Hanna Instruments HI 98201 User Manual

SPECIFICATIONS Range ±999 mV Resolution 1 mV Accuracy (@20°C/68°F) ±5 mV Typical EMC Deviation ±5 mV Environment 0 to 50°C (32 to 122°F);
RH max 95%
Battery Type 4 x 1.5V alkaline Battery Life approx. 700 hours of use Dimensions 175 x 41 x 23 mm
(6.9 x 1.6 x 0.9")
Weight 95 g (3.4 oz.)
ACCESSORIES HI 7020M 200/275 mV test solution, 230
mL bottle
HI 7091M Reducing solution, 230 mL bottle HI 7092M Oxidizing solution, 230 mL bottle HI 70300M Storage solution, 230 mL HI 7061M Cleaning solution, 230 mL bottle HI 73201 Spare electrode
ORP
HI 98201
Pocket-sized ORP Meter
WARRANTY
These meters are guaranteed for one year against defects in workmanship and materials when used for their intended purpose and maintained according to instructions. Probes are guaranteed for six months. This warranty is limited to repair or replacement free of charge. Damages due to acci­dent, misuse, tampering or lack of prescribed maintenance are not covered. If service is required, contact the dealer from whom you purchased the instrument. If under war­ranty, report the model number, date of purchase, serial number and the nature of the failure. If the repair is not covered by the warranty, you will be notified of the charges incurred. If the instrument is to be returned to Hanna Instru­ments, first obtain a Returned Goods Authorization Number from the Customer Service department and then send it with shipment costs prepaid. When shipping any instru­ment, make sure it is properly packaged for complete protection.
The ORP is in compliance with the CE directives.
IST98201 R5 05/05
www.hannainst.com
OPERATION
• Remove the protective cap and turn ORP on with by sliding the ON/ OFF switch on the top of the meter.
• Immerse it into the solution to be tested, without exceeding the maxi­mum immersion level.
• Stir gently and wait for stable reading.
always at least 2 mm -1/8'' over the reference compartment). Repeat the maintenance and conditioning procedure. The cloth junction can be pulled out approxi­mately 20 times. After that, the electrode should be replaced.
ELECTRODE REPLACEMENT
For replacing the electrode, contact your Dealer or the nearest Hanna Service Center.
• After use, rinse the electrode with tap water and replace the protective cap.
MAINTENANCE
• Check the meter by im­mersing it in HI 7020 test solution. The reading should be between 200 and 275 mV at 20ºC (68ºF). Otherwise clean the elec­trode platinum tip by rub­bing it with a lint-free cloth soaked with HI 7061 cleaning solution.
• For more accurate readings, condition the meter by immersing it in a pretreatment reducing (HI 7091) or oxidizing (HI 7092) solution for about 30 minutes before taking measure­ments.
• In case of erroneous readings even after an accurate conditioning and calibration, the ref­erence junction might be contaminated or clogged.
BEPS & BATTERY REPLACEMENT
The meter is provided with BEPS (Battery Error Prevention System) which avoid any erroneous reading due to low battery level, by automati­cally switching the meter off. To replace the batteries, slide off the battery compartment cover and replace all four 1.5V alkaline batteries, while paying attention to the correct polarity.
Batteries should only be replaced in a non­hazardous area and using the battery type specified in this instruction manual.
• Pull out 2 mm (1/8'') of the cloth junction to re­new the electrode refer­ence (it is recommended to cut the cloth leaving