OPTIONAL – WHEN YOU NEED MORE
OPTIONAL – WHEN YOU NEED MORE
Temperature recorder
for continuous recording of the internal temperature on pressuresensitive paper
● No colour ribbon or
ink needed
● Paper roll with 9.2 m length, sufficient for 14
days of continuous operation
● Recorder can be connected and
disconnected
Connection facility
for an external
temperature recorder
for measuring with
Pt 100 or NiCr-Ni
temperature sensor,
connection via
DIN 41524 plug connection
● Documentation of oven temperature
● Sensor permanently installed
Recording of sample
temperature with
display
for measuring the temperature directly at the
sample using flexible
Pt 100 temperature
sensor (to be selected separately, see page 12)
● Digital display
● Switching to recorder connection possible
● Introduction of the sensor into the oven
through an access port or door seal.
Thermicon®P temperature controller
Specification as for Kelvitron®t plus:
● Temperature program with max. 9 segments
● Integral timer for switching the drying oven on
or off, up to maximum 99h 59 min.
● Sensor breakage indicator
Eurotherm 2404/P4 temperature controller
Microprocessor-controlled program controller
with PID control response
● Accurate process control using four freelystored temperature programs of 16 segments
each
Alternatively selectable RS 232 or
RS 422/485 interface
● External temperature recording or
control via PC
● Connection via Sub-D connector
Digicon
®
S temperature controller with
analog interface
● Accurate digital display of the set points or
actual readings
● PD/PID control response for rapid heating,
minimum overshoot and low control deviation
● External settings 0 … 20 mA
● External temperature recording 0 … 10 V
● Connection via DIN 41524 connector
Weekly program timer (168 h)
for automatic repetition of an on-off cycle over a
period of seven days
● Shortest switching interval 2 hours
● 90 hour power reserve in the event of a
power failure
Digital weekly program timer
with digital display for free programming or for
basic program, e.g. "On” at 06.00 hrs and "Off”
at 22.00 hrs.
● Shortest switching interval 10 min.
● 16 hour power reserve in the event of a
power failure
Daily program timer (24 h)
for automatic repetition of an on-off cycle over
a period of one day
● Shortest switching interval 15 min.
● 90 hour power reserve in the event of a
power failure
Adjustable temperature limit controller
Protective Class 3.1 (upper temperature limit protection), completely
independent of the temperature controller
● Automatically takes over the temperature control if the upper
temperature limit set by the operator is exceeded
● No interruption to operation caused by a temperature rise due to
a control system malfunction
Adjustable upper and lower temperature limit controller
Protective Class 3.3 (upper/lower temperature limit protection),
independent of the temperature controller
● Automatically takes over the temperature control as soon as the
temperature range set by the operator is exceeded in either direction.
● No interruption to operation caused by a temperature rise or fall due to
a control system malfunction
09