Hospira GemStar 13088 User manual [gb]

Page 1
®
Infusion Pump
System Operating Manual
For use with the
®
Hospira GemStar Pain Management Pump
List 13088-04-01/02
Hospira, Inc., Lake Forest, IL 60045 430-10984-001 (Rev
. 01/06)
Page 2
Hospira GemStar® Infusion Pump
System Operating Manual
For use with the
Hospira GemStar® Pain Management Pump
List 13088-04-01/02
PLEASE
Read this entire manual before
using the Hospira GemStar® Pump
This manual is designed for use by healthcare professionals, caregivers, and patients. The Hospira Technical Support Operations hotline is available 24 hours a day (in the USA) to provide consultation and technical assistance regarding the Hospira GemStar® Pump.
In the USA: Hospira Technical Support Operations
1-800-241-4002
Outside the USA: Contact your local Hospira
sales office
To order additional copies of this manual
(List 13088) call Hospira Customer Care:
1-877-9HOSPIRA
(1-877-946-7747)
Hospira GemStar Manual 430-10984-001 (Rev. 03/06) i
Page 3
Change History
Part Number Description of Change Pages Changed
430-10984-001 (Rev. 03/06)
Original Issue N/A
ii Hospira GemStar Manual 430-10984-001 (Rev. 03/06)
Page 4
Contents
1. Introduction 1
GemStar® Pump Layout . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
GemStar® Pump Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
GemStar® Pump Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Menu System. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Operating Modes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Indications for Use. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Contraindications for Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Warnings and Cautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
2. Basic Operation 17
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Selecting a Power Source . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Using the GemStar® Docking Station . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Powering On the Pump . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Accessing the Delivery Mode Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Programming the Pump. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Starting a New Container . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
GemStar® Pump Set. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Loading the Cassette. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Releasing the Cassette . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Priming a GemStar® Pump Set. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
3. Pain Management 45
Therapy Features and Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
Pain Management Programming. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
Bolus Delivery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
Rate Titration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
Loading Dose Delivery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
Changing a Pain Management Program . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
Pain Management Programming Worksheet. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
4. Options Menu 55
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
Tips for Using the Options Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
Hospira GemStar Manual 430-10984-001 (Rev. 03/06) iii
Page 5
Contents
Review Program . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
Histories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
Keypad Lock . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
Change Pump Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
Set Clock . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
Print . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
Speed Protocol . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
Connecting to the Data Port . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70
Transferring Data Using a Computer. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
5. Clinician Instructions 77
Clinician Locking Sequence Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78
Program and Deliver a Clinician Activated Loading Dose . . . . . . . . . . . . . . . . . . . 79
Shift Totals and History . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79
6. Troubleshooting 81
Alarm Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 82
Alert Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 91
Other Display Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 94
7. Maintenance 95
Pump Storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 95
Cleaning and Disinfecting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 95
Repair . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 97
8. Operation Test 99
Equipment Required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 99
Test Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100
Performing the Operation Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100
Printing Test Results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 105
Operation Test Checklist . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 107
9. Optional System Components 109
Connecting the Bolus Cord . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 112
Using the GemStar® Lockboxes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 112
Using the Hospira Carrying Cases . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 115
Using the GemStar® Docking Station . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 121
iv Hospira GemStar Manual 430-10984-001 (Rev. 03/06)
Page 6
Contents
10. Pump Sets 123
11. GemStar® Pump Specifications 127
Occlusion Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 130
12. Warranty/Technical Assistance 145
A. Electrical Potentials 147
Guidance on EMC Compatibility . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 147
Information Regarding Electrostatic Discharge (ESD) . . . . . . . . . . . . . . . . . . . . 148
Index 149
IEC Symbols inside back cover
Hospira GemStar Manual 430-10984-001 (Rev. 03/06) v
Page 7
Notes
vi Hospira GemStar Manual 430-10984-001 (Rev. 03/06)
Page 8
Chapter 1
Introduction
The Hospira GemStar® Pain Management Pump is a small and lightweight, single-channel infusion device designed for use in the home, in the hospital, or anywhere electronic infusion is required. The GemStar® Pump can be powered by AC mains adaptor, rechargeable battery pack, docking station, or two disposable AA alkaline batteries. When powered by batteries, the GemStar® Pump is ideal for ambulatory patients.
The following conventions are used throughout this manual:
Convention Use Example
[ALL CAPS] Pump keys Press [YES/ENTER] to continue.
ALL CAPS
WARNING: A WARNING message contains special safety
Display messages The pump displays:
ENTER TO CONFIRM.
information and must be observed at all times. Failure to observe a warning message could be life threatening.
CAUTION: A CAUTION contains information that could prevent
irreversible product damage or hardware failure. Failure to observe a caution could result in serious patient or user injury.
G I G
Note: A NOTE contains information or examples to help explain instructions.
Tip: A TIP contains information that makes the pump easier to use.
Note: Examples are provided only to help explain a pump function and may not be clinically relevant. Graphics are for illustrative purposes only and may not reflect the actual product.
Hospira GemStar Manual 430-10984-001 (Rev. 03/06) 1
Page 9
1. Introduction
GemStar® Pump Layout
Shown approximate size
2 Hospira GemStar Manual 430-10984-001 (Rev. 03/06)
Page 10
1. Introduction
DISPLAY
1
2
3 • Stops infusion.
4 • Starts infusion.
5 • Lights continuously when the pump is powered by AC mains.
6 • Powers the pump on or off. To power-off the pump, press and hold until the
7 • Returns to previous steps during programming.
8
CHANGE
9
10 • Toggles between AM and PM in time-based entries.
11 • Responds “YES” to display questions.
• Displays the pump’s status.
• Silences an alarm.
• This silences an alert for five minutes and an alarm for two minutes.
• Flashes when the pump is powered by external batteries (e.g., battery pack).
• Not lit when using disposable AA batteries.
display clears.
• Exits Help, Change, and Options menus.
• Displays the following menu items from the New Container Change Program New Program
• Resets a numeric value to zero.
• Displays the following Options Menu items from the 1 Review Program 4 Pump Settings 7 Speed Protocol 2 Histories 5 Set Clock 3 Keypad Lock 6 Print
• Scrolls through program review and help messages.
• Toggles between AM and PM in time-based entries.
• DOWN arrow toggles between units of measure when entering concentrations.
• UP arrow enters a decimal point in numeric entries.
• Scrolls back by one screen in program review.
• Accepts entries and goes to the next step.
STOP or RUN mode*:
STOP or RUN mode*:
12 • Responds “NO” to display questions.
13 • Displays information during programming steps and alarm conditions.
• Displays Shift Totals, except during programming steps and alarm conditions.
14
15 [0] - [9] • Selects menu items and enters numeric values. If two numbers appear before a
* Not all menu items are available with all modes.
Hospira GemStar Manual 430-10984-001 (Rev. 03/06) 3
• Clears an
• Primes a GemStar® Pump Set.
menu item (e.g., 02), press both number keys to select that item.
AIR-IN-LINE alarm.
Page 11
1. Introduction
GemStar® Pump Components
Bolus Button Press this button to start a
1
bolus delivery if available in the current program.
Pole Clamp Track Use to connect to an optional
2
pole clamp.
Cassette Release
3
Button
Power Connector Use to connect the AC mains
4
Data Port
5
Connector
Bolus Connector Use to connect a bolus cord.
6
Tubing Channel Place the pump set tubing in
7
Cassette Pocket Load the cassette into this
8
I.V. Tubing Guide Use to hold the I.V. tubing.
9
Battery Pack
10
Connector
Battery Compartment Install two fresh, disposable
11
Push down on this button to release the cassette.
adaptor.
Use to connect the GemStar® serial cable.
this channel.
pocket.
Use to connect the GemStar® Rechargeable Battery Pack.
AA batteries in this compartment.
Top View
Bottom/Side View
4 Hospira GemStar Manual 430-10984-001 (Rev. 03/06)
Page 12
GemStar® Pump Features
Feature Description Reference Page
1. Introduction
Adjustable Settings
Alerts and Alarms
Audible Keypad Response
•Keypad Lock
•New Container
Power-on Self-test
Operation Test
•Program History
•Speed Protocols
Adjusts the distal occlusion pressure limits, air sensor sensitivity, real-time clock, and audible alarm volume.
Signals audible and visual alarms when attention is required.
Sounds a single beep when any key is pressed. If an invalid key is pressed, the pump sounds a flutter tone.
Restricts access to various options. Four lock levels are available.
Repeats the current program without reprogramming the pump.
Performs a self-test each time the pump is powered on.
Performs user-assisted tests to verify the pump is operating properly.
Maintains a history event log, with a date and time stamp of each event for each program. Review the history on the display, or download it to a printer or computer.
Stores up to nine frequently used programs in the pump’s memory for quick access.
Chapter 4 Options Menu
Chapter 6 Troubleshooting
Chapter 1 Introduction,
GemStar® Pump Layout
Chapter 4 Options Menu and Chapter 5 Clinician Instructions
Chapter 2 Basic Operation
Chapter 2 Basic Operation
Chapter 8 Operation Te st
Chapter 4 Options Menu
Chapter 4 Options Menu
Hospira GemStar Manual 430-10984-001 (Rev. 03/06) 5
Page 13
1. Introduction
Menu System
The GemStar® Pump uses a menu system that provides step-by-step instructions through all of the pump’s functions. Refer to the following table for a list of functions in each menu.
Use the keypad to select menu items, respond to display messages, and enter numeric values. For example, press [
2] to select HISTORIES from the Options Menu.
G I
Note: Menu items may vary depending on the pump’s configuration.
Tip: A menu item does not have to be on the display to be selected. For example, when the Options Menu first displays, only the first four menu items are shown. Instead of pressing the down arrow to see the other menu items, press [ directly to SET CLOCK.
5] to go
6 Hospira GemStar Manual 430-10984-001 (Rev. 03/06)
Page 14
1. Introduction
To display this menu:
Delivery Mode Menu
Change Menu Press . 1 NEW CONTAINER
Do this: Menu Items Available
1 Press to display
the Change menu.
2 Select
CHANGE
NEW PROGRAM.
CHANGE
1 CONTINUOUS 2 BOLUS 3 CONT + BOLUS
2 NEW PROGRAM 3 CHANGE PROGRAM
Note: If the current therapy is Weight Dosed or mL/hr Only and the pump is in mode:
3 TITRATE VTBI
will replace
3 CHANGE PROGRAM
on the menu display.
Note: Piggybacking is available only in mL/hr Only and Continuous therapies.
Note: Menu items change depending on when Change Menu is viewed.
Options Menu Press .
1 REVIEW PROGRAM
2 HISTORIES
3 KEYPAD LOCK
4 PUMP SETTINGS
5 SET CLOCK
6 PRINT
7 SPEED PROTOCOL
RUN
For more information on: Refer to the following pages:
Delivery Mode Menu Chapter 3 Pain Management, page 45.
Change Menu For menu item [1], refer to Starting a New Container
in Chapter 2 Basic Operation, on page 30. For menu items [2] and [3], refer to Chapter 3 Pain Management, page 45.
Options Menu Chapter 4 Options Menu, page 55.
Hospira GemStar Manual 430-10984-001 (Rev. 03/06) 7
Page 15
1. Introduction
Operating Modes
The GemStar® Pump has four operating modes:
Mode Description
Help Mode The pump enters the
programming or an alarm condition. The additional information about the current condition.
Programming Mode The pump enters the
powered on or when is pressed.
Run Mode After is pressed, the pump enters the
when not infusing.
Stop Mode When programming is complete, the pump enters the
mode until is pressed.
When the programmed therapy is complete, and during certain alarm conditions, the pump automatically enters the
Press to enter the CHECK CASSETTE alarm.
HELP mode when is pressed during
HELP mode displays
PROGRAMMING mode after the pump is
CHANGE
RUN mode, even
STOP mode.
STOP mode at any time, except during a
STOP
8 Hospira GemStar Manual 430-10984-001 (Rev. 03/06)
Page 16
Indications for Use
Physicians or certified, licensed healthcare professionals should always oversee infusions. Healthcare professionals should always supervise pump users and instruct users in pump operation and troubleshooting. Instruction should include how to prevent intravenous (I.V.) related complications, including how to prevent accidental infusion of air.
The pump is suitable for intravenous, arterial, subcutaneous, short-term epidural infusion of general I.V. fluids and medications. The epidural route is recommended to provide anesthesia or administer analgesia for periods up to 96 hours.
Note: Epidural administration of anesthetics is recommended in the continuous
G
mode only. Epidural administration of analgesics may be delivered by continuous, bolus, or continuous with bolus.
For epidural use, the administration of drugs is restricted to those anesthetic and analgesic drugs approved for continuous epidural administration: e.g., Chloroprocaine Hydrochloride USP, Lidocaine Hydrochloride USP, and Morphine Sulfate Injection USP (Preservative Free).
1. Introduction
WARNING: Delayed respiratory depression following continuous
epidural administration of preservative-free morphine sulfate has been reported.
For epidural administration, the following is recommended:
Nylon or Teflon Pump sets without Y-sites Stickers for the tubing indicating ongoing administration
Note: Facilities practicing epidural anesthesia/analgesia must be staffed and
G
Hospira GemStar Manual 430-10984-001 (Rev. 03/06) 9
equipped to manage cardio-pulmonary resuscitation (CPR). Supplies should include oxygen, naloxone, and other appropriate resuscitative drugs and equipment. Monitor patients continuously (e.g., oximetry) during epidural administration. Frequently observe patient for side effects (for up to 24 hours) following completion of drug administration by the epidural route.
®
catheter
Page 17
1. Introduction
Contraindications for Use
Persons who do not have the mental and physical capability or the emotional stability to properly use the GemStar® Pump should not operate this pump.
DO NOT use the following with the GemStar® Pump System:
Drugs not compatible with silicone rubber or PVC plastic
Drugs not stable under infusion conditions
Warnings and Cautions
Pay attention to the following warnings and cautions when operating the GemStar® Pump. These warnings and cautions are repeated in the manual when relevant to the procedure being discussed. Pay attention to all alert messages.
General Warnings and Cautions
Federal (USA) law restricts this device to sale by or on the order of a physician or
other licensed practitioner.
Manual references to specific values are approximate only, unless indicated
otherwise. Air-in-line sensitivity values are approximate only.
When a proximal occlusion alarm activates, DO NOT power-off the pump.
Closely supervise patients who are likely to be adversely affected by unintended
operations and failures, including interrupted medication or fluid infusion. Ensure provisions for immediate corrective action are available.
Regarding Drugs Used, Pump Sets, and Containers
USE ONLY GemStar® Pump Sets with the GemStar® Pump. Use of
unauthorized sets may result in injury to the patient or damage to the pump.
Some microbore tubing delivery accuracy may be affected when used at extreme
operating temperatures (<10°C, >35°C), relative humidity (<20%) and delivery rates (<4.9 mL/hr, >350 mL/hr). Refer to individual set labels for specific instructions and operating conditions.
Delivery accuracy may be impacted when there is air present in the cassette.
Follow the instructions in the individual set labels to remove any visible air from the cassette.
Ensure slide clamp or lower CAIR® clamp is open prior to inserting cassette.
When using pre-filled vials, ensure vial and injector are securely locked in
lockbox before beginning delivery.
10 Hospira GemStar Manual 430-10984-001 (Rev. 03/06)
Page 18
1. Introduction
A cracked vial might not show evidence of leakage until delivery pressure is
applied.
When using nonvented syringes, the system may underdeliver. System delivery
accuracy could be reduced by as much as 5%.
When using PCA vials (except in a lockbox with a spring assist mechanism), the
system may underdeliver. System delivery accuracy could be reduced by as much as 5%.
Variations in filling head (such as container height) may affect delivery accuracy
by no more than 5%.
When using the GemStar® Spring Assist Mechanism (SAM) Lockbox, USE
ONLY GemStar® Pump sets designated for use with the GemStar® SAM Lockbox. Failure to use designated sets could result in unrestricted flow.
NEVER use drugs that are incompatible with silicone rubber or PVC plastic
(unless these items are specially treated and labeled to be compatible with such drugs).
To reduce the loss of potency for drugs known to be absorbed by PVC plastic and
silicone, begin infusion as soon as practical after priming the pump set. Use of high flow rates during infusion will minimize drug absorption.
DO NOT use medications that are unstable under infusion conditions.
ALWAYS use connections with Luer Lock™ fittings.
To prevent contamination, use aseptic technique with all fluid-path connections.
Remove the protective coverings as assembly progresses.
ALWAYS place the flow stop in the closed position when the set is not in the
pump.
Before installing the cassette in the pump, ensure all slide or lower CAIR®
(roller) clamps are open. Failure to open the slide or lower CAIR® (roller) clamps BEFORE installing the cassette may result in restricted flow.
DO NOT back-prime the pump set using a syringe. Back priming may result in
restricted flow.
Ensure the cassette is properly installed. When properly installed, all four cassette
latches are visibly holding the cassette securely in the cassette pocket.
Ensure the tubing is behind the tubing guide (at the distal end).
Arrange tubing, cords, and cables to minimize the risk of patient strangulation or
entanglement.
NEVER use vented fluid containers (e.g., glass or rigid plastic) unless suspended
Hospira GemStar Manual 430-10984-001 (Rev. 03/06) 11
Page 19
1. Introduction
from a pole.
Stop infusion if signs or symptoms of infiltration occur.
GemStar® Pump Sets contain a pressure-activated anti-siphon valve to reduce the
potential for gravity flow at a typical 36-inch head height.
Failure to use the pressure-activated anti-siphon valve may result in unrestricted
flow if the cassette is not properly installed in the pump.
Always close the slide clamps before removing the cassette from the pump.
Regarding Air-In-Line and Infusion
To reduce the risk of infusing air, use an air-eliminating filter when the air
sensitivity is set to OFF.
ALWAYS prime the pump set to remove all air from the cassette, tubing, and
injection site prior to connecting to the patient.
ALWAYS disconnect the pump set from the patient prior to priming or purging.
Regarding Pump Operation
If the pump does not perform as stated in this manual, remove the pump from
service IMMEDIATELY.
If the pump does not display the self-test or if beeps do not sound at power-on,
check all power connections and install two fresh, disposable AA batteries. If the problem continues, contact Hospira Technical Support Operations before using the pump.
Ensure a cassette is installed in the pump before connecting to a printer or
computer. If the pump is connected to a patient: BEFORE connecting the pump to a computer or printer, press [ cassette. Leave the pump set connected to the patient. Install another cassette in the pump before connecting to a printer or computer.
Disconnect the pump from the patient BEFORE performing the Operation Test.
ALWAYS connect to a grounded AC mains power source when using AC mains
power. If the quality of the grounding source is in doubt, operate the pump only with batteries.
USE ONLY the AC mains adaptors specifically labeled for use with the
GemStar® Pump.
USE ONLY the AC mains adaptor specifically labeled for use with the GemStar®
Pump to charge the battery pack.
STOP], close the slide clamp, and eject the
12 Hospira GemStar Manual 430-10984-001 (Rev. 03/06)
Page 20
1. Introduction
The Hospira GemStar® AC Mains Adaptor is for use with the GemStar® Pump or
GemStar® Rechargeable Battery Pack only. DO NOT mains adaptor with other products.
During charging, the battery pack is warm. If the battery pack becomes hot to the
touch, IMMEDIATELY unplug the AC mains adaptor and contact Hospira Technical Support Operations.
Use only the Li-Ion Docking Station and Li-Ion Battery Pack with the GemStar®
Pump.
With the Li-Ion Battery Pack, the LED indicator flashes yellow/green when an
error condition has occurred or when the temperature is too high to charge the battery pack. In temperatures above 95°F (35°C), cycle AC power to confirm if battery is charged or if error condition exists. If LED flashes green/yellow, do not rely on external battery. Move device to cooler temperature and recharge prior to use. Refer to the battery pack insert for important information and precautions regarding activating and charging the Li-Ion Battery Pack.
When removing the cord (from the docking station, bottom of the battery pack, or
bottom of the pump), firmly grasp the connector and pull straight out. DO NOT twist or bend the cord or connector.
To ensure proper pump operation, ALWAYS replace BOTH disposable batteries
with fresh, disposable AA batteries when a change is required.
ALWAYS fully close the battery door when using disposable batteries.
DO NOT use rechargeable batteries in the battery compartment.
Ensure two fresh, disposable AA batteries are installed for backup power when
using other external power sources (e.g., AC Mains adaptor, docking station, battery pack). The pump will continue to operate on backup power if the external power source fails. In the event that the internal batteries are removed while the pump is operating, the power loss event may not be logged in the history, the programmed therapy may not be retained, and the alarms (audible and visual) may not occur.
ALWAYS avoid sources of high-intensity electromagnetic radiation (e.g., radio
transmitters, MRI scanners, microwave ovens, X-ray machines, and CAT scanners).
Use of radio-frequency emitting devices, such as cellular telephones, 2-way
radios, and Electrical Surgical Devices (ESU), in close proximity to this device may affect its operation.
Possible explosion hazard exists if the pump is used in the presence of flammable
anesthetics. NEVER use the pump in the presence of flammable or explosive
use the GemStar® AC
Hospira GemStar Manual 430-10984-001 (Rev. 03/06) 13
Page 21
1. Introduction
vapors.
Non-hazardous, low-level electrical potentials are commonly observed when
fluids are administered using infusion devices. These potentials are well within accepted safety standards, but may create artifacts on voltage-sensing equipment, such as ECG, EMG, and EEG machines. These artifacts vary at a rate that is associated with the infusion rate. If the monitoring machine is not operating correctly or has loose or defective connections to its sensing electrodes, these artifacts may be accentuated so as to simulate actual physiological signals. To determine if the abnormality in the monitoring equipment is caused by the infusion device instead of some other source in the environment, set the infusion device so that it is temporarily not delivering fluid (if clinically acceptable). Disappearance of the abnormality indicates that it was probably caused by electronic noise generated by the infusion device. Proper setup and maintenance of the monitoring equipment should eliminate the artifact. Refer to the appropriate monitoring system documentation for setup and maintenance instructions.
The GemStar® Pump has been tested and found to comply with EMC/EMI limits
in accordance with:
IEC/EN 60601-1-2 (2001) when using two AA disposable batteries or when
connected to the GemStar® AC Adaptor
IEC/EN 60601-1-2 (2001) (ESD exemption) when using Li-Ion Docking
Station or Li-Ion Battery Pack
When connected to other power sources or optional system components in commercial distribution before November 30, 2004, the GemStar® system complies with EMC/EMI limits in accordance with IEC/EN 60601-1-2 (1993). These limits are designed to provide reasonable protection against harmful interference in a typical medical installation. The equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to other devices in the vicinity. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference with other devices, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures:
Reorient or relocate the receiving device Increase the separation between the equipment Connect the equipment to an outlet on a circuit different from that to which
the other device is connected
Consult the manufacturer or field service technician for help
14 Hospira GemStar Manual 430-10984-001 (Rev. 03/06)
Page 22
1. Introduction
The use of portable and mobile RF equipment may have an impact on this and
other pieces of medical equipment
Prior to using the GemStar® Pump and optional system components, refer to
Appendix A for additional information and precautionary procedures regarding electrical potentials (electromagnetic compatibility and electrostatic discharge).
Regarding Handling and Maintenance
Use proper care during unpacking, installation, and operation of the pump. If the
pump is inadvertently mishandled, check connections and programmed data to ensure there is no damage. Refer to Chapter 8 Operation Test for further information.
Accessory equipment connected to the analog and digital interfaces must be
certified according to the respective IEC standards, (e.g., IEC 60950 for data processing equipment, and ICE 60601-1 for medical equipment). Furthermore, all configurations shall comply with the system standard IEC 60601-1-1. Any person who connects additional equipment to the signal input or output part configures a medical system, and is therefore responsible for ensuring that the system complies with the requirements of IEC 60601-1-1. If in doubt, contact Hospira Technical Support Operations.
ALWAYS avoid dropping or hitting the pump. If the pump is dropped or hit,
ALWAYS check connections and programmed data.
NEVER use sharp objects (e.g., fingernails, pens, pencils, or other probes) to
program or clean the pump.
Use care not to damage the silicone seals around the sensor bodies in the cassette
pocket.
To avoid mechanical or electronic damage, NEVER submerge pump in water or
other fluids and avoid fluid spills. If the pump becomes wet, dry it immediately with a dry, lint-free cloth. Check connections and programmed data.
DO NOT touch connector ports. Connector ports may be susceptible to
Electrostatic Discharge (ESD) damage.
Some cleaning and sanitizing compounds may slowly degrade components made
from some plastic materials. DO NOT use compounds containing combinations of isopropyl alcohol and dimethyl benzyl ammonium chloride (i.e., alcohol wipes).
Hospira GemStar Manual 430-10984-001 (Rev. 03/06) 15
Page 23
1. Introduction
DO NOT sterilize by heat, steam, ethylene oxide (ETO), or radiation. Apply
disinfectants to the outside surface of the pump only. DO NOT use abrasive cleaners or materials on the pump. Using abrasive cleaners or cleaning solutions not recommended by Hospira may result in product damage.
Please comply with local disposal and recycling regulations as appropriate for
disposable batteries, rechargeable battery packs, medical electronic components, and pump sets.
16 Hospira GemStar Manual 430-10984-001 (Rev. 03/06)
Page 24
Chapter 2
Basic Operation
The minimum elements required for using the GemStar® Pump are:
GemStar® Pump
Appropriate power source
Appropriate GemStar® Pump Set
Appropriate fluid container
Patient-access device
Overview
To start a therapy:
1 Determine the program settings. Refer to the appropriate therapy chapter for a
worksheet that may be copied and used to record program settings.
2 Select a power source. Refer to Selecting a Power Source on page 18.
3 Prime the pump set. Refer to GemStar® Pump Set on page 30.
4 Load the cassette into the pump. Refer to Loading the Cassette on page 33.
5 Power-on the pump and go to the Delivery Mode Menu. Refer to Powering On
the Pump on page 25.
6 Select the appropriate optional system components, such as a bolus cord.
7 Select the delivery mode and enter a program. Refer to Chapter 3 Pain
Management, page 45 for detailed instructions.
8 Press [
Hospira GemStar Manual 430-10984-001 (Rev. 03/06) 17
START] to begin delivery.
Page 25
2. Basic Operation
Selecting a Power Source
The GemStar® Pump may be powered by one or more of the following:
Power Source Description Ambulatory Hospital Home
GemStar® AC Mains Adaptor (USA)
GemStar® AC Mains Adaptor (Universal)
GemStar® AC Mains Adaptor (Universal)
GemStar® Docking Station
GemStar® Battery Pack
Disposable Batteries Two disposable AA alkaline
Wall-mount adaptor plugs into a 110 VAC mains supply.
Wall-mount adaptor plugs into a 100-240 VAC mains supply.
Tabletop adaptor plugs into a 100–240 VAC mains supply.
Plugs into a 100–240 VAC Volt mains supply while mounted on an I.V. pole. The docking station’s rechargeable internal batteries power the GemStar® Pump during periods of transport.
The battery pack’s rechargeable internal batteries power the GemStar® Pump during periods of transport or when use of AC power is not desirable.
batteries (available from local sources)
••
••
••
••
••
••
CAUTION: To ensure proper pump operation, ALWAYS replace BOTH
batteries with fresh, disposable AA batteries when a change is required.
DO NOT use rechargeable batteries in the battery compartment.
Ensure two fresh, disposable AA batteries are installed for backup power when using other power sources (e.g., AC Mains adaptor, docking station, battery pack). The pump will continue to operate on backup power if the external power source fails. Refer to the caution regarding Power Loss on page 13.
18 Hospira GemStar Manual 430-10984-001 (Rev. 03/06)
Page 26
Installing Disposable Batteries
1 Flip open the tab of the battery door on
the bottom of the pump.
Note: Turn off the pump before removing or installing batteries.
2 With the pump keypad facing up, hold
onto the tab and turn the battery door to the left (counterclockwise) until it is released from the battery compartment.
3 Pull the door free and remove any
existing batteries.
4 Install two fresh, disposable AA
batteries into the battery compartment. Ensure the batteries are inserted as shown.
5 Return the battery door to the battery
compartment.
6 Turn the battery door to the right
(clockwise). Ensure the battery door is secured under the slot.
2. Basic Operation
7 Flip closed the tab of the battery door to
lock in place.
Hospira GemStar Manual 430-10984-001 (Rev. 03/06) 19
Page 27
2. Basic Operation
Connecting the GemStar® AC Mains Adaptor
1 Connect the cord of the AC mains adaptor to the connector port labeled “3 VDC”
on the bottom of the pump.
2 Plug the AC mains adaptor into a standard wall outlet.
CAUTION: ALWAYS connect to a grounded AC mains power source when
using the AC mains adaptor. USE ONLY AC mains adaptors specifically labeled for use with the GemStar® Pump.
When removing the cord from the bottom of the pump, firmly grasp the connector and pull straight out. DO NOT twist or bend the cord or connector.
Ensure two fresh, disposable AA batteries are installed for backup power when using other power sources (e.g., AC Mains adaptor, docking station, battery pack). The pump will continue to operate on backup power if the external power source fails. Refer to the caution regarding Power Loss on page 13.
G
Note: The green LED on the pump remains lit when connected to AC mains power.
20 Hospira GemStar Manual 430-10984-001 (Rev. 03/06)
Page 28
2. Basic Operation
Using the GemStar® Rechargeable Battery Pack
Use the GemStar® AC Mains Adaptor (List 13026-01 [USA], 13026-03 [Universal] or 13072-01 [Universal]) to recharge the GemStar® Battery Pack.
WARNING: DO NOT touch connector ports. Refer to caution on
page 15.
CAUTION: ALWAYS connect to a grounded AC mains outlet when using AC
mains adaptor.
Use only AC mains adaptors specifically labeled for use with the GemStar® Pump to charge the battery pack.
During charging, the battery pack is warm. If the battery pack becomes hot to the touch, IMMEDIATELY unplug AC mains power and contact Hospira Technical Support Operations.
Ensure two fresh, disposable AA batteries are installed for backup power when using other power sources (e.g., AC Mains adaptor, docking station, battery pack). The pump will continue to operate on backup power if the external power source fails. Refer to the caution regarding Power Loss on page 13.
When removing the cord from the bottom of the battery pack, firmly grasp the connector and pull straight out. DO NOT twist or bend the cord or connector.
Regarding the GemStar® Li-Ion Battery Pack:
The LED indicator on the GemStar® Li-Ion Battery Pack flashes green/yellow when an error condition has occurred or when the temperature is too high to charge the battery pack. In temperatures above 95°F (35°C), cycle AC power to confirm if battery is charged or if error condition exists. If LED flashes green/yellow, do not rely on external battery. Move device to cooler temperature and recharge prior to use.
Refer to the GemStar® Li-Ion Battery Pack insert for important information and precautions regarding activating and charging the Li-Ion Battery Pack.
G
Note: A charged battery pack will last approximately 24 hours when the pump is infusing at rates less than 125 mL/hr.
Hospira GemStar Manual 430-10984-001 (Rev. 03/06) 21
Page 29
2. Basic Operation
To connect the battery pack:
1 Line up the connectors on the top of the battery
pack with the holes on the bottom of the pump.
2 Snap the battery pack into place.
To release the battery pack:
1 Press the release button on the battery pack.
2 Pull the battery pack straight out from the pump.
Note: When the pump is connected to the battery pack, the pump’s green LED flashes and the
G
battery pack light is off.
1
When the pump is connected to the battery pack at power-on, the pump displays
22 Hospira GemStar Manual 430-10984-001 (Rev. 03/06)
USING EXT BATT.
Page 30
G
2. Basic Operation
To charge the battery pack:
1 Plug the pin connector of the AC mains adaptor into the port on the bottom of the
battery pack.
2 Plug the AC mains adaptor into a standard wall outlet.
3 Leave the battery pack connected to the AC mains adaptor for at least 12 hours to
charge the battery pack.
Note: The Li-Ion battery pack (List 13073-03) light is yellow while the battery pack is charging and green when the battery pack is charged or not charging.
Li-Ion batteries cannot be charged when the room temperature is greater than 95°F (35°C) or less than 41°F (5°C)
CAUTION: When removing the cord from the battery pack, firmly grasp the
connector and pull straight out. DO NOT twist or bend the cord or connector.
Hospira GemStar Manual 430-10984-001 (Rev. 03/06) 23
Page 31
2. Basic Operation
Using the GemStar® Docking Station
WARNING: DO NOT touch connector ports. Refer to caution on page
page 15.
CAUTION: Ensure two fresh, disposable AA batteries are installed for backup
power when using other power sources (e.g., AC Mains adaptor, docking station, battery pack). The pump will continue to operate on backup power if the external power source fails. Refer to the caution regarding Power Loss on page 13.
To Connect to the Pump
1 Mount the docking station to the I.V. pole and tighten
the knob.
2 Line up the connectors on the bottom of the pump with
the connectors on the docking station.
3 Slide the pump into the docking station until it snaps
into place.
4 Connect the docking station to AC mains power.
To remove the pump:
1 Press the release button on the top of the docking
station.
2 Slide the pump out.
To charge docking station (List 13075 all codes):
1 Connect the docking station to the AC mains power.
2 Leave the Li-Ion Docking Station connected to AC mains power for at least 6
hours when not in use or to recharge the docking station.
.
Note: The Li-Ion docking station (List 13075-03) light is yellow while the docking
G
station is charging. When the docking station is charged or not charging, the light is
green. The Li-Ion docking station cannot be charged when the room temperature is greater than 104°F (40°C) or less than 41°F (5°C).
24 Hospira GemStar Manual 430-10984-001 (Rev. 03/06)
Page 32
2. Basic Operation
Powering On the Pump
CAUTION: If the pump does not display the self-test or if beeps do not sound at
power-on, check all power connections and install two fresh, disposable AA batteries. If the problem continues, contact Hospira Technical Support Operations before using the pump.
G
G
Note: If a cassette is not properly loaded, the pump will sound an alarm after the power-on test is complete.
1 Press [
ON/OFF] to power-on the pump. The pump
completes a self-test and displays the time and date. No response is required.
GEMSTAR 7
UNIT SELF-TEST
TIME IS 10:00 AM
FRI, JUN 11, 04
Note: The self-test takes approximately 30 seconds to complete. If the pump is locked, lock symbols appear at the beginning and end of the second line:
UNIT SELF-TEST .
2 The pump checks for available power sources. Respond to the display message as
shown in the following table:
Display Message Power Source(s) User Response
No message displays AC power and
disposable batteries.
LOW BATTERIES
ENTER TO CONFIRM
AC power and weak (low voltage) disposable batteries; or weak disposable batteries only.
No response required, the pump goes to the next step.
Press to confirm.
Power-off the pump and install two fresh, disposable AA batteries as a backup power source.
NO INTERNAL BATT
NO POWER LOSS
ALARM AVAILABLE
ENTER TO CONFIRM
Hospira GemStar Manual 430-10984-001 (Rev. 03/06) 25
No disposable batteries or dead disposable batteries; and external power.
Press to confirm.
Power-off the pump and install two fresh, disposable AA batteries as a backup power source.
Page 33
2. Basic Operation
Display Message Power Source(s) User Response
G
USING BATTERIES
ENTER TO CONFIRM
USING EXT BATT
ENTER TO CONFIRM
Disposable batteries only.
External batteries (docking station battery or battery pack) and disposable batteries.
Press to confirm.
If an external power source is in use, ensure the connections are secure.
Press to confirm.
If an AC mains adaptor is in use, ensure the connections are secure.
Note: To power-off the pump, press and hold until the display clears. The pump cannot be powered off during a CHECK CASSETTE alarm (except during
PROGRAMMING mode). Refer to Chapter 6 Troubleshooting for information on
clearing a CHECK CASSETTE alarm.
CAUTION: Ensure two fresh, disposable AA batteries are installed for backup
power when using other power sources (e.g., AC Mains adaptor, docking station, battery pack). The pump will continue to operate on backup power if the external power source fails. Refer to the caution regarding Power Loss on page 13.
For more information on: Refer to the following pages:
Setting the clock Set Clock in Chapter 4 Options Menu, on page 66.
Locking the keypad Keypad Lock in Chapter 4 Options Menu, on page 62.
26 Hospira GemStar Manual 430-10984-001 (Rev. 03/06)
Page 34
2. Basic Operation
Accessing the Delivery Mode Menu
1 After power-on, the pump checks its memory for a stored program. Respond to
the display message as shown in the following table:
Display Message Condition User Response
The pump displays the Delivery Mode Menu.
1 RESUME PROGRAM 2 CLR PROG&SHIFT
3 CLR PROG,SHIFT
AND HISTORY
Pump memory is clear.
A program is currently stored in pump memory.
! To enter a new program: Select a delivery
mode from the Delivery Mode Menu and enter a new program. Refer to Chapter 3 Pain Management for instructions.
! To continue the current program, press [1].
! To reset the current program:
1 PRESS [1]. 2 PRESS [
CHANGE MENU AND SELECT NEW CONTAINER.
! To enter a new program and clear shift totals:
Press [2]. The pump clears the program and shift totals from memory and enters PROGRAMMING mode.
CHANGE] TO DISPLAY THE
! To enter a new program, clear shift totals, and
clear history: Press [3]. The pump clears the program, shift totals, and history from memory and enters PROGRAMMING mode.
Tip: To enter a new program when the pump is already powered on:
I
1 Press [STOP] to place the pump in STOP mode.
2 Press [CHANGE] to display the Change Menu.
3 Select NEW PROGRAM. The pump clears the current program and
displays the Delivery Mode Menu.
Hospira GemStar Manual 430-10984-001 (Rev. 03/06) 27
Page 35
2. Basic Operation
Programming the Pump
Programming Overview
The pump’s menu system provides step-by-step guidance through all of the programming steps. Refer to Chapter 3 Pain Management for detailed programming instructions.
A programming worksheet is available at the end of Chapter 3 that lists the minimum and maximum value (program range) allowed for a Pain Management therapy. If a number outside of the program range is entered, the pump displays the minimum or maximum value allowed.
1 Select the desired delivery mode from the Delivery Mode Menu.
2 Respond to display messages and enter requested values.
Note: Program entries can be changed any time before the Program Review is
G
complete. Press [BACK-UP] to return to previous program displays and make changes as desired.
3 Review the program, if required. Press to begin review. Press [
when done.
4
SAVING PROGRAM displays. The pump stores the program in memory and then
STOP mode.
enters
Tip: If you enter an incorrect number during programming, press [CHANGE]
I
BEFORE pressing [ correct number.
28 Hospira GemStar Manual 430-10984-001 (Rev. 03/06)
YES/ENTER] to reset the number to zero. Then enter the
YES/ENTER]
Page 36
2. Basic Operation
Programming Tips
The following table explains how to use the keypad during programming:
Key Function
[0] - [9] Press the number keys to select menu items and to enter numeric values.
Press to accept entries or to continue when done viewing a display.
Press during programming and alarm conditions to display information
about the current condition.
Press to toggle between units of measure when entering concentrations.
Scrolls through program review and help messages.
Press to insert a decimal point in a numeric value.
CHANGE
Hospira GemStar Manual 430-10984-001 (Rev. 03/06) 29
When entering a numeric value during programming, press to reset the
value to zero.
Press to return to a previous display, to exit the Options Menu, or to exit
the
HELP mode.
CHANGE
Page 37
2. Basic Operation
Starting a New Container
G
Note: The NEW CONTAINER function is not available when the pump is in FULL Lock. Refer to Chapter 4 Options Menu, page 55 for more information on locking the keypad.
To repeat the current program with a new container:
1 From the
2 Select
CONTAINER
Note: In Pain Management therapy, the VTBI can be titrated without adding volume to initial container amount.
3 Review the program. Press to begin review.
4 Press [
STOP mode, press [CHANGE] to display the Change Menu.
NEW CONTAINER from the Change Menu. The pump displays NEW
for several seconds and clears the amount infused from memory.
START] to begin delivery.
GemStar® Pump Set
The GemStar® Pump Set is a sterile, single-use, disposable pump set. It is designed for one-time use only and should be properly discarded after use.
Fluid path and areas beneath undisturbed protective set covers are sterile and nonpyrogenic in the intact unit package.
Contact a Hospira representative for more information on selecting the appropriate pump set configuration. Additional disposable components, such as air-eliminating filters and extension sets, may be added to the line as required. To use a GemStar® Pump Set, follow the instructions included with the set.
STERILE
2
Note: GemStar® Pump Sets contain a pressure-activated anti-siphon valve to reduce
G
the potential for gravity flow at a typical 36-inch head height. Ensure slide clamp or lower CAIR® clamp is open prior to inserting cassette.
The system may underdeliver, and the system delivery accuracy may be reduced by as much as 5% when using the following:
nonvented syringes
PCA vials (except in a lockbox with a spring assist mechanism)
30 Hospira GemStar Manual 430-10984-001 (Rev. 03/06)
Page 38
A typical GemStar® Pump Set is shown below:
2. Basic Operation
1 Piercing Pin 6 Tubing
2 Proximal End 7 Slide Clamp
3 Cassette 8 Secure Lock
4 Flow Stop (open / close ) 9 Pressure-activated Anti-Siphon Valve
5 Distal End
Hospira GemStar Manual 430-10984-001 (Rev. 03/06) 31
Page 39
2. Basic Operation
WARNING: ALWAYS prime the pump set to remove air from the
cassette, tubing, and injection sites prior to connecting to the patient. ALWAYS disconnect the pump set from the patient prior to priming or purging. Arrange tubing, cords, and cables to minimize the risk of patient strangulation or entanglement.
Failure to use the pressure-activated anti-siphon valve may result in unrestricted flow if the cassette is not properly installed in the pump.
CAUTION: USE ONLY GemStar® Pump Sets with the GemStar® Pump. Use
of unauthorized sets may result in injury to the patient or damage to the pump.
To prevent contamination, use aseptic technique with all fluid-path connections. Remove protective coverings as assembly progresses.
Opening the Pump Set
1 Open the GemStar® Pump Set package and
remove the contents.
2 Remove the protective covers from the fluid
container administration port and the pump set.
32 Hospira GemStar Manual 430-10984-001 (Rev. 03/06)
Page 40
Loading the Cassette
1 Identify the four black cassette latches inside the cassette pocket.
2. Basic Operation
2 Ensure slide clamp or lower
CAIR® clamp is open prior to inserting cassette.
3 Line up the cassette as shown
(with the flow stop button facing the pump).
4 Insert the cassette into the cassette
pocket by pushing firmly along the entire cassette until it is firmly seated.
Hospira GemStar Manual 430-10984-001 (Rev. 03/06) 33
Page 41
2. Basic Operation
5 Verify that all four cassette latches
are securely holding the cassette.
CAUTION: Ensure the cassette is properly installed. When properly installed, all four cassette latches are visibly holding the cassette securely in the cassette pocket.
6 Place the tubing in the
tubing channel.
Note: The pump cannot be powered
G
off during a CHECK CASSETTE alarm (except during
PROGRAMMING mode). Refer to
Chapter 6 Troubleshooting for information on clearing a CHECK CASSETTE alarm.
Releasing the Cassette
1 Close the slide clamp on the distal
line.
2 Push down on the cassette release
(large black button) on the top of the pump.
Note: Ensure the tubing is behind
G
34 Hospira GemStar Manual 430-10984-001 (Rev. 03/06)
the tubing guide (at the distal end).
3 Remove the cassette.
Note: When the cassette is
G
released from the pump, the cassette flow stop automatically closes, preventing free flow.
Page 42
2. Basic Operation
Priming a GemStar® Pump Set
Priming Pump Sets WITHOUT a Drip Chamber
WARNING: Failure to use the pressure-activated anti-siphon valve
may result in unrestricted flow if the cassette is not properly installed in the pump.
1 Push down on the flow stop to
close the cassette.
2 Hold the container with the
administration port toward the ceiling and insert the piercing pin.
3 Push down on the flow stop to
open (priming position) the cassette.
4 Invert the cassette approximately
45°.
5 While holding the rotated cassette,
squeeze or roll the container to remove air from the container and cassette.
6 Return the cassette to the upright
position. Continue priming until fluid fills the tubing.
Hospira GemStar Manual 430-10984-001 (Rev. 03/06) 35
Page 43
2. Basic Operation
7 Push down on the flow stop to
close the cassette.
8 Ensure slide clamp is open prior to
inserting cassette.
9 Load the cassette.
G
Note: Ensure the tubing is behind the tubing guide (at the distal end).
CAUTION: Ensure all four
cassette latches are clearly visible after the cassette is installed. Refer to Loading the Cassette on page 33.
36 Hospira GemStar Manual 430-10984-001 (Rev. 03/06)
CAUTION: When priming is complete, ensure no fluid flows at the distal end of the pump set. If flow is observed, DO NOT use the pump set.
10 Connect the pump set to the
patient-access device.
Priming using [ allowed before [ or after one of the following events occur when using Pain Management therapy:
Complete Review of New
Program
New Container Air-in-Line Alarm
PURGE] is only
START] is pressed
Page 44
2. Basic Operation
Priming Pump Sets WITHOUT a Drip Chamber Using the PURGE Key
WARNING: Failure to use the pressure-activated anti-siphon valve
may result in unrestricted flow if the cassette is not properly installed in the pump.
G
Note: The PURGE function is available only if there is a program in the pump.
1 Push down on the flow stop to
close the cassette.
2 Hold the container with the
administration port toward the ceiling and insert the piercing pin.
3 Ensure slide clamp is open prior to
inserting cassette.
4 Load the cassette.
CAUTION: Ensure all four cassette latches are clearly visible after the cassette is installed. Refer to Loading the Cassette on page 33.
Hospira GemStar Manual 430-10984-001 (Rev. 03/06) 37
Page 45
2. Basic Operation
Note: Ensure the tubing is behind
G
the tubing guide (at the distal end).
5 Program desired therapy.
6 Press [
7 The pump displays
8 While holding the container with
PURGE].
PURGE
SET?
Press [YES/ENTER].
the administration port toward the ceiling, press and hold [
HELP
PRIME THE
PURGE].
CAUTION: When priming is complete, ensure no fluid flows at the distal end of the pump set. If flow is observed, DO NOT use the pump set.
9 Connect the pump set to the
patient-access device.
Note: Purge rate is 250 mL/hr.
G
38 Hospira GemStar Manual 430-10984-001 (Rev. 03/06)
The pump purges for up to 2 minutes each time you press and hold [
Continue using the [ key until fluid fills the tubing. Ensure all air is removed from the container, cassette, and tubing.
PURGE].
PURGE]
Page 46
2. Basic Operation
Priming Pump Sets Using Pre-filled Vials
WARNING: Failure to use the pressure-activated anti-siphon valve
may result in unrestricted flow if the cassette is not properly installed in the pump.
A cracked vial might not show evidence of leakage until delivery pressure is applied.
When using the GemStar® Spring Assist Mechanism (SAM) Lockbox, USE ONLY GemStar® Pump sets designated for use with the GemStar® SAM Lockbox. Failure to use designated sets could result in unrestricted flow.
CAUTION: When using PCA vials (except in a lockbox with a spring assist
mechanism), the system may underdeliver. System delivery accuracy could be reduced by as much as 5%.
When using pre-filled vials, ensure vial and injector are securely locked in lockbox before beginning delivery.
USE ONLY purge priming with the GemStar® Lockbox with Spring Assist Mechanism (SAM).
G
Hospira GemStar Manual 430-10984-001 (Rev. 03/06) 39
Note: When using the GemStar® Lockbox with SAM, refer to the SAM pump set labeling and SAM lockbox insert for instructions.
1 If using a pump set with the
Y-extension, use the slide clamp to close off the KVO (Keep Vein Open) side of the tubing.
2 Open the upper and lower slide
clamps.
Page 47
2. Basic Operation
3 Push down on the flow stop to
close the cassette.
4 Connect the pump set to a Hospira
pre-filled vial and injector.
5 Push down on the flow stop to
open (priming position) the cassette.
6 Hold the vial and set upright
(vertically) on a flat surface with the pump set extending from the top.
7 Slowly push down on the injector
until fluid flows at the distal end of the tubing, which indicates that all
air is cleared from the vial, injector, cassette, and tubing.
8 Push down on the flow stop to
close the cassette.
9 If the set includes a KVO
Y-extension, close the slide clamp below the pre-filled vial and open the slide clamp on the KVO side. Prime the KVO Y-extension.
10 Ensure slide clamp is open on
KVO side prior to inserting cassette.
40 Hospira GemStar Manual 430-10984-001 (Rev. 03/06)
Page 48
2. Basic Operation
11 Load the cassette.
CAUTION: Ensure all four cassette latches are clearly visible after the cassette is installed. Refer to Loading the Cassette on page 33.
CAUTION: When priming is complete, ensure no fluid flows at the distal end of the pump set. If flow is observed, DO NOT use the pump set.
12 Connect the pump set to the
patient-access device.
G
Hospira GemStar Manual 430-10984-001 (Rev. 03/06) 41
Note: Ensure the tubing is behind the tubing guide (at the distal end).
Page 49
2. Basic Operation
Priming Pump Sets with Extensions
WARNING: Failure to use the pressure-activated anti-siphon valve
may result in unrestricted flow if the cassette is not properly installed in the pump.
1 Push down on the flow stop to
close the cassette.
2 Hold the container with the
administrative port toward the ceiling and insert the piercing pin.
3 Push down on the flow stop to
open (priming position) the cassette.
4 Invert the cassette approximately
45°.
5 While holding the rotated cassette,
squeeze or roll the container to remove air from the container and cassette.
6 Return the cassette to the upright
position. Continue priming until fluid fills the tubing.
42 Hospira GemStar Manual 430-10984-001 (Rev. 03/06)
Page 50
7 Push down on the flow stop to
close the cassette.
8 Connect the extension set to the
pump set.
9 Ensure the slide clamp is open
before inserting cassette.
G
2. Basic Operation
Note: Ensure the tubing is behind the tubing guide (at the distal end).
10 Load the cassette.
CAUTION: Ensure all four cassette latches are clearly visible after the cassette is installed. Refer to Loading the Cassette on page 33.
Hospira GemStar Manual 430-10984-001 (Rev. 03/06) 43
11 Program the desired therapy.
12 Press [
13 The pump displays
SET?
14 Press and hold [
Note: Purge rate is 250 mL/hr. The
G
pump purges for up to 2 minutes each time you press and hold
[
Continue using the [ until fluid fills the tubing. Ensure
PURGE].
PURGE
Press [YES/ENTER].
PURGE].
HELP
PRIME THE
PURGE].
PURGE] key
Page 51
Notes
all air is removed from the container, cassette, and tubing.
CAUTION: When priming is complete, ensure no fluid flows at the distal end of the pump set. If flow is observed, DO NOT use the pump set.
15 Connect the pump set to the patient-access device.
44 Hospira GemStar Manual 430-10984-001 (Rev. 03/06)
Page 52
Chapter 3
Pain Management
Therapy Features and Specifications
This therapy allows programming of a continuous only, bolus only, or continuous with bolus delivery.
In Continuous + Bolus protocols, you can enter zero for either the continuous rate
or the bolus dose when entering or changing a program. The continuous rate and bolus dose cannot be zero at the same time.
The default bolus delivery rate is 125 mL/hr.
G
G
Note: The delivery rate is delivered in increments of .01.
The maximum continuous delivery rate is 25 mL/hr.
A loading dose may be set during programming, which can be delivered after
programming and priming are complete. The default loading dose rate is 125 mL/hr. Rate, bolus dose, concentration can be programmed in increments of 0.01.
Priming using [PURGE] is only allowed before [START] is pressed or after one of
the following events occur:
Complete Review of New Program New Container Air-in-Line Alarm
VTBI can be titrated.
CAUTION: When using PCA vials (except in a lockbox with a spring assist
mechanism), the system may underdeliver. System delivery accuracy could be reduced by as much as 5%.
Note: New Container is allowed only when at least 1/2 of the container volume has
been infused or 1 hour has elapsed since starting the infusion.
The use of a lockbox is recommended with pain management protocols. Refer to Chapter 9 Optional System Components, page 109, for instructions on using a lockbox with the GemStar® Pump.
Hospira GemStar Manual 430-10984-001 (Rev. 03/06) 45
Page 53
3. Pain Management
Tip: The Programming Worksheet at the end of this chapter lists the minimum and
I
maximum values allowed for this therapy. Use copies of the programming worksheet to record program values before programming the pump.
Pain Management Programming Tips
The following table explains how to use the keypad during pain management programming:
Key Function Key Function
[0] - [9] Selects menu items and
enters numeric values.
Accepts entries or continues when done viewing a display.
Displays information during programming and alarm conditions.
CHANGE
Clears a numeric value (resets to zero) during programming.
DOWN arrow toggles between units of measure when entering concentrations. Scrolls through program review and help messages.
The up arrow inserts a decimal point in a numeric value.
Returns to a previous display, exits the Options Menu, and exits the
HELP mode.
46 Hospira GemStar Manual 430-10984-001 (Rev. 03/06)
Page 54
Pain Management Programming.
1 Select the delivery mode from the Delivery Mode menu.
3. Pain Management
G
G
G
Note: If Continuous + Bolus is selected, the continuous delivery rate or bolus dose can be set to zero through the CHANGE menu without reprogramming the pump.
2 Select the unit of measure and concentration as required.
3 Set the continuous delivery rate.
4 Program a loading dose as required.
5 Program a bolus dose with lockout time and dose limit as required.
Regarding a 1-Hour or 4-Hour Dose Limit: The hour dose limit is the maximum volume (continuous plus bolus or bolus only amount) that can be delivered over the selected period. When the programmed dose limit is reached, the continuous delivery stops and a new bolus is not allowed. As the oldest delivery amounts age out of the 1-hour or 4-hour dose limit record, the pump accepts bolus requests or resumes continuous delivery (if programmed).
Note: If a 1-hour or 4-hour dose limit is programmed, a bolus in progress will stop when the amount infused reaches the dose limit.
Loading doses are included in a 1-hour or 4-hour dose limit. A dose limit in effect could prevent the delivery of a loading dose.
6 If programmed in units of mcg or mg, the line displaying the units flashes. Enter
the container size in proper unit measure.
Note: To enter the total in mL when programming a concentration, press to move to the mL line. Enter the mL value when
mL flashes on the display.
7 Select the air sensitivity, if required.
8 Review the program. Press to begin review.
9 If using the PURGE key to prime the pump set, press [
WARNING: ALWAYS disconnect the pump set from the patient before
priming or purging.
Hospira GemStar Manual 430-10984-001 (Rev. 03/06) 47
PURGE] now.
Page 55
3. Pain Management
Bolus Delivery
Set bolus doses and delivery limits during programming. The default bolus dose delivery rate in a Pain Management therapy is 125 mL/hr.
A bolus cord is available for patient bolus requests. Refer to Connecting the Bolus Cord in Chapter 9 Optional System Components, on page 112, for more information.
Delivering a Bolus Dose
G
From the button on the end of the bolus cord to begin bolus dose delivery. The pump flashes
BOLUS DELIVERY on the display and the amount infused accrues as the delivery
progresses.
When the bolus dose is complete, the pump does the following:
Records the bolus request and delivery amount to the event and bolus history logs
Adds the delivered boluses to the total bolus amount in the program amounts
Resets the bolus lockout time (if applicable) and ignores other bolus requests until
Continues programmed infusion
Note: ALL bolus requests are recorded to the event and bolus history logs.
If the bolus delivery does not start, it could be locked out by one of the following conditions:
Bolus lockout period
Loading dose delivery in progress
Bolus dose delivery in progress
Bolus per-hour limit
1-hour or 4-hour dose limit
RUN mode, press either the bolus button [+] on the top of the pump or the
the lockout time has elapsed
48 Hospira GemStar Manual 430-10984-001 (Rev. 03/06)
Page 56
3. Pain Management
Bolus Dose Interruptions
Bolus deliveries may be interrupted by:
Pressing [STOP]
An alarm condition
To continue bolus delivery after an interruption:
1 Press [START]. The pump displays COMPLETE BOLUS NOW?
2 Respond as follows:
To do this: Press this key: Bolus lockout time is set to:
Deliver the remaining bolus amount
Clear the undelivered bolus amount
Rate Titration
To titrate the rate while the pump is infusing:
1 In the
For example: if the current rate is 1.5 mL/hr, press [ change the rate to 2.5 mL/hr.
Note: You have approximately five seconds after pressing a key to press another
G
numeric key, the decimal key, or [ seconds, the rate does not change and the delivery continues as previously programmed.
2
NEW RATE displays with the new value.
Press [ at the new rate.
Press [
The time of delivery completion
The time the bolus was interrupted
RUN mode, press the number keys for the new rate. Press [YES/ENTER].
2][.] [5] [YES/ENTER] to
YES/ENTER]. If no key is pressed within five
YES/ENTER] within 10 seconds to accept the change and begin delivery
NO] to continue delivery at the previously programmed rate.
Hospira GemStar Manual 430-10984-001 (Rev. 03/06) 49
Page 57
3. Pain Management
Loading Dose Delivery
You can set a loading dose during programming, if desired. A loading dose can be delivered after programming and priming are complete.
G
Note: Loading dose amounts are included in a 1-hour or 4-hour dose limit.
Delivering a Loading Dose
1 After priming and programming, press [
2 The pump displays
begin loading dose.
When the loading dose is complete, the pump does the following:
Automatically begins programmed infusion
Records the dose amount to the history log
Resets the bolus lockout time (if applicable) and ignores other bolus requests until
the lockout time has elapsed
Loading Dose Interruptions
Loading dose deliveries can be interrupted by:
Pressing [STOP]
An alarm condition
To continue a loading dose:
1 Resolve the interruption.
DELIVER THE LOADING DOSE NOW? Press [YES/ENTER] to
START].
2 Press [
50 Hospira GemStar Manual 430-10984-001 (Rev. 03/06)
START]. The pump displays COMPLETE THE LOADING DOSE?
Page 58
3 Respond as follows:
To do this: Press this key:
Deliver the remaining loading dose amount
Clear the remaining loading dose amount
Changing a Pain Management Program
Can be changed at any time Cannot be changed
3. Pain Management
G
• Delivery rate
• Bolus
• Bolus lockout
• Container size
• Air sensitivity
• Dose limit amount
• Delivery mode
• Unit of measure
• Concentration
• Dose limit type
Note: You cannot change a program when the pump is in FULL or CONTAINER Lock. With a
RATE CHANGE Lock, the continuous delivery rate and bolus amount
can be changed within the set ranges.
Although the delivery mode cannot be changed, if Continuous + Bolus is programmed, the continuous rate or bolus amount can be changed to zero. The minimum values set with a
RATE CHANGE Lock could restrict this function.
To change a program:
1 From the
2 As each message displays, make the desired changes or press [
STOP mode, press [CHANGE] and select CHANGE PROGRAM.
YES/ENTER] to
accept the current setting.
3 Review the program. Press to begin review.
4 Press [
START] to begin delivery.
Hospira GemStar Manual 430-10984-001 (Rev. 03/06) 51
Page 59
3. Pain Management
Pain Management Programming Worksheet
GIVE:
GemStar®
Parameters Selections and/or Values to be
Programmable Ranges
Entered
Therapy Pain Management N/A Delivery Mode
Unit of Measure*
Continuous Rate*
Loading Dose
Bolus Dose* 0 - 2500 mg
Bolus Lockout
Optional Dose Limit
Continuous Bolus Only Continuous + Bolus Concentration: mg/mL
Concentration: mcg/mL mL
No
Yes Amount:
minutes 1 - 999 min
4-hour limit Amount:
1-hour limit Amount:
N/A
0.01 - 100 mg/mL
1 - 1000 mcg/mL
0.1 - 2500 mg/hr 1 - 25000 mcg/hr
0.1 - 25.0 mL/hr
0.1 - 2500 mg 1 - 25000 mcg
0.1 - 25.0 mL
0 - 25000 mcg 0 - 25.0 mL
4X rate - 400 mL, 40000 mg, 400000 mcg*
Rate - 100 mL, 10000 mg, 100000 mcg*
# boluses/hourNumber: No dose limit
1 - 12
*For continuous plus bolus delivery.
Note: For Bolus Only, the 1-hour or 4-hour dose limit minimum is 1 bolus dose.
Container Size 0.1 - 99999 mg
1 - 9999000 mcg
0.1 - 9999 mL
52 Hospira GemStar Manual 430-10984-001 (Rev. 03/06)
Page 60
3. Pain Management
Parameters Selections and/or Values to be
Entered
Air Sensitivity** On Alarms at approx. 0.5 mL of air
2 mL Alarms at approx. 2 mL of air Off No alarm
* The delivery rate is programmable and is delivered in increments of .01. ** Air Sensitivity may not show in all setting configurations.
G
Note: Titrating VTBI does not change initial container volume.
Programmable Ranges
N/A
Hospira GemStar Manual 430-10984-001 (Rev. 03/06) 53
Page 61
Notes
54 Hospira GemStar Manual 430-10984-001 (Rev. 03/06)
Page 62
Chapter 4
Options Menu
Overview
1 Press [OPTIONS] from RUN or STOP mode 2 Select the number key for the desired function. 3 Follow the display messages for the selected
function. Refer to the appropriate description in this chapter for more information.
Note: Menu selections may vary depending on the
G
configuration of the pump.
When the pump exits mode it was in when [OPTIONS] was pressed.
OPTIONS, the pump returns to the
Tips for Using the Options Menu
The following keys are helpful when using the Options Menu:
Key Function Key Function
[0] - [7] Selects menu items and
enters numeric values.
Accepts entries or continues when done viewing a display.
1 REVIEW PROGRAM
2 HISTORIES
3 KEYPAD LOCK
4 PUMP SETTINGS
5 SET CLOCK
6 PRINT
7 SPEED PROTOCOL
DOWN arrow toggles between units of measure when entering concentrations. Scrolls through program review and help messages.
The up arrow inserts a decimal point in a numeric value.
Returns to a previous display, exits the Options Menu, and exits the
Tip: A menu item does not have to be on the display to be selected. For example, only
I
items 1 through 4 are shown when the Options Menu is first displayed. To select the Print Menu without scrolling the display, press [
Hospira GemStar Manual 430-10984-001 (Rev. 03/06) 55
HELP mode.
6].
Page 63
4. Options Menu
Select this item from the Options Menu:
REVIEW PROGRAM
HISTORIES
KEYPAD LOCK
PUMP SETTINGS
SET CLOCK
PRINT
SPEED PROTOCOLS
To do this function: Comments:
Display the current program settings.
Display the history for the latest 400 events, shift or volume totals, bolus history (if available), or current configuration settings. Clear the history. Display the Operation Test.
N/A
CAUTION: Ensure two fresh, disposable AA batteries are installed for backup power when using other power sources (e.g., AC Mains adaptor, docking station, battery pack). The pump will continue to operate on backup power if the external power source fails. Refer to the caution regarding Power Loss on page 13.
Lock and unlock the keypad to prevent unauthorized changes to the program.
Change the air sensor, occlusion sensor, and sound level settings.
Set the clock. The clock can be displayed in
Print the history event log, bolus history (if available), current program settings, and speed protocols.
Assign, review/recall, and delete the stored protocols.
The lock levels allow the clinician to restrict access to various levels of pump operation.
N/A
12-hour or 24-hour format. Files may be downloaded to a
printer or personal computer.
Allows up to 9 speed protocols to be saved in pump memory.
56 Hospira GemStar Manual 430-10984-001 (Rev. 03/06)
Page 64
4. Options Menu
Review Program
1 Press [OPTIONS] and select REVIEW PROGRAM. The
pump displays
PROGRAM REVIEW.
2 Press to begin review.
3 Press [
YES/ENTER] or [BACK-UP] when done to exit
OPTIONS
.
Histories
1 Press [OPTIONS] and select HISTORIES.
2 Select the number key for the desired function.
3 Complete the appropriate steps for the selected
function.
1 REVIEW PROGRAM
2 HISTORIES
3 KEYPAD LOCK
4 PUMP SETTINGS
5 SET CLOCK
6 PRINT
7 SPEED PROTOCOL
1 REVIEW PROGRAM
2 HISTORIES
3 KEYPAD LOCK
4 PUMP SETTINGS
5 SET CLOCK
6 PRINT
7 SPEED PROTOCOL
Hospira GemStar Manual 430-10984-001 (Rev. 03/06) 57
Page 65
4. Options Menu
Select this item from the Histories Menu:
HISTORY
SHIFT TOTALS
VOLUME TOTALS
CONFIGURATI ON
CLEAR HISTORY
BOLUS HISTORY
OPERATION TEST
To do this function: Display includes:
Display history event log of the latest 400 events.
• Current date and time
• Date and time the program was cleared, a new container was selected, or a speed protocol was recalled
• History event log of the latest 400 events
CAUTION: Ensure two fresh, disposable AA batteries are installed for backup power when using other power sources (e.g., AC Mains adaptor, docking station, battery pack). The pump will continue to operate on backup power if the external power source fails. Refer to the caution regarding Power Loss on page 13.
Display or clear shift totals.
Display volume totals. Amount infused
Display pump configuration.
Clear history event log and bolus history.
Display bolus history in hourly increments for the last 48 hours.
Perform user-assisted Operation Test.
• Time and date when shift was cleared
• Volume infused in the shift
• Any relevant program-specific information
•VTBI
• Any relevant program-specific information
• Dependent upon pump configuration
CLEARING HISTORY displays for 3 to 4
• seconds
• Current date and time
• Number of boluses delivered
• Number of bolus demands
• Amount delivered through bolus delivery
• Messages that guide the user through each section of the test
G
I
Note: Some menu items may not be available depending on the configuration of the pump.
Tip: Press and HOLD or to scroll through the history displays faster.
58 Hospira GemStar Manual 430-10984-001 (Rev. 03/06)
Page 66
4. Options Menu
History
To display the history event log:
1 Press [
2 Select
OPTIONS] and select HISTORIES.
HISTORY from the Histories Menu.
3 Press or to scroll the display.
4 Press [
BACK-UP] when done to exit OPTIONS.
Shift Totals
To display the shift totals:
1 Press [
2 Select
OPTIONS] and select
HISTORIES
.
SHIFT TOTALS from the Histories Menu.
3 The shift totals display. Press or to scroll
the display.
To clear the shift totals:
1 Follow the steps for displaying shift totals.
2 At the end of the shift totals, the pump displays
CLEAR TOTALS? YES OR NO:
Press [YES/ENTER] to clear the shift totals.
Press [
NO] to exit OPTIONS
without clearing the shift totals.
VERSION N.NNN
S/N XXXXXXXX
TIME IS 09:00 AM
FRI, JUN 11, 05
NEW CONTAINER
08:15AM JUN 11
HISTORY CLEARED
11:00PM JUN 10
EVENT LOG
PM11:18 SET ON
AIR SENSITIVITY
PM11:18 SET CONT
SIZE 100 mL
PM11:18 SET RATE
1.0 mL/hr
PM11:18 SHIFT
CLEARED
END OF EVENT LOG
Sample History Event Log
END OF
SHIFT TOTALS
CLEAR TOTALS?
YES OR NO
I
Tip: Press [HELP] to display shift totals when the pump is not in PROGRAMMING mode or an alarm condition.
Hospira GemStar Manual 430-10984-001 (Rev. 03/06) 59
Page 67
4. Options Menu
Volume Totals
To display the volume totals:
1 Press [
2 Select
3 Press [
OPTIONS] and select HISTORIES.
VOLUME TOTALS from the Histories Menu.
BACK-UP] when done to exit OPTIONS.
To clear the volume totals:
1 Follow the steps for displaying volume totals.
2 At the end of the volume totals, the pump displays
CLEAR TOTALS? YES OR NO:
Press [YES/ENTER] to clear the volume totals.
Press [
NO] to exit OPTIONS
without clearing the volume totals.
Configuration
To display the pump configuration:
1 Press [
OPTIONS] and select HISTORIES.
2 Select CONFIGURATION from the Histories Menu.
3 Press [
BACK-UP] when done to exit OPTIONS.
END OF VOLUME TOTALS CLEAR TOTALS?
YES OR NO
Clear History
To clear the history event log and bolus history:
1 Press [
OPTIONS] and select HISTORIES.
2 Select CLEAR HISTORY from the Histories Menu.
3 The pump display
OPTIONS.
I
Tip: Press [BACK-UP] to exit OPTIONS immediately.
60 Hospira GemStar Manual 430-10984-001 (Rev. 03/06)
CLEARING HISTORY for 3–4 seconds before exiting
Page 68
Bolus History
To display the bolus history:
4. Options Menu
G
1 Press [
2 Select
OPTIONS] and select HISTORIES.
BOLUS HISTORY from the
Histories Menu.
3 Press [
BACK-UP] when done to exit OPTIONS.
TIME IS 09:15 AM
FRI, JUN 11, 05
BOL DEL 30 BOL DEM 37
BOL AMT 30.0 mL
06:10A TO 09:14A
Sample Bolus History
Note: Bolus History is only available if the current program has a bolus programmed, or had a bolus programmed at one time. The Bolus History is
maintained in hourly increments for the latest 48 hours. The Bolus History is cleared when a new program is entered, a Speed Protocol is recalled, or the History is cleared.
Operation Test
The Operation Test checks several pump functions to ensure the GemStar® Pump is working properly. Refer to Chapter 8 Operation Test on page 99 for instructions.
Hospira GemStar Manual 430-10984-001 (Rev. 03/06) 61
Page 69
4. Options Menu
Keypad Lock
1 Press [OPTIONS] and select KEYPAD LOCK from
the Options Menu.
2 Select the desired lock level.
The following lock levels, listed from lowest to highest
1 REVIEW PROGRAM
2 HISTORIES
3 KEYPAD LOCK
4 PUMP SETTINGS
5 SET CLOCK
6 PRINT
7 SPEED PROTOCOL
restrictions, are available:
Lock Level Abbreviation
Rate Change Lock
Container Lock C
Full Lock F
R
1 RATE CHANGE
2 CONTAINER 3 FULL LOCK
When the pump is locked under a particular level, a small lock symbol ( ) appears next to the applied lock level as shown above. When the pump is in
RUN or STOP mode, the lock symbol
appears in the lower left corner of the display and the lock level abbreviation appears in the lower right corner.
R
Clinicians: refer to Chapter 5 Clinician Instructions on page 77 for detailed instructions.
62 Hospira GemStar Manual 430-10984-001 (Rev. 03/06)
Page 70
GemStar® Keypad Lock Level Functions
U = Unlock R = Rate Change Lock
Function U R C F Comments
Lock/Unlock Keypad
Power Pump On or Off
Start or Stop Infusion
Access Display, Help, Print, and Program Review Functions
Deliver a Bolus
Silence an Alarm
Select a New Container
Purge
Change a Program
Review/Recall Speed Protocol
Assign or Delete Speed Protocol
Set Air Sensitivity, Distal Occlusion Level, Sound Level, and Clock
Clear Shift Totals
Program and Deliver a Loading Dose (Outside of Programming Mode)
Enter a New Program
Operation Test
Clear History
C = Container Lock F = Full Lock
• = Function is Available
If programmed.
Under Rate Change Lock, only the delivery rate or bolus dose volume can be changed and must be within the preset values.
4. Options Menu
Hospira GemStar Manual 430-10984-001 (Rev. 03/06) 63
Page 71
4. Options Menu
Change Pump Settings
1 Press [OPTIONS] and select PUMP SETTINGS.
2 Press the number key for the desired pump setting.
3 Follow the display messages. Refer to the
instructions of the desired pump setting for more information.
Select this item from the Pump Settings Menu:
AIR SENSOR
OCCL SENSOR
SOUND LEVEL
To do this function:
Change the air sensor sensitivity.
Change the distal occlusion sensor sensitivity.
Adjust the audible alarm level.
Comments:
The pump alarms when it detects an air bubble of approximately 0.5 mL when ON; or at approximately 2 mL when 2 mL is selected. The air alarm does not sound if OFF is selected.
Note: The pump setting is inactivated if custom-configured (by your facility) to ON or 2mL.
The air alarm can also be set during
programming if the current therapy allows
it.
Default Setting: High
The pump alarms when distal pressure is
greater than the selected occlusion
sensor limit.
Default Setting: Maximum
The alarm cannot be turned off
completely. The alarm automatically
reverts to the maximum volume if the user
does not respond within one minute.
1 REVIEW PROGRAM
2 HISTORIES
3 KEYPAD LOCK
4 PUMP SETTINGS
5 SET CLOCK
6 PRINT
7 SPEED PROTOCOL
G
Note: The current setting flashes on the display.
64 Hospira GemStar Manual 430-10984-001 (Rev. 03/06)
Page 72
4. Options Menu
Air Sensitivity
WARNING: To reduce the risk of infusing air, use an air-eliminating
filter when the air sensitivity is set to OFF.
G
1 Press [
OPTIONS] and select AIR SENSITIVITY. The current air sensitivity
flashes on the display.
2 Select the desired Air Sensitivity level.
Note: If OFF is selected, the pump displays:
ELIMINATING FILTER IS
Air
RECOMMENDED. ENTER TO CONFIRM
You must press [YES/ENTER] to confirm.
3 After selecting the desired Air Sensitivity, the
pump displays:
XXXX TO XXXX
Press [
CHANGE AIR ALARM FROM
:
NO] to keep the current air sensitivity.
The current air alarm status displays and the pump exits
Press [
OPTIONS.
YES/ENTER] to accept the new setting. The new air alarm status displays
and the pump exits OPTIONS.
Occlusion Sensor
1 Press [
2 Select
OPTIONS] and select PUMP SETTINGS.
DISTAL OCCLUSION from
the Pump Settings Menu.
3 Select the desired setting.
4 The pump displays
MAKE XXXXX DEFAULT?
Press [NO] to temporarily change the distal occlusion setting. The new setting will be in effect until a new program is entered or the pump has been off for more than 24 hours.
Press [
YES/ENTER] to permanently change the distal occlusion setting. The
new setting will remain in effect until the setting is changed again through the Options Menu.
AIR SENSITIVITY 1 ON 2 2 mL 3 OFF
CHANGE AIR ALARM
FROM: XXXX
TO: XXXX
YES OR NO
DISTAL OCCLUSION 1 HIGH 2 MEDIUM 3 LOW
Hospira GemStar Manual 430-10984-001 (Rev. 03/06) 65
Page 73
4. Options Menu
Sound Level
1 Press [
2 Select
3 When the pump displays the Sound Level function, it
G
4 Press [YES/ENTER] when done.
Set Clock
1 Press [OPTIONS] and select SET CLOCK.
2 Select 12-Hour or 24-Hour format (24-Hour format is
3 Use the arrow keys to set the weekday and month.
4 Use the number keys to enter the current day, year,
G
OPTIONS] and select PUMP SETTINGS.
SOUND LEVEL from the Pump Settings Menu.
sounds intermittent beeps at the current sound level. The blocks on the top line of the display indicate the current sound level.
Press to increase the sound level.
Press to decrease the sound level.
Press
Note: The alarm cannot be turned off completely. The alarm automatically reverts to the maximum volume if the user does not respond within one minute.
shown in example).
and time.
Note: When entering the time in 12-Hour format, press or to toggle between AM and PM.
[CHANGE] to restore the sound level to the default setting.
-  +
USE TO ADJUST THE SOUND LEVEL ENTER WHEN DONE
1 REVIEW PROGRAM
2 HISTORIES
3 KEYPAD LOCK
4 PUMP SETTINGS
5 SET CLOCK
6 PRINT
7 SPEED PROTOCOL
SET CLOCK TYPE 1 SELECT 12 HOUR 2 SELECT 24 HOUR
5 Press [
G
I
Note: The clock cannot be changed while an intermittent or variable time program is in progress.
Tip: To view the clock from the RUN or STOP mode, press [ [1]. The clock displays until [1] is released.
66 Hospira GemStar Manual 430-10984-001 (Rev. 03/06)
YES/ENTER] to confirm.
YES/ENTER ], then immediately press and HOLD
TIME IS 20:35
MON, JUL 26, 05
ENTER TO CONFIRM
Page 74
Print
WARNING: Ensure a cassette is installed in the pump before
connecting to a printer or computer.
If the pump is connected to a patient: BEFORE connecting the pump to a computer or printer, press
STOP], close the slide clamp, and eject the cassette.
[ Leave the pump set connected to the patient. Install another cassette in the pump before connecting to a printer or computer.
Note: Ensure the printer is connected to the pump BEFORE selecting a print function. Refer to Connecting to the Seiko Printer on page 73 for more
G
information.
4. Options Menu
1 Press [OPTIONS] and select PRINT.
2 Press the number key for the desired print function.
3 When printing is complete, the pump exits
Select this item from the Print Menu:
PRINT HISTORY
PRINT PROGRAM
PRINT PROTOCOL
PRINT BOLUS Hx
STOP PRINT
To do this function:
Print the current program, shift and volume totals, and history event log (400 events).
Print the current program.
Print a specified speed protocol or a complete list of all speed protocols.
Print the current program, shift and volume totals, and the bolus history (Hx) for the latest 48 hours.
Cancel printing. This function is only available if the pump is printing.
OPTIONS.
1 REVIEW PROGRAM
2 HISTORIES
3 KEYPAD LOCK
4 PUMP SETTINGS
5 SET CLOCK
6 PRINT
7 SPEED PROTOCOL
Hospira GemStar Manual 430-10984-001 (Rev. 03/06) 67
Page 75
4. Options Menu
Speed Protocol
1 Press [OPTIONS] and select SPEED PROTOCOL.
2 Press the number key for the desired protocol
function.
3 Follow the display messages. Refer to the
instructions for Speed Protocol functions for more information.
Note: Up to nine speed protocols are stored at one time. The speed protocols stay in memory, even when the pump is powered down, until they are deleted.
Select this item from the Speed Protocol Menu:
ASSIGN
REVIEW/RECALL
DELETE
Assign
1 Press [
2 Select
OPTIONS] and select SPEED PROTOCOL.
ASSIGN from the Speed Protocol menu.
To do this function:
Store a program in pump memory for later use. Review assigned protocols from the RUN or STOP
mode. Recall an assigned protocol from the STOP mode. Delete protocols from pump memory.
1 REVIEW PROGRAM
2 HISTORIES
3 KEYPAD LOCK
4 PUMP SETTINGS
5 SET CLOCK
6 PRINT
7 SPEED PROTOCOL
3 Select [
If the selected speed protocol is already assigned, the pump displays
PROTOCOL ASSIGNMENT?
1]–[9] to assign to the new protocol.
REPLACE
Press [YES/ENTER] to replace the protocol. The pump displays the new speed protocol and exits
Press [
68 Hospira GemStar Manual 430-10984-001 (Rev. 03/06)
NO] to keep the existing protocol and exit OPTIONS.
OPTIONS.
Page 76
Review/Recall
1 Press [
2 Select
OPTIONS] and select SPEED PROTOCOL.
REVIEW/RECALL from the Speed Protocol menu.
3 Select the desired protocol to review/recall.
4 Review the program. Press to begin review.
5 Press [
YES/ENTER]:
If in the
STOP mode, press [1] to recall the protocol; or press [2] to review
another protocol.
If in the
RUN mode, the pump exits OPTIONS.
Delete
4. Options Menu
1 Press [
2 Select
OPTIONS] and select SPEED PROTOCOL.
DELETE from the Speed Protocol Menu.
3 Select the desired protocol to delete.
4 Press [
YES/ENTER] to confirm deletion or [NO] to exit OPTIONS without deleting.
Hospira GemStar Manual 430-10984-001 (Rev. 03/06) 69
Page 77
4. Options Menu
Connecting to the Data Port
Use the data port to connect the GemStar® Pump to a Seiko® DPU 414 or compatible serial printer to print the Operation Test results or the files available from the Print Menu. Using communication software included with most computer operating systems (or available from local sources), these files can also be transferred to a computer.
WARNING: Ensure a cassette is installed in the pump before
connecting to a printer or computer.
If the pump is connected to a patient: BEFORE connecting the pump to a computer or printer, press
STOP], close the slide clamp, and eject the cassette.
[ Leave the pump set connected to the patient. Install another cassette in the pump before connecting to a printer or computer.
Equipment Needed
Connect to Printer Connect to Computer
GemStar® Pump
Hospira Serial Printer Cable List 13078-01
Seiko DPU 414 or Compatible Serial Printer
Serial DB9 Female-to-Female Gender Changer*
Null Modem*
Computer*
*Available from local sources.
••
••
70 Hospira GemStar Manual 430-10984-001 (Rev. 03/06)
Page 78
4. Options Menu
Configuration Settings
Refer to the printer or computer documentation for information on configuration settings. Configure the printer or computer as follows to use the data port:
G
Baud Rate
Data Bits
Stop Bits
Parity
Carriage Return (+Line Feed)
Flow Control =
Note: When using the Seiko DPU 414 printer, the switch settings MUST be configured to work with the GemStar® Pump.
To configure the switch settings on the Seiko DPU 414:
1 Press and hold the ONLINE button while sliding the POWER switch to ON.
2 When the list of the current settings start printing, release the ONLINE button.
3 Set the switches according to the following table.
Press the ONLINE button to continue setting switches.
Press ONLINE to set switch ON.
Press FEED to set switch OFF.
4 Continue setting switches until “Dip SW setting complete!!” is printed.
= 2400
=8
=1
= None
= CR+LF
None
Hospira GemStar Manual 430-10984-001 (Rev. 03/06) 71
Page 79
4. Options Menu
Switch Settings for Seiko DPU-414 Printer
Switch Function Setting Button to Press
SW1-1 Input Method: Serial Off FEED
SW1-2 Printing Speed: High On ONLINE
SW1-3 Auto Loading: On On ONLINE
SW1-4 CR function: CR and LF On ONLINE
SW1-5 DIP SW setting
SW1-6 Printing density: 100% Off FEED
SW1-7 Printing density: 100% On ONLINE
SW1-8 Printing density: 100% On ONLINE
SW2-1 Print Mode: 40 columns On ONLINE
SW2-2 User defined char: On On ONLINE
SW2-3 Ordinary
SW2-4 Zero Font: No slash On ONLINE
SW2-5 Character Set: American On ONLINE
SW2-6 Character Set: American On ONLINE
SW2-7 Character Set: American On ONLINE
SW2-8 Character Set: American Off FEED
SW3-1 Data bit length: 8 bits On ONLINE
SW3-2 Parity: No parity On ONLINE
SW3-3 Parity condition: Odd On ONLINE
SW3-4 Flow control: H/S busy On ONLINE
SW3-5 Baud rate: 2400 bps On ONLINE
SW3-6 Baud rate: 2400 bps Off FEED
SW3-7 Baud rate: 2400 bps Off FEED
SW3-8 Baud rate: 2400 bps On ONLINE
Enabled On ONLINE
command:
On On ONLINE
Characters:
72 Hospira GemStar Manual 430-10984-001 (Rev. 03/06)
Page 80
4. Options Menu
Connecting to the Seiko Printer
WARNING: Ensure a cassette is installed in the pump before
connecting to a printer or computer.
If the pump is connected to a patient: BEFORE connecting the pump to a computer or printer, press [
STOP], close the slide clamp, and eject the cassette.
Leave the pump set connected to the patient. Install another cassette in the pump before connecting to a printer or computer.
Refer to the printer documentation or contact the printer manufacturer for more information regarding Electrostatic Discharge (ESD) susceptibility of printer connector ports.
CAUTION: Do not connect the GemStar® AC Mains Adaptor to a printer.
G
Note: For maximum battery life, operate the pump on AC mains when connected to a printer or computer.
1 Install a cassette into the pump. 2 Connect the GemStar® serial cable to the printer and to the data port on the
bottom of the pump.
3 Ensure the printer is ONLINE and ready to receive data. Refer to printer
documentation for more information.
4 Ensure paper is loaded in the printer.
5 Select the desired print function from the pump menu.
Hospira GemStar Manual 430-10984-001 (Rev. 03/06) 73
Page 81
4. Options Menu
Connecting to a Computer
WARNING: Ensure a cassette is installed in the pump before
connecting to a printer or computer.
If the pump is connected to a patient: BEFORE connecting the pump to a computer or printer, press
STOP], close the slide clamp, and eject the cassette.
[ Leave the pump set connected to the patient. Install another cassette in the pump before connecting to a computer or printer.
Refer to the computer documentation or contact the computer manufacturer for more information regarding Electrostatic Discharge (ESD) susceptibility of computer connector ports.
G
I
Note: For maximum battery life, operate the pump on AC mains when connected to a printer or computer.
1 Ensure slide clamp or lower CAIR® clamp is open prior to inserting cassette.
2 Install a cassette into the pump.
3 Connect the GemStar® serial cable to the null modem.
4 Connect the null modem to the serial gender changer.
5 Connect the serial gender changer to the serial port of the computer.
6 Connect the GemStar® serial cable to the data port on the bottom of the pump.
7 Configure the communication
software on the computer. Refer to Configuration Settings on page 71.
Tip: After the data transfer is complete, use any text editor to view or print the file.
74 Hospira GemStar Manual 430-10984-001 (Rev. 03/06)
Page 82
Transferring Data Using a Computer
4. Options Menu
G
The GemStar® can be connected to a Windows®-compatible computer. The following instructions apply to computers using Microsoft Windows NT
Creating a New Connection
Note: The following procedure is only needed to create a “New Connection” the first time the pump is connected to the computer.
To create a new connection:
1 Follow steps 1–4 in Using an Existing Connection on the following page.
2 If a message appears asking if a modem is necessary for a connection, click NO.
3 Enter a name for the new connection (e.g., GemStar® Print Files) and choose an
icon.
4 Click OK.
5 From the Connect drop-down menu, select the appropriate COM port.
6 Click OK.
7 Verify that the following settings are correct:
®
4.0 or higher.
Baud Rate = 2400
Data Bits =8
Stop Bits =1
®
Windows® 95, Windows® 98, or
Parity = None
Flow Control = None
8 Click OK.
9 Select the Properties icon.
10 Select the Settings tab.
11 Select the ASCII setup button.
12 Ensure ‘Append line feeds to incoming line ends’ is checked.
13 Click OK in the ASCII setup window.
Hospira GemStar Manual 430-10984-001 (Rev. 03/06) 75
Page 83
4. Options Menu
14 Click OK in the Properties window.
15 From the menu, select File > Save.
16 Exit HyperTerminal.
Using an Existing Connection
To transfer data from the GemStar® Pump to a computer:
1 From the Windows
2 Select Programs.
3 Select Accessories.
4 Select HyperTerminal.
5 Double click on the icon with the name entered in step 3, To create a new
connection, on page 70.
6 From the menu, select Call > Connect.
7 From the menu, select Trans fer > Capture Text.
8 Type the path and file name where the printout should be saved.
Ensure the file extension is “.txt”.
The default path is: c:\windows\accessories\hyperterminal.
®
desktop, select the Start Menu.
Note: If the file name already exists, the new printout will be added to the end of
G
76 Hospira GemStar Manual 430-10984-001 (Rev. 03/06)
the existing file. For example, use a patient’s name or id number as the file name. Then select that file for all of that patient’s printouts.
9 Click Start.
10 Select the desired print function from the pump. Refer to the table of printing
functions on page 67.
11 Wait for the transmission to complete.
12 From the menu, select Trans fer > Capture Text > Stop.
13 Exit HyperTerminal.
14 Use any text editor to view or print the file.
Page 84
Chapter 5
Clinician Instructions
The instructions in this section apply to special features available to clinicians. Remove this section from the manual to prevent unauthorized access to these features.
The GemStar® Pump provides quick access to commonly used functions. Instead of accessing these functions by navigating through the menu system, you can enter the key combinations shown in this table:
Function Quick Access Key Combination
Unlock Keypad Place the pump in STOP mode and then enter the lock
sequence to unlock a Full Container or Rate Lock.
Displays Shift Totals Place the pump in STOP mode and then press [HELP] to
display Shift Totals.
Display Clock Place the pump in STOP mode, press [YES], and then press
and hold [1]. The clock displays until you release [1].
Program a Loading Dose
Hospira GemStar Manual 430-10984-001 (Rev. 03/06) 77
From the Pain Management STOP MODE, press [YES/ENTER], and then press [0] to program a Loading Dose.
Page 85
5. Clinician Instructions
Clinician Locking Sequence Instructions
Locking the Pump
I
To lock the pump from the
1 Press [
2 Select
3 Select the desired lock level. For
4 For
5 Press [
6 Repeat steps 4 and 5 for a bolus dose if available in the current program.
7 Enter the lock sequence number [
G
Unlocking the Pump
OPTIONS] to display the Options Menu.
KEYPAD LOCK from the Options Menu.
RATE CHANGE Lock, set the minimum and maximum programmable ranges
for the continuous delivery rate and bolus dose in a Pain Management program.
Tip: Press to toggle between the MIN and MAX values.
MIN flashes on the display. Enter the minimum value.
MAX flashes on the display. Enter the maximum value.
YES/ENTER] when done.
are entered.
Note: Default lock sequence numbers are shown.
RUN or STOP mode:
CONTAINER and FULL Lock, go to step 5.
1][3][0][0][0]. Asterisks display as the numbers
1 Press [
2 Select
3 Select the lock level to be unlocked.
4 Enter the lock sequence number for that level.
5 The pump displays the current lock level. Press [
78 Hospira GemStar Manual 430-10984-001 (Rev. 03/06)
OPTIONS] to display the Options Menu.
KEYPAD LOCK from the Options Menu.
mode the pump was in when [
OPTIONS] was pressed.
YES/ENTER] to return to the
Page 86
5. Clinician Instructions
Program and Deliver a Clinician Activated Loading Dose
This feature is available only when the keypad is UNLOCKED. The default loading dose rate is 125 mL/hr.
G
Note: Loading dose amounts are included in a 1-hour or 4-hour dose limit.
1 From the stop mode, press [
2 The pump displays:
Press [YES/ENTER].
3 Enter the loading dose. Press [
4 The pump displays:
To deliver the loading dose immediately, press [YES/ENTER].
Primary infusion DOES NOT begin automatically when the clinician activated loading dose is complete. Press [
To deliver the loading dose later, press [
The pump returns to the stop mode. The next time [ again displays:
Press [
DELIVER LOADING DOSE NOW?
YES/ENTER] to deliver the loading dose. Primary infusion continues
automatically when the loading dose is complete.
Press [
NO] to cancel the loading dose and continue primary infusion.
Shift Totals and History
Shift Totals
YES/ENTER] [0].
PROGRAM A LOADING DOSE?
YES/ENTER].
DELIVER THE LOADING DOSE NOW?
START] to continue primary infusion.
NO].
START] is pressed, the pump
1 Press [
2 Select
3 The shift totals display. Press
OPTIONS] and select HISTORIES.
SHIFT TOTALS from the Histories Menu.
and to scroll the display. When done, the
pump displays
CLEAR TOTALS? YES OR NO.
END OF
SHIFT TOTALS
CLEAR TOTALS?
YES OR NO
Press [YES/ENTER] to clear the shift totals.
Hospira GemStar Manual 430-10984-001 (Rev. 03/06) 79
Page 87
5. Clinician Instructions
Press [NO] to exit OPTIONS without clearing the shift totals.
I
Tip: Press [HELP] to display shift totals when the pump is not in PROGRAMMING mode or an alarm condition.
Clear History Event Log and Bolus History
1 Press [
2 Select
3
Key Combinations Not on Screens
The table below lists special key combinations to access information or activate an activity which is not listed on the screen.
Screen and/or Situation
If a loading dose has been programmed but not delivered, displays the DELIVER LOADING DOSE NOW? screen
CURRENT DATE AND TIME
displaying
OPTIONS] and select HISTORIES.
CLEAR HISTORY.
CLEARING HISTORY displays and the pump exits OPTIONS.
.
Press Results
[YES/ENTER] + [0] Displays the PROGRAM A LOADING DOSE?
screen
[YES/ENTER] + [1] Displays the real time clock as long as 1 is held
down
80 Hospira GemStar Manual 430-10984-001 (Rev. 03/06)
Page 88
Chapter 6
Troubleshooting
This section contains information on audible and visual alarms that may occur with the GemStar® Pump.
CAUTION: If the pump does not perform as stated in this manual, stop using it
IMMEDIATELY.
For consultation and technical support contact either an authorized Hospira representative or the Hospira Technical Support Operations hotline (available 24 hours a day).
To return the pump for service, first contact Hospira Customer Support between 7:00 AM and 4:00 PM Pacific time. Have the pump serial number available.
In the USA
Outside the USA:
I
Tip: Press [HELP] for additional information during programming and alarm conditions.
Hospira GemStar Manual 430-10984-001 (Rev. 03/06) 81
Hospira Technical Support Operations 1-800-241-4002
Hospira, Inc. Technical Support Operations 755 Jarvis Drive Morgan Hill, CA 95037
Contact your local Hospira sales office.
Page 89
6. Troubleshooting
Alarm Messages
Alarm Display Message
Y = Yes
N = No
AIR IN LINE
• Pump automatically enters
STOP
mode.
Is the pump infusing?
N A/V The specified volume
Alarm Cause/
Condition
A = Audible alarm
V = Visual alarm
of air has been detected in the line distal to the cassette while infusing.
Remedy
1 Press [SILENCE] to quiet
the alarm for 1 minute.
2 Press [STOP] to clear the
alarm.
3 Disconnect the patient
from the pump set.
4 Press and hold [PURGE].
Note: Purge rate is 250 mL/hr. The pump purges for up to 2 minutes each time you press and hold [
PURGE]. Purge
volume may be limited in a Pain Management program.
5 Reconnect the pump set to
the patient-access device.
6 Press [START] to begin
infusion.
82 Hospira GemStar Manual 430-10984-001 (Rev. 03/06)
Page 90
6. Troubleshooting
Alarm Display Message
Y = Yes
N = No
CHANGE BATTERIES N A/V The pump has
Is the pump infusing?
Alarm Cause/
Condition
A = Audible alarm
V = Visual alarm
entered the mode due to battery voltage below acceptable levels.
STOP
Remedy
1 Replace both disposable
batteries with fresh AA batteries and press to restart the pump; or connect the pump to the AC mains adaptor and the pump will automatically restart. Replace both batteries as soon as possible.
CAUTION: Ensure two fresh, disposable AA batteries are installed for backup power when using other power sources (e.g., AC Mains adaptor, docking station, battery pack). The pump will continue to operate on backup power if the external power source fails. Refer to the caution regarding Power Loss on page 13.
2 After power-on, resume the
current program. Refer to
Accessing the Delivery Mode Menu in Chapter 2
Basic Operation for more information.
Hospira GemStar Manual 430-10984-001 (Rev. 03/06) 83
Page 91
6. Troubleshooting
Alarm Display Message
Y = Yes
N = No
CHECK CASSETTE-X N A/V The cassette is
Is the pump infusing?
Alarm Cause/
Condition
A = Audible alarm
V = Visual alarm
improperly installed or not installed.
X indicates where to look for the cause of the alarm:
D = Distal
P = Proximal
Remedy
1 Press [SILENCE] to quiet
the alarm for 1 minute.
2 Press [STOP] to place the
pump in the
3 Ensure slide clamp or lower
CAIR® clamp is open prior to inserting cassette.
4 Reinstall the cassette.
STOP mode.
CAUTION: Ensure all four cassette latches are clearly visible after the cassette is installed.
Note: Ensure the tubing is behind the tubing guide (at the distal end). Refer to Loading
the Cassette on page 33.
5 Press [START] to begin
infusion.
Note: The pump cannot be powered off during a CHECK CASSETTE alarm (except during
PROGRAMMING mode).
CALL
1.800.XXX.XXXX
CODE: NN/MMM/TTT
(SERVICE ALARM)
84 Hospira GemStar Manual 430-10984-001 (Rev. 03/06)
N A/V The pump has
detected a condition that may require service.
1 Write down the telephone
number and code.
2 Power-off the pump. If the
pump does not power-off, disconnect external power and remove the batteries.
3 Call the telephone number
shown on the display message.
Page 92
6. Troubleshooting
Alarm Display Message
Y = Yes
N = No
DIST. OCCLUSION N A/V The pump has
Is the pump infusing?
Alarm Cause/
Condition
A = Audible alarm
V = Visual alarm
detected sustained distal pressure over the user-selected limit.
To avoid a bolus infusion after an occlusion, complete the following steps:
1Press [STOP]. 2 Close the distal
clamp.
3 Press the
cassette release button to remove the cassette from the pump.
4 Open the flow
stop. 5 Wait 10 seconds. 6 Close the flow
stop. 7 Load the
cassette into the
pump. 8 Open the distal
clamp. 9 Review the
program. 10 Press [START]
to start the
pump.
Remedy
1 Press [SILENCE] to quiet
the alarm for 1 minute.
2 Press [STOP] to place the
pump in the stop mode.
3 Check for sources of distal
occlusion: ` Kinked tubing below the
pump
` Clamped patient-
access device
` Clogged I.V. filter ` Other obstructions
4 If the problem continues,
remove and reinstall the cassette.
CAUTION: Ensure all four cassette latches are clearly visible after the cassette is installed.
Note: Ensure the tubing is behind the tubing guide (at the distal end). Refer to Loading
the Cassette on page 33.
5 Press [START] to begin
infusion.
Hospira GemStar Manual 430-10984-001 (Rev. 03/06) 85
Page 93
6. Troubleshooting
Alarm Display Message
Y = Yes
N = No
LOW BATTERIES Y A/V The pump has
Is the pump infusing?
Alarm Cause/
Condition
A = Audible alarm
V = Visual alarm
detected battery voltage below acceptable levels and the batteries are the primary power source.
Note: The pump
allows the user to clear the
LOW BAT-
TERIES
the condition has not been corrected after 15 minutes, the alarm continues.
If the pump is infusing and a
LOW BAT-
TERIES alarm
occurs, the pump con­tinues for approxi­mately 30 minutes* before it can no longer deliver due to insuffi­cient power.
*If an external power source is used, the rate is changed, or the pump is powered-on after a
LOW BAT-
TERIES
those same batteries are still in the pump, this interval may be different.
alarm. If
alarm and
Remedy
1 Press [SILENCE] to quiet
the alarm for 15 minutes.
2 Press [STOP] to place the
pump in the
3 Connect the pump to AC
mains adaptor, rechargeable battery pack, or press the On/Off key to turn off the pump and replace both disposable batteries with fresh AA batteries.
STOP mode.
CAUTION: In the event that the internal batteries are removed while the pump is operating, the power loss event may not be logged in the history, the programmed therapy may not be retained, and the alarms (audible and visual) may not occur. Refer to the caution regarding Power Loss on page 13.
4 Power on the pump and
resume the current program. Refer to Accessing the Program Menu in Section 2 for more information page 13.
5 Press [START] to begin
infusion.
Note: In case of low battery alarm during printing, replace the batteries and reprint or Plug in AC main and reprint.
86 Hospira GemStar Manual 430-10984-001 (Rev. 03/06)
Page 94
6. Troubleshooting
Alarm Display Message
Y = Yes
N = No
POWER LOSS N A/V The pump has
Is the pump infusing?
Alarm Cause/
Condition
A = Audible alarm
V = Visual alarm
detected a power failure due to one of the following:
• Batteries have been removed while the pump was on battery source.
• External power has failed or has been disconnected and no batteries are installed.
• Pump has been powered on with insufficient power source.
Remedy
1 Connect the pump to AC
mains adaptor, rechargeable battery pack, or replace both disposable batteries with fresh AA batteries.
2 Power-on the pump and
resume the current program. Refer to Accessing the Program Menu in Section 2 for more information. Review the program if required.
3 Press [START] to begin
infusion.
CAUTION: Ensure two fresh, disposable AA batteries are installed for backup power when using other power sources (e.g., AC Mains adaptor, docking station, battery pack). The pump will continue to operate on backup power if the external power source fails. Refer to the caution regarding Power Loss on page 13.
Hospira GemStar Manual 430-10984-001 (Rev. 03/06) 87
Page 95
6. Troubleshooting
Alarm Display Message
Y = Yes
N = No
PROX. OCCLUSION N A/V The pump has
Is the pump infusing?
Alarm Cause/
Condition
A = Audible alarm
V = Visual alarm
detected sustained negative pressure on its proximal side.
Remedy
1 Press [SILENCE] to quiet
the alarm for 1 minute.
2 Press [STOP] to place the
pump in the stop mode.
WARNING: When a Proximal Occlusion alarm activates, DO NOT power-off the pump.
3 Check for sources of
proximal occlusion: ` Kinked tubing above
pump
` Blockage in the container
and in the spike
` Closed clamps above
pump
` Empty container ` High torque syringe ` Other obstructions
4 Remove obstructions. 5 Press [START] to begin
delivery.
6 If the problem continues,
ensure slide clamp or lower CAIR® clamp is open. Power-off the pump. Reinstall the cassette. Turn on the pump.
Note: Ensure the tubing is behind the tubing guides (at the distal end).
CAUTION: Ensure all four cassette latches are clearly visible after the cassette is installed.
Refer to Loading the Cassette on page 33
.
88 Hospira GemStar Manual 430-10984-001 (Rev. 03/06)
Page 96
6. Troubleshooting
Alarm Display Message
Y = Yes
N = No
USING BATTERIES Y V This is a normal
Is the pump infusing?
Alarm Cause/
Condition
A = Audible alarm
V = Visual alarm
response when:
• The pump has been powered on with only disposable batteries as a power source; or
• The pump has switched to internal power (disposable batteries) due to a loss of external power or a reduction in external power below acceptable levels.
Remedy
When powering on the pump:
• Press [YES/ENTER] to
confirm disposable batteries are the only power source.
When switching from AC Mains power:
• Press [SILENCE] to clear the
alarm.
CAUTION: Ensure two fresh, disposable AA batteries are installed for backup power when using other power sources (e.g., AC Mains adaptor, docking station, battery pack). The pump will continue to operate on backup power if the external power source fails. Refer to the caution regarding Power Loss on page 13.
Hospira GemStar Manual 430-10984-001 (Rev. 03/06) 89
Page 97
6. Troubleshooting
Alarm Display Message
Y = Yes
N = No
USING EXT BATT Y V This is a normal
Is the pump infusing?
Alarm Cause/
Condition
A = Audible alarm
V = Visual alarm
response when:
• The pump has been powered on with external batteries as a power source; or
• The pump has switched to external battery power due to a loss of AC mains power; or
• The pump has switched to external battery power while running on internal batteries.
Remedy
When powering on the pump:
• Press [YES/ENTER] to
confirm external battery power source.
When switching from AC Mains power:
• Press [SILENCE] to clear the
alarm.
CAUTION: Ensure two fresh, disposable AA batteries are installed for backup power when using other power sources (e.g., AC Mains adaptor, docking station, battery pack). The pump will continue to operate on backup power if the external power source fails. Refer to the caution regarding Power Loss on page 13
90 Hospira GemStar Manual 430-10984-001 (Rev. 03/06)
Page 98
Alert Messages
6. Troubleshooting
Alert Display Message Is the
pump infusing?
Y = Yes
N = No
ALMOST EMPTY Y A/V Delivery will complete in
CHECK PRINTER Y A/V • The printer has not
AlarmCause/Condition Remedy
A = Audible alarm
V = Visual alarm
30 minutes or less.
For bolus only program, delivery will complete during the next bolus delivery.
responded to commands from the pump.
• The printer is not properly connected to the pump.
• Press [STOP] to clear the alarm and put the pump in Stop Mode.
• Press [SILENCE] to clear the alarm.
• No additional alarms sound.
Note: This alert is avail­able only with Pain Man­agement therapy.
1 Press [SILENCE] to
quiet the alarm for 2 minutes.
2 Ensure the printer is
properly connected to the pump by checking the following:
` Printer is powered on
and in ONLINE mode.
` Cable is plugged in to
both the pump and the printer.
` Cable part number
and printer settings are correct.
3 If the problem continues,
select
STOP PRINT
from the Print Menu to cancel all print jobs.
Hospira GemStar Manual 430-10984-001 (Rev. 03/06) 91
Page 99
6. Troubleshooting
Alert Display Message Is the
pump infusing?
Y = Yes
N = No
EMPTY CONTAINER N A/V The pump has delivered
FLASHING DISPLAY
• Line four of the display flashes.
AlarmCause/Condition Remedy
A = Audible alarm
V = Visual alarm
the programmed container size or the pump has delivered the programmed VTBI; AND Auto KVO is delivering (if available in current program).
N A/V A Non-programming
Callback has occurred because the pump has been waiting for more than three minutes for a keypress on a non­programming input or a confirmation display.
1 Press [SILENCE] to
quiet the alarm for 2 minutes.
2 Press [STOP] to place
the pump in the mode and clear the alarm.
3 Use the NEW
CONTAINER function to repeat the current program. Refer to
Starting a New Container in Chapter 2
Basic Operation.
STOP
Note: Titration is allowed at
this time.
Follow the instructions on the display to clear the alarm.
92 Hospira GemStar Manual 430-10984-001 (Rev. 03/06)
Page 100
6. Troubleshooting
Alert Display Message Is the
pump infusing?
Y = Yes
N = No
NO INTERNAL BATT
NO POWER LOSS
ALARM AVAILABLE
ENTER TO CONFIRM
AlarmCause/Condition Remedy
A = Audible alarm
V = Visual alarm
N A/V No disposable batteries
or dead disposable batteries; AND external power.
1 Press [YES/ENTER] to
confirm.
2 Power-off the pump and
install two fresh, disposable AA batteries as a backup power source.
CAUTION:Ensure two fresh, disposable AA batteries are installed for backup power when using other power sources (e.g., AC Mains adaptor, docking station, battery pack). The pump will continue to operate on backup power if the external power source fails. Refer to the caution regarding Power Loss on page 13.
PROGRAMMING
INCOMPLETE
START
• START flashes on the display. Other parts of the display are not affected.
Hospira GemStar Manual 430-10984-001 (Rev. 03/06) 93
N A/V Programming has not
been completed and no keys have been pressed within three minutes.
N A/V A Stop Mode Callback
has occurred because the programmed pump has remained in the STOP mode for more than three minutes.
1 Press any key to clear
the alarm.
2 Complete the program
entries.
1 Press [SILENCE] to
quiet the alarm for 3 minutes.
2 Follow the instructions
on the display to clear the alarm.
Loading...