This manual is designed for use by healthcare professionals, caregivers, and patients.
The Hospira Technical Support Operations hotline is available 24 hours a day (in the
USA) to provide consultation and technical assistance regarding the Hospira
The GemStar® Pump is a small and lightweight, single-channel infusion device
designed for use in the home, in the hospital, or anywhere electronic infusion is
®
required. The GemStar
battery pack, docking station, or two disposable AA alkaline batteries. When powered
by batteries, the GemStar
The following conventions are used throughout this manual:
ConventionUseExample
[ALL CAPS]Pump keysPress [YES/ENTER] to continue.
ALL CAPS
WARNING: A WARNING message contains special safety
information and must be observed at all times. Failure
to observe a warning message could be life
threatening.
Pump can be powered by AC mains adaptor, rechargeable
®
Pump is ideal for ambulatory patients.
Display messagesThe pump displays:
ENTER TO CONFIRM.
CAUTION:A CAUTION contains information that could prevent
irreversible product damage or hardware failure. Failure to
observe a caution could result in serious patient or user injury.
Note: A NOTE contains information or examples to help explain instructions.
Tip: A TIP contains information that makes the pump easier to use.
Note: Examples are provided only to help explain a pump function and may not
be clinically relevant. Graphics are for illustrative purposes only and may not
reflect the actual product.
Hospira GemStar® System Operating Manual 1
Page 9
Chapter 1. Introduction
GemStar® Pump Layout
Shown approximate size
2Hospira GemStar® System Operating Manual
Page 10
Chapter 1. Introduction
DISPLAY
1
2
3•Stops infusion.
4•Starts infusion.
5•Lights continuously when the pump is powered by AC mains.
6•Powers the pump on or off. To power-off the pump, press and hold until the display
7•Returns to previous steps during programming.
8
9
10•Toggles between AM and PM in time-based entries.
11•Responds “YES” to display questions.
•Displays the pump’s status.
•Silences an alarm.
•This silences an alert for five minutes and an alarm for two minutes.
•Flashes when the pump is powered by external batteries (e.g., battery pack). Not
lit when using disposable AA batteries.
clears.
•Exits Help, Change, and Options menus.
•Displays the following menu items from the
New ContainerChange Program (or Titrate VTBI)
New ProgramPiggybacking
•Resets a numeric value to zero.
•Displays the following Options Menu items from the
1 Review Program4 Pump Settings7 Speed Protocol
2 Histories5 Set Clock 8 Next Dose
3 Keypad Lock6 Print
•Scrolls through program review and help messages.
•Toggles between AM and PM in time-based entries.
•DOWN arrow toggles between units of measure when entering concentrations.
•UP arrow enters a decimal point in numeric entries.
•Scrolls back by one screen in program review.
•Accepts entries and goes to the next step.
STOP or RUN mode*:
STOP or RUN mode*:
12•Responds “NO” to display questions.
13•Displays information during programming steps and alarm conditions.
•Displays Shift Totals, except during programming steps and alarm conditions.
14
15 [
0] - [9] •Selects menu items and enters numeric values. If two numbers appear before a
* Not all menu items are available with all modes.
Hospira GemStar® System Operating Manual 3
•Clears an
•Primes a GemStar® Pump Set.
menu item (e.g., 02), press both number keys to select that item.
AIR-IN-LINE alarm.
Page 11
Chapter 1. Introduction
GemStar® Pump Components
Bolus ButtonPress this button to start a
1
bolus delivery if available in the
current program.
Pole Clamp TrackUse to connect to an optional
2
pole clamp.
Cassette Release
3
Button
Power ConnectorUse to connect the AC mains
4
Data Port
5
Connector
Bolus ConnectorUse to connect a bolus cord.
6
Tubing ChannelPlace the pump set tubing in
7
Cassette PocketLoad the cassette into this
8
I.V. Tubing GuideUse to hold the I.V. tubing.
9
Battery Pack
10
Connector
Battery
11
Compartment
Push down on this button to
release the cassette.
adaptor.
Use to connect the GemStar®
serial cable.
this channel.
pocket.
Use to connect the GemStar®
Rechargeable Battery Pack.
Install two fresh, disposable
AA batteries in this
compartment.
Top View
Bottom/Side View
4Hospira GemStar® System Operating Manual
Page 12
Chapter 1. Introduction
Therapies
Seven types of therapy are available for infusion of medications and fluids. The
following table lists possible applications and available features of each therapy.
TherapyPossible Applications
TPNTotal Parenteral Nutrition
fluid delivery
Pain
Management
Infusion of analgesics or
anesthetics for patientcontrolled pain
management (PCA),
epidural infusion and
parenteral administration
of general I.V. fluids
IntermittentInterval-based therapies
such as antibiotics
ContinuousInfusion of mg, mcg, or
mL/hr with a selectable
KVO (Keep Vein Open)
option
Weight Dosed Critical care, pediatrics
mL/hr OnlySimple rate programming
Variable TimeChronotherapy
•= Available
O
E
V
K
A
R
O
T
T
I
U
T
R
A
G
T
-
N
S
Y
I
U
G
E
K
G
T
C
I
A
A
P
B
A
L
O
O
L
B
P
G
N
I
D
E
P
S
A
O
T
D
E
U
T
A
R
N
R
E
I
W
T
B
O
I
T
D
T
V
/
••
••
•••
•••
••••
••
L
A
N
O
I
T
O
P
V
O
K
•
Note: pain management is the only therapy available for List 13088.
Hospira GemStar® System Operating Manual 5
Page 13
Chapter 1. Introduction
GemStar® Pump Features
FeatureDescriptionReference Page
•Adjustable
Settings
•Alerts and
Alarms
•Audible
Keypad
Response
•Keypad Lock
•New
Container
•Piggybacking
•Power-on
Self-test
•Operation
Te st
•Program
History
•Speed
Protocols
Adjusts the distal occlusion pressure limits,
air sensor sensitivity, real-time clock, and
audible alarm volume.
Signals audible and visual alarms when
attention is required.
Sounds a single beep when any key is
pressed. If an invalid key is pressed, the
pump sounds a flutter tone.
Restricts access to various options. Four
lock levels are available.
Repeats the current program without
reprogramming the pump.
Allows delivery from a secondary container
during a Continuous or mL/hr Only
program.
Performs a self-test each time the pump is
powered on.
Performs user-assisted tests to verify the
pump is operating properly.
Maintains a history event log, with a date
and time stamp of each event for each
program. Review the history on the display,
or download it to a printer or computer.
Stores up to nine frequently used programs
in the pump’s memory for quick access.
Chapter 10 Options
Menu
Chapter 1 Introduction,
GemStar® Pump Layout
Chapter 1 Introduction,
GemStar® Pump Layout
Chapter 10 Options
Menu and Chapter 11
Clinician Instructions
Chapter 2 Basic
Operation
Chapter 6 Continuous Not Applicable to List
13088 and Chapter 8
mL/hr Only - Not
Applicable to List 13088
Chapter 2 Basic
Operation
Chapter 14 Operation
Te st
Chapter 10 Options
Menu
Chapter 10 Options
Menu
6Hospira GemStar® System Operating Manual
Page 14
Chapter 1. Introduction
Menu System
The GemStar® Pump uses a menu system that provides step-by-step instructions
through all of the pump’s functions. Refer to the following table for a list of functions
in each menu.
Use the keypad to select menu items, respond to display messages, and enter numeric
values. For example, press [
Menu.
2] to select PAIN MANAGEMENT from the Programming
Note: Menu items may vary depending on the pump’s configuration.
Tip: A menu item does not have to be on the display to be selected. For example,
when the Programming Menu first displays, only the first four therapies are shown.
Instead of pressing the down arrow to see the other menu items, press [7] to go
directly to Variable Time Programming.
in Chapter 2 Basic Operation, on page 32. For
menu items [
therapy chapter.
Options MenuChapter 10 Options Menu, page 107.
1], refer to Starting a New Container
2], [3], and [4], refer to the appropriate
Operating Modes
The GemStar® Pump has four operating modes:
ModeDescription
Help ModeThe pump enters the
programming or an alarm condition. The
additional information about the current condition.
Programming ModeThe pump enters the
powered on or when is pressed.
Run ModeAfter is pressed, the pump enters the
when not infusing.
Stop ModeWhen programming is complete, the pump enters the
mode until is pressed.
When the programmed therapy is complete, and during certain
alarm conditions, the pump automatically enters the
Press to enter the
CHECK CASSETTE alarm.
HELP mode when is pressed during
HELP mode displays
PROGRAMMING mode after the pump is
RUN mode, even
STOP mode.
STOP mode at any time, except during a
STOP
Hospira GemStar® System Operating Manual 9
Page 17
Chapter 1. Introduction
Indications for Use
Physicians or certified, licensed healthcare professionals should always oversee
infusions. Healthcare professionals should always supervise pump users and instruct
users in pump operation and troubleshooting. Instruction should include how to
prevent intravenous (I.V.) related complications, including how to prevent accidental
infusion of air.
The pump is suitable for intravenous, arterial, subcutaneous, short-term epidural
infusion and parenteral administration of general I.V. fluids, medications, nutritional
fluids, and blood/blood products.
Intended parenteral administrations include but are not limited to: intravenous (within
a vein), arterial, subcutaneous (tissue just below the skin), regional (located to specific
areas of the body) and epidural analgesia, intrathecal analgesia, intraspinal analgesia,
combined spinal-epidural analgesia, short term epidural infusion, intraspinal (within
the spinal canal), intrathecal (within the subarachnoid or subdural space),
intraoperative sites (soft tissue, body cavity or surgical wounds), intraarticular (within
a joint), intraperitoneal (within the peritoneal cavity), and intracerebroventricular
infusions. The indications for use include hospital, homecare and ambulatory
environments. The system has also been qualified for use in fixed and rotary wing
aircrafts (e.g., helicopters). The pump must be used with sterile, dedicated, GemStar®
administration sets.
Note: Epidural administration of anesthetics is recommended in the continuous
mode only. Epidural administration of analgesics may be delivered by continuous,
bolus, or continuous with bolus.
For epidural use, the administration of drugs is restricted to those anesthetic and
analgesic drugs approved for continuous epidural administration: e.g., Chloroprocaine
Hydrochloride USP, Lidocaine Hydrochloride USP, and Morphine Sulfate Injection
USP (Preservative Free).
WARNING: Delayed respiratory depression following continuous
epidural administration of preservative-free morphine
sulfate has been reported.
10Hospira GemStar® System Operating Manual
Page 18
For epidural administration, the following is recommended:
Chapter 1. Introduction
Nylon or Teflon
Pump sets without Y-sites
Stickers for the tubing indicating ongoing administration
®
catheter
Note: Facilities practicing epidural anesthesia/analgesia must be staffed and
equipped to manage cardio-pulmonary resuscitation (CPR). Supplies should
include oxygen, naloxone, and other appropriate resuscitative drugs and
equipment. Monitor patients continuously (e.g., oximetry) during epidural
administration. Frequently observe patient for side effects (for up to 24 hours)
following completion of drug administration by the epidural route.
Contraindications for Use
Persons who do not have the mental and physical capability or the emotional stability
to properly use the GemStar
DO NOT use the following with the GemStar
Drugs not compatible with silicone rubber or PVC plastic
Drugs not stable under infusion conditions
Warnings and Cautions
Pay attention to the following warnings and cautions when operating the GemStar
Pump. These warnings and cautions are repeated in the manual when relevant to the
procedure being discussed. Pay attention to all alert messages.
®
Pump should not operate this pump.
®
Pump:
®
General Warnings and Cautions
Federal (USA) law restricts this device to sale by or on the order of a physician or
other licensed practitioner.
Manual references to specific values are approximate only, unless indicated
otherwise. Air-in-line sensitivity values are approximate only.
When a proximal occlusion alarm activates, DO NOT power-off the pump.
Closely supervise patients who are likely to be adversely affected by unintended
operations and failures, including interrupted medication or fluid infusion. Ensure
provisions for immediate corrective action are available.
Hospira GemStar® System Operating Manual 11
Page 19
Chapter 1. Introduction
Regarding Drugs Used, Pump Sets, and Containers
USE ONLY GemStar
sets may result in injury to the patient or damage to the pump.
Some microbore tubing delivery accuracy may be affected when used at extreme
operating temperatures (<10°C, >35°C), relative humidity (<20%) and delivery
rates (<4.9 mL/hr, >350 mL/hr). Refer to individual set labels for specific
instructions and operating conditions.
Delivery accuracy may be impacted when there is air present in the cassette.
Follow the instructions in the individual set labels to remove any visible air from
the cassette.
Ensure slide clamp or lower CAIR
cassette.
When using pre-filled vials, ensure vial and injector are securely locked in
lockbox before beginning delivery.
A cracked vial might not show evidence of leakage until delivery pressure is
applied.
When using nonvented syringes, the system may underdeliver. System delivery
accuracy could be reduced by as much as 5%.
When using PCA vials (except in a lockbox with a spring assist mechanism), the
system may underdeliver. System delivery accuracy could be reduced by as much
as 5%.
When using viscous fluids, the system may underdeliver. System delivery
accuracy could be reduced by as much as 5%.
Variations in filling head (such as container height) may affect delivery accuracy
by no more than 5%.
When using the GemStar
ONLY GemStar® Pump Sets designated for use with the GemStar
Lockbox. Failure to use designated sets could result in unrestricted flow.
NEVER use drugs that are incompatible with silicone rubber or PVC plastic,
(unless these items are specially treated and labeled to be compatible with such
drugs).
To reduce the loss of potency for drugs known to be absorbed by PVC plastic and
silicone, begin infusion as soon as practical after priming the pump set. Use of
high flow rates during infusion will minimize drug absorption.
DO NOT use medications that are unstable under infusion conditions.
ALWAYS use connections with Luer Lock™ fittings.
®
Pump Sets with the GemStar® Pump. Use of unauthorized
®
(roller) clamp is open prior to inserting
®
Spring Assist Mechanism (SAM) Lockbox, USE
®
SAM
12Hospira GemStar® System Operating Manual
Page 20
Chapter 1. Introduction
To prevent contamination, use aseptic technique with all fluid-path connections.
Remove the protective coverings as assembly progresses.
ALWAYS place the flow stop in the closed position when the set is not in the pump.
Before installing the cassette in the pump, ensure all slide or lower CAIR
clamps are open. Failure to open the slide or lower CAIR
®
(roller) clamps
BEFORE installing the cassette may result in restricted flow.
DO NOT back-prime the pump set using a syringe. Back priming may result in
restricted flow.
Ensure the cassette is properly installed. When properly installed, all four cassette
latches are visibly holding the cassette securely in the cassette pocket.
Ensure the tubing is behind the tubing guide (at the distal end).
Arrange tubing, cords, and cables to minimize the risk of patient strangulation or
entanglement.
NEVER use vented fluid containers (e.g., glass or rigid plastic) unless suspended
from a pole. When using a vented drip set, ALWAYS use an air-eliminating filter
OR set the air sensitivity to either ON or 2 mL.
Stop infusion if signs or symptoms of infiltration occur.
GemStar® Pump Sets contain a pressure-activated anti-siphon valve to reduce the
potential for gravity flow at a typical 36-inch head height.
Failure to use the pressure-activated anti-siphon valve may result in unrestricted
flow if the cassette is not properly installed in the pump.
Always close the slide clamps before removing the cassette from the pump.
When using the pump for secondary delivery (piggybacking), ensure the fluids
being infused are chemically and physically compatible.
When infusing short-half-life drugs (i.e. those dosed with units of mcg/kg/min) at
very low rates (0.8 mL/hr or less) the rate may be too slow for that drug. Consider
using a lower drug concentration with these drugs so a pump rate above 0.8 mL/hr
can be used.
®
(roller)
Regarding Air-In-Line and Infusion
To reduce the risk of infusing air, use an air-eliminating filter when the air
sensitivity is set to OFF.
ALWAYS prime the pump set to remove all air from the cassette, tubing, and
injection site prior to connecting to the patient.
ALWAYS disconnect the pump set from the patient prior to priming or purging.
Hospira GemStar® System Operating Manual 13
Page 21
Chapter 1. Introduction
Regarding Pump Operation
If the pump does not perform as stated in this manual, remove the pump from
service IMMEDIATELY.
If the pump does not display the self-test or if beeps do not sound at power-on,
check all power connections and install two fresh, disposable AA batteries. If the
problem continues, contact Hospira Technical Support Operations before using
the pump.
Ensure a cassette is installed in the pump before connecting to a printer or
computer. If the pump is connected to a patient: BEFORE connecting the pump
to a computer or printer, press [
cassette. Leave the pump set connected to the patient. Install another cassette in
the pump before connecting to a printer or computer.
Disconnect the pump from the patient BEFORE performing the Operation Test.
ALWAYS connect to a grounded AC mains power source when using AC mains
power. If the quality of the grounding source is in doubt, operate the pump only
with batteries.
USE ONLY the AC mains adaptors specifically labeled for use with the
GemStar
USE ONLY the AC mains adaptor specifically labeled for use with the GemStar
®
Pump.
Pump to charge the battery pack.
The GemStar
GemStar
®
AC mains adaptor is for use with the GemStar® Pump or
®
Rechargeable Battery Pack only. DO NOT use the GemStar® AC
mains adaptor with other products.
During charging, the battery pack is warm. If the rechargeable battery pack
becomes hot to the touch, IMMEDIATELY unplug the AC mains adaptor and
contact Hospira Technical Support Operations.
Use only the Li-Ion Docking Station and Li-Ion Battery Pack with the GemStar
Pump.
With the Li-Ion Battery Pack, the LED indicator flashes yellow/green when an
error condition has occurred or when the temperature is too high to charge the
battery pack. In temperatures above 95°F (35°C), cycle AC power to confirm if
battery is charged or if error condition exists. If LED flashes green/yellow, do not
rely on external battery. Move device to cooler temperature and recharge prior to
use. Refer to the battery pack insert for important information and precautions
regarding activating and charging the Li-Ion Battery Pack.
When removing the cord (from the docking station, bottom of the battery pack, or
STOP], close the slide clamp, and eject the
®
®
14Hospira GemStar® System Operating Manual
Page 22
Chapter 1. Introduction
bottom of the pump), firmly grasp the connector and pull straight out. DO NOT
twist or bend the cord or connector.
To ensure proper pump operation, ALWAYS replace BOTH disposable batteries
with fresh, disposable AA batteries when a change is required.
ALWAYS fully close the battery door when using disposable batteries.
DO NOT use rechargeable batteries in the battery compartment.
Ensure two fresh, disposable AA batteries are installed for backup power when
using other external power sources (e.g., AC mains adaptor, docking station,
battery pack). The pump will continue to operate on backup power if the external
power source fails. In the event that the internal batteries are removed while the
pump is operating, the power loss event may not be logged in the history, the
programmed therapy may not be retained, and the alarms (audible and visual) may
not occur.
ALWAYS avoid sources of high-intensity electromagnetic radiation (e.g., radio
transmitters, MRI scanners, microwave ovens, X-ray machines, and CAT
scanners).
Use of radio-frequency emitting devices, such as cellular telephones, 2-way
radios, and Electrical Surgical Devices (ESU), in close proximity to this device
may affect its operation.
Possible explosion hazard exists if the pump is used in the presence of flammable
anesthetics. NEVER use the pump in the presence of flammable or explosive
vapors.
Non-hazardous, low-level electrical potentials are commonly observed when
fluids are administered using infusion devices. These potentials are well within
accepted safety standards, but may create artifacts on voltage-sensing equipment,
such as ECG, EMG, and EEG machines. These artifacts vary at a rate that is
associated with the infusion rate. If the monitoring machine is not operating
correctly or has loose or defective connections to its sensing electrodes, these
artifacts may be accentuated so as to simulate actual physiological signals. To
determine if the abnormality in the monitoring equipment is caused by the
infusion device instead of some other source in the environment, set the infusion
device so that it is temporarily not delivering fluid (if clinically acceptable).
Disappearance of the abnormality indicates that it was probably caused by
electronic noise generated by the infusion device. Proper setup and maintenance
of the monitoring equipment should eliminate the artifact. Refer to the appropriate
monitoring system documentation for setup and maintenance instructions.
The GemStar
in accordance with:
®
Pump has been tested and found to comply with EMC/EMI limits
Hospira GemStar® System Operating Manual 15
Page 23
Chapter 1. Introduction
IEC/EN 60601-1-2 (2001) when using two AA disposable batteries or when
connected to the GemStar
IEC/EN 60601-1-2 (2001) (ESD exemption) when using Li-Ion Docking
®
AC Adaptor
Station or Li-Ion Battery Pack
When connected to other power sources or optional system components in
®
commercial distribution before November 30, 2004, the GemStar
Pump
complies with EMC/EMI limits in accordance with IEC/EN 60601-1-2 (1993).
These limits are designed to provide reasonable protection against harmful
interference in a typical medical installation. The equipment generates, uses and
can radiate radio frequency energy and, if not installed and used in accordance
with the instructions, may cause harmful interference to other devices in the
vicinity. However, there is no guarantee that interference will not occur in a
particular installation. If this equipment does cause harmful interference with
other devices, which can be determined by turning the equipment off and on, the
user is encouraged to try to correct the interference by one or more of the
following measures:
Reorient or relocate the receiving device
Increase the separation between the equipment
Connect the equipment to an outlet on a circuit different from that to which
the other device is connected
Consult the manufacturer or field service technician for help
The use of portable and mobile RF equipment may have an impact on this and
other pieces of medical equipment
Prior to using the GemStar
®
Pump and optional system components, refer to
Appendix A for additional information and precautionary procedures regarding
electrical potentials (electromagnetic compatibility and electrostatic discharge).
Regarding Handling and Maintenance
Use proper care during unpacking, installation, and operation of the pump. If the
pump is inadvertently mishandled, check connections and programmed data to
ensure there is no damage. Refer to Chapter 14 Operation Test for further
information.
Accessory equipment connected to the analog and digital interfaces must be
certified according to the respective IEC standards, (e.g., IEC 60950 for data
processing equipment, and ICE 60601-1 for medical equipment). Furthermore, all
configurations shall comply with the system standard IEC 60601-1-1. Any person
who connects additional equipment to the signal input or output part configures a
medical system, and is therefore responsible for ensuring that the system complies
16Hospira GemStar® System Operating Manual
Page 24
Chapter 1. Introduction
with the requirements of IEC 60601-1-1. If in doubt, contact Hospira Technical
Support Operations.
ALWAYS avoid dropping or hitting the pump. If the pump is dropped or hit,
ALWAYS check connections and programmed data.
NEVER use sharp objects (e.g., fingernails, pens, pencils, or other probes) to
program or clean the pump.
Use care not to damage the silicone seals around the sensor bodies in the cassette
pocket.
To avoid mechanical or electronic damage, NEVER submerge pump in water or
other fluids and avoid fluid spills. If the pump becomes wet, dry it immediately
with a dry, lint-free cloth. Check connections and programmed data.
DO NOT touch connector ports. Connector ports may be susceptible to
Electrostatic Discharge (ESD) damage.
Some cleaning and sanitizing compounds may slowly degrade components made
from some plastic materials. DO NOT use compounds containing combinations of
isopropyl alcohol and dimethyl benzyl ammonium chloride (i.e., alcohol wipes).
DO NOT sterilize by heat, steam, ethylene oxide (ETO), or radiation. Apply
disinfectants to the outside surface of the pump only. DO NOT use abrasive
cleaners or materials on the pump. Using abrasive cleaners or cleaning solutions
not recommended by Hospira may result in product damage.
Please comply with local disposal and recycling regulations as appropriate for
disposable batteries, rechargeable battery packs, medical electronic components,
and pump sets.
Fixed and Rotary Wing Aircraft
Do not lay the IV bag flat because air can potentially be introduced into the line
causing an alarm condition. Recommend the use of extension set tubing be
applied prior to patient transport.
The GemStar® pump has only been tested using battery power in fixed and rotary
wing aircraft.
Auditory alarms may not be heard in a noisy flight environment. Care providers
should rely on visual indicators on the pump display to determine if there are any
alarms or other system malfunctions.
Hospira GemStar® System Operating Manual 17
Page 25
Notes
18Hospira GemStar® System Operating Manual
Page 26
Chapter 2
Basic Operation
The following elements are required to use the GemStar® Pump:
6GemStar
6Appropriate power source
6Appropriate GemS tar
6Appropriate fluid container
6Patient-access device
Overview
To start a therapy:
1Determine the program settings. Refer to the appropriate therapy chapter for a
worksheet that may be copied and used to record program settings.
2Select a power source. Refer to Selecting a Power Source on page 20.
3Prime the pump set. Refer to GemStar® Pump Set on page 32.
®
Pump
®
Pump Set
4Load the cassette into the pump. Refer to Loading the Cassette on page 35.
5Power-on the pump and go to the Programming Menu. Refer to Powering On
the Pump on page 27.
6Select the appropriate optional system components, such as a bolus cord.
7Select the therapy type and enter a program. Refer to the appropriate therapy
chapter for detailed instructions.
d
Note: List 13088 only delivers therapy in the Pain management Mode.
8Press [
Hospira GemStar® System Operating Manual 19
START] to begin delivery.
Page 27
Chapter 2. Basic Operation
Selecting a Power Source
The GemStar® Pump may be powered by one or more of the following:
Power SourceDescriptionAmbulatoryHospital Home
GemStar®
AC Mains Adaptor
(USA)
GemStar®
AC Mains Adaptor
(Universal)
GemStar®
AC Mains Adaptor
(Universal)
GemStar® Docking
Station
GemStar® Battery
Pack
Disposable Batteries Two disposable AA alkaline
Wall-mount adaptor plugs into a
1 10 VAC mains supply.
Wall-mount adaptor plugs into a
100-240 VAC mains supply.
Tabletop adaptor plugs into a
100–240 VAC mains supply.
Plugs into a 100–240 VAC Volt
mains supply while mounted on an
I.V. pole. The docking station’s
rechargeable internal batteries
power the GemStar
periods of transport.
The battery pack’s rechargeable
internal batteries power the
GemStar
transport or when use of AC power
is not desirable.
batteries (available from local
sources)
®
Pump during periods of
®
Pump during
••
••
••
•••
•••
•••
CAUTION: To ensure proper pump operation, ALWAYS replace BOTH
batteries with fresh, disposable AA batteries when a change is
required.
DO NOT use rechargeable batteries in the battery compartment.
Ensure two fresh, disposable AA batteries are installed for backup
power when using other power sources (e.g., AC mains adaptor,
docking station, battery pack). The pump will continue to operate
on backup power if the external power source fails. Refer to the
caution regarding Power Loss on page 14.
20Hospira GemS tar® System Operating Manual
Page 28
Installing Disposable Batteries
1Flip open the tab of the battery door o n
the bottom of the pump.
Chapter 2. Basic Operation
d
Note: Turn off the pump before
removing or installing batteries.
2With the pump keypad facing up, hold
onto the tab and turn the battery door to
the left (counterclockwise) until it is
released from the battery compartment.
3Pull the door free and remove any
existing batteries.
4Install two fresh, disposable AA
batteries into the battery compartment.
Ensure the batteries are inserted as
shown.
5Return the battery door to the battery
compartment.
6Turn the battery door to the right
(clockwise). Ensure the battery door is
secured under the slot.
7Flip closed the tab of the battery door to
lock in place.
Hospira GemStar® System Operating Manual 21
Page 29
Chapter 2. Basic Operation
Connecting the GemStar® AC Mains Adaptor
1Connect the cord of the AC mains adaptor to the connector port labeled “3 VDC”
on the bottom of the pump.
2Plug the AC mains adaptor into a standard wall outlet.
CAUTION: ALWAYS connect to a grounded AC mains power source when
using the AC mains adaptor. USE ONLY AC mains adaptors
specifically labeled for use with the GemStar
When removing the cord from the bottom of the pump, firmly
grasp the connector and pull straight out. DO NOT twist or bend
the cord or connector.
Ensure two fresh, disposable AA batteries are installed for backup
power when using other power sources (e.g., AC mains adaptor,
docking station, battery pack). The pump will continue to operate
on backup power if the external power source fails. Refer to the
caution regarding Power Loss on page 14.
®
Pump.
d
Note: The green LED on the pump remains lit when connected to AC mains
power.
22Hospira GemS tar® System Operating Manual
Page 30
Chapter 2. Basic Operation
Using the GemStar® Rechargeable Battery Pack
Use the GemStar
[Universal]) to recharge the GemStar
WARNING: DO NOT touch connector ports. Refer to caution on
CAUTION: ALWAYS connect to a grounded AC mains outlet when using AC
®
AC Mains Adaptor (List 13026-01 [USA], 13026-03 [Universal] or 13072-01
®
Battery Pack (List 13073-03).
page 16.
mains adaptor.
Use only AC mains adaptors specifically labeled for use with the
GemStar
During charging, the battery pack is warm. If the battery pack
becomes hot to the touch, IMMEDIATELY unplug AC mains
power and contact Hospira Technical Support Operations
Ensure two fresh, disposable AA batteries are installed for backup
power when using other power sources (e.g., AC mains adaptor,
docking station, battery pack). The pump will continue to operate
on backup power if the external power source fails. Refer to the
caution regarding Power Loss on page 14
When removing the cord from the bottom of the battery pack,
firmly grasp the connector and pull straight out. DO NOT twist or
bend the cord or connector.
Regarding the GemStar
• The LED indicator on the GemStar
green/yellow when an error condition has occurred or when the
temperature is too high to charge the battery pack. In
temperatures above 95°F (35°C), cycle AC power to confirm if
battery is charged or if error condition exists. If LED flashes
green/yellow, do not rely on external battery. Move device to
cooler temperature and recharge prior to use.
• Refer to the GemStar
information and precautions regarding activating and charging
the Li-Ion Battery Pack.
®
Pump to charge the battery pack.
®
Li-Ion Battery Pack:
®
Li-Ion Battery Pack flashes
®
Li-Ion Battery Pack insert for important
d
Hospira GemStar® System Operating Manual 23
Note: A charged battery pack will last approximately 24 hours when the pump
is infusing at rates less than 125 mL/hr.
Page 31
Chapter 2. Basic Operation
To connect the battery pack:
1Line up the connectors on the top of th e battery
pack with the holes on the bottom of the pump.
2Snap the battery pack into place.
To release the battery pack:
1Press the release button on the battery pack.
2Pull the battery pack straight out from the pump.
Note: When the pump is connected to the battery
d
pack, the pump’s green LED flashes and the battery
pack light is off.
1
When the pump is connected to the battery pack at
power-on, the pump displays
To charge the battery pack:
1Plug the pin connector of the AC mains adaptor into the port on the bottom of the
battery pack.
2Plug the AC mains adaptor into a standard wall outlet.
3Leave the battery pack connected to the AC mains adaptor for at least 12 hours to
charge the battery pack.
Note: The Li-Ion battery pack (List 13073-03) light is yellow while the battery pack is
d
charging and green when the battery pack is charged or not charging.
Li-Ion batteries cannot be charged when the room temperature is greater than 95°F
(35°C) or less than 41°F (5°C).
24Hospira GemS tar® System Operating Manual
USING EXT BATT.
Page 32
Chapter 2. Basic Operation
CAUTION: When removing the cord from the battery pack, firmly grasp the
connector and pull straight out. DO NOT twist or bend the cord or
connector.
®
Using the GemStar
WARNING: DO NOT touch connector ports. Refer to caution on
page 14.
Docking Station
CAUTION: Ensure two fresh, disposable AA batteries are installed for backup
power when using other power sources (e.g., AC mains adaptor,
docking station, battery pack). The pump will continue to operate
on backup power if the external power source fails. Refer to the
caution regarding Power Loss on page 14.
To connect to the Pump
1Mount the docking station to the I.V. pole and tighten
the knob.
2Line up the connectors on the bottom of the pump with
the connectors on the docking station.
3Slide the pump into the docking station until it snaps
into place.
4Connect the docking station to AC mains power.
Hospira GemStar® System Operating Manual 25
Page 33
Chapter 2. Basic Operation
To remove the pump:
1Press the release button on the top of the docking
station.
2Slide the pump out.
To charge docking station (List 13075 all codes):
1Connect the docking station to the AC mains power.
2Leave the Li-Ion Docking Station connected to AC
mains power for at least 6 hours when not in use or to recharge the docking
station.
Note: The Li-Ion docking station (List 13075-03) light is yellow while the docking
d
station is charging. When the docking station is charged or not charging, the light is
green.
The Li-Ion docking station cannot be charged when the room temperature is greater
than 104°F (40°C) or less than 41°F (5°C).
26Hospira GemS tar® System Operating Manual
Page 34
Chapter 2. Basic Operation
Powering On the Pump
CAUTION: If the pump does not display the self-test or if beeps do not sound at
power-on, check all power connections and install two fresh,
disposable AA batteries. If the problem continues, contact Hospira
Technical Support Operations before using the pump.
d
d
Note: If a cassette is not properly loaded, the pump will sound an alarm after the
power-on test is complete.
1Press [
completes a self-test and displays the time and date. No
response is required.
ON/OFF] to power-on the pump. The pump
GEMSTAR 7
UNIT SELF-TEST
TIME IS 10:00 AM
FRI, NOV 10, 06
Note: The self-test takes approximately 30 seconds to complete. If the pump is
locked, lock symbols appear at the beginning and end of the second line:
UNIT SELF-TEST .
2The pump checks for available power sources. Respond to the display message as
shown in the following table:.
Display MessagePower Source(s)User Response
No message displaysAC power and
disposable batteries.
LOW BATTERIES
ENTER TO CONFIRM
AC power and weak
(low voltage)
disposable batteries;
or weak disposable
batteries only.
No response required, the pump goes
to the next step.
Press to confirm.
Power-off the pump and install two
fresh, disposable AA batteries as a
backup power source.
NO INTERNAL BATT
NO POWER LOSS
ALARM AVAILABLE
ENTER TO CONFIRM
Hospira GemStar® System Operating Manual 27
No disposable
batteries or dead
disposable batteries;
and external power.
Press to confirm.
Power-off the pump and install two
fresh, disposable AA batteries as a
backup power source.
Page 35
Chapter 2. Basic Operation
Display MessagePower Source(s)User Response
d
USING BATTERIES
ENTER TO CONFIRM
USING EXT BATT
ENTER TO CONFIRM
Disposable batteries
only.
External batteries
(docking station
battery or battery
pack) and disposable
batteries.
Press to confirm.
If an external power source is in use,
ensure the connections are secure.
Press to confirm.
If an AC mains adaptor is in use,
ensure the connections are secure.
Note: To power-off the pump, press and hold until the display clears. The pump
cannot be powered off during a CHECK CASSETTE alarm (except during
PROGRAMMING mode). Refer to Chapter 12 Troubleshooting for information on
clearing a CHECK CASSETTE alarm.
CAUTION: Ensure two fresh, disposable AA batteries are installed for backup
power when using other power sources (e.g., AC mains adaptor,
docking station, battery pack). The pump will continue to operate
on backup power if the external power source fails. Refer to the
caution regarding Power Loss on page 14.
For more information on:Refer to the following pages:
Setting the clockSet Clock in Chapter 10 Options Menu, on page 118.
Locking the keypadKeypad Lock in Chapter 10 Options Menu, on
page 114.
28Hospira GemS tar® System Operating Manual
Page 36
Chapter 2. Basic Operation
Accessing the Programming Menu
After power-on, the pump checks its memory for a stored program. Respond to the
display message as shown in the following table:
Display MessageConditionUser Response
The pump displays the
Programming Menu.
1 RESUME PROGRAM
2 CLR PROG&SHIFT
3 CLR PROG,SHIFT
AND HISTORY
Pump memory
is clear.
A program is
currently stored
in pump
memory.
>To enter a new program: Select a therapy
from the Programming Menu and enter a
new program. Refer to the appropriate
therapy chapter for instructions.
>To continue the current program, press [1].
>To reset the current program:
1 Press [1].
2 Press [CHANGE] to display the Change
Menu and select
NEW CONTAINER.
f
>To enter a new program and clear shift
totals: Press [
program and shift totals from memory and
enters PROGRAMMING mode.
>To enter a new program, clear shift totals,
and clear history: Press [
clears the program, shift totals, and history
from memory and enters PROGRAMMING
mode.
Tip: To enter a new program when the pump is already powered on:
2]. The pump clears the
1 Press [STOP] to place the pump in STOP mode.
2Press [CHANGE] to display the Change Menu.
3 Select NEW PROGRAM. The pump clears the current program and
displays the Programming Menu.
3]. The pump
Hospira GemStar® System Operating Manual 29
Page 37
Chapter 2. Basic Operation
Programming the Pump
Programming Overview
The pump’s menu system provides step-by-step guidance through all of the
programming steps. Refer to the appropriate therapy chapter for detailed
programming instructions.
Programming worksheets are available at the end of each therapy chapter. Each
worksheet lists the minimum and maximum value (program range) allowed for that
therapy. If a number outside of the program range is entered, the pump displays the
minimum or maximum value allowed.
Note: For Continuous, Intermittent, and Variable time therapies, the pump only
d
delivers in increments of 0.1 mL. These therapies are rounded to the nearest 0.1
mL. For example, at a concentration of 15 mg/mL, a rate of 5 mg rounds to 4.5 mg.
When a value is rounded, the pump sounds four quick beeps, disp lays ROUNDING
for a few seconds, and then displays the rounded value. Press
accept the rounded value and continue to the next step.
Rounding only occurs in Variable, Continuous and Intermittent therapies.
1Select the desired therapy from the Programming Menu.
5 mg = 0.333 mL
4.5 mg = 0.3 mL
[YES/ENTER] to
2Respond to display messages and enter requested values.
Note: Program entries can be changed any time before the Program Review is
d
f
30Hospira GemS tar® System Operating Manual
complete. Press [BACK-UP] to return to previous program displays and make
changes as desired.
3Review the program, if required. Press to begin review. Press [
when done.
4
SAVING PROGRAM displays. The pump stores the program in memory and then
STOP mode.
enters
Tip: If you enter an incorrect number during programming, press [CHANGE]
BEFORE pressing [YES/ENTER] to reset the number to zero. Then enter the
correct number.
YES/ENTER]
Page 38
Chapter 2. Basic Operation
Programming Tips
The following table explains how to use the keypad during programming:
KeyFunction
[
0] - [9]Press the number keys to select menu items and to enter numeric values.
Note: Some menu items require pr essing two number keys to select that item.
For example, when selecting a unit of measure for a weight dosed program, the
menu may display as:
Press to accept entries or to continue when done viewing a display.
Press during programming and alarm conditions to display information
about the current condition.
0][4].
The DOWN arrow toggles between units of measure when entering
concentrations. Scrolls through program review and help messages.
The UP arrow inserts a decimal point in a numeric value.
When entering a numeric value during programming, press to reset the
value to zero.
Press to return to a previous display, to exit the Options Menu, or to exit
HELP mode.
the
Hospira GemStar® System Operating Manual 31
Page 39
Chapter 2. Basic Operation
Starting a New Container
Note: The
Refer to Chapter 10 Keypad Lock, page 114 for more information.
To repeat the current program with a new container:
1From the
2Select
3If Program Review is desired, press [OPTIONS] then press [1] to begin review.
4Press [
NEW CONTAINER
STOP
NEW CONTAINER
CONTAINER
Note: In Weight Dosed, mL, and Pain Management therapies, the VTBI can be
titrated without adding volume to initial container amount.
START
for several seconds and clears the amount infused from memory.
] to begin delivery.
GemStar® Pump Set
The GemStar® Pump Set is a sterile, single-use, disposable pump set. It is designed
for one-time use only and should be properly discarded after use.
Fluid path and areas beneath undisturbed protective set
covers are sterile and nonpyrogenic in the intact unit
package.
Contact a Hospira representative for more information on selecting the appropriate
pump set configuration. Additional disposable components, such as air-eliminating
filters and extension sets, may be added to the line as required. To use a GemStar
Pump Set, follow the instructions included with the set.
function is not available when the pump is in
mode, press [
from the Change Menu. The pump displays
CHANGE
] to display the Change Menu.
STERILE
FULL
NEW
Lock.
2
®
Note: GemStar® Pump Sets contain a pressure-activated anti-siphon valve to reduce
the potential for gravity flow at a typical 36-inch head height.
®
Ensure slide clamp or lower CAIR
False occlusion alarms and fluid delivery inconsistencies may occur when pumping
viscous fluids in microbore tubing at rates greater than 500 mL/hr.
The system may underdeliver, and the system delivery accuracy may be reduced by as
much as 5% when using the following:
•viscous fluids
•nonvented syringes
•PCA vials (except in a lockbox with a spring assist mechanism)
32Hospira GemStar® System Operating Manual
clamp is open prior to inserting cassette.
Page 40
A typical GemStar® Pump Set is shown below:
Chapter 2. Basic Operation
1Piercing Pin6Tubing
2Proximal End7Slide Clamp
3Cassette8Secure Lock
4Flow Stop (open / close ) 9Pressure-activated Anti-Siphon valve
5Distal End
Hospira GemStar® System Operating Manual 33
Page 41
Chapter 2. Basic Operation
WARNING: ALWAYS prime the pump set to remove air from the
cassette, tubing, and injection sites prior to connecting to
the patient. ALWAYS disconnect the pump set from the
patient prior to priming or purging. Arrange tubing, cords,
and cables to minimize the risk of patient strangulation or
entanglement.
Failure to use the pressure-activated anti-siphon valve
may result in unrestricted flow if the cassette is not
properly installed in the pump.
CAUTION: USE ONLY GemStar® Pump Sets with the GemStar® Pump. Use of
unauthorized sets may result in injury to the patient or damage to
the pump.
To prevent contamination, use aseptic technique with all fluid-path
connections. Remove protective coverings as assembly progresses.
Opening the Pump Set
1Open the GemStar
remove the contents.
®
Pump Set package and
2Remove the protective covers from the fluid
container administration port and the pump set.
34Hospira GemS tar® System Operating Manual
Page 42
Chapter 2. Basic Operation
Loading the Cassette
1Identify the four black cassette latches inside the cassette pocket.
2Ensure slide clamp or lower
CAIR
®
clamp is open prior to
inserting cassette.
3Line up the cassette as shown
(with the flow stop button facing
the pump).
4Insert the cassette into the cassette
pocket by pushing firmly along the
entire cassette until it is firmly
seated.
Hospira GemStar® System Operating Manual 35
Page 43
Chapter 2. Basic Operation
5Verify that all four cassette latches
are securely holding the cassette.
CAUTION: Ensure the cassette
is properly installed. When
properly installed, all four
cassette latches are visibly
holding the cassette securely in
the cassette pocket.
6Place the tubing in the
tubing channel.
Note: The pump cannot be
d
powered off during a CHECK
CASSETTE alarm (except during
PROGRAMMING mode). Refer to
Chapter 12 Troubleshooting for
information on clearing a CHECK
CASSETTE alarm.
Releasing the Cassette
1Close the CAIR® (roller) clamp or
slide clamp on the distal line.
2Push down on the cassette release
(large black button) on the top of
the pump.
Note: Ensure the tubing is behind
d
36Hospira GemS tar® System Operating Manual
the tubing guide (at the distal
end).
3Remove the cassette.
Note: When the cassette is
d
released from the pump, the
cassette flow stop automatically
closes, preventing free flow.
Page 44
Chapter 2. Basic Operation
Priming a GemSta r® Pump Set
Priming Pump Sets WITHOUT a Drip Chamber
WARNING: Failure to use the pressure-activated anti-siphon valve
may result in unrestricted flow if the cassette is not
properly installed in the pump.
1Push down on the flow stop to
close the cassette.
2Hold the container with the
administration port toward the
ceiling and insert the piercing pin.
3Push down on the flow stop to
open (priming position) the
cassette.
4Invert the cassette approximately
45°.
5While holding the rotated cassette,
squeeze or roll the container to
remove air from the container and
cassette.
6Return the cassette to the upright
position. Continue priming until
fluid fills the tubing.
Hospira GemStar® System Operating Manual 37
Page 45
Chapter 2. Basic Operation
7Push down on the flow stop to
close the cassette.
8Ensure slide clamp is open prior to
inserting cassette.
9Load the cassette.
CAUTION: Ensure all four
cassette latches are clearly
visible after the cassette is
installed. Refer to Loading the Cassette on page 35.
Note: Ensure the tubing is behind
d
the tubing guide (at the distal
end).
CAUTION: When priming is
complete, ensure no fluid flows
at the distal end of the pump set.
If flow is observed, DO NOT use
the pump set.
10 Connect the pump set to the
patient-access device.
Priming using [
allowed before [
or after one of the following
events occur when using Pain
Management therapy:
6Complete Review of New
Program
6New Container
6Air-in-Line Alarm
PURGE] is only
START] is pressed
38Hospira GemS tar® System Operating Manual
Page 46
Chapter 2. Basic Operation
Priming Pump Sets WITHOUT a Drip ChamberUsing the PURGE Key
WARNING: Failure to use the pressure-activated anti-siphon valve
may result in unrestricted flow if the cassette is not
properly installed in the pump.
d
Note: The PURGE function is available only if there is a program in the pump.
1Push down on the flow stop to
close the cassette.
2Hold the container with the
administration port toward the
ceiling and insert the piercing pin.
3Ensure slide clamp is open prior to
inserting cassette.
4Load the cassette.
CAUTION: Ensure all four
cassette latches are clearly
visible after the cassette is
installed. Refer to Loading the Cassette on page 35.
Hospira GemStar® System Operating Manual 39
Page 47
Chapter 2. Basic Operation
Note: Ensure the tubing is behind
d
the tubing guide (at the distal
end).
5Program desired therapy.
CAUTION: When priming is
complete, ensure no fluid
flows at the distal end of the
pump set. If flow is observed,
DO NOT use the pump set.
9Connect the pump set to the
patient-access device.
6Press [
7The pump displays
8While holding the container with
d
PURGE].
PRIME THE
SET?
the administration port toward the
ceiling, press and hold [
Continue using the [
key until fluid fills the tubing.
Ensure all air is removed
from the container, cassette,
and tubing.
Press [YES/ENTER].
PURGE].
Note: Purge rate is 250 mL/hr.
The pump purges for up to 2
minutes each time you press and
PURGE].
hold [
PURGE]
40Hospira GemS tar® System Operating Manual
Page 48
Chapter 2. Basic Operation
Priming Pump Sets WITH a Drip Chamber
WARNING: Failure to use the pressure-activated anti-siphon valve
may result in unrestricted flow if the cassette is not
properly installed in the pump.
When using a vented chamber, ALWAYS use an
air-eliminating filter OR set the air sensitivity to
either ON or 2 mL.
1Push down on the flow stop to
close the cassette.
2Insert the piercing pin into the
container. Suspend the container
on an I.V. Pole.
3Squeeze the drip chamber until
filled halfway.
4Push down on the flow stop to
open (priming position) the
cassette.
®
5Ensure the CAIR
(roller) clamp
is open prior to inserting cassette.
6Invert the cassette approximately
45°.
Hospira GemStar® System Operating Manual 41
Page 49
Chapter 2. Basic Operation
7While holding the rotated cassette,
remove air from the container and
cassette.
8Return the cassette to the upright
position. Continue priming until
fluid fills the tubing.
9Push down on the flow stop to
close the cassette.
11 Load the cassette.
CAUTION: Ensure all four
cassette latches are clearly
visible after the cassette is
installed. Refer to Loading the Cassette on page 35.
10 Ensure lower CAIR® clamp is
open prior to inserting cassette.
42Hospira GemS tar® System Operating Manual
Page 50
Note: Ensure the tubing is behind
d
the tubing guide (at the distal
end).
CAUTION: When priming is
complete, ensure no fluid flows
at the distal end of the pump set.
If flow is observed, DO NOT use
the pump set.
12 Connect the pump set to the
patient-access device.
Chapter 2. Basic Operation
Hospira GemStar® System Operating Manual 43
Page 51
Chapter 2. Basic Operation
Priming Pump Sets WITH a Drip Chamber Using the PURGE Key
WARNING: Failure to use the pressure-activated anti-siphon valve
may result in unrestricted flow if the cassette is not
properly installed in the pump.
When using a vented set, ALWAYS use an air-eliminating
filter OR set the air sensitivity to either ON or 2 mL.
d
Note: The Purge function is available only if there is a program in the pump.
1Push down on the flow stop to
close the cassette.
2Insert the piercing pin into the
container. Suspend the container
on an I.V. Pole.
3Squeeze the drip chamber until
filled halfway.
4Push down on the flow stop to
open (priming position) the
cassette.
5Ensure the lower CAIR
clamp is open prior to inserting
cassette.
®
(roller)
44Hospira GemS tar® System Operating Manual
Page 52
6Load the cassette.
Chapter 2. Basic Operation
7Program the desired therapy.
d
CAUTION: Ensure all four
cassette latches are clearly
visible after the cassette is
installed. Refer to Loading the Cassette on page 35.
Note: Ensure the tubing is behind
the tubing guide (at the distal
end).
8Press [
9The pump displays
10 Press and hold [
d
PURGE].
SET?
Press [YES/ENTER].
Note: Purge rate is 250 mL/hr.
The pump purges for up to 2
minutes each time you press and
PURGE].
hold [
Continue using the [
until fluid fills the tubing. Ensure
all air is removed from the
container, cassette, and tubing.
CAUTION: When priming is
complete, ensure no fluid flows
at the distal end of the pump set.
If flow is observed, DO NOT use
the pump set.
PRIME THE
PURGE].
PURGE] key
11 Connect the pump set to the
patient-access device.
Hospira GemStar® System Operating Manual 45
Page 53
Chapter 2. Basic Operation
Priming Pump Sets Using Pre-filled Vials
WARNING: Failure to use the pressure-activated anti-siphon valve
may result in unrestricted flow if the cassette is not
properly installed in the pump.
A cracked vial might not show evidence of leakage until
delivery pressure is applied.
d
When using the GemStar
(SAM) Lockbox, USE ONLY GemStar
designated for use with the GemStar
Failure to use designated sets could result in unrestricted
flow.
CAUTION: When using PCA vials (except in a lockbox with a spring assist
mechanism), the system may underdeliver. System delivery
accuracy could be reduced by as much as 5%.
When using pre-filled vials, ensure vial and injector are securely
locked in lockbox before beginning delivery.
USE ONLY purge priming with the GemStar
Spring Assist Mechanism (SAM).
Note: When using the GemStar® Lockbox with SAM, refer to the SAM pump set
labeling and SAM lockbox insert for instructions.
1If using a pump set with the
Y-extension, use the slide clamp to
close off the KVO (Keep Vein
Open) side of the tubing.
®
Spring Assist Mechanism
2Open the upper and lower slide
clamps.
®
Pump sets
®
SAM Lockbox.
®
Lockbox with
46Hospira GemS tar® System Operating Manual
Page 54
Chapter 2. Basic Operation
3Push down on the flow stop to
close the cassette.
4Connect the pump set to a Hospira
pre-filled vial and injector.
5Push down on the flow stop to
open (priming position) the
cassette.
air is cleared from the pre-filled
vial, injector, cassette, and tubing.
8Push down on the flow stop to
close the cassette.
9If the set includes a KVO
Y-extension, close the slide clamp
below the pre-filled vial and open
the slide clamp on the KVO side.
Prime the KVO Y-extension.
6Hold the pre-filled vial and set
upright (vertically) on a flat
surface with the pump set
extending from the top.
7Slowly push down on the injector
until fluid flows at the distal end of
the tubing, which indicates that all
Hospira GemStar® System Operating Manual 47
10 Ensure slide clamp is open on
KVO side prior to inserting
cassette.
Page 55
Chapter 2. Basic Operation
11 Load the cassette.
CAUTION: Ensure all four
cassette latches are
clearly visible after the
cassette is installed.
Refer to Loading the Cassette on page 35
.
CAUTION: When priming is
complete, ensure no fluid flows
at the distal end of the pump set.
If flow is observed, DO NOT use
the pump set.
12 Connect the pump set to the
patient-access device.
Note: Ensure the tubing is behind
d
48Hospira GemS tar® System Operating Manual
the tubing guide (at the distal
end).
Page 56
Chapter 2. Basic Operation
Priming Pump Sets with Extensions
WARNING: Failure to use the pressure-activated anti-siphon valve
may result in unrestricted flow if the cassette is not
properly installed in the pump.
1Push down on the flow stop to
close the cassette.
2Hold the container with the
administrative port toward the
ceiling and insert the piercing pin.
3Push down on the flow stop to
open (priming position) the
cassette.
4Invert the cassette approximately
45°.
5While holding the rotated cassette,
squeeze or roll the container to
remove air from the container and
cassette.
6Return the cassette to the upright
position. Continue priming until
fluid fills the tubing.
Hospira GemStar® System Operating Manual 49
Page 57
Chapter 2. Basic Operation
7Push down on the flow stop to
close the cassette.
8Connect the extension set to the
pump set.
9Ensure the slide clamp is open
prior to inserting cassette.
d
Note: Ensure the tubing is behind
the tubing guide (at the distal
end).
10 Load the cassette.
11 Program the desired therapy.
12 Press [
CAUTION: Ensure all four
cassette latches are clearly
visible after the cassette is
installed. Refer to Loading the Cassette on page 35.
50Hospira GemS tar® System Operating Manual
13 The pump displays
PURGE].
SET?
Press [YES/ENTER].
PRIME THE
Page 58
14 Press and hold [PURGE].
Note: Purge rate is 250 mL/hr.
d
The pump purges for up to 2
minutes each time you press and
PURGE].
hold [
Continue using the [
until fluid fills the tubing. Ensure
all air is removed from the
container, cassette, and tubing.
CAUTION: When priming is
complete, ensure no fluid flows
at the distal end of the pump set.
If flow is observed,
DO NOT use the pump set.
15 Connect the pump set to the
patient-access device.
Chapter 2. Basic Operation
PURGE] key
Hospira GemStar® System Operating Manual 51
Page 59
Notes
52Hospira GemS tar® System Operating Manual
Page 60
Total Parenteral Nutrition (TPN)
- Not Applicable to List 13088
Therapy Features and Specifications
This therapy allows programming of Total Parenteral Nutrition (TPN) protocols.
6Taper Up and Taper Down may be programmed to gradually increase and
decrease the infusion rate at the beginning and end of a therapy. If Taper Down is
selected during programming, the pump allows auto tapering if the continuous
delivery is stopped before the programmed volume has been infused.
6Set an optional KVO (Keep Vein Open) during programming.
Tip: The Programming Worksheet at the end of this chapter lists the minimum and
f
maximum values allowed for this protocol. Use copies of the programming
worksheet to record program values before programming the pump.
TPN Programming Tips
The following table explains how to use the keypad during TPN programming:
Chapter 3
KeyFunctionKeyFunction
[0] - [9]Selects menu items and
enters numeric values.
Accepts entries or continues
when done viewing a display.
Displays information during
programming and alarm
conditions.
Clears a numeric value (resets
to zero) during programming.
Hospira GemStar® System Operating Manual 53
The DOWN arrow toggles
between units of measure
when entering concentrations.
Scrolls through program review
and help messages.
The UP arrow inserts a decimal
point in a numeric value.
Returns to a previous display,
exits the Options Menu, and
exits the
HELP mode.
Page 61
Chapter 3. Total Parenteral Nutrition
TPN Programming
1Select TPN from the Programming Menu.
2Select the TPN delivery mode.
3Enter the total VTBI (Volume To Be Infused).
d
d
d
d
Note: Maximum VTBI is 9600 mL minus any taper amount.
4Enter the Taper Up and Taper Down times (if applicable).
Note: Time is entered in hours:minutes (00:00). For one hour thirty minutes,
press [1] [3] [0]; not [9] [0].
5Enter the total time to infuse.
6Enter the KVO rate. If no KVO is desired, press [
displayed is zero.
Note: KVO runs after Taper Down until the total container is
infused.
7Enter the container size (VTBI + KVO).
Note: If KVO is programmed, increase the container size to accommodate the
desired KVO volume.
8Select the air sensitivity, if required.
9Review the program. Press to begin review.
10 If using the
WARNING: ALWAYS disconnect the pump set from the patient before
PURGE key to prime the pump set, press [PURGE] now.
priming or purging.
YES/ENTER] while the value
54Hospira GemS tar® System Operating Manual
Page 62
Changing a TPN Program
Chapter 3. Total Parenteral Nutrition
d
Can be changed before
is pressed
•VTBI
•Taper times
•Total TPN time
Can be changed at any
time
•KVO rate
•Increase container size
only
•Air sensitivity
Cannot be
changed
•Delivery mode
Note: Y ou cannot change a pr ogram when the pump is in FULL or CONTAINER
Lock.
To change a program:
1From the
2As each message displays, make the desired changes or press [
STOP mode, press [CHANGE] and select CHANGE PROGRAM.
YES/ENTER] to
accept the current setting.
3Review the program. Press to begin review.
4Press [
START] to begin infusion.
Hospira GemStar® System Operating Manual 55
Page 63
Chapter 3. Total Parenteral Nutrition
TPN Auto Taper Down
Auto Taper Down is available only if a Taper Down time was entered during
programming. Auto Taper Down may be activated when the continuous delivery
portion of a TPN with Taper Down protocol is stopped. Auto Taper Down is not
available while the pump is in an alarm condition.
d
1Press [
2Enter the Auto Taper Down time.
3Press [
STOP] during TPN delivery.
The pump displays
•Press [NO] to return to the STOP mode. Then press [START] to continue
therapy.
•Press [
The maximum Auto Taper Down time is three hours or two times the remaining
volume divided by the continuous infusion rate, whichever is less.
Note: Time is entered in hours:minutes (00:00). For one hour thirty minutes,
press [1] [3] [0]; not [9] [0].
Taper Down.
YES/ENTER] to set Auto T aper Down.
YES/ENTER] to return to the RUN mode, then press [START] to begin Auto
DO YOU WANT AUTO TAPERING?
56Hospira GemS tar® System Operating Manual
Page 64
Chapter 3. Total Parenteral Nutrition
GemStar
®
TPN Programming Worksheet
GIVE:
ParametersSelections and/or Values to be
Entered
TherapyTPNN/A
Delivery Mode
VTBI (Volume To Be
Infused)
Taper UpNo
Taper DownNo
Total Time to Infuse
KVO (Keep Vein Open)No KVO Rate: 0.0
Continuous
Continuous + Taper Down
Continuous + Taper Up
Continuous + Taper Up and Down
mL1 - 9600 mL*
Yes hrs:min1 min - 3 hrs
Yes hrs:min1 min - 3 hrs**
hrs:min1 min - 24 hrs
Programmable
Ranges
N/A
Yes KVO Rate:1.0 - 5.0 mL/hr
Container Size
(VTBI + KVO) mL9999 mL max
Air Sensitivity***
* Maximum VTBI is 960 0 mL minus any taper amount.
** Programmable ranges for taper down are 1 minute to 3 hours or 2 times the remaining volume
divided by the continuous delivery rate, whichever is less.
*** Air Sensitivity may not show in all setting configurations.
Hospira GemStar® System Operating Manual 57
OnAlarms at approx. 0.5 mL of air
2 mLAlarms at approx. 2 mL of air
OffNo alarm
N/A
Page 65
Notes
58Hospira GemS tar® System Operating Manual
Page 66
Chapter 4
Pain Management
- Not Applicable to List 13086
Therapy Features and Specifications
This therapy allows programming of a continuous only , bolus only, or continuous with
bolus delivery .
6In Co ntinuous + Bolu s protocols, yo u can enter zero for either the continuous rate
or the bolus dose when entering or changing a program. The continuous rate and
bolus dose cannot be zero at the same time.
6The default bolus delivery rate is 125 mL/hr.
d
Hospira GemStar® System Operating Manual 59
Note: The delivery rate is administered in increments of .01.
6The maximum continuous delivery rate is 25 mL/hr.
6A loading dose may be set during programming, which can be delivered after
programming and priming are complete. The default loading dose rate is 125
mL/hr. Rate, bolus dose, concentration can be programmed in increments of 0.01.
6Priming u si ng [ PURGE] is only allowed before [START] is pressed or after one of
the following events occur:
6Complete Review of New Program
6New Cont ain er
6Air-in-Line Alarm
6VTBI can be titrated.
CAUTION:When using PCA vials (except in a lockbox with a spring assist
mechanism), the system may underdeliver. System delivery
accuracy could be reduced by as much as 5%.
Page 67
Chapter 4. Pain Management
Note: New Container is allowed only when at least 1/2 of the container volume has
d
f
been infused or 1 hour has elapsed since starting the infusion.
The use of a lockbox is recommended with pain management protocols. Refer to
Chapter 15 Optional System Components, page 165, for instructions on using a
lockbox with the GemStar® Pump.
Tip: The Programming W ork sheet at the e nd of this chap ter list s the min imum and
maximum values allowed for this therapy. Use copies of the programming
worksheet to record program values before programming the pump.
Pain Management Programming Tips
The following table explains how to use the keypad during pain management
programming:
KeyFunctionKeyFunction
[0] - [9]Selects menu items and
enters numeric values.
Accepts entries or continues
when done viewing a display.
Displays information during
programming and alarm
conditions.
Clears a numeric value (resets
to zero) during programming.
60Hospira GemStar® System Operating Manual
The DOWN arrow toggles
between units of measure
when entering concentrations.
Scrolls through program review
and help messages.
The UP arrow inserts a decimal
point in a numeric value.
Returns to a previous display,
exits the Options Menu, and
exits the
HELP mode.
Page 68
Pain Management Programming
1Select PAIN MANAGEMENT from the Programming Menu.
2Select the delivery mode.
Note: If Continuous + Bolus is selected, the continuous delivery rate or bolus dose
d
d
can be set to zero without reprogramming the pump. Press [
[CHANGE], then select CHANGE PROGRAM.
3Select the unit of measure and concentration as required.
4Set the continuous delivery rate.
5Program a loading dose as required.
6Program a bolus dose with lockout time and dose limit as required.
Regarding a 1-Hour or 4-Hour Dose Limit: The hour dose limit is the
maximum volume (continuous plus bolus or bolus only amount) that can be
delivered over the selected period. When the programmed dose limit is reached,
the continuous delivery stops and a new bolus is not allowed. As the oldest
delivery amounts age out of the 1-hour or 4-hour dose limit record, the pump
accepts bolus requests or resumes continuous delivery (if programmed).
Note: If a 1-hour or 4-hour dose limit is programmed, a bolus in pr ogr ess will stop
when the amount infused reaches the dose limit.
Loading doses are included in a 1-hour or 4-hour dose limit. A dose limit in effect
could prevent the delivery of a loading dose.
Chapter 4. Pain Management
STOP] and press
7If programmed in units of mcg or mg, the line displaying the units flashes. Enter
the container size in proper unit measure.
d
Hospira GemStar® System Operating Manual 61
Note: To enter the total in mL when programming a concentration, press to
move to the mL line. Enter the mL value when
8Select the air sensitivity, if required.
9Review the program. Press to begin review.
10 If using the PURGE key to prime the pump set, press [
WARNING: ALWAYS disconnect the pump set from the patient before
priming or purging.
mL flashes on the display.
PURGE] now.
Page 69
Chapter 4. Pain Management
Bolus Delivery
Set bolus doses and delivery limits during programming. The default bolus dose
delivery rate in a Pain Management therapy is 125 mL/hr .
A bolus cord is available for patient bolus requests. Refer to Connecting the Bolus Cord in Chapter 15 Optional System Components, on page 168, for more
information.
Delivering a Bolus Dose
d
From the
button on the end of the bolus cord to begin bolus dose delivery. The pump flashes
BOLUS DELIVERY on the display and the amount infused accrues as the delivery
progresses.
When the bolus dose is complete, the pump does the following:
6Records the bolus request and delivery amount to the event and bolus history logs
6Adds the delivered boluses to the total bolus amount in the program amounts
6Resets the bolus lockout time (if applicable) and ignores other bolus requests until
6Continues programmed infusion
Note: ALL bolus requests are recorded to the event and bolus history logs.
If the bolus delivery does not start, it could be locked out by one of the following
conditions:
6Bolus lockout period
6Loading dose delivery in progress
6Bolus dose delivery in progress
6Bolus per-hour limit
61-hour or 4-hour dose limit
RUN mode, press either the bolus button [+] on the top of the pump or the
the lockout time has elapsed
62Hospira GemStar® System Operating Manual
Page 70
Chapter 4. Pain Management
Bolus Dose Interruptions
Bolus deliveries may be interrupted by:
6Pressing [STOP]
6An alarm condition
To continue bolus delivery after an interruption:
1Press [START]. The pump displays COMPLETE BOLUS NOW?
2Respond as follows:
To do this:Press this key:Bolus lockout time is set to:
Deliver the remaining
bolus amount
Clear the undelivered
bolus amount
The time of delivery completion
The time the bolus was
interrupted
Hospira GemStar® System Operating Manual 63
Page 71
Chapter 4. Pain Management
Loading Dose Delivery
You can set a loading dose during programming, if desired. A loading dose can be
delivered after programming and priming are complete.
d
Note: Loading dose amounts are included in a 1-hour or 4-hour dose limit.
Delivering a Loading Dose
1After priming and programming, press [
2The pump displays
to begin loading dose.
When the loading dose is complete, the pump does the following:
6Automatically begins programmed infusion
6Records the dose amount to the history log
6Resets the bolus lockout time (if applicable) and ignores other bolus requests until
the lockout time has elapsed
Loading Dose Interruptions
Loading dose deliveries can be interrupted by:
6Pressing [STOP]
6An alarm condition
To continue a loading dose:
1Resolve the interruption.
DELIVER THE LOADING DOSE NOW? Press [YES/ENTER]
START].
2Press [
64Hospira GemStar® System Operating Manual
START]. The pump display s COMPLETE THE LOADING DOSE?
Page 72
Chapter 4. Pain Management
3Respond as follows:
To do this:Press this key:
Deliver the remaining loading dose amount
Clear the remaining loading dose amount
Changing a Pain Management Program
Can be changed at any timeCannot be ch an ge d
d
•Delivery rate
•Bolus
•Bolus lockout
•Container size
•Air sensitivity
•Dose limit amount
•Delivery mode
•Unit of measure
•Concentration
•Dose limit type
Note: You cannot change a program when the pump is in FULL or CONTAINER
Lock. With a
RATE CHANGE Lock, the continuous delivery rate and bolus amount
can be changed within the set ranges.
Although the delivery mode cannot be changed, if Continuous + Bolus is
programmed, the continuous rate or bolus amount can be changed to zero. The
minimum values set with a
RATE CHANGE Lock could restrict this function.
To change a program:
1From the
2As each message displays, make the desired changes or press [
STOP mode, press [CHANGE] and select CHANGE PROGRAM.
YES/ENTER] to
accept the current setting.
3Review the program. Press to begin review.
4Press [
START] to begin delivery.
Hospira GemStar® System Operating Manual 65
Page 73
Chapter 4. Pain Management
GemStar
®
Pain Management Programming Worksheet
GIVE:
ParametersSelections and/or Values to be
Entered
TherapyPain ManagementN/A
Delivery Mode
Unit of
Measure*
Continuous
Rate*
Loading Dose
Bolus Dose*0 - 2500 mg
Bolus Lockout
Optional Dose
Limit
Continuous
Bolus Only
Continuous + Bolus
Concentration: mg/mL
Concentration: mcg/mL
mL
No
Yes Amount:
minutes1 - 999 min
4-hour limit Amount:
Programmable Ranges
N/A
0.01 - 100 mg/mL
1 - 1000 mcg/mL
0.1 - 2500 mg/hr
1 - 25000 mcg/hr
0.1 - 25.0 mL/hr
0.1 - 2500 mg
1 - 25000 mcg
0.1 - 25.0 mL
0 - 25000 mcg
0 - 25.0 mL
4X rate - 400 mL, 40000 mg,
400000 mcg*
1-hour limit Amount:
# boluses/hour Number:
No dose limit
Rate - 100 mL, 10000
mg, 100000 mcg*
1 - 12
*For continuous plus bolus delivery.
Note: For Bolus Only, the 1-hour or 4-hour dose limit minimum is 1 bolus dose.
Container Size0.1 - 99999 mg
1 - 9999000 mcg
0.1 - 9999 mL
66Hospira GemStar® System Operating Manual
Page 74
Chapter 4. Pain Management
ParametersSelections and/or Values to be
Entered
Air Sensitivity**
* The delivery rate is programmable and is delivered in increments of .01.
** Air Sensitivity may not show in all setting configurations.
d
Note: Titrating VTBI does not change initial container volume.
On Alarms at approx. 0.5 mL of air
2 mLAlarms at approx. 2 mL of air
Off No alarm
Programmable Ranges
N/A
Hospira GemStar® System Operating Manual 67
Page 75
Notes
68Hospira GemStar® System Operating Manual
Page 76
- Not Applicable to List 13088
Therapy Features and Specifications
This therapy allows programming of multiple doses at regular intervals.
6Optional KVO (Keep Vein Open) can be set during programming.
6The pump can be disconnected from the patient between dose deliveries. A call
back alarm can be set to alert the patient or caregiver to reconnect the pump before
the next scheduled delivery begins. The pump must remain powered on for the
call back alarm to function. At the set time before each dose, the pump produces
an audible and visual alarm and flashes
Tip: The Programming Worksheet at the end of this chapter lists the
f
minimum and maximum values allowed for this protocol. Use copies of the
programming worksheet to record program values be fore programming the
pump.
Intermittent Programming Tips
The following table explains how to use the keypad during intermittent programming:
Chapter 5
Intermittent
START on the display.
KeyFunctionKeyFunction
[
0] - [9]Selects menu items and
enters numeric values.
Accepts entries or continues
when done viewing a display.
Displays information during
programming and alarm
conditions.
Clears a numeric value (resets
to zero) during programming.
Hospira GemStar® System Operating Manual69
The DOWN arrow toggles
between units of measure
when entering concentrations.
Scrolls through program review
and help messages.
The UP arrow inserts a decimal
point in a numeric value.
Returns to a previous display,
exits the Options Menu, and
exits the
HELP mode.
Page 77
Chapter 5. Intermittent
Intermittent Programming
1Select INTERMITTENT from the Programming Menu.
2Enter the dose amount in mL.
3Enter the infusion time for each dose in hours:minutes (00:00).
For example, press [
4Enter the frequency of dose deliveries.
For example, press [
5Enter the total number of doses in the container.
To determine the number of doses in the container, divide the volume in the
container by the dose amount.
6Enter the KVO rate in mL.
KVO delivers before a delayed start time, between doses, and after the last dose
until the total container is infused.
4] [0] if each dose is to deliver for 40 minutes.
4] [0] [0] to start a dose every four hours.
d
d
If no KVO is desired, press [
7Enter the container size.
The pump automatically calculates the minimum container size required to deliver
the number of programmed doses and KVO between doses.
Note: If KVO is desired before a delayed start, increase the container size to
include the desired KVO volume.
8Select the air sensitivity, if required.
Note: DO NOT press [YES/ENTER] after selecting the air sensitivity. If
[YES/ENTER] is accidentally pressed and a call back alarm is not desired, press
[
BACK-UP] to return to the beginning of step 9 and press [NO].
9The pump displays
•Press [NO] if you do not want to set the call back alarm.
•Press [
Enter the time before the next dose that the alarm should sound. For example,
press [
scheduled delivery.
YES/ENTER] to set the call back alarm.
1] [0] to set the alarm to sound 10 minutes before the start of the next
SET CALL BACK ALARM FOR NEXT DOSE?
YES/ENTER] while the value displayed is zero.
70Hospira GemStar® System Operating Manual
Page 78
d
Chapter 5. Intermittent
Note: When programming the call back alarm, be sure to allow enough time for the
patient or caregiver to reconnect the pump to the patient and to press [
BEFORE the scheduled dose start time.
10 The pump displays the current time.
•If the displayed time is correct, press [
•To set the clock, press [
OPTIONS].
YES/ENTER].
START]
11 The pump displays the current time and
•Press [YES/ENTER] to begin delivery when the pump enters the RUN mode.
•Press [
Note: When entering the time in 12-hour format, press or to toggle between AM and
d
d
PM.
12 Review the program. Press to begin review.
13 If using the
WARNING: ALWAYS disconnect the pump set from the patient before
Note: The call back alarm works only when the pump is powered on AND in the
STOP mode. DO NOT power-off the pump while it is disconnected from the patient
if you have set the call back alarm.
When the alarm sounds, press [
Then reconnect the pump to the patient.
Tip: To use a delayed start without specifying a start time:
NO] to set a delayed start time. The pump displays SET START TIME.
Enter the desired start time and press [
PURGE key to prime the pump set, press [PURGE] now.
priming or purging.
SILENCE] to silence the alarm for three minutes.
START NOW?
YES/ENTER] when done.
f
1Press [YES/ENTER].
2Complete the program review.
3Power-off the pump.
4When the pump is powered on again, complete the program review, then
press [
START] to begin delivery.
Hospira GemStar® System Operating Manual71
Page 79
Chapter 5. Intermittent
Intermittent Delivery Interruptions
Missed Start Time
6If the programmed start time is missed, it can be changed through the Options
Menu. Refer to Next Dose (Not available for List 13088) in Chapter 10 Options
Menu, on page 121.
6If the pump remains in the STOP mode and the first dose is missed, press
[
START] to begin delivery immediately and all remaining doses will be shifted.
6If the pump is programmed with a delayed start time and is then powered off
before [
programmed start time.
For example, the current time is 10:00 PM. The user sets the delayed start time for
10:30 PM and then powers off the pump. The user does not power on the pump
until 10:40 PM, which is 10 minutes AFTER the scheduled start time. The pump
does not begin delivery until 10:30 PM the FOLLOWING day.
START] is pressed, the pump will skip to the NEXT occurrence of the
d
Dose Interruptions
To understand how interruptions are resolved, compare the normal Intermittent
delivery diagram below for a 1-hour dose every four hours, to the interruption
diagrams shown in the examples:
6If an Intermittent dose is interrupted and the pump is placed in the RUN mode
during the on or off cycle of the current dose, the dose continues delivery from the
72Hospira GemStar® System Operating Manual
Note: The clock cannot be changed while an Intermittent Program is in
progress.
12123456789101112
Dose Delivered -
Page 80
Chapter 5. Intermittent
point where it was interrupted and all remaining doses are shifted.
For example, in the following diagram, an intermittent dose was started at 2:00.
The dose was interrupted during the current dose cycle and was started again
before the end of the current dose cycle.
12123456789101112
Dose Delivered - Dose Interrupted -
6If an Intermittent dose is interrupted and the pump is placed in the RUN mode
after the off cycle of the current dose, the next dose begins delivery . All remaining
doses are shifted if necessary.
For example, in the following diagram, an intermittent dose was started at 2:00. It
was interrupted during the current delivery cycle at 2:30, but was not started again
until after the current delivery cycle at 6:45:
12123456789101112
Dose Delivered - Dose Interrupted -
6When the dose is complete, the start time for the Next Dose displays. To change
the start time refer to Chapter 10 Options Menu, page 107.
6Any missed Intermittent doses, except the first dose, are added to the end of the
delivery schedule.
Hospira GemStar® System Operating Manual73
Page 81
Chapter 5. Intermittent
Changing an Intermittent Program
d
Can be changed at
any time
•Dose parameters
•KVO rate
•Container size
•Air sensitivity
•Call back alarm
Can be changed if no
dose is in progress
•Next dose start time•Current time
Cannot be changed
after is pressed
Note: You cannot change a program when the pump is in FULL or CONTAINER
Lock.
To change a program:
1From the
STOP mode, press [CHANGE] and select CHANGE PROGRAM.
2As each message displays, make the desired changes or press [YES/ENTER] to
accept the current setting.
3Review the program. Press to begin review.
4Press [
START] to begin delivery.
74Hospira GemStar® System Operating Manual
Page 82
Chapter 5. Intermittent
GemStar
®
Intermittent Programming Worksheet
GIVE:
ParametersSelections and/or Values to be EnteredProgrammable
Ranges
TherapyIntermittentN/A
Dose Amount
0.1 - 9600 mL
Infusion Time
Per Dose
Dose
Frequency
Number of
Doses Per
Container
KVO
(Keep Vein
Open)
Container
Size*
Air
Sensitivity**
Call Back
Alarm
Start Time
No KVO Rate: 0.0
Yes KVO Rate:
OnAlarms at approx. 0.5 mL of air
2 mL Alarms at approx. 2 mL of air
OffNo alarm
No
Yes
Alarm: hrs:min before start
Start immediately after programming
hrs:min1 min - 24 hrs
(Infusion time per
dose + 1 min) to 96 hrs
hrs:min
1 - 999
0.1 -5.0 mL/hr
0.1 - 9999 mL
N/A
1 min to time between
doses
Start time:
Specify AM or PM if not 24-hour clock.
* The pump automatically calculates the minimum container size required to deliver the number of
12:00 AM - 11:59 PM or
00:00 - 23:59
programmed doses and KVO between doses. If KVO is desired before a delayed start, increase the
container size to include the desired KVO volume.
** Air sensitivity may not show in all setting configurations.
Hospira GemStar® System Operating Manual75
Page 83
Notes
76Hospira GemStar® System Operating Manual
Page 84
- Not Applicable to List 13088
Therapy Features and Specifications
This therapy allows programming of a continuous delivery.
6Optional KVO (Keep Vein Open) can be set during programming.
6The rate can be titrated while the pump is infusing.
6An optional secondary container can be “piggybacked” on the primary container
during regular delivery using a GemStar
Piggybacking in this chapter for more information.
Tip: The Programming Worksheet at the end of this chapter lists the minimum
f
and maximum values allowed for this protocol. Use copies of the prog ramming
worksheet to record program values before programming the pump.
Continuous Programming Tips
The following table explains how to use the keypad during continuous programming:
Chapter 6
Continuous
®
Piggyback Pump Set. Refer to
KeyFunctionKeyFunction
[0] - [9]Selects menu items and
enters numeric values.
Accepts entries or continues
when done viewing a display.
Displays information during
programming and alarm
conditions.
Clears a numeric value (resets
to zero) during programming.
Hospira GemStar® System Operating Manual 77
The DOWN arrow toggles
between units of measure
when entering concentrations.
Scrolls through program review
and help messages.
The UP arrow inserts a decimal
point in a numeric value.
Returns to a previous display,
exits the Options Menu, and
exits the
HELP mode.
Page 85
Chapter 6. Continuous
Continuous Programming
1Select CONTINUOUS from the Programming Menu.
2Select the unit of measure and concentration as required.
3Enter the rate.
4Enter the VTBI (Volume to be Infused).
5Enter the KVO rate, if desired.
KVO runs after the programmed VTBI until the total container is infused. If no
KVO is desired, press [
6Enter the container size (VTBI + KVO).
VTBI is the default value on the
Note: If KVO is programmed, increase the container size to include the KVO
d
delivery.
To enter the total in mL when programming a concentration, press to move to
the mL line. Enter the mL size when
7Select the air sensitivity, if required.
8Review the program. Press to begin review.
YES/ENTER] while the value displayed is zero.
CONTAINER display.
mL flashes on the display.
9If using the PURGE key to prime the pump set, press [
WARNING: ALWAYS disconnect the pump set from the patient before
priming or purging.
78Hospira GemStar® System Operating Manual
PURGE] now.
Page 86
Rate Titration
To titrate the rate while the pump is infusing:
1In the
For example: if the current rate is 50 mL/hr, press [
change the rate to 100 mL/hr.
Note: You have approximately five seconds after pressing a key to press another
d
numeric key, the decimal key, or [YES/ENTER]. If no key is pressed within five
seconds, the rate does not change and the delivery continues as previously
programmed.
2
NEW RATE displays with the new value.
•Press [
delivery at the new rate.
•Press [
Piggybacking
Piggybacking with the GemStar® Pump requires a GemStar® Primary Piggyback
Pump Set with a proximal Y-site backcheck valve and a secondary piggyback set with
an extension hook. The secondary container must be higher than the primary container. Refer to the instructions included with the piggyback pump set for more
information.
Chapter 6. Continuous
RUN mode, press the number keys for the new rate. Press [YES/ENTER].
1][0] [0] [YES/ENTER] to
YES/ENTER] within 10 seconds to accept the change and begin
NO] to continue delivery at the previously programmed rate.
WARNING: When using the pump for secondary delivery (piggybacking),
ensure the fluids being infused are chemically and physically
compatible.
The piggyback rate and piggyback VTBI display until the secondary container has
been delivered. Then the primary infusion continues.
d
Note: The piggyback infusion amount is NOT added to the primary programmed
infusion amount.
Hospira GemStar® System Operating Manual 79
Page 87
Chapter 6. Continuous
Preparing for Secondary Delivery (Piggybacking)
1Ensure a primary piggyback pump set with a
proximal Y-site backche ck valve is in place.
2Use the extension hook from the secondary set
to suspend the primary container from the I.V.
pole.
3Prepare the secondary set according to the
instructions included with the set.
4Attach a needle or blunt cannula (if appropriate)
and prime the secondary set.
5Suspend the secondary container from the I.V.
pole.
Note: When using a secondary container of
d
500 mL or greater, ensure the bottom of the
secondary container is at least 7 inches (17.8
cm) above the fluid level in the primary
container. Use additional extension hooks if
necessary.
6Connect the secondary set to the proximal
Y-site (or proximal luer port).
80Hospira GemStar® System Operating Manual
Page 88
Chapter 6. Continuous
Programming a Secondary Delivery (Piggybacking)
d
f
d
Note: To distinguish the piggyback delivery from the
primary delivery, PIGGYBACK displays and the piggyback
VTBI counts down to completion. For example, a
piggyback rate of 250 mL/hr with a remaining VTBI of
218.5 mL will display as 219 mL.
Tip: Prepare an d set up cont ain ers BEFORE progr amming the pum p for pigg yback
delivery.
1Press [
2Enter the dose amount for the secondary container.
3Enter the infusion time for the secondary container.
4Open the slide clamp or lower CAIR
5Press [
Note: Primary flow stops until the secondary VTBI is delivered; then primary flow
continues. Any remaining secondary fluid will be delivered at the primary rate. The
maximum piggyback rate is 300 mL/hr. For secondary rates above 300 mL/hr, flow
may occur from the primary container.
Changing a Piggyback Delivery
1Press [
CHANGE] and select PIGGYBACKING from the Change Menu.
®
(roller) clamp on the tubing of the
secondary container.
YES/ENTER] to begin piggyback delivery.
CHANGE] and select PIGGYBACKING from the Change Menu.
PIGGYBACK
RATE 250 mL/hr
VTBI 219 mL
2Press [
3Enter the new piggyback dose amount and press [
4Set the infusion time for the secondary container.
5Press [
d
Note: Y ou must know the amount that ha s alr eady been deliver ed. When the piggyback
dose amount is changed, the pump begins a new VTBI countdown.
Hospira GemStar® System Operating Manual 81
CHANGE] to reset the dose amount to zero.
YES/ENTER].
YES/ENTER] to begin piggyback delivery.
Page 89
Chapter 6. Continuous
Canceling a Piggyback Delivery
1Press [
2Press [
3Press [
CHANGE] and select PIGGYBACKING from the Change Menu.
CHANGE] to reset the dose amount to zero.
YES/ENTER] to cancel piggyback delivery.
Alarms Affecting Piggyback Delivery
The following delivery-related Operational Alarms occur during a Piggyback
Delivery:
6Check Cassette
6Air in Line
6Distal Occlusion
6Proximal Occlusion
If an alarm occurs during the programming sequence, piggyback dose programming is
completed.
Changing a Continuous Program
Can be changed at any timeCannot be changed
•Delivery rate
•VTBI
•KVO rate
•Container size
•Air sensitivity
•Piggybacking
•Unit of measure
•Concentration
d
Note: Y ou cannot change a program when the pump is in FULL or CONTAINER Lock.
To change a program:
1From the
2As each message displays, make the desired changes or press [
STOP mode, press [CHANGE] and select CHANGE PROGRAM.
YES/ENTER] to
accept the current setting.
3Review the program. Press to begin review.
4Press [
82Hospira GemStar® System Operating Manual
START] to begin delivery.
Page 90
Chapter 6. Continuous
GemStar
®
Continuous Programming Worksheet
GIVE:
ParameterSelections and/or Values to be
Entered
TherapyContinuousN/A
Unit of Measure
Rate0.1 - 9999 mg/hr
VTBI
(Volume To Be
Infused)
KVO
(Keep Vein
Open)
Container Size
(VTBI + KVO)
Air Sensitivity*
Piggyback Dose
Amount1 - 7200 mL
Piggyback
Infusion Time
*Air Sensitivity may not show in all setting configurations.
Concentration: mg/mL
Concentration: mcg/mL
mL
NoKVO Rate: 0.0
YesKVO Rate:
OnAlarms at approx. 0.5 mL of air
2 mL Alarms at approx. 2 mL of air
OffNo alarm
hrs:min
Programmable Ranges
0.1 - 1000 mg/mL
1 - 1000 mcg/mL
1 - 999900 mcg/hr
0.1 - 1000 mL/hr
0.1 - 99999 mg
1 - 9999000 mcg
0.1 - 9999 mL
0.1 - 5.0 mL/hr
0.1 - 99999 mg
1 - 9999000 mcg
0.1 - 9999 mL
N/A
1 min or
dose amount / 300 mL/hr to 24
hrs or dose amount / 0.1 mL/hr
Hospira GemStar® System Operating Manual 83
Page 91
Notes
84Hospira GemStar® System Operating Manual
Page 92
Weight Dosed
- Not Applicable to List 13088
Therapy Features and Specifications
This therapy allows programming of delivery rate as a function of body weight.
6The air sensit ivity in a Weight Dosed program is automatically set to ON. This
setting can be changed through the Options Menu after programming is complete.
Refer to Air Sensitivity in Chapter 10 Options Menu, on page 117.
6The pump automatically delivers KVO (Keep Vein Open) after the programmed
VTBI (Volume To Be Infused) has completed.
6The rate can be titrated while the pump is infusing.
6VTBI can be titrated.
Tip: The Programming Worksheet at the end of this chapter lists the minimum and
f
maximum values allowed for this protocol. Use copies of the programming worksheet
to record program values before programming the pump.
WARNING: When using a vented drip set, ALWAYS use an air-
eliminating filter or set the air sensitivity to either ON or
2 mL.
Chapter 7
CAUTION: When infusing short-half-life drugs (i.e., those dosed with units of
mcg/kg/min) at very low rates (0.8 mL/hr or less) the rate may be too
slow for that drug. Consider using a lower drug concentration with
these drugs so a pump rate above 0.8 mL/hr can be used.
Hospira GemStar® System Operators Manual 85
Page 93
Chapter 7. Weight Dosed
Weight Dosed Programming Tips
The following table explains how to use the keypad during weight dosed programming:
KeyFunctionKeyFunction
[0] - [9]Selects menu items and
enters numeric values.
Accepts entries or continues
when done viewing a display.
Displays information during
programming and alarm
conditions.
Clears a numeric value (resets
to zero) during programming.
Weight Dosed Programming
1Select WEIGHT DOSED from the Programming Menu.
2Select the unit of measure.
d
Note: If two numbers appear before a menu time (e.g., 02), press both number
keys to select that item.
3Enter the patient weight, if required.
4Enter the concentration:
•Enter the value for the unit of measure when it flashes on the display.
•Press [
YES/ENTER] to confirm.
The DOWN arrow toggles
between units of measure
when entering concentrations.
Scrolls through program review
and help messages.
The UP arrow inserts a decimal
point in a numeric value.
Returns to a previous display,
exits the Options Menu, and
exits the
HELP mode.
f
d
86Hospira GemStar® System Operators Manual
Tip: Press to toggle between units of measure.
5Enter the rate.
Note: The pump automatically displays the mL/hr equivalent even when another
unit of measure is selected.
6Enter the VTBI.
7Review the program. Press to begin review.
Page 94
Chapter 7. Weight Dosed
8If using the PURGE key to prime the pump set, press [PURGE] now.
WARNING: ALWAYS disconnect the pump set from the patient before
priming or purging.
Auto KVO
When the programmed VTBI has completed delivery, Auto KVO delivery begins
automatically and the empty container alarm sounds.
d
Note: Auto KVO delivers even while the empty container alarm sounds. Press
[SILENCE] to quiet the alarm for 2 minutes.
If the weight dosed rate is:Auto KVO delivers at:
Greater than or equal to 1 mL/hr1 mL/hr
Less than 1 mL/hrThe weight dosed rate
Rate Titration
To titrate the rate while the pump is infusing:
1In the
For example: if the current rate is 7.5 mcg/kg/min, press [
change the rate to 10 mcg/kg/min.
Note: You have approximately five seconds after pressing a key to press another
d
numeric key, the decimal key, or [
seconds, the rate does not change and the delivery continues as previously
programmed.
2
NEW RATE displays with the new value.
•Press [
•Press [
RUN mode, press the number keys for the new rate. Press [YES/ENTER].
1][0] [YES/ENTER] to
YES/ENTER]. If no key is pressed within five
YES/ENTER] within 10 seconds to accept the change and begin
delivery at the new rate.
NO] to continue delivery at the previously programmed rate.
Hospira GemStar® System Operators Manual 87
Page 95
Chapter 7. Weight Dosed
VTBI Titration
To titrate the VTBI while the pump is infusing:
1Press [
2Press the number keys for the new VTBI. Press [
d
3
d
CHANGE] and select TITRATE VTBI.
For example: if the current VTBI is 0 mL, press [
mL.
Note: You have approximately five seconds after pressing a key to press another
numeric key, the decimal key, or [
seconds, the VTBI does not change and the delivery continues as previously
programmed.
NEW VTBI displays with the new rate.
•Press [
•Press [
Note: Titrating VTBI does not change initial container volume.
YES/ENTER] within 10 seconds to accept the change and begin
delivery with the new VTBI.
NO] to continue delivery at the previously programmed VTBI.
YES/ENTER]. If no key is pressed within five
Changing a Weight Dosed Program
Can be changed at any timeCannot be chan ged
•Delivery rate
•VTBI
•Patient weight
YES/ENTER].
1][0] to change the VTBI to 10
•Unit of measure
•Concentration (amount of drug)
d
Note: You cannot change a program when the pump is in FULL or CONTAINER Lock.
To change a program:
1From the
2As each message displays, make the desired changes or press [
accept the current setting.
3Review the program. Press to begin review.
4Press [
88Hospira GemStar® System Operators Manual
STOP mode, press [CHANGE] and select CHANGE PROGRAM.
YES/ENTER] to
START] to begin delivery.
Page 96
Chapter 7. Weight Dosed
GemStar® Weight Dosed Programming Worksheet
GIVE:
ParameterSelections and/or Values to be EnteredProgrammable
Note: Numbers increment by hundredths—for example, 0.1 increments to 0.11, then to 0.12.
** If two numbers appear before a menu item (e.g., 02), press both number keys to select that item.
CAUTION: When infusing short-half-life drugs (i.e., those dosed with units of mcg/kg/min) at very low
rates (0.8 mL/hr or less) the rate may be too slow for that drug. Consider using a lower drug
concentration so a pump rate above 0.8 mL/hr can be used.
Hospira GemStar® System Operators Manual 89
Page 97
Chapter 7. Weight Dosed
Programming a Precedex® (dexmedetomidine
hydrochloride injection) Delivery
If configured and used by your facility, you can program a Precedex® delivery
while programming a Weight Dosed therapy.
Note: Titrate rate, loading dose, and titrate VTBI are available in Precedex® therapy.
d
To program a Precedex
1Select
2Select
3The pump displays
4Enter patient weight in kg, and then press [
5Enter the concentration:
6Enter the rate, and then press [
Note: Rate is entered in mcg/kg/hr, and then displays in mL/hr.
7Enter the VTBI in mL, and then press [
8The pump displays
9If programming a loading dose, specify a dose, time, and rate using the key
10 Review the program using the key; when finished, press [
WEIGHT DOSED from the Programming menu.
MCG/KG/HR as the unit of measure.
6Enter the unit of measure for the value flashing on the display.
6Use the key to toggle between the values, and then press [YES/ENTER].
program a loading dose or [
to toggle between the parameters, and then press [
®
delivery:
IS THIS A PRECEDEX THERAPY? Press [YES/ENTER].
YES/ENTER].
YES/ENTER].
YES/ENTER].
PROGRAM A LOADING DOSE? Press [YES/ENTER] to
NO] to review the program.
YES/ENTER].
YES/ENTER].
WARNING: ALWAYS disconnect the pump set from the patient before
priming or purging.
11 If using the PURGE key to prime the pump set, press [
12 Press [
90Hospira GemStar® System Operators Manual
START] to begin the Precedex
®
delivery.
PURGE] now.
Page 98
Chapter 7. Weight Dosed
GemStar® Precedex® (dexmedetomidine hydrochloride
injection) Programming Worksheet
GIVE:
ParameterSelections and/or Values to be EnteredProgrammable
TherapyWeight DosedN/A
Unit of Measuremcg/kg/hrN/A
Precedex®
Therapy?
Patient Weight
Concentration
(Add values are
required. Use
to toggle between
fields)
Rate
VTBI
(Volume To Be
Infused)
Loading Dose
(2 entries are
required. Use
to toggle between
mg and mL entry
points.)
* Package Insert recommends 0.2 to 0.7 mcg/kg/hr. The delivery rate is programmable and is delivered
in increments of .01.
** Package Insert recommends 1 mcg/kg over10 min . Cannot enter less than 10 min.
[YES/ENTER]N/A
kg
mcg
in
mL
mcg/kg/hr
(mL/hr is automatically calculated)
mL
No
Yes
Dose mcg/kg
Time**
Ranges
2.0 - 200.0 kg
1.0- 1000
0.1 - 1000
0.1 - 1000*
0.1 - 9999 mL
0 - 1.0**
Note: Numbers increment by hundredths—for example, 0.1 increments to 0.11,
then to 0.12.
Hospira GemStar® System Operators Manual 91
Page 99
Notes
92Hospira GemStar® System Operators Manual
Page 100
Chapter 8
mL/hr Only
- Not Applicable to List 13088
Therapy Features and Specifications
This therapy allows quick programming of delivery rate in milliliters (mL) only.
6The air sensitivity is aut omatically set to ON. This setting can be changed through
the Options Menu after programming is complete. Refer to Air Sensitivity in
Chapter 10 Options Menu, on page 117.
6The pump automatically delivers KVO (Keep Vein Open) after the programmed
VTBI (Volume To Be Infused) is complete.
6The rate can be titrated while the pump is infusing.
6An optional secondary container can be “piggybacked” onto the primary container
®
during regular delivery using a GemStar
Piggybacking in this chapter for more information.
6A program review is not required and is available only through the Options Menu.
6VTBI can be titrated.
Piggyback Pump Set. Refer to
Tip: The Programming Worksheet at the end of this chapter lists the minimum and
f
maximum values allowed for this protocol. Use copies of the programming worksheet
to record program values before programming the pump.
WARNING: When using a vented set, ALWAYS use an air-eliminating
filter or set the air sensitivity to either ON or 2 mL.
Hospira GemStar® System Operating Manual 93
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.