HI 98180 HI 98181
HI 98182 HI 98183
HI 98184 HI 98185
Calibration Check
Waterproof
pH/mV/ISE/Temperature
Meters
www.hannainst.com
1
Dear Customer,
Thank you for choosing a Hanna Instruments product.
Please read this instruction manual carefully before using the instruments.
This manual will provide you with the necessary information for correct use of
the instruments, as well as a precise idea of their versatility.
If you need additional technical information, do not hesitate to e-mail us at
tech@hannainst.com or turn to the back cover for our worldwide contact list.
These instruments are in compliance with directives.
WARRANTYWARRANTY
WARRANTY
WARRANTYWARRANTY
The HI 9818x series are guaranteed for two years against defects in workman-
ship and materials when used for their intended purpose and maintained
according to instructions. Electrodes and probes are guaranteed for six months.
This warranty is limited to repair or replacement free of charge.
Damage due to accidents, misuse, tampering or lack of prescribed maintenance
is not covered.
If service is required, contact the dealer from whom you purchased the
instrument. If under warranty, report the model number, date of purchase,
serial number and the nature of the problem. If the repair is not covered by
the warranty, you will be notified of the charges incurred. If the instrument is
to be returned to Hanna Instruments, first obtain a Returned Goods Authorization number from the Technical Service department and then send it with
shipping costs prepaid. When shipping any instrument, make sure it is
properly packed for complete protection.
Remove the instrument from the packing material and examine it carefully to
make sure that no damage has occurred during shipping. If there is any
damage, notify your Dealer or the nearest Hanna Customer Service Center.
Each instrument is supplied with:
• HI 1230B Combination double-junction, gel pH Electrode (HI 98180/1/2)
• HI 7662 stainless steel Temperature Probe with 1 m (3.3') Cable
(HI 98180/1/2)
• HI 72911B Combined pH temperature electrode (HI 98183/4/5)
• pH 4.01 & 7.01 Buffer Solutions (230 mL each)
• 100 mL Plastic Beaker
• 4 x 1.2V AA Rechargeable Batteries (inside the instrument)
• HI 710042 Inductive Recharger with power adaptor
• Instruction Manual
• Rugged carrying case
Note: Save all packing material until you are sure that the instrument
functions correctly. All defective items must be returned in the original
packing with the supplied accessories.
GENERAL DESCRIPTIONGENERAL DESCRIPTION
GENERAL DESCRIPTION
GENERAL DESCRIPTIONGENERAL DESCRIPTION
The HI 9818x instruments are state-of-the-art, heavy-duty pH meters, de-
signed to provide laboratory results and accuracy under harsh industrial
conditions.
They are provided with a series of new diagnostic features which add an
entirely new dimension to the measurement of pH, by allowing the user to
dramatically improve the reliability of the measurement:
• pH calibration up to five calibration points (see instrument specifications).
• Custom calibration with up to five custom buffers (not for HI 98180/1).
• Messages on the graphic LCD for an easy and accurate calibration.
• Diagnostic features to alert the user when the electrode needs cleaning.
• User selectable “Outside Calibration Range” warning.
• Monitoring of the electrode aging.
• User-selectable “calibration time out” to remind when a new calibration is
necessary.
Moreover, they offer an extended temperature range from –20 to 120 ºC
Hanna Instruments reserves the right to modify the design,
construction and appearance of its products without advance notice.
3
(–4 to 248 ºF), using HI 7662 interchangeable temperature probes or
temperature sensor inside pH electrode.
These instruments can also measure with ORP electrodes, thanks to their
capability to measure mV with a resolution up to 0.1 mV (not for HI 98180).HI 98184 and HI 98185 can also measure with ISE electrodes on ppm scale.
The electrode type and unit selection capability (for HI 98185) the ion change
selection capability (for HI 98184) and the ISE calibration in up to five
calibration standard solutions make this instrument very useful for a large
range of concentration solutions measurements.
Other features include:
• Relative mV measurements
• Log on demand up to 300 samples (100 samples on each range - pH,
mV, ISE) (not for HI 98180 and HI 98181)
• Auto Hold feature, to freeze first stable reading on the LCD
• GLP feature, to view last calibration data for pH, Rel mV or ISE
• PC interface
4
Front viewFront view
Front view
Front viewFront view
HI 98180, HI 98181HI 98180, HI 98181
HI 98180, HI 98181
HI 98180, HI 98181HI 98180, HI 98181
FUNCTIONAL DESCRIPTIONFUNCTIONAL DESCRIPTION
FUNCTIONAL DESCRIPTION
FUNCTIONAL DESCRIPTIONFUNCTIONAL DESCRIPTION
1) Liquid Crystal Display (LCD).
2) Functional keys.
3)/ keys to manually increase/decrease the parameters or to scroll between
the parameter list.
4) ESC to escape the current mode, exit calibration, setup, help. etc.
5) CAL key, to enter/exit calibration mode.
6) CLR key, to clean calibration parameters.
7) SETUP key, to enter/exit SETUP mode.
8) ON/OFF key, to turn the instrument ON and OFF.
9) GLP key, to display Good Laboratory Practice information.
10) LIGHT key to toggle between backlighting.
11) MODE key, to change pH resolution or to toggle between mV and Rel mV mode
(HI 98181 only).
12) AutoEnd key, to freeze the first stable reading on the LCD (HI 98180).
RANGE key, to switch between pH and mV range (HI 98181).
13) Help key to enter/exit contextual help.
5
Top viewTop view
Top view
Top viewTop view
14) BNC electrode connector.
15) Temperature socket.
6
Front viewFront view
Front view
Front viewFront view
HI 98182, HI 98183,HI 98182, HI 98183,
HI 98182, HI 98183,
HI 98182, HI 98183,HI 98182, HI 98183,
HI 98184, AND HI 98185HI 98184, AND HI 98185
HI 98184, AND HI 98185
HI 98184, AND HI 98185HI 98184, AND HI 98185
FUNCTIONAL DESCRIPTIONFUNCTIONAL DESCRIPTION
FUNCTIONAL DESCRIPTION
FUNCTIONAL DESCRIPTIONFUNCTIONAL DESCRIPTION
1) Liquid Crystal Display (LCD).
2) Functional keys.
3)/ keys to manually increase/decrease the parameters or to scroll between
the parameter list.
4) ESC to escape the current mode, exit calibration, setup, help, etc.
5) CAL key, to enter/exit calibration mode.
6) RCL key, to enter/exit view logged data mode.
7) SETUP key, to enter/exit SETUP mode.
8) ON/OFF key, to turn the instrument ON and OFF.
9) GLP key, to display Good Laboratory Practice information.
10) LIGHT key to toggle between backlighting.
11) MODE key, to change pH resolution or to toggle between mV and Rel mV mode.
12) RANGE key, to switch between pH and mV range (HI 98182/3) or pH, mV and
ISE range (HI 98184/5).
13) Help key to enter/exit contextual help.
7
Top viewTop view
Top view
Top viewTop view
14) BNC electrode connector.
15) Reference socket (HI 98184/5).
16) Reference socket rubber stopper (HI 98184/5).
17) Temperature socket
18) Temperature socket rubber stopper (HI 98183/4/5).
19) Temperature measurement rubber cork (HI 98183/4/5)
20) Serial communication rubber stopper (HI 98183/4/5).
The instrument is supplied complete with rechargeable batteries. Proceed with
a complete charging process before starting (see page 58).
To prepare the instrument for field measurements close the serial communication socket and all unused connector sockets with proper stopper (to ensure
waterproof protection) (HI 98183/4/5 only). Use the holed temperature
rubber cork for the temperature socket when temperature probe is connected
(HI 98183/4/5 only).
Connect the pH electrode and the temperature probe to the BNC and temperature sockets on the top of the instrument. Push the pH electrode sleeve to cover
the connector accomodation. The temperature probe is used in conjunction
with the pH electrode to utilize the instrument's ATC capability, but it can also
be used independently to take temperature measurements. If the probe is
disconnected, temperature can be set manually with the ARROW keys (see
page 13 for details).
Turn the instrument ON by pressing ON/OFF key.
At start-up the display will show the Hanna logo for a few seconds followed by
the percentage indication of the remaining battery charge, then enters the
measurement mode.
After measurement switch the instrument off, clean the electrode and store it with
a few drops of HI 70300 storage solution in the protection cap (see page 62).
The auto-off feature turns the instrument off after a set period (default 30 min)
with no button pressed to save battery life. To set another period or to disable
this feature, see SETUP menu on page 33.
The auto-off backlight feature turns the backlight off after a set period (default
1 min) with no buttons pressed. To set another period or to disable this
feature, see SETUP menu on page 33.
pH MEASUREMENTS
To take a pH measurement remove the electrode protective cap and simply
submerge the tip of the electrode (4cm/1½”) and the
temperature probe into the sample to be tested.
If necessary, press RANGE until the display changes to
the pH mode. Use Mode key to select the pH resolution.
12
Allow for the electrode to adjust and reading to stabilize (hourglass symbol
turns off).
On the pH screen are displayed:
• pH reading with the selected resolution.
• Temperature reading in the selected unit (ºC or ºF).
• Temperature compensation mode (MTC - manual, ATC - automatic).
While in MTC mode the indicate that the temperature can be
manually changed using ARROW keys.
• Electrode condition during the calibration day.
• The buffers used in last pH calibration (if feature is enabled in SETUP).
• Battery level indicator.
• Available functional keys in according with the model.
In order to take more accurate pH measurements, make sure that the
instrument is calibrated (see page 16 for details).
It is recommended that the electrode is always kept wet and rinsed thoroughly
with the sample to be measured before use.
The pH reading is directly affected by temperature. For accurate pH measurements, temperature must be taken into consideration. If the sample temperature is different from the temperature at which the pH electrode was kept, allow
a few minutes to reach thermal equilibrium.
To use the instrument's Automatic Temperature Compensation feature, submerge
the temperature probe into the sample as close to the electrode as possible and
wait for a few seconds.
If manual temperature compensation (MTC) is desired, the temperature
probe must be disconnected from the instrument.
The display will show the default temperature of 25 ºC, the last measured
temperature reading, or the last set temperature, with the “MTC” indication.
The “MTC” indication and the symbol light up on the LCD to indicate that
the instrument is in MTC mode and the ARROW keys can be used to enter the
desired temperature value.
Note: When in MTC the user can press and hold the ARROW keys, and the
instrument will start incrementing /decrementing the temperature value.
The instrument keeps measuring and the display is updated periodically.
13
ORP MEASUREMENTS (not for HI 98180)
To perform ORP measurements, connect an optional ORP electrode (see “Accessories“
section) to the instrument and turn it ON.
If necessary, enter the mV mode by pressing RANGE
until the display changes to mV.
Submerge the ORP electrode tip (4cm/1½”) into the
sample to be tested and wait a few seconds for the
reading to stabilize.
Measurements are displayed with 0.1 mV resolution.
The “ATC” (or “MTC”) message is turned off because mV readings are not
temperature compensated.
For accurate ORP measurements, the surface of the electrode must be clean
and smooth. Pretreatment solutions are available to condition the electrode
and improve its response time (see “Accessories” section, page 65).
RELATIVE mV MEASUREMENTS (not for HI 98180)
To enter Relative mV mode, press “MODE” while in mV measurement mode. The
relative mV reading will be displayed along with the Absolute mV value and
the current temperature readings.
The relative mV reading is equal to the difference between the absolute mV input
value and relative mV offset established in the relative mV calibration.
Note: If using the pH electrode while in mV mode, the instrument will
measure the mV generated by the pH electrode.
ISE MEASUREMENTS (HI 98184, HI 98185 only)
To perform ion concentration measurements, connect an optional ISE electrode
and the corresponding reference (if necessary) to the instrument and turn it
ON.
14
Enter the ISE mode by pressing RANGE until the display changes to ISE
range.
Submerge the ISE electrode tip (4cm/1½”) into the sample
to be tested and wait a few seconds for the reading to
stabilize.
The ISE reading will be displayed along with the current
temperature reading.
The “ATC” (or “MTC”) message is turned off because ppm readings are not
temperature compensated.
In order to take accurate ISE measurements, make sure that the appropriate ISE
electrode type and ISE unit were set in SETUP menu (HI 98185 only), or the
proper ion charge is set (HI 98184 only), and the instrument was calibrated
(see ISE CALIBRATION for details, page 27).
Notes: • When the reading is out of range, the display will flash the closest
full-scale value.
• The instrument will display “----” on the primary LCD if it is not
calibrated. Perform at least a one-point calibration in order to
take ISE measurements.
• Changing the ISE electrode or the ion charge will need ISE range
calibration.
TEMPERATURE MEASUREMENTS
Connect the HI 7662 temperature probe (HI 98180/1/2) or the temperature
connector (HI 98183/4/5) to the appropriate socket. Immerse the temperature
probe (HI 98180/1/2) or pH electrode (HI 98183/4/5) into the sample and
allow the reading on the secondary LCD to stabilize.
Note: The temperature can be displayed in Celsius degrees (ºC) or in Fahrenheit
degrees (ºF) (see SETUP for details, page 33) (not for HI 98180).
BACKLIGHT FEATURE
The instrument is provided with a Backlight feature, which can be easily
toggled on and off through the keyboard by pressing LIGHT.
Note: The backlight automatically shuts off after a set period (see SETUP
for details, page 33) with no buttons pressed.
15
pp
H CALIBRATIONH CALIBRATION
p
H CALIBRATION
pp
H CALIBRATIONH CALIBRATION
It is recommended to calibrate the instrument frequently, especially if high
accuracy is required.
The pH range should be recalibrated:
• Whenever the pH electrode is replaced.
• At least once a week.
• After testing aggressive chemicals.
• When calibration alarm time out is expired - “CAL DUE” blinks (if feature
is enabled in SETUP).
• If “Outside Cal Range” message blinks during pH measurement (the
measurement range is not covered by current calibration, if feature is
enabled in SETUP).
PROCEDURE
All the familly instruments offers a choice of 7 memorized buffers (1.68, 4.01,
6.86, 7.01, 9.18, 10.01 and 12.45 pH). HI 98182/3/4/5 allow the user to set
up to 5 custom buffers. The set custom buffers are the buffer values at 25 ºC.
When a custom buffer is selected during calibration, the “Custom” functional
key is displayed on the LCD. Press Custom key in order to enter custom buffer
changing mode. Use ARROW keys to change the value in a ±1.00 pH
window, in according with the temperature reading and then Accept. Press
ESC to leave custom buffers value unchanged.
For accurate pH measurements, it is recommended to perform a calibration in
maximum allowed points. However, at least a two-point calibration is suggested.
The instrument will automatically skip the buffers used during calibration and
the buffers which are in a ±0.2 pH window, around one of the calibrated buffers.
• Pour small quantities of selected buffer solutions into clean beakers. For
accurate calibration use two beakers for each buffer solution, the first one for
rinsing the electrode and the second one for calibration.
• Remove the protective cap and rinse the electrode with some of the buffer
solution to be used for the first calibration point.
FIVE-POINT CALIBRATIONFIVE-POINT CALIBRATION
FIVE-POINT CALIBRATION
FIVE-POINT CALIBRATIONFIVE-POINT CALIBRATION
• Immerse the pH electrode and the temperature
probe approximately 4 cm (1½”) into a buffer
solution of your
9.18, 10.01, 12.45 or a custom buffer) and stir
gently. The temperature probe should be close to
the pH electrode.
choice (pH 1.68, 4.01, 6.86, 7.01,
16
• Press CAL. The instrument will display the measured pH, the LCD first
expected buffer and the temperature reading.
• If necessary, press the ARROW keys to select a different buffer value.
• The “ ” tag will blink on the LCD until the reading is stable.
• When the reading is stable and close to the selected buffer, CFM functional
key is displayed.
• Press CFM to confirm first point.
• The calibrated value and the second expected buffer value is then displayed on the LCD.
• After the first calibration point is confirmed, immerse the pH electrode and the
temperature probe approximately 4 cm (1½”) into the second buffer solution
and stir gently. The temperature probe should be close to the pH electrode.
• If necessary, press the ARROW keys to select a different buffer value.
• The “ ” tag will blink on the LCD until the reading is stable.
17
• When the reading is stable and close to the selected buffer, the CFM
functional key is displayed.
• Press CFM to confirm calibration.
• The calibrated value and the third expected buffer value will be displayed.
• After the second calibration point is confirmed, immerse the pH electrode
and the temperature probe approximately 4 cm (1½”) into the third
buffer solution and stir gently. The temperature probe should be close to
the pH electrode.
• If necessary, press the ARROW keys to select a different buffer value.
• The “ ” tag will blink on the LCD until the reading is stable.
• When the reading is stable and close to the selected buffer, the CFM
functional key is displayed.
• Press CFM to confirm calibration.
• The calibrated value and the fourth expected value will be displayed.
18
• After the third calibration point is confirmed, immerse the pH electrode
and the temperature probe approximately 4 cm (1½”) into the fourth
buffer solution and stir gently. The temperature probe should be close to
the pH electrode.
• If necessary, press the ARROW keys to select a different buffer value.
• The “ ” tag will blink on the LCD until the reading is stable.
• When the reading is stable and close to the selected buffer, the CFM
functional key is displayed.
• Press CFM to confirm calibration.
• The calibrated value and the
• After the fourth calibration point is confirmed, immerse the pH electrode and the
temperature probe approximately 4 cm (1½”) into the fifth buffer solution and
stir gently. The temperature probe should be close to the pH electrode.
• If necessary, press the ARROW keys to select a different buffer value.
• The “ ” tag will blink on the LCD until the reading is stable.
• When the reading is stable and close to the selected buffer, the CFM
functional key is displayed.
fifth
expected buffer will be displayed.
19
• Press CFM to confirm calibration.
• The instrument stores the calibration values and returns to normal measurement mode.
FOUR, THREE or TWO-POINT CALIBRATION
• Proceed as described in “FIVE-POINT CALIBRATION” section.
• Press CAL or ESC after the appropriate accepted calibration point. The instruments will return to measurement mode and will memorize the calibration data.
ONE-POINT CALIBRATION
Two SETUP selectable options are available for one-point calibration:
Replace and Offset.
If the Replace option is selected, the slopes between current buffer and
nearest lower and higher buffers will be reevaluated.
If the “Offset” option is selected, an electrode offset correction is performed
keeping unchanged the existing slopes.
• Proceed as described in “FIVE-POINT CALIBRATION” section.
• Press CAL or ESC after the first calibration point was confirmed. The
instruments will memorize the one-point calibration data and will
return to measurement mode.
Notes: • Press MTC or MODE key to toggle between pH buffer selection and
the temperature reading during calibration while temperature
probe is not connected (MTCmode)
• The displayed arrow is moving to the temperature value. Use
ARROW keys in order to change the temperature.
20
ERROR SCREENS
Wrong buffer
The calibration cannot be confirmed.
The pH reading is not close to the selected buffer. Select another buffer using
the ARROW keys or change the buffer.
Electrode Dirty/Broken alternatively with Buffer Contaminated
The calibration cannot be confirmed.
The offset of the electrode is not in the accepted range. Check if the electrode
is broken or clean it following the Cleaning Procedure (see page 65). Check the
quality of the buffer. If necessary, change the buffer.
Wrong slope
The calibration cannot be confirmed.
The evaluated slope is less than the lowest accepted value (80% of default
slope).
The evaluated slope is more than the highest accepted value (110 % of default
slope).
21
Wrong old slope
An inconsistency between new and previous (old) calibration is detected. Clear
old calibration parameters and proceed calibration from the current point. The
instrument will keep all confirmed values during current calibration.
Note: For one-point calibration the electrode condition is not displayed in the
measurement screen.
Each time a buffer is confirmed, the new calibration parameters replace the old
calibration parameters of the corresponding buffer.
If current confirmed buffer has no correspondence in the existing stored calibration
and this is not full, the current buffer is added to the existing stored calibration.
If the existing stored calibration is full (five calibration points), after confirming
the calibration point, the instrument will ask wich buffer will be replaced by
current buffer.
Press ARROW keys to select another buffer to be replaced.
Press CFM to confirm the buffer that will be replaced.
Press CAL or ESC to leave replace mode. In this case, the buffer will not be
memorised.
Note: The replaced buffer is not removed from calibration list and it can be
selected for the next calibration points.
22
Loading...
+ 50 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.