Heraeus PALABOWL Instructions For Use Manual

Bone Cement Mixing System
Instructions for use
EN
PALABOWL
®
1. Device name
1) Device name: PALABOWL® Bone Cement Mixing System
2) Type: CL006-01
3) REF: 5050166
2. Indications
1) The Bone Cement Mixing System is for mixing of the bone cement powder and liquid monomer.
2) This device can mix up to three units (total 120 g powder) of bone cement of any viscosity.
3. Working principles
The Bone Cement Mixing System’s rotating axis paddle allows for a thorough cement mix of all angles without unmixed powder collecting at the edges of the bowl. While mixing, vacuum removes air via a filter from the bowl to reduce porosity of the cement and cement fume.
4. Specification
1) Mixing bowl size (ID×H): 106mm × 72 mm; Vacuum hose size (length): 2.4 m.
2) Net weight: Not more than 0.3 kg.
3) Air tightness: There is no leak under negative pressure 75 kPa ± 1k Pa.
4) Vacuum hose: The vacuum hose shall not shrink under negative pressure 75 kPa ± 1 kPa.
5) Mixing paddle: The Mixing paddle can rotate freely under negative pressure 75 kPa± 1 kPa.
6) Sterilization: Sterilized using ethylene oxide before delivery.
7) Valid period: Two years.
8) The maximum allowable negative pressure value during operating: 73 kPa (550 mmHg).
9) To reach the negative pressure value 73 kPa (550 mmHg)
needed time: about 25 s.
5. Components
1) This device consists of Vacuum Mixing bowl, Mixing paddle, Spatula, Curette, Vacuum hose and Rotating handle.
2) Associated devices: The Bone Cement and Vacuum extractor are not included in this device.
6. Warning and Precautions
1) This device shall only be operated by qualified
professionals.
2) The bone cement powder and liquid monomer are not included in this device but prepared by the hospital. The bone cement must be prepared according to the instructions for use provided by the bone cement
manufacturer.
3) The Vacuum extractor is not included in this device, but prepared by the hospital. The negative pressure within the range of 5 kPa to 75 kPa shall be adjustable.
4) This device is sterilized using ethylene oxide before delivery. Sterilization validity is two years.
5) This device is for single patient use only. Do not reuse and
do not re-sterilize.
6) Follow current local/national regulations to properly dispose the device after use.
7) During transportation of this device, prevent severe shock, vibration, rain/snow, direct sunlight, moist and pressure.
8) This device should be stored in a non-corrosive, cool, dry,
and ventilated environment.
7. Operating instructions
1) Inspect the package prior to each use. Do not use if packaging is damaged.
2) Read the instructions for use, attached documents andlabels.
3) Open the package, take out the product and operate the device according to following figures:
08611
(5) (3) (4)
(6)
(1)
(2)
2
(1) Mixing bowl (2) Mixing paddle
(3) Spatula
(4) Curette (5) Vacuum hose (6) Rotating handle
1.
Place the bowl on a flatsurface and open thelid.
2.
1
Connect the shorter end of the vacuum hose to the vacuum pump and the opposite end (longer end) to the lid. If conditions permit, adjust negative pressure to the optimal range
2
between 65 kPa to 73kPa.
3.
5.
2
7.
2
1
Pour the single compo­nents into the bowl according to the instruc­tions provided by the
cement manufacturer.
Start the vacuum and begin mixing for about 45seconds or according to the instructions for use of the bone cement.
1
Turn the lid counterclock­wise and remove the lid.
4.
1
2
6.
2
8.
Close the lid and twist clockwise to lock.
Once mixing is complete, stop the vacuum and remove the vacuum hose.
1
To scrape the bone cement on the mixing paddle into mixing bowl by using spatula or
curette.
9.
Remove the thoroughly mixed bone cement with the spatula.
3