Dear Customer,
Thank you for choosing a Hanna product. Please read this instruction
manual carefully before using the meter. This manual will provide you
with the necessary information for the correct use of the instrument. If
you need additional technical information, do not hesitate to e-mail
us at tech@hannainst.com. These instruments are in compliance
with directives.
TABLE OF CONTENTSTABLE OF CONTENTS
TABLE OF CONTENTS
TABLE OF CONTENTSTABLE OF CONTENTS
PRELIMINARY EXAMINATION ............. 3
GENERAL DESCRIPTION ..................... 3
SPECIFICATIONS .............................. 4
PRINCIPLE OF OPERATION ................ 4
FUNCTIONAL DESCRIPTION ............... 6
GUIDE TO DISPLAY CODES ................. 7
TIPS FOR AN ACCURATE MEASURE-
MENT ........................................... 10
All rights are reserved. Reproduction in whole or in part is prohibited without the written consent of the
copyright owner, Hanna Instruments Inc., Woonsocket, Rhode Island, 02895 , USA.
Remove the instrument from the packing material and examine it
carefully to make sure that no damage has occurred during shipment. If there is any damage, notify your Dealer.
Each Meter is supplied complete with:
• Two Sample Cuvets and Caps*
• One Transport Cap
• Two 9 V Batteries
• 12 VDC Transformer (HI 710005 or HI 710006)
Note: Save all packing material until you are sure that the instru-
ment functions correctly. Any defective item must be returned
in its original packing with the supplied accessories.
* C99 & C200, C206, C207, C209, C210 and C213 are supplied with 3 cuvets and caps
GENERAL DESCRIPTIONGENERAL DESCRIPTION
GENERAL DESCRIPTION
GENERAL DESCRIPTIONGENERAL DESCRIPTION
C 99 & C 200 Series is a line of 15 different bench, microprocessorbased photometers that measure up to 46 parameters in water and
wastewater. These multipurpose meters are manufactured to measure
the most important parameters of the application they have been
especially designed for:
C 99 Laboratories, with CODC 200 Laboratories
C 203 AquacultureC 205 Boilers & Cooling Towers
C 206 Environmental TestingC 207 Industrial Wastewater
C 208 Water ConditioningC 209 Education
C 210 Pulp & Paper MillsC 211 Chemical Manufacturers
C 212 Power Plant UtilitiesC 213 Municipal Wastewater
C 215 Nutrient AnalysesC 216 Swimming Pool Applic.
C 218 Environmental Applic.
All meters use an exclusive positive-locking system to ensure that the cuvet
is in the same position every time it is placed into the measurement cell.
The reagents are in liquid or powder form and are supplied in bottles
or in packets. The amount of reagent is precisely dosed to ensure the
maximum repeatability.
Display codes aid the user in routine operations.
The meters have an auto-shut off feature, turning the unit off after
10 minutes of non-use.
The C 99 & C 200 Series can be connected to a personal computer via
the HI 920010 three wire RS 232 cable. The HI 92000 Hanna
Windows® Compatible Software aids the user to manage all test
data.
3
SPECIFICATIONSSPECIFICATIONS
SPECIFICATIONS
SPECIFICATIONSSPECIFICATIONS
Light LifeLife of the instrument
Light DetectorSilicon Photocell
Environment0 to 50°C (32 to 122°F);
max 95% RH non-condensing
Power Supply2 x 9 V batteries / 12 to 20 VDC through
voltage adapter
Auto-Shut offAfter 10' of non-use
Dimensions230 x 165 x 70 mm (9.0 x 6.5 x 2.8")
Weight640 g (22.6 oz.)
For specifications related to each single parameter (e.g. range,
accuracy, etc.), refer to the related measurement section.
PRINCIPLE OF OPERATIONPRINCIPLE OF OPERATION
PRINCIPLE OF OPERATION
PRINCIPLE OF OPERATIONPRINCIPLE OF OPERATION
Absorption of Light is a typical phenomenon of interaction between
Electromagnetic Radiation and Matter. When a light beam crosses a
substance, some of the radiation may be absorbed by atoms,
molecules or crystal lattices.
If pure absorption occurs, the fraction of light absorbed depends both
on the optical path length through the matter and on the physicochemical characteristics of substance according to the Lambert-Beer
Law:
-log I/Io = ελ c d
or
A = ε
c d
λ
Where:
-log I/Io=Absorbance (A)
Io=intensity of incident light beam
I =intensity of light beam after absorption
ελ=molar extinction coefficient at wavelength λ
c=molar concentration of the substance
d=optical path through the substance
Therefore, the concentration "c" can be calculated from the absorbance
of the substance as the other factors are known.
Photometric chemical analysis is based on the possibility to develop an
absorbing compound from a specific chemical reaction between sample
and reagents.
4
Given that the absorption of a compound strictly depends on the
wavelength of the incident light beam, a narrow spectral bandwidth
should be selected as well as a proper central wavelength to optimize
measurements.
The optical system of Hanna's C 99 & C 200 multiparameter
photometers is based on special subminiature tungsten lamps and
narrow-band interference filters to guarantee both high performance
and reliable results.
Four measuring channels (at four different wavelengths) allow a wide
range of tests.
C 200 Block diagram (optical layout)
A microprocessor controlled special tungsten lamp emits radiation
which is first optically conditioned and beamed to the sample
contained in the cuvet. The optical path is fixed by the diameter of
the cuvet. Then the light is spectrally filtered to a narrow spectral
bandwidth, to obtain a light beam of intensity Io or I.
The photoelectric cell collects the radiation I that is not absorbed by
the sample and converts it into an electric current, producing a
potential in the mV range.
The microprocessor uses this potential to convert the incoming value
into the desired measuring unit and to display it on the LCD.
The measurement process is carried out in two phases: first the meter
is zeroed and then the actual measurement is performed.
The cuvet has a very important role because it is an optical element
and thus requires particular attention. It is important that both the
measurement and the calibration (zeroing) cuvets are optically identical to provide the same measurement conditions. Whenever possible
use the same cuvet for both.
It is also necessary that the surface of the cuvet is clean and not
scratched. This is to avoid measurement interference due to unwanted
reflection and absorption of light. It is recommended not to touch the
cuvet walls with hands.
Furthermore, in order to maintain the same conditions during the
zeroing and the measuring phases, it is necessary to close the cuvet
to prevent any contamination.
5
FUNCTIONAL DESCRIPTIONFUNCTIONAL DESCRIPTION
FUNCTIONAL DESCRIPTION
FUNCTIONAL DESCRIPTIONFUNCTIONAL DESCRIPTION
FRONT PANEL
1) Cuvet Holder
2) Dual Level Liquid Crystal Display
3) Programs List
4) READ DIRECT, to perform measurement immediately
5) TIMER, to perform measurements after a preprogrammed
countdown
6) ZERO, to zero the meter prior to measurement
7) Program W and V, to select the desired parameter
8) ON/OFF, to turn the meter on and off
REAR PANEL
1) Power Supply 12 VDC 2.5 Watt
2) RS 232 Socket
3) Batteries Compartment
6
GUIDE TO DISPLAY CODESGUIDE TO DISPLAY CODES
GUIDE TO DISPLAY CODES
GUIDE TO DISPLAY CODESGUIDE TO DISPLAY CODES
Note: The secondary LCD below shows a generic "P– –", whereas the
meter will indicate the exact program number (e.g. in C 200,
"P1" for Aluminum).
This indicates that the meter is in a ready
state and zeroing can be performed.
Sampling in progress. This flashing
prompt appears each time the meter is
performing a measurement.
The microprocessor is adjusting the light
level, indicated by a scrolling "SIP".
This indicates that the meter is in a zeroed
state and measurement can be performed.
The light level is accepted. The instrument
is ready to perform a zero reading. This
flashing prompt will only appear when a
second zero reading needs to be performed.
Follow the measurement procedure described in the related chapter.
The instrument is performing an internal
check-up.
The blinking "LOBAT" indicates that the
battery voltage is getting low and the
batteries need to be replaced.
This indicates that the batteries are dead
and must be replaced.
7
ERROR MESSAGESERROR MESSAGES
ERROR MESSAGES
ERROR MESSAGESERROR MESSAGES
a) on zero reading:a) on zero reading:
a) on zero reading:
a) on zero reading:a) on zero reading:
Light over range. The cuvet is not inserted
correctly and an eccess ambient light is
reaching the detector. If the cuvet is
properly inserted, then contact your dealer
or the nearest Hanna Customer Service
Center.
The lamp is not working properly. Contact
your dealer or the nearest Hanna Customer
Service Center.
The lamp is not working properly. Contact
your dealer or the nearest Hanna Customer
Service Center.
This indicates that the meter has lost its
configuration. Contact your dealer or the
nearest Hanna Customer Service Center.
This indicates that the zeroing procedure
failed due to a low signal-to-noise ratio. In
this case press ZERO again.
The instrument cannot adjust the light
level. Please check that the sample does not
contain any debris.
There is not enough light to perform a
measurement. Please check the preparation of the zero cuvet.
There is too much light to perform a
measurement. Please check the preparation of the zero cuvet.
8
b) on sample reading:b) on sample reading:
b) on sample reading:
b) on sample reading:b) on sample reading:
There is too much light for the sample
measurement. Please check if the right
sample cuvet is inserted.
The sample and zero cuvet are inverted.
A zero reading was not taken. Follow the
instruction described in the measurement
procedures for zeroing the meter.
Under range. A blinking "0.00" indicates
that the sample absorbs less light than the
zero reference. Check the procedure and
make sure that you use the same cuvet for
reference (zero) and measurement.
1) A flashing value of the maximum
concentration indicates an over range
condition. The concentration of the sample
is beyond the programmed range: dilute
the sample and rerun the test.
2) A flashing value lower than the
maximum concentration indicates a low
signal-to-noise ratio condition. In this case
accuracy of the result is not guaranteed.
Repeat the reading procedure.
9
TIPS FOR AN ACCURATE MEASUREMENTTIPS FOR AN ACCURATE MEASUREMENT
TIPS FOR AN ACCURATE MEASUREMENT
TIPS FOR AN ACCURATE MEASUREMENTTIPS FOR AN ACCURATE MEASUREMENT
The instructions listed below should be carefully followed during
testing to ensure best accuracy.
• Color or suspended matter in large amounts may cause interference, therefore, these should be removed by treatment with active
carbon and by prior filtration.
• For a correct filling of the cuvet: the
liquid in the cuvet forms a convexity
on the top; the bottom of this convexity must be at the same level of the
10 mL mark.
• In order to measure exactly e.g. 0.5 mL of sample or reagent
with the 1 mL syringe, push the plunger completely into the
syringe and insert the tip into the solution. Pull the plunger up
to above the 0.0 mL mark. Take out the syringe and clean the
outside of the syringe tip. Then, adjust the plunger to the 0.0 mL
mark (the lower edge of the seal must be exactly on the 0.0 mL
mark). Be sure that no drops are hanging on the tip of the
syringe, if so eliminate them. Then, keeping the syringe in vertical
position above the cuvet, push the plunger down until the lower
edge of the seal is exactly on the 0.5 mL mark. Now the exact
amount of 0.5 mL has been added to the cuvet, even if the tip
still contains some reagent or sample.
10 mL
• Proper use of the dropper: to get good
reproducible results, tap the dropper on the
table for several times and wipe the outside
of the dropper tip with a cloth. Always keep
the dropper bottle in a vertical position
while dosing the reagent.
10
• Proper use of the powder reagent packet:
(a) use scissors to open the powder packet;
(b) push the edges of the packet to form a spout;
(c) pour out the content of the packet.
(a)
(b)
(c)
• It is important that the sample does not contain any debris. This
would corrupt the readings.
• Do not let the reacted sample stand too long after reagent is
added, or accuracy will be lost.
• In order to avoid reagent leaking and to obtain
more accurate measurements, it is recommended to
close the cuvet first with the supplied HDPE plastic
stopperand then with the black cap.
• Each time the cuvet is used, the cap must be
tightened to the same degree.
• Whenever the cuvet is placed into
the measurement cell, it must be
dry outside, and completely free of
fingerprints, oil or dirt. Wipe it
thoroughly with HI 731318 or a
lint-free cloth prior to insertion.
• It is possible to take multiple readings in a row, but it is recommended
to take a new zero reading for each sample and to use the same
cuvet for zeroing and measurement.
• It is important to discard the sample immediately after the
reading is taken because the glass might become permanently
stained.
• All the reaction times reported in this manual are referred to 20°C
(68°F). As a general rule of thumb, they should be doubled at
10°C (50°F) and halved at 30°C (86°F).
Plug the 12VDC adapter (HI 710005 - 110VDC, or HI 710006 220VDC) into the DC socket. Plug the adapter into the outlet.
Alternatively, remove the battery cover on the back of the meter;
attach 2 fresh 9V batteries and replace the cover.
Note: Insure the main line is surge protected.
Note: Always turn the meter off before unplugging it to insure no
data is lost.
MEASUREMENT PROCEDURE
• Turn the meter on by pressing ON/OFF.
• The meter will first perform an LCD self
diagnostic test by displaying a full set of
figures.
• Then it will show a scrolling "c --- Hanna Inst" message.
• When the LCD displays “----”, the meter is
ready. On the secondary LCD "P1" will
appear to inform that the first parameter
measurement procedure (e.g. in C200, P1
for Aluminum) can be performed.
• Press the PROGRAM W and PROGRAM V keys to select the
desired parameter.
For the program number, see the tables on page 12 or look at
the list printed on the mask of the meter.
19
• After the desired program number appears on the secondary
display, follow the measurement procedure described in the
related chapter.
• Select a new parameter measurement procedure by pressing the
PROGRAM W and PROGRAM V keys.
Note: in the following measurement sections, a generic "P– –" will be placed on
the secondary LCD instead of the exact
related message (e.g. in C 200, "P1" for
Aluminum).
• Before performing a test read carefully all the instructions related
to the selected parameter.
ABBREVIATIONSABBREVIATIONS
ABBREVIATIONS
ABBREVIATIONSABBREVIATIONS
°C°C
°C:degree Celsius
°C°C
EPAEPA
EPA:US Environmental Protection Agency
EPAEPA
°F°F
°F:degree Fahrenheit
°F°F
g/Lg/L
g/L:grams per liter (ppt)
g/Lg/L
HRHR
HR:High Range
HRHR
LRLR
LR:Low Range
LRLR
mg/Lmg/L
mg/L: milligrams per liter (ppm)
mg/Lmg/L
mLmL
mL:milliliter
mLmL
MRMR
MR:Medium Range
MRMR
µµ
g/Lg/L
µ
g/L: micrograms per liter (ppb)
µµ
g/Lg/L
PANPAN
PAN:1-(2-pyridylazo)-2-naphtol
PANPAN
TPTZTPTZ
TPTZ:2,4,6-tri-(2-pyridyl)-1,3,5-triazine
TPTZTPTZ
20
ALUMINUMALUMINUM
ALUMINUM
ALUMINUMALUMINUM
SPECIFICATIONS
Range0.00 to 1.00 mg/L
Resolution0.01 mg/L
Accuracy±0.02 mg/L ±4% of reading
Typical EMC±0.01 mg/L
Deviation
Light SourceTungsten lamp with narrow band interference
filter @ 525 nm
MethodAdaptation of the aluminon method. The reac-
tion between aluminum and reagents causes a
reddish tint in the sample.
REQUIRED REAGENTS
CodeDescriptionQuantity
HI 93712A-0Ascorbic acid1 packet
HI 93712B-0Aluminon reagent1 packet
HI 93712C-0Bleaching powder1 packet
REAGENT SETS
HI 93712-01 Reagents for 100 tests
HI 93712-03 Reagents for 300 tests
For other accessories see page 141.
MEASUREMENT PROCEDURE
• Select the program number corresponding to Aluminum on the
secondary LCD by pressing PROGRAM W and V.
• Fill a graduated beaker with 50 mL of sample.
• Add the content of one packet of HI 93712A
Ascorbic acid reagent and mix until dissolution is
complete.
• Add the content of one packet of HI 93712B
Aluminon reagent and mix until dissolution is
complete. This is the sample.
21
Aluminum
• Fill two cuvets with 10 mL of sample
each (up to the mark).
• Add the content of one HI 93712C
Bleaching powder packet to one of
the two cuvets. Replace the cap
and shake vigorously until dissolution
is complete. This is the blank.
# 1
# 1
10 mL10 mL
# 2
• Place the blank into the holder and
# 1
ensure that the notch on the cap is
positioned securely into the groove.
• Press TIMER and the display will show the countdown prior to
zeroing the blank. Alternatively wait for 15 minutes and then
press ZERO. "SIP" will blink during zeroing.
• Wait for a few seconds and the display will
show "-0.0-". Now the meter is zeroed and
ready for measurement.
• Remove the blank and insert the other
# 2
cuvet into the instrument.
• Press READ DIRECT.
"SIP" will blink during
measurement.
• The instrument directly displays concentration in mg/L of aluminum
on the Liquid Crystal Display.
INTERFERENCES
Interference may be caused by:
Iron above 20 mg/LAlkalinity above 1000 mg/L
Phosphate above 50 mg/LFluoride must be absent
Aluminum
22
AMMONIA HIGH RANGEAMMONIA HIGH RANGE
AMMONIA HIGH RANGE
AMMONIA HIGH RANGEAMMONIA HIGH RANGE
SPECIFICATIONS
Range0.0 to 50.0 mg/L
Resolution0.1 mg/L
Accuracy±0.5 mg/L ±5% of reading
Typical EMC±0.1 mg/L
Deviation
Light SourceTungsten lamp with narrow band interference
filter @ 420 nm
MethodAdaptation of the ASTM Manual of Water and
Environmental Technology, D1426-92, Nessler
method. The reaction between ammonia and
reagents causes a yellow tint in the sample.
REQUIRED REAGENTS
CodeDescriptionQuantity
HI 93733A-0 Nessler Reagent 4 drops (in fresh and seawater)
HI 93733B-0 Ammonia Reagent 9 mL (in fresh and seawater)
REAGENT SETS
HI 93733-01 Reagents for 100 tests
HI 93733-03 Reagents for 300 tests
For other accessories see page 141.
MEASUREMENT PROCEDURE
• Select the program number corresponding to Ammonia HR on the
secondary LCD by pressing PROGRAM W and V.
• Fill a cuvet with 1 mL of unreacted sample, by
means of the syringe.
1 mL of
sample
• Add 9 mL of HI 93733B
9 mL
Ammonia Reagent, by
means of the 3 mL plastic
pipette. Place the cap and
swirl the solution to mix.
• Place the cuvet into the holder and
ensure that the notch on the cap is
positioned securely into the groove.
23
Ammonia HR
• Press ZERO and “SIP” will blink on the display.
• Wait for a few seconds and the display will
show “-0.0-”. Now the meter is zeroed
and ready for measurement.
• Remove the cuvet.
• Add 4 drops of HI 93733A
x 4
Nessler Reagent. Replace the
cap and mix the solution.
• Reinsert the cuvet into the instrument.
• Press TIMER and the display will show the countdown prior to
measurement or, alternatively, wait for 3 minutes and 30 seconds
and then press READ DIRECT. In both cases the display will show
“SIP” during measurement.
• The instrument directly displays concentration in mg/L of Ammonium ion (NH
+
) on the Liquid Crystal Display.
4
• To convert the reading to mg/L of ammonia (NH3), multiply by a
factor of 0.944.
• To convert the reading to ammonia nitrogen (NH3-N), multiply by
a factor of 0.776.
INTERFERENCES
Interference may be caused by:
acetone, alcohols, aldehydes, glycine, hardness above 1 g/L, iron,
organic chloramines, sulfide, various aliphatic and aromatic amines.
Ammonia HR
24
AMMONIA MEDIUM RANGEAMMONIA MEDIUM RANGE
AMMONIA MEDIUM RANGE
AMMONIA MEDIUM RANGEAMMONIA MEDIUM RANGE
SPECIFICATIONS
Range0.00 to 10.00 mg/L
Resolution0.01 mg/L
Accuracy±0.05 mg/L ±5% of reading
Typical EMC±0.01 mg/L
Deviation
Light SourceTungsten lamp with narrow band interference
filter @ 420 nm
MethodAdaptation of the ASTM Manual of Water and
Environmental Technology, D1426-92, Nessler
method. The reaction between ammonia and
reagents causes a yellow tint in the sample.
REQUIRED REAGENTS
CodeDescriptionQuantity
HI 93715A-0First Reagent4 drops (6 drops in seawater)
HI 93715B-0Second Reagent4 drops (10 drops in seawater)
REAGENT SETS
HI 93715-01 Reagents for 100 tests
HI 93715-03 Reagents for 300 tests
For other accessories see page 141.
MEASUREMENT PROCEDURE
• Select the program number corresponding to Ammonia MR on the
secondary LCD by pressing PROGRAM W and V.
• Fill the cuvet with 10 mL of unreacted sample (up
to the mark) and replace the cap.
10 mL
• Place the cuvet into the holder and
ensure that the notch on the cap is
positioned securely into the groove.
• Press ZERO and “SIP” will blink on the display.
• Wait for a few seconds and the display will
show “-0.0-”. Now the meter is zeroed
and ready for measurement.
25
Ammonia MR
• Remove the cuvet.
• Add 4 drops of the First reagent (6
drops in case of seawater analysis).
Replace the cap and mix the
x 4
solution.
• Add 4 drops of the Second reagent
(10 drops in case of seawater
analysis). Replace the cap and mix
the solution.
x 4
• Reinsert the cuvet into the instrument.
• Press TIMER and the display will show the countdown prior to the
measurement or, alternatively, wait for 3 minutes and 30 seconds
and press READ DIRECT. In both cases “SIP” will blink during
measurement.
• The instrument directly displays concentration in mg/L of ammonia
nitrogen (NH3-N) on the display.
• To convert the reading to mg/L of ammonia (NH3), multiply by a
factor of 1.216.
INTERFERENCES
Interference may be caused by:
acetone, alcohols, aldehydes, glycine, hardness above 1 g/L, iron,
organic chloramines, sulfide, various aliphatic and aromatic amines.
Ammonia MR
26
AMMONIA LOW RANGEAMMONIA LOW RANGE
AMMONIA LOW RANGE
AMMONIA LOW RANGEAMMONIA LOW RANGE
SPECIFICATIONS
Range0.00 to 3.00 mg/L
Resolution0.01 mg/L
Accuracy±0.04 mg/L ±4% of reading
Typical EMC±0.01 mg/L
Deviation
Light SourceTungsten lamp with narrow band interference
filter @ 420 nm
MethodAdaptation of the ASTM Manual of Water and
Environmental Technology, D1426-92, Nessler
method. The reaction between ammonia and
reagents causes a yellow tint in the sample.
REQUIRED REAGENTS
CodeDescriptionQuantity
HI 93700A-0First Reagent4 drops (6 drops in seawater)
HI 93700B-0Second Reagent4 drops (10 drops in seawater)
REAGENT SETS
HI 93700-01 Reagents for 100 tests
HI 93700-03 Reagents for 300 tests
For other accessories see page 141.
MEASUREMENT PROCEDURE
• Select the program number corresponding to Ammonia LR on the
secondary LCD by pressing PROGRAM W and V.
• Fill the cuvet with 10 mL of unreacted sample, up
to the mark, and replace the cap.
• Place the cuvet into the holder and
ensure that the notch on the cap is
positioned securely into the groove.
• Press ZERO and “SIP” will blink on the display.
• Wait for a few seconds and the display will
show “-0.0-”. Now the meter is zeroed
and ready for measurement.
27
10 mL
Ammonia LR
• Remove the cuvet.
• Add 4 drops of the First reagent
(6 drops in case of seawater
analysis). Replace the cap and
mix the solution.
x 4
• Add 4 drops of the Second
reagent (10 drops in case of
seawater analysis). Replace the
cap and mix the solution.
x 4
• Reinsert the cuvet into the instrument.
• Press TIMER and the display will show the countdown prior to the
measurement or, alternatively, wait for 3 minutes and 30 seconds
and press READ DIRECT. In both cases “SIP” will blink during
measurement.
• The instrument directly displays concentration in mg/L of ammonia
nitrogen (NH3-N) on the display.
• To convert the reading to mg/L of ammonia (NH3), multiply the
display by a factor of 1.216.
INTERFERENCES
Interference may be caused by:
acetone, alcohols, aldehydes, glycine, hardness above 1 g/L, iron,
organic chloramines, sulfide, various aliphatic and aromatic amines.
Ammonia LR
28
BROMINEBROMINE
BROMINE
BROMINEBROMINE
SPECIFICATIONS
Range0.00 to 8.00 mg/L
Resolution0.01 mg/L
Accuracy±0.08 mg/L ±3% of reading
Typical EMC±0.01 mg/L
Deviation
Light SourceTungsten lamp with narrow band interference
filter @ 525 nm
MethodAdaptation of the Standard Methods for the
Examination of Water and Wastewater, 18
edition, DPD method. The reaction between bro-
mine and the reagent causes a pink tint in the
sample.
REQUIRED REAGENTS
CodeDescriptionQuantity
HI 93716-0DPD Reagent1 packet
REAGENT SETS
HI 93716-01 Reagents for 100 tests
HI 93716-03 Reagents for 300 tests
For other accessories see page 141.
MEASUREMENT PROCEDURE
• Select the program number corresponding to Bromine on the
secondary LCD by pressing PROGRAM W and V.
th
• Fill the cuvet with 10 mL of unreacted sample (up
to the mark) and replace the cap.
• Place the cuvet into the holder and
ensure that the notch on the cap is
positioned securely into the groove.
• Press ZERO and “SIP” will blink on the display.
29
10 mL
Bromine
• Wait for a few seconds and the display will
show “-0.0-”. Now the meter is zeroed
and ready for measurement.
• Remove the cuvet and add the
content of one packet of HI 93716
reagent. Replace the cap and
shake gently for about 20 seconds
to dissolve most of the reagent.
• Reinsert the cuvet into the instrument.
• Press TIMER and the display will show the countdown prior to the
measurement or, alternatively, wait for 2 minutes and 30 seconds
and press READ DIRECT. In both cases "SIP" will blink during
measurement.
• The instrument directly displays concentration in mg/L of bromine
on the Liquid Crystal Display.
INTERFERENCES
Interference may be caused by: Chlorine, Iodine, Ozone, Oxidized
forms of Chromium and Manganese.
In case of water with hardness greater than 500 mg/L CaCO3, shake
the sample for approximately 2 minutes after adding the reagent.
In case of water with alkalinity greater than 250 mg/L CaCO3 or
acidity greater than 150 mg/L CaCO3, the color of the sample may
develop only partially, or may rapidly fade. To resolve this, neutralize
the sample with diluted HCl or NaOH.
Bromine
30
FREE CHLORINEFREE CHLORINE
FREE CHLORINE
FREE CHLORINEFREE CHLORINE
SPECIFICATIONS
Range0.00 to 2.50 mg/L
Resolution0.01 mg/L
Accuracy±0.03 mg/L ±3% of reading
Typical EMC±0.01 mg/L
Deviation
Light SourceTungsten lamp with narrow band interference
filter @ 525 nm
MethodAdaptation of the EPA DPD method 330.5. The
reaction between free chlorine and the DPD reagent causes a pink tint in the sample.
REQUIRED REAGENTS
POWDER:
CodeDescriptionQuantity
HI 93701-0DPD1 packet
LIQUID:
CodeDescriptionQuantity
HI 93701A-FDPD1 Indicator3 drops
HI 93701B-FDPD1 Buffer3 drops
REAGENT SETS
HI 93701-F Reagents for 300 tests (liquid)
HI 93701-01 Reagents for 100 tests (powder)
HI 93701-03 Reagents for 300 tests (powder)
For other accessories see page 141.
MEASUREMENT PROCEDURE
• Select the program number corresponding to Free Chlorine on the
secondary LCD by pressing PROGRAM W and V.
10 mL
• Fill the cuvet with 10 mL of unreacted sample (up
to the mark) and replace the cap.
• Place the cuvet into the holder and
ensure that the notch on the cap is
positioned securely into the groove.
31
Free Chlorine
• Press ZERO and “SIP” will blink on the display.
• Wait for a few seconds and the display will
show “-0.0-”. Now the meter is zeroed
and ready for measurement.
• Remove the cuvet.
Powder reagents procedure
• Add the content of one packet
of HI 93701 DPD reagent.
Replace the cap and shake
gently for 20 seconds (or 2
minutes in case of seawater
analysis).
• Wait for a minute to allow the
undissolved reagent to precipitate
and reinsert the cuvet into the
instrument.
• Press READ DIRECT and the display will show “SIP” during
measurement.
• The instrument directly displays concentration in mg/L of free
chlorine on the Liquid Crystal Display.
32Free Chlorine
Liquid reagents procedure
• To an empty cuvet add 3
drops of HI 93701A-F DPD1
indicator and 3 drops of
HI 93701B-F DPD1 buffer.
Swirl gently to mix, and
immediately add 10 mL of
x 3
x 3
unreacted sample. Replace
the cap and shake gently
again.
• Reinsert the cuvet into the instrument.
• Press READ DIRECT and the display will show “SIP” during
measurement.
• The instrument directly displays concentration in mg/L of free
chlorine on the Liquid Crystal Display.
INTERFERENCES
Interference may be caused by: Bromine, Iodine, Ozone, Oxidized
forms of Chromium and Manganese.
In case of water with hardness greater than 500 mg/L CaCO3, shake
the sample for approximately 2 minutes after adding the powder
reagent.
In case of water with alkalinity greater than 250 mg/L CaCO3 or
acidity greater than 150 mg/L CaCO3, the color of the sample may
develop only partially, or may rapidly fade. To resolve this, neutralize
the sample with diluted HCl or NaOH.
33Free Chlorine
FREE CHLORINE HIGH RANGEFREE CHLORINE HIGH RANGE
FREE CHLORINE HIGH RANGE
FREE CHLORINE HIGH RANGEFREE CHLORINE HIGH RANGE
SPECIFICATIONS
Range0.0 to 10.0 mg/L
Resolution0.1 mg/L
Accuracy±0.1 mg/L ±3% of reading
Typical EMC±0.1 mg/L
Deviation
Light SourceTungsten lamp with narrow band interference
filter @ 525 nm
MethodAdaptation of the EPA DPD method 330.5. The
reaction between free chlorine and the DPD reagent causes a pink tint in the sample.
REQUIRED REAGENTS
CodeDescriptionQuantity
HI 93701-0DPD1 packet
HI 93734B-0 Free & Total Chlorine HR Reagent5 mL
REAGENT SETS
HI 93734-01Reagents for 100 tests
HI 93734-03Reagents for 300 tests
For other accessories see page 141.
MEASUREMENT PROCEDURE
• Select the program number corresponding to Free Chlorine HR on
the secondary LCD by pressing PROGRAM W and V.
• Add to the cuvet 5 mL of HI 93734B
reagent by means of the 5 mL syringe.
Free Chlorine HR
34
5 mL
Note: To measure exactly 5 mL of reagent with the syringe, push
the plunger completely into the syringe and insert the tip into
HI 93734B reagent bottle. Pull the plunger out until the lower
edge of the seal is on the 5 mL mark of the syringe.
probable level of liquid
taken up by syringe
5 mL
• Fill the cuvet up to the 10 mL mark
of sample
with 5 mL of unreacted sample, using
the 3 mL plastic pipette.
Note: rinse the 3 mL plastic pipette 2
10 mL
or 3 times with sample before adding it
to the cuvet with reagent.
• Replace the cap and shake gently.
• Place the cuvet into the holder and
ensure that the notch on the cap is
positioned securely into the groove.
• Press ZERO and "SIP" will blink on the display.
• Wait for a few seconds and the display will
show “-0.0-”. Now the meter is zeroed
and ready for measurement.
35
Free Chlorine HR
• Remove the cuvet.
• Add the content of one packet
of HI 93701 DPD reagent. Replace the cap and shake gently
for 20 seconds (or 2 minutes in
case of seawater analysis).
• Wait for a minute to allow the
undissolved reagent to precipitate
and reinsert the cuvet into the
instrument.
• Press READ DIRECT and the display will show “SIP” during
measurement.
• The instrument directly displays concentration in mg/L of free
chlorine on the Liquid Crystal Display.
INTERFERENCES
Interference may be caused by: Bromine, Iodine, Ozone, Oxidized
forms of Chromium and Manganese.
Alkalinity above 250 mg/L or acidity above 150 mg/L will not
reliably develop the full amount of color or it may rapidly fade. To
resolve this, neutralize the sample with diluted HCl or NaOH.
In case of water with hardness greater than 1000 mg/L CaCO3,
shake the sample for approximately 1 minute after adding the
powder reagent.
Free Chlorine HR
36
TOTAL CHLORINETOTAL CHLORINE
TOTAL CHLORINE
TOTAL CHLORINETOTAL CHLORINE
SPECIFICATIONS
Range0.00 to 3.50 mg/L
Resolution0.01 mg/L
Accuracy±0.03 mg/L ±3% of reading
Typical EMC±0.01 mg/L
Deviation
Light SourceTungsten lamp with narrow band interference
filter @ 525 nm
MethodAdaptation of the EPA DPD method 330.5. The
reaction between the chlorine and the DPD reagent causes a pink tint in the sample.
REQUIRED REAGENTS
POWDER:
CodeDescriptionQuantity
HI 93711-0DPD1 packet
LIQUID:
CodeDescriptionQuantity
HI 93701A-TDPD1 indicator3 drops
HI 93701B-TDPD1 buffer3 drops
HI 93701CDPD3 solution1 drop
REAGENT SETS
HI 93701-T Reagents for 300 total chlorine tests (liquid)
HI 93711-01 Reagents for 100 total chlorine tests (powder)
HI 93711-03 Reagents for 300 total chlorine tests (powder)
For other accessories see page 141.
MEASUREMENT PROCEDURE
• Select the program number corresponding to Total Chlorine on
the secondary LCD by pressing PROGRAM W and V.
• Fill the cuvet with 10 mL of unreacted sample
(up to the mark) and replace the cap.
• Place the cuvet into the holder and
ensure that the notch on the cap is
positioned securely into the groove.
37
10 mL
Total Chlorine
• Press ZERO and “SIP” will blink on the display.
• Wait for a few seconds and the display will
show “-0.0-”. Now the meter is zeroed
and ready for measurement.
• Remove the cuvet.
Powder reagents procedure
• Add 1 packet of HI 93711
reagent. Replace the cap and
shake gently for 20 seconds
(or 2 minutes in case of seawater analysis).
• Reinsert the cuvet into the
instrument.
• Press TIMER and the display will show the countdown prior to the
measurement or, alternatively, wait for 2 minutes and 30 seconds
and press READ DIRECT. In both cases “SIP” will blink during
measurement.
• The instrument directly displays concentration in mg/L of total
chlorine on the Liquid Crystal Display.
Liquid reagents procedure
• To an empty cuvet
add 3 drops of HI
93701A-T DPD1
indicator, 3 drops of
HI 93701B-T DPD1 buffer
and 1 drop of HI 93701C
x 3
x 3
x 1
DPD solution. Swirl
gently to mix and
immediately add 10
38Total Chlorine
mL of unreacted sample. Replace the cap and shake gently
again.
• Reinsert the cuvet into the instrument.
• Press TIMER and the display will show the countdown prior to the
measurement or, alternatively, wait for 2 minutes and 30 seconds
and press READ DIRECT. In both cases “SIP” will blink during
measurement.
• The instrument directly displays concentration in mg/L of total
chlorine on the Liquid Crystal Display.
Note: free and total chlorine have to be measured separately with
fresh unreacted samples following the related procedure if both
values are requested.
INTERFERENCES
Interference may be caused by: Bromine, Iodine, Ozone, Oxidized
forms of Chromium and Manganese.
In case of water with hardness greater than 500 mg/L CaCO3, shake
the sample for approximately 2 minutes after adding the powder
reagent.
In case of water with alkalinity greater than 250 mg/L CaCO3 or
acidity greater than 150 mg/L CaCO3, the color of the sample may
develop only partially, or may rapidly fade. To resolve this, neutralize
the sample with diluted HCl or NaOH.
39
Total Chlorine
TOTAL CHLORINE HIGH RANGETOTAL CHLORINE HIGH RANGE
TOTAL CHLORINE HIGH RANGE
TOTAL CHLORINE HIGH RANGETOTAL CHLORINE HIGH RANGE
SPECIFICATIONS
Range0.0 to 10.0 mg/L
Resolution0.1 mg/L
Accuracy±0.1 mg/L ±3% of reading
Typical EMC±0.1 mg/L
Deviation
Light SourceTungsten lamp with narrow band interference
filter @ 525 nm
MethodAdaptation of the EPA DPD method 330.5. The
reaction between free chlorine and the DPD reagent causes a pink tint in the sample.
REQUIRED REAGENTS
CodeDescriptionQuantity
HI 93701-0DPD1 packet
HI 93734B-0 Free & Total Chlorine HR Reagent5 mL
HI 93734C-0 Total Chlorine HR Reagent3 drops
REAGENT SETS
HI 93734-01Reagents for 100 tests
HI 93734-03Reagents for 300 tests
For other accessories see page 141.
MEASUREMENT PROCEDURE
• Select the program number corresponding to Total Chlorine HR on
the secondary LCD by pressing PROGRAM W and V.
• Add to the cuvet 5 mL of HI 93734B reagent
by means of the 5 mL syringe.
5 mL
Total Chlorine HR
40
Note: To measure exactly 5 mL of reagent with the syringe, push
the plunger completely into the syringe and insert the tip into
HI 93734B reagent bottle. Pull the plunger out until the lower
edge of the seal is on the 5 mL mark of the syringe.
probable level of liquid
taken up by syringe
5 mL
• Fill the cuvet up to the 10 mL mark
of sample
with 5 mL of unreacted sample, using
the 3 mL plastic pipette.
Note: rinse the 3 mL plastic pipette 2
10 mL
or 3 times with sample before adding it
to the cuvet with reagent.
• Replace the cap and shake gently.
• Place the cuvet into the holder and
ensure that the notch on the cap is
positioned securely into the groove.
• Press ZERO and "SIP" will blink on the display.
• Wait for a few seconds and the display will
show “-0.0-”. Now the meter is zeroed
and ready for measurement.
41Total Chlorine HR
• Remove the cuvet.
• Add 3 drops of HI 93734C reagent to the cuvet.
• Add the content of one
x 3
packet of HI 93701 DPD
reagent to the cuvet.
Replace the cap and
shake gently for 20 seconds (or 2 minutes in
case of seawater analysis).
• Reinsert the cuvet into the instrument.
• Press TIMER and the display will show a countdown prior to the
measurement or, alternatively, wait for 2 minutes and 30 seconds
and press READ DIRECT. The display will show “SIP” during
measurement.
• The instrument directly displays concentration in mg/L of total
chlorine on the Liquid Crystal Display.
INTERFERENCES
Interference may be caused by: Bromine, Iodine, Ozone, Oxidized
forms of Chromium and Manganese.
Alkalinity above 250 mg/L or acidity above 150 mg/L will not
reliably develop the full amount of color or it may rapidly fade. To
resolve this, neutralize the sample with diluted HCl or NaOH.
In case of water with hardness greater than 1000 mg/L CaCO3,
shake the sample for approximately 1 minute after adding the
powder reagent.
Total Chlorine HR
42
CHLORINE DIOXIDECHLORINE DIOXIDE
CHLORINE DIOXIDE
CHLORINE DIOXIDECHLORINE DIOXIDE
SPECIFICATIONS
Range0.00 to 2.00 mg/L
Resolution0.01 mg/L
Accuracy±0.10 mg/L ±5% of reading
Typical EMC±0.01 mg/L
Deviation
Light SourceTungsten lamp with narrow band interference
filter @ 575 nm
MethodAdaptation of the Chlorophenol Red method. The
reaction between chlorine dioxide and reagents
causes a colorless to purple tint in the sample.
REQUIRED REAGENT
CodeDescriptionQuantity
HI 93738A-0Reagent A1 mL
HI 93738B-0Dechlorinating Reagent B1 packet
HI 93738C-0Reagent C1 mL
HI 93738D-0Reagent D1 mL
REAGENT SETS
HI 93738-01 Reagents for 100 tests
HI 93738-03 Reagents for 300 tests
For other accessories see page 141.
MEASUREMENT PROCEDURE
• Select the program number corresponding to Chlorine Dioxide on
the secondary LCD by pressing PROGRAM W and V.
• Fill two graduated mixing cylinders (#1 & #2) up to the
25 mL mark with the sample.
• Add 0.5 mL of HI 93738A chlorine dioxide reagent to each
cylinder (#1 & #2), close them and invert several times to mix.
#1
25 ml
#2
25 ml
#1
43
#1 & #2
#2
Chlorine Dioxide
• Add the content of one packet of HI 93738B dechlorinating
reagent to only one of the two cylinders (#1), close and invert it
several times until it is totally dissolved. This is the blank.
#1
• Add precisely 0.5 mL of HI 93738C chlorine dioxide reagent to
each cylinder (#1 & #2), close them and invert several times to
mix.
#1 & #2
#1
#2
• Add 0.5 mL of HI 93738D chlorine dioxide reagent to each
cylinder (#1 & #2), close them and invert several times to mix.
Cylinder #2 is the reacted sample.
#1 & #2
#1
• Fill a cuvet with 10 mL of the blank
#2
#1
10 mL
(#1) up to the mark and replace the
cap.
• Place the blank (#1) into the holder
#1
and ensure that the notch on the cap
is positioned securely into the groove.
Chlorine Dioxide
44
• Press ZERO and "SIP" will blink on the display.
• Wait for a few seconds and the display will
show "-0.0-". Now the meter is zeroed and
ready for measurement.
• Fill another cuvet with 10 mL of the reacted
sample (#2) up to the mark and replace
the cap.
• Insert the sample into the instrument.
#2
10 mL
#2
• Press READ DIRECT and "SIP" will blink during measurement.
• The instrument directly displays concentration in mg/L of chlorine
dioxide on the Liquid Crystal Display.
SAMPLING PROCEDURE
It is recommended to analyze chlorine dioxide samples immediately
after collection. Chlorine dioxide samples must be stored in dark glass
stoppered bottles, with minimal head space. Excessive heat (above
25°C/78°F), agitation and exposure to light must be avoided.
INTERFERENCES
Interferences may be caused by strong oxidants.
45
Chlorine Dioxide
CHROMIUM VI HIGH RANGECHROMIUM VI HIGH RANGE
CHROMIUM VI HIGH RANGE
CHROMIUM VI HIGH RANGECHROMIUM VI HIGH RANGE
SPECIFICATIONS
Range0 to 1000 µg/L
Resolution1 µg/L
Accuracy±5 µg/L ±4% of reading
Typical EMC±1 µg/L
Deviation
Light SourceTungsten lamp with narrow band interference
filter @ 525 nm
MethodAdaptation of the ASTM Manual of Water and
Environmental Technology, D1687-92,
Diphenylcarbohydrazide method. The reaction
between chromium VI and the reagent causes a
purple tint in the sample.
REQUIRED REAGENTS
CodeDescriptionQuantity
HI 93723-0Powder reagent1 packet
REAGENT SETS
HI 93723-01 Reagents for 100 tests
HI 93723-03 Reagents for 300 tests
For other accessories see page 141.
MEASUREMENT PROCEDURE
• Select the program number corresponding to Chromium VI HR on
the secondary LCD by pressing PROGRAM W and V.
• Fill the cuvet up to the mark with 10 mL of
unreacted sample and replace the cap.
• Place the cuvet into the holder and
ensure that the notch on the cap is
positioned securely into the groove.
• Press ZERO and “SIP” will blink on the display.
• Wait for a few seconds and the display
will show “-0.0-”. Now the meter is
zeroed and ready for measurement.
Chromium VI HR
46
10 mL
• Remove the cuvet and add the
content of one packet of HI
93723 reagent. Replace the cap
and shake vigorously for about
10 seconds.
• Reinsert the cuvet into the instrument.
• Press TIMER and the display will show the countdown prior to the
measurement or, alternatively, wait for 6 minutes and press READ
DIRECT. In both cases "SIP" will blink during measurement.
• The instrument directly displays concentration in µg/L of chromium
VI on the Liquid Crystal Display.
INTERFERENCES
Interference may be caused by:
Vanadium above 1 ppm. However, waiting 10 minutes before
reading, the interference is removed
Iron above 1 ppm
Mercurous and mercuric ions cause slight inhibition of the reaction.
47
Chromium VI HR
CHROMIUM VI LOW RANGECHROMIUM VI LOW RANGE
CHROMIUM VI LOW RANGE
CHROMIUM VI LOW RANGECHROMIUM VI LOW RANGE
SPECIFICATIONS
Range0 to 300 µg/L
Resolution1 µg/L
Accuracy±1 µg/L ±4% of reading
Typical EMC±1 µg/L
Deviation
Light SourceTungsten lamp with narrow band interference
filter @ 525 nm
MethodAdaptation of the ASTM Manual of Water and
Environmental Technology, D1687-92,
Diphenylcarbohydrazide method. The reaction between chromium VI and the reagent causes a
purple tint in the sample.
REQUIRED REAGENTS
CodeDescriptionQuantity
HI 93749-0Powder reagent1 packet
REAGENT SETS
HI 93749-01 Reagents for 100 tests
HI 93749-03 Reagents for 300 tests
For other accessories see page 141.
MEASUREMENT PROCEDURE
• Select the program number corresponding to Chromium VI LR on
the secondary LCD by pressing PROGRAM W and V.
• Fill the cuvet with 10 mL of unreacted sample
(up to the mark) and replace the cap.
• Place the cuvet into the holder and
ensure that the notch on the cap is
positioned securely into the groove.
• Press ZERO and “SIP” will blink on the display.
• Wait for a few seconds and the display will
show “-0.0-”. Now the meter is zeroed
and ready for measurement.
48Chromium VI LR
10 mL
• Remove the cuvet and add the
content of one packet of HI
93749 reagent. Replace the
cap and shake vigorously for
about 10 seconds.
• Reinsert the cuvet into the instrument.
• Press TIMER and the display will show the countdown prior to the
measurement or, alternatively, wait for 6 minutes and press READ
DIRECT. In both cases "SIP" will blink during measurement.
• The instrument directly displays concentration in µg/L of Chromium VI on the Liquid Crystal Display.
INTERFERENCES
Interference may be caused by:
Vanadium above 1 ppm. However, waiting 10 minutes before
reading, the interference is removed
Iron above 1 ppm
Mercurous and mercuric ions cause slight inhibition of the reaction.
49
Chromium VI LR
COLOR OF WATERCOLOR OF WATER
COLOR OF WATER
COLOR OF WATERCOLOR OF WATER
SPECIFICATIONS
Range0 to 500 PCU (Platinum Cobalt Units)
Resolution1 PCU
Accuracy±10 PCU ±5% of reading
Typical EMC± 1 PCU
Deviation
Light SourceTungsten lamp with narrow band interference
filter @ 420 nm
MethodAdaptation of the Standard Methods for the
Examination of Water and Wastewater, 18
edition, Colorimetric Platinum Cobalt method.
REQUIRED ACCESSORIES
0.45 µm membrane for true color measurement.
For other accessories see page 141.
MEASUREMENT PROCEDURE
• Select the program number corresponding to Color of Water on
the secondary LCD by pressing PROGRAM W and V.
th
• Fill one cuvet up to the mark with deionized
water and replace the cap. This is the blank.
• Place the blank (# 1) into the holder
and ensure that the notch on the cap
is positioned securely into the groove.
• Press ZERO and "SIP" will blink on the display.
• Wait for a few seconds and the display will
show "-0.0-". Now the meter is zeroed and
ready for measurement.
• Remove the blank.
Color of Water
50
10 mL
# 1
# 1
• Fill another cuvet up to the mark with
unfiltered sample and replace the cap.
This is the apparent color.
• Filter 10 mL of sample through a filter with a
0.45 µm membrane into the third cuvet, up
to the 10 mL mark and replace the cap. This is
the true color.
10 mL
# 2
# 3
• Insert the apparent color cuvet (# 2)
# 2
into the instrument and ensure that
the notch on the cap is positioned
securely into the groove.
• Press READ DIRECT and "SIP" will blink on the display.
• The meter directly displays the value of apparent color in PCU on
the Liquid Crystal Display.
• Remove the cuvet, insert the true color
# 3
cuvet (# 3) into the instrument and
ensure that the notch on the cap is
positioned securely into the groove.
• Press READ DIRECT and "SIP" will blink on the display.
• The meter directly displays the value of true color in PCU on the
Liquid Crystal Display.
51Color of Water
COPPER HIGH RANGECOPPER HIGH RANGE
COPPER HIGH RANGE
COPPER HIGH RANGECOPPER HIGH RANGE
SPECIFICATIONS
Range0.00 to 5.00 mg/L
Resolution0.01 mg/L
Accuracy±0.02 mg/L ±4% of reading
Typical EMC±0.01 mg/L
Deviation
Light SourceTungsten lamp with narrow band interference
filter @ 575 nm
MethodAdaptation of the EPA method. The reaction
between copper and the bicinchoninate reagent
causes a purple tint in the sample.
REQUIRED REAGENTS
CodeDescriptionQuantity
HI 93702-0Bicinchoninate1 packet
REAGENT SETS
HI 93702-01 Reagents for 100 tests
HI 93702-03 Reagents for 300 tests
For other accessories see page 141.
MEASUREMENT PROCEDURE
• Select the program number corresponding to Copper HR on the
secondary LCD by pressing PROGRAM W and V.
• Fill the cuvet with 10 mL of unreacted
sample (up to the mark) and replace the
cap.
10 ml
• Place the cuvet into the holder and
ensure that the notch on the cap is
positioned securely into the groove.
• Press ZERO and “SIP” will blink on the display.
• Wait for a few seconds and the display will
show “-0.0-”. Now the meter is zeroed
and ready for measurement.
52Copper HR
• Remove the cuvet.
• Add the content of one packet
of HI 93702 reagent. Replace
the cap and shake gently for
about 15 seconds.
• Reinsert the cuvet into the instrument.
• Press TIMER and the display will show the countdown prior to the
measurement or, alternatively, wait for 45 seconds and press
READ DIRECT. In both cases “SIP” will blink during measurement.
• The instrument directly displays concentration in mg/L of copper
on the Liquid Crystal Display.
INTERFERENCES
Interference may be caused by:
Silver
Cyanide
For samples overcoming buffering capacity of reagent (around pH 6.8),
pH should be adjusted between 6 and 8.
53
Copper HR
COPPER LOW RANGECOPPER LOW RANGE
COPPER LOW RANGE
COPPER LOW RANGECOPPER LOW RANGE
SPECIFICATIONS
Range0 to 1000 µg/L
Resolution1 µg/L
Accuracy±10 µg/L ±5% of reading
Typical EMC±1 µg/L
Deviation
Light SourceTungsten lamp with narrow band interference
filter @ 575 nm
MethodAdaptation of the EPA method. The reaction
between copper and the bicinchoninate reagent
causes a purple tint in the sample.
REQUIRED REAGENTS
CodeDescriptionQuantity
HI 93747-0Bicinchoninate1 packet
REAGENT SETS
HI 93747-01 Reagents for 100 tests
HI 93747-03 Reagents for 300 tests
For other accessories see page 141.
MEASUREMENT PROCEDURE
• Select the program number corresponding to Copper LR on the
secondary LCD by pressing PROGRAM W and V.
• Fill the cuvet with 10 mL of unreacted sample (up
to the mark) and replace the cap.
• Place the cuvet into the holder and
ensure that the notch on the cap is
positioned securely into the groove.
• Press ZERO and “SIP” will blink on the display.
• Wait for a few seconds and the display will
show “-0.0-”. Now the meter is zeroed and
ready for measurement.
Copper LR
54
10 mL
• Remove the cuvet.
• Add the content of one packet
of HI 93747 reagent. Replace
the cap and shake gently for
about 15 seconds.
• Reinsert the cuvet into the instrument.
• Press TIMER and the display will show the countdown prior to the
measurement or, alternatively, wait for 45 seconds and press
READ DIRECT. In both cases "SIP" will blink during measurement.
• Multiply the reading on the Liquid Crystal Display by 10 to
obtain the concetration in mg/L of oxygen demand.
INTERFERENCES
Interference may be caused by:
Silver
Cyanide
For samples overcoming buffering capacity of reagent (around pH 6.8),
pH should be adjusted between 6 and 8.
55
Copper LR
CYANIDECYANIDE
CYANIDE
CYANIDECYANIDE
SPECIFICATIONS
Range0.000 to 0.200 mg/L
Resolution0.001 mg/L
Accuracy ±0.005 mg/L ±3% of reading
Typical EMC Dev. ±0.001 mg/L
Light SourceTungsten lamp with narrow band interference
filter @ 610 nm
MethodAdaptation of the Standard Methods for the
Examination of Water and Wastewater, 18
edition, Pyridine-Pyrazolone method. The reac-
tion between cyanide and reagents causes a
blue tint in the sample.
REQUIRED REAGENTS
CodeDescriptionQuantity
HI 93714A-0Reagent A1 spoon
HI 93714B-0Reagent B1 packet
HI 93714C-0Reagent C1 packet
REAGENT SETS
HI 93714-01 Reagents for 100 tests
HI 93714-03 Reagents for 300 tests
For other accessories see page 141.
MEASUREMENT PROCEDURE
• Select the program number corresponding to Cyanide on the
secondary LCD by pressing PROGRAM W and V.
th
• Fill the cuvet with 10 mL of unreacted sample (up
to the mark) and replace the cap.
• Place the cuvet into the holder and
ensure that the notch on the cap is
positioned securely into the groove.
• Press ZERO and “SIP” will blink on the display.
• Wait for a few seconds and the display will
show “-0.0-”. Now the meter is zeroed
and ready for measurement.
Cyanide
56
10 mL
• Remove the cuvet and add 1
level spoon of HI 93714A Cyanide Reagent. Remember to close
the reagent bottle immediately
after use.
Note: Pay attention to the way the spoon is filled:
- do not press the powder;
- do not overfill it.
• Place the HDPE plastic stopper
and cap immediately, to prevent
the escape of chlorine gas which
is developed during the reaction,
and shake gently for 30 seconds.
• Wait for 30 seconds leaving the
cuvet tightly capped and undisturbed, then add the content of
one packet of HI 93714B reagent and shake gently for 10
seconds.
10"
12
9
3
6
• Immediately add the content of
one packet of HI 93714C reagent, replace the cap and shake
vigorously for 20 seconds.
• Reinsert the cuvet into the instrument.
57
Cyanide
• Press TIMER and the display will show the countdown prior to the
measurement or, alternatively, wait for 25 minutes and press
READ DIRECT. In both cases "SIP" will blink during measurement.
Note: Shake gently the cuvet 4 or 5 times during the first 20
minutes of the countdown prior to the measurement. Accuracy is
not affected by undissolved reagent powder.
• The instrument directly displays concentration in mg/L of cyanide
on the Liquid Crystal Display.
• To convert the result in mg/L of Potassium Cyanide (KCN) multiply
by a factor of 2.5.
Note: for most accurate results perform the test at 20-25 °C.
INTERFERENCES
Interference may be caused by large amounts of turbidity that will
cause high readings.
Oxidizing (like chlorine) or reducing agents (such as sulfide or sulfur
dioxide) are known to interfere with the measurement. Distillation will
remove these.
Samples with high pH values should be adjusted to approximately
pH 7 before testing.
CAUTION: cyanides, their solutions, and hydrogen cyanide liberated
by acids, are very poisonous.
Cyanide
58
CYANURIC ACIDCYANURIC ACID
CYANURIC ACID
CYANURIC ACIDCYANURIC ACID
SPECIFICATIONS
Range0 to 80 mg/L
Resolution1 mg/L
Accuracy±1 mg/L ±15% of reading
Typical EMC±1 mg/L
Deviation
Light SourceTungsten lamp with narrow band interference
filter @ 525 nm
MethodAdaptation of the turbidimetric method. The reac-
tion between cyanuric acid and the reagent
causes a white suspension in the sample.
REQUIRED REAGENTS
CodeDescriptionQuantity
HI 93722-0Powder reagent1 packet
REAGENT SETS
HI 93722-01 Reagents for 100 tests
HI 93722-03 Reagents for 300 tests
For other accessories see page 141.
MEASUREMENT PROCEDURE
• Select the program number corresponding to Cyanuric Acid on the
secondary LCD by pressing PROGRAM W and V.
• Fill the cuvet with 10 mL of unreacted sample
(up to the mark) and replace the cap.
10 mL
• Place the cuvet into the holder and
ensure that the notch on the cap is
positioned securely into the groove.
• Press ZERO and “SIP” will blink on the display.
• Wait for a few seconds and the display
will show “-0.0-”. Now the meter is
zeroed and ready for measurement.
59Cyanuric Acid
• Fill a graduated beaker
up to the 25 mL mark
with the sample, add the
content of one packet of
HI 93722 reagent and
swirl gently to mix.
• Fill a second cuvet with 10 mL of the
10 mL
reacted sample up to the mark. Replace
the cap.
• Reinsert the cuvet into the instrument.
• Press TIMER and the display will show the countdown prior to the
measurement or, alternatively, wait for 45 seconds and press
READ DIRECT. In both cases "SIP" will blink during measurement.
• The instrument directly displays concentration in mg/L of cyanuric
acid on the Liquid Crystal Display.
Cyanuric Acid
60
FLUORIDEFLUORIDE
FLUORIDE
FLUORIDEFLUORIDE
SPECIFICATIONS
Range0.00 to 2.00 mg/L
Resolution0.01 mg/L
Accuracy±5% of reading
Typical EMC±0.01 mg/L
Deviation
Light SourceTungsten lamp with narrow band interference
filter @ 575 nm
MethodAdaptation of the Standard Methods for the
Examination of Water and Wastewater, 18
edition, SPADNS method. The reaction between
fluoride and the liquid reagent causes a red tint
in the sample.
REQUIRED REAGENT
CodeDescriptionQuantity
HI 93729-0SPADNS Reagent4 mL
REAGENT SETS
HI 93729-01 Reagents for 100 tests
HI 93729-03 Reagents for 300 tests
For other accessories see page 141.
MEASUREMENT PROCEDURE
• Select the program number corresponding to Fluoride on the
secondary LCD by pressing PROGRAM W and V.
th
• Add 2 mL of HI 93729 SPADNS
reagent to two cuvets.
• Fill one of the cuvets with distilled water
(up to the mark), replace the cap and
invert several times to mix.
• Fill the other cuvet with sample (up to
the mark), replace the cap and invert
several times to mix.
61
2 ml
10 ml
# 1
10 ml
# 2
Fluoride
• Place the cuvet with the reacted distilled
# 1
water (# 1) into the holder and
ensure that the notch on the cap is
positioned securely into the groove.
• Press TIMER and the display will show the countdown prior to
zeroing the blank or, alternatively, wait for two minutes and
press ZERO and "SIP" will blink on the display.
• Wait for a few seconds and the display
will show "-0.0-". Now the meter is
zeroed and ready for measurement.
• Remove the cuvet.
# 2
• Insert the other cuvet (# 2) with the
reacted sample into the instrument.
• Press READ DIRECT and "SIP" will blink on the LCD during
measurement.
• The instrument directly displays concentration in mg/L of fluoride
on the Liquid Crystal Display.
Note: For wastewater or seawater samples, before performing
measurements, distillation is required.
Note: For most accurate results, use two graduated pipettes to
deliver exactly 8 mL of distilled water and 8 mL of sample.
Fluoride
62
INTERFERENCES
Negative interferences may be caused by:
Alkalinity (as CaCO3) above 5000 mg/L
Aluminum above 0.1 mg/L
Iron, ferric above 10 mg/L
Positive interferences may be caused by:
Chloride above 700 mg/L
Phosphate, ortho above 16 mg/L
Sodium hexametaphosphate above 1.0 mg/L
Sulfate above 200 mg/L
Highly colored and turbid samples may require distillation
Highly alkaline samples can be neutralized with nitric acid.
63
Fluoride
CALCIUM HARDNESSCALCIUM HARDNESS
CALCIUM HARDNESS
CALCIUM HARDNESSCALCIUM HARDNESS
SPECIFICATIONS
Range0.00 to 2.70 mg/L
Resolution0.01 mg/L
Accuracy±0.11 mg/L ±5% of reading
Typical EMC±0.01 mg/L
Deviation
Light SourceTungsten lamp with narrow band interference
filter @ 525 nm
MethodAdaptation of the Standard Methods for the
Examination of Water and Wastewater, 18
edition, Calmagite method. The reaction between
calcium and reagents causes a reddish-violet tint
in the sample.
REQUIRED REAGENTS
CodeDescriptionQuantity
HI 93720A-0Ca & Mg indicator0.5 mL
HI 93720B-0Alkali solution0.5 mL
HI 93720C-0EGTA solution1 drop
REAGENT SETS
HI 93720-01 Reagents for 100 tests
HI 93720-03 Reagents for 300 tests
For other accessories see page 141.
MEASUREMENT PROCEDURE
• Select the program number corresponding to Hardness Ca on the
secondary LCD by pressing PROGRAM W and V.
th
• Rinse a graduated beaker
several times with unreacted
sample, before filling it to
the 50 mL mark with the
sample.
• Add 0.5 mL of HI 93720A Calcium
indicator solution and swirl to mix.
• Add 0.5 mL of
HI 93720B Alkali solution and swirl to mix.
Use this solution to rinse
2 cuvets before filling
them up to the 10 mL
mark.
Hardness Ca
64
# 1# 2
• Add 1 drop of HI 93720C EGTA solution to one cuvet
(# 1), replace the cap and invert the cuvet several
times to mix. This is the blank.
# 1
• Place the blank (# 1) into the holder
and ensure that the notch on the cap
is positioned securely into the groove.
• Press ZERO and "SIP" will blink on the display.
• Wait for a few seconds and the display
will show "-0.0-". Now the meter is
zeroed and ready for measurement.
• Remove the blank and insert the
second cuvet (# 2) into the instrument.
• Press READ DIRECT. "SIP" will blink during measurement.
# 1
# 2
• The instrument directly displays concentration in mg/L of calcium
hardness, as CaCO3, on the Liquid Crystal Display.
• To convert the reading to mg/L of calcium (Ca), multiply by a
factor of 0.4.
65Hardness Ca
Note: The test will detect any calcium contamination in the beaker,
measuring syringes or sample cells. To test cleanliness, repeat the test
multiple times until you obtain consistent results.
Note: For better accuracy wash glassware with HCl 6N.
SAMPLE DILUTION
This meter is designed to determine low levels of hardness, typically
found in water purification systems.
When testing some other sources of water, it is not uncommon to come
across levels of hardness that are greater than the range of this
meter.
This problem can be overcome through dilution. Dilutions must be
performed with hardness-free water or the readings will be erroneous.
A dilution to reduce the level of hardness by a factor of one hundred
is performed as follows:
• Fill a 1 mL syringe with the sample.
• Place the syringe in a 50 mL beaker, making sure that the
beaker is clean and empty, and inject 0.5 mL into the beaker.
• Fill the beaker up to the 50 mL mark with hardness-free water.
Now, follow normal measurement procedure. The true value of the
sample is the reading obtained multiplied by a factor of one hundred
(the dilution factor).
The conversion factors to convert readings in mg/L to French degrees
(FD), German degrees (DD) and English degrees (ED) of hardness are
as follows:
1 mg/L = 0.1 FD = 0.0556 DD = 0.07 ED.
INTERFERENCES
Interference may be caused by excessive amounts of heavy metals.
Hardness Ca
66
MAGNESIUM HARDNESSMAGNESIUM HARDNESS
MAGNESIUM HARDNESS
MAGNESIUM HARDNESSMAGNESIUM HARDNESS
SPECIFICATIONS
Range0.00 to 2.00 mg/L
Resolution0.01 mg/L
Accuracy±0.11 mg/L ±5% of reading
Typical EMC±0.02 mg/L
Deviation
Light SourceTungsten lamp with narrow band interference
filter @ 525 nm
MethodAdaptation of the Standard Methods for the
Examination of Water and Wastewater, 18
edition, EDTA colorimetric method. The reaction
between magnesium and reagents causes a reddish-violet tint in the sample.
REQUIRED REAGENTS
CodeDescriptionQuantity
HI 93719A-0Mg indicator0.5 mL
HI 93719B-0Alkali solution0.5 mL
HI 93719C-0EDTA solution1 drop
HI 93719D-0EGTA solution1 drop
REAGENT SETS
HI 93719-01 Reagents for 100 tests
HI 93719-03 Reagents for 300 tests
For other accessories see page 141.
th
MEASUREMENT PROCEDURE
• Select the program number corresponding to Hardness Mg on the
secondary LCD by pressing PROGRAM W and V.
• Rinse a graduated beaker several times
with unreacted sample, before filling it
to the 50 mL mark with the sample.
• Add 0.5 mL of HI 93719A Magnesium
indicator solution, then swirl to mix.
67
Hardness Mg
• Add 0.5 mL of HI 93719B Alkali solution
and swirl to mix. Use this solution to rinse
2 cuvets.
• Fill both cuvets up to the
10 mL10 mL
10 mL mark.
# 1
• Add 1 drop of HI 93719C EDTA
solution to one cuvet (# 1), replace the cap and invert the
cuvet several times to mix. This is
the blank.
• Add 1 drop of HI 93719D EGTA solution to the
second cuvet (# 2), replace the cap and invert
the cuvet several times to mix. This is the
sample.
• Place the blank (# 1) into the holder
and ensure that the notch on the cap
is positioned securely into the groove.
# 2
# 1
# 2
# 1
• Press ZERO and "SIP" will blink on the display.
• Wait for a few seconds and the display
will show "-0.0-". Now the meter is zeroed and ready for measurement.
68Hardness Mg
• Remove the blank (# 1) and insert
the sample (# 2) into the instrument,
# 2
making sure that the notch on the cap
is positioned securely into the groove.
• Press READ DIRECT. “SIP” will blink during measurement.
• The instrument directly displays concentration in mg/L of magnesium hardness, as CaCO3, on the Liquid Crystal Display.
• To convert the result to mg/L of magnesium (Mg), multiply by a
factor of 0.243.
Note: The test will detect any magnesium contamination in the
beakers, measuring syringes or sample cells. To test cleanliness,
repeat the test multiple times until you obtain consistent results.
SAMPLE DILUTION
This meter is designed to determine hardness typically found in water
purification systems. In order to measure samples with high hardness,
follow dilution procedure explained on page 66 (Ca Hardness).
The conversion factors to convert readings in mg/L to French degrees
(FD), German degrees (DD) and English degrees (ED) of hardness are
as follows:
1 mg/L = 0.1 FD = 0.0556 DD = 0.07 ED
INTERFERENCES
Interference may be caused by excessive amounts of heavy metals.
69
Hardness Mg
TOTAL HARDNESS HIGH RANGE TOTAL HARDNESS HIGH RANGE
TOTAL HARDNESS HIGH RANGE
TOTAL HARDNESS HIGH RANGE TOTAL HARDNESS HIGH RANGE
SPECIFICATIONS
Range400 to 750 mg/L
Resolution5 mg/L
Accuracy±10 mg/L ±2% of reading
Typical EMC±5 mg/L
Deviation
Light SourceTungsten lamp with narrow band interference
filter @ 466 nm
MethodAdaptation of the EPA method 130.1. The reac-
tion between calcium, magnesium and reagents
causes a red-violet tint in the sample.
REQUIRED REAGENTS
CodeDescriptionQuantity
HI 93735A-HRHardness Indicator reagent HR 9.5 mL
HI 93735B-0Hardness Buffer reagent2 drops
HI 93735C-0Fixing reagent1 packet
REAGENT SETS
HI 93735-02 Reagents for 100 tests High Range
(400 to 750 mg/L)
HI 93735-0 Reagents for 100 tests (0 to 750 mg/L)
For other accessories see page 141.
MEASUREMENT PROCEDURE
• Select the program number corresponding to Total Hardness HR
on the secondary LCD by pressing PROGRAM W and V.
• Using the syringe, add exactly 0.5 mL of
unreacted sample to the cuvet.
• With the plastic pipette fill the cuvet
up to the 10 mL mark adding HI
93735A-HR Indicator reagent.
Total Hardness HR
70
• Add 2 drops of HI 93735B Buffer reagent.
• Replace the cap, shake
gently to mix and wait
for 15 seconds.
• Place the cuvet into the holder and
ensure that the notch on the cap is
positioned securely into the groove.
• Press ZERO and "SIP" will blink on the display.
• Wait for a few seconds and the display will
show "-0.0-". Now the meter is zeroed and
ready for measurement.
x 2
• Remove the cuvet and add the content of 1 packet
of HI 93735C Fixing reagent.
• Replace the cap and
shake gently till
dissolution is complete.
• Place the cuvet into the holder and
ensure that the notch on the cap is
positioned securely into the groove.
71
Total Hardness HR
• Press TIMER and the display will show the countdown prior to the
measurement or, alternatively, wait for 2 minutes and press READ
DIRECT. In both cases "SIP" will blink during measurement.
• The instrument directly displays the total hardness in mg/L of
CaCO3, on the Liquid Crystal Display.
• The conversion factors to convert readings in mg/L to French
Degrees (FD), German Degrees (DD) and English Degrees (ED) of
hardness are as follows:
1 mg/L= 0.1 FD = 0.0556 DD = 0.07 ED
INTERFERENCES
Interference may be caused by excessive amounts of heavy metals.
Note: if the sample is very acidic, some extra drops of HI 93735B
Buffer reagent may be added.
72Total Hardness HR
TOTAL HARDNESS MEDIUM RANGE TOTAL HARDNESS MEDIUM RANGE
TOTAL HARDNESS MEDIUM RANGE
TOTAL HARDNESS MEDIUM RANGE TOTAL HARDNESS MEDIUM RANGE
SPECIFICATIONS
Range200 to 500 mg/L
Resolution5 mg/L
Accuracy±7 mg/L ±3% of reading
Typical EMC±5 mg/L
Deviation
Light SourceTungsten lamp with narrow band interference
filter @ 466 nm
MethodAdaptation of the EPA method 130.1. The reac-
tion between calcium, magnesium and reagents
causes a red-violet tint in the sample.
REQUIRED REAGENTS
CodeDescriptionQuantity
HI 93735A-MRHardness Indicator reagent MR 9.5 mL
HI 93735B-0Hardness Buffer reagent2 drops
HI 93735C-0Fixing reagent1 packet
REAGENT SETS
HI 93735-01 Reagents for 100 tests Medium Range
(200 to 500 mg/L)
HI 93735-0 Reagents for 100 tests (0 to 750 mg/L)
For other accessories see page 141.
MEASUREMENT PROCEDURE
• Select the program number corresponding to Total Hardness MR
on the secondary LCD by pressing PROGRAM W and V.
• Using the syringe, add exactly 0.5 mL of
unreacted sample to the cuvet.
• With the plastic pipette
fill the cuvet up to the
10 mL mark adding HI
93735A-MR Indicator reagent.
73
Total Hardness MR
• Add 2 drops of HI 93735B
Buffer reagent
x 2
• Replace the cap and shake gently to mix.
Wait for 15 seconds.
• Place the cuvet into the holder and
ensure that the notch on the cap is
positioned securely into the groove.
• Press ZERO and "SIP" will blink on the display.
• Wait for a few seconds and the display will
show "-0.0-". Now the meter is zeroed and
ready for measurement.
• Remove the cuvet and add the content of 1 packet
of HI 93735C Fixing reagent.
• Replace the cap and shake
gently till dissolution is complete.
• Place the cuvet into the holder and
ensure that the notch on the cap is
positioned securely into the groove.
74Total Hardness MR
• Press TIMER and the display will show the countdown prior to the
measurement or, alternatively, wait for 2 minutes and press READ
DIRECT. In both cases "SIP" will blink during measurement.
• The instrument directly displays the hardness in mg/L of CaCO3,
on the Liquid Crystal Display.
• The conversion factors to convert readings in mg/L to French
Degrees (FD), German Degrees (DD) and English Degrees (ED) of
hardness are as follows:
1 mg/L= 0.1 FD = 0.0556 DD = 0.07 ED
INTERFERENCES
Interference may be caused by excessive amounts of heavy metals.
Note: if the sample is very acidic, some extra drops of HI 93735B
Buffer reagent may be added.
75Total Hardness MR
TOTAL HARDNESS LOW RANGE TOTAL HARDNESS LOW RANGE
TOTAL HARDNESS LOW RANGE
TOTAL HARDNESS LOW RANGE TOTAL HARDNESS LOW RANGE
SPECIFICATIONS
Range0 to 250 mg/L
Resolution1 mg/L from 0 to 100 mg/L
5 mg/L from 100 to 250 mg/L
Accuracy±5 mg/L ±4% of reading
Typical EMC±5 mg/L
Deviation
Light SourceTungsten lamp with narrow band interference
filter @ 466 nm
MethodAdaptation of the EPA method 130.1. The reac-
tion between calcium, magnesium and reagents
causes a red-violet tint in the sample.
REQUIRED REAGENTS
CodeDescriptionQuantity
HI 93735A-LRHardness Indicator reagent LR9.5 mL
HI 93735B-0Hardness Buffer reagent2 drops
HI 93735C-0Fixing reagent1 packet
REAGENT SETS
HI 93735-00 Reagents for 100 tests Low Range (0 to 250 mg/L)
HI 93735-0 Reagents for 100 tests (0 to 750 mg/L)
For other accessories see page 141.
MEASUREMENT PROCEDURE
• Select the program number corresponding to Total Hardness LR
on the secondary LCD by pressing PROGRAM W and V.
• Using the syringe, add exactly 0.5 mL of
unreacted sample to the cuvet.
• With the plastic pipette fill the
cuvet up to the 10 mL mark
adding HI 93735A-LR Indicator
reagent.
Total Hardness LR
76
• Add 2 drops of HI 93735B-0
Buffer reagent.
x 2
• Replace the cap and shake gently to
mix. Wait for 15 seconds.
• Place the cuvet into the holder and
ensure that the notch on the cap is
positioned securely into the groove.
• Press ZERO and "SIP" will blink on the display.
• Wait for a few seconds and the display
will show "-0.0-". Now the meter is
zeroed and ready for measurement.
• Remove the cuvet and add the content of 1
packet of HI 93735C Fixing reagent.
• Replace the cap and shake gently till
dissolution is complete.
77
Total Hardness LR
• Place the cuvet into the holder and
ensure that the notch on the cap is
positioned securely into the groove.
• Press TIMER and the display will show the countdown prior to the
measurement or, alternatively, wait for 2 minutes and press READ
DIRECT. In both cases "SIP" will blink during measurement.
• The instrument directly displays the hardness in mg/L of CaCO3,
on the Liquid Crystal Display.
• The conversion factors to convert readings in mg/L to French
Degrees (FD), German Degrees (DD) and English Degrees (ED) of
hardness are as follows:
1 mg/L= 0.1 FD = 0.0556 DD = 0.07 ED
INTERFERENCES
Interference may be caused by excessive amounts of heavy metals.
Note: if the sample is very acidic, some extra drops of HI 93735B
Buffer reagent may be added.
78Total Hardness LR
HYDRAZINEHYDRAZINE
HYDRAZINE
HYDRAZINEHYDRAZINE
SPECIFICATIONS
Range0 to 400 µg/L
Resolution1 µg/L
Accuracy±4% of full scale
Typical EMC±2 µg/L
Deviation
Light SourceTungsten lamp with narrow band interference
filter @ 420 nm
MethodAdaptation of the ASTM Manual of Water and
Environmental Technology, method D1385-88,
p-Dimethylaminobenzaldehyde method. The reaction between hydrazine and the liquid reagent
causes a yellow tint in the sample.
REQUIRED REAGENT
CodeDescriptionQuantity
HI 93704-0Liquid Reagent24 drops
REAGENT SETS
HI 93704-01 Reagents for 100 tests
HI 93704-03 Reagents for 300 tests
For other accessories see page 141.
MEASUREMENT PROCEDURE
• Select the program number corresponding to Hydrazine on the
secondary LCD by pressing PROGRAM W and V.
• Fill one cuvet up to the mark with 10
mL of distilled water.
• Place the cap and insert the cuvet
# 1 into the holder. Ensure that
the notch on the cap is positioned
securely into the groove.
• Press ZERO and SIP will blink while the
instrument is adjusting the light level.
79
10 mL
# 1
# 1
Hydrazine
• Wait for a few seconds and the display will
show a blinking "good". Now the meter is
ready to take a zero measurement.
• Fill a second cuvet with 10 mL of unreacted
sample (up to the mark) and replace the cap.
10 mL
# 2
• Add 12 drops of the
HI 93704 reagent
to each cuvet. Replace the caps and
shake gently to mix.
x 12
# 1# 2
• Place the blank (#1) into the holder
and ensure that the notch on the cap
x 12
# 1
is positioned securely into the groove.
• Press TIMER and the display will show the countdown prior to
zeroing the blank. "SIP" will blink during zeroing.
• Wait for a few seconds and the display will
show “-0.0-”. Now the meter is zeroed
and ready for measurement.
• Remove the blank.
80Hydrazine
• Insert the cuvet with the reacted sample
(# 2) into the instrument. Ensure
# 2
that the notch on the cap is positioned
securely into the groove.
• Press READ DIRECT and “SIP” will blink during measurement.
• The instrument directly displays concentration in µg/L of hydrazine on the Liquid Crystal Display.
INTERFERENCES
Interference may be caused by:
Highly colored samples
Highly turbid samples
Aromatic amines
81
Hydrazine
IODINEIODINE
IODINE
IODINEIODINE
SPECIFICATIONS
Range0.0 to 12.5 mg/L
Resolution0.1 mg/L
Accuracy±0.1 mg/L ±5% of reading
Typical EMC±0.1 mg/L
Deviation
Light SourceTungsten lamp with narrow band interference
filter @ 525 nm
MethodAdaptation of the Standard Methods for the
Examination of Water and Wastewater, 18
edition, DPD method. The reaction between iodine
and the reagent causes a pink tint in the
sample.
REQUIRED REAGENTS
CodeDescriptionQuantity
HI 93718-0DPD Reagent1 packet
REAGENT SETS
HI 93718-01 Reagents for 100 tests
HI 93718-03 Reagents for 300 tests
For other accessories see page 141.
MEASUREMENT PROCEDURE
• Select the program number corresponding to Iodine on the
secondary LCD by pressing PROGRAM W and V.
th
• Fill the cuvet with 10 mL of unreacted sample
(up to the mark) and replace the cap.
• Place the cuvet into the holder and
ensure that the notch on the cap is
positioned securely into the groove.
• Press ZERO and “SIP” will blink on the display.
Iodine
82
10 mL
• Wait for a few seconds and the display will
show “-0.0-”. Now the meter is zeroed
and ready for measurement.
• Remove the cap and add the
content of one packet of HI
93718 reagent. Replace the
cap and shake gently for about
30 seconds to dissolve most of
the reagent.
• Reinsert the cuvet into the instrument.
• Press TIMER and the display will show the countdown prior to the
measurement or, alternatively, wait for 2 minutes and 30 seconds
and press READ DIRECT. In both cases "SIP" will blink during
measurement.
• The instrument directly displays concentration in mg/L of iodine
on the Liquid Crystal Display.
INTERFERENCES
Interference may be caused by: Bromine, Chlorine, Ozone, Oxidized
forms of Chromium and Manganese.
In case of water with hardness greater than 500 mg/L CaCO3, shake
the sample for approximately 2 minutes after adding the reagent.
In case of water with alkalinity greater than 250 mg/L CaCO3 or
acidity greater than 150 mg/L CaCO3, the color of the sample may
develop only partially, or may rapidly fade. To resolve this, neutralize
the sample with diluted HCl or NaOH.
83
Iodine
IRON HIGH RANGEIRON HIGH RANGE
IRON HIGH RANGE
IRON HIGH RANGEIRON HIGH RANGE
SPECIFICATIONS
Range0.00 to 5.00 mg/L
Resolution0.01 mg/L
Accuracy±0.04 mg/L ±2% of reading
Typical EMC±0.01 mg/L
Deviation
Light SourceTungsten lamp with narrow band interference
filter @ 525 nm
MethodAdaptation of the EPA Phenantroline method
315B, for natural and treated waters. The reac-
tion between iron and reagents causes an orange
tint in the sample.
REQUIRED REAGENTS
CodeDescriptionQuantity
HI 93721-0Powder Reagent1 packet
REAGENT SETS
HI 93721-01 Reagents for 100 tests
HI 93721-03 Reagents for 300 tests
For other accessories see page 141.
MEASUREMENT PROCEDURE
• Select the program number corresponding to Iron HR on the
secondary LCD by pressing PROGRAM W and V.
• Fill the cuvet with 10 mL of unreacted sample
(up to the mark) and replace the cap.
• Place the cuvet into the holder and
ensure that the notch on the cap is
positioned securely into the groove.
• Press ZERO and “SIP” will blink on the display.
• Wait for a few seconds and the display will
show “-0.0-”. Now the meter is zeroed
and ready for measurement.
84Iron HR
10 mL
• Remove the cuvet and add the content of one packet of HI 93721
reagent. Replace the cap and shake
until dissolution is complete.
• Reinsert the cuvet into the instrument.
• Press TIMER and the display will show the countdown prior to the
measurement or, alternatively, wait for 3 minutes and press READ
DIRECT. In both cases "SIP" will blink during measurement.
• The instrument directly displays concentration in mg/L of iron on
the Liquid Crystal Display.
Range0 to 400 µg/L
Resolution1 µg/L
Accuracy±10 µg/L ±8% of reading
Typical EMC±1 µg/L
Deviation
Light SourceTungsten lamp with narrow band interference
filter @ 575 nm
MethodAdaptation of the TPTZ Method. The reaction
between iron and the reagent causes a violet tint
in the sample.
REQUIRED REAGENTS
CodeDescriptionQuantity
HI 93746-0TPTZ Reagent2 packets
REAGENT SETS
HI 93746-01 Reagents for 50 tests
HI 93746-03 Reagents for 100 tests
For other accessories see page 141.
MEASUREMENT PROCEDURE
• Select the program number corresponding to Iron LR on the
secondary LCD by pressing PROGRAM W and V.
• Fill one graduated mixing cylinder up to
the 25 mL mark with deionized water.
• Add the content of one
packet of HI 93746 TPTZ
reagent, close the cylinder and shake vigorously
for 30 seconds. This is
the blank.
• Fill a cuvet with 10 mL of the blank up
to the mark and replace the cap.
86Iron LR
10 mL
12
9
6
25 ml
30"
3
• Place the cuvet into the holder and
ensure that the notch on the cap is
positioned securely into the groove.
• Press ZERO and "SIP" will blink on the display.
• Wait for a few seconds and the display
will show "-0.0-". Now the meter is
zeroed and ready for measurement.
• Remove the cuvet.
• Fill another graduated mixing cylinder up to the 25 mL mark with the
sample.
• Add the content of one packet
of HI 93746 TPTZ reagent,
close the cylinder and shake
vigorously for 30 seconds. This
is the reacted sample.
• Fill a cuvet with 10 mL of the reacted
sample up to the mark and replace the
cap.
10 mL
25 ml
12
9
6
30"
3
87
Iron LR
• Insert the sample into the instrument.
• Press TIMER and the display will show the countdown prior to the
measurement or, alternatively, wait for 30 seconds and press
READ DIRECT. In both cases "SIP" will blink during measurement.
• The instrument directly displays concentration in µg/L of iron on
the Liquid Crystal Display.
INTERFERENCES
Interference may be caused by:
Cadmium above 4.0 mg/L
3+
Chromium
Chromium
Cobalt above 0.05 mg/L
Copper above 0.6 mg/L
Cyanide above 2.8 mg/L
Manganese above 50.0 mg/L
Mercury above 0.4 mg/L
Molybdenum above 4.0 mg/L
Nickel above 1.0 mg/L
Nitrite ion above 0.8 mg/L
Sample pH should be between 3 and 4 to avoid developed color to
fade or turbidity formation.
above 0.25 mg/L
6+
above 1.2 mg/L
88Iron LR
MANGANESE HIGH RANGEMANGANESE HIGH RANGE
MANGANESE HIGH RANGE
MANGANESE HIGH RANGEMANGANESE HIGH RANGE
SPECIFICATIONS
Range0.0 to 20.0 mg/L
Resolution0.1 mg/L
Accuracy±0.2 mg/L ±3% of reading
Typical EMC±0.1 mg/L
Deviation
Light SourceTungsten lamp with narrow band interference
filter @ 525 nm
MethodAdaptation of the Standard Methods for the
Examination of Water and Wastewater, 18
edition, Periodate method. The reaction between
manganese and reagents causes a pink tint in
the sample.
REQUIRED REAGENTS
CodeDescriptionQuantity
HI 93709A-0Citrate1 packet
HI 93709B-0Sodium periodate1 packet
REAGENT SETS
HI 93709-01 Reagents for 100 tests
HI 93709-03 Reagents for 300 tests
For other accessories see page 141.
MEASUREMENT PROCEDURE
• Select the program number corresponding to Manganese HR on
the secondary LCD by pressing PROGRAM W and V.
th
• Fill the cuvet with 10 mL of unreacted sample (up
to the mark) and replace the cap.
• Place the cuvet into the holder and
ensure that the notch on the cap is
positioned securely into the groove.
• Press ZERO and “SIP” will blink on the display.
89
10 mL
Manganese HR
• Wait for a few seconds and the display will
show “-0.0-”. Now the meter is zeroed
and ready for measurement.
• Remove the cuvet.
• Add the content of one packet of
HI 93709A Citrate reagent. Replace the cap and shake gently
until dissolution is complete.
• Add the content of one packet of
HI 93709B Sodium Periodate reagent. Replace the cap and shake
gently until dissolution is complete.
• Reinsert the cuvet into the instrument.
• Press TIMER and the display will show the countdown prior to the
measurement or, alternatively, wait for 1 minute and 30 seconds
and press READ DIRECT. In both cases “SIP” will blink during
measurement.
• The instrument directly displays concentration in mg/L of manganese
on the Liquid Crystal Display.
INTERFERENCES
Interference may be caused by:
Calcium above 700 mg/L
Chloride above 70000 mg/L
Iron above 5 mg/L
Magnesium above 100000 mg/L.
Manganese HR
90
MANGANESE LOW RANGEMANGANESE LOW RANGE
MANGANESE LOW RANGE
MANGANESE LOW RANGEMANGANESE LOW RANGE
SPECIFICATIONS
Range0 to 300 µg/L
Resolution1 µg/L
Accuracy±2 µg/L ±3% of reading
Typical EMC±1 µg/L
Deviation
Light SourceTungsten lamp with narrow band interference
filter @ 575 nm
MethodAdaptation of the PAN Method. The reaction
between manganese and the reagents causes an
orange tint in the sample.
REQUIRED REAGENT
CodeDescriptionQuantity
HI 93748A-0Ascorbic acid2 packets
HI 93748B- 0Alkaline-cyanide sol.0.40 mL
HI 93748C-00.1% PAN indicator2 mL
HI 93703-51Dispersing Agent4-6 drops (only when
necessary, see note)
REAGENT SETS
HI 93748-01 Reagents for 50 tests
HI 93748-03 Reagents for 150 tests
For other accessories see page 141.
MEASUREMENT PROCEDURE
• Select the program number corresponding to Manganese LR on
the secondary LCD by pressing PROGRAM W and V.
10 mL
• Fill one cuvet with 10 mL of deionized water
(up to the mark).
• Fill a second cuvet with 10 mL of sample (up to
the mark).
• Add the content of one packet
of HI 93748A Ascorbic acid to
each cuvet, replace the caps
and shake gently until dissolution
is complete.
91
10 mL
Manganese LR
• Add 0.2 mL of the
HI 93748B Alkaline-cyanide reagent solution to
each cuvet, replace the
caps and shake gently.
• Add 1 mL of the
HI 93748C 0.1% PAN indicator solution to each
cuvet, replace the caps
and shake gently.
• Place the cuvet with the reacted deionized water (blank) into the holder and
ensure that the notch on the cap is
positioned securely into the groove.
• Press TIMER and the display will show the countdown prior to
zeroing the blank. Alternatively wait for 2 minutes and then
press ZERO. "SIP" will blink during zeroing.
• Wait for a few seconds and the display will
show "-0.0-". Now the meter is zeroed and
ready for measurement.
• Insert the second cuvet with the reacted
sample into the instrument.
Manganese LR
92
• Press READ DIRECT. "SIP" will blink during measurement.
• The instrument directly displays concentration in µg/L of manganese on the display.
Note: a temperature above 30°C may cause turbidity. In this
case, before zeroing and taking readings, add 2-3 drops of
Dispersing Agent (HI 93703-51) to each cuvet and swirl until
complete dissolution of the turbidity.
INTERFERENCES
Interference may be caused by:
Aluminum above 20 mg/L
Cadmium above 10 mg/L
Calcium above 200 mg/L as CaCO
Cobalt above 20 mg/L
Copper above 50 mg/L
Iron above 10 mg/L
Lead above 0.5 mg/L
Magnesium above 100 mg/L as CaCO
Nickel above 40 mg/L
Zinc above 15 mg/L.
3
3
93
Manganese LR
MOLYBDENUMMOLYBDENUM
MOLYBDENUM
MOLYBDENUMMOLYBDENUM
SPECIFICATIONS
Range0.0 to 40.0 mg/L
Resolution0.1 mg/L
Accuracy±0.3 mg/L ±5% of reading
Typical EMC±0.1 mg/L
Deviation
Light SourceTungsten lamp with narrow band interference
filter @ 420 nm
MethodAdaptation of the mercaptoacetic acid method.
The reaction between molybdenum and the reagents causes a yellow tint in the sample.
REQUIRED REAGENT
CodeDescriptionQuantity
HI 93730A-0Reagent A1 packet
HI 93730B-0Reagent B1 packet
HI 93730C-0Reagent C1 packet
REAGENT SETS
HI 93730-01 Reagents for 100 tests
HI 93730-03 Reagents for 300 tests
For other accessories see page 141.
MEASUREMENT PROCEDURE
• Select the program number corresponding to Molybdenum on the
secondary LCD by pressing PROGRAM W and V.
• Fill the cuvet with 10 mL of unreacted sample
(up to the mark) and replace the cap.
10 mL
• Place the cuvet into the holder and
ensure that the notch on the cap is
positioned securely into the groove.
• Press ZERO and “SIP” will blink on the display.
94Molybdenum
• Wait for a few seconds and the display will
show “-0.0-”. Now the meter is zeroed
and ready for measurement.
• Fill one graduated mixing cylinder up to the 25
mL mark with the sample.
• Add the content of one
packet of HI 93730A molybdenum reagent, close
the cylinder and invert it
several times until dissolution is complete.
• Add the content of one
packet of HI 93730B molybdenum reagent to the
cylinder, close and invert
it several times until dissolution is complete.
• Add the content of one
packet of HI 93730C molybdenum reagent to the
cylinder, close and shake
it vigorously.
25 mL
• Fill an empty cuvet with 10 mL of
sample up to the mark and replace the cap.
• Insert the cuvet into the instrument.
95
10 mL
Molybdenum
• Press TIMER and the display will show the countdown prior to the
measurement or, alternatively, wait for five minutes and press
READ DIRECT. In both cases, "SIP" will blink during measurement.
• The instrument directly displays concentration in mg/L of molybdenum on the Liquid Crystal Display.
INTERFERENCES
Interference may be caused by:
Aluminum above 50 mg/L
Chromium above 1000 mg/L
Copper above 10 mg/L
Iron above 50 mg/L
Nickel above 50 mg/L
Nitrite, as NO
Sulfate above 200 mg/L
Highly buffered samples or with extreme pH may exceed the buffering
capacity of the reagents.
–
2
96Molibdenum
NICKEL HIGH RANGENICKEL HIGH RANGE
NICKEL HIGH RANGE
NICKEL HIGH RANGENICKEL HIGH RANGE
SPECIFICATIONS
Range0.00 to 7.00 g/L
Resolution0.01 g/L
Accuracy±4% of reading
Typical EMC±0.02 g/L
Deviation
Light SourceTungsten lamp with narrow band interference
filter @ 575 nm
MethodAdaptation of the photometric method. The reaction
between nickel and the reagent causes a blue
tint in the sample.
REQUIRED REAGENTS
CodeDescriptionQuantity
HI 93726-0Powder reagent1 packet
REAGENT SETS
HI 93726-01 Reagents for 100 tests
HI 93726-03 Reagents for 300 tests
For other accessories see page 141.
MEASUREMENT PROCEDURE
• Select the program number corresponding to Nickel HR on the
secondary LCD by pressing PROGRAM W and V.
• Fill the cuvet up to the mark with 10 mL of
unreacted sample and replace the cap.
10 mL
• Place the cuvet into the holder and
ensure that the notch on the cap is
positioned securely into the groove.
• Press ZERO and “SIP” will blink on the display.
• Wait for a few seconds and the display will
show “-0.0-”. Now the meter is zeroed
and ready for measurement.
97
Nickel HR
• Remove the cuvet and add the
content of one packet of HI
93726 reagent. Replace the cap
and shake gently until dissolution is complete.
• Reinsert the cuvet into the instrument.
• Press TIMER and the display will show the countdown prior to the
measurement or, alternatively, wait for 1 minute and press READ
DIRECT. In both cases "SIP" will blink during measurement.
• The instrument directly displays concentration in g/L of nickel on
the Liquid Crystal Display.
INTERFERENCES
Interference may be caused by copper.
98Nickel HR
NICKEL LOW RANGENICKEL LOW RANGE
NICKEL LOW RANGE
NICKEL LOW RANGENICKEL LOW RANGE
SPECIFICATIONS
Range0.000 to 1.000 mg/L
Resolution0.001 mg/L
Accuracy ±0.010 mg/L ±7% of reading
Typical EMC±0.001 mg/L
Deviation
Light SourceTungsten lamp with narrow band interference
filter @ 575 nm
MethodAdaptation of the PAN method. The reaction
between nickel and the reagents causes an
orange tint in the sample.
REQUIRED REAGENTS
CodeDescriptionQuantity
HI 93740A-0Phthalate-phosphate2 packets
HI 93740B-00.3% PAN indicator2 mL
HI 93740C-0EDTA2 packets
HI 93703-51Dispersing Agent4-6 drops (only when
necessary, see note)
REAGENT SETS
HI 93740-01 Reagents for 50 tests
HI 93740-03 Reagents for 150 tests
For other accessories see page 141.
MEASUREMENT PROCEDURE
• Select the program number corresponding to Nickel LR on the
secondary LCD by pressing PROGRAM W and V.
Note: for best results perform your tests between 20-24°C.
• Fill one graduated beaker with 25 mL of
deionized water (blank) and another one with
25 mL of sample.
• Add the content of one packet of HI 93740A
Phthalate-phosphate reagent to each beaker.
Cap and swirl gently until the reagent is
dissolved.
Note:If sample contains iron (Fe3+), it is
important that all powder be dissolved completely before continuing with following step.
99
Nickel LR
• Add 1 mL of HI 93740B 0.3% PAN
solution to each beaker, cap and
swirl to mix.
• Press TIMER and the display will show a countdown prior to
adding reagent C or, alternatively, wait for 15 minutes. Then, in
both cases, add one packet of HI 93740C EDTA reagent to each
beaker, cap and swirl to mix until complete dissolution.
• Fill one cuvet up to the mark with 10 mL of
the blank.
• Place the cuvet into the holder and
ensure that the notch on the cap is
positioned securely into the groove.
• Press ZERO and "SIP" will appear on the display.
• Wait for a few seconds and the display will
show "-0.0-". Now the meter is zeroed and
ready for measurement.
• Fill a second cuvet up to the mark with 10 mL
of the reacted sample.
• Insert the second cuvet into the instrument.
10 mL
10 mL
100Nickel LR
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.