Hanna Instruments HI 98201 User Manual

ORP
HI 98201
Pocket-sized Redox Meter
OPERATION:
• Remove the protective cap.
• Turn the meter on with the ON/OFF switch located on the top of the meter.
• Immerse it in the solution to be tested without exceeding the maxi­mum immersion level.
• After use, rinse the electrode with tap water to minimize contamination.
• Always replace the protective cap after use.
MAINTENANCE:
dure. The cloth junction can be pulled out approximately 20 times. After that, the electrode can be replaced (see the Electrode Replace-
ment section).
BEPS & BATTERY REPLACEMENT:
The meter is supplied with BEPS (Battery Error Preventing System) which avoid any erroneous read­ing due to low battery level. When the batteries are too low, the meter automatically switches itself off. To replace the batteries, slide off the battery com­partment cover and replace all four 1.5V batteries while paying attention to their polarity.
Manufacturers since 1978
• Check the meter by immers­ing it in HI7020 ORP solu­tion. The reading should be between 200 and 275 mV at 20ºC (68ºF). Otherwise clean the plati­num tip of the electrode by rubbing it with a soft cot­ton or lint-free cloth soaked with HI7061 cleaning solu­tion.
• For more accurate readings, condition the meter by immersing it in a pretreatment reducing (HI7091) or oxidizing (HI7092) solution for half an our before taking measurements.
• In case of erroneous readings even after an accurate conditioning and calibration, the refer­ence junction might be contaminated or clogged.
• Pull out 2 mm (1/8'') of the cloth junction to renew the electrode reference (it is recommended to cut the cloth leaving always at least 2 mm ­1/8'' over the reference compartment) and re­peat the maintenance and conditioning proce-
Batteries should only be replaced in a nonhazardous area using the battery type specified in this instruction
manual.
ONE YEAR WARRANTY & SERVICEABLE:ONE YEAR WARRANTY & SERVICEABLE:
ONE YEAR WARRANTY & SERVICEABLE:
ONE YEAR WARRANTY & SERVICEABLE:ONE YEAR WARRANTY & SERVICEABLE:
Covered by manship and materials, this tester is now completely serviceable. Contact your dealer for further information.
The
one yearone year
one year warranty against defects in work-
one yearone year
ORP ORP
ORP is in compliance with the CE directives.
ORP ORP
ORP
HI 98201
Pocket-sized Redox Meter
SPECIFICATIONS:
RANGE -999 to +999 mV RESOLUTION 1 mV ACCURACY TYPICAL EMC DEVIATION ±5 mV ENVIRONMENT BATTERY TYPE 4 x 1.5V alkaline (included) BATTERY LIFE approx. 700 hours
DIMENSIONS 175 x 41 x 23 mm
WEIGHT 78 g (2.7 oz.)
(@20°C/68°F) ±5 mV
0 to 50C (32 to 122F); 95% RH
of continuous use
(7.9 x 1.8 x 1")
ELECTRODE REPLACEMENT:
The electrode can be easily replaced in the following way:
• Slide off the battery cover.
• Remove the screw on the back of the below the battery compartment and slide off the whole internal part of the tester.
ORPORP
ORP located
ORPORP
Internal circuit and electrode:
ACCESSORIES:
HI 7020M ORP 200/275 mV solution
(230 mL bottle)
HI 7091M Reducing solution (230 mL bottle) HI 7092M Oxidizing solution (230 mL bottle) HI 70300M Storage solution (230mL) HI 7061M Electrode cleaning solution
(230mL bottle)
HI 73201 Spare electrode
Visit our Internet Home Page:
http://www.hannainst.com
• To remove the electrode proceed as follows: Remove the fastening screw on the front (1) to
loose the electrode. The internal circuit is con­nected to the electrode through two wires (one for the glass sensor and one for the reference). Remove the two small screws (2 & 3) which fasten the four connecting wires to their sock­ets.
• Remove the electrode and replace it with a new
HI73201.
• Connect the new electrode as follows:
first fasten the screw on the front (1) to attach the elec­trode to the circuit. Insert the two small wires into their sockets and fasten the
two screws (2 & 3).
HI73201
• The new electrode is supplied with four new O-rings (4). Make sure they are placed properly (as shown on the above drawing) before rein­serting the circuit into the plastic casing.
• Reinsert the internal circuit with the new electrode into the plastic casing.
• Fasten the screw on the back below the battery com­partment.
• Reinsert the battery com­partment cover.
• Recondition the ORP be­fore using it