COLOR STANDARDS FOR ISM
Instructions for use:
1. Turn the meter on by pressing ON/OFF. When the LCD displays “---”, it is ready.
2. Fill a cuvet up to 1.5 cm (3/4”) below the rim with 10 ml of water, and replace
the cap.
3. Wipe the cuvet with a clean tissue to remove condensation and finger marks.
4. Place the cuvet into the holder and ensure that the notch on the cap is positioned
securely into the groove.
5. Press ZERO and “SIP” will appear on the display. Wait for a few seconds and the
display will show “-0.0-”. Now the meter is zeroed and ready for measurement.
6. Remove the cuvet.
7. Insert the standard into the holder. Press the “READ” key and “SIP” will appear
in the display during measurement.
8. The reading should correspond within the stated tolerance limits.
COLOR STANDARDS FOR ISM
Instructions for use:
1. Turn the meter on by pressing ON/OFF. When the LCD displays “---”, it is ready.
2. Fill a cuvet up to 1.5 cm (3/4”) below the rim with 10 ml of water, and replace
the cap.
3. Wipe the cuvet with a clean tissue to remove condensation and finger marks.
4. Place the cuvet into the holder and ensure that the notch on the cap is positioned
securely into the groove.
5. Press ZERO and “SIP” will appear on the display. Wait for a few seconds and the
display will show “-0.0-”. Now the meter is zeroed and ready for measurement.
6. Remove the cuvet.
7. Insert the standard into the holder. Press the “READ” key and “SIP” will appear
in the display during measurement.
8. The reading should correspond within the stated tolerance limits.
Notes:
1. Checks should be made indoors under normal daylight or artificial lighting
conditions. Avoid working in direct sunlight or directly under powerful artificial
light. Checks should be carried out at a temperature of 20-25°C.
2. Replace standards in the case after use. Store under cool conditions and out of
sunlight. Do not use the standards after the stated expiry date.
3. Hanna color standards are for use with Hanna:
HI 93701, HI 93710, HI 93711 ISM instruments only.
Test values may not correspond with other photometers.
For use as secondary standards only.
DO NOT OPEN THE CUVET !
COLOR STANDARDS FOR ISM
Instructions for use:
1. Turn the meter on by pressing ON/OFF. When the LCD displays “---”, it is ready.
2. Fill a cuvet up to 1.5 cm (3/4”) below the rim with 10 ml of water, and replace
the cap.
3. Wipe the cuvet with a clean tissue to remove condensation and finger marks.
4. Place the cuvet into the holder and ensure that the notch on the cap is positioned
securely into the groove.
5. Press ZERO and “SIP” will appear on the display. Wait for a few seconds and the
display will show “-0.0-”. Now the meter is zeroed and ready for measurement.
6. Remove the cuvet.
7. Insert the standard into the holder. Press the “READ” key and “SIP” will appear
in the display during measurement.
8. The reading should correspond within the stated tolerance limits.
ISTRCLOROR2 03/02
Notes:
1. Checks should be made indoors under normal daylight or artificial lighting
conditions. Avoid working in direct sunlight or directly under powerful artificial
light. Checks should be carried out at a temperature of 20-25°C.
2. Replace standards in the case after use. Store under cool conditions and out of
sunlight. Do not use the standards after the stated expiry date.
3. Hanna color standards are for use with Hanna:
HI 93701, HI 93710, HI 93711 ISM instruments only.
Test values may not correspond with other photometers.
For use as secondary standards only.
DO NOT OPEN THE CUVET !
COLOR STANDARDS FOR ISM
Instructions for use:
1. Turn the meter on by pressing ON/OFF. When the LCD displays “---”, it is ready.
2. Fill a cuvet up to 1.5 cm (3/4”) below the rim with 10 ml of water, and replace
the cap.
3. Wipe the cuvet with a clean tissue to remove condensation and finger marks.
4. Place the cuvet into the holder and ensure that the notch on the cap is positioned
securely into the groove.
5. Press ZERO and “SIP” will appear on the display. Wait for a few seconds and the
display will show “-0.0-”. Now the meter is zeroed and ready for measurement.
6. Remove the cuvet.
7. Insert the standard into the holder. Press the “READ” key and “SIP” will appear
in the display during measurement.
8. The reading should correspond within the stated tolerance limits.
ISTRCLOROR2 03/02
Notes:
1. Checks should be made indoors under normal daylight or artificial lighting
conditions. Avoid working in direct sunlight or directly under powerful artificial
light. Checks should be carried out at a temperature of 20-25°C.
2. Replace standards in the case after use. Store under cool conditions and out of
sunlight. Do not use the standards after the stated expiry date.
3. Hanna color standards are for use with Hanna:
HI 93701, HI 93710, HI 93711 ISM instruments only.
Test values may not correspond with other photometers.
For use as secondary standards only.
DO NOT OPEN THE CUVET !
ISTRCLOROR2 03/02
Notes:
1. Checks should be made indoors under normal daylight or artificial lighting
conditions. Avoid working in direct sunlight or directly under powerful artificial
light. Checks should be carried out at a temperature of 20-25°C.
2. Replace standards in the case after use. Store under cool conditions and out of
sunlight. Do not use the standards after the stated expiry date.
3. Hanna color standards are for use with Hanna:
HI 93701, HI 93710, HI 93711 ISM instruments only.
Test values may not correspond with other photometers.
For use as secondary standards only.
DO NOT OPEN THE CUVET !
ISTRCLOROR2 03/02