SPECIFICATIONS
Range ±999 mV
Resolution 1 mV
Accuracy (@20°C/68°F) ±5 mV
Typical EMC Deviation ±5 mV
Environment 0 to 50°C (32 to 122°F);
RH max 95%
Battery Type 4 x 1.5V alkaline
Battery Life approx. 700 hours of use
Dimensions 175 x 41 x 23 mm
(6.9 x 1.6 x 0.9")
Weight 95 g (3.4 oz.)
ACCESSORIES
HI 7020M 200/275 mV test solution, 230
mL bottle
HI 7091M Reducing solution, 230 mL bottle
HI 7092M Oxidizing solution, 230 mL bottle
HI 70300M Storage solution, 230 mL
HI 7061M Cleaning solution, 230 mL bottle
HI 73201 Spare electrode
ORP
HI 98201
Pocket-sized ORP Meter
WARRANTY
These meters are guaranteed for one year against defects
in workmanship and materials when used for their intended
purpose and maintained according to instructions. Probes
are guaranteed for six months. This warranty is limited to
repair or replacement free of charge. Damages due to accident, misuse, tampering or lack of prescribed maintenance
are not covered. If service is required, contact the dealer
from whom you purchased the instrument. If under warranty, report the model number, date of purchase, serial
number and the nature of the failure. If the repair is not
covered by the warranty, you will be notified of the charges
incurred. If the instrument is to be returned to Hanna Instruments, first obtain a Returned Goods Authorization Number
from the Customer Service department and then send it
with shipment costs prepaid. When shipping any instrument, make sure it is properly packaged for complete
protection.
The ORP is in compliance with the CE directives.
IST98201 R5 05/05
www.hannainst.com
OPERATION
• Remove the protective cap and turn
ORP on with by sliding the ON/
OFF switch on the top of the meter.
• Immerse it into the solution to be
tested, without exceeding the maximum immersion level.
• Stir gently and wait for stable reading.
always at least 2 mm -1/8'' over the reference
compartment). Repeat the maintenance and
conditioning procedure.
The cloth junction can be pulled out approximately 20 times. After that, the electrode
should be replaced.
ELECTRODE REPLACEMENT
For replacing the electrode, contact your Dealer or
the nearest Hanna Service Center.
• After use, rinse the electrode with tap water
and replace the protective cap.
MAINTENANCE
• Check the meter by immersing it in HI 7020 test
solution. The reading
should be between 200
and 275 mV at 20ºC
(68ºF).
Otherwise clean the electrode platinum tip by rubbing it with a lint-free
cloth soaked with HI 7061
cleaning solution.
• For more accurate readings, condition the meter
by immersing it in a pretreatment reducing
(HI 7091) or oxidizing (HI 7092) solution for
about 30 minutes before taking measurements.
• In case of erroneous readings even after an
accurate conditioning and calibration, the reference junction might be contaminated or
clogged.
BEPS & BATTERY REPLACEMENT
The meter is provided with BEPS (Battery Error
Prevention System) which avoid any erroneous
reading due to low battery level, by automatically switching the meter off.
To replace the batteries, slide off the battery
compartment cover and replace all four 1.5V
alkaline batteries, while paying attention to the
correct polarity.
Batteries should only be replaced in a nonhazardous area and using the battery type
specified in this instruction manual.
• Pull out 2 mm (1/8'') of
the cloth junction to renew the electrode reference (it is recommended
to cut the cloth leaving