Hanna Instruments C 125 User Manual

Instruction Manual
C 125
Portable Limpidity Meter
for Rosè & White Wine
www.hannainst.com
1
2
Dear Customer, Thank you for choosing a Hanna product. Please read this instruction manual carefully before using the meter. This
manual will provide you with the necessary information for a correct use. If you need additional technical information, do not hesitate to e-mail us at
tech@hannainst.com or contact your nearest technical office. This instrument is in compliance with the directives.
WARRANTY
All Hanna Instruments meters are warranted for two years against defects in workmanship and materials when used for their intended purpose and maintained according to instructions.
This warranty is limited to repair or replacement free of charge. Damage due to accidents, misuse, tampering or lack of prescribed maintenance are not covered.
If service is required, contact the dealer from whom you purchased the instrument. If under warranty, report the model number, date of purchase, serial number and the nature of the problem. If the repair is not covered by the warranty, you will be notified of the charges incurred. If the instrument is to be returned to Hanna Instruments, first obtain a Returned Goods Authorization number from the Customer Service department and then send it with shipping costs prepaid.
When shipping any instrument, make sure it is properly packaged for complete protection.
TABLE OF CONTENTS
Preliminary examination .............................................................. 4
Principle of operation ................................................................... 4
General description ...................................................................... 5
Functional description .................................................................. 6
Specifications ................................................................................7
Operational guide .......................................................................8
Calibration................................................................................. 10
Tips for accurate operation ......................................................... 12
Battery replacement ..................................................................13
Accessories .................................................................................. 14
CE Declaration of Conformity ....................................................... 15
3
PRELIMINARY EXAMINATION
Remove the instrument from the packing material and examine it to make sure that no damage has occurred during shipping. If there is any damage, notify your Dealer.
C125 is supplied complete with:
Glass Cuvette & Cap (2 pcs.)
Calibration solutions HI93703-0, HI93703-10 and HI93703-5
Tissue for wiping cuvettes (HI731318)
1.5V AA alkaline batteries (4 pcs.)
Instruction manual
Rugged carrying case
Note: Save all packing material until you are sure that the meter functions
correctly. All defective items must be returned in its original packaging with the supplied accessories.
PRINCIPLE OF OPERATION
C125 has been designed to perform measurements according to the ISO 7027 International Standard.
The instrument functions by passing a beam of infrared light through a vial containing the sample being measured.
The light source is a High Emission Infrared LED with a wavelength peaking at 890 nm, ensuring that the interference caused by colored samples is minimum.
A sensor, positioned at 90° with respect to the direction of light, detects the amount of light scattered by the undissolved particles present in the sample.
The microprocessor converts such readings into FTU (=FNU) values.
4
GENERAL DESCRIPTION
C125 is a portable, microprocessor-based limpidity meter, especially designed for application in wine making process for white and rosè wines.
Wine limpidity, like all other liquids, is determined by suspended small particles that deviate the light beams.
During wine making process, the wine turbidity can be caused by several phenomena, which can reduce the limpidity degree and render the product not appreciable for the consumer.
A reduced wine limpidity often goes together with unpleasant fragrance and flavor. The most common causes of wine turbidity are:
• salts (salts of tartaric acid, calcium, iron and copper)
• enzymes
• colloids (proteins, polyphenols)
• biological factors For preventing turbidity, the wine can be treated with particular additives which
act as wine limpidity stabilizers for long time. During wine production the limpidity is measured on three critical steps:
1. Alcoholic fermentation: expected value is below 600 FTU
2. Malolactic fermentation: expected value is below 100 FTU
3. Filtration: expected value is within the 0.3 to 0.5 FTU range Normally the limpidity value for bottled wine should not be greater than 4.00 FTU. The below table gives a correspondence between limpidity values and product
appearance:
Limpidity value (FTU) Appearance
0.10 to 0.50 Very limpid
0.50 to 1.00 Limpid
1.00 to 1.80 Slightly blurred
1.80 to 3.00 Blurred
3.00 to 4.00 Opalescent
> 4.00 Cloudy
5
FUNCTIONAL DESCRIPTION
1) Measurement cell
2) Liquid Crystal Display
3) ON/OFF key, to switch the meter ON and OFF
4) CAL key, to enter the calibration mode
5) READ/ key, to perform measurements and to set the date (day & month) of last calibration
6) DATE/ key, to display the last calibration date and to select either the month or the day of last calibration
6
SPECIFICATIONS
Range 0.00 to 50.00 FTU (FNU)
50 to 700 FTU (FNU)
Resolution 0.01 and 1 FTU (FNU)
Accuracy ±5% full scale (0 to 10 FTU)
±10% full scale (10 to 50 FTU) ±5% full scale (50 to 700 FTU)
Typical EMC Deviation ±2% full scale
Light Source IR LED @890 nm
Light Source Life Life of the instrument
Light Detector Silicon Photocell
Power Source 4 x 1.5V AA alkaline batteries
Battery Life 60 hours or 900 measurements
Auto-off After 5 minutes of non-use
Environment 0 to 50°C (32 to 122°F);
95% RH (non-condensing)
Dimensions 220 x 82 x 66 mm (8.7 x 3.2 x 2.6")
Weight 510 g (18 oz.)
7
OPERATIONAL GUIDE
The instrument is supplied complete with batteries. Before starting operations, install the batteries while paying attention to their polarity (see also “Battery Replacement” section). To maximize the battery life the meter automatically switches off after 5 minutes of non-use. For reactivating it, simply press the ON/OFF key.
TAKING MEASUREMENTS
Turn the meter ON by pressing the ON/OFF key.
At start-up the meter will show the full display for
a few seconds, then changes to the measurement mode.
When the LCD displays “-- --” the meter is ready to measure.
Fill a clean cuvette up to 0.5 cm (one quarter inch) from its rim with the thoroughly agitated sample.
Allow sufficient time for bubbles to escape before securing the cap.
Note: Do not over-tighten the cap.
• Wipe the cuvette thoroughly with a lint-free tissue (HI731318) before inserting into the measure­ment cell. The cuvette must be completely free of fingerprints and other oil or dirt, particularly in the area where the light goes through (approximately the bot­tom 2 cm/1 inch of the cuvette).
• Place the cuvette into the cell and check that the notch on the cap is positioned securely into the groove.
• The mark on the cuvette cap should point towards the LCD.
8
• Press the READ/ key and the LCD will display a blinking “SIP” (Sampling In Progress). The limpidity value will appear after approximately 25 seconds.
SOURCES OF INTERFERENCE
• Presence of floating debris and coarse sediment which settle out rapidly will give false readings.
• The infrared light source used for C125 limpidity meter can effectively minimize errors due to coloured dissolved substances. This effect, named “true colour”, is a common interference for most commercially available instruments operating in the range of visible light.
• Air bubbles and the effect of vibrations that disturb the surface of the sample will give false results.
• Dirty glassware could also affect readings along with scratched or edged cuvette.
9
CALIBRATION
If repeated measurements differ more than an acceptable value, check if the meter calibration status by measuring a standard solution. Anyway, it is recommended to perform a monthly calibration of the meter, or more frequently for greatest accuracy. To check the date of last calibration, press and hold the DATE/ key down for few seconds.
CALIBRATION PROCEDURE
• Turn the meter on and wait for the display to show “----”.
• Press the CAL key once, the “CAL” message will blink on the display for about 6 seconds, then the calibration mode stops.
• While the “CAL” message is still blinking, press CAL again. The instrument is now in the calibration mode and a “CL” will appear on the lower part of the display.
The date of calibration can be edited now by simply pressing the DATE/ key. To scroll to the correct number press the READ/ key.
The first blinking parameter is the month, on the left hand of the display (MM.DD).
• To confirm the displayed data values and to go to the next step, press the CAL key once. A blinking “ZERO” mes­sage will appear.
• Fill the measurement cuvette with HI 93703-0 standard (@ ZERO FTU, or turbidity-free water).
Note: In order to minimize any error introduced by the
cuvette, it is recommended to use the same cuvette during calibration and for measurements.
• Insert the cuvette with the ZERO FTU standard solution (or turbidity-free dilution water) into the measurement cell and press the CAL key.
10
“SIP” and “CL” will start blinking.
After approximately 50 seconds the instrument will ask for the HI 93703-10 standard solution (10 FTU) by displaying “10.0”.
• Fill the cuvette the HI93703-10 standard, insert into the measurement cell and press the CAL key again.
• After the second calibration point (10 FTU) has been accepted, the meter will display “500”, asking for the 500 FTU solution to be placed in the cuvette holder.
Note: At this point the user can exit the calibration mode and save the two-point
calibration by pressing READ.
• To perform a three-point calibration, place a cuvette filled with HI93703-5 standard solution (500 FTU) in the cuvette holder.
• Press the CAL key. “SIP” and “CL” will start blinking.
• After approximately 30 seconds, the display will show “----”.
Now the meter is calibrated and ready for use.
Note: If “Err1” is displayed, the calibration data
are maintained
STANDARD SUSPENSION
Hanna Instruments supplies C125 with the AMCO-AEPA-1 standards which has a long shelf life at all concentrations (approximately six months, if free from contamination). In addition, no special handling or disposal is required and a much higher stability of suspended particles has been observed.
On the other hand, the formazine-based standards are very toxic, carcinogen and unstable (particles flocculate and settle quickly). Lower concentrations change value within a few days or hours after dilution from stock solutions.
The consistency of C125 readings by using both standards has been separately established by Advanced Polymer Systems and Hanna Instruments.
Additional documentation about the formazine standard and more complex calibration procedures is available upon request.
11
VIEWING THE CALIBRATION DATE
• Turn the meter on and wait for the display to show “----”.
• Press and hold the DATE/ key and a “MM.DD” message appears while the key is held.
Note: The displayed date is the date entered by the user at the beginning of the
last calibration.
TIPS FOR ACCURATE OPERATION
The instructions listed below should be carefully followed during testing and during calibration.
• All glassware that comes into contact with standards should be maintained clean. Wash with HI93703-50 cleaning solution and rinse with HI93703-0 or turbidity free water.
• For consistent readings, it is recommended to perform measurement immediately after collecting sample.
• To obtain a representative sample, gently mix the liquid before samples are taken. Do not shake (to prevent air bubbles) and do not let the sample settle.
• Rinse the cuvette twice with 5 ml of the liquid to be tested for removing the effect of dust or any previous liquid. Gently pour the liquid down the side of the cuvette to reduce air bubbles.
• Try to tighten the cuvette cap always to the same degree.
• Before inserting the cuvette into the instrument, wipe it with soft, lint-free tissue. Handle cuvettes so that no fingerprints can get on the areas where light passes (approximately 2 cm/1 inch from the bottom of the cuvette).
• Discard the sample soon after the reading is taken to avoid permanently clouding the glass.
If any problem is experienced during measurements, please contact your dealer or the nearest Hanna Instruments Customer Service Department.
12
BATTERY REPLACEMENT
In order to minimize the battery consumption, the meter is equipped with an auto­shut off function which switches the meter off after 5 minutes of non-use.
The meter is powered by 4 alkaline batteries (1.5V AA), which are constantly monitored by the meter to ensure reliable readings.
A "LO BAT" indication will appear on the display when the batteries become weak and require replacement; the instrument can perform only 50 more measurements.
A "-BA-" message will appear on the display when the batteries level is too low for performing measure­ments; the message appears for a few seconds, then the meter automatically shuts off; immediately replace the batteries.
Battery replacement must only take place in a non hazardous area using 4 alkaline batteries, 1.5V AA size.
Turn the meter off, unscrew the 2 screws on the back of the instrument and remove the battery cover.
Substitute the rundown batteries with new ones, while paying attention to their polarity. Reattach the battery cover and tighten the 2 screws.
13
ACCESSORIES
HI731313 Maintenance kit: rugged carrying case with HI93703-0
and HI93703-10 calibration solutions, HI93703-50 cleaning solution, tissue for wiping cuvettes and 2 measurement cuvettes
HI 731318 Tissue for wiping cuvettes (4 pcs)
HI731321 Spare measurement cuvettes (4 pcs)
HI93703-0 Calibration solution, standard AMCO-AEPA-1 at 0 FTU,
30 mL bottle
HI93703-10 Calibration solution, standard AMCO-AEPA-1 at 10 FTU,
30 mL bottle
HI93703-5 Calibration solution, standard AMCO-AEPA-1 at 500 FTU,
30 mL bottle
HI93703-50 Cleaning solution, 230 mL bottle
Recommendations for Users
Before using this product, make sure that it is entirely suitable for the environment in which it is used. Operation of this instrument in residential area could cause unacceptable interferences to radio and TV equipments,
requiring the operator to take all necessary steps to correct interferences. Any variation introduced by the user to the supplied equipment may degrade the instrument's EMC performance. To avoid electrical shock, do not use this instrument when voltages at the measurement surface exceed 24VAC or
60VDC. To avoid damages or burns, do not perform any measurement in microwave ovens.
Hanna Instruments reserves the right to modify the design, construction and appearance of its products without advance notice.
14
CE DECLARATION OF CONFORMITY
15
SALES AND TECHNICAL SERVICE CONTACTS
Australia:
Tel. (03) 9769.0666 • Fax (03) 9769.0699
China:
Tel. (10) 88570068 • Fax (10) 88570060
Egypt:
Tel. & Fax (02) 2758.683
Germany:
Tel. (07851) 9129-0 • Fax (07851) 9129-99
Greece:
Tel. (210) 823.5192 • Fax (210) 884.0210
Indonesia:
Tel. (21) 4584.2941 • Fax (21) 4584.2942
Japan:
Tel. (03) 3258.9565 • Fax (03) 3258.9567
Korea:
Tel. (02) 2278.5147 • Fax (02) 2264.1729
Malaysia:
Tel. (603) 5638.9940 • Fax (603) 5638.9829
Singapore:
Tel. 6296.7118 • Fax 6291.6906
South Africa:
Tel. (011) 615.6076 • Fax (011) 615.8582
Taiwan:
Tel. 886.2.2739.3014 • Fax 886.2.2739.2983
Thailand:
Tel. (2) 619.0285 • Fax (2) 619.0284
United Kingdom:
Tel. (01525) 850.855 • Fax (01525) 853.668
USA:
Tel. (401) 765.7500 • Fax (401) 765.7575
For e-mail contacts and complete list of Sales and Technical offices, please see www.hannainst.com
MANC125 06/04
16
Loading...