Hamilton Incyte Quick Manual

Page 1
Quick Guide
Validation / Calibration
Incyte
Page 2
Validation/Calibration Incyte – Quick Guide
3
Intended Use
About this Quick Guide
The Quick Guide will help the users to validate and calibrate the Incyte Unit. Every 6 months of running time a verification of the sensor calibration is re­quested. Calibration of a Sensor Unit is required if the Sensor Unit verification did not pass successfully or, if a new sensor respectively a new pre-amplifier is connected to an existing counterpart. The Quick Guide does not replace the Operating Instructions, which are available at www.hamiltoncompany.com. All further information such as Liability, the Safety Precautions, Disposal Instructions and the Operation description are available in the Operating Instructions.
ATTENTION! Complete Instructions are available in the Cell
Density Monitoring System Operating Instructions, Ref 624702.
1 Hardware Compatibility
ATTENTION! Fogale devices are only par tially compatible
with Hamilton products.
Please do not combine Hamilton labelled Pre-Amplifier and sensors with a Control Unit having a software version below 3.0.
Evo200 and ibiomass cannot be updated with the current Hamilton Software.
2 Mount the Incyte Unit
1. Align the pins of the M12 connector and connect the Pre-amplifier to the Arc View Controller using the Sensor Unit Cable.
2. Verify that the sensor’s male VP connector and the Pre-amplifier’s female VP connector are dry and clean.
3. Align the sensor VP head and the Pre-amplifier connector and connect the Sensor in the Pre-amplifier.
4. Hand-tighten the connecting nut of the Pre-amplifier’s VP Connector to ensure a robust and waterproof seal.
HAMILTON WARRANTY
Please refer to the General Terms of Sales (GTS).
IMPORTANT NOTE
Copyright © 2017 Hamilton Bonaduz AG, Bonaduz Switzerland. All rights reser ved. The reproduction of any part of this document in any form is forbidden without the express written agreement of Hamilton Bonaduz AG.
Contents of this manual can be modified without previous announcement. Technical modifications reser ved. Greatest possible care was used on the correctness of the information in this manual. If errors should be discovered nevertheless, Hamilton Bonaduz AG is pleased to be informed about it. Regardless of this, Hamilton Bonaduz AG cannot assume liabilit y for any errors in this manual or for their consequences.
Page 3
Validation/Calibration Incyte – Quick Guide
5
5. Mount the Unit in a stand and place in a 500 mL beaker (Figure 1). Select the most appropriate position for the sensor. The electrodes of the Incyte sensor have to be kept at least 1 cm away from the beaker wall.
1 CM
1 CM
A
B
Figure 1: Mounting restrictions, A: Space requirements / B: Reactor mounting restrictions.
3 Verify the Incyte Sensor Unit
NOTE: The Capacitance Simulator (Ref 243743) and Hamilton 12880 μS/cm Conductivity Standard (Ref 238988), formerly called Solution A, as well as a thermometer are needed to perform the Sensor Unit Verification.
1. Connect the Incyte Unit and power on the Arc View Controller and equili­ brate for at least 30 minutes prior to starting the Verification Procedure.
2. Go to the Main Menu and enter the Calibration Menu. Select the Validation / Check Unit Calibration to start the Verification Procedure.
3. Select the channel. The sensor type is automatically detected.
4. Enter the user name.
5. Connect the Capacitance Simulator and press Test Pre-amplifier Calibration to start the Verification Procedure. The successful verification is indicated. If this procedure fails it is recommended to perform a calibration.
6. Connect the Incyte Sensor to the Pre-amplifier.
7. Select the Manual temperature correction mode. The Automatic mode is disabled.
8. Pour Hamilton 12880 μS/cm Conductivity Standard in a clean beaker. Make sure the sensor mounting instructions are followed.
9. Measure the temperature of the Standard with an external thermometer. Enter the conductivity value according to the temperature correction table on the bottle of the Conductivity Standard.
10. Press Test Sensor Calibration to start the Verification. The successful Verification is indicated. If this procedure fails it is recommended to perform a calibration.
4 Calibrate the Incyte Sensor Unit
The Calibration procedure of an Incyte Sensor Unit is a five-step guided procedure. It takes about an hour to perform this procedure.
Page 4
Validation/Calibration Incyte – Quick Guide
7
ATTENTION! Do not disconnect the Unit from the Arc View Controller, or the Sensor from the Pre-amplifier during the Calibration procedure.
NOTE: Make sure to prepare following equipment in addition to the
Val/Cal Kit: stirrer-plate, thermometer
1. Connect the Unit and power on the Arc View Controller and equilibrate for at least 30 minutes prior to starting the Calibration procedure.
2. Go to the Main Menu and enter the Calibration Menu. Select Calibrate Unit to star t the Calibration procedure.
3. Select the channel. The sensor type is automatically detected.
4. Enter the user name and press start (Figure 2).
Figure 2: Start of the sensor unit calibration.
5. Connect the Capacitance Simulator and press Test Pre-amplifier Calibration to start the Calibration procedure. The successful test is indicated. If this procedure fails it is recommended to contact the local representative.
6. Replace the Simulator by the Incyte Sensor and Press measure in air to calibrate at zero conductivity. If this procedure fails it is recommended to contact the local representative.
7. Select Manual temperature correction mode. The Automatic mode is disabled.
8. Pour Hamilton 12880 μS/cm Conductivity Standard in a clean beaker. Make sure the sensor mounting instructions are followed.
9. Measure the temperature of the Standard with an external thermometer. Enter the conductivity value according to the temperature correction table on the bottle of the Conductivity Standard.
10. Press Calibrate sensor constant to start the Calibration (Figure 3). The successful Calibration is indicated. If this procedure fails it is recommended to contact the local representative.
Figure 3: Calibration of the Sensor Constant.
Page 5
Validation/Calibration Incyte – Quick Guide
9
11. Rinse the Sensor with deionized water.
12. Pour 150 mL deionized water in a clean beaker. In the following steps 1 to 30 mL of Solution B is added gradually to the deionized water. Press start (Figure 4).
13. Add the required amount of Solution B. Press Star t.
Figure 4: In cyte Se nsor Ca libration.
15. The successful Calibration is indicated. If this procedure fails it is recommended to contact the local representative.
16. Choose the directory, where the Calibration Report should be stored and press Generate Calibration Report (Figure 5). The valid calibration is automatically saved in the memory of the Pre-amplifier.
Figure 5: Generation of a Ca libration Re port.
Page 6
Validation/Calibration Incyte – Quick Guide
11
Page 7
© 2017 Hamilton Bonaduz AG. All rights reserved.
624859/02 — 06/2017
Web: www.hamiltoncompany.com
USA: 800-648-5950 Europe: +41-58-610-10-10
Hamilton Americas & Pacific Rim
4970 Energy Way Reno, Nevada 89502 USA Tel: +1-775-858-3000 Fax: +1-775-856-7259 [email protected]
Hamilton Europe, Asia & Africa
Via Crusch 8 CH-7402 Bonaduz, Switzerland Tel: +41-58- 610-10-10 Fax: +41-58-610-00-10 [email protected]
To find a representative in your area, please visit www.hamiltoncompany.com.
This guide may be available in other languages. Visit www.hamiltoncompany.com for more information.
Loading...