Hanna Instruments ELECTRODES User Manual

Code:
ELECTRODE
INSTRUCTION GUIDE
PREPARATION
Remove the protective cap. DO
NOT BE ALARMED IF ANY SALT DEPOSITS ARE PRESENT. This is normal with electrodes and they will disap­pear when rinsed with water.
Shake the electrode down as you would do with a clinical thermom­eter to eliminate any air bubbles inside the glass bulb.
If the bulb and/or junction are dry, soak the electrode in HI70300
Fill Hole
Screw
Reference
Fill Hole
Sensitive
Wire
Reference
Wire
Reference
Junction
Feel Hole
Sleeve Cover
Reference
Fill Hole
Reference
Fill Hole Sensitive
Wire
Reference
Wire
Reference
Junction
Glass BulbGlass Bulb
Storage Solution for at least one hour.
For refillable electrodes:
If the fill solution (electrolyte) is
less than 1 cm (½") below the fill hole, add HI7082 3,5M KCl Elec- trolyte Solution for double junc­tion or HI7071 3,5M KCl+AgCl Electrolyte Solution for single
Sensitive
Wire
Reference
Wire
Reference
Junction
Platinum or
Gold tip
PLASTIC BODY pH ELECTRODE
PLASTIC BODY
ORP ELECTRODE
GLASS BODY
pH ELECTRODE
GLASS BODY
ORP ELECTRODE
Sensitive
Wire
Reference
Wire
Reference
Junction
Platinum or
Gold tip
junction electrodes.
For a faster response lower the rubber sleeve (or unscrew the fill hole screw).
For AmpHel electrodes
If the wires are extended, they should not touch (short circuited).
Maximum recommended length is 15 m (50'), however cables can be extended.
If the electrode does not respond to pH changes, the battery is run down and the electrode should be replaced.
With Process Instrumentation connect the blue wire to the glass terminal and the white wire to the reference terminal of the meter.
With electrodes terminating in spade lugs, make sure the lugs do not touch each other otherwise the battery life will be drastically reduced.
3
MEASUREMENT
Rinse the electrode tip with distilled water.
Immerse the tip in the sample and stir gently for approx. 30 seconds.
For a faster response and to avoid cross contamination of the
samples, rinse the electrode tip with a few drops of the solution to be tested, before taking measurements.
STORAGE
To minimize clogging and ensure a quick response time, the glass
Replace solution in the protective cap with a few drops of HI70300 Storage Solution or, in its absence, Filling Solution (HI7071 for
single junction or HI7082 for double junction electrodes). If none of the above is available, tap water may also be used for a very short period (couple of days). Follow Preparation above before taking measurements.
Note: NEVER STORE THE ELECTRODE IN DISTILLED WATER.
PERIODIC MAINTENANCE
Inspect the electrode for any scratches or cracks. If any present,
replace the electrode.
Rinse off any salt deposits with water.
For refillable electrodes:
Drain the reference chamber with a syringe and refill it with fresh electrolyte (HI7071 for single junction or HI7082 for double junction electrodes). Allow the electrode to stand upright for 1 hour.
Follow the Storage Procedure above.
CLEANING PROCEDURE
General - Soak in Hanna HI7061 General Cleaning Solution for
approximately 1 hour.
Removal of films, dirt or deposits on the membrane/junction:
-
Protein
Soak in Hanna HI7073 Protein Cleaning Solu- tion for 15 minutes.
-
Inorganic
Soak in Hanna HI7074 Inorganic Cleaning Solution for 15 minutes.
-
Oil and grease
Rinse with Hanna HI7077 Inorganic Cleaning Solution.
21
4
IMPORTANT: After performing any of the cleaning procedures rinse the electrode thoroughly with distilled water, drain and refill the reference chamber with fresh electrolyte, (not necessary for GEL filled electrodes) and soak the electrode in HI70300 Storage Solution for at least 1 hour before taking measurements.
TROUBLE SHOOTING
pH Meter: Follow attentively the meter's operating and calibration proce-
dures from the instruction manual. Electrode: Evaluate your electrode performance based on the following
possibilities.
• Noise (Readings fluctuate up and down) could be due to:
- Clogged/Dirty Junction: Refer to Cleaning Procedure above.
- Loss of shielding due to low electrolyte level (in refillable electrodes
only): Empty electrolyte with a syringe and refill with fresh HI7071 for single junction or HI7082 for double junction electrodes.
• Dry Membrane/Junction: Soak in Storage Solution HI70300 for at least 1 hour.
• Drifting: Soak the electrode tip in warm Hanna Solution HI7082 for one hour then flush tip with distilled water. Refill with fresh HI7071 for single junction electrodes and HI7082 for double junction electrodes.
• Low Slope: Refer to Cleaning Procedure above.
• No Slope: Check electrode for cracks in glass stem or bulb and
replace the electrode.
• Slow Response/Excessive Drift: Soak the tip in Hanna Solution HI7061 for 30 minutes, rinse thoroughly in distilled water and then
follow Cleaning Procedure above.
Each solution is available in M or L size (230 or 460 mL). Electrolyte Solutions are supplied in kits 4X50 mL.
Check with your nearest Hanna Dealer for more information.
IstrElect 05/95
Stay in touch with the manifacturer...
ELECTRODE INSTRUCTION GUIDE
Code:
PREPARATION
Remove the protective cap. DO NOT BE ALARMED IF ANY SALT
DEPOSITS ARE PRESENT. This is normal with electrodes and they will disappear when rinsed with water.
Shake the electrode down as you would do with a clinical ther­mometer to eliminate any air bubbles inside the glass bulb.
If the bulb and/or junction are dry, soak the electrode in
Fill Hole
Screw
Reference
Fill Hole
HI70300 Storage Solution for at least one hour.
For refillable electrodes:
If the fill solution (electrolyte) is
less than 1 cm (½") below the fill
Sensitive
Wire
Reference
Wire
Reference
Junction
hole, add HI7082 for double junc­tion or HI7071 for single junction electrodes.
For a faster response lower the rubber sleeve (or unscrew the fill hole screw).
For AmpHel electrodes
If the wires are extended, they
should not touch (short cir­cuited). Maximum recom-
Sensitive
Wire
Reference
Wire
Reference
Junction
Platinum or
Gold tip
PLASTIC BODY pH ELECTRODE
PLASTIC BODY
ORP ELECTRODE
mended length is 15 m (50'), however cables can be extended.
If the electrode does not respond to pH changes, the battery is run down and the electrode should be replaced.
With Process Instrumentation connect the blue wire to the glass terminal and the white wire to the reference terminal of the meter.
With electrodes terminating in spade lugs, make sure the lugs do not touch each other otherwise the battery life will be drastically reduced.
MEASUREMENT
Rinse the electrode tip with distilled water.
Immerse the tip in the sample and stir gently for approx. 30 seconds.
For a faster response and to avoid cross contamination of the
samples, rinse the electrode tip with a few drops of the solution to be tested, before taking measurements.
STORAGE
To minimize clogging and ensure a quick response time, the glass
bulb and the junction should be kept moist and not allowed to dry out.
Replace solution in the protective cap with a few drops of HI70300 Storage Solution or, in its absence, Filling Solution (HI7071 for single junction or HI7082 for double junction electrodes). If none of the above is available, tap water may also be used for a very short period (couple of days). Follow Preparation above before taking measurements.
Note: NEVER STORE THE ELECTRODE IN DISTILLED WATER.
GLASS BODY
pH ELECTRODE
GLASS BODY
ORP ELECTRODE
Feel Hole
Sleeve Cover
Reference
Fill Hole
Reference
Fill Hole Sensitive
Reference
Reference
Junction
Glass BulbGlass Bulb
Sensitive
Wire
Reference
Wire
Reference
Junction
Platinum or
Gold tip
Wire Wire
PERIODIC MAINTENANCE
Inspect the electrode for any scratches or cracks. If any present,
replace the electrode.
Rinse off any salt deposits with water. For refillable electrodes:
Drain the reference chamber with a syringe and refill it with fresh electrolyte (HI7071 for single junction or HI7082 for double junction electrodes). Allow the electrode to stand upright for 1 hour.
Follow the Storage Procedure above.
CLEANING PROCEDURE
General - Soak in Hanna Solution HI7061 for approximately 1
hour.
Removal of films, dirt or deposits on the membrane/junction:
-
Protein
- Soak in Hanna Solution HI7073 for 15 minutes.
-
Inorganic
-
- Soak in Hanna Solution HI7074 for 15 minutes.
Oil and grease
- Rinse with Hanna Solution HI7077.
IMPORTANT: After performing any of the cleaning procedures rinse the electrode thoroughly with distilled water, drain and refill the refer­ence chamber with fresh electrolyte, (not necessary for GEL filled electrodes) and soak the electrode in HI70300 Storage Solution for at least 1 hour before taking measurements.
TROUBLE SHOOTING
pH Meter: Follow attentively the meter's operating and calibration
procedures from the instruction manual. Electrode: Evaluate your electrode performance based on the follow-
ing possibilities.
• Noise (Readings fluctuate up and down) could be due to:
- Clogged/Dirty Junction: Refer to Cleaning Procedure above.
- Loss of shielding due to low electrolyte level (in refillable elec-
trodes only): Empty electrolyte with a syringe and refill with fresh
HI7071 for single junction or HI7082 for double junction electrodes.
• Dry Membrane/Junction: Soak in Storage Solution HI70300 for at
least 1 hour.
• Drifting: Soak the electrode tip in warm Hanna Solution HI7082 for one hour then flush tip with distilled water. Refill with fresh HI7071 for single junction electrodes and HI7082 for double junction elec­trodes.
• Low Slope: Refer to Cleaning Procedure above.
• No Slope: Check electrode for cracks in glass stem or bulb and
replace the electrode.
• Slow Response/Excessive Drift: Soak the tip in Hanna Solution HI7061 for 30 minutes, rinse thoroughly in distilled water and then
follow Cleaning Procedure above.
OPTIONAL ACCESSORIES
HI70300 / HI80300 Storage Solution (230 or 460 mL) HI7061 / HI8061 General Cleaning Solution (230 or 460 mL) HI7073 / HI8073 Protein Cleaning Solution (230 or 460 mL) HI7074 / HI8074 Inorganic Cleaning Solution (230 or 460 mL) HI7077 / HI8077 Oil and Fat Cleaning Solution (230 or 460 mL) HI7071 / HI8071 3,5M KCl + AgCl Electrolyte Solution, 4X50 mL HI7072 / HI8072 1M KNO HI7082 / HI8082 3,5M KCl Electrolyte Solution, 4X50 mL