Hanna Instruments HI 98203 User Manual

SALINTEST
HI98203
Salinity Concentration Pocket-sized Meter
GENERAL INFORMATION:
The SALINTEST utilizes a sodium ion-selective glass electrode to determine the activity of sodium ions in solution.
When dissolved, NaCl ionizes to form Na+ and Cl ions.
Measurement of concentration of the Na+ ions is an indicator of the concentration of NaCl.
The relationship between the SALINTEST reading and the g/L NaCl scale is shown in the logaritmic chart below.
A double junction reference is used to ensure a highly stable reading.
If you suspect that the calibration has drifted, you ca recalibrate by using HI7081 or a solution of known concentration (adjust the reading with the calibration trimmer).
HOW TO USE THE CHART:
• Locate the pNa reading on the horizontal axis of the chart.
• Move vertically upwards to intersect with the 45º line.
• Move horizontally and read the g/L NaCl value.
E.g. Display = 0.42
g/L NaCl = 22
OPERATION:
-
• Remove the protective cap and turn the meter on with the ON/OFF switch located on the top.
• Immerse it into solution to be tested without exceeding the maximum immersion level.
• Stir gently and wait for the read­ing to stabilize.
• Read the value on display and use the chart to convert the reading to g/L NaCl.
• After use, switch the meter off and rinse the electrode with water to minimize contamination. Always re­place the protective cap after use.
• Large differences in pNa readings (±0.1) could be due to a dry electrode or lack in calibration.
• To improve performance, leave the meter in HI7081 solution for a few minutes at least once a week.
Manufacturers since 1978
SALINTEST
HI98203
Salinity Concentration Pocket-sized Meter
SPECIFICATIONS:
RANGE 0.00 to 1.00 pNaCl
(58.4 to 5.84 g/L NaCl)
RESOLUTION 0.01 pNaCl ACCURACY (@20°C/68°F) ±0.02 pNaCl TYPICAL EMC DEVIATION ±0.04 pNaCl ENVIRONMENT 0 to 50C (32 to 122F); 95% RH BATTERY TYPE 4 x 1.5V alkaline (included) BATTERY LIFE approx. 500 hours of continuous use DIMENSIONS 175x41x23 mm (7.9x1.8x1") WEIGHT 78 g (2.7 oz.)
ACCESSORIES:
HI 7081L 30 g/L NaCl solution (460 mL bottle) HI 7081M 30 g/L NaCl solution (230 mL bottle) HI 7061M Electrode cleaning solution
(230 mL bottle)
HI 73202 Spare electrode HI 731326 Calibration screwdriver (20 pcs)
ONE YEAR WARRANTY AND SERVICEABLE:ONE YEAR WARRANTY AND SERVICEABLE:
ONE YEAR WARRANTY AND SERVICEABLE:
ONE YEAR WARRANTY AND SERVICEABLE:ONE YEAR WARRANTY AND SERVICEABLE:
Covered by manship and materials, this tester is now completely serviceable. Contact your dealer for further information.
The
one yearone year
one year warranty against defects in work-
one yearone year
SALINTEST SALINTEST
SALINTEST is in compliance with the CE directives.
SALINTEST SALINTEST
Visit our Internet Home Page:
http://www.hannainst.com
IS98203R2 05/01
CALIBRATION:
• Immerse the tester in HI7081 (30 g/L NaCl) cali­bration solution, without ex­ceeding the maximum im­mersion level.
• Allow the reading to stabi­lize and with a small screw­driver turn the calibration trimmer until the display
shows "0.30".
Calibration is now complete.
MAINTENANCE:
In case of erroneous reading even after an accurate calibration, the reference junction can be contami­nated or clogged. Pull out 2 mm (1/8'') of the cloth junction to renew the electrode reference (it is recom­mended to cut the cloth leav­ing always at least 2 mm -1/ 8'' over the reference compart­ment) and repeat the calibra­tion procedure. The cloth junction can be pulled out approximately 20 times. After that, the electrode can be replaced (see the Electrode Replacement section).
BATTERY REPLACEMENT:
If display fades or the batteries must be replaced. Slide off the battery compartment cover and replace all four 1.5V batter­ies while paying attention to their polarity.
ELECTRODE REPLACEMENT:
The electrode can be easily replaced in the following way:
• Slide off the battery cover.
• Remove the screw on the back of the located below the battery compartment and slide off the whole internal part of the tester.
SALINTESTSALINTEST
SALINTEST switches off, the
SALINTESTSALINTEST
SALINTESTSALINTEST
SALINTEST
SALINTESTSALINTEST
• Remove the fastening screw on the front (1) to loose the electrode. The internal circuit is con­nected to the electrode trough two wires (one for the glass sensor and one for the reference). Remove the two small screws (2 & 3) which fasten the two connecting wires to their sock­ets.
Internal circuit:
• Remove the electrode and replace it with a new HI73202.
• Fasten first the screw on the front (1) to attach the electrode to the circuit. Insert the two small wires into their sockets and fasten the two screws (2 & 3).
• The new electrode is supplied with four O-rings (4). Make sure they are placed properly before reinserting the circuit into the plastic casing.
• Reinsert the internal circuit complete with the new electrode into the plastic casing.
• Fasten the screw on the back, below the battery compartment.
• Reinsert the battery compartment cover.
• Recalibrate the new measurements.