Section 2 General information ............................................................................................................................... 5
2.1.2 Safety symbols used ................................................................................................................................ 5
2.1.3 General safety instructions ....................................................................................................................... 6
2.2 General .............................................................................................................................................................. 7
2.3.2 Principle of operation................................................................................................................................ 8
2.4.1 Identification of the operating instructions ................................................................................................ 8
2.5 Target group for these operating instructions .................................................................................................... 9
2.6 Tasks and obligations of the operating organisation.......................................................................................... 9
2.6.1 Correct use ............................................................................................................................................... 9
2.7 Your service partner, contacting the manufacturer ............................................................................................ 9
2.7.1 Standards and declaration of conformity ................................................................................................ 10
3.1.1 Safety measures prior to unpacking, connecting and operating............................................................. 11
3.1.2 Disposing of the packaging .................................................................................................................... 11
3.1.3 Preparing the installation ........................................................................................................................ 12
3.2 Drawing with dimensions ................................................................................................................................. 13
3.3 Installing the filtration probe sc ........................................................................................................................ 14
3.4 Supply of power ............................................................................................................................................... 19
3.4.1 Connecting power cable ......................................................................................................................... 20
3.5 Connecting filtration probe sc to the data network ........................................................................................... 21
5.1 The operation concept ..................................................................................................................................... 25
5.1.1 Rectifying problems in normal operation ................................................................................................ 25
5.2 Normal operation ............................................................................................................................................. 25
5.2.1 Rectifying problems in operation ............................................................................................................ 25
Section 6 Maintenance and cleaning .................................................................................................................. 27
6.1 Maintenance and cleaning by the operator...................................................................................................... 27
6.1.1 Troubleshooting, error state diagnostics and repair ............................................................................... 27
6.2 Maintenance and cleaning by qualified personnel ........................................................................................... 30
6.2.2 Changing air exhaust.............................................................................................................................. 31
1
Page 4
Table of contents
Section 7 Taking out of operation ....................................................................................................................... 33
7.1 Taking out of operation for a short period ........................................................................................................ 33
7.2 Taking out of operation for an extended period ............................................................................................... 33
7.3 Finally taking out of operation .......................................................................................................................... 33
7.4 Intermediate storage or storage....................................................................................................................... 33
We reserve the right to make changes without notice.
Table 1 Filtration probe sc specifications (in some cases optional extras)
ComponentsFilter module with pump
Dimensions(W × H × D) 330 × 384 × 243 mm (13 × 15.12 × 9.57 inches)
Supply tubing5 m or 10 m, depending on version
EnclosurePlastic enclosure, PPE, flammability class in accordance with UL 94
MassApprox. 9.5 kg filter enclosure, approx. 8 kg tubing
Power supplyVia analysis instrument
Electrical connectionVia analysis instrument
Mean power consumption
Electrical fuse protectionVia analysis instrument, via controller
Operating temperature4 °C...40 °C (39 °F...104 °F), in water
Operating temperature sample tubing–20 °C...45 °C (–4 °F...113 °F); 95 % relative humidity, non-condensing
Storage temperature
Sample temperature4 °C...40 °C (39 °F...104 °F), water temperature
Permissible sample flow speedmax. 3 m/s, from 1 m/s: install only with protection against flow (accessory)
Permissible pH value of the sample5–9
Maximum immersion depth3 m (120 inch)
Fastening to the extension pipe6 screws M5 × 10 mm
Filtrate flow rate
Cables and tubing on the probe pipe
400 VA (brief peak power consumption: 1000 VA),
(with 10 m heated filtration probe tubing)
–20 °C...60 °C (–4 °F...140 °F); 95 % relative humidity, non-condensing
Once the filter modules have become wet, they must be stored wet.
5 ml/minute, 4 out of 5 minutes
Outer sheath of the cables and tubing: UV and weather resistant
Total external diameter of the outer tubing: 32 mm (1.2 inch)
External diameter of the sample tubing 3.2 mm (0.12 inch)
External diameter of the return tubing: 6 mm (0.23 inch)
External diameter of the air tubing: 6 mm (0.23 inch)
External diameter of the data cable: 7.1 mm (0.28 inch)
External diameter of the power cable: 6.1 mm (0.23 inch)
3
Page 6
Specifications
4
Page 7
Section 2General information
2.1Safety instructions
Prior to unpacking, commissioning or operating the
instrument, read this manual in its entirety.
Please pay particular attention to all instructions on hazards
and safety. Otherwise there is a risk of serious injury to the
operator or damage to the instrument, or pollution.
The instrument is only allowed to be installed and used as per
the instructions in this manual.
2.1.1 Instructions on hazards in this manual
DANGER
Indicates a potentially hazardous situation or hazardous
situation that, if not avoided, could result in serious injuries or
fatality.
CAUTION
Indicates a possibly hazardous situation that could result in
light to moderate injuries.
2.1.2 Safety symbols used
This symbol, if present on the instrument, refers to information in the operating instructions on safe
operation and / or instructions that provide safety information.
This symbol, if present on a enclosure or a protective cover for the instrument, identifies the risk of an
electric shock (which may under certain circumstances be fatal). Only personnel qualified for working on
hazardous voltages are allowed to open the enclosure or remove the protective cover.
This symbol, if present on the instrument, identifies the location of a fuse or current limit.
This symbol, if present on the instrument, identifies a part that may become hot and must not be touched
without taking precautions.
Important note: Information that is to be specifically highlighted.
Note: Information with additional aspects on the main text.
Observe all stickers and markings on the instrument. Otherwise
injuries, pollution or damage to the instrument may occur.
This symbol, if present on the instrument, indicates the presence of components that could be damaged by
electrostatic discharge. Appropriate precautions are to be taken.
This symbol, if present on the instrument, indicates the presence of dangerous chemical substances.
Chemicals are only allowed to be handled and maintenance on devices for supplying chemicals to the
instrument is only allowed to be performed by personnel qualified and trained for working with chemicals.
5
Page 8
General information
This symbol, if present on the instrument, identifies the location of the connection for the protective earth
(ground).
As of 12 August 2005, electrical appliances marked with this symbol are no longer allowed to be
disposed of in Europe in unsorted household or industrial waste. As per the applicable
regulations, from this date on consumers in the EU must return old appliances to the
manufacturer for disposal. This disposal is free of charge for the consumer.
Note: You can obtain instructions on the correct disposal of all (marked and unmarked) electrical products
that have been supplied or manufactured by Hach-Lange from your local Hach-Lange sales office.
When carrying or transporting the instrument/instrument components and if the total weight is
18-32 kg (39.7-70.5 lbs)
more than 18 kg, make sure that suitable lifting equipment is used and/or that the
instrument/instrument components are carried by 2 people.
2.1.3 General safety instructions
DANGER
Risk of electric shock!
Work on the electrical installation is only allowed to be
performed by an experienced electrician!
Always isolate the instrument from the mains during
installation work!
CAUTION
Observe applicable health and safety regulations!
Flows of sample of unknown composition can produce
hazards due to traces of chemicals, radiation or biological
effects.
Avoid unnecessary contact with the flows of sample of
unknown composition and take appropriate safety measures.
DANGER
Potential danger in the event of contact with
chemical/biological materials. Handling chemical samples,
standards and reagents can be dangerous. Familiarize
yourself with the necessary safety procedure s an d th e correc t
handling of the chemicals before the work and rea d and follow
all relevant safety data sheets.
Normal operation of this instrument may involve the use of
hazardous chemicals or biologically harmful samples.
•Observe all cautionary information printed on the original
solution containers and safety data sheet prior to their use.
•Dispose of all consumed solutions in accordance with national
regulations and laws.
•Select the type of protective equipment suitable to the
concentration and quantity of the dangerous material at the
respective work place.
6
Page 9
2.2General
2.3Product description
2.3.1 Identification
General information
In this section you will find information on:
•Instrument identification
•Items supplied
•Operating instructions
•Tasks and obligations of the operating organisation
•The manufacturer
•The declaration of conformity
The filtration probe sc is controller-operated filtration system that is
immersed. It supplies an AMTAX sc or PHOSPHAX sc connected in
series with a filtered sample flow. The filtration probe sc features
automatic cleaning and is largely maintenance-free.
Figure1Filtration probesc
1Filtration probe sc3Rating plate
2Supply cable4Connection fastening system
7
Page 10
General information
2.3.1.1 Rating plate for the filtration probe sc
You will find the rating plate at the top end of the supply cable. On
the rating plate you will find
•The model number and the serial number,
•Information on the voltage and frequency for the electrical
connection and
•Information on the brief peak power consumption.
2.3.2 Principle of operation
The filtration probe sc draws in liquid through a filter system and
then pumps this sample flow to an analysis instrument connected in
series. After the analysis of the sample, the analysis instrument
pumps the sample (possibly mixed with chemicals) back to the tank
sampled or to a drain.
A cleaning device ensures the filter modules remain porous. The
filtration probe sc has a compressed air tube that terminates
underneath the filter modules. When compressed air flows out of
the air openings underneath the filter modules, the rising air
bubbles remove soiling from the filter modules. In this way the filter
modules remain porous and manual cleaning is largely
unnecessary.
2.4Items supplied
Standard items supplied:
•Filtration probe sc
•These operating instructions
•2 filter modules
•Small parts, cable, tubing and connector set and
•One cleaning sponge
2.4.1 Identification of the operating instructions
You will find a DOC number on the cover sheet so that you can
re-order these operating instructions. You can identify the issue of
the operating instructions using the date in the DOC number.
8
Page 11
2.5Target group for these operating instructions
These operating instructions are directed at sewage treatment plant
operating organisations and trained and experienced staff sewage
treatment plant supervisors. It is assumed that these persons are
familiar with the operation of personal computers and have a sound
knowledge of
•Measurement and control systems,
•Electrical engineering,
•Water analysis and
•Chemistry
and have been adequately instructed on safety regulations.
Requirements and activities that go beyond the knowledge and
skills held by a sewage treatment plant supervisor must be
entrusted by this person to trained (e. g. electrical) specialists. Such
work is to be planned and co-ordinated in agreement with the
organisation or plant management.
2.6Tasks and obligations of the operating organisation
General information
The organisation operating the plant must ensure only qualified and
trained personnel install, operate or otherwise use this probe. The
operating organisation must ensure the locally applicable
regulations are observed and that these operating instructions are
always available to the plant operators.
2.6.1 Correct use
Incorrect use of the filtration probe sc or its components
and / or its accessories can cause injury, damage or pollution.
Ensure the filtration probe sc and / or its components and / or
its accessories are only used correctly.
The filtration probe sc was developed for the filtration, supply and
disposal of liquid samples and for connection to an analysis
instrument. The primary area of application is water and waste
water analysis.
2.6.1.1 Incorrect use
Any other use or use beyond that defined above is considered
incorrect use and will also render void all claims under the
guarantee and warranty.
2.7Your service partner, contacting the manufacturer
In section Contact on page 39 you will find addresses and
telephone numbers for the various contacts.
9
Page 12
General information
2.7.1 Standards and declaration of conformity
The filtration probe sc and its components comply with the
applicable standards and stipulations.
For the filtration probe sc we confirm compliance.
HACH LANGE GmbH is certified to DIN EN ISO9001.
You can obtain further information on certificates, the declaration of
conformity and standards on request.
2.7.2 Copyright
"Provision to third parties as well as the duplication of these
operating instructions, further processing and provision of their
content to third parties is not allowed unless expressly approved.
Contravention will result in the obligation to pay damages.
All rights reserved in the case of the award of a patent or
registration of a design." (In accordance with DIN 34, ISO 16016)
10
Page 13
Section 3Installation
3.1Safety instructions
In this section you will find information on the following for the
filtration probe sc
•Installing,
•Connecting,
•Connecting to the mains and
•Connecting to the data network.
DANGER
Prior to unpacking, commissioning or operating the
instrument, read all of this manual!
Pay particular attention to all instructions on hazards and
safety. Otherwise there is a risk of serious injury to the
operator or damage to the instrument, or pollution!
The analysis instrument and its peripherals a re only allowed to
be installed and connected by qualified personnel. Only use
appropriate (e. g. insulated) tools and observe the applicable
rules and regulations!
Incorrect connection to the data network or incorrect
configuration can result in damage to the modules and the
customer's network!
3.1.1 Safety measures prior to unpacking, connecting and operating
Ensure only trained and instructed personnel unpack, install,
connect or operate the filtration probe sc.
3.1.2 Disposing of the packaging
Protect the environment when disposing of the packaging and
observe locally applicable regulations
11
Page 14
Installation
3.1.3 Preparing the installation
DANGER
Risk of electric shock!
Plan the mechanical fastening before you position posts or
drill holes. Choose a suitable place to install the instrument.
Plan how the cables and tubing will be laid and their routes.
Note that the cables will need to be removed regularly.
Lay cables and tubing so they will not cause tripping and
without sharp bends. Note that the cables will need to be
removed regularly.
Do not connect the electrical power supply to the mains until
the instrument has been fully wired and fused.
Provide adequate fuse protection for the electrical power
supply.
Always connect a residual current circuit breaker (trigger
current max.: 30 mA) between the mains and the system!
If you install the instrument outside, provide overvoltage
protection between the mains and the system!
12
Page 15
3.2Drawing with dimensions
Installation
Figure 2 Drawing with dimensions
13
Page 16
Installation
3.3Installing the filtration probe sc
Important note: Improper assembly or installation of the filtration
probe sc or its components can result in injury or damage.
The filtration probe sc and its components are only allowed to be
installed by qualified and trained personnel.
Install the filtration probe sc prior to commissioning the analysis
instrument.
3.3.1 Sealing exhaust
If you do not dispose of the used sample via the filtration probe sc,
seal the exhaust.
1. Seal the exhaust using the sealing cap supplied.
14
Figure 3 Sealing exhaust
Page 17
3.3.2 Installing filter module
Installation
Important note: Do not let the filter modules dry out, as dry filter
modules will become unusable immediately and cannot be
repaired.
Never touch the filter surface, as the filter surface will be destroyed
by the grease from the skin.
Never immerse the filtration probe in water without filter modules,
as unfiltered water could block the filtration probe.
1. Insert the filter module in the bottom.
2. Pivot the filter module towards the inside and press onto the
frame until it engages.
Figure 4 Installation of the filter modules
3. Proceed with the second filter module on the other side in the
same way.
15
Page 18
Installation
3.3.3 Installing filtration probe sc
Important note: Only install the filtration probe sc using the rim
mounting system.
Figure 5 Installation parts rim mounting LZX414.00.50000
1Mounting pipe7Cheese head screws M5 × 20 mm (6×)
2Screws M5 × 16 mm (6×)8Flat bar
3Quick-fit anchors (4)9Cheese head screws M8 × 40 mm (4×)
4Base10 Plug
5Clamp half11 Spring cotter
6Clamp half with thread
16
Page 19
Installation
Figure 6 Installation of the rim mounting LZX414.00.5000
1Base6Flat bar
2Quick-fit anchors (4)7Screws
3Plug8Screws M5 × 16 mm (6×)
4Spring cotter9Filtration probe tubing
5Mounting pipe10 Filtration probe sc
Important note: Only install the filtration probe sc in flowing water
with protection against flow.
Install the filtration probe sc so that is it held min. 0.1 m and max.
3 m below the surface of the water.
In case of frost, insulate the first 10 cm of the tubing if the top edge
of the filtration probe sc is not at least 20 cm below the surface of
the water.
17
Page 20
Installation
Important note: Ensure the filtration probe sc is not in contact with
the base of the tank and cannot bang against the side walls.
Figure 7 Installation position for the filtration probe sc
18
Page 21
Installation
3.4Supply of power
Figure 8 Installing with rim mounting
DANGER
Risk of electric shock!
Work on the electrical system is only allowed to be performed
by trained and experienced electricians.
19
Page 22
Installation
3.4.1 Connecting power cable
1. Open the analysis instrument to which the filtration probe sc is
to be connected.
2. Connect the earth wire (Figure 9-3) (see also the operating
instructions for the analysis instrument).
3. Connect the power cable to the filtration probe sc (Figure 9-2).
(see also the operating instructions for the analysis instrument)
4. Connect the power cable for the drain heating (optional
accessory) (Figure 9-1). (see also the operating instructions for
the analysis instrument)
Figure 9 Connecting power supply and heating to the filtration probe sc
1Connection drain heating
(optional)
20
2Filtration probe sc power
cable connection
3Earth wire terminal strip4Filtration probe sc data
cable connection
Page 23
3.5Connecting filtration probe sc to the data network
To ensure correct operation, the filtration probe sc must be
connected to the analysis instrument's data network. The analysis
instrument's data network sends the commands for self-cleaning
and transmits the module pressure value to the analysis instrument.
1. Connect the data cable connector to the socket on the analysis
instrument (Figure 9-4)
(see also the operating instructions for the analysis instrument).
3.6Connecting sample tubing
Note: Pay attention to the markings on the tubing.
1. Connect the black sample return tubing (Figure 10-1) to the
sample outlet on the analysis instrument (see operating
instructions for the analysis instrument).
2. Connect the transparent sample feed tubing (Figure 10-2) to
the sample inlet on the analysis instrument (see operating
instructions for the analysis instrument).
Installation
Figure 10 Connecting sample tubing
1Sample return2Sample feed
21
Page 24
Installation
3.7Connecting air tubing
Note: Pay attention to the markings on the tubing.
1. Connect the white air tubing to the analysis instrument (see
operating instructions for the analysis instrument).
Figure 11 Connecting air tubing
22
Page 25
Section 4System startup
Important note: Correct installation of the system is necessary to
ensure malfunction-free operation. Prior to initial commissioning,
ensure the system has been correctly installed and only author ised
persons have access to the system.
On initial commissioning, ensure all internal and external
expansions are correctly registered in the controller system.
4.1Initial commissioning
1. Ensure the filtration probe sc is completely and correctly
installed.
2. Switch on the analysis instrument and check whether the
filtration probe sc is correctly registered in the system.
3. Follow the instructions in the operating instructions for the
analysis instrument.
The filtration probe sc is then ready for use.
23
Page 26
System startup
24
Page 27
Section 5Operation
5.1The operation concept
No actions are intended to be performed by the operator on the
filtration probe sc (except cleaning and replacement of the filter
modules). All commands and control instructions are sent
automatically by the analysis instrument connected or are made via
the controller. Read the operating instructions for the analysis
instrument and the controller thoroughly to ensure you can operate
the system safely, without errors and correctly.
5.1.1 Rectifying problems in normal operation
In normal operation only feasible, sensible entry options are
displayed. If you then enter incorrect values, these will be rejected.
If you have made incorrect entries, repeat the entry with the correct
values.
Note: If your entry was outside a logical working range, a correction is
made automatically. Note that the value is then still incorrect and must be
corrected.
5.2Normal operation
In normal operation the filtration probe sc does not require
supervision, operation or maintenance.
5.2.1 Rectifying problems in operation
5.2.1.1 Problems with the sensors or the analysis instruments
If a sensor or analysis instrument is reporting errors or warnings, a
corresponding error will be displayed on the controller (see
operating instructions for the controller).
5.2.1.2 Module pressure value too low
If the filter modules block, the module pressure value in the filtration
probe sc drops.
1. Clean or replace the filter modules (see Cleaning / replacing
filter modules" on page 28).
2. Check all relevant settings.
If the problems have not been rectified, please contact customer
service.
25
Page 28
Operation
26
Page 29
Section 6Maintenance and cleaning
CAUTION
When carrying or transporting the instrument/instrument
components and if the total weight is more than 18 kg, make
sure that suitable lifting equipment is used and/or that the
instrument/instrument components are carried by 2 people.
6.1Maintenance and cleaning by the operator
•Regularly check the filtration probe sc for mechanical damage.
•Regularly check the filter modules for mechanical damage.
•Regularly check all connections for leaks and corrosion.
•Regularly check all cables and tubing for mechanical damage.
•Clean the filter modules using the damp sponge from the
accessories.
•Clean the filtration probe sc (not the filter modules) with a
cloth. You can use a brush for stubborn soiling.
6.1.1 Troubleshooting, error state diagnostics and repair
If the filter modules block, the module pressure value in the filtration
probe sc drops.
Clean the filter modules or replace the filter modules
(see Cleaning / replacing filter modules on page 28).
27
Page 30
Maintenance and cleaning
6.1.2 Cleaning / replacing filter modules
Note:
The filter modules are very delicate.
Only clean the filter modules very carefully using the damp sponge
supplied.
Clean the filter modules
•Regularly (min. every 3 months) or
•When the pressure values drop below 0.45 bar.
DANGER
Potential danger in the event of contact with
chemical/biological materials. Handling chemical samples,
standards and reagents can be dangerous. Familiarize
yourself with the necessary safety procedure s an d th e correc t
handling of the chemicals before the work and rea d and follow
all relevant safety data sheets.
Normal operation of this instrument may involve the use of
hazardous chemicals or biologically harmful samples.
•Observe all cautionary information printed on the original
solution containers and safety data sheet prior to their use.
•Dispose of all consumed solutions in accordance with national
regulations and laws.
•Select the type of protective equipment suitable to the
concentration and quantity of the dangerous material at the
respective work place.
NEVER use chlorine bleach together with hydrochloric acid
(Risk of the formation of chlorine gas).
For cleaning use chlorine bleach (5 %–10 %) OR hydrochloric
acid (5 %).
Avoid the contact of cleaning agents and soaps with filter
modules as these may render the filter modules unusable.
1. On the SENSOR SETUP menu on the analysis instrument,
choose the TEST/MAINT, SERVICE MODE command and
accept.
2. Take the filtration probe sc out of the tank.
3. Clean the filter modules carefully using the damp sponge
supplied.
28
4. Carefully lever the filter modules out of the holder using a
screwdriver.
5. Clean or replace the filter modules.
6. Fit the new or cleaned filter modules in the bottom.
Page 31
Maintenance and cleaning
7. Press up the filter modules against the enclosure until they
engage.
Figure 12 Changing filter module
8. Hang the filtration probe sc back in the tank.
9. On the SENSOR SETUP menu on the analysis instrument,
choose the TEST/MAINT and PREPUMPING PROBE
commands and accept.
10. Wait until the pumping operation is complete.
11. Leave the service state using the "Back" arrow.
12. Answer the SERV. HOLD MODE? prompt with NO/MEASURE
START.
13. On the SENSOR SETUP menu choose the MEASURING and
ON commands and accept.
Table 2 List of consumables and warranty periods
DesignationReplace afterSpare part No.
Filter module for filtration probe1 yearLZY140
29
Page 32
Maintenance and cleaning
6.2Maintenance and cleaning by qualified personnel
The filtration probe sc is only allowed to be opened by qualified and
authorised service personnel.
The following consumables in the filtration probe sc should be
changed at regular intervals by the service personnel:
Table 3 List of consumables and warranty periods
DesignationReplace afterWarrantySpare part No.
Set of wear parts for sample pump
(Pump membrane + valves)
Exhaust (copper)1 year1 yearLZY139
Exhaust (2 pcs.) for air cleaning
(Air exhaust set, incl. sealing and screws)
6.2.1 Changing exhaust
1 year, with 5 min. analysis interval
Otherwise 2 years
1 year1 yearLZY138
1 yearLZY130
If you dispose of the used sample via the filtration probe sc, renew
the exhaust after 1 year.
1. Unscrew the exhaust from the enclosure.
30
Figure 13 Changing exhaust
2. Screw the new exhaust into the enclosure and tighten it
hand-tight.
Page 33
6.2.2 Changing air exhaust
Maintenance and cleaning
1. Undo the 12 screws for the outlet.
2. Replace the exhaust and the sealing.
3. Tighten the 12 screws hand-tight.
4. Change the exhaust on the other side of the filtration probe sc.
Figure 14 Changing air exhaust
31
Page 34
Maintenance and cleaning
32
Page 35
Section 7Taking out of operation
7.1Taking out of operation for a short period
No special measures are necessary for taking out of operation for a
short period (up to a week in frost-free ambient conditions).
Important note: If the supply of power to the controller is
interrupted, frost damage may occur.
Ensure the instrument and tubing cannot freeze.
1. Interrupt the measurement and switch the analysis instrument
to the service state.
7.2Taking out of operation for an extended period
If the filter modules may dry out (e.g. on cleaning the tank), in case
of risk of frost or if the filtration probe sc is to be taken out of
operation for an extended period, please work through the following
list:
1. Take the analysis instrument out of operation in accordance
with the operating instructions for the analysis instrument.
2. Isolate the filtration probe sc from the mains.
3. Take the filtration probe sc out of the tank.
Important note: Do not let the filter modules dry out, as dry filter
modules will become unusable immediately and cannot be
repaired.
4. Dismantle the filter modules.
5. Store the filter modules in clean water.
6. Wrap the filtration probe sc in protective film or a dry cloth.
7. Store the filtration probe sc dry.
7.3Finally taking out of operation
When you want to finally take the filtration probe sc out of
operation, proceed as described for intermediate storage.
7.4Intermediate storage or storage
Proceed as for taking out of operation for an extended period.
33
Page 36
Taking out of operation
34
Page 37
Section 8Replacement Parts, consumables,
accessories
8.1Replacement parts and their order numbers
ItemDescription 1Description 2Order No.
Set of seals for filtration probe incl. O-ring, connector press. sensorLZY121
-- Set of wear parts for sample pumpincl. membrane, valves, screwsLZY130
-- Tubing for filtration probeFEP tubing 1 × 3.1 mm, 1.00 mLZY131
-- Air tubing for filtration probe4/6 mm, PUR-HLZY132
1Sample tubing 230 V, 5 mincl. top cover and adaptorLZY112
1Sample tubing 115 V, 5 mincl. top cover and adaptorLZY113
1Sample tubing 230 V, 10 mincl. top cover and adaptorLZY114
1Sample tubing 115 V, 10 mincl. top cover and adaptorLZY115
2Top cover filtration probe (plastic part)LZY116
3Enclose filtration probe (plastic part)without top coverLZY117
5Sealing top coverfor filtration probeLZY119
6Set of screws for filtration probe(28 pieces)LZY120
9Adaptor suction sideLZY128
--Set of screws(4 of each type used)LZY314
10Adaptor physical pageLZY129
12Card filtration probe with press. sensorZBA804 YAB042
13Set of fittings 3.2 mm(4 pieces)LZY111
14T-fitting 4/6 mmLZY133
15Fitting for tubing DN 4/6LZY134
16Elbow fitting 4/6 mmLZY135
17Fitting for exhaust valve incl. O-ringLZY136
18Humidity sensor incl. connectorLZY137
20Exhaust (copper)LZY139
21Adaptor for filtration probewith flange + screwsLZY118
22Safety cover filtration probeincl. pads and screwsLZY122
23Exhaust valve 230 V incl. cable, connector, fittings, screwsLZY123
23Exhaust valve 115 V incl. cable, connector, fittings, screwsLZY124
24Sample pump 230 V incl. sleeve cable and connectorLZY125
24Sample pump 115 V incl. sleeve cable and connectorLZY126
25Sleeve nut with sealingLZY127
26Exhaust (2 pcs.) for air cleaningincl. sealing/screwsLZY138
27Coupling piece incl. screwsLZY265
28Locking screw probe adaptorLZY213
29Seal for filter module O-ring 7,50 × 2,50 (4 pieces)LZY142
30Filter module for filtration probeLZY140
31Adaptor for filtration moduleincluding sealingLZY141
32Connection tubing for filtr. probe4/6 mm, PTFE 25 cmLZY253
35
Page 38
Replacement Parts, consumables, accessories
8.2Accessories
Description 1Description 2Order No.
SpongeEZZ062
Set of small parts 2 spring cotter with chains + 6 screwsLZY245
8.3Optional accessories
DescriptionDescription 2Order No.
Rim mounting for filtration probe scLZX414.00.50000
Rail mounting for filtration probe scLZX414.00.60000
Protection against flowLZY317
Villa 14 – Rue 2 Casa
Plaisance
Quartier Racine Extension
MA-Casablanca 20000
Tél. +212 (0)522 97 95 75
Fax +212 (0)522 36 89 34
info-maroc@hach-lange.com
www.hach-lange.ma
40
Page 43
Section 10Warranty, liability and complaints
HACH LANGE GmbH warrants that the product supplied is free of
material and manufacturing defects and undertakes the obligation
to repair or replace any defective parts at zero cost.
The warranty period for instruments is 24 months. If a service
contract is taken out within 6 months of purchase, the warranty
period is extended to 60 months.
With the exclusion of the further claims, the supplier is liable for
defects including the lack of assured properties as follows: all those
parts that, within the warranty period calculated from the day of the
transfer of risk, can be demonstrated to have become unusable or
that can only be used with significant limitations due to a situation
present prior to the transfer of risk, in particular due to incorrect
design, poor materials or inadequate finish will be improved or
replaced, at the supplier's discretion. The identification of such
defects must be notified to the supplier in writing without delay,
however at the latest 7 days after the identification of the fault. If the
customer fails to notify the supplier, the product is considered
approved despite the defect. Further liability for any direct or
indirect damages is not accepted.
If instrument-specific maintenance and servicing work defined by
the supplier is to be performed within the warranty period by the
customer (maintenance) or by the supplier (servicing) and these
requirements are not met, claims for damages due to the failure to
comply with the requirements are rendered void.
Any further claims, in particular claims for consequential damages
cannot be made.
Consumables and damage caused by improper handling, poor
installation or incorrect use are excluded from this clause.
HACH LANGE GmbH process instruments are of proven reliability
in many applications and are therefore often used in automatic
control loops to provide the most economical possible operation of
the related process.
To avoid or limit consequential damage, it is therefore
recommended to design the control loop such that a malfunction in
an instrument results in an automatic change over to the backup
control system; this is the safest operating state for the environment
and the process.