Hanna Instruments HI 38054 User Manual

Instruction Manual
HI 38054
Ozone
Test Kit
www.hannainst.com
Dear Customer, Thank you for choosing a Hanna Product. Please read the instruction sheet carefully before using the
test kit. It will provide you with the necessary information for correct use of the kit. If you need additional information, do not hesitate to e-mail us at tech@hannainst.com.
Each kit is supplied with:
HI 93711-0 Total Chlorine Reagent, packets (100 pcs);
• Deionized Water, 1 bottle (500 mL);
• 1 checker disc (containing the 38054 disc);
• 2 glass vials with caps;
• 1 plastic pipette (3 mL).
Note: Any damaged or defective item must be returned in
its original packing materials.
SPECIFICATIONSSPECIFICATIONS
SPECIFICATIONS
SPECIFICATIONSSPECIFICATIONS
Range 0 to 2.3 mg/L (ppm) as Ozone Smallest Increment 0.1 mg/L Analysis Method Colorimetric Sample Size 5 mL Number of Tests 100 Case Dimensions 235x175x115 mm (9.2x6.9x4.5") Shipping Weight 966 g (34.1 oz)
SIGNIFICANCE AND USESIGNIFICANCE AND USE
SIGNIFICANCE AND USE
SIGNIFICANCE AND USESIGNIFICANCE AND USE
Ozone is an oxidizing agent and a germicide. It is used for oxidation of organic matter, which produces color or odor in drinking water.
Note: mg/L is equivalent to ppm (parts per million).
CHEMICAL REACTIONCHEMICAL REACTION
CHEMICAL REACTION
CHEMICAL REACTIONCHEMICAL REACTION
The reaction between ozone and the reagent causes a pink tint in the sample which is proportional to the ozone concentration.
ISTR38054 09/00 PRINTED IN ITALY
INSTRUCTIONSINSTRUCTIONS
INSTRUCTIONS
INSTRUCTIONSINSTRUCTIONS
READ THE ENTIRE INSTRUCTIONS BEFORE USING THE KIT
• Use the pipette to fill each glass vial with 5 mL of sample (up to the mark).
5 mL
• Insert one of them into the left hand opening of the checker disc. This is the blank.
• Add to the other vial deionized water up to the 10 mL mark. Replace the cap and shake to mix.
• Remove the cap and add 1 packet of HI 93711-0 reagent. Replace the cap and mix.
• Wait for 2 minutes to allow reaction to occur. This is the reacted sample.
• Remove the cap and insert the reacted sample into the right hand opening of the checker disc.
• Hold the checker disc so that a light source illuminates the samples from the back of the windows.
• Keep the checker disc at a distance of 30-40 cm (12­16") to match the color. Rotate the disc while looking at the color test windows and stop when you find the color match. Read the value in the result window directly in mg/L (or ppm) of Ozone.
For best results: Perform the reading three times and
take the average value (divide by 3 the sum of the three numbers). Intensely colored samples will make the color matching difficult and they should be adequately treated before performing the test. Sus­pended matter in large amounts should be removed by prior filtration.
Caution: Ultraviolet radiation may cause fading of colors.
When not in use, keep the disc protected from light, in a cool and dry place.
Interferences: chlorine, bromine, iodine and oxidized man-
ganese.
REFERENCESREFERENCES
REFERENCES
REFERENCESREFERENCES
Adaptation of the EPA recommended DPD method 330.5.
HEALTH AND SAFETYHEALTH AND SAFETY
HEALTH AND SAFETY
HEALTH AND SAFETYHEALTH AND SAFETY
The chemicals contained in this kit may be hazardous if improperly handled. Read the relevant Health and Safety Data Sheet before performing this test.