BW Technologies LP (BW) warrants the product to be free from defects in material and workmanship under normal use and service for a period of two years,
beginning on the date of shipment to the buyer. This warranty extends only to the sale of new and unused products to the original buyer. BW’s warranty obligation is
limited, at BW’s option, to refund of the purchase price, repair or replacement of a defective product that is returned to a BW authorized service center within the
warranty period. In no event shall BW’s liability hereunder exceed the purchase price actually paid by the buyer for the Product.
This warranty does not include:
a) fuses, disposable batteries or the routine replacement of parts due to the normal wear and tear of the product arising from use;
b) any product which in BW’s opinion, has been misused, altered, neglected or damaged, by accident or abnormal conditions of operation, handling or use;
c) any damage or defects attributable to repair of the product by any person other than an authorized dealer, or the installation of unapproved parts on the
product; or
The obligations set forth in this warranty are conditional on:
a) proper storage, installation, calibration, use, maintenance and compliance with the product manual instructions and any other applicable recommendations of
BW;
b) the buyer promptly notifying BW of any defect and, if required, promptly making the product available for correction. No goods shall be returned to BW until
receipt by the buyer of shipping instructions from BW; and
c) the right of BW to require that the buyer provide proof of purchase such as the original invoice, bill of sale or packing slip to establish that the product is within
Since some countries or states do not allow limitation of the term of an implied warranty, or exclusion or limitation of incidental or consequential damages, the
limitations and exclusions of this warranty may not apply to every buyer. If any provision of this warranty is held invalid or unenforceable by a court of competent
jurisdiction, such holding will not affect the validity or enforceability of any other provision.
Sensor Poisons and Contaminants .................................................................................................................................. 6
Getting Started .................................................................................................................................................................... 8
Parts of the GasAlertMicro 5/PID/IR .................................................................................................................................. 9
Screen Elements ............................................................................................................................................................... 10
Battery Test .............................................................................................................................................................. 12
Pump Test ................................................................................................................................................................ 15
Force Calibration Enabled ........................................................................................................................................ 17
Self Test Fail.................................................................................................................................................................. 19
Battery Test ................................................................................................................................................................... 20
Deactivating the Detector................................................................................................................................................. 21
User Options Menu ........................................................................................................................................................... 21
Exit User Options Menu ................................................................................................................................................. 22
Span Gas Value........................................................................................................................................................ 27
STEL Period ............................................................................................................................................................. 28
TWA Method ............................................................................................................................................................. 28
Language Selection ....................................................................................................................................................... 34
Force Calibration ...................................................................................................................................................... 37
Gas Exposures Computed ............................................................................................................................................. 43
Viewing Gas Exposures................................................................................................................................................. 43
Clearing Gas Exposures ................................................................................................................................................ 44
Gas Alarm Setpoints ...................................................................................................................................................... 44
Viewing the Alarm Setpoints .......................................................................................................................................... 44
Resetting Gas Alarm Setpoints ...................................................................................................................................... 45
Stopping a Gas Alarm .................................................................................................................................................... 46
Performing a Bump Test ................................................................................................................................................ 47
Calibration and Setting Alarm Setpoints ........................................................................................................................ 48
Applying Gas to the Sensors ......................................................................................................................................... 50
Single Gas Calibration Cap ...................................................................................................................................... 50
Auto Span ................................................................................................................................................................. 54
Setting the Calibration Due Date .............................................................................................................................. 57
Identifying the Pump ...................................................................................................................................................... 64
Generation 1: 116885-L3 (yellow) and 118933-L3 (black)........................................................................................ 64
Generation 2: 130916-L3 (yellow) and 130917-L3 (black)........................................................................................ 65
Installing the Pump Module ............................................................................................................................................. 66
Replacing the Pump Filter (Generation 2 Pump) ........................................................................................................... 67
Replacing the Pump Nozzle (Generation 2 Pump) ........................................................................................................ 68
Attaching the Auxiliary Filter .......................................................................................................................................... 68
Confined Space Sampling ............................................................................................................................................. 70
Maximum Hose Length when Confined Space Sampling......................................................................................... 71
Reformatting the MMC/SD Card .................................................................................................................................... 75
Import Datalogs to Fleet Manager II ................................................................................................................................ 76
Minimum PC Requirements ........................................................................................................................................... 76
Importing from MicroDock II to Fleet Manager II............................................................................................................ 76
Import to Fleet Manager II Using a Card Reader ........................................................................................................... 76
View Datalog Files in Spreadsheets ................................................................................................................................ 77
Example of a Datalog Spreadsheet .......................................................................................................................... 78
Charging the Battery ...................................................................................................................................................... 81
Replacing the Alkaline Batteries .................................................................................................................................... 82
Replacing the Lithium Battery Pack ............................................................................................................................... 83
Replacing a Sensor or Sensor Filter .............................................................................................................................. 83
Clean or Replace the Lamp ...................................................................................................................................... 85
Replace the Lamp..................................................................................................................................................... 86
Replace the Electrode Stack .................................................................................................................................... 87
Removal and Disposal of the Battery Pack............................................................................................................... 87
Removal and Disposal of the Coin Cell .................................................................................................................... 87
Replacement Parts and Accessories .............................................................................................................................. 95
General Specifications for Datalogger Units ................................................................................................................ 100
1.Parts of the GasAlertMicro 5/PID/IR...................................................................................................................... 9
2.Screen Elements ................................................................................................................................................. 10
3.Applying Gas to the Sensors ............................................................................................................................... 50
4.Single Gas Calibration Cap ................................................................................................................................. 51
5.Removing the Single Gas Calibration Cap .......................................................................................................... 51
6.Installing the Pump Module ................................................................................................................................. 66
7.Replacing the Pump Filter (Generation 2 Pump)................................................................................................. 67
8.Replacing the Pump Nozzle ................................................................................................................................ 68
9. Attaching the Auxiliary Filter ............................................................................................................................... 68
10. Attaching the Filter Cord..................................................................................................................................... 69
11.Attaching the Sample Probe................................................................................................................................ 70
12.Inserting/Removing the MMC/SD Card ............................................................................................................... 73
13.Replacing the Alkaline Batteries.......................................................................................................................... 82
14.Replacing the Lithium Battery Pack .................................................................................................................... 83
15.Replacing a Sensor or Sensor Filter ................................................................................................................... 84
16.Parts of the PID ................................................................................................................................................... 86
4.Parts of the GasAlertMicro 5/PID/IR...................................................................................................................... 9
5.Screen Elements ................................................................................................................................................. 10
8.Computed Gas Exposures .................................................................................................................................. 43
11.Applying Gas to the Sensors ............................................................................................................................... 50
12.Single Gas Calibration Cap ................................................................................................................................. 51
13.Time Required to Span ....................................................................................................................................... 55
14.Installing the Pump Module ................................................................................................................................. 66
15.Generation 1 Pump Maximum Hose Length ....................................................................................................... 71
16.Generation 2 Pump Maximum Hose Length ....................................................................................................... 71
18.Datalog Status Codes ......................................................................................................................................... 79
19.Datalog Gas and Correction Factor Sensor Codes ............................................................................................. 80
20.Replacing the Alkaline Batteries.......................................................................................................................... 82
21.Replacing a Sensor or Sensor Filter .................................................................................................................. 85
22.Parts of the PID sensor ....................................................................................................................................... 85
24.Replacement Parts and Accessories .................................................................................................................. 95
To ensure personal safety, read the Safety Information -
Read First and Cautions before using the detector.
The GasAlertMicro 5, GasAlertMicro 5 PID, and GasAlertMicro 5 IR
gas detectors (“the detector”) warn of hazardous gas at levels above
user-defined alarm setpoints.
The detector is a personal safety device. It is your responsibility to
respond properly to the alarm.
Note
Unless reference is made to a specific detecto r model, the
GasAlertMicro 5, GasAlertMicro 5 PID, and GasAlertMicro 5 IR
detectors are referred to as GasAlertMicro 5/PID/IR.
The following table lists the gases that are monitored by the detector.
Table 1. Gases Monitored
Gas DetectedUnit of Measure
Oxygen (O2)
Combustible gases (LEL)
Carbon monoxide (CO)parts per million (ppm)
Hydrogen sulfide (H
Phosphine (PH
Sulfur oxide (SO
Chlorine (Cl
Ammonia (NH
Nitrogen dioxide (NO
Hydrogen cyanide (HCN)parts per million (ppm)
Chlorine dioxide (ClO
Ozone (O
Volatile organic compounds
(VOC)
Carbon dioxide (CO
)
2
)
3
S)
2
)
3
)
2
)
3
)
2
)
2
)parts per million or %vol CO
2
% volume
a) percent of lower explosive
limit (%LEL)
b) percent by volume
methane 0-5.0% v/v
parts per million (ppm)
parts per million (ppm)
parts per million (ppm)
parts per million (ppm)
parts per million (ppm)
parts per million (ppm)
parts per million (ppm)
parts per million (ppm)
parts per million (ppm)
CAUTION: FOR SAFETY REASONS, THIS EQUIPMENT MUST BE
OPERATED AND SERVICED BY QUALIFIED PERSONNEL ONLY.
READ AND UNDERSTAND THIS USER MANUAL COMPLETELY
BEFORE OPERATING AND SERVICING.
Safety Information - Read First
Use the detector only as specified in this user manual, otherwise the
protection provided by the detector may be impaired.
International symbols used on the detector and in this user manual are
defined in Table 3.
Read the Cautions on the following pages before using the detector.
ec Warning
This instrument contains batteries. Do not mix with the
solid waste stream. Spent batteries must be disposed of
by a qualified recycler or hazardous materials handler.
Dispose of lithium cells immediately. Do not
disassemble and do not dispose of in fire. Do not mix
with the solid waste stream. Spent batteries must be
disposed of by a qualified recycler or hazardous
materials handler.
• Warning: Substitution of components may impair Intrinsic
Safety.
• Caution: For safety reasons, this equipment must be
operated and serviced by qualified personnel only. Read
and understand this user manual completely before
operating or servicing.
• Do not use the detector if it is damaged. Inspect the
detector before using. Look for cracks and/or missing parts.
• If the detector is damaged or parts are missing, contact BW
Technologies by Honeywell immediately.
• Use only sensor(s) that are specifically designed for the
GasAlertMicro 5/PID/IR detectors. Refer to Replacement
Parts and Accessories.
• Calibrate the detector before first-time use and then on a
regular schedule, depending on use and sensor exposure
to poisons and contaminants. Sensors must be calibrated
regularly and at least once every 180 days (6 months).
• BW recommends to bump test the sensors, before each
day’s use, to confirm their ability to respond to gas by
exposing the detector to a gas concentration that exceeds
the alarm setpoints. Manually verify that the audible and
visual alarms are activated. Calibrate if the readings are not
within the specified limits.
• BW recommends the combustible sensor be checked with a
known concentration of calibration gas after any known
exposure to catalyst contaminants/poisons (sulfur
compounds, silicon vapors, halogenated compounds, etc).
• The combustible sensor is factory calibrated to 50% LEL
methane. If monitoring a different combustible gas in the %
LEL range, calibrate the sensor using the appropriate gas.
• Warning: High off-scale LEL readings may indicate an
explosive concentration.
• Only the combustible gas detection portion of this
instrument has been assessed for performance by CSA
International.
• Protect the combustible sensor from exposure to lead
compounds, silicones, and chlorinated hydrocarbons.
Although certain organic vapors (such as leaded gasoline
and halogenated hydrocarbons) may temporarily inhibit
sensor performance, in most cases, the sensor will recover
after calibration.
aCautions
• Before using common products around sensors, refer to
Sensor Poisons and Contaminants
• High concentrations of certain toxic gases, for example
H2S, may have an adverse effect on the LEL sensor. This
effect, known as inhibition, is usually temporary but in
extreme circumstances can impair the sensitivity of the LEL
sensor.
After any gas exposure that causes an alarm in the toxic
gas sensors, the LEL sensor should be verified with a bump
test, and recalibrated if necessary.
• Any rapid up-scaling reading followed by a declining or
erratic reading may indicate a gas concentration beyond
upper scale limit, which may be hazardous.
• Calibrate only in a safe area that is free of hazardous gas, in
an atmosphere of 20.9% oxygen.
• Use only BW approved batteries for the GasAlertMicro 5/
PID/IR detectors. Refer to Replacement Parts and
Accessories.
• Charge the detector before first-time use. BW recommends
the detector be charged after every workday.
• Charge the battery pack immediately when a low battery
alarm occurs.
• Read and adhere to the battery cautions provided in Battery
Cautions.
• Charge the GasAlertMicro 5/PID/IR batteries using the
recommended charging adapter only. Do not use any other
charging adapter. Failure to adhere to this caution can lead
to fire and/or explosion.
• Read and adhere to all instructions in the charger user
manual. Failure to do so can result in fire, electrical shock,
personal injury, and/or property damage.
• Extended exposure of the GasAlertMicro 5/PID/IR detectors
to certain concentrations of combustible gases and air may
stress a detector element, which can seriously affect its
performance. If an alarm occurs due to high concentration
of combustible gases, recalibration should be performed, or
if needed, the sensor replaced.
aCautions
• Protect the PID sensor from exposure to silicone vapors.
• When calibrating O
the Toxic 2 sensor position, a single gas calibration cap
must be used to ensure accurate calibration. For more
information, refer to Single Gas Calibration Cap
• Replace the CO
area that is free of hazardous gas.
• The optional pump (M5-PUMP) is certified for use with the
GasAlertMicro 5/PID/IR detectors only.
• Do not immerse the detector in liquids.
• Do not test the combustible sensor’s response with a
butane cigarette lighter; doing so will damage the sensor.
• Do not expose the detector to electrical shock or severe
continuous mechanical shock.
• Do not attempt to disassemble, adjust, or service the
detector unless instructions for that procedure are provided
in the user manual and/or that part is listed as a
replacement part. Use only BW Technologies by Honeywell
Replacement Parts and Accessories
• The detector warranty is void if customer, personnel, or
third parties damage the detector during repair attempts.
Repair attempts made by non-BW Technologies by
Honeywell repair/service personnel voids this warranty.
Lithium Battery Packs
• Warning: The lithium battery (M5-BAT08) may present a risk
of fire or chemical burn hazard if misused. Do not
disassemble,
heat above 100°C (212°F), or incinerate.
• Do not use any other lithium batteries with the
GasAlertMicro 5/PID/IR detectors. Use of any other cell can
cause fire and/or explosion. To order and replace the M5BAT08 battery, refer to Replacement Parts and Accessories.
• Warning: Lithium polymer cells exposed to heat at 130°C
(266°F) for 10 minutes can cause fire and/or explosion.
• Dispose of used lithium cells immediately. Do not
disassemble and do not dispose of in fire. Do not mix with
the solid waste stream. Spent batteries must be disposed
of by a qualified recycler or hazardous materials handler.
• Keep lithium cells away from children.
• Products may contain materials that are regulated for
transportation under domestic and international dangerous
goods regulations. Return product in compliance with
appropriate dangerous goods regulations. Contact freight
carrier for further instructions.
Several cleaners, solvents, and lubricants can contaminate and cause
permanent damage to sensors. Before using cleaners, solvents, and
lubricants in close proximity to the detector sensors, read and adhere
to the following caution and table.
a Caution
Use only the following BW Technologies by Honeywell
recommended products and procedures:
• Use water based cleaners.
• Use non-alcohol based cleaners.
• Clean the exterior with a soft, damp cloth.
• Do not use soaps, polishes, or solvents.
The following table lists common products to avoid using around
sensors.
The list below provides the standard items included with the detector.
If the detector is damaged or parts are missing, contact the place of
purchase immediately.
• Batteries: Three replaceable alkaline cells with battery pack, or
one rechargeable lithium battery pack
• Sensors: O
PID, or CO
• Calibration cap and hose
• Single gas calibration cap
• Screwdriver
• Quick reference guide
• Quick reference card
• Technical Reference Guide and training on CD-ROM
• Fleet Manager II (if applicable)
To order parts, refer to Replacement Parts and Accessories
If enabled, the backlight automatically activates for 8 seconds
when there is an alarm condition and whenever there is
insufficient light to view the LCD. Any but ton reactivates the
backlight in low light conditions.
Table 5. Screen Elements
ItemDescription
1Alarm condition
2Automatically span sensor
3Gas cylinder
4Gas type
5Battery life indicator
6Passcode lock
7Data transmission
8Clock
9Stealth mode
10Pump indicator (optional)
11MMC indicator (optional)
Alarm condition (low, high, TWA,
12
13Automatically zero sensor
STEL, or multi alarm) or view TWA,
STEL and peak (MAX) gas exposures
If using the pump module, attach it and the pump accessories prior to
activating the detector.
For illustrations and procedures, refer to the following:
.
•
• Confined Space Sampling
• Replacing the Pump Filter (Generation 2 Pump)
• Replacing the Pump Nozzle (Generation 2 Pump)
• Attaching the Auxiliary Filter
Note
Only activate the detector in a safe atmosphere that is free of
hazardous gas in an atmosphere of 20.9% oxygen.
To activate the detector, press A.
Self-Tests
When the detector is activated, it performs several self-tests. Confirm
the following tests occur.
Note
If an error message displays during the self-test, refer to
Troubleshooting
.
Battery Test
The detector performs a battery test during startup. If the battery has
insufficient power to operate, the following screen displays before
deactivating.
Replace the batteries and reactivate the detector.
1. All of the LCD elements display simultaneously as the detector beeps, flashes, vibrates, and briefly activates the backlight.
If the pump module is installed on the detector, the
following three things must occur during start-up. If any
one of the conditions below does not occur, discontinue
use of the detector and contact BW Technologies by
Honeywell immediately.
• The detector prompts for a pump test during start-up
• The pump module passes the pump test at start-up when
the pump inlet or sample chain inlet is blocked
•The J icon displays on the LCD
8. If the pump module is attached to the detector, the following
screens display.
GasAlertMicro 5/PID/IR
Activating the Detector
When the following screen displays, block the pump inlet.
Successful Bump Test: If the bump test passes, the following screens
display.
Bump Daily Enabled
a Caution
BW recommends that a bump test to all sensors be
performed every 24 hours prior to the beginning of the
work shift.
13. If Bmp Daily (bump daily) is enabled in tech mode, the follow-
ing screens display.
If a bump test of the LEL and O
detector will deactivate.
Apply the LEL gas and then apply the O
percentage than the default 20.9%, such as 18% O
18
sensor is not performed, the
2
(a higher or lower
2
).
2
The detector waits for the sensor(s) to clear (30 seconds) and then
enters normal operation.
Unsuccessful Bump Test: If the bump test is unsuccessful or the bump
test is not performed, the following screen displays and the detector
deactivates.
If additional sensors require a bump test but are not mandatory, the
following screens display.
Press C Yes to accept and proceed to normal operation.
Or
If A No is pressed, or no buttons are pressed, the sensor(s) that is
past due displays as FAI L when the detector enters normal operation.
In the following screen example, only the SO
operational.
The self-test is now complete.
, CO, and O2 sensors are
2
Self-Test Pass
If the detector passes the self-test, it enters normal operation and
displays the ambient gas readings.
The detector begins recording the peak (MAX) gas exposure and calculating the short-term exposure level (STEL) and time-weighted average
(TWA) exposures.
Self Test Fail
If a sensor fails, FAIL flashes above that sensor on the normal operation
screen. For possible problems and solutions, refer to Troubleshooting
The batteries are tested when the detector is activated and continuously
thereafter. The battery power icon displays continually during normal
operation. If battery power is low, flashes.
Datalogger Operation (Optional)
a Caution
Do not remove the battery pack while the detector is
activated. Doing so will prevent the datalogger from
logging correctly.
Datalogger operation is automatic and requires no settings. During
normal operation the card is tested every 20 seconds.
Note
The MMC/SD card icon (S) displays continuously on the LCD
when the card is inserted. The card is not required for operation; however, if the card is not inserted the detector will not
record data.
The Backlght (backlight) option enables
the LCD backlight to activate automatically
in low-light conditions.
If disabled, the backlight activates only
when the detector is in alarm mode.
Note
In stealth mode, the backlight does
not activate.
The detector is shipped with the Backlght
option enabled.
Confidence Beep
The Confibeep (confidence beep) option
provides continuous confirmation that the
detector is operating properly. When confidence beep is enabled, the audible alarm
beeps once every 10 seconds.
Note
Confidence beep is automatically
disabled during a low battery alarm.
The detector is shipped with the
Confibeep option disabled.
Due-Lock
If the Due-lock (calibration user lockout)
option is enabled and a sensor is overdue
for calibration upon startup, the passcode
must be entered to access normal operation.
Note
If the correct passcode is not
entered, the detector deactivates.
The detector is shipped with the Due-lock
option disabled.
If enabled, the Latch (latched alarms)
option causes the low and high gas alarms
(audible, visual, and vibrator) to persist
until they are acknowledgedby pressing C
and the gas concentration is below the low
alarm setpoint. The peak concentrations display continually until the alarm condition no
longer exists. Local regulations may require
Latch option be enabled.
The detector is shipped with the Latch
option disabled.
Passcode Protect
The Passcode option prevents unauthorized
access to the user options menu, the calibration function, and to adjusting the alarm
setpoints.
Note
The passcode is provided separately
in the supplementary booklet.
If passcode protect is enabled and the Enter
passcode: 1000 screen displays, press G or
H to scroll to the correct passcode and then
press C to confirm.
The detector is shipped with the Passcode
option disabled.
If an incorrect passcode is entered or C is not
pressed within 5 seconds to confirm the correct
passcode, Passcode incorrect displays. The
alarm beeps three times and the detector
either
• resumes normal operation
(for user options), or
• deactivates
(for calibration if Due Lock is enabled).
When enabled, the Safe option confirms that
conditions are normal and there are no gas
hazards present. When all gas levels are
normal or below the alarm setpoints, Safe
displays continually on the LCD.
Safe does not display if any active sensor has
failed, if any active sensor has an overdue calibration, and/or if any active sensor has failed
a bump test.
The detector is shipped with the Safe option
disabled.
The Sensor option provides access to additional options and functions
for each sensor.
Depending upon the sensor that is selected, some or all of the following
options are available for configuration:
• Enabling/disabling a sensor
• Setting the span gas value
• Adjusting the STEL period (not applicable to LEL and O
• Selecting the TWA method (not applicable to LEL and O
• Resolution setting (not applicable to CO, LEL, O
sensors)
•% vol CO
•% vol CH
• Selecting the correction factor (LEL and PID sensors only)
1. From the option menu screen, scroll to Sensors and press
C to access the following screen.
2. Press G or H to scroll to the desired sensor. Press C to con-
firm and to access the menu options specific to the selected
sensor.
For all sensor options, if a value is changed but not
confirmed within 5 seconds, the detector emits an
audible alarm and displays the following error message.
The detector retains the previous setting and returns to the
user options menu.
Sensor Enable/Disable
a Warning
Disabling a sensor should only be performed with
extreme caution. The disabled sensor cannot measure
and alarm against the applicable gas.
If a sensor fails, disabling the sensor deactivates the fail alarm. The
sensor should be replaced and enabled as soon as possible. The
detector will function normally with the remaining enabled sensors.
Note
Detectors that are configured for a 1, 2, 3, or 4 gases may contain a dummy sensor in one of the four sensor locations.
After selecting the desired sensor, the following screen displays.
Press C to toggle between enable/disable (sensor can be enabled at
any time).
If disabled, the readings and the gas type for the sensor do not display
when in normal operation.
If a sensor is enabled but it is not installed in the detector, FAIL flashes
above the gas type of the missing sensor.
DisabledEnabled/not installed
If all the sensors are disabled, the following screens display.
Enable one or more sensors to exit and access normal operation.
Span Gas Value
The Span gas option increases or
decreases the gas concentration for
calibration (must match the gas
concentration listed on the gas cylinder).
After selecting the sensor, press H to
scroll to Span gas and press C within
20 seconds to confirm.
Depending upon the sensor selected, a
screen similar to A or B displays. Press
G or H to scroll to the desired value and
press C within 5 seconds to confirm.
If C is not pressed within 5 seconds to
confirm the new value, the detector
retains the previous value and returns to
the user options menu.
Note
BW recommends that span concentration
values be set between specific ranges.
Refer to the Resetting Gas Alarm
The short-term exposure limit (STEL period) option provides
protection for workers from over exposure to high concentrations
of gas, and is based on 5-15 minute intervals.
The STEL period option is available for
every toxic sensor.
After selecting the desired sensor, press
H to scroll to STEL period and press C
within 20 seconds to confirm.
The STEL period can be set from 5 to
15 minutes. Press G or H to scroll to the
required value, and then press C within
5 seconds to confirm.
If C is not pressed within 5 seconds to
confirm the new value, the detector
retains the previous value and returns to
the user options menu.
The detector is shipped with the STELperiod set to 15 minutes.
The TWAmethod (time-weighted average)
option is a safety measure used to calculate
accumulated averages of gases to notify the
user when the maximum average is
accumulated. The options are the OSHA or
ACGIH calculating method.
OSHA Method: 8 hour moving average—oldest value (first hour) is
replaced by the newest value (ninth hour).
ACGIH Method: Infinite accumulated average to 8 hours—total
accumulation, whether it is 2 hours or 8 hours.
After selecting the desired sensor, press H to scroll to TWA method.
Press C within 20 seconds to confirm.
A check displays in the checkbox of the
currently selected method. To select the
other method, press H to move the check
to the other method. Press C to confirm the
selection.
The detector is shipped with the OSHA
method enabled.
Note
If the TWA method has been changed, the TWA, STEL, and
MAX values must be reset to ensure the TWA is calculated
correctly. Refer to Clearing Gas Exposures
The Resolution option displays the gas
measurement using Regular or Extra
resolution.
Regular: Displays gas measurement in
1 ppm increments.
Extra: Displays the gas measurement in
0.1 ppm increments.
Note
Regular resolution for O
extra resolution is 0.01 ppm.
The Resolution option is not available for CO, O
and CO
After selecting the desired sensor, press H to scroll to Resolution.
Press C within 20 seconds to confirm.
A check displays in the checkbox of the
currently selected resolution. To select the
other resolution, press H to move the
check to other resolution. Press C to
confirm the selection.
The detector is shipped with Regular
resolution enabled for applicable sensors.
sensors.
2
and ClO2 sensors is 0.1 ppm, while
3
, LEL, PID,
2
%Vol CO2 (CO2 Sensors Only)
If the %volCO2 is enabled, the detector
displays the carbon dioxide (CO
readings as %vol (0.0).
From the Sensors option menu, select
CO
.
2
Press H to scroll to %vol CO
to toggle between enable and disable.
Confirmation is not required. If no buttons
are pushed, after 20 seconds the detector
returns to the sensor selection screen.
The change is saved automatically.
User Manual
%Vol CH4 (LEL Sensors Only)Correction Factor (CF)
If %vol CH
enabled correction factor is ignored and
the detector operates assuming a
methane (CH
From the Sensors menu, select LEL and
press C to confirm.
Next, press H to scroll to %vol CH
press C within 20 seconds to confirm.
Press C to toggle between enable and
disable.
Confirmation is not required. If no buttons
are pushed within 20 seconds, the
detector returns to the sensor selection
screen. The change is saved
automatically.
If changing the measurement unit
from % LEL to % Vol. or from % Vol.
to % LEL, a calibration must be completed and the alarm setpoints
changed. For calibration information
and for alarm setpoint information
refer to
Setpoints on page 48.
is enabled, any currently
4
) calibration.
4
and
4
Note
Calibration and Setting Alarm
Depending upon the selected sensor, refer to the following sections
This option is used to enter compensation
factors for hydrocarbons other than
methane. The factor can only be applied if
the LEL sensor has been calibrated with
methane.
After selecting the LEL sensor, press H
to scroll to Correction. Press C within
20 seconds to confirm and access the LEL
correction library.
Scroll to the required gas type and
press C. A check displays in the
corresponding checkbox. The detector
automatically applies the correction factor.
To disable the Correction option, press H
to scroll to None or to Methane. A check
displays in the corresponding checkbox. If
required, select a different gas type
correction factor.
Custom: To enter a correction factor that
is not listed in the library, press H to scroll
to Custom and press C within 5 seconds
to confirm.
The Custom LEL correction screen displays. Press G or H to select the
required value, and press C within 5 seconds to confirm.
PID Sensor
This option is used to enter compensation
factors for selected gas types. The factor
can only be applied if the PID sensor has
been calibrated with isobutylene.
After selecting the PID sensor, press H
to scroll to Correction. Press C within
20 seconds to confirm and access the
PID correction library.
Scroll to the required gas type and
press C. A check displays in the
corresponding checkbox. The detector
automatically applies the correction factor.
To disable the Correction option,
press H to scroll to None or to Isobutyl.
A check displays. If required, select a
different gas type correction factor.
Custom: To enter a correction factor for
a custom PID sensor, scroll to Custom
and press C. Press G or H to scroll to
the required value, and press C within
5 seconds to confirm. Refer to the PID
Correction Factor PID Correction Factor
The PID sensor is not sensitive enough to detect VOCs Benzene, Butadiene, and Vynil Chloride before they exceed their
toxic threshold limit value. The MicroDock II is still able to set
the Micro 5PID/IR to one of these VOCs. In this case the
Micro5/PID/IR will warn that th e selection is not supported and
ask if you wish to continue. If the warning is not acknowl edged
the Micro 5/PID/IR will fail the PID Sensor until you reboot and
accept, or choose a different VOC.
Automatic Oxygen (O2) Calibration
When the Autocal option is
enabled, it forces the detector to
automatically calibrate the
oxygen sensor during startup.
If the Autocal option is enabled,
ensure the detector is activated
in a safe area that is free of
hazardous gas in an atmosphere
of 20.9% oxygen.
From the Sensor menu, press H
to scroll to O
within 20 seconds to confirm.
Press H to scroll to Autocal.
Press C
to toggle between enable/
disable.
The detector is shipped with the
Autocal option enabled.
The screen displays showing the month highlighted indicating it is
selected to set.
Press G or H to scroll to the desired month and press C within
20 seconds to confirm. To bypass and retain the current setting, press
C. Continue setting/bypassing the remaining options.
The date/time options are set in the following order:
•month
•day
•year
•hour
• minutes
When the settings have been set or bypassed, the detector beeps twice
and returns to the user options menu.
Note
The detector is shipped with the date and time set to Mountain
Standard Time (MST).
Language Selection
The detector is shipped with English selected as the default language.
The available languages are as follows:
• Français (French)
• Deutsch (German)
• Español (Spanish)
• Prtuguês (Portuguese)
Press H to scroll to Language and press C within 20 seconds to
confirm.
Press G or H to scroll to the desired language and press C. A check
displays in the checkbox of the selected language.
Wait for 20 seconds until the detector returns to the user options menu,
or press G to scroll to Back (English), Retour (French), Zurück
(German), Regreso (Spanish), or Retornar (Portuguese).
All screens now display in the selected language.
Tech Mode
a Warning
Tech mode should only be accessed by authorized
personnel.
Tec h mo de can only be accessed from the Language option. Press H
to scroll to Language. Do not press C until instructed.
In the following order, press and continue to hold each button until
Tech mode displays below the Language option.
1. Press and hold H with the right index finger.
2. Press and hold G with the right middle finger.
3. Press and hold C with the left thumb.
4. Press C to enter Tech m ode. The options are as
To reconfigure the sensor type, physically change the
sensor prior to entering Tech mode.
When a toxic sensor is physically removed and replaced by another toxic
sensor, the detector must be reconfigured to recognize the change.
Note
If a sensor is replaced, the detector will classify the sensor as
overdue for calibration. Calibrate the new sensor immediately.
1.Press H to scroll to Sensors. Press
C within 20 seconds to confirm and
access the toxic sensor menu.
2.Press G or H to scroll to Toxic 1
or Toxic2 and press C within
20 seconds to confirm.
A corresponding list of toxic sensors
displays. A checkbox displays beside
the current toxic sensor.
Note
Toxic 1: List includes the PID and CO
sensors.
Toxic 2: List includes the H
sensor.
3.Press G or H to scroll to the new
sensor and press C to confirm. A
checkbox displays beside the new
sensor. To reconfigure, exit the user
options menu.
The following screen displays. The
detector deactivates and immediately
reactivates. It performs the
reconfiguration during the startup.
The new sensor must be calibrated.
Until the new sensor is calibrated, the
due date displays as OL (over limit)
when in normal operation.
The Initialize option restores the original factory default settings of the detector.
1.Press H to scroll to Initialize and
press C within 20 seconds to
confirm.
2.From the Initialize? screen
perform one of the following options
within 5 seconds.
• Press A No to exit, or
• Press C Yes to initialize.
If No is selected, the following screen
displays and the detector exits the initialize option.
If Yes is selected, the following screen
displays while performing the initializing
process.
When initializing is complete, the
following screen displays.
The detector deactivates and then
immediately reactivates. The detector
then reconfigures to the default settings
while it performs the self-test.
Verify all settings and alarm setpoints,
and then calibrate the sensors.
Force Calibration
If enabled, the Force cal option
automatically forces the detector to enter
calibration if a sensor is overdue upon
startup. Press H to scroll to Force cal.
Press C to toggle between enable and
disable.
The detector is shipped with the Force cal option disabled.
• One beep, one flash, and one
vibrate every 10 seconds
AlarmsScreen
Pump Alarm
• Two fast beeps and alternating flashes
• Vibrator alarm activates
•
L and J flash
Note
If enabled, during an alarm condition the Latch (latching
alarms) option causes the low and high gas alarms (audible,
visual, and vibrator) to persist until the alarm is acknowle dged
by pressing C and the gas concentration is below the low
alarm setpoint. The peak concentrations display continua lly
until the alarm condition no longer exists. Local regulations
may require the Latch Alarms option be enable d.
MMC/SD Fail Alarm
• One beep every 5 seconds
• S flashes
If the Stealth option is enabled, the detector only vibrates
during an alarm; the audible and visual alarms are disabled .
a Caution: If the detector emits a low battery alarm,
charge the lithium battery pack immediately.
To avoid possible personal injury, do not deactivate
the detector during a work shift. TWA and STEL readings
reset if the detector is deactivated for more than
5 minutes.
Table 8. Computed Gas Exposures
Gas ExposureDescription
Time-weighted average
(TWA) based on accumulated
TWA (toxic only)
STEL (toxic only)
*Peak (MAX)
exposure to toxic gases averaged over a work day according to OSHA or ACGIH
method.
Short-term exposure limit
(STEL) to gas based on a
5-15 minute user selectable
period.
Peak (MAX) concentration
exposure during work shift.
*For oxygen, it is the highest or the lowest concentration
encountered.
Viewing Gas Exposures
Press and hold C until the
peak (MAX) gas exposures
displays.
The exposures automatically clear after 5 minutes when the detector
is deactivated.
To clear the MAX, TWA, and STEL exposure readings immediately,
press and hold C and G simultaneously. The detector displays the
following screen during the countdown.
Note
Hold C and G for the entire countdown, otherwise the MAX,
TWA, and STEL exposure readings will not clear.
Gas Alarm Setpoints
The gas alarm setpoints trigger the gas alarms and are described in
Table 9.
.
Table 9. Gas Alarm Setpoints
AlarmCondition
Low alarmToxics and combustibles: Ambient
gas level above low alarm setpoint.
Oxygen: Ambient gas level may be
set above or below 20.9%.
High alarmToxics and combustibles: Ambient
gas level above high alarm setpoint.
Oxygen: Ambient gas level may be
set above or below 20.9%.
TWA alarmToxic only: Accumulated value above
the TWA alarm setpoint.
STEL alarmToxic only: Accumulated value above
the STEL alarm setpoint.
Downscale alarmToxic: If sensor reading is negative
(half of the TWA setpoint).
LEL: If sensor reading is negative
(half of the low alarm setpoint).
Multi alarmTwo or more gas alarm conditions.
Viewing the Alarm Setpoints
To view the current alarm setpoints for all of the sensors, press G
during normal operation.
To change the factory-set alarm setpoints, refer to Calibration and Set-
ting Alarm Setpoints.
Note
To disable an alarm, set the alarm setpoint to 0 (zero).
Stopping a Gas Alarm
The low and high alarms stop when the ambient gas level returns to
below the low alarm setpoint.
Note
If alarms are set to latch, press C to acknowledge the alarm.
The alarms will persist until the alarm is acknowledged an d the
gas concentration is below the low alarm setpoint.
a Caution
Follow all safety procedures as defined by your
employer. Confirm with your supervisor before clearing
TWA and STEL alarms.
The TWA and STEL alarms can be stopped either by
• clearing the MAX, TWA, and STEL exposures (refer to Clearing
Gas Exposures), or
• deactivating the detector for 5 minutes (minimum) and then
reactivating it again.
Sensor Alarm
The detector tests for missing or defective sensors during the activation
self-test. If a sensor fails the self-test, FAIL flashes on the LCD above
the failed sensor. Refer to Troubleshooting
.
Pump Alarm
The external pump draws air over the sensors continually. If the pump
stops operating or becomes blocked, the detector activates the pump
alarm. The following screens display.
The pump alarm continues until the blockage is cleared or it is
acknowledged by pressing C. If C is pressed, the detector automatically launches a pump test to reset the pump module.
Refer to Pump Test for more information. If the pump test is successful,
the detector returns to normal operation, otherwise the pump alarm continues. If the pump alarm persists, refer to the Pump Operation section in
Troubleshooting
.
Low Battery Alarm
The detector tests the batteries upon activation and continuously
thereafter. Battery power is continually displayed during normal
operation. If the battery voltage is low, the detector activates the
low battery alarm.
The low battery alarm continues until the batteries are replaced/charged,
or until the battery power is almost depleted. If the battery voltage
becomes too low, the detector deactivates.
Note
Typ ically, the low battery alarm continues for 30 minutes before
the detector automatically deactivates.
Automatic Deactivation Alarm
If the battery voltage is in immediate danger of falling below the minimum
operating voltage, the audible alarm beeps eight times and the visual
alarm flashes eight times. After 3 seconds, the LCD dims and the
detector deactivates.
To replace or charge the batteries, refer to the following sections:
• Replacing the Alkaline Batteries
• Replacing the Lithium Battery Pack
Bump Test
A bump test is the process of applying a small amount of test gas to
force the detector into alarm. A bump test should be performed regularly
to confirm the sensors are responding correctly to gas, and that the audible, visual, and vibrator alarms activate during an alarm condition. Calibrate if the readings are not within specified limits.
Performing a Bump Test
a Caution
BW recommends to bump test the sensors before each
day’s use to confirm their ability to respond to gas by
exposing the sensors to a gas concentration that
exceeds the alarm setpoints.
To perform a manual bump test refer to
page 50 and the following steps 1-6. Follow this procedure when Bump
Daily is enabled.
The Calibration icon will flash if a sensor has failed a bump
test.
If performing a bump test on a pump unit, connect the calibration hose directly to the pump module.
1. Connect the calibration hose to the 0.5l/min regulator on the
gas cylinder.
Note
Only use the calibration cap for bump tests and calibrations.
2. Connect the calibration hose to the intake inlet on the calibra-
tion cap. Arrows on the calibration cap indicate the direction of
gas flow.
3. Attach the calibration cap and tighten the knob.
Refer to
Applying Gas to the Sensors on page 50.
Note
Ensure the knob is securely fastened before applying gas. Do
not overtighten the calibration cap. Ensure strong air currents
are kept away from the exhaust side of the calibration cap, as
this can affect the bump test accuracy.
4. Apply gas. Verify the visual, audible, and vibrator alarms activate.
5. Close the regulator and remove the calibration cap from the
detector. The detector will temporarily remain in alarm until the
gas clears from the sensors.
6. Disconnect the calibration hose from the cap and regulator.
Calibration and Setting Alarm Setpoints
Guidelines
When calibrating the detector, adhere to the following guidelines:
Recommended gas mixture:
CO: 50 to 500 ppm balance N
H
S: 10 to 100 ppm balance N
2
PH3: 1 to 5 ppm balance N
SO2: 10 to 50 ppm balance N
Cl2: 3 to 25 ppm balance N
NH3: 20 to 100 ppm balance N
NO2: 5 to 50 ppm balance N
HCN: 5 to 20 ppm balance N
ClO2: 0.1 to 1.0 ppm balance N
O3: 0.1 to 1.0 ppm balance N
VOC:100 ppm isobutylene
LEL: 50% LEL or 2.5% by vol. methane balance air
: 5000 ppm balance air
CO
2
O
: 18.0%
2
• To ensure accurate calibration, use a premium-grade calibration
gas. Gases approved by the National Institute of Standards and
Technology (NIST) improves the validity of the calibration.
• Do not use a gas cylinder beyond its expiration date.
• Calibrate a new sensor before use. Install the sensor, activate
the detector, and allow the sensor to stabilize before starting
calibration.
Used sensor: 60 seconds / New sensor: 5 minutes
• Calibrate the detector at least once every 180 days (every 90 days
for HCN sensors) depending upon use and sensor exposure to
poisons and contaminants. For a list of common products that can
damage sensors, refer to Sensor Poisons and Contaminants
• Calibrate the detector if the ambient gas varies during startup.
• Calibrate the sensor before changing the alarm setpoints.
• Calibrate only in a safe area that is free of hazardous gas in an
atmosphere of 20.9% oxygen.
• To disable an alarm, set the alarm setpoint to 0 (zero).
• If the Auto cal option is enabled, the oxygen (O
automatically every time the detector is activated. Activate the
detector in a safe area that is free of hazardous gas in an
atmosphere of 20.9% oxygen.
• After activating the detector, allow it to stabilize for 1 minute before
performing a calibration or bump test.
• If a certified calibration is required, contact BW Technologies by
Honeywell.
Note
A generator must be used for O
and ClO2 sensors.
3
) sensor calibrates
2
.
Diagnostics Testing
The detector tests the ambient air (auto zero) and the test gas that is
applied (auto span) to ensure it meets expected values. Auto zero sets
the zero-gas level of the sensor. If ambient gas is present, the zero
level will be incorrect.
If excessive target gas is present, the detector displays an error
message and lists the affected sensor.
In auto span, if the target gas is not detected or does not meet expected
values, a message displays that the detector is exiting calibration mode.
The detector retains the previous set values.
step #5. If the detector is passcode protected, refer to Pass-
code Protect Activated step #4.
If C Yes is pressed, the following screen displays.
zero and proceed to Auto Span
2
a Warning
Use nitrogen (N2) to zero the CO2 sensor.
Apply N
displays.
to zero the CO2 sensor. The following screen
2
flashes whiles the detector zeros the CO
sensor (approximately 30 seconds).
2
Passcode Protect Activated
4. When auto zero is complete and if the passcode protect option
is enabled, the following screen displays.
The passcode must be entered to proceed.
Press G or H to scroll to the correct passcode and then press
C within 5 seconds to confirm. If entered correctly, the detector beeps twice and proceeds to Auto Span
Incorrect Passcode: If the passcode is incorrect or is not confirmed
within 5 seconds by pressing C, the following screens display.
The current low alarm setpoint displays for the selected sensor.
Press H or G to change the value for the low alarm setpoint. When the
required value displays, press C to confirm.
Setting the High Alarm Setpoint
The current high alarm setpoint displays for the selected sensor.
Press H or G to change the value for the high alarm setpoint. When the
required value displays, press C to confirm.
Proceed to step #9 Setting the Remaining Alarm Setpoints
.
Setting the Remaining Alarm Setpoints
9. Repeat step #8 to set the alarm setpoints for the remaining
sensors. When complete, the detector emits two quick beeps
and proceeds to Finish Calibration
.
Finish Calibration
10. The detector displays the following screen to indicate that
calibration is complete.
1. After calibration is complete and the detector returns
to normal operation, verify the calibration by using a
gas cylinder other than the one used for calibration.
2. The gas concentration should not exceed the sensor's detection range. Confirm that the LCD displays the expected concentration values.
3. To ensure the reading is accurate, apply the verification gas for
the same amount of time as was applied to the sensor when it
was calibrated.
Example: SO
verification gas for 2 minutes.
span time 2 minutes therefore, apply
2
Unsuccessful Span
If the sensor(s) did not span successfully, refer to the following sections
for possible solutions:
• Failed Span
• No Gas Detected
• Did Not Reach Target Span
• Large Span
Failed Span
If a sensor fails the span, the following error message displays.
If the sensor is not calibrated, it displays as FAIL in normal operation the
next time the detector is activated.
If all sensors fail the span, the following screen displays.
1. Ensure the regulator valve is open and that the gas cylinder
is not past the expiration date. Replace the gas cylinder and/
or regulator if required.
2. Replace the sensor(s) and sensor filter if required.
3. Press A to exit and then calibrate again in a safe area that
is free of hazardous gas in an atmosphere of 20.9% oxygen.
4. If the span fails a second time, reactivate the detector to test
the sensors.
If all sensors fail the span, the calibration due dates cannot be reset. The
following screens display.
No Gas Detected
If the detector does not detect any gas within 30 seconds, the following
screens display.
1. Ensure the regulator valve is open and the calibration hose is
free of debris.
2. Replace the gas cylinder and/or regulator if required.
3. Replace the sensor(s) and sensor filter if required.
4. Press G to reapply gas or press H to end the span.
Did Not Reach Target Span
If the target span was not achieved, as defined in the Span Gas Value
option, the following screens display.
Accept Current Span: If the gas cylinder, regulator, and sensor
are operating correctly, press C to accept the current span.
Reject Current Span: Press A to reject. Verify the gas cylinder,
detector settings, and sensor are operating correctly. Calibrate the
sensor again.
Large Span
If the span adjustment is unusually large (more than 15%), the following
screens display.
Ensure the correct gas cylinder being used and that the span concentration value(s) of the detector matches the span concentration value(s) of
the gas cylinder. Refer to Span Gas Value
Adjustment Expected: If the calibration adjustment is expected,
press C to accept the span.
Adjustment Not Expected: If the calibration adjustment is not
expected or the span gas value does not match the gas cylinder,
press A to reject the span and calibrate that sensor again.
When using the pump module, attach it and the pump accessories prior
to activating the detector.
The pump module can be used independently to measure gas or it can
be used with the sample probe to test for gases in confined spaces. To
measure gas using the sample probe, refer to Confined Space Sampling
a Warning
If the pump module is installed on the detector, the
following three things must occur during start-up. If any
one of the conditions below does not occur, discontinue
use of the detector and contact BW Technologies by
Honeywell immediately.
• The detector prompts for a pump test during start-up
• The pump module passes the pump test at start-up when
the pump inlet or sample chain inlet is blocked
•The J icon displays on the LCD
To ensure accurate gas detection, the sensors must be
calibrated immediately when the pump module is
replaced by the diffusion cap and vice versa.
When using the sample probe at -10°C to +0°C (14°F to
32°F), keep the sample probe in your hand.
a Caution
A demand flow regulator must be used to manually
calibrate the GasAlertMicro 5/PID/IR detector when the
pump module is installed.
The calibration cap is designed for use with the diffusion
cap only. It cannot be used with the pump module.
Identifying the Pump
.
There are two generations of pump. The best way to differentiate the
Generation 2 pump from the Generation 1 pump is the inline filter that is
visible on the Generation 2 pump on the left-hand side.
Each model of pump has different operating specifications. Refer to
Maximum Hose Length when Confined Space Sampling on page 71.
Generation 1: 116885-L3 (yellow) and 118933-L3 (black)
To replace the Generation 2 pump filter, refer to Figure 7. and the following steps 1-8.
Note
If using the auxiliary filter on the Generation 1 pump or as an
additional filtration system on the new pump module, refer to
Attaching the Auxiliary Filter
Figure 7. Replacing the Pump Filter (Generation 2 Pump)
.
1. Deactivate the detector.
2. Remove the filter window screw.
3. Remove the filter window.
4. Remove the old filter from the filter window.
Important! Ensure the filter cavity and filter window are clean
and free of debris.
5. Insert a new filter.
6. Replace the filter window and screw.
7. Activate the detector. The detector performs a pump test
during the startup self-tests to verify the pump module is operating correctly.
8. To order additional filters, refer to Replacement Parts and
Accessories.
a Caution
Filters may need to be replaced more frequently in high
particulate areas. BW Technologies by Honeywell
recommends that the auxiliary filter be used as a
additional filtration in high particulate areas.
The filter is designed to protect the pump. For
Generation 1 pumps, the filter must be connected when
the pump is activated. Only remove the filter to perform
calibrations, bump tests, and when using the sample
probe.
For new model pumps, the auxiliary filter can be used as
an additional filtration system in high particulate areas.
To insert the auxiliary filter, complete the following:
1. Loop the filter cord through the ring on the alligator clip and
pull the filter through the loop of the filter cord.
Note
Attaching the filter cord to the alligator clip ensures the filter
remains with the detector when not in use.
2. Pull the filter cord tight. Ensure the loop knot is secure.
3. Align the pump quick connector with the filter quick connector
and turn clockwise. Ensure the connection is secure.
4. Change the filter as required. Environments with heavy airborne particulates require the filter to be changed
frequently.
To order additional filters, refer to Replacement Parts and
The sample probe is used to safely test for gas in confined spaces
before entering.
Attach the sample probe to the pump module prior to activating the
detector.
a Warning
To measure hazardous gas in a confined space, the
sample probe must be used with the pump module.
Figure 11. Attaching the Sample Probe
a Caution
To prevent the Teflon lining inside the Tygon tubing from
causing a blockage when connecting it to the sample
probe, the end of the tubing must be flared. Refer to
steps 3 and 4.
Note
The detector and pump module can also be used with a sintered filter.
1. Deactivate the detector.
2. Attach the pump connector end of the teflon-lined Tygon tubing
to the nozzle on the pump module.
3. Gently insert needle nose pliers into the other end of the teflonlined Tygon tubing. Using a circular motion, flare the end of the
tubing.
4. Connect the flared end of the tubing to the sample probe.
Ensure the teflon lining does not separate from the Tygon tubing, as it will block the tube and generate a pump alarm.
5. Activate the detector. Ensure all connections are secure before
sampling.
6. Insert the sample probe into the confined space.
a Warning
Depending upon the length of the tubing and the type of
gas in the confined space, allow a minimum of 3
Detectors equipped with the datalogger option record information that
can be compiled to create a report. To set how often the detector records
a sample (1-127 seconds), refer to Logger Option
menu.
The following information is recorded in a datalog:
• Date and time
• Serial number of the detector
• Type of gas the detector monitors
• Gas reading(s) that display
• STEL and TWA readings
• Sensor status
• Detector status
• Passcode protect enabled/disabled
• STEL period setting
• Confidence beep enabled/disabled
• Automatic backlight enabled/disabled
• Stealth mode is enabled/disabled
• Latching alarm enabled/disabled
• Calibration past due option enabled/disabled
• Language the detector is set to display
in the user options
MMC/SD Card Compatibility
For a list of compatible memory cards, please contact BW Technologies
by Honeywell.
Inserting the MMC/SD Card
To insert the MMC/SD card into the detector, refer to the following steps
1-5 and Figure 12.
1. Deactivate the detector.
2. Release the latch and remove the battery pack.
3. Insert the MMC/SD card (pins face down).
4. Replace the battery pack and secure the latch.
5. Activate the detector. The MMC/SD card is automatically formatted during the startup self-test.
The MMC/SD card is not required for operation in detectors equipped
with datalogging. However, the following two screens display if the card
is not inserted during startup.
A new MMC/SD card is automatically formatted when it is inserted in the
detector. When the detector is activated, it begins the self-test and then
displays the following screen.
If the MMC/SD card has been accidentally reformatted or erased by the
computer application, the following screens display when the card is
inserted into the detector.
a Caution
Only erased data files can be restored using the detector.
Computer applications sometimes write data over erased
files, and that erased data cannot be restored by the detector.
Always create back up files on the computer.
To restore the logfile, complete the following:
1. From the detector, press H to restore the logfile. The
following screen displays.
If the detector successfully restores the logfile, the
following screen displays and the startup tests continue.
2. From the computer, verify that the logfile has been restored.
When the normal operation screen displays, deactivate the
detector.
3. Remove the MMC/SD card and insert it into the card reader.
4. From the computer desktop, double-click My Computer to
view the list of drives.
5. Double-click the Removable Disk drive to access LOG-FILE0.csv. Open the logfile and verify that the data has been
restored.
If LOGFILE0.csv does not display, ensure that the MMC/SD
card is inserted in the card reader correctly and that all connections are secure.
6. After verifying that the logfile has been restored, re-insert the
MMC/SD card into the detector.
To maintain the detector in good operating condition, perform the
following basic maintenance as required.
• Calibrate, bump test, and inspect the detector at regular intervals.
• Maintain an operations log of all maintenance, calibrations, bump
tests, and alarm events.
• Clean the exterior with a soft damp cloth. Do not use solvents,
soaps, or polishes. Refer to Sensor Poisons and Contaminants
• Do not immerse the detector in liquids.
Battery Cautions
a Warning
To avoid personal injury and/or property damage, adhere
to the following:
• Replace the batteries immediately when the detector emits
a low battery alarm.
• Use only batteries that are recommended by BW
Technologies by Honeywell. Refer to Specifications
• Ensure the alkaline batteries are properly installed in the
detector battery pack.
• Charge batteries using only a recommended BW charger.
Do not use any other charger. Failure to adhere to this
caution
can cause fire and/or explosion.
• Do not calibrate the detector during or immediately after
charging.
• Both the rechargeable lithium battery pack and the alkaline
battery pack are user-changeable in hazardous locations,
but the alkaline battery cells inside the pack can only be
replaced in a safe area that is free of hazardous gas.
• Warning: The battery pack (M5-BAT08) that is equipped with
a lithium polymer cell may present a risk of fire or chemical
.
.
burn hazard if misused. Do not recharge, disassemble, heat
above 212ºF (100ºC), or incinerate.
• Warning: Do not use any other lithium batteries with the
Micro5/PID/IR detectors. Use of any other cell can cause fire
and/or explosion. To order a replacement lithium battery
pack (M5-BAT08), refer to Replacement Parts and
Accessories.
• Warning: Lithium Polymer cells exposed to heat at 266ºF
(130ºC) for 10 minutes can cause fire and/or explosion.
• Dispose of used lithium cells immediately. Do not
disassemble and do not dispose of in fire. Do not mix with
the solid waste stream. Spent batteries must be disposed of
by a qualified recycler or hazardous materials handler.
• Keep lithium cells away from children.
Charging the Battery
To charge the battery, refer to the GasAlertMicro 5/PID/IR Charger
Instruction Sheet.
• The detector must be deactivated to charge the battery.
To replace the lithium battery pack, refer to Figure 14. and the following
steps 1-4.
Figure 14. Replacing the Lithium Battery Pack
1. Deactivate the detector.
2. Open the latch on the bottom of the detector.
3. Remove the battery pack by lifting the bottom of the pack
upward from the detector.
a Warning
Do not disassemble lithium battery packs. Read and
adhere to the cautions in Battery Cautions
4. Replace the battery pack with a fully charged lithium
battery pack.
5. Close the latch.
.
Replacing a Sensor or Sensor Filter
a Warning
To avoid personal injury, only use sensors that are
specifically designed for the detector. Refer to
Specifications
Replace the sensor in a safe and non-hazardous area
that is free of hazardous gas
Each sensor has a high degree of resistance to common vapors and
gases. To clear a sensor, place the detector in a clean environment
and wait 10 to 30 minutes.
Do not expose a sensor to vapors from organic solvents such as paint
thinners and acetone. For a list of common products that can damage
sensors, refer to Sensor Poisons and Contaminants
2. Remove the two machine screws on the rear shell and then
remove the sensor cover or pump module.
3. Remove the sensor filter and/or the sensor(s).
4. Insert the new filter and/or sensor. Ensure the sensor posts are
aligned correctly.
5. Re-assemble the detector.
6. If the sensor is changed, such as SO
must be reconfigured. Refer to Sensors
option.
7. A sensor that has been replaced must be calibrated immediately. Refer to Calibration and Setting Alarm Setpoints
Machine screws (2)
to an H2S, the detector
2
in the Te ch Mo d e
.
Photoionization Detector (PID)
Clean or Replace the Lamp
The PID lamp must be cleaned regularly. Use only the cleaning kit
that is supplied by BW Technologies by Honeywell.
To clean the PID lamp, refer to the illustrations and procedures that
are provided with the PID Lamp Cleaning Kit. To order the kit, refer
to Replacement Parts and Accessories
a Caution
To ensure proper maintenance and continued accurate
readings from the sensor, use only the PID Lamp
Cleaning Kit that is supplied by BW Technologies by
Honeywell.
Replace the electrode stack when it is contaminated. To replace
the electrode stack, refer to Table 22.
procedures.
Ensure your fingers do not make contact with the
diffusion barrier or the electrodes on the underside of
the stack.
1. Remove the sensor cover.
2. Remove the old electrode stack.
3. Insert the new electrode stack.
4. Replace the sensor cover.
, Figure 16., and the following
a Caution
WEEE Battery Directive
Failure to comply with the following battery removal and disposal instructions may result in battery shorting, battery leakage, and/or other damage. Ensure a qualified technician completes the following procedures.
Removal and Disposal of the Battery Pack
To remove the alkaline batteries, refer to steps #1 to #3 in Replacing the
Alkaline Batteries on page 82.
To remove the lithium battery pack, refer to steps #1 and #2 in
the Lithium Battery Pack on page 83.
Dispose of the battery pack according to local laws.