Hach-Lange H170 User Manual [en, de, fr, cs, pl]

DOC022.98.80058
H160, H170
10/2013, Edition 1
Bedienungsanleitung
Uživatelská příručka
Ръководство на потребителя
User Manual
Manuel d'utilisation
Brugsanvisning
English..............................................................................................................................3
Deutsch..........................................................................................................................21
Français.........................................................................................................................40
Čeština...........................................................................................................................59
Dansk..............................................................................................................................77
Polski..............................................................................................................................95
български...................................................................................................................113
2

Table of contents

Specifications on page 3 Startup on page 10
General information on page 4 Operation on page 11
Installation on page 7 Maintenance on page 14
User interface and navigation on page 8 Troubleshooting on page 17

Specifications

Specifications are subject to change without notice.
Specification Details
Dimensions (W x D x H) 9 x 20 x 5 cm (3.5 x 8 x 2 in.)
Enclosure Water resistant
Weight 1300 g (3.0 lb.)
Power requirements 4-ANSI 15 A or IEC-LR6 (AA Alkaline)
Power consumptions 1 W with backlight on
Battery life Up to 200 hours continuous operation, 10-hour advance of low battery warning
Operating temperature 5 to 40 °C (41 to 104 °F)
Storage temperature –20 to +40 °C (4 to 140 °F)
Operating humidity Relative humidity: 50% maximum at 25 ºC (77 ºF), non-condensing
Input connectors 8-pin ISFET, BNC with phono jack temperature, 12-pin conductivity (H170 only)
Languages English
Temperature correction Off, automatic and manual (parameter dependent)
Measurement Continuous measurement
Protection rating IP67
Certifications CE
pH
Calibration Up to five points: 1.68, 4.01, 6.86, 7.00, 9.18, 10.01, 12.45
Accuracy ±0.01 pH
Resolution 0.01 pH
Range 0.00 to 14.00
mV
Calibration None
Accuracy 0.1 mV
Resolution Autoranging, 0.1 and 1
Range Autoranging, ±199.9 mV to ±1999 mV
Temperature
Calibration None
Accuracy ±0.5 °C
English 3
Specification Details
Resolution 0.1 °C (0.1 °F)
Range –0 to 100 °C (32 to 212°F)
Conductivity
Calibration Up to five points
Accuracy ±1% ppt (–2 to +35 °C)
Resolution 0.01 ìS, 0.1 ìS ,1 ìS, 0.01 mS, 0.1 mS
Range Autoranging: 0.00 to 19.99 ìS, 20.0 to 199.9 ìS, 200 to 1999 ìS, 2.00 to 19.99 mS,
TDS
Calibration Up to five points
Accuracy ±1% full scale or ±1 digit
Resolution 0.01 ppm, 0.1 ppm, 1 ppm, 0.01 ppt, 0.1 ppt, 1 ptt, 0.1 mg/L, 1mg/L ,0.01 gal/L,
Range Autoranging: 0.00 to 9.99 ppm, 10.0 to 99.9 pm, 100 to 999 ppm,1.00 to 9.99 ppt,
Salinity
Calibration None (derived from conductivity)
Accuracy ±0.1 ppt, –2 to +35 °C (28.4 to 95 °F )
Resolution 0.1 ppt 1%
Range 0 to 42 ppt
20.0 to 199.9 mS
0.1 gal/L
10.0 to 99.9 ppt, 100 to 200 ppt mg/L: 0.00 to 199.9 mg/L, 200 to 1999 mg/L, 2.00 to 19.99 gal/L, 20.0 to 50 gal/L

General information

In no event will the manufacturer be liable for direct, indirect, special, incidental or consequential damages resulting from any defect or omission in this manual. The manufacturer reserves the right to make changes in this manual and the products it describes at any time, without notice or obligation. Revised editions are found on the manufacturer’s website.

Safety information

N O T I C E
The manufacturer is not responsible for any damages due to misapplication or misuse of this product including, without limitation, direct, incidental and consequential damages, and disclaims such damages to the full extent permitted under applicable law. The user is solely responsible to identify critical application risks and install appropriate mechanisms to protect processes during a possible equipment malfunction.
Please read this entire manual before unpacking, setting up or operating this equipment. Pay attention to all danger and caution statements. Failure to do so could result in serious injury to the operator or damage to the equipment.
Make sure that the protection provided by this equipment is not impaired. Do not use or install this equipment in any manner other than that specified in this manual.
Use of hazard information
Indicates a potentially or imminently hazardous situation which, if not avoided, will result in death or serious injury.
4 English
D A N G E R
W A R N I N G
Indicates a potentially or imminently hazardous situation which, if not avoided, could result in death or serious injury.
Indicates a potentially hazardous situation that may result in minor or moderate injury.
Indicates a situation which, if not avoided, may cause damage to the instrument. Information that requires special emphasis.
C A U T I O N
N O T I C E
Precautionary labels
Read all labels and tags attached to the instrument. Personal injury or damage to the instrument could occur if not observed. A symbol on the instrument is referenced in the manual with a precautionary statement.
This is the safety alert symbol. Obey all safety messages that follow this symbol to avoid potential injury. If on the instrument, refer to the instruction manual for operation or safety information.
This symbol indicates that a risk of electrical shock and/or electrocution exists.
Electrical equipment marked with this symbol may not be disposed of in European public disposal systems after 12 August of 2005. In conformity with European local and national regulations (EU Directive 2002/96/EC), European electrical equipment users must now return old or end-of-life equipment to the Producer for disposal at no charge to the user.
Note: For return for recycling, please contact the equipment producer or supplier for instructions on how to return end­of-life equipment, producer-supplied electrical accessories, and all auxiliary items for proper disposal.
Certification
Canadian Radio Interference-Causing Equipment Regulation, IECS-003, Class A:
Supporting test records reside with the manufacturer. This Class A digital apparatus meets all requirements of the Canadian Interference-Causing
Equipment Regulations. Cet appareil numérique de classe A répond à toutes les exigences de la réglementation canadienne
sur les équipements provoquant des interférences.
FCC Part 15, Class "A" Limits
Supporting test records reside with the manufacturer. The device complies with Part 15 of the FCC Rules. Operation is subject to the following conditions:
1. The equipment may not cause harmful interference.
2. The equipment must accept any interference received, including interference that may cause
undesired operation.
Changes or modifications to this equipment not expressly approved by the party responsible for compliance could void the user's authority to operate the equipment. This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to Part 15 of the FCC rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference, in which case the user will be required to correct the interference at their expense. The following techniques can be used to reduce interference problems:
English
5
1. Disconnect the equipment from its power source to verify that it is or is not the source of the
interference.
2. If the equipment is connected to the same outlet as the device experiencing interference, connect
the equipment to a different outlet.
3. Move the equipment away from the device receiving the interference.
4. Reposition the receiving antenna for the device receiving the interference.
5. Try combinations of the above.

Product overview

N O T I C E
Always disconnect power to the meter before an electrode is changed. Only use the meter as instructed in this manual or instrument damage can occur.
The meter is available in two models:
• H160—pH and ORP
• H170—pH, ORP, conductivity, TDS and salinity
The H-series waterproof handheld meters use glass sensor electrodes with BNC connectors and non-glass probes with ISFET (ion sensitive field effect transistor) silicon chip sensors. The meters use one pH electrode (a BNC pH electrode or an ISFET pH probe) at a time. When a meter is set to on, the meter automatically identifies the type of electrode that is attached.
Figure 1 Handheld meter
1 Rubber dust caps 5 ISFET pH probe connector (8-pin) 2 Conductivity probe connection (12-pin, H170 only) 6 LCD display 3 BNC probe connector 7 Power button 4 3.5 mm phono jack for glass pH electrode, ISE,
ORP or DO temperature sensors
8 Battery cover

Product components

Make sure that all components have been received. Refer to Figure 2. If any items are missing or damaged, contact the manufacturer or a sales representative immediately.
6
English
Figure 2 Handheld meter components
1 Waterproof handheld meter (H160 shown) 2 AA Alkaline batteries (4x)

Installation

C A U T I O N
Multiple hazards. Only qualified personnel must conduct the tasks described in this section of the document.

Electrical installation

Install the batteries
C A U T I O N
Explosion hazard. Incorrect battery installation can cause the release of explosive gases. Be sure that the batteries are of the same approved chemical type and are inserted in the correct orientation. Do not mix new and used batteries.
N O T I C E
Do not tighten the screws too much or instrument damage can occur.
Do not complete this operation with probes attached to the meter. Remove all probes from the meter or instrument damage can occur.
The meter uses AA alkaline or rechargeable NiMH batteries. To conserve the battery life, refer to
Configure the meter on page 10 to configure automatic shutdown of the meter.
Refer to Figure 3 to install the batteries.
Items to collect:
• Phillips screwdriver
• AA Alkaline batteries (4x)
N O T I C E
English
7
Figure 3 Battery installation
Probe and connection types
N O T I C E
Make sure that all the rubber plugs are fully installed before the handheld meter is put in water or the interior of the 3.5 mm phono jack will fill with water. However, the interior of the handheld meter will stay sealed. If water gets into the 3.5 mm phono jack, immediately invert the meter to let water come out of the phone jack, then let the connector dry.
Always set the power to off before a probe connection is changed or damage to the instrument can occur.
The meter can connect to two pH-sensing technologies: ISFET pH probes and glass electrodes. When the meter is set to on, it automatically recognizes the type of probe attached. The meter only uses one type of pH probe at a time.
Note: If a probe is not attached to the meter, the selected mode is inactive.
N O T I C E

User interface and navigation

Keypad description

Figure 4 and Table 1 show the function of the keys for each meter.
8
English
Figure 4 Keypad descriptions
Table 1 Keypad functions
Key Action H160 H170
ON/OFF: Set the power to on or off. x x
pH mode x x
mV mode x x
Set the backlight to on. After 2 minutes without a keystroke, the backlight is set to off.
TDS mode x
Conductivity mode x
Salinity mode x
Hold mode: Operates only in measurement modes. All readings freeze when hold mode is set to on.
Calibration mode x x
Scroll through values, setup screens and options. x x
Setup mode x x
Select an option, setting or value. x x
x x
x x
English 9

Display description

Figure 5 shows measurement modes and values, data connection and storage options, battery
status, temperature values, stabilization lock and connection statuses shown on the display.
Figure 5 Display
1 Setup mode 7 Measurement mode 2 Temperature and data values 8 Measurement units 3 Measured value 9 Automatic Temperature Compensation (ATC) value
4 Stabilization lock 10 Data storage memory location 5 Hold indicator 11 Temperature unit 6 ISFET probe indicator 12 Low battery indicator
(pH, conductivity, TDS or salinity)

Startup

Set the power to on

Push the power button to set the power to on or off. Make sure that the power supply (AC power or battery power) is correctly installed.

Configure the meter

Complete the operations in order:
1. Set the power to on.
2. Push SETUP to put the meter in setup mode.
3. Use the arrow keys to select an option, then push ENTER:
10
English
Option Function Description
READY Stabilization lock Sets the stabilization lock to on or off.
• The "READY" icon shows when an endpoint occurs and the stabilization lock is set to on.
• After stabilization, the display locks the value and the meter ignores slight changes in the measurement of pH, conductivity or TDS.
• Sets the stabilization lock to off during titrations or slight change detection. The display automatically unlocks after a significant measurement change is found.
ºC ºF Temperature
Snd Sound options Sets the sound alerts to on or off. Three sounds tell the user about different
oFF Automatic
CLr ALL CAL
1
This does not erase calibration data for an ISFET probe.
display
shutdown
Clear calibration points
Sets the temperature unit to ºC or ºF.
functions:
One beep: Stabilization occurred with the stabilization lock on.
Two beeps: An error occurred. The error code shows on the display. Refer to Error codes on page 18.
Three beeps: Measurement stability in calibration mode, regardless of the stabilization lock setting.
Sets shutdown parameters from 1 minute to 000 minutes (continuous power). The meter beeps 1 minute before shutdown. Make sure to power cycle for the automatic shutdown to occur. Automatic shutdown disables during: data logging, Bluetooth™ transfers and when connected to the USB port.
Erases all calibration points in storage. Calibrate the meter again.
1
4. Push ENTER to keep the changes and go back to setup mode.
Note: To exit and not keep the changes, push SETUP.

Operation

Override an error

The user can override some errors (e.g., poor electrode condition) until a replacement electrode is ordered and received.
Note: It is not possible to override all errors. The manufacturer does not recommend that the user override error codes. Contact technical support to override error codes.
To cancel or override an error:
1. Find the cause of the error. Refer to Error codes on page 18. If possible, push SETUP or pH, COND, TDS or SAL to cancel the calibration or other error.
2. To override the error, do not complete step 1. Instead, push ENTER.
3. Use the arrow keys until "YES" shows.
4. Push ENTER.

Set the pH operations

Set the calibration alarm, the pH resolution, the calibration points and the calibration methods. Complete the operations in order.
1. Set the power to on.
2. Push SETUP, then pH.
3. Use the arrow keys to select an option, then push ENTER:
English
11
Operation Function
pH resolution Sets the pH resolution to 0.1 or 0.01 pH. The stabilization lock (the "READY" icon) is
Set the 2 or 3-point pH calibration
pH calibration method
faster at a 0.1 pH resolution.
Shows the current number of calibration points. Use the arrow keys to set the calibration points.
• Sets the pH buffer. The default is a 2-point calibration (pH 7.00 then pH 4.00).
• Push ENTER again to set the first calibration buffer. The display shows the next
buffer. Do this procedure until all buffers are set. Previously-selected buffers do not show.
4. Push ENTER to keep the changes and go back to setup mode.
Note: To exit and not keep the changes, push SETUP.
Set the manual temperature compensation
Glass pH electrodes are automatically temperature compensated if a temperature probe is sensed in the 3.5 mm phono jack. If no probe is sensed, the meter defaults to the manual temperature compensation at 25.0 ºC. Manual temperature compensation is always in ºC. Refer to Set the
manual temperature compensation for glass electrodes.
pH buffer values at various temperatures
25 ºC 0 ºC 5 ºC 10 ºC 20 ºC 30 ºC 40 ºC 50 ºC
1.68 1.67 1.67 1.67 1.67 1.68 1.69 1.71
4.01 4.00 4.00 4.00 4.00 4.01 4.03 4.06
6.86 6.98 6.95 6.92 6.87 6.85 6.84 6.83
7.00 7.12 7.09 7.06 7.01 6.99 6.97 6.97
9.18 9.46 9.40 9.33 9.23 9.14 9.07 9.02
10.01 10.32 10.25 10.18 10.06 9.97 9.98 9.83
12.45 13.42 13.21 13.00 12.63 12.29 12.04 11.70
1. Push SETUP, then pH to put into pH setup mode.
2. Use the arrow keys until the manual temperature shows "- - -.- ºC" on the display.
3. Push ENTER to see the manual temperature value.
4. Use the arrow keys to set the manual temperature compensation.
5. Push ENTER to keep the changes and go back to setup mode.
Note: To exit and not keep the changes, push SETUP.
Calibrate for pH measurement
The meter accepts 1,-2-or 3-point calibrations. The default is a 2-point calibration.
1. Set the power to on.
2. Connect the probe to the meter.
3. Push pH to put the meter in pH mode.
4. Put the probe in the buffer solution.
Note: To erase all previous calibration points, slopes and offsets, push CAL, then SELECT. To update a previous calibration with a single-point calibration, ignore this step.
5. Push CAL, then the "CAL" icon flashes on the display.
Note: Stop a calibration when the pH value is in flash mode. Push pH or SETUP to go back to the measurement mode.
6. If the buffer is correct, push ENTER, then the pH buffer flashes until the calibration is complete.
12
English
7. If the pH buffer is not correct, use the arrow keys to select another buffer.
8. Use the arrow keys to see the buffers: 1.68, 4.01, 6.86, 7.00, 9.18, 10.01 and 12.45. During "CAL
2" or "CAL 3," the meter ignores the values used in previous calibrations.
9. The meter beeps three times when a calibration is complete. "CAL 3" and the next pH buffer value (default = 4.01) flashes on the display.
10. Rinse the probe in deionized water. Put the probe in the second buffer solution.
11. If OK, push ENTER, or use the arrow keys to change the value. Push pH to accept a 1-point
calibration and to go back to the pH measurement mode. If ENTER is selected, the buffer value flashes until calibration is complete.
12. If the meter is set for a 2-point calibration, the meter beeps three times when the calibration is complete. The display shows the slope, the mV offset at pH 7.00 and the number of calibration points. Review the calibration data, then push pH or STORE to start reading the pH value of another solution.
13. If the meter is setup for a 3-point calibration, do steps 1 through 12 until all the calibration points are set.

Set the conductivity and TDS options

Set the options for conductivity and TDS: conductivity calibration points, conductivity temperature compensation, conductivity temperature, cell constant, TDS conversion factor, TDS units and salinity units in ppm or as a percentage. Complete the operations in order:
1. Set the power to on.
2. Push SETUP, then COND or TDS to put the meter in conductivity or TDS setup modes.
3. Use the arrow keys to select an option, then push ENTER:
Operation Function
Conductivity calibration point Keeps up to five calibration points.
• Use the arrow keys to select each option, then push
ENTER.
• The first active calibration point shows. Use the arrow keys to see other in-use values.
Note: "- - - -" shows on the display if there are no stored conductivity calibration points.
Conductivity temperature
1
• Shows the conductivity normalization value.
• Use the display settings between 20 to 25 °C (default = 25 °C).
Conductivity probe cell constant The cell constant is automatically found and shown on
TDS conversion factor
the display. Sets the cell constants to:
• K = 0.5 for general procedures
• K = 1 for low conductivity procedures (e.g., drinking water)
• K = 10 for salinity or brackish water
• If no conductivity probe is attached, the display shows "000."
• TDS values are related to conductivity. Calibrate with a TDS standard or with conductivity standards, then set the meter with a conversion factor.
• Sets the conversion factor to 0.40 to 1.00 (default =
0.50).
English 13
Operation Function
TDS units Use the arrow keys to set the TDS units to mg/L (gal/L)
Salinity units in ppm (ppt) or % Use the arrow keys to set the salinity units to ppm (ppt)
1
Conductivity varies a lot with temperature. The conductivity automatic temperature compensation (ATC) adjusts conductivity measurements to calculate the conductivity changes in the reading that are caused by temperature. The readings are referenced to or "normalized" at a standard temperature. The ATC gives the equivalent conductivity or TDS of a solution normalized at 20 or 25 °C.
or ppm (ppt).
or as a percentage.
4. Push ENTER to keep the changes and go back to setup mode.
Note: To exit and not keep the changes, push SETUP.
Calibrate for conductivity and TDS measurement
Calibrate the meter with up to five calibration points with conductivity or TDS standards. If the calibration point is 20% larger than a calibration point in storage, both conductivity points are kept. If the calibration point is 20% smaller than a calibration point in storage, only the new calibration point is kept. If there are five calibration points in storage, the new calibration data replaces values near the same range.
By default, the meter uses the temperature coefficient of 2% per °C. The user can set the coefficient in setup mode. Review calibration points in setup mode.
Note: Salinity measurements are identified from conductivity.
1. Set the power to on.
2. Push COND or TDS.
3. Put the probe in the first standard solution. The meter measures the standard.
4. Use the arrow keys to select the value of the conductivity or TDS standard.
5. Push ENTER. The main display shows, then the "WAit" icon flashes.
Note: Push TDS or COND to stop calibration. The meter stops the current calibration and keeps the previous calibration data.
6. Push CAL. The main value flashes.
7. To erase all previous calibration points, slopes and offset values, push CAL, then DELETE.
Ignore this step to update a previous calibration with a single point calibration.
8. Push COND or TDS to read the conductivity or TDS value of any solution.
9. Do steps 2 through 9 again to complete multiple calibration points (maximum of 5).
10. When the calibration is complete, the meter shows the slope as compared to the nominal
sensitivity for a probe of that cell constant and the total number of conductivity and TDS calibration points in the meter.
Note: The number at the bottom right corner of the display is the calibration points that are in the meter. The value above the slope is the slope relative to the nominal sensitivity.

Maintenance

Replace the batteries

The screen reads "bAt" when the batteries are too low to give a reliable measurement. The low battery icon shows when there is approximately 25 hours of battery power. Measurement errors are possible when the batteries are low. Refer to Install the batteries on page 7 to replace the batteries.

Clean the instrument

Clean the exterior of the instrument with a moist cloth and a mild soap solution and then wipe the instrument dry.
14
English

ISFET pH probe maintenance

C A U T I O N
Multiple hazards. Only qualified personnel must conduct the tasks described in this section of the document.
C A U T I O N
Chemical exposure hazard. Obey laboratory safety procedures and wear all of the personal protective equipment appropriate to the chemicals that are handled. Refer to the current safety data sheets (MSDS/SDS) for safety protocols.
The expected life of an ISFET probe is approximately 18 months. The reference electrode has a KCI gel that is diluted over time. The reference is sealed and is non-refillable. Replace the probe when it becomes difficult to calibrate.
• Keep the probe dry with the protective shroud on when not in use.
• Clean oil, fat, food particles, starch, protein or other materials from the probe tip after use.
• Never use sharp metal objects (e.g., a needle, a pin, etc.) to clean the sensor surface.
• Remove the probe from environments with static electricity. Electrostatic discharge (ESD) can permanently damage the probe.
• Remove the probe from environments that will damage the sensor, such as hydrofluoric acid or abrasive samples.
• Remove the probe from environments that will damage the epoxy materials used in the probe tip (e.g., keep the probe away from acetone, toluene, methylene chloride, xylene and other strong organic solvents).
• Do not use the probe in temperatures more than 60 ºC (140 ºF). Thermal cycling can decrease the life of the probe.
• For semi-solids use, carefully twist the probe to make sufficient contact with the sample to the sensor.
• When semi-solids are tested, make sure that solid objects (i.e., bone or gristle) do not scratch the sensor.
• Cool samples to room temperature to maximize probe life.
• Always use new buffers and new rinse solutions.
Prepare the probe for storage
N O T I C E
Do not use the probe for long-term pH measurement applications.
Note: No electrode storage solution is necessary.
A new probe usually has visible reference gel at the tip of the probe, as well as in the interior of the rubber dust cap. To remove the gel, carefully clean with a soft-bristled toothbrush and mild soapy water (a few drops of dish soap in a warm cup of water). The gel can continue to show for two to five days. Do not put the rubber dust cap on the probe until all of the reference gel is removed. Do not complete the steps that follow until all reference gel is removed from the probe and the interior and exterior of the rubber dust cap.
Prepare the probe for storage
1. Put the new probe (or the probe that was in extended storage) in pH 7 buffer for at least
5 minutes.
2. Stir the probe in pH 7 buffer solution to dislodge air bubbles.
3. Use fresh deionized water to rinse the probe.
4. Dry the probe with a lint-free cloth.
English
15
5. Put the rubber dust cap back on.
6. Keep the probe in dry storage when not in use.
Calibrate the ISFET probe
Refer to Prepare the probe for storage on page 15 before calibration.
N O T I C E
Do not use the probe for applications that cycle between hot and room temperature samples.
Note: Remove other active or non-active measurement devices during pH or conductivity measurement. Other devices, even AC power, can cause interference.
Note: Only do calibrations away from the sun. Direct sunlight can cause unstable readings or difficulty in calibration.
1. Connect the probe to the meter.
2. Set the meter power to on.
3. Clean the probe with new deionized water and dry with a lint-free cloth.
4. Put the probe in the pH 7 buffer.
5. Clean the probe with deionized water and dry with a lint-free cloth.
6. Put the probe in the second buffer (pH 4 or pH 10).
7. Read the results.
8. If the result is not correct, the probe is not correctly hydrated.
Soak the probe for another 5 minutes in pH 7 buffer, then do the calibration again.
Clean the ISFET probe
N O T I C E
Do not use sharp metal objects (a needle, a pin, etc.) to clean the sensor. This can scratch the sensor and cause permanent damage to the probe.
For use in dairy, cheese or meat applications, soak the probe in Pepsin Cleaning Solution for 15 minutes before the probe is cleaned.
Regularly clean a non-glass probe:
1. Remove the rubber dust cap from the probe, then rinse with new deionized water.
2. Use a soft-bristle toothbrush and a mild detergent (a few drops of dish soap in a warm cup of
water) to carefully clean the probe.
3. Rinse with new deionized water to remove all debris from the sensor surface.
4. Dry the probe with a lint-free cloth.
5. Calibrate the probe again. Refer to Calibrate the ISFET probe on page 16, then Prepare the
probe for storage on page 15.
Repair the ISFET probes
N O T I C E
Do not use the probe to find out if the buffer is above 60 °C (140 °F). If the probe is suddenly put into very hot liquid, the probe can be permanently damaged.
Monitor how long the probes are in dry storage. If the probes are in storage for an extended period of time, the KCI gel at the reference junction can crystallize.
1. Heat pH 7 buffer to approximately 45 to 60 °C (115 to 140 °F).
2. Soak the probe for 2 minutes.
3. Put the probe in room temperature pH 7.00 buffer and let cool.
16
English

BNC electrode maintenance

Prevent unstable readings:
• Keep the probe in an electrode storage solution.
• Start measurements with at least a 2-point calibration. Update often with 1, 2 or 3-point calibrations.
• Use new buffers and new deionized rinse solution.
• Use buffer solutions with pH values no larger than 3 pH units apart. Ideally, the buffers bracket the anticipated pH values of the unmeasured samples.
• Use deionized water to rinse residual buffer and sample solutions from the probe after calibration and measurement.
• Calibrate at the same temperature as the sample solution. Although the meter has an ATC, get the best results when the calibration buffers and the sample are the same temperature.
• Keep the connectors clean and dry. Dirty or damp connectors can cause unstable readings.

Troubleshooting

Problem Possible cause Solution
No display Automatic shutdown set the
power to off.
There is no power. Replace the batteries.
Unstable reading The probe is dirty. Clean the probe.
The probe or the meter connectors are dirty.
No flow in the reference junction Clean the warm buffer.
ISFET probe is not properly hydrated.
Interference from other devices Remove other devices from the
Interference from direct sunlight Use protection for the probe from
The probe is in a very low ionic strength solution.
The pH or temperature of the solution changes.
Set the power to on again.
Clean the probe contacts on the cable connector. Clean the meter with methanol on a cotton swab. Let dry completely. Connect the probe to the meter again.
Soak the probe for at least 5 minutes in a pH 7.00 buffer.
solution.
direct sunlight.
A stable reading is not possible.
A stable reading is not possible until pH and temperature are constant.
English 17
Problem Possible cause Solution
The pH or temperature of the solution changes.
The meter continually shows 14.00 or
0.00 with an electrode attached. The "ISFET" icon does not show when an ISFET probe is attached.
The probe sensor surfaces are dirty or the probe needs to be conditioned again.
No flow in the reference junction Clean the warm buffer.
The buffers are contaminated or expired.
Interference from other devices in solution
The battery is low. Replace the batteries if the battery
Interference from direct sunlight. Out of calibration.
The probe is too old. Replace the probe.
The ISFET probe is not sensed by the meter or the probe.
Out of calibration. Calibrate the ISFET probe
The probe is not in the solution. Put the probe in liquid. Carefully
The probe is dirty. Clean the probe.
The probe is damaged. Replace the probe. Contact technical
Clean and condition the probe again.
Calibrate with new buffers again.
Remove other devices from the solution.
icon shows low battery power.
Use protection for the probe from direct sunlight.
Set the meter power to off. If the ISFET icon does not show, replace the ISFET probe.
on page 16
shake the probe to make sure that air bubbles are not caught on the sensor surface.
support.

Error codes

Table 2 shows what codes can occur for various reasons. Error codes can show instrument
malfunction or user error.
Table 2 Error code descriptions
Error code Description Solution
E03 Glass electrode offset error. Clean the probe. If the error
E04 Glass electrode slope error. Clean the probe. If the error
E06 ISFET pH probe slope error. Clean the probe. Soak in 40 ºC
E08 Too long to calibrate. Signal is not
stable during calibration.
18 English
continues, contact technical support to replace the probe.
continues, contact technical support to replace the probe.
(113 ºF) pH 7.00 buffer for 2 minutes. Try calibration again. If the error continues, contact technical support to replace the probe.
Clean the probe. Disconnect stirrers and other AC power sources. Make sure that calibration solution temperature is constant. If the error continues, contact technical support to replace the probe.
Table 2 Error code descriptions (continued)
Error code Description Solution
E13 The ISFET pH probe temperature
sensor is out of range.
E20 The conductivity probe temperature
sensor is out of range.
Replace the probe. Contact technical support for probe replacement information.
Replace the probe. Contact technical support for probe replacement information.

Replacement parts and accessories

W A R N I N G
Personal injury hazard. Use of non-approved parts may cause personal injury, damage to the instrument or equipment malfunction. The replacement parts in this section are approved by the manufacturer.
Note: Product and Article numbers may vary for some selling regions. Contact the appropriate distributor or refer to the company website for contact information.
Replacement parts
Description Quantity Item no.
AA Alkaline batteries 4 1938004
Accessories
Description Quantity Item no.
ISFET pH probes
General purpose round tip probe 1 PH77-SS
Micro pH probe 1 PH17-SS
Piercing tip micro pH probe 1 PH37-SS
pH probe for NMR tubes 1 PH47-SS
Heavy duty piercing tip micro pH probe 1 PH57-SS
Conductivity probes
General purpose conductivity probe, epoxy, K = 0.5 4 graphite contacts CDW97-KP5
LIS conductivity probe, glass, K = 1 4 platinum contacts CDW97-K1
Conductivity probe for salinity and brackish water, epoxy, K = 10
Glass pH probes
Gel-filled combination pH, BNC and 3.5 mm phone 1 5193511
Pt Series combination pH, BNC and 3.5 mm phone 1 5191011
Refillable combination pH, BNC and 3.5 mm phone 1 5194011
4 contacts CDW97-K10
English 19
Consumables
Description Quantity Item no.
Probe cleaning solution for removal of protein deposits on pH probes 500 mL 2964349
Rinse solution, non-ionic surfactant 500 mL 2964449
Buffer solution, red, pH 4.01 500 mL 2283449
Buffer solution, yellow, pH 7.00 500 mL 2283549
Buffer solution, blue, pH 10.01 500 mL 2283649
Conductivity standard solution, 100 μS/cm 500 mL 2971849
Conductivity standard solution, 1000 μS/cm 500 mL 1440049
Conductivity standard solution, 10,000 μS/cm 500 mL 2972249
Conductivity standard solution of 35 ppt salinity, 53 mS/cm 500 mL 2714349
Conductivity standard solution, 100 μS/cm 100 mL 2971842
Conductivity standard solution, 1000 μS/cm 100 mL 1440042
Conductivity standard solution, 10,000 μS/cm 100 mL 2972242
20 English

Inhaltsverzeichnis

Technische Daten auf Seite 21 Inbetriebnahme auf Seite 28
Allgemeine Informationen auf Seite 22 Betrieb auf Seite 29
Installation auf Seite 25 Wartung auf Seite 33
Benutzeroberfläche und Navigation auf Seite 26 Fehlerbehebung auf Seite 36

Technische Daten

Änderungen vorbehalten.
Technische Daten Details
Abmessungen (B x T x H) 9 x 20 x 5 cm (3.5 x 8 x 2 Zoll)
Gehäuse Wasserdicht
Gewicht 1300 g (3,0 lb.)
Stromversorgung 4-ANSI 15 A oder IEC-LR6 (AA-Alkali)
Stromverbrauch 1 W mit eingeschalteter Hintergrundbeleuchtung
Akkulebensdauer Bis zu 200 Stunden kontinuierlicher Betrieb, Warnung 10 Stunden bevor die Batterie
Betriebstemperatur 5 bis 40 °C (41 bis 104 °F)
Lagertemperatur -20 bis +40 °C (4 bis 140 °F)
Luftfeuchtigkeit bei Betrieb Relative Feuchte: maximal 50 % bei 25 ºC (77 ºF), nichtkondensierend
Eingangsanschlüsse ISFET, 8-polig, BNC mit Temperatur über Phono-Buchse, Leitfähigkeit, 12-polig (nur
Sprachen Englisch
Temperaturkorrektur Aus, automatisch und manuell (parameterabhängig)
Messdaten Kontinuierliche Messung
Schutzart IP67
Zertifizierungen CE
pH
Kalibrierung Bis zu fünf Punkte: 1,68, 4,01, 6,86, 7,00, 9,18, 10,01, 12,45
Genauigkeit ±0,01 pH
Auflösung 0,01 pH
Messbereich 0,00 bis 14,00
mV
Kalibrierung Keine
Genauigkeit 0,1 mV
Auflösung Autoranging, 0,1 und 1
Messbereich Autoranging, ±199,9 mV bis ±1.999 mV
Temperatur
Kalibrierung Keine
leer ist.
H170)
Deutsch 21
Technische Daten Details
Genauigkeit ±0,5 °C
Auflösung 0,1 °C (0,1 °F)
Messbereich –0 bis 100 °C (32 bis 212 °F)
Leitfähigkeit
Kalibrierung Bis zu fünf Punkte
Genauigkeit ±1 % ppt (–2 bis +35 °C)
Auflösung 0,01 ìS, 0,1 ìS ,1 ìS, 0,01 mS, 0,1 mS
Messbereich Autoranging: 0,00 bis 19,99 ìS, 20,0 bis 199,9 ìS, 200 bis 1.999 ìS, 2,00 bis
TDS
Kalibrierung Bis zu fünf Punkte
Genauigkeit ±1 % voller Skalenbereich oder ±1 Stelle
Auflösung 0,01 ppm, 0,1 ppm, 1 ppm, 0,01 ppt, 0,1 ppt, 1 ptt, 0,1 mg/l, 1 mg/l, 0,01 gal/l,
Messbereich Autoranging: 0,00 bis 9,99 ppm, 10,0 bis 99,9 pm, 100 bis 999 ppm, 1,00 bis
Salinität
Kalibrierung Keine (abgeleitet von Leitfähigkeit)
Genauigkeit ±0,1 ppt, –2 bis +35 °C (28,4 bis 95 °F )
Auflösung 0,1 ppt 1 %
Messbereich 0 bis 42 ppt
19,99 mS, 20,0 bis 199,9 mS
0,1 gal/l
9,99 ppt, 10,0 bis 99,9 ppt, 100 bis 200 ppt mg/l: 0,00 bis 199,9 mg/l, 200 bis 1.999 mg/l, 2,00 bis 19,99 gal/l, 20,0 bis 50 gal/l

Allgemeine Informationen

Der Hersteller ist nicht verantwortlich für direkte, indirekte, versehentliche oder Folgeschäden, die aus Fehlern oder Unterlassungen in diesem Handbuch entstanden. Der Hersteller behält sich jederzeit und ohne vorherige Ankündigung oder Verpflichtung das Recht auf Verbesserungen an diesem Handbuch und den hierin beschriebenen Produkten vor. Überarbeitete Ausgaben der Bedienungsanleitung sind auf der Hersteller-Webseite erhältlich.

Sicherheitshinweise

H I N W E I S
Der Hersteller ist nicht für Schäden verantwortlich, die durch Fehlanwendung oder Missbrauch dieses Produkts entstehen, einschließlich, aber ohne Beschränkung auf direkte, zufällige oder Folgeschäden, und lehnt jegliche Haftung im gesetzlich zulässigen Umfang ab. Der Benutzer ist selbst dafür verantwortlich, schwerwiegende Anwendungsrisiken zu erkennen und erforderliche Maßnahmen durchzuführen, um die Prozesse im Fall von möglichen Gerätefehlern zu schützen.
Bitte lesen Sie dieses Handbuch komplett durch, bevor Sie dieses Gerät auspacken, aufstellen oder bedienen. Beachten Sie alle Gefahren- und Warnhinweise. Nichtbeachtung kann zu schweren Verletzungen des Bedieners oder Schäden am Gerät führen.
Stellen Sie sicher, dass die durch dieses Messgerät bereitgestellte Sicherheit nicht beeinträchtigt wird. Verwenden bzw. installieren Sie das Messsystem nur wie in diesem Handbuch beschrieben.
22
Deutsch
Bedeutung von Gefahrenhinweisen
G E F A H R
Kennzeichnet eine mögliche oder drohende Gefahrensituation, die, wenn sie nicht vermieden wird, zum Tod oder zu schweren Verletzungen führt.
Kennzeichnet eine mögliche oder drohende Gefahrensituation, die, wenn sie nicht vermieden wird, zum Tod oder zu schweren Verletzungen führen kann.
Kennzeichnet eine mögliche Gefahrensituation, die zu geringeren oder moderaten Verletzungen führen kann.
Kennzeichnet eine Situation, die, wenn sie nicht vermieden wird, das Gerät beschädigen kann. Informationen, die besonders beachtet werden müssen.
W A R N U N G
V O R S I C H T
H I N W E I S
Warnhinweise
Lesen Sie alle am Gerät angebrachten Aufkleber und Hinweise. Nichtbeachtung kann Verletzungen oder Beschädigungen des Geräts zur Folge haben. Im Handbuch wird in Form von Warnhinweisen auf die am Gerät angebrachten Symbole verwiesen.
Dies ist das Sicherheits-Warnsymbol. Befolgen Sie alle Sicherheitshinweise im Zusammenhang mit diesem Symbol, um Verletzungen zu vermeiden. Wenn es am Gerät angebracht ist, beachten Sie die Betriebs- oder Sicherheitsinformationen im Handbuch.
Dieses Symbol weist auf die Gefahr eines elektrischen Schlages hin, der tödlich sein kann.
Elektrogeräte, die mit diesem Symbol gekennzeichnet sind, dürfen ab 12. August 2005 nicht in öffentlichen europäischen Abfallsystemen entsorgt werden. Benutzer von Elektrogeräten müssen in Europa in Einklang mit lokalen und nationalen europäischen Regelungen (EU-Richtlinie 2002/96/EG) Altgeräte kostenfrei dem Hersteller zur Entsorgung zurückgeben.
Hinweis: Mit der Wiederverwertung, der stofflichen Verwertung oder anderen Formen der Verwertung von Altgeräten leisten Sie einen wichtigen Beitrag zum Schutz unserer Umwelt.
Zertifizierung
Kanadische Vorschriften zu Störungen verursachenden Einrichtungen, IECS-003, Klasse A:
Entsprechende Prüfprotokolle hält der Hersteller bereit. Dieses digitale Gerät der Klasse A erfüllt alle Vorgaben der kanadischen Normen für Interferenz
verursachende Geräte. Cet appareil numérique de classe A répond à toutes les exigences de la réglementation canadienne
sur les équipements provoquant des interférences.
FCC Teil 15, Beschränkungen der Klasse "A"
Entsprechende Prüfprotokolle hält der Hersteller bereit. Das Gerät entspricht Teil 15 der FCC­Vorschriften. Der Betrieb unterliegt den folgenden Bedingungen:
1. Das Gerät darf keine Störungen verursachen.
2. Das Gerät muss jegliche Störung, die es erhält, einschließlich jener Störungen, die zu
unerwünschtem Betrieb führen, annehmen.
Änderungen oder Modifizierungen an diesem Gerät, die nicht ausdrücklich durch die für die Einhaltung der Standards verantwortliche Stelle bestätigt wurden, können zur Aufhebung der Nutzungsberechtigung für dieses Gerät führen. Dieses Gerät wurde geprüft, und es wurde festgestellt, dass es die Grenzwerte für digitale Geräte der Klasse A entsprechend Teil 15 der FCC­Vorschriften einhält. Diese Grenzwerte sollen einen angemessenen Schutz gegen gesundheitsschädliche Störungen gewährleisten, wenn dieses Gerät in einer gewerblichen
Deutsch
23
Umgebung betrieben wird. Dieses Gerät erzeugt und nutzt hochfrequente Energie und kann diese auch abstrahlen, und es kann, wenn es nicht in Übereinstimmung mit der Bedienungsanleitung installiert und eingesetzt wird, schädliche Störungen der Funkkommunikation verursachen. Der Betrieb dieses Geräts in Wohngebieten kann schädliche Störungen verursachen. In diesem Fall muss der Benutzer die Störungen auf eigene Kosten beseitigen. Probleme mit Interferenzen lassen sich durch folgende Methoden mindern:
1. Trennen Sie das Gerät von der Stromversorgung, um sicherzugehen, dass dieser die Störungen
nicht selbst verursacht.
2. Wenn das Gerät an die gleiche Steckdose angeschlossen ist wie das gestörte Gerät, schließen
Sie das störende Gerät an eine andere Steckdose an.
3. Vergrößern Sie den Abstand zwischen diesem Gerät und dem gestörten Gerät.
4. Ändern Sie die Position der Empfangsantenne des gestörten Geräts.
5. Versuchen Sie auch, die beschriebenen Maßnahmen miteinander zu kombinieren.

Produktübersicht

H I N W E I S
Trennen Sie das Gerät stets von der Stromversorgung, bevor eine Elektrode gewechselt wird. Verwenden Sie das Messgerät nur gemäß den Anweisungen in diesem Handbuch, da das Gerät andernfalls beschädigt werden kann.
Das Messgerät ist in zwei Modellversionen erhältlich:
• H160 – pH und ORP
• H170 – pH, ORP, Leitfähigkeit, TDS und Salinität
Die wasserdichten tragbaren Messgeräte der H-Serie verwenden Sensorelektroden aus Glas mit BNC-Anschlüssen und Sonden ohne Glas mit ISFET-Silizium-Chipsensoren (ISFET = ionenselektiver Feldeffekttransistor). Die Messgeräte verwenden jeweils nur eine pH-Elektrode (eine BNC-pH-Elektrode oder eine ISFET-pH-Sonde). Wenn das Messgerät eingeschaltet wird, erkennt das Messgerät automatisch den Typ der angeschlossenen Elektrode.
Abbildung 1 Tragbares Messgerät
1 Staubkappen aus Gummi 5 Anschluss für ISFET-Sonde (8-polig) 2 Anschlüsse für Leitfähigkeitssonde (12-polig, nur
H170)
3 Anschluss für BNC-Sonde 7 Ein/Aus-Schalter 4 3,5-mm-Phono-Buchse für pH-Elektrode aus Glas,
ISE-, ORP- oder DO-Temperatursensoren
6 LCD-Display
8 Batteriefachabdeckung
24 Deutsch

Produktkomponenten

Stellen Sie sicher, dass Sie alle Teile erhalten haben. Siehe Abbildung 2. Wenn Komponenten fehlen oder beschädigt sind, kontaktieren Sie bitte den Hersteller oder Verkäufer.
Abbildung 2 Komponenten tragbarer Messgeräte
1 Wasserdichtes tragbares Messgerät (H160 im Bild) 2 Alkali-Batterien 4 x Größe AA

Installation

V O R S I C H T
Mehrere Gefahren. Nur qualifiziertes Personal sollte die in diesem Kapitel des Dokuments beschriebenen Aufgaben durchführen.

Elektrische Installation

Batterien einlegen.
V O R S I C H T
Explosionsgefahr. Das unsachgemäße Einlegen von Batterien kann zur Freisetzung explosiver Gase führen. Vergewissern Sie sich, dass Sie Batterien mit dem zulässigen Chemikalientyp verwenden und dass sie mit der korrekten Polung eingelegt wurden. Verwenden Sie nicht alte und neue Batterien zusammen.
Ziehen Sie die Schrauben nicht zu fest an, da andernfalls Schäden am Instrument auftreten können.
Führen Sie diesen Vorgang nicht aus, wenn Sonden am Messgerät befestigt sind. Entfernen Sie alle Sonden vom Messgerät, da andernfalls Schäden am Instrument auftreten können.
Das Messgerät verwendet AA-Alkali-Batterien oder NiMH-Akkus. Um die Lebensdauer der Batterie zu verlängern, beachten Sie Konfigurieren des Messgeräts auf Seite 28 zum Konfigurieren der Ausschaltautomatik des Messgeräts.
Siehe Abbildung 3 zum Installieren der Batterien.
Zusätzlich erforderliche Artikel:
• Kreuzschlitz-Schraubendreher
• Alkali-Batterien 4 x Größe AA
H I N W E I S
H I N W E I S
Deutsch
25
Abbildung 3 Einlegen der Batterien
Sonde und Verbindungstypen
H I N W E I S
Vergewissern Sie sich, dass alle Gummistopfen korrekt montiert sind, bevor Sie das tragbare Messgerät in Wasser legen. Andernfalls tritt Wasser durch die 3,5-mm-Phono-Buchse ein. Das Innere des tragbaren Messgeräts ist jedoch abgedichtet. Wenn Wasser durch die 3,5-mm-Phono-Buchse eintritt, drehen Sie das Messgerät sofort um, sodass das Wasser aus der Buchse fließen kann. Lassen Sie den Anschluss dann trocknen.
Schalten Sie das Messgerät stets aus, bevor der Sondenanschluss geändert wird. Andernfalls kann das Instrument beschädigt werden.
Das Messgerät kann an zwei Arten von pH-Sensoren angeschlossen werden: ISFET-pH-Sonden und Glaselektroden. Nach dem Einschalten erkennt das Messgerät den Typ der angeschlossenen Sonde automatisch. Das Messgerät verwendet jeweils nur einen pH-Sondentyp.
Hinweis: Wenn keine Sonde an das Messgerät angeschlossen ist, ist der ausgewählte Modus inaktiv.
H I N W E I S

Benutzeroberfläche und Navigation

Beschreibung des Tastenfelds

Abbildung 4 und Tabelle 1 zeigen die Funktionen der Tasten jedes Messgeräts.
26
Deutsch
Abbildung 4 Beschreibungen des Tastenfelds
Tabelle 1 Tastenfeld-Funktionen
Taste Tätigkeit H160 H170
ON/OFF: Schaltet das Messgerät ein oder aus. x x
pH-Modus x x
mV-Modus x x
Zum Einschalten der Hintergrundbeleuchtung. Nach 2 Minuten ohne Tasteneingabe schaltet sich die Hintergrundbeleuchtung aus.
TDS-Modus x
Leitfähigkeitsmodus x
Salinitätsmodus x
Haltemodus: Funktioniert nur in den Messmodi. Bei aktiviertem Haltemodus werden alle Ergebnisse eingefroren.
Kalibriermodus x x
Zum Blättern durch Werte, Setup-Bildschirme und Optionen. x x
Setup-Modus x x
Zum Auswählen einer Option, einer Einstellung oder eines Wertes. x x
x x
x x
Deutsch 27

Displaybeschreibung

Abbildung 5 zeigt Messmodi und -werte, Datenverbindung und Speicheroptionen, Batteriestatus,
Temperaturwerte, Stabilisierungssperre und Verbindungsstatus auf dem Display.
Abbildung 5 Display
1 Setup-Modus 7 Messmodus 2 Temperatur und Datenwerte 8 Maßeinheiten 3 Gemessener Wert 9 Wert für automatische Temperaturkompensation
4 Stabilisierungssperre 10 Datenspeicher 5 Sperranzeige 11 Temperatureinheit 6 ISFET-Sondenanzeige 12 Anzeige für niedrigen Batterieladestand
(ATC) (pH, Leitfähigkeit, TDS oder Salinität)

Inbetriebnahme

Einschalten

Drücken Sie die Einschalttaste, um das Messgerät ein- oder auszuschalten. Vergewissern Sie sich, dass die Stromversorgung (Netzstrom oder Batteriebetrieb) korrekt installiert ist.

Konfigurieren des Messgeräts

Führen Sie die Vorgänge in dieser Reihenfolge durch:
1. Schalten Sie das Gerät ein.
2. Drücken Sie SETUP, um das Messgerät in den pH-Modus zu versetzen.
3. Wählen Sie mit den Pfeiltasten eine Option, und drücken Sie dann ENTER:
28
Deutsch
Optionen Funktion Beschreibung
BEREIT Stabilisierungssperre Schaltet die Stabilisierungssperre ein oder aus.
• Das Symbol „BEREIT“ wird angezeigt, wenn ein Endpunkt auftritt und die Stabilisierungssperre eingeschaltet ist.
• Nach der Stabilisierung sperrt das Display den Wert, und das Messgerät ignoriert Änderungen bei der Messung des pH-, Leitfähigkeits- oder TDS-Werts.
• Stellt die Stabilisierungssperre während Titrierungen oder bei Erkennen leichter Veränderungen aus. Das Display wird automatisch entsperrt, wenn signifikante Veränderungen bei der Messung erkannt werden.
ºC ºF Temperaturanzeige Legt die Temperatureinheit auf °C oder °F fest.
Snd Signaltöne Schaltet die Signaltöne ein oder aus. Drei Töne informieren den
Benutzer über unterschiedliche Funktionen:
Ein Piepton: Stabilisierung ist bei eingeschalteter Stabilisierungssperre aufgetreten.
Zwei Pieptöne: Ein Fehler ist aufgetreten. Der Fehlercode wird auf dem Display angezeigt. Siehe Fehlercodes auf Seite 37.
Drei Pieptöne: Messstabilität im Kalibrierungsmodus, unabhängig von der Einstellung der Stabilisierungssperre.
oFF Ausschaltautomatik Legt die Ausschaltparameter von 1 Minute bis 000 Minuten
CLr ALL CAL
1
Dadurch werden die Kalibrierungsdaten für eine ISFET-Sonde nicht gelöscht.
Kalibrierungspunkte löschen
(kontinuierliche Stromversorgung) fest. Das Messgerät piept 1 Minute vor dem Ausschalten. Schalten Sie das Messgerät aus und wieder ein, um die Ausschaltautomatik zu aktivieren. Die Ausschaltautomatik wird während Datenaufzeichnung, Bluetooth™-Übertragungen und bei Verbindung zum USB-Anschluss deaktiviert.
Löscht alle Kalibrierungspunkte aus dem Speicher. Kalibrieren Sie das Messgerät erneut.
1
4. Drücken Sie ENTER, um die Änderungen zu speichern und zum Setup-Modus zurückzukehren.
Hinweis: Um ohne Speicherung der Änderungen zurückzukehren, drücken Sie SETUP.

Betrieb

Ignorieren eines Fehlers

Der Benutzer kann einige Fehler (z. B. schlechter Elekrodenzustand) ignorieren, bis eine Ersatzelektrode verfügbar ist.
Hinweis: Nicht alle Fehler können ignoriert werden. Der Hersteller empfiehlt das Ignorieren von Fehlercodes nicht. Wenden Sie sich an den technischen Support, um Fehlercodes zu ignorieren.
So brechen Sie einen Fehler ab oder ignorieren ihn:
1. Finden Sie die Ursache des Fehlers heraus. Siehe Fehlercodes auf Seite 37. Falls möglich, drücken Sie SETUP oder pH, COND, TDS oder SAL, um den Kalibrierungsfehler oder einen anderen Fehler abzubrechen.
2. Um den Fehler zu ignorieren, führen Sie Schritt 1 nicht durch. Drücken Sie stattdessen ENTER.
3. Blättern Sie mit den Pfeiltasten, bis „Yes” (Ja) angezeigt wird.
4. Drücken Sie ENTER (EINGABE).

Festlegen der pH-Vorgänge

Legen Sie den Kalibrierungsalarm, die pH-Auflösung, die Kalibrierungspunkte und die Kalibrierungsmethode fest. Führen Sie die Vorgänge in der angegebenen Reihenfolge durch.
Deutsch
29
1. Schalten Sie das Gerät ein.
2. Drücken Sie SETUP und dann pH.
3. Wählen Sie mit den Pfeiltasten eine Option, und drücken Sie dann ENTER:
Betrieb Funktion
pH-Auflösung Legt die pH-Auflösung auf 0,1 oder 0,01 pH fest. Die Stabilisierungssperre
Festlegen der 2- oder 3-Punkt­pH-Kalibrierung
pH-Kalibrierungsmethode
(das Symbol „READY“) ist schneller als eine 0,1-pH-Auflösung.
Zeigt die aktuelle Anzahl der Kalibrierungspunkte an. Verwenden Sie die Pfeiltasten, um die Kalibrierungspunkte festzulegen.
• Legt den pH-Puffer fest. Standardmäßig wird eine 2-Punkt-Kalibrierung durchgeführt (pH 7,00 dann pH 4,00).
• Drücken Sie erneut ENTER, um den ersten Kalibrierungspuffer festzulegen. Auf dem Display wird der nächste Puffer angezeigt. Führen Sie diese Vorgehensweise aus, bis alle Puffer festgelegt sind. Zuvor ausgewählte Puffer werden nicht angezeigt.
4. Drücken Sie ENTER, um die Änderungen zu speichern und zum Setup-Modus zurückzukehren.
Hinweis: Um ohne Speicherung der Änderungen zurückzukehren, drücken Sie SETUP.
Festlegen der manuellen Temperaturkompensation
Bei pH-Elektroden aus Glas findet eine automatische Temperaturkompensation statt, wenn eine Temperatursonde in der 3,5-mm-Phono-Buchse erkannt wird. Wenn keine Sonde erkannt wird, verwendet das Messgerät die standardmäßige manuelle Temperaturkompensation bei 25,0 ºC. Manuelle Temperaturkompensation wird immer in ºC angegeben. Siehe Set the manual temperature
compensation for glass electrodes.
pH-Pufferwerte bei verschiedenen Temperaturen
25 ºC 0 ºC 5 ºC 10 ºC 20 ºC 30 ºC 40 ºC 50 ºC
1,68 1,67 1,67 1,67 1,67 1,68 1,69 1.71
4,01 4,00 4,00 4,00 4,00 4,01 4,03 4,06
6,86 6,98 6,95 6,92 6,87 6.85 6.84 6,83
7,00 7,12 7,09 7,06 7,01 6,99 6,97 6,97
9.18 9,46 9,40 9,33 9,23 9,14 9,07 9,02
10,01 10,32 10,25 10,18 10,06 9,97 9,98 9,83
12,45 13,42 13,21 13,00 12,63 12,29 12,04 11,70
1. Drücken Sie SETUP und dann pH, um den pH-Setup-Modus aufzurufen.
2. Verwenden Sie die Pfeiltasten, bis für manuelle Temperatur „- - -.- ºC“ auf dem Display angezeigt
wird.
3. Drücken Sie ENTER, um den manuellen Temperaturwert anzuzeigen.
4. Verwenden Sie die Pfeiltasten, um die manuelle Temperaturkompensation festzulegen.
5. Drücken Sie ENTER, um die Änderungen zu speichern und zum Setup-Modus zurückzukehren.
Hinweis: Um ohne Speicherung der Änderungen zurückzukehren, drücken Sie SETUP.
Kalibrieren für pH-Messung
Das Messgerät akzeptiert 1-, 2- oder 3-Punkt-Kalibrierungen. Der Standard sind 2-Punkt­Kalibrierungen.
1. Schalten Sie das Gerät ein.
2. Schließen Sie die Sonde an das Messgerät an.
30
Deutsch
Loading...
+ 104 hidden pages