Heraeus Clinifuge Centrifuge User manual

Heraeus
CLINIFUGE
Operating
Instruction
Manual
Kr
LL
READ
The operating,
A
Read
avoid
If a malfunction
call Never
Always
are
Complete
FOR
-
-
the
Never
tube
ME
Clinifuge
follow
FEW
DO'S
all
parts
damage
Heraeus
attempt
use tube
loaded
the the the
10,000
the exceeds
without
the
SAFETY,
lid
is
open
rotor
is
expiration
cycles
maximum
1.2 g - cm~3
rotor
is
operate
manufacturer.
FIRST
centrifuge
AND
Instruments
to
warranty
not
not
is
designed
these
few
simple
DON'Ts
of
this
manual
to
the
unit
or
occurs, consult
at
1-800-441-2554
open
the
lid
adaptors
adaptors.
when
card
NEVER
date
the
OPERATE
properly
stamped
(see
Section
weight
of a tube
for
loaded
symmetrically
Clinifuge
installed
top
to
cautions
carefully
its
accessories.
the
Troubleshooting
manually
loading 5 ml.
shipped
6.2)
if
THE
(see
on
the
including
speed
you
with
rotor
operation
have
perform a variety
and
guidelines.
before
while
attempting
or
FAX
the
rotor
Guide
information
is
and 7 ml.
your
unit
and
CENTRIFUGE
Section
has
passed
sample
(see
Section
not
checked
IF:
3.5) (see
exceeds
4.1)
the
of
separation
to
operate
at
about
spinning!
tubes
return
to
the
Section
24
maximum
tasks
in
the
unit,
the
end
of
the
problem
(see
Section
manufacturer.
6.3)
or
the
grams
or
the
centrifugal
small-scale
to
insure
this
manual.
to
Heraeus
3.6).
10-,
rotor
has
density
speed
laboratories.
smooth
12-
been
of
operation
If
problems
at
1-908-754-9494.
and
15-ml.
used
the
sample
recommended
Before
and
persist,
tubes
more
than
material
by
the
WARNING!
NOTE:
When
or
your
In
case
immediately.
1.
press
2.
less
preferably,
performing
body
fluids,
local
laboratory
of
imbalance
To
accomplish
the
STOP
centrifugation
all
laboratory
safety
(e.g.
key
pull
the
tube
this:
on
the
front
power
and
personnel
officer.
breakage
of
the
plug.
other
functions
must
follow
or
non-symmetrical
unit
immediately
which
universal
loading),
may
expose
laboratory
shut
off
workers
precautions.
to For
power
to
splashed
details,
the
unit
blood
contact
READ
Table
FAST
Fig. Fig.
Fig. Fig. Fig.
ユーーーー ュー
ユー
mama
of
STARTUP
1.
Clinifuge
2.
Closed
3.
Opened
4.
Locking
5.
Tube
DESCRIPTION
Definition
Standard
Optional Mechanical Electrical
Warranty
Npnabra*
Operating
Fig.
Display
.
Speed
.
Time
Wu
pre
.
Standby
ME FIRST
Contents
Control
Clinifuge
Clinifuge
Nut
and
Adaptors
Parts
Equipment
Construction
Controls
6.
Clinifuge
Features
Mode
Clinifuge
REFERENCE
Panel
Assembled
and
Inserts
OF
UNIT
Construction
Operating
Light
Centrifuge
GUIDE
Rotor
(optional)
Controls
and
(Revision
Control
Panel
Operation
05/93)
Manual
ooooums
NI NI
NN NI
NN D D oo oo
oo oo
SETTING
Location
Electrical
Self
よら
mA
PNRNNNNNNN
Check
Standby
Automatic
Opening Emergency Accidental
Fig.
7.
Emergency
ROTOR
3.2.
3.3.
3.4.
Rotor
Fig.
Fig.
Fig.
Removing Taking
Fig.
Fig.
Parts
8.
Fully
9.
Correct
10.
Disassembled
the
11.
Rotor
Assembling
12.
Assembling
UP
THE
Requirements
Cycle
Mode
Lid
Lock
the
Lid
Lid
Opening
Lid
Opening
Lid
Opening
REMOVAL
Assembled
Rotor
Installation
the
Rotor
Rotor Apart
Disassembly - Top the
Rotor
the
CLINIFUGE
during
AND
INSTALLATION
Rotor
Rotor
View
Rotor
Run
3.5.
3.6.
Installing
Fig.
13.
Seating
Fig.
14.
Rotor
Rotor Fig. Fig.
Tube
15.
16.
Rotor
Tube
the
Rotor
Rotor
Installation
Adaptors
Tube
Adaptors
Adaptors
on
Drive Procedure
Loaded
Shaft
with
in
and
without
Rotor
Inserts
PROPER
Positioning Fig.
Positioning Fig.
OPERATING
Opening
Fig.
5.2.
Stop/Start Fig.
5.3.
Quick Run Fig.
5.4.
Speed Fig. Fig.
5.5.
Time
5.5.1.
Timed Fig.
Fig.
5.5.2.
Untimed
Fig.
17.
Correct
18.
Loading
the
19.
Lid
Key
20.
Start/Stop
21.
Quick Run
Selection
22.
Speed
23.
Speed
Selection
Settings
24.
Time
25.
Time
Runs
26.
"hd"
ROTOR
Tubes
Tube
Tube
Adaptors
Tube
AND
Lid
Key
Key
Selection
Display
Selection
Display
(hold)
Time
BALANCING
for
Symmetrical
Positions
Adaptors
PROGRAMMING
Keys
Keys
Display
into
AND
Balancing
Rotor
Top
LOADING
PROCEDURES
15
15
16 16
17
17
17
17
18
18
18
18
18
19
19 19 19 19 19
MAINTENANCE
6.1.
6.2.
6.3.
6.4.
6.5.
6.6.
7.
7.1.
7.2.
7.8.
7.4.
Cycle
Fig.
Rotor
Rotor
Fig.
Routine
Shipping
TECHNICAL
Counter
27.
Cycle
Exchange
Expiration
28.
Expiration
Decontamination
Electrical
Rotor Performance Environmental
Characteristics
TROUBLE-SHOOTING
Checkup
Counter
Date
Date
Cleaning
Requirements
Procedure Requirements
DATA
Line
Supply
Features
Requirements
Stamp
Stamp
GUIDE
20
20 20
20 21
21
21
21 21
22
22
22 22 22
23
FAST
Setting
1.
2.
3.
4.
5.
6.
STARTUP
Set
unit
After
plugging
(See
Electrical Open Remove
If
adaptors
Fig.
16)
Mount
in
the
Up
on a flat,
the
the
drive
instructions)
WARNING!
7.
Load
tubes
8.
Close
lid
if
both
locks
Programming
9.
To
set
time,
10.
To
set
speed
"+"
or
"-"
11.
For
timed
For
untimed
the
Clinifuge
sturdy,
into
Requirements,
lid
by
pressing
all
packing
are
required,
rotor
on
shaft.
If
the
locking
force
the
sure
the rotor
making
making
are
not
the
First
press
for a timed
keypad.
runs,
runs,
REFERENCE
resonance-free
power
the
materials.
remove
the
drive
Screw
sure both
press
nut
rotor
sure
fully
Run
the
SET
START
press
locking
is
the
engaged.
QUICK
supply,
Section
LID
keypad.
Rotor
the
shaft,
aligning
nut
does
not
further
onto
correctly
rotor
is
lid
locks
keypad
run,
press
keypad.
START
GUIDE
surface
turn
on
2.2.)
must
rotor
in
place.
easily
the
seated
the
move
(See
the
engage
drive
on
in a well
main
freely.
Section
groove
(See
shaft.
the
symmetrically
are
engaged.
in
the
TIME
field.
the
SET
keypad
Unit
will
shut
keypad
and
ventilated
power
switch
3.6, Fig. 15),
on
the
Section
the
3.5,
threaded
Instead,
drive
shaft
balanced.
Display
(See
will
Change
in
the
SPEED
off
automatically.
hold
for
area.
located
and
underside
Figures
portion
remove
cross
pin.
Section
read
"OPEN”
time
by
pressing
field.
desired
time.
under
insert
of
the
13
and
of
the
4.1,
and
Change
the
adaptors
rotor
base
14
for
the
drive
rotor
and
Fig.
17
on
centrifuge
"+"
or
"~"
speed
left
side
of
the
(See
Section
with
the
cross-pin
full
rotor
installation
shaft,
do
not
re-install,
making
Positioning
will
not
operate
keypads.
setting
by
pressing
unit
3.6,
try
to
Tubes.)
Preparing
12.
At
end
13.
Check
14.
To
remove
holes
To
prepare
3to
14.
For
more
this
manual.
the
Next
of
cycle,
for
broken
rotor,
and
lift
rotor
the
next
detailed
Run
remove
all
glass
or
remove
straight
run,
follow
tubes
leakage
locking
up.
information
To
the
routine
about
from
rotor.
into
rotor
nut
by
turning
prevent
damage
cleaning
Clinifuge
and/or
chamber.
counter-clockwise.
to
drive
shaft,
procedure
features,
operation and
Insert
do
not
(Section
guidelines,
index
lift
rotor
6.4),
if
fingers
at
an
angle.
necessary,
consult
i
into
then
other
opposite
repeat
Steps
sections
tube
of
Figure
TO Press
in Change pressing
TO
Press
in
Change
pressing
1.
SET
SET
the
TIME
SET
SET
SPEED
Clinifuge
TIME:
keypad
field.
time
by
"+"
or
"-"
SPEED:
keypad
field.
speed
"+"
or
"-"
Control
keypads.
ing
by
keypad.
speed
set
Panel
time
set
+]
|
LID
KEY
Press
to
open
Lid
will
pop
up
automatically.
QUICK
Use Hold Release
RUN
for
untimed
to
start.
to
stop.
KEY
[ia
[quick]
=
0
sano
runs.
+/- Use
and
KEYS
to
change
time
speed
settings
up
or
down.
START/STOP Press
to
start
To
stop
anytime,
press
START/STOP
Panel
will
display
KEY
timed
"br"
runs.
keypad.
message.
Figure
2.
Closed
(Control
Clinifuge
Panel)
=
==
Figure
3.
Open
(down
Clinifuge
view
w/rotor
mounted)
Figure
6
4.
Locking
O
Nut & Assembled
Rotor
Alla
Figure
5.
Tube
Adaptors & Inserts
(optional)
1.
DESCRIPTION
1.1.
Definition
The
Clinifuge
The
following
maintenance-free
microprocessor
no
carbon
membrane
CSA
-
angle (tube
1.2.
Standard
The
Clinifuge
1.
Two-part
2.
Locking
1.3
Optional
Tube
adaptors
features
dust
touch
listed
(Canadian
rotor
designed
adaptors
you
rotor
Nut
Equipment
OF
UNIT
is a laboratory
provide
induction
controlled
build-up
pads
due
for
easy,
drive
easy
Standards
to
carry
available
as
optional
Parts
have
been
shipped
(Figure
(Figure
and
4)
4)
inserts
(See
centrifuge
efficient
motor
to
elimination
cleaning
Association)
10-,
12-
and
equipment
comes
Figure
5.)
used
for a variety
operation:
of
carbon
15-ml.
for 5 ml.
with
brushes
tubes
the
following
of
separation
and 7 ml.
parts:
tubes)
tasks
in
small-scale
laboratories.
1.4
Mechanical
The
Clinifuge
mechanics is
mounted
speed,
1.5 The
pads
time
Electrical
Clinifuge
on
CAUTION:
1.6
Warranty
The
Clinifuge
to
the
customer.
The
Clinifuge
or
the
expiration
frame
which
to
the
and
Construction operates
the
face
The
damage,
is
rotor
Construction
is
constructed
provide
frame
operating
of
the
extra
to
reduce
conditions.
on
one
unit
to
control
microprocessor
only a factory
warranted
is
date
against
warranted
stamped
of
solid steel
security
circuit
noise
in
the
and
board
program
board
is
located
trained
defects
technician
in
materials
against defects
on
the
rotor,
whichever
with
an
armor
event
of
vibration.
using
microprocessor-controlled
settings.
directly
may
and
in
materials
occurs
chamber
rotor
malfunction.
Digital
displays
behind
service
workmanship
and
workmanship
first.
and a suspension
The
(LED's - light
the
front
panel.
this
area.
for a period
for
the
impact-resistant
emitting
digital
To
of
for
the
avoid
LED
displays
electric
one
year
period
diodes)
of
drive
plastic
and
shock
after
shipment
10,000
and
lid
casing
indicate
key
or
cycles
Loading...
+ 19 hidden pages