Heraeus Clinifuge Centrifuge User manual

Page 1
Heraeus
CLINIFUGE
Operating
Instruction
Manual
Kr
LL
Page 2
Page 3
READ
The operating,
A
Read
avoid
If a malfunction
call Never
Always
are
Complete
FOR
-
-
the
Never
tube
ME
Clinifuge
follow
FEW
DO'S
all
parts
damage
Heraeus
attempt
use tube
loaded
the the the
10,000
the exceeds
without
the
SAFETY,
lid
is
open
rotor
is
expiration
cycles
maximum
1.2 g - cm~3
rotor
is
operate
manufacturer.
FIRST
centrifuge
AND
Instruments
to
warranty
not
not
is
designed
these
few
simple
DON'Ts
of
this
manual
to
the
unit
or
occurs, consult
at
1-800-441-2554
open
the
lid
adaptors
adaptors.
when
card
NEVER
date
the
OPERATE
properly
stamped
(see
Section
weight
of a tube
for
loaded
symmetrically
Clinifuge
installed
top
to
cautions
carefully
its
accessories.
the
Troubleshooting
manually
loading 5 ml.
shipped
6.2)
if
THE
(see
on
the
including
speed
you
with
rotor
operation
have
perform a variety
and
guidelines.
before
while
attempting
or
FAX
the
rotor
Guide
information
is
and 7 ml.
your
unit
and
CENTRIFUGE
Section
has
passed
sample
(see
Section
not
checked
IF:
3.5) (see
exceeds
4.1)
the
of
separation
to
operate
at
about
spinning!
tubes
return
to
the
Section
24
maximum
tasks
in
the
unit,
the
end
of
the
problem
(see
Section
manufacturer.
6.3)
or
the
grams
or
the
centrifugal
small-scale
to
insure
this
manual.
to
Heraeus
3.6).
10-,
rotor
has
density
speed
laboratories.
smooth
12-
been
of
operation
If
problems
at
1-908-754-9494.
and
15-ml.
used
the
sample
recommended
Before
and
persist,
tubes
more
than
material
by
the
WARNING!
NOTE:
When
or
your
In
case
immediately.
1.
press
2.
less
preferably,
performing
body
fluids,
local
laboratory
of
imbalance
To
accomplish
the
STOP
centrifugation
all
laboratory
safety
(e.g.
key
pull
the
tube
this:
on
the
front
power
and
personnel
officer.
breakage
of
the
plug.
other
functions
must
follow
or
non-symmetrical
unit
immediately
which
universal
loading),
may
expose
laboratory
shut
off
workers
precautions.
to For
power
to
splashed
details,
the
unit
blood
contact
Page 4
READ
Table
FAST
Fig. Fig.
Fig. Fig. Fig.
ユーーーー ュー
ユー
mama
of
STARTUP
1.
Clinifuge
2.
Closed
3.
Opened
4.
Locking
5.
Tube
DESCRIPTION
Definition
Standard
Optional Mechanical Electrical
Warranty
Npnabra*
Operating
Fig.
Display
.
Speed
.
Time
Wu
pre
.
Standby
ME FIRST
Contents
Control
Clinifuge
Clinifuge
Nut
and
Adaptors
Parts
Equipment
Construction
Controls
6.
Clinifuge
Features
Mode
Clinifuge
REFERENCE
Panel
Assembled
and
Inserts
OF
UNIT
Construction
Operating
Light
Centrifuge
GUIDE
Rotor
(optional)
Controls
and
(Revision
Control
Panel
Operation
05/93)
Manual
ooooums
NI NI
NN NI
NN D D oo oo
oo oo
SETTING
Location
Electrical
Self
よら
mA
PNRNNNNNNN
Check
Standby
Automatic
Opening Emergency Accidental
Fig.
7.
Emergency
ROTOR
3.2.
3.3.
3.4.
Rotor
Fig.
Fig.
Fig.
Removing Taking
Fig.
Fig.
Parts
8.
Fully
9.
Correct
10.
Disassembled
the
11.
Rotor
Assembling
12.
Assembling
UP
THE
Requirements
Cycle
Mode
Lid
Lock
the
Lid
Lid
Opening
Lid
Opening
Lid
Opening
REMOVAL
Assembled
Rotor
Installation
the
Rotor
Rotor Apart
Disassembly - Top the
Rotor
the
CLINIFUGE
during
AND
INSTALLATION
Rotor
Rotor
View
Rotor
Run
Page 5
3.5.
3.6.
Installing
Fig.
13.
Seating
Fig.
14.
Rotor
Rotor Fig. Fig.
Tube
15.
16.
Rotor
Tube
the
Rotor
Rotor
Installation
Adaptors
Tube
Adaptors
Adaptors
on
Drive Procedure
Loaded
Shaft
with
in
and
without
Rotor
Inserts
PROPER
Positioning Fig.
Positioning Fig.
OPERATING
Opening
Fig.
5.2.
Stop/Start Fig.
5.3.
Quick Run Fig.
5.4.
Speed Fig. Fig.
5.5.
Time
5.5.1.
Timed Fig.
Fig.
5.5.2.
Untimed
Fig.
17.
Correct
18.
Loading
the
19.
Lid
Key
20.
Start/Stop
21.
Quick Run
Selection
22.
Speed
23.
Speed
Selection
Settings
24.
Time
25.
Time
Runs
26.
"hd"
ROTOR
Tubes
Tube
Tube
Adaptors
Tube
AND
Lid
Key
Key
Selection
Display
Selection
Display
(hold)
Time
BALANCING
for
Symmetrical
Positions
Adaptors
PROGRAMMING
Keys
Keys
Display
into
AND
Balancing
Rotor
Top
LOADING
PROCEDURES
15
15
16 16
17
17
17
17
18
18
18
18
18
19
19 19 19 19 19
MAINTENANCE
6.1.
6.2.
6.3.
6.4.
6.5.
6.6.
7.
7.1.
7.2.
7.8.
7.4.
Cycle
Fig.
Rotor
Rotor
Fig.
Routine
Shipping
TECHNICAL
Counter
27.
Cycle
Exchange
Expiration
28.
Expiration
Decontamination
Electrical
Rotor Performance Environmental
Characteristics
TROUBLE-SHOOTING
Checkup
Counter
Date
Date
Cleaning
Requirements
Procedure Requirements
DATA
Line
Supply
Features
Requirements
Stamp
Stamp
GUIDE
20
20 20
20 21
21
21
21 21
22
22
22 22 22
23
Page 6
FAST
Setting
1.
2.
3.
4.
5.
6.
STARTUP
Set
unit
After
plugging
(See
Electrical Open Remove
If
adaptors
Fig.
16)
Mount
in
the
Up
on a flat,
the
the
drive
instructions)
WARNING!
7.
Load
tubes
8.
Close
lid
if
both
locks
Programming
9.
To
set
time,
10.
To
set
speed
"+"
or
"-"
11.
For
timed
For
untimed
the
Clinifuge
sturdy,
into
Requirements,
lid
by
pressing
all
packing
are
required,
rotor
on
shaft.
If
the
locking
force
the
sure
the rotor
making
making
are
not
the
First
press
for a timed
keypad.
runs,
runs,
REFERENCE
resonance-free
power
the
materials.
remove
the
drive
Screw
sure both
press
nut
rotor
sure
fully
Run
the
SET
START
press
locking
is
the
engaged.
QUICK
supply,
Section
LID
keypad.
Rotor
the
shaft,
aligning
nut
does
not
further
onto
correctly
rotor
is
lid
locks
keypad
run,
press
keypad.
START
GUIDE
surface
turn
on
2.2.)
must
rotor
in
place.
easily
the
seated
the
move
(See
the
engage
drive
on
in a well
main
freely.
Section
groove
(See
shaft.
the
symmetrically
are
engaged.
in
the
TIME
field.
the
SET
keypad
Unit
will
shut
keypad
and
ventilated
power
switch
3.6, Fig. 15),
on
the
Section
the
3.5,
threaded
Instead,
drive
shaft
balanced.
Display
(See
will
Change
in
the
SPEED
off
automatically.
hold
for
area.
located
and
underside
Figures
portion
remove
cross
pin.
Section
read
"OPEN”
time
by
pressing
field.
desired
time.
under
insert
of
the
13
and
of
the
4.1,
and
Change
the
adaptors
rotor
base
14
for
the
drive
rotor
and
Fig.
17
on
centrifuge
"+"
or
"~"
speed
left
side
of
the
(See
Section
with
the
cross-pin
full
rotor
installation
shaft,
do
not
re-install,
making
Positioning
will
not
operate
keypads.
setting
by
pressing
unit
3.6,
try
to
Tubes.)
Preparing
12.
At
end
13.
Check
14.
To
remove
holes
To
prepare
3to
14.
For
more
this
manual.
the
Next
of
cycle,
for
broken
rotor,
and
lift
rotor
the
next
detailed
Run
remove
all
glass
or
remove
straight
run,
follow
tubes
leakage
locking
up.
information
To
the
routine
about
from
rotor.
into
rotor
nut
by
turning
prevent
damage
cleaning
Clinifuge
and/or
chamber.
counter-clockwise.
to
drive
shaft,
procedure
features,
operation and
Insert
do
not
(Section
guidelines,
index
lift
rotor
6.4),
if
fingers
at
an
angle.
necessary,
consult
i
into
then
other
opposite
repeat
Steps
sections
tube
of
Page 7
Figure
TO Press
in Change pressing
TO
Press
in
Change
pressing
1.
SET
SET
the
TIME
SET
SET
SPEED
Clinifuge
TIME:
keypad
field.
time
by
"+"
or
"-"
SPEED:
keypad
field.
speed
"+"
or
"-"
Control
keypads.
ing
by
keypad.
speed
set
Panel
time
set
+]
|
LID
KEY
Press
to
open
Lid
will
pop
up
automatically.
QUICK
Use Hold Release
RUN
for
untimed
to
start.
to
stop.
KEY
[ia
[quick]
=
0
sano
runs.
+/- Use
and
KEYS
to
change
time
speed
settings
up
or
down.
START/STOP Press
to
start
To
stop
anytime,
press
START/STOP
Panel
will
display
KEY
timed
"br"
runs.
keypad.
message.
Page 8
Figure
2.
Closed
(Control
Clinifuge
Panel)
=
==
Figure
3.
Open
(down
Clinifuge
view
w/rotor
mounted)
Figure
6
4.
Locking
O
Nut & Assembled
Rotor
Alla
Figure
5.
Tube
Adaptors & Inserts
(optional)
Page 9
1.
DESCRIPTION
1.1.
Definition
The
Clinifuge
The
following
maintenance-free
microprocessor
no
carbon
membrane
CSA
-
angle (tube
1.2.
Standard
The
Clinifuge
1.
Two-part
2.
Locking
1.3
Optional
Tube
adaptors
features
dust
touch
listed
(Canadian
rotor
designed
adaptors
you
rotor
Nut
Equipment
OF
UNIT
is a laboratory
provide
induction
controlled
build-up
pads
due
for
easy,
drive
easy
Standards
to
carry
available
as
optional
Parts
have
been
shipped
(Figure
(Figure
and
4)
4)
inserts
(See
centrifuge
efficient
motor
to
elimination
cleaning
Association)
10-,
12-
and
equipment
comes
Figure
5.)
used
for a variety
operation:
of
carbon
15-ml.
for 5 ml.
with
brushes
tubes
the
following
of
separation
and 7 ml.
parts:
tubes)
tasks
in
small-scale
laboratories.
1.4
Mechanical
The
Clinifuge
mechanics is
mounted
speed,
1.5 The
pads
time
Electrical
Clinifuge
on
CAUTION:
1.6
Warranty
The
Clinifuge
to
the
customer.
The
Clinifuge
or
the
expiration
frame
which
to
the
and
Construction operates
the
face
The
damage,
is
rotor
Construction
is
constructed
provide
frame
operating
of
the
extra
to
reduce
conditions.
on
one
unit
to
control
microprocessor
only a factory
warranted
is
date
against
warranted
stamped
of
solid steel
security
circuit
noise
in
the
and
board
program
board
is
located
trained
defects
technician
in
materials
against defects
on
the
rotor,
whichever
with
an
armor
event
of
vibration.
using
microprocessor-controlled
settings.
directly
may
and
in
materials
occurs
chamber
rotor
malfunction.
Digital
displays
behind
service
workmanship
and
workmanship
first.
and a suspension
The
(LED's - light
the
front
panel.
this
area.
for a period
for
the
impact-resistant
emitting
digital
To
of
for
the
avoid
LED
displays
electric
one
year
period
diodes)
of
drive
plastic
and
shock
after
shipment
10,000
and
lid
casing
indicate
key
or
cycles
Page 10
1.7.
Operating
Controls
A.
Speed
B.
Time
C. +
D.
Lid
E.
Quick Run
Figure
1.8.
Display
1.8.1.
Shows
speed
Setting
Setting
and Keys
Opening
6.
Clinifuge
Features
Speed
set
speed
field
window.
and
and
for
Time
Key
Key
Operating
during
Indication
Indication
and
Speed
Controls
acceleration
Field
and
and
B
F. G.
Setting
Control
centrifugation
H.
L
J.
Panel
ο D
Start/Stop Indicator
Main
Window Access
phases.
Key
LED's
Power
holes
Switch
(for
When
E Е
(left
side
manual
the
unit
under
base)
and
emergency
brakes,
"br"
lid
opening)
is
indicated
in
the
1.8.2. Records the
last
1.8.3.
A
red
Time
time
minute
Standby
flashing
in
seconds
counted
Mode
dot
in
the
up
down
Light
time
to
one
in
seconds,
display
minute.
field
indicates
Settings
above
the
unit
is
one
in
minute
standby
are
mode.
counted
(See
Section
down
in
full
2.4,
Standby
minutes,
Mode.)
with
Page 11
2.
SETTING
2.1.
Location
Remove
air
into
Before
during
2.2. Before
1.
2.
A
main
(See
Once
the
2.3.
Each
packing
flow
in
and
the
bottom
operating,
shipment.
Electrical
connecting
the
line
the
line
voltage
power
Figure
the
power
unit
is
ready
Self
Check
time
indicating
humming
be
opened.
UP
materials
around
of
the
To
Requirements
voltage
switch
6.) supply
for
Cycle
the
unit
all
LED's
sound.
THE
CLINIFUGE
and
the
unit.
centrifuge
be
sure
to
open
the
the
unit
to
imprinted
circuit
breaker
located
plug
operation.
is
turned
temporarily.
When
the
place
Proper
also
remove
unit,
you
the
main
on
the
identification
is
a
under
is
connected,
on,
the
Whenever
self-check
the
centrifuge
ventilation
helps
maintain
packing
will
power
15
Amp,
the
left
microprocessor
need
supply,
type
side
the
main
the
is
complete,
materials
on
a
sturdy,
is
required
cool
operation.
plug
inside
it
in.
to
make
plate
is
equal
K,
slow
release
of
the
unit
switch
performs
centrifuge
the
yellow
resonance-free
to
keep
the
the
unit
certain
with
that:
the
caracteristic
may
be
used
is
turned
is
on,
an
automatic
disconnected
LID
motor
used
one
to
and
light
table
cool
during
to
secure
available
commonly
turn
the
unit
the
self-check
self-check
during
will
illuminate,
in
an
area
operation.
and
(120
Volts);
used
on
cycle
run,
the
permitting
allowing
protect
for
instruments.
and
off
is
of
the
electronics,
unit
will
good
A
fan
built
the
rotor
as
desired.
completed,
make
a
the
lid
to
2.4.
Standby
For
convenience,
Mode
microprocessor red
LED
dot
in
function.
2.5.
Automatic
The
lid
is
all-steel
dual
lid
locks.
2.6.
Opening
To
open
Three
conditions
1.
The
unit
2.
To
prevent
stop
before
3.
If a power
failure, the
“br”
completed,
NOTE:
Time or
the
the
lid,
is
not
injury, attempting
failure
the
centrifuge
code
press
and
between
the
Clinifuge
automatically
the
time
display
Lid
Lock
construction
Lid
turn
the
unit
on
will
automatically
plugged
the
lid
to
occurs
into
can
open
during
always
will
be
displayed
key
pad
speed
settings
operating
is
designed
blanks
field.
with
an
and
press
prevent
the
electrical
not
the
operation,
performs
for
desired
are
cycles.
to
operate
out
all
displays
Press
any
integrated
the
LID key
the
lid
power
be
opened
lid.
This
may
the
a
complete
accompanied
function.
retained
in
as
and
key
to
reactivate
strobe
port
pad.
from
being
source
while
the
be
checked
lid
will
remain
braking
by a buzzing
(Refer
to
memory
a
"constant
goes
into
and
tabs
The
lid
will
opened:
and/or
rotor
the
is
spinning.
by
locked
cycle.
Emergency
and
do
not
on.”
a
Standby
the
displays
on
pop
main
looking
until
When
sound.
Lid
need
After
two
minutes
Mode,
and
to
both
sides
to
up
automatically.
switch
through
Opening
Wait
power
power
To
re-start
to
be
is
not
until
the
returns.
is
restored
instructions
re-entered
of
inoperation,
indicated
perform
engage
turned
on.
the rotor
window
in
In
to
after
braking
after
by a flashing
the
centrifugal
the
automatic
comes
case
to a full
the
lid.
of
power
the
centrifuge,
cycle
in
Section
power
failure
the
is
2.7.)
Page 12
2.7.
Emergency
WARNING!
Disconnect
2.8.
Accidental
If
the
lid
opens
that
the
rotor
In
case
of a prolonged
cabinet
1.
2.
3.
lid
Disconnect
Make
sure
Insert
two
ata
time.
Lid
Never Confirm
of
can
attempt
the
lid.
power
Lid
during a run,
has
come
be
opened
the
power
the
rotor
rods
into
Opening
that
cord
before
Opening
to a full
power
cord
is
fully
the
to
open
rotor
has
using
during
the
microprocessor
stop
failure
manually
before
stopped
access
the
lid
stopped
the
Run
by
looking
or
defect,
as
follows:
using
by
looking
holes
on
manually
spinning
following
will
through
the
the
emergency
through
both
while
by
Emergency
flash
the
the
LID
button
the
side
panels
the
rotor
looking
Lid
LID
round
window
can
lid
opening
round
(See
is
still
through
Opening
error
message.
not
procedure.
lid
window.
Figure
spinning.
the
Procedure.
in
the
middle
be
used
7)
and
round
Before
to
open
unlock
window
lifting
of
the
the
the
in
the
lid.
lid.
lid,
the
lid,
The
one
middle
confirm
locked
side
Figure Not
to
7. be
Emergency
used
Access
Lid
while
Hole
Opening
rotor
is still
spinning!
Access
Hole
10
Page 13
3.
ROTOR
The
Clinifuge
WARNING!
3.1.
Rotor
The
rotor
REMOVAL
rotor
is
shipped
Before To by
Each
(dust,
To specially
Parts
is a two-piece
Nut
operating operate, turning
insure
clockwise.
time
the
glass
formed
AND
installed
the
the
rotor
and
rotor
is
splinters,
proper
drive
unit
connected
operation
INSTALLATION
in
the
centrifuge
centrifuge,
drive
For
safety
re-installed,
etc.).
shaft
shaft
and
each
to
the
remove
in
operation,
the
prevent time
drive
must
rotor
it
shaft
and
secured
all
packing
turn
freely
and
check
the
and
chamber
damage,
is
re-installed.
by a locking
the
for
shipping.
material
the
rotor
locking
must
rotor
must
(See
Section
nut.
placed
nut
must
nut
for
be
completely
be
3.1,
inside
the
be
checked
tightness
properly
Fig. 9 and
before
free
unit
for
shipping.
for
tightness.
each
of
contamination
positioned
Section
Rotor
Nut
run
on
the
3.5.)
Figure
8.
Fully
Assembled
É
G
O
RC
NZ
4
Figure
9.
Correct
Rotor
Rotor
Groove
underside
rotor
Threaded
of
drive
Drive
cross
Installation
on
base
shaft
shaft
pin
of
portion
Figure
10.
Disassembled
=
Rotor
<
Rotor
Top
Rotor
Base
(right
side
Cross-pin groove
Base
Rotor
(Bottom
view)
up)
Page 14
3.2.
To
remove
1.
Open
2.
Remove
3.
Remove
4.
To
rotor
Removing
the
the
all
rotor
remove
will
lift
rotor
lid.
tubes
rotor,
out.
the
Rotor
from
and
locking
insert
(Before
the
chamber:
check
for
nut by
turning
index
fingers
inserting
broken
fingers,
glass.
counter-clockwise.
into
holes
check
holes
in
rotor
for
and
pull
broken
straight
glass.)
up.
If
properly
installed,
the
entire
WARNING!
3.3.
Taking
To
separate
1.
Remove
2.
Insert
3.
Using
Never damage
the
the
rotor
index
fingers
thumb,
pull
the
to
the
Rotor
rotor
from
press
Rotor
Centerpiece Press
to
Apart
top
from
unit
and
into
empty
firmly
Hole
here
separate
rotor
motor.
the
place
down
up
at
an
rotor
base:
on a flat
tube
holes
on
centerpiece.
angle.
surface.
on
opposite
Doing
The
so
could
sides
two
of
the
halves
bend
rotor
of
the
the
top.
rotor
drive
should
shaft
separate
and
cause
easily.
permanent
Figure
11--Rotor
Rotor
Base
Disassembly Top
View
Page 15
3.4.
Assembling
1.
Before
2.
Insert
Balancing
3.
To
assemble,
base recessed.
the
each
installation,
tube
adaptors,
and
Loading.)
line
centerpiece.
Rotor
if
up
the
Push
clean
the
required.
tab
in
down
gently
chamber
(See
the
center
until
and
Section
hole
the
unit
rotor
3.6,
of
the
snaps
of
dust,
Rotor
rotor
glass
splinters
Tube
Adaptors
top
with
together.
The
and
and
the
corresponding
center
any
other
Section
of
the
materials.
4.,
notch
base
will
Proper
in
the
be
Rotor
rotor
slightly
Alignment in
centerpiece
rotor
base
Figure
3.5.
A aligned
WARNING!
1.
2.
3. NOTE!
12.
Installing
slot
matching
precisely
Turn
the
Using
the
base
in
When
If
the
sure
performed.)
rotor
the
notch
of
Assembling
the
Rotor
the
cross
when
Position alignment
drive
shaft
tab
in
the
exact
alignment
is
properly
locking
rotor
further
that
the the
the
Rotor
pin
installing
the
rotor
with
the
to
place
rotor
centerpiece
with the
seated,
nut
does
onto
the
rotor
on
the
the
so
that
cross
the
cross
secure
not
drive
is
properly
drive
rotor
the
pin
pin
cross
the
easily
shaft.
shaft
is
grooved
in
the
chamber.
slot
grooved
on
the
drive
in a known
as a guide,
pin
on
the
locking
engage
Instead,
seated.
the
into
shaft.
position,
position
drive
shaft.
nut,
turning
threaded
remove
(To
insure
into
the
the
underside
Failure
to
e.g.
the
grooved
Press
clockwise.
portion
the
locking
correct
underside
of
do so
will
long
bar
pointing
notch
firmly
into
of
the
nut
and
seating,
of
the
rotor
the
rotor
prevent
on
the
place.
drive
shaft,
the
rotor,
be
sure
Alignment
center
of
rotor
base
base
proper
to
front
(See Figure
underside
do
and
re-install,
Step 4 2
tab
opening
top
and
must
seats
in
seating.
of
the
not
try
to
is
properly
in
be
exact
9).
rotor
force
making
RIGHT (threaded completely
centerpiece
Figure
13.
portion
above
Seating
extends
rotor
Rotor
on Drive
Shaft
WRONG
(Threaded below
rotor
centerpiece)
portion
13
Page 16
Figure
3.6.
The Tube
positioned
-
- 7 ml.
- 5
14.
Rotor
Clinifuge-rotor
adaptors
10,
12
and
tubes
ml.
tubes
rotor
(See
(7
ml.
tubes)
Rotor
Tube
Figure
15.
Rotor
Installation
Tube
and
as
follows:
15
ml. must must
Figure
Tab
Adaptor
Rotor
Adaptors
is
designed
inserts
tubes
be
loaded
be
loaded
15)
without
Adaptors
Procedure
to
hold 5 ml., 7 ml.,
(Section
may
be into into
Insert
with
and
1.3,
placed adaptors adaptors
directly
Tab
without
Fig.
5)
into without which
Inserts
10
ml.,
are
available
the rotor
inserts
have
been
Inserts
12
ml.
without
before
fitted
A
and
as
optional
being
with
Rh
Î
15
ml.
tubes.
equipment
adaptors
positioned inserts
Top
(5
ml
tubes)
Rotor
Tube
in
before
|
Insert
Adaptor
the
rotor
being
and
tubes
(See
positioned
with
Insert
must
Figure
into
be
16) the
Figure
14
16.
Tube
Adaptors
Loaded
in
Rotor
Page 17
4.
PROPER
WARNING!
4.1.
Positioning
For
correct
adaptors)
The
maximum
Glass
consideration
CAUTION!
The
In immediately,
To
balance
of
equal
tubes
have
G-forces
Tubes
ROTOR
rotor
must
case
of
imbalance
maintain
Tubes
in
size
and
allowable
different
when
positioning tubes
vary
made
BALANCING
always
(caused
pull
the
plug
symmetrical
for
Symmetrical
loading
difference
from
of
the
weight.
wall
thicknesses
rotor
to
Polystyrol
AND
be
symmetrically
by
tube
or
turn
off
balancing,
Balancing
rotor,
tubes
in
weight
in
rotor.
tend
the
To
to
between
and,
rotor.
prevent
break
LOADING
balanced
breakage
circuit
breaker.
insert
tube
and
adaptors
opposing
therefore,
breakage,
at
very
low
when
loading.
or
non-symmetrical
adaptors
hold
in
all
must
be
tubes
(and
different
observe
g-forces.
empty
placed
adaptors,
volumes.
tube
manufacturer
loading),
opposing
exactly
opposite
if
used)
This
press
the
STOP
slots.
tubes
is
one
gram.
must
be
taken
recommendations.
key
(and
into
Empty
Figure
NOTE:
17.
Correct
When
for
positions
using
proper
Tube
Positions
either
balance.
flared
top,
(See # 3)
12-ml.
or
15-ml.
Filled
tubes,
positions
load
symmmetrically
in
every
other
position
Page 18
4.2.
Positioning
Tube
adaptors
To
position
1.
Before
2.
Turn
the
3.
For 7 ml. the
center
For 5 ml.
on
the
To
remove
WARNING!
for 5 ml.
adaptors:
loading,
rotor
top
tubes,
of
the
tubes,
insert
into
inserts,
When
(See
Section
Tube
Adaptors
tubes
separate
upside
insert
bowl.
push
the
loading
the
each
Secure
the
grooves
push a pencil
adaptors,
4,
and 7 ml.
two
down
and
adaptor
each
flat
end
in
the
Proper
tubes
halves
Rotor
of
place
pair
adaptor
of
the
inserts
adaptor.
or
similar
opposite
Balancing
must
be
the
rotor.
on a flat
pointing
by
pushing
into
Follow
tool
through
sides
of
installed
(See
Section
surface.
the
flat
ends
firmly
the
curved
#3
to
the
the
rotor
and
Loading.)
before
3.3)
down
into
the
openings
position
open
must
the
end
be
loading
with
the
rutted
in
adaptor
in
the
equally
tubes
into
two
tabs
sleeves
the
adaptor,
inside
adaptor.
balanced
the
turned
sliding
the
rotor.
in
rotor.
extending
the
weight.
side
into
tabs
Figure
16
18.
Loading
Tube
Adaptors
into
Inverted
Rotor
Top
Page 19
5.
OPERATING
5.1.
Opening
the
AND
Lid
PROGRAMMING
PROCEDURES
The
process
on.
To
up
automatically.
Once
entire
NEVER TURNING.
cedure.)
NOTE: cannot
Lid
Opening
lid
can
is
completed
open,
press
the
unit
operating
TRY
TO
(See
In
case
be
used
Procedure,
be
opened
the
starts,
cycle.
OPEN
Emergency
of
power
to
open
and
LID
the
only
the
key
LID
key
THE
LID
failure,
the
lid.
Section
when
yellow
pad.
is
disabled
WHILE
Lid
the
Follow
2.7,
the
LID
LED goes
The
lid
will
ROTOR
Opening
LID
key
Emergency
Fig.
7.
braking
pop
for
the
IS
Pro-
pad
Figure
19.
LID
5.2
Start/Stop
Key
When may pad. START/STOP
indicating
When
has LED be
opened
To
stop
pad After
again, continued the
braking
fuge
the
green
be
started
When
the
ended,
will
both
centrifugation,
when
each
indicating
begins
by
the
LED
the
STOP
cycle
is
the
green
illuminate,
or
another
the
red
stop,
at
any
cycle
to
accelerate.
START
pressing
rotor
goes
completed
LED
the that
time. there
LED
starts
off
function
START
indicating
cycle
press is
illuminated.
green
the
If
START
will a delay
is
on,
centrifugation
the
START/STOP
turning,
and
the red
is
available.
and
the
braking
LED
and
that
may
be
started.
the
START/STOP
START
interrupted
LED
cycle
is
pressed
before
the
LED
yellow
the
lights
the
green
goes
cycle
unit
may
during
Clini-
key
on,
LID
may
key
up be
Figure
20.
START/STOP
Key
17
Page 20
5.3.
Quick
Figure
21.
Run
Quick
Run
Key
If
the
green
also
be
in
QUICK
RUN
key
QUICK
gradually
continues the
rotor
If
QUICK
be
re-activated.
The
QUICK
less
than
same The
QUICK
the
speed
started
RUN
for
RUN
to
until
to
brake
RUN
RUN
one
time.)
set
START
the
key
RUN
in
LED
with
the
mode,
required
is
pressed,
its
highest
the
key
immediately.
is
pressed
key
may
minute.
key
programming)
is
on,
QUICK
press
length
speed.
is
released,
again,
be
(Braking
may
also
centrifugation
RUN
key.
To
and
hold
the
of
time.
the
rotor
accelerates
Centrifugation
which
acceleration
used
for
runs
will
still
require
be
used
to
may
operate
QUICK
While
the
causes
will
requiring
the
override
5.4.
Speed
Selection
Speed
per
Speed
during
changing
When
screens
press previously the
IAN
Set
Figure
22.
Speed
Selection
Keys
"="
After in set
complete.
is
minute.
settings
standby
the
for
the
100ths
the
new
keys.
entering
the
SPEED
speed
adjusted
the
centrifuge
TIME
SET set
position
in
in
are
periods,
previously
and
key
speed
in
speed
the
new
field
memory.
gradations
retained
is
in
SPEED
pad will
the
up
speed,
again
in
and
need
set
value.
Standby
are
in
the
appear,
speed
or
down
press
to
replace
The
speed
of
100 revolutions
the
unit's
memory
be
entered
Mode,
the
display
blank.
To
activate,
SPEED
flashing
display
using
field.
the
window.
the
the
SET
the
previously
setting
key
mode
only
The
digit
in
"+"
or
pad
if
is
Figure
18
23.
Speed
Display
Page 21
5.5.
Time
5.5.1.
Timed
Selection
Settings
The
time
setting activated field.
Time
settings
periods,
previously
When
display
key
Press
from will
setting
The
gradations
pads.
and
the
screen
pad
in
the
previous
begin
will
flashing
Runs
is
displayed
by
pressing
are
retained
need
be
set
value.
centrifuge
is
blank.
the
TIME
SET
key
pad
operations, a red
to
flash.
If
set
begin
to
flash.
position
up
may
or
down
be
may
programmed
in
the
the
SET
key
in
memory
entered
is
in
field. A red
previously,
by
standby
To
activate,
again.
be
changed
using
only
If
"hd"
TIME
pad
during
if
mode,
zero
the
(hold)
the
the
from
field
in
the
standby
changing
the
press
the
will
appear.
time
is
message
previous
in
one-minute
"+"
or
"="
1-99
minutes.
and
TIME
not
is
the
time
SET
set
time
key
Figure
25.
Time
5.5.2.
Untimed
Display
Runs
If
continuous
(hold)
function
To
enter several field
until
field. When
An
"hd”
STOP.
untimed
the
centrifuge.
operation
the
times,
"hd"
appears,
run
in
the
"hd”
setting,
then
press
appears
will
continue
time
press
is
required,
display.
press
the
SET
on
the
screen
START.
until
enter
the "-"
key
in in
To
the
operator
the
key the the
end,
"hd"
pad TIME TIME
push
stops
Figure
26.
"hd"
Time
Display
Page 22
6.
MAINTENANCE
6.1.
Cycle
The
Clinifuge
each
time a cycle
regardless
The
at
the
The
remaining
The
cycle
of
number
same
counter
Counter
Centrifuge
is
the
run
of
cycles
time.
rotor
may
Checkup
is
equipped
completed. A cycle
duration.
recorded
cycles
be
will
used
may
be
be
displayed.
to
maintain
with a cycle
is
defined
checked
at
routine
counter
inspection
which
as:
acceleration,
any
time
schedules.
by
closing
the
microprocessor
run
at
speed,
the
lid
automatically
and
deceleration
and
pressing
both
steps
to
SET
up
rest,
keys
Figure
NOTE:
The
6.2.
Rotor
Clinifuge
the
rotor
NOTE:
27.
Cycle
The
microprocessor
replaced
After
each
Exchange
rotor
(See
Fig.
Rotors
cleaned
Counter
(running
10,000
must
28).
with
counts
down
of
cycles,
be
replaced
bleach
the
the
rotor's
contact
after
or
alkaline-based
cycles
working
Baxter
10,000
upward
Scientific
from
life).
Products
cycles
cleaners
zero
or
no
must
to
10,000
for
replacement
later
than
be
at
which
the
Expiration
replaced
point
of
the
after
the rotor
rotor.
Date
every
5,000
must
stamped
cycles.
be
on
20
Page 23
6.3.
Rotor
To
locate
calendar the
expiration
Expiration
the
expiration
are
molded
year.
The
Date
date,
into
point
one
of
Stamp
separate
of
the
the
the
segments
arrow
two
parts
in
indicates
the
of
the
the
rotor
expiration
rotor.
base.
month.
The
The
words
2-digit
(See
Fig.
"Expiry
number
28.)
Date”
in
the
and a circular
center
indicates
Figure
28.
Expiration
6.4.
Routine
1.
Press
the
2.
Unplug
3.
Check
4.
Clean
Remove
Si
6.
Place
7.
8.
WARNING!
6.5. Infectious If
infectious
guidelines.
In which Before Read WARNING!
for
the
holes
and
rotor
(See
Section
Polypropylene Before Check
case
re-installing
that
Decontamination
of
it
must
opening a centrifuge
Me
Date
Cleaning
LID
key
and
the
centrifuge
glass
splinters
exterior
the
pulling
Alkaline
(See
Materials
material
contamination,
First
No
will
infectious.
and
locking
rotor
on a flat
3.3,
Fig.
rotors
the
the
rotor
cleaning
Section
leaks
be
discarded.
Section.
known
transmit
not
nut by
surface and
may
turns
test
Stamp
Procedure
open from
all
straight
(for
the
the
or
spilled
plastic
turning
this
lid.
power
liquid
surfaces
up.
press
11)
be
cleaned
rotor
check
freely
before
reagents
6.2,
Rotor
Exchange.)
Requirements
into
the
unit,
the
rotor
may
suspected
method
infection.
can
example:
source.
following
counter-clockwise,
thumbs
in a dishwasher
that
the
operation.
such
disinfect
be
of
offer
Therefore,
april
laboratory
using a cloth
into
centerpiece
motor
cover
as
bleach
using a non-alkaline
autoclaved
being
contaminated,
complete
human-sourced
all
1997)
safety
moistened
and
at
acceptable
is
correctly
will
affect
for a maximum
assurance
lift
the
refer
with
to
rotor
that
materials
guidelines.
non-corrosive
rotor
out
by
separate
temperature
in
place.
disinfectant,
of
to
products
top
durability.
20
times
the
laboratory
derived
should
liquid.
inserting
and
ranges.
following
at
bottom
121°C.
safety
from
considered
be
index
fingers
rotor
sections.
laboratory
(482°F.),
guidelines
human
safety
sources
potentially
into
after
in
6.6.
Shipping
shipping
When
not
accept
Decontamination
for
return
Requirements
the
liability
authorization
Clinifuge,
for
damage
is
required
and
the
only
use
caused
any
instructions.
time
by
incorrect
the
original
unit
is
packaging
transport
returned
or
packing
for
servicing.
packaging
a
Contact
good
of
Heraeus
quality.
Customer
Heraeus
Service
will
21
Page 24
7.
TECHNICAL
7-1.
Electrical
120
Volts
(10%,
60 Hertz
7.2.
Rotor
Rotor
Material
Maximum
Maximum Maximum Maximum
Maximum
Maximum Rotor
7.3.
Maximum Minimum
Maximum Speed Speed Time
Time
1.
2.
Lid
opening
Quick-run
Load
Tube Speed
Relative
Radius
allowed
Angle
Clinifuge
speed
speed
Relative display selection
display
Selection
"hd"
Mode
Minute
key
Diagnostic
(See
Trouble-Shooting
Drive
Motor.
Data
Storage
Data
Memory
Dimensions
Weight
DATA
Line
Supply - Model
-15%)
Characteristics - Model
Capacity
Length
Centrifugal
imbalance
Performance
Centrifugal
(2
modes)
(hold)
Mode
key
status
messages
Force (RCF)
Force
Guide)
Features
(RCF)
No. C 3538
1.0
Amps
65
Watts
No.
3760
Polypropylene
6 x 15-milliliter
6x
12-milliliter
12 x 10-milliliter
12 x 7-milliliter 12 x 5-milliliter
100
mm x 12
5000
revolutions
2600 x g
9.6
centimeters
1
gram
(.035
ounces)
38
degrees
5000
revolutions
1600
revolutions
2600 x g
5-digit
display
Adjustable
2-digit
Sets
Adjusts
from
For
safety, Accelerates when
the
speed
1.
Lid
2.
Error
3.
Speed
4.
Braking Maintenance-free, Speed With HxWxD
14.5
in
stages
display
for
continuous
time
setting
1-99
minutes
the
rotor
released.
set
open
LID
codes
and
(br)
and
time
NV-RAM
(9-1/4" x 10-1/2" x 14-3/4")
kilograms
tubes
tubes
tubes
tubes
with
tubes
with
tubes,
131
per
minute
per
minute
per
minute
indicates
registers
lid
The
in
(E-XX)
time
settings
(32
revolutions
of
100
seconds
runs
in
one-minute
will
not
to
maximum
QUICK-RUN
programming.
settings
induction
pounds)
optional optional
mm x 6
(rpm)
rpm
for
unlimited
open
while
speed
drive
tube
adaptors
tube
adaptors
tubes
per
minute
or
minutes
time.
gradations
rotor
is
then
key
may
motor
(rpm)
up
or
spinning. brakes also
immediately
be
used
down
to
override
7.4.
Environmental
Allowable
during
during
maximum
*
Do
not
use
operation
storage
where
22
room
highest
high
humidity
Requirements
temperature
or
shipping
sample
causes
condensation
ranges:
temperature
4°C. - 35°C.
-10°C. - 50°C. 15°C.
above
ambient
(39°-
(14° - 1229
95?F)*
temperature
F)
Page 25
8.
TROUBLE-SHOOTING
If
malfunction
If
the
problem
INDICATION
All
displays
remain
Unusually
running
Display
LID
Display
OPEN
Display
a
flashing
dot
in
TIME
Display
E-1
dark
noises
shows
shows
shows
the
field
shows
or
E-12
occurs,
persists,
loud
red
try
to
correct
call
Heraeus
SYMPTOM
1)
Lid
cannot
opened
2)
Drive
rotor
coasts
Often
accompanied
by
heavy
Drive
is
rotor
coasts
Lid
appears
but
no
start
Centrifuge
no
other
are
illuminated.
All
function
are out
GUIDE
the
Instruments
be
at
rest
is
cut
off,
to
vibration
cut
off,
to
rest
closed
is
at
rest;
displays
keys
of
order
problem
rest
using
the
instructions
at
1-800-441-2554
POSSIBLE
1)
Unit plied
2)
Power
during
1)
Locking
2)
Incorrect installation
3)
Rotor
unsymmetrically.
4)
Sudden
balance
tube
Lid
open
Lid
locking
is
not
Centrifuge
Standby
Error
wrong NV-RAM
CAUSE | REMEDY
is
not
sup-
with
power
failure
run
nut
rotor
is
loaded
im-
following
breakage
during
system
activated
in
Mode
on
main
board,
or
defective
loose
run
in
the
or
FAX
details
1)
Check
of
bottom
Emergency Fig.
7)
2)
Check If
in
order,
Press
STOP
when
rotor
1)
Tighten
engage
re-install
2)
Remove
re-install
cross
pin
matching
Replace
ATTENTION:
correctly
(Section
3-4)
Check
must
be
allowable
Rotor
Characteristics
Wait
till
rotor
and
ON.
Heraeus
Check
tightly
remains Technical
The Press operating
Switch error Heraeus
lid
on
Standby
any
power
message
following
main
electrical
has
locking
properly,
rotor. rotor
being
on
the
3.4, rotor
balanced.
Check
Technical
closure.
both
in
the
Support
key
mode.
Technical
Trouble-Shooting
to
1-908-754-9494.
power
switch
plate.
For
lid
unlocking,
Lid
Opening
supply
call
Heraeus
key
immediately.
reached a complete
nut.
from
careful
the
drive
groove
locking
The
placed
Fig.
loading.
imbalance,
stops.
Switch
lid.
Lid sides. display,
Mode
to
reactivate
OFF
remains
cord
Technical
If
nut
remove
drive
shaft
to
shaft
in
the
rotor
nut
and
rotor
top
onto
the
12) Opposite
For
maximum
see
power
If
error
recurs,
Support.
must
If
the
OPEN
call
Heraeus
saves
electrical
and
ON
in
display,
Support.
under
left
see
(Section
connection.
Support.
Open
does
not
rotor
and
and
line
up
the
with
the
base.
tighten.
must
rotor
base.
sides
Section
OFF
call
be
closed
message
energy.
the
normal
again.
If
call
Guide.
side
2.7,
lid
only
stop.
be
7.2,
the
23
Page 26
INDICATION
Display E-3
Display
"br"
shows
shows
SYMPTOM Drive
is
cut
off,
rotor
coasts
Power
centrifuge
not
sound
is
run,
is
turned
will
humming
heard.
to
rest
on,
POSSIBLE
Motor
safety
has
tripped
no
braking
is
flowing
Error
on
main
Motor
defect
Centrifuge
was
lost
during
run.
CAUSE
circuit
because
current
board
power
REMEDY
1)
Lid
open
remedy,
2)
Overheated
Excessively
insufficient
OFF
Check
vent brush.
3)
If
the
later,
current.
Replace
Replace
Allow
braking
see
and
let
all
holes
Then
centrifuge
itis a problem
Call
main
Motor
centrifuge
cycle.
during
LID
motor
high
air
cooling).
the
unit
vent
holes
in
the
switch
will
Heraeus
board
to
Display
braking
display.
(Possible
room
temperature
Switch
stand
and
front
panel
ON
and
not
start
with
Technical
finish
complete
will
return
phase.
(Section
for
clean
the
For
2.8.)
causes:
power
30
minutes.
the
using a paint
start
again.
30
minutes
braking
Support
to
normal.
or
small
24
Page 27
Page 28
esKendro
Laboratory
Products
Kendro
Laboratory Products
Postbox
D-63405
15
63
Hanau / Germany Telephone: (+49)
61
81/
35-300
GmbH
Telefax:
(+49)
61
(+49)
81/ 3 57
55
22/
31
Telephone:
(+49)
55
22/
31
62 62
49, 76,
22,
In
Sales Service
Service
development, make
20056893 Printed
the
interest
changes
in
Germany
of
continuous
we
without
CLINIFUGE
product
reserve
the
express
10/98
right
to
notice.
Loading...