Hospira Plum 360 with Hospira MedNet System Quick Reference Card Software

Page 1
Plum 360® with Hospira MedNet® System
Quick Reference Card Software 15.02 and 15.10
Please refer to the System Operating Manual for complete operating instructions.
Page 2
CONTENTS
Definition of Terms 3
Definition of Operating Keys – Indicators – Display Symbols 3
Set Up PlumSet
Clinical Care Area (CCA) Selection 6
Change CCA During Infusion 6
Programming Line A/B with a Custom Drug Library 7
Programming Line A/B without a Drug List 8
Programming a Bolus (for Plum 360 SW 15.1 and higher) 8
Nurse Callback 9
Titrate Rate On Line A/B 9
Address VTBI Complete Alarm 10
Backprime 10
Stop and Start with 1 Line Pumping 11
Stop and Start when Line A and B Pumping 11
Standby – A/B Delivery Screen 12
Delayed Start 13
Loading Dose 13
Multistep 14
Lock Keypad (Method #1) 14
Lock Keypad (Method #2) 15
®
6
Unlock the Keypad 15
Set the Post Infusion Rate 16
View/Clear Volumes Infused 16
Adjust Display Lighting and Contrast 17
View CCA and Infuser Settings 17
Set Distal Pressure Alarm Limit 18
Discontinue Electronic Flow Control and Setting Gravity Flow 18
Discontinue Fluid Administration 19
Alarms and Troubleshooting 19
High Priority Alarms 20
Medium Priority Alarms 23
Low Priority Alarms 24
Line A – High Priority Alarms 25
Line A – Medium Priority Alarms 27
Line B – High Priority Alarms 28
Line B – Medium Priority Alarms 30
Line A and B – High Priority Alarms 32
Page 3
DEFINITION OF TERMS
Hospira MedNet® Software – Enhances safety at the point of care with highly customizable drug libraries that guide users and help to protect patients by alerting to hard and soft limits, upper and lower dosing limits intended to help prevent infusion errors. These limits are based on a hospital’s specic IV administration practices.
Lower Hard Limit (LHL): The lower limit that cannot be overridden.
Lower Soft Limit (LSL): The lower limit that can be overridden.
Upper Soft Limit (USL): The upper limit that can be overridden.
Upper Hard Limit (UHL): The upper limit that cannot be overridden.
Auto-Program: Auto-programming refers to the ability to receive
a remotely congured therapy from Hospira MedNet
®
software.
DEFINITION OF OPERATING KEYS – INDICATORS – DISPLAY SYMBOLS
[ON/OFF] – Infuser power on and off.
[START] – Is the rst key to press to start a delivery.
For safety reasons, every delivery must be conrmed by checking the programming and then pressing an additional softkey, in response to a prompt.
[STOP] – Stops delivery. If 2 lines are pumping when you press [STOP], you must press one of the following softkeys: s [Stop A], s [Stop B], or s [Stop All] in response to the prompt to specify which line(s) to stop.
3
of 36
Page 4
[SELECT] – Moves the cursor between elds on the display. The top pair of arrows moves the cursor up or to the left.
The bottom pair of arrows moves the cursor down or to the right
[LOCK KEYPAD] – Pressing this key, followed by entering a lock passcode disables all keys on the keypad except
[STOP] until a valid passcode is entered.
[AUDIO PAUSED] – Has two functions, temporarily silencing all audio output for any active alarms for two minutes or temporarily silencing keypad input sound feedback for two minutes if there is no active alarm.
[C] – Clears all values in the currently highlighted eld. [C] also clears the dashes (- - -) that are displayed when an entry is invalid or a drug delivery parameter is beyond the pre-programmed hard limits. NOTE: [C] will NOT clear an entire program.
Caution – Appears on the display to tell the clinician to use
CAUTION because the specied drug has been programmed without rule sets (soft or hard limits), and may have been programmed outside of specied safety limits for that specic drug.
Upper Soft Limit Override – Appears next to the drug name when the dosage of the drug being infused is higher than the upper soft limit set for the drug in the Custom Drug Library.
Page 5
Lower Soft Limit Override – Appears next to the drug name when the dosage of the drug being infused is less than the lower soft limit set for the drug in the Custom Drug Library.
Wireless Connection – Appears when the infuser is communicating with the network using a wireless connection.
Hospira MedNet Connection – Appears when the infuser is communicating with Hospira MedNet software over either a wireless or Ethernet connection.
Battery Capacity – Shows the battery charge level when a battery is installed in the infuser, or indicates that a battery is not installed.
To maintain maximum battery charge and to prolong
battery life, connect the infuser to AC (mains) power whenever possible.
5
of 36
Page 6
SET UP PLUMSET
1
Prime the set according to package instructions.
2
Push in ow regulator to close.
3
Insert primed cassette into the infuser, close door.
4
Press [ON/OFF] to turn on the infuser.
®
CLINICAL CARE AREA (CCA) SELECTION
1
Use to highlight the desired CCA.
2
After you select the desired CCA, press s [Choose].
3
The on-screen message “New Patient” may appear – press s [Yes]
to clear all settings or s [No] to continue.
NOTE: If no settings currently exist or all settings are 0, this screen will be bypassed.
CHANGE CCA DURING INFUSION
1
On the Delivery (A/B) screen, press s [Settings/Vol/CCA].
2
Press s [Change CCA].
• The current CCA is indicated by arrows before and after the CCA.
3
Use the [SELECT] s or t to highlight the new CCA.
4
Press s [Choose].
5
Press s [Previous Screen].
NOTE: New CCA is highlighted at the bottom of the screen.
NOTE: When the Delivery screen displays, the infuser will inform the
user that line is delivering under a prior CCA. Until a VTBI Complete alarm occurs for the line, you can still titrate the infusion on that line under the old CCA.
Page 7
PROGRAMMING LINE A/B WITH A CUSTOM DRUG LIBRARY
1
On the Delivery (A/B) screen, select a line to program. If you are
programming a primary delivery, select Line A. If you are programming a secondary delivery, select Line B.
2
In the drug list, select the drug – You may use the alphanumeric
keypad to search.
3
Select a clinical use (if applicable).
4
Enter Weight (if applicable), Rate and/or Dose, VTBI, and Duration
as appropriate for the displayed program parameters. These values can be entered in any order.
CAUTION: BEFORE STARTING DELIVERY, VERIFY THE VALUES.
5
Press [START].
6
Conrm the program.
7
Press s [Yes] to start the infusion.
• If a maximum hard limit is exceeded, an alert appears. You cannot proceed until the entry is cleared. Press the entry and enter a new value.
• If a soft limit is exceeded, an alert appears when pressed to conrm the program.
• When the alert displays:
— Select s [Yes] to override and continue to the conrmation
screen. or
— Select s [No] to return to the program screen and edit the value.
When programming Line B with a non-Piggybackable drug (as dened in the Custom Drug Library), the delivery mode for Line B defaults to Concurrent and cannot be changed.
If there is a conrmed program on Line A with a drug that is non-interrupt­ible when programming Line B, the delivery mode for Line B will default to Concurrent and cannot be changed.
If Line A is not programmed, or is programmed with a drug that is interrupt­ible, and the drug selected on Line B is Piggybackable, select Piggyback or Concurrent as the delivery mode (the default is Piggyback).
[C] key to clear the
[START] is
7
of 36
Page 8
PROGRAMMING LINE A/B WITHOUT A DRUG LIST
1
On the Delivery (A/B) Screen, select a line to program. If you are
programming a primary delivery, select Line A. If you are programming a secondary delivery, select Line B.
2
On the Program screen, enter the Rate, VTBI and Duration.
Values can be entered in any order.
• A Caution Symbol will appear
on the Conrm Program screen
that alerts you that the infuser is being operated without rule sets.
3
Press [START].
• Before starting delivery, verify the values.
4
Press s [Yes] to start delivery.
PROGRAMMING A BOLUS (FOR PLUM 360_15.1 AND HIGHER WITH HOSPIRA MEDNET SAFETY SOFTWARE)
• A Bolus can be delivered from either Line A or Line B (while in the piggyback mode).
• A Bolus can be completed only if the following conditions are present:
—The line on which the bolus is to be delivered is currently infusing
—Bolus Dose is enabled within the medication’s selected prole
—Rule Sets permit the medication to be delivered by bolus
—There is adequate VTBI of the medication to complete
the bolus dose
—The device is in Piggyback mode (not Concurrent mode)
if the bolus is to be delivered on Line B
Page 9
1
Press s [A].
2
Press s [Bolus]. The Bolus softkey is available only if the medication
to be bolused is enabled in the drug library and is currently infusing.
3
On the Bolus programming screen enter:
Dose: ___________________________________
Duration: ________________________________
The Rate and VTBI will be calculated.
4
Press [START] for conrmation.
5
Select s [Yes] to start the delivery.
NURSE CALLBACK
1
On the programming screen, press s [Add Callback].
2
Press [START] and conrm your order for accuracy.
3
Press s [Yes] to start delivery.
Callback is available for Piggyback, Loading Dose, Multistep and Bolus deliveries.
TITRATE RATE ON LINE A/B
1
On the Delivery (A/B) screen, select s [A] or [B] – Line does not
have to be stopped to titrate Rate.
2
Ensure Rate/Dose eld is highlighted and enter a new Rate/Dose.
3
Press [START] and conrm your order for accuracy.
4
Press s [Yes] to start the titration.
9
of 36
Page 10
ADDRESS VTBI COMPLETE ALARM
Upon completion of delivery, the screen shows a ashing “VTBI completed Line A! – Add more VTBI or Clear Line A” message and the audible alarm sounds.
You can change the default setting to one of the following:
KVO: The infuser continues to deliver uid at a Keep Vein Open (KVO)
rate of 1 mL/hr. If the delivery rate of the infusion that just completed was less than 1 mL/hr , the KVO rate will continue at the same delivery rate.
RATE: The infuser continues to deliver uid at the programmed
rate, maintaining the therapeutic rate until the VTBI Complete alarm can be resolved.
1
Press [AUDIO PAUSED] to stop the alarm sound.
2
Press s [A].
3
Ensure the VTBI eld is highlighted – using the numeric keypad,
enter a new VTBI.
3
Press [START] and conrm your order for accuracy.
4
Press s [Yes] to start the titration.
BACKPRIME
Before you begin a backprime, ensure that there is a line or a syringe on the secondary port to accept the backprimed uid or expelled air. When you release s [Backprime] the infuser performs a cassette test.
1
Press and hold s [Backprime] until uid pumped from Line A
to Line B clears air from the cassette and from Line B (if present).
NOTE: If you press and hold the [Backprime] key for two minutes, a stuck key alarm sounds and the display screen shows Power Off then On. Replace pump if alarm continues.
10 of 36
Page 11
2
If the cassette test detects that there is still air in the line, repeat
Step1 until the cassette test is successful.
3
Press [START] to restart the delivery. If two lines were pumping
when delivery stopped, press the appropriate softkey in response to the prompt.
STOP AND START WITH 1 LINE PUMPING
1
On the Delivery (A/B) screen, press [STOP].
2
Press [START] to resume the infusion.
STOP AND START WHEN LINE A AND B PUMPING
1
On the Delivery (A/B) screen, press [STOP].
2
Press s [Stop All] or s [Stop A] or s [Stop B].
To Resume the infusion:
1
Press [START].
2
Press s [Start All] or s [Start A] or s [Start B].
11
of 36
Page 12
STANDBY – A/B DELIVERY SCREEN AND CONFIRMATION SCREEN
Standby - Enables the clinician to postpone starting the delivery for a period of 24-72 hours. Default is 72 hours.
If a line is in standby and the con gured maximum standby time expires, the program is cleared and the infuser alarms 2 minutes later if there has been no interaction with the pump on either line.
To place the infusion in Standby – from the Delivery Screen:
1
Press s [Standby] and select the
desired line to put in standby mode.
2
On the Con rm Standby screen –
press s [Yes]
To place the infusion in Standby – from the Confi rmation Screen:
After entering program on desired line, press Program Screen:
1
2
[START]. From the Con rm
Press s [Standby] and select the
desired line to put in standby mode.
On the Con rm Standby screen,
press s [Yes] to put the delivery in standby mode.
To restart the infusion from Standby:
1
Press [START].
The Cancel Standby screen displays. If both lines are in standby, the infuser will give you the option to select a line.
2
Select the appropriate softkey. Delivery resumes on the selected line(s).
12 of 36
Page 13
DELAYED START
1
On the Program screen, press s [Delay].
2
Enter time in hours and minutes up to 23:59 hh:mm and press
s [Done]. Delivery screen shows DELAYED and delay time countdown.
3
Press [START] and s [Yes] to conrm to resume the infusion.
4
To clear a delay, choose the line, press s [Delay], then change or
clear the delay settings, then press s [Done].
LOADING DOSE
1
After your medication is selected, press s [Choose].
2
Press s [Loading Dose] before entering any values on the
programming screen. The Program Loading Dose screen appears.
3
Enter Rate and/or Dose, VTBI, and Duration as appropriate for the
parameters displayed for 1 (Loading Dose).
NOTE: On the Program Loading Dose screen, 1 represents the
Loading Dose and 2 represents the Maintenance Dose.
4
After completely programming 1 (Loading Dose) navigate to 2
(Maintenance Dose) and program the Maintenance Dose parameters.
5
When both Loading Dose and Maintenance Dose are programmed,
press
6
Conrm that all programming is correct, and then press s [Yes].
[START].
13
of 36
Page 14
MULTISTEP
1
After your medication is highlighted, press s [Choose].
2
Press s [Multistep] before entering any values on the programming
screen.
3
Enter Rate and/or Dose, VTBI, and Duration as appropriate for the
parameters displayed for Step 1.
4
After programming all values in 1 (Step 1), navigate to the 2 (Step 2),
and program that step.
5
After programming all values in 2 (Step 2), continue with 3 (Step 3)
if desired.
6
Press s [To Steps 4-10] to program additional steps if needed.
7
When all steps are programmed, press [START].
8
Conrm that all programming is correct, and then press s [Yes]
to start the delivery for 1 (Step 1).
LOCK KEYPAD (METHOD #1)
Using the Lock Keypad hardkey.
1
Press the
2
The Passcode data entry screen appears on the display.
3
Enter the passcode using the numeric keypad.
4
Press the s [Enter] softkey.
When the keypad is locked, the lock symbol right corner of the delivery screen.
[LOCK KEYPAD] hardkey.
appears in the bottom
14 of 36
Page 15
LOCK KEYPAD (METHOD #2)
Via Settings/Vols/CCA
1
At the main delivery screen, press [Settings/Vols/CCA] softkey.
2
Press s [Lock Keypad] – the Passcode data entry screen appears
on the display.
3
Enter Passcode to Lock using the numeric keypad.
4
Press s [Enter] softkey.
When the keypad is locked, the lock symbol
appears in the bottom
right corner of the delivery screen.
UNLOCK THE KEYPAD
1
Press any key on the keypad to display the Passcode data entry
screen.
2
Enter the passcode using the numeric keypad.
3
Press the s [Enter] softkey.
NOTE: During an infusion, pressing
[STOP] or opening the cassette door, activates an alarm that cannot be silenced until the keypad is unlocked.
15
of 36
Page 16
SET THE POST INFUSION RATE
1
On the Main Delivery screen, press s [Settings/Vols/CCA].
The Settings/Vols/CCA screen appears, with the Post Infusion Rate highlighted.
2
Press s [Choose]. The Post Infusion Rate screen appears, with
the current setting highlighted.
3
To change the current setting, press s [KVO/Rate]. To return to the
previous setting, press s [KVO/Rate] again.
4
Press s [Done] to save your changes and return to the Settings/
Vols/CCA screen and then press s [Previous Screen] to return to the delivery screen.
NOTE: When changing the Post Infusion Rate from what is defaulted on the device, the new change will remain in place for any subsequent infusions programmed on the device unless manually changed back or, when powering the device back ON and s [Yes] is answered to the “New Patient?” screen.
Verify that the Post Infusion Rate is appro­priate for the Infusion.
VIEW/CLEAR VOLUMES INFUSED
1
On the Main Delivery screen, press s [Settings/Vols/CCA].
2
Use [SELECT] to highlight Volumes Infused and then press s
[Choose].
• The Volumes Infused Screen displays the volumes infused on Line A, Line B and the total infused.
3
Document totals.
4
Press s [Clear A] to clear the Line A total only (and subtract the
total from the Total Volume).
5
Press s [Clear B] to clear the Line B total only (and subtract the
total from the Total Volume).
16 of 36
Page 17
6
Press s [Clear Total] to clear all values, including the Total Volume.
7
To return to the Settings/Vols/CCA screen, press s [Previous
Screen]. If you do not press a key in 30 seconds, the Delivery screen
automatically returns.
ADJUST DISPLAY LIGHTING AND CONTRAST
1
On the Main Delivery screen, press s [Settings/Vols/CCA].
2
Use [SELECT] to highlight Lighting/Contrast and press s [Choose].
The Lighting/Contrast screen appears, with the Backlight Intensity highlighted.
3
Use s [Increase Setting] and s [Decrease Setting] to adjust the
backlight intensity.
4
Press [SELECT] to highlight Display Contrast.
5
Use s [Increase Setting] and s [Decrease Setting] to adjust the
display contrast.
6
Press s [Done] to save the current settings and return to the Settings/
Vols/CCA screen or press s [Previous Screen] to leave this screen without saving changes.
7
Press s [Previous Screen] to return to the Settings/Vols/CCA screen.
VIEW CCA AND INFUSER SETTINGS
1
On the Main Delivery screen, press s [Settings/Vols/CCA].
2
Select s [CCA Settings] and press s [Choose]. The CCA Settings
screen displays.
3
Press s [Page Up] and s [Page Down] to view all the CCA and
infuser settings.
4
When nished, press s [Previous Screen] to return to the Settings/
Vols/CCA screen. Press s [Previous Screen] again to return to the delivery screen.
17
of 36
Page 18
SET DISTAL PRESSURE ALARM LIMIT
NOTE: Prior to changing the Distal Pressure Alarm Limit, the infuser must be stopped.
1
On the Main Delivery screen, press s [Settings/Vols/CCA].
2
Change the Distal Pressure Alarm Limit.
CAUTION –––––––––––––––––––––––––––––––
DO NOT SET THE DISTAL PRESSURE ALARM LEVEL LOWER THAN 3 PSI (155 MMHG) OR HIGHER THAN 12 PSI (624 MMHG). SETTING THE ALARM OUTSIDE OF THAT RANGE MAY RESULT IN UNRELIABLE ALARM FUNCTIONING.
–––––––––––––––––––––––––––––––––––––––
3
Use the keypad to press s [Done] to save your changes and return
to the Settings/Vols/CCA screen or press s [Previous Screen] to view the settings without making changes.
4
Press s [Previous Screen] to return to the Main Delivery screen.
DISCONTINUE ELECTRONIC FLOW CONTROL AND SETTING GRAVITY FLOW
WARNING –––––––––––––––––––––––––––––––
CLOSE ALL CLAMPS BEFORE OPENING THE CASSETTE DOOR.
–––––––––––––––––––––––––––––––––––––––
18 of 36
1
Press [STOP]. If two lines were pumping, press s [Stop All].
2
Press s [ON/OFF] to turn off the infuser.
3
Close all clamps.
4
Open the cassette door and remove the cassette.
Page 19
5
If only 1 line was pumping, open all clamps.
If 2 lines were pumping, you must choose one line for gravity ow.
Open the clamps on that proximal line and on the distal line. Make sure one proximal line stays clamped.
6
Holding the cassette upright, set gravity ow by turning the ow
regulator counter-clockwise.
7
Check the drip chamber to measure the ow rate. Refer to the
administration set package for number of drops/mL.
DISCONTINUE FLUID ADMINISTRATION
1
Press [STOP]. If two lines are pumping, press [Stop All].
2
Press [ON/OFF] to turn off the infuser.
3
Close all clamps.
4
Detach the distal line from the patient access device.
5
Open the cassette door and remove the cassette.
6
Close the cassette door.
7
Discard the set and uid container per hospital policy/procedure.
ALARMS AND TROUBLESHOOTING
The Plum 360 infuser has an intelligent alarm system that handles more than one alarm at a time. Alarms are prioritized as high, medium, or low. You can distinguish the priority by the number of beeps:
Priority Number of Beeps
High 10
Medium 3
Low 2
19
of 36
Page 20
HIGH PRIORITY ALARMS
Alarm Message and Priority
Power Off then On. Replace pump if alarm continues.
High
Replace pump. Audio alarm failure.
High
Replace pump. Backlight failure.
High
Line B VTBI complete in prior CCA! Press STOP key.
High
Distal OCCLUSION! Resolve then Backprime.
High
Possible Cause Corrective Action
Malfunction. Power the infuser off,
and then on. Replace infuser if this does not clear the alarm.
Audio alarm is OFF but sensed ON, or ON but sensed OFF.
Backlight voltage out of range during operation.
Line B delivery VTBI=0; the line was programmed under a different CCA than the CCA currently used; and Line B has a Concurrent delivery programmed.
Distal occlusion detected while attempting to backprime or during cassette check.
Power off the infuser.
Power off the infuser.
Stop the delivery on Line B or open the cassette door.
Examine the distalline for kinks. Resolve the distal occlusion and then either backprime or open and close the cassette door.
20 of 36
Page 21
Distal OCCLUSION! Check IV line and site.
High
A distal occlusion is detected and either the maximum auto-restarts have occurred for the infusion or auto­restart was set to zero.
Examine the distal line for kinks and correct any found. Restart the delivery.
NOTE: The alarm can also be cleared by clearing the conrmed programs on any programmed line or opening the cassette door.
See section in SOM on
Avoiding a Bolus While Resolving a Distal Occlusion.
WARNING –––––––––––––––––––––––––––––
THE PLUM 360 INFUSER DOES NOT HAVE CAPABILITY TO DETECT INFILTRATION TO THE PATIENT
–––––––––––––––––––––––––––––––––––––––
Pump too high above patient. Lower pump or replace set.
High
Distal occlusion detected during delivery due to too much backpressure.
Resolve the occlusion by lowering the infuser on the pole to place it closer to the level of the patient’s heart (see section on Delivery
Accuracy in SOM) and
then press
NOTE: The alarm can also be cleared by clearing the conrmed programs on any programmed line or opening the cassette door.
[START].
Continued on page 22
21
of 36
Page 22
Data was cleared. Re-enter all programming.
High
Distal AIR! Disconnect/reprime. Press START.
High
Corruption of retained delivery parameters is detected. Autoclear of SEEP requires fresh delivery setup.
The single air bolus or the cumulative air detected at the distal sensor exceeds the air detection threshold.
Acknowledge the alarm.
Open the cassette door. Clear programs. See section in SOM on Resolving a Distal
Air-In-Line Alarm.
Door opened! Infusion stopped! Close Door.
High
Cassette test failure! Check set.
High
Depleted Battery! Plug into AC now!
High
Keypad locked. Enter code to disable.
High
The cassette door was opened during a delivery.
Faulty cassette, proximal or distal occlusion, or air was detected in the cassette during the cassette test.
The infuser is running on battery power and the battery voltage is below the depleted battery threshold.
A 3 minute shut down sequence begins.
While the keypad was locked, someone pressed
[STOP] or opened the cassette door during delivery.
Close the cassette door with the cassette inserted.
Resolve occlusion and then open and close the cassette door. Press s [Backprime]. Replace administration set.
Plug into an AC (mains) power source.
After plugging the device in, the pump will automatically reboot.
Enter valid keypad unlock code.
22 of 36
Page 23
MEDIUM PRIORITY ALARMS
Alarm Message and Priority
Door opened! Delayed Start! Close Door.
Medium
Distal OCCLUSION ­Paused! Attempting restart.
Medium
Possible Cause Corrective Action
The cassette door was opened while an infusion was in
Close the cassette door with cassette inserted.
Delayed Start.
A distal occlusion was detected and the maximum number of autoresets is congured but has not occurred for the infusion.
Examine the distal line for kinks and correct any found. No action is necessary if the patient can resolve the alarm condition within 60 seconds of activation (for example, moving an arm to eliminate the occlusion) before the maximum retry number is reached. Open the cassette door.
Continued on page 24
23
of 36
Page 24
LOW PRIORITY ALARMS
Alarm Message and Priority
Keep Plugged into AC! Service battery/replace pump.
Low
Power Off then On. Replace pump if alarm continues.
Low
Low Battery! Plug into AC power!
Low
Programming not complete! Action required!
Low
Possible Cause Corrective Action
The battery or battery charge circuitry needs servicing.
Battery voltage is greater than the expected limit.
The battery charge level is low.
No operator input for two minutes after the infuser is powered on in Clinical mode, except for situations that trigger an alarm.
Power off the infuser. Replace the infuser as soon as possible, so that it can be sent for repair.
Power off the infuser.
Plug into AC (mains) power.
Press any hardkey or softkey except the hardkey that silences alarms.
24 of 36
Page 25
LINE A – HIGH PRIORITY ALARMS
Alarm Message and Priority
No Action Alarm! Start or Clear Line A.
High
Line A VTBI complete in prior CCA! Press STOP key.
High
VTBI Completed Line A! Add more VTBI or Clear A.
High
Possible Cause Corrective Action
No operator action for 2 minutes when Line A has been stopped by the user and is not cleared or restarted.
Press any key on the infuser except
[AUDIO PAUSED]. Select Line A to program or clear it.
NOTE: Will reassert if the condition persists.
Line A was programmed under a different CCA than
Stop the delivery on Line A. Open cassette door.
is currently being used and delivery is complete.
Line A delivery is complete and line was programmed under the
Add VTBI on Line A. Stop Line A. Open cassette door.
CCA that is currently being used.
Continued on page 26
25
of 36
Page 26
Proximal OCCLUSION A! Check Line A.
High
Proximal occlusion or air detected on Line A during delivery.
Examine the proximal line for kinks and correct any found. If the occlusion is caused by a closed clamp, open the clamp. If all clamps are open, the alarm may be caused by excessive air that is creating backpressure in the cassette. To remove air, see section in SOM on
Backpriming. Check
the syringe size. See section in SOM on
Administration Sets and Accessories Guidelines. Restart
Line A.
NOTE: The alarm can also be cleared by clearing the conrmed programs on any programmed line or opening the cassette door.
Proximal AIR Line A! Backprime.
High
26 of 36
The single air bolus detected at the proximal sensor in Line A exceeds the air detection threshold.
Press s [Backprime]. See section in SOM on Backpriming. Check the syringe size. See section in SOM on Administration
Sets and Accessories Guidelines.
Page 27
LINE A – MEDIUM PRIORITY ALARMS
Alarm Message and Priority
Callback to Line A! Silence audio to clear.
Medium
Proximal OCCLUSION A! Resolve then Backprime.
Medium
Bolus complete on Line A!
Medium
(SW 15.1 only)
Possible Cause Corrective Action
A Callback Alarm was programmed for Line
Press
[AUDIO PAUSED]. A, and the VTBI for Line A reaches 0 for a Loading Dose or any step in a multistep therapy except the last step.
Proximal occlusion was detected on Line A during backprime.
Examine Line A for kinks. Resolve the occlusion. Either backprime or open and close the cassette door. Check the syringe size. See section in SOM on
Administration Sets and Accessories Guidelines.
Bolus delivery completes on Line A
Press
[AUDIO PAUSED]. and a Nurse Callback was confgured.
Continued on page 28
27
of 36
Page 28
Proximal OCCLUSION A Startup! Open/close door or Backprime.
Medium
A proximal occlusion was detected on Line A during the cassette integrity test.
LINE B – HIGH PRIORITY ALARMS
Examine Line A for kinks. Resolve the occlusion. Either backprime or open and close the cassette door. See section in SOM on Opening
the Cassette Door Completely. Check
syringe size. See section in SOM on
Administration Sets and Accessories Guidelines.
Alarm Message and Priority
No Action Alarm! Start or Clear Line B.
High
Line B VTBI complete in prior CCA! Clear Line B.
High
28 of 36
Possible Cause Corrective Action
No operator action for 2 minutes when Line B has been stopped by the user and is not cleared or restarted. NOTE: Will reassert if the condition persists.
A Line B Piggyback delivery that was programmed under a different CCA is complete, and no delivery is programmed on Line A.
Press any key on the infuser except
[AUDIO PAUSED].
Select Line B to program or clear it.
Clear the delivery on Line B or open the cassette door.
Page 29
VTBI Completed Line B! Add more VTBI or Clear B.
High
VTBI Completed Line B! Add more VTBI or Clear B.
High
A Piggyback delivery on Line B is complete and the line was programmed under the current CCA and no delivery is programmed to deliver on Line A. NOTE: Piggyback with a Line A delivery will not alarm; it will just transition to Line A delivery.
A Concurrent delivery on Line B is complete and the Line was programmed under the current CCA.
Add VTBI on Line B. Clear program on Line B. Open the cassette door.
Add VTBI on Line B. Stop delivery on Line B. Open the cassette door.
Proximal OCCLUSION B! Resolve then backprime.
High
Continued on page 30
Proximal occlusion detected on Line B during backprime.
Examine Line B for kinks. Make sure a line or syringe is attached to the secondary port and that the line is unclamped or the syringe has enough free space to accept the backprimed uid. Either backprime or open and close the cassette door. See section in SOM on Backpriming to remove the air. Check the syringe size. See section in SOM on
Administration Sets and Accessories Guidelines.
29
of 36
Page 30
Proximal OCCLUSION B! Check Line B.
High
Proximal occlusion detected on Line B during delivery.
Examine the proximal line for kinks and correct any found.
Restart Line B.
NOTE: The alarm can also be cleared by clearing the conrmed program or opening the cassette door.
Proximal AIR Line B! Backprime.
High
The single air bolus detected at the proximal sensor in Line B exceeds the air detection threshold.
LINE B – MEDIUM PRIORITY ALARMS
Alarm Message and Priority
Callback to Line B! Silence audio to clear.
Medium
Possible Cause Corrective Action
A Callback Alarm was programmed for Line B which is in Piggyback mode, Line A is programmed to resume when Line B completes, and the VTBI for Line B reaches 0 for a Piggyback therapy,
Press s [Backprime]. See section in SOM on
Backpriming. Check
the syringe size. See section in SOM on
Administration Sets and Accessories Guidelines.
Press
[AUDIO PAUSED].
30 of 36
Page 31
Callback to Line B! Silence audio to clear.
Medium
Loading Dose, Maintenance Dose or any step in a multistep therapy.
OR
A Callback Alarm was programed for Line B, which is in Piggyback mode, Line A is not programmed to resume when Line B completes, and the VTBI for Line B reaches 0 for Loading Dose or any step in a multistep therapy except the last step.
OR
A Callback Alarm was programed for Line B, which is in Concurrent mode, and the VTBI for Line B reaches 0 for a Loading Dose or any step in a multistep therapy except the last step.
Press
[AUDIO PAUSED].
Continued on page 32
31
of 36
Page 32
Proximal OCCLUSION B. Resolve then Backprime.
Medium
Proximal occlusion detected on Line B during cassette integrity test.
Examine Line B for kinks. Make sure a line or syringe is attached to the secondary port and that the line is unclamped or the syringe has enough free space to accept the backprimed uid. Either backprime or open and close the cassette door. Check the syringe size. See section in SOM on
Administration Sets and Accessories Guidelines.
Bolus complete on Line B!
Silence audio to clear.
Medium
(SW 15.1 only)
Bolus delivery completes on LIne B and a Nurse Callback was congured.
LINE A AND B – HIGH PRIORITY ALARMS
Alarm Message and Priority
No Action Alarm! Acknowledge Alert.
High
32 of 36
Possible Cause Corrective Action
Rate was recalculated; operator has not acknowledged the alert within 30 seconds.
Press
[AUDIO PAUSED].
Press a labeled softkey.
Page 33
Line not in STANDBY! Choose line(s) to Standby:
High
No operator action for 15 seconds when the user has selected s [Standby] when both lines are able to be put in standby, but has not selected a line (A, B or A & B) or selected s [Cancel] to complete the action.
Press
s [Standby All], s [Standby A], s [Standby B],
or s [Cancel].
Delivery was not STOPPED! Choose line(s) to stop.
High
Delivery was not STARTED! Choose line(s) to start.
High
No operator action for 15 seconds when the user has attempted to stop a delivery when both lines are deliv­ering by pressing [STOP], but has not selected a line (A, B or A & B) or selected s [Cancel] to complete the action.
No operator action for 15 seconds when the user has attempted to start a delivery when both lines are conrmed by pressing
[START], but has not selected a line (A, B or A & B) or selected s [Cancel] to complete the action.
Press
s [Stop All], s [Stop A], s [Stop B],
ors [Cancel].
Press
s [Start All], s [Start A], s [Start B], ors [Cancel].
Continued on page 34
33
of 36
Page 34
Yes: Start titration. No: Edit
High
Standby is not possible and no operator action for 30 seconds when a titrated program is waiting to be conrmed.
Press s [Yes] to conrm the program or s [No] to go back to the Program screen.
No Action Alarm!
Yes: Start No: Edit
High
NOTE: Standby
is intentionally not included in the instruction text.
Yes: Start titration! No: Edit
High
NOTE: Standby
is intentionally not included in the instruction text.
No Action Alarm! START: Conrm titration.
High
Standby is possible and no operator action for 2 minutes when a new program is waiting to be conrmed or placed into Standby.
Standby is possible and no operator action for 30 seconds when a titrated program is waiting to be conrmed or placed into Standby.
No operator action for the 30 seconds when a line is titrated during infusion and the
[START] hardkey has not been pressed for a program that can be started.
NOTE: A delivery cannot be started if it is in a concurrency violation. If it is in a concurrency violation, an alarm will occur.
Press s [Yes] to conrm the program or s [No] to go back to the Program screen, or press s [Standby].
Press s [Yes] to conrm the program or s [No] to go back to the Program screen, or press s [Standby].
Press
[START]. Press s [Return to A/B].
34 of 36
Page 35
No Action Alarm! Yes: Override. No: Edit
High
Soft limit override and no operator action for 2 minutes when a new program is waiting to be conrmed.
OR
Soft limit override and no operator action for 30 seconds when a titrated program is waiting to be conrmed.
Press s [Yes] to conrm the program or s [No] to go back to the Program screen.
No Action Alarm! Start or clear lines.
High
No Action Alarm! Yes: Start No: Edit
High
Continued on page 36
No operator action for 2 minutes when both lines have been stopped by the user and not cleared or restarted.
OR
No operator action for 2 minutes after a Bolus has been cancelled (putting both lines into a stopped state).
NOTE: Will reassert if the condition persists.
Standby is not possible and there is no operator action for 2 minutes when a new program is waiting to be conrmed.
Press any hardkey except
[AUDIO PAUSED].
Press s [Yes] or s [No].
35
of 36
Page 36
Alarm Message and Priority
Proximal AIR! Backprime.
High
Possible Cause Corrective Action
The cumulative air de­tected at the proximal sensors in Line A and Line B exceeds the air detection threshold.
Press s [Backprime]. See section in SOM on
Backpriming.
Check the syringe size.
See section in SOM on
Administration Sets and Accessories Guidelines.
Reference Used: 430-98340-002 (A, 2017-03) Plum 360 System Operating Manual © 2017 ICU Medical Inc. | www.icumed.com CS04-0015b-0317R2
Loading...