Hanna Instruments HI 991003 User Manual

Instruction Manual
HI 991003
Portable pH/ORP/Temperature
Meter with Sensor Check
http://www.hannainst.com
WARRANTY
All Hanna Instruments meters are warranted for two years against defects in workmanship
and materials when used for their intended pur­pose and maintained according to instructions. The
probes are warranted for a period of six months.
This warranty is limited to repair or replacement free of charge. Damage due to accidents, misuse, tampering or lack of prescribed maintenance are not covered.
If service is required, contact the dealer from whom you purchased the instrument. If under warranty, report the model number, date of purchase, serial number and the nature of the problem. If the repair is not covered by the warranty, you will be notified of the charges incurred. If the instrument is to be returned to Hanna Instruments, first obtain a Re­turned Goods Authorization Number from the Cus­tomer Service department and then send it with shipment costs prepaid. When shipping any instru­ment, make sure it is properly packaged for com­plete protection.
All rights are reserved. Reproduction in whole or in part is prohibited without the written consent of the copyright owner.
This Instrument is in Compliance
with the CE Directives
Dear Customer, Thank you for choosing a Hanna Instruments
Product. Please read this instruction manual carefully
before using the instrument. This manual will provide you with all the necessary informa­tion for the correct use of the instrument, as well as a precise idea of its versatility in a wide range of applications.
This instrument is in compliance with the directives.
PRELIMINARY EXAMINATION
Remove the instrument from the packing ma­terial and examine it carefully to make sure that no damage has occurred during ship­ment. If noticeable damage is evident, notify your dealer.
Note: Save all packing material until you
are sure that the instrument functions correctly. All defective items must be returned in the original packing to­gether with the supplied accessories.
GENERAL DESCRIPTION
HI 991003 is a portable pH/pH-mV/ORP/tem-
perature meter, with unique Sensor Check feature, that allows the user to determine the electrode status at any time. Main features also include: extended mea­surement ranges; waterproof and compact casing; large dual-line LCD; low battery de­tection; automatic pH calibration at one or two points with two memorized buffer sets; selectable temperature unit (°C or °F). The pH/ORP electrode has a built-in tem­perature sensor for simultaneous pH (or mV) and temperature readings, and also contains a mini amplifier to render mea­surements impervious to noise and electri­cal interferences. Each meter comes together with
HI 1297D pH/ORP/temperature probe with
DIN connector and 1 m (3.3’) cable
3x1.5V AA alkaline batteries
instruction manual
plastic carrying case.
LCD DESCRIPTION
1. Automatic Temperature Compensation indicator
2. Stability indicator
3. Battery percentage
4. pH calibration indicator
5. Low battery indicator
6. Selectable temperature unit
7. Secondary display
8. Primary display
9. Measuring unit for primary display
ACCESSORIES
HI 1297D Combination amplified pH/
ORP/temperature probe with DIN and 1 m (3.3’) cable
HI 1296D Combination amplified pH/tem-
perature probe with DIN con­nector and 1 m (3.3’) cable
HI 7004L pH 4.01 buffer solution, 500 mL HI 7006L pH 6.86 buffer solution, 500 mL HI 7007L pH 7.01 buffer solution, 500 mL HI 7009L pH 9.18 buffer solution, 500 mL HI 7010L pH 10.01 buffer solution, 500 mL HI 70300L Storage solution, 500 mL HI 7061L Cleaning solution, 500 mL HI 77400P Calibration kit (pH 4&7, 20 mL, 5
pcs each)
HI 721312 Hard carrying case HI 710007 Blue shockproof rubber boot HI 76405 Electrode holder
SPECIFICATIONS
Range (*) -2.00 to 16.00 pH
±825 mV (pH-mV)
±1999 mV (ORP)
-5.0 to 105.0°C / 23.0 to 221.0°F
Resolution 0.01 pH
1 mV
0.1°C / 0.1°F
Accuracy ±0.02 pH (@20°C/68°F) ±2 mV
±0.5°C up to 60°C; ±1°C outside
±1°F up to 140°F; ±2°F outside
Typical EMC ±0.02 pH Deviation ±2 mV
±0.2°C or ±0.4°F
Temperature Automatic for pH readings Compensation
pH Calibration Automatic, 1 or 2 point
with 2 sets of standardized buffers
(4.01/7.01/10.01 or 4.01/6.86/9.18)
mV Calibration factory calibrated Probe (included) HI 1297D
amplified pH/ORP/temperature
probe with DIN connector and 1 m (3.3’) cable
Battery Ty pe 3 x 1.5V AA / IEC LR6 Battery Life approx. 1500 hours Auto-off after 8 minutes of nonuse Environment 0 to 50°C (32 to 122°F);
RH 100%
Dimensions150 x 80 x 38 mm (5.9 x 3.2 x 1.5”) Weight 245 g (8.6 oz.)
(*) the temperature range is limited to 80°C (176°F) if
using the HI1296D or HI1297D probes.
To clean the meter, use water only.
Hanna Instruments reserves the right to modify the design, construction and appearance of its products without advance notice.
IST991003 06/03
OPERATIONAL GUIDE
To connect the probe
With the meter turned off, connect the HI 1297D probe to the DIN socket on the top of the meter by aligning the pins and pushing in the plug. Tighten the nut to ensure a good connection. Remove the protective cap from the probe before taking any measurements.
To turn the meter ON
Press the ON/OFF/MODE button until the display lights up. At start-up, all the LCD segments are displayed for 1 second, then the percent indica­tion of the remaining battery life is shown for an­other second (E.g. % 100 BA TT). The meter then enters the normal measuring mode. Note: If the display needs to be checked, keep the
ON button pressed while turning the meter on. The meter will display all segments as long as the button is pressed.
To select the measurement range
While in normal measurement mode, press the SET/HOLD button to select pH, pH-mV or ORP (mV) measurement range. Note: “pH-mV” indicates the mV value equivalent
to the pH.
To freeze the display
While in measurement mode, press the SET/ HOLD button until HOLD appears on the second­ary display; the reading will be frozen on the LCD (E.g. pH 5.73 HOLD). Press any button to return to normal mode.
To turn the meter OFF
While in normal measurement mode, press the ON/OFF/MODE button. OFF will appear on the secondary display. Release the button. Note: the meter is provided with an acoustic sig-
nal feature, which can be disabled using the switch located in the battery compartment.
Note: When the meter detects the absence of a
temperature probe at its input, the Automatic Temperature Compensation is turned off, and the meter uses a default value of 25°C (77°F) for the temperature measurement and com­pensation. In this condition, the secondary LCD shows 25.0°C (77.0°F) blinking. When a probe is connected, the meter automatically returns to the ATC mode, the A TC t ag is turned on, and the temperature is shown on the secondary display .
pH MEASUREMENT & CALIBRATION
• Make sure the meter has been calibrated be­fore use.
• If the probe is dry, soak it in HI 70300 storage solution for one hour to reactivate it.
• Set the meter to pH range.
• Place the probe into the sample to be tested.
• Stir briefly and wait until the symbol on the top left of the LCD is turned off.
• The LCD shows the pH value (automatically compensated for temperature) and the tem­perature of the solution.
• If measurements are taken in different samples successively, rinse the probe tip thoroughly to eliminate cross-contamination. After cleaning, rinse the probe tip with some of the sample to be measured.
pH calibration
• Press and hold the ON/OFF/MODE button until OFF on the secondary display is replaced by CAL.
• Release the button. The LCD enters the cali­bration mode, displaying “pH 7.01 USE” (or “pH
6.86 USE” if the NIST buffer set was selected). After 1 second the meter activates the automatic buffer recognition feature. If a valid buffer is de­tected, then its value is shown on the primary display, and REC appears on the secondary display. If no valid buffer is detected, the meter keeps the USE indication active for 12 seconds, and then it replaces it with WRNG, indicating that the sample being measured is not a valid buffer.
• For a single-point calibration with buffers pH
4.01, 9.18 or 10.01, the meter automatically ac­cepts the calibration when the measurement is stable; the meter will show on the primary display the accepted buffer, with the message “OK 1” on the secondary display, and an au­dible signal is produced. After 1 second, the meter automatically returns to the normal mode. If a single-point calibration with buffers pH 7.01 or 6.86 is desired, then after the calibration point has been accepted the ON/OFF/MODE button must be pressed in order to return to the nor­mal measuring mode. After the button is pressed, the meter shows "7.01" (or "6.86") ­"OK 1", and an audible signal is produced. After 1 second, the meter automatically returns to
the normal measuring mode.
• For a
Note: When the calibration procedure is com-
To quit calibration and to reset to the default
• After entering the calibration mode and before
• To reset to the default values and clear a previ-
two-point calibration, place the probe in pH 7.01 (or pH 6.86) buffer . After the calibration point has been accepted, the “pH 4.01 USE” message appears. The message is held for 12 seconds, unless a valid buffer is recognized. If no valid buffer is recognized, then the WRNG message is shown. If a valid buffer (pH 4.01, pH 10.01 or pH 9.18) is detected, then the meter completes the calibration procedure. When the buffer is accepted, the LCD shows the accepted value with the “OK 2” message on the second­ary display. The meter then returns to the nor­mal measuring mode.
pleted, the CAL tag is turned on.
the first point is accepted, it is possible to quit the procedure and return to the last calibration data by pressing the ON/OFF/MODE button. The secondary LCD displays ESC for 1 second and the meter returns to normal mode.
ous calibration, press the SET/HOLD button after entering the calibration mode and before the first point is accepted. The secondary LCD displays CLR for 1 sec­ond, the meter resets to the default calibration and the “CAL” tag on the LCD disappears.
SENSOR CHECK
The Sensor Check feature allows the user to check the electrode status at any time. This is possible by setting the meter to pH-mV range. The offset value is the reading of the electrode immersed in pH7 buffer (@25°C/77°F). If this read­ing is outside the range ±40 mV, the electrode is considered "poor" and it is recommended to re­place it with a new one. The slope value of the electrode is the difference between readings in pH7 and in pH4 buffers. For a new electrode this is about 174 mV (@25°C/
. With time, the slope value decreases. When
77°F)
it reaches the value of about 150 mV, the elec­trode is considered "poor", and it is recommended to replace it with a new one. Note: Always recalibrate the meter after replacing
the electrode.
Note: To ensure reliable readings, the electrode
must be cleaned with cleaning solution before measuring the offset and the slope.
ORP MEASUREMENT
• Make sure the probe is connected.
• Set the meter to ORP (mV) range.
• Place the probe into the sample to be tested, stir briefly and wait until the symbol on the top left of the LCD is turned off.
• The meter will show the ORP (mV) value and the temperature of the sample.
Note: The ORP range is factory calibrated. Note: ATC tag turns off because the mV readings
are not temperature compensated.
METER SETUP
Setup mode allows the selection of the tempera­ture unit and of the pH buffer set. To enter the Setup mode, press and hold the ON/ OFF/MODE button until CAL on the secondary dis­play is replaced by TEMP and the current tempera­ture unit (E.g. TEMP °C). Then:
for °C/°F selection, use the SET/HOLD button. After the temperature unit has been selected, press ON/OFF/MODE to enter the buffer set se­lection mode; press ON/OFF/MODE twice to re­turn to the normal measuring mode.
to change the calibration buffer set, after setting the temperature unit, the meter will show the current buffer set: “pH 7.01 BUFF” (for 4.01/7.01/
10.01) or “pH 6.86 BUFF” (for 4.01/6.86/9.18). Change the set with the SET/HOLD button, then press ON/OFF/MODE to return to normal mode.
BATTERY REPLACEMENT
The meter displays the remaining battery per­centage when turned on. When the level is below 5%, the the LCD blinks to indicate a low battery condi­tion. If the battery level is low enough to cause erroneous readings, the Battery Error Preven­tion System (BEPS) turns the meter off. Unscrew the 4 screws located on the back of the meter and carefully replace the 3 AA bat­teries located in the battery compartment, while paying attention to their polarity. Reattach the back making sure that the gasket is in place and tighten the 4 screws.
- +
symbol on the bottom left of
Loading...