Hach SIGMA 900 User Manual

Catalog Number 8544
Sigma 900 Composite Sampler
INSTRUMENT MANUAL
© Hach Company, 2003. All rights reserved. Printed in the U.S.A. eac 07/03 2ed
Visit http: //www.hach.com

Table of Contents

2.1 Instrument Description ........................................................................................................................................ 7
2.2 Interface Connectors........................................................................................................................................... 7
2.2.1 Receptacle Caps ....................................................................................................................................... 7
2.3 Front Panel .......................................................................................................................................................... 8
2.3.1 Keypad and Display................................................................................................................................... 8
2.3.2 Humidity Indicator ...................................................................................................................................... 8
2.4 Safety Information ............................................................................................................................................... 9
2.4.1 Use of Hazard Information....................................................................................................................... 10
2.4.2 Precautionary Labels............................................................................................................................... 10
2.4.3 Hazardous Locations ............................................................................................................................... 10
2.4.4 Confined Space Entry ............................................................................................................................. 11
2.4.5 Definition of Confined Space ................................................................................................................... 12
Section 3 Installation......................................................................................................................................... 13
3.1 Unpacking the Instrument ................................................................................................................................. 13
3.2 Selecting the Installation Site ............................................................................................................................ 13
3.3 Mounting the Sampler ....................................................................................................................................... 14
3.4 Attaching the Intake Line................................................................................................................................... 14
3.4.1 Vinyl Intake Tubing................................................................................................................................... 14
3.4.2 Teflon-Lined Polyethylene Intake Line ..................................................................................................... 14
3.5 Setting Up the Intake Line and Strainer ............................................................................................................ 15
3.6 Choosing Bottles ............................................................................................................................................... 15
3.7 Installing the Full Bottle Shut-Off ...................................................................................................................... 15
3.8 Power Connections ........................................................................................................................................... 16
3.9 12 V dc.............................................................................................................................................................. 17
3.10 Auxiliary Connection ....................................................................................................................................... 17
Section 4 Operation........................................................................................................................................... 21
4.1 Keypad and Key Functions................................................................................................................................ 21
4.2 Tips and Techniques for Programming the Instrument ..................................................................................... 22
4.3 Setting the Time and Date ................................................................................................................................ 23
4.4 Bottle Setup Mode ............................................................................................................................................ 23
4.5 Explanations of Program Messages ................................................................................................................. 24
4.6 Data Logging..................................................................................................................................................... 25
4.7 Manual Mode .................................................................................................................................................... 25
4.7.1 Manually Operating the Sample Pump .................................................................................................... 25
4.8 Measurements Triggered by Liquid Levels or Flow........................................................................................... 26
4.9 Programming the Sampler for Level Control ..................................................................................................... 27
4.10 Other Displayed Messages ............................................................................................................................. 27
4.11 Flow Proportional Operation ........................................................................................................................... 28
4.11.1 How to Calculate Pulses/Counts ........................................................................................................... 28
Page 3
8544TOC.fm Table of Contents
Table of Contents
Section 5 Maintenance ..................................................................................................................................... 33
5.1 Cleaning the Sampler ....................................................................................................................................... 33
5.1.1 Cleaning the Sampler Housing................................................................................................................ 33
5.1.2 Cleaning the Sample Bottles ................................................................................................................... 33
5.1.3 Cleaning the Intake Tubing and Pump Tubing ......................................................................................... 33
5.1.4 No Lubrication Required .......................................................................................................................... 33
5.2 Pump Tubing Maintenance................................................................................................................................ 33
5.3 Upgrades, Repairs, General Maintenance ........................................................................................................ 33
5.4 Electrostatic Discharge (ESD) Considerations.................................................................................................. 33
5.5 Internal Maintenance Items............................................................................................................................... 33
5.6 Removing and Opening the Controller.............................................................................................................. 33
5.7 Re-installing the Bottom Panel.......................................................................................................................... 34
5.8 Internal Desiccant Module ................................................................................................................................ 35
5.9 Circuit Board Identification ............................................................................................................................... 35
5.10 Motor/Gear Box............................................................................................................................................... 36
5.11 Internal Case Humidity Indicator..................................................................................................................... 36
5.12 Internal Case Humidity Indicator..................................................................................................................... 36
5.13 Memory Battery .............................................................................................................................................. 36
Section 6 Contact Information for U.S.A. & Outside Europe................................................................... 37
Section 7 Contact Information for Europe ................................................................................................... 38
Section 8 Limited Warranty ............................................................................................................................. 39
Section 9 Parts and Accessories................................................................................................................... 41
Appendix A Programming Flow Chart ................................................................................................................. 43
Program Flow Chart ......................................................................................................................................... 43
Appendix B Batteries and Chargers .................................................................................................................... 45
Lead-Acid (Gel Cell) Batteries .......................................................................................................................... 45
Nickel-Cadmium Batteries ................................................................................................................................ 46
Appendix C Exploded View Drawings ................................................................................................................. 47
Exploded View (1 of 6) ..................................................................................................................................... 47
Exploded View (2 of 6) ..................................................................................................................................... 48
Exploded View (3 of 6) ..................................................................................................................................... 49
Exploded View (4 of 6) ..................................................................................................................................... 50
Exploded View (5 of 6) ..................................................................................................................................... 51
Exploded View (6 of 6) ..................................................................................................................................... 52
Page 4 Table of Contents
8544TOC.fm

Section 1 Specifications

Specifications are subject to change without notice.
Controller Dimensions
Pump/Controller Housing
Sample Pump
Pump Body Impact/corrosion resistant, glass reinforced Delrin
Vertical Lift
Sample Transport Velocity 61 mm/sec. (2 ft/sec.) minimum, at 4.6 m (15 ft) vertical lift in a 3/8 in. ID intake tube
Pump Flow Rate 60 mL/sec at 0.91 m (3 ft) vertical lift in a 3/8 in. ID intake tube
Sample Volume Programmed in milliliters, in one mL increments from 10 to 9999 mL
Sample Volume Repeatability
Sampling Modes Supports 1 bottle. Composite Time, Composite Flow, and Level Actuation
Interval Between Samples
Intake Purge
Control Panel
Height: 34 cm (13.5 in.), Width 28 cm (11 in.), Depth 20.3 cm (8 in.) without wall bracket, 33 cm (13 in.) deep with wall bracket
High impact injection molded ABS: submersible, watertight, dust tight, corrosion and ice resistant; NEMA 4X, 6
High speed peristaltic, dual roller, with 9.53 ID x 15.9 mm OD (3/8 in. x 5/8 in.) medical grade silicone rubber pump tube.
8.23 m (27 ft) maximum. Remote pump option recommended for lifts from 6.7 to
10.7 m (22 to 35 ft)
From 0.3 to 6.7 m (1 to 22 ft) vertical lift +/-5 mL typical for a 100 mL sample (+/­5%)
Selectable in single increments from 1 to 9999 flow pulses (momentary contact closure 25 msec. or 5-12 V dc pulse; 4-20 mA interface optional), or 1 to 9999 minutes in one minute increments.
Air purged automatically before and after each sample; duration automatically compensated for varying intake line lengths.
18 key, 31 function membrane switch keypad with 18 character alphanumeric, liquid crystal display. Self prompting/menu driven program.
Internal Clock Indicated real time and date; 0.007% time base accuracy
Programmable Delay Programmable sampler start time/date
Manual Sample Initiates a sample collection independent of program process
Records program start time and date, stores up to 400 sample collection
Data Logging
Automatic Shutdown
Intake Tubing
Intake Strainers
Power Requirements
ac Power Backup
Internal Battery
Overload Protection 5 amp dc line fuse; ac power pack; internal 1 amp ac line fuse
Temperature Range
times/dates, all program entries, operational status including number of minutes or pulses to next sample, bottle number, number of samples collected, number remaining, sample identification number.
Composite Mode: After preset number of samples have been delivered to composite container, from 1 to 9999 samples, or upon full container.
6.4 mm (1/4 in.) and 9.5 mm (3/8 in.) ID vinyl to 9.5 mm (3/8 in.) ID Teflon-lined polyethylene with protective outer cover
Choice of Teflon and 316 stainless construction, and all 316 stainless steel in standard size and low profile for shallow depth applications
12 V dc-supplied by optional ac power converter or battery. Average current draw with pump running: 2.25 amps dc, without pump running: 4 mA dc
Rechargeable 6 amp-hour gel lead acid battery takes over automatically with ac line power failure. Integral trickle charger maintains battery at full charge
1.5 V dc lithium battery; maintains program logic and real time clock for five years. Internal battery current draws less than 40 micro amps
General use: 0° to 50° C (32° to 122° F). LCD display operated from –10° to 70° C; Storage –40° to 80° C
Page 5
8544specs.fm Specifications
Visit http: //www.hach.com

Section 2 General Information

2.1 Instrument Description

The instrument is designed to automatically collect and preserve samples from a liquid source. The sampler is suitable for collection of conventional and toxic pollutants and suspended solids.
The sampler collects samples on either a timed cycle basis, or in proportion to flow when used in conjunction with a flow meter (flow meter connected to the auxiliary connector). In the timed cycle mode, the interval between samples is controlled by an integral quartz crystal clock. The sample interval may be set from 1 to 9999 minutes, in one minute increments.
In the flow proportional mode, via an external flow meter, the interval between samples may be set from 1 to 9999 flow pulses, in one pulse increments. where each pulse represents a specific flow volume interval.
The number of minutes or flow intervals remaining unit the next sample is shown on the control display. The program can be delayed by entering the desired starting time and date on the sampler keypad.
The sampler utilizes a high speed peristaltic pump, for collection of the sample liquid. The pump body is constructed of impact and corrosion resistant plastic. Tubing is held firmly in place by the pump body haves-no clamps are required. The intake line is air purged before and after each sample. The duration of the pre and post sample purge automatically compensates for different intake line lengths. The pump tubing is 0.95 cm (3/8
in.) ID X 0.32 cm (1/8 in.) wall medical grade silicone rubber. Although the pump is capable of vertical lifts up to 8.2 m (27 ft), lifts greater than 6.7 m (22 ft) are not recommended unless the Remote Pump option is utilized. The pump produces a sample transport velocity of 1 mps (3.3 fps) at 1 m (3 ft) vertical lift. The sample is under pumped flow at all times from the source stream to the sample container, thereby keeping solids in suspension.

2.2 Interface Connectors

2.2.1 Receptacle Caps

The interface connectors are located on the left side of the controller housing. The sampler comes standard with two interface receptacles:
12 V dc (Power Input)
Auxiliary (Multi-purpose input/output port)
Interface receptacles are covered with push-on receptacle caps. These caps protect the connector pins from dirt and moisture and should be attached to any receptacle not in use.
Page 7
8544intro.fm General Information
Section 2
Figure 1 Interface Connectors
1
2
1. 12 V dc 2. Auxiliary

2.3 Front Panel

The front panel consists of the keypad, liquid crystal display, and the internal case humidity indicator (

2.3.1 Keypad and Display

The sampler controller is equipped with a 24 character alphanumeric liquid crystal display which automatically steps the user through the program settings.

2.3.2 Humidity Indicator

The internal case humidity indicator (Cat. No. 2660) turns pink when the internal case humidity exceeds 60 percent.
The sampler is equipped with an internal desiccant module (Cat. No. 8849) to absorb any humidity trapped in the case during final assembly. Under normal operating conditions, this desiccant provides long-term protection against condensed moisture inside the case.
Replacement of the internal desiccant module is only necessary if the indicator turns pink. See to
Figure 2).
section 5.8 on page 35 for details.
Page 8 Front Panel
8544intro.fm
Figure 2 900 Composite Sampler
Section 2
1
7
6
2
3
4
5
1. Function Key 5. Power On Indicator LED
2. Peristaltic Pump 6. Desiccant Indicator
3. Peristaltic Pump Tubing 7. 24-character Alphanumeric Liquid Crystal Display
4. Tubing Clamp

2.4 Safety Information

Please read this entire manual before unpacking, setting up, or operating this instrument.
Pay particular attention to all danger and caution statements. Failure to do so could result in serious injury to the operator or
To ensure the protection provided by this equipment is not impaired, do not use or install this equipment in any manner other than that which is specified in this manual.
damage to the equipment.
Page 9
8544intro.fm Safety Information
Section 2

2.4.1 Use of Hazard Information

If multiple hazards exist, this manual will use the signal word (Danger, Caution, Note) corresponding to the greatest
DANGER Indicates a potentially or imminently hazardous situation which, if not avoided, could result in death or serious injury.
CAUTION Indicates a potentially hazardous situation that may result in minor or moderate injury.
NOTE Information that requires special emphasis.

2.4.2 Precautionary Labels

This symbol, if noted on the instrument, references the instruction manual for operation and/or
safety information.
Read all labels and tags attached to the instrument. Personal injury or damage to the instrument could occur if not
hazard.
observed.
This symbol, when noted on a product enclosure or barrier, indicates that a risk of electrical shock and/or electrocution exists and indicates that only individuals qualified to work with hazardous voltages should open the enclosure or remove the barrier.
This symbol, when noted on the product, identifies the location of a fuse or current limiting device.
This symbol, when noted on the product, indicates that the marked item can be hot and should not be touched without care.
This symbol, when noted on the product, indicates the presence of devices sensitive to Electro-static Discharge and indicates that care must be taken to prevent damage to them.
This symbol, when noted on the product, identifies a risk of chemical harm and indicates that only individuals qualified and trained to work with chemicals should handle chemicals or perform maintenance on chemical delivery systems associated with the equipment.
This symbol, if noted on the product, indicates the need for protective eye wear.
This symbol, when noted on the product, identifies the location of the connection for Protective Earth (ground).

2.4.3 Hazardous Locations

The 900 Composite Sampler is not approved for use in hazardous locations as defined in the National Electrical Code.
Page 10 Safety Information
DANGER Although some Hach products are designed and certified for installation in hazardous locations as defined by the National Electrical Code, many Hach products are not suitable for use in hazardous locations. It is the responsibility of the individuals who are installing the products in hazardous locations to determine the acceptability of the product for the environment. Additionally, to ensure safety, the installation of instrumentation in hazardous locations must be per the manufacturer's control drawing specifications. Any modification to the instrumentation or the installation is not recommended and may result in life threatening injury and/or damage to facilities.
8544intro.fm
Section 2
DANGER
Bien que certains produits Hach soient conçus et certifiés pour être installés dans des endroits dangereux tels que définis par le National Electric Code, de nombreux produits Hach ne conviennent pas pour de tels endroits. Il relève de la responsabilité des personnes qui placent les produits dans des endroits dangereux de déterminer s'ils sont adaptés à cet environnement. En outre, à des fins de sécurité, le placement de machines dans des endroits dangereux doit s'effectuer dans le respect des consignes des schémas de contrôle du fabricant. Toute modification apportée aux machines ou tout déplacement de celles-ci est déconseillé, car susceptible de provoquer des accidents matériels et/ou corporels.
PELIGRO
Aunque algunos productos Hach están diseñados y homologados para su instalación en entornos peligrosos, entendidos éstos conforme a la definición del “National Electrical Code” (Reglamento Eléctrico Nacional), muchos de los productos Hach no son aptos para su utilización en lugares peligrosos. Es responsabilidad de quienes instalen los productos en entornos peligrosos el asegurarse de la idoneidad de dichos productos para este tipo de entorno. Además, para garantizar la seguridad, la instalación de los instrumentos en lugares peligrosos deberá realizarse conforme a las especificaciones del plano del fabricante. Se desaconseja cualquier modificación de los instrumentos o de la instalación, ya que podría provocar lesiones corporales graves, e incluso fatales, y/o daños materiales a los equipos.

2.4.4 Confined Space Entry

GEFAHR
Einige Hach-Produkte sind für den Einbau in explosionsgefährdeten Bereichen gemäß den Festlegungen des National Electrical Code speziell geprüft und zugelassen. Dies gilt jedoch keineswegs für das gesamte Hach-Produktangebot. Die Entscheidung, ob ein Produkt für den Einsatz in explosionsgefährden Bereichen geeignet ist oder nicht, bleibt in die Verantwortung des jeweiligen Installateurs gestellt. Im Interesse der Sicherheit ist es zudem erforderlich, dass ein etwaiger Einbau des Geräts in explosionsgefährdeten Bereichen genau nach den Steuerungsanlagen-Zeichnungen des Herstellers erfolgt. Von der Vornahme von Änderungen an meß- bzw. regeltechnischen Geräten bzw. abweichender Installation wird dringend abgeraten, da hierdurch lebensbedrohliche Personen­und/oder Sachschäden verursacht werden können!
PERICOLO
Nonostante alcuni prodotti Hach, siano predisposti e certificati per l'installazione in ambienti pericolosi, come previsto dal Codice Normativo Nazionale che concerne l'elettricità, è sconsigliabile utilizzare prodotti Hach in ambienti considerati pericolosi. E' diretta responsabilità della persona che installa lo strumento in un luogo ritenuto “pericoloso” appurare se lo strumento è compatibile con tale ambiente. Inoltre, per maggior sicurezza, l'installazione dello strumento in ambienti pericolosi deve seguire le specifiche di progettazione del produttore. Si deve evitare qualunque manomissione allo strumento o all'installazione, tali modifiche possono rappresentare una minaccia per la vita delle persone e creare guasti.
Important Note: The following information is provided to guide users of 900
Composite Samplers on the dangers and risks associated with entry into confined spaces.
On April 15, 1993, OSHA's final ruling on CFR 1910.146, Permit Required Confined Spaces, became law. This new standard directly affects more than 250,000 industrial sites in the United States and was created to protect the health and safety of workers in confined spaces.
Page 11
8544intro.fm Safety Information
Section 2

2.4.5 Definition of Confined Space

A Confined Space is any location or enclosure that presents or has the immediate potential to present one or more of the following conditions:
An atmosphere with less than 19.5% or greater than 23.5% oxygen
and/or more than 10 ppm Hydrogen Sulfide (H
An atmosphere that may be flammable or explosive due to gases, vapors,
mists, dusts, or fibers
Toxic materials which upon contact or inhalation, could result in injury,
impairment of health, or death
Confined spaces are not designed for human occupancy. They have restricted entry and contain known or potential hazards.
Examples of confined spaces include manholes, stacks, pipes, vats, switch vaults, and other similar locations.
Standard safety procedures must always be followed prior to entry into confined spaces and/or locations where hazardous gases, vapors, mists, dusts, or fibers may be present.
S)
2
Before entering any confined space check with your employer for procedures related to confined space entry.
Page 12 Safety Information
8544intro.fm

Section 3 Installation

DANGER Only qualified personnel should conduct the installation tasks described in this section of the manual

3.1 Unpacking the Instrument

Remove the sampler from the shipping carton and inspect it for any damage. Contact Hach Customer Service at 1-800-227-4224 if any items are missing or damaged.

3.2 Selecting the Installation Site

The sampler is completely self-contained and may be located indoors or outdoors without protection, between 0 to 50 °C (32 to 120 °F). Follow the guidelines below and and prevent cross-contamination between samples.
Install the sampler as close to the source as site conditions permit. This
will increase pump tube life and optimize overall sampler performance.
Mount the sampler to a stable flat surface using the wall mounting
bracket, Cat. No. 3596 (
Install the sampler above the sample source, with the intake tubing
sloping downward to the sample.
Figure 3 to allow complete drainage of the intake line
Figure 4).
Make sure the tubing is free of kinks or loops.

Figure 3 Placement of the Sampler

1
2
Flow
1. Slope tubing down to source (no loops, kinks, or excess tubing).
2. Locate strainer in an area of turbulent and well-mixed flow.
Page 13
8544install.fm Installation
Section 3

Figure 4 Wall Mount Bracket

3.3 Mounting the Sampler

Attach the wall mount bracket to a stable panel or wall (Figure 4). Mount the bracket so the sample display and keypad are easily accessible. Mount the sampler to the wall bracket using the four screws provided.

3.4 Attaching the Intake Line

3.4.1 Vinyl Intake Tubing

1. Push one end of the clear flexible PVC (vinyl) tube to the tapered end of
the intake strainer, until the tube is within 0.635 cm (¼ in.) of the solid white Teflon strainer body.
2. Push the other end of the vinyl tube on to the stainless steel fitting
connector end that points away from the pump control housing.

3.4.2 Teflon-Lined Polyethylene Intake Line

The Connection Kit for Teflon-lined Tubing (Cat. No. 2186) contains two identical assemblies, one for connecting the Teflon-lined tubing to the stainless steel tubing connector and the other for connecting the intake strainer.
1. Place the stainless steel hose clamp over the stainless steel fitting until it
abuts the shoulder of the tubing connector. Secure with a tubing clamp.
Page 14 Mounting the Sampler
2. Push the stainless steel fitting into the Teflon-lined tubing until it abuts the
shoulder of the stainless steel fitting.
3. Slide the stainless steel hose clamp over the silicone tubing, then push
the silicone tubing over the stainless steel fitting on the liquid sensor.
8544install.fm
4. Slide the tubing clamp over the silicone tubing until it is over the stainless
steel fitting on the liquid sensor. Tighten to secure.
5. Repeat the procedure for the intake strainer.

3.5 Setting Up the Intake Line and Strainer

For each sampling location, the intake line should be as short as practical, and be free of any sharp bends, coils, or loops. Install the intake line with a downward slope from the sampler to the sample source because:
This will assure the complete drainage of the intake line when it is air-
purged before and after each sample, and will help to prevent cross­contamination of the individual samples.
Complete drainage is important in freezing conditions, as any liquid slugs
that remain could freeze and plug the line.
Place the sample intake and strainer in the mainstream of the sampling source, in an area of turbulent and well-mixed flow.
Also, take in account the vertical location of the intake. A position too near the surface may yield excess lighter materials, while a position too near the bottom may yield excess heavy materials. The constituents of interest must be considered when positioning the intake strainer.
Section 3

3.6 Choosing Bottles

A broad range of bottle configurations are available for the 900 Composite Sampler. Refer to
Size Material
9.5 L (3 gal) Polyethylene
11.4 L (2.5 gal) Glass
15.1 (4 gal) Polyethylene
20.1 L (5.5 gal) Polyethylene
22.7 L (6 gal) Polyethylene

3.7 Installing the Full Bottle Shut-Off

Install the optional full bottle shut-off into the bottle cap using the attached rubber grommet. Install the full bottle shut-off connector to the mating plug which is attached tot he sampler back panel. Insert the sample tubing into the cap of the bottle.
Page 15
8544install.fm Setting Up the Intake Line and Strainer
Section 3

Figure 5 Full Bottle Shut-off

1
2
1. Sample Tubing 2. Full Bottle Shut-off

3.8 Power Connections

The sampler controller operated on 12 V dc which is supplied by an optional ac/dc power converter and/or battery.
Note: Make sure both rubber hold­downs are attached to the clips at each end of the power supply.
The power supply (or battery) is placed in the compartment located directly behind the controller. Pull the rubber hold-downs up and over the clips at each end of the power supply (or battery) to hold it in place.
The short 2-pin cable on the power supply (or battery) connects the controller receptacle labeled 12 V dc.
Batteries available from the manufacturer include a lead-acid gel battery, and nickel-cadmium (Ni-Cad) battery (Cat. No. 1416). The 120 V ac to 12 V dc power converter (Cat. No. 1440) includes a built-in lead-acid battery charger. The Power Converter (Cat. No. 1443) contains a charger for Ni-Cad batteries. 230 V ac and 12 V dc power converters (Cat. No. 1441) are also available but do not come equipped with a built-in battery charger. Stand-alone, wall mount chargers are also available. See APPENDIX for details on batteries and battery charging.
Important: Whenever electricity is present, there is a possibility of electrical shock. Before connecting the sampler to an ac power source, the following safety precautions should be taken:
Page 16 Power Connections
Check the power source to make sure that it satisfies the ac power
requirements of the sampler.
Make sure that all electrical installations and connections are in
accordance with national and local electrical codes.
8544install.fm
Loading...
+ 36 hidden pages