Instruction Manual
C9800
REACTOR
www.hannainst.com
WARRANTYWARRANTY
WARRANTY
WARRANTYWARRANTY
C9800 is warranted for two years against defects in workmanship
and materials when used for its intended purpose and maintained
according to instructions.
This warranty is limited to repair or replacement free of charge.
Damages due to accident, misuse, tampering or lack of prescribed
maintenance are not covered.
If service is required, contact the dealer from whom you purchased the
instrument. If under warranty, report the model number, date of purchase, serial number and the nature of the failure. If the repair is not
covered by the warranty, you will be notified of the charges incurred. If
the instrument is to be returned to Hanna Instruments, first obtain a
Returned Goods Authorization Number from the Customer Service department and then send it with shipment costs prepaid. When shipping
any instrument, make sure it is properly packaged for complete protection.
To validate your warranty, fill out and return the enclosed warranty card
within 14 days from the date of purchase.
All rights are reserved. Reproduction in whole or in part is prohibited
without the written consent of the copyright owner, Hanna Instruments Inc., Woonsocket, Rhode Island, 02895 , USA.
This Instrument is in
Compliance with the CE Directives
Dear Customer,
Thank you for choosing a Hanna product. Please read this
instruction manual carefully before using the reactor. This
manual will provide you with the necessary information for
the correct use of the heater. If you need additional
technical information, do not hesitate to e-mail us at
tech@hannainst.com. These instruments are in compliance
directives.
with
PRELIMINARY EXAMINATIONPRELIMINARY EXAMINATION
PRELIMINARY EXAMINATION
PRELIMINARY EXAMINATIONPRELIMINARY EXAMINATION
Remove the instrument from the packing material and
examine it carefully to make sure that no damage has
occurred during shipment. If there is any damage, notify
your Dealer.
Note: Save all packing material until you are sure that
the instrument functions correctly. Any defective item
must be returned in their original packing with the
supplied accessories.
GENERAL DESCRIPTIONGENERAL DESCRIPTION
GENERAL DESCRIPTION
GENERAL DESCRIPTIONGENERAL DESCRIPTION
C9800 Reactor is a Hanna Instruments thermo-reactor
with a metal dry-block and a metal case construction.
It allows to warm up to 25 samples, contained into 16
mm-diameter tubes, at 105 or 150 ºC.
The reactor is equipped with led indicators, to inform when
power is on and heating is operating, and a 120 minutes
timer, very useful in timed digestion procedures.
It is possible to select time intervals between 0 and 120
minutes and, during these intervals, temperature in the
heater is maintained constant. At the end of the selected
interval an acoustic alarm will sound and the heater will
automatically shut off.
The reactor is also provided with a thermal fuse that
prevents from overheating, by shutting down the heater.
Two models are available:
C9800-01 runs with 115 VAC mains line
C9800-02 runs with 230 VAC mains line.
HEALTH & SAFETYHEALTH & SAFETY
HEALTH & SAFETY
HEALTH & SAFETYHEALTH & SAFETY
Before unpacking, setting up or start using this reactor it is
important to read carefully this manual. Pay particular
attention to all warnings, cautions and notes. Failure to do
so could result in serious injury to the operator or damage
to the equipment.
To ensure a safety use of this instrument, it must not be
installed or used in any manner other than that specified
in this manual.
Reactor Operation: The Reactor incubates glass
vials containing hazardous chemicals at high temperature. Read the Health & Safety Data Sheet
(HSDS) before handling reagent vials.
Handling Reagent Vials: Handle the vials very carefully
and wear protective clothing. After use, dispose of the
reagent vials according to local regulations.
Safety Equipment: When operating with Reactor use
protective clothing, including goggles or face mask, and
gloves.
Use of the optional safety shield is highly recommended.
Reagent Spills: Wipe up spilled reagents immediately. If
reagent contacts skin, rinse the affected area thoroughly
with water. Avoid breathing released vapors. Read the
Health & Safety Data Sheet (HSDS) of each reagent for
complete chemical information.
Fire Hazard: Avoid the presence of flammable liquids or
paper near the operating Reactor. A fire hazard could be
created.
SPECIFICATIONSSPECIFICATIONS
SPECIFICATIONS
SPECIFICATIONSSPECIFICATIONS
Capacity 25 16x100 mm vials and
1 receptacle for a stainless steel probe thermometer
Temperature selectable, 105 or 150 ºC
Accuracy ±2 ºC @ 25 ºC ambient
Temperature stability ±0.5 ºC
Warm-up time 30-40 minutes,
depending on set temperature
Timer 0-120 minutes with acoustic alarm
and automatic shut-off,
or infinity mode feature
Ambient operating temperature +5 to +50 ºC
Storage temperature -20 to +60 ºC
Power supply:
C9800-01 115 VAC; 60Hz; 250W; Fuse 4A
C9800-02 230 VAC; 50Hz; 250W; Fuse 2A
Dimensions 190 x 300 x 95 mm (7.5 x 11.8 x 3.9’’)
Weight approx. 4.8 kg (10.6 lb.)
Hanna Instruments reserves the right to modify the design, construction
and appearance of its products without advance notice.
FUNCTIONAL DESCRIPTIONFUNCTIONAL DESCRIPTION
FUNCTIONAL DESCRIPTION
FUNCTIONAL DESCRIPTIONFUNCTIONAL DESCRIPTION
FRONT PANEL
1) Heater block: 1) Heater block:
1) Heater block: can warm up to 25 samples.
1) Heater block: 1) Heater block:
2) Power LED indicator2) Power LED indicator
2) Power LED indicator: Lights when the power button on
2) Power LED indicator2) Power LED indicator
the back of the instrument is switched ON.
3) Heating LED indicator3) Heating LED indicator
3) Heating LED indicator: Lights only when the heater is
3) Heating LED indicator3) Heating LED indicator
working (warming up). When the indicator cycles ON and
OFF, the heater is maintaining stable the set temperature.
4) Temperature mode switch4) Temperature mode switch
4) Temperature mode switch: Selects 105 or 150 ºC (221
4) Temperature mode switch4) Temperature mode switch
or 302 ºF) operation mode.
5) Timer5) Timer
5) Timer: The knob allows to control the heating procedure.
5) Timer5) Timer
It is possible to choose:
Timed procedure modeTimed procedure mode
Timed procedure mode: when the timer is set
§
Timed procedure modeTimed procedure mode
between 0 and 120 minutes, the heater will maintain constant the selected temperature for the
remaining time. The reactor will shut off at the end
of the elapsed time and will inform the user with an
acoustic alarm.
∞
infinity mode infinity mode
infinity mode: the heating indicator will light
§
infinity mode infinity mode
when the heater is operating. The heater block will
remain at the selected temperature as long as the
instrument power is ON.
REAR PANEL
6) Power switch6) Power switch
6) Power switch: Turns the instrument ON and OFF.
6) Power switch6) Power switch
When the power button is set in the ON position the
power indicator will light.
7) Fuse holder7) Fuse holder
7) Fuse holder: The fuse is placed near the case for easy
7) Fuse holder7) Fuse holder
maintenance operations.
8) Power supply8) Power supply
8) Power supply: 115 or 230 VAC, depending on model.
8) Power supply8) Power supply
OPERATINGOPERATING
OPERATING
OPERATINGOPERATING
Set the power switch (back of the instrument) to ON
position. The power indicator will light. Set the temperature switch to 105 or 150 ºC (according to the procedure
described in the manual of the specific parameter).
Position the timer to the ∞ position by turning the knob
anticlock-wise to allow the heater to operate; the heating
indicator will light. Allow the heater block to warm up for
approximately 30-40 minutes, depending on the set temperature. During this period the heating indicator will
remain ON.
When the heating indicator cycles ON and OFF, it means
that the heater has reached the set temperature and is
maintaining it constant.
At this moment it is possible to start a timed procedure
and select the time interval requested by the specific
parameter digestion procedure by turning the knob clockwise. During this procedure the heating indicator will cycle
ON and OFF.
Attention: to assure a correct digestion it is important to
start the timed procedure as soon as the heater is stabilized at the selected temperature (that is when the
heating indicator starts to cycle ON and OFF).
Notes:
• Use of the optional safety shield is strongly recommended to protect the operator from splattering reagents
in case of a vial should break.
• Do not place the Reactor near heating or cooling source.
Temperature stability could be affected.
• The set temperature is that typically expected into the
tubes. The temperature of the metallic block is a few
degrees higher than the set value, in order to allow the
heat transmission from the block to the liquid inside the
tubes.
• During operation, the block temperature will be slightly
different for block empty or full of vials. If cold tubes are
placed in the block, the temperature will decrease of
several degrees and a few minutes will be required to
recover as the tubes warm.
• An external stainless steel probe thermometer can be
used for a check of the block temperature, simply by
placing it into the appropriate receptacle.
MAINTENANCEMAINTENANCE
MAINTENANCE
MAINTENANCEMAINTENANCE
The instrument should be kept clean and any reagent spill
should be wiped up immediately.
Do not touch the heater block unless you are
certain it has cooled.
If some reagent accidentally spills into the heater
block, disconnect power and allow the instrument
to cool; then remove all tubes and clean the
block.
Should overheating occurs, the reactor automatically shuts down. Both LED indicators turn off.
Contact your dealer or your nearest Hanna Service
Center for maintenance.
Fuse replacement
Replace fuse only with fuse of specified type and
current rating.
• Turn the instrument off before removing the fuse holder.
• Unscrew the fuse holder by rotating it counterclockwise.
• Replace the fuse with one 4A fuse for 115-volt model
or one 2A fuse for 230-volt model.
• Replace the holder before turning the instrument on.
CE DECLARATIONCE DECLARATION
CE DECLARATION
CE DECLARATIONCE DECLARATION
OF CONFORMITYOF CONFORMITY
OF CONFORMITY
OF CONFORMITYOF CONFORMITY
ACCESSORIESACCESSORIES
ACCESSORIES
ACCESSORIESACCESSORIES
HI 151-00 Hanna thermometer (-50.0 to 220 ºC)
HI 151-01 Hanna thermometer (-58.0 to 428 ºF)
HI 740138 2A, 5x20mm fuse
HI 740139 2A, 5x20mm fuse (6 pcs)
HI 740214 4A, 5x20mm fuse
HI 740215 4A, 5x20mm fuse (6 pcs)
HI 740216 Test tube cooling rack (25 holes)
HI 740217 Laboratory bench safety shield
ISTRREACR1 10/01