Hanna Instruments LP 2000 User Manual

Instruction Manual
LP 2000
Precision Bench
Turbidity Meter
www.hannainst.com
Dear Customer, Thank you for choosing a HANNA instruments® product. Please read the instruction manual carefully before using the instru-
ment, to have all the necessary information for a correct use. If you need additional technical information, do not hesitate to e-mail
us at tech@hannainst.com This instrument is in compliance with the directives.
PRELIMINARY EXAMINATION
Remove the instrument from the packing material and examine it to make sure that no damage has occurred during shipping. If there is any damage, notify your dealer immediately.
LP 2000 is supplied complete with:
• Glass cuvet with cap
• Calibration solutions (HI 93703-0 AMCO-EPA-1 @0 FTU and HI 93703-10 AMCO-EPA-1 @10 FTU), 30 mL each
• 12 Vdc power adapter and instructions
Note: Save all packing material until you are sure that the instrument
functions correctly. Any damaged or defective item must be returned in its original packaging with the supplied accessories.
TABLE OF CONTENTS
PRELIMINARY EXAMINATION ...................................................... 3
GENERAL DESCRIPTION .............................................................. 3
MEASURE UNITS ....................................................................... 4
PRINCIPLE OF OPERATION ......................................................... 4
FUNCTIONAL DESCRIPTION ........................................................ 5
SPECIFICATIONS......................................................................... 6
OPERATIONAL GUIDE ................................................................. 7
CALIBRATION ........................................................................... 10
LCD AND ERROR CODES ........................................................... 13
ACCESSORIES ........................................................................... 14
WARRANTY .............................................................................. 14
CE DECLARATION OF CONFORMITY............................................ 15
GENERAL DESCRIPTION
LP 2000 is a bench microprocessor-based meter used to determine the turbidity of water and wastewater. The meter covers the 0 to 1000 FTU* range with 0.01 resolution from 0.00 to 50.00 FTU* and in single digits from 50 to 1000 FTU*. It also auto-ranges between the two scales.
Note: LP 2000 has been designed according to the ISO 7027
International Standard. The turbidity measurement unit is FTU (Formazine Turbidity Unit), which is identical to NTU (Nephelometric Turbidity Unit).
The meter is housed in a rugged and lightweight case, with an easy­to-read LCD.
The meter is simple to use: all operations can be carried out with only four keys and troubleshooting functions can be performed with error codes on the display. An exclusive positive-locking system guarantees that the cuvette is firmly placed in the cell. The keypad is water­resistant and can be wiped clean.
Calibration at 10 FTU* can be easily performed using the supplied standard. In addition, LP 2000 is the first bench turbidity meter that allows the storage and retrieval of the last calibration date.
10 FTU* was chosen as the calibration point because its value best fits the water turbidity measurements in a wide range of applications, from drinking water to wastewater treatment.
* 1 FTU = 1 NTU
32
MEASURE UNITS
CUVET
LIGHT
D
Hanna Instruments uses the primary standard AMCO-AEPA-1 to avoid all formazine-related problems. Formazine is a toxic, unstable substance, which requires particular care. Calibration requiring formazine standards must be preformed immediately after preparation of the standards. Formazine standards may not be reused because of their short life. LP 2000 standards are extremely stable and have a shelf life of up to six months. Uncontaminated standards may be reused. LP 2000 can be used with both AMCO-AEPA and formazine stan­dards.
FTU is equal to the NTU. However, other known measurements include: Jackson Turbidity Unit or JTU (based on the old method of Jackson's candle), and Silica Unit in mg/L of SiO2. See the conversion table below for these units:
JTU FTU/NTU SiO2 (mg/L)
JTU 1 19 2.5
FTU /NTU 0.053 1 0.13
SiO2 (mg / L) 0.4 7.5 1
PRINCIPLE OF OPERATION
FUNCTIONAL DESCRIPTION
1) Cuvette holder
2) Dual level LCD (Liquid Crystal Display)
3) ON/OFF key, to switch the meter ON and OFF
4) CAL key, to enter calibration mode
5) READ/, to perform measurements, and set the day and month of the last calibration
6) DATE/ , to display the last calibration date, and select the
month or day of the last calibration
LP 2000 has been designed to perform measurements according to the ISO 7027 International Standards.
The instrument functions by passing a beam of infrared light through a vial containing the sample being measured.
The light source is a High Emission Infrared LED with a wavelength peaking at 890 nm, which reduces interference caused by colored samples to a minimum.
A sensor, positioned at 90° with respect to the direction of light, detects the amount of light scattered by the undissolved particles present in the sample. The microprocessor converts such readings into FTU values.
DETECTOR
IR LED
EMITTED LIGHT (890 nm)
* 1 FTU = 1 NTU
90° SCATTERE
LIGHT
i) Socket for 12 Vdc power adapter
54