Hospira Gemstar Operations 6 2010 Precalculus

Page 1
Page 2
Hospira GemStar® Infusion System
System Operating Manual
For use with the
Hospira GemStar® Infusion System
List 13086-04 List 13087-04 List 13088-04
PLEASE
Read this entire manual before
using the Hospira GemStar® Pump
This manual is designed for use by healthcare professionals, caregivers, and patients. The Hospira Technical Support Operations hotline is available 24 hours a day (in the USA) to provide consultation and technical assistance regarding the Hospira
®
GemStar
Pump.
In the USA: Hospira Technical Support Operations
1-800-241-4002
Outside the USA: Contact your local Hospira
sales office
To order additional copies of this manual
(List 13086,13087,13088) call Hospira Customer Care:
1-877-9HOSPIRA
(1-877-946-7747)
Hospira GemStar® System Operating Manual i
Page 3
Change History
Part Number Description of Change Pages Changed
430-96527-001 (Rev. 03/06)
430-96527-002 (Rev. 01/07)
430-96527-003 (A, 08/08)
Original Issue N/A
Second Release All
Third Release All
ii Hospira GemStar® System Operating Manual
Page 4
Content s
1. Introduction 1
GemStar® Pump Layout . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
GemStar
Therapies. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
GemStar
Menu System. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Operating Modes. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Indications for Use. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Contraindications for Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Warnings and Cautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
®
Pump Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
®
Pump Features. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
2. Basic Operation 19
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Selecting a Power Source . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Powering On the Pump . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Accessing the Programming Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Programming the Pump. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Starting a New Container . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
GemStar
Loading the Cassette. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Releasing the Cassette . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Priming a GemStar
®
Pump Set . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
®
Pump Set. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
3. Total Parenteral Nutrition (TPN)
- Not Applicable to List 13088 53
Therapy Features and Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
TPN Programming. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
Changing a TPN Program. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
TPN Auto Taper Down . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
TPN Programming Worksheet. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
4. Pain Management
- Not Applicable to List 13086 59
Therapy Features and Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
Pain Management Programming. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
Hospira GemStar® System Operating Manual iii
Page 5
Contents
Bolus Delivery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
Loading Dose Delivery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
Changing a Pain Management Program. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
5. Intermittent
- Not Applicable to List 13088 69
Therapy Features and Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
Intermittent Programming . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70
Intermittent Delivery Interruptions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72
Changing an Intermittent Program . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
Intermittent Programming Worksheet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
6. Continuous
- Not Applicable to List 13088 77
Therapy Features and Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77
Continuous Programming . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78
Rate Titration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79
Piggybacking. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79
Changing a Continuous Program . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 82
Continuous Programming Worksheet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83
7. Weight Dosed
- Not Applicable to List 13088 85
Therapy Features and Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85
Weight Dosed Programming . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 86
Auto KVO. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 87
Rate Titration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 87
VTBI Titration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 88
Changing a Weight Dosed Program . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 88
GemStar Programming a Precedex GemStar
Worksheet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 91
®
Weight Dosed Programming Worksheet . . . . . . . . . . . . . . . . . . . . . . . 89
®
Precedex® (dexmedetomidine hydrochloride injection) Programming
®
(dexmedetomidine hydrochloride injection) Delivery. 90
8. mL/hr Only
- Not Applicable to List 13088 93
Therapy Features and Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 93
mL/hr Only Programming . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 94
iv Hospira GemStar® System Operating Manual
Page 6
Contents
Auto KVO. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 95
Rate Titration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 95
VTBI Titration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 96
Piggybacking. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 96
Changing a mL/hr Only Program. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 99
mL/hr Only Programming Worksheet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100
9. Variable Time
- Not Applicable to List 13088 101
Therapy Features and Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 101
Variable Time Programming . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 102
Variable Time Delivery Interruptions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 103
Changing a Variable Time Program . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 104
GemStar
10. Options Menu 107
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 107
Tips for Using the Options Menu. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 107
Review Program . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 109
Histories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 109
Keypad Lock . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 114
Change Pump Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 116
Set Clock . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 118
Print. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 119
Speed Protocol . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 120
Next Dose (Not available for List 13088). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 121
Connecting to the Data Port . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 122
Transferring Data Using a Computer. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 127
®
Variable Time Programming Worksheet . . . . . . . . . . . . . . . . . . . . . . 105
11. Clinician Instructions 129
Clinician Locking Sequence Instructions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 130
Program and Deliver a Clinician Activated Loading Dose. . . . . . . . . . . . . . . . . . 132
Shift Totals and History . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 133
12. Troubleshooting 135
Alarm Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 136
Alert Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 145
Other Display Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 149
Hospira GemStar® System Operating Manual v
Page 7
Contents
13. Maintenance 151
Pump Storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 151
Cleaning and Disinfecting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 151
Repair . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 153
14. Operation Test 155
Equipment Required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 155
Test Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 156
Performing the Operation Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 156
Printing Test Results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 161
GemStar
Operation Test Checklist . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 163
15. Optional System Components 165
®
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 163
Connecting the Bolus Cord . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 168
Using the GemStar
Using the Hospira Carrying Cases . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 171
Using the GemStar
®
Lockboxes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 168
®
Docking Station . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 177
16. Pump Sets 179
17. GemStar® Pump Specifications 183
Occlusion Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 186
18. Warranty/Technical Assistance 199
A. Electrical Potentials 201
Guidance on EMC Compatibility . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 201
Information Regarding Electrostatic Discharge (ESD) . . . . . . . . . . . . . . . . . . . . 202
Index 203
IEC Symbols inside back cover
vi Hospira GemStar® System Operating Manual
Page 8
Chapter 1

Introduction

The GemStar® Pump is a small and lightweight, single-channel infusion device designed for use in the home, in the hospital, or anywhere electronic infusion is
®
required. The GemStar battery pack, docking station, or two disposable AA alkaline batteries. When powered
by batteries, the GemStar
The following conventions are used throughout this manual:
Convention Use Example
[ALL CAPS] Pump keys Press [YES/ENTER] to continue.
ALL CAPS
WARNING: A WARNING message contains special safety
information and must be observed at all times. Failure to observe a warning message could be life threatening.
Pump can be powered by AC mains adaptor, rechargeable
®
Pump is ideal for ambulatory patients.
Display messages The pump displays:
ENTER TO CONFIRM.
CAUTION: A CAUTION contains information that could prevent
irreversible product damage or hardware failure. Failure to observe a caution could result in serious patient or user injury.
  
Note: A NOTE contains information or examples to help explain instructions.
Tip: A TIP contains information that makes the pump easier to use.
Note: Examples are provided only to help explain a pump function and may not be clinically relevant. Graphics are for illustrative purposes only and may not reflect the actual product.
Hospira GemStar® System Operating Manual 1
Page 9
Chapter 1. Introduction

GemStar® Pump Layout

Shown approximate size
2 Hospira GemStar® System Operating Manual
Page 10
Chapter 1. Introduction
DISPLAY
1
2
3 Stops infusion.
4 Starts infusion.
5 Lights continuously when the pump is powered by AC mains.
6 Powers the pump on or off. To power-off the pump, press and hold until the display
7 Returns to previous steps during programming.
8
9
10 Toggles between AM and PM in time-based entries.
11 Responds “YES” to display questions.
Displays the pump’s status.
Silences an alarm.
This silences an alert for five minutes and an alarm for two minutes.
Flashes when the pump is powered by external batteries (e.g., battery pack). Not lit when using disposable AA batteries.
clears.
Exits Help, Change, and Options menus.
Displays the following menu items from the
New Container Change Program (or Titrate VTBI) New Program Piggybacking
Resets a numeric value to zero.
Displays the following Options Menu items from the
1 Review Program 4 Pump Settings 7 Speed Protocol 2 Histories 5 Set Clock 8 Next Dose 3 Keypad Lock 6 Print
Scrolls through program review and help messages.
Toggles between AM and PM in time-based entries.
DOWN arrow toggles between units of measure when entering concentrations.
UP arrow enters a decimal point in numeric entries.
Scrolls back by one screen in program review.
Accepts entries and goes to the next step.
STOP or RUN mode*:
STOP or RUN mode*:
12 Responds “NO” to display questions.
13 Displays information during programming steps and alarm conditions.
Displays Shift Totals, except during programming steps and alarm conditions.
14
15 [
0] - [9] • Selects menu items and enters numeric values. If two numbers appear before a
* Not all menu items are available with all modes.
Hospira GemStar® System Operating Manual 3
Clears an
Primes a GemStar® Pump Set.
menu item (e.g., 02), press both number keys to select that item.
AIR-IN-LINE alarm.
Page 11
Chapter 1. Introduction

GemStar® Pump Components

Bolus Button Press this button to start a
1
bolus delivery if available in the current program.
Pole Clamp Track Use to connect to an optional
2
pole clamp.
Cassette Release
3
Button
Power Connector Use to connect the AC mains
4
Data Port
5
Connector
Bolus Connector Use to connect a bolus cord.
6
Tubing Channel Place the pump set tubing in
7
Cassette Pocket Load the cassette into this
8
I.V. Tubing Guide Use to hold the I.V. tubing.
9
Battery Pack
10
Connector
Battery
11
Compartment
Push down on this button to release the cassette.
adaptor.
Use to connect the GemStar® serial cable.
this channel.
pocket.
Use to connect the GemStar® Rechargeable Battery Pack.
Install two fresh, disposable AA batteries in this compartment.
Top View
Bottom/Side View
4 Hospira GemStar® System Operating Manual
Page 12
Chapter 1. Introduction

Therapies

Seven types of therapy are available for infusion of medications and fluids. The following table lists possible applications and available features of each therapy.
Therapy Possible Applications
TPN Total Parenteral Nutrition
fluid delivery
Pain Management
Infusion of analgesics or anesthetics for patient­controlled pain management (PCA), epidural infusion and parenteral administration of general I.V. fluids
Intermittent Interval-based therapies
such as antibiotics
Continuous Infusion of mg, mcg, or
mL/hr with a selectable KVO (Keep Vein Open) option
Weight Dosed Critical care, pediatrics
mL/hr Only Simple rate programming
Variable Time Chronotherapy
= Available
O
E
V
K
A
R
O
T
T
I
U
T
R
A
G
T
-
N
S
Y
I
U
G
E
K
G
T
C
I
A
A
P
B
A
L
O
O
L
B
P
G
N
I
D
E
P
S
A
O
T
D
E
U
T
A
R
N
R
E
I
W
T
B
O
I
T
D
T
V
/
••
••
••
••
•••
••
L
A
N
O
I
T
O
P
V
O
K
Note: pain management is the only therapy available for List 13088.
Hospira GemStar® System Operating Manual 5
Page 13
Chapter 1. Introduction
GemStar® Pump Features
Feature Description Reference Page
Adjustable Settings
Alerts and Alarms
•Audible Keypad Response
Keypad Lock
•New Container
Piggybacking
•Power-on Self-test
Operation Te st
Program History
Speed Protocols
Adjusts the distal occlusion pressure limits, air sensor sensitivity, real-time clock, and audible alarm volume.
Signals audible and visual alarms when attention is required.
Sounds a single beep when any key is pressed. If an invalid key is pressed, the pump sounds a flutter tone.
Restricts access to various options. Four lock levels are available.
Repeats the current program without reprogramming the pump.
Allows delivery from a secondary container during a Continuous or mL/hr Only program.
Performs a self-test each time the pump is powered on.
Performs user-assisted tests to verify the pump is operating properly.
Maintains a history event log, with a date and time stamp of each event for each program. Review the history on the display, or download it to a printer or computer.
Stores up to nine frequently used programs in the pump’s memory for quick access.
Chapter 10 Options Menu
Chapter 1 Introduction,
GemStar® Pump Layout
Chapter 1 Introduction,
GemStar® Pump Layout
Chapter 10 Options Menu and Chapter 11 Clinician Instructions
Chapter 2 Basic Operation
Chapter 6 Continuous ­Not Applicable to List 13088 and Chapter 8 mL/hr Only - Not Applicable to List 13088
Chapter 2 Basic Operation
Chapter 14 Operation Te st
Chapter 10 Options Menu
Chapter 10 Options Menu
6 Hospira GemStar® System Operating Manual
Page 14
Chapter 1. Introduction
Menu System
The GemStar® Pump uses a menu system that provides step-by-step instructions through all of the pump’s functions. Refer to the following table for a list of functions in each menu.
Use the keypad to select menu items, respond to display messages, and enter numeric values. For example, press [ Menu.
2] to select PAIN MANAGEMENT from the Programming
Note: Menu items may vary depending on the pump’s configuration.
Tip: A menu item does not have to be on the display to be selected. For example, when the Programming Menu first displays, only the first four therapies are shown. Instead of pressing the down arrow to see the other menu items, press [7] to go directly to Variable Time Programming.
Hospira GemStar® System Operating Manual 7
Page 15
Chapter 1. Introduction
To display this menu:
Programming Menu
Or
Delivery Mode
(FOR LIST
Menu
13088 ONLY)
Do this: Menu Items Available
1 Press to
display the Change menu.
2 Select
NEW
PROGRAM
.
1 TPN 2 PAIN MANAGEMENT 3 INTERMITTENT 4 CONTINUOUS 5 WEIGHT DOSED 6
mL/HR ONLY
7 VARIABLE TIME
1 CONTINUOUS 2 BOLUS 3 CONT + BOLUS
Note: If a program is not currently stored in pump memory at power-on, the Programming Menu automatically displays after the self-test.
Change Menu Press . 1 NEW CONTAINER
2 NEW PROGRAM 3 CHANGE PROGRAM 4 PIGGYBACKING
Note: If the current therapy is Weight Dosed or
mL/hr Only and the pump is in
3 TITRATE VTBI
will replace
3 CHANGE PROGRAM
on the menu display.
RUN mode:
Note: Piggybacking is available only in mL/hr Only and Continuous therapies.
Note: Menu items change depending on when Change Menu is viewed.
Options Menu Press . 1 REVIEW PROGRAM
8 Hospira GemStar® System Operating Manual
2 HISTORIES 3 KEYPAD LOCK 4 PUMP SETTINGS 5 SET CLOCK 6 PRINT 7 SPEED PROTOCOL 8 NEXT DOSE (NOT AVAILABLE FOR LIST 13088)
Page 16
Chapter 1. Introduction
For more information on: Refer to the following pages:
Programming Menu Chapter 2 Basic Operation, page 19.
Change Menu For menu item [
in Chapter 2 Basic Operation, on page 32. For menu items [ therapy chapter.
Options Menu Chapter 10 Options Menu, page 107.
1], refer to Starting a New Container
2], [3], and [4], refer to the appropriate

Operating Modes

The GemStar® Pump has four operating modes:
Mode Description
Help Mode The pump enters the
programming or an alarm condition. The additional information about the current condition.
Programming Mode The pump enters the
powered on or when is pressed.
Run Mode After is pressed, the pump enters the
when not infusing.
Stop Mode When programming is complete, the pump enters the
mode until is pressed.
When the programmed therapy is complete, and during certain alarm conditions, the pump automatically enters the
Press to enter the CHECK CASSETTE alarm.
HELP mode when is pressed during
HELP mode displays
PROGRAMMING mode after the pump is
RUN mode, even
STOP mode.
STOP mode at any time, except during a
STOP
Hospira GemStar® System Operating Manual 9
Page 17
Chapter 1. Introduction

Indications for Use

Physicians or certified, licensed healthcare professionals should always oversee infusions. Healthcare professionals should always supervise pump users and instruct users in pump operation and troubleshooting. Instruction should include how to prevent intravenous (I.V.) related complications, including how to prevent accidental infusion of air.
The pump is suitable for intravenous, arterial, subcutaneous, short-term epidural infusion and parenteral administration of general I.V. fluids, medications, nutritional fluids, and blood/blood products.
Intended parenteral administrations include but are not limited to: intravenous (within a vein), arterial, subcutaneous (tissue just below the skin), regional (located to specific areas of the body) and epidural analgesia, intrathecal analgesia, intraspinal analgesia, combined spinal-epidural analgesia, short term epidural infusion, intraspinal (within the spinal canal), intrathecal (within the subarachnoid or subdural space), intraoperative sites (soft tissue, body cavity or surgical wounds), intraarticular (within a joint), intraperitoneal (within the peritoneal cavity), and intracerebroventricular infusions. The indications for use include hospital, homecare and ambulatory environments. The system has also been qualified for use in fixed and rotary wing aircrafts (e.g., helicopters). The pump must be used with sterile, dedicated, GemStar® administration sets.
Note: Epidural administration of anesthetics is recommended in the continuous
mode only. Epidural administration of analgesics may be delivered by continuous, bolus, or continuous with bolus.
For epidural use, the administration of drugs is restricted to those anesthetic and analgesic drugs approved for continuous epidural administration: e.g., Chloroprocaine Hydrochloride USP, Lidocaine Hydrochloride USP, and Morphine Sulfate Injection USP (Preservative Free).
WARNING: Delayed respiratory depression following continuous
epidural administration of preservative-free morphine sulfate has been reported.
10 Hospira GemStar® System Operating Manual
Page 18
For epidural administration, the following is recommended:
Chapter 1. Introduction
Nylon or Teflon Pump sets without Y-sites Stickers for the tubing indicating ongoing administration
®
catheter
Note: Facilities practicing epidural anesthesia/analgesia must be staffed and
equipped to manage cardio-pulmonary resuscitation (CPR). Supplies should include oxygen, naloxone, and other appropriate resuscitative drugs and equipment. Monitor patients continuously (e.g., oximetry) during epidural administration. Frequently observe patient for side effects (for up to 24 hours) following completion of drug administration by the epidural route.

Contraindications for Use

Persons who do not have the mental and physical capability or the emotional stability to properly use the GemStar
DO NOT use the following with the GemStar
Drugs not compatible with silicone rubber or PVC plastic Drugs not stable under infusion conditions

Warnings and Cautions

Pay attention to the following warnings and cautions when operating the GemStar Pump. These warnings and cautions are repeated in the manual when relevant to the procedure being discussed. Pay attention to all alert messages.
®
Pump should not operate this pump.
®
Pump:
®

General Warnings and Cautions

Federal (USA) law restricts this device to sale by or on the order of a physician or
other licensed practitioner.
Manual references to specific values are approximate only, unless indicated
otherwise. Air-in-line sensitivity values are approximate only.
When a proximal occlusion alarm activates, DO NOT power-off the pump. Closely supervise patients who are likely to be adversely affected by unintended
operations and failures, including interrupted medication or fluid infusion. Ensure provisions for immediate corrective action are available.
Hospira GemStar® System Operating Manual 11
Page 19
Chapter 1. Introduction
Regarding Drugs Used, Pump Sets, and Containers
USE ONLY GemStar
sets may result in injury to the patient or damage to the pump.
Some microbore tubing delivery accuracy may be affected when used at extreme
operating temperatures (<10°C, >35°C), relative humidity (<20%) and delivery rates (<4.9 mL/hr, >350 mL/hr). Refer to individual set labels for specific instructions and operating conditions.
Delivery accuracy may be impacted when there is air present in the cassette.
Follow the instructions in the individual set labels to remove any visible air from the cassette.
Ensure slide clamp or lower CAIR
cassette.
When using pre-filled vials, ensure vial and injector are securely locked in
lockbox before beginning delivery.
A cracked vial might not show evidence of leakage until delivery pressure is
applied.
When using nonvented syringes, the system may underdeliver. System delivery
accuracy could be reduced by as much as 5%.
When using PCA vials (except in a lockbox with a spring assist mechanism), the
system may underdeliver. System delivery accuracy could be reduced by as much as 5%.
When using viscous fluids, the system may underdeliver. System delivery
accuracy could be reduced by as much as 5%.
Variations in filling head (such as container height) may affect delivery accuracy
by no more than 5%.
When using the GemStar
ONLY GemStar® Pump Sets designated for use with the GemStar Lockbox. Failure to use designated sets could result in unrestricted flow.
NEVER use drugs that are incompatible with silicone rubber or PVC plastic,
(unless these items are specially treated and labeled to be compatible with such drugs).
To reduce the loss of potency for drugs known to be absorbed by PVC plastic and
silicone, begin infusion as soon as practical after priming the pump set. Use of high flow rates during infusion will minimize drug absorption.
DO NOT use medications that are unstable under infusion conditions. ALWAYS use connections with Luer Lock™ fittings.
®
Pump Sets with the GemStar® Pump. Use of unauthorized
®
(roller) clamp is open prior to inserting
®
Spring Assist Mechanism (SAM) Lockbox, USE
®
SAM
12 Hospira GemStar® System Operating Manual
Page 20
Chapter 1. Introduction
To prevent contamination, use aseptic technique with all fluid-path connections.
Remove the protective coverings as assembly progresses.
ALWAYS place the flow stop in the closed position when the set is not in the pump.
Before installing the cassette in the pump, ensure all slide or lower CAIR
clamps are open. Failure to open the slide or lower CAIR
®
(roller) clamps
BEFORE installing the cassette may result in restricted flow.
DO NOT back-prime the pump set using a syringe. Back priming may result in
restricted flow.
Ensure the cassette is properly installed. When properly installed, all four cassette
latches are visibly holding the cassette securely in the cassette pocket.
Ensure the tubing is behind the tubing guide (at the distal end). Arrange tubing, cords, and cables to minimize the risk of patient strangulation or
entanglement.
NEVER use vented fluid containers (e.g., glass or rigid plastic) unless suspended
from a pole. When using a vented drip set, ALWAYS use an air-eliminating filter OR set the air sensitivity to either ON or 2 mL.
Stop infusion if signs or symptoms of infiltration occur. GemStar® Pump Sets contain a pressure-activated anti-siphon valve to reduce the
potential for gravity flow at a typical 36-inch head height.
Failure to use the pressure-activated anti-siphon valve may result in unrestricted
flow if the cassette is not properly installed in the pump.
Always close the slide clamps before removing the cassette from the pump. When using the pump for secondary delivery (piggybacking), ensure the fluids
being infused are chemically and physically compatible.
When infusing short-half-life drugs (i.e. those dosed with units of mcg/kg/min) at
very low rates (0.8 mL/hr or less) the rate may be too slow for that drug. Consider using a lower drug concentration with these drugs so a pump rate above 0.8 mL/hr can be used.
®
(roller)

Regarding Air-In-Line and Infusion

To reduce the risk of infusing air, use an air-eliminating filter when the air
sensitivity is set to OFF.
ALWAYS prime the pump set to remove all air from the cassette, tubing, and
injection site prior to connecting to the patient.
ALWAYS disconnect the pump set from the patient prior to priming or purging.
Hospira GemStar® System Operating Manual 13
Page 21
Chapter 1. Introduction
Regarding Pump Operation
If the pump does not perform as stated in this manual, remove the pump from
service IMMEDIATELY.
If the pump does not display the self-test or if beeps do not sound at power-on,
check all power connections and install two fresh, disposable AA batteries. If the problem continues, contact Hospira Technical Support Operations before using the pump.
Ensure a cassette is installed in the pump before connecting to a printer or
computer. If the pump is connected to a patient: BEFORE connecting the pump to a computer or printer, press [ cassette. Leave the pump set connected to the patient. Install another cassette in the pump before connecting to a printer or computer.
Disconnect the pump from the patient BEFORE performing the Operation Test. ALWAYS connect to a grounded AC mains power source when using AC mains
power. If the quality of the grounding source is in doubt, operate the pump only with batteries.
USE ONLY the AC mains adaptors specifically labeled for use with the
GemStar
USE ONLY the AC mains adaptor specifically labeled for use with the GemStar
®
Pump.
Pump to charge the battery pack.
The GemStar
GemStar
®
AC mains adaptor is for use with the GemStar® Pump or
®
Rechargeable Battery Pack only. DO NOT use the GemStar® AC
mains adaptor with other products.
During charging, the battery pack is warm. If the rechargeable battery pack
becomes hot to the touch, IMMEDIATELY unplug the AC mains adaptor and contact Hospira Technical Support Operations.
Use only the Li-Ion Docking Station and Li-Ion Battery Pack with the GemStar
Pump.
With the Li-Ion Battery Pack, the LED indicator flashes yellow/green when an
error condition has occurred or when the temperature is too high to charge the battery pack. In temperatures above 95°F (35°C), cycle AC power to confirm if battery is charged or if error condition exists. If LED flashes green/yellow, do not rely on external battery. Move device to cooler temperature and recharge prior to use. Refer to the battery pack insert for important information and precautions regarding activating and charging the Li-Ion Battery Pack.
When removing the cord (from the docking station, bottom of the battery pack, or
STOP], close the slide clamp, and eject the
®
®
14 Hospira GemStar® System Operating Manual
Page 22
Chapter 1. Introduction
bottom of the pump), firmly grasp the connector and pull straight out. DO NOT twist or bend the cord or connector.
To ensure proper pump operation, ALWAYS replace BOTH disposable batteries
with fresh, disposable AA batteries when a change is required.
ALWAYS fully close the battery door when using disposable batteries. DO NOT use rechargeable batteries in the battery compartment. Ensure two fresh, disposable AA batteries are installed for backup power when
using other external power sources (e.g., AC mains adaptor, docking station, battery pack). The pump will continue to operate on backup power if the external power source fails. In the event that the internal batteries are removed while the pump is operating, the power loss event may not be logged in the history, the programmed therapy may not be retained, and the alarms (audible and visual) may not occur.
ALWAYS avoid sources of high-intensity electromagnetic radiation (e.g., radio
transmitters, MRI scanners, microwave ovens, X-ray machines, and CAT scanners).
Use of radio-frequency emitting devices, such as cellular telephones, 2-way
radios, and Electrical Surgical Devices (ESU), in close proximity to this device may affect its operation.
Possible explosion hazard exists if the pump is used in the presence of flammable
anesthetics. NEVER use the pump in the presence of flammable or explosive vapors.
Non-hazardous, low-level electrical potentials are commonly observed when
fluids are administered using infusion devices. These potentials are well within accepted safety standards, but may create artifacts on voltage-sensing equipment, such as ECG, EMG, and EEG machines. These artifacts vary at a rate that is associated with the infusion rate. If the monitoring machine is not operating correctly or has loose or defective connections to its sensing electrodes, these artifacts may be accentuated so as to simulate actual physiological signals. To determine if the abnormality in the monitoring equipment is caused by the infusion device instead of some other source in the environment, set the infusion device so that it is temporarily not delivering fluid (if clinically acceptable). Disappearance of the abnormality indicates that it was probably caused by electronic noise generated by the infusion device. Proper setup and maintenance of the monitoring equipment should eliminate the artifact. Refer to the appropriate monitoring system documentation for setup and maintenance instructions.
The GemStar
in accordance with:
®
Pump has been tested and found to comply with EMC/EMI limits
Hospira GemStar® System Operating Manual 15
Page 23
Chapter 1. Introduction
IEC/EN 60601-1-2 (2001) when using two AA disposable batteries or when
connected to the GemStar
IEC/EN 60601-1-2 (2001) (ESD exemption) when using Li-Ion Docking
®
AC Adaptor
Station or Li-Ion Battery Pack
When connected to other power sources or optional system components in
®
commercial distribution before November 30, 2004, the GemStar
Pump complies with EMC/EMI limits in accordance with IEC/EN 60601-1-2 (1993). These limits are designed to provide reasonable protection against harmful interference in a typical medical installation. The equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to other devices in the vicinity. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference with other devices, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures:
Reorient or relocate the receiving device Increase the separation between the equipment Connect the equipment to an outlet on a circuit different from that to which
the other device is connected
Consult the manufacturer or field service technician for help The use of portable and mobile RF equipment may have an impact on this and
other pieces of medical equipment
Prior to using the GemStar
®
Pump and optional system components, refer to Appendix A for additional information and precautionary procedures regarding electrical potentials (electromagnetic compatibility and electrostatic discharge).

Regarding Handling and Maintenance

Use proper care during unpacking, installation, and operation of the pump. If the
pump is inadvertently mishandled, check connections and programmed data to ensure there is no damage. Refer to Chapter 14 Operation Test for further information.
Accessory equipment connected to the analog and digital interfaces must be
certified according to the respective IEC standards, (e.g., IEC 60950 for data processing equipment, and ICE 60601-1 for medical equipment). Furthermore, all configurations shall comply with the system standard IEC 60601-1-1. Any person who connects additional equipment to the signal input or output part configures a medical system, and is therefore responsible for ensuring that the system complies
16 Hospira GemStar® System Operating Manual
Page 24
Chapter 1. Introduction
with the requirements of IEC 60601-1-1. If in doubt, contact Hospira Technical Support Operations.
ALWAYS avoid dropping or hitting the pump. If the pump is dropped or hit,
ALWAYS check connections and programmed data.
NEVER use sharp objects (e.g., fingernails, pens, pencils, or other probes) to
program or clean the pump.
Use care not to damage the silicone seals around the sensor bodies in the cassette
pocket.
To avoid mechanical or electronic damage, NEVER submerge pump in water or
other fluids and avoid fluid spills. If the pump becomes wet, dry it immediately with a dry, lint-free cloth. Check connections and programmed data.
DO NOT touch connector ports. Connector ports may be susceptible to
Electrostatic Discharge (ESD) damage.
Some cleaning and sanitizing compounds may slowly degrade components made
from some plastic materials. DO NOT use compounds containing combinations of isopropyl alcohol and dimethyl benzyl ammonium chloride (i.e., alcohol wipes).
DO NOT sterilize by heat, steam, ethylene oxide (ETO), or radiation. Apply
disinfectants to the outside surface of the pump only. DO NOT use abrasive cleaners or materials on the pump. Using abrasive cleaners or cleaning solutions not recommended by Hospira may result in product damage.
Please comply with local disposal and recycling regulations as appropriate for
disposable batteries, rechargeable battery packs, medical electronic components, and pump sets.
Fixed and Rotary Wing Aircraft
Do not lay the IV bag flat because air can potentially be introduced into the line
causing an alarm condition. Recommend the use of extension set tubing be applied prior to patient transport.
The GemStar® pump has only been tested using battery power in fixed and rotary
wing aircraft.
Auditory alarms may not be heard in a noisy flight environment. Care providers
should rely on visual indicators on the pump display to determine if there are any alarms or other system malfunctions.
Hospira GemStar® System Operating Manual 17
Page 25
Notes
18 Hospira GemStar® System Operating Manual
Page 26
Chapter 2

Basic Operation

The following elements are required to use the GemStar® Pump:
6 GemStar 6 Appropriate power source
6 Appropriate GemS tar 6 Appropriate fluid container 6 Patient-access device

Overview

To start a therapy: 1 Determine the program settings. Refer to the appropriate therapy chapter for a
worksheet that may be copied and used to record program settings.
2 Select a power source. Refer to Selecting a Power Source on page 20. 3 Prime the pump set. Refer to GemStar® Pump Set on page 32.
®
Pump
®
Pump Set
4 Load the cassette into the pump. Refer to Loading the Cassette on page 35. 5 Power-on the pump and go to the Programming Menu. Refer to Powering On
the Pump on page 27.
6 Select the appropriate optional system components, such as a bolus cord. 7 Select the therapy type and enter a program. Refer to the appropriate therapy
chapter for detailed instructions.
d
Note: List 13088 only delivers therapy in the Pain management Mode.
8 Press [
Hospira GemStar® System Operating Manual 19
START] to begin delivery.
Page 27
Chapter 2. Basic Operation

Selecting a Power Source

The GemStar® Pump may be powered by one or more of the following:
Power Source Description Ambulatory Hospital Home
GemStar® AC Mains Adaptor (USA)
GemStar® AC Mains Adaptor (Universal)
GemStar® AC Mains Adaptor (Universal)
GemStar® Docking Station
GemStar® Battery Pack
Disposable Batteries Two disposable AA alkaline
Wall-mount adaptor plugs into a 1 10 VAC mains supply.
Wall-mount adaptor plugs into a 100-240 VAC mains supply.
Tabletop adaptor plugs into a 100–240 VAC mains supply.
Plugs into a 100–240 VAC Volt mains supply while mounted on an I.V. pole. The docking station’s rechargeable internal batteries power the GemStar periods of transport.
The battery pack’s rechargeable internal batteries power the GemStar transport or when use of AC power is not desirable.
batteries (available from local sources)
®
Pump during periods of
®
Pump during
••
••
••
••
••
••
CAUTION: To ensure proper pump operation, ALWAYS replace BOTH
batteries with fresh, disposable AA batteries when a change is required.
DO NOT use rechargeable batteries in the battery compartment. Ensure two fresh, disposable AA batteries are installed for backup
power when using other power sources (e.g., AC mains adaptor, docking station, battery pack). The pump will continue to operate on backup power if the external power source fails. Refer to the caution regarding Power Loss on page 14.
20 Hospira GemS tar® System Operating Manual
Page 28

Installing Disposable Batteries

1 Flip open the tab of the battery door o n
the bottom of the pump.
Chapter 2. Basic Operation
d
Note: Turn off the pump before removing or installing batteries.
2 With the pump keypad facing up, hold
onto the tab and turn the battery door to the left (counterclockwise) until it is released from the battery compartment.
3 Pull the door free and remove any
existing batteries.
4 Install two fresh, disposable AA
batteries into the battery compartment. Ensure the batteries are inserted as shown.
5 Return the battery door to the battery
compartment.
6 Turn the battery door to the right
(clockwise). Ensure the battery door is secured under the slot.
7 Flip closed the tab of the battery door to
lock in place.
Hospira GemStar® System Operating Manual 21
Page 29
Chapter 2. Basic Operation
Connecting the GemStar® AC Mains Adaptor
1 Connect the cord of the AC mains adaptor to the connector port labeled “3 VDC”
on the bottom of the pump.
2 Plug the AC mains adaptor into a standard wall outlet.
CAUTION: ALWAYS connect to a grounded AC mains power source when
using the AC mains adaptor. USE ONLY AC mains adaptors specifically labeled for use with the GemStar
When removing the cord from the bottom of the pump, firmly grasp the connector and pull straight out. DO NOT twist or bend the cord or connector.
Ensure two fresh, disposable AA batteries are installed for backup power when using other power sources (e.g., AC mains adaptor, docking station, battery pack). The pump will continue to operate on backup power if the external power source fails. Refer to the caution regarding Power Loss on page 14.
®
Pump.
d
Note: The green LED on the pump remains lit when connected to AC mains power.
22 Hospira GemS tar® System Operating Manual
Page 30
Chapter 2. Basic Operation
Using the GemStar® Rechargeable Battery Pack
Use the GemStar
[Universal]) to recharge the GemStar
WARNING: DO NOT touch connector ports. Refer to caution on
CAUTION: ALWAYS connect to a grounded AC mains outlet when using AC
®
AC Mains Adaptor (List 13026-01 [USA], 13026-03 [Universal] or 13072-01
®
Battery Pack (List 13073-03).
page 16.
mains adaptor. Use only AC mains adaptors specifically labeled for use with the
GemStar During charging, the battery pack is warm. If the battery pack
becomes hot to the touch, IMMEDIATELY unplug AC mains power and contact Hospira Technical Support Operations
Ensure two fresh, disposable AA batteries are installed for backup power when using other power sources (e.g., AC mains adaptor, docking station, battery pack). The pump will continue to operate on backup power if the external power source fails. Refer to the caution regarding Power Loss on page 14
When removing the cord from the bottom of the battery pack, firmly grasp the connector and pull straight out. DO NOT twist or bend the cord or connector.
Regarding the GemStar
• The LED indicator on the GemStar green/yellow when an error condition has occurred or when the temperature is too high to charge the battery pack. In temperatures above 95°F (35°C), cycle AC power to confirm if battery is charged or if error condition exists. If LED flashes green/yellow, do not rely on external battery. Move device to cooler temperature and recharge prior to use.
• Refer to the GemStar information and precautions regarding activating and charging the Li-Ion Battery Pack.
®
Pump to charge the battery pack.
®
Li-Ion Battery Pack:
®
Li-Ion Battery Pack flashes
®
Li-Ion Battery Pack insert for important
d
Hospira GemStar® System Operating Manual 23
Note: A charged battery pack will last approximately 24 hours when the pump is infusing at rates less than 125 mL/hr.
Page 31
Chapter 2. Basic Operation
To connect the battery pack:
1 Line up the connectors on the top of th e battery
pack with the holes on the bottom of the pump.
2 Snap the battery pack into place.
To release the battery pack:
1 Press the release button on the battery pack. 2 Pull the battery pack straight out from the pump.
Note: When the pump is connected to the battery
d
pack, the pump’s green LED flashes and the battery pack light is off.
1
When the pump is connected to the battery pack at power-on, the pump displays
To charge the battery pack:
1 Plug the pin connector of the AC mains adaptor into the port on the bottom of the
battery pack.
2 Plug the AC mains adaptor into a standard wall outlet. 3 Leave the battery pack connected to the AC mains adaptor for at least 12 hours to
charge the battery pack.
Note: The Li-Ion battery pack (List 13073-03) light is yellow while the battery pack is
d
charging and green when the battery pack is charged or not charging.
Li-Ion batteries cannot be charged when the room temperature is greater than 95°F (35°C) or less than 41°F (5°C).
24 Hospira GemS tar® System Operating Manual
USING EXT BATT.
Page 32
Chapter 2. Basic Operation
CAUTION: When removing the cord from the battery pack, firmly grasp the
connector and pull straight out. DO NOT twist or bend the cord or connector.
®
Using the GemStar
WARNING: DO NOT touch connector ports. Refer to caution on
page 14.
Docking Station
CAUTION: Ensure two fresh, disposable AA batteries are installed for backup
power when using other power sources (e.g., AC mains adaptor, docking station, battery pack). The pump will continue to operate on backup power if the external power source fails. Refer to the caution regarding Power Loss on page 14.
To connect to the Pump
1 Mount the docking station to the I.V. pole and tighten
the knob.
2 Line up the connectors on the bottom of the pump with
the connectors on the docking station.
3 Slide the pump into the docking station until it snaps
into place.
4 Connect the docking station to AC mains power.
Hospira GemStar® System Operating Manual 25
Page 33
Chapter 2. Basic Operation
To remove the pump:
1 Press the release button on the top of the docking
station.
2 Slide the pump out.
To charge docking station (List 13075 all codes):
1 Connect the docking station to the AC mains power. 2 Leave the Li-Ion Docking Station connected to AC
mains power for at least 6 hours when not in use or to recharge the docking station.
Note: The Li-Ion docking station (List 13075-03) light is yellow while the docking
d
station is charging. When the docking station is charged or not charging, the light is green.
The Li-Ion docking station cannot be charged when the room temperature is greater than 104°F (40°C) or less than 41°F (5°C).
26 Hospira GemS tar® System Operating Manual
Page 34
Chapter 2. Basic Operation
Powering On the Pump
CAUTION: If the pump does not display the self-test or if beeps do not sound at
power-on, check all power connections and install two fresh, disposable AA batteries. If the problem continues, contact Hospira Technical Support Operations before using the pump.
d
d
Note: If a cassette is not properly loaded, the pump will sound an alarm after the power-on test is complete.
1 Press [
completes a self-test and displays the time and date. No response is required.
ON/OFF] to power-on the pump. The pump
GEMSTAR 7
UNIT SELF-TEST
TIME IS 10:00 AM
FRI, NOV 10, 06
Note: The self-test takes approximately 30 seconds to complete. If the pump is locked, lock symbols appear at the beginning and end of the second line:
UNIT SELF-TEST .
2 The pump checks for available power sources. Respond to the display message as
shown in the following table:.
Display Message Power Source(s) User Response
No message displays AC power and
disposable batteries.
LOW BATTERIES
ENTER TO CONFIRM
AC power and weak (low voltage) disposable batteries; or weak disposable batteries only.
No response required, the pump goes to the next step.
Press to confirm. Power-off the pump and install two
fresh, disposable AA batteries as a backup power source.
NO INTERNAL BATT
NO POWER LOSS
ALARM AVAILABLE
ENTER TO CONFIRM
Hospira GemStar® System Operating Manual 27
No disposable batteries or dead disposable batteries; and external power.
Press to confirm. Power-off the pump and install two
fresh, disposable AA batteries as a backup power source.
Page 35
Chapter 2. Basic Operation
Display Message Power Source(s) User Response
d
USING BATTERIES
ENTER TO CONFIRM
USING EXT BATT
ENTER TO CONFIRM
Disposable batteries only.
External batteries (docking station battery or battery pack) and disposable batteries.
Press to confirm. If an external power source is in use,
ensure the connections are secure. Press to confirm.
If an AC mains adaptor is in use, ensure the connections are secure.
Note: To power-off the pump, press and hold until the display clears. The pump cannot be powered off during a CHECK CASSETTE alarm (except during
PROGRAMMING mode). Refer to Chapter 12 Troubleshooting for information on
clearing a CHECK CASSETTE alarm.
CAUTION: Ensure two fresh, disposable AA batteries are installed for backup
power when using other power sources (e.g., AC mains adaptor, docking station, battery pack). The pump will continue to operate on backup power if the external power source fails. Refer to the caution regarding Power Loss on page 14.
For more information on: Refer to the following pages:
Setting the clock Set Clock in Chapter 10 Options Menu, on page 118. Locking the keypad Keypad Lock in Chapter 10 Options Menu, on
page 114.
28 Hospira GemS tar® System Operating Manual
Page 36
Chapter 2. Basic Operation
Accessing the Programming Menu
After power-on, the pump checks its memory for a stored program. Respond to the display message as shown in the following table:
Display Message Condition User Response
The pump displays the Programming Menu.
1 RESUME PROGRAM 2 CLR PROG&SHIFT
3 CLR PROG,SHIFT
AND HISTORY
Pump memory is clear.
A program is currently stored in pump memory.
> To enter a new program: Select a therapy
from the Programming Menu and enter a new program. Refer to the appropriate therapy chapter for instructions.
> To continue the current program, press [1]. > To reset the current program:
1 Press [1]. 2 Press [CHANGE] to display the Change
Menu and select
NEW CONTAINER.
f
> To enter a new program and clear shift
totals: Press [ program and shift totals from memory and enters PROGRAMMING mode.
> To enter a new program, clear shift totals,
and clear history: Press [ clears the program, shift totals, and history from memory and enters PROGRAMMING mode.
Tip: To enter a new program when the pump is already powered on:
2]. The pump clears the
1 Press [STOP] to place the pump in STOP mode.
2 Press [CHANGE] to display the Change Menu.
3 Select NEW PROGRAM. The pump clears the current program and
displays the Programming Menu.
3]. The pump
Hospira GemStar® System Operating Manual 29
Page 37
Chapter 2. Basic Operation

Programming the Pump

Programming Overview

The pump’s menu system provides step-by-step guidance through all of the programming steps. Refer to the appropriate therapy chapter for detailed programming instructions.
Programming worksheets are available at the end of each therapy chapter. Each worksheet lists the minimum and maximum value (program range) allowed for that therapy. If a number outside of the program range is entered, the pump displays the minimum or maximum value allowed.
Note: For Continuous, Intermittent, and Variable time therapies, the pump only
d
delivers in increments of 0.1 mL. These therapies are rounded to the nearest 0.1 mL. For example, at a concentration of 15 mg/mL, a rate of 5 mg rounds to 4.5 mg.
When a value is rounded, the pump sounds four quick beeps, disp lays ROUNDING for a few seconds, and then displays the rounded value. Press accept the rounded value and continue to the next step.
Rounding only occurs in Variable, Continuous and Intermittent therapies.
1 Select the desired therapy from the Programming Menu.
5 mg = 0.333 mL
4.5 mg = 0.3 mL
[YES/ENTER] to
2 Respond to display messages and enter requested values.
Note: Program entries can be changed any time before the Program Review is
d
f
30 Hospira GemS tar® System Operating Manual
complete. Press [BACK-UP] to return to previous program displays and make changes as desired.
3 Review the program, if required. Press to begin review. Press [
when done.
4
SAVING PROGRAM displays. The pump stores the program in memory and then
STOP mode.
enters
Tip: If you enter an incorrect number during programming, press [CHANGE] BEFORE pressing [YES/ENTER] to reset the number to zero. Then enter the correct number.
YES/ENTER]
Page 38
Chapter 2. Basic Operation

Programming Tips

The following table explains how to use the keypad during programming:
Key Function
[
0] - [9] Press the number keys to select menu items and to enter numeric values.
Note: Some menu items require pr essing two number keys to select that item.
For example, when selecting a unit of measure for a weight dosed program, the menu may display as:
01 mcg/kg/mg 02 mcg/kg/hr 03 mcg/kg/min 04 mg/kg/min
To select mg/kg/min, press [
Press to accept entries or to continue when done viewing a display.
Press during programming and alarm conditions to display information about the current condition.
0][4].
The DOWN arrow toggles between units of measure when entering concentrations. Scrolls through program review and help messages.
The UP arrow inserts a decimal point in a numeric value.
When entering a numeric value during programming, press to reset the value to zero.
Press to return to a previous display, to exit the Options Menu, or to exit
HELP mode.
the
Hospira GemStar® System Operating Manual 31
Page 39
Chapter 2. Basic Operation

Starting a New Container

Note: The Refer to Chapter 10 Keypad Lock, page 114 for more information.
To repeat the current program with a new container:
1 From the
2 Select
3 If Program Review is desired, press [OPTIONS] then press [1] to begin review.
4 Press [
NEW CONTAINER
STOP
NEW CONTAINER
CONTAINER
Note: In Weight Dosed, mL, and Pain Management therapies, the VTBI can be titrated without adding volume to initial container amount.
START
for several seconds and clears the amount infused from memory.
] to begin delivery.

GemStar® Pump Set

The GemStar® Pump Set is a sterile, single-use, disposable pump set. It is designed for one-time use only and should be properly discarded after use.
Fluid path and areas beneath undisturbed protective set covers are sterile and nonpyrogenic in the intact unit package.
Contact a Hospira representative for more information on selecting the appropriate pump set configuration. Additional disposable components, such as air-eliminating
filters and extension sets, may be added to the line as required. To use a GemStar Pump Set, follow the instructions included with the set.
function is not available when the pump is in
mode, press [
from the Change Menu. The pump displays
CHANGE
] to display the Change Menu.
STERILE
FULL
NEW
Lock.
2
®
Note: GemStar® Pump Sets contain a pressure-activated anti-siphon valve to reduce
the potential for gravity flow at a typical 36-inch head height.
®
Ensure slide clamp or lower CAIR False occlusion alarms and fluid delivery inconsistencies may occur when pumping
viscous fluids in microbore tubing at rates greater than 500 mL/hr. The system may underdeliver, and the system delivery accuracy may be reduced by as
much as 5% when using the following:
viscous fluids
nonvented syringes
PCA vials (except in a lockbox with a spring assist mechanism)
32 Hospira GemStar® System Operating Manual
clamp is open prior to inserting cassette.
Page 40
A typical GemStar® Pump Set is shown below:
Chapter 2. Basic Operation
1 Piercing Pin 6 Tubing 2 Proximal End 7 Slide Clamp 3 Cassette 8 Secure Lock 4 Flow Stop (open / close ) 9 Pressure-activated Anti-Siphon valve 5 Distal End
Hospira GemStar® System Operating Manual 33
Page 41
Chapter 2. Basic Operation
WARNING: ALWAYS prime the pump set to remove air from the
cassette, tubing, and injection sites prior to connecting to the patient. ALWAYS disconnect the pump set from the patient prior to priming or purging. Arrange tubing, cords, and cables to minimize the risk of patient strangulation or entanglement.
Failure to use the pressure-activated anti-siphon valve may result in unrestricted flow if the cassette is not properly installed in the pump.
CAUTION: USE ONLY GemStar® Pump Sets with the GemStar® Pump. Use of
unauthorized sets may result in injury to the patient or damage to the pump.
To prevent contamination, use aseptic technique with all fluid-path connections. Remove protective coverings as assembly progresses.

Opening the Pump Set

1 Open the GemStar
remove the contents.
®
Pump Set package and
2 Remove the protective covers from the fluid
container administration port and the pump set.
34 Hospira GemS tar® System Operating Manual
Page 42
Chapter 2. Basic Operation

Loading the Cassette

1 Identify the four black cassette latches inside the cassette pocket.
2 Ensure slide clamp or lower
CAIR
®
clamp is open prior to
inserting cassette.
3 Line up the cassette as shown
(with the flow stop button facing the pump).
4 Insert the cassette into the cassette
pocket by pushing firmly along the entire cassette until it is firmly seated.
Hospira GemStar® System Operating Manual 35
Page 43
Chapter 2. Basic Operation
5 Verify that all four cassette latches
are securely holding the cassette.
CAUTION: Ensure the cassette is properly installed. When properly installed, all four cassette latches are visibly holding the cassette securely in the cassette pocket.
6 Place the tubing in the
tubing channel.
Note: The pump cannot be
d
powered off during a CHECK CASSETTE alarm (except during
PROGRAMMING mode). Refer to
Chapter 12 Troubleshooting for information on clearing a CHECK CASSETTE alarm.

Releasing the Cassette

1 Close the CAIR® (roller) clamp or
slide clamp on the distal line.
2 Push down on the cassette release
(large black button) on the top of the pump.
Note: Ensure the tubing is behind
d
36 Hospira GemS tar® System Operating Manual
the tubing guide (at the distal end).
3 Remove the cassette.
Note: When the cassette is
d
released from the pump, the cassette flow stop automatically closes, preventing free flow.
Page 44
Chapter 2. Basic Operation
Priming a GemSta r® Pump Set

Priming Pump Sets WITHOUT a Drip Chamber

WARNING: Failure to use the pressure-activated anti-siphon valve
may result in unrestricted flow if the cassette is not properly installed in the pump.
1 Push down on the flow stop to
close the cassette.
2 Hold the container with the
administration port toward the ceiling and insert the piercing pin.
3 Push down on the flow stop to
open (priming position) the cassette.
4 Invert the cassette approximately
45°.
5 While holding the rotated cassette,
squeeze or roll the container to remove air from the container and cassette.
6 Return the cassette to the upright
position. Continue priming until fluid fills the tubing.
Hospira GemStar® System Operating Manual 37
Page 45
Chapter 2. Basic Operation
7 Push down on the flow stop to
close the cassette.
8 Ensure slide clamp is open prior to
inserting cassette.
9 Load the cassette.
CAUTION: Ensure all four cassette latches are clearly visible after the cassette is installed. Refer to Loading the Cassette on page 35.
Note: Ensure the tubing is behind
d
the tubing guide (at the distal end).
CAUTION: When priming is complete, ensure no fluid flows at the distal end of the pump set. If flow is observed, DO NOT use the pump set.
10 Connect the pump set to the
patient-access device. Priming using [
allowed before [ or after one of the following events occur when using Pain Management therapy:
6 Complete Review of New
Program
6 New Container 6 Air-in-Line Alarm
PURGE] is only
START] is pressed
38 Hospira GemS tar® System Operating Manual
Page 46
Chapter 2. Basic Operation

Priming Pump Sets WITHOUT a Drip Chamber Using the PURGE Key

WARNING: Failure to use the pressure-activated anti-siphon valve
may result in unrestricted flow if the cassette is not properly installed in the pump.
d
Note: The PURGE function is available only if there is a program in the pump.
1 Push down on the flow stop to
close the cassette.
2 Hold the container with the
administration port toward the ceiling and insert the piercing pin.
3 Ensure slide clamp is open prior to
inserting cassette.
4 Load the cassette.
CAUTION: Ensure all four cassette latches are clearly visible after the cassette is installed. Refer to Loading the Cassette on page 35.
Hospira GemStar® System Operating Manual 39
Page 47
Chapter 2. Basic Operation
Note: Ensure the tubing is behind
d
the tubing guide (at the distal end).
5 Program desired therapy.
CAUTION: When priming is complete, ensure no fluid flows at the distal end of the pump set. If flow is observed, DO NOT use the pump set.
9 Connect the pump set to the
patient-access device.
6 Press [
7 The pump displays
8 While holding the container with
d
PURGE].
PRIME THE
SET?
the administration port toward the ceiling, press and hold [
Continue using the [ key until fluid fills the tubing. Ensure all air is removed from the container, cassette, and tubing.
Press [YES/ENTER].
PURGE].
Note: Purge rate is 250 mL/hr. The pump purges for up to 2 minutes each time you press and
PURGE].
hold [
PURGE]
40 Hospira GemS tar® System Operating Manual
Page 48
Chapter 2. Basic Operation
Priming Pump Sets WITH a Drip Chamber
WARNING: Failure to use the pressure-activated anti-siphon valve
may result in unrestricted flow if the cassette is not properly installed in the pump.
When using a vented chamber, ALWAYS use an air-eliminating filter OR set the air sensitivity to either ON or 2 mL.
1 Push down on the flow stop to
close the cassette.
2 Insert the piercing pin into the
container. Suspend the container on an I.V. Pole.
3 Squeeze the drip chamber until
filled halfway.
4 Push down on the flow stop to
open (priming position) the cassette.
®
5 Ensure the CAIR
(roller) clamp
is open prior to inserting cassette.
6 Invert the cassette approximately
45°.
Hospira GemStar® System Operating Manual 41
Page 49
Chapter 2. Basic Operation
7 While holding the rotated cassette,
remove air from the container and cassette.
8 Return the cassette to the upright
position. Continue priming until fluid fills the tubing.
9 Push down on the flow stop to
close the cassette.
11 Load the cassette.
CAUTION: Ensure all four cassette latches are clearly visible after the cassette is installed. Refer to Loading the Cassette on page 35.
10 Ensure lower CAIR® clamp is
open prior to inserting cassette.
42 Hospira GemS tar® System Operating Manual
Page 50
Note: Ensure the tubing is behind
d
the tubing guide (at the distal end).
CAUTION: When priming is complete, ensure no fluid flows at the distal end of the pump set. If flow is observed, DO NOT use the pump set.
12 Connect the pump set to the
patient-access device.
Chapter 2. Basic Operation
Hospira GemStar® System Operating Manual 43
Page 51
Chapter 2. Basic Operation
Priming Pump Sets WITH a Drip Chamber Using the PURGE Key
WARNING: Failure to use the pressure-activated anti-siphon valve
may result in unrestricted flow if the cassette is not properly installed in the pump.
When using a vented set, ALWAYS use an air-eliminating filter OR set the air sensitivity to either ON or 2 mL.
d
Note: The Purge function is available only if there is a program in the pump.
1 Push down on the flow stop to
close the cassette.
2 Insert the piercing pin into the
container. Suspend the container on an I.V. Pole.
3 Squeeze the drip chamber until
filled halfway.
4 Push down on the flow stop to
open (priming position) the cassette.
5 Ensure the lower CAIR
clamp is open prior to inserting cassette.
®
(roller)
44 Hospira GemS tar® System Operating Manual
Page 52
6 Load the cassette.
Chapter 2. Basic Operation
7 Program the desired therapy.
d
CAUTION: Ensure all four cassette latches are clearly visible after the cassette is installed. Refer to Loading the Cassette on page 35.
Note: Ensure the tubing is behind the tubing guide (at the distal end).
8 Press [
9 The pump displays
10 Press and hold [
d
PURGE].
SET?
Press [YES/ENTER].
Note: Purge rate is 250 mL/hr. The pump purges for up to 2 minutes each time you press and
PURGE].
hold [
Continue using the [ until fluid fills the tubing. Ensure all air is removed from the container, cassette, and tubing.
CAUTION: When priming is complete, ensure no fluid flows at the distal end of the pump set. If flow is observed, DO NOT use the pump set.
PRIME THE
PURGE].
PURGE] key
11 Connect the pump set to the
patient-access device.
Hospira GemStar® System Operating Manual 45
Page 53
Chapter 2. Basic Operation

Priming Pump Sets Using Pre-filled Vials

WARNING: Failure to use the pressure-activated anti-siphon valve
may result in unrestricted flow if the cassette is not properly installed in the pump.
A cracked vial might not show evidence of leakage until delivery pressure is applied.
d
When using the GemStar (SAM) Lockbox, USE ONLY GemStar designated for use with the GemStar
Failure to use designated sets could result in unrestricted flow.
CAUTION: When using PCA vials (except in a lockbox with a spring assist
mechanism), the system may underdeliver. System delivery accuracy could be reduced by as much as 5%.
When using pre-filled vials, ensure vial and injector are securely locked in lockbox before beginning delivery.
USE ONLY purge priming with the GemStar Spring Assist Mechanism (SAM).
Note: When using the GemStar® Lockbox with SAM, refer to the SAM pump set labeling and SAM lockbox insert for instructions.
1 If using a pump set with the
Y-extension, use the slide clamp to close off the KVO (Keep Vein Open) side of the tubing.
®
Spring Assist Mechanism
2 Open the upper and lower slide
clamps.
®
Pump sets
®
SAM Lockbox.
®
Lockbox with
46 Hospira GemS tar® System Operating Manual
Page 54
Chapter 2. Basic Operation
3 Push down on the flow stop to
close the cassette.
4 Connect the pump set to a Hospira
pre-filled vial and injector.
5 Push down on the flow stop to
open (priming position) the cassette.
air is cleared from the pre-filled vial, injector, cassette, and tubing.
8 Push down on the flow stop to
close the cassette.
9 If the set includes a KVO
Y-extension, close the slide clamp below the pre-filled vial and open the slide clamp on the KVO side. Prime the KVO Y-extension.
6 Hold the pre-filled vial and set
upright (vertically) on a flat surface with the pump set extending from the top.
7 Slowly push down on the injector
until fluid flows at the distal end of the tubing, which indicates that all
Hospira GemStar® System Operating Manual 47
10 Ensure slide clamp is open on
KVO side prior to inserting cassette.
Page 55
Chapter 2. Basic Operation
11 Load the cassette.
CAUTION: Ensure all four
cassette latches are clearly visible after the cassette is installed. Refer to Loading the Cassette on page 35
.
CAUTION: When priming is complete, ensure no fluid flows at the distal end of the pump set. If flow is observed, DO NOT use the pump set.
12 Connect the pump set to the
patient-access device.
Note: Ensure the tubing is behind
d
48 Hospira GemS tar® System Operating Manual
the tubing guide (at the distal end).
Page 56
Chapter 2. Basic Operation
Priming Pump Sets with Extensions
WARNING: Failure to use the pressure-activated anti-siphon valve
may result in unrestricted flow if the cassette is not properly installed in the pump.
1 Push down on the flow stop to
close the cassette.
2 Hold the container with the
administrative port toward the ceiling and insert the piercing pin.
3 Push down on the flow stop to
open (priming position) the cassette.
4 Invert the cassette approximately
45°.
5 While holding the rotated cassette,
squeeze or roll the container to remove air from the container and cassette.
6 Return the cassette to the upright
position. Continue priming until fluid fills the tubing.
Hospira GemStar® System Operating Manual 49
Page 57
Chapter 2. Basic Operation
7 Push down on the flow stop to
close the cassette.
8 Connect the extension set to the
pump set.
9 Ensure the slide clamp is open
prior to inserting cassette.
d
Note: Ensure the tubing is behind the tubing guide (at the distal end).
10 Load the cassette.
11 Program the desired therapy. 12 Press [
CAUTION: Ensure all four cassette latches are clearly visible after the cassette is installed. Refer to Loading the Cassette on page 35.
50 Hospira GemS tar® System Operating Manual
13 The pump displays
PURGE].
SET?
Press [YES/ENTER].
PRIME THE
Page 58
14 Press and hold [PURGE].
Note: Purge rate is 250 mL/hr.
d
The pump purges for up to 2 minutes each time you press and
PURGE].
hold [
Continue using the [ until fluid fills the tubing. Ensure all air is removed from the container, cassette, and tubing.
CAUTION: When priming is complete, ensure no fluid flows at the distal end of the pump set. If flow is observed, DO NOT use the pump set.
15 Connect the pump set to the
patient-access device.
Chapter 2. Basic Operation
PURGE] key
Hospira GemStar® System Operating Manual 51
Page 59
Notes
52 Hospira GemS tar® System Operating Manual
Page 60
Total Parenteral Nutrition (TPN)
- Not Applicable to List 13088

Therapy Features and Specifications

This therapy allows programming of Total Parenteral Nutrition (TPN) protocols.
6 Taper Up and Taper Down may be programmed to gradually increase and
decrease the infusion rate at the beginning and end of a therapy. If Taper Down is selected during programming, the pump allows auto tapering if the continuous delivery is stopped before the programmed volume has been infused.
6 Set an optional KVO (Keep Vein Open) during programming.
Tip: The Programming Worksheet at the end of this chapter lists the minimum and
f
maximum values allowed for this protocol. Use copies of the programming worksheet to record program values before programming the pump.

TPN Programming Tips

The following table explains how to use the keypad during TPN programming:
Chapter 3
Key Function Key Function
[0] - [9] Selects menu items and
enters numeric values.
Accepts entries or continues when done viewing a display.
Displays information during programming and alarm conditions.
Clears a numeric value (resets to zero) during programming.
Hospira GemStar® System Operating Manual 53
The DOWN arrow toggles between units of measure when entering concentrations. Scrolls through program review and help messages.
The UP arrow inserts a decimal point in a numeric value.
Returns to a previous display, exits the Options Menu, and exits the
HELP mode.
Page 61
Chapter 3. Total Parenteral Nutrition

TPN Programming

1 Select TPN from the Programming Menu. 2 Select the TPN delivery mode. 3 Enter the total VTBI (Volume To Be Infused).
d
d
d
d
Note: Maximum VTBI is 9600 mL minus any taper amount.
4 Enter the Taper Up and Taper Down times (if applicable).
Note: Time is entered in hours:minutes (00:00). For one hour thirty minutes, press [1] [3] [0]; not [9] [0].
5 Enter the total time to infuse. 6 Enter the KVO rate. If no KVO is desired, press [
displayed is zero.
Note: KVO runs after Taper Down until the total container is infused.
7 Enter the container size (VTBI + KVO).
Note: If KVO is programmed, increase the container size to accommodate the desired KVO volume.
8 Select the air sensitivity, if required. 9 Review the program. Press to begin review. 10 If using the
WARNING: ALWAYS disconnect the pump set from the patient before
PURGE key to prime the pump set, press [PURGE] now.
priming or purging.
YES/ENTER] while the value
54 Hospira GemS tar® System Operating Manual
Page 62

Changing a TPN Program

Chapter 3. Total Parenteral Nutrition
d
Can be changed before
is pressed
•VTBI
Taper times
Total TPN time
Can be changed at any time
•KVO rate
Increase container size only
Air sensitivity
Cannot be changed
Delivery mode
Note: Y ou cannot change a pr ogram when the pump is in FULL or CONTAINER Lock.
To change a program:
1 From the 2 As each message displays, make the desired changes or press [
STOP mode, press [CHANGE] and select CHANGE PROGRAM.
YES/ENTER] to
accept the current setting.
3 Review the program. Press to begin review. 4 Press [
START] to begin infusion.
Hospira GemStar® System Operating Manual 55
Page 63
Chapter 3. Total Parenteral Nutrition

TPN Auto Taper Down

Auto Taper Down is available only if a Taper Down time was entered during programming. Auto Taper Down may be activated when the continuous delivery portion of a TPN with Taper Down protocol is stopped. Auto Taper Down is not available while the pump is in an alarm condition.
d
1 Press [
2 Enter the Auto Taper Down time.
3 Press [
STOP] during TPN delivery.
The pump displays
Press [NO] to return to the STOP mode. Then press [START] to continue
therapy.
Press [
The maximum Auto Taper Down time is three hours or two times the remaining volume divided by the continuous infusion rate, whichever is less.
Note: Time is entered in hours:minutes (00:00). For one hour thirty minutes, press [1] [3] [0]; not [9] [0].
Taper Down.
YES/ENTER] to set Auto T aper Down.
YES/ENTER] to return to the RUN mode, then press [START] to begin Auto
DO YOU WANT AUTO TAPERING?
56 Hospira GemS tar® System Operating Manual
Page 64
Chapter 3. Total Parenteral Nutrition
GemStar
®
TPN Programming Worksheet
GIVE:
Parameters Selections and/or Values to be
Entered
Therapy TPN N/A Delivery Mode
VTBI (Volume To Be Infused)
Taper Up No
Taper Down No
Total Time to Infuse
KVO (Keep Vein Open) No KVO Rate: 0.0
Continuous Continuous + Taper Down Continuous + Taper Up Continuous + Taper Up and Down
mL 1 - 9600 mL*
Yes hrs:min 1 min - 3 hrs
Yes hrs:min 1 min - 3 hrs**
hrs:min 1 min - 24 hrs
Programmable Ranges
N/A
Yes KVO Rate: 1.0 - 5.0 mL/hr
Container Size (VTBI + KVO) mL 9999 mL max
Air Sensitivity***
* Maximum VTBI is 960 0 mL minus any taper amount. ** Programmable ranges for taper down are 1 minute to 3 hours or 2 times the remaining volume
divided by the continuous delivery rate, whichever is less.
*** Air Sensitivity may not show in all setting configurations.
Hospira GemStar® System Operating Manual 57
On Alarms at approx. 0.5 mL of air 2 mL Alarms at approx. 2 mL of air Off No alarm
N/A
Page 65
Notes
58 Hospira GemS tar® System Operating Manual
Page 66
Chapter 4
Pain Management
- Not Applicable to List 13086

Therapy Features and Specifications

This therapy allows programming of a continuous only , bolus only, or continuous with bolus delivery .
6 In Co ntinuous + Bolu s protocols, yo u can enter zero for either the continuous rate
or the bolus dose when entering or changing a program. The continuous rate and bolus dose cannot be zero at the same time.
6 The default bolus delivery rate is 125 mL/hr.
d
Hospira GemStar® System Operating Manual 59
Note: The delivery rate is administered in increments of .01.
6 The maximum continuous delivery rate is 25 mL/hr. 6 A loading dose may be set during programming, which can be delivered after
programming and priming are complete. The default loading dose rate is 125 mL/hr. Rate, bolus dose, concentration can be programmed in increments of 0.01.
6 Priming u si ng [ PURGE] is only allowed before [START] is pressed or after one of
the following events occur:
6 Complete Review of New Program 6 New Cont ain er 6 Air-in-Line Alarm
6 VTBI can be titrated.
CAUTION: When using PCA vials (except in a lockbox with a spring assist
mechanism), the system may underdeliver. System delivery accuracy could be reduced by as much as 5%.
Page 67
Chapter 4. Pain Management
Note: New Container is allowed only when at least 1/2 of the container volume has
d
f
been infused or 1 hour has elapsed since starting the infusion.
The use of a lockbox is recommended with pain management protocols. Refer to Chapter 15 Optional System Components, page 165, for instructions on using a lockbox with the GemStar® Pump.
Tip: The Programming W ork sheet at the e nd of this chap ter list s the min imum and maximum values allowed for this therapy. Use copies of the programming worksheet to record program values before programming the pump.

Pain Management Programming Tips

The following table explains how to use the keypad during pain management programming:
Key Function Key Function
[0] - [9] Selects menu items and
enters numeric values.
Accepts entries or continues when done viewing a display.
Displays information during programming and alarm conditions.
Clears a numeric value (resets to zero) during programming.
60 Hospira GemStar® System Operating Manual
The DOWN arrow toggles between units of measure when entering concentrations. Scrolls through program review and help messages.
The UP arrow inserts a decimal point in a numeric value.
Returns to a previous display, exits the Options Menu, and exits the
HELP mode.
Page 68

Pain Management Programming

1 Select PAIN MANAGEMENT from the Programming Menu. 2 Select the delivery mode.
Note: If Continuous + Bolus is selected, the continuous delivery rate or bolus dose
d
d
can be set to zero without reprogramming the pump. Press [ [CHANGE], then select CHANGE PROGRAM.
3 Select the unit of measure and concentration as required. 4 Set the continuous delivery rate. 5 Program a loading dose as required. 6 Program a bolus dose with lockout time and dose limit as required.
Regarding a 1-Hour or 4-Hour Dose Limit: The hour dose limit is the maximum volume (continuous plus bolus or bolus only amount) that can be delivered over the selected period. When the programmed dose limit is reached, the continuous delivery stops and a new bolus is not allowed. As the oldest delivery amounts age out of the 1-hour or 4-hour dose limit record, the pump accepts bolus requests or resumes continuous delivery (if programmed).
Note: If a 1-hour or 4-hour dose limit is programmed, a bolus in pr ogr ess will stop when the amount infused reaches the dose limit.
Loading doses are included in a 1-hour or 4-hour dose limit. A dose limit in effect could prevent the delivery of a loading dose.
Chapter 4. Pain Management
STOP] and press
7 If programmed in units of mcg or mg, the line displaying the units flashes. Enter
the container size in proper unit measure.
d
Hospira GemStar® System Operating Manual 61
Note: To enter the total in mL when programming a concentration, press to move to the mL line. Enter the mL value when
8 Select the air sensitivity, if required. 9 Review the program. Press to begin review. 10 If using the PURGE key to prime the pump set, press [
WARNING: ALWAYS disconnect the pump set from the patient before
priming or purging.
mL flashes on the display.
PURGE] now.
Page 69
Chapter 4. Pain Management

Bolus Delivery

Set bolus doses and delivery limits during programming. The default bolus dose delivery rate in a Pain Management therapy is 125 mL/hr .
A bolus cord is available for patient bolus requests. Refer to Connecting the Bolus Cord in Chapter 15 Optional System Components, on page 168, for more information.

Delivering a Bolus Dose

d
From the button on the end of the bolus cord to begin bolus dose delivery. The pump flashes
BOLUS DELIVERY on the display and the amount infused accrues as the delivery
progresses. When the bolus dose is complete, the pump does the following:
6 Records the bolus request and delivery amount to the event and bolus history logs 6 Adds the delivered boluses to the total bolus amount in the program amounts 6 Resets the bolus lockout time (if applicable) and ignores other bolus requests until
6 Continues programmed infusion
Note: ALL bolus requests are recorded to the event and bolus history logs.
If the bolus delivery does not start, it could be locked out by one of the following conditions:
6 Bolus lockout period 6 Loading dose delivery in progress 6 Bolus dose delivery in progress 6 Bolus per-hour limit 6 1-hour or 4-hour dose limit
RUN mode, press either the bolus button [+] on the top of the pump or the
the lockout time has elapsed
62 Hospira GemStar® System Operating Manual
Page 70
Chapter 4. Pain Management
Bolus Dose Interruptions
Bolus deliveries may be interrupted by:
6 Pressing [STOP] 6 An alarm condition
To continue bolus delivery after an interruption:
1 Press [START]. The pump displays COMPLETE BOLUS NOW? 2 Respond as follows:
To do this: Press this key: Bolus lockout time is set to:
Deliver the remaining bolus amount
Clear the undelivered bolus amount
The time of delivery completion
The time the bolus was interrupted
Hospira GemStar® System Operating Manual 63
Page 71
Chapter 4. Pain Management

Loading Dose Delivery

You can set a loading dose during programming, if desired. A loading dose can be delivered after programming and priming are complete.
d
Note: Loading dose amounts are included in a 1-hour or 4-hour dose limit.

Delivering a Loading Dose

1 After priming and programming, press [ 2 The pump displays
to begin loading dose.
When the loading dose is complete, the pump does the following:
6 Automatically begins programmed infusion 6 Records the dose amount to the history log 6 Resets the bolus lockout time (if applicable) and ignores other bolus requests until
the lockout time has elapsed

Loading Dose Interruptions

Loading dose deliveries can be interrupted by:
6 Pressing [STOP] 6 An alarm condition
To continue a loading dose: 1 Resolve the interruption.
DELIVER THE LOADING DOSE NOW? Press [YES/ENTER]
START].
2 Press [
64 Hospira GemStar® System Operating Manual
START]. The pump display s COMPLETE THE LOADING DOSE?
Page 72
Chapter 4. Pain Management
3 Respond as follows:
To do this: Press this key:
Deliver the remaining loading dose amount
Clear the remaining loading dose amount
Changing a Pain Management Program
Can be changed at any time Cannot be ch an ge d
d
Delivery rate
Bolus
Bolus lockout
Container size
Air sensitivity
Dose limit amount
Delivery mode
Unit of measure
Concentration
Dose limit type
Note: You cannot change a program when the pump is in FULL or CONTAINER Lock. With a
RATE CHANGE Lock, the continuous delivery rate and bolus amount
can be changed within the set ranges.
Although the delivery mode cannot be changed, if Continuous + Bolus is programmed, the continuous rate or bolus amount can be changed to zero. The minimum values set with a
RATE CHANGE Lock could restrict this function.
To change a program:
1 From the 2 As each message displays, make the desired changes or press [
STOP mode, press [CHANGE] and select CHANGE PROGRAM.
YES/ENTER] to
accept the current setting.
3 Review the program. Press to begin review. 4 Press [
START] to begin delivery.
Hospira GemStar® System Operating Manual 65
Page 73
Chapter 4. Pain Management
GemStar
®
Pain Management Programming Worksheet
GIVE:
Parameters Selections and/or Values to be
Entered
Therapy Pain Management N/A Delivery Mode
Unit of Measure*
Continuous Rate*
Loading Dose
Bolus Dose* 0 - 2500 mg
Bolus Lockout
Optional Dose Limit
Continuous Bolus Only Continuous + Bolus
Concentration: mg/mL
Concentration: mcg/mL mL
No
Yes Amount:
minutes 1 - 999 min
4-hour limit Amount:
Programmable Ranges
N/A
0.01 - 100 mg/mL 1 - 1000 mcg/mL
0.1 - 2500 mg/hr 1 - 25000 mcg/hr
0.1 - 25.0 mL/hr
0.1 - 2500 mg 1 - 25000 mcg
0.1 - 25.0 mL
0 - 25000 mcg 0 - 25.0 mL
4X rate - 400 mL, 40000 mg, 400000 mcg*
1-hour limit Amount:
# boluses/hour Number: No dose limit
Rate - 100 mL, 10000 mg, 100000 mcg*
1 - 12 *For continuous plus bolus delivery.
Note: For Bolus Only, the 1-hour or 4-hour dose limit minimum is 1 bolus dose.
Container Size 0.1 - 99999 mg
1 - 9999000 mcg
0.1 - 9999 mL
66 Hospira GemStar® System Operating Manual
Page 74
Chapter 4. Pain Management
Parameters Selections and/or Values to be
Entered
Air Sensitivity**
* The delivery rate is programmable and is delivered in increments of .01. ** Air Sensitivity may not show in all setting configurations.
d
Note: Titrating VTBI does not change initial container volume.
On Alarms at approx. 0.5 mL of air 2 mLAlarms at approx. 2 mL of air Off No alarm
Programmable Ranges
N/A
Hospira GemStar® System Operating Manual 67
Page 75
Notes
68 Hospira GemStar® System Operating Manual
Page 76
- Not Applicable to List 13088

Therapy Features and Specifications

This therapy allows programming of multiple doses at regular intervals.
6 Optional KVO (Keep Vein Open) can be set during programming. 6 The pump can be disconnected from the patient between dose deliveries. A call
back alarm can be set to alert the patient or caregiver to reconnect the pump before the next scheduled delivery begins. The pump must remain powered on for the call back alarm to function. At the set time before each dose, the pump produces an audible and visual alarm and flashes
Tip: The Programming Worksheet at the end of this chapter lists the
f
minimum and maximum values allowed for this protocol. Use copies of the programming worksheet to record program values be fore programming the pump.

Intermittent Programming Tips

The following table explains how to use the keypad during intermittent programming:
Chapter 5
Intermittent
START on the display.
Key Function Key Function
[
0] - [9] Selects menu items and
enters numeric values.
Accepts entries or continues when done viewing a display.
Displays information during programming and alarm conditions.
Clears a numeric value (resets to zero) during programming.
Hospira GemStar® System Operating Manual 69
The DOWN arrow toggles between units of measure when entering concentrations. Scrolls through program review and help messages.
The UP arrow inserts a decimal point in a numeric value.
Returns to a previous display, exits the Options Menu, and exits the
HELP mode.
Page 77
Chapter 5. Intermittent

Intermittent Programming

1 Select INTERMITTENT from the Programming Menu. 2 Enter the dose amount in mL. 3 Enter the infusion time for each dose in hours:minutes (00:00).
For example, press [
4 Enter the frequency of dose deliveries.
For example, press [
5 Enter the total number of doses in the container.
To determine the number of doses in the container, divide the volume in the container by the dose amount.
6 Enter the KVO rate in mL.
KVO delivers before a delayed start time, between doses, and after the last dose until the total container is infused.
4] [0] if each dose is to deliver for 40 minutes.
4] [0] [0] to start a dose every four hours.
d
d
If no KVO is desired, press [
7 Enter the container size.
The pump automatically calculates the minimum container size required to deliver the number of programmed doses and KVO between doses.
Note: If KVO is desired before a delayed start, increase the container size to include the desired KVO volume.
8 Select the air sensitivity, if required.
Note: DO NOT press [YES/ENTER] after selecting the air sensitivity. If [YES/ENTER] is accidentally pressed and a call back alarm is not desired, press [
BACK-UP] to return to the beginning of step 9 and press [NO].
9 The pump displays
Press [NO] if you do not want to set the call back alarm.
Press [ Enter the time before the next dose that the alarm should sound. For example,
press [ scheduled delivery.
YES/ENTER] to set the call back alarm.
1] [0] to set the alarm to sound 10 minutes before the start of the next
SET CALL BACK ALARM FOR NEXT DOSE?
YES/ENTER] while the value displayed is zero.
70 Hospira GemStar® System Operating Manual
Page 78
d
Chapter 5. Intermittent
Note: When programming the call back alarm, be sure to allow enough time for the patient or caregiver to reconnect the pump to the patient and to press [ BEFORE the scheduled dose start time.
10 The pump displays the current time.
If the displayed time is correct, press [
To set the clock, press [
OPTIONS].
YES/ENTER].
START]
11 The pump displays the current time and
Press [YES/ENTER] to begin delivery when the pump enters the RUN mode.
Press [
Note: When entering the time in 12-hour format, press or to toggle between AM and
d
d
PM.
12 Review the program. Press to begin review. 13 If using the
WARNING: ALWAYS disconnect the pump set from the patient before
Note: The call back alarm works only when the pump is powered on AND in the
STOP mode. DO NOT power-off the pump while it is disconnected from the patient
if you have set the call back alarm.
When the alarm sounds, press [ Then reconnect the pump to the patient.
Tip: To use a delayed start without specifying a start time:
NO] to set a delayed start time. The pump displays SET START TIME.
Enter the desired start time and press [
PURGE key to prime the pump set, press [PURGE] now.
priming or purging.
SILENCE] to silence the alarm for three minutes.
START NOW?
YES/ENTER] when done.
f
1 Press [YES/ENTER]. 2 Complete the program review. 3 Power-off the pump. 4 When the pump is powered on again, complete the program review, then
press [
START] to begin delivery.
Hospira GemStar® System Operating Manual 71
Page 79
Chapter 5. Intermittent

Intermittent Delivery Interruptions

Missed Start Time

6 If the programmed start time is missed, it can be changed through the Options
Menu. Refer to Next Dose (Not available for List 13088) in Chapter 10 Options Menu, on page 121.
6 If the pump remains in the STOP mode and the first dose is missed, press
[
START] to begin delivery immediately and all remaining doses will be shifted.
6 If the pump is programmed with a delayed start time and is then powered off
before [ programmed start time.
For example, the current time is 10:00 PM. The user sets the delayed start time for 10:30 PM and then powers off the pump. The user does not power on the pump until 10:40 PM, which is 10 minutes AFTER the scheduled start time. The pump does not begin delivery until 10:30 PM the FOLLOWING day.
START] is pressed, the pump will skip to the NEXT occurrence of the
d

Dose Interruptions

To understand how interruptions are resolved, compare the normal Intermittent delivery diagram below for a 1-hour dose every four hours, to the interruption diagrams shown in the examples:
6 If an Intermittent dose is interrupted and the pump is placed in the RUN mode
during the on or off cycle of the current dose, the dose continues delivery from the
72 Hospira GemStar® System Operating Manual
Note: The clock cannot be changed while an Intermittent Program is in progress.
12123456789101112
Dose Delivered -
Page 80
Chapter 5. Intermittent
point where it was interrupted and all remaining doses are shifted. For example, in the following diagram, an intermittent dose was started at 2:00.
The dose was interrupted during the current dose cycle and was started again before the end of the current dose cycle.
12123456789101112
Dose Delivered - Dose Interrupted -
6 If an Intermittent dose is interrupted and the pump is placed in the RUN mode
after the off cycle of the current dose, the next dose begins delivery . All remaining doses are shifted if necessary.
For example, in the following diagram, an intermittent dose was started at 2:00. It was interrupted during the current delivery cycle at 2:30, but was not started again until after the current delivery cycle at 6:45:
12123456789101112
Dose Delivered - Dose Interrupted -
6 When the dose is complete, the start time for the Next Dose displays. To change
the start time refer to Chapter 10 Options Menu, page 107.
6 Any missed Intermittent doses, except the first dose, are added to the end of the
delivery schedule.
Hospira GemStar® System Operating Manual 73
Page 81
Chapter 5. Intermittent

Changing an Intermittent Program

d
Can be changed at any time
Dose parameters
KVO rate
Container size
Air sensitivity
Call back alarm
Can be changed if no dose is in progress
Next dose start time Current time
Cannot be changed after is pressed
Note: You cannot change a program when the pump is in FULL or CONTAINER Lock.
To change a program:
1 From the
STOP mode, press [CHANGE] and select CHANGE PROGRAM.
2 As each message displays, make the desired changes or press [YES/ENTER] to
accept the current setting.
3 Review the program. Press to begin review. 4 Press [
START] to begin delivery.
74 Hospira GemStar® System Operating Manual
Page 82
Chapter 5. Intermittent
GemStar
®
Intermittent Programming Worksheet
GIVE:
Parameters Selections and/or Values to be Entered Programmable
Ranges
Therapy Intermittent N/A
Dose Amount
0.1 - 9600 mL
Infusion Time Per Dose
Dose Frequency
Number of Doses Per Container
KVO (Keep Vein Open)
Container Size*
Air Sensitivity**
Call Back Alarm
Start Time
No KVO Rate: 0.0 Yes KVO Rate:
On Alarms at approx. 0.5 mL of air 2 mL Alarms at approx. 2 mL of air Off No alarm
No Yes
Alarm: hrs:min before start
Start immediately after programming
hrs:min 1 min - 24 hrs
(Infusion time per dose + 1 min) to 96 hrs
hrs:min
1 - 999
0.1 -5.0 mL/hr
0.1 - 9999 mL N/A
1 min to time between doses
Start time:
Specify AM or PM if not 24-hour clock.
* The pump automatically calculates the minimum container size required to deliver the number of
12:00 AM - 11:59 PM or 00:00 - 23:59
programmed doses and KVO between doses. If KVO is desired before a delayed start, increase the container size to include the desired KVO volume.
** Air sensitivity may not show in all setting configurations.
Hospira GemStar® System Operating Manual 75
Page 83
Notes
76 Hospira GemStar® System Operating Manual
Page 84
- Not Applicable to List 13088

Therapy Features and Specifications

This therapy allows programming of a continuous delivery.
6 Optional KVO (Keep Vein Open) can be set during programming. 6 The rate can be titrated while the pump is infusing. 6 An optional secondary container can be “piggybacked” on the primary container
during regular delivery using a GemStar Piggybacking in this chapter for more information.
Tip: The Programming Worksheet at the end of this chapter lists the minimum
f
and maximum values allowed for this protocol. Use copies of the prog ramming worksheet to record program values before programming the pump.

Continuous Programming Tips

The following table explains how to use the keypad during continuous programming:
Chapter 6
Continuous
®
Piggyback Pump Set. Refer to
Key Function Key Function
[0] - [9] Selects menu items and
enters numeric values.
Accepts entries or continues when done viewing a display.
Displays information during programming and alarm conditions.
Clears a numeric value (resets to zero) during programming.
Hospira GemStar® System Operating Manual 77
The DOWN arrow toggles between units of measure when entering concentrations. Scrolls through program review and help messages.
The UP arrow inserts a decimal point in a numeric value.
Returns to a previous display, exits the Options Menu, and exits the
HELP mode.
Page 85
Chapter 6. Continuous

Continuous Programming

1 Select CONTINUOUS from the Programming Menu. 2 Select the unit of measure and concentration as required. 3 Enter the rate. 4 Enter the VTBI (Volume to be Infused). 5 Enter the KVO rate, if desired.
KVO runs after the programmed VTBI until the total container is infused. If no KVO is desired, press [
6 Enter the container size (VTBI + KVO).
VTBI is the default value on the
Note: If KVO is programmed, increase the container size to include the KVO
d
delivery. To enter the total in mL when programming a concentration, press to move to
the mL line. Enter the mL size when
7 Select the air sensitivity, if required. 8 Review the program. Press to begin review.
YES/ENTER] while the value displayed is zero.
CONTAINER display.
mL flashes on the display.
9 If using the PURGE key to prime the pump set, press [
WARNING: ALWAYS disconnect the pump set from the patient before
priming or purging.
78 Hospira GemStar® System Operating Manual
PURGE] now.
Page 86

Rate Titration

To titrate the rate while the pump is infusing:
1 In the
For example: if the current rate is 50 mL/hr, press [ change the rate to 100 mL/hr.
Note: You have approximately five seconds after pressing a key to press another
d
numeric key, the decimal key, or [YES/ENTER]. If no key is pressed within five seconds, the rate does not change and the delivery continues as previously programmed.
2
NEW RATE displays with the new value.
Press [ delivery at the new rate.
Press [
Piggybacking
Piggybacking with the GemStar® Pump requires a GemStar® Primary Piggyback Pump Set with a proximal Y-site backcheck valve and a secondary piggyback set with an extension hook. The secondary container must be higher than the primary container. Refer to the instructions included with the piggyback pump set for more information.
Chapter 6. Continuous
RUN mode, press the number keys for the new rate. Press [YES/ENTER].
1][0] [0] [YES/ENTER] to
YES/ENTER] within 10 seconds to accept the change and begin
NO] to continue delivery at the previously programmed rate.
WARNING: When using the pump for secondary delivery (piggybacking),
ensure the fluids being infused are chemically and physically compatible.
The piggyback rate and piggyback VTBI display until the secondary container has been delivered. Then the primary infusion continues.
d
Note: The piggyback infusion amount is NOT added to the primary programmed infusion amount.
Hospira GemStar® System Operating Manual 79
Page 87
Chapter 6. Continuous

Preparing for Secondary Delivery (Piggybacking)

1 Ensure a primary piggyback pump set with a
proximal Y-site backche ck valve is in place.
2 Use the extension hook from the secondary set
to suspend the primary container from the I.V. pole.
3 Prepare the secondary set according to the
instructions included with the set.
4 Attach a needle or blunt cannula (if appropriate)
and prime the secondary set.
5 Suspend the secondary container from the I.V.
pole.
Note: When using a secondary container of
d
500 mL or greater, ensure the bottom of the secondary container is at least 7 inches (17.8 cm) above the fluid level in the primary container. Use additional extension hooks if necessary.
6 Connect the secondary set to the proximal
Y-site (or proximal luer port).
80 Hospira GemStar® System Operating Manual
Page 88
Chapter 6. Continuous

Programming a Secondary Delivery (Piggybacking)

d
f
d
Note: To distinguish the piggyback delivery from the primary delivery, PIGGYBACK displays and the piggyback VTBI counts down to completion. For example, a piggyback rate of 250 mL/hr with a remaining VTBI of
218.5 mL will display as 219 mL.
Tip: Prepare an d set up cont ain ers BEFORE progr amming the pum p for pigg yback delivery.
1 Press [ 2 Enter the dose amount for the secondary container. 3 Enter the infusion time for the secondary container. 4 Open the slide clamp or lower CAIR
5 Press [
Note: Primary flow stops until the secondary VTBI is delivered; then primary flow continues. Any remaining secondary fluid will be delivered at the primary rate. The maximum piggyback rate is 300 mL/hr. For secondary rates above 300 mL/hr, flow may occur from the primary container.

Changing a Piggyback Delivery

1 Press [
CHANGE] and select PIGGYBACKING from the Change Menu.
®
(roller) clamp on the tubing of the
secondary container.
YES/ENTER] to begin piggyback delivery.
CHANGE] and select PIGGYBACKING from the Change Menu.
PIGGYBACK
RATE 250 mL/hr
VTBI 219 mL
2 Press [ 3 Enter the new piggyback dose amount and press [ 4 Set the infusion time for the secondary container. 5 Press [
d
Note: Y ou must know the amount that ha s alr eady been deliver ed. When the piggyback dose amount is changed, the pump begins a new VTBI countdown.
Hospira GemStar® System Operating Manual 81
CHANGE] to reset the dose amount to zero.
YES/ENTER].
YES/ENTER] to begin piggyback delivery.
Page 89
Chapter 6. Continuous

Canceling a Piggyback Delivery

1 Press [ 2 Press [ 3 Press [
CHANGE] and select PIGGYBACKING from the Change Menu. CHANGE] to reset the dose amount to zero. YES/ENTER] to cancel piggyback delivery.

Alarms Affecting Piggyback Delivery

The following delivery-related Operational Alarms occur during a Piggyback Delivery:
6 Check Cassette 6 Air in Line 6 Distal Occlusion 6 Proximal Occlusion
If an alarm occurs during the programming sequence, piggyback dose programming is completed.
Changing a Continuous Program
Can be changed at any time Cannot be changed
Delivery rate
•VTBI
KVO rate
Container size
Air sensitivity
Piggybacking
Unit of measure
Concentration
d
Note: Y ou cannot change a program when the pump is in FULL or CONTAINER Lock.
To change a program:
1 From the 2 As each message displays, make the desired changes or press [
STOP mode, press [CHANGE] and select CHANGE PROGRAM.
YES/ENTER] to
accept the current setting.
3 Review the program. Press to begin review. 4 Press [
82 Hospira GemStar® System Operating Manual
START] to begin delivery.
Page 90
Chapter 6. Continuous
GemStar
®
Continuous Programming Worksheet
GIVE:
Parameter Selections and/or Values to be
Entered
Therapy Continuous N/A
Unit of Measure
Rate 0.1 - 9999 mg/hr
VTBI (Volume To Be
Infused) KVO
(Keep Vein Open)
Container Size (VTBI + KVO)
Air Sensitivity*
Piggyback Dose Amount 1 - 7200 mL
Piggyback Infusion Time
*Air Sensitivity may not show in all setting configurations.
Concentration: mg/mL
Concentration: mcg/mL mL
No KVO Rate: 0.0
Yes KVO Rate:
On Alarms at approx. 0.5 mL of air 2 mL Alarms at approx. 2 mL of air Off No alarm
hrs:min
Programmable Ranges
0.1 - 1000 mg/mL
1 - 1000 mcg/mL
1 - 999900 mcg/hr
0.1 - 1000 mL/hr
0.1 - 99999 mg 1 - 9999000 mcg
0.1 - 9999 mL
0.1 - 5.0 mL/hr
0.1 - 99999 mg 1 - 9999000 mcg
0.1 - 9999 mL N/A
1 min or dose amount / 300 mL/hr to 24 hrs or dose amount / 0.1 mL/hr
Hospira GemStar® System Operating Manual 83
Page 91
Notes
84 Hospira GemStar® System Operating Manual
Page 92
Weight Dosed
- Not Applicable to List 13088

Therapy Features and Specifications

This therapy allows programming of delivery rate as a function of body weight.
6 The air sensit ivity in a Weight Dosed program is automatically set to ON. This
setting can be changed through the Options Menu after programming is complete. Refer to Air Sensitivity in Chapter 10 Options Menu, on page 117.
6 The pump automatically delivers KVO (Keep Vein Open) after the programmed
VTBI (Volume To Be Infused) has completed.
6 The rate can be titrated while the pump is infusing. 6 VTBI can be titrated.
Tip: The Programming Worksheet at the end of this chapter lists the minimum and
f
maximum values allowed for this protocol. Use copies of the programming worksheet to record program values before programming the pump.
WARNING: When using a vented drip set, ALWAYS use an air-
eliminating filter or set the air sensitivity to either ON or 2 mL.
Chapter 7
CAUTION: When infusing short-half-life drugs (i.e., those dosed with units of
mcg/kg/min) at very low rates (0.8 mL/hr or less) the rate may be too slow for that drug. Consider using a lower drug concentration with these drugs so a pump rate above 0.8 mL/hr can be used.
Hospira GemStar® System Operators Manual 85
Page 93
Chapter 7. Weight Dosed

Weight Dosed Programming Tips

The following table explains how to use the keypad during weight dosed programming:
Key Function Key Function
[0] - [9] Selects menu items and
enters numeric values.
Accepts entries or continues when done viewing a display.
Displays information during programming and alarm conditions.
Clears a numeric value (resets to zero) during programming.
Weight Dosed Programming
1 Select WEIGHT DOSED from the Programming Menu. 2 Select the unit of measure.
d
Note: If two numbers appear before a menu time (e.g., 02), press both number keys to select that item.
3 Enter the patient weight, if required. 4 Enter the concentration:
Enter the value for the unit of measure when it flashes on the display.
Press [
YES/ENTER] to confirm.
The DOWN arrow toggles between units of measure when entering concentrations. Scrolls through program review and help messages.
The UP arrow inserts a decimal point in a numeric value.
Returns to a previous display, exits the Options Menu, and exits the
HELP mode.
f
d
86 Hospira GemStar® System Operators Manual
Tip: Press to toggle between units of measure.
5 Enter the rate.
Note: The pump automatically displays the mL/hr equivalent even when another unit of measure is selected.
6 Enter the VTBI. 7 Review the program. Press to begin review.
Page 94
Chapter 7. Weight Dosed
8 If using the PURGE key to prime the pump set, press [PURGE] now.
WARNING: ALWAYS disconnect the pump set from the patient before
priming or purging.
Auto KVO
When the programmed VTBI has completed delivery, Auto KVO delivery begins automatically and the empty container alarm sounds.
d
Note: Auto KVO delivers even while the empty container alarm sounds. Press [SILENCE] to quiet the alarm for 2 minutes.
If the weight dosed rate is: Auto KVO delivers at:
Greater than or equal to 1 mL/hr 1 mL/hr Less than 1 mL/hr The weight dosed rate

Rate Titration

To titrate the rate while the pump is infusing:
1 In the
For example: if the current rate is 7.5 mcg/kg/min, press [ change the rate to 10 mcg/kg/min.
Note: You have approximately five seconds after pressing a key to press another
d
numeric key, the decimal key, or [ seconds, the rate does not change and the delivery continues as previously programmed.
2
NEW RATE displays with the new value.
Press [
Press [
RUN mode, press the number keys for the new rate. Press [YES/ENTER].
1][0] [YES/ENTER] to
YES/ENTER]. If no key is pressed within five
YES/ENTER] within 10 seconds to accept the change and begin
delivery at the new rate.
NO] to continue delivery at the previously programmed rate.
Hospira GemStar® System Operators Manual 87
Page 95
Chapter 7. Weight Dosed
VTBI Titration
To titrate the VTBI while the pump is infusing:
1 Press [ 2 Press the number keys for the new VTBI. Press [
d
3
d
CHANGE] and select TITRATE VTBI.
For example: if the current VTBI is 0 mL, press [ mL.
Note: You have approximately five seconds after pressing a key to press another numeric key, the decimal key, or [ seconds, the VTBI does not change and the delivery continues as previously programmed.
NEW VTBI displays with the new rate.
Press [
Press [
Note: Titrating VTBI does not change initial container volume.
YES/ENTER] within 10 seconds to accept the change and begin
delivery with the new VTBI.
NO] to continue delivery at the previously programmed VTBI.
YES/ENTER]. If no key is pressed within five
Changing a Weight Dosed Program
Can be changed at any time Cannot be chan ged
Delivery rate
•VTBI
Patient weight
YES/ENTER].
1][0] to change the VTBI to 10
Unit of measure
Concentration (amount of drug)
d
Note: You cannot change a program when the pump is in FULL or CONTAINER Lock.
To change a program:
1 From the 2 As each message displays, make the desired changes or press [
accept the current setting.
3 Review the program. Press to begin review. 4 Press [
88 Hospira GemStar® System Operators Manual
STOP mode, press [CHANGE] and select CHANGE PROGRAM.
YES/ENTER] to
START] to begin delivery.
Page 96
Chapter 7. Weight Dosed

GemStar® Weight Dosed Programming Worksheet

GIVE:
Parameter Selections and/or Values to be Entered Programmable
Ranges
Therapy Weight Dosed N/A Unit of Measure**
mcg/kg/min mcg/kg/hr mcg/min mg/kg/min mg/kg/hr mg/min mL/hr mg/hr
mcg/hr grams/hr ng/kg/min U/kg/hr U/min U/hr mUn/min mEq/hr
N/A
Patient Weight (When programming in kg/min or kg/hr)
2.0 - 200.0 kg Concentration (2 entries are
required. Use to toggle)
(For all units of measure EXCEPT mL/hr)
mg U mEq in in in
mL mL mL
0.1 - 9999000 mg
0.1 - 99999999 U
0.1 - 9999 mEq
0.1 - 9999 mL Rate
(see below *)
VTBI (Volume To Be Infused)
* Programmable Ranges for Rate (not all units of measure are available in all pump configurations):
0.001 - 833 mg/kg/min 0.1 - 1000 mL/hr 0.001 - 99999999 U/min
0.001 - 500000 mg/kg/hr 0.001 - 999999 mg/hr 0.001 - 99999999 U/hr
0.001 - 16666 mg/min 0.001 - 9999999 mcg/hr 0.001 - 999999999 mUn/min
0.001 - 166666 mcg/kg/min 0.001 - 999 grams/hr 0.001 - 999 mEq/hr
0.001 - 9999999 mcg/kg/hr 0.009 - 999 ng/kg/hr
0.002 - 166666 mcg/min 0.001 - 99999999 U/kg/hr
0.1 - 9999 mL
Note: Numbers increment by hundredths—for example, 0.1 increments to 0.11, then to 0.12.
** If two numbers appear before a menu item (e.g., 02), press both number keys to select that item.
CAUTION: When infusing short-half-life drugs (i.e., those dosed with units of mcg/kg/min) at very low rates (0.8 mL/hr or less) the rate may be too slow for that drug. Consider using a lower drug concentration so a pump rate above 0.8 mL/hr can be used.
Hospira GemStar® System Operators Manual 89
Page 97
Chapter 7. Weight Dosed

Programming a Precedex® (dexmedetomidine hydrochloride injection) Delivery

If configured and used by your facility, you can program a Precedex® delivery while programming a Weight Dosed therapy.
Note: Titrate rate, loading dose, and titrate VTBI are available in Precedex® therapy.
d
To program a Precedex
1 Select 2 Select 3 The pump displays 4 Enter patient weight in kg, and then press [ 5 Enter the concentration:
6 Enter the rate, and then press [
Note: Rate is entered in mcg/kg/hr, and then displays in mL/hr.
7 Enter the VTBI in mL, and then press [ 8 The pump displays
9 If programming a loading dose, specify a dose, time, and rate using the key
10 Review the program using the key; when finished, press [
WEIGHT DOSED from the Programming menu. MCG/KG/HR as the unit of measure.
6 Enter the unit of measure for the value flashing on the display. 6 Use the key to toggle between the values, and then press [YES/ENTER].
program a loading dose or [
to toggle between the parameters, and then press [
®
delivery:
IS THIS A PRECEDEX THERAPY? Press [YES/ENTER].
YES/ENTER].
YES/ENTER].
YES/ENTER].
PROGRAM A LOADING DOSE? Press [YES/ENTER] to
NO] to review the program.
YES/ENTER].
YES/ENTER].
WARNING: ALWAYS disconnect the pump set from the patient before
priming or purging.
11 If using the PURGE key to prime the pump set, press [ 12 Press [
90 Hospira GemStar® System Operators Manual
START] to begin the Precedex
®
delivery.
PURGE] now.
Page 98
Chapter 7. Weight Dosed
GemStar® Precedex® (dexmedetomidine hydrochloride
injection) Programming Worksheet
GIVE:
Parameter Selections and/or Values to be Entered Programmable
Therapy Weight Dosed N/A Unit of Measure mcg/kg/hr N/A Precedex®
Therapy? Patient Weight
Concentration (Add values are
required. Use to toggle between fields)
Rate
VTBI (Volume To Be Infused)
Loading Dose (2 entries are
required. Use to toggle between mg and mL entry points.)
* Package Insert recommends 0.2 to 0.7 mcg/kg/hr. The delivery rate is programmable and is delivered in increments of .01.
** Package Insert recommends 1 mcg/kg over10 min . Cannot enter less than 10 min.
[YES/ENTER] N/A
kg mcg
in mL
mcg/kg/hr
(mL/hr is automatically calculated)
mL
No Yes
Dose mcg/kg Time**
Ranges
2.0 - 200.0 kg
1.0- 1000
0.1 - 1000
0.1 - 1000*
0.1 - 9999 mL
0 - 1.0**
Note: Numbers increment by hundredths—for example, 0.1 increments to 0.11, then to 0.12.
Hospira GemStar® System Operators Manual 91
Page 99
Notes
92 Hospira GemStar® System Operators Manual
Page 100
Chapter 8
mL/hr Only
- Not Applicable to List 13088

Therapy Features and Specifications

This therapy allows quick programming of delivery rate in milliliters (mL) only.
6 The air sensitivity is aut omatically set to ON. This setting can be changed through
the Options Menu after programming is complete. Refer to Air Sensitivity in Chapter 10 Options Menu, on page 117.
6 The pump automatically delivers KVO (Keep Vein Open) after the programmed
VTBI (Volume To Be Infused) is complete.
6 The rate can be titrated while the pump is infusing. 6 An optional secondary container can be “piggybacked” onto the primary container
®
during regular delivery using a GemStar Piggybacking in this chapter for more information.
6 A program review is not required and is available only through the Options Menu. 6 VTBI can be titrated.
Piggyback Pump Set. Refer to
Tip: The Programming Worksheet at the end of this chapter lists the minimum and
f
maximum values allowed for this protocol. Use copies of the programming worksheet to record program values before programming the pump.
WARNING: When using a vented set, ALWAYS use an air-eliminating
filter or set the air sensitivity to either ON or 2 mL.
Hospira GemStar® System Operating Manual 93
Loading...