Henny Penny HHC-990 User Manual

SECTION 5. TROUBLESHOOTING
5-1. TROUBLESHOOTING GUIDE
PROBLEM CAUSE CORRECTION
Product not • Doors left open • Keep doors closed except holding temperature to load and serve product
• Product held too long • Hold product only for recom­mended times
• Door gasket torn or worn • Replace bad door gaskets
Model HHC-990
Cabinet steaming • Humidity setpoint too high • Decrease humidity setpoint product soggy (SP-4) in Special Program Mode
Product dry • Humidity setpoint too low • Increase humidity setpoint
(SP-4) in Special Program Mode
• No water in pan • Pour water in water pan
Unit not heating • Blown fuse • Change15 amp fuse
Unit not reaching set • Doors left open • Keep doors closed except temperature to load and serve product
• Door gasket torn or worn • Replace bad door gaskets
More detailed troubleshooting information is available in the T echnical Manual, available at www .hennypenny .com,
or 1-800-417-8405 or 1-937-456-8405.
5-1 703
Model HHC-990
5-2. ERROR CODES AND The display shows the following error codes and warnings when a WARNINGS fault is detected, along with an alarm sound. Both the heat and
humidity systems shut down, except when specified otherwise.
Display Cause Panel Board Correction
“E-4 CPU TOO HOT” • Control board too hot;
unit overheating or louvers clogged
“E-5 AIR TEMP TOO HOT” • Faulty relay, PC board,
or air probe
“E-54A CPU TEMP SENSOR • Faulty PC board
OPEN”
• Turn switch to OFF position, then back to ON; if display still shows “E-4”, the PC board is getting too hot; clean louvers and check cooling fan; if cooling fan is not working, have it replaced; once panel cools down, the controls should return to normal; if “E-4” persists, have the PC board replaced
• Turn switch to OFF position, then back to ON; if display shows “E-5”, the heating circuits and temperature probe should be checked; once the unit cools down, the controls should return to normal; if “E-5” persists, have the PC board replaced
• Turn switch to OFF position, then back to ON; if display shows “E-54A”, the control should be re-initialized (see Pro­gramming Section); if the error code persists, have PC board replaced
“E-54B CPU TEMP SENSOR • Faulty PC board
SHOR TED”
• Turn switch to OFF position, then back to ON; if display shows “E-54B”, the control should be re-initialized (see Pro­gramming Section); if the error code persists, have PC board replaced
“E-6A AIR TEMP SENSOR • Faulty air probe
F AILED OPEN”
• Turn switch to OFF position, then back to ON; if the display shows “E-6”, the temperature probe should be checked; once the temperature probe is repaired, or replaced, the controls should return to normal; if “E-6” persists, have the PC board replaced
303 4-7
Loading...
+ 2 hidden pages