Hikvision DS-2TP21B-6AVFW User Manual

042120NA
How To Configure Highly Accurate
DS-2TP21B-6AVFW Thermographic Handheld Camera
1. General Parameters
86°−113° F (30.0°−45.0° C)
Temperature Measurement Accuracy
0.9° F (±0.5° C)
Camera Resolution
Thermal: 160 × 120, Optical: 640 × 480
Operating Environment A stable indoor environment without wind or direct sunlight Working temperature: 50°−95° F (10°−35° C)
2. Installation
Installation Cautions
The performance of this temperature-screening scheme is greatly affected by the environment. This scheme would apply only to indoor environments or scenarios with calm air and consistent temperature. Besides, the relative installation location of devices also greatly affects the temperature measurement accuracy. In order to improve measurement accuracy, the installation environment has to meet certain requirements:
- Select installation environments with one-direction path to ensure that cameras
capture the full faces of all passing persons. It is recommended that people stop movement during measurements.
- Select indoor environments with calm air and consistent temperature condition.
Outdoor environments with rapid temperature changes are not recommended.
- If this scheme is used in entrance scenes that connect indoor and outdoor
environments, it is suggested that the installation location be kept at a certain distance from the entrance (such as customs or security checkpoints). Persons coming in from outdoors should stay indoors for more than five minutes before the measurement. In these ways, the influence of the outdoor temperature on measured body surface temperature could be reduced.
18639 Railroad Street, City of Industry, CA 91748 Phone: +1 909-895-0400 Fax: +1 909-595-0788
Questions: techsupport.usa@hikvision.com www.hikvision.com 1
© 2020 Hikvision USA Inc. All Rights Reserved.
- Avoid objects with high temperature placed in the scene.
Camera Installation
- The camera should be set in front of the one-direction path, capturing the full faces
of passing persons, as shown below.
042120NA
Figure 1, Camera Placement and Positioning
- Recommended installation height is 4 ft 11 in5 ft 7 in (1.5 m1.7 m), namely the
camera should be set at the same height as human faces, or even be set at a lower elevation angle, thereby reducing error caused by other heat sources near the ground.
- The distance between the camera and measured object is recommend to be set to
3 ft 4 in6 ft 7 in (1 m2 m) normally.
- Avoid any other object whose temperature is higher than a human body in the
camera view background.
- This handheld thermographic camera can be mounted on a tripod without extra
adapters. There is a tripod offered by Hikvision for fixed placement, but it requires additional purchase. As domestic suppliers have not fully restored production capacity recently, there are multiple suppliers, and there may be slight differences under the same model, which will not affect the use. If you are concerned about this, we recommend that, except for tripod adapters that should be purchased together with the products, tripods that meet the standards could be purchased locally.
18639 Railroad Street, City of Industry, CA 91748 Phone: +1 909-895-0400 Fax: +1 909-595-0788
Questions: techsupport.usa@hikvision.com www.hikvision.com 2
© 2020 Hikvision USA Inc. All Rights Reserved.
3. Configuration
Figure 2, Physical Features
042120NA
User Interface
The User Interface of this handheld thermographic camera is as shown below.
Figure 3, Interface
Configuration Notes
1. Go to Settings >> Local Settings >> Thermography Settings.
2. Set Emissivity to 0.98.
3. Set Distance to the actual distance between the camera and measured object.
18639 Railroad Street, City of Industry, CA 91748 Phone: +1 909-895-0400 Fax: +1 909-595-0788
Questions: techsupport.usa@hikvision.com www.hikvision.com 3
© 2020 Hikvision USA Inc. All Rights Reserved.
Figure 4, Local Settings > Thermography Settings
042120NA
Figure 5, Local > Thermography > Distance > Custom
4. Go to Thermography.
5. Set Hot Spot, Center Spot, and other rules as required.
Figure 6, Hot Spot
6. Go to Palettes.
7. Set Alarm Temperature (Above) as required. When the measured value exceeds the
set value, the displayed Max Temp. Value turns red and flashes, as shown below.
18639 Railroad Street, City of Industry, CA 91748 Phone: +1 909-895-0400 Fax: +1 909-595-0788
Questions: techsupport.usa@hikvision.com www.hikvision.com 4
© 2020 Hikvision USA Inc. All Rights Reserved.
8. Adjust Focus Knob, and make sure the live view is clear during measurement.
4. Other Notes
Figure 7, Alarm Temperature
042120NA
Before this handheld device is used for actual body temperature measurement, run for five minutes for preheating.
The measured human body temperature is the same as the measured Max Temp. Value in this scheme; make sure that the Max Temp. Point is on the human face in the camera view during measurement.
This product is used for preliminary screening of people with elevated skin temperature. If an alarm occurs, use a specialized medical thermometer as a further check.
All of the above Installation and Configuration descriptions are based on the device being used with fixed placement. Other uses as a portable handheld device or for outdoor measurements are not recommended normally. Under these circumstances, the temperature measurement accuracy may not meet the same standard.
18639 Railroad Street, City of Industry, CA 91748 Phone: +1 909-895-0400 Fax: +1 909-595-0788
Questions: techsupport.usa@hikvision.com www.hikvision.com 5
© 2020 Hikvision USA Inc. All Rights Reserved.
042120NA