Hanna Instruments HI 981411-0 User Manual

Instruction Manual
HI 981411-0 HI 981411-1
Panel-Mounted
pH Indicators &
Controllers
Manufactors since 1978
WARRANTY
HI 981411-0 and HI 981411-1 are warranted for two years against defects in workmanship and materials when used for their intended purpose and maintained according to instructions.
This warranty is limited to repair or replacement free of charge. Damages due to accident, misuse, tampering or lack of prescribed
maintenance are not covered. If service is required, contact the dealer from whom you purchased the
instrument. If under warranty, report the model number, date of pur­chase, serial number and the nature of the failure. If the repair is not covered by the warranty, you will be notified of the charges incurred. If the instrument is to be returned to Hanna Instruments, first obtain a Returned Goods Authorization Number from the Customer Service de­partment and then send it with shipment costs prepaid. When shipping any instrument, make sure it is properly packaged for complete protec­tion.
To validate your warranty, fill out and return the enclosed warranty card within 14 days from the date of purchase.
All rights are reserved. Reproduction in whole or in part is prohibited without the written consent of the copyright owner, Hanna Instruments Inc., Woonsocket, Rhode Island, 02895 , USA.
These Instruments are in
Compliance with the CE Directives
Dear Customer, Thank you for choosing a Hanna product. This manual will
provide you with the necessary information for the correct operation of the meter. Please read it carefully before using the meter. If you need additional technical information, do not hesitate to e-mail us at tech@hannainst.com.
These instruments are in compliance with the .
PRELIMINARY EXAMINATION
Remove the instrument from the packing material and examine it carefully. If any damage has occurred during shipment, immediately notify your Dealer or the nearest Hanna Customer Service Center.
The meter is supplied with:
- Mounting brackets;
- Instruction manual. Note: Conserve all packing material until the instrument
has been observed to function correctly. Any defective item must be returned in its original packing.
GENERAL DESCRIPTION
HI 981411-0 and HI 981411-1 are pH indicators and controllers with a relay output designed for simplicity of use in a wide range of applications.
The models are panel-mounted with membrane keypads on the front panel and an easy-to-read LCD display.
Power supply, wiring and selection are made via the plug­in terminal blocks on the rear panel. The instruments are equipped with a BNC socket and accept input from conven­tional combination pH electrode.
LED indicators on the front panel identify whether the controller is in set or reading mode, and if dosing device is activated.
SPECIFICATIONS
HI 981411
Range 0.0 to 14.0 pH Resolution 0.1 pH Accuracy (@ 20°C/68°F) ±0.2 pH Offset Calibration Manual, through trimmer Setpoint Adjustable, from 0.0 to 14.0 pH
Dosing Selection Acid or Alkaline, selectable on the back panel
Contact Open=Acid dosage=Relay ON if Measure > Setpoint Contact Close=Alkaline dosage=Relay ON if Measure < Setpoint
Power supply: External
HI 981411-0 12 VDC HI 981411-1 115/230 VAC; 50/60Hz
Dimensions 79 x 49 x 95 mm (3.1 x 1.9 x 3.7”) Weight:
HI 981411-0 140 g (4.96 oz) HI 981411-1 250 g (8.85 oz)
CE DECLARATION
OF CONFORMITY
HI981411-0 and HI981411-1
14-4-99
Recommendations for Users Before using this product, make sure that it is entirely suitable for the environment in which it is used. Operation of this instrument in residential areas could cause unacceptable interferences to radio and TV equipment. The glass bulb at the end of the electrode is sensitive to electrostatic dis­charges. Avoid touching this glass bulb at all times. During operation, ESD wrist straps should be worn to avoid possible damage to the electrode by electrostatic discharges. Any variation introduced by the user to the supplied equipment may degrade the instrument’s EMC performance. To avoid electrical shock, do not use this instrument when voltages at the measurement surface exceed 24 VAC or 60 VDC. To avoid damages or burns, do not perform any measurement in microwave ovens.
ISTR981411R5 06/2001
http://www.hannainst.com
HI 981411
pH CONTROLLER
MEAS
SET
DOSING
pH
0.0 ...14.0 pH
OPERATIONAL GUIDE
95mm MIN
3.74"
76mm
3"
0.25/4mm
0.01/0.160"
79mm
3.11"
ADJUSTABLE LOCATION BRACKET
POWER CONNECTION HI 981411-0
Connect a 2-wire power cable to the terminal strip while paying attention to the correct positive and negative con­tacts (12VDC).
HI 981411-1
Connect a 3-wire power cable to the terminal strip while paying attention to the correct earth, neutral and line contacts (115/230 VAC).
Notes:
• All external cables connected to the rear panel should end with wire lugs.
• It is recommended to cover the unused terminals with insulating tape.
• At start up the meter needs a few seconds to stabilize. Wait until a stable reading is displayed.
DOSING CONTACT
This contact (maximum 5A, 240VAC, 30VDC) is used for connection to an alarm or dosing system. The relay contact acts as a power switch for the controlling device.
OPERATING THE METER
All settings of the parameters can be controlled via front panel keys and trimmers. “SET” and “MEAS” LEDs light up to indicate which is the operating function.
Make sure that the meter is calibrated and that the Setpoint is properly adjusted before proceeding.
Attach the pH-electrode to the meter with a BNC connector. Install the electrode in the fittings or immerse it in the solution to be monitored. Press the "MEAS" key.
The LCD will show the pH value. Any initial variation on readings may be due to temperature compensation.
The “DOSING” LED will light up when the dosing contact is closed.
CALIBRATION
Make sure the meter is in the measurement mode (the “MEAS” LED lights on). Immerse the electrode in HI 7007 (pH7.01) buffer solution (or any buffer solution close to the operating pH value). Shake briefly and wait for reading to stabilize. Adjust the calibration trimmer until “7.0” is displayed on the LCD.
SETPOINT
Press the "SET" key. The display will show the default or previously adjusted value. Using a small screwdriver adjust
FUNCTIONAL DESCRIPTION
FRONT PANEL
Keypad
SET To display
setpoint
MEAS To display
measurement
Trimmers
CAL For pH Offset
Calibration
SET To adjust the
Setpoint
SET
LEDs
SET ON when LCD
displays the set value
MEAS ON when the LCD displays the measurement
value
DOSING ON when dosing contact is activated
REAR PANEL
1. Power Supply:
HI 981411-0 HI 981411-1
12VDC 115/230 VAC L1: Positive 230 VAC L2: Negative 115 VAC L3: ------ Neutral L4: ------ Protection Earth Note: the power input is internally protected by a 400 mA fuse.
2.Dosing Selection Contacts: Leave S1 and S2 open to select ACID dosage; Short S1 and S2 to select ALKALINE dosage.
3.The dosing (alarm) contact acts as a switch. The contact has to be protected outside by the user.
CAL
HI981411 LAYOUT
Front view of the panel-mounted unit
Side view of the panel-mounted unit
Supplied adjustable lo­cation brackets allow the controller to slide into the cutout and will hold the unit securely in place. 95mm (3.74") is the minimum amount of space required to install the controller.
Panel-mounted unit assembling view
42mm
1.65"
49mm
1.93"
73mm
2.87"
HI 740146
ACCESSORIES
pH CALIBRATION SOLUTIONS HI 7007M pH 7.01 Buffer Solution, 230mL bottle HI 7007L pH 7.01 Buffer Solution, 460mL bottle
ELECTRODE STORAGE SOLUTION HI 70300M Storage Solution, 230mL bottle
ELECTRODE CLEANING SOLUTIONS HI 7061M General Cleaning Solution, 230mL bottle HI 7061L Oil & Fat Cleaning Solution, 230mL bottle
79mm
3.11"
pH ELECTRODES
HI1002/5 Heavy-duty BNC-type double junction, plastic-
body with external thread and 5m (16.5’) cable
HI1090T Screw type, double junction, glass-body with
external thread
HI1110S Screw type, single junction, glass-body HI1210T Screw type, double junction, plastic-body with
external thread
HI2114P/2 BNC-type, double junction plastic-body with
2m (6.6’) cable
HI1210B/5 BNC-type cable, double junction, plastic-body,
with 5 m (16.5') cable
HI2910B/5 BNC-type, double junction, plastic-body, built-
in amplifier, with 5 m (16.5') cable
EXTENSION CABLES FOR SCREW-TYPE ELECTRODES TO BNC CONNECTOR
HI7855/5 Extension cable 5m (16.5') long HI7855/10 Extension cable 10m (33') long
OTHER ACCESSORIES
BL PUMPS Dosing Pumps with flow rate from 1.5 to 20 LPH HI6050 Submersible electrode holder, 60cm (24”)
HI 6051 Submersible electrode holder, 110cm (43”) HI 6054B Electrode Holders for in-line applications HI 6054T Electrode Holders for in-line applications HI 710005 12 VDC power adapter, US plug HI 710006 12 VDC power adapter, European plug HI 710012 12 VDC power adapter, Australian plug HI 710013 12 VDC power adapter,Southern Africa plug HI 710014 12 VDC power adapter, UK plug
HI731326 Calibration Screwdriver (20 pcs) HI 7871 Level Controller (min and max)
HI 7873 Level Controller (min, max and overflow)