Hanna Instruments HI 112, HI 111, HI 113 Instruction Manual

1
pH/mV/°C
Bench Meters
with Custom Buffers
HI 110 HI 111 HI 112 HI 113
Instruction Manual
MR
MEM
SET
C
C
RANGE
GLP
CFM
CAL
pH / ORP Meter
HI 113
Select Setup Item
Select pH/ORP
RANGE
SET
Set Parameters
GLP
Show GLPData
MEM
Store Reading
MR
MemoryRecall
Functions
Confirm Calibration Point
CFM
Select pH Buffer
/
Enter/Exit Calibration
CAL
pH Calibration
www.hannainst.com
2
WARRANTY
HI 110, HI 111, HI 112 and HI 113 are guaranteed for two years against defects in workmanship and materials when used for their intended purpose and maintained according to instructions. Electrodes and probes are guaranteed for six months. This warranty is limited to repair or replacement free of charge.
Damage due to accidents, misuse, tampering or lack of prescribed maintenance is not covered.
If service is required, contact the dealer from whom you purchased the instrument. If under warranty, report the model number, date of purchase, serial number and the nature of the problem. If the repair is not covered by the warranty, you will be notified of the charges incurred. If the instrument is to be returned to Hanna Instruments, first obtain a Returned Goods Authori­zation number from the Technical Service department and then send it with shipping costs prepaid. When shipping any instrument, make sure it is properly packed for complete protection.
Dear Customer, Thank you for choosing a Hanna Instruments product. Please read this instruction manual carefully before using the instruments. This manual will provide you with the necessary information for correct use of
the instruments, as well as a precise idea of their versatility. If you need additional technical information, do not hesitate to e-mail us at
tech@hannainst.com or turn to the back cover for our worldwide contact list. These instruments are in compliance with directives.
TABLE OF CONTENTS
WARRANTY ......................................................................................................................... 2
PRELIMINARY EXAMINATION .............................................................................................. 3
GENERAL DESCRIPTION ..................................................................................................... 3
FUNCTIONAL DESCRIPTION HI 110 AND HI 111 ............................................................. 4
HI 110 AND HI 111 SPECIFICATIONS .............................................................................. 5
FUNCTIONAL DESCRIPTION HI 112 AND HI 113 ............................................................. 6
HI 112 AND HI 113 SPECIFICATIONS .............................................................................. 7
OPERATIONAL GUIDE ......................................................................................................... 8
pH CALIBRATION ................................................................................................................ 10
GOOD LABORATORY PRACTICE (GLP) ...................................................................................15
SETUP .....................................................................................................................................17
TEMPERATURE CALIBRATION (for technical personnel only) ........................................... 21
mV CALIBRATION (for technical personnel only, HI 111 AND HI 113) ........................... 23
RELATIVE mV CALIBRATION (HI 111 AND HI 113) ....................................................... 24
PC INTERFACE (HI 112 AND HI 113) ............................................................................ 25
pH BUFFER TEMPERATURE DEPENDENCE ..................................................................... 27
ELECTRODE CONDITIONING & MAINTENANCE .................................................................. 28
TROUBLESHOOTING GUIDE ............................................................................................... 31
TEMPERATURE CORRELATION FOR pH SENSITIVE GLASS ............................................... 32
ACCESSORIES ...................................................................................................................... 33
3
PRELIMINARY EXAMINATION
Remove the instrument from the packing material and examine it carefully to make sure that no damage has occurred during shipping. If there is any damage, notify your Dealer or the nearest Hanna Customer Service Center.
Each instrument is supplied with:
HI 1131B Glass-body Combination pH Electrode with 1 m (3.3') Cable  HI 7669/2W Temperature Probe  HI 76404 Electrode Holder  pH 4.01 & 7.01 Buffer Solutions (20 mL each)HI 7071S Electrolyte Solution  12VDC Power Adapter  Instruction Manual
Note: Save all packing material until you are sure that the instrument
functions correctly. All defective items must be returned in the original packing with the supplied accessories.
The HANNA HI 110, HI 111, HI 112 and HI 113 are microprocessor-based pH and Temperature bench meters.
HI 111 and HI 113 can also measure Oxidation Reduction Potential (ORP) in mV range. Relative mV feature is also provided.
pH measurements are compensated for temperature effect manually or automatically with the HI 7669/2W temperature probe.
Up to three-point (HI 110, HI 111) or up to five-point (HI 112, HI 113) calibration is possible. In addition, two custom buffers can be set and used during calibration.
The GLP feature provides a guarantee of data consistency. An alarm time-out is available to alert the user that too much time elapsed
since the last pH calibration. Through the RS232 serial port data can be transferred to a PC (HI 112, HI 113).
Hanna Instruments reserves the right to modify the design,
construction and appearance of its products without advance notice.
GENERAL DESCRIPTION
4
1) Liquid Crystal Display (LCD).
2) CFM/GLP key, to confirm different values or to display Good Laboratory
Practice information.
3) CAL key, to enter and exit/escape calibration mode.
4) /ºC key, to manually decrease temperature value or other parameters.
5) /ºC key, to manually increase temperature value or other parameters.
6) MR key, to recall the stored value.
7) MEM key, to store a value into memory.
8) SET key, to enter/exit SETUP mode.
9) RANGE key, to select measurement range or switch to focused data in SETUP (HI 111 only).
SEL key, to switch to focused data in SETUP menu (HI 110 only).
10) ON/OFF switch.
11) Power supply socket.
12) BNC electrode connector.
13) Electrode reference socket.
14) Temperature probe socket.
FUNCTIONAL DESCRIPTION
HI 110 AND HI 111
1
3
2
7
4
6
5
Front Panel
Primary LCD
Secondary LCD
POWER 12VDC
INPUT
REF
TEMP
11
12 13 14
Rear Panel
9
10
8
pH
MR
MEM
SET
C
C
RANGE
GLP
CFM
CAL
pH / ORP Meter
HI 111
Select Setup Item
Select pH/ORP
RANGE
SET
Set Parameters
GLP
Show GLPData
MEM
Store Reading
MR
MemoryRecall
Functions
Confirm Calibration Point
CFM
Select pH Buffer
/
Enter/Exit Calibration
CAL
pH Calibration
5
HI 110 AND HI 111
SPECIFICATIONS
EGNAR
Hp00.61ot00.2
)ylno111IH,PRO&ESI(Vm9.996±
)ylno111IH,PRO&ESI(Vm0002±
Cº0.021ot9.9
NOITULOSER
Hp10.0
)ylno111IH,PRO&ESI(Vm1.0
)ylno111IH,PRO&ESI(Vm1
Cº1.0
YCARUCCA
F°86/C°02@
Hp10.0±
)Vm9.996±(Vm2.0±
)Vm0002±(Vm1±
)rorreeborpgnidulcxe(Cº4.0±
egnartesffoVmleR )ylno111IH(Vm0002±
noitarbilaCHp
,noitarbilactniop3ro2,1 elbaliavasreffubdradnats5
,)10.01,81.9,10.7,68.6,10.4(
sreffubmotsuc2dna
erutarepmeT
noitasnepmoc
:morfcitamotuArolaunaM
)Fº0.842ot2.41(Cº0.021ot9.9
edortcelEHp B1311IH
eborperutarepmeT W2/9667IH
ecnadepmitupnI 01
21
smho
ylppusrewoP retpadaCDV21
snoisnemiD )9.2x1.7x4.9(mm47x281x042
thgieW
;)bl5.2(gK1.1
)bl5.5(gK5.2redlohhtiwtik
tnemnorivnE
)Fº22123(Cº050
gnisnednoc-nonHR%59.xam
ytnarraW sraey2
6
1) Liquid Crystal Display (LCD).
2) CFM/GLP key, to confirm different values or to display Good Laboratory Practice information.
3) CAL key, to enter and exit/escape calibration mode.
4) /ºC key, to manually decrease temperature value or other parameters.
5) /ºC key, to manually increase temperature value or other parameters.
6) MR key, to recall the stored value.
7) MEM key, to store a value into memory.
8) SET key, to enter/exit SETUP mode.
9) RANGE key, to select measurement range or switch to focused data in SETUP.
10) ON/OFF switch.
11) Power supply socket.
12) BNC electrode connector.
13) Electrode reference socket.
14) Temperature probe socket.
15) RS232 serial communication connector.
FUNCTIONAL DESCRIPTION
HI 112 AND HI 113
1
9
3
2
7
4
6
5
10
Front Panel
Primary LCD
Secondary LCD
RS232
15
POWER 12VDC
INPUT
REF
TEMP
11
12 13 14
Rear Panel
8
pH
MR
MEM
SET
C
C
RANGE
GLP
CFM
CAL
pH / ORP Meter
HI 113
Select Setup Item
Select pH/ORP
RANGE
SET
Set Parameters
GLP
Show GLPData
MEM
Store Reading
MR
MemoryRecall
Functions
Confirm Calibration Point
CFM
Select pH Buffer
/
Enter/Exit Calibration
CAL
pH Calibration
7
HI 112 AND HI 113
SPECIFICATIONS
EGNAR
Hp00.61ot00.2
Hp000.61ot000.2
)ylno311IH,PRO&ESI(Vm9.999±
)ylno311IH,PRO&ESI(Vm0002±
Cº0.021ot9.9
NOITULOSER
Hp10.0
Hp100.0
)ylno311IH,PRO&ESI(Vm1.0
)ylno311IH,PRO&ESI(Vm1
Cº1.0
YCARUCCA
F°86/C°02@
Hp10.0±
Hp200.0±
)Vm9.999±(Vm5.0±
)Vm0002±(Vm1±
)rorreeborpgnidulcxe(Cº4.0±
egnartesffoVmleR )ylno311IH(Vm0002±
ecafretniretupmoC 232SRdetalosi-otpo
noitarbilaCHp
,stniopnoitarbilac5otpU
elbaliavasreffubdradnats7
)54.21,10.01,81.9,10.7,68.6,10.4,86.1( ,
sreffubmotsuc2dna
erutarepmeT
noitasnepmoc
:morfcitamotuArolaunaM
)Fº0.842ot2.41(Cº0.021ot9.9
edortcelEHp B1311IH
eborperutarepmeT W2/9667IH
ecnadepmitupnI 01
21
smho
ylppusrewoP retpadaCDV21
snoisnemiD )9.2x1.7x4.9(mm47x281x042
thgieW
;)bl5.2(gK1.1
)bl5.5(gK5.2redlohhtiwtik
tnemnorivnE
)Fº22123(Cº050
gnisnednoc-nonHR%59.xam
ytnarraW sraey2
8
OPERATIONAL GUIDE
POWER CONNECTION
Plug the 12 VDC adapter into the power supply socket. Note: These instruments use non volatile memory to retain the pH, mV,
temperature calibrations and all other settings, even when unplugged.
Note: Make sure a fuse protects the main line.
ELECTRODE AND PROBE CONNECTIONS
For pH or ORP measurements connect an electrode with internal reference to the BNC connector on the back of the instrument.
For electrodes with a separate reference connect the electrodes BNC to the BNC connector and the reference electrode plug to the reference socket.
For temperature measurements and automatic temperature compensation connect the temperature probe to the appropriate socket.
INSTRUMENT START-UP
 Turn the instrument on by pressing the ON/OFF switch.  All LCD tags are displayed and a beep is sounded while
the instruments perform a self test.
HI 110 and HI 111
HI 112 and HI 113
pH MEASUREMENTS
Make sure the instrument has been calibrated before taking pH measurements.
 Submerge the electrode tip and the temperature probe
approximately 4 cm (1½) into the sample to be tested and stir gently. Allow time for the electrode to stabilize.
 The pH is displayed on the primary LCD and the temperature on the
secondary LCD.
 If the reading is out of range, the closest full-scale value will be displayed
blinking on the primary LCD.
pH
BUFFER pH
INTVDATETIME%
3
2
1
BUF
WRONG
READY
CFM
MEM
pHmVCAL
BUFFER pH
INTVDATETIME%
BUF
WRONG
READY
CFM
MEM
pHmVCAL
9
If measurements are taken successively in different samples, it is recommended to rinse the electrode thoroughly with deionized water or tap water and then with some of the next sample to prevent cross-contamination.
The pH reading is affected by temperature. In order to measure the pH accurately, the temperature effect must be compensated for. To use the Automatic Temperature Compensation feature, connect and submerge the
HI 7669/2W temperature probe into the sample as close as possible to the electrode and wait for a few seconds.
If the temperature of the sample is known, simple manual compensation can be performed by disconnecting the temperature probe.
The display will then show the default temperature of 25 ºC or the last temperature reading with the ºC tag blinking.
The temperature can now be adjusted with the ARROW keys (from 9.9 ºC to 120.0 ºC).
mV/ORP MEASUREMENTS (HI 111 and HI 113)
An optional ORP electrode must be used to perform ORP measurements (see Accessories).
Oxidation-Reduction Potential (REDOX) measurements provide the quantification of the oxidizing or reducing power of the tested sample.
To correctly perform a REDOX measurement, the surface of the ORP electrode must be clean and smooth.
 Press RANGE to enter mV range.  Submerge the tip of the ORP electrode 4 cm (1½) into
the sample to be tested and allow a few seconds for the reading to stabilize.
 The instrument displays the mV reading on the primary LCD and the
temperature on the secondary LCD.
 If the reading is out of range, the closest full-scale value will be displayed
blinking on the primary LCD.
C C
mV
mV
10
RELATIVE mV MEASUREMENTS (HI 111 and HI 113)
 Press RANGE until rEL message will be displayed on the secondary LCD line
for about one second and mV tag will blink. After one second the temperature will be displayed on the secondary LCD.
The reading displayed by the instrument is equal to the difference between the current mV input value and relative mV offset established in the relative mV calibration.
TEMPERATURE MEASUREMENTS
Connect the HI 7669/2W temperature probe to the appropriate socket and turn the instrument on.
Immerse the temperature probe into the sample and allow the reading on the secondary LCD to stabilize.
MEMORY FUNCTION
Press and hold down MEM. The last reading will appear on the LCD along with MEM tag until MEM is released.
Press MR and the previously memorized reading will be displayed on the LCD with MEM tag.
pH CALIBRATION
Calibrate the instrument frequently, especially if high accuracy is required. The instrument should be recalibrated:  Whenever the pH electrode is replaced.  At least once a week.  After testing aggressive chemicals.  If CAL INTV tags are blinking during measurement.
Every time you calibrate the instrument use fresh buffers and perform an electrode Cleaning Procedure (see page 30).
MR
mV
MEM
pH
MEM
11
PREPARATION
Pour small quantities of the buffer solutions into clean beakers. If possible, use plastic or glass beakers to minimize any EMC interferences.
For accurate calibration and to minimize cross-contamination, use two beakers for each buffer solution. One for rinsing the electrode and one for calibration.
If you are measuring in the acidic range, use pH 7.01 or 6.86 as first buffer and pH 4.01 or 1.68 (HI 112, HI 113) as second buffer. If you are measuring in the alkaline range, use pH 7.01 or 6.86 as first buffer and pH 10.01, 9.18 or 12.45 (HI 112, HI 113) as second buffer.
PROCEDURE
Calibration can be performed in three (HI 110, HI 111) or five (HI 112, HI 113 ) calibration points.
For accurate measurements, it is recommended to perform a three (HI 110, HI
111) or five-point calibration (HI 112, HI 113). However, at least a two-point
calibration is suggested. The calibration buffer can be selected from the calibration buffer list that
includes the defined custom buffers in SETUP menu and the memorized standard buffers:
 pH 4.01, 6.86, 7.01, 9.18, 10.01 (HI 110, HI 111)  pH 1.68, 4.01, 6.86, 7.01, 9.18, 10.01, 12.45 (HI 112, HI 113) The custom buffers are a special option that allows the user to calibrate in a
buffer solution different from a standard one. Up to two custom buffers can be set in SETUP menu (see page 17). Each custom buffer value can be changed in a ±1.0 pH window around the set value, during calibration, when it is selected; the BUFFER pH tag will blink.
The instruments will automatically skip the buffer used during calibration and the buffers which are in a ±0.2 pH window, around one of the calibrated buffers.
All new calibrations will override existing stored calibration data in a ±0.2 pH window. The slopes adjacent to the new points will be reevaluated.
If the new calibration point has no correspondence in the existing stored calibration data, it is added to it if this is not full (3 or 5 points), or the instrument will ask which buffer will be replaced by the current buffer.
If at least a two-point calibration has been performed and an offset correction of the electrode is wanted, keeping unchanged the existing slopes, perform a one­point calibration with OFFS option selected in SETUP menu. If Pnt option is selected, the slopes adjacent to the calibration points will be reevaluated.
12
THREE- (HI 110, HI 111) / FIVE-POINT (HI 112, HI 113) CALIBRATION
 Immerse the pH electrode and the temperature probe
approximately 4 cm (1½) into a buffer solution and stir gently. The temperature probe should be close to the pH electrode.
 Press CAL. The CAL and 
BUF
 tags will appear and the 7.01 buffer will be displayed on the secondary LCD.
 If necessary, press the ARROW keys to select a different buffer value.  The   tag will blink on the LCD until the reading is
stable.
 When the reading is stable and close to the selected buffer, the
READY tag will be displayed and the CFM tag will blink.
 Press CFM to confirm calibration.  The calibrated value will be displayed on the primary LCD
and the second expected buffer value on the secondary LCD.
 After the first calibration point is confirmed, immerse the pH electrode and the
temperature probe approximately 4 cm (1½) into the second buffer solution and stir gently. The temperature probe should be close to the pH electrode.
 If necessary, press the ARROW keys to select a different buffer value.  The   tag will blink on the LCD until the reading is
stable.
 When the reading is stable and close to the selected buffer, the
READY tag will be displayed and the CFM tag will blink.
 Press CFM to confirm calibration.  The calibrated value is then displayed on the primary
LCD and and the third expected buffer value on the secondary LCD.
 After the second calibration point is confirmed, immerse the pH electrode
and the temperature probe approximately 4 cm (1½) into the next buffer solution and stir gently. The temperature probe should be close to the pH electrode.
BUFFER pH
2
BUF
pHCAL
BUFFER pH
1
BUF
pHCAL
13
 If necessary, press the ARROW keys to select a different buffer value.  The   tag will blink on the LCD until the reading is
stable.
 When the reading is stable and close to the selected buffer, the
READY tag will be displayed and the CFM tag will blink.
 Press CFM to confirm calibration.  If the fifth calibration point is reached, the instrument stores
the calibration value and returns to normal measurement mode, or else it goes to the next calibration point.
TWO-POINT CALIBRATION
 Proceed as described in THREE-/FIVE-POINT CALIBRATION section.  Press CAL after the second calibration point was confirmed. The instruments
will memorize the two-point calibration data and will return to measurement mode.
ONE-POINT CALIBRATION
Two SETUP selectable options are available: Pnt and OFFS. If the Pnt option is selected, the new calibration point overrides an existing
one. The adjacent slopes will be reevaluated. If the OFFS option is selected, an electrode offset correction is performed
keeping unchanged the existing slopes.  Proceed as described in THREE-/FIVE-POINT CALIBRATION section.  Press CAL after the first calibration point was confirmed. The instruments will
memorize the one-point calibration data and will return to measurement mode.
Note:  If the value measured by the instrument is not close to the
selected buffer, WRONG   and WRONG   tags will blink alternately. In this case check if the correct buffer has been used, or regenerate the electrode by following the Cleaning Procedure (see page 30). If necessary, change the buffer or the electrode.
 When a custom buffer is displayed, the BUFFER pH tag blinks.
To change the custom buffer value in according with the buffer temperature proceed as described in WORKING WITH CUSTOM BUFFERS paragraph (see page 14).
 If the buffer temperature or the manual temperature exceeds the
temperature limits of the buffer, WRONG tag and temperature reading will blink.
 If WRONG, BUFFER pH tags and OLD message are displayed
blinking on the secondary LCD line, an inconsistency between new and previous (old) calibration is detected. Clear calibration parameters and proceed with calibration from the current calibration point. The instrument will keep all confirmed values during current calibration.
14
 To clear calibration parameters for all uncalibrated buffers starting
with current buffer, press and hold down CFM, then press CAL. If this procedure is performed while calibrating in the first calibration point, the instrument returns to measurement mode, else calibration can be continued from the current point.
 Press RANGE (SEL for HI 110) to toggle
between pH buffer and temperature reading (HI 110, HI 111).
Press RANGE to toggle between pH buffer,
buffer number and temperature reading during calibration (HI 112, HI 113 only).
 Each time a buffer is confirmed, the new calibration parameters
replace the old calibration parameters of the coresponding buffer. If current confirmed buffer has no correspondence in the existing stored calibration and this is not full, (three buffers for HI 110, HI 111; five buffers for HI 112, HI 113), the current buffer is added to the existing stored calibration. If the existing stored calibration is full, the instrument asks which buffer will be replaced by current buffer.
Press the ARROW keys to select another buffer to be replaced. Press CFM to confirm the buffer that will be replaced. Press CAL to leave calibration without replacing.
Note: If the replaced buffer is outside the ±0.2 pH window, around each of
the calibrated buffers, it is possible to select this buffer for next calibra­tion during current calibration.
WORKING WITH CUSTOM BUFFERS
If at least one custom buffer was set in SETUP menu, it can be selected for calibration by pressing the ARROW keys. The BUFFER pH tag will blink.
Press SET if you want to adjust the buffer value. The buffer value will start blinking.
Use the ARROW keys to change the buffer value. After about 5 seconds since you performed the last change, the buffer value is
updated. Press SET if you want to change it again. Note: Custom buffer value can be adjusted in a ±1.00 pH window, around
the set value.
CFM
CAL
BUFFER pH
pH
BUF
CAL
BUFFER pH
pH
15
GOOD LABORATORY PRACTICE (GLP)
GLP is a set of functions that allows storage and retrieval of data regarding the maintenance and status of the electrode.
All data regarding the last calibration (one, two, three, four or five-point) is stored for the user to review when necessary. This data includes the following: calibration time stamp, offset (mV), average of slope (%), calibration buffers and the time until a new calibration is required.
CALIBRATION ALARM TIME-OUT
These instruments allow the user to set the number of days before the next required calibration. This value can be set from 1 to 7 days. The default setting is OFF (disabled).
The instruments check if the time-out time has expired. If the time elapsed, the CAL INTV tags will blink as a reminder.
Note: If the instruments were not calibrated, the CAL INTV tags will be
displayed even if the feature is disabled in SETUP menu.
LAST CALIBRATION DATA
Last calibration data are stored automatically after a successful calibration. To view the pH calibration data, press GLP while in measurement mode. The instruments will display the time (hh:mm) of the last calibration.
Press the ARROW keys to view the next logged calibration parameter (pressing the ºC key):
 The date (yyyy.mm.dd).
 The pH calibration offset.
 The pH calibration slope (the GLP slope is the average of the calibration
slopes; the percentage is referred to the ideal value of 59.16 mV/pH).
DATE
CAL
mVCAL
%
CAL
16
 The pH calibration buffers in calibrating order.
The first pH calibration buffer:
The second pH calibration buffer:
The third pH calibration buffer:
Note:  For HI 112 and HI 113, the fourth and the fifth calibration
buffers are also displayed in GLP.
 The OLd message displayed beside the pH value means that
this buffer was not used during last calibration. Press and hold down SET if you want to see calibration date (or time, if old calibration was made in the same day with current calibration).
 For each custom buffer used in calibration, the BUFFER pH tag
will blink.
 If no bUF message appears on the LCD, the instrument informs
you that calibration was performed in less than three (HI 110, HI
111) or five points (HI 112, HI 113).
 pH resolution during calibration.
 Calibration Alarm Time-Out status:
if disabled.
BUF
CAL
BUFFER pH
BUF
pHCAL
BUFFER pH
BUF
pHCAL
BUF
pHCAL
BUFFER pH
CAL
17
Setup mode allows viewing and modifying the following parameters:  Calibration Alarm Time-Out  Custom buffer 1  Custom buffer 2  One-point calibration behaviour  Instrument ID (HI 112 & HI 113)  Current Time  Current Date  Beep Status  Baud Rate (serial communication, HI 112 & HI 113)  Command prefix (serial communication, HI 112 & HI 113) To enter SETUP mode press SET while the instrument is in measurement mode. Select a parameter with the ARROW keys. Press CAL if you want to change a parameter value. The selected parameter
will start blinking. Press RANGE (SEL for HI 110) to toggle between displayed parameters. Press the ARROW keys to increase or decrease the displayed value. Press CFM to save the modified value or CAL to escape. Press the ARROW keys to select the next/previous parameter.
INTV
CAL
CAL
or the number of days until the calibration alarm will be displayed.
or if expired (7 days ago).
 The instrument ID.
Note:  Press GLP at any moment and the instruments will return to
measurement mode.
 If calibration has not been performed, the instruments display
no CAL message blinking.
SETUP
18
CALIBRATION ALARM TIME-OUT SET
Press CAL when the calibration time-out is displayed. Calibration time-out (OFF or 1 to 7 days) will start blinking.
Press the ARROW keys to change the calibration time-out value. Press CFM to save the modified calibration time-out value. Press CAL to escape without saving.
FIRST CUSTOM BUFFER SET
Press CAL when custom buffer 1 is displayed. The custom buffer (disabled  no cb1 or 0 to 16 pH) will start blinking.
Press the ARROW keys to change the custom buffer value. Press CFM to save the modified custom buffer value. Press CAL to escape without saving.
SECOND CUSTOM BUFFER SET
Press CAL when custom buffer 2 is displayed. The custom buffer (disabled  no cb2 or 0 to 16 pH) will start blinking.
Press the ARROW keys to change the custom buffer value. Press CFM to save the modified custom buffer value. Press CAL to escape without saving.
ONE-POINT CALIBRATION BEHAVIOUR
Press CAL when 1 Pnt message is displayed on the secondary LCD. One of the two options (Pnt or OFFS) will start blinking (see pH CALIBRATION PROCEDURE for details, page 11).
Press the ARROW keys to toggle between Pnt and OFFS options. Press CFM to save the behaviour for one-point calibration. Press CAL to escape without saving.
CAL
BUFFER pH
CAL
BUFFER pH
INTV
CAL
CAL
19
INSTRUMENT ID PARAMETER SET (HI 112, HI 113 only)
Press CAL when the instrument ID is displayed. The instrument ID (0000 to
9999) will start blinking.
Press the ARROW keys to change the instrument ID value. Press CFM to save the modified instrument ID value. Press CAL to escape without saving.
CURRENT TIME SET
Press CAL when the current time is displayed. The hour will start blinking.
Press the ARROW keys to change the hour. Press RANGE (SEL for HI 110). The minutes will start blinking.
Press the ARROW keys to change the minutes. Press CFM to save the modified value. Press CAL to escape without saving.
CURRENT DATE SET
Press CAL when the current date is displayed. The year will start blinking.
Press the ARROW keys to change the year. Press RANGE. The month will start blinking.
Press the ARROW keys to change the month.
TIME
TIME
DATE
DATE
20
Press RANGE. The day will start blinking.
Press the ARROW keys to change the day. Press CFM to save the modified value. Press CAL to escape without saving.
BEEP STATUS SET
Press CAL when the beep status is displayed. Beep status (On or OFF) will start blinking.
Press the ARROW keys to change the beep status (On or OFF). Press CFM to save the modified beep status. Press CAL to escape without saving.
BAUD RATE SET (HI 112, HI 113 only)
Press CAL when the baud rate is displayed. The baud rate (600, 1200, 2400, 4800 or 9600) will start blinking.
Press the ARROW keys to change the baud rate value. Press CFM to save the modified baud rate value. Press CAL to escape without saving.
SERIAL COMMUNICATION COMMAND PREFIX SET (HI 112, HI 113 only)
Press CAL when the command prefix is displayed. The command prefix (0 to
47) will start blinking.
Press the ARROW keys to change the command prefix. Press CFM to save the modified command prefix value. Press CAL to escape without saving.
Note:  See the PC interface section on page 25 for a complete explanation.
 The command prefix must be 16 if HI 92000 PC software is used.
DATE
21
All the instruments are factory calibrated for temperature. Hannas temperature probes are interchangeable and no temperature
calibration is needed when they are replaced. If the temperature measurements are inaccurate, temperature recalibration
should be performed. For an accurate recalibration, contact your dealer or the nearest Hanna Customer Service Center, or follow the instructions bellow.  Prepare a vessel containing ice and water and another one containing hot
water (at a temperature of around 50 ºC). Place insulation material around the vessels to minimize temperature changes.
 Use a calibrated thermometer with a resolution of 0.1 ºC as a reference
thermometer.
 With the instrument off, press and hold down the CAL&MEM keys, then
power on the instrument. The CAL tag will appear and the secondary LCD will show 0.0 ºC.
 Immerse the temperature probe in the vessel with ice and water as near
as possible to the reference thermometer. Allow a few seconds for the probe to stabilize.
 Use the ARROW keys to set the reading on the secondary LCD to that of ice
and water, measured by the reference thermometer. When the reading is stable and close to the selected calibration point, READY tag will appear and CFM tag will blink.
 Press CFM to confirm. The secondary LCD will show 50.0 ºC.
 Immerse the temperature probe in the second vessel as near as possible to
the reference thermometer. Allow a few seconds for the probe to stabilize.
TEMPERATURE CALIBRATION
(for technical personnel only)
CAL
GLP
CFM
CAL
22
 Use the ARROW keys to set the reading on the secondary LCD to that of
the hot water.
 When the reading is stable and close to the selected
calibration point, READY tag will appear and CFM tag will blink.
 Press CFM to confirm. The instrument returns to
measurement mode.
Note: If the reading is not close to the selected calibration point, WRONG
tag will blink. Change the temperature probe and restart calibration.
C C
23
mV CALIBRATION
(for technical personnel only)
All the instruments are factory calibrated for mV. Hannas ORP electrodes are interchangeable and no mV calibration is
needed when they are replaced. If the mV measurements are inaccurate, mV recalibration should be performed. For an accurate recalibration, contact your dealer or the nearest Hanna
Customer Service Center, or follow the instructions below. A two or three-point calibration can be performed at 0.0 mV, 600.0 mV and
1800.0 mV.  Attach to the BNC connector a mV simulator with an accuracy of ±0.1 mV.  With the instrument off, press and hold down the MR&CFM keys, then
power on the instrument. The CAL tag will appear and the secondary LCD will show 0.0 mV.
 Set 0.0 mV on the simulator.
When the reading is stable and close to the selected calibration point, READY tag will appear and CFM tag will blink.
 Press CFM to confirm. The secondary LCD will display 600 mV.  Set 600.0 mV on the simulator.
When the reading is stable and close to the selected calibration point, READY tag will appear and CFM tag will blink.
 Press CFM to confirm. The secondary LCD will display 1800 mV.  Set 1800.0 mV on the simulator.
When the reading is stable and close to the selected calibration point, READY tag will appear and CFM tag will blink.
 Press CFM to confirm. The instrument returns to measurement mode. Note:  If the reading is not close to the selected calibration point,
WRONG tag will blink. Verify calibration condition or contact your vendor if you can not calibrate.
 Press CAL in any moment of the calibration process. The instrument
will return to measurement mode. If calibration process is stopped after 600 mV is confirmed, the 600 mV range is calibrated and calibration parameters are memorized.
24
RELATIVE mV CALIBRATION (HI 111
AND HI 113)
 Press CAL when the instrument is in RELATIVE mV measurement mode.
The mV and   tags will blink. Absolute mV is displayed on the primary LCD and AbS message is displayed on the secondary LCD.
 When the absolute reading is stable and in measurement range, the
instrument asks for confirmation.
 If the reading is out of range, WRONG tag will be displayed.  Press CFM to confirm the absolute value. The instrument will display 0.0
mV on the primary LCD and rEL message on the secondary LCD. In this moment the relative mV offset is equal to absolute mV reading.
 Use the ARROW keys if you want to change the displayed relative mV value.  Press CFM to confirm the relative mV value. The relative mV offset is
displayed on the primary LCD and OFF message on the secondary LCD for a few seconds. The instrument returns to measurement mode.
Note: The relative mV value can be changed only inside the relative mV offset
window (± 2000 mV).
25
Data transmission from the instrument to the PC can be done with the HI 92000 Windows® compatible software (optional). HI 92000 also offers graphing and on-line help feature.
Data can be exported to the most popular spreadsheet programs for further analysis.
To connect your instrument to a PC, use the optional Hanna HI 920010 cable connector. Make sure that your instrument is switched off and plug one connector to the instrument RS232C socket and the other into the serial port of your PC.
Note:  Other cables than HI 920010 may use a different configuration.
In this case communication between instrument and PC may not be possible.
 If you are not using Hanna Instruments HI 92000 software,
please see the following instructions.
SENDING COMMANDS FROM PC
It is also possible to remotely control the instrument with any terminal program. Use HI 920010 cable to connect the instrument to a PC, start the terminal program and set the communication options as follows: 8, N, 1, no flow control.
COMMAND TYPES
To send a command to the instrument the scheme is:
<command prefix> <command> <CR>
where: <command prefix> is a selectable ASCII character
between 0 and 47. <command> is the command code (3 characters).
Note: Either small or capital letters can be used.
SIMPLE COMMANDS
PHR Sets the range to pH 0.001 resolution MVR Sets the range to mV (HI 113 only) MRS Sets the range to relative mV (HI 113 only) RNG Is equivalent to pressing RANGE CAL Is equivalent to pressing CAL CFM Is equivalent to pressing CFM/GLP UPC Is equivalent to pressing the UP arrow key DWC Is equivalent to pressing the DOWN arrow key MEM Is equivalent to pressing MEM MRR Is equivalent to pressing MR SET Is equivalent to pressing SET
PC INTERFACE (HI 112 AND HI 113)
26
CHR n Change the instrument range according with the parameter
value (n):
 n=0 pH range/0.001 resolution  n=1 pH range/0.01 resolution  n=3 mV range  n=4 Relative mV range
COMMANDS REQUIRING AN ANSWER
pH? Causes the instrument to send the pH reading. If the range is set
to mV, Err6 is received.
MV? Causes the instrument to send the mV reading. If the range is set
to pH, Err6 is received.
MR? Causes the instrument to send the relative mV reading. If the
range is set to pH or mV, Err6 is received.
TM? Causes the instrument to send the temperature reading. RAS Causes the instrument to send a complete set of readings in
according with the current range:
 pH, mV and temperature reading on pH range.  mV and temperature reading on mV range.  Rel mV, absolute mV and temperature reading on Rel
mV range.
MDR Requests the instrument model name and firmware code. PAR Requests the setup parameters setting (instrument ID, calibra-
tion alarm time-out, pH range, beep status).
GLP Requests the calibration data record. Note:  Err8 is sent if instrument is not in measurement mode.
 Err7 is sent if mV are asked during pH calibration mode.  NAK (21) character is sent when the instrument receives an
unknown or a corrupted command.
Note: Invalid commands will be ignored.
27
The temperature has an effect on pH. The calibration buffer solutions are affected by temperature changes to a lesser degree than normal solutions. During calibration the instrument will automatically calibrate to the pH value corresponding to the measured or set temperature.
During calibration the instrument will display the pH buffer value at 25 ºC.
pH BUFFER TEMPERATURE
DEPENDENCE
PMET SREFFUBHp
86.1 10.4 68.6 10.7 81.9 10.01 54.21
0 23 76.1 10.4 89.6 31.7 64.9 23.01 24.31
5 14 76.1 00.4 59.6 01.7 93.9 42.01 02.31
01 05 76.1 00.4 29.6 70.7 33.9 81.01 00.31
51 95 76.1 00.4 09.6 40.7 72.9 21.01 18.21
02 86 86.1 00.4 88.6 30.7 22.9 60.01 36.21
52 77 86.1 10.4 68.6 10.7 81.9 10.01 54.21
03 68 86.1 20.4 58.6 00.7 41.9 69.9 92.21
53 59 96.1 30.4 48.6 99.6 01.9 29.9 31.21
04 401 96.1 40.4 48.6 89.6 70.9 88.9 40.21
54 311 07.1 50.4 38.6 89.6 40.9 58.9 48.11
05 221 17.1 60.4 38.6 89.6 10.9 28.9 07.11
55 131 27.1 70.4 48.6 89.6 99.8 97.9 75.11
06 041 27.1 90.4 48.6 89.6 79.8 77.9 44.11
56 941 37.1 11.4 58.6 99.6 59.8 67.9 -
07 851 47.1 21.4 58.6 99.6 39.8 57.9 -
57 761 67.1 41.4 68.6 00.7 19.8 47.9 -
08 671 77.1 61.4 68.6 10.7 98.8 37.9 -
58 581 87.1 71.4 78.6 20.7 78.8 47.9 -
09 491 97.1 91.4 88.6 30.7 58.8 57.9 -
59 302 18.1 02.4 98.6 40.7 38.8 67.9 -
28
PREPARATION PROCEDURE
Remove the protective cap of the pH electrode. DO NOT BE ALARMED IF SALT DEPOSITS ARE PRESENT. This is normal with
electrodes. They will disappear when rinsed with water. During transport, tiny bubbles of air may form inside the glass bulb affecting
proper functioning of the electrode. These bubbles can be removed by shaking down the electrode as you would do with a glass thermometer.
If the bulb and/or junction is dry, soak the electrode in HI 70300 or HI 80300 Storage Solution for at least one hour.
ELECTRODE CONDITIONING
& MAINTENANCE
29
For refillable electrodes:
If the filling solution (electrolyte) is more than 2½ cm (1) below the fill hole, add HI 7082 or HI 8082 3.5M KCl Electrolyte Solution for double junction or HI 7071 or HI 8071 3.5M KCl+AgCl Electrolyte Solution for single junction electrodes.
For faster response, unscrew the fill hole screw during measurements.
For AMPHEL® electrodes:
If the electrode does not respond to pH changes, the battery is run down and the electrode should be replaced.
MEASUREMENT
Rinse the electrode tip with distilled water. Immerse the tip (bottom 4 cm /1½) in the sample and stir gently for a few seconds.
For a faster response and to avoid cross-contamination of the samples, rinse the electrode tip with a few drops of the solution to be tested, before taking measurements.
STORAGE PROCEDURE
To minimize clogging and assure a quick response time, the glass bulb and the junction should be kept moist and not allowed to dry out.
Replace the solution in the protective cap with a few drops of HI 70300 or
HI 80300 Storage Solution or, in its absence, Filling Solution (HI 7071 or HI 8071 for single junction and HI 7082 or HI 8082 for double junction
electrodes). Follow the Preparation Procedure on page 28 before taking measurements.
Note: NEVER STORE THE ELECTRODE IN DISTILLED OR DEIONIZED WATER.
PERIODIC MAINTENANCE
Inspect the electrode and the cable. The cable used for connection to the instrument must be intact and there must be no points of broken insulation on the cable or cracks on the electrode stem or bulb. Connectors must be perfectly clean and dry. If any scratches or cracks are present, replace the electrode. Rinse off any salt deposits with water.
For refillable electrodes:
Refill the reference chamber with fresh electrolyte (HI 7071 or HI 8071 for single junction and HI 7082 or HI 8082 for double junction electrodes). Allow the electrode to stand upright for 1 hour.
Follow the Storage Procedure above.
30
CLEANING PROCEDURE
 General Soak in Hanna HI 7061 or HI 8061 General
Cleaning Solution for approximately ½ hour.
 Protein Soak in Hanna HI 7073 or HI 8073 Protein
Cleaning Solution for 15 minutes.
 Inorganic Soak in Hanna HI 7074 Inorganic Cleaning
Solution for 15 minutes.
 Oil/grease Rinse with Hanna HI 7077 or HI 8077 Oil and Fat
Cleaning Solution.
IMPORTANT: After performing any of the cleaning procedures, rinse the electrode thoroughly with distilled water, refill the reference chamber with fresh electrolyte (not necessary for gel-filled electrodes) and soak the electrode in HI 70300 or HI 80300 Storage Solution for at least 1 hour before taking measurements.
31
TROUBLESHOOTING GUIDE
SMOTPMYS MELBORP NOITULOS
evissecxe/esnoperwolS
.tfird
.edortceleHpytriD nipitedortceleehtkaoS
1608IHro1607IH
setunim03rofnoitulos
ehtnaelcnehtdna
.edortcele
puetautculfsgnidaeR
.)esion(nwoddna
.noitcnujytrid/deggolC
leveletylortcelewoL
sedortceleelballifer(
.)ylno
.edortceleehtnaelC
noituloshserfhtiwllifeR
sedortceleelballiferrof(
.)ylno
ehtniegnarfotuO
.elacsVm
yrD
.noitcnuj/enarbmem
ro00307IHnikaoS
egarots00308IH
.noitulos
tonseodretemehT
reffubehttpecca rofnoitulos .noitarbilac
HpredrofotuO
.edortcele
gninaelcehtwolloF
onllitsfI.erudecorp
ehtecalper,stluser
.edortcele
:swohsyalpsidehtfI
ro"00.2-"dna"Hp"
.gniknilb"00.61"
HpehtniegnarfotuO
.elacs
.retemehtetarbilaceR)a
HpehterusekaM)b
deificepsehtnisielpmas
.egnar
etylortceleehtkcehC)c larenegehtdnalevel .edortceleehtfoetats
:swohsyalpsidehtfI
ro"0002-"dna"Vm"
gniknilb"0002+"
ehtniegnarfotuO
.elacsVm
.detcennoctonedortcelE
tonseodretemehT
ehthtiwkrow
.eborperutarepmet
erutarepmetnekorB
.eborp
erutarepmetehtecalpeR
.eborp
otsliafretemehT
sevigroetarbilac
.sgnidaerytluaf
.edortceleHpnekorB .edortceleehtecalpeR
retemehtputratstA
sgatDCLllasyalpsid
.yltnenamrep
sisyekehtfoenO
.dekcolb
rodraobyekehtkcehC
.rodnevehttcatnoc
egassemrorre"xxrrE"
.deyalpsid
.rorrelanretnI dnaretemehtfforewoP
ehtfI.notirewopneht
ehttcatnoc,stsisreprorre
.rodnev
32
Typical Electrode Life
Ambient Temperature 1  3 years 90 ºC Less than 4 months 120 ºC Less than 1 month
Alkaline Error
High concentrations of sodium ions interfere with readings in alkaline solutions. The pH at which the interference starts to be significant depends upon the composition of the glass. This interference is called alkaline error and causes the pH to be underestimated. Hannas glass formulations have the indicated characteristics.
The resistance of glass electrodes partially depends on the temperature. The lower the temperature, the higher the resistance. It takes more time for the reading to stabilize if the resistance is higher. In addition, the response time will suffer to a greater degree at temperatures below 25 ºC.
Since the resistance of the pH electrode is in the range of 50  200 Mohm, the current across the membrane is in the pico Ampere range. Large currents can disturb the calibration of the electrode for many hours.
For these reasons high humidity environments, short circuits and static discharges are detrimental to a stable pH reading.
The pH electrodes life also depends on the temperature. If constantly used at high temperatures, the electrode life is drastically reduced.
TEMPERATURE CORRELATION
FOR pH SENSITIVE GLASS
1.0 Mol L-1 Na
+
0.1 Mol L-1 Na
+
Sodium Ion Correction for the Glass at 20-25 ºC Concentration pH Error
13.00
13.50
14.00
12.50
13.00
13.50
14.00
0.10
0.14
0.20
0.10
0.18
0.29
0.40
33
pH BUFFER SOLUTIONS
HI70004P pH 4.01 Buffer Sachets, 20 mL, 25 pcs HI70007P pH 7.01 Buffer Sachets, 20 mL, 25 pcs HI70010P pH 10.01 Buffer Sachets, 20 mL, 25 pcs HI7001L pH 1.68 Buffer Solution, 500 mL HI7004L pH 4.01 Buffer Solution, 500 mL HI7006L pH 6.86 Buffer Solution, 500 mL HI7007L pH 7.01 Buffer Solution, 500 mL HI7009L pH 9.18 Buffer Solution, 500 mL HI7010L pH 10.01 Buffer Solution, 500 mL HI7001L pH 1.68 Buffer Solution in FDA approved bottle, 500 mL HI8004L pH 4.01 Buffer Solution in FDA approved bottle, 500 mL HI8006L pH 6.86 Buffer Solution in FDA approved bottle, 500 mL HI8007L pH 7.01 Buffer Solution in FDA approved bottle, 500 mL HI8009L pH 9.18 Buffer Solution in FDA approved bottle, 500 mL HI8010L pH 10.01 Buffer Solution in FDA approved bottle, 500 mL
ELECTRODE STORAGE SOLUTIONS
HI70300L Storage Solution, 460 mL HI80300L Storage Solution in FDA approved bottle, 460 mL
ELECTRODE CLEANING SOLUTIONS
HI70000P Electrode Rinse Sachets, 20 mL, 25 pcs HI7061L General Cleaning Solution, 460 mL HI7073L Protein Cleaning Solution, 460 mL HI7074L Inorganic Cleaning Solution, 460 mL HI7077L Oil & Fat Cleaning Solution, 460 mL HI8061L General Cleaning Solution in FDA approved bottle, 460 mL HI8073L Protein Cleaning Solution in FDA approved bottle, 460 mL HI8077L Oil & Fat Cleaning Solution in FDA approved bottle, 460 mL
ELECTRODE REFILL ELECTROLYTE SOLUTIONS
HI7071 3.5M KCl + AgCl Electrolyte, 4x30 mL, for single junction
electrodes HI7072 1M KNO3 Electrolyte, 4x30 mL HI7082 3.5M KCl Electrolyte, 4x30 mL, for double junction electrodes HI8071 3.5M KCl + AgCl Electrolyte in FDA approved bottle, 4x30 mL,
for single junction electrodes HI8072 1M KNO3 Electrolyte in FDA approved bottle, 4x30 mL HI8082 3.5M KCl Electrolyte in FDA approved bottle, 4x30 mL, for
double junction electrodes.
ACCESSORIES
34
ORP PRETREATMENT SOLUTIONS
HI7091L Reducing Pretreatment Solution, 460 mL HI7092L Oxidizing Pretreatment Solution, 460 mL
pH ELECTRODES
All electrodes part numbers ending in B are supplied with a BNC connector and 1 m (3.3') cable, as shown below :
HI1043B
Glass-body, double junction, refillable, combination pH electrode. Use: strong acid/alkali.
HI1053B
Glass-body, triple ceramic, conic shape, refillable, combination pH electrode. Use: emulsions.
HI1083B
Glass-body, micro, Viscolene, non-refillable, combination pH electrode. Use: biotechnology, micro titration.
HI1131B
Glass-body, single junction, refillable, combination pH electrode. Use: general purpose.
120 mm
4.7"
12 mm
0.5"
9.5mm DIA
0.37"
"S" VERSION
HI 1043
120 mm
4.7"
12 mm
0.5"
"S" VERSION
HI 1053
120 mm
12 mm
0.5"
5 mm
0.2"
3 mm
0.12"
3.0 mm DIA
0.12"
HI 1083
120 mm
4.7"
12 mm
0.5"
9.5mm DIA
0.37"
"S" VERSION
HI 1131
35
HI1330B
Glass-body, semimicro, single junction, refillable, combination pH electrode. Use: laboratory, vials.
HI1331B
Glass-body, semimicro, single junction, refillable, combination pH electrode. Use: flasks.
HI1230B
Plastic-body (Ultem®), double junction, gel-filled, combination pH electrode. Use: general, field.
HI2031B
Glass-body, semimicro, conic, refillable, combination pH electrode. Use: semisolid products.
HI1332B
Plastic-body (Ultem®), double junction, refillable, combination pH electrode. Use: general purpose.
120 mm
4.7"
5mm
0.2"
5mm DIA
0.2"
"S" VERSION
HI 1330
120 mm
4.7"
12 mm
0.5"
HI 1230
75 mm
2.95"
6 mm
0.25"
HI 2031
120 mm
4.7"
12 mm
0.5"
"S" VERSION
HI 1332
210 mm
8.25"
8 mm
0.3"
7.5mm DIA
0.29"
"S" VERSION
HI 1331
36
120 mm
4.7"
12 mm
0.5"
FC 100
75 mm
2.95"
6 mm
0.25"
FC 200
120 mm
4.7"
12 mm
0.5"
FC 210
120 mm
4.7"
12 mm
0.5"
9.5mm DIA
0.37"
FC 220
110 mm
4.3"
12 mm
0.5"
FC 911
FC100B
Plastic-body (Kynar®), double junction, refillable, combination pH electrode. Use: general purpose for food industry.
FC200B
Plastic-body (Kynar®), open junction, conic, Viscolene, non-refillable, combination pH electrode. Use: meat & cheese.
FC210B
Glass-body, double junction, conic, Viscolene, non-refillable, combination
pH electrode. Use: milk, yogurt.
FC220B
Glass-body, triple-ceramic, single junction, refillable, combination pH electrode. Use: food processing.
FC911B
Plastic-body (Kynar®), double junction, refillable with built-in amplifier, combination pH electrode. Use: very high humidity.
37
Kynar® is registered Trademark of Pennwalt Corp.
Ultem® is registered Trademark of General Electric Co.
HI1413B
Glass-body, single junction, flat tip, Viscolene, non-refillable, combination pH electrode. Use: surface measurement.
ORP ELECTRODES
HI3131B
Glass-body, refillable, combination platinum ORP electrode. Use: titration.
HI3230B
Plastic-body (Ultem®), gel-filled, combination platinum ORP electrode. Use: general purpose.
HI4430B
Plastic-body (Ultem®), gel-filled, combination gold ORP electrode. Use: general purpose.
Consult the Hanna General Catalog for more electrodes with screw-type or BNC connectors.
110 mm
4.3"
12 mm
0.5"
HI 1413
150 mm
5.9"
12 mm
0.5"
"S" VERSION
HI 3131
120 mm
4.7"
12 mm
0.5"
HI 3230
"S" VERSION
120 mm
4.7"
12 mm
0.5"
HI 4430
"S" VERSION
38
EXTENSION CABLE FOR SCREW-TYPE ELECTRODES (SCREW TO BNC ADAPTER)
HI 7855/1 Extension cable 1 m (3.3') long HI 7855/3 Extension cable 3 m (9.9') long
OTHER ACCESSORIES
HI 710005 Voltage adapter from 115 VAC to 12 VDC (USA plug) HI 710006 Voltage adapter from 230 VAC to 12 VDC (European plug) HI 710012 Voltage adapter from 240 VAC to 12 VDC (UK plug) HI 710013 Voltage adapter from 230 VAC to 12 VDC (South Africa plug) HI 710014 Voltage adapter from 230 VAC to 12 VDC (Australia plug)
HI 76405 Electrode holder
HI 8427 pH and ORP electrode simulator with 1 m (3.3') coaxial cable
ending in female BNC connectors
HI 931001 pH and ORP electrode simulator with LCD and 1 m (3.3')
coaxial cable ending in female BNC connectors HI 7669/2W Temperature probe with 1 m (3.3') cable HI 92000 Windows® compatible software HI 920010 9 to 9-pin RS232 cable
CONNECT TO SCREW TYPE ELECTRODES
CONNECT TO THE
BNC SOCKET
OF THE METER
HI 7855 SERIES CABLE CONNECTORS
CONNECTOR AND 3.0 mm (0.12") CABLE WITH BNC
39
RECOMMENDATIONS FOR USERS
Before using these products, make sure they are entirely suitable for the environment in which they are used.
Operation of these instruments in residential areas could cause unacceptable interferences to radio and TV equipment, requiring the operator to follow all necessary steps to correct interferences.
The glass bulb at the end of the pH electrode is sensitive to electrostatic discharges. Avoid touching this glass bulb at all times.
During operation, ESD wrist straps should be worn to avoid possible damage to the electrode by electrostatic discharges.
Any variation introduced by the user to the supplied equipment may degrade the instruments EMC performance.
To avoid electrical shock, do not use these instruments when voltages at the measurement surface exceed 24 VAC or 60 VDC.
To avoid damage or burns, do not perform any measurement in microwave ovens.
40
SALES AND TECHNICAL SERVICE CONTACTS
Australia:
Tel. (03) 9769.0666  Fax (03) 9769.0699
China:
Tel. (10) 88570068  Fax (10) 88570060
Egypt:
Tel. & Fax (02) 2758.683
Germany:
Tel. (07851) 9129-0  Fax (07851) 9129-99
Greece:
Tel. (210) 823.5192  Fax (210) 884.0210
Indonesia:
Tel. (21) 4584.2941  Fax (21) 4584.2942
Japan:
Tel. (03) 3258.9565  Fax (03) 3258.9567
Korea:
Tel. (02) 2278.5147  Fax (02) 2264.1729
Malaysia:
Tel. (603) 5638.9940  Fax (603) 5638.9829
Singapore:
Tel. 6296.7118  Fax 6291.6906
South Africa:
Tel. (011) 615.6076  Fax (011) 615.8582
Taiwan:
Tel. 886.2.2739.3014  Fax 886.2.2739.2983
Thailand:
Tel. (662) 619.0708.11  Fax (662) 619.0061
United Kingdom:
Tel. (01525) 850.855  Fax (01525) 853.668
USA:
Tel. (401) 765.7500  Fax (401) 765.7575
For e-mail contacts and a complete list of Sales and Technical offices, please see www.hannainst.com.
MAN113
03/05
Loading...