Hanna Instruments C 9800 User Manual

Instruction Manual
C9800
REACTOR
www.hannainst.com
WARRANTYWARRANTY
WARRANTY
WARRANTYWARRANTY
C9800 is warranted for two years against defects in workmanship and materials when used for its intended purpose and maintained according to instructions.
This warranty is limited to repair or replacement free of charge. Damages due to accident, misuse, tampering or lack of prescribed
maintenance are not covered. If service is required, contact the dealer from whom you purchased the
instrument. If under warranty, report the model number, date of pur­chase, serial number and the nature of the failure. If the repair is not covered by the warranty, you will be notified of the charges incurred. If the instrument is to be returned to Hanna Instruments, first obtain a Returned Goods Authorization Number from the Customer Service de­partment and then send it with shipment costs prepaid. When shipping any instrument, make sure it is properly packaged for complete protec­tion.
To validate your warranty, fill out and return the enclosed warranty card within 14 days from the date of purchase.
All rights are reserved. Reproduction in whole or in part is prohibited without the written consent of the copyright owner, Hanna Instru­ments Inc., Woonsocket, Rhode Island, 02895 , USA.
This Instrument is in
Compliance with the CE Directives
Dear Customer, Thank you for choosing a Hanna product. Please read this
directives.
with
PRELIMINARY EXAMINATIONPRELIMINARY EXAMINATION
PRELIMINARY EXAMINATION
PRELIMINARY EXAMINATIONPRELIMINARY EXAMINATION
Remove the instrument from the packing material and examine it carefully to make sure that no damage has occurred during shipment. If there is any damage, notify your Dealer.
Note: Save all packing material until you are sure that
the instrument functions correctly. Any defective item must be returned in their original packing with the supplied accessories.
GENERAL DESCRIPTIONGENERAL DESCRIPTION
GENERAL DESCRIPTION
GENERAL DESCRIPTIONGENERAL DESCRIPTION
C9800 Reactor is a Hanna Instruments thermo-reactor with a metal dry-block and a metal case construction.
It allows to warm up to 25 samples, contained into 16 mm-diameter tubes, at 105 or 150 ºC.
The reactor is equipped with led indicators, to inform when power is on and heating is operating, and a 120 minutes timer, very useful in timed digestion procedures.
It is possible to select time intervals between 0 and 120 minutes and, during these intervals, temperature in the heater is maintained constant. At the end of the selected interval an acoustic alarm will sound and the heater will automatically shut off.
The reactor is also provided with a thermal fuse that prevents from overheating, by shutting down the heater.
Two models are available:
C9800-01 runs with 115 VAC mains line C9800-02 runs with 230 VAC mains line.
HEALTH & SAFETYHEALTH & SAFETY
HEALTH & SAFETY
HEALTH & SAFETYHEALTH & SAFETY
Before unpacking, setting up or start using this reactor it is important to read carefully this manual. Pay particular attention to all warnings, cautions and notes. Failure to do so could result in serious injury to the operator or damage to the equipment.
To ensure a safety use of this instrument, it must not be installed or used in any manner other than that specified in this manual.
Reactor Operation: The Reactor incubates glass vials containing hazardous chemicals at high tem­perature. Read the Health & Safety Data Sheet (HSDS) before handling reagent vials.
Handling Reagent Vials: Handle the vials very carefully and wear protective clothing. After use, dispose of the reagent vials according to local regulations.
Safety Equipment: When operating with Reactor use protective clothing, including goggles or face mask, and gloves.
Use of the optional safety shield is highly recommended.
Reagent Spills: Wipe up spilled reagents immediately. If reagent contacts skin, rinse the affected area thoroughly with water. Avoid breathing released vapors. Read the Health & Safety Data Sheet (HSDS) of each reagent for complete chemical information.
Fire Hazard: Avoid the presence of flammable liquids or paper near the operating Reactor. A fire hazard could be created.
SPECIFICATIONSSPECIFICATIONS
SPECIFICATIONS
SPECIFICATIONSSPECIFICATIONS
Capacity 25 16x100 mm vials and
1 receptacle for a stainless steel probe thermometer
Temperature selectable, 105 or 150 ºC Accuracy ±2 ºC @ 25 ºC ambient Temperature stability ±0.5 ºC Warm-up time 30-40 minutes,
depending on set temperature
Timer 0-120 minutes with acoustic alarm
and automatic shut-off,
or infinity mode feature
Ambient operating temperature +5 to +50 ºC Storage temperature -20 to +60 ºC Power supply: C9800-01 115 VAC; 60Hz; 250W; Fuse 4A C9800-02 230 VAC; 50Hz; 250W; Fuse 2A Dimensions 190 x 300 x 95 mm (7.5 x 11.8 x 3.9’’) Weight approx. 4.8 kg (10.6 lb.)
Hanna Instruments reserves the right to modify the design, construction and appearance of its products without advance notice.
FUNCTIONAL DESCRIPTIONFUNCTIONAL DESCRIPTION
FUNCTIONAL DESCRIPTION
FUNCTIONAL DESCRIPTIONFUNCTIONAL DESCRIPTION
FRONT PANEL
1) Heater block: 1) Heater block:
1) Heater block: can warm up to 25 samples.
1) Heater block: 1) Heater block:
2) Power LED indicator2) Power LED indicator
2) Power LED indicator: Lights when the power button on
2) Power LED indicator2) Power LED indicator the back of the instrument is switched ON.
3) Heating LED indicator3) Heating LED indicator
3) Heating LED indicator: Lights only when the heater is
3) Heating LED indicator3) Heating LED indicator working (warming up). When the indicator cycles ON and OFF, the heater is maintaining stable the set temperature.
4) Temperature mode switch4) Temperature mode switch
4) Temperature mode switch: Selects 105 or 150 ºC (221
4) Temperature mode switch4) Temperature mode switch or 302 ºF) operation mode.
5) Timer5) Timer
5) Timer: The knob allows to control the heating procedure.
5) Timer5) Timer It is possible to choose:
Timed procedure modeTimed procedure mode
Timed procedure mode: when the timer is set
§
Timed procedure modeTimed procedure mode between 0 and 120 minutes, the heater will main­tain constant the selected temperature for the remaining time. The reactor will shut off at the end of the elapsed time and will inform the user with an acoustic alarm.
infinity mode infinity mode
infinity mode: the heating indicator will light
§
infinity mode infinity mode when the heater is operating. The heater block will remain at the selected temperature as long as the instrument power is ON.
REAR PANEL
6) Power switch6) Power switch
6) Power switch: Turns the instrument ON and OFF.
6) Power switch6) Power switch When the power button is set in the ON position the power indicator will light.
7) Fuse holder7) Fuse holder
7) Fuse holder: The fuse is placed near the case for easy
7) Fuse holder7) Fuse holder maintenance operations.
8) Power supply8) Power supply
8) Power supply: 115 or 230 VAC, depending on model.
8) Power supply8) Power supply
OPERATINGOPERATING
OPERATING
OPERATINGOPERATING
Set the power switch (back of the instrument) to ON position. The power indicator will light. Set the tempera­ture switch to 105 or 150 ºC (according to the procedure described in the manual of the specific parameter).
Position the timer to the position by turning the knob
anticlock-wise to allow the heater to operate; the heating indicator will light. Allow the heater block to warm up for approximately 30-40 minutes, depending on the set tem­perature. During this period the heating indicator will remain ON.
When the heating indicator cycles ON and OFF, it means that the heater has reached the set temperature and is maintaining it constant.
At this moment it is possible to start a timed procedure and select the time interval requested by the specific parameter digestion procedure by turning the knob clock­wise. During this procedure the heating indicator will cycle ON and OFF.
Attention: to assure a correct digestion it is important to start the timed procedure as soon as the heater is stabi­lized at the selected temperature (that is when the heating indicator starts to cycle ON and OFF).
Notes:
• Use of the optional safety shield is strongly recom­mended to protect the operator from splattering reagents in case of a vial should break.
• Do not place the Reactor near heating or cooling source. Temperature stability could be affected.
• The set temperature is that typically expected into the tubes. The temperature of the metallic block is a few degrees higher than the set value, in order to allow the heat transmission from the block to the liquid inside the tubes.
• During operation, the block temperature will be slightly different for block empty or full of vials. If cold tubes are placed in the block, the temperature will decrease of several degrees and a few minutes will be required to recover as the tubes warm.
• An external stainless steel probe thermometer can be used for a check of the block temperature, simply by placing it into the appropriate receptacle.
MAINTENANCEMAINTENANCE
MAINTENANCE
MAINTENANCEMAINTENANCE
The instrument should be kept clean and any reagent spill should be wiped up immediately.
Do not touch the heater block unless you are certain it has cooled.
If some reagent accidentally spills into the heater block, disconnect power and allow the instrument to cool; then remove all tubes and clean the block.
Should overheating occurs, the reactor automati­cally shuts down. Both LED indicators turn off. Contact your dealer or your nearest Hanna Service Center for maintenance.
Fuse replacement
Replace fuse only with fuse of specified type and current rating.
• Turn the instrument off before removing the fuse holder.
• Unscrew the fuse holder by rotating it counterclockwise.
• Replace the fuse with one 4A fuse for 115-volt model or one 2A fuse for 230-volt model.
• Replace the holder before turning the instrument on.
CE DECLARATIONCE DECLARATION
CE DECLARATION
CE DECLARATIONCE DECLARATION
OF CONFORMITYOF CONFORMITY
OF CONFORMITY
OF CONFORMITYOF CONFORMITY
ACCESSORIESACCESSORIES
ACCESSORIES
ACCESSORIESACCESSORIES
HI 151-00 Hanna thermometer (-50.0 to 220 ºC) HI 151-01 Hanna thermometer (-58.0 to 428 ºF) HI 740138 2A, 5x20mm fuse HI 740139 2A, 5x20mm fuse (6 pcs) HI 740214 4A, 5x20mm fuse HI 740215 4A, 5x20mm fuse (6 pcs) HI 740216 Test tube cooling rack (25 holes) HI 740217 Laboratory bench safety shield