PARK position to the HOME position. The cassette diaphragm is engaged. The stepper motor rotates approximately 1 2/3 revolutions per pump cycle to permit a 0.33 mL fluid displacement every pump cycle. The stepper motor then reverses and the plunger returns to HOME position. This cycle repeats for the duration of fluid administration.
The screw/coupler assembly links the motor and the plunger. This assembly includes a flag that passes through an interrupter module. This screw/coupler, flag/interrupter module combination is used in conjunction with predetermined factory calibration data to determine the plunger position. During operation, if the screw/coupler flag passes through the interrupter module at the incorrect time sequence, a motor phase loss is detected.
Technical Service Manual 4 - 33 430-95424-002 (Rev. 01/06)
Page 62
SECTION 4 THEORY OF OPERATION
This page intentionally left blank.
430-95424-002 (Rev. 01/06) 4 - 34 Plum A+
®
3 Infusion System
Page 63
Section 5

MAINTENANCE AND SERVICE TESTS

A complete maintenance program promotes infusion system longevity and trouble-free operation. Such a program should include routine maintenance, periodic maintenance inspection, and following any repair procedure, performance verification testing.
5.1

ROUTINE MAINTENANCE

Routine maintenance consists of basic inspection and cleaning procedures. As a minimum requirement, inspect and clean the infuser after each use. In addition, establish a regular cleaning schedule for the device.
5.1.1

CLEANING

Accumulation of dust or spilled fluids on the cassette door and housing can affect proper operation. The following cleaning procedures are designed to sustain longevity and promote trouble-free operation.
Follow hospital protocol for establishing the infuser cleaning schedule.
WARNING:
CAUTION: Do not immerse the infuser in liquids. Immersion could damage the device. Do not allow liquids to enter the electronics compartment. Do not spray cleaning solutions toward any openings in the device.
CAUTION: Certain cleaning and sanitizing compounds may slowly degrade components made from some plastic materials. Using abrasive cleaners or cleaning solutions not recommended by Hospira may result in product damage and, potentially, void the product warranty. Do not use compounds containing combinations of isopropyl alcohol and dimethyl benzyl ammonium chloride. Do not use solvents that are harmful to plastic.
CAUTION: To avoid damage to the device, cleaning solutions should be used only as directed in Table 5-1. The disinfecting properties of cleaning solutions vary; consult the manufacturer for specific information.
DISCONNECT THE INFUSER FROM AC POWER PRIOR TO CLEANING THE DEVICE. FAILURE TO COMPLY WITH THIS WARNING COULD RESULT IN ELECTRICAL SHOCK.
Technical Service Manual 5 - 1 430-95424-002 (Rev. 01/06)
Page 64
SECTION 5 MAINTENANCE AND SERVICE TESTS
Table 5-1. Cleaning Solutions
Cleaning Solution Manufacturer Preparation
Coverage
Dispatch
Manu-Klenz
Precise
Sporicidin
Household bleach Various Per hospital procedures;
TM
HB Steris Corporation Per manufacturer's
recommendation
TM
®
TM
®
Caltech Industries Per manufacturer's
recommendation
Steris Corporation Per manufacturer's
recommendation
Caltech Industries Per manufacturer's
recommendation
Sporicidin International Per manufacturer’s
recommendation
do not exceed one part bleach in ten parts water
1. Clean the exposed surfaces of the infusion system with a soft, lint-free cloth
dampened with one of the cleaning solutions listed in Table 5-1, or a mild solution of soapy water.
2. Remove soap residue with clear water.
5.1.2

SANITIZING

Sanitize the external surfaces of the infuser using a cleaning solution listed in Table 5-1.
Note: Not all cleaning solutions are sanitizers. Check product labeling.
CAUTION: Do not sterilize the infuser using heat, steam, ethylene oxide (ETO), or radiation. These methods may cause the device to malfunction.
5.2

PERFORMANCE VERIFICATION TEST

The performance verification test (PVT) consists of the tests described in the following sections. The PVT can be used for diagnostic purposes during the troubleshooting of a malfunctioning infuser. The PVT should be used for performance verification before an infuser is placed back in service after repair. If any malfunction is detected as a result of the PVT, see Table 6-4.
Note: Perform the PVT exactly as described in this manual to assure effective
and reliable product evaluation information.
Note: When performing the PVT, all lines must be tested. However, if appropriate,
the test may be performed on all lines concurrently.
®
430-95424-002 (Rev. 01/06) 5 - 2 Plum A+
3 Infusion System
Page 65
5.2 PERFORMANCE VERIFICATION TEST
5.2.1

EQUIPMENT REQUIRED

The PVT requires the following equipment, or equivalents:
- Graduated cylinder, 25 mL, with 0.2 mL graduations (Type A)
- Sterile water or tap water in an IV bag/container
- Digital pressure meter (DPM), 0 to 50 psi (Fluke
- Three-way stopcock, latex-free (List No. 3233-01)
- IV Set (List No. 11419)
- 21-gauge butterfly needle, latex-free (List No. 4492-01), or 18-gauge blunt cannula
®
- Safety analyzer (Fluke
- Digital multimeter (DMM), (Fluke
- Nurse call test cable (P/N 561-88416-001) (optional)
Biomedical 232D)
®
187) (optional)
®
Biomedical DPM3)
5.2.2

INSPECTION

Inspect the infusion system periodically for signs of defects such as worn accessories, broken connections, or damaged cables. In addition, inspect the infusion system after repair or during cleaning. Replace any damaged or defective external parts.
Inspect the following areas for missing or damaged parts:
-Labels
- AC power cord, retainer, and straps
- Rubber foot pads
- Door assembly and handle
- Keypad and display
-LEDs
- External screws
- Pole clamp assembly
- Front and rear enclosures
- Battery doors
- Peripheral interface assembly and components
Technical Service Manual 5 - 3 430-95424-002 (Rev. 01/06)
Page 66
SECTION 5 MAINTENANCE AND SERVICE TESTS
5.2.3

TEST SETUP

WARNING:
To set up the infuser for the PVT, proceed as follows:
1. Confirm the infuser and appropriate accessories are assembled.
2. Hang two sterile water containers at a height of 18 ± 6 inches above the pumping chamber of the infuser.
3. Connect the infuser to AC power, and press [ON/OFF] to turn on the device.
4. Verify the infuser is in the unlocked mode. Toggling the [LOCKOUT] switch alternates between unlocked [DOWN] and locked [UP] modes.
5. Turn off the infuser.
A PATIENT SHOULD NEVER BE CONNECTED TO THE INFUSER DURING DEVICE TESTING.
5.2.4

SELF TEST

CAUTION: Do not place the infuser in service if the self test fails.
Note: Conduct all tests with the infuser connected to AC power unless otherwise
specified.
Note: If an alarm condition occurs during the self test, cycle the power and repeat the
self test. If the alarm condition recurs, note the message and take corrective action
(see Section 6). Repeat the self test. If the alarm condition continues to recur, remove
the infuser from service and contact Hospira.
To perform the self test, see Figure 5-1, and proceed as follows:
1. Connect the AC power cord to a grounded AC outlet. Verify the charge/line indicator CHARGE illuminates and an alarm beep sounds.
2. Without a cassette installed, turn on the infuser.
3. The LCD screen briefly displays the SELF TEST screen (see Figure 5-1).
Note: If the SELF TEST screen does not appear, contact Hospira.
4. After the self test is complete, the message “INSERT PLUM SET CLOSE LEVER” appears.
5. Verify the time and date. To set the time and date, see Section 1.8.3.
6. Open the cassette door and insert a primed cassette. Close the cassette door. The cassette test is complete when the “CASSETTE TEST IN PROGRESS” message disappears.
Note: The message “MECHANISM INITIALIZATION IN PROGRESS” may briefly
appear prior to the “CASSETTE TEST IN PROGRESS” message.
7. The “CLEAR SETTINGS?” message may appear. Press [YES], then turn off the infuser.
430-95424-002 (Rev. 01/06) 5 - 4 Plum A+
®
3 Infusion System
Page 67
5.2 PERFORMANCE VERIFICATION TEST
LINE FLOW
INDICATORS
A B
HOSPIRA Plum A+
STATUS REGION
MESSAGE REGION
SOFTKEY LABEL REGION
LINE INDICATOR
Version X.XX MM/DD/YY
START
STOP
CHARGE
ON/OFF
Copyright Hospira Inc.
2005
System Self Test
In Progress
1 4
2 5
78
CLEAR
0
3
6
9
.
WORKING
REGION
SOFTKEYS
SELECT
KEYPAD
SILENCE
05K01002
Figure 5-1. Display and Keypad
Technical Service Manual 5 - 5 430-95424-002 (Rev. 01/06)
Page 68
SECTION 5 MAINTENANCE AND SERVICE TESTS
5.2.5

CASSETTE ALARM TEST

To perform the cassette alarm test, proceed as follows:
1. Verify the infuser is on. Insert an empty cassette and close the door.
2. Verify the “CASSETTE TEST FAIL” message is flashing on the display and the alarm sounds after the cassette test is complete.
3. Open the door and remove the cassette.
4. Turn off the infuser.
5.2.6

FREE FLOW TEST

To perform the free flow test, proceed as follows:
1. With a primed cassette installed, turn on the infuser.
2. After the self test, press [YES] to clear settings.
3. Place the distal end of tubing into a collection container a minimum of 36 inches below the cassette.
4. With the cassette door closed, check the distal end of the tubing for fluid flow. Verify a minimal flow of fluid occurs (a few drops maximum).
5. Open the cassette door and check the distal end of the tubing for fluid flow. Verify a minimal flow of fluid occurs (a few drops maximum).
Note: A small amount of fluid may be expelled from the cassette when opening
or closing the door.
6. Close the cassette door.
5.2.7

DISPLAY TEST

To perform the display test, see Figure 5-1, then proceed as follows:
1. Verify the LCD backlight is illuminated and the display is clearly legible at eye level from approximately 18 inches.
2. With the infuser in the DELIVERY screen, press the [OPTIONS/VOL INF] softkey to select the OPTIONS screen.
3. Select Lighting/Contrast, and press [CHOOSE].
4. Press the [DECREASE SETTING] and [INCREASE SETTING] softkeys to change backlight intensity. Verify intensity decreases and increases.
5. Select Display Contrast.
6. Press [DECREASE SETTING] and [INCREASE SETTING] to change display contrast. Verify the display contrast decreases and increases.
7. Press the [CANCEL] softkey to return to the OPTIONS screen.
8. Press the [BACK] softkey to return to the DELIVERY screen.
430-95424-002 (Rev. 01/06) 5 - 6 Plum A+
®
3 Infusion System
Page 69
5.2 PERFORMANCE VERIFICATION TEST
5.2.8

KEYPAD VERIFICATION/FUNCTIONAL TEST

To perform the keypad verification/functional test, see Figure 5-1, then proceed as follows:
1. With the infuser in the DELIVERY screen, press the [A] softkey to select line A.
2. Verify the PROGRAM screen is displayed.
3. Enter a rate of 123 mL/hr and VTBI of 4567 mL.
4. Press [START] and verify the “CONFIRM PROGRAM?” message is displayed. If rate and VTBI are correct, press [YES].
5. Verify fluid is pumping, the message “PUMPING” is displayed in the line A status bar, and the line A LED flashes.
6. Press [STOP], then press and hold the [BACKPRIME] softkey.
7. Verify the “BACKPRIMING” and “RELEASE BACKPRIME TO STOP” messages are displayed, and verify the infuser is actually backpriming.
8. Release the [BACKPRIME] softkey, press [START], and verify normal pumping operation.
9. Press the [B] softkey.
10. Verify PIGGYBACK is the displayed delivery mode. If necessary, change the delivery
mode by pressing the [CHANGE MODE] softkey.
11. Enter a rate of 890 mL/hr and VTBI of 2 mL.
12. Press [START] and verify the “CONFIRM PROGRAM?” message is displayed. If rate and VTBI are correct, press [YES].
13. Verify fluid is pumping, the message “PUMPING” is displayed in the line B status bar, and the line B LED flashes.
14. After 20 seconds, verify pumping has switched to line A.
15. Press [STOP].
16. Press [OPTIONS/VOL INF]. Select Volume Infused, and press [CHOOSE].
17. Select line A.
18. Press [CLEAR]. Verify the line A volume is 0 mL and press [ENTER].
5.2.9

ALARM LOUDNESS TEST

To perform the alarm loudness test, see Figure 5-2, then proceed as follows:
1. Press the [A] softkey to select line A.
2. Enter a rate of 400 mL/hr and VTBI of 1 mL.
3. Press [START] and verify the “CONFIRM PROGRAM?” message is displayed. If rate and VTBI are correct, press [YES].
4. Verify fluid is pumping, the message “PUMPING” is displayed in the line A status bar, and the line A LED flashes.
5. Verify the alarm sounds when the dose has been delivered.
6. Turn the volume control knob between HIGH and LOW (see Figure 5-2). Verify the alarm loudness changes.
7. Press [SILENCE], and verify the alarm is silenced.
8. Press [STOP].
Technical Service Manual 5 - 7 430-95424-002 (Rev. 01/06)
Page 70
SECTION 5 MAINTENANCE AND SERVICE TESTS
VOLUME CONTROL KNOB
LOCKOUT SWITCH
NURSE CALL JACK
ROUND
SEAL (2)
POLE CLAMP
ASSEMBLY
RECTANGLE SEAL
DATAPORT CONNECTOR
PERIPHERAL INTERFACE ASSEMBLY
EQUIPOTENTIAL
POST
05K07016
Figure 5-2. Rear View
5.2.10

LOCKOUT SWITCH TEST

To perform the lockout switch test, proceed as follows:
1. Press the [A] softkey to select line A.
2. Enter a rate of 400 mL/hr and VTBI of 50 mL.
3. Press [START] and verify the “CONFIRM PROGRAM?” message is displayed. If rate and VTBI are correct, press [YES].
4. Verify fluid is pumping, the message “PUMPING” is displayed in the line A status bar, and the line A LED flashes.
5. Toggle the lockout alarm switch up (ON) to engage the alarm (see Figure 5-2).
6. Press any key except [STOP], and verify an alarm sounds and the “HARD LOCKOUT ENABLED” message is displayed. Verify the infuser continues to operate until [STOP] is pressed.
7. Press [STOP] and verify the “HARD LOCKOUT VIOLATION” message appears.
8. Toggle the lockout alarm switch down (OFF). Verify the “HARD LOCKOUT VIOLATION” message disappears and the alarm stops.
9. Press [START].
10. Open the door and verify the “DOOR OPEN WHILE PUMPING” message is displayed and the audio alarm activates.
11. Close the cassette door.
12. Press [NO] at the “CLEAR SETTINGS?” prompt.
430-95424-002 (Rev. 01/06) 5 - 8 Plum A+
®
3 Infusion System
Page 71
5.2 PERFORMANCE VERIFICATION TEST
5.2.11

PROXIMAL OCCLUSION TEST

To perform the proximal occlusion test, proceed as follows:
1. Press the [A] softkey to select line A.
2. Enter a rate of 400 mL/hr and VTBI of 50 mL.
3. Press [START] and verify the “CONFIRM PROGRAM?” message is displayed. If rate and VTBI are correct, press [YES].
4. Verify fluid is pumping, the message “PUMPING” is displayed in the line A status bar, and the line A LED flashes.
5. After several pumping cycles, clamp line A tubing proximal to the cassette. Verify the “PROX OCCL A/AIR” message flashes and the alarm sounds before three pumping cycles are completed.
6. Press [SILENCE] and verify the alarm stops while the message on the display continues to flash.
7. Unclamp the proximal line and press [START]. Verify pumping resumes.
8. Press [STOP].
5.2.12

PROXIMAL AIR-IN-LINE TEST

To perform the proximal air-in-line alarm test, see Figure 5-3, then proceed as follows:
1. Install the special cassette marked proximal, and close the cassette door.
Note: Confirm the special cassette proximal bubble sensor tips are removed.
2. Press [YES] to clear settings.
3. Press the [A] softkey to select line A.
4. Enter a rate of 400 mL/hr and VTBI of 50 mL.
5. Press [START] and verify the “CONFIRM PROGRAM?” message is displayed. If rate and VTBI are correct, press [YES].
6. Verify fluid is pumping, the message “PUMPING” is displayed in the line A status bar, and the line A LED flashes.
7. Before 1 mL of fluid is delivered, verify the alarm sounds and the “PROX AIR A. BACKPRIME” message is flashing on the display.
8. Open the door, and remove the special cassette.
Technical Service Manual 5 - 9 430-95424-002 (Rev. 01/06)
Page 72
SECTION 5 MAINTENANCE AND SERVICE TESTS
PROXIMAL BUBBLE SENSOR BULB TIPS (REMOVED FOR PROXIMAL-AIR-IN-LINE ALARM TEST)
(REMOVED FOR DISTAL-AIR-IN-LINE ALARM TEST)
CASSETTE CENTERING DEVICE
DISTAL BUBBLE SENSOR BULB TIPS
98G01024
Figure 5-3. Special Cassettes with Bubble Sensor Tips Removed
5.2.13

DISTAL AIR-IN-LINE TEST

To perform the distal air-in-line alarm test, see Figure 5-3, then proceed as follows:
1. Install the special cassette marked distal, and close the cassette door.
Note: Confirm the special cassette proximal bubble sensor tips are removed.
2. Press [YES] to clear settings.
3. Press the [A] softkey to select line A.
4. Enter a rate of 400 mL/hr and VTBI of 50 mL.
5. Press [START] and verify the “CONFIRM PROGRAM?” message is displayed. If rate and VTBI are correct, press [YES].
6. Verify fluid is pumping, the message “PUMPING” is displayed in the line A status bar, and the line A LED flashes.
7. Before 1 mL of fluid is delivered, verify the alarm sounds and the “DISTAL AIR” message is flashing on the display.
8. Open the door, and remove the special cassette.
430-95424-002 (Rev. 01/06) 5 - 10 Plum A+
®
3 Infusion System
Page 73
5.2 PERFORMANCE VERIFICATION TEST
5.2.14

DISTAL OCCLUSION TEST

To perform the distal occlusion test, see Figure 5-4, then proceed as follows:
1. Install the cassette and connect the distal tubing to the DPM through a three-way stopcock as illustrated in Figure 5-4. Close the cassette door.
Note: A reflux valve may be attached between the stopcock and the DPM
to keep moisture out of the DPM.
Note: The height of the DPM must be 0 ± 12 inches from the midline of the
pumping chamber.
2. Turn on the infuser.
3. Press [YES] to clear settings.
4. Press [OPTIONS/VOL INF] to select the OPTIONS screen.
5. Select Pressure/Post Infusion Rate, and press [CHOOSE].
6. Verify the distal pressure limit is set at 6 psi. If the pressure limit is not 6 psi, enter 6 and press [ENTER].
7. Press the [A] softkey to select line A.
8. Enter a rate of 40 mL/hr and VTBI of 50.0 mL.
9. Open the three-way stopcock to air.
10. Press [START] and verify the “CONFIRM PROGRAM?” message is displayed. If rate and VTBI are correct, press [YES].
11. Verify fluid is pumping, the message “PUMPING” is displayed in the line A status bar, and the line A LED flashes.
12. Set the three-way stopcock to measure pressure.
13. Verify the distal occlusion audible alarm occurs at 6 ± 3 psi. Verify the DISTAL OCCLUSION message is flashing on the screen.
14. Open the three-way stopcock to air.
15. Open and close the door. Press [NO] at the “CLEAR SETTINGS?” prompt.
16. Press [OPTIONS/VOL INF] to select the OPTIONS screen.
17. Select Pressure/Post Infusion Rate, and press [CHOOSE].
18. Select Distal Pressure Limit. Enter 10 psi, and press [ENTER].
19. Set the three-way stopcock to measure pressure, then press [START].
20. Verify the distal occlusion audible alarm occurs at 10 ± 3 psi. Verify the DISTAL OCCLUSION message is flashing on the screen.
21. Open the door and remove the cassette.
Technical Service Manual 5 - 11 430-95424-002 (Rev. 01/06)
Page 74
SECTION 5 MAINTENANCE AND SERVICE TESTS
OFF
mmHg
cmOFHO
-13.5T O1 5
-13.5T O7 5
PSI
INCHES OFH O
UNIVERSAL
PRESSUREMETER
PRESSUREINPUT
OFF
mmHg
cmOFHO
-13.5T O1 5
-13.5T O7 5
PSI
INCHES OFH O
UNIVERSAL
PRESSUREMETER
PRESSUREINPUT
FROM FLUID CONTAINERS
1
23
THREE-WAY STOPCOCK
UNIVERSAL
UNIVERSAL
PRESSUREMETER
PRESSUREMETER
cmOFHO
2
cmOFHO
2
mmHg
mmHg
OFF
OFF
PRESSUREINPUT
PRESSUREINPUT
0.00
0.00
INCHES OFH O
INCHES OFH O
-13.5T O1 5
-13.5T O1 5
-13.5T O7 5
-13.5T O7 5
2
2
PSI
PSI
DPM
01K07018
Figure 5-4. Distal Occlusion Test Setup
5.2.15

DELIVERY ACCURACY TEST

Note: Accuracy testing is for informational purposes only, and is not to be used
as a re-release test. If there is any concern as to infuser accuracy, contact Hospira.
CAUTION: Do not remove the protective cover from the 21-gauge needle.
To perform the delivery accuracy test, proceed as follows:
1. Open the cassette door and insert a primed cassette. Close the cassette door.
2. Install an 18-gauge blunt cannula or a 21-gauge needle to the distal end of the tubing. Verify the fluid container is 18 to 24 inches above the pumping chamber. Verify all lines are unclamped.
3. Place the distal output end of tubing into the graduated cylinder.
4. Press the [A] softkey to select line A.
5. Enter a rate of 200 mL/hr and VTBI of 10 mL.
6. Press [START] and verify the “CONFIRM PROGRAM?” message is displayed. If rate and VTBI are correct, press [YES].
7. Verify fluid is pumping, the message “PUMPING” is displayed in the line A status bar, and the line A LED flashes.
8. Press the [B] softkey to select line B.
9. Verify the infuser is in the PIGGYBACK delivery mode. If necessary, press [CHANGE MODE] to change the delivery mode.
10. Enter a rate of 200 mL/hr and VTBI of 10 mL.
430-95424-002 (Rev. 01/06) 5 - 12 Plum A+
®
3 Infusion System
Page 75
5.2 PERFORMANCE VERIFICATION TEST
11. Press [START] and verify the “CONFIRM PROGRAM?” message is displayed. If rate and VTBI are correct, press [YES].
12. Verify fluid is pumping, the message “PUMPING” is displayed in the line B status bar, and the line B LED flashes.
13. Verify the “KVO” message flashes on the display and an audible alarm sounds when total delivery is complete on line A.
14. Press [STOP] and verify the volume delivered is 20 ± 1 mL.
5.2.16

NURSE CALL TEST

Note: The nurse call test may be bypassed if the nurse call function is not used.
To perform the nurse call test, attach the nurse call test cable and proceed as follows:
1. Set the primary delivery rate to 400 mL/hr, and the primary dose limit to 1 mL.
2. Connect the DMM to the nurse call test cable.
3. Press [START] and verify pumping action.
4. After “DOSE END” and “KVO” appear on the display, observe a short circuit on the DMM (approximately 1 Ω on a scale of 0 to 100 Ω).
5.2.17

ELECTRICAL SAFETY TEST

To perform the electrical safety test, proceed as follows:
1. Connect the AC power cord to a safety analyzer.
2. Connect the safety analyzer ground lead to the ground test-point located on the rear of the infuser.
3. Check the leakage current with the safety analyzer. Leakage current (both open and closed ground) must not exceed 100 microamperes AC
4. Measure the resistance of the AC connector ground lug with the safety analyzer. Resistance should not exceed 0.1 Ω.
rms
.
5.2.18

END OF THE PVT

If all performance verification tests have been successful, proceed as follows:
1. Press [OPTIONS/VOL INF]. Select Volume Infused, and press [CHOOSE].
2. Press [CLEAR] to clear the volume infused.
3. Press [ENTER].
4. Press the [A] softkey.
5. Press [YES] at the “CLEAR LINE A SETTINGS?” prompt.
6. Press [CANCEL/BACK] to return to the delivery screen.
7. Press the [B] softkey.
8. Press [YES] at the “CLEAR LINE B SETTINGS?” prompt
9. Reset the infuser to the original configuration.
10. Turn off the infuser and return the device to service.
Note: If any tests fail, see Section 6, or contact Hospira.
Technical Service Manual 5 - 13 430-95424-002 (Rev. 01/06)
Page 76
SECTION 5 MAINTENANCE AND SERVICE TESTS
5.3

PERIODIC MAINTENANCE INSPECTION

Periodic maintenance inspections should be performed per hospital procedures for compliance to accreditation requirements. It is recommended that JCAHO and/or hospital protocol be followed for establishing a periodic maintenance inspection schedule. Product specifications for this inspection are listed in Section 8.
To perform the periodic maintenance inspection, complete the PVT in Section 5.2.
5.4

BATTERY OPERATION OVERVIEW

The infusion system is intended to operate on battery power on an exception basis only, such as emergency backup or temporary portable operation. Examples of emergency backup include AC power failure or inadvertent disconnection of the AC power cord. An instance of temporary portable operation includes patient transfer from one location to another.
The device should be connected to AC power whenever possible to allow the battery to remain fully charged. The line power indicator turns off when the infuser is operating on battery power. The backlight extinguishes after approximately one minute of operation on battery power.
Factors that most commonly affect battery life are the depth and frequency of discharge and the length of the recharge period. As a general rule, the more often the battery is discharged and recharged, the sooner it will need replacement. The primary cause of damage is leaving the battery in a less than fully charged state for any period of time. Battery damage can occur in a matter of hours and cause a permanent loss of battery capacity. The amount of lost capacity depends on the degree of discharge, the storage temperature, and the length of time the battery was stored in a discharged state.
Note: A permanently damaged battery cannot be recharged to full capacity.
When the battery discharges below the acceptable level while the infuser is operating, the audio indicator is activated and the “WARNING: LOW BATTERY” message displays. Although it is not recommended to continue operating the infuser on battery power at this point, the battery continues providing power until it is depleted. When the battery is depleted, delivery stops, a continuous alarm tone sounds, and, after three minutes, the infuser automatically turns off.
CAUTION: As soon as the low battery alarm occurs, connect the infuser to AC power.
When the infuser detects that the battery has reduced capacity, it will register a Replace Battery condition. For the first two occurrences of a Replace Battery condition, the “WARNING: LOW BATTERY” message will appear and the audio indicator will activate. The message and audio indicator can be cleared only when the device is plugged in or turned off. For the third and subsequent occurrences, the “WARNING: REPLACE BATTERY” message will appear, and the audio indicator will activate and persist over power cycles. The message and audio indicator are cleared by replacing the battery, accessing the biomed settings screen, and pressing the [CHANGE BATTERY] softkey.
430-95424-002 (Rev. 01/06) 5 - 14 Plum A+
®
3 Infusion System
Page 77
5.4 BATTERY OPERATION OVERVIEW
Recharging can occur any time the infuser is connected to AC power. It is recommended that the infuser be connected to AC power whenever practical to maximize available battery charge during transport or ambulation. The infuser does not have to be on for the battery to recharge.
Note: The infuser should be operated on battery power for three continuous hours
at least once every six months for optimum battery performance and life.
Technical Service Manual 5 - 15 430-95424-002 (Rev. 01/06)
Page 78
SECTION 5 MAINTENANCE AND SERVICE TESTS
This page intentionally left blank.
430-95424-002 (Rev. 01/06) 5 - 16 Plum A+
®
3 Infusion System
Page 79
Section 6

TROUBLESHOOTING

This section contains information on technical assistance, warning messages, alarm messages and error codes, and troubleshooting procedures for the Plum A+ infusion system.
®
6.1

TECHNICAL ASSISTANCE

For technical assistance, product return authorization, and to order parts, accessories, or manuals within the United States, contact Hospira Technical Support Operations.
1-800-241-4002
For additional technical assistance, technical training, and product information, visit the website at www.hospira.com.
Send all authorized, prepaid returns within the United States to the following address:
Hospira, Inc.
Technical Support Operations
755 Jarvis Drive
Morgan Hill, California 95037
For technical assistance, product return authorization, and to order parts, accessories, or manuals from outside the United States, contact the nearest Hospira sales office.
3
6.2

WARNING MESSAGES

Table 6-1 lists warning messages, possible causes, and corrective actions. These warning
messages are captured in the Error Log.
Note: When the infuser detects that the battery has reduced capacity, it will register
a Replace Battery condition. For the first two occurrences of a Replace Battery condition, the “WARNING: LOW BATTERY” message will appear and the audio indicator will activate. The message and audio indicator can be cleared only when the device is plugged in or turned off. For the third and subsequent occurrences, the “WARNING: REPLACE BATTERY” message will appear, and the audio indicator will activate and persist over power cycles. The message and audio indicator are cleared by replacing the battery, accessing the biomed settings screen, and pressing the [CHANGE BATTERY] softkey.
Note: If the device is not plugged in, and turned on with a previously depleted battery,
the infuser will display a “DEPLETED BATTERY” message for 16 seconds, then power off.
Technical Service Manual 6 - 1 430-95424-002 (Rev. 01/06)
Page 80
SECTION 6 TROUBLESHOOTING
Message Possible Cause Corrective Action
Table 6-1. Warning Messages
Stop delivery, then turn off Attempting to turn off the
infuser while a delivery is in progress
Warning: Low Battery Battery is discharged so
that only approximately 30 minutes of battery life remains
Warning: Replace Battery Battery service needed
Battery voltage is less than the depleted threshold and the charge level is higher than the low charge threshold
Warning: Charger Service A hardware problem with
the battery charging circuit is detected; charging circuitry is not behaving as expected
Stop all lines, then turn off the infuser
Plug into AC power
Replace the battery
Press [SILENCE]
6.3

ALARM MESSAGES AND ERROR CODES

Under most alarm conditions the infuser ceases normal operation, generates an audible alarm, and displays an alarm message or error code on the LCD screen.
There are two types of alarm conditions:
- alarm codes that can be cleared by the operator
- error codes that require qualified service personnel
6.3.1

OPERATIONAL ALARM MESSAGES

Table 6-2 lists infuser alarm codes that can be cleared by the operator. Also listed in Table 6-2 are the alarm messages, descriptions, possible causes, and corrective actions.
Note: Operational alarm messages are displayed on the LCD screen. Associated error
codes are displayed in the alarms log (see Section 1.8.2).
430-95424-002 (Rev. 01/06) 6 - 2 Plum A+
®
3 Infusion System
Page 81
6.3 ALARM MESSAGES AND ERROR CODES
Table 6-2. Operational Alarm Messages and Corrective Actions
Alarm
Code
N100
(URC)
N101
(NAA)
N102
(RL)
N103
(SEEP
CRC)
N104
(NC2)
Alarm Description Possible Cause
Unrecognizable cassette
Incorrect cassette type
An incorrect cassette is inserted
No action No operator action
and no delivery for two minutes
Interruption or a partial change to
a program during delivery parameters entry
Infuser idle 2 minutes
NV RAM lost thrpy data
Infuser in reset or idle for over two minutes
Programming
set without start
for two minutes
Therapy data is lost Infuser did not
complete the
previous
non-volatile
memory write
successfully
Nurse callback B Delivery line B
has changed (if alarm is enabled)
End of delivery
step on line B
other than VTBI
complete while
callback is
enabled
Corrective
Action
Insert proper cassette
Complete programming of the infuser
Press [START]
Re-enter all programmed data
Press [SILENCE]
N105
(NC1)
N160
or
E160
(VTB2)
N161
or
E161
(VTB1)
N180
or
E180
(OD1)
N181
or
E181
(OD1)
Nurse callback A Delivery line A
has changed (if alarm is enabled)
Line B VTBI complete
Programmed volume to be infused completed on line B
Line A VTBI complete
Programmed volume to be infused completed on line A
Distal Occl Peak distal
occlusion, non-delivery
Distal Occl Negative distal
occlusion, non-delivery
End of delivery
step on line A
other than VTBI
complete while
callback is
enabled
VTBI is complete
on line B
VTBI is complete
on line A
Distal occlusion
detected during
non-delivery
Distal occlusion
detected during
non-delivery
Press [SILENCE]
Press [SILENCE] and replace IV bag, and restart line B
Press [SILENCE] and replace IV bag, and restart line A
Backprime the cassette and restart the infuser
Backprime the cassette and restart the infuser
Technical Service Manual 6 - 3 430-95424-002 (Rev. 01/06)
Page 82
SECTION 6 TROUBLESHOOTING
Table 6-2. Operational Alarm Messages and Corrective Actions
Alarm
Code
N182
or
E182
(OP2)
N183
or
E183
(OP2)
N184
or
E184
(OP1)
Alarm Description Possible Cause
Prox. Occl B, Air or Prox. Occl B
Negative proximal occlusion B, non-delivery
Proximal occlusion detected on line B during non-delivery
Prox. Occl B, Air or Prox. Occl B
Peak proximal occlusion B, non-delivery
Proximal occlusion detected on line B during non-delivery
Prox. Occl A, Air or Prox. Occl A
Negative proximal occlusion A, non-delivery
Proximal occlusion detected on line A during non-delivery
Corrective
Action
Backprime the cassette and restart line B or Stop all lines, backprime the cassette, and restart all lines
Backprime the cassette and restart line B or Stop all lines, backprime the cassette, and restart all lines
Backprime the cassette and restart line A or Stop all lines, backprime the cassette, and restart all lines
N185
or
E185
(OP1)
N186
or
E186
(OD1)
N187
or
E187
(OD1)
N188
or
E188
(OP2)
Prox. Occl A, Air or Prox. Occl A
Peak proximal occlusion A, non-delivery
Distal Occl Peak distal
occlusion, delivery
Distal Occl Negative distal
occlusion, delivery
Prox. Occl B, Air Negative proximal
occlusion B, delivery
Proximal occlusion detected on line A during non-delivery
Distal occlusion detected during delivery
Distal occlusion detected during delivery
Proximal occlusion detected during delivery on line B
Backprime the cassette and restart line A or Stop all lines, backprime the cassette, and restart all lines
Fix occlusion and restart the infuser
Fix occlusion and restart the infuser
Fix occlusion and restart line B or Stop all lines, fix occlusion and restart the infuser
430-95424-002 (Rev. 01/06) 6 - 4 Plum A+
®
3 Infusion System
Page 83
6.3 ALARM MESSAGES AND ERROR CODES
Table 6-2. Operational Alarm Messages and Corrective Actions
Alarm
Code
N189
or
E189
(OP2)
N190
or
E190
(OP1)
N191
or
E191
(OP1)
Alarm Description Possible Cause
Prox. Occl B, Air Peak proximal
occlusion B, delivery
Proximal
occlusion
detected during
delivery on line B
Prox. Occl A, Air Negative proximal
occlusion A, delivery
Proximal
occlusion
detected during
delivery on line A
Prox. Occl A, Air Peak proximal
occlusion A, delivery
Proximal
occlusion
detected during
delivery on line A
Corrective
Action
Fix occlusion and restart line B or Stop all lines, fix occlusion and restart the infuser
Fix occlusion and restart line A or Stop all lines, fix occlusion and restart the infuser
Fix occlusion and restart line A or Stop all lines, fix occlusion and restart the infuser
N230
or
E230
(APT)
N231
or
E231
(APB)
N232
or
E232
(APA)
Prox. Air Total Proximal
air-in-line total
Prox. Air on B, backprime
Proximal air-in-line on line B
Prox. Air on A, backprime
Proximal air-in-line on line A
500 µL of air
has entered
the cassette
500 µL of air
has entered
the cassette
on line B
500 µL of air
has entered
the cassette
on line A
Backprime the cassette and restart the infuser or Remove and manually reprime the cassette, and restart the infuser
Backprime the cassette and restart line B or Remove and manually reprime the cassette and restart the infuser
Backprime the cassette and restart line A or Remove and manually reprime the cassette and restart the infuser
Technical Service Manual 6 - 5 430-95424-002 (Rev. 01/06)
Page 84
SECTION 6 TROUBLESHOOTING
Table 6-2. Operational Alarm Messages and Corrective Actions
Alarm
Code
N233
or
E233
(ADC)
N234
or
E234
(ADB)
N250
or
E250
(DCO1)
N251
or
E251
(CS1)
Alarm Description Possible Cause
Distal air cumulative
Distal air cumulative
500 µL of air detected in the last 5.3 mL of fluid delivered
Distal air bolus Distal air bolus 100 µL bolus
of air detected at distal sensor
Door opened while pumping
Valve/cass test fail
Door opened while pumping
Valve/cassette test failure
Door opened while pumping
Valve/cassette fails the leak test
Corrective
Action
Remove and manually reprime the cassette and restart the infuser
Remove and manually reprime the cassette and restart the infuser
Turn off the infuser or Insert the cassette and close the door
Replace cassette and retest or Backprime and retest
N252
or
E252
(BDP)
N253
or
E253
(LOV)
N254
or
E254
(FPL)
Depleted battery Low battery The battery
terminal voltage is less than
5.45 V
Lockout violation Hard lockout
violation
The use of the [STOP] key or an attempt to open the door while lockout switch is locked
Lockout Enabled Keypad locked Any action not
resulting in stopping of the delivery while the lockout switch is locked
Connect the infuser to AC power or Recharge or replace the battery
Unlock the lockout switch
Unlock the lockout switch
430-95424-002 (Rev. 01/06) 6 - 6 Plum A+
®
3 Infusion System
Page 85
6.3 ALARM MESSAGES AND ERROR CODES
Table 6-2. Operational Alarm Messages and Corrective Actions
Alarm
Code
N255
(SLV)
N256
(SLE)
Alarm Description Possible Cause
Lockout violation Soft lockout
violation
Lockout enabled Soft lockout
enabled
The use of the
[STOP] key or an
attempt to open
the door while the
lockout switch
is locked
Any action not
resulting in
stopping of
delivery while
the lockout
switch is locked
Corrective
Action
Unlock the software lockout switch
Unlock the software lockout switch
6.3.2

ERROR CODES REQUIRING TECHNICAL SERVICE

Table 6-3 lists infusion system error codes that require technical service. Also listed
in Table 6-3 are malfunction descriptions, possible causes, and corrective actions.
Table 6-3. Error Codes Requiring Technical Service
Error
Code Malfunction Possible Cause Corrective Action
E300 ADC failure Analog to digital
converter failure
E301 Audio alarm
failure
E302 Backlight failure Backlight (CCFT tube) is
Piezo is off but sensed on or Piezo is on but sensed off
not at the expected range
Replace CPU PWA
(see Section 7.2.14.6)
Reset time and date, if required
(see Section 1.8.3)
Turn power off, then on, to reset the infuser
Replace piezo alarm
(see Section 7.2.14.7)
Replace CPU PWA
(see Section 7.2.14.6)
Reset time and date, if required
(see Section 1.8.3)
Turn power off, then on, to reset the infuser
Replace display
(see Section 7.2.14.3)
Reset time and date, if required
(see Section 1.8.3)
Technical Service Manual 6 - 7 430-95424-002 (Rev. 01/06)
Page 86
SECTION 6 TROUBLESHOOTING
Table 6-3. Error Codes Requiring Technical Service
Error
Code
Malfunction Possible Cause Corrective Action
E320 Battery charge
current out of range
E321 Battery not
charging
E322 Battery current
calibration value out of range
E323 Battery trickle
charge current out of range
E324 Supply
overvoltage
E325 Battery
overvoltage
Battery charge current is out of range after 8 hours
Battery charging timed out Complete battery discharge
has occurred
Battery integrator calibration value is out of range
Battery trickle charge current is out of range
An overvoltage condition is detected in the charging circuit
An overvoltage condition is detected in the battery
Replace battery
(see Section 7.2.4)
Replace power supply PWA
(see Section 7.2.14.1)
Reset time and date, if required
(see Section 1.8.3)
Charge battery for additional 8 hours
Replace battery
(see Section 7.2.4)
Replace power supply PWA
(see Section 7.2.14.1)
Reset time and date, if required
(see Section 1.8.3)
E326 Battery
disconnected
E327 Brownout
condition
E340 Critical
instruction failure
E341 Critical data
memory failure
Battery disconnected while the infuser is on
Check for loose battery connections
Replace battery
(see Section 7.2.4)
Reset time and date, if required
(see Section 1.8.3)
Brownout condition detected Replace power supply PWA
(see Section 7.2.14.1)
Reset time and date, if required
(see Section 1.8.3)
Power-up CPU register test failed (no malfunction message displayed)
Replace CPU PWA
(see Section 7.2.14.6)
Reset time and date, if required
(see Section 1.8.3)
Critical data memory failure Replace mechanism
assembly
(see Section 7.2.14.8)
Reset time and date, if required
(see Section 1.8.3)
430-95424-002 (Rev. 01/06) 6 - 8 Plum A+
®
3 Infusion System
Page 87
6.3 ALARM MESSAGES AND ERROR CODES
Table 6-3. Error Codes Requiring Technical Service
Error
Code
Malfunction Possible Cause Corrective Action
E342 Display failure Defective display Replace display
(see Section 7.2.14.3)
Reset time and date, if required
(see Section 1.8.3)
E343 Distal air sensor
failure 1
E344 Distal air sensor
failure 2
E345 Distal pressure
sensor failure 1
E346 Distal pressure
sensor failure 2
E347 Hardware
watchdog failure
E378 I/O valve
phase loss
With the cassette removed, the distal air sensor self test detects liquid
With the cassette inserted, the distal air sensor self test detects sensor out of range
Replace mechanism assembly
(see Section 7.2.14.8)
Reset time and date, if required
(see Section 1.8.3)
Distal air sensor failed while the infuser is off
Distal air sensor failed while the infuser is on
Hardware watchdog failure Replace CPU PWA
(see Section 7.2.14.6)
Reset time and date, if required
(see Section 1.8.3)
Generic I/O valve failure Turn power off, then on,
to reset the infuser Replace mechanism
assembly
(see Section 7.2.14.8)
Reset time and date, if required
(see Section 1.8.3)
E379 L/S valve
phase loss
E380 Plunger motor
phase loss
Generic L/S valve failure Turn power off, then on,
to reset the infuser
Generic plunger motor failure
Replace mechanism assembly
(see Section 7.2.14.8)
Reset time and date, if required
(see Section 1.8.3)
Technical Service Manual 6 - 9 430-95424-002 (Rev. 01/06)
Page 88
SECTION 6 TROUBLESHOOTING
Table 6-3. Error Codes Requiring Technical Service
Error
Code
Malfunction Possible Cause Corrective Action
E430 Proximal
air sensor failure 1
E431 Proximal
air sensor failure 2
E432 Proximal
pressure
Proximal air sensor ongoing test detects liquid with cassette removed
Proximal air sensor self test detects liquid with cassette removed
Proximal pressure sensor failed while the infuser is off
Replace mechanism assembly
(see Section 7.2.14.8)
Reset time and date, if required
(see Section 1.8.3)
sensor 1
E433 Proximal
pressure
Proximal pressure sensor failed while the infuser is on
sensor 2
E434 RAM failure RAM failure Turn power off, then on,
to reset the infuser Replace peripheral
assembly
(see Section 7.2.7)
Reset time and date, if required
(see Section 1.8.3)
E435 RTC failure Real-time clock failure Turn power off, then on,
to reset the infuser Replace CPU PWA
(see Section 7.2.14.6)
Reset time and date, if required
(see Section 1.8.3)
E436 ROM failure ROM checksum failure Turn power off, then on,
to reset the infuser Replace peripheral
assembly
(see Section 7.2.7)
Reset time and date, if required
(see Section 1.8.3)
E437 Software failure Generic software failure Turn power off, then on,
to reset the infuser
E438 Stack
out-of-range failure
Stack out-of-range failure
Replace CPU PWA
(see Section 7.2.14.6)
Reset time and date, if required
(see Section 1.8.3)
430-95424-002 (Rev. 01/06) 6 - 10 Plum A+
®
3 Infusion System
Page 89
Error
Code
6.3 ALARM MESSAGES AND ERROR CODES
Table 6-3. Error Codes Requiring Technical Service
Malfunction Possible Cause Corrective Action
E439 Stuck key A key is sensed as pressed
for over two minutes
E440 Power hold stuck Power hold signal stuck
Power cannot be turned off
Replace keypad
(see Section 7.2.14.2)
Reset time and date, if required
(see Section 1.8.3)
E443 LCD failure LCD bias is out of range Replace display
(see Section 7.2.14.3)
Reset time and date, if required
(see Section 1.8.3)
E444 CPU timebase
inaccurate
CPU timer 2 and RTC measured times disagree
Turn power off, then on, to reset the infuser
Replace CPU PWA
(see Section 7.2.14.6)
Reset time and date, if required
(see Section 1.8.3)
E445 RTC memory
failure
Real-time clock memory is corrupt
Turn power off, then on, to reset the infuser
Reset time and date, if required
(see Section 1.8.3)
E446 CPU timer failure CPU timer 1 and timer 2
measured times disagree
E447 Battery ADC
reading failure
16 consecutive readings have been either all zero or the max value
E448 SEEP write
SEEP data write failed Replace mechanism
failure
E449 SEEP calibration
data corrupted
E450 MMIO port
Calibration data block corrupted
I/O port read/write failure Replace CPU PWA
read/write failure
Replace CPU PWA
(see Section 7.2.14.6)
Reset time and date, if required
(see Section 1.8.3)
assembly
(see Section 7.2.14.8)
Replace CPU PWA
(see Section 7.2.14.6)
Replace CPU/driver cable
(see Section 7.2.14.4)
Reset time and date, if required
(see Section 1.8.3)
(see Section 7.2.14.6)
Reset time and date, if required
(see Section 1.8.3)
Technical Service Manual 6 - 11 430-95424-002 (Rev. 01/06)
Page 90
SECTION 6 TROUBLESHOOTING
Table 6-3. Error Codes Requiring Technical Service
Error
Code
Malfunction Possible Cause Corrective Action
E451 Inaccurate
delivery
E452 Software failure Miscellaneous
E453 Two SEEP
CRC errors
E454 NVRAM over
capacity
E455 Invalid device
configuration
E456 Invalid drug
library
Over/under delivery detected Turn power off, then on,
software failures
NVRAM data block corrupted Replace mechanism
Software trying to write into non-existent NVRAM space
Incorrect flash memory on peripheral PWA
A drug library install was started but not completed successfully
to reset the infuser Reset time and date,
if required
(see Section 1.8.3)
If error codes recur,
contact Hospira
assembly
(see Section 7.2.14.8)
Replace CPU PWA
(see Section 7.2.14.6)
Replace CPU/driver cable
(see Section 7.2.14.4)
Reset time and date, if required
(see Section 1.8.3)
Turn power off, then on, to reset the infuser
Replace peripheral PWA
(see Section 7.2.8)
Attempt to reinstall the library install
(see the system operating manual)
Replace peripheral assembly
(see Section 7.2.7)
E457 Drug library
corrupted
CRC failure on drug library Reload the library
(see the system operating manual)
Note: The following error codes are not generated in the biomed service mode:
E320
E321
E322
E323
E324
E325
E326
E343
E345
E346
E371
E372
E373
E374
E375
E376
E377
E378
E379
E380
E430
E431
E432
E433
E441
E447
Note: Some error codes include sub-ID codes. These sub-ID codes are intended for
Hospira internal use only, and should be included when contacting Hospira Technical
Support Operations (see Section 6.1).
®
430-95424-002 (Rev. 01/06) 6 - 12 Plum A+
3 Infusion System
Page 91
6.4 TROUBLESHOOTING PROCEDURES
6.4

TROUBLESHOOTING PROCEDURES

This section details recommended procedures for problems not associated with malfunction alarms. Before performing any troubleshooting procedure, turn the infuser off, then on.
Allow the self test to complete and proceed as follows:
1. If a malfunction exists, carefully inspect the infuser for damage as described in Section 5.2.2.
2. If an infuser inspection has not disclosed a malfunction, perform the PVT in Section 5.2. See Table 6-4 for section reference, probable cause, and corrective actions.
3. If, after completing step 1 and step 2, a malfunction has not been located, or if the infuser persistently fails, contact Hospira (see Section 6.1).
Table 6-4. Troubleshooting with the PVT
Test Failure Probable Cause Corrective Action
Self test
Section 5.2.4
Cassette alarm test
Section 5.2.5
Free flow test
Section 5.2.6
Display test
Section 5.2.7
Keypad verification/ functional test
Section 5.2.8
Cassette not properly installed
Defective CPU PWA Replace CPU PWA
Cassette not properly seated
Defective cassette Replace cassette
Cassette not properly seated
Defective cassette Replace cassette
Defective or dirty valve pins Clean valve pins
Defective display assembly Replace display
Defective keypad Replace keypad
Reseat cassette
(see Section 7.2.14.6)
Reseat cassette
Reseat cassette
Replace mechanism assembly
(see Section 7.2.14.8)
(see Section 7.2.14.3)
(see Section 7.2.14.2)
Alarm loudness test
Section 5.2.9
Technical Service Manual 6 - 13 430-95424-002 (Rev. 01/06)
Defective CPU Replace CPU PWA
(see Section 7.2.14.6)
Defective peripheral PWA Replace peripheral PWA
(see Section 7.2.8)
Defective piezo alarm assembly
Replace piezo alarm assembly
(see Section 7.2.14.7)
Page 92
SECTION 6 TROUBLESHOOTING
Table 6-4. Troubleshooting with the PVT
Test Failure Probable Cause Corrective Action
Lockout switch test
Section 5.2.10
Proximal occlusion test
Section 5.2.11
Proximal air-in-line test
Section 5.2.12
Distal air-in-line test
Section 5.2.13
Defective peripheral PWA Replace peripheral PWA
(see Section 7.2.8)
Closed proximal clamp Open clamp
Cassette not properly
Re-prime cassette
primed
Defective cassette Replace cassette
Dirty sensor pin Clean sensor pin
Defective APP PWA Replace mechanism assembly
(see Section 7.2.14.8)
Defective special cassette Replace special cassette
Dirty sensors Clean sensors
Defective APP PWA Replace mechanism assembly
(see Section 7.2.14.8)
Defective special cassette Replace special cassette
Dirty sensors Clean sensors
Defective APP PWA Replace mechanism assembly
(see Section 7.2.14.8)
Distal occlusion test
Section 5.2.14
Delivery accuracy test
Section 5.2.15
Electrical safety test
Section 5.2.16
Cassette not properly
Re-prime cassette
primed
Defective cassette Replace cassette
Dirty sensor pin Clean sensor pin
Defective APP PWA Replace mechanism assembly
(see Section 7.2.14.8)
Set not properly primed Re-prime cassette
Damaged or faulty cassette Replace cassette
Defective mechanism assembly
Replace mechanism assembly
(see Section 7.2.14.8)
Defective AC power cord Replace AC power cord
(see Section 7.2.5)
430-95424-002 (Rev. 01/06) 6 - 14 Plum A+
®
3 Infusion System
Page 93
Section 7

REPLACEABLE PARTS AND REPAIRS

This section itemizes all parts and subassemblies of the infusion system that are repairable within the scope of this manual. In addition, this section details replacement procedures for all listed parts.
7.1

REPLACEABLE PARTS

Replaceable parts for the infusion system are itemized in the spare parts price list and are identified in Figure 9-1. Table 9-2 identifies each part by an index number that correlates to Figure 9-1.
To request a copy of the current spare parts price list, contact Hospira Technical Support
Operations (see Section 6.1), or to view the catalog online, visit the website at:
www.hospiraparts.com
For convenient reference, insert a copy of the spare parts price list here.
Technical Service Manual 7 - 1 430-95424-002 (Rev. 01/06)
Page 94
SECTION 7 REPLACEABLE PARTS AND REPAIRS
This page intentionally left blank.
430-95424-002 (Rev. 01/06) 7 - 2 Plum A+
®
3 Infusion System
Page 95
7.2 REPLACEMENT PROCEDURES
7.2

REPLACEMENT PROCEDURES

This section contains safety and equipment precautions, required tools and materials, and step-by-step procedures for replacing parts in the infuser. Unless otherwise stated, always perform the PVT after a replacement procedure.
7.2.1

SAFETY AND EQUIPMENT PRECAUTIONS

Before opening the front enclosure of the infuser, take all necessary precautions for working on high-voltage equipment.
WARNING: POSSIBLE EXPLOSION HAZARD EXISTS IF THE INFUSER
WARNING: UNLESS OTHERWISE INDICATED, DISCONNECT THE INFUSER
CAUTION: Use proper ESD grounding techniques when handling components. Wear an antistatic wrist strap and use an ESD-protected workstation. Store PWAs in antistatic bags before placing them on any surface.
IS SERVICED IN THE PRESENCE OF FLAMMABLE ANESTHETICS.
FROM AC POWER BEFORE PERFORMING ADJUSTMENT OR REPLACEMENT PROCEDURES.
7.2.2

REQUIRED TOOLS AND MATERIALS

The following tools and materials, or equivalents, are required for the replacement procedures in this section. In addition, the beginning of each procedure lists tools and materials required for that specific procedure.
- Set of flat blade screwdrivers
- Set of Phillips
- Set of standard and metric nutdrivers
®
screwdrivers
- Wide head pliers
- Diagonal cutters
-X-acto
®
knife
- Metric 10 mm wrench
- Custom nutdriver (P/N 519-95056-001)
- Long needle nose pliers
Technical Service Manual 7 - 3 430-95424-002 (Rev. 01/06)
- Mild solvent
- Lint-free cloth
Page 96
SECTION 7 REPLACEABLE PARTS AND REPAIRS
7.2.3

RUBBER FOOT PAD REPLACEMENT

The recommended tool for this procedure is a #2 Phillips screwdriver.
The replacement parts for this procedure are:
Pad, Rubber Foot Screw, 6-32 x 1/2, Pan Head, Phillips
To replace a rubber foot pad, see Figure 7-1, then proceed as follows:
1. Turn off the infuser, and disconnect the device from AC power.
2. Place the infuser face down on a soft flat surface.
3. Using the Phillips screwdriver, remove the screw that secures the rubber foot pad.
4. Install the replacement rubber foot pad in the exact reverse order of removal.
Replacement of a rubber foot pad is routine maintenance and no verification procedure is normally required. However, if the infuser may have been damaged during rubber foot pad replacement, perform the PVT in Section 5.2.
Figure 7-1. Bottom View
RUBBER FOOT PAD (4)
BATTERY DOOR
PAN HEAD SCREW (4)
05K06006
(3)
6-32 x 1/2
430-95424-002 (Rev. 01/06) 7 - 4 Plum A+
®
3 Infusion System
Page 97
7.2 REPLACEMENT PROCEDURES
7.2.4

BATTERY, BATTERY DOOR, AND DOOR PAD REPLACEMENT

Recommended tools for this procedure are a medium size flat blade screwdriver, mild solvent, and a lint-free cloth.
The replacement parts for this procedure are:
Assembly, Battery, with Wire Harness Door, Battery Pad, Door Ring, Retaining Screw, 6-32 x 1/2, Hex Head, Slotted, with Washer
To replace a battery, battery door, and door pad, see Figure 7-2, then proceed as follows:
1. Turn off the infuser, and disconnect the device from AC power.
2. Place the infuser face down on a soft flat surface.
3. Using the flat blade screwdriver, remove the screw and retaining ring that attach the battery door to the infuser, and remove the door.
4. Inspect the battery door and door pad for damage. Replace the door, if required.
5. If the battery door pad is defective, remove it and clean the door with mild solvent. Dry the battery door thoroughly, and install the replacement pad on the door.
6. Disconnect the battery harness from the charger circuit cable. Carefully pull the battery harness wires and connector outside the enclosure, and remove the battery.
7. Connect the replacement battery harness to the charger circuit cable, and insert the replacement battery into the enclosure.
Note: The cable connectors are keyed so that cables cannot be connected
incorrectly.
Note: Confirm the battery harness is not pinched between the battery and the
enclosure.
8. Replace the battery door using the screw and retaining ring that were removed in step 3.
9. Press [ON/OFF] with the infuser disconnected from AC power, and verify the front panel battery symbol illuminates.
10. Access the BIOMED SETTINGS screen and press [CHANGE BATTERY].
Note: The [CHANGE BATTERY] softkey will not appear on earlier versions
of the Plum A+
Replacement of the battery door and door pad is routine maintenance and no verification procedure is normally required. However, if the infuser may have been damaged during these procedures, perform the PVT in Section 5.2.
®
3.
Technical Service Manual 7 - 5 430-95424-002 (Rev. 01/06)
Page 98
SECTION 7 REPLACEABLE PARTS AND REPAIRS
AC POWER CORD
BATTERY ASSEMBLY (3)
DOOR PAD (3)
RETAINING RING (3)
BATTERY DOOR (3)
VELCRO STRAP
EQUIPOTENTIAL
TERMINAL
POWER CORD
6-32 x 1/2 HEX HEAD SCREW (3)
Figure 7-2.
AC Power Cord, Retainer, Velcro Strap, and Battery Assembly
4-40 x 3/8
FLAT HEAD
SCREW
RETAINER
05K06001
430-95424-002 (Rev. 01/06) 7 - 6 Plum A+
®
3 Infusion System
Page 99
7.2 REPLACEMENT PROCEDURES
7.2.5

AC POWER CORD, RETAINER, AND VELCRO STRAP REPLACEMENT

The recommended tool for this procedure is a #2 Phillips screwdriver.
The replacement parts for this procedure are:
Cordset, AC Power, Hospital Grade, Detachable Retainer, AC Power Cord Strap, Velcro, AC Power Cord Screw, 4-40 x 3/8, Pan Head, Phillips
To replace the AC power cord, retainer, or Velcro strap, see Figure 7-2, then proceed as follows:
1. Turn off the infuser, and disconnect the device from AC power.
2. Remove the batteries as described in Section 7.2.4.
3. Using the Phillips screwdriver, remove the screw from the AC power cord retainer. Turn the power cord retainer approximately 1/8 turn counterclockwise.
4. Unplug the power cord, and slide the plug through the retainer.
Note: Remove the AC power cord from its receptacle by grasping the plug.
Do not pull the cord.
5. Remove the Velcro strap from the power cord. Inspect the Velcro strap for wear and replace the strap, if required. Attach the strap to the replacement power cord.
6. Install the replacement AC power cord in the exact reverse order of removal.
7. Reinstall the batteries and connect the infuser to AC power.
8. Press [ON/OFF] and verify the infuser powers on.
Replacement of the AC power cord, retainer, and Velcro strap is routine maintenance and no verification procedure is normally required. However, if the infuser may have been damaged during these procedures, perform the PVT in Section 5.2.
Technical Service Manual 7 - 7 430-95424-002 (Rev. 01/06)
Page 100
SECTION 7 REPLACEABLE PARTS AND REPAIRS
7.2.6

SEPARATING THE FRONT ENCLOSURE, REAR ENCLOSURE, AND MAIN CHASSIS ASSEMBLY

The recommended tool for this procedure is a #2 Phillips screwdriver.
Note: The front enclosure consists of an upper assembly and a lower assembly.
The main chassis assembly consists of an upper chassis and a lower chassis.
The replacement parts for this procedure are:
Enclosure, Upper Front Enclosure, Lower Front Enclosure, Rear Chassis, Upper Chassis, Lower Screw, 6-32 x 1/2, Pan Head, Phillips Screw, 6-32 x 1 1/4, Pan Head, Phillips Screw, 6-32 x 2 3/4, Pan Head, Phillips Screw, 8-32 x 3 1/2, Pan Head, Phillips Washer, Flat, #6 Washer, Flat #8
CAUTION: Use proper ESD grounding techniques when handling components. Wear an antistatic wrist strap and use an ESD-protected workstation. Store PWAs in antistatic bags before placing them on any surface.
To separate the front enclosure, rear enclosure, and main chassis assembly, see Figure 7-3, then proceed as follows:
1. Turn off the infuser, and disconnect the device from AC power.
2. Remove the battery doors and batteries as described in Section 7.2.4.
3. Remove the AC power cord and retainer as described in Section 7.2.5.
4. Using the Phillips screwdriver, remove the screws from the rear enclosure.
5. Remove the rear enclosure by lifting it up and to the side.
6. Disconnect the three internal power connectors.
7. Using the Phillips screwdriver, remove the screws from the bottom corners of the center mechanism.
8. Set the infuser upright and remove the front upper enclosure by pulling it away from the upper chassis.
9. Remove the lower front enclosure by tilting the infuser back approximately 10 and pull the lower front enclosure away from the lower chassis.
10. Reassemble the front enclosure, rear enclosure, and main chassis assembly in the exact reverse order of disassembly. Follow the screw placement sequence as illustrated in Figure 7-4.
° ,
Note: When reassembling the upper front enclosure, lift all three door
handles first.
To verify successful assembly, perform the PVT in Section 5.2.
430-95424-002 (Rev. 01/06) 7 - 8 Plum A+
®
3 Infusion System
Loading...