Instruction Manual
C 219
Maple Syrup
Transmittance Analyzer
www.hannainst.com
WARRANTY
C 219 is warranted for two years against defects in workmanship and materials when used for its intended purpose and
maintained according to instructions.
This warranty is limited to repair or replacement free of charge.
Damages due to accident, misuse, tampering or lack of
prescribed maintenance are not covered.
If service is required, contact the dealer from whom you
purchased the instrument. If under warranty, report the model
number, date of purchase, serial number and the nature of the
failure. If the repair is not covered by the warranty, you will be
notified of the charges incurred. If the instrument is to be
returned to Hanna Instruments, first obtain a Returned Goods
Authorization Number from the Customer Service department
and then send it with shipment costs prepaid. When shipping
any instrument, make sure it is properly packaged for complete
protection.
All rights are reserved. Reproduction in whole or in part is
prohibited without the written consent of the copyright owner,
Hanna Instruments Inc.
Dear Customer,
Thank you for choosing a Hanna product. This manual will
provide you with the necessary information for the correct
operation of the meter. Please read it carefully before using
the meter. If you need additional technical information, do not
hesitate to e-mail us at tech@hannainst.com or call
1-800-426-6287. This instrument is in compliance with
directives EN 50081-1 & EN 50082-1.
Remove the instrument from the packing material and
examine it carefully to make sure that no damage has
occurred during shipment. If there is any damage, notify
your Dealer.
Each C 219 instrument is supplied complete with
• Two 9V Batteries
• One Light Shield Cap
• One Transport Cap
• 12 VDC Transformer (HI 710005 or HI 710006)
Note
: save all packing material until you are sure that the
instrument works correctly. Any defective item must be
returned in its original packing.
PRELIMINARY EXAMINATION
GENERAL DESCRIPTION
The C 219 instrument measures the percent light transmittance
of maple syrup as compared to analytical reagent glycerol. The
transmittance value allows identification of syrup quality class.
Measurements are done by using matched square optical
cuvets having a 10 mm light path.
Display codes aid the user in routine operations.
The meters have an auto-shut off feature that will turn the
instrument off after 10 minutes of non-use.
SPECIFICATIONS
Range 0.0 to 100.0 % Transmittance
Resolution 0.1 in % Transmittance
Accuracy ±1% of reading @ % Transmittance=75.0
Typical EMC ±0.1 in % Transmittance
Deviation
Light Source Tungsten lamp with narrow band
interference filter @ 560 nm
Method Direct Measurement
Light Detector Silicon Photocell
Environment 0 to 50°C (32 to 122°F);
max 95% RH non-condensing
Battery Type 2 x 9 volt
Auto-Shut off After 10' of non-use
Dimensions 230 x 165 x 70 mm (9.0 x 6.5 x 2.8")
Weight 640 g (22.6 oz.)
REQUIRED ST
ANDARD
4 mL Glycerol (see accessories)
This Instrument is in Compliance
with the CE Directives
FRONT P
ANEL
1) Cuvet Holder
2) Liquid Crystal Display
3) Color Class/% Transmittance Table
4) READ, to perform measurement
5) CAL, to calibrate the meter prior to measurement
6) ON/OFF, to turn the meter on and off
REAR P
ANEL
1) Power Supply 12 VDC 2.5 Watt
2) RS 232 Socket
3) Batteries Compartment
FUNCTIONAL DESCRIPTION
SIGNIFICANCE AND USE
The primary characteristic for classifying maple syrup is color.
When syrup is very light in color, as matching the minimum light
transmittance standards, the grade assigned to syrup is high.
When syrup color is dark, the grade is low. Color classes are
expressed in percent of light transmission as compared to an analytical glycerol standard fixed at one hundred percent transmission. All syrup is considered to be at minimum density, 66.0
degrees Brix at 68 degrees F (20
0
C), and to have no flavors
nor other characteristics extraneous to pure maple syrup.
COLOR CLASS RANGE (% Transmittance)
Extra light 75.0 to 100.0
Light 60.5 to 74.9
Medium 44.0 to 60.4
Amber 27.0 to 43.9
Dark less than 27.0
GUIDE TO DISPLAY CODES
23
POWER SUPPLY
12 VDC 2.5 Watt
RS232
ooooo
oooo
1
+
-
Hanna Instruments reserves the right to modify the design,
construction and appearance of its products without advance notice.
This indicates that the meter is in a ready state and
calibration can be performed.
Sampling in Progress. This flashing prompt appears
each time the meter is performing a measurement.
The microprocessor is adjusting the light level, indicated by a scrolling “SIP”.
This indicates that the meter has been calibrated
and measurement can be performed.
Calibration was not performed. Follow the instructions
described in the measurement procedures for
calibrating the meter.
The instrument is performing an internal check-up.
The instrument cannot adjust the light level. Please
check that the sample does not contain any debris.
There is not enough light to perform a measurement.
Please check the preparation of the Glycerol cuvet.
There is too much light to perform a measurement.
Please, check the preparation of the Glycerolcuvet.
There is too much light for the sample measurement.
Please check if the right sample cuvet is inserted.
The sample and Glycerol cuvet are inverted.
The blinking "LOBAT" indicates that the battery voltage
is getting low and the battery needs to be replaced.
This indicates that the battery is dead and must be
replaced.
Light over range. The cuvet is not inserted correctly
and an eccess ambient light is reaching the detector.
If the cap is properly installed, contact your
dealer or the nearest Hanna Customer Service Center.
This indicates that the calibration procedure failed
due to a low signal-to-noise ratio. In this case press
CAL again. Or, it may indicate over range: a flashing 100.0% value as result of a reading
indicates that the sample absorbs less light than
the glycerol standard.
Excel©Copyright of "Microsoft Co."
Lotus 1-2-3©Copyright of "Lotus Co."
TELIX®is registered Trademark of "Deltacomm"
Windows®and Windows Terminal®are registered Trademark of "Microsoft Co."
CE DECLARATION
OF CONFORMITY
• Press CAL and
“SIP” will appear
on the display.
• Wait for a few seconds and the
display will show “100.0%”. Now
the meter is calibrated and ready
for measurement.
• Remove the Glycerol cuvet.
• Add to a second cuvet 4 ml of maple syrup,
up to 5 mm (0.2") below the rim, using a
second syringe. This is the
sample.
• Place the sample cuvet into the
holder, then put the black cap on.
• Press READ and “SIP”
will appear during
measurement.
• The instrument directly displays percent of light transmittance as compared to glycerol standard (fixed at one hundred percent).
• Check in the manual or the front panel to which color class
the transmittance value corresponds.
INTERFERENCES
Interference may be caused by air bubbles or turbidity in the
sample. Scratched or dirty cuvets will also affect readings.
Always check clearness of cuvets prior to use.
• It is recommended to take and analyze samples at room
temperature.
• It is recommended to take the sample of maple syrup in
the middle of the container.
• Do not use the same cuvet with glycerol for more than
5-10 times to perform calibration and dispose of it after 1
day has passed.
• Never re-use maple syrup sample cuvets, dispose of them
after one use.
• Use 1 syringe for glycerol, and the other one for syrup in
order to avoid contamination.
• Samples should be completely free of air bubbles. If bubbles are present, let the samples stand until air bubbles
have completely disappeared (not more thant 30 minutes,
see note below).
NOTE: Maple syrup samples exposed to direct sunlight and/or
air will darken over time. This is caused by microorganisms.
ISTRC219R1 11/01
OPERATIONAL GUIDE
POWER CONNECTION
Plug the 12VDC adapter (HI 710005 - 110VDC, or HI 710006
- 220VDC) into the DC socket. Plug the adapter into the outlet.
Alternatively, remove the battery cover on the back of the
meter; attach two fresh 9 V batteries and replace the cover.
Note: insure the main line is surge protected.
Note: always turn the meter off before unplugging it to
insure no data is lost.
MEASUREMENT PROCEDURE
• Turn the meter on by pressing
ON/OFF.
• The meter will first perform an LCD
self diagnostic test by displaying a full
set of figures.
• Then it will show a scrolling
"C219 Hanna Inst"
message.
• When the LCD displays “----”,
the meter is ready.
• To perform calibration, use a first syringe
to fill one cuvet with 4 mL of Glycerol, up to
5 mm (0.2") below the rim.
• Place the Glycerol cuvet into the
holder paying attention to the
direction of the light indicated by
the arrow, then put the black cap
on.
BATTERY REPLACEMENT
Battery replacement must only take place in a non-hazardous
area using two 9V alkaline batteries.
Remove the battery cover on the back of the meter, attach two
fresh 9V batteries and replace the cover while paying attention
to the correct polarity.
INTERFACE WITH PC
ACCESSORIES
C219/C220 Kit Kit for maple syrup analysis,
including, 82 cuvets, 30 mL of
Glycerol & two 5 mL syringe
(80 tests average)
HI 721310 9V battery (10 pcs)
HI 731318 Tissue for wiping cuvets (4 pcs)
HI 731325 Caps (4 pcs)
Light under range. The standard is to dark for proper
calibration. If this is not the case, contact your dealer
or the nearest Hanna Customer Service Center.
Light over range. The standard is to light for proper
calibration. If this is not the case, contact your dealer
or the nearest Hanna Customer Service Center.
This indicates that the meter has lost its configuration. Contact your dealer or the nearest Hanna
Customer Service Center.
To connect your meter to the PC use the optional HI 920010
cable (available from your Hanna Dealer). Make sure that your
meter is switched off and plug the connectors, one into the meter
RS 232C socket, the other into the serial port of your PC.
Note
: cables other than HI 920010 may use a different configuration, in which case, communication between the meter and
the PC may not be possible.
Data transmission from the instrument to the PC is now much
easier with HI 92000 Windows®compatible application software offered by Hanna Instruments.
HI 92000 allows you to use the powerful capabilities of most
spread sheet programs (e.g. Excel®, Lotus 1-2-3®). Simply
open your file downloaded by HI 92000 from your spread
sheet program and then it is possible to make any elaboration
available with your software (e.g. graphics, statistic analysis).
User friendly, HI 92000 offers a variety of features and has an
on line help to support you throughout any situation.
To install HI 92000 you need a 3.5” drive and a few minutes to
follow the instructions conveniently printed on the disk label.
For acquiring data with PC from the C219 instrument select
“Real-Time Logging”. Click on “Settings” and then select the
communication port (1 to 4) and set the baud rate. The
transmission speed (baud rate) of the meter and the external
device must be identical. The meter is factory set to 2400.
If you wish to change this value, please contact your nearest
Hanna Center.
Data are recorded on PC by pressing “log sample” after
measurement has been performed. Follow the on line help for
more details.
TIPS FOR AN ACCURATE
MEASUREMENT
The instruction listed below should be carefully followed during
testing to ensure best accuracy.
• Do not touch the cuvet walls with hands.
• Whenever the cuvet is placed into the measurement cell, it
must be completely free of fingerprints, oil or dirt. Wipe it
thoroughly with HI 731318 or a lint-free cloth prior to
insertion.
• It is important that the sample does not contain any debris.
This would corrupt the readings.
• Use matched square cuvets, having 4 clear faces and a
10 mm path length.
Recommendations for Users
Before using these products, make sure that they are entirely suitable for
the environment in which they are used.
Operation of these instruments in residential area could cause unacceptable
interferences to radio and TV equipments, requiring the operator to take all
necessary steps to correct interferences.
Any variation introduced by the user to the supplied equipment may degrade the instruments' EMC performance.
To avoid damages or burns, do not perform any measurement in microwave ovens.
LOBAT
%
DECLARATION OF CONFORMITY
We
Hanna Instruments Italia Srl
via E.Fermi, 10
35030 Sarmeola di Rubano - PD
ITALY
herewith certify that the bench meter:
has been tested and found to be in compliance with EMC Directive 89/336/EEC and Low
Voltage Directive 73/23/EEC according to the following applicable normatives:
EN 50082-1:Electromagnetic Compatibility - Generic Immunity Standard
IEC 801-2 Electrostatic Discharge
IEC 801-3 RF Radiated
IEC 801-4 Fast Transient
EN 50081-1:Electromagnetic Compatibility - Generic Emission Standard
EN 55022 Radiated, Class B
EN61010-1:Safety requirements for electrical equipment for measurement,
Date of Issue: 10-11-2001
C219
control and laboratory use
D.Volpato - Engineering Manager
On behalf of
Hanna Instruments S.r.l.