Hach-Lange HQD Portable Meter User Manual

DOC022.52.80017
HQd Portable Meter
06/2013, Edition 4
User Manual
Table of Contents
Specifications..................................................................................................................................................................................5
General information.....................................................................................................................................................................5
Safety information..............................................................................................................................................................................5
Precautionary labels..........................................................................................................................................................................6
Product overview...............................................................................................................................................................................6
Product components..........................................................................................................................................................................6
Installation.........................................................................................................................................................................................7
Install the batteries.............................................................................................................................................................................7
Connect to AC power.........................................................................................................................................................................8
User interface and navigation...............................................................................................................................................9
User interface.....................................................................................................................................................................................9
Display description.............................................................................................................................................................................9
Navigation........................................................................................................................................................................................11
Startup...............................................................................................................................................................................................11
Turn the meter on and off................................................................................................................................................................11
Change the language.......................................................................................................................................................................11
Change the date and time................................................................................................................................................................12
Connect a probe..............................................................................................................................................................................12
Standard operation....................................................................................................................................................................12
About calibration..............................................................................................................................................................................12
About sample measurements..........................................................................................................................................................13
About check standards....................................................................................................................................................................13
Use a sample ID..............................................................................................................................................................................13
Use an operator ID...........................................................................................................................................................................13
Data management.......................................................................................................................................................................13
About stored data.............................................................................................................................................................................13
View stored data..............................................................................................................................................................................14
View stored probe data.............................................................................................................................................................14
1
Table of Contents
Print stored data...............................................................................................................................................................................14
Change the report options........................................................................................................................................................15
Send data to a USB storage device.................................................................................................................................................15
Open data files on a PC...................................................................................................................................................................16
Data file description.........................................................................................................................................................................16
Remove column headers.................................................................................................................................................................18
Send data directly to a computer.....................................................................................................................................................18
Advanced operation..................................................................................................................................................................18
Security options...............................................................................................................................................................................18
Turn Security Options on..........................................................................................................................................................18
Full access options menu.........................................................................................................................................................19
Restricted operator access options menu................................................................................................................................19
Set the display options.....................................................................................................................................................................20
Set the sound options......................................................................................................................................................................20
Set the probe calibration reminder............................................................................................................................................21
Change the temperature units.........................................................................................................................................................21
Set the measurement mode.............................................................................................................................................................21
Set auto measurement intervals...............................................................................................................................................21
Start interval measurements.....................................................................................................................................................21
Prevent data log overflow in interval mode...............................................................................................................................22
View instrument information.............................................................................................................................................................22
Update the meter software...............................................................................................................................................................22
Download software updates.....................................................................................................................................................23
Transfer method settings.................................................................................................................................................................23
Bi-directional Communication between the meter and a PC...........................................................................................................23
About meter control..................................................................................................................................................................23
About meter configuration.........................................................................................................................................................24
Maintenance...................................................................................................................................................................................24
Clean the meter...............................................................................................................................................................................24
Replace the batteries.......................................................................................................................................................................24
Troubleshooting..........................................................................................................................................................................25
2
Table of Contents
Replacement parts and accessories...............................................................................................................................26
Examples of printed reports.................................................................................................................................................
28
Index....................................................................................................................................................................................................33
3
Table of Contents
4

Specifications

Specifications are subject to change without notice.
Specification Details
Dimensions 19.7 x 9.5 cm (7.75 x 3.75 in.)
Weight 335 g (0.75 lb) without batteries; 430 g (0.95 lb) with
Meter enclosure IP67, waterproof to 1 meter for 30 minutes
Battery enclosure Water resistant to 0.6 m (2 ft) for 15 seconds
Power requirements (internal)
Power requirements (external)
Meter protection class Class I
Storage temperature –20 to +60 °C (–4 to +140 °F)
Operating temperature 0 to +60 °C (32 to 140 °F)
Operating humidity 90% (non-condensing)
5-pin input connector M12 connector for IntelliCAL™ probes
8-pin input connector The 8-pin connector enables USB and external AC
USB/DC adapter Peripheral and host
Data memory (internal) 500 results
Data storage Automatic in Press to Read mode and Interval
Data export USB connection to PC or USB storage device
four AA alkaline batteries
AA Alkaline or rechargeable Nickel Metal Hydride (NiMH) batteries (4); battery life: up to 200 hours
Class II, external power adapter: 100–240 VAC, 50/60 Hz input; 4.5 to 7.5 VDC (7 VA) output
power connectivity
Mode. Manual in Continuous Read Mode.
(limited to the storage device capacity). Transfer entire data log or as readings are taken.
Specification Details
Connections Integrated USB type A (for USB flash memory
Temperature correction Off, automatic and manual (parameter dependent)
Measurement display lock Continuous measurement, Interval or Press to Read
Keyboard External PC keyboard connector via USB/DC
device, printer, keyboard) and Integrated USB type B (for PC)
mode. Averaging function for LDO probes.
adapter

General information

Revised editions are found on the manufacturer’s website.

Safety information

N O T I C E
The manufacturer is not responsible for any damages due to misapplication or misuse of this product including, without limitation, direct, incidental and consequential damages, and disclaims such damages to the full extent permitted under applicable law. The user is solely responsible to identify critical application risks and install appropriate mechanisms to protect processes during a possible equipment malfunction.
Please read this entire manual before unpacking, setting up or operating this equipment. Pay attention to all danger and caution statements. Failure to do so could result in serious injury to the operator or damage to the equipment.
Make sure that the protection provided by this equipment is not impaired. Do not use or install this equipment in any manner other than that specified in this manual.

Use of hazard information

Indicates a potentially or imminently hazardous situation which, if not avoided, will result in death or serious injury.
D A N G E R
English 5
W A R N I N G
Indicates a potentially or imminently hazardous situation which, if not avoided, could result in death or serious injury.
Indicates a potentially hazardous situation that may result in minor or moderate injury.
Indicates a situation which, if not avoided, may cause damage to the instrument. Information that requires special emphasis.
C A U T I O N
N O T I C E

Precautionary labels

Read all labels and tags attached to the instrument. Personal injury or damage to the instrument could occur if not observed. A symbol on the instrument is referenced in the manual with a precautionary statement.
This symbol, if noted on the instrument, references the instruction manual for operation and/or safety information.
This symbol indicates that the marked item can be hot and should not be touched without care.
Measurement data can be stored and transferred to a printer, PC or USB storage device.
The HQd series meters are available in 4 models:
HQ11d—pH/mV/ORP
• HQ14d—conductivity, salinity, total dissolved solids (TDS), resistivity
• HQ30d—all IntelliCAL probes, 1 probe connector
• HQ40d—all IntelliCAL probes, 2 probe connectors
Features common to all models:
• Automatic probe and parameter recognition
• Instrument guided calibration procedures
• Calibration data stored in the probe
• Probe specific method settings for regulatory compliance and Good Laboratory Practice (GLP)
• Security Options
• Real-time data logging via USB connection
• USB connectivity to PC/printer/USB storage device/keyboard
• Bi-directional communication with PC-based systems via a virtual serial port connection
• Sample ID and Operator ID for data traceability
• Adjustable automatic shut-off
Electrical equipment marked with this symbol may not be disposed of in European public disposal systems after 12 August of 2005. In conformity with European local and national regulations (EU Directive 2002/96/EC), European electrical equipment users must now return old or end-of-life equipment to the Producer for disposal at no charge to the user.
Note: For return for recycling, please contact the equipment producer or supplier for instructions on how to return end-of-life equipment, producer-supplied electrical accessories, and all auxiliary items for proper disposal.

Product overview

The HQd series portable meters are used with digital IntelliCAL™ probes to measure various parameters in water. The meter automatically recognizes the type of probe that is connected to the meter.
6 English

Product components

Refer to Figure 1 and Figure 2 to make sure that all components have been received. If any items are missing or damaged, contact the manufacturer or a sales representative immediately.
Figure 1 Meter components (HQ40d model)
1 Meter 4 AC-DC power supply
2 AA batteries (pk/4) 5 USB/DC adapter
3 AC power cord
Figure 2 Meter components (HQ11d, HQ14d and HQ30d models)

Installation

C A U T I O N
Multiple hazards. Only qualified personnel must conduct the tasks described in this section of the document.

Install the batteries

W A R N I N G
Explosion hazard. Incorrect battery installation can cause the release of explosive gases. Be sure that the batteries are of the same approved chemical type and are inserted in the correct orientation. Do not mix new and used batteries.
The battery compartment is not waterproof. If the battery compartment becomes wet, remove and dry the batteries and dry the interior of the compartment. Check the battery contacts for corrosion and clean them if necessary.
When using nickel metal hydride (NiMH) batteries, the battery icon will not indicate a full charge after freshly charged batteries have been inserted (NiMH batteries are 1.2 V versus 1.5 V for alkaline batteries). Even though the icon does not indicate complete charge, 2300 mAH NiMH batteries will achieve 90% of instrument operation lifetime (before recharge) versus new alkaline batteries.
To avoid potential damage to the meter from battery leakage, remove the meter batteries prior to extended periods of non-use.
The meter can be powered with AA alkaline or rechargeable NiMH batteries. To conserve battery life, the meter will power off after 5 minutes of inactivity. This time can be changed in the Display Options menu.
For battery installation refer to Figure 3.
N O T I C E
N O T I C E
N O T I C E
1 Meter 2 AA batteries (pk/4)
English 7
1. Pull the release tab on the battery cover and the remove the cover.
2. Install 4 AA alkaline or 4 AA nickel metal hydride (NiMH) batteries.
Make sure that the batteries are installed in the correct polarity.
3. Replace the battery cover.
Figure 3 Battery installation

Connect to AC power

D A N G E R
Electrocution Hazard. AC power outlets in wet or potentially wet locations MUST ALWAYS be provided with a Ground Fault Circuit Interrupting (GFCI/GFI) circuit breaker. The AC-DC power adapter for this product is not sealed and must not be used on wet benches or in wet locations without GFCI protection.
The meter can be powered by AC power with an AC power adapter kit. The kit includes an AC-DC power supply, USB/DC adapter and AC power cord.
1. Turn the meter off.
2. Plug the AC power cord into the AC-DC power supply (Figure 4).
3. Connect the AC-DC power supply to the USB/DC adapter.
4. Connect the USB/DC adapter to the meter.
5. Plug the AC power cord into an AC receptacle.
6. Turn the meter on.
1 Batteries 2 Release tab 3 Battery cover
8 English
Figure 4 AC power connection

User interface and navigation

User interface

Figure 5 Keypad description
1 USB storage device/printer/Qwerty
keyboard connection (USB peripheral)
2 Personal computer connection
(USB host)
3 USB/DC adapter
4 AC-DC power supply
5 AC power cord
1 ON/OFF: turn on or turn off the
meter
2 OPERATOR ID: associate data with
an individual
3 BACKLIGHT: illuminate the display
screen
4 SAMPLE ID: associate data with a
sample location
5 BLUE/LEFT key: calibrates, cancels
or exits the current menu
6 UP and DOWN key: scroll through
menus, enter numbers and letters or change the reading screen view
7 GREEN/RIGHT key: reads, selects,
confirms or stores data
8 DATA LOG: recall or transfer stored
data
9 METER OPTIONS: change
settings, run check standards, view meter information

Display description

Measurement screen
The meter display shows the concentration, units, temperature, calibration status, operator ID, sample ID, date and time (Figure 6).
English 9
Figure 6 Single screen display
Figure 7 Single-screen display—big-screen mode
1 Calibration status indicator 9 Time
2 Main measurement value and unit 10 Date
3 IntelliCAL probe type and port
indicator
4 Battery status 12 Display size icon
5 Power source 13 Calibrate (Cancel, Exit)
6 Sample temperature (ºC or ºF) 14 Sample and operator identification
7 Secondary measurement unit 15 Stability or display lock indicator
8 Tertiary units (for some probes)
11 Read (OK, Select)
Big-screen mode
The font size of the sample reading can be increased or decreased with the key (Figure 7).
Note: When two probes are connected, push and hold the big-screen mode. The big-screen mode can also be selected in the Display Options menu (Refer to Set the display options on page
20).
key to select the
10 English
1 Calibration status indicator 5 Main measurement unit
2 IntelliCAL probe type and port
indicator
3 Power source or battery status 7 Sample temperature (ºC or ºF)
4 Main measurement value 8 Stability or display lock indicator
6 Display size icon
Dual-screen mode (HQ40d model only)
When two probes are connected to the HQ40d meter, the display can show the reading from both probes simultaneously or show just one probe (Figure 8).
Note: For probe calibration, change the screen mode to the single screen mode.
To change the screen mode to single or dual screen, use the and
keys. In dual screen mode, the key will select the probe on the
left and the key will select the probe on the right.
Figure 8 Dual-screen display
1 Probe that is connected to port on
left
2 Probe that is connected to port on
right
3 Measurement information for probe
on left
4 Measurement information for probe
on right

Navigation

The meter contains menus to change various options. Use the
keys to highlight different options. Push the GREEN/RIGHT key to
select an option. There are two ways to change options:
1. Select an option from a list: Use the and keys to select an option. If check boxes are shown, more than one option can be selected. Push the BLUE/LEFT key under Select.
Note: To deselect check boxes, push the
BLUE/LEFT key under Deselect.
2. Enter an option value using the arrow keys: Push the
and keys to enter or change a value.
3. Push the GREEN/RIGHT key to advance to the next space.
4. Push the GREEN/RIGHT key under OK to accept the value.
and

Startup

Turn the meter on and off

Push the key to turn on or turn off the meter. If the meter does not turn on, make sure that the batteries are properly installed or that the AC power supply is properly connected to an electrical outlet.

Change the language

The display language is selected when the meter is powered on for the first time. The language can also be changed from the Meter Options menu.
Access to the language menu can be restricted with the Security Options. Refer to Security options on page
1. Push the
key and select Language.
2. Select a language from the list.
Note: While turning the meter on, the language can also be changed when the power key is pushed and held.
18.
English 11

Change the date and time

The date and time can be changed from the Date & Time menu.
1. Push the key and select Date & Time.
2. Update the time and date information:
Option Description
Format Select one of the formats below for the date and time. Use the
and keys to select from the format options.
dd-mm-yyyy 24h
dd-mm-yyyy 12h
mm/dd/yyyy 24h
mm/dd/yyyy 12h
dd-mmm-yyyy 24h
dd-mmm-yyyy 12h
yyyy-mm-dd 24h
yyyy-mm-dd 12h
Date Use the and keys to enter the current date.
Time Use the and keys to enter the current time.
Figure 9 Probe connection
The current date and time will be shown on the display.
Connect a probe after the date and time setup, so that the meter is ready to take a measurement.

Connect a probe

1. Make sure that the display shows the current time and date.
Note: The time stamp for a probe is set when the probe is first connected to the meter. This time stamp makes it possible to record the probe history and record the time when measurements are made.
2. Plug the probe into the meter (Figure 9).
3. Push and turn the locking nut to tighten.
12 English
1 Probe connection port
(HQ40d only)
2 USB/DC adapter port 3 Probe connection port

Standard operation

About calibration

Each probe uses a different type of calibration solution. Make sure to calibrate the probes frequently to maintain the highest level of accuracy.
Note: For step-by-step instructions, refer to the documents that are included with each probe.
The calibration icon can indicate that:
the calibration timer has expired
• the LDO sensor cap should be replaced
the calibration is out of range
• the calibration results are outside acceptance criteria settings

About sample measurements

Each probe has specific preparation steps and procedures for taking sample measurements. For step-by-step instructions, refer to the documents that are included with the probe.

About check standards

Run Check Standards verifies equipment accuracy by measuring a solution of a known value. The meter will indicate if the Check Standard
passed or failed. If the Check Standard fails, the calibration icon
is
shown until the probe is calibrated.
The meter can be set to automatically show a reminder for check standard measurement at a specified interval with a specified acceptance criteria. The reminder, value of the check standard, and acceptance criteria can be changed. For step-by-step instructions, refer to the documents that are included with the probe.

Use a sample ID

The sample ID tag is used to associate measurements with a particular sample location. If assigned, stored data will include the sample ID.
1. Push the key.
2. Select, create or delete a sample ID:
Option Description
Current ID Select an ID from a list. The current ID will be
Create a New SampleIDEnter a name for a new sample ID.
Delete Sample ID Delete an existing sample ID.
associated with sample data until a different ID is selected.

Use an operator ID

The operator ID tag associates measurements with an individual operator. All stored data will include the operator ID.
1. Push the key.
2. Select, create or delete an operator ID:
Option Description
Current ID Select an ID from a list. The current ID will be
Create a New Operator ID
Delete Operator ID Delete an existing operator ID.
associated with sample data until a different ID is selected.
Enter a name for a new operator ID (maximum 10 names can be entered).

Data management

About stored data

The following types of data are stored in the data log:
Sample measurements: stored automatically each time a sample is measured in the Press to Read or Interval Mode. When the continuous measurement mode is used, data is stored only when Store is selected.
• Calibrations: stored only when Store is selected at the end of a calibration. Calibration data is also stored in the IntelliCAL (R) probe.
• Check standard measurements: stored automatically each time a check standard is measured (in the Press to Read or Interval Mode).
When the data log becomes full (500 data points), the oldest data point is deleted when a new data point is added. The entire data log can be deleted to remove data that has already been sent to a printer or PC
key > Delete Data Log). To prevent deletion of the data log by a
( user, use the Security Options menu.
English 13

View stored data

The data log contains sample, calibration and check standard data. The most recent data point in the data log is tagged as Data Point 001.
1. Push the key.
2. Select View Data Log to view the stored data. The most recent data
point is shown. The top of the screen shows whether the data is from a sample reading, a calibration or a check standard. Push the key to view the next most recent data point.
Option Description
Reading Log Reading Log—shows sample measurements including
Calibration Log Calibration Log—shows calibration data. Select Details
Check Standard Log

View stored probe data

Make sure that a probe is connected to the meter. If two probes are connected, select the appropriate probe when prompted.
1. To view the calibration data that is stored in a probe, push the key and select View Probe Data. The current calibration and calibration history for the probe can be viewed.
Option Description
View Current Calibration
View Calibration History
the time, date, operator and sample ID. Select Details to view the associated calibration data.
to view additional information about the calibration.
Check Standard Log—shows check standard measurements. Select Details to view the calibration data that was associated with the measurement.
The current calibration information shows the calibration details for the most recent calibration. If the probe has not been calibrated by the user, the factory calibration data is shown.
The calibration history shows a list of the times when the probe was calibrated. Select a date and time to view a summary of the calibration data.

Print stored data

The meter must connect to AC power to start the USB connection. Make sure that the connection to AC power is made before the meter is powered on. All data can be sent to a printer. The PD-24 Citizen Printer is compatible with the HQd meters and is FCC Part 15B, Class B compliant with the HQ meters. Other printers may not be compatible. Compatible printers should support a minimum of 72 columns of data, be capable of printing up to 500 continuous data-stream events in 1, 2 and 3 lines of text and fully support code page 437 and code page 850.
1. Turn off the meter. Make sure that the meter is connected to AC power. Refer to Connect to AC power on page 8.
2. Connect the printer to the meter with a USB cable type A. Refer to
Figure 10.
3. Turn on the meter.
4. Push the key.
5. Select Send Data Log. Wait for the display to show “Transfer
Complete” and wait for the printer to stop printing. Disconnect the printer.
14 English
Figure 10 Connection to the printer

Change the report options

Printed reports for sample data can contain 1, 2 or 3 lines of information. Refer to Examples of printed reports on page 28 information.
for further
1 Meter 7 USB cable
2 AC-DC power supply 8 USB/DC adapter
3 AC power cord 9 Port for probe connection
4 AC power outlet 10 Port for USB/DC adapter
5 Power supply for printer (optional) 11 Port for probe connection
6 Citizen Printer, FCC Part 15B,
Class B compliant
1. Push the
key. Select Report Options.
2. Select Report Type and select one of the options.
Option Description
Basic report One line of data.
Advanced report Two lines of data. The first line contains the same
Total report Three lines of data. The first two lines contain the same
information as in the basic report.
information as in the advanced report.

Send data to a USB storage device

N O T I C E
The transfer of a large number of data points will take some time. DO NOT disconnect the USB storage device until the transfer is complete.
Data can be transferred to a USB storage device for storage or transfer to a computer.
1. Turn off the meter. Make sure that the meter is connected to AC power.
2. Plug the USB storage device into the meter before the meter is powered on.
3. Turn on the meter.
4. Push the key.
5. Select Send Data Log. Wait for the display to show “Transfer
Complete” and for any lights on the USB storage device to stop flashing. Then remove the USB device.
Note: If the data transfer is slow, reformat the USB storage device to use the file allocation table (FAT) format for the next use.
English 15

Open data files on a PC

Data that has been downloaded to a USB storage device can be transferred to a computer. The data is sent in a text (.txt) file format.
1. Plug the USB storage device into the computer.
2. Find the data file. The file will have the following format: “Meter Serial
Number-Data File Type-Date Time”. Example: 9999NN000000­SENDDATA-0603131624.TXT
3. Save the data file to a location on the computer.
4. Open a spreadsheet program such as Microsoft® Excel
®
spreadsheet software.
5. Open the data file from the spreadsheet program. Select the delimited option with comma as the delimiter. The data will be shown in the spreadsheet program.
Note: If an application is used that is not compatible with column headings, the headings can be omitted. Refer to Remove column headers on page 18
.

Data file description

Data that is saved to a USB storage device and then opened in a spreadsheet application will have multiple columns of data. A description for each of the columns is shown in Table 1.
Table 1 Spreadsheet column descriptions
Column header name Data description and example values
Type Type of data: RD = Reading; CL = Calibration; CK =
Parameter Type Parameter: LDO, pH, CD (conductivity), ORP,
Date Date of reading: stored in user-defined date format
Time Time of reading: stored in user-defined format
Operator ID Operator ID that was used when the data was
Check Standard; CH = Calibration History; IC = Current Calibration
ISE
recorded. Shows “- - -” if no operator ID is used.
Table 1 Spreadsheet column descriptions (continued)
Column header name Data description and example values
Probe Model Model number of probe, for example pHC101,
Probe SN Probe serial number
Method name User-defined name of the probe settings that were
Sample ID Sample ID that was used when the data was
Primary Reading Value Measured value. Shows “—" if the value was out of
Primary Reading Units Measurement units, for example pH or µS/cm
Supp Reading 1 First supplemental reading (example: temperature), if
Supp Units 1 Units for first supplemental reading, if applicable.
Supp Reading 2 Second supplemental reading (example: “mV” for pH),
Supp Units 2 Units for second supplemental reading, if applicable
Supp Reading 3 Third supplemental reading, if applicable
Supp Units 3 Units for third supplemental reading, if applicable.
Reading Setting 1–4 Any settings that affect the reading, for example
Any settings that affect the reading, for example “NaCl/Non-Linear”
CDC401, LDO101
If two probes are connected to the HQ40d meter, the serial number shows “<“ or “>” to identify the port (left or right) the probe was connected to during the reading.
used for the reading.
recorded. Shows "Sample ID" if the default sample ID is used.
range.
applicable
if applicable
“NaCl/Non-Linear”
Reading Message 1–4 Any message that was shown during the measurement, for example “Out of limits”.
16 English
Table 1 Spreadsheet column descriptions (continued)
Column header name Data description and example values
Check Std Value Value of the check standard that was used to verify
Check Std Units Check standard units, for example μS/cm.
Check Std Graph Bar-graph showing the measurement in relation to the
Check Std Status Status of the check standard reading. Example:
Calibration Status
Cal Date Date of calibration reading: stored in user-defined date
Cal Time Time of calibration reading: stored in user-defined
Cal Operator ID The operator ID specified when the probe was
Cal Slope Name Slope (pH or LDO) or cell constant (conductivity)
Cal Slope The slope value for the calibration
Cal Slope Aux Used by pH to give the percent of nominal slope
Cal Slope Units Units of the calibration slope. Example: “mV/pH” for
Cal Offset Calibration offset value
Cal Offset Units Calibration offset units. Example: “mV” for pH.
Cal r² Calibration correlation coefficient without a unit (may
accuracy, for example: 7.00 pH–25 ºC (pH, temp­compensated); 7.01 pH (pH, custom)
Note: pH is not displayed here as it is included in the previous column.
acceptance limits. Example: “6.901 <—|—> 7.101”.
“Reading within limits”, “Reading outside limits”
Status of the calibration that is in use. = current calibration is valid;
format
time format
calibrated. Shows “- - -” if undefined.
pH
be blank)
= calibration has expired.
Table 1 Spreadsheet column descriptions (continued)
Column header name Data description and example values
Cal Number of Std’s Number of standards used during calibration, for
Cal Std 1 Known value of the first calibration standard
Cal Std 1 Units Units of the first calibration standard
Cal Std 1 Primary Value Measured value of the first calibration standard
Cal Std 1 Primary Units Associated units for the calibration measurement
Cal Std 1 Supp Value Value of supplemental measurement, for example
Cal Std 2–7 Known value of additional calibration standards, if
Cal Std 2–7 Units Units of additional calibration standards, if used
Cal Std 2–7 Primary Value
Cal Std 2–7 Primary Units
Cal Std 2–7 Supp Value Value of supplemental measurement, for example
Cal Std Supp Units Units applicable to all secondary calibration readings.
Cal Message 1–4 Any messages about the calibration
Date/Time POSIX Date and time of reading stored in POSIX format
Cal Date/Time POSIX Date and time of calibration stored in POSIX format
Meter SN Meter serial number used to take the measurement
example 5. May be blank depending on record type, parameter type, and method settings.
temperature
used
Measured values of additional calibration standards, if used
Associated units for additional calibration measurements, if used
temperature
Example: “ºC” or “ºF” for temperature
(number of seconds from January 1, 1970) Example: 1149234913
(number of seconds from January 1, 1970). Example: 1111320348
English 17

Remove column headers

When transferred data is viewed in a spreadsheet program, the first row of data contains headings to identify the type of data in each column. If an application or post-processing method is used that is incompatible with the headers, the column headers can be omitted.
1. Push the key.
2. Select Column Headers.
3. Set the column headers to off.

Send data directly to a computer

Data can be transferred from any HQd series meter directly to a computer when the HQ40d PC Application is installed. The data can be sent in real time during data collection, or the entire data log can be transferred.
To download the most current version of the software, refer to the applicable product page on the manufacturer's website.
1. Install the HQ40d PC Application on the computer.
2. Turn off the meter. Make sure that the meter is connected to AC
power.
3. Connect the PC to the meter with a USB type B cable.
4. Turn on the meter.
5. Open the HQ40d PC Application on the computer. Click on the green
triangle in the menu bar to start a connection.
6. Collect the data in real time or transfer the data from the data log:
Real time—when a data point is stored in the meter, the result is sent simultaneously to the PC Application (refer to Set the
measurement mode on page 21).
• Data log—push the
the display to show “Transfer Complete.” The data is sent as a comma separated values (.csv) file.
The data is shown in the HQ40d PC Application window.
key and select Send Data Log. Wait for

Advanced operation

Security options

The Security Options menu is used to protect the meter setup and method settings from unwanted changes. This menu is available in the Full Access Options menu.
The Setup Measurement Mode, Date and Time, Temperature Units, Language, Probe settings, Delete data log and Security Options screens are disabled in the Operator Access Options menu. All menu options are enabled in the Full Access Options menu.
Note: The Full Access Options menu is shown when the key is pushed when Security Options is OFF, whether or not a password has been set.
When the meter is powered on for the first time and Security Options is selected, the display prompts the user to set a password. Until the meter
is shut off, pushing the menu, even after Security Options is turned on and a password has been set. After the meter is shut off and powered on again with Security Options on, the Operator Access Options menu is displayed until a valid password is entered.
Store the password in a safe and accessible place. If the specified password is forgotten and Security Options is turned on, the operator is locked out of the restricted menus. Contact technical support if the password is lost.

Turn Security Options on

The Security Options and the Set Password options are used together to prevent access to restricted menus.
key will still display the Full Access Options
18 English
1. Push the key and select Security Options.
2. Change the settings as needed to allow or prevent menu access.
Option Description
Security Options
Set Password Set a password that must be entered to enable the Full
When Security Options is on, and a password has been specified, the password is required to enable the Full Access Options menu. If the meter is turned off while Security Options is on, the password is required to enable the Full Access Options menu again when the meter is turned on.
Access Options menu. The requirement for password entry is controlled by setting Security Options on or off.

Full access options menu

The Full Access Options menu is displayed when Security Options is OFF or when Security Options is ON and a valid password is entered (Table 2). These options do not need to be changed if the factory default settings are used.
Table 2 Full Access Options
Option Description
(Probe model) settings Settings such as measurement options, calibration
Run check standard Measure standard solution (available for pH,
Measurement mode Press to Read
Instrument information Probe information
options, check standard options, units and resolution. Refer to the probe documentation for more information.
Note: A probe must be connected to the meter.
conductivity, ORP and ISE probes)
Interval: Duration and Interval
Continuous
Meter information
Table 2 Full Access Options (continued)
Option Description
Security options ON or OFF
Set password
Display options Contrast
Auto shutoff
Backlight
Mode
Sounds Key press
Stability alert
Calibration reminder
Date and time Format
Date
Time
Temperature units Set temperature units
Language Select language

Restricted operator access options menu

The Operator Access Options menu is shown at meter startup when Security Options is ON (Table 3). When a valid password is entered, the menu changes to Full Access Options.
English 19
Table 3 Operator access options
Option Description
(Probe model) settings Only methods (if methods exist) can be selected. Refer
Run check standard Measure standard solution (available for pH,
Instrument information Probe information
Access password Enter password
Display options Contrast
Sounds Key press
to the probe documentation.
Note: A probe must be connected to the meter.
conductivity, ORP and ISE probes)
Note: A probe must be connected to use this option.
Meter information
Auto shutoff
Backlight
Mode
Stability alert
Calibration reminder

Set the display options

Use Display Options to change the display contrast, battery saving auto­shutoff options, the backlight option or the detailed or big reading screen mode.
1. Push the key and select Display Options.
2. Select which display option to change.
Option Description
Contrast Adjust the contrast of the display. The lightest setting is 0 and
the darkest setting is 9.
Option Description
Auto­shutoff
Backlight The display backlight is turned off when the key is pushed.
Mode Select Detailed or Big screen size. Detailed will show more
To maximize battery life, set a time period after which the meter will automatically power off if no key is pushed (1, 2, 5, 10, 30 min, 1 h, 2 h or never). Auto-shutoff is not active when the meter is connected to AC power or in the Interval Reading Mode.
Is it possible to set a time period after which the backlight will automatically power off if no key is pushed.
information with smaller numbers and text. Big will show less information with larger numbers and text.
Note: The screen size can also be selected from the measure mode (refer to Display description on page 9).

Set the sound options

The meter can make an audible sound when a key is pushed, when stability is reached or when the calibration reminder is due. The meter also makes an audible sound when it begins transferring data to a USB storage device and again when the data transfer is complete.
1. Push the key and select Sound.
2. Choose which events will produce an audible sound. Multiple items
can be selected.
Option Description
Key Press The meter will make an audible sound whenever a key is
Stability Alert The meter will make an audible sound whenever
Cal reminder The meter will make an audible sound when calibration is
pushed.
measurement stability is reached.
due.
Note: Refer to Set the probe calibration reminder to set the calibration reminder to on or off.
20 English

Set the probe calibration reminder

Make sure that a probe is connected to the meter.
1. Push the key and select the probe settings.
2. Select Modify Current Settings.
3. Select Calibration Options.
4. Select Calibration Reminder.
Option Description
Calibration reminder
Reminder repeat: Off, 2 h, 4 h, 8 h, 2 d, 5 d, 7d
Expires: Immediately, Reminder + 30 min, Reminder + 1 hr, Reminder + 2 hr, Continue Reading
The meter can be set to make an audible sound when calibration is due. Calibration expires after a specified time set by the user.
Note: The meter cannot be used to read samples after calibration has expired unless Continue Reading is selected.

Change the temperature units

To select degrees Celsius or Fahrenheit:
1. Push the key and select Temperature Units.
2. Select the Celsius or Fahrenheit option.

Set the measurement mode

One of three modes can be used to specify when measurements are taken and how the data is stored. When a data point is stored, the result is sent simultaneously to any device (PC/printer/ USB storage device) that is connected to the meter.
1. Push the key and select Measurement Mode.
2. Select Mode.
3. Select one of the measurement modes.
Option Description
Press to Read The sample is measured only when the GREEN/RIGHT key
Interval The sample is measured at regular intervals for a specified
Continuous The sample is measured continuously. Data is stored in the
under Read is pushed. Data is stored in the data log automatically when the stability criteria are met.
duration (refer to Set auto measurement intervals on page 21). Data is stored in the data log automatically.
data log only when the GREEN/RIGHT key under Store is pushed.

Set auto measurement intervals

When the measurement mode is set to Interval, the time intervals and duration must be specified. Measurements are stored at the user-defined intervals whether or not stability criteria are met.
Note: Use of an external USB storage device or direct printer connection while in Interval Measurement mode prevents data from being over-written in the data log. Data points are over-written on a First In/First Out basis. Refer to Prevent data log
overflow in interval mode on page
1. Push the
key and select Measurement Mode.
22.
2. Select Mode.
3. Select Interval as the Measurement Mode.
4. Select Duration and select the total time that measurements will be
taken for (15 min, 30 min, 1 h, 4 h, 8 h, 24 h, 48 h or no limit).
5. Select Interval and select how often measurements will be taken (every 10 s, 30 s, 1 min, 5 min, 15 min or 30 min).

Start interval measurements

During interval measurements, the meter goes into a standby state between readings to conserve power. The auto-shutoff option is disabled. Measurements stop when the selected interval duration has passed. The auto-shutoff option then becomes active.
English 21
Interval measurements are suspended for calibrations, check standard measurements or when the key is pushed. Interval measurements
resume when returning to the measurement screen.
1. From the Main Measurement screen, select Start to begin interval measurements. The screen will show “Recording” and the remaining time of the duration. The sample number automatically advances when each reading is taken.
2. To stop interval measurements, select Stop.
3. To repeat the interval measurement after it has been stopped or
completed, select Start.

Prevent data log overflow in interval mode

When measurements are taken at specified intervals (Table 4), each result is automatically stored. The meter can store up to 500 data records. When 500 records have been stored, data is replaced on a first­in, first-out basis. To prevent loss of data, connect the meter to a PC/printer/USB storage device.
Note: Stop interval measurements before changes are made to a method or to meter settings.
Table 4 Recommended interval/duration pairs
Interval Duration
10 seconds 1 hour
30 seconds 4 hours
1 minute 8 hours
5 minutes 24 hours
Note: When 2 probes are connected to the meter, use the next lowest recommended duration time. For example, for a 30-second interval, set the duration to 1 hour to prevent data log overload with 2 probes.

View instrument information

The instrument information menu shows specific information such as the serial number for the meter or IntelliCAL (R) probe(s).
1. Push the key and select Instrument Information.
2. Select (Probe model) Information or Meter Information.
Option Description
Probe information
Meter information
The Probe Information screen shows the probe model number, serial number, software version and date of first use. For LDO and LBOD probes, the lot code for the sensor cap and the remaining time before sensor cap replacement is shown.
Note: A probe must be connected to the meter.
The Meter Information screen shows the meter model number, serial number, software version and memory information. The amount of memory used and the number of available user method settings, operator IDs and sample IDs is shown.

Update the meter software

A USB storage device that contains software update files is used to update the meter software.
Note: The meter must be turned off and then on again before the software update will begin. The software update initiates upon meter startup after the USB device is correctly inserted.
N O T I C E
Do not remove the USB device until the “Update complete” message is shown. The meter can become damaged if the USB device is removed before the update process is complete.
1. Save stored data from the data log to a USB storage device or to a PC. Refer to Send data to a USB storage device on page 15
Send data directly to a computer on page 18.
2. Turn off the meter.
3. Connect the USB/DC adapter, AC-DC power supply and cord
(Figure 4 on page 9).
and
22 English
4. Insert the USB storage device that contains the software update files
into the USB/DC adapter.
5. Turn on the meter. The update process starts. The display will show “Updating meter to <firmware version>”. After an interval, the display changes to “Updating files, please wait...” In addition, the display will show a rotating flask and emit a periodic audio signal during the update process.
Note: A large capacity USB storage device increases the time required for completion of the update process, even if most of the device memory is empty.
6. Wait for the meter to finish the software update. When the update process is complete, the message “Update complete. Remove USB device” is shown. The meter will turn off after the USB device has been removed.
7. Repeat steps 1 through 6 to update the software in other HQd meters as necessary.

Download software updates

To download the most current version of the software, refer to the applicable product page on the manufacturer's website.
1. Transfer the update files to a USB storage device.
2. Follow the instructions in Update the meter software on page
update the software in the meter.
22 to

Transfer method settings

Probe settings that have been changed by the user for measurements, calibrations or check standards (Meter Options > (Probe Model) Settings > Modify Current Settings) can be copied to a USB storage device. The USB device can then be used to transfer the method settings to other HQd meters that accept the same probes.
N O T I C E
Make sure the USB storage device does not contain HQd meter software update files to prevent unintentional updates.
1. Turn off the meter.
2. Connect the meter to AC power (Figure 4 on page
9).
3. Plug the USB storage device into the USB/DC adapter before the meter is powered on.
4. Turn on the meter.
5. Push the
already contains a method settings file, an option to export or import methods is shown. Select Export Methods.
6. In the Select Methods to Export screen, select one or more methods to copy to the USB device. A check mark is shown next to each selected method.
7. Select OK. The settings are copied to the USB storage device. When complete, the Transfer Summary screen is shown.
8. Connect the AC power and USB device to a meter that will receive the method settings. Turn the meter on.
9. Push the key and select Transfer Methods. If the USB device already contains a method settings file, an option to export or import methods is shown. Select Import Methods.
10. In the Select Methods to Import screen, select one or more methods to transfer to the meter. A check mark is shown next to each selected method.
11. Select OK. The user method settings are transferred from the USB storage device to the meter. When complete, the Transfer Summary screen is shown. Select details to view additional information about the transfer.
12. Disconnect the USB storage device from the meter.
key and select Transfer Methods. If the USB device

Bi-directional Communication between the meter and a PC

For measurement automation the meter can be used to implement a command set for meter remote control or automated data transfer. The command set can be used to perform minimal configuration and to control the meter. To set up the meter for communication and control, refer to About meter configuration on page 24. For additional information and the command set contact Technical Support.

About meter control

The virtual serial connection can be used to control meter functions from a PC. For example, the functions include starting a measurement cycle,
English 23
turning off the meter and sending the entire measurement (including calibrations) to the PC or other information management system.

About meter configuration

To use the meter communication and control from the PC, an INF file must be installed.
1. The meter software must be version 2.0.0.710 or higher. To download the most current version of the software, refer to the applicable product page on the manufacturer's website.
2. Open the Zip file.
3. Copy the INF file from the software upgrade package to a convenient
location on the PC.
Note: The INF file must be installed to use the meter manual control from a PC.
4. Turn on the meter.
5. Push the key and select Instrument Information.
6. Select USB Device Type and then select Virtual Serial to use the
virtual serial port on the meter.
7. Push OK. The meter will automatically restart to complete the setting change.
8. Connect the meter with the USB cable to the PC and turn on the meter. Windows XP starts the "Found New Hardware Wizard".
9. Select "No, not at this time" to the query "Can Windows connect to Windows Update to search for software?"
10. Click Next. The next wizard screen will prompt.
11. Select "Install from a list or specific location (Advanced)" to the query
"What do you want the wizard to do?"
12. Click Next. The next wizard screen will prompt.
13. Select the option "Search for the best driver in these locations."
14. Uncheck the "Search removable media (floppy, CD-ROM)" option
and select the "Include this location in the search:" and click the "Browse" button.
15. Select the INF folder or location and click OK.
16. Click Next. The new software will be installed.
17. Click Finish to complete the Found New Hardware Wizard for: HQd
Meter - Virtual Serial Port.
18. To make sure that the installation succeeded, go to Computer Management>Device Manager>Ports. The new installed port is listed as HQd Meter - Virtual Serial Port (COM#).
19. The meter is now ready for communication with PC-based systems using the Virtual Serial port. A program interface must be developed by the user for the command set used to control the meter functions from the PC. Contact Technical Support for more information and command set documentation.

Maintenance

C A U T I O N
Multiple hazards. Only qualified personnel must conduct the tasks described in this section of the document.

Clean the meter

The meter is designed to be maintenance-free and does not require regular cleaning for normal operation. Exterior surfaces of the meter may be cleaned as necessary.
1. Wipe the surface of the meter with a damp cloth.
2. Use a cotton-tipped applicator to clean or dry the connectors.

Replace the batteries

W A R N I N G
Explosion hazard. Incorrect battery installation can cause the release of explosive gases. Be sure that the batteries are of the same approved chemical type and are inserted in the correct orientation. Do not mix new and used batteries.
For battery replacement, refer to Figure 11. Make sure that the cover is tightly closed to maintain the IP67 enclosure rating.
24 English
1. Pull the release tab on the battery cover and the remove the cover.
2. Remove the batteries.
3. Install 4 AA alkaline or 4 AA nickel metal hydride (NiMH) batteries.
Make sure that the batteries are installed in the correct polarity.
4. Replace the battery cover.
Figure 11 Battery replacement
1 Batteries 2 Release tab 3 Battery cover
Error/Warning Description Solution
Connect a Probe Probe disconnected
Probe Not Supported Probe disconnected
Bootloader X.X.XX.XX error
or connected improperly
Software not updated to most current version
Problem with probe Connect a different IntelliCAL
or connected improperly
Software not updated to most current version
Problem with probe Connect a different IntelliCAL
HQd meter does not support IntelliCAL probe
Software not updated to most current version.
Tighten the locking nut on the probe connector.
Disconnect the probe and then connect the probe again
To download the most current version of the software, refer to the applicable product page on the manufacturer's website.
probe to verify if problem is with probe or meter
Tighten the locking nut on the probe connector.
Disconnect the probe and then connect the probe again.
To download the most current version of the software, refer to the applicable product page on the manufacturer's website.
probe to the meter to verify if problem is with the meter or the probe.
Contact Technical Support.
To download the most current version of the software, refer to the applicable product page on the manufacturer's website.

Troubleshooting

Refer to the following table for common problem messages or symptoms, possible causes and corrective actions.
English 25
Error/Warning Description Solution
0 days remaining message (For LDO and LBOD only)
Meter not configured Software error(s) If the meter starts up correctly,
LDO or LBOD sensor cap used for 365 days
There are 0 days remaining in the life of the LDO sensor cap.
Meter set to incorrect date and time
Software not updated to most current version
Replace the LDO or LBOD sensor cap and iButton®.
Replace the LDO sensor cap. Calibration will be allowed. However, the calibration icon and question mark will appear on the measurement screen even if the calibration has passed.
1. Disconnect the probe from the meter.
2. Remove the meter batteries.
3. Install the meter batteries
properly. Follow the polarity makings.
4. Set correct date and time in the meter.
5. Connect the probe and verify that message has been removed.
To download the most current version of the software, refer to the applicable product page on the manufacturer's website.
back up the Data Log and Method files. To download the most current version of the software, refer to the applicable product page on the manufacturer's website.
Error/Warning Description Solution
Meter will not power on or powers on intermittently
Unable to access Full Access Options screen
Unable to access Full or Operator Access Options screen
Batteries are not installed correctly
Software not updated to most current version
Damaged meter Contact Technical Support.
Correct password has not been entered
Software not updated to most current version
Examine battery orientation to make sure the batteries follow the polarity markings. Test again.
Clean the battery terminals, then install new batteries.
Connect AC power adapter and test again.
To download the most current version of the software, refer to the applicable product page on the manufacturer's website.
Contact Technical Support.
To download the most current version of the software, refer to the applicable product page on the manufacturer's website.

Replacement parts and accessories

Note: Product and Article numbers may vary for some selling regions. Contact the appropriate distributor or refer to the company website for contact information.
Replacement parts
Description Item no.
AC Power/USB Adapter Kit, 115 VAC 5826300
AC Power/USB Adapter Kit, 230 VAC 5834100
Batteries, Alkaline AA 1938004
Battery cover 9245500
Battery Contact, dual fixed 5188400
26 English
Replacement parts (continued)
Description Item no.
Battery Contact, dual spring 5188800
Cable, USB 6 ft (1.8 m), Type A male, Type B male 5924000
Field Kit (includes Protective Glove Kit for meter and five 120-mL sample cups)
Field Case for 2 probes with up to 5 m cables (10 m total). Includes empty case, insert for meter and probe storage, 4 containers for sample collection.
Field Case for 3 probes with up to 5 m cables (15 m total). Includes empty case, insert for meter and probe storage, 4 containers for sample collection.
Field Case for 2 probes with greater than 5 m cables (30 m total). Includes empty case, insert for meter with protective glove.
Keyboard (QWERTY), USB type LZV582
Printer, Citizen PD-24 USB printer, 120–220 VAC
Printer Paper for Citizen PD-24, thermal, 5/pk 5836000
Probe Clips, color coded (5 colors, 2 clips of each color), 10/pk
Probe Depth Marker (rugged cables) 5828610
Probe Holder, standard (fits on protective glove) 5829400
Protective Glove Kit for meter 5828700
Universal Probe Stand for standard IntelliCAL Probes 8508850
Meter Stand for Portable HQd meters 4754900
5825800
8505500
8505501
8505600
US: 2960100
EU: 5835900
5818400
English 27

Examples of printed reports

Printed reports contain a report header and all stored data for samples, check standards and calibrations.
Report header
The first line of a report shows the report header (Figure 12).
Figure 12 Report header
1 Meter serial number
2 Report label 4 File type extension
1
The probe serial number is shown on calibration history and current calibration reports.
1
3 Date and time, 24 h (YYMMDDhhmm)
Data reports
Sample data can be printed in a basic, advanced or total report format.
• Total report—refer to Figure 13.
• Advanced report—refer to Figure 14.
• Basic report—refer to Figure 15.
28 English
Figure 13 Basic report for sample data—1 line
1 Data type (RD=reading) 6 Units
2 Parameter (pH, LDO, etc.) 7 Sample ID: user-defined; shows “SAMPLE ID” if undefined
3 Time (hh:mm:ss in 24 h or user-defined format) 8 Sample ID counter
4 Date (DD-MM-YY or user-defined format) 9 Operator ID: user-defined; shows “- - -” if undefined
5 Measured value
Figure 14 Advanced report for sample data—2 lines
1 Probe model 4 Additional units: shows all additional units associated with the reading.
2 Error message (if applicable) 5 Probe settings: shows the highest-priority setting associated with the reading
3 Probe serial number (a “<“ or “>” on the HQ40d meter indicates the probe
position)
English 29
Figure 15 Total report for sample data—3 lines
1 Method name for probe settings 4 Calibration slope/ratio/constant
2 Time of calibration, prefaced by “CAL” and displayed as hh:mm in 24 h (or user-
defined) format
3 Date of calibration (DD-MM-YY or user-defined format) 6 Operator ID: user-defined; shows “- - -” if undefined
5 Offset—contents vary depending on type of parameter and user settings. May
be blank.
Check standard reports
Check standard data is printed with 1 line of information (Figure 16)
Figure 16 Check standard report
1 Report type (ST = check standard) 5 Measured value
2 Report type (ST = check standard) 6 Units
3 Time (hh:mm:ss in 24 h or user-defined format) 7 Check standard status: Pass/Fail based on the acceptance criteria
4 Date (DD-MM-YY or user-defined format) 8 Operator ID: user-defined; shows “- - -” if undefined
30 English
Calibration reports
Calibration data is printed when the data log is sent to the printer or when probe data is sent to the printer. Calibration data is printed with 2 lines of information (Figure 17).
Figure 17 Calibration report
1 Report type (CL = calibration, IC = current calibration) 7 Probe serial number (a “<“ or “>” on the HQ440d meter indicates the probe
2 Parameter (pH, LDO, etc.) 8 Calibration slope/ratio/constant
3 Method name for probe settings 9 Offset—contents vary depending on type of parameter and user settings. May
4 Time of calibration, prefaced by “CAL” and displayed as hh:mm in 24 h (or user-
defined) format
5 Probe model 11 Operator ID: user-defined; shows “- - -” if undefined
6 Date of calibration (DD-MM-YY or user-defined format)
position)
be blank.
10 r²: contents vary depending on type of parameter being read, configuration of
specific method and number of calibration standards used. May be blank.
The calibration history can be printed from the probe data menu. Refer to Figure 18.
English 31
Figure 18 Calibration history report
1 Report type (CH=calibration history) 5 Date of calibration (DD-MM-YY or user-defined format)
2 Parameter (pH, LDO, etc.) 6 Probe serial number (a “<“ or “>” on the HQ40d meter indicates the probe
3 Time of calibration, prefaced by “CAL” and displayed as hh:mm in 24 h (or user-
defined) format
4 Probe model 8 Offset—contents vary depending on type of parameter and user settings. May
position).
7 Calibration slope/ratio/constant
be blank.
32 English
A
AC power connection
auto-shutoff ............................................................................................20
...............................................................................8
B
backlight .................................................................................................20
Bi-directional Communication ................................................................23
big screen mode ..............................................................................10, 20
C
calibration ...............................................................................................12
dual-screen mode .............................................................................10
probe reminder
reminder ............................................................................................20
check standards .....................................................................................13
contrast, display .....................................................................................20
..................................................................................21
D
data file descriptions ..............................................................................16
data log ..................................................................................................23
prevent data loss ...............................................................................22
printing ..............................................................................................14
transfer data to PC ......................................................................18, 24
data log, stored data ..............................................................................14
date and time
changing ............................................................................................ 12
detailed screen mode .............................................................................20
display description .................................................................................20
dual-screen mode
calibration ..........................................................................................10
I
interval measurements
...........................................................................21

Index

M
measurement modes
menu navigation .....................................................................................11
modes
screen ...............................................................................................20
O
operator ID .............................................................................................13
P
power .......................................................................................................8
printing ...................................................................................................14
probes
connecting .........................................................................................12
stored data
S
sample ID
screen modes ..........................................................................................9
security options ......................................................................................18
software
sounds, turn on or off .............................................................................20
stored data .......................................................................................13, 14
...............................................................................................13
restrict menu access .........................................................................19
install PC Application software ..........................................................18
update meter software
file descriptions .................................................................................16
probe .................................................................................................14
.............................................................................21
........................................................................................14
......................................................................22
33
Index T
transfer data
to PC ...........................................................................................16, 18
to USB storage device
transfer data to USB device
data log .............................................................................................23
troubleshooting ......................................................................................25
......................................................................15
U
USB storage device
file descriptions .................................................................................16
transfer data to ..................................................................................15
34
HACH COMPANY World Headquarters
P.O. Box 389, Loveland, CO 80539-0389 U.S.A. Tel. (970) 669-3050 (800) 227-4224 (U.S.A. only) Fax (970) 669-2932 orders@hach.com www.hach.com
©
Hach Company/Hach Lange GmbH, 2010, 2013. All rights reserved.
HACH LANGE GMBH
Willstätterstraße 11 D-40549 Düsseldorf, Germany Tel. +49 (0) 2 11 52 88-320 Fax +49 (0) 2 11 52 88-210 info@hach-lange.de www.hach-lange.de
HACH LANGE Sàrl
6, route de Compois 1222 Vésenaz SWITZERLAND Tel. +41 22 594 6400 Fax +41 22 594 6499
Loading...