Hanna Instruments C 214 User Manual

TECHNICAL SERVICE CONTACTS
Australia:
Tel. (03) 9769.0666 • Fax (03) 9769.0699
e-mail: hannains@hannainst.com.au
China:
Tel. (10) 88570068 • Fax (10) 88570060
e-mail: hannachina@vip.sina.com
Egypt:
Tel. & Fax (02) 2758.683
e-mail: hannaegypt@go.com.eg
Germany:
Tel. (07851) 9129-0 • Fax (07851) 9129-99
e-mail: hannager@aol.com
Greece:
Tel. (210) 823.5192 • Fax (210) 884.0210
e-mail: hannagr@otenet.gr
Indonesia:
Tel. (21) 4584.2941 • Fax (21) 4584.2942
e-mail: transit@dnet.net.id
Japan:
Tel. (03) 3258.9565 • Fax (03) 3258.9567
e-mail: sales@hanna.co.jp
Korea:
Tel. (02) 2278.5147 • Fax (02) 2264.1729
e-mail: mccoyhan@chollian.net
Malaysia:
Tel. (603) 5638.9940 • Fax (603) 5638.9829
e-mail: hannamal@tm.net.my
Norway:
Tel. (23) 3811.00 • Fax (23) 3811.01
e-mail: hanna@hannainst.no
Singapore:
Tel. 6296.7118 • Fax 6291.6906
e-mail: hannaap@pacific.net.sg
South Africa:
Tel. (011) 615.6076 • Fax (011) 615.8582
e-mail: hannasa@mweb.co.za
United Kingdom:
Tel. (01525) 850.855 • Fax (01525) 853.668
e-mail: salesteam@hannainst.co.uk
USA:
Tel. (401) 765.7500 • Fax (401) 765.7575
e-mail: sales@hannainst.com
For additional Technical Support in your local lan­guage, see www.hannainst.com
MANC214R3 10/04
Instruction Manual
C 214
Multiparameter Bench
Photometer
for
Wastewater Treatment
Application
www.hannainst.com
Compliance with the CE Directives
Dear Customer, Thank you for choosing a Hanna product. Please read this instruction
manual carefully before using the meter. This manual will provide you with the necessary information for the correct use of the instrument. If you need additional technical information, do not hesitate to e-mail us at tech@hannainst.com. These instruments are in compliance with
directives.
TABLE OF CONTENTS
PRELIMINARY EXAMINATION ........... 3
ABBREVIATIONS ............................. 3
GENERAL DESCRIPTION ................... 4
SPECIFICATIONS ............................. 5
PRECISION AND ACCURACY ............. 5
PRINCIPLE OF OPERATION ............... 6
FUNCTIONAL DESCRIPTION ..............8
GUIDE TO DISPLAY CODES .............. 9
TIPS FOR AN ACCURATE
MEASUREMENT ............................12
PARAMETERS REFERENCE TABLES ..17
HEALTH & SAFETY ........................19
OPERATIONAL GUIDE .................... 20
AMMONIA LOW RANGE ................. 22
AMMONIA HIGH RANGE ................ 25
CHLORINE, FREE ...........................28
CHLORINE, TOTAL ......................... 31
NITRATE ....................................... 34
NITROGEN, TOTAL LR .................... 37
NITROGEN, TOTAL HR ................... 42
OXYGEN DEMAND, CHEMICAL LR ....47
OXYGEN DEMAND, CHEMICAL MR .. 50 OXYGEN DEMAND, CHEMICAL HR ...53
PHOSPHORUS, REACTIVE .............. 56
PHOSPHORUS, ACID HYDROLYZABLE ...
PHOSPHORUS, TOTAL ................... 63
PHOSPHORUS, REACTIVE HR .........67
PHOSPHORUS, TOTAL HR ..............70
INTERFACE WITH PC ..................... 74
STANDARD METHODS ................... 76
BATTERY REPLACEMENT ................ 76
ACCESSORIES ............................... 77
CE DECLARATION OF CONFORMITY .... 78
WARRANTY ...................................79
HANNA LITERATURE ......................79
WARRANTY
All Hanna Instruments meters are warranted for two years against defects in workmanship and materials when used for its intended purpose and maintained according to the instructions.
This warranty is limited to repair or replacement free of charge. Damages due to accident, misuse, tampering or lack of prescribed
maintenance are not covered. If service is required, contact your dealer. If under warranty, report the
model number, date of purchase, serial number and the nature of the failure. If the repair is not covered by the warranty, you will be notified of the charges incurred.
If the instrument is to be returned to Hanna Instruments, first obtain a Returned Goods Authorization Number from the Customer Service Department and then send it with shipment costs prepaid. When shipping any instrument, make sure it is properly packaged for complete
59
protection. To validate your warranty, fill out and return the enclosed warranty
card within 14 days from the date of purchase.
Hanna Instruments reserves the right to modify the design, construction and appearance of its products without advance notice.
HANNA LITERATURE
Hanna publishes a wide range of catalogs and handbooks for an equally wide range of applications. The reference literature currently covers areas such as:
All rights are reserved. Reproduction in whole or in part is prohibited without the written consent of the copyright owner, Hanna Instruments Inc., Woonsocket, Rhode Island, 02895 , USA.
2
Water Treatment
Process
Swimming Pools
Agriculture
Food
Laboratory
and many others. New reference material is constantly being added to the library.
For these and other catalogs, handbooks and leaflets contact your dealer or the Hanna Customer Service Center nearest to you. To find the Hanna Office in your vicinity, check our home page at www.hannainst.com.
79
CE DECLARATION OF CONFORMITY
Recommendations for Users
Before using these products, make sure that they are entirely suitable for your specific application and for the environment in which they are used.
Operation of these instruments may cause unacceptable interferences to other electronic equip­ments, this requiring the operator to take all necessary steps to correct interferences.
Any variation introduced by the user to the supplied equipment may degrade the instruments' EMC performance.
To avoid damages or burns, do not put the instrument in microwave ovens. For yours and the instrument safety do not use or store the instrument in hazardous environments.
PRELIMINARY EXAMINATION
Please examine this product carefully. Make sure that the instrument is not damaged. If any damage occured during shipment, please notify your Dealer.
Each Meter is supplied complete with:
• Five Sample Vials
• Two 9 V Batteries
• 12 VDC Transformer (HI 710005 or HI 710006)
• Instruction Manual
Note: Save all packing material until you are sure that the instrument
works correctly. Any defective item must be returned in its original packing with the supplied accessories.
ABBREVIATIONS
°C: degree Celsius COD: Chemical Oxygen Demand EPA: US Environmental Protection Agency °F: degree Fahrenheit g/L: grams per liter. g/L is equivalent to ppt (part per thoushand) HR: high range LR: low range mg/L: milligrams per liter. mg/L is equivalent to ppm (part per million) mL: milliliter MR: medium range µg/L: micrograms per liter. µg/L is equivalent to ppb (part per billion)
78
3
GENERAL DESCRIPTION
ACCESSORIES
C 99 & C 200 Series is a line of 17 different bench, microprocessor­based photometers to measure
more than 50 different parameters in water and wastewater. These multipurpose meters are manufactured to measure the most important parameters of the application they have been especially designed for: C 99 Laboratories, with COD C 200 Laboratories
C 203 Aquaculture C 205 Boilers & Cooling Towers C 206 Environmental Testing C 207 Industrial Wastewater C 208 Water Conditioning C 209 Education C 210 Pulp & Paper Mills C 211 Chemical Manufacturers C 212 Power Plant Utilities C 213 Municipal Wastewater C 214 Wastewater Treatment Applic. C 215 Nutrient Analyses C 216 Pool & Spa Applic. C 218 Environmental Applic. C 226 Pool & Spa Applic.
The reagents are in liquid or powder form and are supplied in bottles and ready-to-use vials or in packets. The amount of reagent is precisely dosed to ensure the maximum repeatability.
Display codes aid the user in routine operations. The meters have an auto-shut off feature that will turn off the instrument
after 10 minutes of non-use. The C 99 & C 200 Series can be connected to a personal computer via
the HI 920010 three wire RS 232 cable. The HI 92000 Hanna
®
Windows
Compatible Software aids the user to manage all test data.
REAGENT SETS
HI 93701-01 100 free chlorine tests HI 93701-03 300 free chlorine tests HI 93711-01 100 total chlorine tests HI 93711-03 300 total chlorine tests HI 93754A-25 25 COD LR tests HI 93754B-25 25 COD MR tests HI 93754C-25 25 COD HR tests HI 93758A-50 50 reactive phosphorus tests HI 93758B-50 50 acid hydrolyzable phosphorus tests HI 93758C-50 50 total phosphorus tests HI 93763A-50 50 reactive phosphorus HR tests HI 93763B-50 50 total phosphorus HR tests HI 93764A-25 25 ammonia LR tests HI 93764B-25 25 ammonia HR tests HI 93766-50 50 nitrate tests HI 93767A-50 50 total nitrogen LR tests HI 93767B-50 50 total nitrogen HR tests
OTHER ACCESSORIES C115-00300 5 mL graduated syringe C9800-01 Hanna Reactor (115 VAC) C9800-02 Hanna Reactor (230 VAC) HI 710005 110VAC to 12VDC voltage adapter HI 710006 220VAC to 12VDC voltage adapter HI 721310 9V battery (10 pcs)
HI 731311 glass vials with cap (5 pcs) HI 731318 Tissue for wiping cuvets (4 pcs) HI 731340 200 µL automatic pipette
HI 731341 1000 µL automatic pipette HI 731342 2000 µL automatic pipette HI 731350 tips for 200 µL automatic pipette (25 pcs) HI 731351 tips for 1000 µL automatic pipette (25 pcs) HI 731352 tips for 2000 µL automatic pipette (4 pcs)
HI 740142 1 mL graduated syringe HI 740143 1 mL graduated syringe (6 pcs) HI 740144 Pipette tip (6 pcs)
HI 740157 Plastic refilling pipette (20 pcs) HI 740216 Test tube cooling rack (25 holes) HI 740217 Laboratory bench safety shield HI 92000 Windows
®
Compatible Software
HI 920010 PC connection cable HI 93703-50 Cuvets cleaning solution (230 mL)
4
77
STANDARD METHODS
SPECIFICATIONS
Description Range Method
Ammonia LR 0.00 to 3.00 mg/L Nessler Ammonia HR 0 to 100 mg/L Nessler Chlorine, Free 0.00 to 5.00 mg/L DPD Chlorine, Total 0.00 to 5.00 mg/L DPD Nitrate 0.0 to 30.0 mg/L Chromotropic Acid Nitrogen, Total 0.0 to 25.0 mg/L Chromotropic Acid Nitrogen, Total HR 10 to 150 mg/L Chromotropic Acid COD LR 0 to 150 mg/L Dichromate, Mercuric Sulphate COD MR 0 to 1500 mg/L Dichromate, Mercuric Sulphate COD HR 0 to 15000 mg/L Dichromate, Mercuric Sulphate Phosphorus, Reactive 0.00 to 5.00 mg/L Ascorbic Acid Phosphorus, Acid Hydrolyzable 0.00 to 5.00 mg/L Ascorbic Acid Phosphorus, Total 0.00 to 3.50 mg/L Ascorbic Acid Phosphorus, Reactive HR 0.0 to 100.0 mg/L Vanadomolybdophosphoric Acid
Phosphorus, Total HR 0.0 to 100.0 mg/L Vanadomolybdophosphoric Acid
BATTERY REPLACEMENT
Besides the 12 VDC supply, this meter can also be fed by two 9V batteries.
To prolong battery life, switch your meter off after use. However, the meter has an auto-shut off feature that will turn itself off after 10 minutes of non-use.
A blinking "LOBAT" on the display, while a measurement is in progress, indicates low voltage and the batteries should be replaced.
If the batteries are not replaced immediately, in order to prevent erroneous readings due to low voltage, "-BA-" is displayed soon afterwards. At this point the batteries must be changed.
Battery replacement must only take place in a non-hazardous area using two 9V alkaline batteries.
Remove the battery cover on the back of photometer, attach two fresh 9V batteries, while paying attention to the correct polarity, and replace the cover.
The meter will turn on automatically when a new battery is connected. You can turn it off by pressing ON/OFF.
Light Life Life of the instrument Light Detector Silicon Photocell Environment 0 to 50°C (32 to 122°F);
max 90% RH non-condensing
Power Supply 2 x 9 V batteries / 12 to 20 VDC through
voltage adapter
Auto-Shut off After 10' of non-use Dimensions 230 x 165 x 70 mm (9.0 x 6.5 x 2.8") Weight 640 g (22.6 oz.)
For specifications related to each single parameter (e.g. range, precision, etc.), refer to the related measurement section.
PRECISION AND ACCURACY
Precision is how closely repeated measurements agree with each other. Precision is usually expressed as standard deviation (SD).
Accuracy is defined as the nearness of a test result to the true value. Although good precision suggests good accuracy, precise results can be
inaccurate. The figure explains these definitions. For each parameter, the precision is expressed in the related measurement
section as standard deviation at a specific concentration value of the analite. The standard deviation is obtained with a single instrument using
a representative lot of reagents.
76
5
PRINCIPLE OF OPERATION
Absorption of Light is a typical phenomenon of interaction between electromagnetic radiation and matter. When a light beam crosses a substance, some of the radiation may be absorbed by atoms, molecules or crystal lattices. If pure absorption occurs, the fraction of light absorbed depends both on the optical path length through the matter and on the physical­chemical characteristics of substance according to the Lambert-Beer Law:
-log
I
A ελc d
Where:
-log
=Absorbance (A)
I/I
o
=intensity of incident light beam
I
o
=intensity of light beam after absorption
I
ελ=molar extinction coefficient at wavelength λ
c =molar concentration of the substance d =optical path through the substance
Therefore, the concentration "c" can be calculated from the absorbance of the substance as the other factors are known.
Photometric chemical analysis is based on the possibility to develop an absorbing compound from a specific chemical reaction between sample and reagents.
Given that the absorption of a compound strictly depends on the wavelength of the incident light beam, a narrow spectral bandwidth should be selected as well as a proper central wavelength to optimize measurements.
The optical system of Hanna's C 99 & C 200 Series multiparameter photometers is based on special subminiature tungsten lamps and narrow-band interference filters to guarantee both high performance and reliable results.
Four measuring channels (at four different wavelengths) allow a wide range of tests.
ελc d
/I
o
or
/PUP - Program Up /PDN - Program Down /PTM - Turn Test Mode On /Brx - Set the baud rate
1 - 300 2 - 600
3 - 1200 4 - 2400 /KBL - Lock Keyboard /KBU - Unlock Keyboard ?PR# - Send Current Program Number ?BRQ - Send current baud rate
1 - 150 2 - 300
3 - 600 4 - 1200
5 - 2400 ?CNQ - Send Concentration (three bytes)
conc
decimal point
unit
m - ppm b - ppb t - ppt u - pcu h - pH
?ERR - send error / status information
0 - No error
1 - CAP
2 - HI
3 - ZERO
4 - LO
5 - IDLE
6 - ZERO DONE
7 - TIMED READ
C 200 Series block diagram (optical layout)
6
75
INTERFACE WITH PC
To connect your meter to the PC use the optional HI 920010 (available from your Hanna Dealer). Make sure that your meter is switched off and plug the connectors, one into the meter RS 232C socket, the other into the serial port of your PC.
Note: Cables other than HI 920010 may use a different configuration,
in which case, communication between the meter and the PC may not be possible.
SETTING THE BAUD RATE
The transmission speed (baud rate) of the meter and the external device must be identical. The meter is factory set to 2400.
If you wish to change this value, please contact your nearest Hanna Center.
SENDING COMMANDS FROM PC
With terminal programs such as, for example, Telix®, Windows Termi­nal®, it is possible to remotely control your Hanna Instruments bench meter. Use HI 920010 cable to connect the meter to the PC, start the terminal program and set the communication options as follows: 8, N, 1, no flow control.
Command Types
To send a command to the meter the scheme is:
<DLE> <command> <CR>
This line makes the computer send a Data Link Escape character (/ or ?), the command expressed as a number or a 3-character sequence and a CR character.
Note: Windows Terminal® and all the other terminal programs that
support the ANSI escape sequence, represent the DLE character by the string ‘^P’ and the CR character by the string ‘^M’.
Type of Commands
/OFF - Turn the meter OFF /PDR - Press Read Direct /PTR - Press Timer /PZR - Press Zero
A microprocessor controlled special tungsten lamp emits radiation which is first optically conditioned and beamed to the sample contained in the vial. The optical path is fixed by the diameter of the vial. Then the light is spectrally filtered to a narrow spectral bandwidth, to obtain a light beam of intensity
or
I
I
.
o
The photoelectric cell collects the radiation I that is not absorbed by the sample and converts it into an electric current, producing a potential in the mV range.
The microprocessor uses this potential to convert the incoming value into the desired measuring unit and to display it on the LCD.
The measurement process is carried out in two phases: first the meter is zeroed and then the actual measurement is performed.
The vial has a very important role because it is an optical element and thus requires particular attention. It is important that both the measurement and the calibration (zeroing) vials are optically identical to provide the same measurement conditions. Most of parameters use the same vial for both. In each case, insert the test vials always in the same position into the holder.
It is necessary that the surface of the vial is clean and not scratched. This is to avoid measurement interference due to unwanted reflection and absorption of light. It is recommended not to touch the vial walls with hands.
Furthermore, in order to maintain the same conditions during the zeroing and the measuring phases, it is necessary to close the vial to prevent any contamination.
Excel© Copyright of "Microsoft Co."
Lotus 1-2-3© Copyright of "Lotus Co."
Windows® and Windows Terminal® are registered Trademark of "Microsoft Co."
TELIX® is registered Trademark of "Deltacomm"
74
7
FUNCTIONAL DESCRIPTION
FRONT PANEL
1) Vial Holder
2) Dual Level Liquid Crystal Display (LCD)
3) Programs List
4) READ DIRECT, to perform measurement immediately
5) TIMER, to perform measurement after a preprogrammed countdown
6) ZERO, to zero the meter prior to measurement
7) Program t and s, to select the desired parameter
8) ON/OFF, to turn the meter on and off
REAR PANEL
1) Power Supply 12 VDC 2.5 Watt
2) RS 232 Socket
3) Batteries Compartment
Wait for a few seconds and the
display will show “-0.0-”. Now the meter is zeroed and ready for measurement.
Remove the blank vial.
Place the sample vial into the holder
and push it completely down.
• Press READ DIRECT and "SIP" will blink on the display.
• The instrument directly displays concentration in mg/L of phosphate
3-
(PO
) on the Liquid Crystal Display.
4
The method detects free (orthophosphate), condensed inorganic forms (meta-, pyro- and other polyphosphates) and organic forms of phosphates present in the sample.
• To convert the reading to mg/L of P2O5 , multiply by a factor of
0.748.
• To convert the reading to mg/L of phosphorus (P) concentration, multiply by a factor of 0.326.
Note: for accurate measurements
1) wash glassware only with phosphate-free detergents
2) clean all glassware with 1 : 1 hydrochloric acid solution and rinse with
deionized water.
INTERFERENCES
Arsenate: positive error pH: the sample should have a neutral pH Temperature: the method is temperature sensitive.
It is recommended to add the Molybdovanadate Reagent and to run measurements at T = 20 to 25 °C: T < 20 °C causes a negative error T > 25 °C causes a positive error
Turbidity: turbidity and suspended matter in large amounts may
cause interference because the strongly acidic reaction conditions may dissolve suspended matter or cause desorption of phosphates from particles. Before measurement, turbidity or suspended matter should be removed by treatment with active carbon and by prior filtration.
8
73
Phosphorus, Total HR
At the end of the digestion place the
vials carefully in the test tube rack and allow to cool to room temperature. Warning: as the vials are still hot, be careful in handling them.
• Select the program number corresponding to Total Phosphorus High Range on the secondary LCD by pressing PROGRAM t and s.
Remove the cap from the vials and add exactly 2.0 mL of Sodium Hydroxide (NaOH) Solution 1,54 N to each vial, while keeping the vials at a 45-degree angle. Replace the cap tightly and mix by inverting the vials a couple of times.
Remove the cap from the vials and add exactly 0.5 mL of HI 93763B-0 Molybdovanadate Reagent to each vial, while keeping the vial at a 45-degree angle. Replace the cap tightly and mix by inverting the vials a couple of times.
GUIDE TO DISPLAY CODES
Note: The secondary LCD below shows a generic "P– –", whereas the
meter will indicate the exact program number (e.g. "P1" for Ammonia LR).
This indicates that the meter is in a ready state and zeroing can be performed.
Sampling in progress. This flashing prompt appears each time the meter is performing a measurement.
The microprocessor is adjusting the light level, indicated by a scrolling "SIP".
This indicates that the meter is in a zeroed state and measurement can be performed.
The instrument is performing an internal check-up.
Place the blank vial into the holder and push it completely down.
Press TIMER and the display will show the countdown prior to the measurement. Alternatively, wait for 7 minutes and press ZERO. In both cases "SIP" will blink during measurement.
Phosphorus, Total HR
72
The blinking "LOBAT" indicates that the battery voltage is getting low and the batteries need to be replaced.
This indicates that the batteries are dead and must be replaced.
Light over range. The vial is not inserted correctly and an eccess ambient light is reaching the detector. If the vial is properly inserted, then contact your dealer or the nearest Hanna Customer Service Center.
9
ERROR MESSAGES a) on zero reading:
The lamp is not working properly. Contact your dealer or the nearest Hanna Customer Service Center.
The lamp is not working properly. Contact your dealer or the nearest Hanna Customer Service Center.
This indicates that the meter has lost its configuration. Contact your dealer or the nearest Hanna Customer Service Center.
This indicates that the zeroing procedure failed due to a low signal-to-noise ratio. In this case press ZERO again.
The instrument cannot adjust the light level. Please check that the vial does not contain any debris.
Reagent Blank Correction: This method needs a reagent blank correction. A single blank vial may be used more than once; the blank vial is stable up to one day (room temperature). For most accurate measurement, run a blank for each set of measurements and always use the same lot of reagents for blank and samples.
Preheat the Hanna Reactor C 9800 to 150 °C (302°F). For correct
use of the reactor follow Reactor Instruction Manual. Use of the optional HI 740217 safety shield is strongly recommended. Do not use an oven or microwave because leaking samples can generate a corrosive and possibly explosive atmosphere.
• Remove the cap from two Reagent Vials.
• Add exactly 5.0 mL of sample to one vial (sample vial), and 5.0 mL of deionized water to the other vial (blank vial), while keeping the vials at a 45-degree angle.
b) on sample reading:
There is not enough light to perform a measurement. Please check the preparation of the zero vial.
There is too much light to perform a measurement. Please check the preparation of the zero vial.
There is too much light for the sample measurement. Please check if the right sample vial is inserted.
The sample and zero vial are inverted.
10
• Add the content of one packet of Potassium Persulfate for Phosphorus analysis to each vial. Replace the cap tightly and shake gently the vials until all the powder is
completely dissolved.
Insert the vials into the reactor and heat them for 30 minutes at 150°C.
71
Phosphorus, Total HR
PHOSPHORUS, TOTAL HIGH RANGE
SPECIFICATIONS
Range 0.0 to 100.0 mg/L (as PO Resolution 0.1 mg/L Precision ±3.0 @ 75.0 mg/L Typical EMC ±0.1 mg/L
Deviation Light Source Tungsten lamp with narrow band interference filter
@ 420 nm
Method Adaptation of the Standard Methods for the
Examination of Water and Wastewater, 20 edition, 4500-P C, vanadomolybdophosphoric acid
method. A persulfate digestion converts organic and condensed inorganic forms of phosphates to orthophosphate. Then the reaction between orthophosphate and the reagents causes a yellow tint in the sample.
REQUIRED REAGENTS
Code Description Q.ty/test Q.ty/set
* Reagent Vial 1 vial 50 vials
--- Deionized Water 5 mL 1 bottle
--- Potassium Persulfate 1 packet 50 packets
--- NaOH Solution 1.54 N 2 mL 1 bottle HI 93763B-0 Molybdovanadate Reagent 0.5 mL 1 bottle
* Reagent Vial identification: P, white cap. Note: Store the unused vials in their container in a cool and dark place.
3-
)
4
A zero reading was not taken. Follow the instruction described in the measurement procedures for zeroing the meter.
Under range. A blinking "0.00" indicates that the sample absorbs less light than the zero reference. Check the procedure and make sure that you use the same vial for reference (zero) and measurement (unless a reagent blank correction procedure is prescribed).
th
1) A flashing value of the maximum concentration indicates an over range condition. The concentration of the sample is beyond the programmed range: dilute the sample and rerun the test.
2) A flashing value lower than the maximum concentration indicates a low signal-to­noise ratio condition. In this case accuracy of the result is not guaranteed. Repeat the reading procedure.
REAGENT SET
HI 93763B-50 Reagents for up to 49 tests
For other accessories see page 77.
REQUIRED ACCESSORIES
C 9800-01 Hanna reactor (115 VAC) C 9800-02 Hanna reactor (230 VAC) HI 740216 test tube cooling rack (25 holes) HI 740217 Laboratory bench safety shield
For other accessories see page 77.
MEASUREMENT PROCEDURE
Before starting to use the reagent kit it is recommended to read carefully
all the instructions and the Health & Safety Data Sheet (HSDS). Pay particular attention to all warnings, cautions and notes. Failure to do so may result in serious injury to the operator.
Phosphorus, Total HR
70
11
TIPS FOR AN ACCURATE MEASUREMENT
The instructions listed below should be carefully followed during testing to ensure most accurate results.
COLLECTING AND MEASURING SAMPLES
• For adding the exact amount of sample or liquid reagent to the reagent vials it is strongly recommended to use the available Hanna automatic fixed-volume pipettes or class A laboratory pipettes (symbolized like a generic pipette tip in the following parameter related chapters):
pipette code volume
HI 731340 200 µL HI 731341 1000 µL HI 731342 2000 µL
Place the sample vial into the holder
and push it completely down.
• Press READ DIRECT and "SIP" will blink on the display.
• The instrument directly displays concentration in mg/L of phosphate
3-
(PO
) on the Liquid Crystal Display.
4
• To convert the reading to mg/L of P2O5 , multiply by a factor of
0.748.
• To convert the reading to mg/L of phosphorus (P) concentration, multiply by a factor of 0.326.
For correct use of the automatic pipette, please follow the related Instruction Sheet.
Alternatively, it is possible to use the optional HI 740142 1 mL graduated syringe or C115-00300 5 mL graduated syringe. For the correct use of the syringes, see instructions below.
1. In order to measure exactly e.g. 5 mL of reagent with the
syringe: (a) push the plunger completely into the syringe and insert the
tip into the solution.
(b) pull the plunger up until the lower edge of the seal is exactly
on the 5 mL mark.
(c) take out the syringe and clean the outside of the syringe tip.
Be sure that no drops are hanging on the tip of the syringe, if so eliminate them. Then, keeping the syringe in vertical position above the vial, push the plunger completely down into the syringe. Now the exact amount of 5 mL has been added to the vial.
( a
)
( b
)
(c)
5 mL
Note: for accurate measurements
1) wash glassware only with phosphate-free detergents
2) clean all glassware with 1 : 1 hydrochloric acid solution and rinse with
deionized water.
INTERFERENCES
Interference may be caused by: Bismuth: negative error Fluoride: negative error pH: the sample should have a neutral pH Sulfide: negative error
To eliminate sulfide: add Bromine Water drop-wise until a pale yellow color develops; remove Bromine Water excess by adding Phenol solution drop-wise.
Temperature: the method is temperature sensitive.
It is recommended to run measurements at T = 20 to 25 °C: T < 20 °C causes a negative error T > 25 °C causes a positive error
Turbidity: turbidity and suspended matter in large amounts may
cause interference because the strongly acidic reaction conditions may dissolve suspended matter or cause desorption of phosphates from particles. Before measurement, turbidity or suspended matter should be removed by treatment with active carbon and by prior filtration.
.
12
69
Phosphorus, Reactive HR
Loading...
+ 28 hidden pages