This document and the subject matter disclosed herein are proprietary
information. Hospira retains all the exclusive rights of dissemination,
reproduction, manufacture and sale. Any party using this document
accepts it in confidence, and agrees not to duplicate it in whole or in
part nor disclose it to others without the written consent of Hospira.
430-95215-003 (Rev. 10/05)
Page 7
Plum A+ Infusion Systems1- 1
1)Descriptive Information
NOTE: In this manual, references to Plum A+ apply to
both systems unless otherwise noted.
Plum A+
®
Lists # 12391-04
and 11971-04
Plum A+®3
Lists # 12618-04
and 12348-04
CAUTION: THIS
ISTOBEUSED
DEVICE
AN IV POLEWITH
WITH
6-WHEELBASEAND
A
SHELF.
A
430-95215-003 (Rev. 10/05)
Page 8
1- 21) Descriptive Information
The Plum A+ and Plum A+3 Volumetric Infusion Systems are
designed to meet the fluid delivery requirements of today’s
evolving healthcare environments. Both are cassette based multifunction infusion systems. The Plum A+ allows two lines in and
one line out while the Plum A+3 allows six lines in and three lines
out. Each pump can be used for standard, piggyback, or
concurrent delivery. Therapy modes include:
•Standard Infusions
•Multistep Programming
•Loading Dose
•Dose Calculation
The Plum A+ and Plum A+3 are designed to deliver parenteral,
enteral, or epidural infusions over a broad range of infusion rates
from multiple fluid container types.
Both are designed to be used in most areas of patient care,
including, but not limited to:
•General Floor•Labor/Delivery/
Post Partum
•Medical/Surgical•OR/Anesthesia•Hemodialysis
•ICU/CCU•Post Op/
Recovery
•Pediatrics•Cardiac Cath
Lab
•Neonatolgy•Emergency•Nutritional
•Burn Unit
•Oncology
•Mobile Intensive
Care
1.1 Product Description
Each system includes a pumping module (hereafter called the
infuser) and an assortment of disposable IV sets (hereafter called
a set), optional accessories, and this operator’s manual.
430-95215-003 (Rev. 10/05)
Page 9
Plum A+ Infusion Systems1- 3
A computer dataport is also provided to interface with a host
computer.
Accessory equipment connected to the analog and digital
interfaces must be certified according to the respective IEC
Standards (e.g. IEC 60950 for data processing equipment and
IEC 60601-1 for Medical Equipment). Furthermore, all
configurations shall comply with the system standard IEC 606011-1. Any person who connects additional equipment to the signal
input or output part configures a medical system, and is therefore
responsible for ensuring that the system complies with the
requirements of the system Standard IEC 60601-1-1. If in doubt,
consult the technical service department or your local
representative.
NOTE: Do not connect dataport while infusing.
1.2 Indications for Use
USER QUALIFICATION
The Plum A+ is intended for use at the direction or under the
supervision of licensed physicians or certified healthcare
professionals who are trained in the use of the infuser and the
administration of parenteral, enteral, and epidural fluids and
drugs and whole blood or red blood cell components. The training
should emphasize preventing related IV complications, including
appropriate precautions to prevent accidental infusion of air. The
epidural route can be used to provide anesthesia or analgesia.
430-95215-003 (Rev. 10/05)
Page 10
1- 41) Descriptive Information
WARNING
ADMINISTER ONLY ANESTHETICS/ANALGESICS
APPROVED FOR EPIDURAL ADMINISTRATION (AS
INDICATED OR ALLOWED BY THE DRUGS’ FDA
APPROVED LABELING). EPIDURAL ADMINISTRATION
OF DRUGS OTHER THAN THOSE INDICATED FOR
EPIDURAL USE COULD RESULT IN SERIOUS INJURY TO
THE PATIENT.
1.3 Conventions
This section describes the conventions used throughout this
manual, as follows:
CONVENTIONAPPLICATIONEXAMPLE
ItalicReference to a
section, figure, or
table
Function or mode
specific instructions
[BRACKETED
ALL CAPS]
[Italic]Softkey Options
Initial Caps
lowercase
Keys or buttons on
the device are
displayed in
[BRACKETED ALL
CAPS] or with a
graphic.
Screen displays and
device labels (as
appropriate)
(See Figure 3-1,
Priming
Cassette)
Primary Only:
Attach an empty
container.
[START]
or
[Choose]
Therapy
Dose Calculation
430-95215-003 (Rev. 10/05)
Page 11
Plum A+ Infusion Systems1- 5
CONVENTIONAPPLICATIONEXAMPLE
BoldEmphasis...sets are
supplied Sterile
and are for....
WARNINGS, CAUTIONS, AND NOTES
Alert messages used throughout this manual are described
below. Pay particular attention to these messages.
WARNING
A WARNING MESSAGE CONTAINS SPECIAL SAFETY
EMPHASIS AND MUST BE OBSERVED AT ALL TIMES.
FAILURE TO OBSERVE A WARNING MESSAGE IS
POTENTIALLY LIFE THREATENING.
CAUTION: A CAUTION
PROCEDUREORSTATEMENT. ITCONTAINSINFORMATION
A
COULDPREVENTIRREVERSIBLEPRODUCTDAMAGEOR
THAT
HARDWARE
COULDRESULTINSERIOUSPATIENTORUSERINJURY.
FAILURE. FAILURETOOBSERVEA CAUTION
USUALLYAPPEARSINFRONTOF
NOTE: A Note highlights information that helps explain a
concept or procedure.
This symbol directs the user to consult accompanying
documents.
NOTE: Figures are rendered as graphic representations
to approximate the actual product. Therefore, figures
may not exactly reflect the product.
430-95215-003 (Rev. 10/05)
Page 12
1- 61) Descriptive Information
1.4 Precautions
The Plum A+ has been designed and manufactured to be safe,
reliable, and easy to use. This section details precautions and
possible hazards.
For safe operation of the Plum A+, observe the following
precautions and hazards.
ARTIFACTS
•Nonhazardous, low-level electrical potentials are
commonly observed when fluids are administered using
infusion devices. These potentials are well within
accepted safety standards, but may create artifacts on
voltage-sensing equipment such as ECG, EMG, and
EEG machines. These artifacts vary at a rate that is
associated with the infusion rate. If the monitoring
machine is not operating correctly or has loose or
defective connections to its sensing electrodes, these
artifacts may be accentuated so as to simulate actual
physiological signals. To determine if the abnormality in
the monitoring equipment is caused by the infusion
device instead of some other source in the environment,
set the infusion device so that it is temporarily not
delivering fluid. Disappearance of the abnormality
indicates that it was probably caused by the electronic
noise generated by the infusion device. Proper setup and
maintenance of the monitoring equipment should
eliminate the artifact. Refer to the appropriate monitoring
equipment system documentation for setup and
maintenance instructions.
•The Plum A+ system is designed to operate normally in
the presence of most encountered electromagnetic
interference (EMI) conditions. In the event of extreme
levels of interference, such as encountered next to an
electrosurgical generator, it is possible that the normal
operation of a sensor or microcomputer might be
430-95215-003 (Rev. 10/05)
Page 13
Plum A+ Infusion Systems1- 7
disrupted. Even in this event, the outcome would likely be
a false alarm or detected system malfunction and would
not result in a hazard to patient or operator.
•Use of radio frequency emitting devices such as cellular
telephones and 2-way radios in close proximity of this
device may affect its operation.
HEALTHCARE PROFESSIONALS AND PATIENT
RELATED
•Setting the primary rate greater than the secondary rate
will result in a more rapid infusion of any residual
secondary drug remaining in the line and the cassette.
•Consult drug labeling to confirm drug compatibility,
concentration, delivery rates, and volumes are all
suitable for secondary, concurrent and piggyback
delivery modes.
•Arrange tubing, cords, and cables to minimize the risk of
patient strangulation or entanglement.
•Before opening the door, close clamp on the primary line
or remove the secondary container from the secondary
port to prevent mixing of primary and secondary fluids.
•Although unlikely, failure of certain robust mechanical
components such as the anti-free flow mechanism or
valve control springs could cause fluid delivery limited to
the contents of the fluid container. Single fault failure of
certain electronic/motor control components would result
in no more than 5 mL of unexpected fluid delivery.
•A small amount of fluid is expelled from the set (less than
0.05 ml) each time the door is opened or closed with a
set installed. If potent drugs are being used, take
appropriate action to guard against overmedication of the
patient.
•Before disconnecting a syringe from the cassette, pull up
the plunger slightly to avoid spilling the fluid. For rigid
containers, close the upper slide clamp, open the
cassette door, then remove and invert the cassette (ports
down).
430-95215-003 (Rev. 10/05)
Page 14
1- 81) Descriptive Information
•Air bubbles may form distal to the cassette as result of
normal out-gassing of dissolved air in the fluid. This may
occur if chilled solution is in use, if the infuser is mounted
significantly above the patient, or when using certain
fluids known to routinely outgas. In these cases, an air
eliminating filter may be used.
•Repeated opening and closing of the door may defeat the
proximal air-in-line alarm and may cause a distal air-inline alarm, requiring repriming.
•In vitro studies have suggested that packed red blood
cells with unusually high hematocrit be diluted with bloodcompatible fluids, such as 0.9% sodium chloride
injection, to decrease hemolysis and increase flow rate.
•The screen displays the VTBI (volume to be infused) in
integers when value is above 99.9. Any fraction of a
milliliter delivered is not displayed, but is retained in
memory.
•For Plum A+3 users, be aware that changing the weight
on one device does NOT change the weight on the other
two devices. Patient weight must be changed on each
device when delivering weight-based therapy dependent
on medication requirements.
CONCURRENT FLOW
GUIDELINES
When delivering short half-life critical drugs (see Critical Drugs,
this section) using the Plum A+ in the Concurrent mode, the
following delivery rate guidelines should be observed:
•If the critical drug (with half-life less than 6 minutes)
is to be infused at less than 2 mL/hr, the other
infusion should be no faster than 5 times the critical
drug’s rate. Dopamine, for example, delivered at 1.5
mL/hr should not be accompanied by an infusion
programmed any faster than 7.5 mL/hr.
•If the critical drug (with half-life less than 6 minutes)
is to be infused at 2 to 5 mL/hr the other infusion
430-95215-003 (Rev. 10/05)
Page 15
Plum A+ Infusion Systems1- 9
should be no faster than ten times the critical drug’s
rate. Dopamine, for example, delivered at 3.5 ml/hr
should not be accompanied by an infusion
programmed any faster than 35 mL/hr.
•If the critical drug (with half-life less than 6 minutes)
is to be infused at 5.1 mL/hr or greater, the other
infusion can be programmed at any desired rate.
NOTE: The total of the primary rate plus the secondary
rate cannot exceed 500 mL/hr.
430-95215-003 (Rev. 10/05)
Page 16
1- 101) Descriptive Information
These guidelines apply only when infusing short half-life critical
drugs in Concurrent mode. Individual patient responses may vary
requiring adjustment of delivery rates.
DELIVERY RATE GUIDELINES
SHORT HALF-LIFE (LESS
THAN 6 MINUTES) CRITICAL
RUG INFUSION RATE
D
0.5 - 1.9 mL/hr5 Times the Critical Drug Rate
2 - 5 mL/hr10 Times the Critical Drug Rate
5.1 or GreaterAny Desired Ratio
CRITICAL DRUGS
Examples of drugs with a short half-life (approximately 6 minutes
or less when given IV) include:
DobutamineLidocaine
DopamineNitroglycerin
EpinephrineNitroprusside
EpoprostenolNorepinephrine
EsmololOxytocin
IsoproterenolProcainamide
MAXIMUM RATEOF
CCOMPANYING INFUSION
A
For these drugs, the concurrent flow guidelines should be
followed when the infusion rate of the drug will be 5 mL/hr or less.
NOTE: This list of critical drugs is not intended to be allinclusive of critical drugs or drugs with a short half-life.
The clinician should become familiar with the pharmacodynamics
of any critical drug before administration.
This information is presented to inform clinicians of a rare
situation that could be misinterpreted if they are unfamiliar with
this phenomenon.
430-95215-003 (Rev. 10/05)
Page 17
Plum A+ Infusion Systems1- 11
EPIDURAL ADMINISTRATION
•Recommended use of the epidural route is to provide
anesthesia or analgesia for periods up to 96 hours.
•This device can be used to administer only those
anesthetics/analgesics approved for epidural
administration (as indicated or allowed by the drugs’ FDA
approved labeling). Epidural administration of drugs
other than those indicated for epidural use could result in
serious injury to the patient.
•For epidural administration, the use of Hospira catheters,
Plum sets without Y-sites, and "epidural" stickers
indicating ongoing epidural administration are
recommended.
•Administration of drugs via the epidural route should be
limited to personnel familiar with associated techniques
and patient management problems. Proper epidural
placement of the catheter is essential since catheter
migration could result in intravascular or intrathecal
administration. Facilities practicing epidural
administration must be equipped with resuscitative
equipment, oxygen, naloxone, and other resuscitative
drugs. Adequate monitoring equipment (e.g., Oximetry)
is recommended for continuous monitoring of the patient
during epidural administration. Patients must be
observed frequently for side effects in a fully-equipped
and staffed environment for at least 24 hours following
completion of drug administration by the epidural route.
DELAYED RESPIRATORY DEPRESSION FOLLOWING
CONTINUOUS EPIDURAL ADMINISTRATION OF
PRESERVATIVE-FREE MORPHINE SULFATE HAS
BEEN REPORTED.
•The epidural space has 58 openings through which fluid
can exit. Pressure buildup during administration is
transient. However, if a large volume of fluid is
administered over a short time period, the pressure will
take longer to return to normal. If overdelivery occurs
during administration, observe the patient closely for
430-95215-003 (Rev. 10/05)
Page 18
1- 121) Descriptive Information
signs of spinal cord compression (disorientation,
headache, transient neuralgias) and drug overdose.
BATTERY OPERATION
•Do not operate the Plum A+ on patients when the battery
is removed. Use of a properly maintained and charged
battery helps confirm proper operation.
•The battery may not be fully charged upon receipt.
Connect the infuser to AC power for at least six hours.
•Use AC power whenever possible. Connect to AC power
during storage to ensure a fully charged battery for
emergencies. If quality earth grounding source is in
doubt, use battery power.
•If the low-battery alarm sounds, connect the infuser to AC
power immediately.
SETS AND ACCESSORIES
•Only compatible LifeCare Plum Sets can be used with
the Plum A+. See individual set instructions for additional
information.
•Administration sets should be changed per CDC
guidelines or healthcare provider policy. Discard after
use.
•LifeCare IV infusion sets with integral nonblood filters are
not for use in the administration of blood, blood products,
emulsions, suspensions, or any medications not totally
soluble in the solution being administered. These
medications may be administered through the lower Yinjection site, below the filter.
•When infusing at low delivery rates (5 mL/hr or less) the
use of thick-walled microbore Plum sets is
recommended. This will reduce the amount of the fluid
bolus that may be delivered when a distal line occlusion
is released.
•Syringes must be larger than 3 cc. Use syringe adapter
(List 11986-48) when using syringes smaller than 10cc.
Some 10cc syringes may require use of a syringe
430-95215-003 (Rev. 10/05)
Page 19
Plum A+ Infusion Systems1- 13
adapter. Syringes larger than 10cc may be attached
directly to the secondary port of the cassette. Use of a
syringe adapter may decrease the occurrence of
proximal occlusion alarms.
•Use a 19-gauge or larger needle or catheter at the
venipuncture site for viscous fluids if operating at rates
greater than 500 ml/hr.
See Section 10for information on sets and accessories.
BACKPRIMING
•Backpriming is not recommended for reconstituting
secondary containers containing dry powders.
•To avoid pressurization when backpriming into a
syringe, the user must confirm there is sufficient
empty space to accept the backprimed fluid.
GENERAL
•Possible explosion hazard exists if used in the
presence of flammable anesthetics.
•Do not place Plum A+ in service if it fails the self-test.
•Do not operate the Plum A+ with the case opened.
•Keep the cassette door securely closed while the
infuser is not in use, to avoid cassette door damage.
•The Plum A+3 is to be used with an IV pole with a 6wheel base and a shelf.
CLEANING
For more information on cleaning the infuser, see Section 8.
•To avoid mechanical or electronic damage, do not
immerse the Plum A+ in any fluids or cleaning solutions.
•Do not spray cleaning solutions toward any opening in
the instrument.
•Certain cleaning and sanitizing solutions may slowly
degrade components made from some plastic materials.
430-95215-003 (Rev. 10/05)
Page 20
1- 141) Descriptive Information
Using abrasive cleaners or cleaning solutions not
recommended by Hospira may result in product damage.
Do not use compounds containing combinations of
isopropyl alcohol and dimethyl benzyl ammonium
chloride.
•Never use sharp objects such as fingernails, paper clips,
or needles to clean any part of the infuser.
•Do not sterilize by heat, steam, ethylene oxide (ETO), or
radiation.
•To avoid infuser damage, cleaning solutions should only
be used as directed. The disinfecting properties of
cleaning solutions vary; consult the manufacturer for
specific information.
Use the following procedure to avoid the administration of a bolus
following a distal occlusion (i.e., a closed distal clamp):
•If a secondary container is in use, clamp proximal
tubing before opening cassette door.
•Open cassette door and remove the cassette.
•Open the flow regulator briefly to dissipate the
pressure and then close it.
•Eliminate the source of occlusion (closed clamp).
•Reinsert the cassette and close the cassette door.
•Open all clamps and resume infusion.
NOTE: When troubleshooting an occlusion where all
clamps are in the OPEN position, use care to avoid
delivery of a bolus by opening the flow regulator to
release any built-up pressure. Close the clamp between
the cassette and the patient before opening the flow
regulator to relieve the pressure. See Section 7, Alarms
and Troubleshooting, for more information.
430-95215-003 (Rev. 10/05)
Page 21
Plum A+ Infusion Systems2- 1
2)Principles of Operation
2.1 Features
The Plum A+ is a dual-line volumetric infusion system designed
to meet the growing demand for hospital-wide, as well as
alternate site and home healthcare, standardization. With it’s
primary line, secondary line, and piggyback fluid delivery
capability, the Plum A+ is suited for a wide range of medical/
surgical and critical care applications. Full compatibility with
LifeCare Plum Series administration sets and accessories and
the LifeShield® and CLAVE® needleless protection systems,
makes the Plum A+ a convenient and cost-effective infusion
system.
PROGRAMS
•Dose Calculation
•Loading Dose
•Multistep Programming
LINE PROGRAMMING OPTIONS
•Drug List
•Nurse Call Back
•Delayed Start Setting
•Concurrent delivery
•Titration
•Micro 0.1-99.9 mL/hr (in 0.1 mL increments) flow
rate range for both lines
•Macro 100-999 mL/hr (in 1 mL increments) flow
rate range for both lines
•Automated Secondary drug delivery (Piggyback)
•Standby Setting
430-95215-003 (Rev. 10/05)
Page 22
2- 22) Principles of Operation
PLUMSET CAPABILITIES
•Anti Free-Flow Protection
•Direct Connection for syringe delivery
AIR MANAGEMENT
•Air Trap
•Air Removal/Backpriming
•Air Detection-Proximal
•Air Detection-Distal
BATTERY
NOTE: The Plum A+3 utilizes three batteries.
•Battery Gauge
•Six Hour Battery Recharge Time
•Long battery life (6 hours) for emergency backup
and temporary portable operation
BIOMEDICAL
•Serial Communication
•Upgradability (Field- for List 12391 only)
•Variable Rate Cap
•Alarm History
•Nurse Call Relay Connector
OPTIONS
•Volumes Infused (A, B, Total Volume)
•KVO at dose end (1 ml/hr or less depending on
delivery rate) or Continue Rate (CR) to continue
at the current rate
•Variable Distal Pressure Settings
430-95215-003 (Rev. 10/05)
Page 23
Plum A+ Infusion Systems2- 3
OTHER FEATURES
•Nonpulsatile volumetric accuracy
•Microprocessor control
•Large liquid crystal display (LCD) screen
•Panel back illumination on mains power
•Lockout switch
•Standard syringe use
•Parenteral and nonparenteral (enteral) fluid
delivery
•Blood and blood products delivery
•Wide range of Standard and Specialty
administration sets
•Password protected keypad lock
2.2 Definitions
TERMDEFINITION
Administration
Set
AlarmA condition requiring operator
AlertWhile entering a program, a visual
The sterile, disposable assembly with
flexible tubing that connects to a
source fluid container for input to the
Cassette pumping chamber and to
some output device for administration
to the patient.
intervention that invokes both audible
and visible alarm indicators.
Confirmation of the condition usually
is required.
warning that informs the clinician of
the violation of a hospital-defined rule
set, e.g., a dose limit for a specific
drug.
430-95215-003 (Rev. 10/05)
Page 24
2- 42) Principles of Operation
TERMDEFINITION
Alternate UnitsDose Rate units that may be selected.
Alternate Units are any units other than
mL/hr.
Alternate Units
Parameters
BackpressureResistance to fluid flow on the Distal or
Drug Amount, Diluent, Patient Weight
(if applicable), and Dose Rate.
output portion of the Administration
Set, usually expressed as mmHg.
proximal tubing and Cassette air trap
by rapid pumping of fluid from Line A
to an output receptacle on Line B. No
fluid delivered distal to the cassette
during a Backprime.
VTBI, and duration only. Drug name
may or may not be specified.
service mode of Infuser operation.
of fluid delivered over a discrete period
of time.
Cassette that contains the actual
pumping chamber with inlet and outlet
valves. It also contains access ports to
diaphragms for proximal and distal line
pressure sensing and chambers for air
trapping and air quantity
measurement.
prevent hardware failure, irreversible
damage to equipment, or loss of data.
430-95215-003 (Rev. 10/05)
Page 25
Plum A+ Infusion Systems2- 5
TERMDEFINITION
Cleared
Program
Cleared
Settings
ConcentrationFor Dose Calculation based Line A or B
Concurrent
Delivery
Concurrent
Mode
Continue Rate
(CR)
Default Drug
Library (DDL)
Deliver
Together
Same as Cleared Settings, but
restricted to the program delivery
settings for an individual line.
Programmed delivery settings for both
lines and the general control
parameters are at their default values.
delivery programming of premixed
drugs, the concentration consists of
Drug Amount (in micrograms [mcg],
milligrams [mg], grams,
milliequivalents [mEq], millimols
[mmols], or units [USP Units]) and
Diluent (Volume in milliliters [mL]).
Abbreviated on screen displays as
“Conc”.
Simultaneous delivery of two fluids at
independent flow rates.
The Line B control selection mode that
allows either Line A or Line B (or both)
of the Mechanism to be programmed
for delivery that completes with KVO or
CR. Line A operation is unaffected by
Line B delivery.
An option used to provide fluid
delivery at the current fluid delivery
rate when the VTBI becomes zero.
A drug list of 99 drug names, default
units and default concentrations, plus
‘No Drug Selected’.
A Biomed Delivery option setting,
which selects default of Piggyback or
Concurrent delivery mode.
430-95215-003 (Rev. 10/05)
Page 26
2- 62) Principles of Operation
TERMDEFINITION
Delivery ModeThe normal operational mode for the
Infuser, either Concurrent or
Piggyback, during which fluid delivery
may occur.
Delivery RateThe delivery rate (as Volume/Time)
assigned by the operator or calculated
from other delivery parameters.
DeviceThe Plum A+ IV Infusion Device
(Infuser), not including the disposable
administration sets.
Diluent
(Volume)
DistalDownstream (output) from the
DoorThe lever-operated, bottom-hinged,
Dose (Rate)Delivery rate times drug concentration.
Dose
Calculation
Drug AmountThe mass or quantity of the drug
DurationThe time period used to deliver a VTBI
EnteralDelivery via an intestinal route.
The volume of fluid used with the
Diluent Unit (e.g., 250 mL), which
dilutes the drug to be infused.
Cassette portion of the Administration
Set.
Cassette holder on the front of the
Mechanism.
Allows programming Dose Rates in
alternative units of measure. Dose
Calculation can be used in Simple
Delivery, Loading Dose, and Multistep.
premixed with a Diluent to express the
Concentration for the Line being
programmed.
at an existing Delivery Rate or as set by
the operator.
430-95215-003 (Rev. 10/05)
Page 27
Plum A+ Infusion Systems2- 7
TERMDEFINITION
ErrorA condition resulting in an audible
warning and often an accompanying
message, such as an invalid key press,
resulting from an operator attempt to
make an input that is not valid for the
current operational context. Not a
Malfunction.
Filling Head
Height (FHH)
Flow RateThe resulting rate of fluid flow. (See
Fluid Delivery
Begin
InfuserA Plum A+ device, not including the
Initialized
Position
KeyAny of the marked locations on the
KVO(Keep Vein Open) Internally selected
The gravity induced Proximal Line
pressure due to fluid height in source
container above the Distal Line output
level.
Rate and Dose.)
The time required, measured from the
initial start of a keypad press, for a
digit in the third decimal place of a
scale measuring grams to change
when pumping water, the delivery tube
is fluid filled, and the end of the
delivery tube is immersed.
disposable administration set.
The plunger is in the Plunger Park
Position, and each valve is in the Valve
Home Position.
front panels intended for user input via
a pressing action.
minimal delivery rate, the lesser of 1.0
mL/Hr or actual Rate, intended to
provide sufficient fluid flow to
decrease the potential for clotting at
the IV infusion site.
430-95215-003 (Rev. 10/05)
Page 28
2- 82) Principles of Operation
TERMDEFINITION
Line AInlet tubing on the top center of
Cassette when viewed after it has been
loaded in Mechanism Door.
Line BInlet connector on the top right of
Cassette when viewed after it has been
loaded in Mechanism Door.
Loading DoseAllows programming of an initial
infusion rate for a specific volume,
followed automatically by a
maintenance rate from the same
container (e.g., a fluid challenge). If
Dose Calc is used, the Loading Dose
and Maintenance Dose are in the same
unit of measure, over the same period
of time (mcg/min), from the same
container.
Macro CassetteThe Plum Cassette as part of an
Administration Set whose attainable
delivery performance is for general
purpose uses.
MalfunctionOne of a number of conditions that
indicate a failure of the Infuser. These
conditions usually require a repair
action.
MechanismThe provider of the physical activities
of: 1) valve control for Line selection
and inlet to or outlet from the Cassette;
2) Plunger extension and retraction for
changing fluid volume of the Cassette
pumping chamber.
430-95215-003 (Rev. 10/05)
Page 29
Plum A+ Infusion Systems2- 9
TERMDEFINITION
MultistepAllows a sequential program to deliver
up to 10 steps; fluid volumes and
delivery rates may be programmed for
each step. The program may be
entered based on Rate and Volume or
Volume and Time. If Dose Calc is used,
the delivery steps are in the same unit
of measure, over the same period of
time, from the same container.
NoteA Note explains something new, gives
tips on performing an activity, or
provides a reminder of necessary
preparations.
ParenteralDelivery via other than an intestinal
route, such as intravenous (IV)
injection.
Patient WeightThe Alternate Units Parameter whose
value entry is required with Dose rates
(quantity/time) having the third
parameter of weight, such as mcg/kg/
min.
Piggyback
Delivery
Piggyback
Mode
Plunger Park
Position
Infusion will stop Line A and infuse
Line B until VTBI completes.
The Line B control selection mode that
requires Line A delivery suspension
while Line B delivers and then its
resumption when Line B VTBI
completes. If Line A is not programmed
and started, Line B delivery just stops
when its VTBI completes. Line A
operation is delayed for duration of its
interruption by Line B.
The position of the pumping plunger
that allows the cassette to be
withdrawn from the Infuser without
restriction, with an opened Door.
430-95215-003 (Rev. 10/05)
Page 30
2- 102) Principles of Operation
TERMDEFINITION
POSTPower On Self Tests. Startup
initialization and test of critical
hardware modules and software
functions, to determine Infuser
readiness for its Intended Use.
Power OnWhen the power is first applied to the
Infuser processor. It occurs when
Infuser is in the off condition and the
AC power is applied or during battery
operation when the ON/OFF key is
pressed.
PressureThe measured fluid pressure for
Proximal Line A, Proximal Line B, or
the Distal line when a fully primed
administration set is loaded into a
Mechanism with the Cassette Door
closed.
ProgramsA Concentration-based extension to
basic Volume/Rate delivery for Dose
Rate applications. Softkey-selected
types are: 1) Standard Program for
single rate delivery; 2) Loading Dose
for up to two steps of delivery at
differing rates; 3) Multistep for up to
ten steps of delivery at differing rates.
Loading Dose and Multistep include
the basic Delivery Rate option.
ProtocolIn the medical field, hospital-approved
procedures. For devices with an
electronic communications interface,
the format and content of messages
being exchanged.
ProximalUpstream (input, as Line A and/or B)
with respect to the Cassette portion of
the Administration Set.
RateSee Delivery Rate
430-95215-003 (Rev. 10/05)
Page 31
Plum A+ Infusion Systems2- 11
TERMDEFINITION
RoundingTo round-off to the closest whole
number (e.g., 25.5 would round to 26,
but 25.4 would round to 25).
ScreenThat active portion of the LCD
assigned to a delivery or other user
interactive display operation.
SoftkeyA front panel key whose function label
is on the bottom portion of the display
screen just above its location, so that
label may be changed to indicate
differing key functions within the
operational context of the defined
screens.
Standard
Programming
StandbyA delivery program state that
SystemThe Plum A+ Infuser system with its
TruncateTo round-down to the last whole
UnitsA Drug Amount (without a quantity of
VIVolume Infused.
(see Basic Delivery)
resembles a delayed start for a line in
all respects, but status display of
STANDBY instead of DELAYED. A
press of the START key (and a Start
Softkey if necessary) will begin the
programmed delivery.
associated accessories and
Administration Sets.
number (e.g., 103.7 would be 103).
mass referent) adopted as the United
States Pharmacopoeia (USP) standard
of measure for use in a Concentration.
430-95215-003 (Rev. 10/05)
Page 32
2- 122) Principles of Operation
TERMDEFINITION
VTBIVolume To Be Infused – The volume of
fluid or IV solution (remaining) for
delivery by a program or Therapy step
from a Line.
WarningAudible indication at the defined
Warning sound level, often
accompanied by a brief message, to
advise operator that: 1. Input was
attempt to use a function in wrong
sequence, with wrong values, or at
wrong time (resulting in an Invalid Key
Warning); 2. Input was properly related
to a function (resulting in a Valid Key
Warning). (The Low Battery Warning is
an exception, using the Alarm
Indication sound level.)
WeightThe Alternate Units Parameter whose
value entry is required with Alternate
Units for Dose that include /kg/.
430-95215-003 (Rev. 10/05)
Page 33
Plum A+ Infusion Systems3- 1
3)Equipment Description
Line A Flow
Indicator
Status
Region
Working
Region
Message
Region
Softkey Label
Softkey
Selector
Start and
Stop Keys
Charging/
Line
Indicator
ON/OFF
Key
Status Region
Working Region
Message Region
Back
Prime
AB
Options/
Vol Inf
Line B Flow
Indicator
LCD
Battery
Charge
Indicator
Numeric
Keys
(Up/Left)
Direction
Keys
(Down/
Right)
Silence
Key
Numeric Field
Clear Key
Basic Layout of Front Panel Display and
Decimal Key
Keypad
430-95215-003 (Rev. 10/05)
Page 34
3- 23) Equipment Description
3.1 Operating Keys
NOTE: For Plum A+3, all
three pumps act as
individual Plum A+’s.
The [ON/OFF] key is
used to control the
power of the Plum A+.
NOTE: All deliveries must
be stopped in order to turn
the power off.
NOTE: To confirm proper
shutdown of infuser, do
not unplug until “double beep” is heard
(approximately 5
seconds). If “double beep” is not heard, the infuser will
turn itself back ON sensing the AC power was lost and
revert to battery back-up power.
The [START] key is used to
begin infusion and as final
confirmation of
programming.
430-95215-003 (Rev. 10/05)
Page 35
Plum A+ Infusion Systems3- 3
The [STOP] key is used to
stop the current delivery
on one line; used in
conjunction with soft keys
to stop delivery on both
lines.
The [SILENCE] key is used
to silence an audible alarm
during actions to correct
its cause.
The [SELECT] keys
(or ) are used to
move the highlighting
cursor between the
programming fields.
430-95215-003 (Rev. 10/05)
Page 36
3- 43) Equipment Description
The numeric keys are used
to enter values for any
highlighted field requiring
numeric data.
The [CLEAR] key will
cause a highlighted
numeric field to be cleared
in preparation for data
entry.
The [Decimal] key is used
to manually enter numbers
other than whole numbers
(i.e., 1.2 mL).
.
Softkeys are located at the
bottom of the main display
and serve a variety of
functions. What each key
does is indicated by the
text in the display above
the actual softkey.
Drug
List
Standby Enter
Cancel/
Back
430-95215-003 (Rev. 10/05)
Page 37
Plum A+ Infusion Systems3- 5
3.2 Indicators
The Charge/Line Indicator
is illuminated to indicate
the battery is currently
charging.
NOTE: If the device is plugged
into AC power, with a battery
installed, and the Charge
Indicator is not illuminated,
contact technical support.
The fluid drop symbol,
when flashing, indicates
an active infusion on Line
A and/or Line B.
Battery Capacity Symbol
(located on right side of
message region on
display) indicates relative
charge with number of
white boxes from left to
right.
430-95215-003 (Rev. 10/05)
Page 38
3- 63) Equipment Description
3.3 Rear Case Controls
NOTE: For Plum A+3, each control applies to all three
pumps
The audio level rotary
knob, located on the back,
adjusts the alarm audio
level.
The toggle switch, located
below the audio level
control, activates the
lockout function when the
lever is placed in the “up”
position, disabling all front
panel keys except [STOP].
Password Protected
Keypad Lock may also be
used. It both locks and
unlocks the keypad by
pressing the Decimal key,
followed by 9, 6, & 3.
NOTE: For Plum A+3, keypad lock applies to individual
pump in use.
NOTE: Pressing of a key while lockout is active results in the
display of the lockout enabled message. This action is
recommended to confirm each lockout activation.
430-95215-003 (Rev. 10/05)
.
Page 39
Plum A+ Infusion Systems3- 7
NOTE: If the lockout switch is put in the “Locked” position
while the infuser is off, the [ON/OFF] key allows the unit to
turn on.
The connector located
below the lockout switch
is the Nurse Call Relay
Connector, for providing
remote notification of a
nurse call event.
The RS-232 DataPort is
used to interface with a
host computer.
NOTE: Do not connect dataport while infusing.
The Equipotential Post
(located on the back of
Plum A+3) is a terminal for
connection of an
equipotential conductor.
430-95215-003 (Rev. 10/05)
Page 40
3- 83) Equipment Description
3.4 Administration Sets
The Plum A+ is compatible with a wide range of LifeCare Plum
series administration sets. Become familiar with the set
components before preparing the administration set.
Primary Line
Flow Regulator
Pumping Chamber
Secondary
Inlet Port
(Capped)
Finger Grip
Patient Line
Air Trap
PREPARING THE ADMINISTRATION SET
NOTE: For detailed instructions, see administration set
packaging.
WARNING
ARRANGE TUBING, CORDS, AND CABLES TO
MINIMIZE THE RISK OF PATIENT STRANGULATION
OR ENTANGLEMENT.
Use the aseptic technique to prepare the administration set for
priming, then proceed as follows:
Make sure flow regulator is closed.
430-95215-003 (Rev. 10/05)
Page 41
Plum A+ Infusion Systems3- 9
Insert piercing pin into container outlet with
a twisting motion.
Fill Drip chamber to about 1/2 full or to the
score mark. Do not completely fill.
NOTE: To avoid getting Distal Air Alarms,
confirm the fluid bag is positioned higher
than the infuser. This should be set up
prior to priming the cassette.
PRIMING THE ADMINISTRATION SET
Invert the cassette.
Turn the flow regulator counter-clockwise
until a drop of fluid is seen in the pumping
chamber.
Turn the cassette upright, then prime the
remainder of the administration set.
430-95215-003 (Rev. 10/05)
Page 42
3- 103) Equipment Description
Push in the flow regulator to close. Confirm
there is no flow.
LOADING THE CASSETTE
To load the primed cassette into the plum A+, proceed as follows.
Open the cassette door by lifting up the
handle.
Insert cassette into door guides. Close the
cassette door. Confirm there is no flow and
no kinks appear in tubing.
NOTE: If flow is observed, close the tubing
clamps and replace the administration set.
NOTE: The administration set should be
changed per CDC guidelines or healthcare
provider policy. Discard after use.
PREPARING THE SECONDARY LINE
CAUTION: CONSULTTHEDRUGLABELINGTOCONFIRM
COMPATIBILITY, CONCENTRATION, DELIVERYRATES,
DRUG
ANDVOLUMESAREALLSUITABLEFORCONCURRENT
DELIVERY
AUTOMATICALLY
The Plum A+ features concurrent or piggyback delivery modes
when therapy requires administering more than one fluid.
ORPIGGYBACKDELIVERY (SECONDARYFOLLOWED
BYPRIMARY) MODES.
430-95215-003 (Rev. 10/05)
Page 43
Plum A+ Infusion Systems3- 11
In addition to standard containers, the Plum A+ can use syringes
on the secondary port. The secondary line can be prepared
without removing or repriming the cassette.
Before preparing the secondary line, observe the following
guidelines:
•Review the backpriming function.
•Attach the secondary container using an 18or 19- gauge, 31 mm (or shorter) needle, if the
cassette secondary inlet port has a piercing
reseal.
•Syringes must be larger than 3 cc. Use
syringe adapter (List 11986-48) when using
syringes smaller than 10cc. Some 10cc
syringes may require use of a syringe
adapter. Syringes larger than 10cc may be
attached directly to the secondary port of the
cassette. Use of a syringe adapter may
decrease the occurrence of proximal
occlusion alarms.
•When using a syringe adapter, retract the
plunger to draw approximately 1 mL of fluid
into the syringe to clear air from the adapter
filter.
To prepare the secondary line, use aseptic technique and
proceed as follows:
With cassette door closed: Loosen and
remove white cap, then discard. Add
secondary tubing or syringe.
Open all proximal clamps before pressing
[START].
430-95215-003 (Rev. 10/05)
Page 44
3- 123) Equipment Description
Set up per operating instructions. (See
Sections 4 - 6 of this manual.)
Syringe: Secure the container to the
cassette door using the optional container
support arm (List 12095-03).
3.5 Discontinuing Electronic Flow Control
& Setting Gravity Flow
CAUTION: BEFOREOPENINGDOOR, CLOSECLAMPON
PRIMARY
CONTAINER
PRIMARY
If electronic flow needs to be stopped, the precision flow regulator
on the cassette can be used to manually set and maintain fluid
flow in a gravity mode (see the following graphics).
ORSECONDARYSETORREMOVESECONDARY
FROMSECONDARYPORTTOPREVENTMIXINGOF
ANDSECONDARYFLUIDS.
1
PRESS STOP
STOP
3
SET GRAVITY RATE BY
TURNING FLOW REGULATOR
2
OPEN DOOR & REMOVE CASSETTE
NOTE: FLOW REGULATOR CLOSES
AUTOMATICALLY WHEN DOOR IS OPENED
TO PREVENT ACCIDENTAL FREE FLOW
NOTE: IF DOOR OPEN ALARM OCCURS, TURN PUMP OFF
COUNTER-CLOCKWISE
NOTE: CASSETTE SHOULD BE
IN UPRIGHT POSITION
430-95215-003 (Rev. 10/05)
Page 45
Plum A+ Infusion Systems3- 13
3.6 Discontinuing Fluid Administration
1
PRESS STOP
STOP
3
CLOSE DOOR
4
DISCARD SET AND FLUID
CONTAINER PER HOSPITAL
2
OPEN DOOR, REMOVE CASSETTE
PROCEDURE
430-95215-003 (Rev. 10/05)
Page 46
3- 143) Equipment Description
NOTES
430-95215-003 (Rev. 10/05)
Page 47
Plum A+ Infusion Systems4- 1
4)Basic Operation
4.1 Getting Started
UNPACKING
CAUTION: PRODUCTDAMAGEMAYOCCURUNLESS
PROPER
INSTALLATION
DAMAGED
UPON
Inspect the Plum A+ packaging for possible shipping damage. If
damage is found, contact the delivery company immediately.
Use care when unpacking the infuser. Retain the packing slip and
save all packing material in case the infuser is damaged or fails
the infuser self-test and has to be returned to Hospira.
CAREISEXERCISEDDURINGUNPACKINGAND
. DONOTUSETHE PLUM A+ IFITAPPEARS
INANYWAY. THEBATTERYMAYNOTBECHARGED
RECEIPT.
Inspect thoroughly for damage.
CAUTION: I
CONTACT HOSPIRA.
To set up, connect the AC (mains) cord to a properly grounded
receptacle, unless temporary battery operation is desired.
NOTE: Use AC (mains) power whenever possible. Store
the infuser connected to AC (mains) to confirm a fully
charged battery for emergencies.
NOTE: If the quality of the earth grounding source is in
doubt, use battery power.
NOTE: The Plum A+3 is NOT to be used with the tandem
carrier. It is only to be used with the Plum A+ and other
single-channel devices.
The Plum A+ may be safely and conveniently mounted on an IV
stand or on a tandem carrier mounted on an IV stand.
Always mount the tandem carrier(List#
12270-01) to the pole before attaching
the infuser(s) to the tandem carrier.
Make sure tandem carrier clamp is
attached firmly to pole.
Make sure infusers are firmly attached
to tandem carrier.
This section details the Plum A+ instrument setup procedures.
SYSTEM SELF-TESTS
The infuser performs a suite of System Self-Tests when power is
applied to confirm readiness for use.
Failure during the Self-Tests is reported as a Malfunction
Condition.
CAUTION: D
THE SYSTEM SELF-TESTS.
FAILS
ONOTPLACETHE PLUM A+ INSERVICEIFIT
430-95215-003 (Rev. 10/05)
Page 49
Plum A+ Infusion Systems4- 3
Connect the AC power (mains) cord to an AC power receptacle,
then confirm that the Charging/Line indicator illuminates (above
the [ON/OFF] key on the front panel).
A systematic self-testing of the processing, delivery, and safety
systems is performed whenever the Plum A+ is turned on, to
verify readiness for operation.
NOTE: If the quality of earth grounding source is in doubt,
use battery power.
Press the [ON/OFF] key to turn the power on. Check screen
display and listen for a beep to indicate the audio is working. Wait
for the self-tests to complete. If successful, put a fully primed
macro cassette into the cassette door and close the door. See
Loading Cassette in Section 3 for details of correct procedure.
When the Charging/Line indicator is off and the battery symbol on
the display is flashing, this indicates that the Plum A+ is operating
on low battery power and should be recharged.
CAUTION: D
THEBATTERYREMOVED. USEOFAPROPERLY
WITH
MAINTAINED
PROPER
OPERATION.
ONOTOPERATETHE PLUM A+ ONPATIENTS
ANDCHARGEDBATTERYHELPSCONFIRM
To confirm battery is fully charged, connect the Plum A+ to AC
(mains) power for a minimum of six hours while in the OFF mode.
If an alarm occurs during the power on self-test, identify the alarm
message, then take corrective action (see Section 7, Alarms and Troubleshooting).
Turn infuser OFF, then back ON. If the self-test alarm recurs,
remove the infuser from service and contact the local Hospira
representative.
The infuser performs a suite of System Self-Tests when power is
applied to confirm readiness for use.
430-95215-003 (Rev. 10/05)
Page 50
4- 44) Basic Operation
Failure during the Self-Tests is reported as a Malfunction
Condition.
DATA RETENTION
Delivery program settings and programming option selections are
retained in non-volatile memory. If the Plum A+ has been turned
off for longer than four hours, all delivery settings are cleared and
programming option selections are restored to their default
selections for next use.
4.2 Power On
1) Press to turn on the
Plum A+ and you will see
the System Self-Tests
screen.
HOSPIRA Plum A+
Release XX.XX
Copyright Hospira
2005
System Self Test
In Progress
CAUTION: THEINFUSERMUSTBE
INITIALLY
OTHERWISETHEDATAPORTMAYNOT
OPERATE
After self-testing
successfully completes,
the display will proceed to
STARTEDON AC POWER,
PROPERLY.
**SCREENSSHOWNINTHISMANUAL
AREEXA MPL ES ONLYANDDON’T
NECESSARILYREFLECTACTUAL
CLINICALPRACTICE**
the next screen.
Programming the infuser may begin during the
cassette check phase of the start-up testing.
NOTE: If battery is missing or defective, a “Battery not
installed” message with a “Continue without battery” prompt
will appear (AC power only).
430-95215-003 (Rev. 10/05)
Page 51
Plum A+ Infusion Systems4- 5
CAUTION: DONOTOPERATETHE PLUM A+ ONPATIENTS
THEBATTERYREMOVED. USEOFAPROPERLY
WITH
MAINTAINED
PROPER
ANDCHARGEDBATTERYHELPSCONFIRM
OPERATION.
2) To set up the Plum A+,
insert selected Plum Set
and close cassette holder
door lever.
NOTE: If cassette is already
inserted, this screen is skipped.
Example of Clear Settings
Screen
3) The on-screen message
“Clear Settings?” may
appear. Select [YES] to
clear all settings.
Clear ALL settings when
changing patients or
multiple programming
parameters.
SETUP
Insert PLUM Set
Close Lever
SETUPA
Clear Settings?
Yes clears ALL settings
YesNo
B
NOTE: This clears all delivery parameters, such as Rate and
VTBI on both lines, as well as setting items, such as the KVO
and Nurse Callback options, to their default values.
NOTE: If no settings currently exist or all settings are zero,
this screen will be bypassed.
430-95215-003 (Rev. 10/05)
Page 52
4- 64) Basic Operation
4.3 Simple Delivery
Example of Main Delivery
STOPPED STOPPED
A
screen (with cleared
settings).
**SCREENSSHOWNINTHIS
MANUALAREEXA MPL ESONLYAND
DON’TNECESSARILYREFLECT
ACTUALCLINICALPRACTICE**
NOTE: For clarity, whole number
font sizes will appear larger than partial number font sizes
(numbers on the right-side of the comma).
Back
Prime
Rate
0
mL/hr
Vol Inf
00
mL
B
A
1) To Program a delivery rate, select [A].
2) To enter delivery rate, use
A
PROGRAM
numeric keypad.
Rate
3) Use key to move
down to the VTBI field.
VTBI
Duration
Enter Value using keypad
Program
Options
00:00
Therapy
0
0
0
Options/
Vol Inf
mL/hr
mL
hr : min
Cancel/
Back
B
430-95215-003 (Rev. 10/05)
Page 53
Plum A+ Infusion Systems4- 7
4) To enter VTBI amount,
use keypad.
NOTE: While the VTBI is being
entered, the duration is
automatically computed. If the
computed value is greater than
99:59, dashes (- -) appear in place
of hours.
5) Use key to move
down to the DURATION field.
6) Use keypad to change the
hours part of duration if
desired.
NOTE: When all parameters are
displayed, changing any one
parameter will result in other
parameters automatically being
computed. To understand the
Automatic Calculation feature, see
Section 6.8 Examples of Automatic
Calculation.
A
PROGRAM
Rate
VTBI
Duration
Enter Value using keypad
Program
Options
A
Rate
VTBI
Duration
Enter Value using keypad
Program
Options
5.
00:00
Therapy
PROGRAM
5.
0.
00:05
Therapy
5
0
5
5
mL/hr
mL
hr : min
Cancel/
Back
mL/hr
mL
hr : min
Cancel/
Back
7) Use to move from hours to minutes.
NOTE: The maximum hours allowed is 99 and the maximum
minutes is 59.
8) Use keypad to change minutes.
CAUTION: VERIFY ALL VALUES BEFORESTARTING
INFUSION
.
430-95215-003 (Rev. 10/05)
Page 54
4- 84) Basic Operation
9) Press to begin
infusion.
Example of Main Delivery
Screen just after being
programmed and started.
NOTE: The Line A Fluid Flow
Indicator also will flash to indicate
pumping.
PUMPING STOPPED
A
Rate
5.
5
mL/hr
Vol Inf
0.
1
mL
A
B
0
0
Options/
B
Vol Inf
430-95215-003 (Rev. 10/05)
Page 55
Plum A+ Infusion Systems4- 9
4.4 Titration
1) Select Line [A]
(or [B]) at the Main
Delivery Screen.
2) Enter the desired Rate
using the keypad.
NOTE: When Rate is changed, the
duration is automatically
recomputed using the current
VTBI. To understand the Automatic
Calculation feature, see Section
6.8 Examples of Automatic
Calculation.
CAUTION: V
INFUSION
ERIFY ALL VALUES BEFORESTARTING
.
PUMPING PUMPING
A
25
4120
A
Rate
VTBI
Duration
Enter Value using keypad
Program
Options
Rate
mL/hr
Vol Inf
mL
A
PROGRAM
03:20
Therapy
15
50
B
50
Options/
Vol Inf
mL/hr
mL
hr : min
Cancel/
Back
B
3) After any change,
press to initiate new
Rate and return to the
Main Delivery Screen.
NOTE: Volume Infused is an
incrementing value which is
cleared in
[OPTIONS/VOL INF].
PUMPING PUMPING
A
15
41.
Rate
mL/hr
Vol Inf
6
mL
B
A
50
20
Options/
Vol Inf
B
430-95215-003 (Rev. 10/05)
Page 56
4- 104) Basic Operation
The following 3 screens are examples of
changed values.
Example of changed rate.
4) From the Main Delivery
Screen, select [A].
A
Rate
VTBI
Duration
PROGRAM
15
50
03:20
5) Use keypad to change the
RATE value.
NOTE: The display time is only
accurate to ± one minute. As
shown in this example, the delivery shows 3 hours and 20
minutes (not 3 minutes and 20 seconds).
NOTE: [START] accepts the change. Delivery is changed to
the new RATE and DURATION and display returns to Main
Delivery Screen.
Example of Changed VTBI
Enter Value using keypad
Program
Options
A
Therapy
PROGRAM
value.
6) Use key to highlight
the VTBI field.
Rate
VTBI
Duration
15
100
06:40
mL/hr
mL
hr : min
Cancel/
Back
mL/hr
mL
hr : min
Enter Value using keypad
7) Use the keypad to change
the VTBI value.
Program
Options
Therapy
Cancel/
Back
NOTE: When the VTBI is changed,
the Duration is recomputed with respect to the current Rate.
To understand the Automatic Calculation feature, see Section
6.8 Examples of Automatic Calculation.
NOTE: [START] accepts the change. Delivery is changed to
the new VTBI and DURATION and display returns to Main
Delivery Screen.
430-95215-003 (Rev. 10/05)
Page 57
Plum A+ Infusion Systems4- 11
Example of changed
A
PROGRAM
DURATION field.
Rate
8) Use to highlight the
Duration field, then use
the keypad to change it’s
value.
VTBI
Duration
Enter Value using keypad
Program
Options
NOTE: When DURATION is
changed, RATE is automatically
recomputed using the current VTBI value. To understand the
Automatic Calculation feature, see Section 6.8 Examples of
Automatic Calculation.
NOTE: [START] accepts the change. Delivery is changed to
the new DURATION and RATE and display returns to Main
Delivery Screen.
CAUTION: V
INFUSION
ERIFY ALL VALUES BEFORESTARTING
.
15
100
00:40
Therapy
4.5 Piggyback Delivery
mL/hr
mL
hr : min
Cancel/
Back
PIGGYBACK DELIVERYInfusion will stop Line A
and infuse Line B until
VTBI completes. Line A will
automatically restart.
1) With Line A pumping, to
program a Piggyback
delivery, select
[B].
430-95215-003 (Rev. 10/05)
PUMPING STOPPED
A
Rate
5.
5
mL/hr
Vol Inf
200
mL
A
B
5
Options/
B
Vol Inf
Page 58
4- 124) Basic Operation
The Delivery Mode field is
highlighted.
NOTE: If an entry was previously
programmed, RATE is highlighted.
If no previous entries, MODE is
highlighted.
2) If field reads
“Concurrent”, select
[CHANGE MODE] to
switch to “Piggyback”.
3) Select to highlight the
RATE field.
4) To enter delivery rate, use
numeric keypad.
5) Use key to move
down to the VTBI field.
PROGRAM
Mode
Piggyback
Rate
VTBI
Duration
Change using Change Mode
Program
Change
Options
Mode
PROGRAM
Mode
Rate
VTBI
Duration
Enter value using keypad
Program
Options
0
0
00:00
Therapy
Piggyback
200
0
00:00
Therapy
mL/hr
mL
hr : min
Cancel/
Back
mL/hr
mL
hr : min
Cancel/
Back
B
B
6) To enter VTBI amount,
PROGRAM
B
use keypad.
Mode
Piggyback
NOTE: While the VTBI is being
entered, the duration is
automatically computed. If the
computed value is greater than
99:59, dashes (- -) appear in place
of hours. To understand the
Rate
VTBI
Duration
Enter value using keypad
Program
Options
200
100
00:30
Therapy
mL/hr
mL
hr : min
Cancel/
Back
Automatic Calculation feature, see
Section 6.8 Examples of Automatic
Calculation.
430-95215-003 (Rev. 10/05)
Page 59
Plum A+ Infusion Systems4- 13
7) Use key to move
PROGRAM
down to the DURATION
Mode
field.
(Optional) Use keypad to
change duration.
Enter value using keypad
Program
Options
Rate
VTBI
Duration
Piggyback
133
100
00:45
Therapy
When all parameters are
displayed, changing any
one parameter will result in
other parameters
automatically being
computed.
CAUTION: VERIFY ALL VALUES BEFORESTARTING
INFUSION
.
8) Press to begin infusion.
NOTE: If Line A has a Delay Start remaining, a message will
appear warning of a possible non-delivery (see section 6.4).
mL/hr
mL
hr : min
Cancel/
Back
B
9) Select [YES] to continue.
Example of Main Delivery
Screen just after being
programmed and started.
430-95215-003 (Rev. 10/05)
DELAYED PUMPING
A
Rate
5.
5
mL/hr
Vol Inf
200.
mL
A
B
133
1
Options/
B
Vol Inf
Page 60
4- 144) Basic Operation
4.6 Concurrent Delivery
CONCURRENT DELIVERY-
PUMPING STOPPED
A
B
Simultaneous delivery of two
fluids at independent flow rates.
NOTE: In concurrent mode with certain alarm conditions, one
line may stop while the other line may continue to deliver.
1) With pumping started on line A, to program a
Concurrent delivery rate, select [B].
The Delivery Mode field is
PROGRAM
B
highlighted.
Mode
Piggyback
NOTE: If an entry was previously
programmed, RATE is highlighted.
If no previous entries, MODE is
highlighted.
If field reads “Piggyback”,
Rate
VTBI
Duration
Change using Change Mode
Program
Change
Options
Mode
0
0
00:00
Therapy
mL/hr
mL
hr : min
Cancel/
Back
select [CHANGE MODE] to
switch to “Concurrent”.
430-95215-003 (Rev. 10/05)
Page 61
Plum A+ Infusion Systems4- 15
Verify change to
PROGRAM
“Concurrent” mode.
Mode
Concurrent
2) Use to highlight Rate
field.
3) To enter delivery rate, use
Rate
VTBI
Duration
Change using Change Mode
Program
Options
00:00
Change
Therapy
Mode
PROGRAM
numeric keypad.
Mode
Concurrent
NOTE: When Rate for Line A plus
Rate for Line B is greater than valid
rate ranges, a “Concurrency
violation” message will explain why
second line did not start. See
Section 9.3 for valid delivery rate
ranges.
NOTE: “Concurrency violation” may not be displayed if unit is
displaying “VTBI Complete”. The device will not start and an
invalid keypress sound will occur when START is pressed.
Rate
VTBI
Duration
Enter value using keypad
Program
Options
00:00
Therapy
10
0
0
0
mL/hr
mL
hr : min
Cancel/
Back
mL/hr
mL
hr : min
Cancel/
Back
B
B
4) Use key to move down to the VTBI field.
5) To enter VTBI amount,
PROGRAM
use keypad.
Mode
Concurrent
NOTE: While the VTBI is being
entered, the duration is
automatically computed. While the
computed value is greater than
99:59, dashes (- -) appear in place
of hours. To understand the
Rate
VTBI
Duration
Enter value using keypad
Program
Options
10
100
10:00
Therapy
mL/hr
mL
hr : min
Cancel/
Back
430-95215-003 (Rev. 10/05)
B
Page 62
4- 164) Basic Operation
Automatic Calculation feature, see Section 6.8 Examples of
Automatic Calculation.
6) Use key to move down to the DURATION
field.
(Optional) Enter values for desired hours and
minutes.
NOTE: When Duration is entered, Rate is automatically
computed.
CAUTION: V
INFUSION
ERIFY ALL VALUES BEFORESTARTING
.
7) Press to begin infusion and return to the
Main Delivery screen.
Example of a concurrent
PUMPING PUMPING
A
B
delivery.
Rate
20.
5.
5
mL/hr
Vol Inf
4
A
mL
10
1
Options/
B
Vol Inf
430-95215-003 (Rev. 10/05)
Page 63
Plum A+ Infusion Systems4- 17
4.7 Stop and Start with only 1 Line
Pumping
1) Press at the Main
Delivery Screen.
NOTE: If [A] (or [B]) is
selected prior to [STOP], the line is
stopped, the display will show the
programming screen. Select
[CANCEL/BACK] to return to the
Main Delivery screen.
Example of STOPPED
delivery.
2) To discontinue infusion,
press .
3) Clamp secondary tubing
if connected.
4) Open door, remove set
PUMPING STOPPED
A
Rate
mL/hr
Vol Inf
5
8.
mL
A
STOPPED STOPPED
Rate
mL/hr
Vol Inf
5
8.
mL
A
Back
Prime
A
Back
Prime
100
100
B
0
0
Options/
B
Vol Inf
B
0
0
Options/
B
Vol Inf
and close door. Or, press to resume
infusion.
430-95215-003 (Rev. 10/05)
Page 64
4- 184) Basic Operation
4.8 Stop and Start with Both Lines
pumping
1) Press at the Main
Delivery screen.
2) Select [STOP A],
[STOP B], or
[STOP ALL].
NOTE: If [A] (or [B]) is
selected prior to [STOP], the
display will show the programming
screen. Select
to return to the Main Delivery
screen.
[CANCEL/BACK]
PUMPING PUMPING
A
25
AB
25
Choose line(s) to stop:
Stop
All
Rate
mL/hr
Vol Inf
320.
PUMPING PUMPING
320.
mL
B
A
Rate
mL/hr
Vol Inf
mL
StopAStop
B
50
4
Options/
Vol Inf
50
4
Cancel
B
430-95215-003 (Rev. 10/05)
Page 65
Plum A+ Infusion Systems4- 19
Example of stopped
STOPPED STOPPED
A
infusion.
Rate
mL/hr
Vol Inf
320.
mL
A
STOPPED STOPPED
NOTE: This is an example using
[STOP ALL].
CAUTION: V
BEFORESTARTINGINFUSION.
ERIFY ALL VALUES
3) To resume infusion,
25
Back
Prime
AB
press .
Rate
mL/hr
Vol Inf
320.
mL
StartAStart
4) Select [START A],
[START B], or
[START ALL].
Or, to discontinue
25
Choose line(s) to start:
Start
All
infusion, press .
5) Clamp secondary tubing if connected.
B
50
4
Options/
B
Vol Inf
50
4
Cancel
B
6) Open door, remove set and close door.
430-95215-003 (Rev. 10/05)
Page 66
4- 204) Basic Operation
4.9 Backpriming
BACKPRIMING- Removal of air
STOPPED STOPPED
A
or fluid from the proximal line
and cassette airtrap.
1) To initiate a backprime,
Rate
5.
5
mL/hr
Vol Inf
8.
9
mL
0
0
select [BACK PRIME].
NOTE: Backpriming is available
only when all delivery is stopped.
Secondary container is necessary
to receive the air.
2) Select and hold
[BACK PRIME] until fluid
pumped from line A to
line B has expelled the air
from cassette air trap into
Back
Prime
A
5.
8.
A
BACKPRIMING
Rate
5
mL/hr
Vol Inf
9
mL
Options/
B
Vol Inf
0
0
Secondary container.
NOTE: Fluid flow is at maximum
rate possible.
Release Backprime to Stop
Back
A
Prime
Options/
B
Vol Inf
3) Upon release of [BACK PRIME], the cassette
test is performed to confirm readiness for
B
B
pumping. Pressing will resume
programmed delivery.
430-95215-003 (Rev. 10/05)
Page 67
Plum A+ Infusion Systems4- 21
4.10 Clearing Program Settings
1) Press at the Main
Delivery Screen.
Clear ALL settings when
changing patients or
multiple programming
parameters.
Use [CLEAR PROGRAM]
when changing ONLY Line A or B.
2) Select Line A (or B if
active) from the stopped
Main Delivery Screen.
3) Select
[CLEAR PROGRAM].
NOTE: If Line A or B is
programmed with Dose
Calculation, Rate, VTBI and
Duration will not display values.
The values to be cleared may be
viewed in applicable programming
screen under
[DRUG LIST].
PUMPING STOPPED
A
Rate
5.
5
mL/hr
Vol Inf
0.
1
mL
A
A
PROGRAM
Rate
VTBI
Duration
Enter Value using keypad
Program
Options
Clear
Program
5.
03:20
Therapy
22
B
5
0
0
Options/
Vol Inf
mL/hr
mL
hr : min
Cancel/
Back
B
430-95215-003 (Rev. 10/05)
Page 68
4- 224) Basic Operation
4) Select [YES].
Clear ALL settings when
changing patients or
multiple programming
parameters.
Use [CLEAR PROGRAM]
when changing ONLY Line
A or B.
Example screen of
“Cleared Program”.
5) Select [CANCEL/BACK] to
return to the Main
Delivery screen or
program new settings.
PROGRAMA
Clear Line A Settings?
YesNo
A
PROGRAM
Rate
VTBI
Duration
Enter Value using keypad
Program
Options
0
0
00:00
Therapy
mL/hr
mL
hr : min
Cancel/
Back
430-95215-003 (Rev. 10/05)
Page 69
Plum A+ Infusion Systems5- 1
5)Therapies
Allows programming Dose Rates in
Dose
Calculation
Loading
Dose
Multistep
alternative units of measure. Dose
Calculation can be used in Simple Delivery,
Loading Dose, and Multistep.
Allows programming of an initial infusion
rate for a specific volume, followed
automatically by a maintenance rate from
the same container (e.g., a fluid challenge). If
Dose Calc is used, the Loading Dose and
Maintenance Dose are in the same unit of
measure, over the same period of time
(mcg/min), from the same container.
Allows a sequential program to deliver up to
10 steps; fluid volumes and delivery rates
may be programmed for each step. The
program may be entered based on Rate and
Volume or Volume and Time. If Dose Calc is
used, the delivery steps are in the same unit
of measure, over the same period of time,
from the same container.
NOTE: During programming, user must identify desired drug
with softkey drug list prior to setting program parameters.
Use drug manufacturer recommendations for IV
administration when using the Plum A+ infuser.
430-95215-003 (Rev. 10/05)
Page 70
5- 25) Therapies
5.1 Dose Calc (mcg/kg/min on A)
OSE CALCULATION- Allows programming
D
Dose Rates in alternative units of measure.
Dose Calculation can be used in Simple
Delivery, Loading Dose, and Multistep.
Clear ALL values when changing patients or
multiple programming parameters.
Use
[CLEAR PROGRAM] when changing ONLY
Line A or B.
1) Select [A] at the Main
Delivery screen.
**SCREENSSHOWNINTHIS
MANUALAREEXA MPL ESONLYAND
DON’TNECESSARILYREFLECT
ACTUALCLINICALPRACTICE**
The Line A Programming
screen appears.
2) Select
[THERAPY]for
Dose Calculation
programming.
STOPPED STOPPED
A
Rate
0
mL/hr
Vol Inf
00
Back
Prime
A
Rate
VTBI
Duration
Enter Value using keypad
Program
Options
mL
B
A
PROGRAM
0
0
00:00
Therapy
0
Options/
Vol Inf
mL/hr
mL
hr : min
Cancel/
Back
B
430-95215-003 (Rev. 10/05)
Page 71
Plum A+ Infusion Systems5- 3
The Drug List screen
appears, in mainly
alphabetical order.
NOTE: Default entry is “No Drug
Selected”. This selection may be
used later to remove an incorrect
or unwanted drug name from the
display.
3) Use to highlight
desired drug name.
When list does not contain
desired generic drug name
or group name select
[PAGE DOWN] (or [PAGE
UP]) until it shows, then
use to highlight
choice.
A
Program Drug List
No Drug Selected
Abciximab
Aldesleukin
Alteplase (rt-PA)
Select, then Enter
PageUpPage
Down
A
Program Drug List
Dexmedetomidine HCI
Dobutamine
Docetaxel
Dopamine
Doxorubicin
Epinephrine
Select, then Enter
Epoprostenol
PageUpPage
Down
Acyclovir
Albumin
Alfentanil
Amikacin
Enter
Diltiazem
Enter
Cancel/
Back
Cancel/
Back
4) Select [ENTER] to indicate selection and
advance to the Program Selection screen.
The Program Selection
screen for Line A appears.
5) Select [CHOOSE] to
select Dose Calculation
A
Dopamine
PROGRAM
Dose Calculation
Loading Dose
Multistep
and continue.
Select, then Choose
Choose Back
430-95215-003 (Rev. 10/05)
Page 72
5- 45) Therapies
The Program Dose Calc
screen appears.
NOTE: Cursor defaults to package
insert recommended dose units
associated with the selected drug.
6) Select [CHOOSE] to
continue. Or, use to
change the units to use in
the delivery.
The Container
Concentration Selection
screen appears whenever
the selected Dose Units
are gram-based (e.g.,
grams, mg, mcg, or ng)
otherwise this screen is
skipped.
NOTE: Cursor defaults to package
insert recommended concentration
units associated with the selected
drug.
7) Select [CHOOSE] to continue. Or, use
to change the Container Concentration units
to use in the delivery.
430-95215-003 (Rev. 10/05)
Page 73
Plum A+ Infusion Systems5- 5
The next page of the
Program Dose Calc screen
appears to provide for
entry of delivery program
parameters.
8) Use keypad to enter
desired values. Once
Program Dose Calc
A
Dopamine
Conc
Weight
Enter value using keypad
Program
Options
Dose
VTBI
Duration
Rate
400
70mgkg
00: 00
each value is entered, use
to highlight the next
programming field.
NOTE: The Concentration elements and weight value must
be entered before the [SELECT] keys allow access to
remaining fields.
9) Continue to enter values
using the keypad, and
advancing to the next
field using .
NOTE: As the Dose value is
entered, the Rate is automatically
calculated. To understand the
Automatic Calculation feature, see
Section 6.8 Examples of Automatic
Calculation.
Program Dose Calc
A
Dopamine
Conc
Weight
Dose
VTBI
Duration
Rate
Enter value using keypad
Program
Options
40070mgkg250 mL
00: 00
26.
0
0
0
10
0
3
250 mL
mcg/kg/min
mL
hr : min
mL/hr
Cancel/
Back
mcg/kg/min
mL
hr : min
mL/hr
Cancel/
Back
430-95215-003 (Rev. 10/05)
Page 74
5- 65) Therapies
10) Use to scroll down to
the VTBI field.
11) Enter a value for VTBI
using the keypad.
NOTE: As the VTBI value is
entered or changes, Duration is
Program Dose Calc
A
Dopamine
Conc
Weight
Dose
VTBI
Duration
Rate
Enter value using keypad
Program
Options
40070mgkg250 mL
250
09: 30
26.
10
3
mcg/kg/min
mL
hr : min
mL/hr
Cancel/
Back
automatically updated.
NOTE: At this point, programming is complete unless
[PROGRAM OPTIONS] are desired. See Section 6, Additional
Features, for more information.
12) After entering the desired VTBI value and any
desired [PROGRAM OPTIONS], press to
begin program.
This will bring up the
PROGRAM CONFIRMATION
screen which appears for
all units of measure except
mL/hr.
CAUTION: VERIFY ALL VALUES
BEFORESELECTING
STARTINGINFUSION.
[YES]AND
Program Dose Calc
A
Delivery will be:
Dose
Drug
Conc
Weight
YesNo
10
mcg/kg/min
Dopamine
400
mg
70
Confirm Program?
kg
250 mL
NOTE: The Concentration drug
amount or Diluent volume cannot be changed while a
delivery is in progress.
430-95215-003 (Rev. 10/05)
Page 75
Plum A+ Infusion Systems5- 7
The Main Delivery screen
appears displaying the
information just entered.
NOTE: The Therapy Type and
Drug are displayed as well as the
Dose Rate and Dose Units.
PUMPING STOPPED
A
Dopamine
Dose Calculation
26.
mcg/kg/min
Rate
3
mL/hr
Vol Inf
00
mL
10
A
0
Options/
B
Vol Inf
B
430-95215-003 (Rev. 10/05)
Page 76
5- 85) Therapies
5.2 Dose Calc (mg/min on B)
OSE CALCULATION- Allows programming
D
Dose Rates in alternative units of measure.
Dose Calculation can be used in Simple
Delivery, Loading Dose, and Multistep.
Clear ALL values when changing patients or
multiple programming parameters.
Use [CLEAR PROGRAM] when changing ONLY
Line A or B.
1) Select [B] at the Main
Delivery screen.
The Line B Programming
screen appears.
2) Select
[THERAPY].
PUMPING STOPPED
A
Dopamine
Dose Calculation
26.
mcg/kg/min
Mode
Rate
VTBI
Duration
Change using Change Mode
Program
Options
Rate
3
mL/hr
Vol Inf
00
mL
10
A
PROGRAM
Piggyback
00:00
Change
Mode
B
0
0
Therapy
0
Options/
Vol Inf
mL/hr
mL
hr : min
Cancel/
Back
B
B
430-95215-003 (Rev. 10/05)
Page 77
Plum A+ Infusion Systems5- 9
The Drug List screen
appears, in mainly
alphabetical order.
No Drug Selected
Abciximab
Aldesleukin
Alteplase (rt-PA)
Select, then Enter
PageUpPage
Down
Acyclovir
Albumin
Alfentanil
Amikacin
Enter
3) Use to highlight
desired drug name.
When list does not contain
desired generic drug name
or group name select
[PAGE DOWN] (or
[PAGE UP]) until it shows,
Select, then Enter
PageUpPage
Leucovorin
Levofloxacin
Lidocaine
Lipids
Lorazepam
Magnesium Infusion
Mannitol
Mesna
Enter
Down
then use to highlight
choice.
4) Select [ENTER] to indicate selection and
advance to the Therapy Selection screen.
Cancel/
Back
Cancel/
Back
BProgram Drug List
BProgram Drug List
The Program Select screen
for Line B appears.
5) Select
[CHOOSE] to
continue.
PROGRAM
Dose Calculation
Loading Dose
Multistep
Select, then Choose
Choose Back
B
Lidocaine
430-95215-003 (Rev. 10/05)
Page 78
5- 105) Therapies
The Program Dose Calc
screen appears.
NOTE: Cursor defaults to
manufacturer package insert
recommended dose units
associated with the selected drug.
6) Select [CHOOSE] to
continue. Or, use to
change the units to use in
the delivery.
The Container
Concentration Selection
screen appears whenever
the selected Dose Units
are gram-based (e.g.,
grams, mg, mcg, or ng)
otherwise this screen is
skipped.
NOTE: Cursor defaults to
manufacturer package insert
recommended concentration units
associated with the selected drug.
7) Select [CHOOSE] to continue. Or, use
to change the Container Concentration units
to use in the delivery.
430-95215-003 (Rev. 10/05)
Page 79
Plum A+ Infusion Systems5- 11
The next page of the
Program Dose Calc screen
appears to provide for
entry of delivery program
parameters.
8) Use keypad to enter
desired values. Once
Conc00grams
ModePiggyback
Dose
VTBI
Duration
Rate
Enter value using keypad
Program
Options
mL
00: 00
0
0
0
mg/min
mL
hr : min
mL/hr
entered, use to
highlight the next field.
9) To override the Piggyback Delivery Mode
default, scroll down to Mode using , then
select [CHANGE MODE].
NOTE: Delivery Mode option is available on Line B and its
default is Piggyback. The weight field is independent of Line
A unless Line A selected drug is weight-based. In that
scenario, the Line A weight information applies to Line B.
10) Continue entering
desired values using the
keypad. Use to
advance to the next field.
NOTE: When entering the RATE,
with a VTBI value entered, the
DOSE and DURATION fields are
automatically updated. To
understand the Automatic
Calculation feature, see Section 6.8 Examples of Automatic
Calculation.
Conc2
ModeConcurrent
Dose
VTBI
Duration
Rate
Enter value using keypad
Program
Options
500
grams
mL
500
16 : 40
30
2
mg/min
mL
hr : min
mL/hr
BProgram Dose Calc
Lidocaine
Cancel/
Back
BProgram Dose Calc
Lidocaine
Cancel/
Back
NOTE: The Duration field temporarily changes to dashes
when the computed value is larger than what can be
displayed.
430-95215-003 (Rev. 10/05)
Page 80
5- 125) Therapies
11) Continue entering the RATE value, if desired.
NOTE: As Rate is entered, the Duration and Dose are
automatically adjusted as individual numbers are entered.
NOTE: If the Duration value is changed, the Rate and Dose
are automatically adjusted.
NOTE: At this point, programming is complete unless
[Program Options] are desired. See Section 6, Additional
Features, for more information.
12) Press to begin
program.
This will bring up the
PROGRAM CONFIRMATION
screen which appears for
all units of measure except
Program Dose Calc
Delivery will be:
Dose
Drug
Conc
Weight
Confirm Program?
YesNo
mL/hr.
CAUTION: VERIFY ALL VALUES
BEFORESELECTING
STARTINGINFUSION.
NOTE: The Concentration drug
amount or Diluent volume cannot be changed while a
delivery is in progress.
[YES]AND
2
Lidocaine
2
70
mg/min
grams
kg
B
500 mL
430-95215-003 (Rev. 10/05)
Page 81
Plum A+ Infusion Systems5- 13
Upon , the Main
Delivery screen shows
delivery status for both A
and B lines.
NOTE: See Delivery Rate Range
in Section 9. The ratio of
concurrent rates has
recommendations specified in
Section 1.4.
STOPPED PUMPING
A
Dopamine
Dose Calculation
26.
3
10.
mcg/kg/min
A
Rate
mL/hr
Vol Inf
mL
Dose
Lidocaine
Dose Calculation
30
1
2
mg/min
Options/
B
Vol Inf
B
430-95215-003 (Rev. 10/05)
Page 82
5- 145) Therapies
5.3 Loading Dose
OADING DOSE- Allows programming of an
L
initial infusion rate for a specific volume,
followed automatically by a maintenance rate
from the same container (e.g., a fluid
challenge). If Dose Calc is used, the Loading
Dose and Maintenance Dose are in the same
unit of measure, over the same period of time
(mcg/min), from the same container.
Clear ALL values when changing patients or
multiple programming parameters.
Use [CLEAR PROGRAM] when changing ONLY
Line A or B.
1) Select [A] (or [B]) at
STOPPED STOPPED
A
B
the Main Delivery screen.
Rate
0
mL/hr
Vol Inf
00
Back
Prime
mL
A
0
Options/
B
Vol Inf
430-95215-003 (Rev. 10/05)
Page 83
Plum A+ Infusion Systems5- 15
The selected Line (A or B)
Programming screen
appears.
2) Select [THERAPY].
The Drug List Screen
appears, in mainly
alphabetical order.
NOTE: Default entry is “No Drug
Selected”. This process will guide
you through a “fluid challenge”, so
“No Drug Selected” will be entered.
3) Select [ENTER] to
indicate selection and
advance to the Therapy
selection screen.
A
PROGRAM
Rate
VTBI
Duration
Enter Value using keypad
Program
Options
A
Select, then Enter
PageUpPage
00:00
Therapy
Program Drug List
No Drug Selected
Abciximab
Acyclovir
Albumin
Aldesleukin
Alfentanil
Alteplase (rt-PA)
Amikacin
Enter
Down
0
0
mL/hr
mL
hr : min
Cancel/
Back
Cancel/
Back
The Select Therapy screen
for Line A (or B) appears.
NOTE: This example shows line A
being programmed.
430-95215-003 (Rev. 10/05)
A
Select, then Choose
PROGRAM
Dose Calculation
Loading Dose
Multistep
Choose Back
Page 84
5- 165) Therapies
4) Use to scroll down to
Loading Dose.
5) Select [CHOOSE] to
continue.
The next page of the
Program Loading Dose
screen appears.
6) Select [CHOOSE] to
continue. Or, use to
change the units to use in
the delivery.
The Loading Dose
Programming screen
appears.
7) The default field is Rate.
Since a “fluid Challenge”
is being performed in this
10) Enter Duration using
keypad. The Rate value is
automatically calculated.
To understand the
Automatic Calculation
feature, see Section 6.8
Examples of Automatic
Calculation.
Program Loading Dose
A
Rate
VTBI
mL/hr
mL
0
1
2
Enter value using keypad
Program
Options
Program Loading Dose
A
1
2
Enter value using keypad
Program
Options
Rate
mL/hr
500
250
0
0
VTBI
mL
250
0
0
Duration
hr : min
00 : 00
00 : 00
Cancel/
Duration
hr : min
00 : 30
00 : 00
Cancel/
Back
Back
11) After the values for
loading dose Step-1 are
entered, use to
highlight the maintenance
dose, Step-2.
NOTE: The Loading Dose therapy
has two steps. The first is the
loading dose followed by the
maintenance dose. These words
will appear on the main screen as
each step is delivered.
Program Loading Dose
A
Rate
VTBI
mL
250
Duration
hr : min
00 : 30
0
00 : 00
Cancel/
Back
mL/hr
500
1
75
2
Enter value using keypad
Program
Options
430-95215-003 (Rev. 10/05)
Page 86
5- 185) Therapies
12) Enter the set of desired
values for Step-2 of the
program.
13) Select [PROGRAM
OPTIONS] to select a
Delayed Start or Standby.
NOTE: At this point, programming
is complete unless
Options] are desired. See Section
6, Additional Features, for more
information.
NOTE: Nurse Callback enables an
alarm when a change in the
delivery on a line occurs such as
completing Step-1 and starting
Step-2. Or on line B, a piggyback
delivery is completed and line A is
starting its delivery.
[Program
Program Loading Dose
A
Rate
VTBI
mL
250
750
Enter
Duration
hr : min
00 : 30
10 : 00
mL/hr
500
1
75
2
Enter value using keypad
Program
Options
A
Program Options
Delay Start for 00:00 hr:min
Callback No
Enter Value using keypad
Standby
Cancel/
Back
Cancel/
Back
14) Enter a delay time (less
A
Program Options
than or equal to 24 hours)
using the keypad.
Delay Start for 00:05 hr:min
NOTE: Not available if a delivery is
Callback Yes
taking place on the line.
NOTE: Nurse Callback cannot be
changed while a delivery is taking
Change using Yes/No
Yes/No
Enter
Cancel/
Back
place on the line.
15) Select [ENTER] to accept the changes and
430-95215-003 (Rev. 10/05)
Page 87
Plum A+ Infusion Systems5- 19
return to the Program Loading Dose screen.
NOTE: [CANCEL/BACK] returns the fields to their original
value and displays the previous screen.
CAUTION: V
INFUSION.
ERIFY ALL VALUES BEFORESTARTING
16) Press to begin program.
The Main Delivery Screen
A
appears.
NOTE: The words “DELAYED” and
“Loading Dose” appear in the
upper-left corner. These indicate
the options which have been
selected.
After the programmed
A
delay period has expired,
the Plum A+ starts
“Pumping”. This is
indicated where “Delayed”
previously was displayed.
DELAYED STOPPED
Loading Dose
500
Loading Dose
500
Rate
mL/hr
Vol Inf
00
mL
B
A
PUMPING STOPPED
Rate
mL/hr
Vol Inf
00
mL
0
Options/
Vol Inf
0
B
B
A
Vol Inf
Options/
B
430-95215-003 (Rev. 10/05)
Page 88
5- 205) Therapies
After the Step-1 has
completed, the Main
Delivery screen changes to
the second step.
PUMPING
A
Maintenance Dose
75
2500
STOPPED
Rate
mL/hr
Vol Inf
mL
0
NOTE: A flashing “Callback Line A”
message and the audio alarm
sound indicates activation of that
option.
N105 Callback Line A
A
Options/
B
Vol Inf
17) The user should acknowledge the Callback
message and stop the audio alarm by
pressing .
Upon completion of
delivery, the screen shows
a flashing “Line A VTBI
A
Maintenance Dose
1
complete” message and
the audible alarm sounds.
NOTE: The KVO rate will change
to the rate of 1 mL/hr or less
depending on delivery rate and
“Pumping” mode changes to
“KVO”. If Continue Rate option was selected, pumping
continues at original rate.
10000
N161 Line A VTBI complete
KVO
mL/hr
Vol Inf
A
Rate
STOPPED
mL
B
0
Options/
Vol Inf
B
B
18) Press to stop alarm sound.
NOTE: Warning message continues to flash and audio alarm
will return after a two minute period of silence, unless delivery
is stopped or VTBI of last line is changed.
430-95215-003 (Rev. 10/05)
Page 89
Plum A+ Infusion Systems5- 21
5.4 Multistep Programming
ULTISTEP- Allows a sequential program to
M
deliver up to 10 steps; fluid volumes and
delivery rates may be programmed for each
step. The program may be entered based on
Rate and Volume or Volume and Time. If Dose
Calc is used, the delivery steps are in the same
unit of measure, over the same period of time,
from the same container.
Clear ALL values when changing patients or
multiple programming parameters.
Use [CLEAR PROGRAM] when changing ONLY
Line A or B.
1) Select [A] at the Main
Delivery screen.
430-95215-003 (Rev. 10/05)
STOPPED STOPPED
A
Rate
0
mL/hr
Vol Inf
00
Back
Prime
mL
B
A
0
Options/
Vol Inf
B
Page 90
5- 225) Therapies
The Line A Programming
screen appears.
2) Select [THERAPY].
The Drug List Screen
appears, in mainly
alphabetical order.
3) Use to highlight
desired drug name.
When list does not contain
desired generic drug name
or group name, select
[PAGE DOWN] (or [PAGE
P]) until it shows, then
U
A
PROGRAM
Rate
VTBI
Duration
Enter Value using keypad
Program
Options
A
Select, then Enter
PageUpPage
A
Select, then Enter
PageUpPage
00:00
Therapy
Program Drug List
No Drug Selected
Abciximab
Acyclovir
Albumin
Aldesleukin
Alfentanil
Alteplase (rt-PA)
Amikacin
Enter
Down
Program Drug List
Ticar-clavulanate
Tirofiban
Tobramycin
TPN
Urokinase
Vancomycin
Vecuronium
Verapamil
Enter
Down
0
0
mL/hr
mL
hr : min
Cancel/
Back
Cancel/
Back
Cancel/
Back
use to highlight
choice.
4) Select [ENTER] to indicate selection and
advance to the Therapy selection screen.
430-95215-003 (Rev. 10/05)
Page 91
Plum A+ Infusion Systems5- 23
The Select Therapy screen
for Line A (or B) appears.
NOTE: This example shows line A
being programmed.
5) Use to scroll down to
Multistep.
6) Select [CHOOSE] to
continue.
The Program Multistep
screen appears.
NOTE: Cursor defaults to package
insert recommended dose units
associated with the selected drug.
7) Use key to highlight
the units to use in the
delivery.