Visual Quick Start Guide
Meter Setup
Electrode Storage
• After use , rinse the elec trode wit h water and repla ce
the prote ctive cap fil led with a few dro ps of HI70300
storage solution.
DO NOT USE DIST ILLED OR DEIONIZED WATER FO R
STORAGE PURPOSES.
Electrode Cleaning
• Soak the el ectrode in HI7 00643P electr ode cleaning
solutio n for approxima tely 15 minutes t o dissolve
deposit s and other coat ings. After cl eaning, place i n
storage solution for 1 hour before use.
ON/OFF/MODE Button
SET/HOLD Button
Freeze Display
Press any bu tton
to return t o
measurement
mode.
3
sec.
Battery Replacement
To replace the bat teries:
• Open the ba ttery com partment c ap (on the bottom o f
the instrument).
• Remove old batteries.
• Replace new b atteries , observin g the polarit y.
• Close th e battery co mpartmen t cap.
HI99164
Hanna Instr uments reser ves the right to m odify the de sign, constru ction,
or appearance of its products without advance notice.
6
sec.
Care and Maintenance
HI70300
storage
solution
15
min.
1
hour
HI700643P
electrode
cleaning solution
HI70300
storage
solution
QR99164 10/ 15
Pressing t he
ON/OFF/MODE
button a fter
buer sele ction
returns t o
measurement
mode.

pH Calibration
†
3
sec.
pH 7.0 1 pH 4.01 pH 10.01
pH 7.0 1pH 4.01 pH 10.01
For one- point calibr ation using pH 4,
pH 7, or pH 10 buer go to pr ocedure B
For two-point calibration using
pH 7 buer go t o procedur e A
A
B
Two-Point Calibration u sing pH 7 Buer
A
B
One-Point Cali bration using pH 4, pH 7, or pH 10 Buer
Continue w ith
pH 4.01 or pH
10.01 buer
solution
†
†
The probe ti p should be rin sed with puri ed water (rever se
osmosis, dis tilled, or dei onized) befo re and after p lacing in
any solutio n (buer, storag e or sample).