Hanna Instruments HI 38077 User Manual

Instruction Manual
HI 38077
Phosphate Test Kit for
Irrigation Water
www.hannainst.com
Dear Customer, Thank you for choosing a Hanna Product. Please read the instruction sheet carefully before using the
test kit. It will provide you with the necessary information for correct use of the kit. If you need additional information, do not hesitate to e-mail us at tech@hannainst.com.
Each kit is supplied with:
HI 93713-0 Reagent, packets (100 pcs);
• 1 checker disc (containing the 38073 disc);
• 2 glass vials with caps; 1 plastic pipette (3 mL).
Note: Any damaged or defective item must be returned in
its original packing materials.
SPECIFICATIONSSPECIFICATIONS
SPECIFICATIONS
SPECIFICATIONSSPECIFICATIONS
Range 0 to 5 mg/L (ppm) Phosphate (PO Smallest Increment 0.1 mg/L (ppm) Phosphate Analysis Method Colorimetric Sample Size 7.5 mL Number of Tests 100 Case Dimensions 235x175x115 mm (9.2x6.9x4.5") Shipping Weight 429 g (15.1 oz.)
SIGNIFICANCE AND USESIGNIFICANCE AND USE
SIGNIFICANCE AND USE
SIGNIFICANCE AND USESIGNIFICANCE AND USE
Phosphorus occurs in natural waters and in wastewaters almost entirely as phosphates. Large quantity of phosphate arise from laundering industries as it is used in many cleaning prepara­tions, from soil runoff and sewage.
Phosphorus is essential to plants since it contributes to the formation of buds, roots and blooming as well as lignification and its lack results in stifled plants or pale green color with reddish pigmentation on leaves' edges.
On the other hand, an extensive discharge of phosphorus in water is the major cause of eutrophication, which is an abnormal and excessive growth of aquatic plants.
Note:
mg/L is equivalent to ppm (parts per million).
CHEMICAL REACTIONCHEMICAL REACTION
CHEMICAL REACTION
CHEMICAL REACTIONCHEMICAL REACTION
Adaptation of the ascorbic acid method. The reaction be­tween phosphate and the reagent causes a blue tint in the sample.
ISTR38077 02/00 PRINTED IN ITALY
INSTRUCTIONSINSTRUCTIONS
INSTRUCTIONS
INSTRUCTIONSINSTRUCTIONS
3-
)
4
READ THE ENTIRE INSTRUCTIONS BEFORE USING THE KIT
• Using the pipette, add to each glass vial 7.5 mL of sample (up to the mark).
• Insert one of them into the left hand opening of the checker disc. This is the blank.
• Add to the other glass vial 1 packet of HI 93713-0 Reagent. Replace the cap and shake vigorously for 20 seconds. This is the reacted sample.
• Wait for 3 minutes.
• Remove the cap and insert the reacted sample into the right hand opening of the checker disc.
• Hold the checker disc so that a light source illuminates the samples from the back of the windows.
• Keep the checker disc at a distance of 30-40 cm (12­16") from the eyes to match the color. Rotate the disc while looking at the color test windows and stop when you find the color match. Read the value in the result window, divide it by 10 and record it in mg/L (or ppm)
7.5 mL
of Phosphate. To convert this result to mg/L (ppm) of Phosphorus, divide it by 3,1.
For best results: Perform the reading three times and take
the average value (divide by 3 the sum of the three numbers). Intensely colored samples will make the color matching determination difficult and they should be adequately treated before performing the test. Sus­pended matter in large amounts should be removed by prior filtration.
Caution: Ultraviolet radiation may cause fading of colors.
When not in use, keep the disc protected from light, in a cool and dry place.
Interferences: iron above 5 ppm; copper above 10 ppm;
hydrogen sulfide, arsenate and highly buffered samples.
REFERENCESREFERENCES
REFERENCES
REFERENCESREFERENCES
APHA, AWWA, WEF, Standard Methods for the Examina­tion of Water and Wastewater, 18
HEALTH AND SAFETYHEALTH AND SAFETY
HEALTH AND SAFETY
HEALTH AND SAFETYHEALTH AND SAFETY
The chemicals contained in this kit may be hazardous if improperly handled. Read the relevant Health and Safety Data Sheet before performing this test.