User interface "Sepacontrol"
User interface "Sepacontrol"
The user interface "Sepacontrol" from Heraeus is designed for programming and using the operating functions. The present description substitutes the corresponding section of your operating instructions.
The figure below shows the user interface of the Megafuge 1.0RS, which is described in detail below. On
the last page of the Appendix you will find the same figure enlarged to full-page size.
Key-operated switch
left and right digit selection, respectively
setpoint
display,
temperature,
overtemperature
setpoint setting setpoint setting
setpoint
change
set
+
–
enter
copying
into working
memory
open lid
lid
quick
run
start
stop
INSTRUMENTS
start/stop key
lid release
Quick Run
release
Quick Run
"ready for
start"
display
stop
release
system
check
programming
error
over-
temperature
key-operated switch
progr.mode
C
BA
Megafuge 1.0RS
program number selection
diagnostic
LEDs
lid open
check lid
imbalance
system check
progr.error
overtemp.
storage in memory
program display
unbalance
display
display
acceleration/
braking
curves
store
mode
accel
decel
select
call
program
storage
mode
program
execute select select select
RCF,
rotor type,
radius,
special
messages
speed
rcf delay ∆t
rotor
radius
"select" keys
run time,
start
delay
time temp
run time
start delay
time temp
operating temperature
overtemperature
acceleration
deceleration
accel
decel
speed
RCF
rotor type
radius
speed
rcf delay ∆t
rotor
radius
assignment LEDs
You can use the key-operated switch to set the operating mode of "Sepacontrol".
Position A: Unlimited programming. You can call or change all programs.
Position B: The program memory is protected against modifications. However, the current centrifugation
program in the main memory can still be modified.
Position C: Program and main memory are protected against modifications. The preset centrifugation
program can be run an unlimited number of times by actuating the control keys.
The setpoints can be displayed any time by pressing "set" or the select key ("select").
You can pull out the key in any position. The respective setting is then valid until the next modification.
If you are in a hurry:
Setting the centrifugation conditions without storing the parameters.
Prerequisites: The centrifuge is properly installed, the rotor correctly loaded and balanced, the keyoperated switch in position A or B.
2
User interface "Sepacontrol"
General procedure for all "select" fields:
• Press the "set" key. The display selected for the previous run flashes.
• Press the "select" key under the desired function field until the red LED next to the parameter to be
changed (e.g. "temp") is lit. Choose the digit to be altered by briefly pressing one of the direction keys
to the left and right of the "set" key (
+
count by pressing the
- or – key until the desired number is displayed.
set
) (repeatedly if necessary), and increase or decrease the
• By briefly pressing the key you increase or decrease the chosen digit by one step at a time. If you
press and hold the chosen key, the display changes continuously upward and downward, respectively. Release the key as soon as you have reached the desired value, and fine-tune if necessary by
repeatedly pressing the key.
• Acknowledge each entry by pressing "enter".
To enter parameters:
1. Press first the "set" and then the "select" key under the four-function field ("speed", "rcf", "rotor", "radius") until the red LED next to "rotor" is lit, and enter the order number of the rotor (only possible when
centrifuge is at rest).
Entering the rotor order number is mandatory. The built-in electronics uses this information to rec-
ognize the maximum permissible speed and sees to it that it is not exceeded.
2. Optional: Choose one of the accelerating and braking curves ("accel" and "decel", respectively, possible choices no. 1–9 for both, plus 0 for brakeless deceleration with "decel", see Appendix).
NOTE: If with the braking curve preset to 0, the set speed is lowered during the run, the newly set speed is
triggered with braking curve 1; the preset curve 1 remains unchanged.
3. Optional: Enter the centrifugation radius. You can set the radius in steps of 10 mm and 1 mm, respectively, from the nominal radius (maximum value) to the smallest radius compatible with the design.
When preselecting an RCF value (see below), the set RCF value corresponds to the radius entered here.
4. Set the speed, or alternatively the relative centrifugal force (RCF, in multiples of the earth gravity g). To
do this, choose the respective parameter ("speed" or "rcf") in the four-function field ("speed", "rcf", "rotor", "radius"), and enter the desired value.
5. Set the run time. Select "time", and enter the desired run time. The time measured is the one between
the start and the end of the braking process.
h.min: Entry and display in hours and minutes, between 0.00 and 9.59
h: Entry and display in hours, between 10 and 99
hld Continuous operation (hld = "hold")
6. Optional: Enter the desired start delay (1 min to 99 hours, display: "delay").
7. Only refrigerated centrifuges (MEGAFUGE series RS): Set the temperature to a value between 0 °C
and 40 °C.
8. Optional (only refrigerated centrifuges – MEGAFUGE series RS): Set the overtemperature control to
the desired value (minimum 5 K, upper limit set by the built-in microprocessor). The centrifuge switches
automatically off when the overtemperature is exceeded (setpoint + ∆t).
NOTE: Altered parameters are accepted by the main memory only if acknowledged with "enter" (or "start").
Otherwise, following a timeout of about 2 min (or about 20 s with the centrifuge running), the instrument
switches the display to the actual values without accepting the modifications.
Operation without program
1. Close the centrifuge lid if still open. The yellow LED next to the "lid" key is lit, the "ready for start" display next to the start/stop key (green LED) is activated, and the centrifuge is ready for use.
2. Press the start/stop key. The centrifuge starts according to the factory-set default value with the acceleration curve no. 9 (acceleration curves see Appendix). The "lid" LED goes out, the lid cannot be
opened any more. The very moment the centrifuge starts, the stop function is released (red LED next to
the start/stop key is lit).
3. Wait until the instrument stops automatically after completing the set run time ( braking curve, see Appendix). However, you may stop the centrifugation any time by pressing the start/stop key. At this point
the green LED is lit anew, e. g. you can continue a manually interrupted run at any time.
3
User interface "Sepacontrol"
Turning off the continuous tone after centrifugation: any key for current run, "enter" key in actual value
mode for all subsequent runs (in this case brief display "Alert off"; renewed "enter" or mains supply
ON/OFF results in "Alert on").
Display
Actual values
If none of the display panels flashes, actual values are displayed; this is true both during a centrifuge run
and at rest. The display values changes as soon as the instrument registers new actual values (e. g.
speed during a run, temperature at rest, etc.). Following a centrifuge run, the time display reads the remaining run time (which is different from 0 only in case of premature STOP) or the run time passed (if "hld"
has been preselected). This gives the operator an opportunity to check whether the run has been prematurely interrupted e. g. by a short-term power failure. The actual time display is overwritten with the new
actual values only following a renewed start or power OFF/ON.
By pressing the respective "select" key you can switch between the parameters (acceleration or deceleration, speed, RCF etc.). The LED for the currently displayed parameter is lit (see figure under "assignment
LEDs"). Further explanations see in the following section under "setpoint values".
Setpoint values
In order to display the current setpoint values, always press first the "set" key (display panels flash) and
then the key listed below as required.
Program display displays the number of the currently active program after pressing "store
mode" key (01–32)
acceleration/
braking curves
speed, RCF, rotor type,
radius
displays the number of the acceleration curve (1–9), upon pressing the
"select" key the number of the braking curve (0–9) (see Appendix)
displays the set speed, and upon pressing the "select" key successively
the relative centrifugal force (in multiples of the earth gravity g), the rotor
type (identical to the rotor or bucket order no.) and the radius setting
run time, start delay displays the set run time and upon pressing the "select" key the start
delay (if set)
temperature, overtemperature displays the temperature setpoint, and upon pressing the "select" key the
overtemperature control (if set; lowest value 5 K)
4