Hach-Lange GANIMEDE N User Manual

DOC022.52.90472
GANIMEDE P and GANIMEDE N laboratory machine
User manual
06/2013, Edition 1A
© HACH-LANGE GmbH. , 2013. All rights reserved. Printed in Germany
Table of contents
Section 1 Technical data ........................................................................................................................5
Section 2 General information...............................................................................................................7
2.1.1 Hazard notes in this manual ...................................................................................................... 7
2.1.2 Warning labels ........................................................................................................................... 7
2.1.3 Chemical and biological safety ..................................................................................................8
2.1.4 Safety around source lamps ......................................................................................................8
2.2 Product overview ................................................................................................................................ 9
2.2.1 Product overview of the Ganimede P ........................................................................................ 9
2.2.2 Product overview of the Ganimede N ........................................................................................ 9
2.3 Scope of delivery .............................................................................................................................. 10
2.4 Front and rear view........................................................................................................................... 10
2.4.1 Sample changer....................................................................................................................... 10
2.4.2 Open-loop control unit.............................................................................................................. 11
2.4.3 Analysis unit............................................................................................................................. 12
Section 3 Installation............................................................................................................................ 17
3.1 Installation of the components..........................................................................................................17
3.1.1 Operating environment ............................................................................................................ 17
3.1.2 Installation of the devices......................................................................................................... 17
3.1.3 Installation of the cables .......................................................................................................... 19
3.1.4 Installation of the tubes............................................................................................................ 20
3.1.4.1 Installation of the tubes for the Ganimede P................................................................... 20
3.1.4.2 Installation of the tubes for the Ganimede N................................................................... 23
3.1.5 Installation of the sample tray .................................................................................................. 26
3.2 Open-loop control unit ...................................................................................................................... 27
3.2.1 Overview of the display and keys of the open-loop control unit............................................... 27
3.2.2 Switch-on of the open-loop control unit.................................................................................... 28
3.3 Reagents .......................................................................................................................................... 28
3.3.1 Reagents for Ganimede P ....................................................................................................... 28
3.3.2 Reagents for Ganimede N ....................................................................................................... 29
3.3.3 Storage .................................................................................................................................... 29
3.3.4 Reagents reserve..................................................................................................................... 29
3.4 Ganimede P sample preparation...................................................................................................... 29
3.4.1 Load the sample tray ............................................................................................................... 29
3.5 Ganimede N sample preparation...................................................................................................... 31
3.5.1 Load the sample tray ............................................................................................................... 31
Section 4 Operation.............................................................................................................................. 35
4.1 Operation of the open-loop control unit ............................................................................................ 35
3
Table of contents
4.2 Menu structure of the open-loop control unit.....................................................................................36
4.2.1 Menu structure .........................................................................................................................36
4.2.2 Start of a measurement series .................................................................................................40
4.2.3 Calibration (recording of multi-point calibrations).....................................................................41
4.2.3.1 Ganimede P.....................................................................................................................41
4.2.3.2 Ganimede N ....................................................................................................................42
4.2.4 Rinse ........................................................................................................................................42
4.2.5 Stop..........................................................................................................................................43
4.2.6 Data/current..............................................................................................................................43
Section 5 Maintenance..........................................................................................................................45
5.1 Sample changer ................................................................................................................................45
5.2 Open-loop control unit.......................................................................................................................45
5.3 Analysis unit ......................................................................................................................................45
5.3.1 Mount the cover........................................................................................................................45
5.3.2 Downtime .................................................................................................................................46
5.3.3 Tube change ............................................................................................................................46
5.3.3.1 Insertion of the tube in a pump........................................................................................46
5.3.4 Clean the analysis cuvette .......................................................................................................47
5.3.5 Cleaning of the analysis unit ....................................................................................................49
Section 6 Troubleshooting ...................................................................................................................51
6.1 Error messages on the open-loop control unit ..................................................................................51
6.2 Error messages on the analysis unit .................................................................................................51
Section 7 Replacement parts and accessories ................................................................................53
7.1 Replacement parts ............................................................................................................................53
7.2 Accessories.......................................................................................................................................53
Section 8 Contact information ............................................................................................................55
Section 9 Warranty and liability ...........................................................................................................57
Appendix A Interface protocol.............................................................................................................59
4

Section 1 Technical data

Subject to change.
Ganimede P analysis unit
Ganimede P for the automatic serial analysis of orthophosphate
Typ e
Photometric method
Total P digestion Rapid digestion at 150 °C and 6 bar
Sample pH pH 2 – 11
and total phosphate in the analytical laboratory, consisting of analysis unit, sample changer and control unit
By means of ammonium molybdate in accordance with DIN EN 1189 of December 1996
Measurement range 0.01 – 3.8 mg/L PO
Measurement accuracy ±0.03 mg/L at 1 mg/L
Measurement wavelength 880 nm
Calibration 2-point calibration and 10-point calibration
Processing time per sample 4 minutes for single determination
Ganimede N analysis unit
Ganimede N for the automatic serial analysis of inorganically and
Typ e
Photometric method
Digestion Alkaline digestion at 150 °C and 8 bar
Measurement range
Reproducibility of the measurement results < 2 % at 15 mg/L
Carryover < 1 % within a measurement range
Measurement wavelengths 210 nm and 228 nm
Calibration
organically bound nitrogen in the analytical laboratory, consisting of analysis unit, sample changer and open-loop control unit
By means of UV self-absorption at 210 nm and 228 nm in accordance with DIN EN ISO 11 905-1
Low measurement range 0.5 – 30 mg/L TN High measurement range 30 – 150 mg/L TN
2-point calibration and multiple-point calibration (2–11 measurement points)
-P
4
B
B
Processing time per sample 7 minutes for single determination
Sample changer
Sample tray for 36 × 25 mL or
Sample tray
Environmental conditions
Temperature range 15 – 35 °C
Relative humidity < 75 %, non-condensing
Air pressure 75 – 106 kPa
Equipment properties
Analysis unit
Particle mobility
Sample changer
Sample tray for 53 × 15 mL Tray diameter 240 mm
According to DIN 38402; the prerequisites of EN 1484 TOC of > 100 µm are significantly exceeded
According to DIN 38402; the prerequisites of EN 1484 TOC of > 100 µm are significantly exceeded with 1 mm (sample needle)
5
Technical data
Tube material Analysis unit
Display
Power connector
Protection class
Housing protection rating Sample changer IP 20
Overvoltage category
Power consumption
Open-loop control unit
Analysis unit 230 V +5%/-15%, 50 Hz
Sample changer 100–240 V, +5%/-15%, 50/60 Hz
Open-loop control unit
Analysis unit Class I
Sample changer Class I
Analysis unit II
Sample changer II
Analysis unit 150 VA
Sample changer 48 VA
Open-loop control unit
Ismaprene, thermoplastic elastomer on polypropylene basis, almost opaque beige, suitable for the food and medical sectors.
5.7-inch color display
External power supply: Input: 100–240 V, 50/60 Hz Output: 15 VDC, 2 A Or rechargeable battery operation: Rechargeable battery pack:
15 V, 6 rechargeable batteries, type HHR-200A
20 VA
Solution class Sample changer 1
Analysis unit 370 × 710 × 360 mm (14.57 × 27.95 × 14.17 inches)
Dimensions (W×H×D)
Weight
Other
Interface
Sample changer 280 × 210 × 390 mm (11.02 × 8.27 × 15.35 inches)
Open-loop control unit
Analysis unit 40 kg
Sample changer 5 kg (packed)
Open-loop control unit
Analysis unit
Sample changer RS 232 (9-pole Sub-D connector) for analysis unit
Open-loop control unit
65 × 245 × 190 mm (2.56 × 9.65 × 7.48 in.)
2.7 kg (packed)
2 x IR interfaces to the open-loop control unit on the front side 1 x RS 232 (9-pole Sub-D socket) for sample changer 2 x RS 232 C for PC/LIMS 1 x Centronics for printer
Wireless connection to LIMS or printer for the transfer of the results with IR interfaces
Serial RS 232 interface for PC or CCD scanner 8-pole round socket for PS/2 compact keyboard
Certification CE compliant
6

Section 2 General information

WARNING
It is forbidden to remove, bypass or override safety devices, safety functions and monitoring devices.
WARNING
The manufacturer is not responsible for any damages due to misapplication or misuse of this product including, without limitation, direct, incidental and consequential damages, and totally excludes such damages as permitted under applicable laws.
The user is solely responsible to identify critical application risks and install appropriate mechanisms to protect processes during a possible equipment malfunction.

2.1 Safety information

Read this entire manual carefully before unpacking, setting up or operating the device. Pay attention to all hazard and warning notes. Failure to do so could result in serious injury to the operator or damage to the device.
To prevent damage to or impairment of the device's protection equipment, the device may only be used or installed as described in this manual.

2.1.1 Hazard notes in this manual

Indicates a potentially or imminently hazardous situation that, if not avoided, can result in death or serious injury.
Indicates a potentially or imminently dangerous situation that, if it is not avoided, can lead to death or to serious injuries.
Indicates a possible dangerous situation that can have minor or moderate injuries as the result.
Indicates a situation that, if it is not avoided, can lead to damage to the device. Information that requires special emphasis.
Note: Information that supplements points in the main text.

2.1.2 Warning labels

DANGER
WARNING
CAUTION
NOTICE
Read all labels and tags attached to the device. Failure to do so could result in personal injury or damage to the device.
This symbol, if noted on the instrument, references the instruction manual for operation and/or safety information.
7
General information
18-32 kg (39.7-70.5 lbs
This symbol may be found on an enclosure or barrier within the product and indicates a risk of electric shock and/or death by electrocution.
If shown on the device, this symbol indicates the location of a fuse or current limiter.
If shown on the device, this symbol indicates a component that can be hot and must not be touched without precautionary measures.
This symbol indicates that a UV lamp is used in the device.
If shown on the device, this symbol indicates dangerous chemical substances. The handling of chemicals and the execution of maintenance on chemical feed equipment of the device may only be carried out by qualified personnel who are trained to work with chemicals.
If shown on the device, this symbol indicates dangerous places on the device that can result in crushing or surface damage. Only operate the device with closed safety covers.
Electrical equipment marked with this symbol may no longer be disposed of in unsorted European domestic or public disposal systems after 12 August 2005. In conformity with the applicable provisions, consumers in the EU must return old electrical equipment to the manufacturer for disposal from this date. This is at no charge to the user.
Note: Contact the manufacturer or supplier for instructions on how to return end-of-life devices, manufacturer-supplied electrical accessories and all auxiliary items for pro per disposal.
When carrying or transporting the instrument components or instrument and if the total weight is more than 18 kg, make sure that suitable lifting equipment is used and/or that the instrument components or instrument are/is carried by two people.

2.1.3 Chemical and biological safety

Potential danger with contact with chemical/biological substances. Working with chemical samples, standards and reagents can be dangerous. Make yourself
familiar with the necessary safety procedures and the correct handling of the chemicals before the work and read and follow all relevant safety data sheets.
Normal operation of this instrument may involve the use of hazardous chemicals or biologically harmful samples.
Before handling these substances observe all danger notes and safety information printed on the containers of the original solutions and in the safety data sheet.
DANGER
Dispose of all consumed solutions in accordance with the national regulations and
laws.
Select the type of protective equipment accordingly

2.1.4 Safety around source lamps

The source lamps are operated at high temperatures.
8
To avoid the risk of electrocution, make sure the instrument is disconnected from the power source before changing the lamps.
Health hazard caused by ozone. Hazardous levels of ozone can be generated when the UV lamp is not cooled.
Health risk due to UV light. UV light can cause irritation to the eyes and skin. Protect eyes and skin from direct exposure to
UV light. Do not look directly at an energized lamp without UV safety glasses.
Burn hazard. Allow the lamp(s) to cool for at least 30 minutes before maintaining/replacing them.

2.2 Product overview

General information
CAUTION
WARNING
CAUTION
The instrument may not be used in dangerous environments. The manufacturer and its suppliers reject any express or indirect guarantee for use with high-risk
activities.
Any use other than use in accordance with requirements defined in the user manual leads to the loss of the warranty claims and can lead to personal injury and property damage, for which the manufacturer assumes no liability.

2.2.1 Product overview of the Ganimede P

In combination with the ready-to-use GaniChem P GCA 100 reagent kit, the Ganimede P enables a rapid and automated serial analysis of PO orthophosphate in accordance with DIN EN 1189 of December 1996.
The digestion of the total phosphate takes place as high-temperature digestion.
The phosphate content is determined as phosphorus molybdenum blue at 880 nm.

2.2.2 Product overview of the Ganimede N

In combination with the ready-to-use GaniChem N GCA 200 reagent kit, the Ganimede N enables a quick and automated serial analysis of inorganically and organically bound nitrogen in aqueous samples in accordance with DIN EN ISO 11 905-1.
WARNING
NOTICE
and PO4-P as total phosphate or
4
Inorganically and organically bound nitrogen is oxidized to nitrate with peroxodisulfate by means of alkaline digestion at 150
The nitrate content is photometrically measured by means of UV self-absorption in a differential measurement at 210
°C and 8 bar.
nm and 228 nm.
9
General information

2.3 Scope of delivery

Please check that the order is complete. If anything is missing or damaged, please contact the manufacturer or retailer.
Sample changer
Open-loop control unit
Analysis unit P or analysis unit N

2.4 Front and rear view

2.4.1 Sample changer

Figure 1 Front view of sample changer
1 Sample changer arm 5 Reset button 2 Sample needle 6 Sample tray (here, 36 positions for 30-mL sample 3 Rinsing vessel for sample needle
4 Sample containers 7 Light-emitting diode
10
containers; a sample tray with 53 positions for 18-mL sample containers is optionally available)
General information
Figure 2 Rear view of sample changer
1 Sample needle 6 Suction tube for rinse water 2 Fuse (2× T 1.25 A L, 250 VAC) 7 Drain hose 3 Power switch 8 Type plate 4 Power connector 9 Rinsing vessel for sample needle 5 Interface for the analysis unit connection cable

2.4.2 Open-loop control unit

Figure 3 Front view of the open-loop control unit
1 Shortcut keys 4 Alphanumeric keys 2 Graphical display 5 Menu keys 3 Cursor keys
11
General information
Figure 4 Detailed view of the open-loop control unit
1 IR interface 4 Power switch 2 RS 232 interface for PC and hand-held scanner 5 Power connector 3 Interface for compact keyboard

2.4.3 Analysis unit

Figure 5 Front view of analysis unit
1 Analysis cuvette monitoring window 4 IR interfaces 2 LCD display of the current sample 5 Power switch 3 Digestion cuvette monitoring window
12
Figure 6 Interior view of analysis unit P
General information
1 Digestion solution A 7 Acid solution D 2 Color solution B 8 Reduction solution C 3 Waste bottle 9 Distilled water 4 Valve 10 IR interfaces 5 Pump 11 Power switch 6 Safety cover
13
General information
Figure 7 Interior view of analysis unit N
1 Acid solution A 6 Safety cover 2 Digestion solution B 7 Distilled water 3 Waste bottle 8 IR interfaces 4 Val ve 9 Power switch 5 Pump
Figure 8 Analysis unit power connector
1 Type plate 4 Power connector 2 230 V/115 V voltage selector switch 5 Sample changer power connector 3 T3A L fuse, 250 V
14
Figure 9 Analysis unit interfaces
1 Printer interface 3 RS 232 socket 2 Sample changer interface 4 RS 232 connector
General information
15
General information
16

Section 3 Installation

3.1 Installation of the components

The instrument may not be used in dangerous environments. The manufacturer and its suppliers reject any express or indirect guarantee for use with high-risk
activities.
Any use other than use in accordance with requirements defined in the user manual leads to the loss of the warranty claims and can lead to personal injury and property damage, for which the manufacturer assumes no liability.
Unpack all supplied parts carefully, as they are highly sensitive in part to shock and impact. Read the user manual prior to installation and proceed exactly as described.

3.1.1 Operating environment

Observe the following points to allow the instrument to function normally and give a long operating life.
Position the instrument securely on a flat surface taking care to remove any objects from under the device.
WARNING
NOTICE
The ambient temperature must be 15 °C to 35 °C (60 °F to 95 °F).
The relative humidity must be lower than 75% (no condensate formation).
A minimum clearance of 15 cm (5.9inches) must be ensured above and on all sides of
the device; this allows the air to circulate and prevents overheating of the electric components. The dosage arm of the sample changer must be able to move freely.
Do not operate or store the instrument in extremely dusty, humid or wet locations.
Make sure that no liquids enter the instrument and immediately wipe away any liquids
that contact the instrument.
Protect the instrument against vibrations, direct sunlight, corrosive gases as well as strong magnetic and/or electric fields.
Only use genuine replacement parts and accessories.
The instrument must be serviced by the manufacturer's service department at the
prescribed intervals.

3.1.2 Installation of the devices

When carrying or transporting the instrument components or instrument and if the total weight is more than 18 kg, make sure that suitable lifting equipment is used and/or that the instrument components or instrument are/is carried by two people.
WARNING
The carrying strap provided must be used to be able to carry the analysis unit safely over short distances.
17
Installation
Figure 10 Use of the carrying strap
1 Slip the carrying strap over the analysis unit as shown in
the illustration.
2 Carefully tip the analysis unit to the side and guide the
outer parts of the carrying strap behind the feet of the unit. Proceed in the same way on the opposite side.
1. Place the analysis unit on a stable and level tabletop. Level out any unevenness with the rear adjustable feet.
2. Remove the transport locking screws on the bottom of the sample changer. To do so, tip the sample changer to the side and remove the screws above the corresponding sign.
3. Position the sample changer to the left of the analysis unit on a stable and level tabletop.
4. Place the open-loop control unit near (approx. 0.5 m 1.64 ft) to the analysis unit.
Note: Communication between the analysis unit and the open-loop control unit takes place via IR interfaces.
3 Tightly pull the carrying belt upwards at the front and
back sides.
4 The analysis unit must now be transported safely over a
short distance by two people.
18
Loading...
+ 42 hidden pages