Hospira GemStar 13086 User manual [gb]

Page 1
Front Cover.fm Page i Thursday, February 3, 2005 4:58 PM
430-00897-A02, C (Rev. 04/2007)
Page 2
430-00897-002.book Page i Friday, October 22, 2004 2:38 PM
Hospira GemStar® Infusion Pump
System Operating Manual
Hospira GemStar Pump
List 13000-04-07/08 List 13100-04-07/08
PLEASE
Read this entire manual before
using the Hospira GemStar Pump
This manual is designed for use by healthcare professionals, caregivers, and patients. The Hospira Technical Support Operations hotline is available 24 hours a day (in the USA) to provide consultation and technical assistance regarding the Hospira GemStar Pump.
In the USA: Hospira Technical Support Operations
1-800-241-4002
Outside the USA: Contact your local Hospira
sales office
To order additional copies of this manual
(List 13981) call Hospira Customer Care:
1-877-9HOSPIRA
(1-877-946-7747)
Hospira GemStar Manual 430-00897-002 (Rev. 10/04) i
Page 3
Title.fm Page ii Wednesday, March 9, 2005 3:30 PM
Change History
Part Number Description of Change Pages Changed
430-00897-001 (Rev. 9/02)
430-00897-A01 (Rev. 12/03)
430-00897-002 (Rev. 10/04)
430-00897-A02 (Rev. 02/05)
Original Issue N/A
Updated Cover Updated Change History Updated Battery Pack Information
Second Release Inclusive
Updated Cover Updated Change History Aligned Note and Tip icons
Corrected typographical error in note
Reformatted dashed line with scissors icon
Front and Back ii 19
Front and Back ii 21, 61, 69, 88, 97, 100–102,
106–107, 111, 122, 125, 144, 149, 178, 190
178
121–126
Added UL/CSA information Updated CSA information
171 Back
Note: Change page manuals assembled by Hospira include a change page identifier
in the part number on the cover page and on the bottom of each changed page. This change page manual is identified as 430-00897-A02.
ii Hospira GemStar Manual 430-00897-A02 (Rev. 02/05)
Page 4
Contents.fm Page iii Friday, November 12, 2004 8:28 AM
1. Introduction 1
GemStar Pump Layout . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
GemStar Pump Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Therapies. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
GemStar Pump Features. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Menu System. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Operating Modes. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Indications for Use. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Contraindications for Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Warnings and Cautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
2. Basic Operation 17
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Selecting a Power Source . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Connecting to the GemStar Docking Station. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Powering On the Pump . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Accessing the Programming Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Programming the Pump. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Starting a New Container . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
GemStar Pump Set . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Loading the Cassette. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Releasing the Cassette . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Priming a GemStar Pump Set. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Contents
3. Total Parenteral Nutrition (TPN) 47
Therapy Features and Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
TPN Programming. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
Changing a TPN Program. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
TPN Auto Taper Down . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
TPN Programming Worksheet. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
Hospira GemStar Manual 430-00897-002 (Rev. 10/04) iii
Page 5
Contents.fm Page iv Thursday, November 11, 2004 7:43 AM
Contents
4. Pain Management 53
Therapy Features and Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
Pain Management Programming. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
Bolus Delivery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
Loading Dose Delivery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
Changing a Pain Management Program. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
Pain Management Programming Worksheet. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
5. Intermittent 61
Therapy Features and Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
Intermittent Programming . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
Intermittent Delivery Interruptions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
Changing an Intermittent Program . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
Intermittent Programming Worksheet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
6. Continuous 69
Therapy Features and Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
Continuous Programming . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70
Rate Titration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
Piggybacking. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
Changing a Continuous Program . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
Continuous Programming Worksheet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
7. Weight Dosed 77
Therapy Features and Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77
Weight Dosed Programming . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78
Auto KVO. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79
Rate Titration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79
VTBI Titration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80
Changing a Weight Dosed Program . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80
Weight Dosed Programming Worksheet. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81
iv Hospira GemStar Manual 430-00897-002 (Rev. 10/04)
Page 6
Contents.fm Page v Thursday, November 11, 2004 7:43 AM
8. mL/hr Only 83
Therapy Features and Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83
mL/hr Only Programming . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 84
Auto KVO. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85
Rate Titration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85
VTBI Titration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 86
Piggybacking. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 86
Changing a mL/hr Only Program. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 89
mL/hr Only Programming Worksheet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90
9. Variable Time 91
Therapy Features and Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 91
Variable Time Programming . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 92
Variable Time Delivery Interruptions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 93
Changing a Variable Time Program . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 94
Variable Time Programming Worksheet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 95
Contents
10. Options Menu 97
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 97
Tips for Using the Options Menu. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 97
Review Program . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 99
Histories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 99
Keypad Lock . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 104
Change Pump Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 106
Set Clock . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 108
Print. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 109
Speed Protocol . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 110
Next Dose . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 111
Connecting to the Data Port . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 112
Transferring Data Using a Computer. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 117
11. Clinician Instructions 121
Clinician Locking Sequence Instructions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 122
Program and Deliver a Clinician Activated Loading Dose. . . . . . . . . . . . . . . . . . 124
Shift Totals and History . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 125
Hospira GemStar Manual 430-00897-002 (Rev. 10/04) v
Page 7
Contents.fm Page vi Wednesday, November 10, 2004 12:05 PM
Contents
12. Troubleshooting 127
Alarm Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 128
Alert Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 135
Other Display Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 138
13. Maintenance 139
Pump Storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 139
Cleaning and Disinfecting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 139
Repair . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 141
14. Operation Test 143
Equipment Required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 143
Test Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 144
Performing the Operation Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 144
Printing Test Results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 149
Operation Test Checklist . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 151
15. Optional System Components 153
Connecting the Bolus Cord . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 156
Using the GemStar Lockboxes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 156
Using the Hospira Carrying Cases . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 159
Using the GemStar Docking Station . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 165
16. Pump Sets 167
17. GemStar Pump Specifications 171
Occlusion Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 174
18. Warranty/Technical Assistance 189
Index 191
IEC Symbols inside back cover
vi Hospira GemStar Manual 430-00897-002 (Rev. 10/04)
Page 8
430-00897-002.book Page 1 Friday, October 22, 2004 2:38 PM
The GemStar® Pump is a small and lightweight, single-channel infusion device designed for use in the home, in the hospital, or anywhere electronic infusion is required. The GemStar Pump can be powered by AC mains adaptor, rechargeable battery pack, docking station, or two disposable AA alkaline or lithium batteries. When powered by batteries, the GemStar Pump is ideal for ambulatory patients.
The following conventions are used throughout this manual:
Convention Use Example
Chapter 1
Introduction
  
[ALL CAPS] Pump keys Press [YES/ENTER] to continue.
ALL CAPS
WARNING: A WARNING message contains special safety
CAUTION: A CAUTION contains information that could prevent
Note: A NOTE contains information or examples to help explain instructions.
Tip: A TIP contains information that makes the pump easier to use.
Note: Examples are provided only to help explain a pump function and may not be clinically relevant. Graphics are for illustrative purposes only and may not reflect the actual product.
Display messages The pump displays:
ENTER TO CONFIRM.
information and must be observed at all times. Failure to observe a warning message could be life threatening.
irreversible product damage or hardware failure. Failure to observe a caution could result in serious patient or user injury.
Hospira GemStar Manual 430-00897-002 (Rev. 10/04) 1
Page 9
Introduction.fm Page 2 Thursday, November 11, 2004 3:04 PM
1. Introduction
GemStar Pump Layout
Shown approximate size
2 Hospira GemStar Manual 430-00897-002 (Rev. 10/04)
Page 10
Introduction.fm Page 3 Thursday, November 11, 2004 3:04 PM
DISPLAY
1
2
Displays the pump’s status.
Silences an alarm.
3 Stops infusion.
4 Starts infusion.
1. Introduction
5
Lights continuously when the pump is powered by AC mains.
Flashes when the pump is powered by external batteries (e.g., battery pack).
Power Indicator
Not lit when using disposable AA batteries.
6 Powers the pump on or off. To power-off the pump, press and hold until the
display clears.
7 Returns to previous steps during programming.
Exits Help, Change, and Options menus.
8
Displays the following menu items from the
STOP or RUN mode*:
New Container Change Program (or Titrate VTBI) New Program Piggybacking
Resets a numeric value to zero.
9
Displays the following Options Menu items from the
STOP or RUN mode*:
1 Review Program 5 Set Clock 2 Histories 6 Print 3 Keypad Lock 7 Speed Protocol 4 Pump Settings 8 Next Dose
10 Scrolls through display messages.
Toggles between AM and PM in time-based entries.
UP arrow enters a decimal point in numeric entries.
DOWN arrow toggles between units of measure when entering concentrations.
11 Responds “YES” to display questions.
Accepts entries and goes to the next step.
12 Responds “NO” to display questions.
13 Displays information during programming steps and alarm conditions.
Displays Shift Totals, except during programming steps and alarm conditions.
14
Clears an
AIR-IN-LINE alarm.
Primes a Gemstar Pump Set.
15 [0] - [9] • Selects menu items and enters numeric values. If two numbers appear before a
menu item (e.g., 02), press both number keys to select that item.
* Not all menu items are available with all modes.
Hospira GemStar Manual 430-00897-002 (Rev. 10/04) 3
Page 11
Introduction.fm Page 4 Thursday, November 11, 2004 3:04 PM
1. Introduction
GemStar Pump Components
Bolus Button Press this button to start a
1
bolus delivery if available in the current program.
Pole Clamp Track Use to connect to an
2
optional pole clamp.
Cassette Release
3
Button
Power Connector Use to connect the AC
4
Data Port
5
Connector
Bolus Connector Use to connect a bolus cord.
6
Tubing Channel Place the pump set tubing in
7
Cassette Pocket Load the cassette into this
8
I.V. Tubing Guide Use to hold the I.V. tubing.
9
Battery Pack
10
Connector
Battery Compartment Install two fresh, disposable
11
Push down on this button to release the cassette.
mains adaptor.
Use to connect the GemStar serial cable.
this channel.
pocket.
Use to connect the GemStar Rechargeable Battery Pack.
AA batteries in this compartment.
Top View
Bottom/Side View
4 Hospira GemStar Manual 430-00897-002 (Rev. 10/04)
Page 12
L
Introduction.fm Page 5 Thursday, November 11, 2004 3:04 PM
Therapies
Seven types of therapy are available for infusion of medications and fluids. The following table lists possible applications and available features of each therapy.
1. Introduction
Therapy Possible
Applications
TPN Total Parenteral
Nutrition fluid delivery
Pain Management
Infusion of analgesics or anesthetics for patient­controlled pain management (PCA)
Intermittent Interval-based
therapies such as antibiotics
Continuous Infusion of mg, mcg,
or mL/hr with a selectable KVO (Keep Vein Open) option
Weight Dosed Critical care,
pediatrics
mL/hr Only Simple rate
programming
Variable Time Chronotherapy
= Available
O
E
V
K
A
R
O
T
T
I
U
T
R
A
G
T
-
N
S
Y
I
U
G
E
K
G
T
C
I
A
A
P
B
A
L
O
O
L
B
P
G
N
I
D
E
P
S
A
O
T
D
E
U
T
A
R
N
R
E
I
W
T
B
O
I
T
D
T
V
/
••
••
••
••
•••
••
A
N
O
I
T
O
P
V
O
K
Note: Not all therapies are available in all pump configurations.
Hospira GemStar Manual 430-00897-002 (Rev. 10/04) 5
Page 13
Introduction.fm Page 6 Thursday, November 11, 2004 3:04 PM
1. Introduction
GemStar Pump Features
Feature Description Reference Page
Adjustable Settings
Alerts and Alarms
•Audible Keypad Response
Keypad Lock
•New Container
Piggybacking
•Power-on Self-test
Operation Te st
Program History
Speed Protocols
Adjusts the distal occlusion pressure limits, air sensor sensitivity, real-time clock, and audible alarm volume.
Signals audible and visual alarms when attention is required.
Sounds a single beep when any key is pressed. If an invalid key is pressed, the pump sounds a flutter tone.
Restricts access to various options. Four lock levels are available.
Repeats the current program without reprogramming the pump.
Allows delivery from a secondary container during a Continuous or mL/hr Only program.
Performs a self-test each time the pump is powered on.
Performs user-assisted tests to verify the pump is operating properly.
Maintains a history event log, with a date and time stamp of each event for each program. Review the history on the display, or download it to a printer or computer.
Stores up to nine frequently used programs in the pump’s memory for quick access.
Chapter 10 Options Menu
Chapter 12 Troubleshooting
Chapter 1 Introduction,
GemStar Pump Layout
Chapter 10 Options Menu and Chapter 11 Clinician Instructions
Chapter 2 Basic Operation
Chapter 6 Continuous and Chapter 8 mL/hr Only
Chapter 2 Basic Operation
Chapter 14 Operation Te st
Chapter 10 Options Menu
Chapter 10 Options Menu
6 Hospira GemStar Manual 430-00897-002 (Rev. 10/04)
Page 14
Introduction.fm Page 7 Thursday, November 11, 2004 3:04 PM
Menu System
The GemStar Pump uses a menu system that provides step-by-step instructions through all of the pump’s functions. Refer to the following table for a list of functions in each menu.
Use the keypad to select menu items, respond to display messages, and enter numeric values. For example, press [ Menu.
2] to select PAIN MANAGEMENT from the Programming
1. Introduction
Note: Menu items may vary depending on the pump’s configuration.
Tip: A menu item does not have to be on the display to be selected. For example, when the Programming Menu first displays, only the first four therapies are shown. Instead of pressing the down arrow to see the other menu items, press [ directly to Variable Time Programming.
7] to go
Hospira GemStar Manual 430-00897-002 (Rev. 10/04) 7
Page 15
Introduction.fm Page 8 Thursday, November 11, 2004 3:04 PM
1. Introduction
To displ a y this menu:
Programming Menu
Do this: Menu Items Available
1 Press to display
the Change menu.
2 Select
NEW PROGRAM.
1 TPN 2 PAIN MANAGEMENT 3 INTERMITTENT 4 CONTINUOUS 5 WEIGHT DOSED
mL/HR ONLY
6 7 VARIABLE TIME
Note: If a program is not currently stored in pump memory at power-on, the Program­ming Menu automatically displays after the self-test.
Change Menu Press . 1 NEW CONTAINER
2 NEW PROGRAM
3 CHANGE PROGRAM
4 PIGGYBACKING
Note: If the current therapy is Weight Dosed or mL/hr Only and the pump is in mode:
3 TITRATE VTBI
will replace
3 CHANGE PROGRAM
on the menu display.
RUN
Note: Piggybacking is available only in mL/hr Only and Continuous therapies.
Options Menu Press . 1 REVIEW PROGRAM
8 Hospira GemStar Manual 430-00897-002 (Rev. 10/04)
2 HISTORIES
3 KEYPAD LOCK
4 PUMP SETTINGS
5 SET CLOCK
6 PRINT
7 SPEED PROTOCOL
8 NEXT DOSE
Page 16
Introduction.fm Page 9 Thursday, November 11, 2004 3:04 PM
For more information on: Refer to the following pages:
Programming Menu Chapter 2 Basic Operation, page 17.
Change Menu For menu item [1], refer to Starting a New Container
Options Menu Chapter 10 Options Menu, page 97.
Operating Modes
The GemStar Pump has four operating modes:
Mode Description
1. Introduction
in Chapter 2 Basic Operation, on page 29. For menu items [2], [3], and [4], refer to the appropriate therapy chapter.
Help Mode The pump enters the
programming or an alarm condition. The additional information about the current condition.
Programming Mode The pump enters the
powered on or when is pressed.
Run Mode After is pressed, the pump enters the
when not infusing.
Stop Mode When programming is complete, the pump enters the
mode until is pressed.
When the programmed therapy is complete, and during certain alarm conditions, the pump automatically enters the
Press to enter the CHECK CASSETTE alarm.
HELP mode when is pressed during
HELP mode displays
PROGRAMMING mode after the pump is
RUN mode, even
STOP mode.
STOP mode at any time, except during a
STOP
Hospira GemStar Manual 430-00897-002 (Rev. 10/04) 9
Page 17
Introduction.fm Page 10 Thursday, November 11, 2004 3:04 PM
1. Introduction
Indications for Use
Physicians or certified, licensed healthcare professionals should always oversee infusions. Healthcare professionals should always supervise pump users and instruct users in pump operation and troubleshooting. Instruction should include how to prevent intravenous (I.V.) related complications, including how to prevent accidental infusion of air.
The pump is suitable for intravenous, arterial, subcutaneous, short-term epidural infusion and parenteral administration of general I.V. fluids, medications, nutritional fluids, and blood/blood products. The epidural route is recommended to provide anesthesia or administer analgesia for periods up to 96 hours.
Note: Epidural administration of anesthetics is recommended in the continuous
mode only. Epidural administration of analgesics may be delivered by continuous, bolus, or continuous with bolus.
For epidural use, the administration of drugs is restricted to those anesthetic and analgesic drugs approved for continuous epidural administration: e.g., Chloroprocaine Hydrochloride USP, Lidocaine Hydrochloride USP, and Morphine Sulfate Injection USP (Preservative Free).
WARNING: Delayed respiratory depression following continuous
epidural administration of preservative-free morphine sulfate has been reported.
For epidural administration, the following is recommended:
Nylon or Teflon Pump sets without Y-sites Stickers for the tubing indicating ongoing administration
Note: Facilities practicing epidural anesthesia/analgesia must be staffed and equipped to manage cardio-pulmonary resuscitation (CPR). Supplies should include oxygen, naloxone, and other appropriate resuscitative drugs and equipment. Monitor patients continuously (e.g., oximetry) during epidural administration. Frequently observe patient for side effects (for up to 24 hours) following completion of drug administration by the epidural route.
®
catheter
10 Hospira GemStar Manual 430-00897-002 (Rev. 10/04)
Page 18
Introduction.fm Page 11 Thursday, November 11, 2004 3:04 PM
Contraindications for Use
Persons who do not have the mental and physical capability or the emotional stability to properly use the GemStar Pump should not be operate this pump.
DO NOT use the following with the GemStar Pump System:
Drugs not compatible with silicone rubber or PVC plastic
Drugs not stable under infusion conditions
Warnings and Cautions
Pay attention to the following warnings and cautions when operating the GemStar Pump. These warnings and cautions are repeated in the manual when relevant to the procedure being discussed. Pay attention to all alert messages.
General Warnings and Cautions
Federal (USA) law restricts this device to sale by or on the order of a physician or
other licensed practitioner.
Manual references to specific values are approximate only, unless indicated
otherwise. Air-in-line sensitivity values are approximate only.
When a Proximal Occlusion alarm activates, DO NOT power-off the pump.
Closely supervise patients who are likely to be adversely affected by unintended
operations and failures, including interrupted medication or fluid infusion. Ensure provisions for immediate corrective action are available.
1. Introduction
Regarding Drugs Used, Pump Sets, and Containers
USE ONLY GemStar Pump Sets with the GemStar Pump. Use of unauthorized
sets may result in injury to the patient or damage to the pump.
NEVER use drugs that are incompatible with silicone rubber or PVC plastic.
To reduce the loss of potency for drugs known to be absorbed by PVC plastic and
silicone, begin infusion as soon as practical after priming the pump set. Use of high flow rates during infusion will minimize drug absorption.
Hospira GemStar Manual 430-00897-002 (Rev. 10/04) 11
Page 19
Introduction.fm Page 12 Thursday, November 11, 2004 3:04 PM
1. Introduction
DO NOT use medications that are unstable under infusion conditions.
ALWAYS use connections with Luer Lock™ fittings.
To prevent contamination, use aseptic technique with all fluid-path connections.
Remove the protective coverings as assembly progresses.
ALWAYS place the flow stop in the closed position when the set is not in the
pump.
Make sure the cassette is properly installed. When properly installed, all four
cassette latches are visibly holding the cassette securely in the cassette pocket.
Ensure the tubing is behind the tubing guide (at the distal end).
Arrange tubing, cords, and cables to minimize the risk of patient strangulation or
entanglement.
NEVER use vented fluid containers (e.g., glass or rigid plastic) unless suspended
from a pole. When using a vented drip set, ALWAYS use an air-eliminating filter OR set the air sensitivity to either ON or 2 mL.
Stop infusion if signs or symptoms of infiltration occur.
GemStar Pump Sets contain an anti-siphon valve to reduce the potential for
gravity flow at a typical 36-inch head height.
Failure to use the anti-siphon valve may result in unrestricted flow.
ALWAYS close the slide clamps before removing the cassette from the pump.
When using the pump for secondary delivery (piggybacking), ensure the fluids
being infused are chemically and physically compatible.
When infusing short-half-life drugs (i.e. those dosed with units of mcg/kg/min) at
very low rates (0.8 mL/hr or less) the rate may be too slow for that drug. Consider using a lower drug concentration with these drugs so a pump rate above 0.8 mL/hr can be used.
When using pre-filled vials, the system may underdeliver at low flow rates.
12 Hospira GemStar Manual 430-00897-002 (Rev. 10/04)
Page 20
Introduction.fm Page 13 Thursday, November 11, 2004 3:04 PM
Regarding Air-In-Line and Infusion
To reduce the risk of infusing air, use an air-eliminating filter when the air
sensitivity is set to OFF.
ALWAYS prime the pump set to remove all air from the cassette, tubing, and
injection site prior to connecting to the patient.
ALWAYS disconnect the pump set from the patient prior to priming or purging.
Regarding Pump Operation
If the pump does not perform as stated in this manual, remove the pump from
service IMMEDIATELY.
If the pump does not display the self-test or if beeps do not sound at power-on,
check all power connections and install two fresh, disposable AA batteries. If the problem continues, contact Hospira Technical Support Operations before using the pump.
Ensure a cassette is installed in the pump before connecting to a printer or
computer. If the pump is connected to a patient: BEFORE connecting the pump to a computer or printer, press [ cassette. Leave the pump set connected to the patient. Install another cassette in the pump before connecting to a printer or computer.
Disconnect the pump from the patient BEFORE performing the Operation Test.
ALWAYS connect to a grounded AC mains power source when using AC mains
power. If the quality of the grounding source is in doubt, operate the pump only with batteries.
USE ONLY the AC mains adaptors specifically labeled for use with the GemStar
Pump.
USE ONLY the AC mains adaptor specifically labeled for use with the GemStar
Pump to charge the battery pack.
The GemStar AC Mains Adaptor is for use with the GemStar Pump or GemStar
Rechargeable Battery Pack only. DO NOT with other products.
During charging, the battery pack is warm. If the battery pack becomes hot to the
touch, IMMEDIATELY unplug the AC mains adaptor and contact Hospira Technical Support Operations.
1. Introduction
STOP], close the slide clamp, and eject the
use the GemStar AC mains adaptor
Hospira GemStar Manual 430-00897-002 (Rev. 10/04) 13
Page 21
Introduction.fm Page 14 Thursday, November 11, 2004 3:04 PM
1. Introduction
When removing the cord from the bottom of the pump, firmly grasp the connector
and pull straight out. DO NOT twist or bend the cord or connector.
To ensure proper pump operation, ALWAYS replace BOTH disposable batteries
with fresh, disposable AA batteries when a change is required.
ALWAYS fully close the battery door when using disposable batteries.
DO NOT use rechargeable batteries in the battery compartment.
Ensure two fresh, disposable AA batteries are installed for backup power when
using other power sources (e.g, docking station, battery pack). The pump will continue to operate on backup power if the external power source fails.
ALWAYS avoid sources of high-intensity electromagnetic radiation (e.g., radio
transmitters, MRI scanners, microwave ovens, X-ray machines, and CAT scanners).
Use of radio-frequency emitting devices, such as cellular telephones, 2-way
radios, and Electrical Surgical Devices (ESU), in close proximity to this device may affect its operation.
Possible explosion hazard exists if the pump is used in the presence of flammable
anesthetics. NEVER use the pump in the presence of flammable or explosive vapors.
Non-hazardous, low-level electrical potentials are commonly observed when
fluids are administered using infusion devices. These potentials are well within accepted safety standards, but may create artifacts on voltage-sensing equipment, such as ECG, EMG, and EEG machines. These artifacts vary at a rate that is associated with the infusion rate. If the monitoring machine is not operating correctly or has loose or defective connections to its sensing electrodes, these artifacts may be accentuated so as to simulate actual physiological signals. To determine if the abnormality in the monitoring equipment is caused by the infusion device instead of some other source in the environment, set the infusion device so that it is temporarily not delivering fluid (if clinically acceptable). Disappearance of the abnormality indicates that it was probably caused by electronic noise generated by the infusion device. Proper setup and maintenance of the monitoring equipment should eliminate the artifact. Refer to the appropriate monitoring system documentation for setup and maintenance instructions.
14 Hospira GemStar Manual 430-00897-002 (Rev. 10/04)
Page 22
Introduction.fm Page 15 Thursday, November 11, 2004 3:04 PM
The GemStar pump, when connected to the GemStar AC adaptor, has been tested
and found to comply with EMC/EMI limits in accordance with IEC/EN 60601-1-2 (2001). When connected to other power sources and/or optional system components, the GemStar system complies with EMC/EMI limits in accordance with IEC/EN 60601-1-2 (1993). These limits are designed to provide reasonable protection against harmful interference in a typical medical installation. The equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to other devices in the vicinity. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference with other devices, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures:
Reorient or relocate the receiving device Increase the separation between the equipment Connect the equipment to an outlet on a circuit different from that to which
the other device is connected
Consult the manufacturer or field service technician for help The use of portable and mobile RF equipment may have an impact on this and
other pieces of medical equipment
1. Introduction
Regarding Handling and Maintenance
Use proper care during unpacking, installation, and operation of the pump. If the
pump is inadvertently mishandled, check connections and programmed data to ensure there is no damage. Refer to Chapter 14 Operation Test for further information.
ALWAYS avoid dropping or hitting the pump. If the pump is dropped or hit,
ALWAYS check connections and programmed data.
NEVER use sharp objects (e.g., fingernails, pens, pencils, or other probes) to
program or clean the pump.
Use care not to damage the silicone seals around the sensor bodies in the cassette
pocket.
To avoid mechanical or electronic damage, NEVER submerge pump in water or
other fluids and avoid fluid spills. If the pump becomes wet, dry it immediately with a dry, lint-free cloth. Check connections and programmed data.
Hospira GemStar Manual 430-00897-002 (Rev. 10/04) 15
Page 23
Introduction.fm Page 16 Thursday, November 11, 2004 3:04 PM
1. Introduction
Some cleaning and sanitizing compounds may slowly degrade components made
from some plastic materials. DO NOT use compounds containing combinations of isopropyl alcohol and dimethyl benzyl ammonium chloride (i.e., alcohol wipes).
DO NOT sterilize by heat, steam, ethylene oxide (ETO), or radiation. Apply
disinfectants to the outside surface of the pump only. DO NOT use abrasive cleaners or materials on the pump. Using abrasive cleaners or cleaning solutions not recommended by Hospira may result in product damage.
Please comply with local disposal and recycling regulations as appropriate for
disposable batteries, rechargeable battery packs, medical electronic components, and pump sets.
16 Hospira GemStar Manual 430-00897-002 (Rev. 10/04)
Page 24
430-00897-002.book Page 17 Friday, October 22, 2004 2:38 PM
The minimum elements required for using the GemStar Pump are:
GemStar Pump
Appropriate power source
Appropriate GemStar Pump Set
Appropriate fluid container
Patient-access device
Overview
Chapter 2
Basic Operation
To start a therapy:
1 Determine the program settings. Refer to the appropriate therapy chapter for a
worksheet that may be copied and used to record program settings.
2 Select a power source. Refer to Selecting a Power Source on page 18.
3 Prime the pump set. Refer to GemStar Pump Set on page 29.
4 Load the cassette into the pump. Refer to Loading the Cassette on page 32.
5 Power-on the pump and go to the Programming Menu. Refer to Powering On
the Pump on page 24.
6 Select the appropriate optional system components, such as a bolus cord.
7 Select the therapy type and enter a program. Refer to the appropriate therapy
chapter for detailed instructions.
8 Press [
START] to begin delivery.
Hospira GemStar Manual 430-00897-002 (Rev. 10/04) 17
Page 25
Basic_Operation.fm Page 18 Wednesday, November 10, 2004 8:40 AM
2. Basic Operation
Selecting a Power Source
The GemStar Pump may be powered by one or more of the following:
Power Source Description Ambulatory Hospital Home
GemStar AC Mains Adaptor (USA)
GemStar AC Mains Adaptor (Universal)
GemStar Docking Station
GemStar Battery Pack
Disposable Batteries Two disposable AA alkaline or
Wall-mount adaptor plugs into a 1 10 VAC mains supply.
Tabletop adaptor plugs into a 100–240 VAC mains supply.
Plugs into a 100–240 VAC Volt mains supply while mounted on an I.V. pole. The docking station’s internal batteries power the GemStar Pump during periods of transport.
The battery pack’s internal batteries power the GemStar Pump during periods of transport or when use of AC power is not desirable.
lithium batteries (available from local sources)
••
••
••
••
••
CAUTION: To ensure proper pump operation, ALWAYS replace BOTH
batteries with fresh, disposable AA batteries when a change is required.
DO NOT use rechargeable batteries in the battery compartment. Ensure two fresh, disposable AA batteries are installed for backup
power when using other power sources (e.g, docking station, battery pack). The pump will continue to operate on backup power if the external power source fails.
18 Hospira GemStar Manual 430-00897-002 (Rev. 10/04)
Page 26
430-00897-002.book Page 19 Friday, October 22, 2004 2:38 PM
Installing Disposable Batteries
1 Flip open the tab of the battery door on
the bottom of the pump.
2 With the pump keypad facing up, hold
onto the tab and turn the battery door to the left (counterclockwise) until it is released from the battery compartment.
3 Pull the door free and remove any
existing batteries.
4 Install two fresh, disposable AA
batteries into the battery compartment. Ensure the batteries are inserted as shown.
5 Return the battery door to the battery
compartment.
2. Basic Operation
6 Turn the battery door to the right
(clockwise). Ensure the battery door is secured under the slot.
7 Flip closed the tab of the battery door to
lock in place.
Hospira GemStar Manual 430-00897-002 (Rev. 10/04) 19
Page 27
430-00897-002.book Page 20 Friday, October 22, 2004 2:38 PM
2. Basic Operation
Connecting the GemStar AC Mains Adaptor
1 Connect the cord of the AC mains adaptor to the connector labeled “3 VDC” on
the bottom of the pump.
2 Plug the AC mains adaptor into a standard wall outlet.
CAUTION: ALWAYS connect to a grounded AC mains power source when
using the AC mains adaptor. USE ONLY AC mains adaptors specifically labeled for use with the GemStar Pump.
When removing the cord from the bottom of the pump, firmly grasp the connector and pull straight out. DO NOT twist or bend the cord or connector.
Note: The green LED on the pump remains lit when connected to AC mains power.
20 Hospira GemStar Manual 430-00897-002 (Rev. 10/04)
Page 28
Basic_Operation.fm Page 21 Wednesday, February 2, 2005 2:24 PM
Using the GemStar Rechargeable Battery Pack
Use the GemStar AC Mains Adaptor (List 13026-01 [USA] or 13072-01 [Universal]) to recharge the GemStar Battery Pack.
CAUTION: ALWAYS connect to a grounded AC mains outlet when using AC
mains adaptor.
Use only AC mains adaptors specifically labeled for use with the GemStar Pump to charge the battery pack.
During charging, the battery pack is warm. If the battery pack becomes hot to the touch, IMMEDIATELY unplug AC mains power and contact Hospira Technical Support Operations.
2. Basic Operation
Note: A charged battery pack will last approximately 24 hours when the pump is infusing at rates less than 125 mL/hr.
To connect the battery pack:
1 Line up the connectors on the top of the battery
pack with the holes on the bottom of the pump.
2 Snap the battery pack into place.
1
To release the battery pack:
1 Press the release button on the battery pack.
2 Pull the battery pack straight out from the pump.
Note: When the pump is connected to the battery pack, the pump’s green LED flashes and the battery pack light is off.
When the pump is connected to the battery pack at power-on, the pump displays
Hospira GemStar Manual 430-00897-A02 (Rev. 02/05) 21
USING EXT BATT.
Page 29
430-00897-002.book Page 22 Friday, October 22, 2004 2:38 PM
2. Basic Operation
To recharge the battery pack:
1 Plug the pin connector of the AC mains adaptor into the port on the bottom of the
battery pack.
2 Plug the AC mains adaptor into a standard wall outlet.
3 Leave the battery pack connected to the AC mains adaptor for at least 8 hours to
recharge the battery pack
CAUTION: When removing the cord from the bottom of the pump, firmly
grasp the connector and pull straight out. DO NOT twist or bend the cord or connector.
Note: The battery pack light is yellow while the battery pack is charging. When the battery pack is charged, the battery pack light is green.
22 Hospira GemStar Manual 430-00897-002 (Rev. 10/04)
Page 30
430-00897-002.book Page 23 Friday, October 22, 2004 2:38 PM
Connecting to the GemStar Docking Station
Connecting the Pump
1 Mount the docking station to the I.V. pole and tighten
the knob.
2 Line up the connectors on the bottom of the pump with
the connectors on the docking station.
3 Slide the pump into the docking station until it snaps
into place.
4 Connect the docking station to AC mains power.
2. Basic Operation
Note: The docking station light is yellow while the docking station is charging. When the docking station is charged, the light is green.
Disconnecting the Pump
To remove the pump:
1 Press the release button on the top of the docking
station.
2 Slide the pump out.
Hospira GemStar Manual 430-00897-002 (Rev. 10/04) 23
Page 31
430-00897-002.book Page 24 Friday, October 22, 2004 2:38 PM
2. Basic Operation
Powering On the Pump
CAUTION: If the pump does not display the self-test or if beeps do not sound at
power-on, check all power connections and install two fresh, disposable AA batteries. If the problem continues, contact Hospira Technical Support Operations before using the pump.
Note: If a cassette is not properly loaded, the pump will alarm after the power-on test is complete.
1 Press [ON/OFF] to power-on the pump. The pump
completes a self-test and displays the time and date. No response is required.
GEMSTAR 7
UNIT SELF-TEST
TIME IS 10:00 AM
FRI, JUN 11, 04
Note: The self-test takes approximately 30 seconds to complete. If the pump is locked, lock symbols appear at the beginning and end of the second line:
UNIT SELF-TEST .
2 The pump checks for available power sources. Respond to the display message as
shown in the following table:.
Display Message Power Source(s) User Response
No message displays AC power and
disposable batteries.
LOW BATTERIES
ENTER TO CONFIRM
NO INTERNAL BATT
NO POWER LOSS
ALARM AVAILABLE
ENTER TO CONFIRM
AC power and weak (low voltage) disposable batteries; or weak disposable batteries only.
No disposable batteries or dead disposable batteries; and external power.
No response required, the pump goes to the next step.
Press to confirm.
Power-off the pump and install two fresh, disposable AA batteries as a backup power source.
Press to confirm.
Power-off the pump and install two fresh, disposable AA batteries as a backup power source.
YES
ENTER
YES
ENTER
24 Hospira GemStar Manual 430-00897-002 (Rev. 10/04)
Page 32
Basic_Operation.fm Page 25 Wednesday, November 10, 2004 12:27 PM
Display Message Power Source(s) User Response
2. Basic Operation
USING BATTERIES
ENTER TO CONFIRM
USING EXT BATT
ENTER TO CONFIRM
Disposable batteries only.
External batteries (docking station battery or battery pack) and disposable batteries.
Press to confirm. If an external power source is in use,
ensure the connections are secure. Press to confirm.
If an AC mains adaptor is in use, ensure the connections are secure.
YES
ENTER
YES
ENTER
Note: To power-off the pump, press and hold until the display clears. The pump cannot be powered off during a CHECK CASSETTE alarm (except during
PROGRAMMING mode). Refer to Chapter 12 Troubleshooting for information on
clearing a CHECK CASSETTE alarm.
For more information on: Refer to the following pages:
Setting the clock Set Clock in Chapter 10 Options Menu, on page 108. Locking the keypad Keypad Lock in Chapter 10 Options Menu, on
page 104.
Hospira GemStar Manual 430-00897-002 (Rev. 10/04) 25
Page 33
Basic_Operation.fm Page 26 Friday, November 12, 2004 8:32 AM
2. Basic Operation
Accessing the Programming Menu
After power-on, the pump checks its memory for a stored program. Respond to the display message as shown in the following table:
Display Message Condition User Response
The pump displays the Programming Menu.
1 RESUME PROGRAM 2 CLR PROG&SHIFT
3 CLR PROG,SHIFT
AND HISTORY
Pump memory is clear.
A program is currently stored in pump memory.
> To enter a new program: Select a therapy
from the Programming Menu and enter a new program. Refer to the appropriate therapy chapter for instructions.
> To continue the current program, press [1]. > To reset the current program:
1 Press [1]. 2 Press [CHANGE] to display the Change
Menu and select
NEW CONTAINER.
f
> To enter a new program and clear shift
totals: Press [2]. The pump clears the program and shift totals from memory and enters PROGRAMMING mode.
> To enter a new program, clear shift totals,
and clear history: Press [3]. The pump clears the program, shift totals, and history from memory and enters PROGRAMMING mode.
Tip: To enter a new program when the pump is already powered on:
1 Press [STOP] to place the pump in STOP mode.
2 Press [CHANGE] to display the Change Menu.
3 Select NEW PROGRAM. The pump clears the current program and
displays the Programming Menu.
26 Hospira GemStar Manual 430-00897-002 (Rev. 10/04)
Page 34
430-00897-002.book Page 27 Friday, October 22, 2004 2:38 PM
Programming the Pump
Programming Overview
The pump’s menu system provides step-by-step guidance through all of the programming steps. Refer to the appropriate therapy chapter for detailed programming instructions.
Programming worksheets are available at the end of each therapy chapter. Each worksheet lists the minimum and maximum value (program range) allowed for that therapy. If a number outside of the program range is entered, the pump displays the minimum or maximum value allowed.
1 Select the desired therapy from the Programming Menu.
2 Respond to display messages and enter requested values.
Note: Program entries can be changed any time before the Program Review is
complete. Press [BACK-UP] to return to previous program displays and make changes as desired.
2. Basic Operation
3 Review the program, if required. Press to begin review. Press [
when done.
SAVING PROGRAM displays. The pump stores the program in memory and then
4
STOP mode.
enters
Tip: If you enter an incorrect number during programming, press [CHANGE]
BEFORE pressing [ correct number.
Hospira GemStar Manual 430-00897-002 (Rev. 10/04) 27
YES/ENTER] to reset the number to zero. Then enter the
YES/ENTER]
Page 35
430-00897-002.book Page 28 Friday, October 22, 2004 2:38 PM
2. Basic Operation
Programming Tips
The following table explains how to use the keypad during programming:
Key Function
[0] - [9] Press the number keys to select menu items and to enter numeric values.
Note: Some menu items require pressing two number keys to select that items.
For example, when selecting a unit of measure for a weight dosed program, the menu may display as:
01 mcg/kg/mg 02 mcg/kg/hr 03 mcg/kg/min 04 mg/kg/min
To select mg/kg/min, press [
0][4].
Press to accept entries or to continue when done viewing a display.
YES
ENTER
Press during programming and alarm conditions to display information
about the current condition.
Press or to scroll through display messages.
Press to insert a decimal point in a numeric value.
When entering a numeric value during programming, press before
pressing to reset the value to zero.
YES
ENTER
Press to return to a previous display, to exit the Options Menu, or to exit
the
HELP mode.
28 Hospira GemStar Manual 430-00897-002 (Rev. 10/04)
Page 36
Basic_Operation.fm Page 29 Wednesday, November 10, 2004 8:41 AM
Starting a New Container
Note: The NEW CONTAINER function is not available when the pump is in FULL Lock.
Refer to Chapter 10 Options Menu, page 97 for more information on locking the keypad.
To repeat the current program with a new container:
2. Basic Operation
1 From the 2 Select
CONTAINER
3 Review the program. Press to begin review. 4 Press [
STOP mode, press [CHANGE] to display the Change Menu.
NEW CONTAINER from the Change Menu. The pump displays NEW
START] to begin delivery.
GemStar Pump Set
The GemStar Pump Set is a sterile, single-use, disposable pump set. It is designed for one-time use only and should be properly discarded after use.
Fluid path and areas beneath undisturbed protective set covers are sterile and nonpyrogenic in the intact unit package.
Contact a Hospira representative for more information on selecting the appropriate pump set configuration. Additional disposable components, such as air-eliminating filters and extension sets, may be added to the line as required. To use a GemStar Pump Set, follow the instructions included with the set.
Note: GemStar Pump Sets contain an anti-siphon valve to reduce the potential for
gravity flow at a typical 36-inch head height.
False occlusion alarms and fluid delivery inconsistencies may occur when pumping viscous fluids in microbore tubing at rates greater than 500 mL/hr.
for several seconds and clears the amount infused from memory.
STERILE
2
Hospira GemStar Manual 430-00897-002 (Rev. 10/04) 29
Page 37
430-00897-002.book Page 30 Friday, October 22, 2004 2:38 PM
2. Basic Operation
A typical GemStar Pump Set is shown below:
1 Piercing Pin 6 Tubing
2 Proximal End 7 Slide Clamp
3 Cassette 8 Secure Lock
4 Flow Stop (open / close ) 9 Anti-Siphon valve
5 Distal End
30 Hospira GemStar Manual 430-00897-002 (Rev. 10/04)
Page 38
430-00897-002.book Page 31 Friday, October 22, 2004 2:38 PM
WARNING: ALWAYS prime the pump set to remove air from the
cassette, tubing, and injection sites prior to connecting to the patient. ALWAYS disconnect the pump set from the patient prior to priming or purging. Arrange tubing, cords, and cables to minimize the risk of patient strangulation or entanglement.
Failure to use the anti-siphon valve may result in unrestricted flow.
CAUTION: USE ONLY GemStar Pump Sets with the GemStar Pump. Use of
unauthorized sets may result in injury to the patient or damage to the pump.
To prevent contamination, use aseptic technique with all fluid-path connections. Remove protective coverings as assembly progresses.
2. Basic Operation
Opening the Pump Set
1 Open the GemStar Pump Set package and
remove the contents.
2 Remove the protective covers from the fluid
container administration port and the pump set.
Hospira GemStar Manual 430-00897-002 (Rev. 10/04) 31
Page 39
Basic_Operation.fm Page 32 Friday, November 12, 2004 8:57 AM
2. Basic Operation
Loading the Cassette
1 Identify the four black cassette latches inside the casse tte pocket.
2 Line up the cassette as shown
(with the flow stop button facing the pump).
3 Insert the cassette into the cassette
pocket by pushing firmly along the entire cassette until it is firmly seated.
32 Hospira GemStar Manual 430-00897-002 (Rev. 10/04)
Page 40
430-00897-002.book Page 33 Friday, October 22, 2004 2:38 PM
2. Basic Operation
4 Verify that all four cassette latches
are securely holding the cassette.
CAUTION: Ensure the cassette is properly installed. When properly installed, all four cassette latches are visibly holding the cassette securely in the cassette pocket.
5 Place the tubing in the
tubing channel.
Releasing the Cassette
1 Close the CAIR® (roller) clamp or
slide clamp on the distal line.
2 Push down on the cassette release
(large black button) on the top of the pump.
3 Remove the cassette.
Note: When the cassette is
Note: Ensure the tubing is behind
Hospira GemStar Manual 430-00897-002 (Rev. 10/04) 33
the tubing guide (at the distal end).
Note: The pump cannot be powered off during a CHECK CASSETTE alarm (except during
PROGRAMMING mode). Refer to
Chapter 12 Troubleshooting for information on clearing a CHECK CASSETTE alarm.
released from the pump, the cassette flow stop automatically closes, preventing free flow.
Page 41
430-00897-002.book Page 34 Friday, October 22, 2004 2:38 PM
2. Basic Operation
Priming a GemStar Pump Set
Priming Pump Sets WITHOUT a Drip Chamber
WARNING: Failure to use the anti-siphon valve may result in
unrestricted flow.
1 Push down on the flow stop to
close the cassette.
2 Hold the container with the
administration port toward the ceiling and insert the piercing pin.
3 Push down on the flow stop to
open (priming position) the cassette.
4 Invert the cassette approximately
45°.
5 While holding the rotated cassette,
squeeze or roll the container to remove air from the container and cassette.
6 Return the cassette to the upright
position. Continue priming until fluid fills the tubing.
34 Hospira GemStar Manual 430-00897-002 (Rev. 10/04)
Page 42
430-00897-002.book Page 35 Friday, October 22, 2004 2:38 PM
2. Basic Operation
7 Push down on the flow stop to
close the cassette.
8 Load the cassette.
CAUTION: Ensure all four cassette latches are clearly visible after the cassette is installed. Refer to Loading the Cassette on page 32.
Note: Ensure the tubing is behind
the tubing guide (at the distal end).
CAUTION: When priming is complete, ensure no fluid flows at the distal end of the pump set. If flow is observed, DO NOT use the pump set.
9 Connect the pump set to the
patient-access device.
Hospira GemStar Manual 430-00897-002 (Rev. 10/04) 35
Page 43
430-00897-002.book Page 36 Friday, October 22, 2004 2:38 PM
2. Basic Operation
Priming Pump Sets WITHOUT a Drip Chamber Using the PURGE Key
WARNING: Failure to use the anti-siphon valve may result in
unrestricted flow.
Note: The PURGE function is available only if there is a program in the pump.
1 Push down on the flow stop to
close the cassette.
2 Hold the container with the
administration port toward the ceiling and insert the piercing pin.
3 Load the cassette.
CAUTION: Ensure all four cassette latches are clearly visible after the cassette is installed. Refer to Loading the Cassette on page 32.
Note: Ensure the tubing is behind the tubing guide (at the distal end).
4 Program desired therapy.
36 Hospira GemStar Manual 430-00897-002 (Rev. 10/04)
Page 44
430-00897-002.book Page 37 Friday, October 22, 2004 2:38 PM
5 Press [PURGE].
2. Basic Operation
PURGE
6 The pump displays
SET?
Press [YES/ENTER].
HELP
PRIME THE
7 While holding the container with
the administration port toward the ceiling, press and hold [
PURGE].
Note: Purge rate is 250 mL/hr. The pump purges for up to 2 minutes each time you press and hold [
Continue using the [
PURGE].
PURGE]
key until fluid fills the tubing. Ensure all air is removed from the container, cassette, and tubing.
CAUTION: When priming is complete, ensure no fluid flows at the distal end of the pump set. If flow is observed, DO NOT use the pump set.
8 Connect the pump set to the
patient-access device.
Hospira GemStar Manual 430-00897-002 (Rev. 10/04) 37
Page 45
430-00897-002.book Page 38 Friday, October 22, 2004 2:38 PM
2. Basic Operation
Priming Pump Sets WITH a Drip Chamber
WARNING: Failure to use the anti-siphon valve may result in
unrestricted flow.
When using a vented chamber, ALWAYS use an air-eliminating filter OR set the air sensitivity to either ON or 2 mL.
1 Push down on the flow stop to
close the cassette.
2 Insert the piercing pin into the
container. Suspend the container on an I.V. Pole.
3 Squeeze the drip chamber until
filled halfway.
4 Push down on the flow stop to
open (priming position) the cassette.
®
5 Ensure the CAIR
(roller) clamp
is open.
6 Invert the cassette approximately
45°.
38 Hospira GemStar Manual 430-00897-002 (Rev. 10/04)
Page 46
430-00897-002.book Page 39 Friday, October 22, 2004 2:38 PM
2. Basic Operation
7 While holding the rotated cassette,
remove air from the container and cassette.
8 Return the cassette to the upright
position. Continue priming until fluid fills the tubing.
9 Push down on the flow stop to
close the cassette.
CAUTION: Ensure all four cassette latches are clearly visible after the cassette is installed. Refer to Loading the Cassette on page 32.
Note: Ensure the tubing is behind the tubing guide (at the distal end).
10 Load the cassette.
Hospira GemStar Manual 430-00897-002 (Rev. 10/04) 39
CAUTION: When priming is complete, ensure no fluid flows at the distal end of the pump set. If flow is observed, DO NOT use the pump set.
11 Connect the pump set to the
patient-access device.
Page 47
430-00897-002.book Page 40 Friday, October 22, 2004 2:38 PM
2. Basic Operation
Priming Pump Sets WITH a Drip Chamber Using the PURGE Key
WARNING: Failure to use the anti-siphon valve may result in
unrestricted flow.
When using a vented set, ALWAYS use an air-eliminating filter OR set the air sensitivity to either ON or 2 mL.
Note: The Purge function is available only if there is a program in the pump.
1 Push down on the flow stop to
close the cassette.
2 Insert the piercing pin into the
container. Suspend the container on an I.V. Pole.
3 Squeeze the drip chamber until
filled halfway.
4 Push down on the flow stop to
open (priming position) the cassette.
5 Ensure the CAIR
is open.
®
(roller) clamp
40 Hospira GemStar Manual 430-00897-002 (Rev. 10/04)
Page 48
430-00897-002.book Page 41 Friday, October 22, 2004 2:38 PM
2. Basic Operation
6 Load the cassette.
CAUTION: Ensure all four cassette latches are clearly visible after the cassette is installed. Refer to Loading the Cassette on page 32.
Note: Ensure the tubing is behind the tubing guide (at the distal end).
7 Program the desired therapy.
8 Press [
9 The pump displays
10 Press and hold [
11 Connect the pump set to the
PURGE].
PURGE
PRIME THE
SET?
Press [YES/ENTER].
PURGE].
Note: Purge rate is 250 mL/hr. The pump purges for up to 2 minutes each time you press and
PURGE
hold [
Continue using the [ until fluid fills the tubing. Ensure all air is removed from the container, cassette, and tubing.
CAUTION: When priming is complete, ensure no fluid flows at the distal end of the pump set. If flow is observed, DO NOT use the pump set.
patient-access device.
].
PURGE] key
HELP
Hospira GemStar Manual 430-00897-002 (Rev. 10/04) 41
Page 49
430-00897-002.book Page 42 Friday, October 22, 2004 2:38 PM
2. Basic Operation
Priming Pump Sets Using Pre-filled Vials
WARNING: Failure to use the anti-siphon valve may result in
unrestricted flow.
CAUTION: When using pre-filled vials, the system may underdeliver at low
flow rates.
1 If using a pump set with the
Y-extension, use the slide clamp to close off the KVO (Keep Vein Open) side of the tubing.
2 Open the upper and lower slide
clamps.
3 Push down on the flow stop to
close the cassette.
4 Connect the pump set to a Hospira
pre-filled vial and injector.
5 Push down on the flow stop to
open (priming position) the cassette.
42 Hospira GemStar Manual 430-00897-002 (Rev. 10/04)
Page 50
430-00897-002.book Page 43 Friday, October 22, 2004 2:38 PM
2. Basic Operation
6 Hold the vial and set upright
(vertically) on a flat surface with the pump set extending from the top.
7 Slowly push down on the injector
until fluid flows at the distal end of the tubing, which indicates that all air is cleared from the vial, injector, cassette, and tubing.
8 Push down on the flow stop to
close the cassette.
9 If the set includes a KVO
Y-extension, close the slide clamp below the pre-filled vial and open the slide clamp on the KVO side. Prime the KVO Y-extension.
10 Load the cassette.
CAUTION: Ensure all four cassette latches are clearly visible after the cassette is installed. Refer to Loading the Cassette on page 32.
Hospira GemStar Manual 430-00897-002 (Rev. 10/04) 43
Page 51
430-00897-002.book Page 44 Friday, October 22, 2004 2:38 PM
2. Basic Operation
Note: Ensure the tubing is behind
the tubing guide (at the distal end).
CAUTION: When priming is complete, ensure no fluid flows at the distal end of the pump set. If flow is observed, DO NOT use the pump set.
11 Connect the pump set to the
patient-access device.
44 Hospira GemStar Manual 430-00897-002 (Rev. 10/04)
Page 52
430-00897-002.book Page 45 Friday, October 22, 2004 2:38 PM
Priming Pump Sets with Extensions
WARNING: Failure to use the anti-siphon valve may result in
unrestricted flow.
2. Basic Operation
1 Push down on the flow stop to
close the cassette.
2 Hold the container with the
administrative port toward the ceiling and insert the piercing pin.
3 Push down on the flow stop to
open (priming position) the cassette.
4 Invert the cassette approximately
45°.
5 While holding the rotated cassette,
squeeze or roll the container to remove air from the container and cassette.
6 Return the cassette to the upright
position. Continue priming until fluid fills the tubing.
Hospira GemStar Manual 430-00897-002 (Rev. 10/04) 45
Page 53
430-00897-002.book Page 46 Friday, October 22, 2004 2:38 PM
2. Basic Operation
7 Push down on the flow stop to
close the cassette.
8 Connect the extension set to the
pump set.
9 Load the cassette.
CAUTION: Ensure all four cassette latches are clearly visible after the cassette is installed. Refer to Loading the Cassette on page 32.
Note: Ensure the tubing is behind
the tubing guide (at the distal end).
10 Program the desired therapy.
11 Press [
12 The pump displays
13 Press and hold [
PURGE].
PURGE
SET?
Press [YES/ENTER].
Note: Purge rate is 250 mL/hr. The pump purges for up to 2 minutes each time you press and
PURGE].
hold [
Continue using the [ until fluid fills the tubing. Ensure all air is removed from the container, cassette, and tubing.
HELP
PRIME THE
PURGE].
PURGE] key
CAUTION: When priming is complete, ensure no fluid flows at the distal end of the pump set. If flow is observed, DO NOT use the pump set.
14 Connect the pump set to the
patient-access device.
46 Hospira GemStar Manual 430-00897-002 (Rev. 10/04)
Page 54
430-00897-002.book Page 47 Friday, October 22, 2004 2:38 PM
Total Parenteral Nutrition (TPN)
Therapy Features and Specifications
This therapy allows programming of Total Parenteral Nutrition (TPN) protocols.
Taper Up and Taper Down may be programmed to gradually increase and
decrease the infusion rate at the beginning and end of a therapy. If Taper Down is selected during programming, the pump allows auto tapering if the continuous delivery is stopped before the programmed volume has been infused.
Set an optional KVO (Keep Vein Open) during programming.
Tip: The Programming Worksheet at the end of this chapter lists the minimum and
maximum values allowed for this protocol. Use copies of the programming worksheet to record program values before programming the pump.
Chapter 3
TPN Programming Tips
The following table explains how to use the keypad during TPN programming:
Key Function Key Function
[0] - [9] Selects menu items and
enters numeric values.
Accepts entries or continues when done viewing a display.
Displays information during programming and alarm conditions.
Hospira GemStar Manual 430-00897-002 (Rev. 10/04) 47
Scrolls through program review and help messages. The up arrow inserts a decimal point in a numeric value.
Clears a numeric value (resets to zero) during programming.
Returns to a previous display, exits the Options Menu, and exits the
HELP mode.
Page 55
430-00897-002.book Page 48 Friday, October 22, 2004 2:38 PM
3. Total Parenteral Nutrition (TPN)
TPN Programming
1 Select TPN from the Programming Menu.
2 Select the TPN delivery mode.
3 Enter the total VTBI (Volume To Be Infused).
Note: Maximum VTBI is 9600 mL minus any taper amount.
4 Enter the Taper Up and Taper Down times (if applicable).
Note: Time is entered in hours:minutes (00:00). For one hour thirty minutes, press [1] [3] [0]; not [9] [0].
5 Enter the total time to infuse.
6 Enter the KVO rate. If no KVO is desired, press [
displayed is zero.
Note: KVO runs after Taper Down until the total container is infused.
7 Enter the container size (VTBI + KVO).
Note: If KVO is programmed, increase the container size to accommodate the desired KVO volume.
8 Select the air sensitivity, if required.
9 Review the program. Press to begin review.
10 If using the
WARNING: ALWAYS disconnect the pump set from the patient before
PURGE key to prime the pump set, press [PURGE] now.
priming or purging.
YES/ENTER] while the value
48 Hospira GemStar Manual 430-00897-002 (Rev. 10/04)
Page 56
TPN.fm Page 49 Wednesday, November 10, 2004 8:42 AM
Changing a TPN Program
3. Total Parenteral Nutrition (TPN)
Can be changed before is pressed
•VTBI
Taper times
Total TPN time
START
Can be changed at any time
KVO rate
Increase container size only
Air sensitivity
Cannot be changed
Delivery mode
Note: Y ou cannot change a pr ogram when the pump is in FULL or CONTAINER Lock.
To change a program:
1 From the 2 As each message displays, make the desired changes or press [
STOP mode, press [CHANGE] and select CHANGE PROGRAM.
YES/ENTER] to
accept the current setting.
3 Review the program. Press to begin review. 4 Press [
START] to begin infusion.
Hospira GemStar Manual 430-00897-002 (Rev. 10/04) 49
Page 57
430-00897-002.book Page 50 Friday, October 22, 2004 2:38 PM
3. Total Parenteral Nutrition (TPN)
TPN Auto Taper Down
Auto Taper Down is available only if a Taper Down time was entered during programming. Auto Taper Down may be activated when the continuous delivery portion of a TPN with Taper Down protocol is stopped. Auto Taper Down is not available while the pump is in an alarm condition.
1 Press [
2 Enter the Auto Taper Down time.
3 Press [
STOP] during TPN delivery.
The pump displays
Press [NO] to return to the STOP mode. Then press [START] to continue therapy.
Press [
The maximum Auto Taper Down time is three hours OR two times the remaining volume divided by the continuous infusion rate, whichever is less.
Note: Time is entered in hours:minutes (00:00). For one hour thirty minutes, press [1] [3] [0]; not [9] [0].
Taper Down.
YES/ENTER] to set Auto Taper Down.
YES/ENTER] to return to the RUN mode, then press [START] to begin Auto
DO YOU WANT AUTO TAPERING?
50 Hospira GemStar Manual 430-00897-002 (Rev. 10/04)
Page 58
TPN.fm Page 51 Wednesday, November 10, 2004 8:42 AM
3. Total Parenteral Nutrition (TPN)
GemStar
®
TPN Programming Worksheet
GIVE:
Parameters Selections and/or Values to be
Entered
Therapy TPN N/A Delivery Mode
VTBI (Volume To Be Infused)
Taper Up
Taper Down No
Total Time to Infuse
Continuous Continuous + Taper Down Continuous + Taper Up Continuous + Taper Up and Down
mL 1 - 9600 mL*
No Yes hrs:min 1 min - 3 hrs
Yes hrs:min 1 min - 3 hrs**
Programmable Ranges
N/A
hrs:min 1 min - 24 hrs
KVO (Keep Vein Open)
Container Size
(VTBI + KVO) mL 9999 mL max Air Sensitivity
* Maximum VTBI is 9600 mL minus any taper amount. ** Programmable ranges for taper down are 1 minute to 3 hours OR 2 times the
remaining volume divided by the continuous delivery rate, whichever is less.
Hospira GemStar Manual 430-00897-002 (Rev. 10/04) 51
No KVO Rate: 0.0 Yes KVO Rate: 1.0 - 5.0 mL/hr
On Alarms at approx. 0.5 mL of air 2 mL Alarms at approx. 2 mL of air Off No alarm
N/A
Page 59
430-00897-002.book Page 52 Friday, October 22, 2004 2:38 PM
Notes
52 Hospira GemStar Manual 430-00897-002 (Rev. 10/04)
Page 60
PainMgmt.fm Page 53 Friday, November 12, 2004 8:24 AM
Therapy Features and Specifications
This therapy allows programming of a continuous only , bolus only, or continuous with bolus delivery .
6 In Continuo us + Bolus protocol s, you can enter zero for either the continuous rate
or the bolus dose when entering or changing a program. The continuous rate and bolus dose cannot be zero at the same time.
6 The default bolus delivery rate is 125 mL/hr. 6 A loading dose may be set during programming, which can be delivered after
programming and priming are complete. The default loading dose rate is 125 mL/hr.
6 Priming using [PURGE] is only allowed before [START] is pressed or after one of
the following events occur:
6 Complete Review of New Program 6 New Container 6 Air-in-Line Alarm
Chapter 4
Pain Management
CAUTION: When using pre-filled vials, the system may underdeliver at low
flow rates.
Note: New Container is allowed only when at least 1/2 of the container volume has
d
f
been infused or 1 hour has elapsed since starting the infusion.
The use of a lockbox is recommended with pain management protocols. Refer to Chapter 15 Optional System Components, page 153, for instructions on using a lockbox with the GemStar Pump.
Tip: The Programming W or ksheet at the e nd of this chap ter list s the min imum and maximum values allowed for this protocol. Use copies of the programming worksheet to record program values before programming the pump.
Hospira GemStar Manual 430-00897-002 (Rev. 10/04) 53
Page 61
430-00897-002.book Page 54 Friday, October 22, 2004 2:38 PM
4. Pain Management
Pain Management Programming Tips
The following table explains how to use the keypad during pain management programming:
Key Function Key Function
[0] - [9] Selects menu items and
enters numeric values.
Accepts entries or continues when done viewing a display.
Displays information during programming and alarm conditions.
Pain Management Programming
1 Select PAIN MANAGEMENT from the Programming Menu.
2 Select the delivery mode.
Note: If Continuous + Bolus is selected, the continuous delivery rate or bolus dose can be set to zero through the CHANGE menu without reprogramming the pump.
3 Select the unit of measure and concentration as required.
Scrolls through program review and help messages. The up arrow inserts a decimal point in a numeric value. The down arrow toggles units of measure when entering concentrations.
Clears a numeric value (resets to zero) during programming.
Returns to a previous display, exits the Options Menu, and exits the
HELP mode.
54 Hospira GemStar Manual 430-00897-002 (Rev. 10/04)
Page 62
430-00897-002.book Page 55 Friday, October 22, 2004 2:38 PM
4 Set the continuous delivery rate.
Note: The pump only delivers in increments of 0.1 mL. Values entered in a Pain
Management therapy may be rounded to the nearest 0.1 mL. For example, at a concentration of 15 mg/mL, a rate of 5 mg rounds to 4.5 mg.
5 mg = 0.333 mL
4.5 mg = 0.3 mL
When a value is rounded, the pump sounds four quick beeps, displays ROUNDING for a few seconds, and then displays the rounded value. Press accept the rounded value and continue to the next step.
5 Program a loading dose as required.
6 Program a bolus dose with lockout time and dose limit as required.
Regarding a 1-Hour or 4-Hour Dose Limit: The hour dose limit is the maximum volume (continuous plus bolus or bolus only amount) that can be delivered over the selected period. When the programmed dose limit is reached, the continuous delivery stops and a new bolus is not allowed. As the oldest delivery amounts age out of the 1-hour or 4-hour dose limit record, the pump accepts bolus requests or resumes continuous delivery (if programmed).
4. Pain Management
[YES/ENTER] to
Note: If a 1-hour or 4-hour dose limit is programmed, a bolus in progress will stop
Hospira GemStar Manual 430-00897-002 (Rev. 10/04) 55
when the amount infused reaches the dose limit.
Loading doses are included in a 1-hour or 4-hour dose limit. A dose limit in effect could prevent the delivery of a loading dose.
7 If programmed in units of mcg or mg, the line displaying the units flashes. Enter
the container size in proper unit measure.
Note: To enter the total in mL when programming a concentration, press to move to the mL line. Enter the mL value when
8 Select the air sensitivity, if required.
9 Review the program. Press to begin review.
10 If using the PURGE key to prime the pump set, press [
WARNING: ALWAYS disconnect the pump set from the patient before
priming or purging.
mL flashes on the display.
PURGE] now.
Page 63
430-00897-002.book Page 56 Friday, October 22, 2004 2:38 PM
4. Pain Management
Bolus Delivery
Set bolus doses and delivery limits during programming. The default bolus dose delivery rate in a pain management protocol is 125 mL/hr.
A bolus cord is available for patient bolus requests. Refer to Connecting the Bolus Cord in Chapter 15 Optional System Components, on page 156, for more information.
Delivering a Bolus Dose
From the
RUN mode, press either the bolus button [+] on the top of the pump or the
button on the end of the bolus cord to begin bolus dose delivery. The pump flashes
BOLUS DELIVERY on the display and the amount infused accrues as the delivery
progresses.
When the bolus dose is complete, the pump does the following:
Records the bolus request and delivery amount to the event and bolus history logs
Adds the delivered boluses to the total bolus amount in the program amounts
Resets the bolus lockout time (if applicable) and ignores other bolus requests until
the lockout time has elapsed
Continues programmed infusion
Note: ALL bolus requests are recorded to the event and bolus history logs.
If the bolus delivery does not start, it could be locked out by one of the following conditions:
Bolus lockout period
Loading dose delivery in progress
Bolus dose delivery in progress
Bolus per-hour limit
1-hour or 4-hour dose limit
56 Hospira GemStar Manual 430-00897-002 (Rev. 10/04)
Page 64
430-00897-002.book Page 57 Friday, October 22, 2004 2:38 PM
Bolus Dose Interruptions
Bolus deliveries may be interrupted by:
Pressing [STOP]
An alarm condition
To continue bolus delivery after an interruption:
1 Press [START]. The pump displays COMPLETE BOLUS NOW?
2 Respond as follows:
To do this: Press this key: Bolus lockout time is set to:
4. Pain Management
Deliver the remaining bolus amount
Clear the undelivered bolus amount
Loading Dose Delivery
You can set a loading dose during programming, if desired. A loading dose can be delivered after programming and priming are complete.
Note: Loading dose amounts are included in a 1-hour or 4-hour dose limit.
Delivering a Loading Dose
1 After priming and programming, press [
2 The pump displays
begin loading dose.
When the loading dose is complete, the pump does the following:
Automatically begins programmed infusion
Records the dose amount to the history log
Resets the bolus lockout time (if applicable) and ignores other bolus requests until
the lockout time has elapsed
The time of delivery completion
The time the bolus was interrupted
START].
DELIVER THE LOADING DOSE NOW? Press [YES/ENTER] to
Hospira GemStar Manual 430-00897-002 (Rev. 10/04) 57
Page 65
430-00897-002.book Page 58 Friday, October 22, 2004 2:38 PM
4. Pain Management
Loading Dose Interruptions
Loading dose deliveries can be interrupted by:
Pressing [STOP]
An alarm condition
To continue a loading dose:
1 Resolve the interruption.
2 Press [
START]. The pump displays COMPLETE THE LOADING DOSE?
3 Respond as follows:
To do this: Press this key:
Deliver the remaining loading dose amount
Clear the remaining loading dose amount
Changing a Pain Management Program
Can be changed at any time Cannot be changed
Delivery rate
Bolus
Bolus lockout
Note: You cannot change a program when the pump is in FULL or CONTAINER Lock. With a
RATE CHANGE Lock, the continuous delivery rate and bolus amount
can be changed within the set ranges.
Although the delivery mode cannot be changed, if Continuous + Bolus is programmed, the continuous rate or bolus amount can be changed to zero. The minimum values set with a
Container size
Air sensitivity
Dose limit amount
RATE CHANGE Lock could restrict this function.
Delivery mode
Unit of measure
Concentration
Dose limit type
58 Hospira GemStar Manual 430-00897-002 (Rev. 10/04)
Page 66
430-00897-002.book Page 59 Friday, October 22, 2004 2:38 PM
To change a program:
1 From the
STOP mode, press [CHANGE] and select CHANGE PROGRAM.
4. Pain Management
2 As each message displays, make the desired changes or press [
accept the current setting.
3 Review the program. Press to begin review.
4 Press [
START] to begin delivery.
YES/ENTER] to
Hospira GemStar Manual 430-00897-002 (Rev. 10/04) 59
Page 67
430-00897-002.book Page 60 Friday, October 22, 2004 2:38 PM
4. Pain Management
GemStar
®
Pain Management Programming Worksheet
GIVE:
Parameters Selections and/or Values to be
Entered
Therapy Pain Management N/A Delivery Mode
Unit of Measure
Continuous Rate
Loading Dose
Bolus Dose 0 - 2500 mg
Bolus Lockout
Optional Dose Limit
Continuous Bolus Only Continuous + Bolus
Concentration: mg/mL
Concentration: mcg/mL mL
No
Yes Amount:
minutes 5 - 999 min
4-hour limit Amount:
Programmable Ranges
N/A
0.1 - 100 mg/mL
1 - 1000 mcg/mL
0.1 - 2500 mg/hr 1 - 25000 mcg/hr
0.1 - 25.0 mL/hr
0.1 - 2500 mg 1 - 25000 mcg
0.1 - 25.0 mL
0 - 25000 mcg 0 - 25.0 mL
4X rate - 400 mL, 40000 mg, 400000 mcg*
1-hour limit Amount:
# boluses/hourNumber: No dose limit
Rate - 100 mL, 10000 mg, 100000 mcg*
1 - 12
*For continuous plus bolus delivery.
Note: For Bolus Only, the 1-hour or 4-hour dose limit minimum is 1 bolus dose.
Container Size 0.1 - 99999 mg
1 - 9999000 mcg
0.1 - 9999 mL
Air Sensitivity
60 Hospira GemStar Manual 430-00897-002 (Rev. 10/04)
On Alarms at approx. 0.5 mL of air 2 mL Alarms at approx. 2 mL of air Off No alarm
N/A
Page 68
Intermittent.fm Page 61 Wednesday, February 2, 2005 2:24 PM
Therapy Features and Specifications
This therapy allows programming of multiple doses at regular intervals.
Optional KVO (Keep Vein Open) can be set during programming.
The pump can be disconnected from the patient between dose deliveries. A call
back alarm can be set to alert the patient or caregiver to reconnect the pump before the next scheduled delivery begins. The pump must remain powered on for the call back alarm to function. At the set time before each dose, the pump produces an audible and visual alarm and flashes
Tip: The Programming Worksheet at the end of this chapter lists the
minimum and maximum values allowed for this protocol. Use copies of the programming worksheet to record program values before programming the pump.
Chapter 5
Intermittent
START on the display.
Intermittent Programming Tips
The following table explains how to use the keypad during intermittent programming:
Key Function Key Function
[0] - [9] Selects menu items and
enters numeric values.
Accepts entries or continues when done viewing a display.
Displays information during programming and alarm conditions.
Hospira GemStar Manual 430-00897-A02 (Rev. 02/05) 61
Scrolls through program review and help messages. Toggles between AM and PM in time-based entries. The up arrow inserts a decimal point in numeric values.
Clears a numeric value (resets to zero) during programming.
Returns to a previous display, exits the Options Menu, and exits the
HELP mode.
Page 69
430-00897-002.book Page 62 Friday, October 22, 2004 2:38 PM
5. Intermittent
Intermittent Programming
1 Select INTERMITTENT from the Programming Menu.
2 Enter the dose amount in mL.
3 Enter the infusion time for each dose in hours:minutes (00:00).
For example, press [
4 Enter the frequency of dose deliveries.
For example, press [
5 Enter the total number of doses in the container.
To determine the number of doses in the container, divide the volume in the container by the dose amount.
6 Enter the KVO rate in mL.
KVO delivers before a delayed start time, between doses, and after the last dose until the total container is infused.
4] [0] if each dose is to deliver for 40 minutes.
4] [0] [0] to start a dose every four hours.
If no KVO is desired, press [
7 Enter the container size.
The pump automatically calculates the minimum container size required to deliver the number of programmed doses and KVO between doses.
62 Hospira GemStar Manual 430-00897-002 (Rev. 10/04)
Note: If KVO is desired before a delayed start, increase the container size to include the desired KVO volume.
8 Select the air sensitivity, if required.
Note: DO NOT press [YES/ENTER] after selecting the air sensitivity. If [YES/ENTER] is accidentally pressed and a call back alarm is not desired, press [
BACK-UP] to return to the beginning of step 9 and press [NO].
9 The pump displays
Press [NO] if you do not want to set the call back alarm.
Press [
Enter the time before the next dose that the alarm should sound. For example, press [ scheduled delivery.
YES/ENTER] to set the call back alarm.
1] [0] to set the alarm to sound 10 minutes before the start of the next
SET CALL BACK ALARM FOR NEXT DOSE?
YES/ENTER] while the value displayed is zero.
Page 70
Intermittent.fm Page 63 Wednesday, November 10, 2004 8:43 AM
Note: When programming the call back alarm, be sure to allow enough time for the
patient or caregiver to reconnect the pump to the patient and to press [ BEFORE the scheduled dose start time.
10 The pump displays the current time.
If the displayed time is correct, press [
To set the clock, press [
5. Intermittent
START]
YES/ENTER].
OPTIONS].
11 The pump displays the current time and
P ress [YES/ENTER] to begin delivery when the pump enters the RUN mode.
Press [
Note: When entering the time in 12-hour format, press or to toggle between AM and PM.
12 Review the program. Press to begin review. 13 If using the
WARNING: ALWAYS disconnect the pump set from the patient before
Note: The call back alarm works only when the pump is powered on AND in the
STOP mode. DO NOT power-off the pump while it is disconnected from the patient
if you have set the call back alarm. When the alarm sounds, press [
Then reconnect the pump to the patient.
Tip: To use a delayed start without specifying a start time:
NO] to set a delayed start time. The pump displays SET START TIME.
Enter the desired start time and press [
PURGE key to prime the pump set, press [PURGE] now.
priming or purging.
SILENCE] to silence the alarm for three minutes.
START NOW?
YES/ENTER] when done.
1 Press [YES/ENTER]. 2 Complete the program review. 3 Power-off the pump. 4 When the pump is powered on again, complete the program review, then
press [
START] to begin delivery.
Hospira GemStar Manual 430-00897-002 (Rev. 10/04) 63
Page 71
430-00897-002.book Page 64 Friday, October 22, 2004 2:38 PM
5. Intermittent
Intermittent Delivery Interruptions
Missed Start Time
If the programmed start time is missed, it can be changed through the Options
Menu. Refer to Next Dose in Chapter 10 Options Menu, on page 111.
If the pump remains in the STOP mode and the first dose is missed, press [START]
to begin delivery immediately and all remaining doses will be shifted.
If the pump is programmed with a delayed start time and is then powered off
before [ programmed start time.
For example, the current time is 10:00 PM. The user sets the delayed start time for 10:30 PM and then powers off the pump. The user does not power on the pump until 10:40 PM, which is 10 minutes AFTER the scheduled start time. The pump does not begin delivery until 10:30 PM the FOLLOWING day.
START] is pressed, the pump will skip to the NEXT occurrence of the
Dose Interruptions
To understand how interruptions are resolved, compare the normal Intermittent delivery diagram below for a 1-hour dose every four hours, to the interruption diagrams shown in the examples:
12123456789101112
Dose Delivered -
If an Intermittent dose is interrupted and the pump is placed in the RUN mode
during the on or off cycle of the current dose, the dose continues delivery from the point where it was interrupted and all remaining doses are shifted.
For example, in the following diagram, an intermittent dose was started at 2:00. The dose was interrupted during the current dose cycle and was started again before the end of the current dose cycle.
64 Hospira GemStar Manual 430-00897-002 (Rev. 10/04)
Page 72
430-00897-002.book Page 65 Friday, October 22, 2004 2:38 PM
12123456789101112
Dose Delivered - Dose Interrupted -
If an Intermittent dose is interrupted and the pump is placed in the RUN mode
after the off cycle of the current dose, the next dose begins delivery. All remaining doses are shifted if necessary.
For example, in the following diagram, an intermittent dose was started at 2:00. It was interrupted during the current delivery cycle at 2:30, but was not started again until after the current delivery cycle at 6:45:
5. Intermittent
12123456789101112
Dose Delivered - Dose Interrupted -
When the dose is complete, the start time for the Next Dose displays. To change
the start time refer to Chapter 10 Options Menu, page 97.
Any missed Intermittent doses, except the first dose, are added to the end of the
delivery schedule.
Hospira GemStar Manual 430-00897-002 (Rev. 10/04) 65
Page 73
Intermittent.fm Page 66 Wednesday, November 10, 2004 8:44 AM
5. Intermittent
Changing an Intermittent Program
Can be changed at any time
Dose parameters
KVO rate
Container size
Air sensitivity
Call back alarm
Can be changed if no dose is in progress
Next dose start time Current time
Cannot be changed
START
after is pressed
Note: You cannot change a program when the pump is in FULL or CONTAINER Lock.
To change a program:
1 From the
STOP mode, press [CHANGE] and select CHANGE PROGRAM.
2 As each message displays, make the desired changes or press [YES/ENTER] to
accept the current setting.
3 Review the program. Press to begin review. 4 Press [
START] to begin delivery.
66 Hospira GemStar Manual 430-00897-002 (Rev. 10/04)
Page 74
430-00897-002.book Page 67 Friday, October 22, 2004 2:38 PM
5. Intermittent
GemStar
®
Intermittent Programming Worksheet
GIVE:
Parameters Selections and/or Values to be Entered Programmable
Ranges
Therapy Intermittent N/A
Dose Amount
0.1 - 9600 mL
Infusion Time Per Dose
Dose Frequency
Number of Doses Per Container
KVO (Keep Vein Open)
Container Size*
Air Sensitivity
Call Back Alarm
Start Time
No KVO Rate: 0.0
Yes KVO Rate:
On Alarms at approx. 0.5 mL of air 2 mL Alarms at approx. 2 mL of air Off No alarm
No Ye s
Alarm: hrs:min before start
Start immediately after programming
hrs:min 1 min - 24 hrs
(Infusion time per dose + 1 min) to 96 hrs
hrs:min
1 - 999
0.1 -5.0 mL/hr
0.1 - 9999 mL N/A
1 min to time between doses
Start time:
Specify AM or PM if not 24-hour clock.
* The pump automatically calculates the minimum container size required to deliver the number of
12:00 AM - 11:59 PM or 00:00 - 23:59
programmed doses and KVO between doses. If KVO is desired before a delayed start, increase the container size to include the desired KVO volume.
Hospira GemStar Manual 430-00897-002 (Rev. 10/04) 67
Page 75
430-00897-002.book Page 68 Friday, October 22, 2004 2:38 PM
Notes
68 Hospira GemStar Manual 430-00897-002 (Rev. 10/04)
Page 76
Continuous.fm Page 69 Wednesday, February 2, 2005 2:38 PM
Therapy Features and Specifications
This therapy allows programming of a continuous delivery.
Optional KVO (Keep Vein Open) can be set during programming.
The rate can be titrated while the pump is infusing.
An optional secondary container can be “piggybacked” on the primary container
during regular delivery using a GemStar Piggyback Pump Set. Refer to Piggybacking in this chapter for more information.
Tip: The Programming Worksheet at the end of this chapter lists the minimum
and maximum values allowed for this protocol. Use copies of the programming worksheet to record program values before programming the pump.
Chapter 6
Continuous
Continuous Programming Tips
The following table explains how to use the keypad during continuous programming:
Key Function Key Function
[0] - [9] Selects menu items and
enters numeric values.
Accepts entries or continues when done viewing a display.
Displays information during programming and alarm conditions.
Hospira GemStar Manual 430-00897-A02 (Rev. 02/05) 69
Scrolls through program review and help messages. The up arrow inserts a decimal point in a numeric value.
Clears a numeric value (resets to zero) during programming.
Returns to a previous display, exits the Options Menu, and exits the
HELP mode.
Page 77
430-00897-002.book Page 70 Friday, October 22, 2004 2:38 PM
6. Continuous
Continuous Programming
1 Select CONTINUOUS from the Programming Menu.
2 Select the unit of measure and concentration as required.
3 Enter the rate.
4 Enter the VTBI (Volume to be Infused).
5 Enter the KVO rate, if desired.
KVO runs after the programmed VTBI until the total container is infused. If no
KVO is desired, press [
6 Enter the container size (VTBI + KVO).
VTBI is the default value on the
Note: If KVO is programmed, increase the container size to include the KVO delivery.
To enter the total in mL when programming a concentration, press to move to the mL line. Enter the mL size when
7 Select the air sensitivity, if required.
YES/ENTER] while the value displayed is zero.
CONTAINER display.
mL flashes on the display.
8 Review the program. Press to begin review.
9 If using the PURGE key to prime the pump set, press [
WARNING: ALWAYS disconnect the pump set from the patient before
priming or purging.
70 Hospira GemStar Manual 430-00897-002 (Rev. 10/04)
PURGE] now.
Page 78
430-00897-002.book Page 71 Friday, October 22, 2004 2:38 PM
Rate Titration
To titrate the rate while the pump is infusing:
1 In the
2
RUN mode, press the number key for the new rate. Press [YES/ENTER].
For example: if the current rate is 100 mL/hr, press [ the rate to 50 mL/hr.
Note: You have approximately five seconds after pressing a key to press another numeric key, the decimal key, or [ seconds, the rate does not change and the delivery continues as previously programmed.
NEW RATE displays with the new value.
Press [
Press [
YES/ENTER] within 10 seconds to accept the change and begin delivery
at the new rate.
NO] to continue delivery at the previously programmed rate.
6. Continuous
5] [0] [YES/ENTER] to change
YES/ENTER]. If no key is pressed within five
Piggybacking
Piggybacking with the GemStar Pump requires a GemStar Primary Piggyback Pump Set with a proximal Y-site backcheck valve and a secondary piggyback set with an extension hook. The secondary container must be higher than the primary container. Refer to the instructions included with the piggyback pump set for more information.
WARNING: When using the pump for secondary delivery (piggybacking),
ensure the fluids being infused are chemically and physically compatible.
The piggyback rate and piggyback VTBI display until the secondary container has been delivered. Then the primary infusion continues.
Note: The piggyback infusion amount is NOT added to the primary programmed infusion amount.
Hospira GemStar Manual 430-00897-002 (Rev. 10/04) 71
Page 79
430-00897-002.book Page 72 Friday, October 22, 2004 2:38 PM
6. Continuous
Preparing for Secondary Delivery (Piggybacking)
1 Ensure a primary piggyback pump set with a
proximal Y-site backcheck valve is in place.
2 Use the extension hook from the secondary set
to suspend the primary container from the I.V. pole.
3 Prepare the secondary set according to the
instructions included with the set.
4 Attach a needle or blunt cannula (if appropriate)
and prime the secondary set.
5 Suspend the secondary container from the I.V.
pole.
Note: When using a secondary container of
500 mL or greater, ensure the bottom of the secondary container is at least 7 inches (17.8 cm) above the fluid level in the primary container. Use additional extension hooks if necessary.
6 Connect the secondary set to the proximal Y-site
(or proximal luer port).
72 Hospira GemStar Manual 430-00897-002 (Rev. 10/04)
Page 80
430-00897-002.book Page 73 Friday, October 22, 2004 2:38 PM
Programming a Secondary Delivery (Piggybacking)
6. Continuous
Note: To distinguish the piggyback delivery from the primary delivery, PIGGYBACK displays and the piggyback VTBI counts down to completion. For example, a piggyback rate of 250 mL/hr with a remaining VTBI of
218.5 mL will display as shown.
Tip: Prepare and set up containers BEFORE programming the pump for piggyback delivery.
1 Press [
2 Enter the dose amount for the secondary container.
3 Enter the infusion time for the secondary container.
4 Open the slide clamp or CAIR
5 Press [
Note: Primary flow stops until the secondary VTBI is delivered; then primary flow continues. Any remaining secondary fluid will be delivered at the primary rate. The maximum piggyback rate is 300 mL/hr. For secondary rates above 300 mL/hr, flow may occur from the primary container.
Changing a Piggyback Delivery
1 Press [
CHANGE] and select PIGGYBACKING from the Change Menu.
®
(roller) clamp on the tubing of the secondary
container.
YES/ENTER] to begin piggyback delivery.
CHANGE] and select PIGGYBACKING from the Change Menu.
PIGGYBACK
RATE 250.0 mL/hr
VTBI 218.5 mL
2 Press [
3 Enter the new piggyback dose amount and press [
4 Set the infusion time for the secondary container.
5 Press [
Note: You must know the amount that has already been delivered. When the piggyback dose amount is changed, the pump begins a new VTBI countdown.
Hospira GemStar Manual 430-00897-002 (Rev. 10/04) 73
CHANGE] to reset the dose amount to zero.
YES/ENTER].
YES/ENTER] to begin piggyback delivery.
Page 81
430-00897-002.book Page 74 Friday, October 22, 2004 2:38 PM
6. Continuous
Canceling a Piggyback Delivery
1 Press [
2 Press [
3 Press [
CHANGE] and select PIGGYBACKING from the Change Menu.
CHANGE] to reset the dose amount to zero.
YES/ENTER] to cancel piggyback delivery.
Changing a Continuous Program
Can be changed at any time Cannot be changed
Delivery rate
•VTBI
KVO rate
Note: You cannot change a program when the pump is in FULL or CONTAINER Lock.
To change a program:
1 From the
STOP mode, press [CHANGE] and select CHANGE PROGRAM.
2 As each message displays, make the desired changes or press [YES/ENTER] to
accept the current setting.
3 Review the program. Press to begin review.
4 Press [
START] to begin delivery.
Container size
Air sensitivity
Piggybacking
Unit of measure
Concentration
74 Hospira GemStar Manual 430-00897-002 (Rev. 10/04)
Page 82
430-00897-002.book Page 75 Friday, October 22, 2004 2:38 PM
6. Continuous
GemStar
®
Continuous Programming Worksheet
GIVE:
Parameter Selections and/or Values to be
Entered
Therapy Continuous N/A
Unit of Measure
Concentration: mg/mL
Concentration: mcg/mL mL
Rate 0.1 - 9999 mg/hr
VTBI
(Volume To Be Infused)
KVO
No KVO Rate: 0.0
(Keep Vein Open)
Yes KVO Rate:
Container Size
(VTBI + KVO)
Air Sensitivity
On Alarms at approx. 0.5 mL of air 2 mL Alarms at approx. 2 mL of air Off No alarm
Piggyback Dose Amount 1 - 7200 mL
Piggyback Infusion Time
hrs:min
Programmable Ranges
0.1 - 1000 mg/mL
1 - 1000 mcg/mL
1 - 999900 mcg/hr
0.1 - 1000 mL/hr
0.1 - 99999 mg
1 - 9999000 mcg
0.1 - 9999 mL
0.1 - 5.0 mL/hr
0.1 - 99999 mg
1 - 9999000 mcg
0.1 - 9999 mL
N/A
1 min or dose amount / 300 mL/hr to 24 hrs or dose amount / 0.1 mL/hr
Hospira GemStar Manual 430-00897-002 (Rev. 10/04) 75
Page 83
430-00897-002.book Page 76 Friday, October 22, 2004 2:38 PM
Notes
76 Hospira GemStar Manual 430-00897-002 (Rev. 10/04)
Page 84
430-00897-002.book Page 77 Friday, October 22, 2004 2:38 PM
Therapy Features and Specifications
This therapy allows programming of delivery rate as a function of body weight.
The air sensitivity in a Weight Dosed program is automatically set to ON. This
setting can be changed through the Options Menu after programming is complete. Refer to Air Sensitivity in Chapter 10 Options Menu, on page 107.
The pump automatically delivers KVO (Keep Vein Open) after the programmed
VTBI (Volume To Be Infused) has completed.
The rate can be titrated while the pump is infusing.
Chapter 7
Weight Dosed
Tip: The Programming Worksheet at the end of this chapter lists the minimum and
maximum values allowed for this protocol. Use copies of the programming worksheet to record program values before programming the pump.
WARNING: When using a vented drip set, ALWAYS use an air-
eliminating filter OR set the air sensitivity to either ON or 2 mL.
CAUTION: When infusing short-half-life drugs (i.e., those dosed with units of
mcg/kg/min) at very low rates (0.8 mL/hr or less) the rate may be too slow for that drug. Consider using a lower drug concentration with these drugs so a pump rate above 0.8 mL/hr can be used.
Hospira GemStar Manual 430-00897-002 (Rev. 10/04) 77
Page 85
430-00897-002.book Page 78 Friday, October 22, 2004 2:38 PM
7. Weight Dosed
Weight Dosed Programming Tips
The following table explains how to use the keypad during weight dosed programming:
Key Function Key Function
[0] - [9] Selects menu items and
enters numeric values.
Accepts entries or continues when done viewing a display.
Displays information during programming and alarm conditions.
Weight Dosed Programming
1 Select WEIGHT DOSED from the Programming Menu. 2 Select the unit of measure.
Note: If two numbers appear before a menu time (e.g., 02), press both number keys to select that item.
3 Enter the patient weight, if required. 4 Enter the concentration:
Enter the value for the unit of measure when it flashes on the display.
Press [
Tip: Press to toggle between units of measure.
YES/ENTER] to confirm.
Scrolls through program review and help messages. The up arrow inserts a decimal point in a numeric value.
Clears a numeric value (resets to zero) during programming.
Returns to a previous display, exits the Options Menu, exits the
HELP mode.
5 Enter the rate.
78 Hospira GemStar Manual 430-00897-002 (Rev. 10/04)
Note: The pump automatically displays the mL/hr equivalent even when another unit of measure is selected.
6 Enter the VTBI. 7 Review the program. Press to begin review.
Page 86
WeightDosed.fm Page 79 Wednesday, November 10, 2004 8:45 AM
8 If using the PURGE key to prime the pump set, press [PURGE] now.
WARNING: ALWAYS disconnect the pump set from the patient before
priming or purging.
Auto KVO
When the programmed VTBI has completed delivery, Auto KVO delivery begins automatically and the empty container alarm sounds.
7. Weight Dosed
Note: Auto KVO delivers even while the empty container alarm sounds. Press
SILENCE] to quiet the alarm for 2 minutes.
[
If the weight dosed rate is: Auto KVO delivers at:
Greater than or equal to 1 mL/hr 1 mL/hr Less than 1 mL/hr The weight dosed rate
Rate Titration
To titrate the rate while the pump is infusing:
1 In the
For example: if the current rate is 4.0 mg/kg/hr, press [ the rate to 5.0 mg/kg/hr.
Press to enter the rate in mL/hr. When using units other than mL/hr , the pump automatically adjusts the mL/hr value.
Note: You have approximately five seconds after pressing a key to press another
numeric key, the decimal key, or [YES/ENTER]. If no key is pressed within five seconds, the rate does not change and the delivery continues as previously programmed.
2
NEW RATE displays with the new value.
Press [
P ress [
RUN mode, press the number keys for the new rate. Press [YES/ENTER].
5] [YES/ENTER] to change
YES/ENTER] within 10 seconds to accept the change and begin delivery
at the new rate.
NO] to continue delivery at the previously programmed rate.
Hospira GemStar Manual 430-00897-002 (Rev. 10/04) 79
Page 87
430-00897-002.book Page 80 Friday, October 22, 2004 2:38 PM
7. Weight Dosed
VTBI Titration
To titrate the VTBI while the pump is infusing:
1 Press [
2 Press the number keys for the new VTBI. Press [
3
CHANGE] and select TITRATE VTBI.
For example: if the current VTBI is 50 mL, press [ the VTBI to 100 mL.
Note: You have approximately five seconds after pressing a key to press another numeric key, the decimal key, or [ seconds, the VTBI does not change and the delivery continues as previously programmed.
NEW VTBI displays with the new rate.
Press [
Press [
YES/ENTER] within 10 seconds to accept the change and begin delivery
with the new VTBI.
NO] to continue delivery at the previously programmed VTBI.
YES/ENTER]. If no key is pressed within five
Changing a Weight Dosed Program
Can be changed at any time Cannot be changed
Delivery rate
•VTBI
Patient weight
Note: You cannot change a program when the pump is in FULL or CONTAINER Lock.
YES/ENTER].
1][0][0][YES/ENTER] to change
Unit of measure
Concentration (amount of drug)
To change a program:
1 From the
2 As each message displays, make the desired changes or press [
accept the current setting.
3 Review the program. Press to begin review.
4 Press [
80 Hospira GemStar Manual 430-00897-002 (Rev. 10/04)
STOP mode, press [CHANGE] and select CHANGE PROGRAM.
YES/ENTER] to
START] to begin delivery.
Page 88
WeightDosed.fm Page 81 Wednesday, November 10, 2004 8:46 AM
GemStar® Weight Dosed Programming Worksheet
GIVE:
7. Weight Dosed
Parameter Selections and/or Values to be Entered Programmable
Ranges
Therapy Weight Dosed N/A Unit of Measure**
mcg/kg/min mcg/kg/hr mcg/min mg/kg/min mg/kg/hr mg/min mL/hr mg/hr
mcg/hr grams/hr ng/kg/min U/kg/hr U/min U/hr mU/min mEq/hr
N/A
Patient Weight (When programming in kg/min or kg/hr)
2.0 - 200.0 kg Concentration (2 entries are
required. Use to toggle)
(For all units of measure EXCEPT mL/hr)
mg U mEq in in in
mL mL mL
0.1 - 9999000 mg
0.1 - 99999999 U
0.1 - 9999 mEq
0.1 - 9999 mL Rate
(see below *)
VTBI (Volume To Be Infused)
* Programmable Ranges for Rate (not all units of measure are available in all pump configurations):
0.001 - 833 mg/kg/min 0.1 - 1000 mL/hr 0.001 - 99999999 U/min
0.001 - 500000 mg/kg/hr 0.001 - 999999 mg/hr 0.001 - 99999999 U/hr
0.001 - 16666 mg/min 0.001 - 9999999 mcg/hr 0.001 - 999999999 mU/min
0.001 - 166666 mcg/kg/min 0.001 - 999 grams/hr 0.001 - 999 mEq/hr
0.001 - 9999999 mcg/kg/hr 0.009 - 999 ng/kg/hr
0.002 - 166666 mcg/min 0.001 - 99999999 U/kg/hr
** If two numbers appear before a menu item (e.g., 02), press both number keys to select that item.
CAUTION: When infusing short-half-life drugs (i.e., those dosed with units of mcg/kg/min) at very low rates (0.8 mL/hr or less) the rate may be too slow for that drug. Consider using a lower drug
0.1 - 9999 mL
concentration so a pump rate above 0.8 mL/hr can be used.
Hospira GemStar Manual 430-00897-002 (Rev. 10/04) 81
Page 89
430-00897-002.book Page 82 Friday, October 22, 2004 2:38 PM
Notes
82 Hospira GemStar Manual 430-00897-002 (Rev. 10/04)
Page 90
430-00897-002.book Page 83 Friday, October 22, 2004 2:38 PM
Therapy Features and Specifications
This therapy allows quick programming of delivery rate in milliliters (mL) only.
The air sensitivity is automatically set to ON. This setting can be changed through
the Options Menu after programming is complete. Refer to Air Sensitivity in Chapter 10 Options Menu, on page 107.
The pump automatically delivers KVO (Keep Vein Open) after the programmed
VTBI (Volume To Be Infused) is complete.
The rate can be titrated while the pump is infusing.
An optional secondary container can be “piggybacked” onto the primary container
during regular delivery using a GemStar Piggyback Pump Set. Refer to Piggybacking in this chapter for more information.
A program review is not required and is available only through the Options Menu.
Chapter 8
mL/hr Only
Tip: The Programming Worksheet at the end of this chapter lists the minimum and
maximum values allowed for this protocol. Use copies of the programming worksheet to record program values before programming the pump.
WARNING: When using a vented set, ALWAYS use an air-eliminating
filter OR set the air sensitivity to either ON or 2 mL.
Hospira GemStar Manual 430-00897-002 (Rev. 10/04) 83
Page 91
mLhr.fm Page 84 Wednesday, November 10, 2004 9:39 AM
8. mL/hr Only
mL/hr Programming Tips
The following table explains how to use the keypad during mL/hr programming:
Key Function Key Function
[0] - [9] Selects menu items and
enters numeric values.
Accepts entries or continues when done viewing a display.
Displays information during programming and alarm conditions.
Scrolls through program review and help messages. The up arrow inserts a decimal point in a numeric value.
Clears a numeric value (resets to zero) during programming.
Returns to a previous display, exits the Options Menu, exits the mode.
HELP
mL/hr Only Programming
1 Select mL/hr Only from the Programming Menu. 2 Enter the rate. 3 Enter the VTBI. 4 If using the PURGE key to prime the pump set, press [
WARNING: ALWAYS disconnect the pump set from the patient before
priming or purging.
PURGE] now.
Note: A program review is not required for mL/hr Only programs.
84 Hospira GemStar Manual 430-00897-002 (Rev. 10/04)
Page 92
430-00897-002.book Page 85 Friday, October 22, 2004 2:38 PM
Auto KVO
When the programmed VTBI has completed delivery, Auto KVO delivery begins automatically and the empty container alarm sounds.
8. mL/hr Only
Note: Auto KVO delivers even while the empty container alarm sounds. Press [SILENCE] to quiet the alarm for 2 minutes.
If the weight dosed rate is: Auto KVO delivers at:
Greater than or equal to 1 mL/hr 1 mL/hr
Less than 1 mL/hr The programmed rate
Rate Titration
To titrate the rate while the pump is infusing:
1 In the
For example, if the current rate is 100 mL/hr, press [ the rate to 50 mL/hr.
Note: You have approximately five seconds after pressing a key to press
2
NEW RATE displays with the new value.
Press [
Press [
RUN mode, press the number keys for the new rate. Press [YES/ENTER].
5] [0] [YES/ENTER] to change
another numeric key, the decimal key, or [ pressed within five seconds, the rate does not change and the delivery continues as previously programmed.
YES/ENTER] within 10 seconds to accept the change and begin delivery
at the new rate.
NO] to continue delivery at the previously programmed rate.
YES/ENTER]. If no key is
Hospira GemStar Manual 430-00897-002 (Rev. 10/04) 85
Page 93
430-00897-002.book Page 86 Friday, October 22, 2004 2:38 PM
8. mL/hr Only
VTBI Titration
To titrate the VTBI while the pump is infusing:
1 Press [
2 Press the number keys for the new VTBI. Press [
For example: if the current VTBI is 50 mL, press [ the VTBI to 100 mL.
Note: You have approximately five seconds after pressing a key to press
NEW VTBI displays with the new rate.
3
Press [
Press [
Piggybacking
Piggybacking with the GemStar Pump requires a GemStar Primary Piggyback Pump Set with a proximal Y-site backcheck valve and a secondary piggyback set with an extension hook. The secondary container must be higher than the primary container. Refer to the instructions included with the piggyback pump set for more information.
WARNING: When using the pump for secondary delivery
CHANGE] and select TITRATE VTBI.
YES/ENTER].
1][0][0][YES/ENTER] to change
another numeric key, the decimal key, or [YES/ENTER]. If no key is pressed within five seconds, the VTBI does not change and the delivery continues as previously programmed.
YES/ENTER] within 10 seconds to accept the change and begin delivery
with the new VTBI.
NO] to continue delivery at the previously programmed VTBI.
(piggybacking), ensure the fluids being infused are chemically and physically compatible.
The piggyback rate and piggyback VTBI display until the secondary container has been delivered. Then the primary infusion continues.
Note: The piggyback infusion amount is NOT added to the primary programmed infusion amount.
86 Hospira GemStar Manual 430-00897-002 (Rev. 10/04)
Page 94
430-00897-002.book Page 87 Friday, October 22, 2004 2:38 PM
Preparing for Secondary Delivery (Piggybacking)
1 Ensure a primary piggyback pump set with a
proximal Y-site backcheck valve is in place.
2 Use the extension hook from the secondary set
to suspend the primary container from the I.V. pole.
3 Prepare the secondary set according to the
instructions included with the set.
4 Attach a needle or blunt cannula (if appropriate)
and prime the secondary pump set.
5 Suspend the secondary container from the I.V.
pole.
Note: When using a secondary container of
500 mL or greater, ensure the bottom of the secondary container is at least 7 inches (17.8 cm) above the fluid level in the primary container. Use additional extension hooks if necessary.
8. mL/hr Only
6 Connect the secondary set to the proximal Y-site
(or proximal luer port).
Hospira GemStar Manual 430-00897-002 (Rev. 10/04) 87
Page 95
mLhr.fm Page 88 Thursday, February 3, 2005 11:31 AM
8. mL/hr Only
Programming a Secondary Delivery (Piggybacking)
Note: To distinguish the piggyback delivery from the primary delivery, counts down to completion. For example, a piggyback rate of 250 mL/hr with a remaining VTBI of 218.5 mL will display as shown.
Tip: Prepare and set up containers BEFORE programming the pump for piggyback delivery.
1 Press [
2 Enter the dose amount for the secondary container.
3 Enter the infusion time for the secondary container.
4 Open the slide clamp or CAIR
5 Press [
Note: Primary flow stops until the secondary VTBI is delivered; then primary flow continues. Any remaining secondary fluid will be delivered at the primary rate. The maximum piggyback rate is 300 mL/hr. For secondary rates above 300 mL/hr, flow may occur from the primary container.
Changing a Piggyback Delivery
1 Press [
2 Press [
PIGGYBACK displays and the piggyback VTBI
CHANGE] and select PIGGYBACKING from the Change Menu.
®
(roller) clamp on the tubing of the secondary set.
YES/ENTER] to begin piggyback delivery.
CHANGE] and select PIGGYBACKING from the Change Menu.
CHANGE] to reset the dose amount to zero.
PIGGYBACK RATE 250.0 VTBI 218.5 mL
mL/hr
3 Enter the new piggyback dose amount and press [
4 Set the infusion time for the secondary container.
5 Press [
Note: You must know the amount that has already been delivered. When the piggyback dose amount is changed, the pump begins a new VTBI countdown.
88 Hospira GemStar Manual 430-00897-A02 (Rev. 02/05)
YES/ENTER] to begin piggyback delivery.
YES/ENTER].
Page 96
430-00897-002.book Page 89 Friday, October 22, 2004 2:38 PM
Canceling a Piggyback Delivery
8. mL/hr Only
1 Press [
2 Press [
3 Press [
CHANGE] and select PIGGYBACKING from the Change Menu.
CHANGE] to reset the dose amount to zero.
YES/ENTER] to cancel piggyback delivery.
Changing a mL/hr Only Program
Can be changed at any time
Delivery rate
Container size
Note: You cannot change a program when the pump is in FULL or CONTAINER Lock.
To change a program:
1 From the
2 As each message displays, make the desired changes or press [
accept the current setting.
3 Press [
Note: A program review is not required and is available only through the Change or Options Menu.
STOP mode, press [CHANGE] and select CHANGE PROGRAM.
START] to begin delivery.
Piggybacking
•VTBI
YES/ENTER] to
Hospira GemStar Manual 430-00897-002 (Rev. 10/04) 89
Page 97
430-00897-002.book Page 90 Friday, October 22, 2004 2:38 PM
8. mL/hr Only
GemStar
®
mL/hr Only Programming Worksheet
GIVE:
Parameter Selections and/or Values
to be Entered
Therapy mL/hr Only N/A
Rate
VTBI
(Volume To Be Infused) 0.1 - 9999 mL
Piggyback Dose Amount
Piggyback Infusion Time
hrs:min
Programmable Ranges
0.1 - 1000 mL/hr
1 - 7200 mL
1 min or dose amount / 300 mL/hr to 24 hrs or dose amount / 0.1 mL/hr
90 Hospira GemStar Manual 430-00897-002 (Rev. 10/04)
Page 98
430-00897-002.book Page 91 Friday, October 22, 2004 2:38 PM
Therapy Features and Specifications
This therapy allows programming of multiple doses at specified times.
Optional KVO (Keep Vein Open) can be set during programming.
The default bolus delivery rate is 400 mL/hr.
Tip: The Programming Worksheet at the end of this chapter lists the minimum and
maximum values allowed for this protocol. Use copies of the programming worksheet to record program values before programming the pump.
Chapter 9
Variable Time
Variable Time Programming Tips
The following table explains how to use the keypad during variable time programming:
Key Function Key Function
[0] - [9] Selects menu items and
enters numeric values.
Accepts entries or continues when done viewing a display.
Displays information during programming and alarm conditions.
Scrolls through program review and help messages. Toggles between AM and PM in time-based entries.The up arrow inserts a decimal point in a numeric value.
Clears a numeric value (resets to zero) during programming.
Returns to a previous display, exits the Options Menu, and exits the
HELP mode.
Hospira GemStar Manual 430-00897-002 (Rev. 10/04) 91
Page 99
430-00897-002.book Page 92 Friday, October 22, 2004 2:38 PM
9. Variable Time
Variable Time Programming
1 Select VARIABLE TIME from the Programming Menu.
2 Select the phase type.
3 Select the unit of delivery.
4 Change the displayed time if necessary.
5 Enter the optional base rate with start and stop time.
6 If programming in Rates or Doses, go to step 7. If programming in Percentages,
set the daily amount. Then go to step 8.
Note: If programming in percentages, the total daily amount of all phases pro-
grammed must equal 100%.
7 Enter the phase information for each phase. Up to 24 phases can be programmed.
Note: When entering time in 12-hour format, press or to toggle between AM and PM.
8 Enter the optional bolus dose and its lockout.
9 Enter the optional KVO.
10 Enter the container size.
The pump calculates the minimum container size required to deliver the phases, base rate, and KVO for one day or in a 24-hour period. If delivering the protocol
for more than one day, multiply this amount by the number of days of therapy.
Note: Increase the container size to include bolus deliveries.
11 Select the air sensitivity, if required.
92 Hospira GemStar Manual 430-00897-002 (Rev. 10/04)
Page 100
VariableTime.fm Page 93 Wednesday, November 10, 2004 8:48 AM
12 Review the program. Press to begin review.
9. Variable Time
13 If using the PURGE key to prime the pump set, press [
WARNING: ALWAYS disconnect the pump set from the patient before
priming or purging.
Note: With variable time programming, the pump keeps track of where it is in the
program each minute of the entire 24-hour period. For example, if Phase 1 is missed, and the pump is placed in
RUN mode in the middle of Phase 2, it starts delivery
according to the program parameters from the middle of Phase 2.
Variable Time Delivery Interruptions
If a phase is interrupted and the pump is placed in the
Before the end of the phase. The phase runs to its scheduled
After the end of the phase, but before the beginning of the next start time.
In the middle of a phase and the phase start was missed because of a program interruption or a missed program start.
RUN mode:
PURGE] now.
Then:
stop time. The next phase begins delivery at
the scheduled start time. All applicable Base Rate or KVO
delivery infuses while waiting for the start of the next phase.
The pump immediately begins delivery of the current phase at its current time-based point.
Hospira GemStar Manual 430-00897-002 (Rev. 10/04) 93
Loading...