Sensor
Unit
Range
Definition
Breath Rate
Bth/m (breaths per minute)
0 to 120
Heart Rate
Bts/m (beats per minute)
0 to 240
Activity
VMU (Vector Magnitude Units)
±16
<0.2 = Low Activity <0.8 = Walking/Jogging >0.8 = Running
Posture
Degrees
0 to 180
0 = Vertical 90 = Horizontal 180 = Inverted
Temperature
Degrees F or C
50° to 140° F 10° to 60° C
BioHarness 3
QUICK REFERENCE
IMPORTANT! Follow all instructions here before operating the BioHarness 3. Make sure your strap is the correct size before using it (available in two sizes, Small XS-M or Large M-XL). For more details, refer to the BioHarness 3, RAELink3 (or RAELink3 Z1, which has the same functions), and ProRAE Guardian User’s Guides or Quick References.
The BioHarness 3 measures the following parameters:
Operation
1. Moisten the gray sensor pads on the back of the strap with water.
2. Snap the BioHarness Echo Module onto the BioHarness 3 by clicking the bottom in first and then the top.
3. Power on the BioHarness Echo Module by pressing the button in the center until all four LEDs flash.
4. Link the BioHarness Echo Module to the RAELink3.
Note: It links automatically after the initial Bluetooth pairing.
5. Attach the shoulder strap with Velcro in the back and fit the clasp to the loop in the front, and
then do the same with the clasp and loop in the front.
Note: BioHarness Echo Module should be under the left armpit when properly positioned.
1
By Honeywell
Pull tighter or add slack to change length of strap
Tensioner
BioHarness 3
QUICK REFERENCE
6. Adjust the shoulder strap and main strap tension by increasing or decreasing
the length of the loop behind the tensioners.
7. Launch ProRAE Guardian on your computer. Units come online automatically.
Charging The RAELink3
Before using the RAELink3, charge its Li-ion battery. For more details, consult the RAELink3 User’s Guide or Quick Reference.
Charging The BioHarness Echo Module
1. Snap the BioHarness Echo Module into the Charging Cradle and use a USB cable to connect the
Charging Cradle to a computer or the included USB charger.
2. Snap the BioHarness Echo Module into the Charging Cradle and allow it to charge. Typical
charging time for a full charge is 3 hours. A charge of 90% is typically reached in 1 hour. While charging, the BioHarness Echo Module’s orange LED flashes. When fully charged, the orange LED glows continuously.
3. Once it is charged, remove the BioHarness Echo Module from the Charging Cradle.
Pairing The BioHarness Echo Module And The RAELink3 Via Bluetooth
After an initial pairing of the BioHarness Echo Module with the RAELink3, the two will automatically find each other every time they are turned on. For initial pairing, follow the instructions below.
Note: When doing initial pairing, never turn on more than one BioHarness Echo Module and one RAELink3 at the same time.
1. Enter Programming Mode on the RAELink3 by holding the [N/-] and [MODE] keys together.
Press [N/-] twice until you see “OP Mode” (Operation Mode). Set the mode of operation to Remote. Set the BT (Bluetooth) to BioHarness 3. Press [N/-] until you confirm BT (Bluetooth) power is on.