Hanna Instruments HI 98186 User Manual

Instruction Manual
HI 98186
Dissolved Oxygen
BOD/OUR/SOUR
Meter
HI 98186
www.hannainst.com
1
Dear Customer, Thank you for choosing a Hanna Instruments product. Please read this instruction manual carefully before using the instrument. This manual will provide you with the necessary information for correct use of the instrument, as well as a precise idea of its versatility. If you need additional technical information, do not hesitate to e-mail us at tech@hannainst.com or turn to the back cover for our worldwide contact list. These instruments are in compliance with directives.
WARRANTYWARRANTY
WARRANTY
WARRANTYWARRANTY
The HI 98186 instrument is guaranteed for two years against defects in workmanship and materials when used for its intended purpose and maintained according to instructions. Electrodes and probes are guaranteed for six months. This warranty is limited to repair or replacement free of charge.
Damage due to accidents, misuse, tampering or lack of prescribed maintenance is not covered.
If service is required, contact the dealer from whom you purchased the instrument. If under warranty, report the model number, date of purchase, serial number and the nature of the problem. If the repair is not covered by the warranty, you will be notified of the charges incurred. If the instrument is to be returned to Hanna Instruments, first obtain a Returned Goods Authori­zation number from the Technical Service department and then send it with shipping costs prepaid. When shipping any instrument, make sure it is properly packed for complete protection.
2
TABLE OF CONTENTSTABLE OF CONTENTS
TABLE OF CONTENTS
TABLE OF CONTENTSTABLE OF CONTENTS
WARRANTY ......................................................................................... 2
PRELIMINARY EXAMINATION ................................................................ 4
GENERAL DESCRIPTION ........................................................................ 4
FUNCTIONAL DESCRIPTION .................................................................. 6
SPECIFICATIONS.................................................................................. 8
PROBE FUNCTIONAL DESCRIPTION ....................................................... 9
PROBE CONNECTION AND PREPARATION............................................. 10
OPERATIONAL GUIDE......................................................................... 11
OUR MEASUREMENT ......................................................................... 21
SOUR MEASUREMENT ....................................................................... 24
TEMPERATURE MEASUREMENT ............................................................ 27
DO CALIBRATION PROCEDURE ............................................................ 27
GOOD LABORATORY PRACTICE (GLP) ................................................... 32
SETUP .............................................................................................. 33
LOGGING .......................................................................................... 45
AutoEnd ........................................................................................... 48
PRESSURE CALIBRATION .................................................................... 49
TEMPERATURE CALIBRATION .............................................................. 51
PC INTERFACE ................................................................................... 54
BATTERY RECHARGER/REPLACEMENT .................................................... 62
PROBE MAINTENANCE ....................................................................... 64
TROUBLESHOOTING GUIDE ................................................................ 65
ACCESSORIES .................................................................................... 66
3
PRELIMINARY EXAMINATIONPRELIMINARY EXAMINATION
PRELIMINARY EXAMINATION
PRELIMINARY EXAMINATIONPRELIMINARY EXAMINATION
Remove the instrument from the packing material and examine it carefully to make sure that no damage has occurred during shipping. If there is any damage, notify your Dealer or the nearest Hanna Customer Service Center.
Each instrument is supplied with:
HI 76407/4F Polarographic probe with built-in temperature sensors and
4 m (13.4”) cable
HI 76407A Membrane cap (2 pcs)
HI 7041S Electrolyte solution (30 mL)
• 4 x 1.2V AA, 1300 mAh Rechargeable Batteries (inside the instrument)
HI 710042 Inductive Recharger with power adaptor
• Instruction Manual
• Rugged carrying case Note: Save all packing material until you are sure that the instrument
functions correctly. All defective items must be returned in the original packing with the supplied accessories.
GENERAL DESCRIPTIONGENERAL DESCRIPTION
GENERAL DESCRIPTION
GENERAL DESCRIPTIONGENERAL DESCRIPTION
The HI 98186 is a state of art, microprocessor based, dissolved oxygen instrument with many automated and specific application features, designed to provide laboratory results and accuracy under harsh industrial conditions. All measurements are automatically compensated for temperature. Salinity compensation in water allows direct determination of dissolved oxygen in saline waters. With its internal barometer, the instrument is able to automati­cally compensate for changes in barometric pressure so there is no need for charts, altitude information or external barometric pressure information. The instrument contains built-in application software for the calculation of Biochemical Oxygen Demand (BOD), Oxygen Uptake Rate (OUR) and Specific Oxygen Uptake Rate (SOUR). HI 98186’s calibration has been greatly simplified when compared to other dissolved oxygen instruments. It is provided with a series of new features which add entirely new dimensions to the measurement of DO, by allowing the user to dramatically improve the reliability of the measurement:
• One or two points automatic dissolved oxygen calibration.
• One point manual dissolved oxygen calibration using a value in milligrams per liter or percent saturation entered by the user.
• One or two points user temperature calibration.
• Messages on the graphic LCD for an easy and accurate calibration.
• User-selectable “calibration time out” to remind when a new calibration is necessary.
4
Moreover, the meter offers an extended temperature range from –20 ºC to 120 ºC (–4 ºF to 248 ºF).
Other features include:
• Log on demand up to 400 samples.
• Auto Hold feature, to freeze first stable reading on the LCD.
• GLP feature, to view last calibration data.
• PC interface.
Hanna Instruments reserves the right to modify the design,
construction and appearance of its products without advance notice.
5
Front viewFront view
Front view
Front viewFront view
FUNCTIONAL DESCRIPTIONFUNCTIONAL DESCRIPTION
FUNCTIONAL DESCRIPTION
FUNCTIONAL DESCRIPTIONFUNCTIONAL DESCRIPTION
1) Liquid Crystal Display (LCD).
2) F1, F2, F3 functional keys.
/ keys to manually increase/decrease the parameters or to scroll the pa-
3)
rameter list.
4) ESC key to leave current mode, exit calibration, setup, help, etc.
5) CAL key, to enter/exit calibration mode.
6) RCL key, to enter/exit view logged data mode.
7) SETUP key, to enter/exit SETUP mode.
8) ON/OFF key, to turn the instrument ON and OFF.
9) GLP key, to display Good Laboratory Practice information.
10) LIGHT key, to toggle display backlighting.
11) MODE key to change DO measuring unit when in DO measure screen, or to
toggle between standard and pressure in DO calibration.
12) RANGE key, to switch between DO, BOD, OUR AND SOUR.
13) Help key to enter/exit contextual help.
6
Top viewTop view
Top view
Top viewTop view
15
14) Electrode DIN connector.
15) USB connector.
14
DOUSB
7
SPECIFICATIONSSPECIFICATIONS
SPECIFICATIONS
SPECIFICATIONSSPECIFICATIONS
EGNAR
NOITULOSER
YCARUCCA
Fº86/Cº02@
noitarbilaCOD
noitarbilacerutarepmeTeulaverutarepmetegnarniynatatniop2ro1
noitarbilacerusserPeulaverusserpegnarniynatatniop1
noitasnepmocerutarepmeT
noitasnepmoCerusserPgHmm058ot054morfcitamotuA
L/gm00.05ot00.0
noitarutas%0.006ot0.0
gHmm058ot054
)Fº0.842ot0.4–(Cº0.021ot0.02–
L/gm10.0
noitarutas%1.0
gHmm1
)Fº1.0(Cº1.0
tigid1±gnidaer%5.1
noitarbilacehtmorf%51±nihtiwgHmm3±
tniop
)Fº4.0±(Cº2.0±
)rorreeborpgnidulcxe(
tanoitarbilaccitamotuastniop2roeno•
.)L/gm0(%0dna)L/gm62.8(%001
ybderetneeulavagnisulaunamtniop1•
.L/gmronoitarutas%niresueht
morfcitamotuA
)Fº0.221ot0.23(Cº0.05ot0.0
noitasnepmoCytinilaSL/g07ot0morfcitamotuA
eborPOD F4/70467IH cihpargoraloP
GOLselpmas004,dnamednO
efiL&epyTyrettaB
regrahceR 240017IH )dedulcni(regrahcerevitcudnI
ffo-otuA
ecafretnICPdetalosi-otpo BSU
snoisnemiD)''2x57.3x9.8(mm25x59x5.622
)ylnoretem(thgieWg525
tnemnorivnE
8
seirettabelbaegrahcerAAV2.1x4
tuohtiwesusuounitnocfosruoh002.xorppa
)thgilkcabhtiwsruoh05(thgilkcab
rosetunim06,03,01,5:elbatcelesresU
delbasid
)Fº221–23(Cº05–0
%001HR.xam
PROBE FUNCTIONAL DESCRIPTIONPROBE FUNCTIONAL DESCRIPTION
PROBE FUNCTIONAL DESCRIPTION
PROBE FUNCTIONAL DESCRIPTIONPROBE FUNCTIONAL DESCRIPTION
PROBE FUNCTIONAL DESCRIPTIONPROBE FUNCTIONAL DESCRIPTION
PROBE FUNCTIONAL DESCRIPTION
PROBE FUNCTIONAL DESCRIPTIONPROBE FUNCTIONAL DESCRIPTION
1. D.O. Probe
2. Protective Cap
3. Watertight Shielded Cable
4. Polypropylene Probe Body
5. Temperature Sensor
6. O-Ring Seal
7. Silver Chloride Anode
8. Platinum Cathode (sensor)
9. Oxygen Permeable Teflon® Membrane
10. Membrane Cap
9
PROBE CONNECTION AND PREPARATIONPROBE CONNECTION AND PREPARATION
PROBE CONNECTION AND PREPARATION
PROBE CONNECTION AND PREPARATIONPROBE CONNECTION AND PREPARATION
To take measurements, connect the D.O. probe to the meter securely by aligning the pins with the socket located on the top of the meter, pushing the plug in and tightening the threaded ring.
Probes shipped from Hanna Instruments are dry. To hydrate the probe and prepare it for use, connect it to the meter and proceed as follows:
1. Remove the red and black plastic cap.
This cap is for shipping purposes and can be thrown away.
2. Wet the sensor by soaking the bottom
2½ cm (1") of the probe in electrolyte (HI 7041S) for 5 minutes.
3. Rinse the membrane cap (HI 76407A)
supplied in the kit with the meter) with electrolyte solution while shaking it gently. Refill with clean electrolyte solution.
4. Tap gently the sides of the membrane
cap with your finger tip to ensure that no air bubbles are trapped. To avoid damag­ing the membrane, do not tap it directly on the bottom.
5. Make sure that the rubber O-ring sits
properly inside the membrane cap.
6. With the sensor facing down, slowly
screw the cap clockwise. Some electrolyte will overflow.
When not in use and during polarization (see PROBE CONDITIONING page 11), use the protective transparent cap supplied in the kit with the meter.
10
OPERATIONAL GUIDEOPERATIONAL GUIDE
OPERATIONAL GUIDE
OPERATIONAL GUIDEOPERATIONAL GUIDE
INITIAL PREPARATION
The instrument is supplied complete with rechargeable batteries. Proceed with a complete charging process before starting (see page 60).
To prepare the instrument for field measurements close the serial communica­tion socket and all unused connector sockets with proper stopper (to ensure waterproof protection).
Connect the DO probe to the 7-pin connector. Make sure that the probe sleeve is properly inserted and tighten the threaded ring.
Turn the instrument ON by pressing ON/OFF key. At start-up the display will show the Hanna logo for a few seconds , followed by
the percentage indication of the remaining battery charge, then enters the probe conditioning mode. The probe will be conditioned for one minute, and afterwards the instrument will enter measurement mode.
If the DO probe is not connected or is damaged, the conditioning period is skipped.
To save battery life, the auto-off feature turns the instrument off after a set period (default 30 min) if no button is pressed. To set another period or to disable this feature, see SETUP menu on page 33.
The auto-off backlight feature turns the backlight off after a set period (default 1 min) with no buttons pressed. To set another period or to disable this feature, see SETUP on page 33.
PROBE CONDITIONING
At startup, the probe is under polarization with a fixed voltage of approxi­mately 800 mV for 1 minute.
Probe polarization is essential for stable measurements with the same recur­ring degree of accuracy.
With the probe properly polarized, oxygen is continually consumed when it passes through the sensitive diaphragm and dissolves in the electrolyte solu­tion contained in the probe.
Whenever measurements are taken with a non-polarized probe, the oxygen level revealed is both that of the tested solution, as well as that present in the electrolyte solution. This reading is incorrect.
Keep the protective cap on during polarization time and remove it for calibra­tion and measurements.
11
SALINITY COMPENSATION
If the sample contains significant concentration of salinity, the read out values must be corrected, taking into account the lower degree of oxygen solubility in this situation. Before taking any DO measurements remember to set the salinity value from the SETUP menu (page 33).
The salinity affects the D.O. concentration, decreasing its value. The table below shows the maximum oxygen solubility at various temperatures and salinity levels.
l/g0l/g01l/g02l/g03l/g53
006.4146.3147.2109.1105.110.23
218.3119.2170.2192.1119.015.63
490.3152.2174.1137.0183.012.93
644.2156.1119.0122.0198.98.24
838.1190.1104.0157.944.94.64
0182.1185.0139.923.930.90.05
2177.0111.0105.929.856.86.35
4192.0186.901.955.803.82.75
6168.982.937.812.879.78.06
8154.909.893.809.766.74.46
0280.965.870.806.783.70.86
2237.832.877.733.721.76.17
4204.839.794.770.778.62.57
5242.897.763.759.657.60.77
6290.856.732.738.646.68.87
8218.783.789.616.624.64.28
0345.741.757.693.622.60.68
2392.709.645.691.630.66.98
4350.786.633.610.658.52.39
6328.674.641.638.586.58.69
8316.682.669.566.515.54.001
0414.690.697.505.563.50.401
2422.639.536.553.522.56.701
4440.677.584.512.590.52.111
6478.516.533.570.579.48.411
8407.574.502.559.458.44.811
0545.533.570.538.457.40.221
leveLaeSta)l/g(ytinilaS
NoteNote
Note: The relationship between salinity and chlorinity for sea water is given by the equation
NoteNote
below:
Salinity (g/l) = 1.80655 Chlorinity (g/l)
12
BAROMETRIC PRESSURE COMPENSATION
The dissolved oxygen saturation value varies with pressure, so it is important to compensate the effect that pressure has on DO measurements.
leveLaeSevobasreteM,edutitlA
0m003m006m009m0021m0051m0081m0012m0042m0072m0003m0033m0063m0093m0004
06.411.416.311.316.211.217.112.118.014.010.017.93.90.99.80.23
28.313.318.214.219.115.110.116.012.019.95.92.98.85.84.86.53
41.316.212.217.113.119.015.011.017.93.90.97.84.80.89.72.93
64.210.215.111.117.013.019.96.92.99.86.82.89.76.75.78.24
88.114.110.116.012.018.95.91.98.84.81.88.75.73.72.74.64
013.119.015.011.017.94.90.97.84.81.88.75.72.79.68.60.05
218.014.010.016.93.99.86.83.80.87.74.71.79.66.65.66.35
413.019.96.92.99.85.82.89.76.74.71.78.66.63.62.62.75
619.95.92.98.85.82.89.76.73.70.78.65.63.61.60.68.06
815.91.98.85.81.88.76.73.70.78.65.63.60.68.57.54.46
021.98.84.81.88.75.73.70.77.65.62.60.68.56.55.50.86
227.84.81.88.75.72.70.77.65.62.60.68.56.54.53.56.17
424.81.88.75.72.70.77.65.62.60.68.56.54.52.51.52.57
523.80.87.74.71.78.66.64.61.69.57.55.53.51.50.50.77
621.88.75.72.70.77.65.62.60.68.56.54.52.50.59.48.87
828.75.73.70.77.65.62.60.68.56.54.52.50.58.47.44.28
036.73.70.78.65.63.60.68.56.54.52.50.58.46.46.40.68
233.70.78.65.63.61.68.56.54.52.50.58.47.45.44.46.98
431.78.66.63.61.69.56.54.52.50.59.47.45.43.43.42.39
638.66.63.61.69.57.55.53.51.59.47.45.44.42.41.48.69
836.64.61.69.57.55.53.51.59.47.45.44.42.41.40.44.001
044.62.69.57.55.53.51.59.47.46.44.42.41.49.39.34.401
242.60.68.56.53.52.50.58.46.44.43.41.40.48.38.36.701
440.68.56.54.52.50.58.46.45.43.41.40.48.37.37.32.111
648.56.54.52.50.58.47.45.43.42.40.49.37.36.35.38.411
847.55.53.51.59.47.45.44.42.40.49.37.36.35.34.34.811
055.53.51.59.47.46.44.42.41.49.38.36.35.34.33.30.221
m
13
The HI 98186 meter contains a built-in barometer, and it is able to automati- cally compensate for changes in barometric pressure. If another pressure value than the barometer’s reading is to be used, then the manual pressure feature must be enabled from the SETUP menu (see page 33), and afterwards the pressure value can be set using the ARROW keys.
The table below contains a conversion altitude (m) to pressure (mmHg) for the altitude values from the previous table.
edutitlA
000300600900210051008100120042007200030033006300930004
)m(
erusserP
067237507976456036706485365245225305484764164
)gHmm(
DO MEASUREMENTS
Make sure the probe’s protective cap has been removed. In order to take accurate dissolved oxygen measurements make sure that the instrument has been calibrated (see page 27 for details). Press RANGE to access the DO measure screen. If necessary press MODE to change the measuring unit. Immerse the tip of the probe into the sample to be tested. Allow approxi­mately one minute for the reading to stabilize (hourglass symbol turns off).
On the screen are displayed:
Dissolved Oxygen reading in the selected unit (% saturation or mg/L)
Temperature reading in the selected unit (°C or °F)
Pressure reading in the selected unit (mmHg, inHg, atm, psi, kPa, mbar).
If the Manual Pressure option is enabled ( displayed in front of the pressure value) the pressure value can be changed using the ARROW keys. For accurate dissolved oxygen measurements, a water movement of
0.3 m/s is required. This is to ensure that the oxygen-depleted membrane
surface is constantly replenished. A moving stream will provide adequate circulation.
14
BOD MEASUREMENTS
Biochemical oxygen demand (BOD) is an indicator for the concentration of biodegradable organic matter present in a sample of water. It can be used to infer the general quality of the water and its degree of pollution. BOD measures the rate of oxygen uptake by microorganisms in a sample of water at a fixed temperature and over a given period of time. To ensure that all other conditions are equal, a very small amount of microorganism seed is added to each sample being tested. This seed is typically generated by diluting activated sludge with deionized water. The samples are kept at 20 ºC in the dark and tested for dissolved oxygen (DO) after five days. The loss of dissolved oxygen in the sample, once correction have been made for the degree of dilution, is called the BOD
5
Before measuring BOD, remember to set the BOD configuration from the SETUP menu (see page 33).
Press RANGE to access the BOD measure screen.
Press BOD Data to view the BOD initial data management screen. Press Compute to evaluate the BOD for a specified sample (available only
when the measurement is stable and at least one initial BOD data record has been memorized).
BOD initial data management screen
BOD Data is pressed while in BOD measurement screen.
Press Add new to add a new BOD initial data record (the key is available only when the measurement is stable). A 200 records memory space is available for BOD initial data.
Press View ID to view the saved BOD initial values (the key is available only
15
when at least one initial BOD data record has been memorized). By pressing Add New the BOD Parameters screen will be displayed:
BOD Parameters:
Bottle ID: a number used to identify a specific bottle.
Range: 0000 to 9999.
The type of the Sample: Sample or Seed.
Bottle Volume: the total volume of the BOD bottle.
Range: 0.1 to 300.0 mL.
Sample Volume: the volume of sample in the BOD bottle.
Range: 0.1 to 300.0 mL (for a seed sample this value is 0.0 mL and
cannot be set).
Seed Volume: the volume of seed in the BOD bottle.
Range: 0.0 to 300.0 mL. Press Prev/Next to select a different parameter on the screen. Press ARROW keys to modify the selected parameter’s value. Press Save to save the BOD parameters and the initial DO, temperature,
pressure and salinity values for the specified bottle.
If a bottle with the same ID already exists, the instrument will ask for replacement confirmation.
Press Replace to replace the existing record, or ESC to return to the previous screen without replacing.
When a new record is saved the meter will display a message indicating the remaining free BOD initial data space in %.
16
By pressing View ID a list of all the saved BOD initial data records will be displayed. The seed records will have the symbol “*” displayed after the bottle ID.
Use the ARROW keys to scroll the list of BOD initial data records. Press More to view detailed information for the selected record.
Press PgUp/Pg/Down to view the next/previous screen of information. Use the ARROW keys to view the detailed information about the
next/previous record. If Delete is pressed.
Use ARROW keys to focus on the record to be deleted and then press CFM. Press ESC to exit.
If Delete All is pressed the instrument asks for confirmation. Press CFM to confirm or ESC to exit without deleting.
17
BOD evaluation
From the BOD measure screen press Compute to evaluate the BOD for a specified sample. The following screen will be displayed.
If the date of the current measurement is previous to the date of the selected measurement then the Eval. BOD key will not be displayed.
Press MORE to view detailed information for the selected record. Use ARROW keys to select the bottle for BOD evaluation. Press Eval. BOD to compute the BOD for the selected bottle. If the time
difference between the current reading and the selected reading is less than 1 day the instrument will ask for record replacement confirmation, and the BOD can’t be evaluated.
Press CFM to replace the selected record’s DO, temperature, pressure and salinity values with the current values.
Press ESC to return to the previous screen without replacing. If the conditions regarding the time difference are met, after pressing the
Eval.BOD key, the instrument will display the computed BOD value.
Press LOG to save the BOD result. Press ESC to return to the BOD measure screen. If the final DO reading or DO loss don’t meet the criteria for BOD measure-
ments set from the instrument’s setup, a warning message will be displayed.
18
Press any key to clear the warning message from the display or press HELP to view detailed information about the warning.
Note: If the Autodelete BOD start data option is enabled in SETUP
(see page 33), when the BOD result is saved (LOG key pressed) the corresponding BOD initial data record will be automatically deleted from the instrument’s memory.
Seed Correction
In case that the BOD was evaluated for a seeded sample and the list of the saved seed BOD values is not empty, the Correct functional key will be displayed.
Press Correct to view the list of the saved seed BOD values.
Select the desired seed BOD and then press Correct to compute the corrected BOD value. The instrument will display the corrected BOD value.
If the information about the BOD of a certain seed doesn’t exist at the moment of the BOD evaluation for a seeded sample, the sample’s BOD can be corrected at a later time from the BOD recall menu (view logged BOD data).
In order to perform a seed correction from the BOD recall, press RCL key from the BOD measurement screen to enter BOD recall, select the desired BOD record and press More. The instrument will display a complete set of informa­tion about the selected record.
Press Correct to view the list of the seed values. Select the desired seed BOD and then press Correct to compute the corrected
BOD value. The new BOD value will be displayed.
19
Note: If the final DO value is greater than the initial DO value an error
message will be displayed.
20
OUR MEASUREMENTOUR MEASUREMENT
OUR MEASUREMENT
OUR MEASUREMENTOUR MEASUREMENT
The OUR is used to determine the oxygen consumption or respiration rate. It is defined as the mg/L of oxygen consumed per hour.
The following equation is used for OUR determination:
where: DO
START = Dissolved oxygen level at start of test
DOEND = Dissolved oxygen level at end of test
tELAPSED = Elapsed time of test in seconds
total volume/sample volume = Dilution factor of sample Before starting an OUR test remember to set the OUR configuration from the
SETUP menu (page 33).
OUR measure screen:
Press Start to begin a new OUR test. If the DO value is less than the minimum start DO value the meter will display an error message, and the test cannot be started.
If the minimum start DO condition is met the instrument will display the instantaneous oxygen consumption rate and the amount of time that has passed from the beginning of the test.
21
Loading...
+ 47 hidden pages