Heraeus HeraSafe Safety Cabinet User manual

Page 1
Operating Instructions
Safety Cabinet HERAsafe KS / KSP
Valid: 12.2004 / 50077476 C
Page 2
Operating Instructions Safety Cabinet HERAsafe KS / KSP
Copyright
These operating instructions are protected by copyright. Rights resulting the­reof, particularly reprint, photomechanical or digital postprocessing or reproduc­tion, even in part, are only allowed with the written consent of Kendro Laboratory Products GmbH. This regulation does not apply to reproductions for in-plant use.
Trademarks
®
HERAsafe
and Heraeus are registered trademarks of Kendro Laboratory Pro­ducts GmbH. All other trademarks mentioned in the operating instructions are the exclusive property of the respective manufacturers.
Kendro Laboratory Products GmbH Robert-Bosch-Straße 1 D - 63505 Langenselbold Germany
2
Page 3
Operating Instructions Safety Cabinet HERAsafe KS / KSP
Contents
1. General notes ......................................................................................... 6
1.1 Safety instructions .......................................................................... 6
1.2 Warranty ......................................................................................... 7
1.3 Explanation of symbols................................................................... 7
1.3.1 Symbols used in the operating instructions: ............................ 7
1.3.2 Symbols on the device: ............................................................ 8
1.4 Use of the device ............................................................................ 9
1.4.1 Correct use ............................................................................... 9
1.4.2 Incorrect use: ............................................................................ 9
1.5 Standards and safety regulations ................................................. 10
2. Delivery ................................................................................................. 11
2.1 Standard components ................................................................... 11
2.2 Acceptance inspection .................................................................. 11
3.1 Ambient conditions ....................................................................... 12
3.2 Room ventilation ........................................................................... 12
3.3 Correct location ............................................................................. 13
3.4 Installation in series ...................................................................... 13
3.5 Transport....................................................................................... 13
4. Description ........................................................................................... 14
4.1 Overall view .................................................................................. 14
4.2 Safety system ............................................................................... 17
4.3 Filter system ................................................................................. 18
4.4 Controls and display ..................................................................... 20
4.5 Sample chamber access .............................................................. 21
4.6 Device interfaces .......................................................................... 22
4.7 UV lamp unit ................................................................................. 24
4.8 Working area................................................................................. 24
5.1 Initial operation ............................................................................. 25
5.2 Installing unit and accessories ..................................................... 25
5.3 Levelling the cabinet ..................................................................... 28
5.4 Activating the remote control ........................................................ 28
5.5 Power supply connection ............................................................. 29
5.6 RS 232 interface connection ........................................................ 31
5.7 UV connection .............................................................................. 31
5.8 Installation test .............................................................................. 32
6. Handling and control ........................................................................... 33
6.1 Display .......................................................................................... 33
6.1.1 Functions of the display components ..................................... 33
6.1.2 Display during the calibration routine ..................................... 34
6.1.3 Display in OFF mode .............................................................. 34
6.1.4 Display in work mode ............................................................. 34
6.1.5 Power interruption .................................................................. 34
6.1.6 Display and functions after a power failure ............................ 34
6.1.7 Failure messages ................................................................... 34
6.2 Description of the operating modes ............................................. 35
6.3 Remote control ............................................................................. 37
6.3.1 Basic functions ....................................................................... 37
6.3.2 Moving the front window to the working position ................... 38
6.3.3 Silencing the audible alarm signal ......................................... 39
6.3.5 Activating and deactivating the internal power supply ........... 40
6.3.6 Displaying the UV disinfection time ........................................ 41
3
Page 4
Operating Instructions Safety Cabinet HERAsafe KS / KSP
Contents
6.3.7 Activating and deactivating the potential-free contact (optional)41
6.3.8 Switching the cabinet to OFF mode ....................................... 41
6.3.9 Setting the time ....................................................................... 42
6.3.10 Displaying the downflow velocity ........................................... 43
6.3.11 Displaying the operating hours of the HEPA filters ................ 43
6.3.12 Displaying the Kendro Performance Factor ........................... 44
6.3.13 Setting and activating the timer .............................................. 45
6.3.14 Deactivating the timer ............................................................. 46
6.3.15 Setting the UV disinfection time ............................................. 47
6.3.16 Starting the UV disinfection .................................................... 47
6.3.17 Cancelling the UV disinfection ............................................... 48
6.3.18 Activating the stop watch ....................................................... 48
6.4 Pilot switch .................................................................................... 49
6.4.1 Moving the front window: ....................................................... 49
6.4.2 Silencing the audible alarm signal: ........................................ 50
6.4.3 Switching the device to OFF mode ........................................ 50
7. Operation .............................................................................................. 51
7.1 Hygiene preparations for the sample chamber ............................ 51
7.2 Preparing the sample chamber .................................................... 51
7.3 Response to failure messages ..................................................... 52
7.4 Work rules ..................................................................................... 53
8. Shut-down............................................................................................. 54
8.1 Interrupting an operation .............................................................. 54
8.2 Shutting the unit down .................................................................. 54
9. Cleaning and decontamination .......................................................... 55
9.1 Decontamination procedure ......................................................... 55
9.2 Wipe/spray disinfection................................................................. 55
9.3 UV disinfection after a wipe/spray disinfection ............................ 57
9.3.1 UV disinfection using the integral UV lamps .......................... 57
9.3.2 UV disinfection using a mobile UV device ............................. 57
9.4 Disinfection with formaldehyde .................................................... 57
9.5 Cleaning the exterior surfaces ...................................................... 58
9.6 Cleaning the floorpan ................................................................... 58
10. Maintenance ......................................................................................... 59
10.1 Inspection...................................................................................... 59
10.2 Service .......................................................................................... 59
10.2.1 UV lamps ................................................................................ 59
10.2.2 Sample chamber illumination ................................................. 60
10.3 Retrofitting and repairs ................................................................ 61
11. Disposal ................................................................................................ 62
11.1 Disposal procedure ...................................................................... 62
12. Technical data ...................................................................................... 63
13. Device log ............................................................................................. 68
14. Certificate of decontamination ........................................................... 69
4
Page 5
Operating Instructions Safety Cabinet HERAsafe KS / KSP
Figures
Fig. 1 Device locations in a room ...........................................................13
Fig. 2 Lift points ...................................................................................... 13
Fig. 3-a Overall view / Version KS ............................................................. 15
Fig. 3-b Overall view / Version KSP ..........................................................16
Fig. 4-a Filter system with prefilters, downflow filter and exhaust
air filter / Version KS.....................................................................18
Fig. 4-b Filter system with prefilters, downflow filter and exhaust
air filter / Version KSP ..................................................................19
Fig. 5 Controls and display ..................................................................... 20
Fig. 6 Display with remote control sensor ..............................................20
Fig. 7-a Access through front cover ..........................................................21
Fig. 7-b Access through front window .......................................................21
Fig. 8 Supply interfaces .......................................................................... 22
Fig. 9 UV lamp unit .................................................................................24
Fig. 10 Working area on the workplate .................................................... 24
Fig. 11-a Rack installation / Version KS ......................................................25
Fig. 11-b Installation / Version KSP ............................................................. 26
Fig. 11-c Drain valve installation ................................................................. 27
Fig. 12 Inserting the batteries ................................................................... 28
Fig. 13 RS 232 interface connection ........................................................31
Fig. 14 UV connection ..............................................................................31
Fig. 15 Functions of the display components ..........................................33
Fig. 16 Basic functions of the remote control ........................................... 37
Fig. 17 Basic functions of the pilot switch ................................................49
Fig. 18 Sitting posture .............................................................................. 53
Fig. 19 Front cover seal replacement ...................................................... 60
5
Page 6
Operating Instructions Safety Cabinet HERAsafe KS / KSP
1. General notes
The following are the addresses of the international Kendro Sales Organisations.
Postal address Germany
Kendro Laboratory Products GmbH Robert-Bosch-Straße 1 D - 63505 Langenselbold
Anfragen aus Deutschland Telefon
Vertrieb 0800 1 536376 Service 0800 1 112110
Fax
Vertrieb/Service 0800 1 112114
E-Mail info@kendro.de
Enquiries from Europe, Middle East and Africa Tel. + 49 (0) 6184 / 90-6940 Fax + 49 (0) 6184 / 90-7474 E-Mail info@kendro.com
Postal address USA
Kendro Laboratory Products Inc. 275 Aiken Road Asheville, NC 28804 USA
Enquiries from North America Phone +1 800-879 7767 Fax +1 828-658 0363 E-Mail info@kendro.com
Enquiries from Latin America Phone +1 828-658 2711 Fax +1 828-645 9466 E-Mail info@kendro.com
Enquiries from Asia Pacific Phone +852-2711 3910 Fax +852-2711 3858 E-Mail info@kendro.com
1.1 Safety instructions
These safety instructions describe the safety features of the HERAsafe series and apply to the models KS 9/12/15/18. The safety cabinet has been manufactured in keeping with the latest techno­logical developments and has been tested before delivery for its correct func­tion. It may, however, present potential hazards if it is not used according to the intended purpose or outside of operating parameters. Therefore, the following procedures must always be observed:
The safety cabinet must be operated only by trained and authorized personnel.
For any operation of this unit, the operator must prepare clear and concise written instructions in the language of the operating and cleaning personnel based on these operating instructions, applicable safety data sheets, plant hygiene guidelines, and technical regulations, in particular:
which decontamination measures are to be applied for the cabinet and
accessories,
which protective measures apply while specific agents are used,
which measures are to be taken in the case of an accident.
Repairs to the device must be carried out only by trained and authorized expert personnel.
• The contents of the operating instructions are subject to change without further notice.
Concerning translations into foreign languages, the German version of these operating instructions is binding lease.
Keep these operating instructions close to the unit so that safety instructions and important information are always accessible.
Should you encounter problems that are not detailed adequately in these ope­rating instructions, please contact Kendro Laboratory Productsimmediate­ly for your own safety.
6
Page 7
1. General notes
1.2 Warranty
Kendro Laboratory Products warrant the operational safety and functions of the safety cabinet only under the condition that:
the device is operated and serviced exclusively in accordance with its intended purpose and as described in these operating instructions,
the device is not modified,
only original spare parts and accessories that have been approved by Kendro Laboratory Products are used,
inspections and maintenance are performed at the specified intervals,
an installation test is performed prior to the initial operation of the device and that a repeat test is performed on the occasion of all inspections and repairs.
The warranty is valid from the date of delivery of the device to the operator.
1.3 Explanation of symbols
1.3.1 Symbols used in the operating instructions:
Operating Instructions Safety Cabinet HERAsafe KS / KSP
WARNING!
is used if non-observance may cause serious or even lethal injuries.
CAUTION!
is used if non-observance may cause medium to minor injuries or damage.
NOTE!
is used for hints and useful information.
RECYCLING!
Valuable raw materials can be reused.
7
Page 8
Operating Instructions Safety Cabinet HERAsafe KS / KSP
1. General notes
1.3.2 Symbols on the device:
Betriebsanleitung beachten (Decke Schaltraum)
Observe operating instructions (cover electrical box)
Cytostatic agents (left device front)
Checked safety (Cover light dome)
T5A note (sample chamber fusing)
RS 232 interface (connection label)
Armrest installation (right side of light dome)
8
Page 9
1. General notes
1.4 Use of the device
1.4.1 Correct use
KS version:
The safety cabinet is a laboratory device for installation and operation in micro­biological and biotechnical laboratories of safety levels 1, 2, and 3. It has been designed as a Class II microbiological safety cabinet, in accordance with EN 12469 / 2000. Depending on the hazard level of the agents involved, the operator must pre­pare in writing appropriate decontamination procedures for the device and the accessories used in the sample chamber.
KSP version:
The safety cabinet also meets the requirements from DIN 12980 of 2/2004 and can be used for the production of cytostatic agents. For cytostatic agent applications, use the lowered one-piece working plate only.
Operating Instructions Safety Cabinet HERAsafe KS / KSP
Prior to the initial operation of the cabinet, the operator must perform an instal­lation test. The test result must be documented by a test report. The cabinet must only be released for operation if it is in compliance with the operating parame­ters specified by Kendro Laboratory Products.
After any changes to the installation conditions or after any modification to the technical system, a repeat test must be performed and the test result must be documented by a test report that shows that all operating parameters are in com­pliance with those specified by Kendro Laboratory Products.
1.4.2 Incorrect use:
The safety cabinet must not be used in laboratories that do not comply with the requirements of safety levels 1, 2, and 3. The unit must not be operated as a Class II safety cabinet, if:
no repeat test is performed after changes to the installation conditions or after modifications to the technical system,
the alarm system of the device has issued a failure message and the cau­se for the failure has not been repaired.
The alarm system must not be tampered with or disabled. If alarm system com­ponents heve been removed or disabled for service or repairs, the unit must only be released for operation if all alarm system components are functioning again properly.
The filters installed in the device are not capable of separating gaseous sub­stances. Therefore, do not work with or store substances in the device:
who in quantity or concentration are toxic,
if a reaction with other substances may result in hazardous toxic concentrations or formation of toxic gases,
that may form combustible or explosive mixtures in combination with air.
9
Page 10
Operating Instructions Safety Cabinet HERAsafe KS / KSP
1. General notes
1.5 Standards and safety regulations
The device complies with the safety requirements of the following standards and guidelines:
IEC 1010-1 / EN 61010-1
EN 12469 / 2000
DIN 12980 / 1996 (E DIN 12980 / 2'2004)
Low Voltage Directive 73/23 EWG
EMC Directive 89/336 EWG
10
Page 11
2. Delivery
2.1 Standard components
Delivery for the safety cabinet includes the following:
Version KS:
safety cabinet,
remote control,
drain valve.
Version KSP:
safety cabinet
remote control
armrests
rack
All versions:
Device documentation: — operating instructions, — factory test report.
Operating Instructions Safety Cabinet HERAsafe KS / KSP
Optional components and accessories are listed as separate items in the deli­very document.
2.2 Acceptance inspection
After the device has been delivered, immediately check the device:
for completeness,
for possible damage.
If the delivery is incomplete or if you detect any transport damage to the device, contact immediately.
the forwarding agency and Kendro Laboratory Products
11
Page 12
Operating Instructions Safety Cabinet HERAsafe KS / KSP
3. Installation
3.1 Ambient conditions
The operational safety and correct function of the unit depend on the location where it is to be operated. The safety cabinet must be operated only at locati­ons that meet the ambient conditions listed below.
Location requirements:
The electrical system of the device has been designed for an operating height of up to 2000 m above sea level.
The mains power supply outlets should be out of casual reach to prevent ac­cidental shut-off. Ideally, the outlets should be installed above the safety cabinet.
The flooring of the location must be adequately strong and not flammable.
The room in which the device is installed must be of adequately height. For units not connected to an exhaust system, the distance between the exhaust air opening and the room ceiling must be at least 200 mm (8 in).
The location must be equipped with an appropriate ventilation system (see Section 3.2.).
The temperature within the room must be between 15° C and 40° C (49° F and 104° F).
The relative humidity in the vicinity of the device must not exceed 90 %.
NOTE - Ambient conditions!
If ambient conditions vary from those descri­bed above, please contact Kendro Laborato­ry Products for assistance in installing the device.
NOTE - Temporary storage!
If the device is stored only temporarily (up to four weeks), the ambient temperature may be between -20° C and +60° C (-4° F and +140° F) at a relative air humidity of up to 90 %. For lon­ger storage periods, the location requirements apply.
3.2 Room ventilation
The room ventilation should preferably be a ventilation system that complies with the national requirements for the application.
The inlet air and exhaust air openings of the room ventilation must be located so that drafts are prevented from impairing the function of the safety cabi­net air system.
12
Page 13
3. Installation
3.3 Correct location
Choose a draft-free location where the safety cabinet does not interfere with the plant traffic. Fig. 1: This figure shows preferred locations for safety cabi­nets and unsuiable locations, not in accordance with the safety requirements. Unsuitable locations: The locations [1], [2], and [3] are not suitable because they are exposed to drafts from windows and doors. Location [5] is undesirable because it is in range of plant traffic and within the exhaust air range of a ventilation system [4]. Preferred locations [6], [7], and [8] are correct because they are in a draft-free section of the room and not exposed to plant traffic.
3.4 Installation in series
Operating Instructions Safety Cabinet HERAsafe KS / KSP
When several devices are to be installed in series, please ob­serve the following:
Make sure that vibrations cannot be transferred between adjacent units.
Exterior surfaces of the cabinets must always be accessible for cleaning and disinfection.
3.5 Transport
Version KS: Fig. 2: To prevent tilting, always transport the device using a
suitable carrier, even for a transport within a building, and se­parate it from the rack.
NOTE – Lift points
For transport, lift the unit only at the lift points shown in Figure 2. Do not allow the weight of the cabinet to rest on the floorpan!
Version KSP:
If required, the device frame must be removed from the rack to be reinstalled onto the rack for the initial start-up after the transport (see Section 5.2).
Locations in a room
Fig. 1
Fig. 2
Lift points
13
Page 14
Operating Instructions Safety Cabinet HERAsafe KS / KSP
4. Unit description
4.1 Overall view
Version KS:
Fig. 3a: Plenum assembly [5] with plenum for downflow blower [24] and plenum for exhaust air blower [2]. The downflow filter and the exhaust air filter are installed directly to the pertaining blower. The exhaust air is released into the environment around the device through the opening [1].
At the side of the plenum, the optional media supply lines [25] are routed into the sample chamber [20].
Electrical box [4] with power supply cable [3]. An RS 232 connection to a PC and two fuse holders are installed at the front of the electrical box.
Front cover [6] with integral, electrically movable front window [7], operated by:
the pilot switch [26],
the remote control [28]. The gas struts [22] secure the front cover in the open state. Optionally, two safety latches [27] can be installed to the front cover so that the door can be protected against unauthorized opening.
The light dome [23] with two tubes is part of the front cover assy.
Side slates [9] with two sealed access openings [19]. These can be equip­ped with media valves [13].
Internal outlets [10] for power supply of accessories and adapter [11] for mobile UV device.
Optional UV lamp unit [8] consisting of two UV lamps per side.
Optional stands [12], adjustable in height (KFS 2) and with fixed height (KFS 1).
Workplate segments [15] with optional arm rests [16]. A one-piece workplate and special workplates are available options.
Lockable drain valve [17] for installation into floorpan.
Display [21] with remote control sensor and alarm system LEDs.
Test hoses for the supply unit [18] at the left side of the sample chamber and for the exhaust unit [14] at the right side of the sample chamber.
14
NOTE – Test hoses!
Do not remove the two test hoses for checking downflow and exhaust air.
Page 15
4. Unit description
Operating Instructions Safety Cabinet HERAsafe KS / KSP
Overall view / Version KS
Fig. 3
15
Page 16
Operating Instructions Safety Cabinet HERAsafe KS / KSP
4. Unit description
Version KSP:
Fig. 3b: Rack [5] with integral prefilter housing [6].
• Filter inserts [2] for the prefilter housing. The number of supplied filter in­serts depends on the width of the device. To protect from liquid residues, the filter inserts are oriented slightly sloping from the device backpanel to form a drain duct to the floorpan [3]. The filter plates [4] have tongues and grooves and are inserted in an over­lapping pattern from right to left.
Membrane sleeve or opening for filling with aerosole for testing the filter of the downflow unit [1] at the left side of the sample chamber and of the ex­haust air unit [7] at the right side of the sample chamber.
16
Overall view / Version KSP
Fig. 3b
Page 17
4. Unit description
4.2 Safety system
The safety system comprises a combination of protective and alarm systems that ensure maximum personal and material protection.
Safety systems:
Vacuum-sealed air system A vacuum-sealed air system in combination with HEPA filters for downflow and exhaust air forms the basis of the safety system for personal and ma­terial protection.
Personal protection
Air aspired from the exterior along the entire working opening at a constant high velocity prevents that:
• agents may leak through the working opening of the chamber. As the exterior air pressure around the unit exceeds the pressure of the in­ternal air system (vacuum sealing), it is ensured that:
• agents cannot be released to the exterior in the case of a leak in the
cabinet housing.
Material protection
A steady airflow within the air system ensures that:
a constant downflow allows the HEPA filters to remove contaminants
so that the samples are always surrounded by ultrapure air,
harmful particles are not carried over through the sample chamber
(protection from cross-contamination).
HEPA filters The downflow (i.e. the air circulating within the device) and the exhaust air (air that is released to the exterior) are cleaned by HEPA filters (HEPA = High Efficiency Particulate Air Filter). For version KSP, a prefiltering system is used to increase the filter efficien­cy and to protect the exhaust air and downflow systems.
Safety lockout
To protect from UV radiation, the optional UV disinfection routine can be run only if the front opening is closed. During UV disinfection, the front opening safety lockout is activated and prevents harmful UV radiation from being emitted from the sample chamber.
Operating Instructions Safety Cabinet HERAsafe KS / KSP
Warning system:
Airflow monitoring
Airflow monitoring determines the velocity of the airflow in the sample chamber as well as the inflow velocity of the air aspired from the exterior through the working opening. As soon as airflow velocities move above or below a spe­cified safety value, a signal is transmitted to the alarm system.
Visual and audible alarm system
The warning system constantly monitors the safety-relevant device functions:
Inflow velocity of the air aspired from the exterior,
downflow velocity,
working position of the front window. If the warning system detects changes to one of these device functions, it issues:
an audible and a visual alarm signal.
17
Page 18
Operating Instructions Safety Cabinet HERAsafe KS / KSP
4. Unit description
Position monitoring The position sensors monitor the position of the front co­ver as well as the movement of the front window; it will in­dicate when the front window is in the working position.
Kendro Performance Factor The Kendro Performance Factor (PER) is a value that in­dicates the safety state of the safety cabinet. This value is calculated from data determined by the safety system and from values captured empirically by service personnel during safety checks. This data is entered into a parameter list of the control software and interconnected. The result can be indicated by the display.
4.3 Filter system
Version KS: Fig. 4: The filter system consists of two HEPA filters [2] and
[5] for the circulating air and for the exhaust air and of a coarse filter for the aspired air.
HEPA filters: Room air [10] is drawn into the sample cham­ber through the working opening. In the air duct, room air and the downflow within the chamber [7] are then blended to make up the blend air [9]. The blend air is then:
filtered proportionally by the downflow filter [5] and supp­lied as ultrapure air [6] evenly into the sample chamber of the device,
filtered by the exhaust air filter [2] and released as ultra­pure air [1] to the exterior of the device.
Inlet air protection: The air duct between the sample chamber and the device plenum has an inlet air protection [8] below the working surface to prevent coarse particles from entering the plenum where they may impair blower [3] und [4] and filter functions.
Filter system with prefilters,
Fig. 4a
downflow filter and exhaust air
filter / Version KS
18
Page 19
4. Unit description
Version KSP: Fig. 4b: The filter system consists of a set of HEPA prefilters
[10] (the number of filters depends on the sample chamber width) and two HEPA filters [2] and [5] for downflow air and exhaust air.
HEPA prefilters:
The compact size of the filter inserts allows filter replacement with minimal contamination hazard. The use of the prefilters protects the downflow and exhaust air filters considerably. The filter inserts used are HEPA filters of Class H14 (accor­ding to DIN EN 1822). Alternatively, activated carbon filters can be inserted into the housing to protect from smells. Room air [11] is drawn into the sample chamber through the working opening. In the air duct above the prefilter, room air and the downflow within the chamber [7] are then blended to make up the blend air [8]. The blend air is then:
aspirated into the prefilter [10], filtered and fed into the air duct [9].
Operating Instructions Safety Cabinet HERAsafe KS / KSP
HEPA filters: From the air duct, the prefiltered air is:
filtered proportionally by the downflow filter [5] and supp­lied as ultrapure air [6] evenly into the sample chamber of the device,
filtered by the exhaust air filter [2] and released as ultrapure air [1] to the exterior of the device.
Fig. 4b
Filter system with prefilters,
downflow filter and exhaust air
filter / Version KSP
19
Page 20
Operating Instructions Safety Cabinet HERAsafe KS / KSP
4. Unit description
4.4 Controls and display
Fig. 5: The safety cabinet is equipped with two separate control
elements that operate independently of each other:
remote control [3],
pilot switch [1].
The status indicators of the display [2] indicate control opera­tions initiated with the control elements.
Remote control: All device functions can be activated and deactivated easily using the remote control.
Pilot switch: Use to control all basic functions required for the operation. This switch allows you to operate the safety cabi­net if the remote control is not operational or available.
Fig. 6: The display [1] can show text or numeric values and has 12 LEDs [2] and [4] to indicate the current operational state of the unit. The display module also houses the sensor system for the re­mote control. The pulses transmitted are best received by the sensor if the distance A between the remote control [5] and the sensor [3] does not exceed 1m and if the radio signal beam does not deviate more than 15° horizontally. The transmission range of the remote control also depends on the battery state of charge.
Fig. 5
Controls and display
20
Display with remote control
Fig. 6
sensor
Page 21
4. Unit description
4.5 Sample chamber access
Front cover:
The sample chamber of the device is accessible via two modes: Fig. 7a: Manual opening of the front cover [2] allows access to the complete sample chamber width with an opening height C. It is generally needed for decontamination and introduction of larger accessories.
NOTE – Front cover lockout!
The front cover is equipped with a safety lock­out and can only be opened in standby mode, i.e. when the front window is completely closed.
Operating Instructions Safety Cabinet HERAsafe KS / KSP
Optional equipment:
As an option, the front cover of both versions can be equip­ped with a safety feature: Two safety latches at the left and right frame struts secure the front cover from unauthorized opening. With this equipment, the function of the pilot switch is overridden by a safety bezel so that only the remote control can be used to access the sample chamber. If the remote control is faulty, the safety bezel can be unscrewed and the device can be operated using the pilot switch (see Sec­tion 6.4).
Front window: Fig. 7b: The electrically operated front window [1] is made of
multi-layer safety glass and integral to the front cover frame. It can be raised to a maximum opening height B. To access the sample chamber during the work process, the front win­dow must stay in the work position with opening height A.
CAUTION – Front window movement!
Do not attempt to move the front window manually as otherwise the motor drive may be damaged.
Access through front cover
Fig. 7a
Lowering the front window when the device is deenergized:
The safety feature (optional) is backed up by a battery. Should a power failure occur, the pilot switch (see Section 6.4) can be used to lower the front window completely.
Access through front window
Fig. 7b
21
Page 22
Operating Instructions Safety Cabinet HERAsafe KS / KSP
4. Unit description
4.6 Device interfaces
Fig. 8: The standard equipment includes the outlets [10] for internal power supply
as well as the openings [8] on both sides for routing of cables and hoses. All other supply connections are available as options and may also be retrofitted.
Power supply connection: The connection to the power supply system is achie­ved through a cable with grounding plug [2] at the rear of the electrical box.
Contact connection: The front of the box has an RS 232 interface [3] for the connection to a PC as well as two fuse holders for 5A miniature fuses: [4] for (L), [5] for (N). Internal power supply: There are two electrical outlets (5A) [10] and one UV disinfection adapter (optional) [9] located in the side walls.
Media valves: There are two sealed feed throughs [8] on each side. These may be used for installation of media valves [7].
Media supply lines: Additionally, media can be supplied into the sample chamber through three pipes. The inlets [1] are located on top of the housing, the out­lets [6] are placed at the sample chamber backpanel.
22
Supply interfaces
Fig. 8
Page 23
4. Unit description
Disinfection adapter (optional): The disinfection adapter [9] is used to con-
nect a mobile UV radiation device. The adapter is connected to the device control, the UV disinfection routine with a mobile UV device may be controlled with the remote control.
Media connections (optional): The media supply unit consists of three pipes that are routed into the sample chamber through the top of the unit. The inlet connections [1] and outlet connections [6] with thread (R 3/8"), e.g. for media valves [7], are preinstalled and equipped with a sealing plug. The media connections are universal-type connections. Two equipotential bonding connections are installed at the top of the unit and at the stand.
Operating Instructions Safety Cabinet HERAsafe KS / KSP
External systems: A failure detection systems or gas supply solenoid valves may be connected to the safety cabinet control. The unit may also be connec­ted to an external ventilation system. These external systems can then be en­abled or disabled in the secured operating mode using the remote control.
Caution – Combustible gas!
If a gas burner is to be operated in the samp­le chamber, an appropriate shut-off device for the gas supply system (shut-off valve, solenoid valve) must be installed. Use only laboratory safety burners in the sam­ple chamber.
23
Page 24
Operating Instructions Safety Cabinet HERAsafe KS / KSP
4. Unit description
4.7 UV lamp unit
Fig. 9: The UV lamp unit consists of two lamp housings [2] with
two UV lamps each [1] that are integral to the side walls. Both lamp housings are protected by a stainless steel cover [3]. By cross-radiation of the UV units, all surfaces will be disinfected as the shadow zone is reduced. The operating time of the UV lamps is preset. The UV disin­fection routine can be activated quickly by using the remote control.
NOTE – Protection from UV radiation!
As a protection from harmful UV radiation, the UV lamps can only be activated if the front window is completely closed.
4.8 Working area
The standard equipment consists of the segmented workplates. One-piece or special workplates are available as options. The workplates or workplate segments are placed onto the frame above the sample chamber floorpan using two submer­ging wire straps as handles.
Fig. 10: The working area A for perfect material protection ex­tends over the entire width B and depth C of the workplate. The two arm rests [3] are positioned at a distance D (20 cm) to each other centrically on the workplate [1] or on the workplate segments. The armrests are installed to the second perforation line [2] of the workplate.
Fig. 9
UV lamp unit
24
Working area on the workplate,
Fig. 10
armrests
Page 25
5. Start-up
5.1 Initial operation
Prior to initial operation, the safety cabinet must be subjected to an installation test. Correct assembly and installation per­formed by the operator are essential for good start-up.
Version KSP:
Upon the initial connection to the power supply system, the device control of version KSP starts an automatic calibration routine to determine the parameters for the safety system of the device.
Operating Instructions Safety Cabinet HERAsafe KS / KSP
NOTE – Calibration routine
The calibration routine is no substitite for the installation check performed by the service personnel.
5.2 Installing unit and accessories
Version KS:
Device without rack:
Place the device without rack onto a sufficiently stable substructure so that the weight of the device frame does not rest upon the floorpan.
Remove the protective foil from the floorpan.
Device with rack: Assemble the rack (accessory) and install the device frame onto the rack:
1. Fig. 11a: Install the two side members [3] to the rear support [4] using the screws [5].
2. Place the safety cabinet on the stand with the guide pins [1] positioned properly in their receptacles in the stand.
3. Install the housing to the vertical posts of the stand with the retaining screws [8].
4. Align the safety cabinet working area. Place a bubble le­vel onto the workplate and use a wrench (size 24 mm) to rotate the rack stands [6] so that the workplate is perfect­ly horizontal in all directions. When adjusting the rack stand height, proceed from left to right and from rear to front.
Rack installation / Version KS
Fig. 11a
25
Page 26
Operating Instructions Safety Cabinet HERAsafe KS / KSP
5. Start-up
Version KSP: Fig. 11b: The rack [3] and the prefilter housing [2] form an as-
sembly. To facilitate the installation of the device frame [1], the rack is equipped with a lifting mechanism [5]. The four columns [7] of the rack have threaded rods [6] onto which the device frame can be placed and lowered.
NOTE – Installation
The cabinet must be installed only by autho­rized service personnel.
NOTE – Leveling the device
To level the device, do not use the lifting me­chanism but only the adjustable stands of the device.
Align the working area of the safety cabinet:
Place a bubble level onto the workplate and rotate the stands [4] of the rack using a 24-mm wrench until the workplate is exactly horizontal in all directions. For the vertical alignment of the device stands, proceed from left to right and from rear to front.
Installation / Version KSP
Fig. 11b
26
Page 27
5. Start-up
Drain valve (optional) Fig. 11c: The drain valve [2] is installed into the floorpan ope-
ning [1] at the floor at the left front area of the sample cham­ber.
Operating Instructions Safety Cabinet HERAsafe KS / KSP
NOTE – Installation
The optional drain valve must be installed only by authorized service personnel.
WARNING – Drain valve lock!
To prevent contaminated liquid from being accidentally drained from the floorpan, the drain valve must be secured using a conven­tional padlock.
Drain valve installation
Fig. 11c
27
Page 28
Operating Instructions Safety Cabinet HERAsafe KS / KSP
5. Start-up
5.3 Levelling the cabinet
The cabinet should be levelled only after it has been positioned.
1. Remove transport protection (foil) from the workplate or from the workplate segments.
2. Lift the workplate or the workplate segments by the wire hooks and place it/them onto the front and rear rails in the sample chamber with the wide line of holes facing forward.
3. Place a bubble level onto the workplate and use the four levellers to effect a level state in all planes.
5.4 Activating the remote control
The remote control operates on two batteries with the following specifications:
1.5 V alkaline cell (AAA,Type LR 03)
Installing the batteries:
1. Fig. 12: Open the lid [4] of the battery housing at the bot­tom of the remote control by inserting a pointed instrument into the notch [1] and prying the lid off.
2. Insert the batteries [3]. The positive and negative poles are marked at the bottom of the battery housing.
3. Check the position of the coding switch [2]. If the switch is not set to position 1, rotate it to that position.
4. Insert the two hinges of the lid into the joints at the batte­ry housing and slightly press onto the lid so that the retai­ning clip engages.
Functional check:
After the safety cabinet has been connected to the power supply system, the function of the remote control can be checked by switching on the light within the sample chamber. Point the re­mote control toward the display at the sample chamber back­panel:
Press key
Contamination protection:
While the sample chamber is used, protect the remote con­trol against dirt and contamination by using the disposable transparent cover.
Inserting the batteries
Fig. 12
28
Page 29
5. Start-up
5.5 Power supply connection
WARNING – High voltage!
Contact with current-carrying components may cause a lethal electric shock. Before connecting the device to the power sup­ply system, check plug and power supply cable for possible damage. Do not use damaged components to connect the device to the power supply system!
Establishing the power supply connection:
1. Before connecting the device to the power supply system, check to see if the voltage of the outlet corresponds with the specifications on the nameplate of the device. If the ratings given for voltage (V) and maximum current (A) are not correct, the device must not be connected to the power supply system.
2. Connect the grounding plug of the device to a properly grounded and fused outlet.
The outlet must be fused separately using a fusible link T 16 A or using
a circuit breaker B 16.
3. Make sure that the power supply line is not subjected to tensile or compressive force.
Operating Instructions Safety Cabinet HERAsafe KS / KSP
Installation of the power supply connection:
The mains power supply outlets should be out of reach to prevent accidental shutting off. Ideally, the outlets should be installed above the safety cabinet.
Connecting the equipotential bonding:
If the sample chamber is supplied with media (gas, water, etc.), the on-site equipo­tential bonding must be connected to one of the premounted threaded bushings either at the top of the housing or at the stand.
Initialization routine:
After the unit has been connected to the power supply system, the device control runs through a start-up initialization routine and switches the functions to the OFF mode. The safety cabinet is now operational and can be operated using the re­mote control or the pilot switch
Presetting the alarm limits upon initial operation, KSP:
NOTE – Presetting the alarm linits!
The alarm limits must be set only upon the initial operation of the device.
29
Page 30
Operating Instructions Safety Cabinet HERAsafe KS / KSP
5. Start-up
After the initialization routine has been completed, the calibration routine is run:
The display shows "CAL".
The routine starts automatically when the device control is in the work mode (sees Section 6.2) and runs for approx 30 minutes. If the airflow is disturbed during the run time, the routine is cancelled and restarted automatically.
At the end of the routine, the determined parameters for the alarm limits are saved.
NOTE – Initial operation!
According to applicable national standards and regulations, the calibration routine is no substitute for a start-up performed by an au­thorized service technician.
Setting the clock:
After the initialization routine has been run, the clock should be set to the ap­propriate time zone (see Section 6, Operation).
NOTE – Power supply connection!
The safety cabinet should remain connected to the power supply system at all times to en­sure that settings for the individual unit con­figuration remain active in the memory. If the power supply is interrupted for more than 5 mi­nutes, the time must be reset correctly. After the power supply connection has been reestablished, the system switches to the ope­rating mode that had been active last.
30
Page 31
5. Start-up
5.6 RS 232 interface connection
The RS 232 interface has been designed for a cable connection with 9-pin connectors and a contact assignment of 1:1.
Connection of the device:
1. Turn PC off.
2. Fig. 13: Connect the connector of the serial interface ca­ble (not comprised in the scope of delivery) to the socket [1] at the supply interface at the front of the electri­cal box.
3. Connect the serial interface cable to an unassigned slot COM 1/COM 2 etc. at the PC.
4. Turn PC on.
Transfer protocol:
The interface must be configured as follows: Baud: 9600 Data bits: 8 Parity: none Stop bit: 1 Protocol: none FIFO–puffer(extended modulation): enabled
Operating Instructions Safety Cabinet HERAsafe KS / KSP
Fig. 13
RS 232 interface connection
Occupancy of conductors:
Type of connector [X] : 9-Pin SUB-D Pin 2: TxD Pin 3: RxD Pin 5: GND
5.7 UV connection
Fig. 14: UV disinfection adapter (optional) for an external UV
disinfection unit. Voltage: 230 V Current: max. 1,1 A Connectors: [1], [2], [3] and PE-sign
Fig. 14
UV connection
31
Page 32
Operating Instructions Safety Cabinet HERAsafe KS / KSP
5. Start-up
5.8 Installation test
Do not operate the device before the installation test has been completed.
The installation test must be performed in accordance with the specificati­ons of EN 12469 / 2000. The cabinet may be operated as a Class II micro­biological safety cabinet, in accordance with EN 12469 / 2000, if the device functions or function patterns listed below were checked and if the test re­sults are within the safety value tolerances specified in Annex F:
Electrical safety test
Inflow velocity test
Downflow velocity test
Leakage test of HEPA filters
Airflow control test
A repeat test must also be performed after repairs to the device or after consi­derable changes (more than 5 cm) to the location of the device.
The operator must prepare a test report or request a written test report from the authorized test service.
NOTE – Safety warranty!
The operational safety of the device, particu­larly the personal and material protection, are guaranteed only if all safety functions of the device have been tested and approved. Kendro Laboratory Products will not warrant the operational safety if the device is operated without performance of the required installa­tion test or if the installation test and repeat test are not performed by adequately trained and authorized personnel!
NOTE – Device hygiene!
The initial start-up with subsequent installation test does not include any decontamination measures. For operation in the work process, the sample chamber of the device and the accessories required for the work process must be disinfected and cleaned in accordance with the hygiene guidelines set forth for the application.
32
Page 33
6. Handling and control
6.1 Display
Fig. 15: The display at the sample chamber backpanel shows
status messages,
parameter input and putput.
6.1.1 Functions of the display components
[1] Display segment for numbers and text
Value displays
The LEDs 2-5 illuminate when the corresponding value is called up: [2] Display time (yellow LED), usually the active standard display [3] Display downflow velocity (yellow LED) [4] Operating hours after last filter replacement (yellow LED) [5] Display Kendro Performance Factor (yellow LED)
Function displays
The LEDs 6-9 illuminate only when the pertaining value is called up: [6] Ventilation reduced (yellow LED) [7] Potential-free contact activated (yellow LED) [8] Internal power supply activated (yellow LED) [9] UV disinfection routine activated (yellow LED)
Operating Instructions Safety Cabinet HERAsafe KS / KSP
Status displays
The LEDs 10/13 and 11/12 show the operating condition of the device as eit­her/or conditions: [10] Front window is not in working position (red LED) [13] Front window is in working position (green LED) [12] Airflow is steady (green LED) [11] Airflow is not steady (red LED)
Fig. 15
Functions of the display
components
33
Page 34
Operating Instructions Safety Cabinet HERAsafe KS / KSP
6. Handling and control
6.1.2 Display during the calibration routine
The routine runs for approx 30 minutes. During this time, the display shows alternatingly cal and the descending time value. If the calibration routine can­not be started due to faults, the display shows cal permanently.
6.1.3 Display in OFF mode
In the OFF mode, the display shows the current time. For the initial start-up of the device, the clock must be set to the correct time zone and to the corresponding time output (CET mode or AM/PM mode) (see Section 6.3.9).
6.1.4 Display in work mode
In the work mode, the display shows the values of the device data that had been shown last (see Section 6.3.9.ff):
Time (hours and minutes)
Downflow velocity
Operating hours after last filter replacement
Kendro Performance Factor
6.1.5 Power interruption
If the master PCB of the device detects a power failure or a failure of the po­wer supply, a warning is issued.
Note – Power failure warning
In case of a power failure, the display goes off after 10 seconds, then an audible alarm signal sounds for about 30 seconds. After this warning, the device is no longer operative.
The safety feature (optional) allows the complete lowering of the front window after a power failure by pressing the pilot switch.
6.1.6 Display and functions after a power failure
After the power supply has been reestablished after a power failure, the display shows the values and functions that were last shown prior to the failure. The functions that had been selected last will be continued where they had been interrupted.
6.1.7 Failure messages
Failure messages are shown on the display as text/number combinations with the codes ER 1 to ER 6. If one of these codes appears on the display, contact Technical Service immediately.
34
Error code Fault cause
ER 1 Pressure sensor 1 / supply ER 2 Pressure sensor 2 / exhaust ER 4 BUS error ER 5 RAM error ER 6 Fernbedienung defekt
Page 35
6. Handling and control
6.2 Description of the operating modes
The following operating modes exist for the device:
Calibration mode (after initial start-up, version KSP)
OFF mode
Work mode
Standby mode
UV mode
OFF mode: The device is at "idle". Utilized for charging the sample chamber or for cleaning and disinfection, the front cover can be opened when the front window is completely closed.
The air system blowers are switched off.
The sample chamber illumination is available.
The internal power supply within the sample chamber is available:
If the internal power supply is activated, the yellow status indicator IN-
TERNAL POWER SUPPLY ACTIVATED
The display shows the time. If the timer has been activated, the display al­ternately shows the current time and the preset switching time for the timer.
The contact for the external device connection is not available.
Operating Instructions Safety Cabinet HERAsafe KS / KSP
is illuminated.
Work mode: Ensures personal and material protection. In this operating mode,
the work process is run within the sample chamber. The device is in the work mode when the front window has been moved into the working position and the airflow is steady.
The front window is in the working position:
The green status indicator
FRONT WINDOW IS IN WORKING POSITION is
illuminated.
No audible alarm signal.
The air system blowers are switched on to ensure steady airflow:
The green status indicator
AIRFLOW STEADY is illuminated.
The sample chamber illumination is available.
The power supply for the sample chamber outlets is available:
If the internal power supply is ON, the yellow status indicator
POWER SUPPLY ACTIVATED
is illuminated.
INTERNAL
The monitor-contact for the external device connection is available:
If the contact is activated, the yellow status indicator
CONTACT ACTIVATED
is illuminated.
POTENTIAL-FREE
Standby mode: For an interruption of the work process, the front window can
be lowered and the sample chamber sealed aerosol-tight. The air system out­put has been reduced to match the lower air requirement.
The front window is closed:
The air system operates at reduced output.
The red status indicator
FRONT WINDOW IS NOT IN WORKING POSITON is
illuminated.
The yellow status indicator
AIRFLOW REDUCED is illuminated.
The sample chamber illumination is available.
The internal power supply in the sample chamber is available:
If the internal power supply is ON, the yellow status indicator
POWER SUPPLY ACTIVATED
is illuminated.
INTERNAL
35
Page 36
Operating Instructions Safety Cabinet HERAsafe KS / KSP
6. Handling and control
UV mode: For running the UV disinfection routine, the front window is completely
lowered to protect against UV radiation. The routine cannot be run until the front window is in the "closed" position.
The UV disinfection routine is activated:
The yellow status indicator
minated until the preset time for the routine has elapsed. Then, the UV lamps are switched off automatically, and the status indicator is swit­ched off.
The sample chamber illumination is available.
The internal power supply in the sample chamber is not available.
The integral UV lamp (optional) is available.
The power supply for the UV disinfection adapter (optional) is available.
UV DISINFECTION ROUTINE ACTIVATED is illu-
36
Page 37
6. Handling and control
6.3 Remote control
The device control software monitors the selected operating mode of the safety cabinet and automatically deactivates ope­rating functions that are not in accordance with the safety re­quirements of the mode. The remote control has priority over the pilot switch.
6.3.1 Basic functions
Fig.16: The simple operation of the remote control allows the
user to control all basic functions using a minimum of opera­ting steps to switch the safety cabinet into work mode.
Switching the device into work mode:
Keep the key depressed until the ready signal
sounds
Operating Instructions Safety Cabinet HERAsafe KS / KSP
Moving the front window up:
Press the key for approx. 1 second
When the front window reaches the working positi­on, the movement stops automatically. When the movement starts above the working position, the front window stops at the maximal opening position.
Stopping the upward movement:
Press, then release the key
This function can be selected with any key except
the key.
Lowering the front window:
Keep the key depressed
Stopping the downward movement:
Release the key
Basic functions of the remote
Fig. 16
control
Silencing the audible alarm signal, (see section 6.3.3):
Press the key
Switching the device to OFF mode:
Keep the key depressed until the ready
signal sounds
37
Page 38
Operating Instructions Safety Cabinet HERAsafe KS / KSP
6. Handling and control
Availability of functions in the different operating modes: x = available
Key
/$
OFF mode Work mode
XXXX
XXX
XXXX
XXXX
XX
Mode
Standby
mode
XX
XX
XX X
X
6.3.2 Moving the front window to the working position
UV mode
1. Raising or lowering the dront window:
Press, then release the key for approx. 1 second The red status indicator on the display is illuminated
The audible alarm signal is on.
2. When the front window reaches the working position, the movement is au­tomatically stopped.
The green status indicator on the display is illuminated
The audible alarm signal is off if the airflow is steady.
3. If the movement starts above the working position, the front window must first be lowered below the working position and then be raised again. To lower the front window:
Keep the key depressed
4. To stop the downward movement:
Release the key
38
Page 39
6. Handling and control
6.3.3 Silencing the audible alarm signal
When the front window is moved out of the working position or when the pres­sure sensors detect a safety-relevant change of the airflow velocities, the cor­responding visual and audible alarm signals are issued. To silence the audible alarm:
Press, then release the key
NOTE – Quitting the optical alarm signal!
The optical alarm signals can notbe quitted. They change her status indication only, if the necessary functions of the device are operational.
NOTE – Silencing the audible alarm signal!
Operating Instructions Safety Cabinet HERAsafe KS / KSP
The audible alarm signal only can be silenced, if the front window is either completely closed or has been moved to the maximum upper ope­ning position. In the working position the audible alarm signal can not be silenced.
Changing the factory setting: You can change the factory setting so that the audible alarm signals can be silenced (switched off) in any operating mode at any position of the front window. The status is at the display as follows:
P7 0 signal can be silenced
P7 1 signal cannot be silenced
Switch the device to OFF mode:
Keep the key depressed until the ready signal sounds
Deactivate the factory settings:
Keep the key depressed for 5 seconds
Silence audible alarm signals:
Press the key
39
Page 40
Operating Instructions Safety Cabinet HERAsafe KS / KSP
6. Handling and control
Reestablish the factory settings:
Switch the device to OFF mode:
Keep the key depressed until the ready signal sounds
Activate the factory settings:
Keep the key depressed for 5 seconds
6.3.4Switching the illumination on and off
In each operating mode, the sample chamber illumination can be switched on or off.
1. To switch the illumination on or off:
Press, then release the key
6.3.5 Activating and deactivating the internal power supply
All outlets in the sample chamber can be activated (power supply on) or deac­tivated (power supply off) simultaneously.
1. To activate the power supply:
Press, then release the key The yellow status indicator is illuminated.
2. To deactivate the power supply:
Press, then release the key The yellow status indicator goes off.
40
Page 41
6. Handling and control
6.3.6 Displaying the UV disinfection time
This value refers to the operating hours of the set run time of the UV disinfec­tion or of the power supply for the UV disinfection adapter (optional). This dis­play function is only available when the front window is not closed. The device must be switched to work mode.
1. To display the value:
• Press, then release the key
The time value is output in segments of 30 minutes.
2. To deactivate the value display:
The value display is deactivated automatically after 2-3 seconds.
6.3.7 Activating and deactivating the potential-free contact (optional)
An external solenoid valve or alarm system that is connected to the safety ca­binet control system can generally be enabled only if the device is operated in the secure work mode. Enable or disable an external system:
Operating Instructions Safety Cabinet HERAsafe KS / KSP
1. To activate the contact:
Press, then release the key
The yellow status display L is illuminated.
2. To deactivate the contact:
Press, then release the key
The yellow status display L goes off.
NOTE – LED for potential-free contact
If this button is pressed on the remote control, the LED also illuminates if no external alarm sy­stem is connected to the device.
6.3.8 Switching the cabinet to OFF mode
The unit can be switched to OFF mode from any other operating mode:
• Keep the key depressed until the ready signal sounds
41
Page 42
Operating Instructions Safety Cabinet HERAsafe KS / KSP
6. Handling and control
6.3.9 Setting the time
The current time of the time zone in which the unit operates must be set at the start-up of the safety cabinet. Two different display modes can be selected:
CET mode (24:00 hours)
AM/PM mode (12:00 hours)
When the time is to be set, the device must be in the OFF mode.
1. Set the time:
Keep the . key depressed until the two-digit hour disply flashes.
The minute display shows either A, P or no value (CET time display). The time zone is set at the same time as the hour value: First, set the time zone (sequence: CET, A, P), then set the exact hour value.
2. While increasing or decreasing the hour value in increments, set the time zone:
Press, then release the or the key
3. Scroll through values:
Keep the or the key depressed
If the keys are depressed for approx. 2 or 3 seconds, a higher scroll speed is selected.
4. Store the hour and time zone setting:
Press the . key
The function switches to minute display (flashing).
5. Set the minutes:
Press, then release the or the key
6. Scroll through values:
Keep the or the key depressed
If the keys are depressed for approx. 2 or 3 seconds, a higher scroll speed is selected.
42
7. Store the minute value:
Press, then release the . key
The display shows the time.
Page 43
6. Handling and control
NOTE – Calling up device data!
The following data can be called up in succession:
• Time (hours/minutes)
• Downflow velocity
• Operating hours after last filter replacement
• Kendro Performance Faktor
To call up values in succession:
• Press, then release the . key for each
value
The following three sections contain detailed information about displaying values.
Operating Instructions Safety Cabinet HERAsafe KS / KSP
6.3.10 Displaying the downflow velocity
The sensor system of the device continuously monitors the downflow velocity of the airstream in the sample chamber. The currently determined value (m/s) can only be called up in the work mode.
To display the velocity value:
Press the . key repeatedly until the yellow E LED illuminates.
6.3.11 Displaying the operating hours of the HEPA filters
The filter total operating hours after the last filter replacement can be display­ed. Upon each filter replacement, the hours are reset to zero.
1. To display the total operating hours:
Press the key repeatedly, until the yellow status indicator
on the display is illuminated.
2. The value indicates the full hours.
43
Page 44
Operating Instructions Safety Cabinet HERAsafe KS / KSP
6. Handling and control
6.3.12 Displaying the Kendro Performance Factor
The Kendro Performance Factor (PER) is a value that indicates the safety state of the safety cabinet. This value is calculated from data determined by the cabinet safety system and from values captured empirically by service personnel during safety checks. This data is entered into a parameter list of the control software and interconnected. The result can be indicated by the display.
1. To call up the PER:
Press the . key repeatedly until the yellow status indicator on the display is illuminated.
The PER is displayed as an integer.
2. To evaluate the PER:
Number in the 100 to 60 range: The safety cabinet is operationally safe. Personal and material protection is ensured.
Number in the 59 to 30 range: The safety cabinet is still operatio­nally safe. Personal and material protection is ensured.The safety system should be checked.
Number in the 29 to 0 range: The safety of the device may be im­paired. The failure causes must only be repaired by authorized ser­vice personnel. Contact Technical Service.
NOTE – Calling up value!
The value should only be called up, when the airflows have stabilized itself after a lead time of approx. 20 min.
44
Page 45
6. Handling and control
6.3.13 Setting and activating the timer
The timer function allows you to switch the safety cabinet from the OFF mode to the work mode at a predetermined time. The timer can only be set when the device is in the OFF mode.
1. To set the switching time:
Keep the key depressed until the ready signal sounds
The display flashes the two-digit hour display.
2. Increase or decrease the hour value in increments:
Press, then release the or the key
3. Scroll through the value display:
Operating Instructions Safety Cabinet HERAsafe KS / KSP
Keep the or the depressed
If the keys are depressed for approx. 2 or 3 seconds, the higher scroll speed is selected.
4. Store the hour setting:
Press, then release the . key
The function switches to minute display (flashing).
5. Set the minute value:
Press, then release the or the key
6. Scroll through the value display:
Keep the or the depressed
If the keys are depressed for approx. 2 or 3 seconds, the higher scroll speed is selected.
7. Store the minute value and activate the timer:
Press, then release the . key The status display flashes. After a moment, the display shows
the current time again. When the value is stored, the timer is auto­matically activated.
If the setting is not stored, the switching time will be reset to the original value after approx. 15 seconds.
45
Page 46
Operating Instructions Safety Cabinet HERAsafe KS / KSP
6. Handling and control
NOTE – Activating the timer!
If the timer is to be activated without a prior change of the time values, the procedure abo­ve is performed and the existing time values are confirmed:
Press, then release the . key for each
value.
After the minute value has been confirmed, the timer has been activated.
NOTE – Timer function!
The timer can not be used as a start routine that calls up its function automatically. It must be reactivated separately for each de­ferred device start.
6.3.14 Deactivating the timer
If the device was started at a preset time, the timer function is also deactivated automatically. The deferred start can therefore be cancelled only while the de­vice is still in the OFF mode.
To deactivate the timer function for starting the device:
Keep the key depressed until the ready signal sounds
The safety cabinet changes to the work mode.
The status indicator illuminates continuously, and the display
shows the current time.
46
Page 47
6. Handling and control
6.3.15 Setting the UV disinfection time
Depending on the equipment option of the cabinet, this setting is used to:
determine the disinfection time of the optional UV lamps in the side walls or
set the time for the power supply of a mobile UV device at the UV adapter. Factory setting is one hour. The time can be set within a range between 0 and 24 hours in increments of 30 minutes each. The unit must be in the work mode (the front window must not be closed). For each following start of the UV dis­infection, the routine is run with this preset time value.
1. Select the function:
Keep the key depressed until the ready signal sounds
The display flashes the run time that had been selected last.
2. Set or change the disinfection time. To increase the value in incre­ments:
Operating Instructions Safety Cabinet HERAsafe KS / KSP
Press, then release the or the key
3. Scroll through the value display in increments of 30 minutes:
Keep the or the key depressed
4. Store the setting:
Press, then release the . key
If the setting is not stored, the disinfection time will be reset to the original va­lue after approx. 15 seconds.
6.3.16 Starting the UV disinfection
The UV disinfection can only be started if the front window is completely lowered (standby mode). Depending on the equipment of the unit version, this function is used to:
switch on the optional UV lamps in the side walls or
activate the power supply for the outlets of the mobile UV device.
Start the routine:
Keep the key depressed until the ready signal sounds
The display alternately shows the Text fection time in hours and minutes.
The yellow status indicator is illuminated.
After the disinfection time has elapsed, the current time is displayed.
dis and the remaining disin-
47
Page 48
Operating Instructions Safety Cabinet HERAsafe KS / KSP
6. Handling and control
6.3.17 Cancelling the UV disinfection
While the UV disinfection routine is run, it can be interrupted at any time.
1. Cancelling the routine:
Press, then release the key The status indicator goes off.
2. The display shows the current time.
6.3.18 Activating the stop watch
The stop watch function starts a countdown for a preset time (max. 99 min and 59 s) and issues an audible signal when the set time has elapsed. The signal cannot sound if some other device function has already caused an alarm. The function can only be avtivated if the device is in the work mode.
1. Select the stop watch function:
Keep the key depressed until the ready signal sounds
2. Set the minutes (0 - 99) in increments:
Press, then release the or the key
3. Scroll through the value display:
Keep the or the depressed
If the keys are depressed for approx. 2 or 3 seconds, the higher scroll speed is selected.
4. Store the minute setting:
Press, then release the . key
The function switches to second display (flashing).
5. Set the second value (0 - 59):
Press, then release the or the key
48
6. Scroll through the value display:
Keep the or the depressed
If the keys are depressed for approx. 2 or 3 seconds, the higher scroll speed is selected.
7. Store the setting and start the stop watch:
Press, then release the . key
Page 49
6. Handling and control
8. Indication of the function:
The predefined time value counts to zero.
6.4 Pilot switch
Fig. 17: If the remote control is not available, the basic func-
tions required for the operation of the device can be controlled with the pilot switch:
switch the device on,
Raising and lowering the front window,
silencing the audible alarm signal,
switching the device to OFF mode.
6.4.1 Moving the front window:
The movement of the front window is controlled by depressing the pilot switch control element with the corresponding arrow symbol.
1. To raise the window, press the control element [1]. When the front window reaches the working position, the move­ment stops automatically. If the movement starts above the working position, the front window stops at the maximal opening position.
2. To stop the upward movement, release the control element.
3. To lower the window, keep the control element [2] depressed.
4. To stop the downward movement, release the control element.
Operating Instructions Safety Cabinet HERAsafe KS / KSP
Fig. 17
Basic functions of the
pilot switch
NOTE – Switching functions (on/off)
If the device is started with the pilot switch, the chamber illumination is switched on. If the device is switched off with the pilot switch, the chamber illumination is also swit­ched off.
5. If the front window is not in the working position,
the red status indicator
KING POSITION
is illuminated.
FRONT WINDOW IS NOT IN WOR-
the audible alarm signal can be silenced (i.e. swit-
ched off) if the front window is either completely closed or has been moved to the maximum upper opening position.
6. When the front window reaches its working position, the movement stops automatically:
The green status indicator
KING POSITION
is illuminated.
FRONT WINDOW IS IN WOR-
The audible alarm signal is switched off.
49
Page 50
Operating Instructions Safety Cabinet HERAsafe KS / KSP
6. Handling and control
6.4.2 Silencing the audible alarm signal:
1. Move the front window to the upper and lower end positions.
2. Release the control element.
3. Press, then release the control element. The audible alarm signal remains off.
6.4.3 Switching the device to OFF mode
1. Move the front window to the upper and lower end positions.
2. Release the control element.
3. Keep the previously actuated control element depressed until the ready si­gnal sounds.
NOTE – Switch-off function
If the device is switched to the OFF mode with the pilot switch, the chamber illumination is switched off. The device-internal power sup­ply remains in the last functional state that it had been switched to.
50
Page 51
7. Operation
7.1 Hygiene preparations for the sample chamber
The sample chamber surfaces and the accessories required for the work pro­cess must be disinfected and cleaned in accordance with the hygiene guideli­nes set forth for the application.
7.2 Preparing the sample chamber
Installing the accessories:
1. Lower the front window completely.
2. Open the front cover or move the front window into the maximum opening position.
3. Position the accessories in the working area of the workplate.
4. Close the front cover.
5. Move the front window to its working position and wait until the airflow has stabilized.
Operating Instructions Safety Cabinet HERAsafe KS / KSP
CAUTION – Operational safety!
The personal and material protection is ensu­red only if the airflow system of the device is working properly. If the alarm system issues failure messages when the front window is in the working posi­tion, stop all applications that may release harmful aerosols!
NOTE – Application of cystostatic agents
For the application of cystostatic agents, the lowered single-part working plate must be used.Application of cystostatic agents
6. Then place samples into the chamber.
7. To interrupt the work process or for extensive experiment cycles without manual interference, switch the device to standby mode. When the working opening is completely closed, the safety cabinet is sealed aerosol-tight.
51
Page 52
Operating Instructions Safety Cabinet HERAsafe KS / KSP
7. Operation
7.3 Response to failure messages
Failure messages are displayed in form of text/number combinations with a code between ER 1 and ER 6 (see chapter 6.1.6). If one of these codes is display­ed, contact the Technical Service immediately. To isolate the cause of the failure, the operating personnel must perform only the following tests and measures:
Check to see if the exhaust air opening on top of the cabinet is blocked.
Ensure that the on-site exhaust air system is activated.
Close doors and windows in the laboratory to prevent drafts.
Switch off devices in the vicinity of the safety cabinet that cause air turbu­lence or emit excessive heat.
Open flames in the sample chamber may impair airflow conditions.
52
Page 53
7. Operation
7.4 Work rules
The observance of work rules ensures a minimum of operational safety when handling the safety cabinet.
Before starting an operation:
Take off jewelry.
Put on required personal protective gear, e.g. hand, face, or body protection.
Clean and disinfect sample chamber surfaces at regular intervals.
During operationan:
Place samples only with in the defined work area of the workplate.
Do not place unnecessery items into the sample chamber.
Use only disinfected and cleaned accessories for the work process.
Do not cause air turbulence, by quick hand, arm or body movement in the sample chamber or in front of the work opening.
Do not place accessories into the sample chamber that cause air turbulence or emit excessive heat.
Do not block air circulation at the ventilation slots of the workplate.
Operating Instructions Safety Cabinet HERAsafe KS / KSP
Sitting posture during work: Fig. 18: To prevent risks to health, a height-adjustable
working chair with an adjustable seat back should be used during extended work periods at the safety ca­binet. A When the forearm rests on the armrest, is should
be in an almost horizontal position.
B When the thigh is in a horizontal position, the angle
between thigh and lower leg should exceed 90°.
To ensure a compensation between floor and sitting height, a footrest (DIN 4556) should be used. The mi­nimal effective surface of the footrest should be 45 x 35 cm. C The slope should be adjustable within a range of
5° to 15°.
D The adjustable height should extend to a minimum
of 11 cm above the floor.
After finishing an operation:
Remove samples from the sample chamber and store them properly.
Clean and disinfect the sample chamber surfaces, including the workplate and the floorpan. Clean and disinfect all accessories.
Fig. 18
Sitting posture
53
Page 54
Operating Instructions Safety Cabinet HERAsafe KS / KSP
8. Shut-down
8.1 Interrupting an operation
To interrupt a work process, the device is switched to the OFF mode.
1. Remove all samples from the safety cabinet and store them properly.
2. Remove accessories from the sample chamber and clean and disinfect them.
3. Clean and disinfect the sample chamber surfaces, the workplate, and the floorpan.
8.2 Shutting the unit down
If the unit is not to be used or stored for an extended period of time, it must be completely decontaminated.
CAUTION – Decontamination measures!
To shut the device down, the sample chamber must be disinfected completely and the ple­num, including the filters, must be sterilized using formaldehyde.
1. After the device has been decontaminated, close the front window completely.
2. Disconnect the device from the power supply system.
NOTE – Removing the batteries!
To prevent the batteries from leaking and to prevent their contacts from oxidizing during ex­tensive operation intermissions, remove the batteries from the remote control and store them in a dry place. The batteries for the safety feature "Lower front window" are leakproof and do not need to be removed. The nickel-metalhybrid battery for the safety feature "Lower front window" is leakproof and can remain installed.
54
Page 55
9. Cleaning and decontamination
9.1 Decontamination procedure
Several procedures can be applied for decontaminating the safety cabinet. Which procedure is selected, depends on:
the potential risk imminent in the agents,
the degree of purity required by an experiment or by a work process.
Possible decontamination procedures: Wipe/spray disinfection: is the standard disinfection procedure for cabinets
used for microbiological experiments.
UV disinfection: is particularly suited as an intensifying additional disinfection after a wipe/spray disinfection.
Sterilization with steam: can be used for treating the removable stainless steel components. Examples of autoclavable components are the bezels of the UV lamps, the workplate or workplate segments and the armrests.
Operating Instructions Safety Cabinet HERAsafe KS / KSP
Disinfection with formaldehyde: can be performed if a sterile scmple cham­ber is required for the work process. This sterilization procedure is mandatory:
when filters are replaced,
when the device is shut down,
when the device is discarded.
9.2 Wipe/spray disinfection
The wipe/spray disinfection is performed in three stages:
Predisinfection,
cleaning,
final disinfection
Recommended disinfectants:
NOTE – Compatibility!
Chloride-containing disinfectants may damage some surfaces; use only chloride-free disinfec­tants or a disinfectant with a low enough chlo­ride content to have been proved harmless for stainless steel finishes! Disinfectants with an alcohol content of more than 70 % may cause embrittlement of plastic components after extended esposure. Use only disinfectants with a low alcohol content. When using a disinfectant with an alcohol content of more than 70 %, the release limit of 200 g within 2 hours must not be exceeded. Also suited are disinfectants based on quater­nary ammonium compounds.
55
Page 56
Operating Instructions Safety Cabinet HERAsafe KS / KSP
9. Cleaning and decontamination
Predisinfection:
1. Remove all samples from the sample chamber and store them properly.
2. Remove accessories from the safety cabinet and disinfect them using the disinfection procedure recommended by the manufacturer.
3. The workplate and stainless steel components can be removed from the sam­ple chamber and disinfected separately.
4. For predisinfection, spray disinfectant on all sample chamber surfaces or wipe the surfaces using disinfectant.
5. Do not remove the optional UV lamps from the sockets; wipe them thorou­ghly using a damp cloth.
6. Switch the device to work mode, move front window to the working position.
7. Allow disinfectant to react as recommended by manufacturer, then opera­te the safety cabinet for at least 15 to 20 minutes in the work mode so that released aerosols can be absorbed by the filters.
Cleaning:
1. Wipe the surfaces clean using a clean cloth and plenty of clear water.
2. Remove dirt residues and deposits thoroughly.
3. Remove the cleaning liquid from the floorpan and wipe all sample chamber surfaces dry.
Final disinfection:
1. Again, spray disinfectant on all sample chamber surfaces or wipe the sur­faces clean with disinfectant.
2. Allow disinfectant to react as recommended by manufacturer.
56
Page 57
9. Cleaning and decontamination
9.3 UV disinfection after a wipe/spray disinfection
A UV disinfection can be performed either by using the optional integral UV lamps or by using a mobile UV device. The run time of the routine can be preset with the remote control (see Section 6.3.16).
9.3.1 UV disinfection using the integral UV lamps
1. Remove the stainless steel covers from the lamp mountings in the side walls.
2. Lower the front window completely.
3. Start the UV disinfection using the remote control.
9.3.2 UV disinfection using a mobile UV device
To control the disinfection routine of a mobile UV device with the safety cabi­net software, ensure that the connecting plug of the UV device is compatible with the UV disinfection adapter of the safety cabinet.
1. Position the mobile UV device centered on the work area and connect to the UV disinfection adapter.
2. Lower the front window completely.
3. Start the UV disinfection using the remote control.
Operating Instructions Safety Cabinet HERAsafe KS / KSP
9.4 Disinfection with formaldehyde
Procedure:
For gas disinfection, formaldehyde is evaporated in the tightly sealed sample chamber. The quantity of the formaldehyde used depends on the sample chamber volume of the cabinet version to be disinfected (see Technical Data). Per cubic meter of sample chamber volume, at least 5 g formaldehyde must be evaporated with 20 ml water (corresponds with 25 ml of a 20 % formaldehyde solution). The formaldehyde evaporates immediately when its boiling point is reached. The required reaction time is at least 6 hours. After the required reaction time, the formaldehyde should be neutralized by eva­porating a 25 % ammonium solution (10 ml per cubic meter of sample cham­ber volume).
Ambient conditions and accessories:
The temperature at the cabinet location should be approx 21° C, the relative humidity should be between 60 and 85 %. To evaporate the solution, a heating device with a container is required.
NOTE – Disinfection procedure!
A disinfection with formaldehyde must be performed in accordance with the specificati­ons of NSF 49/1992, Annex G. As this procedure has considerable risks, it must only be performed by specially trained and authorized service personnel!
57
Page 58
Operating Instructions Safety Cabinet HERAsafe KS / KSP
9. Cleaning and decontamination
9.5 Cleaning the exterior surfaces
Wipe the exterior surfaces of the device clean using a solution of tepid water and commercial dishwasher solution. Then, wipe exterior surfaces dry using a soft, clean cloth.
9.6 Cleaning the floorpan
Clean the floorpan using a solution of tepid water and commercial dishwasher.
1. Remove the working plates from the sample chamber.
2. Remove dirt residues and deposits thoroughly.
3. Wipe the floorpan clean using a clean cloth and plenty of clear water.
4. Remove the cleaning liquid from the floorpan and wipe the floorpan surfaces thoroughly clean.
NOTE – Material residues
After cleaning, make sure that all cleaning materials have been removed completely from the floorpan.
5. Reinstall the working plates.
58
Page 59
10. Maintenance
10.1 Inspection
PER is the value that indicates the safety state of the safety cabinet by moni­toring various device parameters.
If PER shows a value below 60, the device should be inspected.
Independent of the PER value, the safety cabinet must be inspected on an annual basis.
The annual inspection comprises the following checks:
Electrical safety in accordance with national regulations.
Functional test of the device.
Checking all components for possible damage.
Checking the filter state.
NOTE – Perforated plate!
Operating Instructions Safety Cabinet HERAsafe KS / KSP
The Perforated plate at the blanket of the samp­le chamber serves for the protection of the downflow filter and prevents refluxing. While scanning the filter surface for leak test the perforated plate must be built-in.
Checking the airflow conditions.
Repeat test in accordance with NSF 49/1992, Annex F.
10.2 Service
HEPA filters:
As the filter replacement is an interference with the safety system of the device, filters must only be replaced by Kendro Laboratory Products or by adequately trained and authorized service personnel.
NOTE – Repeat test!
Before filters are replaced the disinfection with formaldehyde is mandatory. After the filter replacement, a repeat test must be performed in accordance with EN 12469 /
2000.
10.2.1 UV lamps
The UV lamps should be replaced after 1500 operating hours:
1. Remove cover from lamp housing.
2. Rotate the lamp in the socket so that the lamp contacts can be removed from the groove in the socket.
3. Insert the new lamp into the socket and rotate until the contacts engage.
4. Replace cover.
59
Page 60
Operating Instructions Safety Cabinet HERAsafe KS / KSP
10. Maintenance
10.2.2 Sample chamber illumination
The tubes are installed in the front cover bezel.
1. Remove the bezel retaining screws and the bezel.
2. The luminescent tubes are retained by rotatable sockets. Rotate the tube carefully to the removal position and remove it from the socket.
3. Insert the replacement tube and rotate it to the working position.
4. Reinstall the bezel and secure it with the screws.
10.2.3 Replacing the front cover seal
Fig. 19: When the front cover seal shows visible sign of wear, the seal must be
replaced. The sealing is slipped onto the housing frame and secured by an additional re­taining strip at the lower edge of the sample chamber opening.
60
Front cover seal replacement
Fig. 19
1. Lower the front window completely and open the front cover.
2. Remove the retaining screws [3] and remove the retaining strip [2].
3. Remove the seal [1] from the housing frame retaining lip.
4. Slip the groove of the new seal onto the housing frame retaining lip and push it on so that the seal fits evenly along the entire sealing surface.
5. Secure the retaining strip with the screws.
Page 61
10. Maintenance
10.3 Retrofitting and repairs
External communication systems, e.g. failure report systems or components for supplying media such as gas solenoid valves can be retrofitted and integrated into the device control.
NOTE – Retrofitting and repairs!
All retrofitting and repair work are interferences with the safety system of the unit. Particular­ly modifications to the filter system and resul­ting changes of the airflow may impair personal and material protection. Such work must be carried out only by authorized service personnel.
Operating Instructions Safety Cabinet HERAsafe KS / KSP
61
Page 62
Operating Instructions Safety Cabinet HERAsafe KS / KSP
g
11. Disposal
11.1 Disposal procedure
Discarded cabinets or unit components contain reusable materials. All compon­ents with the exception of the HEPA filters can be disposed of after having been thoroughly cleaned and disinfected. The HEPA filters must be disposed of in ac­cordance with the applicable national and state regulations for special solid waste.
CAUTION – Contamination hazard!
As the device can be used for processing and treating infectious substances, it may be contaminated. Prior to disposal, the complete device with filters must be decontaminated by performing a formaldehyde sterilization!
Recyclable materials!
Component Material
Thermal insulation components
Printed circuit boards
Plastic components, general Note material labeling Exterior housing Steel, painted Device backpanel Stainless steel/steel, painted Front cover seal EMPP Front window, side windows Multi-layer safety glass Operatin foil Workplates Stainless steel UV bezels Stainless steel Armrests Stainless steel Batteries Alkaline cells Batteris (safety feature: Lower Front W indow)
panel and indicator
Polystyrene foam, EPS/PPS compound
Enclosed electrical components coated with various plastics, mounted on epoxy resin-bound boards.
Polyethylene
Nickel/metal hydrid
62
Page 63
12. Technical data
~200
780/860
~250
HERAsa fe
Width mm 1000 1300 1600 1900 Depth mm Height mm
Width mm 900 1200 1500 1800 Depth mm Height mm
Front window
Working position mm Max. opening mm
Operating Instructions Safety Cabinet HERAsafe KS / KSP
Dimensions
KS/KSP 9 KS/ KSP 12 KS/KSP 15 KS/KSP 18
Exterior dimensions KS/KSP
1580
Interior dimensions KS/KSP
630 780
Front opening KS
780
Front window
Working position mm Max. opening mm
KFS 1, seat pos. (KS only) KSF 2, heightable mm
KFS 1 mm KFS 2 mm
KFS 1 (KS only) mm KFS 2 mm
Diameter mm Distance lower edge mm Distance from backpanel Bushing 1 mm Bushing 2 mm
mm
Front opening KSP
780
Working area height KS/KSP
750
750 - 950
Stand height KS
680
680 - 880
Unit height with stand KS/KSP
2260
max. 2460
Side wall feed throughs KS/KSP
23
130
160 250
63
Page 64
Operating Instructions Safety Cabinet HERAsafe KS / KSP
g
Current
g
s
12. Technical data
Optional media connections in chamber backpanel
HERAsa fe
Diameter Zoll R 3/8 R 3/8 R 3/8 R 3/8 Outer distance left Connection 1 mm Connection 2 mm Connection 3 mm Outer distance right Connection 4 mm Connection 5 mm Connection 6 mm
HERAsa fe Volta
Rated voltage V Blower voltage V
e
KS/KSP 9 KS/KSP 12 KS/KSP 15 KS/KSP 18
370 470 570
570 470 370
Electrical data
KS/KSP 9 KS/KSP 12 KS/KSP 15 KS/KSP 18
1/N/PE AC, 230 V - 50/60 Hz
48 V / DC
Power consumption A 7,3 7,3 8,7 8,7 Leakage current IEC1010, EN 61010 Motherboard fusing A Outlet fusing A On-site fusing A
Power
Power input W 1379 1379 2000 2000
Protection
Protection class Protection type
Overvoltage category (IEC 1010, EN 61010)
Contamination degree (IEC 1010, EN 61010)
Connectin
Connecting lines
HERAsa fe
Pressure media valves bar
line
mA
Circuit breaker B16 / Fuse T 16 A
Pre ssur e
KS/KSP 9 KS/KSP 12 KS/KSP 15 KS/KSP 18
< 3,5
2 x T 16 A
2 x T 5 A
I
IP 20
II
2
Cable (3,0 m)
max. 6
64
Page 65
12. Technical data
HERAsafe KS 9 KS 12 KS 15 KS 18
Unit volume m
3
1,11,41,72,1
Floorpan l 30 40 50 60
Cabinet kg 170 200 230 280 Stand KSF1 kg 26 26 27 27 Stand KSF2 kg 30 30 31 31
kg
kg 50 75 75 75
KSP 9 KSP 12 KSP 15 KSP 18
Unit volume m
3
1,41,72,12,7
Floorpan l 4 5 6 7,5
Cabinet kg 200 240 280 330 Stand KSF1 kg 26 26 27 27 Stand KSF2 kg 30 30 31 31
kg
kg 50 75 75 75
Volume, weights, and loads
Volume, weights, and loads
HERAsa fe
Max. load on overall working area
s
s
Max. load per working area module
ghts
s
Max. load on overall working area
Max. load per working area module
Volume
Wei
Load
Operating Instructions Safety Cabinet HERAsafe KS / KSP
25
Volume
Weight
Load
25
65
Page 66
Operating Instructions Safety Cabinet HERAsafe KS / KSP
12. Technical data
HERAsa fe
Airflow velocities
Inflow m/s Downflow m/s
Air volume
Overall volume flow m³/h 1050 1430 1790 2060 Downflow volume flow m³/h 680 930 1170 1320 Exhaust air volume flow m³/h 370 500 620 740
HERAsafe KSP 9 KSP 12 KSP 15 KSP 18
Airflow velocities
Inflow m/s Downflow m/s
Air volume
Overall volume flow m³/h 900 1230 1540 1870 Downflow volume flow m³/h 600 830 1040 1170 Exhaust air volume flow m³/h 300 400 500 600
Ai rf l ow sy stem
KS 9 KS 12 KS 15 KS 18
Airflow s
0,45 0,36
Ai rf l ow sy stem
Airflow s
0,45 0,32
Filters
HERAsa fe
Type Material Separability in MPPS % Separability at 0,3 µm particle size
Dow nflow
Width mm 915 1220 1525 1830 Depth mm Height mm
Abluft
Width mm 457 915 Depth mm 457 457 Height mm 117 117
HERAsa fe
Type Material Separability in MPPS % Separability at 0,3 µm particle size Width mm Depth mm Height mm
% 99,999
KS/KSP 9 KS/KSP 12 KS/KSP 15 KS/KSP 18
HEPA (H 14 according to DIN EN 1822)
Glas fiber fleece
99,995
99,999%
457
93
610 457 117
Prefilters
KSP 9 KSP 12 KSP 15 KSP 18
HEPA (H 14 according to DIN EN 1822)
Glas fiber fleece
99,995
865 257 575
66
Page 67
12. Technical data
KSP 9 KSP 12 KSP 15 KSP 18
Max. ambient temperature during operation
°C
Min. ambient temperature during operation
°C
Max. humidity / operation % r.H. Max. gumidity / storage % r.H.
Noise level 57dB
Room temperature
20° C / 68° F
Temperature
Heat dissipation to environment
°K < 2 < 2 < 2
Ambient conditions
95
90
Humidi ty
40
10
HERAsa fe
Ergonomics
The noise level was determined in accordance with EN ISO3744. The sound press ure level was measured at a distance of 1 m in front of the working opening. The measurement uncertainty is within a range of ± 2 dB.
Room temperature rise
kJ/s 0,18 0,18 0,36 0,36
< 2
Above room temperature with window closed
Operating Instructions Safety Cabinet HERAsafe KS / KSP
67
Page 68
Operating Instructions Safety Cabinet HERAsafe KS / KSP
13. Device log
Device type: Serial number: Service number: Location: Operator's note:
Work carried out Notes
Part number:
Date Signature
68
Page 69
14. Certificate of decontamination
Operating Instructions Safety Cabinet HERAsafe KS / KSP
69
Page 70
Internet: http://www.kendro.com
Loading...