Use this manual to get acquainted with your centrifuge and its accessories.
The manual helps you to avoid inappropriate handling. Make sure to keep it always close to the centrifuge.
A manual that is not kept handy cannot provide
protection against improper handling and thus
against damage to persons and objects.
The manual comprises chapters on
• Safety regulations
• Instrument description
• Rotor program and accessories
• Transportation and hook-up
• Use of the centrifuge
• Maintenance and care
• Troubleshooting
• Technical data
• Index
Overleaf you will find a graphic repre-
sentation of the control panel of the
Biofuge haemo with a survey of the
most important functions
Please fold out
12000
speedtime
10
-
+
set
Biofuge
haemo
set
lid
quick
run
start
stop
INSTRUMENTS
The control panel of the
Biofuge haemo
Display
Speed
Resting state: upon pressing any key: display of
preselected speed, after 2 min switch
to standby mode (flashing red light)
During run: preselected speed
Braking: "br"
End: "0"
Lid open: "OPEN"
Resting/end: error codes (if present)
Run time ("time")
Resting/end: upon pressing any key: display of
preselected run time, after 2 min
switch to standby mode (flashing red
light)
During run: remaining run time or (with quick start)
run time passed
Keys
lid: open lid (only possible if connected
with mains supply)
quick run: short-term acceleration as long as key
is pressed, with indication of run time
passed
start/stop: normal start or manual stop
set speed/time: selection of the parameter to be
changed (speed or run time)
+/- stepwise increase/decrease of preset
values for the selected parameter
speed or run time, accelerated
change when pressed permanently
Error codes
(troubleshooting see chapter "Troubleshooting")
E-1/E-12: keys do not respond (main board defective)
E-3: no braking current: lid popped open or
manually opened during run; excess tem-
perature in the drive
br: power turned off during run or power failure
lid: lid opened during run
OPEN: lid lock mechanism out of order
Contents
For your safety........................................... 3
Heraeus centrifuges are manufactured according to
current technical standards and regulations. Nonetheless, centrifuges may pose dangers if
• they are not used as designed
• they are operated by untrained personnel
• their design is improperly changed
• the safety instructions are not heeded
Therefore anybody concerned with operation and
maintenance of the centrifuge must read and follow
the safety instructions.
In addition, the pertinent regulations for prevention of
accidents must be strictly followed.
This manual is an integral part of the centrifuge assembly and must be kept close at
hand at all times.
Proper use
The centrifuge is designed to separate liquidsuspended materials having different densities and
particle size, respectively. The maximum sample density is 1.2 g/cm
3
at maximum speed.
Improper use
During a run, a safety zone of 30 cm around the centrifuge must be maintained where neither persons nor
hazardous materials may be stationed.
The centrifuge may cause harm to you or other persons and may damage material goods if you do not
respect the following safety measures:
Centrifuging hazardous materials
• The centrifuge is neither made inert, nor is it explo-
sion-proof. Therefore never use the centrifuge in an
explosion-prone environment.
• Explosive or flammable substances must not be
centrifuged. The same holds for substances prone
to react briskly with each other.
3
For your safety
• Do not centrifuge toxic or radioactive substances or
pathogenic microorganisms unless you have taken
proper precautions.
Such precautions can e.g. consist of biological
seals.
• Should toxins or pathogenic substances enter the
centrifuge or its parts, you must carry out the proper
procedures for disinfection (see "Maintenance and
care – Disinfection").
• Strongly corrosive substances that may cause dam-
age to materials and impair the mechanical strength
of the rotor may be centrifuged only inside protective vessels.
Handling
• Never use the centrifuge unless the rotor is properly
mounted.
• Never manually open the lid if the rotor still turns.
• Use only original parts for the centrifuge. The only
exception are common glass or plastic centrifuge
tubes if these are approved for the rotor speed and
RCF values of your rotor, respectively.
• Never use the centrifuge with the lid open.
• Never use the centrifuge if the paneling has been
partially or totally removed.
• Changes in mechanical or electrical components
may be carried out only by persons authorized to
this effect by KENDRO Laboratory Products.
• You may use the centrifuge only with a properly
loaded rotor. You must not overload the rotor.
• If the rotor or the lid shows visible traces of corro-
sion or wear, you must stop using it.
• Strictly follow the rules and regulations for cleaning
and disinfection.
4
For your safety
Conformity to current standards
Heraeus centrifuges are manufactured and tested
according to the following standards and regulations:
for all voltages:
• IEC 1010-1 / EN 61010-1
• IEC 1010-2 / EN 61010-2-020
− Pollution degree 2
− Overvoltage category II
for 120 V only:
• CAN/CSA-C22.2 No. 1010.1-92
• CAN/CSA-C22.2 No. 1010.2.020-94
Safety instructions in this manual
This symbol denotes potential hazards to
persons.
This symbol denotes potential damage to the
centrifuge or parts in its immediate surroundings.
General hints are marked with this symbol.
In addition, you are asked to adhere to the pertinent
regulations, in Germany
• Regulations for prevention of accidents VBG 4
• Regulations for prevention of accidents VBG 5
• Regulations for prevention of accidents VBG 7z
5
For your safety
for your notes
6
The Biofuge haemo
The Biofuge haemo
The figure below shows the Biofuge haemo with the lid
opened. In this state, the standard display at the end of
a run and after opening the lid is "OPEN". After a couple of minutes the display changes to the standby
mode, and all display panels turn dark.
ACHTUNG
Rotor vorschriftsmäßig
auf Welle befestigen
ATTENTION
Fixer le rotor sur
l'arbre conformément
aux préscriptions
ATTENTION
Rotor must be fitted in
accordance with the
instruction manual
ATENCIÓN
Fijar el rotor en el
arbol según las
prescripciones
Safety systems
The Biofuge haemo is equipped with a number of
safety systems.
Rotor chamber
The rotor chamber is integrated into the steel case.
When the lid is closed, the rotor chamber is sealed
against the surroundings by a rubber ring with a special profile that allows cooling air to pass through the
rotor chamber.
Warning if lid is manually opened during a run, or if
lid lock mechanism is out of order
If the lid is manually opened during a run, or if the lid
lock mechanism is out of order, a corresponding message appears in the display ("Lid" and "OPEN",
respectively).
Lid lock mechanism
You can open the lid only when the power is turned on
and the rotor has practically come to a halt (< 80 rpm).
You can start the centrifuge only if the lid is properly
closed.
Emergency lid release
In order to permit you to remove samples even after a
power failure, the centrifuge is equipped with an emergency lid release (see „Operation“).
7
The Biofuge haemo
Properties
The Biofuge haemo is an air-cooled benchtop centrifuge for the determination of the hematocrit.
It can also be used for the preparation of sensitive
samples in the biochemical and medical laboratory.
The rotor accepts common centrifuge tubes up to a
volume of 2 ml.
The preset speed is reached in seconds. You can also
spin samples for only a few seconds using the "quick
run" key if this is required for the task in question. The
long-lived, maintenance-free induction motor provides
quiet and vibration-free operation even at high speeds.
The user-friendly control panel permits easy operation.
At the start the preset speed and run time are displayed. During a run, the remaining run time is displayed: above 1 min in minutes and below 1 min in
seconds.
If you press the "+" or "-" key repeatedly, you increase
the corresponding preset parameter (speed or run
time) stepwise. If you press and hold down the chosen
key, the respective value changes continuously, at first
slowly and after a few seconds at an accelerated pace.
Air cooling of the Biofuge haemo
During a run, the spinning rotor creates frictional heat.
This leads to a temperature increase of the rotor, the
tubes and finally of the samples. The extent of warming
depends on:
• run time
• temperature of the environment
• location of the centrifuge
• rotor speed
The Biofuge haemo is equipped with air cooling to
prevent overheating of the samples. The cooling is
particularly effective in the cold room (lowest permissible temperature +4 °C).
"Quick run" operation
If you press and hold the "quick run" key, the rotor is
accelerated with maximum force up to the maximum
speed.
8
Accessories
1
3
4
5
6
7
8
9
1
0
1
1
1
2
1
3
1
4
1
5
1
6
1
7
1
8
1
9
2
0
2
1
2
2
2
3
2
4
H
E
R
A
E
U
S
I
N
S
T
R
U
M
E
N
T
S
C
a
t
-
N
o
1
0
6
7
I
I
1
2
0
0
0
/
R
P
M
I
I
1
2
3
4
5
6
7
8
9
1
0
1
1
1
2
1
3
1
4
1
5
1
6
1
7
1
8
1
9
2
0
2
1
2
2
2
3
2
4
Accessories
Standard accessories
Hematocrit dish and reading harp
The Biofuge haemo is delivered with a special rotor
(hematocrit dish or rotor) equipped with 24 guide notches for receiving blood capillaries.
A reading harp for determining hematocrit values after
removing the capillaries is delivered with the instrument.
Reading spiral
As an option, you can order a reading spiral which is
placed directly onto the rotor for reading hematocrit
values in situ.
maximum capacity 24 capillaries with 1.2 to 1.8 mm diameter and 75 + mm length
maximum speed n
maximum RCF value at n
10
12000
max
14926
max
Accessories
Additional accessories
Fixed-angle rotor 7500 3325 for microliter tubes
Optionally you can order for the Biofuge haemo a
fixed-angle rotor with 24 holes for placing microliter
tubes with a volume of 1.5 or 2.0 ml.
In addition, three sets of adapters with 24 reduction
sleeves each are available on request. With the adapters you can centrifuge all commercially available microliter tubes with volumes of 0.2 ml to 0.6 ml as well
as 0.2 ml PCR reaction vessels.
Please refer to our sales documents for detailed information on accessories including technical data and
order numbers.
3
1
1
2
4
1
1
1
5
1
6
a
1
d
7
2
4
x
1
4
8
g
1
9
p
p
2
1
0
/
9
6
2
#
1
3
3
2
2
2
2
3
1
S
m
T
a
N
0
x
E
1
M
l
o
U
R
T
9
S
N
I
S
8
U
E
A
7
R
E
H
6
m
p
r
5
0
0
4
0
3
1
4
x
a
m
3
2
4
2
1
11
Accessories
Fixed-angle rotor: technical data
Table 1: Fixed-angle rotor order no. 7500 3325 for Biofuge haemo
no. of places/volume 24 x 1.5 / 2 ml
maximum permissible load 24 x 4 g
temperature range -4 °C to +40 °C
maximum speed n
maximum RCF value at n
minimum speed n
minimum RCF value at n
[min-1] 13000
max
16060
max
[min-1] 2000
min
380
min
maximum radius [cm] 8.5
minimum radius [cm] 5.9
angle of incidence 40°
maximum kinetic energy [s] 1.65 kNm
12
Adapters for rotor order no. 7500 3325
Table 2: Adapters for the fixed-angle rotor of the Biofuge haemo
Adapter Dimensions
(∅ x H)
reduction sleeve PCR 6,2 x 20 mm 0.2 ml 24 gray 7600 3750
reduction sleeve 8 x 43.5 mm 0.5 / 0.6 ml 24 turquoise 7600 3758
reduction sleeve 6 x 46 mm 0.25 / 0.4 ml 24 red 7600 3759
Other accessories
Capacity Number per
Set
Color Order No.
• Reading harp 7600 0938
• Reading spiral 7500 1236
For use with the fixed-angle rotor also available:
• fixed-angle rotor 24 x 1.5 / 2 ml 7500 3325
• special cap nut for fixing the rotor
• 10 mm tubular hexagon wrench for tightening the
cap nut
cap nut
order no.
70056208
tubular socket
wrench
order no.
2036 0072
The printed documents consist of the delivery notes
and this Manual.
Accessories
13
Accessories
for your notes
14
Before use
Before use
Where to install the centrifuge
The centrifuge must be operated in a place meeting the
following criteria:
• A safety zone of 30 cm around the centrifuge must
be maintained. Hazardous materials must not be
kept within this zone during centrifugation.
• The substructure must be stable and resonance-
free. A good support is provided by a plane laboratory bench or a large laboratory carriage with casters that may be locked.
• To ensure sufficient air circulation, a minimum dis-
tance from the wall of 10 cm at the back and of
15 cm on each side must be kept.
• The centrifuge must be protected from heat and
direct sunshine.
• The location should be well ventilated.
Mains connection
Make sure that your mains voltage and frequency
match the specifications on the instrument.
Removing the transport protection
Turn the instrument on. The display panel shows for
about 6 s the routine internal software check sequence.
Open the lid by pressing the "open lid" key and remove
the transport protection for the rotor.
Check that the rotor moves freely by lightly turning it,
and make sure the rotor is tightly screwed on.
15
Before use
for your notes
16
Loading...
+ 44 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.