Instruction Manual
(HI 981402, HI 981403)
On-line, Waterproof
pH meter with Alarm
http://www.hannainst.com
WARRANTYWARRANTY
WARRANTY
WARRANTYWARRANTY
HI 981402 and HI 981403 are warranted for two years against
defects in workmanship and materials when used for their intended
purpose and maintained according to instructions. The electrode is
warranted for a period of one year. This warranty is limited to
repair or replacement free of charge.
Damages due to accident, misuse, tampering or lack of prescribed
maintenance are not covered.
If service is required, contact the dealer from whom you purchased the
instrument. If under warranty, report the model number, date of purchase, serial number and the nature of the failure. If the repair is not
covered by the warranty, you will be notified of the charges incurred. If
the instrument is to be returned to Hanna Instruments, first obtain a
Returned Goods Authorization Number from the Customer Service department and then send it with shipment costs prepaid. When shipping
any instrument, make sure it is properly packaged for complete protection.
To validate your warranty, fill out and return the enclosed warranty card
within 14 days from the date of purchase.
All rights are reserved. Reproduction in whole or in part is prohibited
without the written consent of the copyright owner, Hanna Instruments Inc., Woonsocket, Rhode Island, 02895 , USA.
Hanna Instruments reserves the right to modify the design, construction
and appearance of its products without advance notice.
These Instruments are in
Compliance with the CE Directives
Dear Customer,
Thank you for choosing a Hanna product. This manual will
provide you with the necessary information for the correct
operation of the meter. Please read it carefully before using
the meter. If you need additional technical information, do
not hesitate to e-mail us at tech@hannainst.com.
These instruments are in compliance with the
EN 50081-1 and EN 50082-1.
PRELIMINARY EXAMINATIONPRELIMINARY EXAMINATION
PRELIMINARY EXAMINATION
PRELIMINARY EXAMINATIONPRELIMINARY EXAMINATION
Remove the instrument from the packing material and
examine it carefully. If any damage has occurred during
shipment, immediately notify your Dealer or the nearest
Hanna Customer Service Center.
The meter is supplied with:
• HI 1286 pH electrode for HI981402;
• HI 2114P/2 pH electrode for HI981403;
• Calibration screwdriver;
• 12 VDC power adapter.
Note:Conserve all packing material until the instrument has
been observed to function correctly. Any defective item
must be returned in its original packing.
GENERAL DESCRIPTIONGENERAL DESCRIPTION
GENERAL DESCRIPTION
GENERAL DESCRIPTIONGENERAL DESCRIPTION
HI 981402 and HI 981403 are pH meters specially designed
to meet the needs of simple continuous monitoring of pH.
The housing has been completely sealed against vapors and
humidity with IP54 rating.
You can simply hang the meter right above the sample to be
tested for continuous measurement.
The HI 1286 and HI 2114P/2 gel-filled pH electrodes are
interchangeable and the BNC connector is protected behind
a waterproof sheath. The unique design of the electrode
provides longer life even in aggressive solutions.
You can even select your own setpoint and be alerted of an
abnormal situation with a flashing LED alarm.
Measurements are highly accurate and the meter can be
calibrated at one or two points.
You no longer need to worry about battery changes either:
the unit runs without interruption on 12 VDC power supply.
directives
FUNCTIONAL DESCRIPTIONFUNCTIONAL DESCRIPTION
FUNCTIONAL DESCRIPTION
FUNCTIONAL DESCRIPTIONFUNCTIONAL DESCRIPTION
1. Molded eye
2. Alarm LED
3. pH 4.0 calibration trimmer
4. BNC connector
5. HI 1286 or HI 2114P/2 pH electrode
6. 12 VDC power adapter
7. Power supply connector
8. Protective sheath
9. pH 7.0 calibration trimmer
10. Liquid Crystal Display
SPECIFICATIONSSPECIFICATIONS
SPECIFICATIONS
SPECIFICATIONSSPECIFICATIONS
HI 981402
Electrode HI 1286 interchangeable pH electrode
Setpoint 3.0 to 11.0 pH
Hysteresis ±0.5 pH around setpoint
Alarm LED blinks when pH is outside hysteresis range
HI 981403
Electrode HI 2114P/2 interchangeable pH electrode
Setpoint 6.0 to 9.5 pH
Hysteresis 0.5 pH below setpoint
Alarm LED starts blinking when pH is over setpoint
COMMON SPECIFICATIONS
Range 0.0 to 14.0 pH
Resolution 0.1 pH
Accuracy (@ 25°C/77°F) ±0.2 pH
Typical EMC Deviation ±0.2 pH
Calibration Manual with two trimmers
for offset and slope
Casing IP54
Power supply External 12 VDC (included)
Dimensions 86 x 94 x 33 mm (3.4 x 3.7 x 1.3”)
Weight 150 g (5.3 oz.)
OPERATIONAL GUIDEOPERATIONAL GUIDE
OPERATIONAL GUIDE
OPERATIONAL GUIDEOPERATIONAL GUIDE
pH ELECTRODE CONNECTION & MAINTENANCE
In order to protect the instrument against
vapors and humidity, the BNC connector is
shielded behind a waterproof sheath.
• Slide the protective sheath down. Connect the pH
electrode to the BNC connector and then slide the
protective sheath back up all the way up to the casing.
Since the protective sheath is rubberized to ensure
maximum waterproof protection, make sure the connector is completely covered.
• Do not be alarmed if white crystals appear around the
electrode protective cap. This is normal with pH electrodes and they dissolve when rinsed with water.
• When not in use, rinse the electrode with water to
minimize contamination and store it with a few drops of
storage (HI 70300) or pH 7 (HI 7007) solution in the
protective cap. Always replace the protective cap after use.
DO NOT USE DISTILLED OR DEIONIZED WATER FOR
STORAGE PURPOSES.
• If the electrode has been left dry, soak the tip in a
storage (HI 70300) or pH 7 (HI 7007) solution for at
least one hour to reactivate it.
• To minimize clogging and provide longer life for the pH
electrode, it is recommended to clean it monthly. Immerse the tip of the electrode in HI 7061 for one hour
and then rinse it with tap water.
TAKING pH MEASUREMENTS
• Turn the meter on by connecting the
12 VDC power adapter to the meter
and to the mains.
• Remove the protective cap from the
pH electrode and immerse the tip
(4 cm/1½”) of pH electrode in the
sample.
• The LCD will show the pH value. Allow the
reading to stabilize and the instrument
will start continuous monitoring.
Note: to prevent damages to the electrode, remove the pH
electrode from the sample before turning the meter off.
ADJUSTING THE SETPOINT
With HI 981402 and HI 981403 you can select your own
setpoint and be alerted with a visual LED alarm when an
abnormal situation arises.
• Unscrew and remove the rear panel and gasket seal to
access the MEASURE/SET switch and the setpoint adjustment trimmer.
• Move the switch to the left (SET).
• With a small screwdriver adjust the setpoint trimmer to
display the desired value in the range 3 to 11 pH (HI
981402) or 6 to 9.5 pH (HI 981403).
• Make sure the switch is moved back to the
right (MEASURE Mode).
• Replace the rear panel and the gasket, ensuring the
unit is properly closed.
• In HI 981402, whenever the pH
reading varies by more than ±0.5
pH from the setpoint, the red alarm
LED blinks to warn the user.
In HI 981403 the alarm LED
starts blinking when the pH reading becomes greater than setpoint
and stops blinking only when
the reading becomes 0.5 pH lower
than setpoint.
CALIBRATIONCALIBRATION
CALIBRATION
CALIBRATIONCALIBRATION
For the greatest accuracy, frequent calibration of the instrument is recommended. In addition, the instrument must be
recalibrated whenever:
a) The pH electrode is replaced.
b) After testing aggressive chemicals.
c) Where extreme accuracy is required.
d) At least once a month.
PREPARATION
4
0
0
I 7
H
Pour small quantities of pH
7.0 (HI 7007) and pH 4.0
(HI 7004) solution into two
HI 7004 HI 7007HI 7004 HI 7007
clean beakers.
For accurate calibration use two beakers for each buffer
solution, the first one for rinsing the tip of the electrode and
the second one for calibration. This way, contamination of
the buffers is minimized.
RINSE
HI 7007
CALIBRATION
HI 7007
pH 4.0 (HI 7004) is recommended for measuring acidic
samples. Use pH 10.0 (HI 7010) if subsequent samples are
alkaline.
CALIBRATION PROCEDURE
• Turn the meter on and make sure that
the MEASURE/SET switch is on the MEASURE mode.
• Remove the protective cap from the
electrode, rinse and immerse it in a pH
7.0 buffer. Stir gently and then wait a
couple of minutes for the reading to
stabilize.
Note: the electrode should be sub-
merged approximately 4 cm
(1½”) in the solution.
• Adjust the right hand trimmer
with the calibration screwdriver
until the LCD shows pH 7.0.
• Rinse and immerse the pH electrode in
pH 4.0 (or pH 10.0) buffer and stir
gently.
• Wait a couple of minutes and
then adjust the left hand trimmer until the LCD shows the
value of the second buffer.
The pH calibration is now complete.
(1½
")
ACCESSORIESACCESSORIES
ACCESSORIES
ACCESSORIESACCESSORIES
HI 1286 Double junction, plastic body pH electrode
with 2 m (6.6’) cable and BNC connector
HI 2114P/2 Double junction, plastic body pH electrode
with 2 m (6.6’) cable and BNC connector
HI 70004P pH 4.01 solution, 20 mL sachet (25 pcs)
HI 70007P pH 7.01 solution, 20 mL sachet (25 pcs)
HI 7004L pH 4.01 solution, 460 mL bottle
HI 7007L pH 7.01 solution, 460 mL bottle
HI 7061L Electrode cleaning solution, 460 mL bottle
HI 710005 12 VDC power adapter, US plug
HI 710006 12 VDC power adapter, European plug
HI 710012 12 VDC power adapter, Australian plug
HI 710013 12 VDC power adapter, Southern Africa plug
HI 710014 12 VDC power adapter, UK plug
4 cm
HI 77400P pH 4 & 7 solutions, 20 mL sachet (5 each)
CE DECLARATIONCE DECLARATION
CE DECLARATION
CE DECLARATIONCE DECLARATION
OF CONFORMITYOF CONFORMITY
OF CONFORMITY
OF CONFORMITYOF CONFORMITY
Recommendations for Users
Before using this product, make sure that it is entirely suitable for the environment in
which it is used. Operation of this instrument in residential areas could cause unacceptable interferences to radio and TV equipment.
The glass bulb at the end of the electrode is sensitive to electrostatic discharges. Avoid touching
this glass bulb at all times. During operation, ESD wrist straps should be worn to avoid possible
damage to the electrode by electrostatic discharges.
Any variation introduced by the user to the supplied equipment may degrade the instrument’s
EMC performance.
To avoid electrical shock, do not use this instrument when voltages at the measurement surface
exceed 24 VAC or 60 VDC. To avoid damages or burns, do not perform any measurement in
microwave ovens.
ISTR9814R1 07/98PRINTED IN PORTUGAL