Menu Map �������������������������������������������������������������������������������������������������������������16
Installing a Chemcassette cartridge ��������������������������������������������������������������������18
Wiring and tubing �����������������������������������������������������������������������������������������������������20
The SPM Flex gas detector is an extractive gas monitoring system that draws gas
®
samples locally or from a remote point to a Chemcassette
detection system� A wide range of toxic gas Chemcassette cartridges are available
that enable detection of gases used or generated in semiconductor manufacturing
and industrial environments�
The SPM Flex gas detector, available in wall mounted and portable versions,
locally displays gas concentration, alarm, fault and status information via its backlit
color LCD and LEDs� A simple to use 4-button keypad adjacent to the display
provides the ability to set-up, review, operate and make changes to the detector’s
configuration� The intuitive display and menu structure are designed to require
minimal training� The SPM Flex has a local audio alarm with user-configurable
output levels� The detector can be used both indoors and outdoors in a wide range
of weather conditions�
The detector has flexible power and communications capabilities� These include
3 on-board relays, 4-20 mA analog output and Modbus/TCP outputs for signal
and service connectivity� The gas detector is equipped with a USB port for
configuration-sharing firmware updates and data downloads� For web-enabled
devices, web pages are available via the Ethernet port�
tape-based optical gas
Standard operation conditions
The SPM Flex gas detector is designed for use in temperatures between 32ºF
and 104ºF (0ºC and 40ºC) and relative humidities between 0 and 100% (the
relative humidities are limited by tape and calibration)� The sample line will require
additional hardware to remove moisture in high relative humidity conditions where
condensing may occur (the sample must be non-condensing)� Dry conditions may
require humidification�
High-altitude applications
The SPM Flex pump is optimized for operation at altitudes between –1000 feet
(–305 meters) and 3000 ft� (945 m) above sea level� At altitudes above 3,000 ft�
(915 m), up to a maximum of 6,000 ft� (1,830 m), contact Honeywell Analytics for
calibration� (At 6,000 feet, pump capacity is reduced 18% and a flow-system
adjustment to the bypass valve is required� This must be performed by a Honeywell
Analytics certified technician� Contact Customer Service�)
5
SPM Flex Gas Detector
DANGER
!
Safety Information
Danger notices contain information that could prevent death or serious injury�
!
WARNING
Warnings contain information that could prevent injury or equipment damage�
!
CAUTION
Caution notices contain information that could prevent equipment damage�
loop to indicate concentration readings and fault
status�
An easy-to-install case that carries Honeywell’s
Chemcassette tape� Cartridges are specific to
different gas types/families�
Commonly used network technology for wired Local
Area Networks�
The unit is actively monitoring the specified gas or
family of gases
When configured, the SPM Flex will hold the alarm or
fault status active until the user performs an alarm/
fault reset�
Liquid Crystal Display
Light Emitting Diode
Communications protocol running over Ethernet that
is commonly used for communicating with industrial
devices�
An alert in which the SPM Flex will reset
automatically when the condition is no longer
present (i�e�, the detector does not have to be reset
by the user)�
An electrical-operation output switch that can be
used to indicate the presence of alarm and fault
conditions�
Time Weighted Average, the average exposure to a
harmful gas, usually calculated over a period of eight
hours (a the typical workday)�
Universal Serial Bus is an industry standard
communications protocol and bus that is commonly
available on personal computers�
Volts of Direct Current
8
SPM Flex Gas Detector
WARNING
!
Product Overview
The SPM Flex gas detector is available in two configurations, portable and fixed�
The portable model is equipped with a handle assembly and a shoulder strap�
The fixed unit is supplied with a mounting bracket� If replacing an old SPM
detector, a retrofit bracket is available (these are shown on pages 29 and 30)�
1
All units are supplied with a power adaptor
manual, and a printed quick start guide�
Operate and service the SPM Flex gas detector only as specified in this manual
and the accompanying quick start guide� Failure to do so may impair the
protection provided by the detector and may also void the warranty�
Initial setup
Prior to use, the detector requires some minimal setup:
• Unpack the unit (save the packaging to re-use for service requests)
• Read the quick start guide and this manual
• For most gases, install a filter at the unit or at the end of a sample line (see
the Detectable Gases table on page 60)
and cable, a CD containing the user
Portable detectors
• Connect the power adaptor (in a dry, indoor location)
• Charge for at least 4 hours (the detector can be used while being charged)
• Open the detector’s cover
• Remove the rocker switch cover
• Turn the rocker switch to the on position
• Replace the rocker switch cover
• Remove the paper optics card from the gate (save the card for later use)
• Install a chemcassette
• Configure the detector and begin to monitor (see pages 15-16 for basic
display usage)
• See detailed Operation (pages 34-45) and Optional accessories (pages 62-64)
1 FSP Group Model FSP135-AAAN rated 1 5.62A / 24V (UL Listed QQGQ (E190414))
9
SPM Flex Gas Detector
Fixed detectors
• Install the wall-mount bracket and mount the detector to it (see pages 29-30)
• Wire in accordance with local electrical codes utilizing a trained electrician
(see pages 20-26)
• Input
• Honeywell-supplied power adaptor (indoor, dry location use only) or
• 24 VDC power supply
• Signal (as desired)
• 4-20 mA
• Relays
• Ethernet
• Open the detector’s cover
• Remove the rocker switch cover
• Turn the rocker switch to the on position
• Replace the rocker switch cover
• Remove the paper optics card from the gate (save the card for later use)
• Install a Chemcassette cartridge
• Configure the detector and begin to monitor (see pages 15-16 for basic display
usage)
• See detailed Operation (pages 34-45) and Optional accessories (pages 62-64)
Additional accessories
• External dust filters (required on inlet for most gases; see the Detectable Gases
table on page 60)
• Tubing
• Sampling wand
• Mating signal connectors for fixed installations
The connectors and ports of a typical installation are shown in the figure below�
Sample in port
Ethernet
connector
Exhaust out port
Power connector
Relays or 4-20 mA connector
Figure 1. Connectors and ports
10
SPM Flex Gas Detector
Battery operation
The SPM Flex gas detector is supplied with a power supply/charger than can
connect the detector’s power connection and a standard wall outlet� (Ensure that
the correct power cable has been ordered for local operation�) The power supply/
charger is designed for indoor, dry-location use only� The lithium-ion battery pack
provides continuous operation of 6+ hours, depending on sampling conditions
and unit configuration� The battery’s charge time is typically 4 hours� The
detector can be operated while its battery is being charged (this may increase
the charge time)� Because batteries discharge slowly when not in use, Honeywell
Analytics recommends keeping the detector connected to the power supply/
charger when not in use�
!
WARNING
• The battery is not field-replaceable� Return the unit to Honeywell Analytics if
a battery replacement is necessary�
• Risk of fire and burns� Do not open, crush, heat above 140ºF (60ºC), or
incinerate the battery� Follow manufacturer’s instructions�
How to extend battery life
Power consumption is optimized to extend battery life when not in monitor
mode� In addition, the unit should not be exposed to extreme temperatures
which shortens battery life� The time before recharging is required can be
extended by taking the detector out of monitor mode when it is not in use�
How to recharge an overly-discharged battery
If the battery indicates 0% charge and does not appear to be charging when
plugged into the power supply, it may be overly discharged� Follow these steps:
1� Verify that the latest software has been installed� The detector’s software
version can be found by navigating to the
menu� If necessary, the current software can be downloaded from the SPM
Flex section of the Honeywell Analytics website (see page 7)� Install the
new software from a USB flash drive by following the prompts in the
Program
2� Turn the detector off and then power it up again� This will reinitiate the
charging circuit� Initially (until the battery reaches the normal charge range),
the detector will slow charge� The process can take 6 hours�
option on the
Maintenance
Software
menu�
option on the
Review
Update
11
SPM Flex Gas Detector
The handle (portable model)
The detector can be conveniently carried with the optional handle, which is
mounted onto the cover at three locations� If necessary, the handle can be
removed by the user by removing two bolts� All of the detector’s functions
can be performed with the handle attached� The handle swings out of the
way for access to the Chemcassette cartridge area�
Opening the cover
Figure 2. Opening the detector cover
Release the cover by first unscrewing the handle pin� Pivot the handle up as
shown in the illustration and push down the four latches (two on each side)�
The detector cover will then be free to swing open, allowing a Chemcassette
cartridge to be inserted or replaced, the power switch to be turned on or off,
or the USB data port to be accessed�
12
Main power rocker switch
SPM Flex Gas Detector
Rocker switch cover
Main power rocker switch
Figure 3. SPM Flex main power rocker switch
After the detector’s cover is open, use a Phillips screwdriver to remove the
rocker switch cover� Turn the main power rocker switch to the on position and
replace the rocker switch cover�
The detector can now be turned on�
13
SPM Flex Gas Detector
Turning the detector on and off
Press and hold the Power/Cancel button until the green LED begins blinking�
The detector will begin a startup sequence that lasts about 30 seconds�
The four LEDs provide at-a-glance information about the current state of the
detector:
LEDs (red, green, yellow, blue)
Figure 4. SPM Flex LEDs
LEDDescription
Red
GreenblinkingThe detector is active
Yellow
BluesolidConnected to external power
solidAlarm 1
blinkingAlarm 2
solidMaintenance fault
blinkingInstrument fault
To turn the detector off, press and hold the Power/Cancel button for 5
seconds or select “Power Off” from the menu�
14
Navigation
SPM Flex Gas Detector
Display
Power/Cancel
Accept/Select
Figure 5. SPM Flex controls
Up arrow
Down arrow
All SPM Flex menus are navigated by the four buttons shown in Figure 5�
Pressing either of the arrows or the Accept/Select button will access the
menu from the main display screen� The arrows also are used to scroll up and
down through lists of options� The Accept/Select button is used to initiate a
highlighted option� During navigation, the Power/Cancel button will cancel a
command or, when pressed for more than 3 seconds, it will exit to the main
display�
15
SPM Flex Gas Detector
Menu Map
Main Menu
Login/Logout
Reset Alarms & Faults
Monitor Mode
Change Chemcassette
Review Mode
Maintenance
Setup
Test
Power Off
Reset all
Enter
monitor mode
Event
History
Inhibit
General
Optics
Verification
Silence
audio alarm
Out of
monitor mode
Chem-
cassette
4-20 mA
Current Loop
Calibration
MonitoringLatching
Simulation
Settings
Flow Char-
acterization
Force
Relay
Gas
Gate
Force
Additional
Status
Update
Program
Network
Open/Close
OutputsNetwork
4-20 mA
Output
State
Security
Trend/Plot
Configuration
Manager
Software
The SPM Flex detector’s menus are easily navigated� As an example, this is
how a user could review the detector’s gas settings when starting from the In
Monitor screen:
1� Press the Up Arrow to open the Main Menu�
2� Press the Down Arrow 4 times to highlight the Review Mode option�
3� Press the Accept/Select button to enter Review Mode�
4� Press the Up Arrow or Down Arrow until the Gas settings option is
highlighted�
5� Press the Accept/Select button to display the Gas Summary�
To return to the detector to the In Monitor display, press the Power/Cancel
button three times to back out of the Main Menu options�
16
SPM Flex Gas Detector
The display
Figure 6 shows the elements of the display that will be seen in various
situations�
Status bar
Mute
Alarm 1/
Alarm 2/
Inhibit
indicator
Instrument
Maintenance
Fault
Fault
Concentration
trend
indicator
Menu
Monitor mode indicator
or
Duty cycle indicator:
Clock
Battery indicator
title
15:
42:
01
0
123
.
-
H2S
Hydrogen Sulfide
In monitor
Gas name
Gas concentration
2
Menu
ppm
5.678
1.234
options
Bar graph
Units
w/upper, lower
alarm limits
42:
01
15:
Login / Logout
Reset alarms/faults
Monitor mode
Change Chemcassette
Main menu
Highlighted
Scroll bar
option
Monitor mode indicator
or
Duty cycle waiting state:
01:59
min sec
Figure 6. Elements of typical SPM Flex detector displays
The bar graph on the right shows the current concentration relative to the Alarm
1 and Alarm 2 setpoints� (The bar’s range is relative to the alarm levels, not to
the full scale of the selected gas)�
The color of the status bar changes according to the system status (green =
OK/in monitor, yellow = fault, red = alarm, blue = Out of Monitor mode)�
Display and navigation
Under normal operation the LCD and LED display system status, gas
concentrations, and alarms� In set-up, review, calibration, and test modes, the
LCD shows the relevant menu options and system status bar� The interface is
navigated using the four buttons:
ControlFunction
Up arrowScroll up through lists of options
Down arrowScroll down through lists of options
Accept/SelectAccept or select a highlighted option
Power/Cancel
In the main display mode, the display shows the current gas name and
concentration, the monitoring states and faults and/or alarm status, if
applicable�
Power/Cancel will turn the unit on if it’s in the off state, will exit back to the previous menu level, and will
enter Reset mode (quick press), or Power Off mode (long press and hold) from the main display.
17
SPM Flex Gas Detector
Installing a Chemcassette cartridge
Close the Chemcassette cartridge door before putting the detector into
Monitor mode (this holds the Chemcassette cartridge in place)� Take the
detector out of Monitor mode before opening the door�
NOTE
Chemcassette cartridges must be stored according to the manufacturer’s
guidelines when not in use�
Chemcassette
cartridge
Take-up spindle
Encoder
Optics gate card
Figure 7. The optics gate card and Chemcassette cartridge
1. If the detector’s handle is installed, release it by unscrewing the thumbscrew.
For fixed installations, skip to Step 3.
2. Tilt the handle up as far as possible.
3. Unfasten the latches on the sides of the detector (two on each side) that secure
the Chemcassette cartridge access cover.
4. The detector has slotted hinges. Lift up the cover slightly and then open it all
the way to the right.
5. Select the Change Chemcassette menu option and follow the on screen
instructions.
6. If required, remove the old Chemcassette cartridge. When prompted, snap the
new Chemcassette cartridge into place. It will t in only one orientation. The
take-up spindle will rotate slightly to allow the cartridge to snap into place.
7. To close the cover, position it over the base and press so that it contacts the
gasket uniformly.
8. Close the latches on the sides of the detector to secure the cover.
18
SPM Flex Gas Detector
If there is an error (i�e�, the same cartridge was reinserted, the new cartridge
has expired, or an unqualified user is attempting to insert a cartridge that
detects a different gas than the last one), an error message will be displayed
and the installation will be canceled� If the installation is successful, a summary
of the current configuration will be displayed and the user can then choose
whether to enter Monitor Mode or exit the Change Chemcassette Mode out of
monitor�
Removing a Chemcassette cartridge
1. Navigate to Maintenance > Open/Close gate.
2. Select “Open Gate.”
3. Remove the cartridge.
4. Insert the optics gate card.
5. Select “Close Gate.”
Changing the target gas within a gas family
Refer to the Detectable Gases table on page 60 for details about gas families�
1. If necessary, install a Chemcassette cartridge.
2. While out of monitor mode, navigate to Set-up > Monitoring > Gas.
3. Select “Current gas” and pick the desired gas from the selection box
4. Select “Save.”
Selecting a target gas from a different gas family
Refer to the Detectable Gases table on page 60 for details about gas families�
1. Remove the Chemcassette cartridge.
2. Clean the tape-contact surfaces (see page 57).
3. Install the new Chemcassette cartridge.
4. Select the desired target gas from the menu.
Entering monitor mode
From the main menu, navigate to the “Monitor Mode” option and select “Enter
monitor mode�”
1. If a Chemcassette cartridge has been installed, the detector will enter Monitor
Mode.
2. If a Chemcassette cartridge has not been installed, a user with the necessary
passcode can inititate the Chemcassette wizard. Otherwise, the detector will not
enter monitor mode.
19
SPM Flex Gas Detector
CAUTION
!
Wiring and tubing
Typical fixed installation topologies
The SPM Flex gas detector has flexible installation options that allow the
user to select the one most suitable for a specific application� The detector
is supplied with weather-sealed connectors for power, Ethernet, and
communications (for relays or 4-20 mA)� The Ethernet port can be replaced
with an appropriate connection in accordance with local codes (allowing
the user to wire directly to the terminal block)� Install each detector near a
dedicated circuit breaker�
• The safety of any system incorporating the SPM Flex gas detector is
the responsibility of the assembler of the system�
• Position a permanently-installed SPM Flex gas detector so that it does
not interfere with access to the dedicated circuit breaker�
• Use only the power adaptor described in the Specifications section
(see page 61)�
• Use the power adaptor only in indoor applications�
This table shows the default wiring configuration for the IP-rated
communication connector when installed by Honeywell Analytics� The relays
are labeled for the factory default but the configuration can be modified to
have a single alarm and separate faults�
Relay/mA Terminal Connections
Relay/mA Terminal Connections
Front ViewNo.Color
mA SinkmA Source Isolated mA*
Alarm2 NOAlarm2 NOAlarm2 NO
Alarm2 COM Alarm2 COMAlarm2 COM
Alarm1 NOAlarm1 NOAlarm1 NO
Alarm1 COM Alarm1 COMAlarm1 COM
Fault NOFault NOFault NO
Fault COMFault COMFault COM
24 VDC +4-20 mA -4-20 mA -
4-20 mA +24 VDC -4-20 mA +
1
2
3
7
6
8
5
4
*default pin configuration
1Red
2Black
3White
4Green
5Brown
6Blue
7Yellow
8 White/Black
Do not wire relays and 4-20 mA in the same wire bundle�
Electrical connection is made via conduit directly to ports or via the
connector (see Figure 1 on page 10)� The terminals used are suitable for
conductors of 20 to 14 AWG (0�8 to 1�6 mm dia�)�
20
NOTE
SPM Flex Gas Detector
The SPM Flex Ethernet connector is removable and can be replaced by
the customer with a suitable conduit connector� The opening dimensions
are shown in the following illustration� When selecting a conduit connector,
consider its ability to seal to the SPM Flex case, the depth of the thread
(enough to attach to case without interfering with internal components), and
adequate space for the intended wiring� The IP rating can be maintained only
with the original factory connectors�
1.068 in.
(27.13 cm)
Figure 8. Ethernet opening dimensions
Ethernet connector
0.386 in.
(9.81 cm)
P8
P9
+
24 VDC
-
Fault
Relay 2
NO
NC
COM
COM
Relay 1
NO
COM
NC
NC
+
-
4-20 mA
NO
4-20 mA
connections
Input power
connections
Relay
connections
top of
detector
Figure 9. SPM Flex connections
Terminal Module
The terminal module is located on the main PCB inside the gas detector� All
power, relay, and 4-20 mA connections to SPM Flex are made via this module�
Wire entry to the terminal module area is via the cable entry/conduit entry
located at the top of the detector�
21
SPM Flex Gas Detector
Wiring diagrams
Controller
R
L
Signal
Figure 10.
Controller
Signal
R
L
+VE
1
2
-VE
4-20 mA sink wiring diagram
+VE
-VE
3
1
2
Current
Flow
3
SPM Flex
24V+
4-20 mA+
Current
Flow
24V-
SPM Flex
24V+
4-20 mA -
24V-
Figure 11.
Controller
Figure 12.
4-20 mA source wiring diagram
+V1
+V2
-V2
-V1
4-20 mA isolated wiring diagra
SPM Flex
24V+
4-20 mA+
4-20 mA-
24V-
NOTE
Do not wire relays and 4-20 mA in the same wire bundle�
22
Loading...
+ 50 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.