Hach-Lange CDC401-03 User Manual

Instruction Sheet
CDC401-01 or CDC401-03 Probe

Safety

To ensure that the protection provided by this equipment is not impaired, do not use or install this equipment in any manner other than that specified in this document.
This symbol, if noted on the instrument, references the instruction manual for operation and/or safety information.
Electrical equipment marked with this symbol may not be disposed of in European public disposal systems after 12 August of 2005. In conformity with European local and national regulations (EU Directive 2002/96/EC), European electrical equipment users must now return old or end-of life equipment to the Producer for disposal at no charge to the user.
Note: For return for recycling, please contact the equipment producer or supplier for instructions on how to return end-of-life equipment, producer-supplied electrical accessories, and all auxiliary items for proper disposal.

Overview

with either a 1- or 3-meter cable. The probe is designed for fast, accurate measurement of conductivity, salinity, resistivity, or total dissolved solids (TDS) in wastewater, drinking water, and general applications.
DOC302.53.00755
Figure 1CDC401-01 or CDC401-03 Probe
1 Sensor Array, including Temperature Sensor and
4-pole Conductivity Sensor
2 Color-coded clip locations
Each electrode is supplied with color-code clips for easy identification. There are two each of five colors (black, yellow, light green, light blue, and magenta), one for the electrode and one for the cable connector.
3 1- or 3-meter Cable
1

Assembly

Calibration

Maintenance

Specifications

1. Attach a pair of color-code clips to the electrode and cable connector (Figure 1, item 2).
2. Rinse probe with deionized water prior to use.
For calibration steps, refer to the HQ Portable Meters User Manual. The IntelliCAL™ probes store the current calibration and calibration history in the probe memory. It is not necessary to recalibrate when moving a calibrated electrode from one meter to another.
General Maintenance
Between samples, wipe the probe with a clean towel.
If the connection to the meter is poor, readings may be unstable. To ensure accurate
readings, check the connection to the meter.
Storage
Between uses, store the probe dry in ambient conditions.
Specifications are subject to change without notice.
Conductivity Range 0.01 µS/cm to 200.0 mS/cm
0.0–19.99 µS/cm: 0.01 µS/cm
20.0–199.9 µS/cm: 0.1 µS/cm
Conductivity Resolution
Conductivity Accuracy ± 0.5% of Reading
TDS Range 0 to 50,000 mg/L as NaCl
TDS Accuracy ± 0.5% of Reading
TDS Resolution
Salinity Range 0 to 42 ppt (‰)
Salinity Accuracy ± 0.1 ppt (‰)
Salinity Resolution 0.01 ppt (‰)
Temperature Range –10.0 to 110.0 ºC
Temperature Accuracy ± 0.3 ºC
200.0–1999.0 µS/cm: 1.0 µS/cm
2.0–19.99 mS/cm: 0.01 mS/cm
20.0–200.0 mS/cm: 0.1 mS/cm
0.0–199.9 mg/L: 0.1 mg/L
200.01999.0 mg/L: 1.0 mg/L
2.0–19.99 g/L: 0.01 g/L
20.0–50.0 g/L: 0.1 g/L
Warranty Sensor is covered by a one-year warranty.
FOR TECHNICAL ASSISTANCE, PRICE INFORMATION AND ORDERING:
In the U.S.A. – Outside the U.S.A. – On the Worldwide Web – ; E-mail –
© Hach Company, 2005–2006. All rights reserved. Printed in the U.S.A. January 2006, Edition 3
Call 800-227-4224
toll-free
Contact the HACH office or distributor serving you.
www.hach.com techhelp@hach.com
HACH COMPANY
WORLD HEADQUARTERS Telephone: (970) 669-3050 FAX: (970) 669-2932
Loading...