Hospira, Inc., Lake Forest, IL 60045
430-10984-001 (Rev
. 01/06)
Page 2
Hospira GemStar® Infusion Pump
System Operating Manual
For use with the
Hospira GemStar® Pain Management Pump
List 13088-04-01/02
PLEASE
Read this entire manual before
using the Hospira GemStar® Pump
This manual is designed for use by healthcare professionals, caregivers, and patients.
The Hospira Technical Support Operations hotline is available 24 hours a day (in the
USA) to provide consultation and technical assistance regarding the Hospira
GemStar® Pump.
The Hospira GemStar® Pain Management Pump is a small and lightweight,
single-channel infusion device designed for use in the home, in the hospital, or
anywhere electronic infusion is required. The GemStar® Pump can be powered by
AC mains adaptor, rechargeable battery pack, docking station, or two disposable AA
alkaline batteries. When powered by batteries, the GemStar® Pump is ideal for
ambulatory patients.
The following conventions are used throughout this manual:
ConventionUseExample
[ALL CAPS]Pump keysPress[YES/ENTER] to continue.
ALL CAPS
WARNING: A WARNING message contains special safety
Display messagesThe pump displays:
ENTER TO CONFIRM.
information and must be observed at all times. Failure
to observe a warning message could be life
threatening.
CAUTION:A CAUTION contains information that could prevent
irreversible product damage or hardware failure. Failure to
observe a caution could result in serious patient or user injury.
G
I
G
Note: A NOTE contains information or examples to help explain instructions.
Tip: A TIP contains information that makes the pump easier to use.
Note: Examples are provided only to help explain a pump function and may not
be clinically relevant. Graphics are for illustrative purposes only and may not
reflect the actual product.
Adjusts the distal occlusion pressure limits,
air sensor sensitivity, real-time clock, and
audible alarm volume.
Signals audible and visual alarms when
attention is required.
Sounds a single beep when any key is
pressed. If an invalid key is pressed, the
pump sounds a flutter tone.
Restricts access to various options. Four
lock levels are available.
Repeats the current program without
reprogramming the pump.
Performs a self-test each time the pump is
powered on.
Performs user-assisted tests to verify the
pump is operating properly.
Maintains a history event log, with a date
and time stamp of each event for each
program. Review the history on the display,
or download it to a printer or computer.
Stores up to nine frequently used programs
in the pump’s memory for quick access.
Chapter 4 Options Menu
Chapter 6
Troubleshooting
Chapter 1 Introduction,
GemStar® Pump Layout
Chapter 4 Options Menu
and Chapter 5 Clinician
Instructions
The GemStar® Pump uses a menu system that provides step-by-step instructions
through all of the pump’s functions. Refer to the following table for a list of functions
in each menu.
Use the keypad to select menu items, respond to display messages, and enter numeric
values. For example, press [
2] to select HISTORIES from the Options Menu.
G
I
Note: Menu items may vary depending on the pump’s configuration.
Tip: A menu item does not have to be on the display to be selected. For example,
when the Options Menu first displays, only the first four menu items are shown.
Instead of pressing the down arrow to see the other menu items, press [
directly to SET CLOCK.
Physicians or certified, licensed healthcare professionals should always oversee
infusions. Healthcare professionals should always supervise pump users and instruct
users in pump operation and troubleshooting. Instruction should include how to
prevent intravenous (I.V.) related complications, including how to prevent accidental
infusion of air.
The pump is suitable for intravenous, arterial, subcutaneous, short-term epidural
infusion of general I.V. fluids and medications. The epidural route is recommended to
provide anesthesia or administer analgesia for periods up to 96 hours.
Note: Epidural administration of anesthetics is recommended in the continuous
G
mode only. Epidural administration of analgesics may be delivered by continuous,
bolus, or continuous with bolus.
For epidural use, the administration of drugs is restricted to those anesthetic and
analgesic drugs approved for continuous epidural administration: e.g., Chloroprocaine
Hydrochloride USP, Lidocaine Hydrochloride USP, and Morphine Sulfate Injection
USP (Preservative Free).
1. Introduction
WARNING: Delayed respiratory depression following continuous
epidural administration of preservative-free morphine
sulfate has been reported.
For epidural administration, the following is recommended:
Nylon or Teflon
Pump sets without Y-sites
Stickers for the tubing indicating ongoing administration
Note: Facilities practicing epidural anesthesia/analgesia must be staffed and
equipped to manage cardio-pulmonary resuscitation (CPR). Supplies should
include oxygen, naloxone, and other appropriate resuscitative drugs and
equipment. Monitor patients continuously (e.g., oximetry) during epidural
administration. Frequently observe patient for side effects (for up to 24 hours)
following completion of drug administration by the epidural route.
®
catheter
Page 17
1. Introduction
Contraindications for Use
Persons who do not have the mental and physical capability or the emotional stability
to properly use the GemStar® Pump should not operate this pump.
DO NOT use the following with the GemStar® Pump System:
Drugs not compatible with silicone rubber or PVC plastic
Drugs not stable under infusion conditions
Warnings and Cautions
Pay attention to the following warnings and cautions when operating the GemStar®
Pump. These warnings and cautions are repeated in the manual when relevant to the
procedure being discussed. Pay attention to all alert messages.
General Warnings and Cautions
Federal (USA) law restricts this device to sale by or on the order of a physician or
other licensed practitioner.
Manual references to specific values are approximate only, unless indicated
otherwise. Air-in-line sensitivity values are approximate only.
When a proximal occlusion alarm activates, DO NOT power-off the pump.
Closely supervise patients who are likely to be adversely affected by unintended
operations and failures, including interrupted medication or fluid infusion. Ensure
provisions for immediate corrective action are available.
Regarding Drugs Used, Pump Sets, and Containers
USE ONLY GemStar® Pump Sets with the GemStar® Pump. Use of
unauthorized sets may result in injury to the patient or damage to the pump.
Some microbore tubing delivery accuracy may be affected when used at extreme
operating temperatures (<10°C, >35°C), relative humidity (<20%) and delivery
rates (<4.9 mL/hr, >350 mL/hr). Refer to individual set labels for specific
instructions and operating conditions.
Delivery accuracy may be impacted when there is air present in the cassette.
Follow the instructions in the individual set labels to remove any visible air from
the cassette.
Ensure slide clamp or lower CAIR® clamp is open prior to inserting cassette.
When using pre-filled vials, ensure vial and injector are securely locked in
Stop infusion if signs or symptoms of infiltration occur.
GemStar® Pump Sets contain a pressure-activated anti-siphon valve to reduce the
potential for gravity flow at a typical 36-inch head height.
Failure to use the pressure-activated anti-siphon valve may result in unrestricted
flow if the cassette is not properly installed in the pump.
Always close the slide clamps before removing the cassette from the pump.
Regarding Air-In-Line and Infusion
To reduce the risk of infusing air, use an air-eliminating filter when the air
sensitivity is set to OFF.
ALWAYS prime the pump set to remove all air from the cassette, tubing, and
injection site prior to connecting to the patient.
ALWAYS disconnect the pump set from the patient prior to priming or purging.
Regarding Pump Operation
If the pump does not perform as stated in this manual, remove the pump from
service IMMEDIATELY.
If the pump does not display the self-test or if beeps do not sound at power-on,
check all power connections and install two fresh, disposable AA batteries. If the
problem continues, contact Hospira Technical Support Operations before using
the pump.
Ensure a cassette is installed in the pump before connecting to a printer or
computer. If the pump is connected to a patient: BEFORE connecting the pump
to a computer or printer, press [
cassette. Leave the pump set connected to the patient. Install another cassette in
the pump before connecting to a printer or computer.
Disconnect the pump from the patient BEFORE performing the Operation Test.
ALWAYS connect to a grounded AC mains power source when using AC mains
power. If the quality of the grounding source is in doubt, operate the pump only
with batteries.
USE ONLY the AC mains adaptors specifically labeled for use with the
GemStar® Pump.
USE ONLY the AC mains adaptor specifically labeled for use with the GemStar®
The Hospira GemStar® AC Mains Adaptor is for use with the GemStar® Pump or
GemStar® Rechargeable Battery Pack only. DO NOT
mains adaptor with other products.
During charging, the battery pack is warm. If the battery pack becomes hot to the
touch, IMMEDIATELY unplug the AC mains adaptor and contact Hospira
Technical Support Operations.
Use only the Li-Ion Docking Station and Li-Ion Battery Pack with the GemStar®
Pump.
With the Li-Ion Battery Pack, the LED indicator flashes yellow/green when an
error condition has occurred or when the temperature is too high to charge the
battery pack. In temperatures above 95°F (35°C), cycle AC power to confirm if
battery is charged or if error condition exists. If LED flashes green/yellow, do not
rely on external battery. Move device to cooler temperature and recharge prior to
use. Refer to the battery pack insert for important information and precautions
regarding activating and charging the Li-Ion Battery Pack.
When removing the cord (from the docking station, bottom of the battery pack, or
bottom of the pump), firmly grasp the connector and pull straight out. DO NOT
twist or bend the cord or connector.
To ensure proper pump operation, ALWAYS replace BOTH disposable batteries
with fresh, disposable AA batteries when a change is required.
ALWAYS fully close the battery door when using disposable batteries.
DO NOT use rechargeable batteries in the battery compartment.
Ensure two fresh, disposable AA batteries are installed for backup power when
using other external power sources (e.g., AC Mains adaptor, docking station,
battery pack). The pump will continue to operate on backup power if the external
power source fails. In the event that the internal batteries are removed while the
pump is operating, the power loss event may not be logged in the history, the
programmed therapy may not be retained, and the alarms (audible and visual) may
not occur.
ALWAYS avoid sources of high-intensity electromagnetic radiation (e.g., radio
transmitters, MRI scanners, microwave ovens, X-ray machines, and CAT
scanners).
Use of radio-frequency emitting devices, such as cellular telephones, 2-way
radios, and Electrical Surgical Devices (ESU), in close proximity to this device
may affect its operation.
Possible explosion hazard exists if the pump is used in the presence of flammable
anesthetics. NEVER use the pump in the presence of flammable or explosive
Non-hazardous, low-level electrical potentials are commonly observed when
fluids are administered using infusion devices. These potentials are well within
accepted safety standards, but may create artifacts on voltage-sensing equipment,
such as ECG, EMG, and EEG machines. These artifacts vary at a rate that is
associated with the infusion rate. If the monitoring machine is not operating
correctly or has loose or defective connections to its sensing electrodes, these
artifacts may be accentuated so as to simulate actual physiological signals. To
determine if the abnormality in the monitoring equipment is caused by the
infusion device instead of some other source in the environment, set the infusion
device so that it is temporarily not delivering fluid (if clinically acceptable).
Disappearance of the abnormality indicates that it was probably caused by
electronic noise generated by the infusion device. Proper setup and maintenance
of the monitoring equipment should eliminate the artifact. Refer to the appropriate
monitoring system documentation for setup and maintenance instructions.
The GemStar® Pump has been tested and found to comply with EMC/EMI limits
in accordance with:
IEC/EN 60601-1-2 (2001) when using two AA disposable batteries or when
connected to the GemStar® AC Adaptor
IEC/EN 60601-1-2 (2001) (ESD exemption) when using Li-Ion Docking
Station or Li-Ion Battery Pack
When connected to other power sources or optional system components in
commercial distribution before November 30, 2004, the GemStar® system
complies with EMC/EMI limits in accordance with IEC/EN 60601-1-2 (1993).
These limits are designed to provide reasonable protection against harmful
interference in a typical medical installation. The equipment generates, uses and
can radiate radio frequency energy and, if not installed and used in accordance
with the instructions, may cause harmful interference to other devices in the
vicinity. However, there is no guarantee that interference will not occur in a
particular installation. If this equipment does cause harmful interference with
other devices, which can be determined by turning the equipment off and on, the
user is encouraged to try to correct the interference by one or more of the
following measures:
Reorient or relocate the receiving device
Increase the separation between the equipment
Connect the equipment to an outlet on a circuit different from that to which
the other device is connected
Consult the manufacturer or field service technician for help
The use of portable and mobile RF equipment may have an impact on this and
other pieces of medical equipment
Prior to using the GemStar® Pump and optional system components, refer to
Appendix A for additional information and precautionary procedures regarding
electrical potentials (electromagnetic compatibility and electrostatic discharge).
Regarding Handling and Maintenance
Use proper care during unpacking, installation, and operation of the pump. If the
pump is inadvertently mishandled, check connections and programmed data to
ensure there is no damage. Refer to Chapter 8 Operation Test for further
information.
Accessory equipment connected to the analog and digital interfaces must be
certified according to the respective IEC standards, (e.g., IEC 60950 for data
processing equipment, and ICE 60601-1 for medical equipment). Furthermore, all
configurations shall comply with the system standard IEC 60601-1-1. Any person
who connects additional equipment to the signal input or output part configures a
medical system, and is therefore responsible for ensuring that the system complies
with the requirements of IEC 60601-1-1. If in doubt, contact Hospira Technical
Support Operations.
ALWAYS avoid dropping or hitting the pump. If the pump is dropped or hit,
ALWAYS check connections and programmed data.
NEVER use sharp objects (e.g., fingernails, pens, pencils, or other probes) to
program or clean the pump.
Use care not to damage the silicone seals around the sensor bodies in the cassette
pocket.
To avoid mechanical or electronic damage, NEVER submerge pump in water or
other fluids and avoid fluid spills. If the pump becomes wet, dry it immediately
with a dry, lint-free cloth. Check connections and programmed data.
DO NOT touch connector ports. Connector ports may be susceptible to
Electrostatic Discharge (ESD) damage.
Some cleaning and sanitizing compounds may slowly degrade components made
from some plastic materials. DO NOT use compounds containing combinations of
isopropyl alcohol and dimethyl benzyl ammonium chloride (i.e., alcohol wipes).
DO NOT sterilize by heat, steam, ethylene oxide (ETO), or radiation. Apply
disinfectants to the outside surface of the pump only. DO NOT use abrasive
cleaners or materials on the pump. Using abrasive cleaners or cleaning solutions
not recommended by Hospira may result in product damage.
Please comply with local disposal and recycling regulations as appropriate for
disposable batteries, rechargeable battery packs, medical electronic components,
and pump sets.
The GemStar® Pump may be powered by one or more of the following:
Power SourceDescriptionAmbulatoryHospital Home
GemStar®
AC Mains Adaptor
(USA)
GemStar®
AC Mains Adaptor
(Universal)
GemStar®
AC Mains Adaptor
(Universal)
GemStar® Docking
Station
GemStar® Battery
Pack
Disposable Batteries Two disposable AA alkaline
Wall-mount adaptor plugs into a
110 VAC mains supply.
Wall-mount adaptor plugs into a
100-240 VAC mains supply.
Tabletop adaptor plugs into a
100–240 VAC mains supply.
Plugs into a 100–240 VAC Volt
mains supply while mounted on an
I.V. pole. The docking station’s
rechargeable internal batteries
power the GemStar® Pump during
periods of transport.
The battery pack’s rechargeable
internal batteries power the
GemStar® Pump during periods of
transport or when use of AC power
is not desirable.
batteries (available from local
sources)
••
••
••
•••
•••
•••
CAUTION: To ensure proper pump operation, ALWAYS replace BOTH
batteries with fresh, disposable AA batteries when a change is
required.
DO NOT use rechargeable batteries in the battery compartment.
Ensure two fresh, disposable AA batteries are installed for backup
power when using other power sources (e.g., AC Mains adaptor,
docking station, battery pack). The pump will continue to operate
on backup power if the external power source fails. Refer to the
caution regarding Power Loss on page 13.
1Connect the cord of the AC mains adaptor to the connector port labeled “3 VDC”
on the bottom of the pump.
2Plug the AC mains adaptor into a standard wall outlet.
CAUTION: ALWAYS connect to a grounded AC mains power source when
using the AC mains adaptor. USE ONLY AC mains adaptors
specifically labeled for use with the GemStar® Pump.
When removing the cord from the bottom of the pump, firmly
grasp the connector and pull straight out. DO NOT twist or bend
the cord or connector.
Ensure two fresh, disposable AA batteries are installed for backup
power when using other power sources (e.g., AC Mains adaptor,
docking station, battery pack). The pump will continue to operate
on backup power if the external power source fails. Refer to the
caution regarding Power Loss on page 13.
G
Note: The green LED on the pump remains lit when connected to AC mains
power.
Use the GemStar® AC Mains Adaptor (List 13026-01 [USA], 13026-03 [Universal]
or 13072-01 [Universal]) to recharge the GemStar® Battery Pack.
WARNING: DO NOT touch connector ports. Refer to caution on
page 15.
CAUTION: ALWAYS connect to a grounded AC mains outlet when using AC
mains adaptor.
Use only AC mains adaptors specifically labeled for use with the
GemStar® Pump to charge the battery pack.
During charging, the battery pack is warm. If the battery pack
becomes hot to the touch, IMMEDIATELY unplug AC mains
power and contact Hospira Technical Support Operations.
Ensure two fresh, disposable AA batteries are installed for backup
power when using other power sources (e.g., AC Mains adaptor,
docking station, battery pack). The pump will continue to operate
on backup power if the external power source fails. Refer to the
caution regarding Power Loss on page 13.
When removing the cord from the bottom of the battery pack,
firmly grasp the connector and pull straight out. DO NOT twist or
bend the cord or connector.
Regarding the GemStar® Li-Ion Battery Pack:
The LED indicator on the GemStar® Li-Ion Battery Pack flashes
green/yellow when an error condition has occurred or when the
temperature is too high to charge the battery pack. In temperatures
above 95°F (35°C), cycle AC power to confirm if battery is charged
or if error condition exists. If LED flashes green/yellow, do not rely
on external battery. Move device to cooler temperature and
recharge prior to use.
Refer to the GemStar® Li-Ion Battery Pack insert for important
information and precautions regarding activating and charging the
Li-Ion Battery Pack.
G
Note: A charged battery pack will last approximately 24 hours when the pump is
infusing at rates less than 125 mL/hr.
1Plug the pin connector of the AC mains adaptor into the port on the bottom of the
battery pack.
2Plug the AC mains adaptor into a standard wall outlet.
3Leave the battery pack connected to the AC mains adaptor for at least 12 hours to
charge the battery pack.
Note: The Li-Ion battery pack (List 13073-03) light is yellow while the battery pack is
charging and green when the battery pack is charged or not charging.
Li-Ion batteries cannot be charged when the room temperature is greater than 95°F
(35°C) or less than 41°F (5°C)
CAUTION: When removing the cord from the battery pack, firmly grasp the
connector and pull straight out. DO NOT twist or bend the cord or
connector.
WARNING: DO NOT touch connector ports. Refer to caution on page
page 15.
CAUTION: Ensure two fresh, disposable AA batteries are installed for backup
power when using other power sources (e.g., AC Mains adaptor,
docking station, battery pack). The pump will continue to operate
on backup power if the external power source fails. Refer to the
caution regarding Power Loss on page 13.
To Connect to the Pump
1Mount the docking station to the I.V. pole and tighten
the knob.
2Line up the connectors on the bottom of the pump with
the connectors on the docking station.
3Slide the pump into the docking station until it snaps
into place.
4Connect the docking station to AC mains power.
To remove the pump:
1Press the release button on the top of the docking
station.
2Slide the pump out.
To charge docking station (List 13075 all codes):
1Connect the docking station to the AC mains power.
2Leave the Li-Ion Docking Station connected to AC mains power for at least 6
hours when not in use or to recharge the docking station.
.
Note: The Li-Ion docking station (List 13075-03) light is yellow while the docking
G
station is charging. When the docking station is charged or not charging, the light is
green.
The Li-Ion docking station cannot be charged when the room temperature is greater
than 104°F (40°C) or less than 41°F (5°C).
CAUTION: If the pump does not display the self-test or if beeps do not sound at
power-on, check all power connections and install two fresh,
disposable AA batteries. If the problem continues, contact Hospira
Technical Support Operations before using the pump.
G
G
Note: If a cassette is not properly loaded, the pump will sound an alarm after the
power-on test is complete.
1Press [
ON/OFF] to power-on the pump. The pump
completes a self-test and displays the time and date. No
response is required.
GEMSTAR 7
UNIT SELF-TEST
TIME IS 10:00 AM
FRI, JUN 11, 04
Note: The self-test takes approximately 30 seconds to complete. If the pump is
locked, lock symbols appear at the beginning and end of the second line:
UNIT SELF-TEST .
2The pump checks for available power sources. Respond to the display message as
shown in the following table:
Display MessagePower Source(s)User Response
No message displaysAC power and
disposable batteries.
LOW BATTERIES
ENTER TO CONFIRM
AC power and weak
(low voltage)
disposable batteries;
or weak disposable
batteries only.
No response required, the pump goes
to the next step.
Press to confirm.
Power-off the pump and install two
fresh, disposable AA batteries as a
backup power source.
No disposable
batteries or dead
disposable batteries;
and external power.
Press to confirm.
Power-off the pump and install two
fresh, disposable AA batteries as a
backup power source.
Page 33
2. Basic Operation
Display MessagePower Source(s)User Response
G
USING BATTERIES
ENTER TO CONFIRM
USING EXT BATT
ENTER TO CONFIRM
Disposable batteries
only.
External batteries
(docking station
battery or battery
pack) and disposable
batteries.
Press to confirm.
If an external power source is in use,
ensure the connections are secure.
Press to confirm.
If an AC mains adaptor is in use,
ensure the connections are secure.
Note: To power-off the pump, press and hold until the display clears. The pump
cannot be powered off during a CHECK CASSETTE alarm (except during
PROGRAMMING mode). Refer to Chapter 6 Troubleshooting for information on
clearing a CHECK CASSETTE alarm.
CAUTION: Ensure two fresh, disposable AA batteries are installed for backup
power when using other power sources (e.g., AC Mains adaptor,
docking station, battery pack). The pump will continue to operate
on backup power if the external power source fails. Refer to the
caution regarding Power Loss on page 13.
For more information on:Refer to the following pages:
Setting the clockSet Clock in Chapter 4 Options Menu, on page 66.
Locking the keypadKeypad Lock in Chapter 4 Options Menu, on page 62.
The pump’s menu system provides step-by-step guidance through all of the
programming steps. Refer to Chapter 3 Pain Management for detailed programming
instructions.
A programming worksheet is available at the end of Chapter 3 that lists the minimum
and maximum value (program range) allowed for a Pain Management therapy. If a
number outside of the program range is entered, the pump displays the minimum or
maximum value allowed.
1Select the desired delivery mode from the Delivery Mode Menu.
2Respond to display messages and enter requested values.
Note: Program entries can be changed any time before the Program Review is
G
complete. Press [BACK-UP] to return to previous program displays and make
changes as desired.
3Review the program, if required. Press to begin review. Press [
when done.
4
SAVING PROGRAM displays. The pump stores the program in memory and then
STOP mode.
enters
Tip: If you enter an incorrect number during programming, press [CHANGE]
When entering a numeric value during programming, press to reset the
value to zero.
Press to return to a previous display, to exit the Options Menu, or to exit
the
HELP mode.
CHANGE
Page 37
2. Basic Operation
Starting a New Container
G
Note: The NEW CONTAINER function is not available when the pump is in FULL Lock.
Refer to Chapter 4 Options Menu, page 55 for more information on locking the keypad.
To repeat the current program with a new container:
1From the
2Select
CONTAINER
Note: In Pain Management therapy, the VTBI can be titrated without adding
volume to initial container amount.
3Review the program. Press to begin review.
4Press [
STOP mode, press [CHANGE] to display the Change Menu.
NEW CONTAINER from the Change Menu. The pump displays NEW
for several seconds and clears the amount infused from memory.
START] to begin delivery.
GemStar® Pump Set
The GemStar® Pump Set is a sterile, single-use, disposable pump set. It is designed
for one-time use only and should be properly discarded after use.
Fluid path and areas beneath undisturbed protective set
covers are sterile and nonpyrogenic in the intact unit
package.
Contact a Hospira representative for more information on selecting the appropriate
pump set configuration. Additional disposable components, such as air-eliminating
filters and extension sets, may be added to the line as required. To use a GemStar®
Pump Set, follow the instructions included with the set.
STERILE
2
Note: GemStar® Pump Sets contain a pressure-activated anti-siphon valve to reduce
G
the potential for gravity flow at a typical 36-inch head height.
Ensure slide clamp or lower CAIR® clamp is open prior to inserting cassette.
The system may underdeliver, and the system delivery accuracy may be reduced by as
much as 5% when using the following:
•nonvented syringes
•PCA vials (except in a lockbox with a spring assist mechanism)
WARNING: ALWAYS prime the pump set to remove air from the
cassette, tubing, and injection sites prior to connecting to
the patient. ALWAYS disconnect the pump set from the
patient prior to priming or purging. Arrange tubing, cords,
and cables to minimize the risk of patient strangulation or
entanglement.
Failure to use the pressure-activated anti-siphon valve
may result in unrestricted flow if the cassette is not
properly installed in the pump.
CAUTION: USE ONLY GemStar® Pump Sets with the GemStar® Pump. Use
of unauthorized sets may result in injury to the patient or damage
to the pump.
To prevent contamination, use aseptic technique with all fluid-path
connections. Remove protective coverings as assembly progresses.
CAUTION: Ensure the cassette
is properly installed. When
properly installed, all four
cassette latches are visibly
holding the cassette securely in
the cassette pocket.
6Place the tubing in the
tubing channel.
Note: The pump cannot be powered
G
off during a CHECK CASSETTE
alarm (except during
PROGRAMMING mode). Refer to
Chapter 6 Troubleshooting for
information on clearing a CHECK
CASSETTE alarm.
The pump purges for up to 2
minutes each time you press
and hold [
Continue using the [
key until fluid fills the tubing.
Ensure all air is removed
from the container, cassette,
and tubing.
PURGE].
PURGE]
Page 46
2. Basic Operation
Priming Pump Sets Using Pre-filled Vials
WARNING: Failure to use the pressure-activated anti-siphon valve
may result in unrestricted flow if the cassette is not
properly installed in the pump.
A cracked vial might not show evidence of leakage until
delivery pressure is applied.
When using the GemStar® Spring Assist Mechanism
(SAM) Lockbox, USE ONLY GemStar® Pump sets
designated for use with the GemStar® SAM Lockbox.
Failure to use designated sets could result in unrestricted
flow.
CAUTION: When using PCA vials (except in a lockbox with a spring assist
mechanism), the system may underdeliver. System delivery
accuracy could be reduced by as much as 5%.
When using pre-filled vials, ensure vial and injector are securely
locked in lockbox before beginning delivery.
USE ONLY purge priming with the GemStar® Lockbox with
Spring Assist Mechanism (SAM).
This therapy allows programming of a continuous only, bolus only, or continuous with
bolus delivery.
In Continuous + Bolus protocols, you can enter zero for either the continuous rate
or the bolus dose when entering or changing a program. The continuous rate and
bolus dose cannot be zero at the same time.
The default bolus delivery rate is 125 mL/hr.
G
G
Note: The delivery rate is delivered in increments of .01.
The maximum continuous delivery rate is 25 mL/hr.
A loading dose may be set during programming, which can be delivered after
programming and priming are complete. The default loading dose rate is 125
mL/hr. Rate, bolus dose, concentration can be programmed in increments of 0.01.
Priming using [PURGE] is only allowed before [START] is pressed or after one of
the following events occur:
Complete Review of New Program
New Container
Air-in-Line Alarm
VTBI can be titrated.
CAUTION:When using PCA vials (except in a lockbox with a spring assist
mechanism), the system may underdeliver. System delivery
accuracy could be reduced by as much as 5%.
Note: New Container is allowed only when at least 1/2 of the container volume has
been infused or 1 hour has elapsed since starting the infusion.
The use of a lockbox is recommended with pain management protocols. Refer to
Chapter 9 Optional System Components, page 109, for instructions on using a
lockbox with the GemStar® Pump.
1Select the delivery mode from the Delivery Mode menu.
3. Pain Management
G
G
G
Note: If Continuous + Bolus is selected, the continuous delivery rate or bolus dose
can be set to zero through the CHANGE menu without reprogramming the pump.
2Select the unit of measure and concentration as required.
3Set the continuous delivery rate.
4Program a loading dose as required.
5Program a bolus dose with lockout time and dose limit as required.
Regarding a 1-Hour or 4-Hour Dose Limit: The hour dose limit is the
maximum volume (continuous plus bolus or bolus only amount) that can be
delivered over the selected period. When the programmed dose limit is reached,
the continuous delivery stops and a new bolus is not allowed. As the oldest
delivery amounts age out of the 1-hour or 4-hour dose limit record, the pump
accepts bolus requests or resumes continuous delivery (if programmed).
Note: If a 1-hour or 4-hour dose limit is programmed, a bolus in progress will stop
when the amount infused reaches the dose limit.
Loading doses are included in a 1-hour or 4-hour dose limit. A dose limit in effect
could prevent the delivery of a loading dose.
6If programmed in units of mcg or mg, the line displaying the units flashes. Enter
the container size in proper unit measure.
Note: To enter the total in mL when programming a concentration, press to
move to the mL line. Enter the mL value when
mL flashes on the display.
7Select the air sensitivity, if required.
8Review the program. Press to begin review.
9If using the PURGE key to prime the pump set, press [
WARNING: ALWAYS disconnect the pump set from the patient before
Set bolus doses and delivery limits during programming. The default bolus dose
delivery rate in a Pain Management therapy is 125 mL/hr.
A bolus cord is available for patient bolus requests. Refer to Connecting the Bolus Cord in Chapter 9 Optional System Components, on page 112, for more information.
Delivering a Bolus Dose
G
From the
button on the end of the bolus cord to begin bolus dose delivery. The pump flashes
BOLUS DELIVERY on the display and the amount infused accrues as the delivery
progresses.
When the bolus dose is complete, the pump does the following:
Records the bolus request and delivery amount to the event and bolus history logs
Adds the delivered boluses to the total bolus amount in the program amounts
Resets the bolus lockout time (if applicable) and ignores other bolus requests until
Continues programmed infusion
Note: ALL bolus requests are recorded to the event and bolus history logs.
If the bolus delivery does not start, it could be locked out by one of the following
conditions:
Bolus lockout period
Loading dose delivery in progress
Bolus dose delivery in progress
Bolus per-hour limit
1-hour or 4-hour dose limit
RUN mode, press either the bolus button [+] on the top of the pump or the
START]. The pump displays COMPLETE THE LOADING DOSE?
Page 58
3Respond as follows:
To do this:Press this key:
Deliver the remaining loading dose amount
Clear the remaining loading dose amount
Changing a Pain Management Program
Can be changed at any timeCannot be changed
3. Pain Management
G
• Delivery rate
• Bolus
• Bolus lockout
• Container size
• Air sensitivity
• Dose limit amount
• Delivery mode
• Unit of measure
• Concentration
• Dose limit type
Note: You cannot change a program when the pump is in FULL or CONTAINER
Lock. With a
RATE CHANGE Lock, the continuous delivery rate and bolus amount
can be changed within the set ranges.
Although the delivery mode cannot be changed, if Continuous + Bolus is
programmed, the continuous rate or bolus amount can be changed to zero. The
minimum values set with a
RATE CHANGE Lock could restrict this function.
To change a program:
1From the
2As each message displays, make the desired changes or press [
STOP mode, press [CHANGE] and select CHANGE PROGRAM.
Display the history for the latest
400 events, shift or volume totals,
bolus history (if available), or
current configuration settings.
Clear the history. Display the
Operation Test.
N/A
CAUTION: Ensure two
fresh, disposable AA
batteries are installed for
backup power when using
other power sources (e.g.,
AC Mains adaptor, docking
station, battery pack). The
pump will continue to
operate on backup power if
the external power source
fails. Refer to the caution
regarding Power Loss on
page 13.
Lock and unlock the keypad to
prevent unauthorized changes to
the program.
Change the air sensor, occlusion
sensor, and sound level settings.
Set the clock.The clock can be displayed in
Print the history event log, bolus
history (if available), current
program settings, and speed
protocols.
Assign, review/recall, and delete
the stored protocols.
The lock levels allow the clinician
to restrict access to various levels
of pump operation.
N/A
12-hour or 24-hour format.
Files may be downloaded to a
printer or personal computer.
Allows up to 9 speed protocols to
be saved in pump memory.
Display history event
log of the latest 400
events.
• Current date and time
• Date and time the program was cleared, a new
container was selected, or a speed protocol was
recalled
• History event log of the latest 400 events
CAUTION: Ensure two fresh, disposable
AA batteries are installed for backup
power when using other power sources
(e.g., AC Mains adaptor, docking station,
battery pack). The pump will continue to
operate on backup power if the external
power source fails. Refer to the caution
regarding Power Loss on page 13.
Display or clear shift
totals.
Display volume totals.•Amount infused
Display pump
configuration.
Clear history event log
and bolus history.
Display bolus history
in hourly increments
for the last 48 hours.
Perform user-assisted
Operation Test.
• Time and date when shift was cleared
• Volume infused in the shift
• Any relevant program-specific information
•VTBI
• Any relevant program-specific information
• Dependent upon pump configuration
CLEARING HISTORY displays for 3 to 4
•
seconds
• Current date and time
• Number of boluses delivered
• Number of bolus demands
• Amount delivered through bolus delivery
• Messages that guide the user through each
section of the test
G
I
Note: Some menu items may not be available depending on the configuration of the
pump.
Tip: Press and HOLD or to scroll through the history displays faster.
Note: Bolus History is only available if the current
program has a bolus programmed, or had a bolus
programmed at one time. The Bolus History is
maintained in hourly increments for the latest 48 hours.
The Bolus History is cleared when a new program is
entered, a Speed Protocol is recalled, or the History is
cleared.
Operation Test
The Operation Test checks several pump functions to ensure the GemStar® Pump is
working properly. Refer to Chapter 8 Operation Test on page 99 for instructions.
2Press the number key for the desired pump setting.
3Follow the display messages. Refer to the
instructions of the desired pump setting for more
information.
Select this item from
the Pump Settings
Menu:
AIR SENSOR
OCCL SENSOR
SOUND LEVEL
To do this
function:
Change the air
sensor sensitivity.
Change the distal
occlusion sensor
sensitivity.
Adjust the audible
alarm level.
Comments:
The pump alarms when it detects an air
bubble of approximately 0.5 mL when ON;
or at approximately 2 mL when 2 mL is
selected. The air alarm does not sound if
OFF is selected.
Note: The pump setting is inactivated if
custom-configured (by your facility) to ON
or 2mL.
WARNING: To reduce the risk of infusing air, use an air-eliminating
filter when the air sensitivity is set to OFF.
G
1Press [
OPTIONS] and select AIR SENSITIVITY. The current air sensitivity
flashes on the display.
2Select the desired Air Sensitivity level.
Note: If OFF is selected, the pump displays:
ELIMINATING FILTER IS
Air
RECOMMENDED. ENTER TO CONFIRM
You must press [YES/ENTER] to confirm.
3After selecting the desired Air Sensitivity, the
pump displays:
XXXX TO XXXX
•Press [
CHANGE AIR ALARM FROM
:
NO] to keep the current air sensitivity.
The current air alarm status displays and the
pump exits
•Press [
OPTIONS.
YES/ENTER] to accept the new setting. The new air alarm status displays
and the pump exits OPTIONS.
Occlusion Sensor
1Press [
2Select
OPTIONS] and select PUMP SETTINGS.
DISTAL OCCLUSION from
the Pump Settings Menu.
3Select the desired setting.
4The pump displays
MAKE XXXXX DEFAULT?
•Press [NO] to temporarily change the distal
occlusion setting. The new setting will be in effect
until a new program is entered or the pump has been off for more than 24
hours.
•Press [
YES/ENTER] to permanently change the distal occlusion setting. The
new setting will remain in effect until the setting is changed again through the
Options Menu.
instructions for Speed Protocol functions for more
information.
Note: Up to nine speed protocols are stored at one
time. The speed protocols stay in memory, even when
the pump is powered down, until they are deleted.
Select this item from
the Speed Protocol Menu:
ASSIGN
REVIEW/RECALL
DELETE
Assign
1Press [
2Select
OPTIONS] and select SPEED PROTOCOL.
ASSIGN from the Speed Protocol menu.
To do this function:
Store a program in pump memory for later use.
Review assigned protocols from the RUN or STOP
mode. Recall an assigned protocol from the STOP mode.
Delete protocols from pump memory.
1REVIEW PROGRAM
2HISTORIES
3KEYPAD LOCK
4PUMP SETTINGS
5SET CLOCK
6PRINT
7SPEED PROTOCOL
3Select [
If the selected speed protocol is already assigned, the pump displays
PROTOCOL ASSIGNMENT?
1]–[9] to assign to the new protocol.
REPLACE
•Press [YES/ENTER] to replace the protocol. The pump displays the new speed
protocol and exits
Use the data port to connect the GemStar® Pump to a Seiko® DPU 414 or compatible
serial printer to print the Operation Test results or the files available from the Print
Menu. Using communication software included with most computer operating
systems (or available from local sources), these files can also be transferred to a
computer.
WARNING: Ensure a cassette is installed in the pump before
connecting to a printer or computer.
If the pump is connected to a patient: BEFORE
connecting the pump to a computer or printer, press
STOP], close the slide clamp, and eject the cassette.
[
Leave the pump set connected to the patient. Install
another cassette in the pump before connecting to a
printer or computer.
Refer to the printer or computer documentation for information on configuration
settings. Configure the printer or computer as follows to use the data port:
G
Baud Rate
Data Bits
Stop Bits
Parity
Carriage Return (+Line Feed)
Flow Control =
Note: When using the Seiko DPU 414 printer, the switch settings MUST be configured
to work with the GemStar® Pump.
To configure the switch settings on the Seiko DPU 414:
1Press and hold the ONLINE button while sliding the POWER switch to ON.
2When the list of the current settings start printing, release the ONLINE button.
3Set the switches according to the following table.
Press the ONLINE button to continue setting switches.
•Press ONLINE to set switch ON.
•Press FEED to set switch OFF.
4Continue setting switches until “Dip SW setting complete!!” is printed.
WARNING: Ensure a cassette is installed in the pump before
connecting to a printer or computer.
If the pump is connected to a patient: BEFORE
connecting the pump to a computer or printer, press
[
STOP], close the slide clamp, and eject the cassette.
Leave the pump set connected to the patient. Install
another cassette in the pump before connecting to a
printer or computer.
Refer to the printer documentation or contact the printer
manufacturer for more information regarding
Electrostatic Discharge (ESD) susceptibility of printer
connector ports.
CAUTION: Do not connect the GemStar® AC Mains Adaptor to a printer.
G
Note: For maximum battery life, operate the pump on AC mains when connected to a
printer or computer.
1Install a cassette into the pump.
2Connect the GemStar® serial cable to the printer and to the data port on the
bottom of the pump.
3Ensure the printer is ONLINE and ready to receive data. Refer to printer
documentation for more information.
4Ensure paper is loaded in the printer.
5Select the desired print function from the pump menu.
WARNING: Ensure a cassette is installed in the pump before
connecting to a printer or computer.
If the pump is connected to a patient: BEFORE
connecting the pump to a computer or printer, press
STOP], close the slide clamp, and eject the cassette.
[
Leave the pump set connected to the patient. Install
another cassette in the pump before connecting to a
computer or printer.
Refer to the computer documentation or contact the
computer manufacturer for more information regarding
Electrostatic Discharge (ESD) susceptibility of computer
connector ports.
G
I
Note: For maximum battery life, operate the pump on AC mains when connected to a
printer or computer.
1Ensure slide clamp or lower CAIR® clamp is open prior to inserting cassette.
2Install a cassette into the pump.
3Connect the GemStar® serial cable to the null modem.
4Connect the null modem to the serial gender changer.
5Connect the serial gender changer to the serial port of the computer.
6Connect the GemStar® serial cable to the data port on the bottom of the pump.
7Configure the communication
software on the computer. Refer
to Configuration Settings on
page 71.
Tip: After the data transfer is complete, use any text editor to view or print the file.
the existing file. For example, use a patient’s name or id number as the file name.
Then select that file for all of that patient’s printouts.
9Click Start.
10 Select the desired print function from the pump. Refer to the table of printing
functions on page 67.
11 Wait for the transmission to complete.
12 From the menu, select Trans fer > Capture Text > Stop.
13 Exit HyperTerminal.
14 Use any text editor to view or print the file.
Page 84
Chapter 5
Clinician Instructions
The instructions in this section apply to special features available to clinicians.
Remove this section from the manual to prevent unauthorized access to these
features.
The GemStar® Pump provides quick access to commonly used functions. Instead of
accessing these functions by navigating through the menu system, you can enter the
key combinations shown in this table:
FunctionQuick Access Key Combination
Unlock KeypadPlace the pump in STOP mode and then enter the lock
sequence to unlock a Full Container or Rate Lock.
Displays Shift TotalsPlace the pump in STOP mode and then press [HELP] to
display Shift Totals.
Display ClockPlace the pump in STOP mode, press [YES], and then press
and hold [1]. The clock displays until you release [1].
This section contains information on audible and visual alarms that may occur with
the GemStar® Pump.
CAUTION: If the pump does not perform as stated in this manual, stop using it
IMMEDIATELY.
For consultation and technical support contact either an authorized Hospira
representative or the Hospira Technical Support Operations hotline (available 24
hours a day).
To return the pump for service, first contact Hospira Customer Support between 7:00
AM and 4:00 PM Pacific time. Have the pump serial number available.
In the USA
Outside the USA:
I
Tip: Press [HELP] for additional information during programming and alarm conditions.
entered the
mode due to battery
voltage below
acceptable levels.
STOP
Remedy
1Replace both disposable
batteries with fresh AA
batteries and press to
restart the pump; or connect
the pump to the AC mains
adaptor and the pump will
automatically restart.
Replace both batteries as
soon as possible.
CAUTION: Ensure two
fresh, disposable AA
batteries are installed
for backup power when
using other power
sources (e.g., AC Mains
adaptor, docking
station, battery pack).
The pump will continue
to operate on backup
power if the external
power source fails.
Refer to the caution
regarding Power Loss
on page 13.
detected battery
voltage below
acceptable levels and
the batteries are the
primary power
source.
Note: The pump
allows the user to
clear the
LOW BAT-
TERIES
the condition has not
been corrected after
15 minutes, the alarm
continues.
If the pump is infusing
and a
LOW BAT-
TERIES alarm
occurs, the pump continues for approximately 30 minutes*
before it can no longer
deliver due to insufficient power.
*If an external power
source is used, the
rate is changed, or the
pump is powered-on
after a
LOW BAT-
TERIES
those same batteries
are still in the pump,
this interval may be
different.
alarm. If
alarm and
Remedy
1Press [SILENCE] to quiet
the alarm for 15 minutes.
2Press [STOP] to place the
pump in the
3Connect the pump to AC
mains adaptor, rechargeable
battery pack, or press the
On/Off key to turn off the
pump and replace both
disposable batteries with
fresh AA batteries.
STOP mode.
CAUTION: In the event
that the internal batteries
are removed while the
pump is operating, the
power loss event may not
be logged in the history,
the programmed therapy
may not be retained, and
the alarms (audible and
visual) may not occur.
Refer to the caution
regarding Power Loss on
page 13.
4Power on the pump and
resume the current program.
Refer to Accessing the Program Menu in Section 2
for more information
page 13.
5Press [START] to begin
infusion.
Note: In case of low battery
alarm during printing, replace
the batteries and reprint
or
Plug in AC main and reprint.
detected a power
failure due to one of
the following:
• Batteries have
been removed
while the pump
was on battery
source.
• External power
has failed or has
been
disconnected and
no batteries are
installed.
• Pump has been
powered on with
insufficient power
source.
Remedy
1Connect the pump to AC
mains adaptor, rechargeable
battery pack, or replace
both disposable batteries
with fresh AA batteries.
2Power-on the pump and
resume the current program.
Refer to Accessing the Program Menu in Section 2
for more information. Review
the program if required.
3Press [START] to begin
infusion.
CAUTION: Ensure two
fresh, disposable AA
batteries are installed
for backup power when
using other power
sources (e.g., AC Mains
adaptor, docking station,
battery pack). The pump
will continue to operate
on backup power if the
external power source
fails. Refer to the
caution regarding Power
Loss on page 13.
• The pump has
been powered on
with only
disposable
batteries as a
power source; or
• The pump has
switched to
internal power
(disposable
batteries) due to a
loss of external
power or a
reduction in
external power
below acceptable
levels.
Remedy
When powering on the pump:
• Press [YES/ENTER] to
confirm disposable batteries
are the only power source.
When switching from AC Mains
power:
• Press [SILENCE] to clear the
alarm.
CAUTION: Ensure two
fresh, disposable AA
batteries are installed for
backup power when
using other power
sources (e.g., AC Mains
adaptor, docking station,
battery pack). The pump
will continue to operate
on backup power if the
external power source
fails. Refer to the caution
regarding Power Loss on
page 13.
• The pump has
been powered on
with external
batteries as a
power source; or
• The pump has
switched to
external battery
power due to a
loss of AC mains
power; or
• The pump has
switched to
external battery
power while
running on
internal batteries.
Remedy
When powering on the pump:
• Press [YES/ENTER] to
confirm external battery
power source.
When switching from AC Mains
power:
• Press [SILENCE] to clear the
alarm.
CAUTION: Ensure two
fresh, disposable AA
batteries are installed for
backup power when
using other power
sources (e.g., AC Mains
adaptor, docking station,
battery pack). The pump
will continue to operate
on backup power if the
external power source
fails. Refer to the caution
regarding Power Loss on
page 13
the programmed
container size or the
pump has delivered the
programmed VTBI; AND
Auto KVO is delivering
(if available in current
program).
NA/VA Non-programming
Callback has occurred
because the pump has
been waiting for more
than three minutes for a
keypress on a nonprogramming input or a
confirmation display.
1Press [SILENCE] to
quiet the alarm for 2
minutes.
2Press [STOP] to place
the pump in the
mode and clear the
alarm.
3Use the NEW
CONTAINER function to
repeat the current
program. Refer to
Starting a New
Container in Chapter 2
Basic Operation.
STOP
Note: Titration is allowed at
this time.
Follow the instructions on the
display to clear the alarm.
install two fresh,
disposable AA batteries
as a backup power
source.
CAUTION:Ensure
two fresh, disposable
AA batteries are
installed for backup
power when using
other power sources
(e.g., AC Mains
adaptor, docking
station, battery pack).
The pump will
continue to operate
on backup power if
the external power
source fails. Refer to
the caution regarding
Power Loss on
page 13.
PROGRAMMING
INCOMPLETE
START
• START flashes on
the display. Other
parts of the display
are not affected.