Hach-Lange User Manual [en, ru, de, es, fr, it, cs, pl]
Specifications and Main Features
Frequently Asked Questions
User Manual
DOC023.98.90162
FP 360 sc
01/2010, Edition 1B
en 53 Basic User Manual
de 72 Basis-Bedienungsanleitung
it 57 Manuale utente di base
fr 77 Manuel d'utilisation de base
es 61 Manual básico del usuario
pt 79 Manual do utilizador básico
cs 85 Základní uživatelská příručka
da 58 Grundlæggende brugervejledning
nl 56 Basisgebruikershandleiding
pl 60 Podstawowa instrukcja obsługi
sv 59 Grundläggande bruksanvisning
fi 63 Peruskäyttöohje
bg 89 Основно ръководство за потребителя
hu 86 Alap felhasználói kézikönyv
ro 87 Manual de bază al utilizatorului
ru 62 Основное руководство пользователя
tr 94 Temel Kullanıcı Kılavuzu
sk 88 Základný návod na použitie
sl 44 Osnovni uporabniški priročnik
hr 43 Osnovni korisnički priručnik
sr 47 Osnovni priručnik za korišćenje
el 48 Βασικό ΕγχειρίδιοΧρήστη
English ...............................................................................................................................................................3
Italiano .............................................................................................................................................................35
Français ...........................................................................................................................................................51
Español ............................................................................................................................................................67
Dansk .............................................................................................................................................................115
Nederlands ....................................................................................................................................................131
Polski .............................................................................................................................................................147
Svenska .........................................................................................................................................................165
Suomi .............................................................................................................................................................181
Български .....................................................................................................................................................197
Magyar ...........................................................................................................................................................215
Română .........................................................................................................................................................231
Türkçe ............................................................................................................................................................263
Slovenský jazyk .............................................................................................................................................279
Hrvatski ..........................................................................................................................................................311
Srpski .............................................................................................................................................................327
Specifications are subject to change without notice.
Measurement
Measurement method
Measuring range
Display unitsppb, ppm, µg/L, mg/L
Reproducibility2.5% of measurement value at a constant temperature
Measurement accuracy5% of measurement value ±2% from measurement range limit at a constant temperature
Limit of detection1.2 ppb (PAH)
Response time10 s (T90)
CalibrationFactory calibration with UV fluorescence calibration standard, custom adaptation possible
Sensor software
Software versionFrom 1.14
Equipment properties
Weight
Pressure range
Dimensions
Low measurement range
High measurement range
Measurement sensorStainless steel version 2.8 kg, titanium version 1.8 kg
Flow cellApproximately 0.6 kg, including installation board approximately 2.0 kg
Measurement sensorMax. 30 bar
Flow cell Max. 1 bar
Measurement sensor
(Ø × length)
Flow cell (L × W × D)
UV fluorescent measurement process for polycyclic aromatic hydrocarbons (PAH)
Excitation wavelength: 254 nm Emission (measurement) wavelength: 360 nm
0 to 50 ppb and 0 to 500 ppb in relation to PAH calibration standard, corresponding to
0.1 to 1.5 ppm and 0.1 to 15 ppm of oil calibration standard
0 to 500 ppb and 0 to 5000 ppb in relation to PAH calibration standard, corresponding to
0.1 to 15 ppm and 0.1 to 150 ppm of oil calibration standard
68 mm × 306 mm (2.68 in. × 12.05 in.) (without plugs or suspension pin)
68 mm × 399 mm (2.68 in. × 15.71 in.) (including suspension pin)
68 mm × 413 mm (2.68 in. × 16.26 in.) (with additional cleaning option)
98 mm × 98 mm × 150 mm (3.86 in. × 3.86 in. × 5.91 in.) (without fittings),
installation board: 600 × 300 × 10 mm (23.62 in. × 11.81 in. × 0.39 in.)
English
English 3
Equipment properties (continued)
Housing: stainless steel 1.4571 or titanium
Optic bracket: POM
Measurement sensor
Materials
Flow cell
ShackleStainless steel 1.4301
Environment considerations
Measuring medium temperature1 to 40 °C (34 to 104 °F)
Ambient temperature
Sensor distance - wall/groundMinimum 100 mm (3.94 in.) (recommended)
Miscellaneous
Cable length1.5 or 10 m, extension cable up to total maximum length of 40 m
Connection information
Inspection intervalEvery 2 years; 1/year service agreement by request, with warranty extension up to 5 years
Maintenance requirementsClean the measurement window, if necessary. Intervals are dependent on the measuring medium.
–5 to +45 °C (23 to 113 °F)
Measurement sensor wetted by at least half from the measuring medium:
–25 to +55 °C (-13 to 131 °F)
English 4
Dimensions
Figure 1 Sensor without and with cleaning unit
1Sensor without cleaning unit2Sensor with cleaning unit
English 5
General information
Safety information
Please read this entire manual before unpacking, setting up or operating
this equipment. Pay attention to all danger and caution statements.
Failure to do so could result in serious injury to the operator or damage
to the equipment.
Make sure that the protection provided by this equipment is not
impaired, do not use or install this equipment in any manner other than
that specified in this manual.
Use of hazard information
DANGER
Indicates a potentially or imminently hazardous situation which, if
not avoided, will result in death or serious injury.
WARNING
Indicates a potentially or imminently hazardous situation which, if
not avoided, could result in death or serious injury.
CAUTION
Indicates a potentially hazardous situation that may result in minor
or moderate injury.
NOTICE
Indicates a situation which, if not avoided, may cause damage to
the instrument. Information that requires special emphasis.
Note: Information that supplements points in the main text.
Precautionary labels
Read all labels and tags attached to the instrument. Personal injury or
damage to the instrument could occur if not observed. A symbol on the
instrument is referenced in the manual with a precautionary statement.
This is the safety alert symbol. Obey all safety messages that follow
this symbol to avoid potential injury. If on the instrument, refer to the
instruction manual for operation or safety information.
This symbol indicates that a risk of electrical shock and/or
electrocution exists.
This symbol shows that a UV lamp is used in the equipment.
Electrical equipment marked with this symbol may not be disposed of
in European domestic or public disposal systems after 12 August
2005. In conformity with European local and national regulations
(EU Directive 2002/96/EC), European electrical equipment users
must now return old or end-of life equipment to the manufacturer for
disposal at no charge to the user.
Note: For return for recycling, please contact the equipment
producer or supplier for instructions on how to return end-oflife equipment, producer-supplied electrical accessories, and
all auxiliary items for proper disposal.
English 6
Product overview
DANGER
This product is not suitable for use in potentially explosive
atmospheres.
WARNING
The UV rays from the flash bulb are harmful to eyes and skin.
Do not look directly through the measurement window during
operation under any circumstances. Remove the measurement
sensor from operation before carrying out any maintenance or
installation work.
The FP 360 sc sensor is a UV fluorimeter used to continuously measure
the concentration of PAH (polycyclic aromatic hydrocarbons) in water.
The measurement values can be converted to reflect the total oil content
for mineral oils using lab data.
The sensor may need to be installed with additional accessories
depending on its area of application.
Area of application Installation withSensor variants
Open channels, shafts, tanks
(solid matter max. 200 mg/L)
Measurement media without
solid matter with continuous,
low sample throughput
Do not apply the sensor to any hard mechanical effects.
Chain mount kit
Flow cell
Sensor with or without
cleaning unit
Sensor without
cleaning unit
Product components
The sensor is available in a variety of types.
Refer to Figure 2 to make sure that all components have been received.
If any of these items are missing or damaged, contact the manufacturer
or a sales representative immediately.
Figure 2 FP 360 sc sensor
1FP 360 sc measurement sensor 56 mm fitting for cleaning unit
2Basic user manual with CD6Safety sleeve
3Measurement window7Connector cable
4Sensor with cleaning unit
(depending on model)
(depending on model)
8Shackle
English 7
Function test
Connect sensor cable
CAUTION
Before power is applied, refer to the controller operation
instructions.
After the components are removed from the package, do a function test.
1.Connect the connector cable to the sensor (8-pin polarized
connector) and an appropriate sc controller (5-pin polarized
connector) (refer to Connect sensor cable on page 8).
2.Apply power to the sc controller. The display is activated and the
sensor goes to measurement mode.
The sensor ticks quietly and regularly.
3.Cover the sensor measurement window with a sheet of white paper
(do not use recycled paper).
4.Vary the distance between the measurement window and the
paper. The measurement value on the display will change
accordingly.
Note: In air, the measurement value displayed is not exactly zero due to
reflections on the window surface (refer to Check the zero point on
page 14).
Installation
DANGER
Personal injury hazard. Only qualified personnel should conduct
the tasks described in this section of the manual.
NOTICE
If the sensor is not fully inserted, sun protection is recommended
in high ambient temperatures and intense solar radiation to
protect against thermal and UV effects.
CAUTION
Always put the cables and hoses in a position that does not bend
or cause a trip hazard.
CAUTION
Before power is applied, refer to the controller operation
instructions.
1.Connect the polarized socket on the connector cable to the sensor
plug (8-pin plug) (refer to Figure 3 step 1).
2.Push the safety sleeve on the plug (step 2).
3.Hand-tighten the safety sleeve in position (step 3).
Figure 3 Connect the sensor cable to the sensor
English 8
4.remove the protective cap on the controller socket and keep it to
seal the connector opening in case the sensor must be removed.
5.Connect the sensor to the controller using the keyed quick-connect
fitting. Hand-tighten (refer to Figure 4).
Note: Connector cables are available in various lengths. Maximum
overall cable length is 40 m (131,23 ft).
Figure 4 Connect the sensor to the controller
Installation options
The sensor may need to be installed with additional optional accessories
depending on the area of application.
Note: Refer to the documentation supplied with the accessories for
detailed installation instructions.
Installation with the chain mount kit
The FP 360 sc sensor is installed with the chain mount kit in open
channels, shafts and tanks.
Figure 5 FP 360 sc measurement sensor with chain mount kit
1Sensor without cleaning unit2Sensor with cleaning unit
English 9
Installation of chain mount kit for sensors with cleaning unit
Install the cleaning unit hose
Note: An air hose is required, to operate the sensor with the optional
cleaning unit. An oil-free compressed air (6 bar) and a solenoid valve or
the HOAB compressed air cleaning system is also required.
With the HOAB compressed air cleaning system, replace the end of the
hose connection that is connected to the compressed air on the
underside of the instrument with the straight 6 mm fitting (refer to
Figure 2 on page 7, item 5) supplied with the probe.
Figure 6 Install the cleaning unit hose
Figure 7 FP 360 sc measurement sensor with cleaning unit and
chain mount kit
English 10
Installation with flow cell
The FP 360 sc sensor with flow cell is installed for samples free of solids
and particulates and limited sample flows.
Figure 8 Installation with flow cell
Operation
Sensor setup
When the sensor is connected for the first time, the sensor serial
number is displayed as the name of the sensor. To change the sensor
name:
1.Open the MAIN MENU.
2.Select SENSOR SETUP and confirm.
3.Select the corresponding sensor and confirm.
4.Select CONFIGURE and confirm.
5.Select EDIT NAME and confirm.
6.Edit the name and confirm to return to the SENSOR SETUP menu.
Complete sensor configuration in the same manner, with the following
menu options selected:
•SET PARAMETER
•SELECT UNITS
•AVERAGE
•LOG SETUP
•GAIN VALUE
Menu structure
SENSOR STATUS
SELECT SENSOR (if there is more than one sensor)
ERROR LISTPossible error messages: SENSOR ERROR
WARNING
LIST
Possible warning messages: TEST/MAINT, BULB CHANGE,
LAST CONFIGUR, TARGET VALUE
Note: Refer to Troubleshooting on page 17 for a list of all possible error
and warning messages together with a description of all necessary
countermeasures to be taken.
English 11
SENSOR SETUP
SELECT SENSOR (if there is more than one sensor)
CALIBRATE
SET OUTMODE
HOLD
ACTIVE
SET TRANSFER
SENSOR MEASURECurrent, uncorrected measurement value
CONFIGURE
FACTOR
OFFSET
2 POINT
3 POINT
4 POINT
5 POINT
FACTOR
OFFSET
2-POINT
3-POINT
4-POINT
Behavior of the outputs during calibration and zero
point adjustment
FACTOR: 0.1 to 100
OFFSET: –1000 to +1000
Is shown when FACTOR has been selected under
CONFIGURE. Refer to Calibration on page 13 for
detailed information.
Is shown when OFFSET has been selected under
CONFIGURE. Refer to Calibration on page 13 for
detailed information.
Is shown when 2 POINT has been selected under
CONFIGURE. Refer to Calibration on page 13 for
detailed information.
Is shown when 3 POINT has been selected under
CONFIGURE. Refer to Calibration on page 13 for
detailed information.
Is shown when 4 POINT has been selected under
CONFIGURE. Refer to Calibration on page 13 for
detailed information.
SENSOR SETUP (Continued)
SELECT SENSOR (if there is more than one sensor)
CALIBRATE
5-POINT
SET CAL DEFLTSecurity query, reset to (FACTOR=1, OFFSET=0)
CONFIGURE
EDIT NAME
SET PARAMETER
SELECT UNITS
AVE RAG E
LOG SETUP
GAIN VALUE
SET DEFAULTS
Is shown when 5 POINT has been selected under
CONFIGURE. Refer to Calibration on page 13 for
detailed information.
Name can include up to 16 characters
DEFAULT CONFIG: sensor serial number
PAH: Measurement value related to PAH calibration
standard
OIL: Measurement value related to oil calibration
standard
DEFAULT CONFIG: PAH
Range 0 to 500 ppb: AUTO, 0.01 to 50, 0.01 to 500,
Range 0 to 5000 ppb: AUTO, 0.01 to 500, 0.01 to
5000
DEFAULT CONFIG: AUTO
Security query,
reset to default configuration for all menu options
listed above.
English 12
SENSOR SETUP (Continued)
SELECT SENSOR (if there is more than one sensor)
DIAG/TEST
PROBE INFO
SENSOR NAMEDevice name
EDIT NAME
SERIAL NUMBERDevice serial number
RANGE0 to 500 or 0 to 5000
MODEL NUMBERItem no. Sensor
CODE VERSIONSensor software
COUNTER
OPERATING
HOURS
MAINTENANCECounter counting down days
BULB CHANGECounter counting down days
TEST/MAINT
SET OUTMODEEquipment output behavior in the SERVICE menu
HOLD
ACTIVE
SET
TRANSFER
SIGNALS
LAMP CURRFlash lamp intensity
DIAG / TESTZero point and slope check with external standards
READING
OFFSET
CUBE CAL
Operating hours counter
Password-protected access for the service
Calibration
Factory calibration
The calibration curve zero point and slope are preset. Retrospective
calibration of these basic settings is generally not required outside of the
inspection intervals.
Do regular zero point checks to make sure that impurities or faults are
being detected (refer to Check the zero point on page 14).
If the zero signal increases due to the measuring medium components
or the installation conditions, you can compensate for this influence via
an offset correction. Carry out a lab analysis of the sample to do this. If
there is no PAH/oil contamination in the measuring medium, enter the
measurement value shown by the device as the offset (refer to Adjust
the zero point (OFFSET) on page 15).
Process calibration/adjustment
The sensor is pre-calibrated with various concentrations of a special
calibration standard in ultra-pure water. These ideal measurement
conditions rarely occur in reality. The measurement values shown are
qualitative trend indicators if no adjustment is made to the on-site
measurement conditions.
If you require quantitatively correct measurement values, either a
contrast adjustment or a multi-point calibration needs to be carried out.
Both of these operations must be carried out on site using lab analysis
data. Basic prerequisites for quantitative measurements are precise
knowledge about the oil type in occurrence and constant measurement
conditions, e.g. in cool water in a heat exchanger. If several oils are
present in varying quantities, it is generally not possible to carry out a
quantitative measurement.
If the measurement conditions change, you must check the accuracy of
the results again through lab analyses and make any adjustments where
necessary.
English 13
The measurement conditions can change in terms of
•Composition of PAH or oil impurities
•Distribution of impurities in water
•Te mp
•Measuring medium composition
•Measurement sensor and measurement window
Determination of factors and adjustment of slope
To adjust the slope:
1.At the sensor installation site, draw a lab sample of the measuring
medium and promptly analyze the sample for PAH and oil content.
2.Make a note of the measurement value shown on the controller at
the time of the sample being taken. Make sure the correct unit is
shown for the measurement value, e.g. as ppm oil.
3.Repeat steps 1 and 2 several times.
4.Use the sample value and the value shown on the controller at the
time of sampling to calculate a factor.
5.Find an average value from the factors.
6.Enter the factor as the slope (refer to Adjust the slope (FACTOR)
on page 15).
Example for engine oilExample for naphthalene
Lab value: 4.0 ppm oil
Measurement value shown
2.4 ppm oil
Calculated factor: 1.67
1 At the time of sampling
1
:
It is advisable to adjust the slope if the following conditions apply:
•If the measuring sample is PAH-/oil free, the measurement value
must almost be zero.
•The factors calculated from the lab values must enable a sensible
average value to be derived.
If these conditions do not apply, do a multi-point calibration.
Note: If both the zero point and the slope must be changed, use a 2-point
calibration (refer to Multi-point calibration (2 to 5-point calibration) on
page 15).
Multi-point calibration
In the event of a multi-point calibration, enter the lab value as the target
value and the value shown as the actual value. Make sure that all values
are entered in the same unit, e.g. oil in ppm. (refer to Multi-point
calibration (2 to 5-point calibration) on page 15).
Check the zero point
•Medium: ultra-pure water
•Target value: < 1 ppb. Clean the window in the event of deviations.
Use a glass container (not plastic) large enough to enable the
measurement to be taken with an 8 to 10 cm distance between the
measurement window and the base (e.g. a 1000 mL glass beaker).
Place a black, non-reflective underlay under the container and switch off
artificial lights during the measurement process.
Note: In air, the measurement value displayed is not exactly zero due to
reflections on the window surface. This is standard sensor behavior and
not an indicator of malfunction.
Note: Always use ultra-pure water. Distilled water and demineralized
water are not suitable as these can contain organic compounds.
English 14
Adjustment of slope and zero point; multi-point calibration
Adjust the slope (FACTOR)
1.Open the MAIN MENU.
2.Select SENSOR SETUP and confirm.
3.Select the corresponding sensor and confirm.
4.Select CALIBRATE and confirm.
5.Select CONFIGURE and confirm.
6.Select FACTOR and confirm.
7.Enter the calculated factor and confirm.
8.Go back to the MAIN MENU or the Measurement mode display.
Adjust the zero point (OFFSET)
1.Open the MAIN MENU.
2.Select SENSOR SETUP and confirm.
3.Select the corresponding sensor and confirm.
4.Select CALIBRATE and confirm.
5.Select CONFIGURE and confirm.
6.Press OFFSET and confirm.
7.Enter the required offset and confirm.
8.Go back to the MAIN MENU or the Measurement mode display.
Multi-point calibration (2 to 5-point calibration)
Note: Multi-point calibration means that the pairs must be input in
ascending order.
1.Open the MAIN MENU.
2.Select SENSOR SETUP and confirm.
3.Select the corresponding sensor and confirm.
4.Select CALIBRATE and confirm.
5.Select CONFIGURE and confirm.
6.Select the type of calibration, for example 2 POINT and confirm.
7.Select 1PAIR and confirm.
8.Edit the TARGET VALUE and confirm.
9.Edit the ACTUAL VALUE and confirm.
10. Repeat the process for 2PAIR and confirm.
11. Go back to the MAIN MENU or the Measurement mode display.
English 15
Maintenance
The inside of the sensor is maintenance-free.
The cleanliness of the measurement window in the sensor head has an
impact on the accuracy of measurements. Check the measurement
window at regular intervals to make sure it is clean. The required
frequency of these checks is dependent on the measuring medium. Also
do a check in the event of unusually high measurement values and
clean the measurement window if necessary (refer to Cleaning the
measurement windows).
For sensors with a cleaning unit, adjust the frequency of cleaning to the
measurement conditions. The inspection interval must be shortened to
reflect any increase in solid matter.
NOTICE
Individual components of the suspension device (shackle and
chain of the chain mount kit) are made of stainless steel and may
corrode.
Maintenance schedule
IntervalMaintenance task
Visual inspectionApplication-dependent
System inspectionEvery 2 years
Calibration checkEvery 2 yearsCheck calibration
1 When operated according to factory settings and appropriate use
Contact the manufacturer's service department every 2 years to arrange
sensor inspection, testing, calibration and seal replacement. The flash
bulb is also replaced every 4 years.
Check for contamination
and corrosion.
Check plugs and flash
bulb.
1
Cleaning the measurement windows
CAUTION
Cleaning agents can be hazardous to health.
Wear protective equipment and avoid direct contact with cleaning
fluids.
NOTICE
Other cleaning agents can damage the material. Damage caused
by cleaning carried out incorrectly is not covered by the warranty.
1.Rinse the sensor with fresh water until all attached solid matter has
been removed.
2.Use pure acetone and a soft, clean cloth (e.g. camera lens cleaning
paper) to carefully remove the deposits on the measurement
window. Avoid sharp objects on the measurement window surface.
3.Rinse the residue from the cleaning agent with fresh water.
English 16
Troubleshooting
Error messages
Possible sensor errors displayed by the sc controller.
Displayed errorsDefinitionResolution
SENSOR ERRORElectronic defect
Warnings
Possible sensor warning messages displayed by the sc controller.
Displayed warnings DefinitionResolution
DIAG/TESTCounter expired
BULB CHANGECounter expired
LAST CONFIGUR
TAR GET VALUE
Changed configuration
was not accepted
With multi-point
calibration, values not
entered in ascending
order
Replacement parts
DesignationQuantityService life
Flash bulb 14 years
O-Rings42 years
Call manufacturer customer
service
Call manufacturer customer
service
Call manufacturer customer
service
Send configuration again
Enter calibration values in
ascending order
English 17
English 18
Technische Daten
Änderungen vorbehalten.
Messung
Messmethode
Messbereich
Anzeigeeinheitenppb, ppm, µg/L, mg/L
Reproduzierbarkeit2,5 % vom Messwert bei konstanter Temperatur
Messgenauigkeit5 % vom Messwert ±2 % vom Messbereichsendwert bei konstanter Temperatur
Nachweisgrenze1,2 ppb (PAH)
Ansprechzeit10 s (T90)
KalibrierungWerkskalibrierung mit UV-Fluoreszenz-Kalibrierstandard, kundenspezifische Anpassung möglich
Sensor-Software
SoftwareversionAb 1.14
Geräteeigenschaften
Masse
Druckbereich
Abmessungen
Niedriger Messbereich
Hoher Messbereich
MesssensorEdelstahlversion 2,8 kg, Titanversion 1,8 kg
DurchflusszelleCa. 0,6 kg, incl Montagetafel ca. 2,0 kg
0 bis 50 ppb und 0 bis 500 ppb bezogen auf PAK-Kalibrierstandard,
entsprechend 0,1 bis 1,5 ppm und 0,1 bis 15 ppm Öl-Kalibrierstandard
0 bis 500 ppb und 0 bis 5000 ppb bezogen auf PAK-Kalibrierstandard,
entsprechend 0,1 bis 15 ppm und 0,1 bis 150 ppm Öl-Kalibrierstandard
68 mm × 306 mm (2.68 in. × 12.05 in.) (ohne Stecker und Aufhängungsstift)
68 mm × 399 mm (2.68 in. × 15.71 in.) (inkl. Aufhängungsstift)
68 mm × 413 mm (2.68 in. × 16.26 in.) (mit zusätzlicher Reinigungsoption)
98 mm × 98 mm × 150 mm (3.86 in. × 3.86 in. × 5.91 in.) (ohne Fittings),
Montagetafel: 600 × 300 × 10 mm (23.62 in. × 11.81 in. × 0.39 in.)
Deutsch
Deutsch 19
Geräteeigenschaften (Fortsetzung)
Gehäuse: Edelstahl 1.4571 oder Titan
Optikträger: POM
Sensor
Werkstoffe
Durchflusszelle
SchäkelEdelstahl 1.4301
Umgebungsbedingungen
Temperatur des Messmediums1 bis 40 °C (34 bis 104 °F)
Umgebungstemperatur
Abstand Sensor - Wand/BodenMin. 100 mm (3.94 in.) (Empfehlung)
Sonstiges
Kabellänge1,5 oder 10 m, Verlängerungskabel bis maximal 40 m Gesamtlänge
Anschlüsse
InspektionsintervallAlle 2 Jahre; auf Wunsch 1/Jahr Servicevertrag, mit Garantieverlängerung auf 5 Jahre
WartungsaufwandReinigung des Messfensters bei Bedarf, Intervalle sind abhängig vom Messmedium
–5 bis +45 °C (23 bis 113 °F)
bei mindestens zur Hälfte vom Messmedium benetztem Messsensor:
–25 bis +55 °C (-13 bis 131 °F)
Deutsch 20
Abmessungen
Abbildung 1 Sensor ohne und mit Reinigungseinheit
1Sensor ohne Reinigungseinheit2Sensor mit Reinigungseinheit
Deutsch 21
Allgemeine Informationen
Sicherheitshinweise
Lesen Sie das gesamte Handbuch sorgfältig durch, bevor Sie das Gerät
auspacken, aufbauen oder in Betrieb nehmen. Achten Sie auf alle
Gefahren- und Warnhinweise. Nichtbeachtung kann zu schweren
Verletzungen des Bedieners oder zu Beschädigungen am Gerät führen.
Um sicherzustellen, dass die Schutzvorrichtungen des Geräts nicht
beeinträchtigt werden, darf dieses Gerät in keiner anderen als der in
diesem Handbuch beschriebenen Weise verwendet oder installiert
werden.
Gefahrenhinweise in diesem Handbuch
Warnschilder
Beachten Sie alle Kennzeichen und Schilder, die an dem Gerät
angebracht sind. Nichtbeachtung kann Personenschäden oder
Beschädigungen am Gerät zur Folge haben.
Dieses Symbol ist ein Warndreieck. Befolgen Sie alle
Sicherheitshinweise, die diesem Symbol folgen, um mögliche
Verletzungen zu vermeiden. Wenn sich dieses Symbol auf dem
Gerät befindet, verweist es auf Informationen in den Betriebs- und/
oder Sicherheitshinweisen der Betriebanleitung.
Dieses Symbol kann an einem Gehäuse oder einer Absperrung im
Produkt angebracht sein und zeigt an, dass Stromschlaggefahr und/
oder das Risiko einer Tötung durch Stromschlag besteht.
GEFAHR
Zeigt eine potenziell oder unmittelbar gefährliche Situation an,
die, wenn sie nicht vermieden wird, zum Tod oder zu schweren
Verletzungen führt.
WARNUNG
Zeigt eine potenziell oder unmittelbar gefährliche Situation an,
die, wenn sie nicht vermieden wird, zum Tod oder zu schweren
Verletzungen führen kann.
VORSICHT
Kennzeichnet eine mögliche Gefahrensituation, die geringfügige
oder mittelschwere Verletzungen zur Folge haben kann.
ACHTUNG
Kennzeichnet eine Situation, die, wenn sie nicht vermieden wird,
zu Schäden am Gerät führen kann. Informationen, die besonders
hervorgehoben werden sollen.
Hinweis: Informationen, die Aspekte aus dem Haupttext ergänzen.
Deutsch 22
Dieses Symbol zeigt an, dass eine UV-Lampe im Gerät benutzt wird.
Mit diesem Symbol gekennzeichnete elektrische Geräte dürfen ab
dem 12. August 2005 europaweit nicht mehr im unsortierten Hausoder Gewerbemüll entsorgt werden. Gemäß geltenden
Bestimmungen (EU-Direktive 2002/96/EG) müssen ab diesem
Zeitpunkt Verbraucher in der EU elektrische Altgeräte zur Entsorgung
an den Hersteller zurückgeben. Dies ist für den Verbraucher
kostenlos.
Hinweis: Anweisungen zur fachgerechten Entsorgung aller
(gekennzeichneten und nicht gekennzeichneten)
elektrischen Produkte, die von Hach-Lange geliefert oder
hergestellt wurden, erhalten Sie bei Ihrem zuständigen
Hach-Lange-Verkaufsbüro.
Produkt Überblick
GEFAHR
Dieses Produkt ist nicht zum Einsatz in explosionsgefährdeten
Bereichen geeignet.
WARNUNG
Die UV-Strahlung der Blitzlampe schädigt Augen und Haut.
Schauen Sie während des Betriebs keinesfalls direkt in das
Messfenster. Nehmen Sie den Messsensor außer Betrieb, bevor
Sie Wartungs- oder Montagearbeiten durchführen.
Der FP 360 sc-Sensor ist ein UV-Fluorometer zur kontinuierlichen
Messung der PAK (Polyzyklische Aromatische Kohlenwasserstoffe)
Konzentration in Wasser. Die Messwerte können bei Mineralölen auf
den Gesamtölgehalt umgerechnet werden.
Je nach Einsatzgebiet muss der Sensor mit weiterem optionalem
Zubehör installiert werden.
Messmedien ohne Feststoffe
mit kontinuierlichem
geringem Probendurchsatz
Setzen Sie den Sensor keinen mechanischen Stößen aus.
Kettenhalterung
Durchflusszelle
Sensor mit oder ohne
Reinigungseinheit
Sensor ohne
Reinigungseinheit
Lieferumfang
Der Sensor ist in verschiedenen Varianten lieferbar.
Kontrollieren Sie die Lieferung auf Vollständigkeit (siehe Abbildung 2).
Wenn etwas fehlt oder beschädigt ist, wenden Sie sich bitte sofort an
den Hersteller oder die Vertriebsstelle.
Abbildung 2 FP 360 sc-Sensor
1FP 360 sc Messsensor56 mm Fitting für Reinigungs-
2Basic User Manual mit CD6Sicherheitshülse
3Messfenster7Anschlusskabel
4Sensor mit Reinigungseinheit
(modellabhängig)
einheit (modellabhängig)
8Schäkel
Deutsch 23
Funktionskontrolle
Installation
VORSICHT
Vor dem Einschalten der Spannungsversorgung unbedingt die
Hinweise in der Betriebsanleitung des Controllers beachten!
Prüfen Sie die Komponenten nach dem Auspacken auf eventuelle
Transportschäden und führen Sie eine kurze Funktionskontrolle durch.
1.Schließen Sie das Anschlusskabel an den Sensor (8-poliger
verpolungssicherer Stecker) und an einen geeigneten sc Controller
(5-poliger verpolungssicherer Stecker) an (siehe Sensorkabel
anschließen auf Seite 25).
2.Schalten Sie die Spannungsversorgung des sc Controllers ein.
Die Anzeige wird aktiviert und der Sensor geht in den Messbetrieb
über.
Sie hören ein regelmäßiges leises Ticken des Sensors.
3.Verdecken Sie das Messfenster des Sensors mit einem weißen
Blatt Papier (kein Recyclingpapier).
4.Variieren Sie den Abstand zwischen Messfenster und Papier.
Der auf dem Display angezeigte Messwert muss entsprechend
variieren.
Hinweis: In Luft ist der anzeigte Messwert wegen Reflexionen an der
Fensteroberfläche nicht exakt bei Null (siehe Nullpunkt kontrollieren auf
Seite 31).
GEFAHR
Die Installation dieses Systems muss von qualifiziertem
Fachpersonal unter Einhaltung aller lokal gültigen
Sicherheitsvorschriften ausgeführt werden.
ACHTUNG
Falls der Sensor nicht komplett eintaucht, soll bei hoher
Umgebungstemperatur und starker Sonneneinstrahlung ein
Sonnenschutz zum Schutz vor thermischen und UV-Belastungen
verwendet werden.
Deutsch 24
Sensorkabel anschließen
VORSICHT
Verlegen Sie Kabel und Schläuche stets stolperfrei und ohne
Knick.
VORSICHT
Vor dem Einschalten der Spannungsversorgung unbedingt die
Hinweise in der Betriebsanleitung des Controllers beachten!
1.Stecken Sie die verpolungssichere Buchse des Anschlusskabels
auf den Stecker des Sensors (8-poliger Stecker) (siehe Abbildung 3
Schritt 1).
2.Schieben Sie die Sicherheitshülse auf den Stecker (Schritt 2).
3.Schrauben Sie die Sicherheitshülse handfest an (Schritt 3).
Abbildung 3 Anschluss des Sensorkabels an den Sensor
4.Schrauben Sie die Schutzkappe der Controllerbuchse ab und
bewahren Sie diese auf.
5.Achten Sie auf die Führung im Stecker und schieben Sie den
Stecker in die Buchse ( siehe Abbildung 4).
6.Schrauben Sie die Mutter handfest an.
Hinweis: Verbindungskabel sind in verschiedenen Längen erhältlich. Die
maximale Gesamtkabellänge beträgt 40 m (131,23 ft).
Abbildung 4 Anschluss des Sensors an den Controller
Deutsch 25
Optionale Installationsvarianten
Je nach Einsatzgebiet muss der Sensor mit weiterem optionalem
Zubehör installiert werden.
Hinweis: Genaue Details zur Installation der Varianten entnehmen Sie
bitte den entsprechenden Anleitungen.
Installation mit Kettenhalterung
In offenen Rinnen, Schächten und Tanks wird der FP 360 sc-Sensor mit
der Kettenhalterung installiert.
Abbildung 5 FP 360 sc Messsensor mit Kettenhalterung
Installation der Kettenhalterung bei Sensoren mit
Reinigungseinheit
Schlauch der Reinigungseinheit montieren
Hinweis: Zum Betrieb des Sensors mit optionaler Reinigungseinheit wird
ein Luftschlauch benötigt. Weiter erforderlich ist entweder ölfreie
Druckluft (6 bar) und ein Magnetventil, oder das Druckluftreinigungssystem HOAB.
Bei der Verwendung des Druckluftreinigungssystems HOAB muss der
druckseitige Schlauchanschluss an der Unterseite des Geräts gegen
das mit der Sonde mitgelieferte gerade 6 mm Fitting (Abbildung 2 auf
Seite 23, Punkt 5) getauscht werden.
Abbildung 6 Schlauch der Reinigungseinheit montieren
1Sensor ohne Reinigungseinheit2Sensor mit Reinigungseinheit
Deutsch 26
Abbildung 7 FP 360 sc Messsensor mit Reinigungseinheit und
Kettenhalterung
Installation mit Durchflusszelle
Für Messmedien ohne Feststoffgehalt und kleinere Probenströme wird
der FP 360 sc-Sensor mit der Durchflusszelle installiert.
Abbildung 8 Installation mit Durchflusszelle
Deutsch 27
Bedienung
Sensor Setup
Wenn Sie den Sensor zum ersten Mal anschließen, wird die
Seriennummer des Sensors als Sensorname angezeigt. Den
Sensornamen können Sie wie folgt ändern:
1.Öffnen Sie das HAUPTMENÜ.
2.Wählen Sie SENSOR SETUP und bestätigen Sie.
3.Wählen Sie den entsprechenden Sensor und bestätigen Sie.
4.Wählen Sie KONFIGURIEREN und bestätigen Sie.
5.Wählen Sie NAME MESSORT und bestätigen Sie.
6.Editieren Sie den Namen und bestätigen Sie, um zum SENSOR
SETUP Menü zurückzukehren.
Vervollständigen Sie in gleicher Weise Ihre Systemkonfiguration mit der
Einstellung folgender Menüpunkte:
•PAR AMETE R
•EINHEIT WÄHLEN
•MITTELWERT
•LOGGER
•VERSTÄRK.
Menü Struktur
SENSOR STATUS
WÄHLE SENSOR (bei mehreren Sensoren)
FEHLERMögliche Fehlermeldungen: SENSOR FEHLER
WARNUNGEN
Hinweis: Im Abschnitt Störungen, Ursachen, Beseitigung auf Seite 33
finden Sie neben der Auflistung sämtlicher Fehler- und Warnmeldungen
auch die Beschreibung aller notwendigen Maßnahmen.
Mögliche Warnmeldungen: WARTUNG, LAMPENWECHSEL,
LETZTE KONFIG, SOLLWERT
SENSOR SETUP
WÄHLE SENSOR (bei mehreren Sensoren)
KALIBRIEREN
AUSGANGSMODUS
HALTEN
MITLAUFEN
ERSATZWERT
MESSWERT SENSaktueller, unkorrigierter Messwert
KONFIGURIEREN
FAKTOR
OFFSET
2 PUNKTE
3 PUNKTE
4 PUNKTE
5 PUNKTE
FAKTOR
OFFSET
2 PUNKTE
3 PUNKTE
4 PUNKTE
Verhalten der Ausgänge während der Kalibrierung
oder Nullpunkteinstellung
FAKTOR: 0.1 bis 100
OFFSET: –1000 bis +1000
Erscheint, wenn FAKTOR unter KONFIGURIEREN
ausgewählt wurde. Detaillierte Beschreibung siehe
Kalibrierung auf Seite 30.
Erscheint, wenn OFFSET unter KONFIGURIEREN
ausgewählt wurde. Detaillierte Beschreibung siehe
Kalibrierung auf Seite 30.
Erscheint, wenn 2 PUNKTE unter
KONFIGURIEREN ausgewählt wurde. Detaillierte
Beschreibung siehe Kalibrierung auf Seite 30.
Erscheint, wenn 3 PUNKTE unter
KONFIGURIEREN ausgewählt wurde. Detaillierte
Beschreibung siehe Kalibrierung auf Seite 30.
Erscheint, wenn 4 PUNKTE unter
KONFIGURIEREN ausgewählt wurde. Detaillierte
Beschreibung siehe Kalibrierung auf Seite 30.
Deutsch 28
SENSOR SETUP (Fortsetzung)
WÄHLE SENSOR (bei mehreren Sensoren)
KALIBRIEREN
5 PUNKTE
WERKS-KAL
KONFIGURIEREN
NAME MESSORT
PAR AME TE R
EINHEIT WÄHLEN
MITTELWERT
LOGGER
VERSTÄRK.
WERKS-KONFIG
Erscheint, wenn 5 PUNKTE unter
KONFIGURIEREN ausgewählt wurde. Detaillierte
Beschreibung siehe Kalibrierung auf Seite 30.
Sicherheitsabfrage, Rücksetzen auf (FAKTOR=1,
OFFSET=0)
bis 16-stelliger Name möglich,
WERKS-KONFIG: Seriennummer des Sensors
PAK: Messwert bezogen auf PAK-Kalibrierstandard
OIL: Messwert bezogen auf Öl-Kalibrierstandard
WERKS-KONFIG: PAK
0.01 bis 500,
Messbereich 0 bis 5000 ppb: AUTO, 0.01 bis 500,
0.01 bis 5000
WERKS-KONFIG: AUTO
Sicherheitsabfrage,
Rücksetzen auf Werkskonfiguration bei allen oben
aufgeführten Menüpunkten.
SENSOR SETUP (Fortsetzung)
WÄHLE SENSOR (bei mehreren Sensoren)
DIAG/TEST
SENSOR INFO
NAME SENSORGerätename
NAME MESSORT
GERÄTENUMMERSeriennummer des Gerätes
MESSBEREICH0 bis 500 oder 0 bis 5000
TYPENBEZ.Artikel-Nr. Sensor
SOFTWARE-VERSSensor-Software
ZÄHLER
BETRIEBSSTUNDEN
WARTUNGZählwerk Tage rückwärts
LAMPENWECHSELZählwerk Tage rückwärts
SERVICE
AUSGANGSMODUS Verhalten der Geräteausgänge im SERVICE-Menü
HALTEN
MITLAUFEN
ERSATZWERT
SIGNALE
I LAMPEIntensität der Blitzlampe
DIAG / TEST
OFFSET MESSW
KAL
TROCKENSTD
Betriebsstundenzählwerk
Nullpunkt und Steilheitskontrolle mit externen
Standards
passwortgeschützter Zugang für den Service
Deutsch 29
Kalibrierung
Werkskalibrierung
Nullpunkt und Steilheit der Kalibrierkurve sind ab Werk eingestellt. Eine
Nachkalibrierung dieser Grundeinstellungen ist in der Regel außerhalb
der Inspektionsintervalle nicht erforderlich.
Führen Sie regelmäßige Nullpunkt-Kontrollen durch, um sicher
Verschmutzungen oder Störungen zu erkennen (siehe Nullpunkt
kontrollieren auf Seite 31).
Führen die Bestandteile des Messmediums oder die
Installationsbedingungen bei der Messung zu einem erhöhten
Nullsignal, können Sie diesen Einfluss über eine Offset-Korrektur
kompensieren. Führen Sie hierzu eine Laboranalyse der Probe durch.
Wenn tatsächlich keine PAK-/Ölverunreinigung im Messmedium vorliegt,
geben Sie den vom Gerät angezeigten Messwert als Offset ein (siehe
Anpassung des Nullpunkts (OFFSET) auf Seite 31).
Prozesskalibrierung/Anpassung
Der Sensor ist werkseitig mit unterschiedlichen Konzentrationen eines
speziellen Kalibrierstandards in Reinstwasser kalibriert. Diese optimalen
Messbedingungen sind in der Praxis nur selten gegeben. Die
angezeigten Messwerte sind ohne Anpassung an die vor Ort gegebenen
Messbedingungen qualitative Trendanzeigen.
Wenn Sie quantitativ richtige Messwerte benötigen, ist entweder eine
Anpassung der Steilheit oder eine Mehrpunktkalibrierung notwendig.
Beides muss vor Ort unter Verwendung von Laboranalysedaten
durchgeführt werden. Grundvoraussetzung für quantitative Messungen
sind genaue Kenntnisse über die auftretende Ölsorte und gleich
bleibende Messbedingungen, z. B. im Kühlwasser eines
Wärmetauschers. Wenn mehrere Öle in variierenden Anteilen
vorhanden sein können, ist eine quantitative Messung in den meisten
Fällen nicht möglich.
Wenn sich die Messbedingungen ändern, müssen Sie die Richtigkeit der
Messwerte erneut mit Laboranalysen prüfen und gegebenenfalls erneut
anpassen.
Deutsch 30
Die Messbedingungen können sich ändern in Bezug auf
•Zusammensetzung der PAK- oder Öl-Verunreinigungen
•Verteilung der Verunreinigungen im Wasser
•Temperatur
•Zusammensetzung des Messmediums
•Messsensor und Messfenster
Faktorermittlung und Anpassung der Steilheit
Zur Anpassung der Steilheit des Sensors gehen Sie wie folgt vor:
1.Nehmen Sie am Installationsort des Sensors eine Probe des
Messmediums und analysieren Sie diese zeitnah auf den PAKoder Ölgehalt.
2.Notieren Sie den angezeigten Messwert am Controller zum
Zeitpunkt der Probenahme. Achten Sie auf die richtige Einheit des
Messwerts, z. B. als ppm Öl.
3.Wiederholen Sie Punkt 1 und 2 mehrmals.
4.Errechnen Sie aus jedem Laborwert und dem am Controller zum
Zeitpunkt der Probenahme angezeigten Wert einen Faktor.
5.Errechnen Sie einen Mittelwert aus den Faktoren.
6.Geben Sie den gemittelten Faktor als Steilheit ein (siehe