Hioki BT3554, BT3554-01, BT3554-10, BT3554-11 Instruction Manual

BT3554
BT3554-10
BT3554-11
BATTERY TESTER

Instruction Manual
Oct. 2018 Revised edition 3 BT3554A961-03 18-10H
EN

Battery Tester Quick Guide

Let's Start by Measuring a New Battery
To determine if a battery is worn out, measure the data of a new battery. As a battery wears out, its internal resistance increases to approximately 1.5 to 2 times (reference value) that of a new one. Use these values as guidelines when determining the battery wear judgment values.
Example: Changes in the internal resistance and voltage due to battery wear
Resistance has doubled...
Worn out
1
2
3
4
New battery
Internal resistance: 0.5 m Voltage: 2.0 V (The above values are examples.)
Resistance wear out limit
FAIL
Default value × 2
WARNING
Default value × 1.5
PASS
Default value
BT3554A961-03
Battery requiring replacement
1.0 m
Ω
(twice the value of a
Ω
new battery)
1.8 V (90% of the value of a new battery)
PassedPassedPassedPassed
Failed
5
6
Internal resistance
7
Appx. Index
Discharge capacity
Basic Usage
Connect the test leads to the instrument.
1
Turn the instrument power on.
2
Check the clock settings when using the
instrument for the rst time. (p. 36)
Changing the ranges. (p. 39)
3
Turn on the auto-hold function and auto-
4
memory function. (
Auto-hold function: Holds measurement values automatically when they
become stable. (p. 75)
Auto-memory function: Automatically stores measurement values immediately after they are held. (p. 76)
Connect the test leads to the battery.
5
Press the key to read out the
6
measurement values. (p. 69)
and are displayed.)
Data is stored in internal memory.
(Black)
Hold at least 1 sec.
(Black)(Red)
+
(Red)
Handy Features
Comparator function
With the comparator function, the threshold values can be set to determine if the battery is worn out. (p. 55)
1
Setting examples of the battery wear judgment values
PASS
(Passing limit)
New battery
Resistance: 0.5 m Voltage: 2.0 V (The above values are examples.)
Ω
WARNING
(Warning limit)
In use
0.75 mΩ (1.5 times the value of a new battery)
1.8 V (90% of the value of a new battery)
(Failure limit)
Battery requiring replacement
1.0 mΩ (twice the value of a new battery)
FAIL
2
3
4
5
6
7
Appx. Index
Downloading measurement values to a computer
By connecting the instrument to a computer with the included USB cable, the measurement data can be downloaded to the computer. (p. 83)
Viewing measurement values on a mobile device (only
for BT3554-01)
With the Bluetooth® Communication function, the measurement data can be viewed on a smart phone or tablet. (p. 84)

Contents

Introduction .........................................................................1
Verifying Package Contents ..............................................2
Options ................................................................................ 4
Safety Information .............................................................. 7
Operating Precautions ..................................................... 12
1 Overview 19
1.1 Measuring Battery Wear ............................... 19
1.2 Overview ........................................................21
1.3 Features .........................................................22
1.4 Names and Functions of Parts .................... 24
1.5 Dimensions .................................................... 29
2 Measurement Preparations 31
2.1 Attaching the Neck Strap ............................. 31
2.2 Installing/Replacing Alkaline Batteries ....... 32
2.3 Connecting the Test Lead .............................33
Connecting a Pin Type Lead and the Model
9466 Remote Control Switch .................................... 34
2.4 Turning the Power ON/OFF ..........................35
2.5 Clock Function .............................................. 36
Turning the Date and Time Display ON/OFF ............ 36
Adjusting the Date and Time .....................................36
1
2
3
4
5
6
7
3 Measurement 37
3.1 Pre-operation Inspection .............................. 38
3.2 Setting the Measurement Range ................. 39
3.3 Adjusting Zero Value
(Zero Adjustment) .........................................40
Appx. Index
8
9
i
Contents
Shorting Methods for Various Test Leads ................. 41
Performing Zero Adjustment ..................................... 44
Zero Adjustment Troubleshooting ............................. 46
Canceling the Zero Adjustment Operation ................46
3.4 Retaining the Displayed Values ................... 47
Canceling the Retaining State ..................................47
When retaining with the Model 9466 Remote
Control Switch ........................................................... 48
3.5 Determining Battery-wear Judgment
Values .............................................................49
3.6 Measuring a Battery (Inspection) ................ 50
Error Measurements ................................................. 52
Warning Display ........................................................52
3.7 Measuring the Temperature .........................53
Changing the temperature display unit ..................... 54
4 Comparator Function (Evaluation by
Threshold Values) 55
4.1 Overview ........................................................55
4.2 Turning On the Comparator Function .........56
4.3 Setting Threshold Values for the
Comparator .................................................... 57
Comparison Table for the Comparator ...................... 61
4.4 Setting the Comparator Buzzer ................... 63
4.5 Canceling the Comparator Function ........... 64
5 Memory Function 65
5.1 Overview ........................................................65
Memory Structure .....................................................65
5.2 Storing Data in the Memory ......................... 66
5.3 Canceling the Memory Function .................. 68
5.4 Reading Out Stored Data .............................. 69
ii
Contents
5.5 Clearing Stored Data ..................................... 70
Clearing a Single Set of Data ...................................70
Clearing Data from Each Unit ................................... 71
Clearing All Data ....................................................... 72
6 Other Features 73
6.1 Noise Frequency Avoidance Function ........73
6.2 Auto-hold Function ....................................... 75
6.3 Auto-memory Function ................................. 76
6.4 Auto Power Save Function (APS) ................ 77
6.5 Backlight ........................................................ 78
6.6 System Reset ................................................. 79
Default Settings (Factory Defaults) ........................... 80
6.7 Battery Level Warning ..................................81
7 Communications Function 83
7.1 Communicating with a Computer ................ 83
7.2 Communicating with a Smart Phone or
Tablet (Only for BT3554-01) ..........................84
Installing the smartphone app GENNECT Cross ...... 85
Turning ON/OFF the Bluetooth® function .................. 86
Pairing the app with the battery tester (BT3554-01) . 87 Making measurements with the Bluetooth
®
function . 88
Appx.
1
2
3
4
5
6
8 Specications 89
8.1 GeneralSpecications ................................. 89
8.2 BasicSpecications ..................................... 91
8.3 AccuracySpecications ............................... 93
8.4 FunctionalSpecications ............................. 95
8.5 CommunicationSpecications .................. 102
iii
7
8
9
Contents
9 Maintenance and Service 103
9.1 Repair, Inspection, Cleaning ...................... 103
9.2 Troubleshooting ..........................................105
Before Returning for Repair .................................... 105
9.3 Error Messages ........................................... 107
9.4 Frequently Asked Questions ...................... 108
9.5 Replacing the Fuse ..................................... 109
9.6 Replacing the Test Lead’s Tip Pin ............. 110
9.7 Disposing the Instrument (Removing
the Lithium Battery) .................................... 113
Appendix Appx.1
Appx. 1 Effect of Extending the Test Lead
and Induced Voltage ......................... Appx.1
Reducing Induced Voltage ............................Appx.1
Appx. 2 Effect of Eddy Currents.................... Appx.2
Appx. 3 AC 4-terminal Measurement Method .
Appx.3
Appx. 4 Effects of Current Density ............... Appx.5
When the Measurement target is Wide or
Thick .............................................................Appx.5
Appx. 5 Synchronous Detection System...... Appx.7
Appx. 6 Calibration ......................................... Appx.8
Calibrating the Resistance Measurement
Component ...................................................Appx.8
Calibrating the Voltage Measurement Unit ...Appx.9
iv

Introduction

Introduction
Thank you for purchasing the Hioki BT3554, BT3554‑01, BT3554‑10, BT3554‑11 Battery Tester. To obtain maximum performance from the instrument, please read
this manual rst, and keep it handy for future reference.
In this document, the model is indicated as BT3554 or BT3554‑01 (as printed on the instrument), shown below.
Model
BT3554 BT3554 9465‑10
BT3554‑01 BT3554‑01
BT3554‑10 BT3554 L2020
BT3554‑11 BT3554‑01
Model printed on
the instrument
Bluetooth
Standard accessory:
®
9465‑10
L2020
: Yes,
Pin Type Lead
– : No
Trademarks
• Bluetooth® is a registered trademark of Bluetooth SIG, Inc.(USA). The trademark is used by HIOKI E.E. CORPORATION under license.
• Android and Google Play are trademarks of Google, Inc.
• IOS is a registered trademark of Cisco Systems, Inc. and/or its
afliates in the United States and certain other countries.
• iPhone, iPad, iPad miniTM, iPad Pro, and iPod touch are trademarks of Apple Inc.
• The App Store is a service mark of Apple Inc.
1
2
3
4
5
6
7
AppendixIndex
1

Verifying Package Contents

Verifying Package Contents
When you receive the instrument, inspect it carefully to ensure that no damage occurred during shipping. In particular, check the accessories, panel switches, and connectors. If damage is evident,
or if it fails to operate according to the specications, contact your
authorized Hioki distributor or reseller.
Conrm that these contents are provided.
(4)
(5)
(3)
(2)
(6)
(7)
(1)
(8)
(9)
2
Verifying Package Contents
(1) Model BT3554 or BT3554‑01 Battery Tester × 1,
Protector × 1 (attached)
(2) Zero adjustment board × 1
(3) Instruction manual*
Emits Radio Waves (only for BT3554‑01) × 1, Application software*
(4) Carrying Case × 1
(5) Spare fuse × 1
(6) LR6 (AA) alkaline batteries × 8
(7) USB cable × 1
(8) Model 9465‑10 or L2020 Pin Type Lead × 1
(9) Neck strap × 1
*1 Instruction manuals may also be available in other languages.
Please visit our website at http://www.hioki.com. *2 The latest version can be downloaded from our website. *3 Apply to battery cover or other location as desired.
1
× 1, Precautions Concerning Use of Equipment That
2
CD × 1, Power‑on option sticker*3 × 1
1
2
3
4
5
6
7
AppendixIndex
3

Options

Options
The following options are available for the instrument. Contact your authorized Hioki distributor or reseller when ordering.
Model 9772 Pin Type Lead
The lead's pins are arranged parallel to each other. It is a strong, wear‑resistant lead.
(Red)
(Red)
(Black)
(Black)
Model L2020 Pin Type Lead
This pin type lead has a four-terminal structure and can be used in conned spaces where the measurement target is difcult to reach.
(Red)
(Red)
(Black)
(Black)
Model 9465-10 Pin Type Lead
This pin type lead has a four‑terminal structure.
(Red)
(Black)
(Black)
(Red)
4
Model 9772-90 Tip Pin
The replacement tip pin for the model 9772 Pin Type Lead.
9.15 mm
1.8 mm
2.5 mm
4.3 mm
φ
Options
1
2
Model 9465-90 Tip Pin
The replacement tip pin for the model 9465‑10 and L2020 Pin Type Lead.
2.7 mm φ1.27 mm
φ
Model 9466 Remote Control Switch
When this switch is attached to the test lead, the instrument can hold the values while measuring them.
Wearable models: • Model 9465‑10 Pin Type Lead
• Model 9772 Pin Type Lead
• Model L2020 Pin Type Lead
Switch
Miniplug φ2.5 mm (EXT.HOLD)
3
4
5
6
7
AppendixIndex
5
Options
Model Z5038 0 ADJ Board
Hook and loop fastener is necessary seperately to stick to the carrying case and use model Z5038.
(For models 9772, L2020, and 9465‑10)
Model 9460 Clip Type Lead with Temperature Sensor
Resistance, voltage, and temperature can be measured simultaneously with this lead.
(Red)
Temperature sensor
(Black)
Miniplug (TEMP SENSOR)
(Red)
(Black)
Model 9467 Large Clip Type Lead
These can clip the test lead to the measurement target with a thick bar. Four‑ terminal measurement can be conducted just by clipping the lead to the target.
(Red)
(Red)
Approx.
29 mm
φ
6
(Black)
Approx. 850 mm
Approx. 250 mm
(Black)

Safety Information

Safety Information
This instrument is designed to conform to IEC 61010 Safety Standards, and has been thoroughly tested for safety prior to shipment. However, using the instrument in a way not described in this manual may negate the provided safety features. Before using the instrument, be certain to carefully read the following safety notes:
DANGER
Mishandling during use could result in injury or death, as well as damage to the instrument. Be certain that you understand the instructions and precautions in the manual before use.
WARNING
With regard to the electricity supply, there are risks of
electric shock, heat generation, re, and arc ash due
to short circuits. If persons unfamiliar with electricity measuring instrument are to use the instrument, another person familiar with such instruments must supervise operations.
Protective gear
WARNING
This instrument is measured on a live line. To prevent electric shock, use appropriate protective insulation and adhere to applicable laws and regulations.
1
2
3
4
5
6
7
AppendixIndex
7
Safety Information
Notation
In this document, the risk seriousness and the hazard levels are
classied as follows.
DANGER
WARNING
CAUTION
IMPORTANT
Indicates an imminently hazardous situation that will result in death or serious injury to the operator.
Indicates a potentially hazardous situation that may result in death or serious injury to the operator.
Indicates a potentially hazardous situation that may result in minor or moderate injury to the operator or damage to the instrument or malfunction.
Indicates information related to the operation of the instrument or maintenance tasks with which the operators must be fully familiar.
Indicates a high voltage hazard. If a particular safety check is not performed or the instrument is mishandled, this may give rise to a hazardous situation; the operator may receive an electric shock, may get burnt or may even be fatally injured.
Indicates prohibited actions.
Indicates the action which must be performed.
8
HOLD
[HOLD]
Indicates a control key.
Indicates the display on the screen.
Symbols on the instrument
Safety Information
Indicates cautions and hazards. When the symbol is printed on the instrument, refer to a corresponding topic in the Instruction Manual.
Indicates a fuse.
Indicates a grounding terminal.
Indicates DC (Direct Current).
Symbols for various standards
Indicates the Waste Electrical and Electronic Equipment Directive (WEEE Directive) in EU member states.
Indicates that the product conforms to regulations set out by the EU Directive.
Indicates that the product incorporates Bluetooth® wireless technology.
FCC ID
FCC ID Indicates the ID number of the wireless module certied
by the U.S. Federal Communications Commission (FCC).
Indicates the identication number of a wireless module
IC
approved by Industry Canada (IC).
1
2
3
4
5
6
7
AppendixIndex
9
Safety Information
Screen Display
The instrument screen displays the alphanumeric characters as follows.
A B C D E F G H I J K L M N O P Q R S T U V W X Y Z
1 2 3 4 5 6 7 8 9 0
A different display is used in the case below:
Displays when setting the Bluetooth
:
communication.
: Displays when clearing the stored data.
Displays when setting the comparator buzzer
:
to FAIL.
Accuracy
We dene measurement tolerances in terms of f.s. (full scale), rdg.
(reading) and dgt. (digit) values, with the following meanings:
(maximum display value or range)
f.s.
The maximum displayable value. This is usually the name of the currently selected range.
(reading or displayed value)
rdg.
The value currently being measured and indicated on the measuring instrument.
(resolution) The smallest displayable unit on a digital measuring instrument,
dgt.
i.e., the input value that causes the digital display to show a "1" as
the least-signicant digit.
®
10
Safety Information
Measurement Categories
To ensure safe operation of measuring instruments, IEC 61010 establishes safety standards for various electrical environments, categorized as CAT II to CAT IV, and called measurement categories.
DANGER
• Using a measuring instrument in an environment designated with a higher category than the rating of the instrument could result in a severe accident, and must be carefully avoided.
• Using a measuring instrument without categories in an environment categorized as CAT II to CAT IV could result in a severe accident, and must be carefully avoided.
CAT II: When directly measuring the electrical outlet receptacles of the
primary electrical circuits in equipment connected to an AC electrical outlet by a power cord (portable tools, household appliances, etc.).
CAT III: When measuring the primary electrical circuits of heavy equipment
(xed installations) connected directly to the distribution panel, and
feeders from the distribution panel to outlets.
CAT IV: When measuring the circuit from the service drop to the service
entrance, and to the power meter and primary overcurrent protection device (distribution panel).
Distribution panel
Service entrance
Service drop
CAT IV
Power meter
Fixed installation The instrument’s labeling does not indicate its suitability for use in any particular measurement category.
Internal wiring
CAT III
CAT II
T
Outlet
1
2
3
4
5
6
7
AppendixIndex
11

Operating Precautions

Operating Precautions
Follow these precautions to ensure safe operation and to obtain the
full benets of the various functions. Ensure that your use of the product falls within the specications not
only of the instrument itself, but also of any accessories, options, batteries, and other equipment being used.
Installing the instrument
CAUTION
Installing the instrument in inappropriate locations may cause a malfunction of instrument or may give rise to an accident. Avoid the following locations.
• Exposed to direct sunlight or high temperature
• Exposed to corrosive or combustible gases
• Exposed to a strong electromagnetic eld or electrostatic charge
• Near induction heating systems (such as high‑frequency induction heating systems and IH cooking equipment)
• Susceptible to vibration
• Exposed to water, oil, chemicals, or solvents
• Exposed to high humidity or condensation
• Exposed to high quantities of dust particles
Do not place the instrument on an unstable table or an inclined place. Dropping or knocking down the instrument can cause injury or damage to the instrument.
12
Preliminary checks
Operating Precautions
DANGER
If the test lead or the instrument is damaged, there is a risk of electric shock. Before using the instrument perform the following inspection:
• Before using the instrument check that the coating of the test leads are neither ripped nor torn and that no metal parts are exposed. Using the instrument under such conditions could result in electric
shock. Replace the test leads with those specied
by our company.
• Verify that the instrument operates normally to ensure that no damage occurred during storage
or shipping. If you nd any damage, contact your
authorized Hioki distributor or reseller.
Precautions for Transportation
During shipment of the instrument, handle it carefully so that it is not damaged due to a vibration or shock.
Handling the instrument
DANGER
To avoid electric shock, do not remove the instrument's case. The internal components of the instrument carry high voltages and may become very hot during operation.
1
2
3
4
5
6
7
AppendixIndex
13
Operating Precautions
CAUTION
To avoid damage to the instrument, protect it from physical shock when transporting and handling. Be especially careful to avoid physical shock from dropping.
Precautions for measurement
DANGER
To avoid electric shock, be careful to avoid shorting live lines with the test leads.
WARNING
• Do not use the instrument and test lead with circuits
that exceed its ratings or specications. Doing so
may cause damage, resulting in an electric shock.
• Do not measure any voltage that would exceed the instrument’s maximum input voltage (terminal-to­terminal) or maximum rated terminal-to-ground voltage of 60 V.
Maximum input voltage (terminal-to-terminal) 60 V DC
Maximum rated terminal­to-ground voltage 60 V DC
14
• Do not measure AC voltage.
Operating Precautions
WARNING
• Be sure to connect the test lead correctly.
• Wear gloves of rubber or similar material during measurement.
Ensure sufcient ventilation when measuring batteries in the measurement room to prevent explosions. Sparks may occur when the test leads are connected to batteries, which can ignite any
accumulated inammable gases such as hydrogen.
1
2
CAUTION
• After measuring a high-voltage battery, rst short the test leads together to discharges the DC elimination capacitor connected across the leads before continuing to measure a low‑voltage battery. Otherwise an excess voltage may be applied to the low‑voltage battery causing damage to the battery.
• To avoid damage to the instrument, do not apply voltage to the EXT.HOLD and TEMP.SENSOR terminal.
IMPORTANT
Do not place the test leads in contact with the measurement
terminals of a battery that is leaking uid. Doing so may cause
a degradation in instrument functionality due to exposure to electrolyte from the leaking battery.
3
4
5
6
7
AppendixIndex
15
Operating Precautions
Handling the test leads
CAUTION
Do not apply force when the pin type lead tip is in contact with the battery at a tilted angle.
OK NO
Avoid subjecting the temperature probe tip to physical shock, and avoid sharp bends in the leads. These may damage the probe or break a wire.
IMPORTANT
When using the instrument, use only the test leads with those
specied by our company. Using other test leads may result
in incorrect measurements due to loose connections or other reasons.
Zero adjustment board
WARNING
To prevent short-circuit accidents, do not place the zero adjustment board on top of the battery.
16
Batteries and fuses
Operating Precautions
WARNING
• To avoid electric shock when replacing the batteries
and fuse, rst disconnect the test leads from the
object to be measured, and then remove the case.
• To prevent instrument damage or electric shock, use only the screw for securing the battery cover in place that are originally installed. If you have lost
a screw or nd that a screw is damaged, please
contact your Hioki distributor for a replacement.
Replace the fuse only with one of the specied type, characteristics, rated current, and rated voltage. Do
not use fuses other than those specied (especially,
do not use a fuse with higher-rated current) or do not short circuit and use the fuse holder. Doing so may damage the instrument and result in bodily injury. Fuse type: 216.630, Littelfuse Inc., fast-acting, rating 250 V / F 630 mAH, circuit breaker rating 1500 A
• Battery may explode if mistreated. Do not short-
circuit, recharge, disassemble or dispose of in re.
1
2
3
4
5
6
7
AppendixIndex
17
Operating Precautions
CAUTION
Poor performance or damage from battery leakage could result. Observe the cautions listed below:
• Do not mix old and new batteries, or different types of batteries.
• Be careful to observe the battery polarity during installation.
• Do not use batteries after their recommended expiry date.
• Do not allow weak batteries to remain in the instrument.
• Replace batteries only with the specied type.
• Remove the batteries from the instrument if it is to be stored for a long time.
Handle and dispose of batteries in accordance with local regulations.
CD disc precautions
• Exercise care to keep the recorded side of discs free of dirt and scratches. When writing text on a disc's label, use a pen or marker with a soft tip.
• Keep discs inside a protective case and do not expose to direct sunlight, high temperature, or high humidity.
• Hioki is not liable for any issues your computer system experiences in the course of using this disc.
18
1

Overview

1.1 Measuring Battery Wear

IMPORTANT
To determine if a battery is worn out, rst measure the internal
resistance of a new or good battery.
When the battery is worn out, the internal resistance rises to approximately 1.5 to 2 times its default value (reference values).
The graph below shows the relation between storage capacity and default value of internal resistance in a lead-acid battery. "CS", "HS", and "MSE" denote JIS (Japanese Industrial Standard) lead­acid battery types. The internal resistance of an MSE (sealed stationary lead-acid battery) can be read at approximately 1 m
approximately 0.13 m
]
[m
Ω
10
1
0.1
Internal resistance value
(1000 Ah).
Ω
CS
HS
MSE
CS Clad type stationary lead-acid
HS High discharge rate stationary
MSE Sealed stationary lead-acid
(100 Ah) and
Ω
battery
lead-acid battery
battery
1
2
3
4
5
6
7
Appx. Index
0.01 10 100 1000 10000
Lead-acid battery capacity
[Ah]
19
Measuring Battery Wear
• The warning limit (WARNING) for an MSE (sealed stationary lead-acid battery) is when the internal resistance reaches approximately 1.5 times its default value. The failure limit (FAIL) varies for each manufacturer.
• Default value of internal resistance may vary among batteries with the same capacity, depending on the model or manufacturer. Use the graph on the previous page as reference.
• Internal resistance warning limit (WARNING) and failure limit (FAIL) varies for each manufacturer.
Source: Battery technician certication textbook, Battery Association of Japan (BAJ)
Comparator function (p. 55)
With the comparator function, the battery measurement values can be compared with the present threshold values to determine which ranges those values fall within: PASS, WARNING, or FAIL.
In open (liquid) stationary lead-acid batteries such as CS, HS, and alkaline lead-acid batteries, variations in internal resistance are small compared with sealed stationary lead-acid batteries, and sometimes it is difcult to determine worn out state of the batteries.
20
Overview

1.2 Overview

This instrument measures internal resistance, voltage, and terminal temperature* of lead-acid, nickel-cadmium, nickel-hydrogen, and other types of batteries, enabling you to determine if the battery is
worn out.
1
2
3
* Temperature measurement requires the optional model 9460 Clip
Type Lead with Temperature Sensor.
The measurement data can be copied to a computer by connecting the instrument to a computer after measurement with the USB cable provided. Also, users can view the data on their smart phones or tablets with the Bluetooth® Communication function.
Measurement data
Specifying
threshold values
Measurement data
21
4
5
6
7
Appx. Index
Features

1.3 Features

Enables measurement without shutting down UPS
systems
This instrument uses high-precision AC resistance measurement technology, and noise reduction technology. Time required for measurement is reduced since the instrument is capable of measuring live wires without requiring the UPS system to be shut
down.
Reliable measurement values
This instrument is capable of obtaining reliable measurement values without being affected by lead or connector resistance because it uses the AC 4-terminal method to measure internal resistance.
Simultaneous display of resistance, voltage, and
temperature
Without changing functions, this instrument can display battery internal resistance, voltage, and terminal temperature simultaneously. Temperature measurement requires the optional model 9460 Clip Type Lead with Temperature Sensor.
Comparator function
The comparator function enables you to set threshold values for internal resistance and voltage. This can determine battery wear more easily.
Large memory capacity
This instrument can store up to 6000 sets of data combining presently measured values (resistance, voltage, temperature, and comparator measurement results). It can be used to measure up to 12 units of 500-cell cubicles.
22
Features
Auto-memory function
Turning this function on stores the measurement values in the instrument's internal memory automatically, the instant when each set of data is held. This can lead to increased operational efciency.
PC interface
Measurement data can be loaded into a computer.
Bluetooth® Communication function
Users can view measurement values on their smart phones and tablets.
Model L2020 Pin Type Lead (optional)
The model L2020 pin type lead is L-shaped and handy to measure in conned location.
Model 9772 Pin Type Lead (optional)
Using the model 9772 Pin Type Lead with a pin tip designed to t in holes of φ5 mm enables measurement without removing terminal covers. Measurement is possible in virtually any location because the pin can be inserted diagonally in hard-to-reach places.
Model 9466 Remote Control Switch (optional) for
storing measurement values
The model 9466 Remote Control Switch makes it possible to hold and store measurement values by pressing a key. This is useful when both hands of the operator are busy.
1
2
3
4
5
6
7
Appx. Index
23
Names and Functions of Parts

1.4 Names and Functions of Parts

Front
Display (p. 27)
Measurement terminals
Control keys
(BT3554-01)
Measurement terminals
(1)
(p. 14, p. 33)
(1) SOURCE
terminals
(2) SENSE
terminals
24
(2)
The SOURCE side of the banana plug on the test lead is connected to this terminal.
The SENSE side of the banana plug on the test lead is connected to this terminal.
Control keys
Names and Functions of Parts
Pressing
Turns the comparator buzzer ON/OFF. Turns the comparator ON/ OFF.
Turns the auto-hold and auto-memory features ON/ OFF.
Displays the clock. Adjusts the clock.
Selects a conguration
setting.
Changes the value. Uses the right or left key to select a digit.
Turns on the memory
storage.
Stores measurement value when it is retained.
Conrms the settings.
Holds or cancels the measurement values.
Reads or cancels stored measurement values.
Deletes the settings.
Pressing and
holding (at
least 1 second)
Turns the power ON/OFF.
Sets comparator threshold values.
Turns off the
memory storage.
Deletes the last
stored data.
Pressing and
holding the key
while turning on
the power
Displays the setup screen for the
disconnection detection function.
Turns the Bluetooth Communication function ON/OFF (for BT3554-01). Displays the Serial No.
Displays the APS setup screen.
Displays the system
reset screen.
1
2
3
4
5
®
6
7
Appx. Index
25
Names and Functions of Parts
Pressing
Turns the backlight ON/ OFF.
Changes the resistance
range.
Changes the voltage
range.
Pressing and
holding (at
least 1 second)
Starts or cancels
zero adjustment.
(press and hold for at least 2 seconds.)
Pressing and
holding the key
while turning on
the power
Displays the setup screen for the noise frequency avoidance function.
Displays all LCD screen elements.
26
Display
Names and Functions of Parts
Data No.
Resistance measurement value and units
Voltage measurement value and units
Zero adjustment ON
Clock display and setting
Comparator buzzer ON
Now communicating
Bluetooth® function ON (only for BT3554-01)
Auto Power Save ON
Battery level
Auto-hold ON
Auto-memory ON
Noise frequency avoidance function ON
Input overow
Retain measurement value
Indicators other than those shown above may also light up when all the indicators on the display are shown, but only the above indicators are used by the instrument.
Temperature measurement value and units
Saved memory number
Read memory number
Comparator number
Selected memory
number is in use
Setting each function
For setting comparator threshold values
For noise frequency avoidance function
Comparator ON
PASS result
WARNING result
FAIL result
1
2
3
4
5
6
7
Appx. Index
27
Names and Functions of Parts
Top view
(1) (2) (3)
(1) EXT.HOLD terminal Connects the model 9466 Remote
(2) USB terminal Connects the USB cable.
(3) TEMP.SENSOR terminal Connects the miniplug of the model 9460
Control Switch (optional).
Clip Type Lead with Temperature Sensor (optional).
Rear view
Fuse cover
Tightening screw
Serial No.*
Bluetooth (only for BT3554-01)
®
label
Battery cover
* The serial number consists of 9 digits. The rst two (from the left)
indicate the year of manufacture, and the next two indicate the month of manufacture. Required for production control. Do not peel off the label.
(p. 17)
(p. 17)
28

1.5 Dimensions

Dimensions
1
60.6±3 mm
132±3 mm
2
199±3 mm
3
4
5
6
7
Appx. Index
29
Dimensions
30

Measurement Preparations

2

2.1 Attaching the Neck Strap

Operators can hang the instrument around their neck by attaching the neck strap. Attach the neck strap as described below.
Switch off the instrument and
1
remove the test leads.
Pass the neck strap through
2
the 2 attachments and fasten it in place with the buckles (2 on each side of the instrument).
Adjust the length of the neck
3
strap.
The instrument can be placed in the carrying case even with the neck strap attached.
1
2
3
4
5
6
7
Appx. Index
31
Installing/Replacing Alkaline Batteries

2.2 Installing/Replacing Alkaline Batteries

When using the instrument for the rst time, insert 8 LR6 (AA) alkaline batteries. Before attempting measurement, check to make sure that the battery level is sufcient. If the battery level is low,
replace the batteries with new ones.
• The indicator ashes when alkaline battery voltage
becomes low. Replace the batteries as soon as possible.
• In this document, the "alkaline battery" means the LR6 (AA)
battery to run the instrument, and the "battery" refers to the
measurement target.
Switch off the instrument and
1
remove the test leads.
1
Open the alkaline battery cover
2
on the rear of the instrument.
Insert 8 alkaline batteries, taking
3
care of the proper polarities.
Replace the alkaline battery
4
cover.
32
3
4
2
Connecting the Test Lead

2.3 Connecting the Test Lead

WARNING
To avoid electric shocks, be sure to connect the test leads properly.
Connect the test leads to the instrument. Be sure to connect all 4
terminals: SOURCE (+,−) and SENSE (+,−).
(Black)
(Red)
Align the  marks of the same color.
When using the optional model 9460 Clip Type Lead with Temperature Sensor, connect the miniplug to the TEMP.SENSOR terminal. For more information, see “Measuring the Temperature” (p.
53).
1
2
3
4
5
6
7
Appx. Index
33
Connecting the Test Lead

Connecting a Pin Type Lead and the Model 9466 Remote Control Switch

The Pin Type Lead (Models 9465-10, 9772, and L2020) and the optional model 9466 Remote Control Switch can be combined
together as shown below.
Connect the remote control switch to the probe of the lead, and join
the 2 cables using the supplied spiral tube.
(1) (2) (3) (4)
(L2020)
(4)(5)
(1) Model 9446 Remote Control Switch
(2) Probe
(3) Spiral tube (small)
Bundle up the center of the lead between the probe and junction
with a spiral tube.
(4) Junction
(5) Spiral tube (large)
Arbitrarily bundle up the lead between junctions.
34
(9465-10)
Turning the Power ON/OFF

2.4 Turning the Power ON/OFF

Press and hold the key (for at least 1 second) to turn the
power on or off. Check the clock settings when using the instrument
for the rst time.
1
Power: ON Power: OFF
The indicator ashes when alkaline battery voltage
becomes low. Replace the batteries as soon as possible.
2
3
4
5
6
7
Appx. Index
35
Clock Function

2.5 Clock Function

The date and time can be displayed by pressing the DATE key.
Check the clock settings when using the instrument for the rst time. The time is displayed using a 24-hour clock. The instrument's
calendar recognizes leap years automatically.

Turning the Date and Time Display ON/OFF

Press the DATE key to switch date-and-time display on or off.
Date and time display: ON Date and time display: OFF
Year
Month
Time
Day
(00:00 on January 1, 2016)

Adjusting the Date and Time

1
2 3
36
(Press the key for at least 1 second.) Enters the date and time setup mode.
Select the values.
Conrm the specied
values.
Date and time will not be set if you exit the clock setup screen without pressing the ENTER key.

Measurement

3
To ensure safe operation, be sure to read "Operating Precautions" (p.
12) before starting the measurements.
• Internal battery resistance varies considerably depending on charge or discharge status. To increase measurement accuracy, make measurements under similar conditions (for example, a fully charged battery).
• Lead-acid batteries (measuring objects) have high levels of terminal resistance. For this reason, resistance values may differ between the case and the tip of the terminal. Be sure to
connect the test lead to the terminals at a xed location.
For more information, see "Effects of Current Density" (p. Appx.5).
• Use the optional model 9460 Clip Type Lead with Temperature Sensor to measure the battery temperature. Or, use a non­contact thermometer, such as a radiation thermometer, for safety.
• Measurement may not be possible for insulated terminals, due
to insufcient ow of current for measurement. In such a case,
clean the terminal (remove the insulation) before measurement.
1
2
3
4
5
6
37
7
Appx. Index
Pre-operation Inspection

3.1 Pre-operation Inspection

Before using the instrument, verify that it operates normally to ensure that no damage occurred during storage or shipping. If
you nd any damage, contact your authorized Hioki distributor or
reseller.
Inspection item Method of checking
Is the fuse burned out?
Is the test lead disconnected?
Touch the test lead to the zero adjustment board.
If the resistance display still shows [−−−−], the fuse might be burned out or the test lead may be disconnected. If so, replace it with a new one.
Is the battery level
sufcient?
Inspecting batteries
38
At the upper right-hand area of the screen, indicates the present alkaline battery status. If
is displayed, the alkaline batteries need to be replaced soon. Be sure to have spare alkaline batteries available.
Measurement may not be possible for insulated
terminals, due to insufcient ow of current for
measurement. In such a case, clean the terminal (remove the insulation) before measurement.
Setting the Measurement Range

3.2 Setting the Measurement Range

Set resistance and voltage measurement ranges as described below.
Resistance range 3 m
Voltage range 6 V/ 60 V
Temperature range
/30 mΩ/300 mΩ/3
Ω
(Single range) Because temperature measurement uses a signal range, range setting is not required.
Ω
1
2
Press the Ω key or V key to display the present settings. Press the key repeatedly to cycle through the ranges.
Resistance range
3.000 m
Voltage range
6.000 V
When there is no activity for approximately 1 second, the settings on
the display get conrmed and returns to the measurement screen.
3.000
Ω
Ω
30.00 m
300.0 m
60.00 V
Ω
Ω
39
3
4
5
6
7
Appx. Index
Adjusting Zero Value (Zero Adjustment)

3.3 Adjusting Zero Value (Zero Adjustment)

The zero adjustment function displays subsequent measurement
results using the measured value (correction value) obtained when
it was performed as zero. The dened accuracy does not require zero adjustment to be
performed when using accessory or optional test leads, but the process should be performed in the following instances:
• When you wish to increase measurement accuracy*
• When using test leads that are not an accessory or option, or when using test leads whose length has been extended
* For the 3 m
on whether zero adjustment has been performed. For more information, see "8 Specications" (p. 89).
• Performing zero adjustment adjusts the zero points of all ranges.
• Even after the instrument is turned off, the correction values
are retained and the zero adjustment function is not canceled.
• After replacing the test lead, be sure to perform zero adjustment prior to measurement.
• Be sure to use the included or optional zero adjustment board
when performing zero adjustment.
• Be sure to keep the test lead shorted during zero adjustment.
• Keep the tip of the test lead away from the metal components.
range, the accuracy specications differ depending
Ω
40
Adjusting Zero Value (Zero Adjustment)

Shorting Methods for Various Test Leads

For Pin Type Leads
Use the included or optional zero adjustment board. The zero
adjustment can be achieved using the AC 4-terminal method.
Select 2 holes on the zero
1
adjustment board which are at the same distance as the terminals on the battery to be measured.
Push the test leads in a direction perpendicular to the
2
holes so that it is symmetrical to the central plus sign (+) on the zero adjustment board.
Model 9465-10 or L2020 Pin Type Lead
1
2
3
4
Center the plus sign (+).
Model 9772 Pin Type Lead
Insert the marked (engraved) side into the hole.
5
6
7
Appx. Index
41
Adjusting Zero Value (Zero Adjustment)
• Keep the zero adjustment board at least 10 centimeters away from the instrument.
• Be sure to use the included or optional zero adjustment board
when performing zero adjustment.
Be sure to connect each of the SOURCE and SENSE terminals by inserting the tip of the pin into the holes on the zero adjustment board. (See the gure below.)
SOURCE
SENSE
• Do not place the zero adjustment board on top of the battery or any metal. Electromagnetic induction effect could result in
unstable measurement values. In such a case, keep the zero
adjustment board away from any metal.
• Performing zero adjustment by connecting the tips of pin-
type leads or using a metal sheet other than the included zero
adjustment board will result in inaccurate adjustment of the
zero point.
• When the distance between the terminals on the battery (measurement target) is more than the distance between the
holes on the zero adjustment board, use the holes at both corners to perform zero adjustment.
• Consider the zero adjustment board to be a consumable. Replacing it with a new one after using it around 700 times is recommended.
42
Adjusting Zero Value (Zero Adjustment)
For Clip Type Leads
Perform zero adjustment by engaging red and black clips together.
Model 9460 Clip Type Lead with Temperature Sensor
SOURCE
SENSE
Model 9467 Large Clip Type Lead
SOURCE
SENSE
1
2
3
4
5
6
7
Appx. Index
43
Adjusting Zero Value (Zero Adjustment)

Performing Zero Adjustment

Check to ensure that the test
1
leads are connected properly.
Disconnect any leads connected to the measurement target.
Press the 0ADJ key for at least
2
2 seconds.
This enables the standby state for acquiring the correction values.
While [0AdJ] is ashing, short
3
the test leads using the zero adjustment board.
For more information, see "Shorting Methods for Various Test Leads" (p. 41).
If the test leads are not shorted while the display is blinking, it will result in an error.
(Black) (Red)
1
2
Press for at least 2 sec.
Flashes for approx. 10 secs.
For pin type leads
44
Adjusting Zero Value (Zero Adjustment)
The instrument automatically begins obtaining correction values. When the zero adjustment operation is complete, [0ADJ] is lit up and the instrument returns to the measurement mode:
1
2
3 mΩ range 30 mΩ range
Zero adjustment
complete
• Keep the test leads shorted until the zero adjustment operation is complete.
• The zero adjustment starts even if a key is pressed after the test lead has been shorted.
300 mΩ range
3
range and voltage
Ω
range
3
4
5
6
7
Appx. Index
45
Adjusting Zero Value (Zero Adjustment)

Zero Adjustment Troubleshooting

Check items Solution
Is the fuse burned out? If so, replace it with a new fuse. (p. 109)
Do the correction values obtained exceed 300 counts in either resistance or voltage range?
Did you short the test leads properly while the instrument is in standby for correction values?
Ensure that the test lead is properly connected to the instrument.
The test lead may be disconnected. If so, replace it with a new one.
Try cleaning the zero adjustment board.
While the instrument is in standby for correction values (for approximately 10 seconds), short
the test leads using the zero adjustment board to perform zero adjustment.

Canceling the Zero Adjustment Operation

Pressing the 0ADJ key for at least 2 seconds while the zero adjustment function is active, cancels the zero adjustment operation.
Zero adjustment: ON Zero adjustment: OFF
Press for
at least 2
seconds.
46
Retaining the Displayed Values

3.4 Retaining the Displayed Values

• When the warning display or voltage is displayed as [−−−−], the values cannot be retained.
• Changing any of the settings cancels retaining.
• Turning off the power cancels retaining.
Measurement values displayed on the screen can be retained. Pressing the HOLD key will light up [HOLD], and retain the measurement values.
Retaining function: OFF Retaining function: ON

Canceling the Retaining State

Pressing the HOLD key again cancels the retaining state.
The auto-hold function can be used to automatically recognize
the stability of measurement values and retain them. For more information, see "6.2 Auto-hold Function" (p. 75).
1
2
3
4
5
6
7
Appx. Index
47
Retaining the Displayed Values
NO
NONO

When retaining with the Model 9466 Remote Control Switch

The optional model 9466 Remote Control Switch is available for the operation in the same way as when using the HOLD key.
Disconnect the test leads from
1
the battery (measurement target).
Insert the miniplug of the model
2
9466 Remote Control Switch into the EXT.HOLD terminal.
Connect the connectors of the
3
test leads to the instrument.
Press the PRESS button on the
4
model 9466 Remote Control Switch.
The measurement value is retained.
Model 9466 Remote Control Switch
Top of the instrument
(Black) (Red)
Canceling the retaining state
Press the PRESS button on the model 9466 Remote Control Switch, or the HOLD key on the instrument.
48
Determining Battery-wear Judgment Values

3.5 Determining Battery-wear Judgment Values

To determine if a battery is worn out, rst measure the internal
resistance of a new or good battery, and decide the limits for judging battery wear. As a battery wears out, its internal resistance increases to approximately 1.5 to 2 times (reference value) that of a new or good battery, and voltage values decreases to 90% of the default value. Use these values as guidelines when determining the battery wear judgment values.
Example of the battery wear judgment values
Default value
(PASS)
Warning limit
(WARNING)
Failure limit
(FAIL)
1
2
3
4
5
Resistance:
Voltage:
The above values are examples.
The above values vary depending on the manufacturer and battery model. For more information, see "1.1 Measuring Battery Wear" (p. 19).
0.5 m
Ω
2.0 V 1.8 V
0.75 m
Ω
1.0 m
Ω
6
7
Appx. Index
49
Measuring a Battery (Inspection)

3.6 Measuring a Battery (Inspection)

Prepare the measurement. (p. 31)
1
Set the resistance and voltage ranges. (p. 39)
2
Perform zero adjustment. (p. 40)
3
Connect the test leads to the battery.
4
Ensure that all pins are connected to the battery.
(Red)
+
Read the measurement values.
5
50
(Black)
Internal resistance of the battery
Voltage
Measuring a Battery (Inspection)
Use the measurement values to judge whether the battery
6
is worn out. Example:
Voltage
1
Default
value
(2.0 V)
Current
value
(1.8 V)
As shown above, this battery needs to be replaced.
To retain measurement values
To store measurement values
To load the stored data to a computer
To set threshold values and judge whether the battery is worn out
PASS
WARNING
Default value
(0.5 mΩ)
Refer to "3.4 Retaining the Displayed
Values" (p. 47).
Measurement values can be stored by pressing the MEMO key while the values are being retained.
Refer to "5.2 Storing Data in the Memory" (p. 66).
Refer to "Communications Function" (p.
83).
Based on the wear judgment values, threshold values can be set for judging whether the battery is worn out.
Refer to "Comparator Function (Evaluation by Threshold Values)" (p.
55).
Present value
(1.0 m
FAIL
FAIL
Resistance
)
Ω
2
3
4
5
6
7
Appx. Index
51
Measuring a Battery (Inspection)

Error Measurements

If [−−−−] is displayed and [OVER] ashes on the screen (at the same time, the maximum display values ashes), this does not
indicate an error.
[
] If [
−−−−
[OVER] display
and maximum display value
ashes
• The test lead is not connected correctly to the
• The resistance of the measurement target signicantly
• This indicates that it is measuring an exceeded
] is shown in the resistance display, the test
−−−−
lead is open.
Or, a failure such as abnormal current ow due to a
disconnected test lead prevents measurements.
measurement target.
exceeds the measurement range.
measurement range of either resistance, voltage, or temperature.
Note the maximum open-circuit terminal voltage of the instrument (approximately 5 V maximum) when measuring the resistance of a relay or a connector. There is a possibility that such
measurement may damage the oxidized coating on the connector
of the measurement target, leading to incorrect measurements.

Warning Display

In the event of an overvoltage input error, the [OVER] display and
maximum display value ash, the red backlight lights up, and the buzzer sounds.
52
Measuring the Temperature
NO
NONO

3.7 Measuring the Temperature

Use the optional model 9460 Clip Type Lead with Temperature Sensor to measure the battery temperature.
Connect the connector of
1
the model 9460 Clip Type Lead with Temperature Sensor to the instrument.
Connect the miniplug of the
2
model 9460 Clip Type Lead with Temperature Sensor to the TEMP.SENSOR terminal.
The instrument detects the temperature sensor and automatically displays the temperature.
1
2
(Black) (Red)
3
4
5
Top of the instrument
6
7
Appx. Index
53
Measuring the Temperature

Changing the temperature display unit

The units of temperature (°C or °F) can be changed.
1 2
3
4
The setting of the temperature unit is retained even after the power is turned off.
Turn off the instrument.
Display the setting of the temperature unit.
Press the key for at least 3 seconds.
Conrm the setting.
The instrument is restarted. The settings will not change if the power is turned off before the settings are applied.
54
Comparator Function
4
(Evaluation by Threshold Values)

4.1 Overview

The battery measurement values can be compared with the present threshold values using the comparator function to determine the ranges in which the values fall within: PASS, WARNING, or FAIL. Up to 200 comparator conditions can be set. Refer to “Measuring Battery Wear” (p. 19) for more information on how to determine the threshold values. A buzzer sounds when a measurement falls within the range of WARNING or FAIL under the default settings. For more information, see “4.4 Setting the Comparator Buzzer” (p. 63). A resistance warning limit, resistance failure limit, and voltage warning limit are available as the threshold values.
Voltage
PASS
Default value
Voltage
warning limit
WARNING
Default
value
Resistance
warning
limit
FAIL
FAIL
Resistance failure limit
Resistance
1
2
3
4
5
6
7
Appx. Index
55
Turning On the Comparator Function

4.2 Turning On the Comparator Function

1
2
3
Press the key.
The comparator number ashes.
Press the COMP key again to return to normal measurements.
Select the comparator number.
(A number from 1 to 200 can be selected.)
Conrm the settings.
The comparator function is now turned on.
When the comparator function is turned on, the instrument
changes to the range specied by the comparator settings.
56
Setting Threshold Values for the Comparator

4.3 Setting Threshold Values for the Comparator

Threshold values can be set for the comparator (resistance warning limit, resistance failure limit, and voltage warning limit).
1
Example: Threshold values for a battery with the default values* 0.4
Resistance warning limit: 0.6 Ω (1.5 times of the default value) Resistance failure limit: 0.8 Ω (2 times of the default value) Voltage warning limit: 1.8 V * The default value refers to the value of resistance and voltage for
a new battery or a battery in good condition.
Selecting a comparator number
1
2
3
and 2 V.
Ω
Press the key for at least 1 second.
The comparator number ashes.
Press the COMP key again to return to normal measurements.
Select the comparator number.
(A number from 1 to 200 can be selected.)
Conrm the settings.
The range setup screen is displayed.
2
3
4
5
6
7
Appx. Index
57
Setting Threshold Values for the Comparator
Setting the range
Select the resistance range.
1
2
3
Setting the threshold values
1
(To shift the decimal point)
Select the voltage range.
(To shift the decimal point)
Conrm the settings.
The resistance warning limit and
[WARNING] ashes.
Set the resistance warning limit.
2
3
58
Conrm the settings.
The resistance failure limit and
[FAIL] ashes.
Set the resistance failure limit.
4
5
Conrm the settings.
The voltage warning limit and
[WARNING] ashes.
Set the voltage warning limit.
Setting Threshold Values for the Comparator
1
6
• Voltage is determined using absolute values. Comparison is
• If you set the resistance failure limit to a value that is less than
Conrm the settings.
Returns to the measurement screen, with the comparator function turned on. The settings are now saved.
possible even if the positive and negative electrodes of the test lead are connected to negative and positive terminals, respectively. (Data is stored with the sign.)
the resistance warning limit as previously set, the warning limit will be set to the same value as the new failure limit.
2
3
4
5
6
7
Appx. Index
59
Setting Threshold Values for the Comparator
When a measurement value is determined as "PASS"
Voltage
PASS
Default value
Voltage
warning limit
Measurement
value
WARNING
FAIL
FAIL
Default value
Resistance warning limit
Resistance failure limit
When a measurement value is determined as "WARNING"
Voltage
Default value
Voltage
warning limit
PASS
WARNING
Default value
Measurement
value
Resistance warning limit
FAIL
FAIL
Resistance failure limit
60
Resistance
Resistance
Setting Threshold Values for the Comparator
When a measurement value is determined as "FAIL"
Voltage
Default value
Voltage
warning limit
PASS
WARNING
Default value
Resistance warning limit
FAIL
Measurement
value
FAIL
Resistance failure limit
Resistance

Comparison Table for the Comparator

The result is determined by the display and buzzer as shown in the following table:
Resistance
warning limit
Resistance
(low)
Voltage (high) PASS WARNING FAIL
Voltage warning limit
Voltage (low) WARNING WARNING FAIL
The boundary conditions are as follows:
Resistance
PASS
Voltage
WARNING
Resistance
warning limit
<
warning limit
Voltage
Resistance
<
WARNING
Voltage
PASS
Resistance failure limit
Resistance
(medium)
Resistance
failure limit
Resistance
(high)
Resistance
<
FAIL
1
2
3
4
5
6
7
Appx. Index
61
Setting Threshold Values for the Comparator
Examples of how to read the comparator output table
Example 1:
If the measured resistance is less than or equal to the resistance warning limit, and the measured voltage is more than or equal to the voltage warning limit, [PASS] is displayed.
Example 2:
If the measured resistance is more than the resistance warning limit and is less than or equal to the resistance failure limit, and the measured voltage is more than the voltage warning limit,
[WARNING] is displayed and the buzzer sounds.
When the resistance warning limit and resistance failure limit are set to the same value, the boundary conditions are as shown below:
Resistance
PASS
Resistance
warning limit
=
Resistance failure limit
Resistance
<
FAIL
62
Setting the Comparator Buzzer

4.4 Setting the Comparator Buzzer

The buzzer can be enabled in accordance with comparison results when the comparator function is used. The buzzer can be set to
sound in the following states. By default, the buzzer is congured to
sound when the comparison result is WARNING or FAIL. In addition to the buzzer, when the comparison result is WARNING or FAIL, the backlight is turned on red.
1
2
OFF
PASS (ON) The buzzer sounds when comparison results are PASS.
FAIL (ON)
When you press the key, the present comparator buzzer
setting is displayed. Press the key repeatedly to cycle through the settings.
When there is no activity for approximately 1 second, the settings
on the display will be conrmed and returns to the measurement
screen.
The key tone settings cannot be changed.
The buzzer will not sound regardless of comparison results.
The red backlight turns on, along with the buzzer sounds when comparison results are WARNING or FAIL.
OFF PASS
FAIL
3
4
5
6
7
Appx. Index
63
Canceling the Comparator Function

4.5 Canceling the Comparator Function

Pressing the COMP key when the comparator is turned on cancels the comparator function.
Comparator: ON Comparator: OFF
• The range keys cannot be used while the comparator function is turned on.
• If there are no measurement values, [−−−−] is displayed and a comparator judgment cannot be performed.
• Even when the power is turned off, the comparator settings are saved and the comparator will be restored to on when the power is again turned on.
64

Memory Function

5

5.1 Overview

This instrument can store up to 6000 sets of data combining presently measured values*. After measurement, saved data can be displayed or transferred to a computer. The structure of the internal memory is as follows: * Date and time, resistance, voltage, temperature, comparator
threshold values, and results of judgment

Memory Structure

Unit name
(12 units)
A 1 2 3 ... 498 499 500
B 1 2 3 ... 498 499 500
C 1 2 3 ... 498 499 500
D 1 2 3 ... 498 499 500
E 1 2 3 ... 498 499 500
F 1 2 3 ... 498 499 500
G 1 2 3 ... 498 499 500
H 1 2 3 ... 498 499 500
J 1 2 3 ... 498 499 500
L 1 2 3 ... 498 499 500
N 1 2 3 ... 498 499 500
P 1 2 3 ... 498 499 500
Memory number (500 cells)
1
2
3
4
5
6
7
Appx. Index
65
Storing Data in the Memory

5.2 Storing Data in the Memory

Pressing the MEMO key stores the present measurement values. Handy function: Auto-memory function (p. 76)
1 2
3 4
Turn on the memory function.
Select the memory number.
After a certain period of time of no activity, the settings on the display
are conrmed and the instrument
returns to the measurement screen. When the memory function is on, you can select the memory number at any time.
Conrm the settings.
Retain the measurement values.
When [−−−−] is displayed, the values cannot be retained.
66
5
Store the measurement values.
The measurement values are stored in the cell with the selected memory number. After the data is stored, the next available memory number is displayed.
The retaining is now canceled.
Storing Data in the Memory
1
2
• Pressing and holding the CLEAR key for 2 or more seconds can clear
the last stored data. However, this operation is possible only immediately after the data is stored.
• If [USED] is displayed with a memory number, that number will be
overwritten.
3
4
5
6
7
Appx. Index
67
Canceling the Memory Function

5.3 Canceling the Memory Function

To cancel the memory function when it is on, press and hold the
MEMO key for at least 1 second. [oFF] is displayed, and the
instrument returns to the normal mode.
Memory function: ON Memory function: OFF
(Press and
hold for
at least 1
second.)
Normal mode
68
Reading Out Stored Data

5.4 Reading Out Stored Data

The stored measurement values are displayed after they are read out.
1
2
3
• Press the DATE key to check the date and time when the data
• The comparator result for the data being read out is also
• You cannot select the number of the memory location in which
• If no data has been stored, [−−−−] is displayed in the memory
Display the memory readout screen.
Select the memory number to be read out.
The measurement values for the selected memory number are displayed.
To return to the measurement screen, press the READ key.
was stored.
displayed.
no data is stored.
number display area and the instrument returns to the measurement screen.
1
2
3
4
5
6
• For data measured using the optional model 9460 Clip Type Lead with Temperature Sensor, the temperature is also displayed.
69
7
Appx. Index
Clearing Stored Data

5.5 Clearing Stored Data

Clearing a Single Set of Data

1
2
3
4
Display the memory readout screen.
Select the number of the memory cell in which you want to clear data.
The measurement values for the selected memory number are displayed.
Press the key once.
If there is no activity for approximately 3 seconds, returns to the readout screen.
Conrm the settings.
The data in the cell with the selected memory number is now cleared.
70

Clearing Data from Each Unit

Clearing Stored Data
1
2
3
4
Display the memory readout screen.
Select the unit in which you want to clear data.
Press the key twice.
1
2
3
If there is no activity for approximately 3 seconds, returns to the readout screen.
Conrm the settings.
All data stored in the selected unit (500 cells) are cleared.
4
5
6
7
Appx. Index
71
Clearing Stored Data

Clearing All Data

1
2
3
Display the memory readout screen.
Press the key 3 times.
If there is no activity for approximately 3 seconds, returns to the readout screen.
Conrm the settings.
All data (12 units/6000 sets) is now cleared.
72

Other Features

6

6.1 Noise Frequency Avoidance Function

When the noise frequency avoidance function is used, internal resistance with reduced noise is measured automatically.
Turning the Noise Frequency Avoidance Function ON/OFF
1 2
3
4
Turn off the instrument.
Display the setup screen for the noise frequency avoidance function.
Select either [oFF] or [on].
Conrm the settings.
The instrument is restarted. The settings will not change if the power is turned off before the settings are applied.
(when off is selected)
(when on is selected)
1
2
3
4
5
6
7
Appx. Index
: Noise frequency avoidance function
73
Noise Frequency Avoidance Function
Canceling the Noise Frequency Avoidance Function
The function will be canceled when the power is switched off and turned on again.
• When the noise frequency avoidance function is turned on, the
time required for measurement may take longer. [FrEq] will
ash.
• It may not be possible to avoid all noise depending on the type of noise.
74
Auto-hold Function

6.2 Auto-hold Function

This function automatically recognizes the stability of measurement values and retains them. Press the A HOLD/MEMO key several times to display [A.HOLD].
1
Auto-hold
OFF
To cancel the retaining, press the HOLD key or the PRESS button on the model 9466 Remote Control Switch.
• When resistance is displayed as [−−−−], data is not retained automatically.
• Data will not be retained automatically when [OVER] and the
maximum display value of the resistance are ashing.
• The instrument retains and stores measurement values automatically when the auto-memory function is used together with this function.
• Use the auto-hold function together with the comparator function to determine if [OVER] (and the maximum display
value) is ashing due to a range setting error. Setting the
comparator buzzer to [FAIL] is also recommended. For more information, see “4.4 Setting the Comparator Buzzer” (p. 63).
Canceling the Auto-hold function
Press the A HOLD/MEMO key several times to hide [A.HOLD].
Auto-memory
Auto-hold and
Auto-memory
2
3
4
5
6
7
Appx. Index
75
Auto-memory Function

6.3 Auto-memory Function

This function automatically stores measurement values in memory immediately after they are retained. Press the A HOLD/MEMO key several times to display [A.MEMO]. At this point, the memory function is also turned on.
Auto-hold
OFF
Use the cursor keys to select the memory number of the location in which the data is to be stored. [USED] is displayed if a memory number containing data is selected.
• The instrument retains and stores measurement values automatically when the auto-memory function is used together with the auto-hold function.
• Use the auto-memory function together with the comparator function to determine if [OVER] (and the maximum display
value) are ashing due to a range setting error. Setting the
comparator buzzer to [FAIL] is also recommended. For more information, see “4.4 Setting the Comparator Buzzer” (p. 63).
Auto-memory
Auto-hold and
Auto-memory
Canceling the Auto-memory Function
Press the A HOLD/MEMO key several times to hide [A.MEMO].
76
Auto Power Save Function (APS)

6.4 Auto Power Save Function (APS)

The auto power save function can be used to reduce power consumption of the instrument. The instrument is turned off automatically if any of the following conditions continues for approximately 10 minutes with no key operation:
[−−−−] is displayed as a resistance value.
• Data is being retained. (Measurement is suspended.)
• A state other than measurement state. (Any of the setup screens or the data readout screen is displayed.)
• After completion of communication.
[APS] starts ashing 1 minute before the power turns off.
Turning the Auto Power Save function ON/OFF
1 2
Turn off the instrument.
Display the auto power save setup screen.
1
2
3
4
5
3
4
Select either [oFF] or [on].
Conrm the settings.
The instrument is restarted. The settings will not change if the power is turned off before the settings are applied.
(when on is selected)
6
7
Appx. Index
(when off is selected)
77
Backlight
• For continuous use of the instrument, set the function to off. (The default setting is on.)
• When the auto power save setup screen is displayed unintentionally, turn off the power and turn it on again. The instrument settings are restored unchanged.

6.5 Backlight

The instrument’s backlight can be turned on and off.
Backlight OFF Backlight ON
78
System Reset

6.6 System Reset

This can be used to restore the instrument to its default settings.
However, note that the following settings will not be cleared:
• Date and time
• Stored measurement data (6000 data sets)
• Comparator threshold values (200 sets)
• Temperature display unit
1 2
3
4
Turn off the instrument.
Display the system reset screen.
Select [yES].
Select [no] to cancel the system reset operation.
Conrm the settings.
The instrument is restarted.
1
2
3
4
5
6
7
79
Appx. Index
System Reset

Default Settings (Factory Defaults)

Resistance range 3.000 m
Voltage range 6.000 V
Zero adjustment function Disabled
Auto-hold function OFF
Auto-memory function OFF
Comparator function OFF
Comparator buzzer setting WARNING/FAIL (ON)
Auto Power Save function ON
Temperature display unit °C
Ω
• When the system reset screen is displayed unintentionally, turn off the power and turn on again. The instrument settings are restored without resetting the system.
• For more information on how to clear stored measurement data, see “5.5 Clearing Stored Data” (p. 70).
80
Battery Level Warning

6.7 Battery Level Warning

The battery level of the instrument is displayed at the upper-right corner of the screen.
1
2
Battery level
indicator
Battery fully charged.
Black charge bars start disappearing from the left as the battery gets drained.
Battery level is low. Replace the battery as soon as possible.
(Flashing) Battery is completely drained. Replace with new batteries.
• Using manganese batteries signicantly reduces the continuous operating time of the instrument.
• The battery level indicator does not function accurately when nickel-metal hydride batteries are used.
• The battery level indicator serves as an approximate guideline for the continuous operating time.
Battery status
81
3
4
5
6
7
Appx. Index
Battery Level Warning
82

Communications Function

7

7.1 Communicating with a Computer

By using the USB cable included with the instrument, you can send data to a computer or control the instrument from a computer. Users can manage the saved data on their smart phones and tablets. (Only for BT3554-01) For more information, refer to the manual of the CD shipped with the instrument.
The virtual COM port on the computer is used as the USB interface.
Send data
Control the instrument
• Insert the connector in the proper direction when connecting the USB cable.
[PC] is displayed on the screen when communication is in progress through the USB cable.
• Do not disconnect the USB cable when communication is in progress. The dedicated computer application software displays a warning message when communication is lost due to cable disconnection. Connect the disconnected USB cable again.
1
2
3
4
5
6
7
Appx. Index
83
Communicating with a Smart Phone or Tablet (Only for BT3554-01)

7.2 Communicating with a Smart Phone or Tablet (Only for BT3554-01)

The BT3554-01 is a battery tester with Bluetooth® low energy support. When the Bluetooth® function is enabled, you can review measurement data and create measurement reports on mobile devices (iPhone, iPad, iPad miniTM, iPad Pro, iPod touch, and Android™ devices). For more information about this functionality, see the help function in the application software GENNECT Cross.
Install the GENNECT Cross on
1
your mobile device. (p. 85)
Enable the Bluetooth® function
2
on the BT3554-01. (p. 86)
Launch the GENNECT Cross
3
and pair it with the BT3554-01. (p. 87)
Select the general
4
measurement or battery function. (p. 88)
84
Communicating with a Smart Phone or Tablet (Only for BT3554-01)

Installing the smartphone app GENNECT Cross

Search for “GENNECT Cross” on the App Store from your iPhone, iPad or other Apple device, or on Google Play™ from your Android device. Then download and install the GENNECT Cross. You will need an Apple ID to download the app on the App Store, or a Google account to download the app on Google Play. For more information about how to register an account, contact the store at which you purchased your device.
• Because the BT3554-01 emit radio waves, use in a country or region where they have not been approved may be
subject to nes or other penalties as a violation of applicable
laws or regulations. For more information, see the attached “Precautions Concerning Use of Equipment That Emits Radio Waves” or go to our website.
• BT3554-01 availability is limited to certain countries. For more information, contact your authorized Hioki distributor or reseller.
• Bluetooth® communications range varies greatly with distance from obstructions (walls, metal obstruction, etc.) as
well as distance from the oor or ground. To ensure stable
measurement, verify adequate signal strength.
• Although this app is provided free of charge, downloading or use of the app may incur Internet connection charges. Such charges are the sole responsibility of the user.
• This app is not guaranteed to operate on all mobile devices.
1
2
3
4
5
6
7
Appx. Index
85
Communicating with a Smart Phone or Tablet (Only for BT3554-01)

Turning ON/OFF the Bluetooth® function

1 2
3
4
is displayed when the Bluetooth® function is on.
ashes when the instrument is connected to a mobile device.
Turn off the instrument.
Display the Bluetooth® setup screen.
Select [on].
To turn off Bluetooth®, select [oFF].
Conrm the settings.
The instrument is restarted. The settings will not change if the power is turned off before the settings are applied.
(when off is selected)
(when on is selected)
86
Communicating with a Smart Phone or Tablet (Only for BT3554-01)

Pairing the app with the battery tester (BT3554-01)

2
3
1
2
3
1
• When the app is launched for the rst time (before being paired with any instrument), the connection setup screen will be displayed.
• While the mobile device is displaying the connection setup screen, simply move it close to the BT3554-01 to automatically pair it with the instrument (the app can be paired with up to 8 instruments).
• Allow about 5 to 30 seconds for the instrument to pair with the app after being turned on. If the instrument fails to pair within 1 minute, relaunch GENNECT Cross and cycle the instrument’s power.
87
4
5
6
7
Appx. Index
Communicating with a Smart Phone or Tablet (Only for BT3554-01)

Making measurements with the Bluetooth® function

Select either the general measurement or battery measurement function on the home screen and measure. For more information about each function, see the help function in the GENNECT Cross.
General measurement
function
88
Battery function
(List display function)
Battery function
(Graph display screen)
Specications
8
8.1 General Specications
Operating environment
Operating temperature and humidity
Storage temperature and humidity
Standards Safety EN 61010
Dielectric strength 1.5 kV AC (cutoff current 5 mA, 1 minute)
Power supply LR6 (AA) alkaline batteries × 8
Continuous operating time
Backup battery Approx. 10 years (at 23°C)
Interface USB, Bluetooth
Dimensions Approx. 199W × 132H × 60.6D mm
Mass BT3554: Approx. 937 g (33.1 oz.) (including
Product warranty period
Indoors, Pollution Degree 2, altitude up to 2000 m (6562 ft.)
Temperature 0°C to 40°C (32°F to 104°F) Humidity 80% RH or less (no condensation)
Temperature −10°C to 50°C (14°F to 122°F)
Humidity 80% RH or less (no condensation)
EMC EN 61326
Between all measurement terminals and USB terminal
Rated supply voltage: 1.5 V DC × 8 Nickel-metal hydride batteries can be used. (However, the battery level indicator is not supported.)
Approx. 8.5 hours (when alkaline batteries are used, may vary depending on usage conditions)
®
(only for BT3554-01)
(7.83"W × 5.20"H × 2.39"D) (protector attached)
BT3554-01: Approx. 947 g (33.4 oz.) (including
3 years
batteries and protector)
batteries and protector)
1
2
3
4
5
6
7
8
Index
89
General Specications
Fuse 250 V / F 630 mAH (216.630, Littelfuse, Inc.
Accessories p. 2
Options p. 4 to p. 6
Display LCD (monochrome, 182 segments)
product)
90
Loading...