
Product Notice 
Gas Detection Bump Test and Calibration Frequency 
Regions Affected: Global  
This document is intended to provide a general guideline for bump testing and calibration of Honeywell 
Analytics gas detection products. The following gas detector tests should only be conducted in a fresh air 
atmosphere containing 20.9% oxygen and free of any toxic or combustible gas vapors. 
Corporate, local, regional or national regulations, policies and procedures may take pre cedence over these 
recommendations. 
Bump Test 
Function Check: A qualitative function check where a challenge gas is passed over the sensor(s) at a  
concentration and exposure time sufficient to activate all alarm indicators to a minimum of their lower alarm 
setting. The purpose of this check is not to confirm the accuracy of sensor response, just that gas can get 
to the sensor(s) and that all the alarms present are functional.  
Accuracy Check: A quantitative function check using a known concentration of gas that is traceable to a 
recognized standard to confirm that the alarms are functional and that the sensor(s) response is within 
acceptable limits. Typically, a +/-10% to +/-20% of applied gas value is considered acceptable unless 
otherwise specified by the manufacturer, internal company policy, or a regulatory agency. 
Calibration 
Zero calibration: Establishes sensor(s) baseline reading, 20.9% oxygen and free of any toxic o r combustible 
gas vapors. To maintain a stable baseline, it is essential to perform zero calibration in fresh air. In the 
event, fresh air cannot be confirmed, use zero grade air or N 
a zero calibration to reestablish baseline as required, and before a span calibration. 
Span calibration: The adjustment of the sensor(s) response to match the value of a known concentration 
of applied gas that is traceable to a recognized standard. The calibration procedure should be done in 
accordance with the equipment manufacturer’s instructions. 
NOTE: Consult the product manual for product specific recommendations. 
Recommended Frequency 
Fixed gas monitors: 
  *Perform an Accuracy Check on a 4 to 6-month basis based on region at a minimum 
o Consult local regional codes for proper bump test intervals 
  Perform a Zero calibration and *Span calibration prior to first time use and then on an annual basis 
at a minimum 
o **Factory calibrated sensors may not require a Span calibration at first time use  
o Infrared (sealed construction only) and ultrasonic based products typically do not require 
span calibrations and should be avoided where possible 
Direct Reading Portable Gas Monitors: 
  Perform a Function or Accuracy check prior to each day’s use at a minimum. 
 Perform a Zero and Span Calibration prior to first time use. Calibration frequency can vary from 
every 30 to 180 days, or, calibrate following a failed bump test. 
Non-direct Reading Portable Gas Monitors: 
 to perform a detector zero calibration. Perform 
2
BULLETIN NUMBER:  2017115-2  6/19/2018 

Product Notice 
  Perform a Function check periodically 
  Perform a Zero Calibration as required; Span Calibration is typically not required over the working 
life of the monitor. 
Flame Detection Testing 
Electro-optical flame detectors: 
 Perform a lamp test on a semi-annual basis. Perform a complete fire protection system function 
test annually.  
Information/Reference Materials 
  Honeywell Analytics:  
o http://www.honeywellanalytics.com/  
o http://www.raesystems.com/customer-care/resource-center/ebooks  
o http://www.raesystems.com/customer-care/resource-center/tech-notes  
o http://www.raesystems.com/sites/default/files/content/resources/pid_handbook_1002-
02.pdf 
  International Safety Equipment Association (ISEA): 
o https://safetyequipment.org/product-groups/instruments/ 
  Occupational Safety and Health Administration (OSHA): 
o https://www.osha.gov/dts/shib/ 
FactoryMutual(FMClass3260): 
 
o http://www.fmapprovals.com/products-we-certify/understanding-the-benefits/fm-
approved-flame-detection 
*Corporate, local, regional or national regulations, policies and procedures may take precedence over the se 
recommendations. 
**Factory calibrated sensors are calibrated to NIOSH standards and may not require span calibrations at 
initial use. 
 Zero calibrations on fixed instruments is recommended at time of start-up to ensure the best 
possible operational results. 
 Allow for a minimum of 15-minute warm-up period to allow for the sensor to acclimate to the 
environment (i.e. Altitude, humidity and temperature).   
BULLETIN NUMBER:  2017115-2  6/19/2018