Hantek DSO8060 Software Manual

Hantek
®
Hantek
No.177 zhuzhou road(huite industry city), QingDao,China
Electronic co.,Ltd.
DSO8000 SERIES
USER’S
MANUAL
8060
www.hantek.com.cn
DSO8000 SERIES
USER’S MANUAL
8060
8060
Content
General Safety Notice …………………………………….…..3 Digital Scope Meters………………………..…………………4 CHAPTER 1: Getting Start…………………………………... 6
General Check………………………………………....7
User’s Interface……………………………………...…7
Input Connections…………………………………
Function Check……………………………………….13 To compensate probes…………………………
To display a signal automatically…………………....15
Using the Scope Meter……………………………
CHAPTER 2: Operating Scope……………………………..19
Set Vertical System…………………………… Set Horizontal System………………………… Set Trigger System……………………………… Save / Recall Waveforms and Setups……………...48 Utility Function…………………………… Signal Measurement…………………………… Cursor Measurement…………………………
IO Set.......................................................................70
CHAPTER 3: Using examples…………………………
CHAPTER 4: Multimeter………………………………….....81
CHAPTER 5 Waveform Generator....................................95
CHAPTER 6: Troubleshooting…………………………….106
CHAPTER 7: Specications……………
CHAPTER 8: Appendix...................................................110
…………
………………
....
…..
...
…….
…….
…..
…..
….......
…...73
….107
.12
...14
.16
..20
.36 .40
...52
..63
66
DSO8000 SERIES
General Safety Notice
1. Safety Terms and Symbols
Terms in this manual:
These terms may appear in this manual:
WARNING: Warning statements identify conditions or practices that could result in injury or loss of life.
CAUTION: Caution statements identify conditions or practices that could result in damage to this product or other property.
Terms on the Product:
These terms may appear on the product:
DANGER: indicates an injury hazard may be immediately accessible.
WARNING: indicates an injury hazard may be not immediately accessible.
CAUTION: indicates that a potential damage to the instrument or other property
might occur.
Symbols on the Product:
These symbols may appear on the product:
Hazardous Refer to Protective Grounding Test Voltage Instructions Earth Terminal Terminal Grounding of Chassis Terrninal
2. General Safety Information
Carefully read the following safety information in order to avoid any personal injury or damage on this product or any products connected with it. Review the following safety precautions carefully before operate the device to avoid any personal injuries or damages to the device and any products connected to it. To avoid potential hazards use
the device as specied by this user’s manual only.
■ To Avoid Fire or Personal Injury.
■ Use Proper Power adapter.
and certied for the country of use.
Use only the power cord specied for this product
8060
■ Connect and Disconnect Properly.
leads while they are connected to a voltage source.
■ Connect and Disconnect Properly.
measurement device before connecting the probe to the circuit under test. Disconnect the probe input and the probe reference lead from the circuit under test before disconnecting the probe from the measurement device.
■ Observe All Terminal Ratings.
and markings on the product. Consult the product manual for further ratings information before making connections to the product.
■ Use Proper Probe.
measurement.
To avoid shock hazard, use a properly rated probe for your
■ Avoid Circuit or Wire Exposure.
components when power is on.
■ Do Not Operate With Suspected Failures.
device, have it inspected by qualied service personnel before further operations.
■ Provide Proper Ventilation.
ventilation of the device.
Refer to the installation instructions for proper
Do not connect or disconnect probes or test
Connect the probe output to the
To avoid re or shock hazard, observe all ratings
Do not touch exposed connections and
If suspected damage occurs with the
■ Do not operate in Wet/Damp Conditions.
■ Do not operate in an Explosive Atmosphere.
■ Keep Product Surfaces Clean and Dry.
Digital Scope Meters
DSO8000 Series
measurements in a compact, lightweight package. DSO8000 series is ideal for production test, eld service, research, design, education and training involving
applications of analog circuit tests and troubleshooting.
digital scope meters offer exceptional waveform viewing and
Product features:
■ Dual Channel, Bandwidth:
60MHz (DSO8060)
■ Maximum real-time sampling rate: 150MSa/s (DSO8060)
DSO8000 SERIES
■ Memory depth:
32K points (Single Channel), 16K points (Dual Channels)
■ Color TFT LCD, 320×240 pixels resolution.
■ USB storage and printing supports, rmware upgrade via USB interface.
■ Adjustable waveform intensity, more effective waveform viewing.
■ One-touch automatic setup, ease of use (AUTOSET).
■ 15 Waveforms, 15 setups, supports CSV and bitmap format.
■ 22 Automatic measurements.
■ Automatic cursor tracking measurements.
■ Waveform recorder, record and replay dynamic waveforms.
■ User selectable fast offset calibration.
■ Built-in FFT function, Frequency Counter.
■ Pass/Fail Function.
■ Addition, Subtraction, Multiplication and Division Mathematic Functions.
■ Advanced trigger types include: Edge, Pulse width,ALT,Video.
■ Multiple Language User Interface.
■ Pop-up menu makes it easy to read and easy to use.
■ Built-in multi-language help system.
■ Easy-to-use le system supports Multi-Language characters le name input.
■ Selectable bandwidth limit:
DSO8060:20MHz
8060
CHAPTER 1: Getting Start
This chapter covers the following topics:
General Check
Functional Check
The User Interface
Input Connections
To compensate probes
To display a signal automatically
Using the Scope Meter
DSO8000 SERIES
General Check
When you have got a new DSO8000 series scope meter, it is suggested that you should
perform a general inspection on the instrument according to the following steps:
■ Check the shipping container for damage:
Keep the damaged shipping container or cushioning material until the contents of the shipment have been checked for completeness and the instrument has been checked mechanically and electrically.
■ Check the accessories:
Accessories supplied with the instrument are listed in “Accessories” in this guide. If the
contents are incomplete or damaged,please notify our distributor at your local area or
the overseas sales dept.
■ Check the instrument:
In case there is any mechanical damage or defect, or the instrument does not operate properly or fails performance tests, please notify our distributor at your local area or
the overseas sales dept.
User’s Interface
The rst thing to do with a new scope meter is to know its front panel. This chapter
helps to be familiar with the layout of the keys and how to use them. Read this chapter carefully before further operations.
Front Panel (Figure 1-1):
The buttons not only allow you to use some of the functions directly but also bring up the
manual keys on the screen, which enable the access to many measurement features associated with advanced functions, mathematics, and reference or to run control
features.
8060
v
Figure 1-1 Scope Meter Front Panel
DSO8000 SERIES
Figure 1-2 Front Panel Description
Description
1. LCD Display
2. F1~F5: Sets or switch options for the menu
3. Arrow Keys
4. HORI: Shows Horizontal menu
5. TRIG: Shows Trigger menu
6. LEVEL: Adjust the trigger level
7. RUN/STOP: key for running or stopping the operation
8. AUTO: Be used for auto setting under the oscilloscope operation mode
8060
9. TIME/DIV: Decrease or Increase the time base
10. POSITION: Adjust the horizontal trigger position
11. CH2: Shows the CH2 menu
12. VOLTS: Decrease or Increase the voltage/div
13. CH1: Shows the CH1 menu
14. M/R: Shows the Math or REF menu
15. DMM Buttons: The DMM control buttons
16. DMM/SCOPE: Switch DMM , Scope or RPM function
17. MENU ON/OFF: Turn on/off the menu
18. SAVE RECALL: Shows SAVE or RECALL menu
19. MEAS: Shows Measurement menu
20. UTILITY: Shows Utility menu
21. CURSOR: Shows Cursor menu
Display screen
1 15 3 4 5 6 7 8
21
20
19
18 17
16
Figure 1-3 LCD Display screen
Description
1. Shows Brand
2. Shows horizontal trigger time
9
10
11
12
13
2
14
0
3. Shows location of the current waveform in the memory
4. Shows the trigger position in the memory
5. Shows the trigger mode
6. Shows the trigger source
7. Shows the trigger level
8. Shows the power
9. The center of current waveform window
10. CH1 waveform
11. The trigger level symbol
12. The grid
13. CH2 waveform
14. Shows the time base
15. Shows running staturs
16. CH1/CH2
17. Shows the coupling
18. Shows the voltage/div
19. CH2 mark
20. CH1 mark
21. Shows the trigger position in current waveform window
DSO8000 SERIES

8060
Input Connections
See the following Figure 1-4:
4
5
1
3
2
Figure 1-4 Input Connections
Description:
1. The power adapter is supplied for AC power supply and battery recharging.
2. Multi-meter input jacks, including four circular banana jacks .The four circular jacks are used for voltage,resistance,mA range current and A range current inputs.
3. Multi-meter test lead.
4. Scope Meter probes.
5. Scope Meter input channels.

DSO8000 SERIES
Function Check
Perform this quick functional check to verify that your scope meter is operating correctly.
1. Turn on the instrument.
Use the power adapter designed for your scope meter only. Use a power source that
delivers 100 to 240 VACRMS, 50Hz. Turn on the scope meter.
2. Input a signal to a channel.
Set the attenuation switch on the probe to X10 and connect the probe to CH1 on the scope meter. To do this:
Align the slot in the probe connector with the key on the CH1 BNC connector.
Push to connect, and twist to the right to lock the probe in place.
Attach the probe tip and ground lead to the OUTPUT port.(Before this,you must press
OSC/DMM key to Enter Waveform Generator mode and output a 1kHz 2V peak -to-peak square wave.And then Enter the Oscilloscope mode.)
Set the probe attenuation of the scope meter to 10X.To do this, push CH1→Probe→10X
.
Figure 1-5 Set the probe
3. Push the (approximately 1 kHz 2 V peak- to- peak).
Turn off CH1 and turn on CH2, repeat steps 2 and 3.
button. Within a few seconds, a square wave will display
AUTO

8060
To compensate probes
Perform this adjustment to match the characteristics of the probe and the channel input.
This should be performed whenever attaching a probe to any input channel the rst time.
1. From CH1 menu, set the probe attenuation to 10X (press
Set the switch to X10 on the probe and connect it to CH1 of the scope meter.
When usiing the probe hook-tip, insert the tip onto the probe rmly to ensure proper
connection. Attach the probe tip to the probe compensator connector and the reference lead to the
ground pin, Select CH1, and then press AUTO.
2. Check the shape of the displayed waveform.
Correctly Compensated
CH1→Probe→10X
).
Over compensated

DSO8000 SERIES
Under Compensated
3. If necessary, use a non-metallic tool to adjust the trimmer capacitor of the probe for the attest square wave being displayed on the scope meter.
4. Repeat if necessary.
WARNNING: To avoid electric shock while using the probe, be sure the perfection of the insulated cable, and do not touch the metallic portions of the probe head while it is
connected with a voltage source.
To display a signal automatically
The scope meter has an Auto feature that automatically sets up the scope meter to best
display the input signal. Using Auto requires signals with a frequency greater than or equal to 50 Hz and a duty cycle greater than 1%.
Press the AUTO button, the scope meter turns on and scales all channels that have signals applied, and it selects a time base range based on the trigger source. The trigger
source selected is the lowest numbered input that has a signal applied. The scope meter has two-channels input . Connect a signal to the CH1 input.
1.Connect a signal to the scope meter.
2.Press AUTO.
The scope meter may change the current setting to display the signal. It will automatically
adjust the vertical and horizontal scaling, the trigger coupling, type, position, slope, level,
and mode settings.

8060
Using the Scope Meter
This part provides a step-by-step introduction to the scope functions. The introduction does not cover all of the capabilities of the scope functions but gives basic examples to show how to use the menus and perform basic operations.
Turn on the scope meter
Connect scope meter to AC power via a power adapter. (The scope meter can still work with built-in Li-ion battery even without AC power supply.)Turn the scope meter on by pressing down the power on/off key.The instrument then performs Self-checking after
power on. The welcome picture will display on the screen when the system nishes Self-
checking.
The scope meter is powered on in its last setup conguration.
Figure 1-6 The Login window
Menu Operation
The following example shows how to use the tools menus to select a function, as shown in the following gure.

DSO8000 SERIES
Figure 1-7 The Menu
1. Press the screen and the corresponding optional settings on the bottom. Press
OFF
2. Choose one key from F1 to F5 and press it to change function setting.
MENU ON/OFF
again to hide the Function Menu.
key to display the Function Menu on the bottom of the
MENU ON/
Set up the vertical system
1. Change the vertical setup and notice that each change affects the
status bar differently.
Change the vertical sensitivity with or button and notice the change in the
status bar.
2. Move the signal vertically. The / button moves the signal vertically. Also notice that the channel symbol
on the left side of the display moves in conjunction with the button.
To set up the horizontal system
1. Change the time base.

8060
The or button changes the time base in a 1-2-5 step sequence, and displays
thevalue in the status bar.
2. Move signal horizontally.
The or button moves displayed signal horizontally on waveform window. It sets
the trigger point position.
Set up the trigger system
1. Change the trigger Level
The or button changes the trigger level. The trigger level value is displayed
at the top-right of the screen and a line is displayed showing the location of the trigger level.
2. Change the trigger setup and notice these changes in the status bar.
Press TRIG button in the key panel.Choose one key from F1 to F5 and press it to
change function setting.

DSO8000 SERIES
CHAPTER 2: Operating Scope
The end user should know how to determine the system setup from the status bar of a Scope Meter. This chapter will detail the scope meter function of the test tool.
Set Vertical System
Set Horizontal System
Set Trigger System
Save and Recall waveforms and setups
Utility Function
Measure Signal
Cursor Measure

8060
Set Vertical System
Each channel of Scope meter has its own independent operation menu and it will pop up after pressing CH1 or CH2 button. The settings of all items in the menu are shown in the table below.
To make vertical CH1 and CH2 settings, do the following:
1. Press the CH1 or CH2 button and then the function menu appears at the bottom of the
screen.
2. Select and press key from F1 to F5 keys to make different settings.
Now, you can nd the menu that looks like the following gure 2-1.
The CH1 menu (Page 1/2)
The CH1 menu (Page 2/2)
The CH1 menu (Page 1/3)(not include DSO8060)
The CH1 menu (Page 2/3)(not include DSO8060)
The CH1 menu (Page 3/3)(not include DSO8060)
0
DSO8000 SERIES
Figure 2-1 The CH1 Menu
The following table describes the channel menu:
Menu Setting Description
Enable
Coupling
Probe
Reset Set the vertical display position back to middle window
1/2
ON
OFF
AC
DC
GND
10X
100X
1000X
Turn on Channel Turn off Channel
The dc component in the input signal is blocked The ac and dc components of the input signal are allowed Disconnect the input signal
1X
Select one according to the probe attenuation factor to ensure correct vertical scale reading
Go to the next menu page
2/2
Volt/Div
Invert
BW 20M
2/2
3/3
BW 100M
3/3
Coarse
Fine
ON
OFF
ON
OFF
Back to the previous menu page
Set Voltage range coarsely
Set Voltage range nely
Turn on the invert function Turn off the invert function
Turn on the BW 20M(not include DSO8060) Turn off the BW 20M
Back to the previous menu page
Go to the next menu (not include DSO8060)
Turn on the BW 100M(not include DSO8060) Turn off the BW 100M
Go to the rst menu(not include DSO8060)

8060
BW Limit
Turn on the BW Limit Turn off the BW Limit
1. Change Volt/DIV
It is the default setting of Volts/Div in a 1-2-5-step sequence from 1mV/div,2mV/Div,5mV/ div or 10mV/div, 20mV/div, 50mV/div, ... , to 1V/div,2V/div,5V/div.
The Volt/DIV will be displayed in the status bar on the bottom of the screen.
Figure 2-2 Channel Volt/Div
Press CH1→Volt/Div→Coarse/Fine to Set Volt/Div range.
2. Set Channel Coupling
With CH1 taken for example, input a sine wave signal containing a dc offset. Press CH1→Coupling→AC to set “AC” coupling. It will pass AC component but block the
DC component of the input signal. The waveform is displayed as Figure 2-3

DSO8000 SERIES
Figure 2-3 Waveform Display
Press CH1→Coupling→DC, to set “DC” coupling. It will pass both AC and DC
components of the input signal. The waveform is displayed as Figure 2-4.

8060
Figure 2-4 Waveform Display
Press
CH1→Coupling→GND
, to set “GND” coupling, it disconnects the input signal.
The screen displays as Figure 2-5:

DSO8000 SERIES
Figure 2-5 Waveform Display
3. Set Probe Attenuation
The scope meter allows adjusting the probe attenuation scale factor correspondingly in the channel operation menu in order to comply with the probe attenuation scale. The attenuation factor changes the vertical scaling of the scope meter so that the
measurement results reect the actual voltage levels at the probe tip. To change (or check) the probe attenuation setting, press the CH1 or CH2 button
(according to which channel in using). Toggle the Probe soft button to match the attenuation factor of the probe.
This setting remains in effect until changes again.
Figure 2-6 Probe Setting

8060
4. Invert a waveform
The displayed waveform reverses 180 degrees relatively to the ground potential. Press
CH1 or CH2 → F5 → F3
, to turn on/off the
Invert.
Figure 2-7 Turn Invert off

DSO8000 SERIES
Figure 2-8 Turn Invert on
5. Set Band-Width Limit
With CH1 taken for example, input a signal containing high frequency component.
Press CH1→F5→F4→OFF ,to set up bandwidth limit to “OFF” status. The scope meter is set to full bandwidth and passing the high frequency component in the signal.
The waveform is displayed as Figure 2-9.

8060
Figure 2-9 Turn BW limit Off
Press CH1→F5→F4→ON, to set up bandwidth limit to “ON” status. It will reject the frequency component higher than 20MHz.
Press CH1→F5→F5→F2→ON, to set up bandwidth limit to “ON” status. It will reject the frequency component higher than 100MHz.This function is not used in DS08060.
The waveform is displayed as Figure 2-10:

Loading...
+ 84 hidden pages