Hanna Instruments HI 98198/20, HI 98198/10 User manual

Page 1
HI98198
INSTRUCTION MANUAL
opdo
Optical dissolved oxygen meter
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 2
Dear Customer,
Thank you for choosing a Hanna Instruments product. Please read this instruction manual carefully before using the meter. This manual will provide you with the of the meter, as well as a precise idea of its versatility. If you need additional technical information, do not hesitate to e‑mail us at or view our worldwide contact list at
necessary information for correct use
All rights are reserved. Reproduction in whole or in part is prohibited without the written consent of the copyright owner, Hanna Instruments Inc.,
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 3
PRELIMINARY EXAMINATION ..............................................................................................
GENERAL DESCRIPTION ......................................................................................................
FUNCTIONAL DESCRIPTION ................................................................................................
SPECIFICATIONS ...............................................................................................................
THEORY OF OPERATION .....................................................................................................
PROBE COMPONENTS ........................................................................................................
GETTING STARTED .............................................................................................................
SETUP ..............................................................................................................................
DO CALIBRATION ............................................................................................
PRESSURE CALIBRATION.....................................................................................................
TEMPERATURE CALIBRATION...............................................................................................
GOOD LABORATORY PRACTICE (GLP) ....................................................................................
MEASUREMENT..................................................................................................................
LOGGING...........................................................................................................................
RETRIEVING DATA..............................................................................................................
OPERATIONAL GUIDE....................................................................................................
BOD MEASUREMENT..........................................................................................................
OUR MEASUREMENT .........................................................................................................
SOUR MEASUREMENT .......................................................................................................
BATTERIES REPLACEMENT...................................................................................................
PROBE MAINTENANCE........................................................................................................
TROUBLESHOOTING GUIDE.................................................................................................
ACCESSORIES.....................................................................................................................
CERTIFICATION.....................................................................................................................
RECOMMENDATIONS FOR USERS............................................................................................
WARRANTY.....................................................................................................................
TABLE OF CONTENTS
4
5
6
8
9
10
11
13
27
32
33
35
36
37
39
41
42
46
48
52
53
55
56
58
59
59
3
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 4
Remove HI98198 opdo™ meter from the packing material and examine it carefully to make sure that no damage has occurred during shipment. If noticeable damage is evident, contact your local Hanna Instruments office.
Each meter is supplied with:
HI764113 Optical DO probe with built‑in temperature sensor, protective shield and 4 m (13’4) cable
HI764113‑1 Smart Cap™ with o‑ring
HI7040 Bicomponent Zero Oxygen Solution
• Calibration/storage vessel
• 100 mL plastic beaker (2 pcs.)
• 1 syringe with silicon grease
• 1 lens wipe
PRELIMINARY EXAMINATION
• 1.5V AA batteries (4 pcs.)
• Instruction manual
• Meter quality certificate
• Probe quality certificate
• Cap quality certificate
• USB Type A to C cable
Note: Save all packing material until you are sure that the meter functions correctly. All defective items must be returned in the original packing with the supplied accessories.
4
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 5
The HI98198 opdo™ meter is a rugged, portable dedicated dissolved oxygen (DO) meter designed for fresh and saltwater measurements of dissolved oxygen. This professional, waterproof meter com‑ plies with IP67 standards and measures DO, barometric pressure, and temperature. The HI98198 is supplied with a HI764113 digital optical dissolved oxygen probe in a custom thermoformed durable carrying case with accessories. It is compact and ergonomically designed to provide ready access to the materials required for routine sampling.
The HI98198 opdo™ meter is only compatible with the Hanna digital dissolved oxygen probe (HI764113).
Concentration measurements are automatically compensated for barometric pressure, temperature and salinity. Barometric pressure and temperature are automatically measured and compensated. Salinity is automatically compensated by setting manually the salinity concentration of the water being measured. The meter also has a built in application to measure and calculate BOD (Biological Oxygen Demand), OUR (Oxygen uptake rate), and SOUR (Specific Oxygen Update Rate).
Other features include:
• Text messages on the graphic LCD display to warn and guide the user.
• Back lighted display.
• One or two points calibration at 0 or/and 100% saturation (with auto recognition).
• A single point manual calibration in mg/L or % saturation using a reference method for calibration value.
• A dedicated HELP key for assistance anytime.
• A user selectable “Calibration due” warning.
• A dedicated GLP key that includes at last 5 calibrations with time, date, calibration points as well
as barometric pressure, temperature and salinity setting.
• AutoEnd freezes the next stable measurement value on the display.
• Log on demand with 4000 records capability.
• USB‑C port for easy data transfer to memory stick, PC or other compatible device.
GENERAL DESCRIPTION
5
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 6
FRONT VIEW TOP VIEW
FUNCTIONAL DESCRIPTION
1) Liquid Crystal Display (LCD).
2) F1, F2, F3 functional keys.
3) / keys to manually increase/decrease the parameters or to scroll through the menu.
4) ON/OFF ( ) key, to turn the meter ON and OFF.
5) LIGHT ( ) key, to toggle display back‑lighting.
6) GLP key, to display Good Laboratory Practice information.
7) CAL key, to enter/exit calibration menu.
8) SETUP key, to enter/exit Setup menu.
9) RCL key, to enter/exit view logged data mode (RCL means RECALL).
10) MODE key to switch between DO, BOD, OUR and SOUR applications.
11) RANGE key, to change between % saturation or concentration in DO measurement mode.
12) HELP key to enter/exit contextual help.
13) ESC key to leave current mode, exit calibration, setup, help, etc.
14) Quick connect DIN connector.
15) USB-C connector.
16) Junction for barometric pressure sensor.
6
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 7
HI98198 Meter with HI764113 Probe Specifications
Range 0.00 to 50.00 mg/L / 0.0 to 500.0% saturation
Dissolved Oxygen
Barometric Pressure
Temperature
Resolution 0.01 mg/L / 0.1% saturation
1.5% of reading ± 0.01mg/L for 0.00‑20.00mg/L
Accuracy
Range 420 to 850 mmHg
Resolution 1 mmHg
Accuracy ±3 mmHg within ±15% from the calibration point
Range
Resolution 0.1 °C (0.1 °F)
Accuracy ±0.3 °C (±0.4 °F)
5% of reading for 20.00‑50.00mg/L
1.5% of reading ±0.1% for 00‑200.0%
SPECIFICATIONS
5% of reading for 200.0‑500.0%
‑5.0 to 50.0 °C (23 to 122 °F)
DO Calibration
Temperature Calibration Single point anywhere within temperature range
Pressure Calibration Single point anywhere within pressure range
Temperature
Compensation
Pressure Compensation Automatic from 420 to 850 mmHg
Salinity Compensation Automatic from 0 to 70 PSU (manually set)
DO Probe HI764113 Optical Probe
LOG On demand with 4000 records capability
Battery Type / Life
Auto Power Off User selectable: 5, 10, 30, 60 minutes or disabled
PC Connectivity USB-C
Dimensions 185 x 93 x 35.2 mm (7.3 x 3.6 x 1.4”)
Meter Weight
(with batteries)
Case Ingress
Protection Rating
Environment 0 to 50 °C (32 to 122 °F) max. RH 100%
• One or two points automatic calibration at 100% (8.26 mg/L) and 0% (0 mg/L).
• Single point manual using a value entered by the user in % saturation or mg/L.
Automatic from ‑5.0 to 50.0 °C (23.0 to 122.0 °F)
1.5V AA batteries (4 pcs.) / approx. 200 hours of continuous use
without backlight (50 hours with backlight)
450 g (15.9 oz)
IP67
7
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 8
HI764113 Probe Specifications
Probe body material ABS
Smart Cap™ material Polypropylene
Cable jacket material PVC
Cable length 4 m (13.1 ft.), 10 m (32.8 ft.), and 20 m (65.6 ft.) options
SPECIFICATIONS
Probe guard 316 Stainless Steel Temperature
Measurement
Pressure 20 m (29 PSI)
Probe Dimensions
(with Guard)
Response Time (t95) 45 seconds
Probe Weight
(with Guard)
Probe Ingress
Protection Rating
Sensor type Optical; Luminescence Quenching
Origin Assembled in USA
Thermistor
174 X 25 mm (6.8 X 1”)
400 g (14.2 oz)
4 m (13.1 ft.) cable length
IP68
8
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 9
THEORY OF OPERATION
The Hanna HI764113 optical DO sensing probe is based on the principle of fluorescence quenching. The sensing method features an immobilized Pt based luminophore that is excited by the light of a blue LED and emits a red light. Dissolved oxygen quenches this excitation. When there is no oxygen present, the lifetime of the signal is the greatest; as oxygen hits the sensing surface, the lifetime becomes shorter. The intensity and lifetime are inversely proportional to the amount of oxygen present; as oxygen interacts with the luminophore it reduces the intensity and lifetime of the luminescence. The lifetime of the luminescence is measured by a photodetector, and is used to calculate the dissolved oxygen concentration. This is in turn reported by the meter as a % saturation or mg/L reading of Dissolved Oxygen.
The major components of the probe include a blue LED for excitation, a red LED that is used as a reference light, and a photodetector. The Smart Cap™ is locked in place on the optical probe and includes the immobilized O
Over time, the sensor’s optical components can age but are compensated for by using the reference signal to compensate the measuring path. As a result, the sensor provides accurate DO measurements over long periods of time without the need for frequent calibration.
sensitive luminophore with rugged insoluble black oxygen permeable protective layer.
2
THEORY OF OPERATION
9
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 10
PROBE COMPONENTS
1. Strain relief
2. ABS Probe body
3. Temperature Sensor
4. O‑Ring Seal
5. Optical window
6. Alignment key
7. Smart Cap™
8. RFID Tag
9. Embedded O
10. Protective shield
sensitive luminophore with black protective layer
2
10
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 11
INITIAL PREPARATION
Install the supplied batteries into the meter. See Batteries Replacement for details, page 52. To prepare the meter for field measurements close the USB communication socket with the attached
stopper. Turn the meter ON by pressing ON/OFF key. At start‑up the meter will show the Hanna Instruments logo for a few seconds, followed by the
percentage indication of the remaining battery life. Before connecting the probe for the first time, press SETUP and using
Date/Time. Press Modify and set the current date and time.
Note: Verify time and date are properly set on meter prior to probe initialization.
1. Remove the HI764113 from the carrying case. Remove stainless steel guard from probe body if it was installed. Use care not to leave fingerprints on the optical window.
2. Sparingly lubricate the O‑ring with a thin film of silicone grease. Use care to prevent grease from contacting the optical window.
3. Remove the HI764113‑1 optical cap from the container. Align the notched cutout arrow on the Smart Cap™ with the matching guide on the probe body.
/ arrows navigate to
GETTING STARTED
4. Slide and press the Smart Cap™ onto the HI764113 body until the cap snaps in place. Once the cap is installed, it should not be removed unless a new cap is required.
5. Connect the HI764113 probe to the HI98198 meter by plugging the DIN connector to the socket located on the top of the meter.
6. Power the meter to initiate the cap timer.
Note: Turn off the meter before connecting or disconnecting
the probe.
11
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 12
A probe info screen will be displayed momentarily before defaulting to the measurement screen:
No Probe” message is displayed on meter if the probe connector is not plugged in properly.
GETTING STARTED
The “No cap info detected. Install the cap and press Continue.“ message on the meter indicates that the Smart Cap™ is missing or not properly installed on the probe.
No conditioning period is required when using a HI764113 DO probe. Store the HI764113 probe in the HI98198 carrying case when not in use. For frequent use and short term storage it is recommended to remove the stainless steel guard and
replace with the storage vessel that has a small amount of deionized water. The probe can also be stored with the stainless steel guard on in a beaker containing deionized water.
For longer term storage, remove the stainless steel guard and replace with the storage vessel.
12
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 13
Setup menu allows viewing and modifying the measurement parameters. The following table lists the Setup parameters, their valid range and the factory default settings.
Item Description Valid value Default
Log Separator Log file column separator Comma, Semicolon Comma Salinity The solution’s salt content 0 to 70 PSU 0 PSU Manual Pressure Used to manually set the pressure
and disable the automatic barometer measurement
Pressure Unit mmHg, inHg, atm,
Temperature Unit °C or °F ° C Calibration Timeout Calibration due timer Disabled, 10 to 70 days Disabled
BOD Configuration
Sample min delta DO
Sample min end DO Seed min delta DO
Seed min end DO
OUR Configuration
Min time Max time Min start DO
Min end DO
Total volume
Sample volume
SOUR configuration
Min time Max time Min start DO
The minimum diff. between the start and the end DO val. The minimum end DO value The minimum diff. between the start and the end DO val. The minimum end DO value
The minimum time for the OUR test The maximum time for the OUR test The minimum DO value for starting the OUR test The minimum DO value at the end of the OUR test The total volume of the solution to be tested The volume of sample in the solution to be tested
The minimum time for the SOUR test The maximum time for the SOUR test The minimum DO value for starting the SOUR test
Enabled or disabled Disabled
mmHg
mbar, psi, kPa
0.00 to 50.00 mg/L
0.00 to 50.00 mg/L
0.00 to 50.00 mg/L
0.00 to 50.00 mg/L
1 to 3600s 1 to 3600s
0.01 to 50.00 mg/L
0.00 to 50.00 mg/L
0.1 to 300.0 mL
0.1 to 300.0 mL
1 to 3600s 1 to 3600s
0.01 to 50.00 mg/L
0.00 mg/L
0.00 mg/L
0.00 mg/L
0.00 mg/L
3600s
0.01 mg/L
0.00 mg/L
0.1 mL
0.1 mL
3600s
0.01 mg/L
SETUP
1s
1s
13
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 14
SETUP
14
Item Description Valid value Default
Min end DO
Total volume
Sample volume
Solids weight
SOUR @ 20 °C Autodelete BOD
start data Backlight Backlight Level 1 to 7 4 Contrast Contrast Level 0 to 20 10 Auto Light OFF Time backlight remains ON 1, 5, 10, 30 minutes 1 Auto Power OFF Time until meter is powered OFF Disabled or 5, 10, 30, 60
Date / Time 01.01.2006 to
Time Format AM/PM or 24 hours 24 hours Date Format DD / MM / YYYY
Language Message display language Up to 3 languages English Beep ON Beeper Status Enabled or Disabled Disabled Instrument ID Meter identification 0000 to 9999 0000 Meter Info Displays general information Probe Info Displays probe and cap information
The minimum DO value at the end of the test The total volume of the solution to be tested The volume of sample in the solution to be tested Total solids or Volatile Suspended solids weight Correct the SOUR value to 20 °C
Automatically delete BOD start data, after BOD compute
0.00 to 50.00 mg/L
0.1 to 300.0 mL
0.1 to 300.0 mL
0.1 to 300.0 g/L
Enabled or Disabled Enabled or Disabled Disabled
minutes
12.31.2099 00 :00 to 23 :59
MM / DD / YYYY YYYY / MM / DD YYYY / MM / DD YYYY ‑ MM ‑ DD Mon DD, YYYY DD ‑ Mon ‑ YYYY YYYY‑Mon‑DD
0.00 mg/L
0.1 mL
0.1 mL
0.1 g/L
Disabled
30
current
date/ time
YYYY/MM/
DD
SETUP
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 15
PARAMETER SCREENS
SETUP
SETUP
Log Separator
Log separator or CSV file separator is a special character used to separate columns in the CSV log file. There are two options available: Comma (,) or Semicolon (;). The field separator depends upon regional preferences.
Press the function key to modify the CSV file separator.
Salinity
Dissolved oxygen solubility decreases if water contains salts. This parameter is used to compensate concentration measurements (mg/L) made in brackish or sea water samples. The salinity is expressed in PSU scale and is manually entered by the user. The DO concentration will be automatically compensated to improve the accuracy during calibration and measurement. Enter the salinity value to be close to the known salt level of the sample. Seawater typically has a salinity of 35 and the oxygen solubility is 18% less as compared to fresh water at 25 °C. By entering the approximate salinity value, the calibration and subsequent concentration measurement will be compensated to display the correct oxygen concentration. An 18% error would result if the salinity value is not entered.
SETUP
Highlight Salinity. Press Modify.
/ keys to change the salinity value. Press Accept to confirm or ESC to exit without saving.
Use
15
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 16
Manual Pressure
The HI98198 has a built in barometer for automatic compensation of ambient pressure for oxygen
SETUP
measurement. The user may enable Manual Pressure to disregard the barometer measurement and manually enter a pressure value which will be used for oxygen measurements. Once enabled, pressure values are entered from the measurement screen using the
Highlight Manual pressure. Press the displayed functional key to enable or disable the feature.
Pressure Unit
The HI98198 has the ability to convert and display pressure measurements in user selected units. Automatic and Manual Pressure will utilize these units.
To change the pressure measurement unit highlight Pressure unit and press Modify.
/ keys.
SETUP
Using / keys highlight the desired pressure measurement unit.
Press Accept to confirm or ESC to exit without saving.
16
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 17
Temperature Unit
The HI98198 has the ability to convert and display temperature measurements in degree Celsius or
SETUP
Fahrenheit.
Highlight Temperature Unit. To change the temperature unit, press the displayed functional key.
Calibration Timeout
The HI764113 dissolved oxygen probe utilizes optical technology which offers significant advantages over polarographic or galvanic type dissolved oxygen probes. Some advantages include reduced maintenance, simplified handling and infrequent calibrations. However if a user wishes to validate or calibrate on a SOP schedule, a timer may be set as a reminder.
The Calibration time out parameter can be set from 10 to 70 days or can be disabled. The meter is provided with a real time clock (RTC) and is used to monitor the time elapsed since the
last DO calibration. The calibration timer is reset every time the meter is calibrated and the “CAL DUE” status is triggered
when the meter detects a calibration time out. The “CAL DUE” tags will start blinking to warn the user that the meter should be recalibrated.
If the calibration timeout is changed (e.g. to 20 days), then the timer will be immediately reset.
Notes:
- Before the DO calibration is cleared (default values loaded) the display always shows the “CAL DUE”.
Before an abnormal condition in the RTC is detected, the meter forces the “CAL DUE”
­status.
- After a user temperature or pressure calibration is made (or cleared), the “CAL DUE” message will be activated.
SETUP
17
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 18
Press SETUP key.
SETUP
Highlight Calibration timeout.
Press Modify. Use the / keys to set the desired value. Press Accept to confirm or ESC to return without saving.
Note: If enabled, a ”CAL DUE” warning will be shown on the display after the Calibration timeout period has expired.
SETUP
BOD Configuration
When making BOD measurements, the BOD method configuration parameters must be filled in. These parameters will be used to flag error limits for the determination. Ignore this parameter if not making BOD measurements.
Highlight BOD configuration. Press Select.
18
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 19
Press / keys to modify the selected parameter’s value. Press Prev/Next to select a different parameter.
SETUP
Press Save to save the new BOD configuration. Press ESC to leave without changing.
Parameters:
Sample min DO ‑ the minimum acceptable difference between the initial and final DO values
for a sample. If the difference is less than this value the meter will show a warning message when evaluating the BOD.
Range: 0.00 to 50.00 mg/L.
Sample min end DO ‑ the minimum acceptable final DO value for a sample. If the final DO value
is less than this value the meter will show a warning message when evaluating the BOD.
Range: 0.00 to 50.00 mg/L.
Seed min DO ‑ the minimum acceptable difference between the initial and final DO values
for a seed sample. If the difference is less than this value the meter will show a warning message when evaluating the BOD.
Range: 0.00 to 50.00 mg/L.
Seed min end DO ‑ the minimum acceptable final DO value is less than this value the meter will
show a warning message when evaluating the BOD.
Range: 0.00 to 50.00 mg/L.
OUR Configuration
The OUR method configuration allows the user to set the parameters related to the OUR measurement. OUR is used to determine the oxygen consumption or respiration rate. It is defined as the mg/L of
oxygen consumed per hour. These parameters will be used to flag error limits and make dilution calculations for the determination. Ignore this parameter if not making OUR measurements. Highlight OUR configuration and press Select.
SETUP
19
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 20
Press / keys to modify the selected parameter’s value. Press Prev/Next to select a different parameter. Press Save to save the new OUR configuration.
SETUP
Press ESC to leave without changing.
Parameters:
Min time ‑ the minimum time for the OUR test. Range: 1 to 3600 seconds.
Max time ‑ the maximum time for the OUR test. The test will stop automatically when the maximum
time has elapsed.
Range: 1 to 3600 seconds.
Min start DO ‑ the minimum accepted DO value for starting the OUR test. If the DO reading is
less than this value the test cannot be started.
Range: 0.01 to 50.00 mg/L.
Min end DO ‑ the minimum accepted DO value at the end of the test. If the DO reading at the
end of the OUR test is less than this value a warning message will be displayed.
Range: 0.00 to 50.00 mg/L.
Total volume ‑ the volume of the diluted mixture. Range: 0.1 to 300.0 mL
Sample volume ‑ the volume of sample in the diluted mixture. Range: 0.1 to 300.0 mL.
SETUP
20
SOUR Configuration
Specific Oxygen Uptake Rate (SOUR), also known as the oxygen consumption or respiration rate, is defined as the milligram of oxygen consumed per gram of volatile suspended solids (VSS) per hour. These parameters will be used to flag error limits and make dilution calculations for the determination. Ignore this parameter if not making SOUR measurements. Highlight SOUR configuration and press Select.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 21
Press
/ keys to modify the selected parameter’s value.
Press Prev/Next to select a different parameter.
SETUP
Press Save to save the new SOUR configuration. Press ESC to leave without changing.
Min time ‑ the minimum time for the SOUR test. Range: 1 to 3600 seconds.
Max time ‑ the maximum time for the SOUR test. The test will stop automatically when the
maximum time has elapsed.
Range: 1 to 3600 seconds.
Min start DO ‑ the minimum accepted DO value for starting the SOUR test. If the DO reading is
less than this value the test cannot be started.
Range: 0.01 to 50.00 mg/L.
Min end DO ‑ the minimum accepted DO value at the end of the test. If the DO reading at the
end of the SOUR test is less than this value a warning message will be displayed.
Range: 0.00 to 50.00 mg/L.
Total volume ‑ the volume of the diluted mixture. Range: 0.1 to 300.0 mL
Sample volume ‑ the volume of sample in the diluted mixture. Range: 0.1 to 300.0 mL.
Solids weight: Total solids or Volatile suspended solids weight. Range: 0.1 to 300.0 g/L.
SOUR @ 20°C: If this option to enabled the SOUR value is corrected to 20 °C.
SETUP
Autodelete BOD start data
Highlight Autodelete BOD start data.
Press the displayed functional key to enable/disable the feature. If enabled the BOD initial data record used in BOD result evaluation is deleted automatically after the
BOD result has been saved into the meter’s memory (LOG key pressed). If disabled, the user has to delete BOD initial data records that were used in BOD result evaluation,
entering View initial BOD data mode.
21
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 22
Backlight
The display backlight can be adjusted to increase visibility in various lighting environments.
SETUP
To adjust the backlight, highlight Backlight and press Modify.
Use / keys to change intensity, then press Accept to confirm.
Press ESC to leave without changing.
Contrast
The display contrast ratio adjusts the ratio of light areas to dark areas to improve readability in various environments.
To adjust the contrast, highlight Contrast and press Modify.
SETUP
Use / keys to change intensity, then press Accept to confirm.
Press ESC to leave without changing.
22
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 23
Auto Light OFF
The backlight must be manually enabled by pressing the light bulb key on the keypad. When the
SETUP
selected time has expired, back lighting turns off automatically. To change the duration, highlight Auto Light OFF than press the desired functional key to change
the option. The backlight display setting should be set to the shortest time to preserve the battery life.
Auto Power OFF
This selection can be used to preserve the battery life in the event that the meter is accidentally left on.
Highlight Auto Power OFF. Press Modify.
SETUP
Use / keys to select interval, then press Accept.
Press ESC to leave without changing.
23
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 24
Date / Time
This parameter should be configured to the current date and time prior to connecting a HI764113
SETUP
probe for the first time. To set the time and / or date, from the Setup menu, highlight Date / Time, then press Modify.
SETUP
Use /keys to select item. Use
Press Accept to confirm the new settings, or ESC to leave without changing.
Time Format
Your desired time format may be selected in this parameter. Highlight Time Format from Setup menu. Select between AM/PM or 24 hours.
Press functional key to change the option.
Date Format
Your desired date format may be selected in this parameter. Highlight Date Format from Setup menu and press Modify.
/ keys to change the emphasized values.
Highlight desired date format using Press ESC to leave without changing.
24
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
/ keys then press Accept.
Page 25
Language
This option allows the user to choose the desired language in which all information will be displayed.
SETUP
To modify the language, highlight Language from the Setup menu and press the desired virtual key to make selection.
Beep On
This option allows the user to turn an acoustic warning signal on or off. If enabled, a short tone will occur to signal a condition is correct (key press, calibration) or a long tone will occur when an incorrect key is pressed.
Disabling will silence audible signals.
Instrument ID
This parameter allows the user to set a four digit code to easily identify the meter.
SETUP
Press Modify, then use without saving.
Find Quality Products Online at: sales@GlobalTestSupply.com
/ keys to change the meter’s ID. Press Accept to confirm or ESC to exit
25
www.GlobalTestSupply.com
Page 26
Meter Info
This parameter captures the firmware version, language version, pressure calibration date, meter serial
SETUP
number as well as the battery capacity. Highlight the Meter Info then press Select.
Probe Info
This parameter displays probe and Smart Cap™ information. Press Select to view cap information.
Press Next to view the Cap information.
SETUP
Press Prev to return to first screen. Press ESC to return to Setup. Press ESC two times to return to Measurement.
26
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 27
There are three calibration functions and procedures that can be accessed through the CAL button. These are calibration of the probe (DO), calibration of the barometer (Pressure) and calibration of the temperature (Temp). Prepare the dissolved oxygen probe according to the instructions from GETTING STARTED (page 11). No conditioning period is required when using a HI764113 DO probe. Store the HI764113 probe in the HI98198 carrying case when not in use. For short term storage the
CALIBRATION
probe can also be stored in a beaker of deionized water or in storage cap (when shield is removed).
DO CALIBRATION
General Calibration Guidelines:
1. Before calibration, rinse the probe with clean water to remove debris from the probe body; wipe with a lint free cloth.
2. Remove the protective shield and set aside.
3. Inspect. Visually inspect the Smart Cap for biofouling. If necessary, use a mild detergent and a soft bristled
toothbrush to clean the probe and Smart Cap. A scratch in the black protective layer on the Smart Cap will affect the calibration (and measurement). Replace the Smart Cap if the sensing surface has been compromised.
4. Rinse the cap with water after cleaning and dry with a laboratory tissue.
5. Discard zero oxygen solution in an appropriate manner after use (do not return to bottle). Follow local disposal regulations.
6. Confirm all water droplets have been removed from the cap surface as well as the temperature element before performing the calibration procedure in water saturated air.
CALIBRATION
Calibration of the HI764113 optical dissolved oxygen probe may be performed several different ways:
Single point automatic zero calibration at 0% saturation or 0 mg/L
Single point automatic slope calibration at 100% saturation or 8.26 mg/L
Two points automatic calibration at 0% saturation (0 mg/L) and 100% saturation (8.26 mg/L)
Single point manual calibration using a standard value set by the user in % saturation or mg/L.
Any 0% (or mg/L) calibration is made exposing the probe to an environment with the absence of oxygen (such as Hanna HI7040 solution), a nitrogen sparging, or other oxygen scavengers. A 100% calibration is best made in water saturated air, however air‑saturated water is also acceptable. A single point manual calibration may be made by comparing the displayed value to a determination made by a reference method (such as the Winkler titration), or a reference probe in the same sample.
Notes: Temperature and Pressure calibration (if required) should be made prior to the probe calibration. Before attempting the calibration, the probe and standards must be prepared. Remove the protective guard from the probe.
27
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 28
For 100% calibration, water‑saturated air is the simplest method (and recommended method). It involves suspending the probe over a contained surface of water or moistened absorbent material. The temperature element should also be inside the container. Under equilibrium conditions, the partial pressure of oxygen in air‑saturated water is equal to the partial pressure of oxygen in water‑saturated air; a probe calibrated in water‑saturated air will correctly read the partial pressure of oxygen in water samples.
CALIBRATION
Note: performing the calibration in dry air will introduce an error because reference compensations are based on air containing 100% relative humidity.
Place a moistened sponge into the bottom of the Hanna storage/calibration vessel, or place deionized water into the bottom of the small bottle or flask, then suspend the probe into the vessel. Warning: DO NOT tighten the calibration vessel on the probe thread as the water vapor will become pressurized. If using water, verify the probe Smart Cap™ has not been moistened. Wait a minimum of 15 minutes for the air to become saturated with water vapor.
Calibrating the probe over a large water surface, such as a lake or aeration basin used in wastewater treatment should also suffice. Air‑saturated water can be produced by bubbling air into a water sample for a prolonged period. The point at which the water is completely saturated is difficult to estimate so there is a risk of under or over saturation (if the temperature changes).
Single point automatic zero calibration
Submerge the probe into HI7040 temperature and probe values to become stable. Press CAL. The calibration menu will be displayed.
zero oxygen solution and stir gently for 2‑3 minutes. Wait for the
CALIBRATION
Press DO to select the DO calibration. The meter will automatically recognize the 0% standard.
28
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 29
The DO calibration screen will be displayed and the standard 0% saturation (or 0 mg/L, depending on the currently selected measuring unit), will be automatically selected. When the reading is stable within the acceptable range, the CFM functional key is displayed. Press CFM to confirm the calibration point.
CALIBRATION
Press ESC to leave calibration or continue for a two point calibration. The probe must be rinsed thoroughly with purified water to remove all traces of the zero calibration solution.
Single point automatic slope calibration at 100%
Suspend the probe into the water saturated air container and wait for the probe and sample to reach thermal equilibrium. Press CAL. The calibration menu will be displayed.
CALIBRATION
Press DO to select the DO calibration. The meter will automatically recognize the 100% saturated standard (or equivalent mg/L).
The DO calibration screen will be displayed and the standard 100% saturation (or 8.26 mg/L, depending on the currently selected measuring unit), will be automatically displayed. When the reading is stable within the acceptable range, the CFM functional key is displayed. Press CFM to confirm the calibration point.
29
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 30
Two points automatic calibration at 0% saturation (0 mg/L) and 100% saturation (8.26 mg/L)
Note: if performing the two point calibration, the 0 point must be done first.
Submerge the probe into HI7040 zero oxygen solution and stir gently for 2‑3 minutes. Press CAL. The calibration menu will be displayed. Wait for the temperature and probe values to
become stable.
CALIBRATION
Press DO to select the DO calibration. The meter will automatically recognize the 0 standard. The DO calibration screen will be displayed and the standard 0% saturation (or 0 mg/L, depending on the currently selected measuring unit), will be automatically selected. When the reading is stable within the acceptable range, the CFM functional key is displayed.
Press CFM to confirm the calibration point. Remove probe from the standard and rinse it off completely in running water to remove all traces of zero standard. Blot any water off the cap with a lint free tissue. Suspend the probe into the water saturated air container and wait for the probe and sample to reach thermal equilibrium. The meter will automatically recognize the 100% saturated standard (or equivalent mg/L). The standard 100% saturation (or 8.26 mg/L), will be automatically displayed.
CALIBRATION
When the reading is stable within the acceptable range, the CFM functional key is displayed. Press CFM to confirm the calibration point. The meter will return to the measurement screen and the probe will store the calibration data.
30
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 31
Single point manual calibration
Determine the dissolved oxygen value of the sample using a reference method such as the Winkler titration. Place the HI764113 probe in a sample, lake, basin, stream etc. that a recent dissolved oxygen determination has been made. The probe should be at thermal equilibrium with the sample. Press CAL. The calibration menu will be displayed.
CALIBRATION
Press DO to select the DO calibration.
Press the Manual functional key.
CALIBRATION
Using the
When the reading is stable, the CFM functional key is displayed. Press CFM to confirm the calibration point. The meter will return to the measurement screen and the probe will store the calibration data.
/ keys, adjust the standard value to the value determined previously.
31
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 32
Pressure calibration
The HI98198 meter incorporates a pressure transducer that accurately tracks local barometric pressure. Oxygen partial pressure is influenced by the local barometric pressure as dictated by Henry's Law. The HI98198 automatically compensates dissolved oxygen measurements for these changes. The HI98198’s pressure transducer is factory calibrated and no user calibration should be necessary. If the pressure reading appears out of tolerance, the pressure reading can be checked and a calibration
CALIBRATION
performed if required. To calibrate pressure follow the instructions below. A reference barometer with at least 1 mmHg resolution is required. Press CAL from any measure mode (DO, BOD, OUR or SOUR). The calibration menu will be displayed.
Press Pressure functional key to select the pressure calibration. The pressure calibration screen will be displayed.
CALIBRATION
32
Using the / keys, enter the true local barometric pressure reading from the reference barometer. DO NOT use the pressure reported by the weather bureau. Weather bureaus correct pressures to sea level. When the reading is stable within range of the entered barometric pressure the CFM functional key is displayed.
Press CFM to confirm the calibration. The meter will return to the measurement screen and will store the calibration data. ”CAL DUE” will be displayed and a probe calibration must follow.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 33
To clear the pressure calibration and restore the factory one, press CAL from any measure mode (DO, BOD, OUR or SOUR).
The calibration menu will be displayed. Press the Pressure functional key and then Clear. The user calibration will be removed and the factory calibration restored. Press ESC key at any time to exit the calibration without changing values.
CALIBRATION
Temperature calibration
The HI764113 has been factory calibrated for temperature. Dissolved oxygen values are based upon temperature compensation so accurate temperature measurements are required. Should a thermistor go out of tolerance, your measurement will be compromised. An additional user temperature calibration of the HI764113 optical dissolved oxygen probe may be performed if desired. The probe has a built‑in temperature sensor and the thermal transfer is made via the stainless steel contact located on the probe body. The contact point must be fully immersed in solution and in thermal equilibrium with the solution measured during calibration. The greater the difference between the temperatures at which the probe was stored and the temperature of the sample, the longer it will take to reach thermal equilibrium. When the probe is connected to the meter and the meter is powered, the temperature is displayed. Observe when temperature is stable before conducting a temperature calibration. A reference thermometer with 0.1 °C (or better) accuracy is suggested.
Note: Calibration data for temperature is stored in the probe memory.
Procedure:
Remove the protective shield from probe. Place HI764113 probe and the reference thermometer into a stirred container of water. Ensure the temperature contact on the probe is submerged in the water. Observe the temperature on display until it stops changing (this may take several minutes). Press CAL from any measure mode (DO, BOD, OUR or SOUR). The calibration menu will be displayed.
CALIBRATION
Press the Temp functional key to select the temperature calibration.
33
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 34
Use the / keys to set the calibration point values to that of the reference thermometer. When available, CFM will appear on the display. Press CFM to complete the calibration.
CALIBRATION
To clear the user temperature calibration at any time, press CAL from any measure mode (DO, BOD, OUR or SOUR). The calibration menu will be displayed. Press the Temp functional key then Clear.
The previous user calibration has been removed and the factory calibration restored. Press ESC button at any time to escape from the calibration without changing values. ”CAL DUE” will be displayed and a probe calibration must follow.
CALIBRATION
34
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 35
GLP is a set of functions that allows storage and retrieval of data regarding the calibration of the DO probe.
All data regarding DO calibration is stored for the user to review when necessary.
Last DO calibration data
Data from the last 5 DO calibrations are stored automatically after a successful calibration. To view the last calibration data, press GLP when the meter is in the DO, BOD, OUR or SOUR measurement mode. The last calibration will appear at the top of the list.
The meter will display GLP data related to the calibration including calibration standards, salinity, pressure and temperature.
to scroll through previous calibration records. A total of five calibrations are stored for reference.
Use
GOOD LABORATORY PRACTICE (GLP)
GOOD LABORATORY PRACTICE (GLP)
35
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 36
MEASUREMENT
Measurements can be made simply and quickly with a calibrated HI764113 and HI98198. The AutoEnd key can freeze data with a single touch, and the measurement units changed by pressing the Range key. Power the opdo™ meter with the probe connected. After initialization, the meter enters in measurement
MEASUREMENT
mode. Place the probe in the sample to be measured and allow it to reach thermal equilibrium with the sample. Direct measurements of dissolved oxygen (as % saturation or mg/L), temperature and atmospheric pressure are shown on the opdo™ meter display.
Range Key
Pressing the RANGE key while in measurement toggles the measurement between a percent satura‑ tion or concentration (mg/L) measurement. The concentration measurement is calculated for oxygen dissolved in water and is based upon air saturated water. If measuring other fluids (not water), the mg/L measurements will be in error. In those cases the % saturated reading should be used.
AutoEnd
AutoEnd allows the user to display a measurement with a hold function. This is useful when you are taking a measurement in a location where it might be difficult to write down the measurement. Pressing AutoEnd will suspend the reading on the display until it is cleared.
To suspend a stable reading on the display press AutoEnd while the meter is in DO measurement. The “Wait” symbol will blink until the reading is stable.
MEASUREMENT
When the reading is stable “Hold” indicator will be displayed.
Press Continue to resume live readings.
36
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 37
Mode Key
Repetitive presses of the MODE key changes the direct measurement into one of three available dissolved oxygen applications: BOD, OUR, and SOUR. See page 41 for BOD, page 46 for OUR, and page 49 for SOUR for application specifics.
LOGGING
This feature allows the user to log DO, BOD,OUR and SOUR measurements. All logged data can be transfered to a PC using the USB Type C port.
Transfer is simple as the computer recognizes the USB connection. A spreadsheet program such as Excel makes integration of the computer with the meter simple.
The maximum number of records is 4000.
LOGGING THE CURRENT DATA
To log the current measurement press the LOG key.
LOGGING
MEASUREMENT LOGGING
The meter will display for a couple of seconds the record number and the amount of free space (in %). If the maximum number of stored measurements is achieved, the “Log space is full” message will
be displayed for a few seconds when LOG key is pressed.
Enter RCL mode and delete records in order to free log space.
37
37
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 38
VIEW LOGGED DATA
Press RCL to retrieve the stored data while in the measurement mode (DO).
LOGGING
The list of records is displayed in the range it was measured (% sat. or mg/L). Use
/ to select the desired record.
Press Delete All to enter Delete All screen. Press CFM to clear all logged records.
Press Delete to enter Delete the selected screen. Press CFM to delete record.
LOGGING
38
Press More to view more information of the selected record. Press Pg Down to see additional information.
To export data to an USB drive or PC, see the next sections.
If there is no logged data, the meter will display “No Records”.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 39
DO recall:
LOGGING
RETRIEVING DATA
METER TO PC
Logged data on opdo™ can be transferred from the meter to a PC by following these simple directions. Suitable operating systems include Windows (XP or newer), OS X or Linux.
1. Connect opdo
2. Turn on opdo™.
3. The meter will display “Connected to HOST”
The PC should detect the USB as a removable drive. Open the drive to view the stored files. Log files are formatted as Comma separated values (*.CSV) and can be opened with any text editor or spreadsheet program.
(Note: Field separator may be set as comma or semicolon depending upon region preferences, see SETUP.)
Note:
1. Western Europe (ISO-88859-1) character set and English language are suggested settings.
2. Other files may be visible depending upon computer settings. All files stored will appear in this folder.
3. Adjust Font or column width appropriately.
to the PC using the supplied USB‑A to USB‑C cable.
RETRIEVING DATA
39
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 40
METER TO USB DRIVE
The logged data can be transfered from the meter to a USB Flash Drive. In order to transfer all the logged data onto a USB Flash Drive insert USB‑C into opening on the top of meter. Press the RCL key and using
/ keys highlight any log.
RETRIEVING DATA
Press the More function key then press Export.
If the file is already saved on the flash drive, opdo™ will ask for confirmation of overwriting the existing file. The message “FILE EXISTS! OVERWRITE?”.
LOGGING
40
Press CFM for overwriting the existing file.
Note: Do not remove the USB Flash Drive during the file transfer.
If the USB is not found the following screen will appear.
Try reseating USB drive and pressing Export again.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 41
DISSOLVED OXYGEN MEASUREMENTS:
1. Select measurement units of convenience. Press SETUP to change temperature or pressure mea‑ surement units (see page 16 and 17).
2. Press RANGE to access the preferred measurement units; mg/L or % saturation.
3. Routinely Inspect probe for biofouling. Routinely clean off the probe with clean water (between measurements). Biologically active waters may require more frequent cleanings. Ensure weeds, debris, or other materials are not coating the protective guard as this will block sample circulation preventing sample from reaching the active measurement surface.
4. Verify pressure and temperature measurements are reading correctly and that the probe has been
OPERATIONAL GUIDE
calibrated in accordance with sampling protocols.
5. When measuring across a temperature gradient, (i.e. the car is hot and the waters cool), allow the probe to come thermal equilibrium with the water being sampled.
6. The HI98198 with HI764113 have been designed for dissolved oxygen water quality measurements in urban and natural waters. It may be used for discrete spot sampling using the meters Log on Demand function. The maximum depth rating for the probe is 20 m and should be adhered to.
7. Select a water sampling site that will have a representative sampling. Position the probe if possible, facing flow to minimize bubble collection or fluid cavitation. The probe should be measuring the partial pressure of the dissolved oxygen in water. Gas bubbles have a greater partial pressure due to the surface tension of the bubble. Noisy or erratic measurement or higher measurements are possible.
BOD MEASUREMENTS
Biochemical oxygen demand (BOD) is an indicator for the concentration of biodegradable organic matter present in a sample of water. It can be used to infer the general quality of the water and its degree of pollution. BOD measures the rate of oxygen uptake by microorganisms in a sample of water at a fixed temperature and over a given period of time. To ensure that all other conditions are equal, a very small amount of microorganism seed is added to each sample being tested. This seed is typically generated by diluting activated sludge with deionized water. The samples are kept at 20 °C in the dark and tested for dissolved oxygen (DO) after five days. The loss of dissolved oxygen in the sample, once corrections have been made for the degree of dilution, is called the BOD Before measuring BOD, remember to set the BOD configuration from the Setup menu (see page 18).
BOD PROCEDURE
DAY 0 (CAL DO)
Go to Setup and configure BOD configuration parameter which sets up expected sample and seed concentrations.
Determine if you wish to save all BOD data or automatically delete starting data. Go to Setup and configure Autodelete BOD start data.
.
5
DO MEASUREMENT BOD MEASUREMENT
41
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 42
Press MODE to access the BOD measurement screen.
BOD MEASUREMENT
Press BOD Data. The following screen will open. Place the probe into the BOD sample and allow the measurement to stabilize. The sample should be well mixed. Follow all standard operating procedures.
Press Add new. The following screen will open and all parameters have to be filled out.
BOD Parameters:
Bottle ID: a number used to identify a specific bottle.
Range: 0000 to 9999.
The type of the Sample: Sample or Seed.
Bottle Volume: the total volume of the BOD bottle.
Range: 0.1 to 300.0 mL.
Sample Volume: the volume of sample in the BOD bottle.
Range: 0.1 to 300.0 mL (for a seed sample this value is 0.0 mL and cannot be set).
Seed Volume: the volume of seed in the BOD bottle.
Range: 0.0 to 300.0 mL. Press Prev/Next to select a different parameter on the screen. Press
/ keys to modify the selected parameter’s value.
Press Save to save the BOD parameters and the initial DO, temperature, pressure and salinity values for the specified bottle. If a bottle with the same ID already exists, the meter will ask for replacement confirmation.
Press Replace to replace the existing record, or ESC to return to the previous screen without replacing.
42
BOD MEASUREMENT
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 43
When a new record is saved the meter will display a message indicating the remaining free BOD initial data space in %.
Repeat procedure with additional samples: Rinse off probe between samples. Place the probe into the BOD sample and allow the measurement to stabilize. The sample should be well mixed. Follow
BOD MEASUREMENT
all standard operating procedures.
Press Add new and complete BOD Parameters screen (see above). Press Save log initial data from this sample. Prepare all sample bottles for incubation.
Remove probe from sample and fill and cover to prepare bottle for incubation. Rinse probe off with purified water. If desired, move probe to next sample. Press and repeat this
procedure on additional samples and Seed samples. At the conclusion of Day 0 the probe should be cleaned and stored and all samples should be incubated
following Operational Procedures.
Day 5 (Final DO)
Remove samples and Seed samples from the incubator for analysis. Press MODE to display BOD. Press CAL then DO to calibrate the DO probe. Place cleaned and dry probe in sample to be evaluated.
BOD MEASUREMENT
Press Compute. The list of the Day 0 samples and seeds will be present. Use the list of BOD initial data records. The seed records will have the symbol “*” displayed after the bottle ID.
Press More to view detailed information for the selected record.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
/ keys to scroll the
43
Page 44
BOD MEASUREMENT
Select the Bottle ID that is currently being measured by the probe and press EvalBOD.
The BOD will be calibrated.
BOD MEASUREMENT
Press RCL while in the BOD application. BOD recall:
Note: “S.C. “ message in the title bar means seed corrected. “not S.C.” message in the title bar means seed not corrected.
The Correct functional key will be displayed if the BOD result was not seed corrected.
Note: For a seed corrected sample, the last page will show the Seed bottle ID used for correction.
To export to PC
Connect cable to PC and meter while in BOD mode. Meter will display “Connected to Host”. A CSV file will appear on PC with BOD data.
To export to USB drive
While in BOD mode press RCL. Press More. Place the USB into the meter and press Export key. A BOD CSV file will be exported to USB drive.
44
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 45
Seed Correction
In case that the BOD was evaluated for a seeded sample and the list of the saved seed BOD values is not empty, the Correct functional key will be displayed. Press Correct to view the list of the saved seed BOD values. Select the desired seed BOD and then press Correct to compute the corrected BOD value. The meter will display the corrected BOD value.
If the information about the BOD of a certain seed doesn’t exist at the moment of the BOD evaluation for a seeded sample, the sample’s BOD can be corrected at a later time from the BOD recall menu
BOD MEASUREMENT
(view logged BOD data).
In order to perform a seed correction from the BOD recall, press RCL key from the BOD measurement screen to enter BOD recall, select the desired BOD record and press More. The meter will display a complete set of information about the selected record. Press Correct to view the list of the seed values. Select the desired seed BOD and then press Correct to compute the corrected BOD value. The new BOD value will be displayed. The BOD results with SEED correction will be displayed. Press LOG to save and replace the previous sample with this bottle ID or press to keep the previous sample BOD with no seed correction. Clean the probe off thoroughly between samples and continue the evaluation with the next sample.
Note: Various error message will pop up if limits are exceeded.
BOD MEASUREMENT
Also if the final DO value is greater than the initial DO value an error message will be displayed.
45
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 46
The OUR is used to determine the oxygen consumption or respiration rate in water. It is defined as the mg/L of oxygen consumed per hour.
The following equation is used for OUR determination:
where: DO
= Dissolved oxygen level at start of test
START
DO
= Dissolved oxygen level at end of test
END
OUR MEASUREMENT
t
= Elapsed time of test in seconds
ELAPSED
total volume/sample volume = Dilution factor of sample Before starting an OUR test remember to set the OUR configuration from the Setup menu (page 19). This a an example of possible setting. Follow your normal SOP.
Make sure the DO probe has been calibrated before using. Clean off probe between samples. Press MODE to select the OUR measurement mode. Place the calibrated probe into the airtight container
of sample. Stir the sample. Press Start to begin taking measurements. During the analysis the meter will display the ongoing calculation.
OUR MEASUREMENT
46
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 47
OUR recall:
OUR MEASUREMENT
At the end of the analysis the meter will display the computed OUR value, the duration of the measurement, the pressure and the temperature values. Press Log to record a record of this analysis.
OUR MEASUREMENT
47
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 48
To analyze another sample, clean and dry off probe. Place probe into the next prepared sample.
Press Start.
OUR Error messages will be displayed anytime a measurement exceeds the limits that were configured. These will have to be reconfigured and the analysis restarted.
If the DO value is less than the minimum start DO value the meter will display an error message, and the test cannot be started
OUR MEASUREMENT
SOUR MEASUREMENT
48
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 49
The Specific Oxygen Uptake Rate (SOUR), also known as the oxygen consumption or respiration rate, is defined as the milligram of oxygen consumed per gram of volatile suspended solids (VSS) per hour. This quick test has many advantages: rapid measure of influent organic load and biodegradability, indication of the presence of toxic or inhibitory wastes, degree of stability and condition of a sample, and calculation of oxygen demand rates at various points in the aeration basin. The following equation is used for SOUR determination:
SOUR = OUR / Solids Weight
where: OUR is the Oxygen Uptake Rate (see equation on page 46)
SOUR MEASUREMENT
Solids Weight is the Total solids or the Volatile suspended solids weight in g/L.
Temperature correction:
The SOUR value is corrected to 20 °C (68 °F) according to the Farrel and Bhide equation:
SOUR
=SOURTxΘ
20
Where T is the measured temperature in °C and Θ is a temperature dependent variable:
Θ= 1.05 for T above 20 °C Θ= 1.07 for T below 20 °C
This calculation is valid only for temperature values in the range 10 to 30 °C. Temperature correction is performed only if the option SOUR @ 20 °C is enabled.
Before starting a SOUR test remember to set the SOUR configuration from the Setup menu. Make sure DO probe has been calibrated before using. Clean off probe between samples. This is an example of possible setting. Follow your normal SOP.
(20-T)
SOUR MEASUREMENT
Make sure the DO probe has been calibrated before using. Clean off probe between samples. Press MODE to select the SOUR measurement mode. Place the calibrated probe into the airtight
container of sample. Stir the sample. Press Start to begin taking measurements.
49
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 50
During the analysis the meter will display the ongoing calculation.
At the end of the analysis the meter will display the computed SOUR value, the duration of the measurement, and the pressure and temperature values. Press Log to record a record of this analysis.
SOUR MEASUREMENT
To analyze another sample, clean and dry off probe. Place probe into the next prepared sample. Press Start.
SOUR Error messages will be displayed anytime a measurement exceeds the limits that were configured. These will have to be reconfigured and the analysis restarted. Examples follow.
If the DO reading is less than the minimum end DO value set during SOUR configuration, a warning icon will be displayed and a beep will be heard every two seconds. Press Stop to stop the test and the beeper. In case that the SOUR value is corrected to 20 °C (68 °F) and the measured temperature isn’t in the range 10 to 30 °C the temperature value will blink to alert that the temperature correction isn’t valid.
SOUR MEASUREMENT
50
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 51
Press LOG to save a complete set of data regarding the SOUR test.
Press Start to begin a new SOUR test.
Notes: If the DO reading is less than the minimum end DO value set during SOUR configuration, a warning message will be displayed. Press any key to clear the message from the screen, or press HELP to view detailed information about the warning.
SOUR MEASUREMENT
If the DO value is greater than the DO value from the beginning of the test an error message will be displayed.
Press Start to begin a new SOUR test or ESC to return to the SOUR measure screen.
SOUR recall:
SOUR MEASUREMENT
Note: In case that the SOUR value was corrected to 20 °C the message “(@20 °C)” will be displayed before the SOUR value.
51
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 52
To replace the batteries, follow the next steps:
• Turn OFF the meter.
• Open the battery compartment by removing the four screws from the back of the meter.
• Remove the old batteries.
• Insert four new 1.5V AA batteries in the battery compartment while paying attention to the correct
polarity.
• Close the battery compartment using the four screws.
If the battery capacity is less than 10% the backlight feature is not available.
BATTERIES REPLACEMENT
BATTERIES REPLACEMENT
Note: The meter is provided with the BEPS (Battery Error Prevention System) feature, which automatically turns the meter off when the batteries level is too low to ensure reliable readings.
52
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 53
PROBE MAINTENANCE Cleaning the HI764113 Probe Body
• Rinse the probe with clean water to remove debris from around the probe body; wipe with a soft
cloth. Unscrew and remove the protective shield and set aside. Ensure weeds, debris, or other materials are not coating the protective guard as this will block sample circulation preventing sample from reaching the active measurement surface.
• The exterior of the HI764113 can be cleaned by wiping the exterior surface with a aqueous‑soapy
mixture and gently rubbing the exterior surface. Rinse with clean water.
PROBE MAINTENANCE
The Stainless steel guard can be cleaned and polished with a cleaner suitable for stainless steel cookware or appliances. Avoid chlorine bleach containing products.
Inspection
• Routinely Inspect probe for biofouling. Carefully inspect the probe body and Smart Cap™.
A scratch in the black protective layer on the Smart Cap™ will affect the calibration (and measurement). Replace the Smart Cap™ if the sensing surface has been compromised.
Cleaning the Smart Cap™
• Use a mild detergent and a soft bristled toothbrush to clean the Smart Cap™.
Rinse the cap with water after cleaning and dry with a laboratory tissue.
Yearly Replacement of the Smart Cap™ .
• Easy to use smart caps contain pre‑loaded calibration coefficients that are automatically transmitted
to the probe. The Smart Cap™ stores data in a RFID tag. If caps are switched between probes, no information will be lost. Smart Cap™ installation dates are easily tracked via the Probe info screen and when powering , the meter displays remaining cap life.
• Rinse the probe with clean water to remove debris from around the probe body; wipe with a soft
tissue.
opdo™ Probe Replacement Cap Kit
HI764113‑1
Contains: 1 opdo™ probe Smart Cap™ 1 syringe with silicon grease 1 lens wipe 1 certification/ instruction sheet
Note: Verify time and date are properly set on meter prior to new cap initialization.
PROBE MAINTENANCE
53
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 54
1. Unplug the HI764113 from the HI98198 meter. Remove the stainless
steel guard from probe body and set aside. Wipe off the probe and cap with a soft tissue.
2. Remove the expired “Smart Cap™” from the probe; Squeeze the used
cap at the notched V and pull it off the probe body (without twisting).
PROBE MAINTENANCE
3. Remove the used O‑ring from the O‑ring groove by rolling it off the probe.
4. Wipe the O‑ring groove with a soft tissue followed by a cleaning with
the supplied lens cleaner wipe.
5. Remove the new O‑ring from the HI764113‑1 kit and slide on probe
body using care not to roll or twist the O‑ring.
6. Sparingly lubricate the O‑ring with a thin film of the supplied grease. Use
care to prevent grease or fingerprints from contacting the optical window.
PROBE MAINTENANCE
54
7. Clean the lens with an unused portion of the supplied lens cleaner wipe.
8. Remove the new optical cap from the box. Align the notched cutout
arrow on the Smart Cap™ with the matching guide on the probe body.
9. Slide and press the Smart Cap™ onto the HI764113 body until the
cap snaps in place. Once the cap is installed, it should not be removed unless a new cap is required.
10. Connect the HI764113 probe to the HI98198 meter by connecting
the connector to the DIN jack located on the top of the meter.
11. Power the meter to initiate the cap timer.
12. Calibrate.
13. Reinstall the stainless steel guard onto the probe body.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 55
No conditioning period is required when using a HI764113 DO probe. Store the HI764113 probe in the HI98198 carrying case when not in use. For frequent use and short
term storage it is recommended to remove the stainless steel guard and replace with the storage vessel that has a small amount of deionized water. The probe can also be stored with the stainless steel guard on in a beaker containing deionized water.
For longer term storage, remove the stainless steel guard and replace with the storage vessel.
TROUBLESHOOTING GUIDE
SYMPTOMS PROBLEM SOLUTION
PROBE MAINTENANCE
TROUBLESHOOTING GUIDE
Display shows DO reading blinking.
Meter shuts off. Dead batteries; Auto Power
The meter does not start when pressing ON/OFF.
Reading out of range. Recalibrate the meter;
Off feature is enabled: in this case, meter shuts off after selected period of non use.
Initialization error. Press and hold down
Check the sample is within measurable range.
Replace batteries; Press
ON/OFF.
ON/OFF for about
20 seconds or disconnect and then connect the batteries.
55
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 56
SYMPTOMS PROBLEM SOLUTION
TROUBLESHOOTING GUIDE
Display shows “No Cap Detected” .
Display shows “No Cap Info Detected” with Cap.
Display shows “Cap Damaged”.
TROUBLESHOOTING GUIDE
Display shows “Cap Expired”.
Display “No Probe”. Probe not present/
Meter shows “Probe Err xx” message.
Cap not on correctly. Check Cap/Reseat Cap.
Cannot read Cap Info.
Sensor Damaged. Try reseating cap. If this doesn’t
Cap reached lifetime use expiration.
Not connected properly.
Probe Internal Error. Disconnect and reconnect probe.
Check Cap and Retry. Replace Cap if it continues.
resolve issue, replace Cap.
Continue OR Replace Cap (Continuing use may result in incorrect measurements).
Connect/Disconnect and reconnect probe OR Turn meter OFF then ON.
If problem persists replace the probe.
56
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 57
Code Description
ACCESSORIES
HI7040
HI40036P 100 mL plastic beaker (10 pcs.)
ACCESSORIES
HI740027P 1.5V AA batteries (12 pcs.)
HI764113
HI764113-1
HI764113-2 Calibration/storage vessel
HI764113-3 Stainless steel protective shield
HI764113/10 HI764113 with 10 m cable
HI764113/20 HI764113 with 20 m cable
HI920016 USB Type A to C cable
Bi‑component Zero
Optical DO 4 m (13’4) cable
Smart Cap™ with O‑ring
probe with built‑in temperature sensor, shield and
57
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 58
All Hanna Instruments conform to the CE European Directives.
RoHS
compliant
Disposal of Electrical & Electronic Equipment. The product should not be treated as household waste. Instead hand it over to the appropriate collection point for the recycling of electrical and
CERTIFICATION
electronic equipment which will conserve natural resources.
Disposal of waste batteries. This product contains batteries, do not dispose of them with other household waste. Hand them over to the appropriate collection point for recycling.
Ensuring proper product and battery disposal prevents potential negative consequences for the environment and human health. For more information, contact your city, your local household waste disposal service, the place of purchase or go to
MAN98198 07/18
58
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 59
Recommendations for users
Before using this product, make sure it is entirely suitable for your specific application and for the environment in which it is used. Any variation introduced by the user to the supplied equipment may degrade the meters’ performance. For yours and the meter’s safety do not use or store the meter in hazardous environments.
Warranty
Hanna Instruments reserves the right to modify the design, construction or appearance of its
The HI98198 and HI764113 are warranted for two years against defects in workmanship and materials when used for their intended purpose and maintained according to instructions. HI764113‑1 is warranted for 1 year. This warranty is limited to repair or replacement free of charge.
Damage due to accidents, misuse, tampering or lack of recommended maintenance is not covered.
If service is required, contact your local Hanna Instruments Office. If under warranty, report the model number, date of purchase, serial number and the nature of the problem. If the repair is not covered by the warranty, you will be notified of the charges incurred. If the meter is to be returned to Hanna Instruments, first obtain a Returned Goods Authorization (RGA) number from the Technical Service department and then send it with shipping costs prepaid. When shipping any meter, make sure it is properly packed for complete protection.
products without advanced notice.
59
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Loading...